Tektronix TCP400, TCP300, TCPA400, TCPA300 Service manual

Instruction Manual
TCPA300/400 Amplifiers & TCP300/400 Series AC/DC Current Probes
071-1183-03
This document applies for firmware version 1.0 and above.
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
www.tektronix.com
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
H In North America, call 1-800-833-9200.
H Worldwide, visit www.tektronix.com to find contacts in your area.
Warranty 2
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A P ARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Table of Contents

Getting Started
Operating Basics
Preface xv........................................................
System Configuration 1--1.............................................
Options 1--3........................................................
Standard Accessories 1--4.............................................
Optional Accessories 1--5.............................................
Probe Covers 1--6....................................................
Travel Case 1--7.....................................................
Connecting the Amplifier to an Oscilloscope 1--8..........................
Power on the Amplifier 1--8...........................................
Connecting a Current Probe to the Amplifier 1-- 9..........................
Operating the Current Probe Slide 1--10...................................
Degaussing and Autobalancing the Current Probe 1--11......................
DC Measurements 1--12...............................................
AC Measurements 1--14...............................................
Control Summary 2--1.........................................
TCPA300 and TCPA400 Controls 2--2...................................
PROBE DEGAUSS AUTOBALANCE Button and Indicator 2--2..........
MANUAL BALANCE Buttons and Indicator 2--3......................
PROBE OPEN Indicator 2--3.......................................
OVERLOAD Indicator 2--3........................................
NOT TERMINATED INTO 50 Indicator 2--4........................
NONCOMPATIBLE PROBE TYPE Indicator 2--4......................
ON/STANDBY Button 2--4........................................
RANGE Button (TCPA300 only) 2--4................................
COUPLING Button and Indicators 2--4...............................
PROBE INPUT Connector 2--5.....................................
OUTPUT Connector 2--5..........................................
Probe DC Gain Adjust (located on probes) 2--5.........................
GPIB Operation 2--6..............................................
Reference
Reference Notes 3--1...........................................
Degaussing a Probe with an Unpowered Conductor in the Jaws 3--1............
Measuring Differential Current 3--2.....................................
AC and DC Coupling 3--3.............................................
Maximum Current Limits 3--4..........................................
Measuring Noncontinuous Current with the TCP404XL Probe 3--6............
Extending Current Range 3--8..........................................
Increasing Sensitivity 3--10.............................................
Application Notes 3--11..........................................
Automobile Charging Systems 3--11......................................
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
i
Table of Contents
Inductance Measurements 3--13.........................................
Continuity Test of Multiple-Conductor Cable 3--15..........................
Measuring Inductor Turns Count 3--16....................................
Power Measurement and Analysis Software 3--17...........................
Troubleshooting and Error Codes 3--19............................
Displaying Error Codes with the Probe Degauss Autobalance Button 3--22.......
Correcting the Cause of an Error Code 3--24...............................
Shutdown Error 3--24..................................................
Specifications
Warranted Specifications 4 --1..........................................
Nominal and Typical Characteristics 4--2.................................
Mechanical Characteristics 4--3........................................
Environmental Characteristics 4--5......................................
TCP404XL Maximum Measurement Times 4--12...........................
Performance Verification
Performance V erification Overview 5--1...........................
Performance Verification and Functional Checks 5--1.......................
Test Procedure Conditions 5--2.........................................
Equipment Preparation 5--2............................................
TCPA300 and TCPA400 Performance Verification 5--3..............
Equipment Required 5--3..............................................
Making DC Current Loops 5--4.........................................
Front-Panel Display 5--5..............................................
DC Gain Accuracy 5--6...............................................
Bandwidth 5--9......................................................
AC Coupling 5--12....................................................
Degauss 5--13........................................................
Current Overload Test 5--14............................................
TCPA300 Amplifier Test Record 5--16....................................
TCPA400 Amplifier Test Record 5--17....................................
TCP305 and TCP312 Performance Verification 5--19.................
Required Test Equipment 5--20..........................................
DC Gain Accuracy 5--21...............................................
Rise Time 5--24......................................................
Bandwidth 5--26......................................................
TCP305 Current Probe Test Record 5--29..................................
TCP312 Current Probe Test Record 5--30..................................
TCP303 and TCP404XL Performance Verification 5--31..............
Equipment Required 5--32..............................................
DC Gain Accuracy 5--33...............................................
Rise Time 5--36......................................................
Bandwidth (TCP303) 5--39.............................................
Bandwidth (TCP404XL) 5--41..........................................
TCP303 Current Probe Test Record 5--42..................................
TCP404XL Current Probe Test Record 5--43...............................
ii
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Adjustment Procedures
Maintenance
Table of Contents
Adjustment Procedures Overview 6--1............................
TCPA300 and TCPA400 Amplifier Adjustments 6--3................
Required Test Equipment 6--3..........................................
Accessing the Adjustments 6--3.........................................
TCPA300 Amplifier 6--4..............................................
TCPA400 Amplifier 6--5..............................................
TCP305 and TCP312 Adjustments 6--7...........................
Required Test Equipment 6--7..........................................
Accessing the Adjustments 6--7.........................................
TCP305 and TCP312 DC Gain Adjustment 6--8............................
TCP303 and TCP404XL Adjustments 6--11........................
Required Test Equipment 6--11..........................................
Adjustment Locations 6--12.............................................
TCP303 and TCP404XL Transient Response and Coarse Gain Adjustment 6--13..
DC Gain Adjustment 6--18.............................................
Replaceable Parts
Customer Maintenance 7--1.....................................
Service Strategy 7--1.................................................
Preventive Maintenance 7--1...........................................
Disassembly 7--3..............................................
Amplifiers 7--3......................................................
Equipment Required 7--3..............................................
Removing the Outer Case and Internal Covers 7--4.........................
Replacing the Front Panel Assembly 7--7.................................
Replacing the Amplifier Board 7 --8.....................................
Replacing the Power Supply 7--10.......................................
Replacing the Fan 7--11................................................
Probes 7--13.........................................................
Equipment Required 7--13..............................................
TCP305 and TCP312 7--14.............................................
TCP303 and TCP404XL 7--18..........................................
Isolating Hardware Faults 7--21..................................
Determining the Amplifier Firmware Version Number 7--22...................
Replaceable Parts 8--1..........................................
Parts Ordering Information 8--1.........................................
Using the Replaceable Parts List 8--1....................................
Amplifier Replaceable Parts 8--3.................................
Current Probes Replaceable Parts 8--9............................
Glossary and Index
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
iii
Table of Contents
iv
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

List of Figures

Table of Contents
Figure 1--1: Typical TCPA300/400 current measurement system 1--1...
Figure 1--2: Using the probe holders 1--6..........................
Figure 1--3: Equipment locations in the travel case 1--7..............
Figure 1--4: Connecting and disconnecting a current probe
to the amplifier 1--9........................................
Figure 1--5: TCP312 and TCP305 slide operation 1--10...............
Figure 1--6: TCP303 and TCP404XL slide operation 1--11............
Figure 1--7: Current probe polarity 1--13...........................
Figure 2--1: The TCPA300 front panel 2--1........................
Figure 3--1: Measuring differential current and nulls 3--3............
Figure 3--2: Effect of AC or DC coupling on low-frequency signals 3--3.
Figure 3--3: Applying the amp-second product rule 3--5..............
Figure 3--4: Duty cycle calculation 3--6............................
Figure 3--5: Increasing the DC measurement range 3--9..............
Figure 3--6: Increasing probe sensitivity 3--10.......................
Figure 3--7: Setup for measuring charging current 3--12..............
Figure 3--8: Charge Current Waveforms 3--12.......................
Figure 3--9: Measuring inductance with a low-impedance source 3--13..
Figure 3--10: Linear current vs. time ramp 3--14.....................
Figure 3--11: Measuring inductance with a high-impedance source 3--15.
Figure 3--12: High-impedance source current ramp 3--15.............
Figure 3--13: Measuring the number of turns in a coil 3--16...........
Figure 3--14: Turns measurement using reference coil 3--17............
Figure 3--15: Error code display 3--22..............................
Figure 3--16: Interpreting the error code display 3--23................
Figure 4--1: Probe jaw dimensions (nominal) 4--4...................
Figure 4--2: Frequency derating--TCP312 4--8......................
Figure 4--3: Frequency derating--TCP305 4--8......................
Figure 4--4: Frequency derating--TCP303 4--9......................
Figure 4--5: Frequency derating--TCP404XL 4--9...................
Figure 4--6: Insertion impedance graphs for the current probes 4--10...
Figure 4--7: Specified operating area of the probes 4--11..............
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
v
Table of Contents
Figure 4--8: Measuring 750A noncontinuous at 50 °C
ambient temperature 4--12....................................
Figure 4--9: Measuring 600A noncontinuous at 50 °C
ambient temperature 4--12....................................
Figure 4--10: Measuring 750A noncontinuous at 23 °C
ambient temperature 4--13....................................
Figure 5--1: Check LED functionality 5--5.........................
Figure 5--2: Equipment setup for DC gain accuracy test 5--6..........
Figure 5--3: Bandwidth test setup 5--9.............................
Figure 5--4: AC coupling test setup 5--12...........................
Figure 5--5: Setup for degaussing the current probe 5 -- 13.............
Figure 5--6: Overload test setup 5--14..............................
Figure 5--7: DC gain accuracy test setup for TCP305 and TCP312 5--21.
Figure 5--8: Rise time test setup for the TCP305 and TCP312 5--24.....
Figure 5--9: Bandwidth test setup for TCP305 and TCP312 5--26.......
Figure 5--10: DC gain accuracy test setup for the
TCP303 and TCP404XL 5--33.................................
Figure 5--11: Rise time test setup for the TCP303 and TCP404XL 5--36.
Figure 5--12: Bandwidth test setup for TCP303 5--39.................
Figure 6--1: Amplifier adjustments 6--4...........................
Figure 6--2: Gain adjustment locations 6--5........................
Figure 6--3: TCP305 and TCP312 DC gain adjustment location 6--7...
Figure 6--4: DC gain adjustment setup for TCP305 and TCP312 6--9..
Figure 6--5: TCP303 and TCP404XL adjustment locations 6--12.......
Figure 6--6: TCP303 and TCP404XL transient response and
coarse gain adjustment setup 6--14.............................
Figure 6--7: DC gain accuracy adjustment setup 6--18................
Figure 7--1: Removing the case from the amplifier 7--4...............
Figure 7--2: Removing the right-side inner panel 7--5................
Figure 7--3: Removing the left-side inner panel 7--6.................
Figure 7--4: Disconnecting the front panel assembly 7--7.............
Figure 7--5: Removing the three heat sink clips 7--8.................
Figure 7--6: Removing the amplifier board 7--9.....................
Figure 7--7: Replacing the power supply 7--10.......................
Figure 7--8: Removing the AC power connector 7--11.................
Figure 7--9: Removing the fan 7--12...............................
vi
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Table of Contents
Figure 7--10: Removing the strain relief boot 7--14...................
Figure 7--11: Removing the top half of the probe 7--15................
Figure 7--12: Removing the probe slide 7--15........................
Figure 7--13: Removing the current transformer 7--16................
Figure 7--14: Removing the circuit board and cable assembly 7--17.....
Figure 7--15: Removing the handle 7--18...........................
Figure 7--16: Removing the current transformer 7--19................
Figure 7--17: Removing the circuit board and cable 7--20.............
Figure 7--18: Displaying the amplifier firmware version number 7--22..
Figure 8--1: TCPA300 and TCPA400 replaceable parts 8--3...........
Figure 8--2: TCPA300 and TCPA400 standard accessories 8--6........
Figure 8--3: TCPA300 and TCPA400 optional accessories 8--7........
Figure 8--4: TCP305 and TCP312 replaceable parts 8--9.............
Figure 8--5: TCP303 replaceable parts 8--11........................
Figure 8--5: TCP404XL replaceable parts 8--13......................
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
vii
Table of Contents
viii
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

