Tektronix products are covered by U.S. and foreign patents, issued and
pending. Information in this publication supercedes that in all previously
published material. Specifications and price change privileges reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
WARRANTY
Tektronix warrants that the products that it manufactures and sells will be free from defects
in materials and workmanship for a period of one (1) year from the date of purchase from
an authorized Tektronix distributor. If any such product proves defective during this
warranty period, Tektronix, at its option, either will repair the defective product without
charge for parts and labor, or will provide a replacement in exchange for the defective
product. Batteries are excluded from this warranty.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect
before the expiration of the warranty period and make suitable arrangements for the
performance of service. Customer shall be responsible for packaging and shipping the
defective product to the service center designated by Tektronix, shipping charges prepaid,
and with a copy of customer proof of purchase. Tektronix shall pay for the return of the
product to Customer if the shipment is to a location within the country in which the
Tektronix service center is located. Customer shall be responsible for paying all shipping
charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or
improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish
service under this warranty a) to repair damage resulting from attempts by personnel other
than Tektronix representatives to install, repair or service the product; b) to repair damage
resulting from improper use or connection to incompatible equipment; c) to repair any
damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a
product that has been modified or integrated with other products when the effect of such
modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE
LISTED PRODUCTS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR
IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE
DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY
PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.
TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY
INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES
IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Contacting Tektronix
Phone1-800-833-9200*
AddressTektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
Web sitewww.tektronix.com
Sales
support
Service
support
Technical
support
*This phone number is toll free in North America. After office
hours, please leave a voice mail message.
Outside North America, contact a Tektronix sales office or
distributor; see the Tektronix web site for a list of offices.
1-800-833-9200, select option 1*
1-800-833-9200, select option 2*
Email: techsupport@tektronix.com
1-800-833-9200, select option 3*
6:00 a.m. -- 5:00 p.m. Pacific time
Table of Contents
General Safety Summaryiii............................
Review the following safety precautions to avoid injury and prevent
damage to this product or any products connected t o it.
Injury Precautions
Do Not Operate in Wet/Damp C onditions. To avoid electric shock, do not
operate this product in wet or damp conditions.
Do Not Operate in an Explosive Atmosphere. To avoid injury or fi re
hazard, do not operate this product in an explosive atmosphere.
Observe Maximum Working Voltage. Do not use this product on bare
wires above 300 V (DC + peak AC).
Product Damage Precautions
Do Not D rop the Probe Head. The probe head contains fragile
components that can be damaged by a high impact. Take care to
prevent the probe head from dropping on the floor or other hard
surface. Secure the probe head in a safe location when not in use.
Do Not Operate With Suspected Failures. If you suspect there is damage
to this product, have it i nspected by qualified service personnel.
Do Not Immerse in Liquids. Clean the probe using only a damp cloth.
Refer to cleaning instructions.
Safety Terms and Symbols
Terms in This Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or pract ices that
could result in injury or loss of life.
CAUTION. Caution statements identify condit ions or practices that
could result in damage to this product or other property.
TCP202 Instruction Manual
v
General Safety Summary
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you
read the marking.
WARNING indicates an injury hazard not immediately accessible as
you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. These symbols may appear on the product:
CAUTION
Refer to Manual
Do not connect to or
remove from an
uninsulated conductor t hat
is HAZARDOUS LIVE.
WARNING
High Voltage
Breakable.
Do not drop.
Double
Insulated
Use only on an
insulated wire.
Protective Ground
(Earth) Terminal
Certifications and Compliances
Refer to the specifications section for a listing of certifications and
compliances that apply to this product.
vi
TCP202 Instruction Manual
Getting Started
This section describes the TCP202 Current Probe and gives
instructions on how to install and functiona lly test the probe.
Features and Accessories
The TCP202 Current Probe (Figure 1) is a 50 MHz current probe for
use with Tektronix oscilloscopes that have the TekProbe interface.
The probe can measure up to 15 A (DC + peak AC).
