The print history shown below lists the printing dates of all Revisions and Addenda created
for this manual. The Revision Level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered
sequentially. When a new Revision is created, all Addenda associated with the previous Revision
of the manual are incorporated into the new Revision of the manual. Each new Revision includes
a revised copy of this print history page.
Revision A (Document Number 2016-900-01)................................................................ May 1999
Revision B (Document Number 2016-900-01) ....................................................... December 2001
Revision C (Document Number 2016-900-01) ............................................................ August 2003
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation.
Although some instruments and accessories would normally be used with non-hazardous voltages, there are
situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety
precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance
information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring
that the equipment is operated within its specifications and operating limits, and for ensuring that operators are
adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper
use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example,
setting the line voltage or replacing consumable materials. Maintenance procedures are described in the user
documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be
performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly
trained service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are rated Measurement Category I
and Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC
60664. Most measurement, control, and data I/O signals are Measurement Category I and must not be directly
connected to mains voltage or to voltage sources with high transient over-voltages. Measurement Category II
connections require protection for high transient over-voltages often associated with local AC mains connections.
Assume all measurement, control, and data I/O connections are for connection to Category I sources unless
otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks
or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage
levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous
voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that
operators are prevented access and/or insulated from every connection point. In some cases, connections must be
exposed to potential human contact. Product operators in these circumstances must be trained to protect
themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive
part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedancelimited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards,
install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power recep tacle.
Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate
main input power disconnect device must be provided in close proximity to the equipment and within easy reach of
the operator.
11/07
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the
!
circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before:
connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes,
such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line
(earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of
withstanding the voltage being measured.
The instrument and accessories must be used in accordance with specifications and operating instructions, or the
safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and
operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground
connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation
requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
The symbol on an instrument indicates that the user shoul d refer to the operating instructions located in the
documentation.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined
effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these
voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp
must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death.
Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such
damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protecti on from electri c shock and fire, re placement components in mains circuits - including the power
transformer, test leads, and input jacks - must be purchased from Keithley Instruments. Standard fuses with
applicable national safety approvals may be used if the rating and type are the same. Other components that are
not safety-related may be purchased from other suppliers as long as they are equivalent to the original component
(note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and
functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley
Instruments office for information.
T o clean an instrument, use a damp cloth or mild, water-based cleaner . Clean the exterior of the instrument only . Do
not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist
of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never
require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected,
the board should be returned to the factory for proper cleaning/servicing.
This section contains general information about the Model 2016 THD Multimeter. The
information is organized as follows:
•Feature overview
•Warranty information
•Manual addenda
•Safety symbols and terms
•Specifications
•Inspection
•Options and accessories
If you have any questions after reviewing this information, please contact your local
Keithley representative or call one of our Applications Engineers at 1-800-348-3735 (U.S.
and Canada only). Worldwide phone numbers are listed at the front of this manual.
Feature overview
The Model 2016 is a 6½-digit high-performance digital multimeter. It has 0.002% 90-day
basic DC voltage accuracy and 0.008% 90-day basic resistance accuracy. At 6
multimeter delivers 50 triggered readings/sec over the IEEE-488 bus. At 4
up to 2000 readings/sec into its internal buffer. The Model 2016 has broad measurement ranges:
½
-digits, the
½
-digits, it can read
•THD, THD+n, and SINAD from 20Hz to 50kHz with 0.0001% (0.00001 dB) resolution.
•DC voltage from 0.1
•AC (RMS) voltage from 0.1
•DC current from 10nA to 3A.
•AC (RMS) current from 1
•Two and four-wire resistance from 100µ
•Frequency from 3Hz to 500kHz.
•Thermocouple temperature from -200°C to +1372°C.
Some additional capabilities of the Model 2016 include:
•Built-in function generator — Provides 10Hz to 20kHz sinewave or pulse outputs.
Maximum sinewave amplitude is 2V rms (50
•Full range of functions — In addition to those listed above, the Model 2016 functions
include period, dB, dBm, continuity, diode testing, mX+b, and percent.