List of Tables

Table of Contents
Table 1--1: Amplifier options 1--3................................
Table 1--2: Service options 1--3..................................
Table 3--1: Unpowered circuit degauss limits 3--1...................
T able 3--2: Automobile charging systems test setup 3--11.............
Table 3--3: Troubleshooting 3--19.................................
Table 3--4: Amplifier error codes 3--23.............................
Table 4--1: Warranted TCPA300 and TCPA400 specifications 4--1....
Table 4--2: Nominal and typical amplifier characteristics 4--2........
Table 4--3: TCPA300 and TCPA400 mechanical characteristics 4--3...
Table 4--4: Probe mechanical characteristics 4--4...................
Table 4--5: Shipping weights and dimensions 4--4...................
T able 4--6: Environmental characteristics 4--5.....................
Table 4--7: Certifications and compliances 4--6.....................
Table 5--1: Amplifier checks 5--1.................................
T able 5--2: Probe performance verification checks 5--1..............
Table 5--3: Required test equipment 5--3.........................
Table 5--4: Equipment settings 5--7..............................
Table 5--5: DC gain accuracy test for the TCPA300 and TCPA400 5--7.
Table 5--6: Equipment settings for bandwidth check 5--9............
Table 5--7: TCPA300 bandwidth measurements 5 -- 11................
Table 5--8: TCPA400 bandwidth measurements 5 -- 11................
Table 5--9: TCPA300 AC coupling measurements 5--13...............
Table 5--10: Maximum current ratings for
TCPA300-compatible probes 5--15............................
Table 5--11: Required test equipment 5--20.........................
T able 5--12: Equipment settings for DC gain accuracy 5--21...........
Table 5--13: DC gain accuracy test for the TCP305 and TCP312 5--23..
Table 5--14: Equipment settings for rise time 5--25..................
Table 5--15: Equipment settings for bandwidth 5--27.................
Table 5--16: Bandwidth test for the TCP305 and TCP312 5--28........
Table 5--17: Required test equipment 5 -- 32.........................
T able 5--18: Equipment settings for DC gain accuracy 5--33...........
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
ix
Table of Contents
T able 5--19: DC gain accuracy test worksheet for the TCP303 and
TCP404XL 5--35...........................................
Table 5--20: Equipment settings for rise time 5--37..................
Table 5--21: Equipment settings for bandwidth 5--40.................
Table 5--22: Bandwidth test for the TCP303 5--41...................
Table 6--1: Amplifier and probe adjustments 6--1...................
Table 6--2: Required test equipment 6--3..........................
Table 6--3: TCPA300 gain adjustments 6--4........................
Table 6--4: TCPA400 gain adjustments 6--5........................
Table 6--5: Required test equipment 6--7..........................
Table 6--6: Settings for DC gain adjustment 6--8...................
Table 6--7: DC gain accuracy adjustments for the
TCP305 and TCP312 6--9...................................
Table 6--8: Required test equipment 6--11..........................
Table 6--9: Settings for transient response and
preliminary gain adjustments 6--14............................
T able 6--10: Equipment settings for DC gain accuracy 6--19...........
Table 6--11: DC gain accuracy adjustments for the
TCP303 and TCP404XL 6--19................................
Table 7--1: Equipment required 7--3.............................
Table 7--2: Equipment required 7--13.............................
Table 7--3: Amplifier hardware faults 7--21.........................
Table 7--4: Probe hardware faults 7--22............................
Table 8--1: Power cord identification 8--5.........................
x
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
ToAvoidFireor
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Connect and Disconnect Properly. Connect the probe output to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe ground from the circuit under test before disconnecting the probe from the measurement instrument.
Ground the Product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Connect the ground lead of the probe to earth ground only.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
xi
General Safety Summary
Provide Proper Ventilation. Refer to the manual’s installation instructions for details on installing the product so it has proper ventilation.
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
CAUTION
Refer to Manual
Do not connect to or
remove from an
uninsulated conductor that
is HAZARDOUS LIVE.
Protective Ground
(Earth) Terminal
Breakable.
Do not drop.
Do not connect
to or discon-
nect from unin-
sulated Haz-
ardous Live conductors.
WARNING
Hot Surface
Use only on an
insulated wire.
xii
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

Service Safety Summary

Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then disconnect the power cord from the mains power.
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
xiii
Service Safety Summary
xiv
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

Preface

This Instruction Manual supports the operation and basic maintenance of the TCPA300 and TCPA400 Current Probe Amplifiers, and the TCP300/400 Series AC/DC current probes that mate with the amplifiers. The current probes covered in this manual are listed below:
H TCP312 (30 amps, 100 MHz, compatible with TCPA300)
H TCP305 (50 amps, 50 MHz, compatible with TCPA300)
H TCP303 (150 amps, 15 MHz, compatible with TCPA300)
H TCP404XL (500 amps*, 2 MHz, compatible with TCPA400)
*750 amps DC derated with duty cycle
If you are not familiar with these products, please refer to the Getting Started and Operating Basics chapters of this manual for basic operating information.
If you are an advanced user, the Reference section contains information on advanced applications as well as user diagnostic and troubleshooting informa­tion.
The Performance Verification and Adjustment Procedure sections support the qualification and calibration of the probes when used with either amplifier.
Manual Conventions
The Maintenance section supports the routine maintenance and repair of mechanical parts associated with the amplifiers.
The Glossary and Index are provided as quick reference locators for important information.
The term “amplifier” is used to refer to either the TCPA300 or TCPA400 when referring to common attributes. If a subject is unique to either amplifier, the amplifier will be referred to directly by model.
The terms “current probe” and “probe” are used to refer to any of the TCP300/400 Series current probes when referring to common attributes. If a subject is unique to a particular probe, the probe will be referred to directly by model.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
xv
Preface
xvi
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Getting Started

Getting Started

The TCPA300 and TCPA400 current probe amplifiers let you use one probe to simultaneously measure AC and DC current. The amplifiers convert the sensed current into a proportional voltage signal that you can measure directly with an oscilloscope.
The TCPA300 and TCPA400 current probe amplifiers provide better linearity than other current measurement systems because of a current feedback process used with the probe. DC measurement capability and high bandwidth allow the amplifiers to accurately represent square waves and fast-rise signals.
The TCPA300 and TCPA400 and associated probes provide these features:
H Simultaneous DC and AC current measurements up to 750 A peak
H High sensitivity
H One-button autobalancing and probe degaussing
H No adjustments needed to match a current probe to an individual amplifier
H AC or DC coupling of signal

System Configuration

H Direct scaling and unit readout on compatible TEKPROBE level II
oscilloscopes
A complete current measurement system consists of a current probe amplifier, a compatible current probe, and an appropriate oscilloscope. Refer to Figure 1--1.
AmplifierTest oscilloscope
Current probe
Input
Output
50 Ω oscilloscope input -- use the TEKPROBE Interface Cable or use a 50 Ω cable. (Add 50 Ω termination here if oscilloscope has only high-impedance input.)
Figure 1- 1: Typical TCPA300/400 current measurement system
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 1
Getting Started
TCPA300 and TCPA400
Current Probe Amplifiers
Current Probes
The amplifier amplifies the current sensed by the probe and converts the current to a proportional voltage that is displayed on an oscilloscope or other similar measuring device.
The following Tektronix current probes are compatible with the TCPA300 Amplifier:
H TCP312 (30 amps, 100 MHz)
H TCP305 (50 amps, 50 MHz)
H TCP303 (150 amps, 15 MHz)
The following Tektronix current probes are compatible with the TCPA400 Amplifier:
H TCP404XL (750 amps*, 2 MHz)
* 500 amps continuous, 750 amps DC derated with duty cycle
You can also use the CT-4 High-Current Transformer with the TCP305 and TCP312 current probes to extend the AC current measurement range to 20,000 peak amps.
Oscilloscope
An oscilloscope displays the output from the current measuring system. A 50 Ω cable is included to connect the amplifier to the oscilloscope input channel. A TEKPROBE-to-TEKPROBE interface cable is also included for connecting to TEKPROBE level II oscilloscopes.
If the oscilloscope does not have an input that can be set to 50 Ω impedance, you need a feedthrough 50 Ω termination. This termination is included as a standard accessory with your TCPA300 and TCPA400 Current Probe Amplifiers.
1- 2
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

Options

Getting Started
Table 1--1 lists options that are available for the TCPA300 and TCPA400 amplifiers.
Table 1- 1: Amplifier options
Option Description
A1 Universal Euro power cord
A2 United Kingdom power cord
A3 Australia power cord
A5 Switzerland power cord
A6 Japan power cord
AC China power cord
A99 No power cord
L5 Japanese Instruction Manual
Table 1--2 lists the Tektronix service options you can order for your amplifiers and probes. Designed to support tracking of calibration to requirements of ISO9000 and to provide for extended repair coverage, these options help fix your long-term maintenance costs and eliminate unplanned expenditures. Tektronix Service Options are available at the time you order your instrument. Contact your local Tektronix Sales Office for more information.
Table 1- 2: Service options
Option Description
D1 Provides the initial Test Data Report from the factory on delivery.
C3 Provides factory calibration certification on delivery, plus two more years of
calibration coverage. Throughout the coverage period, the instrument will be calibrated according to its Recommended Calibration Interval.
D3 Provides test data on delivery plus a Test Data Report for every calibration
performed during three years of coverage (requires Option C3).
R3 Extends product repair warranty to a total of three years.
C5 Provides factory calibration certification on delivery, plus four more years of
calibration coverage. Throughout the coverage period, the instrument will be calibrated according to its Recommended Calibration Interval.
D5 Provides test data on delivery plus a Test Data Report for every calibration
performed during five years of coverage (requires Option C5).
R5 Extends product repair warranty to a total of five years.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 3
Getting Started

Standard Accessories

The following accessories are shipped with the amplifiers and probes. Refer to the Replaceable Parts List beginning on page 8--1 for Tektronix part numbers to use in ordering accessories.
Amplifiers
Probes
The following accessories are shipped with the TCPA300 and TCPA400 amplifiers.
H Power Cord (customer-chosen option)
H BNC Cable
H Termination, 50 Ω, 2W
H TEKPROBE Interconnect Cable
H Instruction Manual (English or Japanese; customer-chosen language option)
H Certificate of Traceable Calibration
When you order a current probe, you will receive these accessories:
H Probe cover
H Probe ground lead, 6 inch length (TCP305 and TCP312 only)
H Instruction Sheet
H Certificate of Traceable Calibration
1- 4
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

Optional Accessories

Getting Started
You can order the following optional accessories for the amplifiers and probes. Refer to the Replaceable Parts List beginning on page 8--1 for Tektronix part numbers to use in ordering accessories.
H One-turn 50 Ω current loop. The current loop is used in the performance
verification procedure for checking the performance of the TCPA300 Amplifier and the compatible probes.
H CT-4 High-Current Transformer. If you need to measure high-amplitude AC
currents, consider using the CT-4 with the TCP303 and TCP312 probes. The CT-4 provides step-down ratios of 20:1 or 1000:1. For more information about the CT-4, consult your Tektronix sales representative.
H TCPA Calibration Adapter. Use the TCPA Calibration Adapter to verify the
amplifier(s) performance independent of the current probes.
H Travel Case. The travel case includes room to store one amplifier and two
current probes, along with related cables and adapters.
H Deskew Fixture. This fixture converts the PROBE COMPENSATION output
or TRIGGER OUTPUT of the TDS5000 or TDS7000 into a set of test point connections that allow you a convenient way to compensate for timing differences between voltage and current probes.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 5
Getting Started

Probe Covers

The TCP300/400 Series Current Probes come with a probe cover that stores the probe when not in use. Use the probe cover to hold your probe in a convenient place at your bench or workstation when you are not using it. You can attach the probe cover to the side of the bench to keep the probe off of your work surface. See Figure 1--2.
TCP305/312
TCP303/TCP404XL
Figure 1- 2: Using the probe covers
1- 6
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

Travel Case

Getting Started
The travel case is a recommended accessory for the TCP A300/400 Amplifiers. The travel case includes room to store one amplifier and two TCP300/400 Series Current Probes, one of each size. (For example, you can store a TCP305 and a TCP303 probe.) A compartment is included to store associated cables and terminations. See Figure 1--3 for the proper location of the equipment.
Instruction
manual
Large current
probe
Probe holders
Small current
probe
Figure 1- 3: Equipment locations in the travel case
Cables & terminations
Amplifier
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 7
Getting Started

Connecting the Amplifier to an Oscilloscope

You will need an oscilloscope to display the TCPA300 and TCPA400 measure­ment output. To use the full dynamic range of the probe/amplifier combination, the oscilloscope must be capable of displaying a vertical scale factor of 1 mV/div to 1V/div.
If you are using a TEKPROBE II-compatible oscilloscope, use the TEKPROBE­to-TEKPROBE interface cable. Otherwise, use the supplied 50 Ω BNC cable to connect the amplifier OUTPUT connector to your oscilloscope (see Figure 1--1 on page 1--1).
The input impedance of the oscilloscope channel must be 50 Ω, or you will encounter slowed pulse response, increased aberrations, or incorrect DC measurement amplitudes. If your oscilloscope provides only 1 MΩ inputs, you need to attach a 50 Ω feed-through termination between the oscilloscope input and the BNC cable. Do not install this termination at the amplifier end of the BNC cable.
To utilize the full bandwidth capability of the TCPA300 and TCPA400 and attached current probe, the oscilloscope bandwidth must be approximately five times that of the current probe. For example, when using a TCP312 Current Probe, the oscilloscope bandwidth must be at least 500 MHz. When using a TCP305 Current Probe, the oscilloscope bandwidth must be at least 250 MHz.