Figure 1: TCP202 Current Probe
TCP202 Instruction Manual
1
Getting Started
Probe Head. The probe head contai ns a Hall Effect
device for measuring current. T he j aw of the probe
clamps on any wire (including insulation) that is less
than or equal to 3.8 mm (0.15 inches) in diameter.
TekProbe Interface. The TekProbe interface provides
power, signal, and probe characteristic data.
If your oscilloscope does not support the TekProbe
interface, you can use the optional 1103 probe power
supply as an effective interface. Contact your local
Tektronix representative for more information.
Degauss. The degauss button re moves residual
magnetism from the probe core that would otherwise
cause measurement errors.
Balance Thumbwheel. The balance t humbwheel
compensates for minor DC offsets of the probe output.
Coarse Balance Adjustment. Thecoarsebalance
adjustment centers the range of the balance thumbwheel. Use this adjustment onl y if the thumbwhee l
does not have enough range.
Calibrator (optional). With some TDS oscilloscopes, you
can compensate the probe for maximum accuracy using
the optional calibrator. Use the calibrator only if your
oscilloscope displays the AΩ symbol when the current
probe is connected. (For ordering information, see
Replaceable Parts startingonpage44.)
2
TCP202 Instruction Manual
Installation
Install the TCP202 Current Probe as follows:
1. Set the input channel to a zero reference point.
2. Connect the output of the probe to the TekProbe interface of the
If your TDS oscilloscope does not display amperes/division, interpret
volts/division as amperes/division.
If you are using the 1103 TekProbe Power Supply, you must set the
oscilloscope input coupling to DC and the input impedance t o 50 Ω.
Interpret V/division as 10 A/division. For example, interpret
100 mV/division as 1 A/division.
Getting Started
oscilloscope or other measurement instrument. The measurement
instrument input must have a ground reference.
3. With the probe jaw empty, push the slide on the probe until it
locks in the CLOSED position.
4. Press the DEGAUSS button.
5. Balance the probe as follows:
a. Set the oscilloscope vertical scale to 10 mA/division.
b. Open and close the probe jaw.
c. Adjust the BALANCE thumbwheel until the displayed
signal is zero.
6. If you cannot zero the display as described in step 5c, use the
coarse balance adjustment (Figure 2) to center the range of the
thumbwheel.
Coarse balance
Figure 2: Coarse Balance Adjustment
TCP202 Instruction Manual
3
Getting Started
Compensating the Probe (Optional)
With some TDS oscilloscopes, you can compensate the probe for
maximum accuracy using the optional calibrator. Perform this
procedure only if your oscilloscope displays the AΩ symbol when
you connect the probe output to the oscilloscope input.
1. To connect the calibrator, press the release button (Figure 3) and
connect the calibrator to the probe compensation output of the
oscilloscope.
Release button
Figure 3: Connecting the Calibrator
2. Close and lock the probe jaw over the calibrator loop.
If the probe passes the compensation routine, the probe is accurate to
within ± 1%from50mAto5Aand± 2% from 5 A to 15 A.
If the probe fails the compe nsation routine, refer to the procedure for
checking the DC accuracy of the probe on page 41 and the procedure
for accessing the internal gain adjustment on page 32.
Repeat this procedure whenever you move the current probe to
another input.
4
TCP202 Instruction Manual
Operating Basics
This section contains important precautions and techniques for
clamping the probe on a circuit and taking basic measurements.
Measurement Limits
Before you clamp the probe on a circuit to measure curre nt, observe
the precautions for uninsulated wires and the maximum current
ratings.
Precautions for Uninsulated Conductors
WARNING. To avoid ele ctrical shock from uninsulat ed conductors,
observe t he following precautions:
Disconnect the power to the uninsulated conductor before inserting
the conductor in or removing the conductor from t he probe jaw.
Do not apply a voltage higher than 300 V (DC + peak AC) between
earth ground and an uninsulated conductor that is in the probe jaw.