•Optional scanning — For external scanning, Model 2016 is compatible with Keithley’s
Model 7001 and 7002 switch matrices and cards.
•Programming language and remote interfaces — The Model 2016 has the SCPI
programming language and two remote interface ports (IEEE-488/GPIB and RS-232C).
•Reading and setup storage — Up to 1024 readings and two setups (user and factory
defaults) can be stored and recalled.
•Closed-cover calibration — The instrument can be calibrated either from the front panel
or remote interface.
µ
V to 1000V.
µ
µ
A to 3A.
V to 750V, 1000V peak.
to 120MΩ.
Ω
Ω
or 600Ω) or 9.5V rms (HI Z).
Warranty information
Warranty information is located at the front of this instruction manual. Should your
Model 2016 require warranty service, contact the Keithley representative or authorized
repair facility in your area for further information. When returning the instrument for repair,
be sure to fill out and include the service form at the back of this manual to provide the
repair facility with the necessary information.
Manual addenda
Any improvements or changes concerning the instrument or manual will be explained in
an addendum included with the manual. Be sure to note these changes and incorporate them
into the manual.
Safety symbols and terms
General Information1-3
The following symbols and terms may be found on the instrument or used in this manual.
Thesymbol on the instrument indicates that the user should refer to the operating
instructions located in the manual.
The
symbol
Use standard safety precautions to avoid personal contact with these voltages.
The
WARNING
injury or death. Always read the associated information very carefully before performing the
indicated procedure.
The
instrument. Such damage may invalidate the warranty.
Specifications
Full Model 2016 specifications are included in Appendix A.
!
CAUTION
on the instrument shows that high voltage may be present on the terminal(s).
heading used in this manual explains dangers that might result in personal
heading used in this manual explains hazards that could damage the
1-4General Information
Inspection
The Model 2016 was carefully inspected electrically and mechanically before shipment.
After unpacking all items from the shipping carton, check for any obvious signs of physical
damage that may have occurred during transit. (Note: There may be a protective film over the
display lens, which can be removed.) Report any damage to the shipping agent immediately.
Save the original packing carton for possible future reshipment. The following items are
included with every Model 2016 order:
•Model 2016 THD Multimeter with line cord.
•Safety test leads (Model 1751).
•Accessories as ordered.
•Certificate of calibration.
•Product Information CD-ROM that contains a PDF of the Model 2016 User's Manual.
If an additional manual is required, order the appropriate manual package. The manual
packages include a manual and any pertinent addenda.
Options and accessories
The following options and accessories are available from Keithley for use with the
Model 2016.
General purpose probes
Model 1754 Universal Test Lead Kit:
lugs, two banana plugs, two hooks, and two alligator clips.
Model 8605 High Performance Modular Test Leads:
test probes and leads. The test leads are terminated with a banana plug with retractable sheath
on each end.
Model 8606 High Performance Probe Tip Kit:
clips, and two spring hook test probes. (The spade lugs and alligator clips are rated at 30V RMS,
42.4V peak; the test probes are rated at 1000V.) These components are for use with high
performance test leads terminated with banana plugs, such as the Model 8605.
The following test leads and probes are rated at 30V RMS, 42.4V peak:
Models 5805 and 5805-12 Kelvin Probes:
with banana plug termination. Designed for instruments that measure 4-terminal resistance. The
Model 5805 is 0.9m long; the Model 5805-12 is 3.6m long.
Model 5806 Kelvin Clip Lead Set:
plug termination. Designed for instruments that measure 4-terminal resistance. A set of eight
replacement rubber bands is available as Keithley P/N GA-22.
Consists of one set of test leads (0.9m), two spade
Consists of two high voltage (1000V)
Consists of two spade lugs, two alligator
Consists of two spring-loaded Kelvin test probes
Includes two Kelvin clip test leads (0.9m) with banana
General Information1-5
Model 8604 SMD Probe Set:
surface mount device “grabber” clip on one end and a banana plug with a retractable sheath on
the other end.