Power on the Amplifier

After you have connected the amplifier to the oscilloscope, allow the equipment to warm up to a stable temperature; usually 20 minutes is required.
Connect the power cord to the power input connector on the rear of the amplifier, and then connect the power cord to your local mains supply (100 VAC to 240 VAC, 50 Hz to 400 Hz). To allow for proper ventilation, place the rear panel of the amplifier at least 2 inches away from any obstructions. Set the amplifier on the bottom rubber feet, and keep papers and other items away from the bottom of the amplifier which could restrict airflow and cause overheating.
Power on the amplifier by pressing the ON/STANDBY button at the lower-left corner of the front panel. The amplifier goes through a self-test and cycles the front-panel LEDs.
NOTE. The amplifier stores the power state it is in when the power cord is unplugged. If you do not put the amplifier into STANDBY mode before unplug­ging it, the amplifier will power on immediately when you plug it in again.
When you connect a probe to the amplifier, the amplifier uses detection circuitry to indicate probe conditions such as noncompatible probe type and probe open.
1- 8
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

Connecting a Current Probe to the Amplifier

To connect a current probe to the amplifier input connector, align the red dot on the probe connector with the red dot on the amplifier front panel, as shown in Figure 1--4. Push the probe connector in to lock the connector. Do not twist the connector. The connector locks into place when you push it all the way into the front panel.
If you connect a probe to the wrong amplifier, (for example, a TCP312 to a TCPA400), the NONCOMPATIBLE PROBE TYPE LED illuminates. Discon­nect the probe and use the correct amplifier. The TCPA400 amplifier accepts TCP3XX probes, but will only operate properly with TCP4XX probes.
To disconnect the probe from the amplifier, pull back on the connector housing and pull the connector straight off the front panel.
CAUTION. Handle current probes with care. Do not drop a probe or subject it to impact, or the core may crack.
Getting Started
Do not connect or disconnect a current probe while the probe is clamped around a live conductor, or the probe may suffer electrical damage.
Align the red dots
Pull back collar,
then pull out connector
Connecting Disconnecting
Figure 1- 4: Connecting and disconnecting a current probe to the am plifier
Each current probe is calibrated before it is shipped, and should not require further adjustment. If a probe requires adjustment, refer to the Adjustment Procedure Overview on page 6--1 or contact your nearest Tektronix Service Center. The adjustment procedure must be performed only by qualified service personnel.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 9
Getting Started

Operating the Current Probe Slide

The current probes each have a slide mechanism that opens and closes the probe jaw. This allows you to clamp the probe around a conductor under test. The slide must be locked closed to accurately measure current or to degauss the probe. If a probe is unlocked, the PROBE OPEN indicator on the amplifier will light.
WARNING. Do not clamp the TCP305 or TCP312 current probes around uninsulated wires. Damage to the probe or personal injury may result. Only use the TCP305 or TCP312 current probes on INSULATED wires.
The TCP303 and TCP404XL current probes can be used to measure current on uninsulated wires. However , the circuit must be de-energized when connecting or removing the current probe.
Figure 1--5 illustrates the slide operation of the TCP305 and TCP312 current probes. To open the probe, pull the slide back until the jaw is open. To lock the probe, push the slide forward until the detent snaps into place.
Probe open Probe locked
Figure 1- 5: TCP312 and TCP305 slide operation
Figure 1--6 on page 1--11 illustrates the slide operation of the TCP303 and TCP404XL current probes. To open the probe, press the bottom of the lock button and squeeze the handle until the core is open. To lock the probe, release the squeeze handle and press the top of the lock button.
1- 10
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
(1) Unlock the probe
(2) Squeeze
the handle
(a) Opening the probe
Getting Started
(2) Lock the probe
(1) Release the handle
(b) Closing and locking the probe
Figure 1- 6: TCP303 and TCP404XL slide operation

Degaussing and Autobalancing the Current Probe

Degaussing the probe removes any residual magnetization from the probe core. Such residual magnetization can induce measurement error. Autobalancing removes unwanted DC offsets in the amplifier circuitry. Failure to degauss the probe is a leading cause of measurement errors. The DEGAUSS LED flashes until you degauss the probe.
To degauss the probe, disconnect the probe from the test circuit, or ensure that the conductor under test has no power, close and lock the slide, and then press the amplifier PROBE DEGAUSS AUTOBALANCE button on the front panel of the amplifier. To maintain measurement accuracy, degauss your probe in each of these cases:
H After you turn on the amplifier and allow a 20-minute warm-up period.
H Before you connect the probe to a conductor.
H Whenever a current or thermal overload condition occurs.
H Whenever you connect a new probe.
H Whenever you subject the probe to a strong external magnetic field.
H Periodically during normal use.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 11
Getting Started
To degauss and autobalance a current probe, perform these steps:
1. Verify that the current probe is connected to the amplifier.
2. Remove the current probe from the conductor under test.
3. Lock the probe slide closed (see Figures 1--5 and 1--6).
4. Press the amplifier PROBE DEGAUSS AUTOBALANCE button.
5. Wait about five seconds for the degauss procedure to complete.
The PROBE DEGAUSS AUTOBALANCE LED glows green when the operation has successfully completed. If the LED is blinking orange, the degauss operation is still in progress. If the LED is red, the operation failed, and the cause of the failure needs to be found and fixed. For more information, refer to the PROBE DEGAUSS AUTOBALANCE button discussion on page 2--2.
NOTE. The degauss procedure will fail if the amplifier is not properly connected to an oscilloscope having 50 Ω input impedance. If this occurs, the NOT TERMINATED INTO 50 Ω LED lights on the amplifier front panel.

DC Measurements

After you have completed the oscilloscope adjustments and the amplifier degauss/autobalance procedure, your system is ready to measure current.
To measure DC current, perform these steps:
1. Verify that the amplifier and the oscilloscope input coupling are set to DC,
and the input impedance is set to 50 Ω.
2. Lock the probe closed without a conductor passing through it.
3. Adjust the ground reference of the oscilloscope to move the trace to the
desired graticule line.
4. Press the amplifier PROBE DEGAUSS AUTOBALANCE button.
The NOT TERMINATED INTO 50 Ω LED is lighted if impedance is not 50 Ω. If this is the case, make necessary changes. (For example, use a 50 Ω termina­tion.)
5. After the degauss/autobalance routine completes, adjust the ground reference (if necessary) using the amplifier MANUAL BALANCE controls.
1- 12
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Getting Started
WARNING. Do not clamp the TCP305 or TCP312 current probes around uninsulated wires. Damage to the probe or personal injury may result. Only use the TCP305 or TCP312 current probes on INSULATED wires.
The TCP303 and TCP404XL current probes can be used to measure current on uninsulated wires. However , the circuit must be de-energized when connecting or removing the current probe.
6. Open the probe slide, place the probe around the conductor under test, and then lock the slide. For correct measurement polarity, make sure the probe arrow is pointing in the direction of conventional (positive to negative) current flow. Reversing the flow will display the current waveform upside­down on the oscilloscope.
7. Adjust the oscilloscope time base, trigger, and gain as needed.
Figure 1--7 shows a current probe connected to a power supply line. Notice that the probe arrow points toward the negative terminal of the power supply to conform to the conventional current flow of positive (+) to negative (--).
Current probe
Conventional current arrow
Figure 1- 7: Current probe polarity
Power supply
Load
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
1- 13
Getting Started

AC Measurements

To measure AC current only, and remove the DC component of the current being measured, follow the instructions below. These are identical to the instructions for DC current measurements except that the amplifier coupling in step 2 is set to AC.
1. Verify that the oscilloscope input coupling is set to DC.
2. Verify that the amplifier input coupling is AC, and the input impedance is set
to 50 Ω. (The NOT TERMINATED INTO 50 Ω LED is on if impedance is not 50 Ω.)
3. Adjust the ground reference of the oscilloscope to move the trace to the desired graticule line.
4. Lock the probe closed without a conductor passing through it, and then press the amplifier PROBE DEGAUSS AUTOBALANCE button.
WARNING. Do not clamp the TCP305 or TCP312 current probes around uninsulated wires. Damage to the probe or personal injury may result. Only use the TCP305 or TCP312 current probes on INSULATED wires.
The TCP303 and TCP404XL current probes can be used to measure current on uninsulated wires. However , the circuit must be de-energized when connecting or removing the current probe.
5. Open the probe slide, place the probe around the conductor under test, and then lock the slide. For correct measurement polarity, make sure the probe arrow is pointing in the direction of conventional (positive to negative) current flow. Reversing the flow will invert the displayed current waveform on the oscilloscope.
NOTE. Even when making AC current measurements, leave the oscilloscope coupling on DC. Change only the amplifier coupling to AC. Using the oscillo­scope AC coupling may cause the amplifier to exceed its output dynamic range.
6. Adjust the oscilloscope time base and trigger as needed.
1- 14
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Operating Basics

Control Summary

This section describes the function of each TCP A300 and TCPA400 front panel control and connector. The overview in Figure 2--1 is followed by a detailed description.
Some seldom-used functions do not appear in Figure 2--1. These functions are completely discussed in the detailed descriptions that follow Figure 2--1.
The PROBE DEGAUSS AUTOBALANCE button removes residual magnetism from the attached current probe. A multi-color LED indicates the status of the degauss circuit.
The MANUAL BALANCE buttons allow you to fine-adjust DC offset from the amplifier. The adjacent LED lights when one of the buttons has been pressed.
The four probe error lights indicate the following faults: PROBE OPEN, OVERLOAD (current or temperature), NOT TERMINATED INTO 50 and NONCOMPATIBLE PROBE TYPE.
The ON/STANDBY button turns on power to the amplifier.
Figure 2- 1: The TCPA300 front panel
The RANGE button toggles between the two scale factors that are available for the attached probe (TCPA300 only). LEDs indicate the selected range.
The COUPLING button selects AC or DC probe coupling, as indicated by the LEDs.
The current probes connect to the TCPA300 and TCPA400 at the PROBE INPUT connector.
The TCPA300 and TCPA400 output appears at the OUTPUT connector. Connect this to a 50 Ω input of your oscilloscope.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
2- 1
Control Summary

TCPA300 and TCPA400 Controls

These front panel controls and indicators are common to both the TCPA300 and TCPA400 current probe amplifiers, unless otherwise indicated.
PROBE DEGAUSS
AUTOBALANCE Button
and Indicator
When pressed, this button performs two functions that maximize measurement accuracy. First, the amplifier generates a degauss signal to remove any residual magnetism from the attached current probe. Second, the amplifier initiates an operation to remove any undesired DC offsets from the circuitry. During the degauss process, the amplifier is busy and cannot be used to measure current.
The indicator light next to the PROBE DEGAUSS AUTOBALANCE button blinks red whenever the amplifier detects that the current probe needs degaus­sing. The amplifier cannot detect all circumstances that require probe degaus­sing, so you may need to degauss the probe at times when the PROBE DEGAUSS AUTOBALANCE light is not blinking red. The red blinking light serves as a reminder to degauss the current probe when one of the following conditions occurs:
H The amplifier has just been turned on with a current probe connected.
H The current probe has been changed.
H An overload was detected.
To perform the probe degauss/autobalance function, remove the probe from all conductors (or ensure that the conductor under test has no power), make sure the probe is locked closed, and then press the PROBE DEGAUSS AUTOBA­LANCE button. The probe degauss/autobalance routine will not pass if the current probe is disconnected from the amplifier input, or if it is unlocked (the PROBE OPEN LED is on).
2- 2
The indicator blinks orange during the time the amplifier is busy performing the probe degauss functions. When the degauss and autobalance procedure is complete, the indicator light turns green.
The PROBE DEGAUSS AUTOBALANCE indicator light will be orange if the MANUAL BALANCE buttons have been pressed after a degauss has been successfully completed. This indicates that the DC offset value has been manually changed from the original value set during the degauss routine. Depending on the amount of offset (balance) you have entered with the MANUAL BALANCE buttons, another degauss operation may be necessary to ensure accurate measurements. Generally, if you change the DC offset by more than 5 divisions, you should de-energize the circuit under test and perform another degauss routine. Then, re-energize the circuit and take your measure­ments.
If the degauss operation has failed, and the AC and DC COUPLING LEDs are alternately flashing, this indicates the amplifier is displaying an error code with the four status LEDs on the lower-left front panel. If this occurs, refer to
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Control Summary
Displaying Error Codes with the Probe Degauss Autobalance Button on page 3--22.
NOTE. The NOT TERMINATED INTO 50 Ω LED is on when the amplifier output is not properly terminated into a 50 Ω load. Make sure your amplifier OUTPUT is connected to an oscilloscope input using a 50 Ω BNC cable, and that the oscilloscope input is set to 50 Ω impedance. See Figure 1--1 on page 1--1 for proper cabling.
If your oscilloscope does not have 50 Ω impedance settings for inputs, you can place a 50 Ω feed-through termination on the oscilloscope input and connect the amplifier output cable to the termination. Do not place the feed-through termination at the amplifier end of the BNC connecting cable.
MANUAL BALANCE
Buttons and Indicator

PROBE OPEN Indicator

OVERLOAD Indicator

The MANUAL BALANCE buttons allow you to fine-adjust the DC offset that appears at the amplifier OUTPUT connector. The manual balance adjustment only functions when the amplifier is set to DC coupling, and the MANUAL BALANCE indicator is only lighted after you press one of the MANUAL BALANCE buttons in DC coupling mode.
When lit, this indicator informs you that the current probe is unlocked. You must have the probe slide locked to degauss the probe or to accurately measure current.
When this LED is red, it informs you that the measurement you are taking exceeds the continuous amplitude limit of the TCPA300 and TCPA400. The detection circuits detect only low frequency and DC overloads. Since overloads can magnetize the probe, always degauss the probe after an overload. When the OVERLOAD indicator flashes red, it indicates the pulsed waveform is exceeding the amplitude limit of the amplifier.
When this LED is orange, it indicates that the safe operating temperature of the probe, and possibly the amplifier, has been exceeded. Disconnect the probe from the current source and allow time for the probe head and amplifier to cool.
When this LED blinks red and orange, it indicates that both the safe operating temperature of the probe and the current limit have been exceeded.
WARNING. To avoid personal injury or equipment damage, do not exceed the specified electrical limits of the TCPA300 and TCPA400 or any applicable accessories.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
2- 3
Control Summary
NOT TERMINATED
INTO 50 Ω Indicator
NONCOMPATIBLE
PROBE TYPE Indicator