Maximum Current Rating
There are two basic current ratings: maximum pe ak current and
maximum continuous current. In addition, the maximum peak
current is limited by the A/second rating. For a graph of these limits,
refertoFigure19onpage27.
The maximum peak current is 50 A with a pulse width ≤ 10 s
(500 Aseconds). This 500 Aseconds rating applies to any peak
current over 15 A. The product of the peak current and pulse width
(at 50% of peak) must be 500 Aseconds or le ss. For example, the
maximum pulse width is 20 s for a pulse of 25 A peak.
TCP202 Instruction Manual
5
Operating Basics
The maximum continuous current that this probe can measure is
15 A (DC + peak AC). This limit derates with frequency; as the
frequency increases, the maximum current rating decreases. For a
graph of this relationship, see Figure 14 on page 23.
Multiple Current Probes
Up to four TCP202 current probes may be used simultaneously on
one TDS Series oscilloscope if the total in-phase current measured
by all of the probes does not exceed 40 amperes. Above 40 amperes,
the TekProbe interface may overload and cause measurement or
display errors.
When one or two TCP202 current probes are used on a TDS Serie s
oscilloscope, the ratings are as shown in Figure 19 on page 27.
Degaussing and Balancing the P robe
For accurate measurements, you must occasionally degauss and
balance the probe. Degaussing removes residual magnetization from
the probe core that would otherwise shift the zero point and cause
measurement errors. Balancing the probe compensates for any DC
offset that remains on the probe output after degaussing.
Degauss and balanc e your probe in each of the following cases:
HAfter turning on the oscilloscope and allowing a 20-minute
warm-up period
HWhenever an overload condition occurs
HWhenever the probe is subjected to a strong external magnetic
field
HWhenever there is a measurable DC offset that does not come
from the conductor under test
To degauss and balance the probe, follow the installation procedure
on page 3.
You may degauss the probe with a conductor in the jaw if you first
disconnect the power to the conductor. This technique compensates
for any offsets caused by stray DC magnetic fields around the circuit
under test.
6
TCP202 Instruction Manual
CAUTION. While degauss occurs, the probe will induce a small
voltage in the unpowered circuit (33 mV in series with 1.5 MΩ). Your
circuit must be able to absorb this induced voltage. With low
impedance circuits, several amperes may be induced in the circuit
being measured. This may be of concern if you are using very small
conductors.
Measurement Procedures and Techniques
This section presents procedures and techniques for basic current
measurements.
Operating Basics
Basic Procedure
Measure AC or DC current in a single conductor as follows:
1. Observe the safety precautions and operating limits.
2. If necessary, degauss and balance the probe.
3. Close and lock the probe jaw over the conductor. Conventional
current flows from positive to negative. For the correct polarity
reading, connect the probe so that the current flow, from positive
to negative, is aligned with the arrow on the probe jaw (see
Figure 4).
4. Read the measurement. Adjust the display of the measurement
instrument as necessary.
5. If necessary, adjust the oscilloscope offset control to offset a DC
level. (AC coupling is automatically disabled on TDS oscilloscopes when the probe is connected.)
TCP202 Instruction Manual
7
Operating Basics
Figure 4: Polarity of Current Flow
Measuring Differential Current
You can place two conductors in a current probe to provide a
differential or null current measurement. This avoids the necessity of
using two current measurement systems with a differential
oscilloscope amplifier.
1. Orient the two conductors under test so that the pol arities oppose
each other. Clamp the current probe around the two conductors as
shown in Figure 5.
2. Measure the current. A waveform above the baseline indicates
the conductor with the conventional current flow in the direction
of the probe arrow is carrying the greater current.
3. To adjust for a current null, adjust the current in one of the
conductors until the displayed measurement is zero.
8
TCP202 Instruction Manual
Conductor 1
Operating Basics
Conductor 2
Current
Current
Current probe
Figure 5: Measuring Differential Current and Nulls
TCP202 Instruction Manual
9
Loading...
+ 43 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.