Low thermal probes
Model 8610 Low Thermal Shorting Plug:
1-inch square circuit board, interconnected to provide a short circuit among all plugs.
Cables and adapters
Models 7007-1 and 7007-2 Shielded GPIB Cables:
bus using shielded cables and connectors to reduce electromagnetic interference (EMI). The
Model 7007-1 is 1m long; the Model 7007-2 is 2m long.
Models 8501-1 and 8501-2 Trigger Link Cables:
instruments with Trigger Link connectors (e.g., Model 7001 Switch System). The Model 8501-1
is 1m long; the Model 8501-2 is 2m long.
Model 8502 Trigger Link Adapter:
of the Model 2016 to instruments that use the standard BNC trigger connectors.
Model 8504 DIN to BNC Trigger Cable:
(Voltmeter Complete) and two (External Trigger) of the Model 2016 to instruments that use
BNC trigger connectors. The Model 8504 is 1m long.
Consists of two test leads (0.9m), each terminated with a
Consists of four banana plugs mounted to a
Connect the Model 2016 to the GPIB
Connect the Model 2016 to other
Allows you to connect any of the six Trigger Link lines
Allows you to connect Trigger Link lines one
Rack mount kits
Model 4288-1 Single Fixed Rack Mount Kit:
19-inch rack.
Model 4288-2 Side-by-Side Rack Mount Kit:
486, 487, 2000, 2001, 2002, 2010, 2015, 2016, 6514, 6517, 7001) side-by-side in a standard
19-inch rack.
Model 4288-3 Side-by-Side Rack Mount Kit:
side-by-side in a standard 19-inch rack.
Model 4288-4 Side-by-Side Rack Mount Kit:
instrument (Models 195A, 196, 220, 224, 230, 263, 595, 614, 617, 705, 740, 775, etc.)
side-by-side in a standard 19-inch rack.
Carrying case
Model 1050 Padded Carrying Case:
shoulder strap.
Mounts a single Model 2016 in a standard
Mounts two instruments (Models 182, 428,
Mounts a Model 2016 and a Model 199
Mounts a Model 2016 and a 5.25-inch
A carrying case for a Model 2016. Includes handles and
1-6General Information
Basic Mea-
surements
2
Basic
Measurements
2-2Basic Measurements
Introduction
This section summarizes front panel operation of the Model 2016. It is organized as follows:
•
Front panel summary —
connections.
•
Rear panel summary —
•
Power-up —
the warm-up time, and default conditions.
•
Display —
instrument.
•
Measuring voltage —
level voltage considerations.
•
Measuring current —
fuse replacement.
•
Measuring resistance —
shielding considerations.
•
Measuring frequency and period —
connections.
•
Measuring temperature —
measurements.
•
Math —
readings.
•
Measuring continuity —
•
Testing diodes —
•
Measuring distortion —
calculations used.
Includes an illustration and summarizes keys, display, and
Includes an illustration and summarizes connections.
Describes connecting the instrument to line power, the power-up sequence,
Discusses the display format and messages that may appear while using the
Covers DC and AC voltage measurement connections and low
Covers DC and AC current measurement connections and current
Details two and four-wire measurement connections and
Covers frequency and period measurement
Describes the use of thermocouples for temperature
Covers the mX+b, percent, dBm, and dB math functions performed on single
Explains setting up and measuring continuity of a circuit.
Describes testing general-purpose and zener diodes.
Explains how to measure total harmonic distortion and the
Front panel summary
F
M
f
The front panel of the Model 2016 is shown in Figure 2-1. This figure includes important
abbreviated information that should be reviewed before operating the instrument.