ON/STANDBY Button

RANGE Button
(TCPA300 only)
When lit, this indicator informs you that the TEKPROBE interface cable or BNC cable from the OUTPUT of the amplifier is not connected to a 50 ohm input on the oscilloscope. You need to switch the termination setting on the oscilloscope to 50 Ω,orusea50Ω termination on the oscilloscope input.
NOTE. NOT TERMINATED INTO 50 Ω is only detected during the DEGAUSS AUTOBALANCE operation.
When lit, this indicator informs you that the probe that is connected to the amplifier is not designed to work with the amplifier. TCP3XX probes only work with the TCPA300 Amplifier, and the TCP404XL probe only works with the TCPA400 Amplifier.
Use this button to power on the amplifier. When the amplifier is in STANDBY mode, the amplifier is in a limited-power mode. Most of the secondary circuitry is disabled, but the line voltage remains connected to the amplifier power supply.
Press the RANGE button to toggle between the scale factors (sensitivity settings) of the probe attached to the TCPA300. If no RANGE LEDs are lit, this indicates a probe is not connected to the amplifier.
COUPLING Button and
Indicators
The COUPLING button determines the coupling between the TCPA300/TCPA400 and the oscilloscope. Press the COUPLING button to toggle between AC and DC coupling. To couple the amplifier for DC plus AC measurements, use DC coupling. For AC measurements only, use AC coupling. When the amplifier is set to AC coupling, the Manual Balance adjustment is disabled since the DC offset component is not visible on the output waveform.
NOTE. Even when making AC current measurements, leave the oscilloscope coupling on DC. Change only the amplifier coupling to AC. Using the oscillo­scope AC coupling may cause the amplifier to exceed its output dynamic range.
Under normal operation, the AC and DC COUPLING LEDs indicate the coupling mode of the amplifier. If they alternately flash after a degauss opera­tion, this indicates the amplifier is displaying an error code with the four status LEDs on the lower-left front panel. If this occurs, refer to Displaying Error Codes with the Probe Degauss Autobalance Button on page 3--22.
2- 4
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Control Summary

PROBE INPUT Connector

OUTPUT Connector

All current probes compatible with the TCPA300 and TCPA400 attach at the PROBE INPUT connector, which is a multi-pin female connector. For informa­tion about connecting a probe, see Connecting a Current Probe to the Amplifier on page 1--9.
The amplifier current measurement output is accessed at the OUTPUT connector, which should be connected to the oscilloscope input. Attach one end of a 50 Ω BNC cable to this connector and the other end to a 50 Ω vertical input of your oscilloscope. The output impedance of the amplifier is 50 Ω.
To get a direct readout of current on your Tektronix oscilloscope, use the TEKPROBE interface cable to connect the amplifier to your oscilloscope.
NOTE. To obtain accurate measurements, the input impedance of your oscillo­scope must be 50 Ω. Make sure your amplifier OUTPUT is connected to an oscilloscope input using a 50 Ω BNC cable, and that the oscilloscope input is set to 50 Ω impedance. See Figure 1--1 on page 1--1 for proper cabling.
If your oscilloscope does not have 50 Ω impedance settings for inputs, you can place a 50 Ω feedthrough termination on the oscilloscope input and connect the amplifier output cable to the termination. Do not place the feedthrough termination at the amplifier end of the BNC connecting cable.
Probe DC Gain Adjust
(located on probes)
After the PROBE DEGAUSS AUTOBALANCE routine has been run, the probe and amplifier system will meet all published specifications. However, if you want to improve the tolerance of the system accuracy, or to intentionally offset the accuracy to make up for total system errors, you can manually adjust the gain of the probe. Refer to Figures 6--3 and 6--5 on pages 6--7 and 6--12 for the adjustment locations.
NOTE. You should be careful to note the existing position of the DC Gain Adjustment before you alter it, so that you may return it to the initial, calibrated position. By altering the DC Gain Adjustment, you may cause the probe to not meet the warranted DC Accuracy specification.
For example, the typical accuracy of the TCP312 probe on the 1A/V range is 1%. If you want to measure a 3A p-p, 1kHz square wave and need increased accuracy (better than 0.25% is attainable), first adjust the TCP312 on the 1A/V range setting using a calibrated 3A p-p, 1kHz square wave source. Then, attach the probe to your circuit and take your measurement. Remember, altering the DC Gain Adjustment may cause the probe to not meet the warranted DC Accuracy specification.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
2- 5
Control Summary

GPIB Operation

The TCPA300 and TCPA400 Current Probe Amplifiers do not have direct GPIB connections. However, you can use a computer to control the oscilloscope that the amplifier is connected to, enabling you to change the time and scale factors of your current measurements over the oscilloscope GPIB bus.
Refer to your oscilloscope manual for instructions on using the GPIB bus.
2- 6
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Reference

Reference Notes

These notes are provided to help you utilize the full potential of the TCPA300 and TCPA400 current probe systems.

Degaussing a Probe with an Unpowered Conductor in the Jaws

Under almost all conditions, you can degauss your current probe while a conductor of an unpowered circuit is clamped in the jaws. The advantage of degaussing with an unpowered circuit is that any offset from stray DC magnetic fields are compensated. In an automated environment, degaussing with the conductor in the probe jaws eliminates the need to manually remove the probe.
NOTE. Be certain that the conductor in the probe jaws is completely unpowered. Any current flowing through the conductor will cause a residual offset in the current probe, and the amplifier may report an inaccurate result.
If the impedance of your circuit is higher than that shown in Table 3--1, the degauss procedure will succeed because the amplifier will be able to saturate the probe core. While degauss occurs, the probe will induce a voltage in the unpowered circuit. This also appears in Table 3--1. Your circuit must be able to absorb this induced voltage. With low impedance circuits, several amperes may be induced in the circuit being measured. This may be of concern when you are using very small conductors.
Table 3- 1: Unpowered circuit degauss limits
Probe type Minimum circuit resistance Maximum induced voltage
TCP312 10 mΩ 40 mV at 200 Hz
TCP305 5mΩ 40 mV at 200 Hz
TCP303 5mΩ 30 mV at 200 Hz
TCP404XL 1mΩ 15 mV at 100 Hz
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 1
Reference Notes

Measuring Differential Current

You can place two conductors in a current probe to provide differential or null current measurement. This avoids the necessity of using two current measure­ment systems with a differential oscilloscope.
WARNING. Do not put more than one uninsulated conductor at a time in the TCP303 or TCP404XL probes. Do not put any uninsulated conductors in the TCP305 or TCP312 probes. An uninsulated conductor is defined as any conductor without insulation or without insulation rated for the voltage present on the conductor under test.
An insulated conductor is defined as any conductor that is surrounded by an insulating material that is capable of isolating the voltage present on the conductor. Note that lacquer coatings like those typically found on transformer windings do not provide sufficient, reliable insulation for use with current probes. The lacquer coating can be easily nicked or damaged, compromising the insulating capabilities of the lacquer coating.
Do not force the slide closed. Damage to the probe may result. If you cannot close the slide around the conductor(s), either reduce the number of conductors you are measuring, or, if possible, take your measurement on a smaller conductor.
1. Orient the two conductors under test so that the polarities oppose each other. Clamp the current probe around the two conductors as shown in Figure 3--1. Be careful not to pinch a conductor in the probe jaws.
2. Measure the current. A waveform above the baseline indicates the conductor with the conventional current flow in the direction of the probe arrow is carrying the greater current. Conventional current flows from positive to negative.
3. To adjust for a current null, adjust the current in one of the conductors until the displayed measurement is zero.
3- 2
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

AC and DC Coupling

Reference Notes
Conductor #2
Conductor #1
Current
Current
Current probe
Figure 3- 1: Measuring differential current and nulls
You can couple the signal input to the TCPA300 and TCPA400 with either DC or AC coupling. DC coupling shows the DC and AC measurement components while AC coupling removes the DC component from the displayed signal. When you use AC coupling, make sure that the input DC current does not exceed the probe specifications.
AC coupling will affect waveforms at frequencies higher than the AC Coupling Low-Frequency Bandwidth. For example, pulsed currents may exhibit rolloff or decreased amplitude. Figure 3--2(a) shows a low-frequency square wave using AC coupling. The signal exhibits low-frequency rolloff. By changing the amplifier coupling to DC, the pulse is displayed as truly square, as shown in Figure 3--2(b).
(a) AC-Coupled signal
(b) DC-Coupled signal
Figure 3- 2: Effect of AC or DC coupling on low-fr equency signals
If you are trying to examine a low-frequency signal that is superimposed on a comparatively large DC component, you can resolve the signal by performing these steps:
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 3
Reference Notes
1. Select the range setting that will display the maximum detail without
2. Adjust the oscilloscope V/div sensitivity (A/div if using the TEKPROBE

Maximum Current Limits

Current probes have three maximum current ratings: continuous, pulsed, and Ampere-second product. Exceeding any of these ratings can saturate the probe core, magnetizing the core and causing measurement errors. See Specifications beginning on page 4--1 for the maximum current ratings of compatible probes.
H Maximum Continuous Current refers to the maximum current that can be
H Maximum Pulsed Current refers to the maximum peak value of pulsed
exceeding the dynamic range of the signal.
interface), to display maximum signal detail.
continuously measured at DC or at a specified AC frequency. The maximum continuous current value is derated with frequency; as the frequency increases, the maximum continuous current rating decreases.
current the probe can accurately measure, regardless of how short (within bandwidth limitations) the pulse duration is.
Procedure A
H Ampere-Second Product defines the maximum width of pulsed current that
you can measure when the pulse amplitude is between the maximum continuous and maximum pulsed current specifications. The maximum continuous specification itself varies by frequency.
NOTE. Always degauss the probe after measuring a current that exceeds the maximum continuous current, maximum pulsed current, or Ampere-second product rating of the probe. Exceeding these ratings can magnetize the probe and cause measurement errors.
To determine if your measurement exceeds the Ampere-second product, perform either Procedure A or Procedure B.
To determine the maximum allowable pulse width, measure the peak current of the pulse (see Figure 3--3a). Divide the Ampere-second (or Ampere-microse­cond) specification of your probe by the measured peak current of the pulse. The quotient is the maximum allowable pulse width; the pulse width at the 50% point of the measured signal must be less than this value.
For example, the TCP312 Current Probe has a maximum Ampere-second product of 500 As in the 10 A/V range setting. If a pulse measured with a TCP312 has a peak current of 40 A, the maximum allowable pulse width would be 500 As divided by 40 A, or 12.5 s.
3- 4
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Reference Notes
I
max
p
Pulse width at 50%
50%
I
max
c
0A
(a) Maximum allowable pulse width (b) Maximum allowable pulse amplitude
Figure 3- 3: Applying the amp-second product rule
Do Not Exceed
Pulse width at 50%
50%
Maximum pulsed current
Maximum continuous current
Procedure B
To determine the maximum allowable pulse amplitude, measure the pulse width at the 50% points (see Figure 3--3b). Divide the Ampere-second (or Ampere­microsecond) specification of your probe by the pulse width. The quotient is the maximum allowable current; the peak amplitude of the measured pulse must be less than this value.
For example, the TCP312 Current Probe has a maximum Ampere-second product of 500 As in the 10 A/V range setting. If a pulse measured with a TCP312 probe has a width of 15 s, the maximum allowable peak current would be 500 As divided by 15 s, or 33.3 A.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 5
Reference Notes

Measuring Noncontinuous Current with the TCP404XL Probe

When you measure a noncontinuous current with the TCP404XL probe, you need to take into consideration several factors to ensure that you make accurate measurements and do not trip the thermal overload circuit. The amplitude and duty cycle of the continuous and noncontinuous current, and the ambient temperature, all affect the maximum amount of time allowed for the measure­ment, which defines the safe operating area of the probe.
You can calculate the duty cycle of the continuous and noncontinuous current when you know the duration of the noncontinuous current (defined in Figure 3--4 as measurement time) and the measurement period. Figure 3--4 illustrates a continuous and noncontinuous current and how to calculate the duty cycle.
Measured
current (A)
Noncontinuous
current level (A)
Continuous
current level (A)
Measurement
time (min)
Measurement
period (min)
Duty cycle (%) = 100x
Measurement time
Measurement period
Time (min)
Figure 3- 4: Duty cycle calculation
After you calculate the duty cycle of the noncontinuous current you are measuring, you can use the other factors (continuous and noncontinuous current amplitude, etc.) to compare your measurement to those shown in Figures 4--8 through 4--10 in the Specifications section.
You can see how duty cycle affects the measurement time by looking at any of the three graphs in Figures 4--8 through 4--10. As the duty cycle increases on the x-axis, the measurement time (of the noncontinuous current) on the y-axis decreases.
Figures 4--8 and 4--9 on page 4--12 illustrate the effect of measuring two different noncontinuous current amplitudes, 750 amps and 600 amps respectively, with
3- 6
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Reference Notes
varying duty cycles and levels of continuous current. The ambient temperature in these two examples is kept constant at 50 °C.
WARNING. When using the probe near the upper current limit and maximum ambient temperature for extended lengths of time, the probe head surface can become hot to the touch. To prevent injury, keep your hands away from the probe head until it has had time to cool after disconnecting the probe from the circuit.
To see how noncontinuous current amplitude affects measurement time, look at the curves for measurements of 200 amps continuous between the two graphs in Figures 4--8 and 4--9 on page 4--12. Compare the maximum measurement time allowed for a duty cycle of 20%: At 750 amps, you have 3 minutes of safe measurement time, versus 17 minutes for a smaller noncontinuous-current amplitude of 600 amps.
By looking at any of the three graphs, you can also see that when you measure a noncontinuous current having the same amplitude and duty cycle, the measure­ment time decreases as the continuous-current amplitude increases.
Finally, compare the two graphs in Figures 4--8 and 4--10 on pages 4--12 and 4--13. Here, the effect of ambient temperature on measurement time is illustrated. Given a continuous current of 200 A with a noncontinuous current of 750 A, and having a 20% duty cycle, a 27 °C increase in temperature yields a 12 minute decrease in maximum measurement time.
Keep these factors into account when taking measurements to ensure accuracy and to protect both yourself from injury and the equipment from damage.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 7
Reference Notes

