Tektronix TCP300,TCP400 Series User manual

TCP312, TCP305, TCP303, and TCP404XL AC/DC Current Probes
TCP312, TCP305, and TCP303 Current Probes are designed to work with the TCPA300 Current Probe Amplier. The TCP404XL Current Probe is designed to work with the TCPA400 Current Probe Amplier.
Getting Started
To connect the current probe to the amplier, align the red dots on the two connectors and gently push the probe connector in. Do not twist the connector.
www.tektronix.com Copyright © Tektronix
*P071118501*
071-1185-01
TCP312, TCP305, TCP303, and TCP404XLAC/DC Current Probes
Specication
s
Standard Accessories
TCP312: DC–100 MHz, 30 A DC
TCP305: DC–50 MHz, 50 A DC
TCP303: DC–15 MHz, 150 A DC
TCP404XL: DC–2 MHz, 500 A DC (750 A DC derated with duty cycle)
The following accessories are shipped with the probes:
Accessory Tektronix part number
Probe cover
(TCP305 and TCP312)
(TCP303 and TCP404XL)
Probe ground lead, 6 inch length (TCP305 and TCP312 only)
Instruction Sheet
Certicate of Traceable Calibration
016-1923-00
016-1924-00
119-3120-01
071-1185-01 —
Recommended Accessories
The following accessories can be ordered for your probes:
Accessory Tektronix part number
Deskew Fixture 067-1478-00
Travel Case
Related Documentation
WARN
TCP300/400 Series AC/DC Current Probes User Manual, Tektronix part number 077-1183-XX, before using the probe.
The TCPA300/400 Ampliers & TCP300/400 Series AC/DC Current Probes User Manual contains a complete list of specications, accessories, and operating information for the probes.
Warranty Information
016-1922-00
ING. To reduce risk of shock or re, refer to the TCPA300/400 Ampliers &
For warranty information, go to www.tektronix.com/warranty.
2
Loading...