Basic Measurements2-3
igure 2-1
odel 2016
ront panel
5
1
3
SHIFT
LOCAL
POWER
1Function keys
STEPCH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
SCAN
TALK
LSTN
SRQ
SHIFT
TIMER
MATH
DCV
EX TRIG
SAVE SETUP
SOURCE
CH1REM
HOLD TRIG FAST MED SLOWAUTO ERR
THD
ACV
HOLD
TRIG
MEAS
THD
dBm
DCI
LIMITS ON/OFFDELAY
STORE
RECALL
CONFIG HALT
STEP SCAN
dB
ACI
REL FILT
CONT
Ω2Ω
TEST
GPIB
DIGITS RATE
(shifted and unshifted)
RELFILTER
RS232
CAL
BUFFER
4
PERIOD TCOUPL
FREQ
MATH
REAR
4W
STAT
2016 THD MULTIMETER
TEMP
RANGE
AUTO
EXIT ENTER
RANGE
4
350V
PEAK
F
SENSE
Ω 4 WIRE
INPUTS
FRONT/REAR
782
INPUT
HI
1000V
!
PEAK
LO
500V
PEAK
R
3A 250V
AMPS
6
Select measurement function (DC and AC voltage, DC and AC current, 2-wire and 4-wire
resistance, frequency, period, temperature with thermocouples), math function (mX+b, %,
dBm, dB), THD (total harmonic distortion) or special function (continuity, diode test).
2Operation keys
EX TRIGSelects external triggers (front panel, bus, trigger link) as the trigger
source.
TRIG Triggers a measurement from the front panel.
STORE Enables reading storage.
RECALLDisplays stored readings and buffer statistics (maximum, minimum, aver-
FILTERDisplays digital filter status for present function and toggles filter on/off.
RELEnables/disables relative reading on present function (not applicable for
age, standard deviation). Use ▲ and ▼ to scroll through buffer; use
and to toggle between reading number and reading.
distortion measurements).
and Moves through selections within functions and operations. If scanner
card installed, manually scans channels.
THD SOURCESelects and configures the internal function generator.
THD MEASConfigures distortion measurements.
STEPSteps through channels; sends a trigger after each channel.
SCANScans through channels; sends a trigger after last channel.
DIGITSChanges number of digits of resolution.
RATEChanges reading rate: fast, medium, slow.
EXITCancels selection, moves back to measurement display.
ENTERAccepts selection, moves to next choice or back to measurement display.
SHIFTUsed to access shifted keys.
LOCALCancels GPIB remote mode.
2-4Basic Measurements
3Shifted operation keys
DELAYSets user delay between trigger and measurement.
HOLDHolds reading when the selected number of samples is within the selected
LIMITSSets upper and lower limit values for readings.
ON/OFFEnables/disables limits; selects beeper operation for limit testing.
TESTSelects built-in tests, diagnostics, display test.
CALAccesses calibration.
SAVESaves present configuration for power-on user default.
SETUPRestores factory or user default configuration.
CONFIGSelects minimum/maximum channels, timer, and reading count for step/scan.
HALTTurns off step/scan.
GPIBEnables/disables GPIB interface; selects address and language.
RS232Enables/disables RS-232 interface; selects baud rate, flow control, terminator.
4Range keys
▲Moves to higher range; increments digit; moves to next selection.
▼Moves to lower range; decrements digit; moves to previous selection.
AUTOEnables/disables autorange. SHIFT-AUTO displays the most recent error
5Annunciators
* (asterisk)Reading being stored.
(diode)Instrument is in diode testing function.
)))
(speaker)Beeper on for continuity or limits testing.
(more)Indicates additional selections are available.
4W4-wire resistance reading displayed.
AUTOAutoranging enabled.
BUFFERRecalling stored readings.
ERRQuestionable reading; invalid cal step.
FASTFast reading rate.
FILTDigital filter enabled.
HOLDInstrument is in hold mode.
LSTNInstrument addressed to listen over GPIB.
MATHMath function (mX+b, %, dB, dBm) enabled.
MEDMedium reading rate.
REARReading acquired from rear inputs.
RELRelative reading displayed.
REMInstrument is in GPIB remote mode.
SCANInstrument is in scan mode.
SHIFTAccessing shifted keys.
SLOWSlow reading rate.
SRQService request over GPIB.
STATDisplaying buffer statistics.
STEPInstrument is in step mode.
TALKInstrument addressed to talk over GPIB.
TIMERTimed scans in use.
TRIGIndicates external trigger (front panel, bus, trigger link) selected.
tolerance.
message.