Extending Current Range

You may encounter situations where your measurement exceeds the maximum current rating of the connected probe. This section discusses methods for extending AC and DC current ranges without exceeding specified limits.
WARNING. To avoid personal injury or equipment damage, do not exceed the specified electrical limits of the TCPA300 and TCPA400 or any applicable accessories. When using multiple conductors, do not exceed current limits on either conductor.
Extending DC Range
If you want to measure a low-amplitude AC component that is superimposed on an extremely large steady state DC component (such as in a power supply), or if you want to extend the DC current range of your probe, you can add offset (bucking) current with a second conductor.
WARNING. Do not put more than one uninsulated conductor at a time in the probe jaws. An uninsulated conductor is defined as any conductor without insulation or without insulation rated for the voltage present on the conductor under test.
To supply additional bucking current, place a second conductor that has a pure DC component of known value in the probe jaw with the conductor under test, as shown in Figure 3--5(a). Orient the second conductor so that the bucking current flows in the opposite direction of the DC flow in the conductor under test.
You can increase the value of the bucking current by winding multiple turns of the second conductor around the probe, as shown in Figure 3--5(b). The bucking current is equal to the current flowing in the conductor, multiplied by the number of turns wound around the probe. For example, if the second conductor has a current of 100 mA DC and is wrapped around the probe five times, the DC bucking current is 100 mA multiplied by 5, or 500 mA DC.
3- 8
To determine measurement values, add the value of the bucking current to the displayed measurement.
NOTE. Adding a second conductor to the probe increases the insertion imped­ance and reduces the upper bandwidth limit of the probe. Winding multiple turns further increases the insertion impedance, further reducing the upper bandwidth limit.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Reference Notes
Current
Current
Conductor
under test
Bucking current
supplied by
second conductor
Current probe
(a) Adding a second conductor (b) Adding multiple turns
Figure 3- 5: Increasing the DC measurement range
Extending AC Range
You can extend the AC amplitude limit of the TCPA300 by using the Tektronix CT-4 High-Current Transformer. The CT-4, designed for use with the TCP305 and TCP312 current probes, extends the current probe range by a factor of 20:1 or 1000:1. The CT-4 can provide external steady-state DC bucking current up to 300 A. For more information about the CT-4, consult your Tektronix sales representative.
Current
Current
Current probe
Conductor
under test
Extra turns added
to increase
bucking current
Because the CT-4 has a lower AC bandwidth limit than the TCP305 and TCP312 current probes, set the amplifier coupling to DC when using the CT-4.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 9
Reference Notes

Increasing Sensitivity

If you are measuring DC or low-frequency AC signals of very small amplitudes, you can increase measurement sensitivity of your Current Probe by winding several turns of the conductor under test around the probe as shown in Figure 3--6. The signal is multiplied by the number of turns around the probe.
WARNING. Do not put more than one uninsulated conductor at a time in the probe jaws. An uninsulated conductor is defined as any conductor without insulation or without insulation rated for the voltage present on the conductor under test.
When viewing the signal on the oscilloscope screen, divide the displayed amplitude by the number of turns to obtain the actual current value. For example, if a conductor is wrapped around the probe five times and the oscilloscope shows a reading of 5 mA DC, the actual current flow is 5 mA divided by 5, or 1mADC.
NOTE. Winding multiple turns around the probe increases insertion impedance and reduces the upper bandwidth limit of the probe.
Conductor
under test
Extra turns for
increased sensitivity
Current probe
Figure 3- 6: Increasing probe sensitivity
3- 10
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

Application Notes

This section describes some of the typical measurement applications of the TCPA300 and TCPA400 Current Probe Amplifiers:
H Automobile Charging Systems
H Inductance Measurements
H Continuity Test of Multiple-Conductor Cable
H Measuring Inductor Turns Count
H Power Measurement and Analysis Software Applications

Automobile Charging Systems

Most automotive charging systems are three-phase alternators with a diode rectifier network. A meter averages current from all three phases, and cannot detect a single-phase diode problem. Observing the charge current waveform can quickly reveal if one diode is shorted or open.
The diagram in Figure 3--7 shows the equipment setup. Refer to Table 3--2 for the test equipment setup. A TCP303 Current Probe was used for this high-cur­rent, low-voltage application. The TCP303 is degaussed and clamped around the positive battery lead from the alternator. The probe arrow is pointed away from the alternator and toward the battery side of the circuit, to reflect conventional current. The automobile is started and the lights are turned on to add a significant load to the circuit.
Table 3- 2: Automobile charging systems test setup
Instrument Control Setting
TCPA300 COUPLING DC
RANGE 50 A/V
Oscilloscope Coupling DC
Amps/Division (Volts/Division if not using TEK­PROBE interface cable)
Zero-Current Reference Center graticule line
Time Base 200 s/division
10 A/Div (200 mV/Div)
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 11
Application Notes
TCP303/404XL
+
---
Battery Alternator
To
ground
To current probe amplifier
Figure 3- 7: Setup for measuring charging current
The waveform in Figure 3--8(a) shows the three-phase ripple frequency. The average charge current is approximately 27 A with a minimum peak of approxi­mately 23 A and a maximum peak of approximately 31 A. The waveform shows a continuous cycle with no dropouts, so the alternator circuit appears to be functioning properly. A single-phase diode failure normally appears as an extreme drop in charge current every third cycle, as shown in Figure 3--8(b).
+
---
To ground
(a) Normal waveform (b) Waveform with one bad phase
Figure 3- 8: Charge current waveforms
3- 12
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
40A
0A

Inductance Measurements

You can use the TCPA300 and TCPA400 to measure inductance of coils. Two different methods can be used: one for low-impedance pulse sources and another for high-impedance pulse sources of known value.
Application Notes
Low-Impedance Pulse
Sources
Figure 3--9 shows a measurement setup using a constant-voltage pulse generator of extremely low output impedance. The inductor is connected across the output terminals of the pulse source. The current probe is attached to one of the source leads and the current ramp is measured.
The inductance is effectively defined by the slope of the current ramp, shown in Figure 3--10, and is mathematically expressed by the following formula:
E
L =
di dt
where L is the inductance in henries, E is the voltage of the pulse generator, dt is the change in time, and di is the change in current.
Current
Pulse
generator
flow (i)
Inductor
Current
probe
Figure 3- 9: Measuring inductance with a low-impedance source
NOTE. If the probe impedance is a significant part of the total circuit inductance, measurement accuracy will be affected. Refer to the probe specifications for probe insertion impedance.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 13
Application Notes
Current
flow (i)
d
i
Time (t)
d
t
Figure 3- 10: Linear current vs. time ramp
High-Impedance Pulse
Sources
If the pulse source has a higher impedance of known resistance, such that the output voltage drops as the current increases, the inductance of a coil can be calculated by the time constant of the charge curve. Figure 3--11 shows the setup diagram, which is similar to the previous example. The current ramp represented in Figure 3--12 shows how the values for the inductance formula are obtained.
Use this formula to calculate the inductance based on the current measurement:
L = τ R
where L is the inductance in henries, τ is the time required for the current to rise or fall 63.2% of the total current value, and R is the source resistance of the pulse generator.
3- 14
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Application Notes
Resistance
Pulse
generator
Current
probe
L
Current
flow (i)
Inductance
Figure 3- 11: Measuring inductance with a high-impedance source
Current
flow (i)
100%
63.2%
36.8%
0
Figure 3- 12: High-impedance source current ramp

Continuity Test of Multiple-Conductor Cable

Single conductors in a multiconductor cable can be tested with the TCPA300 and TCPA400. To check a conductor, clamp the current probe around the cable bundle and check for a specific, known current signal. If there is no current or the current is abnormally low, then the conductor has a continuity problem. If the current is abnormally high, then the conductor may have a short to ground.
ττ
Time (t)
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 15
Application Notes

Measuring Inductor Turns Count

To obtain an approximate turns count of an inductor, connect the inductor to a current limited source, as shown in Figure 3--13. Measure the input current on one of the inductor leads, then clamp the current probe around the inductor and note the current value. The number of turns is equal to the ratio of coil current to input current. The accuracy of this method is limited by the current measurement accuracy. The following method allows more precise turns measurement.
For a more precise turns count, you need a coil with a known number of turns to use as a reference. The measurement setup is similar to the previously described one, except the reference coil and the test coil are inserted into the current probe so that the currents oppose each other (see Figure 3--14). You must observe the polarity of coil current to determine whether the test coil has more or fewer turns than the reference coil. The turns are calculated by using the formula:
N2= N1×
where N2is the number of turns in the test coil, N1is the number of turns in the reference coil, I
I I
m
1
is the measured coil current, and I1is the input current.
m
Measure input current here
Current flow (i)
Coil
Figure 3- 13: Measuring the number of turns in a coil
Clamp probe around coil to measure current from coil turns
3- 16
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCP303 Current probe
Current flow
in coil #1
Input current
Current flow
in coil #2
Figure 3- 14: Turns measurement using reference coil
Application Notes

Power Measurement and Analysis Software

Power measurement software that is available for some Tektronix oscilloscopes transform your current measurement system into a sophisticated analysis tool that quickly measures and analyzes many different applications, such as:
H Power dissipation at switching devices and magnetic components in
switching power supplies
H Precompliance test for EN61000--3--2 standard
H Power quality
H Modulation analysis
After making the measurements, the software generates detailed test reports in customizable formats. When the software is used with a Tektronix TDS5000 Series or TDS7054/TDS7104 digital phosphor oscilloscope and differential voltage and current probes, it forms a complete measurement system for power supply design and test.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 17
Application Notes
3- 18
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

