TCP312, TCP305, TCP303, and TCP404XL
AC/DC Current Probes
TCP312, TCP305, and TCP303 Current Probes are designed to work with the
TCPA300 Current Probe Amplifier. The TCP404XL Current Probe is designed
to work with the TCPA400 Current Probe Amplifier.
Getting Started
To connect the current probe to the amplifier, align the red dots on the two
connectors and gently push the probe connector in. Do not twist the connector.
Degauss the probe before you take measurements: Press the PROBE DEGAUSS
AUTOBALANCE button and wait 5 seconds for the LED to turn green. See your
amplifier manual for complete details.
www.tektronix.com
Copyright © Tektronix
*P071118501*
071-1185-01

TCP312, TCP305, TCP303, and TCP404XLAC/DC Current Probes
Specification
s
Standard Accessories
TCP312: DC–100 MHz, 30 A DC
TCP305: DC–50 MHz, 50 A DC
TCP303: DC–15 MHz, 150 A DC
TCP404XL: DC–2 MHz, 500 A DC (750 A DC derated with duty cycle)
The following accessories are shipped with the probes:
Accessory Tektronix part number
Probe cover
(TCP305 and TCP312)
(TCP303 and TCP404XL)
Probe ground lead, 6 inch length (TCP305 and TCP312 only)
Instruction Sheet
Certificate of Traceable Calibration
016-1923-00
016-1924-00
119-3120-01
071-1185-01
—
Recommended Accessories
The following accessories can be ordered for your probes:
Accessory Tektronix part number
Deskew Fixture 067-1478-00
Travel Case
Related Documentation
WARN
TCP300/400 Series AC/DC Current Probes User Manual, Tektronix part number
077-1183-XX, before using the probe.
The TCPA300/400 Amplifiers & TCP300/400 Series AC/DC Current Probes User
Manual contains a complete list of specifications, accessories, and operating
information for the probes.
Warranty Information
016-1922-00
ING. To reduce risk of shock or fire, refer to the TCPA300/400 Amplifiers &
For warranty information, go to www.tektronix.com/warranty.