Basic Measurements2-5
6Input connections
INPUT HI and LOUsed for making DC volts, AC volts, 2-wire resistance measurements.
AMPSUsed in conjunction with INPUT LO to make DC current and AC
SENSE
Ω4 WIREUsed with INPUT HI and LO to make 4-wire resistance measure-
HI and LOments.
current measurements. Also holds current input fuse (3A, 250V, fast
blow, 5×20mm).
7INPUTS
Selects input connections on front or rear panel.
8Handle
Pull out and rotate to desired position.
2-6Basic Measurements
Rear panel summary
The rear panel of the Model 2016 is shown in Figure 2-2. This figure includes important
abbreviated information that should be reviewed before operating the instrument.
Figure 2-2
Model 2016
rear panel
234
1
HI
!
LO
42V PEAK
INPUT
1000V
PEAK
!
INV/PULSE
SOURCE
OUTPUT
500V
PEAK
350V
PEAK
SOURCE
OUTPUT
SENSE
Ω 4W
7
6
FUSE
500 mAT
(SB)
250 mAT
(SB)
TRIGGER
LINK
!
3 5
1
VMC
4 6
2
EXT TRIG
!
LINE
LINE RATING
100 VAC
50, 60Hz
120 VAC
40 VA MAX
220 VAC
240 VAC
8
7
5
2
1
#2
EXTERNAL TRIGGER INPUT
Trigger Reading
>10µsec
TTL HI
TTL LO
MADE IN
U.S.A.
RS232
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
6
34
#1
VOLT METER COMPLETE OUTPUT
Reading
Complete
>10µsec
TTL HI
TTL LO
IEEE-488
120
5
Basic Measurements2-7
1Input connections
INPUT HI and LOUsed for making DC volts, AC volts, 2-wire resistance measurements.
SENSE
Ω4 WIREUsed with INPUT HI and LO to make 4-wire resistance measurements.
HI and LO
2TRIGGER LINK
One 8-pin micro-DIN connector for sending and receiving trigger pulses among other
instruments. Use a trigger link cable or adapter, such as Models 8501-1, 8501-2, 8502, 8504.
3RS-232
Connector for RS-232 operation. Use a straight-through (not null modem) DB-9 cable.
4IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as Models 7007-1 and
7007-2.
5Power module
Contains the AC line receptacle, power line fuse, and line voltage setting. The Model 2016
can be configured for line voltages of 100V/120V/220V/240VAC at line frequencies of 45Hz
to 66Hz.
6INV/PULSE SOURCE OUTPUT
A BNC connector that provides an inverted sine wave or pulsed output for exciting
devices under test during distortion measurement or for general purpose use. This
connector also may provide a square wave pulse for triggering or synchronizing other
systems to the SOURCE OUTPUT sine wave.
7SOURCE OUTPUT
A BNC connector that provides a sine wave output for exciting devices under test during
distortion measurement or for general purpose use.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
2-8Basic Measurements
Power-up
Line power connection
Follow the procedure below to connect the Model 2016 to line power and turn on the
instrument.
1.Check to see that the line voltage selected on the rear panel (see Figure 2-3) is correct
for the operating voltage in your area. If not, refer to the next procedure, “Setting line
voltage and replacing fuse.”
CAUTIONOperating the instrument on an incorrect line voltage may cause damage to
2.Before plugging in the power cord, make sure that the front panel power switch is in the
off (0) position.
3.Connect the female end of the supplied power cord to the AC receptacle on the rear
panel. Connect the other end of the power cord to a grounded AC outlet.
WARNINGThe power cord supplied with the Model 2016 contains a separate ground
the instrument, possibly voiding the warranty.
wire for use with grounded outlets. When proper connections are made,
instrument chassis is connected to power line ground through the ground
wire in the power cord. Failure to use a grounded outlet may result in
personal injury or death due to electric shock.
4.Turn on the instrument by pressing the front panel power switch to the on (1) position.
Figure 2-3
Power module
350V
PEAK
SOURCE
OUTPUT
Model 2016
HI
1000V
TRIGGER
PEAK
!