Troubleshooting and Error Codes

Table 3--3 lists possible problems that you may encounter when measuring current with the TCPA300 and TCPA400. Use this as a quick troubleshooting reference.
Table 3- 3: Troubleshooting
Problem Remedy
Amplifier will not power on Check that the amplifier is plugged into a working AC outlet.
Defective amplifier. See HW Troubleshooting on page 7--21 to help determine which module is defective, or refer the instrument to qualified service personnel for repair.
Front panel displays an error status. Error Status LEDs are on the lower-left section of the amplifier front panel
Note the error status and correct the condition. The error status is labeled on the front panel. For example, if the PROBE OPEN LED is lit, close and lock the probe jaws.
The AC and DC COUPLING LEDs are flashing alternately
All LEDs are flashing All LEDs flashing indicates a thermal shutdown. Power-cycle the system and let it
OVERLOAD LED remains lit red after removing probe from circuit.
OVERLOAD LED remains lit orange Degauss the probe. If the OVERLOAD LED remains lit orange, disconnect the probe
Probe does not degauss successfully and LEDs are lighted in an unfamiliar pattern
Cannot degauss the probe Current Probe is not locked. Lock the current probe.
The Error Status LEDs are displaying an error code. Error codes are described on page 3--22. Note the error code and turn the instrument off, then on. If the error condition persists, refer the instrument to qualified service personnel for repair.
If the RANGE and DEGAUSS LED are both off, and one of the COUPLING LEDs is lit, this indicates the amplifier is not detecti ng a probe. Reconnect the probe.
All LEDs flashing indicate a thermal shutdown. Power-cycle the system and let it cool for 15 minutes before taking measurements. In most cases, 15 minutes is sufficient.
cool before taking measurements. In most cases, 15 minutes is sufficient.
Probe transformer or amplifier main board is defective.
from the circuit and let it cool for 15 minutes before degaussing again. If this does not clear the LED, the probe transformer or probe Hall device may be defective. Connect another probe to the amplifier, or connect the probe in question to another amplifier.
Amplifier is displaying an error code or is out of calibration or defective. Refer to Interpreting the Error Code display on page 3--23. If another amplifier is available, try to degauss the probe on the other amplifier, to help isolate the cause.
Component failure on probe circuit board: broken adjustment/wire/etc. Disassemble probe, repair/replace component or refer the instrument to qualified service personnel for repair.
Current Probe is improperly connected to the amplifier input connector.
Amplifier output is not terminated into 50 Ω load. Set input impedance of oscilloscope to 50 Ω or connect a 50 Ω feed-through termination at the oscilloscope input, not at the amplifier output.
Current Probe is defective or not compatible with the amplifier type.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 19
Troubleshooting and Error Codes
Table 3- 3: Troubleshooting (Cont.)
Problem Remedy
Degauss takes longer than 10 seconds Probe is attached to an energized circuit. Disconnect probe from ci rcuit and retry.
Probe is faulty -- Probe transformer (defective Hall device with excessive noise or drift) may cause this symptom. May also be caused by a shorted or open wire in the probe cable assembly.
Defective main board -- If the probe is not the cause, then it is most likely a defective main board in the amplifier.
Cannot make a current measurement -- no measurement output from amplifier
MANUAL BALANCE will not adjust Oscilloscope or amplifier input coupling not set to DC. MANUAL BALANCE buttons
Current Probe is not locked. Lock the current probe.
Current Probe is improperly connected to the amplifier input connector.
Amplifier COUPLING is set to AC. Set the COUPLING to DC.
A Degauss/Autobalance routine has not been completed successfully on the system. Degauss the probe.
Oscilloscope/Amplifier is not set to an appropriate sensitivity setting.
Defective interface cable (between oscilloscope and amplifier).
only function when the amplifier is in the DC coupling mode.
Oscilloscope/Amplifier is not set to an appropriate sensitivity setting.
Stray DC component in measurement Degauss the probe using the PROBE DEGAUSS AUTOBALANCE button. (The
Current Probe has been overloaded, dropped, or exposed to magnetic field.)
You can use the MANUAL BALANCE buttons to null out any residual DC offset after per­forming a Probe Degauss Autobalance routine.
Measurements are inaccurate Degauss the probe.
The amplifier output is not terminated into 50 Ω load. Set the input impedance of oscilloscope to 50 Ω or connect a 50 Ω feedthrough termination at the oscilloscope input. Do not attach the termination at the amplifier output.
The measurement exceeds the maximum continuous current or Ampere-second product ratings of the Current Probe. If possible, upgrade to a probe with a higher current rating or use a CT-4 transformer.
Amplifier or current probe out of calibration. Refer to Performance Verification on page 5--1.
The jaw mechanism is dirty -- disassemble probe, clean and lubricate. Refer to the probe disassembly procedures on page 7--13.
Defective current probe transformer.
Measurements roll off at high frequencies The oscilloscope bandwidth limit is turned on. Verify that the bandwidth limit switch on
the oscilloscope is set to the desired bandwidth position.
(Be careful not to exceed the frequency limit of the probe used. Frequencies above the probe design may cause overheating and damage to the probe.)
3- 20
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Troubleshooting and Error Codes
Table 3- 3: Troubleshooting (Cont.)
Problem Remedy
Measurements exhibit excessive noise Current Probe is not locked. Lock the current probe.
Current Probe is improperly connected to the amplifier input.
Amplifier output is not terminated into 50 Ω load. Set input impedance of oscilloscope to 50 Ω or connect a 50 Ω feed-through termination at the oscilloscope input, not at the amplifier output.
Defective current probe transformer.
Measurement aberrations exceed the specified limit
Excessive low frequency droop in pulse response or low DC gain accuracy
Measurements exhibit excessive delay or slowed pulse response
Probe jaw not opening and closing freely The jaw mechanism is dirty -- disassemble probe, clean and lubricate. Refer to the
The amplifier output is not terminated into 50 Ω load. Set the input impedance of the oscilloscope to 50 Ω or connect a 50 Ω feedthrough termination (see the replaceable parts list on page 8--6) at the oscilloscope input. Do not attach the termination to amplifier output.
The measurement exceeds the maximum continuous current or Ampere-second product ratings of the Current Probe. (For more informat ion, see Current Limits on page 3--4.) If possible, upgrade to a probe with a higher current rating or use a CT-4 transformer.
Check that the probe slider is fully closed and locked.
Check and clean the probe transformer and lid.
Degauss the probe.
The amplifier or the oscilloscope may be AC coupled. Set the COUPLING to DC.
The amplifier output is not terminated into 50 Ω load. Set the input impedance of the oscilloscope to 50 Ω or connect a 50 Ω feedthrough termination (see the replaceable parts list on page 8--6) at the oscilloscope input. Do not attach the termination to the amplifier output.
The measurement exceeds the Ampere-second product of the Current Probe. If possible, upgrade to a probe with a higher current rating or use a CT-4 transformer.
The oscilloscope bandwidth limit is turned on. Verify that the bandwidth limit switch on the oscilloscope is set to the desired bandwidth position.
probe disassembly procedures on page 7--13.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 21
Troubleshooting and Error Codes

Displaying Error Codes with the Probe Degauss Autobalance Button

This section describes the error codes that the amplifiers display using the function indicator LEDs.
When an internal error condition exists, the amplifiers may generate error codes when you press the PROBE DEGAUSS AUTOBALANCE button. The AC and DC Coupling LEDs will flash alternately to indicate that error codes are being displayed instead of normal conditions.
The four fault indicator LEDs above the ON/STANDBY switch are used to form a four-bit binary error code. The PROBE OPEN LED indicates the most significant bit (MSB), and the NONCOMPATIBLE PROBE TYPE LED indicates the least significant bit (LSB). See Figure 3--15.
To continue past an error code, press any button except ON/STANDBY. However, the degauss will fail until the internal error condition is corrected and the degauss operation is run again.
Press the PROBE DEGAUSS AUTO
BALANCE button
AC and DC LEDs flash alternately to indicate
Four--bit error code
is displayed here
MSB
LSB
error codes are being displayed
Figure 3- 15: Error code display
For example, after you press the PROBE DEGAUSS AUTOBALANCE button, if the AC and DC Coupling LEDs are flashing, and the NOT TERMINATED INTO 50 Ω LED is illuminated, this indicates an Error Code 2 -- a null error in the DC offset circuit occurred. See Figure 3--16 for an illustrated example.
3- 22
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Four--bit error code
hei
alD
Cof
f
lif
i
I
fth
c
c
lif
i
vic
p
p
plylev
)
thepowersupplylevelsforth
e
1)Removetheprobefromthecircui
t
asadj
O
verloa
dtri
I
fth
c
c
c
ircui
89Unuse
d
hei
alD
Cof
f
lif
i
I
fth
c
c
c
ircui
1
3
w
ireint
h
cab
lea
bly
is displayed here
Figure 3- 16: Interpreting the error code display
Table 3--4 shows the complete list of error codes for the amplifier.
Table 3- 4: Amplifier error codes
Code Description of error Action to take
Troubleshooting and Error Codes
MSB
=
LSB
= binary 0010 = code 2 =
a null error in the DC offset occurred
1 The amplifier detected a probe, but a
valid Hall device was not detected.
2 An error occurred while nulling out
t
3
ntern
er.
set oftheamp
4 TCPA300 ONLY -- An error occurred
while the the amplifier was adjusting the
ower su
5
different Range settings.
els for the
6 An error occurred while the amplifier
w
7
points.
usting the
p
8 Unused
10 An error occurred while nulling out
11
12
ntern
er.
The amplifier cannot null out the
set oftheamp
t
probe DC offset voltage.
14
Disconnect and reattach the probe to the amplifier. Power cycle the amplifier. Check probe transformer (defective Hall device). If these steps do not resolve the error, the amplifier needs service. May also be caused by a shorted or open wire in the probe cable assembly.
Power cycle the amplifier and run the Degauss/Autobalance adjustment routine again.
-
e error reo
urs, then theamp
er needsser
e.
Power cycle the amplifier and run the Degauss/Autobalance adjustment routine again. If the error reoccurs, then perform the following steps: 1
Removetheprobefrom thecircuit.
.
2) Check probe -- The signal path through the probe transformer may be open or shorted, which can be caused by defective probe transformer or cable assembly. If the above steps do not resolve the error, the amplifier needs servi c e.
Power cycle the amplifier and run the Degauss/Autobalance adjustment routine again.
e error reo
urs, then removetheprobefrom the
t.
If this does not resolve the error, the amplifier needs service.
Power cycle the amplifier and run the Degauss/Autobalance adjustment routine again.
-
e error reo
urs, then removetheprobefrom the
t.
If this does not resolve the error, the amplifier needs service.
Remove the probe from the circuit. Error 12 will most likely occur when the probe is connected to an energized circuit when a Degauss/Autobalance routine is initiated. Check probe transformer (defectiveHall device).
This error may also be caused by a Hall device with excessive noise or drift in the probetransformer, orashortedor open
eprobe
ssem
. Power cycle the amplifier and run the Degauss/Autobalance adjustment routine again. If these steps do not resolve the error, the amplifier needs service.
15 Unspecified error (internal software
error)
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Power cycle the amplifier and run the Degauss/Autobalance adjustment routine again. If the error reoccurs, then the amplifier needs service.
3- 23
Troubleshooting and Error Codes

Correcting the Cause of an Error Code

After interpreting the error code, disconnect the probe from the circuit and degauss the probe again. Take the measurement again. If the error code persists, contact your Tektronix Service Center.

Shutdown Error

If all LEDs flash at the same time, a shutdown condition exists. If this occurs, turn off the amplifier and disconnect the probe from the circuit under test. Turning off the amplifier clears the error register.
A shutdown condition can be caused by exceeding the frequency/current specifications. Verify that you are not exceeding the specifications and let the probe and amplifier cool before using them again. In most cases, 15 minutes is sufficient. If the measurement you were taking was within the probe and amplifier specifications, degauss the probe and take the measurement again. If the shutdown error persists, contact your Tektronix Service Center.
3- 24
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Specifications

Specifications

These specifications are valid only under the following conditions:
H The probe and amplifier have been calibrated at an ambient temperature of
23 _±5 _C.
H The probe and amplifier are operating within the environmental limits
described in Table 4--6 on page 4--5 (the operating temperature limits are 0 _Cto+50_C, unless otherwise stated).
H The probe and amplifier have had a warm-up period of at least 20 minutes.
H The probe degauss/autobalance routine has been performed after the
20-minute warm-up period, and thereafter whenever the PROBE DEGAUSS/AUTOBALANCE light blinks.
H The amplifier output is properly terminated into 50 Ω.
Specifications are separated into two categories: warranted specifications and nominal or typical characteristics.

Warranted Specifications

Warranted specifications, Table 4--1, are guaranteed performance specifications unless specifically designated as typical or nominal.
Table 4- 1: Warranted TCPA300 and TCPA400 specifications
Amplifier TCPA300..................... TCPA400
DC Gain Accuracy 1% 1%
Installed probe TCP312 TCP305 TCP303 TCP404XL
Bandwidth (--3 dB) DC to 100 MHz DC to 50 MHz DC to 15 MHz DC to 2 MHz
Rise Time, 10% to 90% 3.5 ns 7ns 23 ns 175 ns
DC Gain Accuracy: Warranted 3% 3% 3%
2
.............. ≤1% ≤1% ≤1% ≤1%
1
Warranted from 10 _Cto50_C. For temperature range of 0 _Cto<10 _C, spec is +3%/- 6%.
2
23 _C ±5 _C
Typical
1
3%
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
4- 1
Specifications
(
)
(
)
(
)

Nominal and Typical Characteristics

Nominal and typical characteristics, listed in Table 4--2, are not guaranteed. They are provided to characterize the configuration, performance, or operation of typical probe/amplifier combinations.
Table 4- 2: Nominal and typical amplifier characteristics
Installed probe
Parameter
Ranges, nominal 1A/V,
Maximum Amp-Second Product (See Figures 4--2 through 4--5 for frequency derating)
Input Coupling, nominal AC, DC AC, DC AC, DC AC, DC
AC Coupling Low­Frequency Bandwidth, (low-pass --3dB point), typical
Displayed Noise, typical 75 A
Signal Delay, typical, (includes probe, amplifier, and TEKPROBE­to-TEKPROBE interface cable, or 012-0117-00 BNC cable)
Maximum Bare Wire Voltage Use with insulated
Lowest Measurable Current 1mA 5mA 5mA 1A
Insertion Impedance (see Figure 4--6 for plots)
TCP312
10 A/V
1 A/V -- 50A*s 10 A/V -- 500A*s
<7 Hz <7 Hz <7 Hz <7 Hz
RMS
Limit measurement bandwidth to 20 MHz.
approximately 17 ns approximately 19 ns approximately 53 ns approximately 103 ns
wires only
1 MHz 0.08Ω 10 MHz 0.15Ω 50 MHz 0.27Ω 100 MHz 0.7Ω
TCP305 TCP303 TCP404XL
5A/V, 10 A/V
5 A/V -- 500A*s 10 A/V -- NA
500 A Limit measurement bandwidth to 20 MHz.
Use with insulated wires only
1 MHz 0.035Ω 10 MHz 0.12Ω 50 MHz 0.04Ω
RMS
5A/V, 50 A/V
5 A/V -- 3000A*s 50 A/V -- 15000A*s
500 A Limit measurement bandwidth to 20 MHz.
600VRMSCATI&II 300 VRMS CAT III
1 MHz 0.01Ω 5 MHz 0.025Ω 15 MHz 0.1Ω
RMS
1A/mV
NA
70 mA Limit measurement bandwidth to 20 MHz.
600VRMSCATI&II 300 VRMS CAT III
10 kHz 0.1 mΩ 100 kHz 0.6 mΩ 1MHz 8.0mΩ 2MHz 16.0mΩ
RMS
Current Transfer Ratio 1 V/Amp and
100 mV/Amp
Maximum Current Ratings at Sensitivity Ranges
High Current Sensitivity Range 10 A/V Range 10 A/V Range 50 A/V Range 1 A/ mV Range
DC (continuous).......
DC (noncontinuous).......
RMS (sinusoidal).......
Peak Pulse.......
4- 2
30 A
N/A
21 A
50 A 50 A 500 A 750 A
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
200 mV/Amp and 100 mV/Amp
50 A 150 A 500 A
N/A N/A 750 A
35 A 150 A 500 A
200 mV/Amp and 20 mV/Amp
1 mV/Amp
Table 4- 2: Nominal and typical amplifier characteristics (Cont.)
(
)
(
)
(
)
Specifications
Installed probe
Parameter
Parameter
Low Current Sensitivity Range 1 A/V Range 5 A/V Range 5 A/V Range N/A
DC (continuous).......
DC (noncontinuous).......
RMS (sinusoidal).......
Peak Pulse.......
Input Voltage 100--240 VAC (±10%), 47 Hz to 440 Hz, single phase
Maximum Power 50 Watts maximum
Fuse Rating 3.15 AH, 250 V (Not operator-replaceable.)
5A
N/A
3.5 A
50 A
25 A 25 A -- -- --
N/A N/A -- -- --
17.7 A 17.7 A -- -- --
50 A 500 A -- -- --