500V
PEAK
LO
SENSE
INPUT
Ω 4W
42V PEAK
INV/PULSE
SOURCE
OUTPUT
35
1
46
2
!
!
FUSELINE
100 VAC
500 mAT
120 VAC
(SB)
220 VAC
250 mAT
240 VAC
(SB)
LINK
!
VMCEXT TRIG
LINE RATING
50, 60
25 VA MAX
MADE IN
U.S.A.
RS232
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
120
Fuse
Spring
120
Line Voltage Selector
Window
Fuse Holder Assembly
Setting line voltage and replacing fuse
A rear panel fuse located next to the AC receptacle protects the power line input of the
instrument. If the line voltage setting needs to be changed or the line fuse needs to be replaced,
perform the following steps.
WARNINGMake sure the instrument is disconnected from the AC line and other
equipment before changing the line voltage setting or replacing the line
fuse.
1.Place the tip of a flat-blade screwdriver into the power module by the fuse holder
assembly (see Figure 2-3). Gently push in and to the left. Release pressure on the
assembly and its internal spring will push it out of the power module.
2.Remove the fuse and replace it with the type listed in Table 2-1.
CAUTIONFor continued protection against fire or instrument damage, only replace
fuse with the type and rating listed. If the instrument repeatedly blows
fuses, locate and correct the cause of the trouble before replacing the fuse.
See the Model 2016 Service Manual for troubleshooting information.
3.If configuring the instrument for a different line voltage, remove the line voltage selector
from the assembly and rotate it to the proper position. When the selector is installed into
the fuse holder assembly, the correct line voltage appears inverted in the window.
4.Install the fuse holder assembly into the power module by pushing it in until it locks in
place.
Basic Measurements2-9
Table 2-1
Fuse ratings
Line voltageFuse ratingKeithley P/N
100/120V
220/240V
0.5A, 250V, slo-blo, 5 × 20 mm
0.25A, 250V, slo-blo, 5 × 20 mm
FU-71
FU-96-4
2-10Basic Measurements
Power-up sequence
On power-up, the Model 2016 performs self-tests on its EPROM and RAM and momentarily
lights all segments and annunciators. If a failure is detected, the instrument momentarily displays
an error message and the ERR annunciator turns on. (Error messages are listed in Appendix B.)
NOTEIf a problem develops while the instrument is under warranty, return it to Keithley
If the instrument passes the self-tests, the firmware revision levels are displayed. An example
of this display is:
REV: A01 A02
where: A01 is the main board ROM revision.
NOTEThe unit will display USER SETUP during power-up if a user setup has been saved.
After the power-up sequence, the instrument begins its normal display of readings.
Instruments, Inc., for repair.
A02 is the display board ROM revision.
High energy circuit safety precautions
To optimize safety when measuring voltage in high energy distribution circuits, read and use
the directions in the following warning.
WARNINGS Dangerous arcs of an explosive nature in a high energy circuit can cause
severe personal injury or death. If the multimeter is connected to a high
energy circuit when set to a current range, low resistance range, or any
other low impedance range, the circuit is virtually shorted. Dangerous
arcing can result even when the multimeter is set to a voltage range if the
minimum voltage spacing is reduced in the external connections.
• When making measurements in high energy circuits, use test leads that
meet the following requirements:
• Test leads should be fully insulated.
• Only use test leads that can be connected to the circuit (e.g., alligator
clips, spade lugs, etc.) for hands-off measurements.
• Do not use test leads that decrease voltage spacing. These diminish arc
protection and create a hazardous condition.
Use the following sequence when testing power circuits:
Basic Measurements2-11
1. De-energize the circuit using the regular installed connect-disconnect
device, such as a circuit breaker, main switch, etc.
2. Attach the test leads to the circuit under test. Use appropriate safety
rated test leads for this application.
3. Set the multimeter to the proper function and range.
4. Energize the circuit using the installed connect-disconnect device and
make measurements without disconnecting the multimeter.