Mechanical Characteristics

Mechanical characteristics of the amplifiers are listed in Table 4--3.
Table 4- 3: TCPA300 and TCPA400 mechanical characteristics
Parameter, nominal Characteristic
Length 173 mm (6.8 in)
Width 91.4 mm (3.6 in)
Height 167 mm (6.6 in)
Weight 1.14 kg (2.5 lb)
TCP404XLTCP303TCP305TCP312
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
4- 3
Specifications
Mechanical characteristics of the probes are listed in Table 4--4.
Table 4- 4: Probe mechanical characteristics
TCP305 and TCP312 TCP303 and TCP404XL
Probe dimensions
Length:..............
Width:..............
Height:..............
20 cm (7.77 inches)
1.6 cm (0.625 inches)
3.2 cm (1.25 inches)
268 cm (10.55 inches)
41 cm (1.60 inches)
156 cm (6.13 inches)
Cable length 1.5 m (5 feet) TCP303: 2 m (6.6 feet)
TCP404XL: 8 m (26.25 feet)
Weight 0.15 kg (0.33 lb) TCP303: 0.66 kg (1.45 lb)
TCP404XL: 0.88 kg (1.90 lb)
41 mm (1.6 in)
11.9 mm
(0.470 in)
18.9 mm
(0.745 in)
Maximum wire size
3.8 mm (0.15 in)
TCP305 and TCP312
67 mm (2.7 in)
TCP303 and TCP404XL
Maximum wire size
21 mm (0.83 in)
21 mm (0.83 in)
25 mm (1.0 in)
Figure 4- 1: Probe jaw dimensions (nominal)
Shipping weights and dimensions are listed in Table 4--5.
Table 4- 5: Shipping weights and dimensions
TCPA300 and TCPA400 Amplifiers
Dimensions
Length:.......
Width:.......
Height:.......
298.5 mm (11.75 inches)
298.5 mm (11.75 inches)
254 mm (10.00 inches)
Weight 2.7 kg (6.00 lb) 0.585 kg (1.29 lb) 1.33 kg (2.93 lb) 1.55 kg (3.42 l b)
4- 4
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCP305 and TCP312
Current Probes
330 mm (13.00 inches)
254 mm (10.00 inches)
108 mm (4.25 inches)
TCP303 Current Probe
375 mm (14.75 inches)
318 mm (12.50 inches)
127 mm (5.00 inches)
TCP404XL Current Probe
375 mm (14.75 inches)
318 mm (12.50 inches)
127 mm (5.00 inches)

Environmental Characteristics

The environmental characteristics in Table 4--6 are warranted performance specifications. Unlike the warranted characteristics in Table 4--1, the environ­mental characteristics are type tested; therefore there are no performance verification procedures provided to test these characteristics. Unless otherwise noted, these characteristics apply to all probes and amplifiers.
Table 4- 6: Environmental characteristics
Parameter Characteristic
Ambient Temperature
Operating 0 _Cto+50_C(32_ F to 122 _F)
Nonoperating -- 4 0 _Cto+75_C(--40_ F to 167 _F)
Humidity
Operating 5 to 95% R.H. to +30 _ C(86_F)
Specifications
5 to 85% R.H. +30 _Cto+50_ C(86_F to 122 _F)
Nonoperating 5 to 95% R. H. t o +30 _C(86_F)
5 to 85% R.H. +30 _Cto+75_ C(86_F to 167 _F)
Altitude
Operating 2000 m (6800 ft) maximum
Nonoperating 12,192 m (40,000 ft) maximum
Random Vibration, Amplifiers
Operating 0.31 g
Nonoperating 2.46 g
Random Vibration, Probes
Operating 2.66 g
Nonoperating 3.48 g
Shock, Amplifiers 50 g, 11 ms duration, hal f-sine pul ses
Electro-Magnetic Compliance Meets FCC Part 15, Subpart B, Class A
, 5 Hz to 500 Hz, 10 minutes each axis
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
RMS
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
4- 5
Specifications
Declarati
f
Conf
g
y
g
Table 4--7 lists the certifications and compliances for the amplifiers and probes.
Table 4- 7: Certifications and compliances
Category Standards or description
EC Declaration of Conformity -­EMC
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities:
EN 61326 EMC requirements for Class A electrical equipment for
measurement, control and laboratory use.
1
IEC 61000--4--2 Electrostatic discharge immunity (Performance criterion B) IEC 61000--4--3 RF electromagnetic f ield immunity (Performance criterion A) IEC 61000--4--4 Electrical fast transient / burst immunity (Performance criterion B) IEC 61000--4--5 Power line surge immunity (Performance criterion B) IEC 61000--4--6 Conducted RF immunity (Performance criterion A) IEC 61000--4--11 Voltage dips and interruptions immunity (Performance criterion B)
EN 61000--3--2 AC power line harmonic emissions
Australia / New Zealand
on o
ormity--
EMC
Complies with EMC provision of Radiocommunications Act per the following standard(s):
AS/NZS 2064.1/2 Industrial, Scientific, and Medical Equipment: 1992
EMC Compliance Meets the intent of Directive 89/336/EEC for Electromagnetic Compatibility when it is used with the
product(s) stated in the specifications table. Refer to the EMC specification published for the stated products. May not meet the intent of the directive if used with other products.
FCC Compliance Emissions comply with FCC Code of Federal Regulations 47, Part 15, Subpart B, Class A Limits.
EC Declaration of Conformity -­Low Voltage
2
Compliance was demonstrated to the following specification as listed in the Official Journal of the European Communities:
Low Voltage Directive 73/23/EEC, amended by 93/68/EEC
EN 61010-1/A2:1995 Safety requirements for electrical equipment for measurement
control and laboratory use.
EN 61010-2-032:1995 Particular requirements for hand-held current clamps for electrical
measurement and test equipment.
U.S. Nationally Recognized UL3111-1 Standard for electrical measuring and test equipment. Testing Laboratory Listing
UL3111--2--032 Standard f or hand-held current clamps for electrical measurement
and test.
Canadian Certification CAN/CSA C22.2 No. 1010.1 Safety requirements for electrical equipment for measurement,
control, and laboratory use.
CAN/CSA C22.2 No. 1010.2.032:1996
Particular requirements for hand-held probe assemblies for electrical measurement and test equipment.
1
Emissions that exceed the levels required by this standard may occur when this equipment is connected to a test object.
2
The TCP305 and TCP312 Current Probes are exempt from the Low Voltage Directive and are not third-party listed. However, they have been evaluated to applicable safety standards.
4- 6
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Specifications
gyp
Table 4- 7: Certifications and compliances (cont.)
Category Standards or description
Additional Compliance UL 61010B-1 Safety requirements for electrical equipment for measuring,
controlling and laboratory use.
IEC61010-1/A2:1995 Safety requirements for electrical equipment for measurement,
control, and laboratory use.
IEC 61010-2-032:1995 Particular requirements for hand-held current clamps for electrical
measurement and test.
IEC 61010-1:2001 Safety requirements for electrical equipment for measurement
control and laboratory use.
Installation (Overvoltage) Category Descriptions
Pollution Degree Descriptions A measure of the contaminates that could occur in the environment around and within a product.
Terminals on this product may have different installation (overvoltage) category designations. The installation categories are:
CAT III Distribution-level mains (usually permanently connected). Equipment at this level is
typically in a fixed industrial location.
CAT II Local-level mains (wall sockets). Equipment at this level includes appliances, portable
tools, and similar products. Equipment is usually cord-connected.
CAT I Secondary (signal level) or battery operated circuits of electronic equipment.
Typically the internal environment inside a product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution Degree 1 No pollution or only dry, nonconductive pollution occurs. Products in
this category are generally encapsulated, hermetically sealed, or located in clean rooms.
Pollution Degree 2 Normally only dry, nonconductive pollution occurs. Occasionally a
temporary conductivity that is caused by condensation must be expected. This location is a typical office/home environment. Temporary condensation occurs only when the product is out of service.
Pollution Degree 3 Conductive pollution, or dry, nonconductive pollution that becomes
conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct wind.
Pollution Degree 4 Pollution that generates persistent conductivity through conductive
dust, rain, or snow. Typical outdoor locations.
Equipment Type Test and measuring
Safety Class Class 1 (as defined in IEC 61010-1, Annex H) -- grounded product
Overvoltage Category Overvoltage Category II (as defined in IEC 61010-1, Annex J)
Pollution Degree Pollution Degree 2 (as defined in IEC 61010-1). Note: Rated for indoor use only.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
4- 7
Specifications
TCP312 derating curve
35
30
25
20
15
Peak current (A)
10
5
0
1k 10k 100k
Frequency (Hz)
Figure 4- 2: Frequency derating- TCP312
60
50
40
TCP305 derating curve
50_C ambient 25_C ambient
1M 10M
50_C ambient 25_C ambient
100M
30
Peak current (A)
20
10
0
1k 10k 100k
Frequency (Hz)
Figure 4- 3: Frequency derating- TCP305
1M 10M
100M
4- 8
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Specifications
250
200
150
100
Peak current (A)
50
0
100
1k 10k 100k
TCP303 derating curve
Frequency (Hz)
Figure 4- 4: Frequency derating- TCP303
800
700
600
500
TCP404XL derating curve
50_C ambient 25_C ambient
1M 10 M
50_C ambient 25_C ambient
100M
400
300
Peak current (A)
200
100
0
0 1k 10k 100k
Frequency (Hz)
Figure 4- 5: Frequency derating- TCP404XL
1M 10M
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
4- 9
Specifications
1.0
TCP312
.1
.01
Insertion Z (ohms)
.001
10k 100k 1M 10M 100M
1.0
Frequency (Hz)
TCP305
.1
.01
Insertion Z (ohms)
.001
10k 100k 1M 10M 100M
Frequency (Hz)
1.0
TCP303
.1
.01
Insertion Z (ohms)
.001
10k 100k 1M 10M 100M
Frequency (Hz)
100m
10m
1m
Insertion Z (ohms)
100µ
10µ
TCP404XL
10 100 1k 10k 100k
Frequency (Hz)
1M
10M
4- 10
Figure 4- 6: Insertion impedance graphs for the curr ent probes
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Specifications
60
TCP312 Max peak pulse vs PW
50
Maximum peak pulse 50A
40
500 A*μs
30
Amperes (Peak)
20
50 A*
10
µs
0
0 5 10 15 20
Allowable pulse width (µseconds)
TCP305 Max peak pulse vs PW
60
50
40
Maximum peak pulse 50A
Range setting
500 A*
µs
5A/V 10 A/V
Amperes (Peak)
30
20
Range setting
1A/V 10 A/V
30 A continuous
Any width
5 A continuous
25
Any width
25 A continuous
10
0
0
5 10152025
Allowable pulse width (µseconds)
TCP303 Max peak pulse vs PW
600
Maximum peak pulse 500A
Range setting
500
400
300
15000 A*µs
Amperes (Peak)
200
3000 A*µs
212 A peak (150 A RMS continuous)
100
25A continuous
0
0 50 100 150
Allowable pulse width (µseconds)
Figure 4- 7: Specified operating area of the probes
5A/V 50 A/V
Any width
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
4- 11
Specifications

TCP404XL Maximum Measurement Times

Figures 4--8 through 4--10 show the maximum measurement time for the TCP404XL probe, with varying duty cycles and temperatures. For more detail on the relationship between measurement factors, refer to Measuring Noncontin- uous Current with the TCP404XL Probe on page 3--6.
12
10
8
6
time (min)
4
Maximum measurement
2
300A Continuous
0
0 5 10 15 20 25 30 35
Measuring 750A at 50 _C ambient
0A Continuous
100A Continuous
200A Continuous
Duty cycle (%)
Figure 4- 8: Measuring 750A noncontinuous at 50 °C ambient temperature
30
25
20
15
time (min)
10
Measuring 600A at 50 _C ambient
0A Continuous
200A Continuous
300A Continuous
4- 12
Maximum measurement
5
400A Continuous
0
0102030405060
Duty cycle (%)
70
Figure 4- 9: Measuring 600A noncontinuous at 50 °C ambient temperature
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Specifications
25
20
time (min)
15
10
5
Maximum measurement
0
010203040 5060
500A Continuous
Measuring 750A at 23 _C ambient
0A Continuous
200A Continuous
300A Continuous
400A Continuous
Duty cycle (%)
Figure 4- 10: Measuring 750A noncontinuous at 23 °C ambient temperature
At 23 degrees ambient temperature, 600 A can be measured continuously with the TCP404XL probe.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
4- 13
Specifications
4- 14
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
WARNING
The following servicing instructions are for use only by qualified personnel. To avoid injury, do not perform any servicing other than that stated in the operating instructions unless you are qualified to do so. Refer to all safety summaries before performing any service.
Performance Verification

Performance Verification Overview

The Performance Verification tests allow you to demonstrate that the Current Probe Amplifier and Current Probes meet their specified levels of performance. Since the TCPA300 and TCPA400 cannot operate without a current probe, a calibration adapter is used to verify the amplifier(s) separately. The TCPA Calibration Adapter is an optional accessory--see Replaceable Parts for part numbers. After you have verified the amplifier performance independent of the current probe(s), you can check the performance of current probes with confidence.
Tolerances that are specified in these procedures apply to the amplifiers and current probes and do not include test equipment error. The recommended calibration interval is one year.

Performance Verification and Functional Checks

The warranted specifications for the amplifiers that are checked in the perfor­mance verification procedures are listed in Table 5--1. See Table 4--1 in Specifi- cations for warranted values. Functional checks are included for you to verify proper amplifier operation.
Table 5- 1: Amplifier checks
Performance verification Functional checks
DC gain
Amplifier
TCPA300
TCPA400
1
accuracy
With probe attached
Bandwidth
Front-panel display
AC coupling
Degauss
1
Overload
The warranted specifications for the probes that are checked in the performance verification procedures are listed in Table 5--2. See Table 4--1 in Specifications for warranted values.
Table 5- 2: Probe performance verification checks
Probe DC gain accuracy Bandwidth Risetime
TCP312
TCP305
1
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 1
Performance Verification Overview
Table 5- 2: Probe performance verification checks (Cont.)
Probe RisetimeBandwidthDC gain accuracy
TCP303
TCP404XL

Test Procedure Conditions

These procedures are valid only under the following conditions:
H The system has been calibrated at an ambient temperature of 23_ ±5 _C.
H The system is operating in an environment whose limits are described in
Table4--6onpage4--5.
H The system, including probe, has had a warm-up period of at least
20 minutes.

Equipment Preparation

H The probe degauss/autobalance routine has been performed after the
20-minute warm-up period.
Before performing the verification tests, turn all equipment on and allow the entire system to warm up for a minimum of 20 minutes.
NOTE. Before performing any verification procedure, properly degauss the probe. Remove the probe from any current-carrying conductor, lock the probe, and press the amplifier PROBE DEGAUSS AUTOBALANCE button. The degauss/autobalance routine is complete when the indicator light turns green.
The amplifier front panel will light the NOT TERMINATED INTO 50 Ω LED during the degauss/ autobalance routine if the amplifier is not properly terminated into 50 Ω. Verify that the oscilloscope input is 50 Ω and set to DC coupling. Use a 50 Ω feedthrough termination, attached at the oscilloscope input, if necessary.
5- 2
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

TCPA300 and TCPA400 Performance Verification

This section covers the performance verification for the TCPA300 and TCPA400 Current Probe Amplifiers. Refer to the list of equipment required in Table 5--3. Power on the amplifier and test equipment and let them warm up for 20 minutes.
Before starting these procedures, photocopy the appropriate test record beginning on page 5--16 to record the performance test results. The recommended calibration interval is one year.

Equipment Required

The performance verification procedures require the test equipment listed in Table 5--3. The test equipment must meet or exceed the specifications listed. You may need to modify the test procedures if you do not use the recommended equipment.
Table 5- 3: Required test equipment
Recommended
Qty Item Description
equipment
1 Oscilloscope 500 MHz bandwidth Tektronix TDS5000
1 Leveled Sine Wave Generator 3 MHz to 100 MHz Wavetek 9100 with >250 MHz Oscilloscope
Calibration Option
1 Digital Multimeter DCV: 0.1% Accuracy, 51/2digit resolution Keithley 2700
1 Current Source DCA: 0.1% Accuracy, 0 to ±100 mA,
floating outputs
1 Current Source DCA: 0.1% accuracy, 0 to ±15 A Wavetek 9100
1 Square Wave Generator ACV 1.5 Vp-p, 28 Hz square wave Wavetek 9100
1 HF Current Loop 50 Ω, BNC Connector Tektronix part number 015-0601-50
1 Precision Termination 50 Ω, ±0.1%, 0.5 W Tektronix part number 011-0129-00
2 BNC Cables 50 Ω, 0.76 m (30 in) long Tektronix part number 012-0117-00
1 Cable TEKPROBE Interface Cable Tektronix part number 012-1605-00
2 BNC to Dual Banana Adapters Tektronix part number 103-0090-00
1 DC Current Loop 5 turns 18 AWG coated wire Refer to page 5--4
1 DC Current Loop 50 turns 13 AWG coated wire Refer to page 5--4
1 Calibration Adapter TCPA Calibration Adapter Tektronix part number 174-4765-00
1 Current Probe
1
Required for functional test only.
1
TCPA300/400-compatible probe Tektronix TCP303, 305, or 312 (TCPA300) or
Fluke 5100 or HP 6612C
TCP404XL (TCPA400)
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 3
TCPA300 and TCPA400 Performance Verification

Making DC Current Loops

You need to construct simple current loops to complete some tests for the amplifiers. The current loops are also used in the DC gain accuracy tests for the probes.
5-Turn DC Current Loop
50-Turn DC Current Loop
Construct the loop using a cylindrical form approximately 3 inches in diameter:
1. Wind exactly 5 turns of #18 coated wire around the form.
2. Scrape about a half-inch of coating off of the ends of the wire.
NOTE. Ensure that the current loop has exactly 5 turns. A significant error will result for each turn variance from 5 turns.
The 5-turn DC current loop is also used in the current probe performance verification and adjustment procedures.
Construct the loop using a cylindrical form approximately 3 inches in diameter:
1. Wind exactly 50 turns of #13 coated wire around the form.
2. Scrape about a half-inch of coating off of the ends of the wire.
NOTE. Ensure that the current loop has exactly 50 turns. A significant error will result for each turn variance from 50 turns.
5- 4
The 50-turn DC current loop is also used in the current probe performance verification procedures.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

Front-Panel Display

TCPA300 and TCPA400 Performance Verification
The status and error LEDs on the front panel are illuminated briefly during the amplifier power-on sequence. Verify that all LEDs are functional by doing the following procedure:
1. Power-cycle the amplifier and check that the LEDs shown in Figure 5--1 light briefly. Note that the PROBE DEGAUSS AUTOBALANCE and OVERLOAD LEDs are multi-color:
a. The PROBE DEGAUSS AUTOBALANCE LED lights in an
orangeredgreen sequence.
b. The OVERLOAD LED lights in an orangeredgreen sequence.
c. The MANUAL BALANCE LED is orange, and lights briefly.
d. The RANGE and COUPLING LEDs are green, and light briefly.
2. Record the results (pass/fail) on the test record.
Multi--color LED
Verify LEDs light
Multi--color LED
Figure 5- 1: Check LED functionality
Verify LEDs light
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 5
TCPA300 and TCPA400 Performance Verification

DC Gain Accuracy

Use the following procedure to verify the DC gain accuracy of the amplifier.
Test Equipment Setup
Refer to Figure 5--2 and set up the test equipment as follows:
1. Connect the output of the amplifier to the DMM:
a. Connect a BNC Cable from the amplifier OUTPUT to the 50 Ω
precision feedthrough termination.
b. Attach the termination to a BNC-to-dual banana adapter.
c. Observe polarity and insert the dual banana adapter into the digital
multimeter DC voltage input.
2. Connect the probe input to the current source:
a. Connect the BNC end of the Calibration Adapter to a BNC-to-dual
banana adapter.
b. Observe polarity and insert the dual banana adapter into the current
source DC output. Do not connect the Calibration Adapter to the amplifier at this time.
Amplifier
DMM
Current source
5- 6
BNC-to-dual
banana adapter
50 Ω Precision
termination
+--
50 Ω Coaxial cable
Calibration adapter
BNC-to-dual
banana adapter
+--
Figure 5- 2: Equipment setup for DC gain accuracy test
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCPA300 and TCPA400 Performance Verification
3. Make or verify the equipment settings in Table 5--4:
Table 5- 4: Equipment settings
Digital multimeter
Measurement Type DC volts Range Autoranging
Current source
Voltage 6V Current 100 mA Output Off
TCPA300 and TCPA400 amplifier
Coupling DC
Procedure
After the equipment is set up, proceed as follows:
4. Connect the Calibration Adapter to the PROBE INPUT of the amplifier.
5. For each of the Range settings in Table 5--5, perform the following steps:
a. If you are checking a TCPA300, set the amplifier to the desired Range
setting in Table 5--5.
b. Enable the output of the current source.
c. Record the exact measurement of the digital multimeter as M
.
1
Table 5- 5: DC gain accuracy test for the TCPA300 and TCPA400
Current source out-
Amplifier Range, A/V
TCPA300 1 100 mA 5.0000
5 100 mA 2.5000
10 100 mA 1.2375
50 100 mA 0.4835
50 (COMP)1100 mA 0.4585
put
Expected output, V (VDC)
E
Measured output, M (VDC)
1
% Error, calculated
TCPA400 1A/mV 100 mA 2.5000
1
Both 10 A/V and 50 A/V RANGE LEDs light.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 7
TCPA300 and TCPA400 Performance Verification
d. Calculate the %Error between the measured output, M
expected output (V
M
%Error =
1
)%Errorasfollows:
E
V
E
V
× 100
E
, and the
1
For example, using an expected output VEof 2.5000 V and a measured output M
%Error =
of 2.510, the %Error would be:
1
2.5100 2.5000
2.5000
× 100 =+0.4%
NOTE. It is important that the correct polarity be used to calculate the %Error.
6. Verify that the measured DC gain accuracy is within the limits specified for all settings in the test record on page 5--16 for the TCPA300 (page 5--17 for the TCPA400). Record the %Error in the test record.
NOTE. If the %Error on any of the settings is greater than the specified limit, you must perform the adjustment procedure that begins on page 6--3.
7. Disconnect the DMM cable from the amplifier and the calibration adapter from the current source.
5- 8
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual

Bandwidth

TCPA300 and TCPA400 Performance Verification
This procedure tests the bandwidth of the TCPA300 and TCPA400 amplifiers. In this test you measure a signal at a relatively low frequency and again at the upper test frequency. The two measurements are compared to verify that the signal amplitude does not fall below a certain limit. Refer to Figure 5--3 when making equipment connections.
Amplifier
Test oscilloscope
TekProbe interface cable or
50 Ω coaxial cable
50 Ω oscilloscope input -- use the TekProbe InterfaceCableorusea50Ω cable. (Add 50 Ω termination here if oscilloscope has only high-impedance input.)
Calibration adapter
Note: If using a Wavetek 9100, use the Signal Out
BNC connector on the rear of the instrument.
Leveled
sine wave
generator
Output
Figure 5- 3: Bandwidth test setup
1. If you are using a Tektronix oscilloscope that supports the TekProbe Level 2 Interface, use the TekProbe interface cable to connect the amplifier OUT- PUT to the oscilloscope input. If you are not using a Tektronix oscilloscope that supports the TekProbe Level 2 Interface, use a 50 Ω BNC cable. If the input impedance of your oscilloscope is 1 MΩ, connect a 50 Ω feedthrough termination at the oscilloscope input. Do not connect the termination at the amplifier output.
2. Connect the Calibration Adapter to the amplifier PROBE INPUT.
3. Make or verify the equipment settings in Table 5--6:
Table 5- 6: Equipment settings for bandwidth check
Oscilloscope
Vertical input impedance 50 Ω Time base TCPA300
TCPA400
40 ns/division
80 ns/division
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 9
TCPA300 and TCPA400 Performance Verification
Table 5- 6: Equipment settings for bandwidth check (Cont.)
Oscilloscope
Record length 500 Coupling DC Offset 0 V (mid-scale) Trigger type Edge Trigger mode Auto Trigger position 50% Acquisition mode Average Number of waveforms to average 16 Measurement type Peak-to-Peak
Leveled sine wave generator
Frequency 3MHz Amplitude TCPA300
TCPA300 and TCPA400
Coupling DC
TCPA400
~3.0 V
1.0 V
p-p
p-p
4. Connect the Calibration Adapter to the output of the leveled sine wave generator.
5. Enable the output of the leveled sinewave generator.
6. Verify the amplifier output is what is listed for the TCPA300 amplifier in
Table 5--7 on page 5--11. If you are checking a TCPA400 amplifier, use Table5--8onpage5--11.
7. Using the peak-peak measurement capability of the oscilloscope, measure and record the peak-peak reading as M
in Table 5--7 or Table 5--8 on
1
page 5--11.
8. If you are checking a TCPA300, for each range setting and bandwidth filter in Table 5--7 on page 5--11, perform the following steps. If you are checking a TCPA400, use Table 5--8 on page 5--11.
a. Set the oscilloscope time base to 4 or 5 ns/division. Increase the signal
generator frequency to 50 or 100 MHz, depending on the range setting and bandwidth.
b. When you check the 100 MHz frequency, press either MANUAL
BALANCE button until the orange MANUAL BALANCE LED lights. This engages a higher-frequency filter in the amplifier. After you check the bandwidth at 100 MHz, press either MANUAL BALANCE button again to turn off the filter (the LED goes off.)
5- 10
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Loading...