5. De-energize the circuit using the installed connect-disconnect device.
6. Disconnect the test leads from the circuit under test.
The maximum common-mode voltage (voltage between INPUT LO and the
chassis ground) is 500V peak. Exceeding this value may cause a breakdown
in insulation, creating a shock hazard.
The maximum common-made voltage (voltage between SOURCE
OUTPUT and the chassis ground, and INV/PULSE SOURCE OUTPUT
and the chassis ground) is 42V peak. Exceeding this value may cause a
breakdown in insulation, creating a shock hazard.
2-12Basic Measurements
Power-on defaults
Power-on defaults are the settings the instrument assumes when it is turned on. The
Model 2016 offers two choices for the settings: factory and user. The power-on default will be
the last configuration you saved. The SAVE and SETUP keys select the two choices of power-on
defaults.
To save present configuration as user settings:
1.Configure the instrument as desired for USER default.
2.Press SHIFT then SAVE.
3.Use the ▲ and ▼ keys to select FACTory or USER.
4.Press ENTER.
NOTEThe unit will display USER SETUP during power-up if a user setup has been saved.
To restore factory or user settings:
1.Press SHIFT then SETUP.
2.Use the ▲ and ▼ keys to select FACTory or USER.
3.Press ENTER.
Since the basic measurement procedures in this manual assume the factory defaults, reset the
instrument to the factory settings when following step-by-step procedures. Table 2-2 lists the
factory default settings.
Table 2-2
Factory defaults
SettingFactory default
Basic Measurements2-13
Autozero
Buffer
Continuity
Beeper
Digits
Rate
Threshold
Current (AC and DC)
Digits (AC)
Digits (DC)
Filter
Count
Mode
Range
Relative
Value
Rate (AC)
Rate (DC)
Diode test
Digits
Range
Rate
Distortion
Measurement type
Frequency
Number of harmonics
Units
Shaping filter
Fundamental frequency
Bandpass filter
Low cutoff
State
High cutoff
State
Frequency and Period
Digits
Range
Relative
Value
Rate
Function
*DETector:BANDwidth 30
On
No effect
On
4H
Fast (0.1 PLC)
10Ω
5½
6½
On
10
Moving average
Auto
Off
0.0
Medium*
Medium (1 PLC)
6½
1mA
Medium (1 PLC)
THD
Auto
2
Percent
None
60Hz
20Hz
Off
50kHz
Off
6½
10V
Off
0.0
Slow (1 sec)
DCV
2-14Basic Measurements
Table 2-2 (cont.)
Factory defaults
SettingFactory default
GPIB
Address
Limits
Beeper
High limit
Low limit
mX+b
Scale factor
Offset
Percent
References
Resistance (2-wire and 4-wire)
Digits
Filter
Count
Mode
Range
Relative
Value
Rate
RS-232
Baud
Flow
Tx term
Scanning
Source output
Sine wave frequency
Sine wave output impedance
Sine wave amplitude
Sine wave channel 2 shape
Temperature
Digits
Filter
Count
Mode
Junction
Temperature
Relative
Value
Rate
Thermocouple
Units
*DETector:BANDwidth 30
No effect
(16 at factory)
Off
Never
+1
-1
Off
1.0
0.0
Off
1.0
½
On
10
Moving average
Auto
Off
0.0
Medium (1 PLC)
Off
No effect
No effect
No effect
Off
Off
60Hz
50Ω
0.5Vrms with 50Ω impedance
Inverted sine
5½
On
10
Moving average
Simulated
23°C
Off
0.0
Medium (1 PLC)
J
°C
Table 2-2 (cont.)
Factory defaults
SettingFactory default
Triggers
Continuous
Delay
Source
Voltage (AC and DC)
dB reference
dBm reference
Digits (AC)
Digits (DC)
Filter
Count
Mode
Range
Relative
Value
Rate (AC)
Rate (DC)
*DETector:BANDwidth 30
On
Auto
Immediate
No effect
75Ω
5½
½
On
10
Moving average
Auto
Off
0.0
Medium*
Medium (1 PLC)
Basic Measurements2-15
GPIB primary address
The GPIB primary address of the instrument must be the same as the primary address you
specify in the controller’s programming language. The default primary address of the instrument
is 16, but you can set the address to any value from 0 to 30 by using the following step by step
instructions.
1.Press SHIFT then GPIB.
2.Use the ▲ and ▼ keys to select ADDRess. Or, press ENTER. Once you have pressed
ENTER, the unit automatically displays the address selection.
3.Use the and keys to toggle from ADDRess to the numeric entry. Notice the
values are blinking.
4.Use the ▲ and ▼ keys to change the numeric entries to the desired address.
5.Press ENTER.
See Section Four — Remote Operation for more GPIB information.
Warm-up time
The Model 2016 is ready for use as soon as the power-up sequence has completed. However,
to achieve rated accuracy, allow the instrument to warm up for one hour. If the instrument has
been subjected to extreme temperatures, allow additional time for internal temperatures to
stabilize.
2-16Basic Measurements
Display
The display of the Model 2016 is primarily used to display readings, along with the units and
type of measurement. Annunciators are located on the top, bottom, right, and left of the reading
or message display. The annunciators indicate various states of operation. See Figure 2-1 for a
complete listing of annunciators.
Status and error messages
Status and error messages are displayed momentarily. During Model 2016 operation and
programming, you will encounter a number of front panel messages. Typical messages are either
of status or error variety, as listed in Appendix B.
Measuring voltage
The Model 2016 can make DCV measurements from 0.1µV to 1000V and ACV
measurements from 0.1µV to 750V RMS, 1000V peak.
Connections
Assuming factory default conditions, the basic procedure is as follows:
1.Connect test leads to the INPUT HI and LO terminals. Either the front or rear inputs can
2.Select the measurement function by pressing DCV or ACV.
3.Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
4.Connect test leads to the source as shown in Figure 2-4.
WARNING
CAUTION
5.Observe the display. If the “OVERFLOW” message is displayed, select a higher range
6.Take readings from the display.
be used; place the INPUTS button in the appropriate position.
autoranging. If you want manual ranging, use the RANGE
measurement range consistent with the expected voltage.
Maximum common mode voltage (voltage between LO and chassis ground)
is 500V peak. Exceeding this value may cause a shock hazard.
Do not apply more than 1000V peak to the input or instrument damage
may occur. The voltage limit is subject to the 8
until a normal reading is displayed (or press AUTO for autoranging). Use the lowest
possible range for the best resolution.
▲
and ▼ keys to select a
×
107V•Hz product.
Fi
4
D
Basic Measurements2-17
gure 2-
C and AC voltage
measurements
SHIFT
LOCAL
POWER
TALK
LSTN
SRQ
SHIFT
TIMER
MATH
DCV
EX TRIG
SAVE SETUP
SOURCE
THD
Model 2016
CH1REM
STEPCH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
SCAN
dBm
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
RECALL
REL FILT
dB
ACI
HOLD TRIG FAST MED SLOWAUTO ERR
THD
ACV
HOLD
TRIG
MEAS
CONT
Ω2 Ω4
TEST
GPIB
DIGITS RATE
RELFILTER
RS232
CAL
BUFFER
PERIOD TCOUPL
FREQ
EXIT ENTER
MATH
REAR
4W
STAT
2016 THD MULTIMETER
TEMP
RANGE
RANGE
SENSE
INPUT
Ω 4 WIRE
HI
350V
PEAK
AUTO
INPUTS
F
FRONT/REAR
1000V
!
PEAK
LO
500V
PEAK
R
3A 250V
AMPS
DC Voltage
Source
Input Impedance = 10M on 1000V and 100V ranges;
> 10G on 10V, 1V and 100mV ranges
Caution: Maximum Input = 1010V peak
Maximum Common Mode = 500V peak
Model 2016
SENSE
INPUT
Ω 4 WIRE
350V
PEAK
INPUTS
F
FRONT/REAR
HI
1000V
!
PEAK
LO
500V
PEAK
R
3A 250V
AMPS
AC Voltage
Source
CAL
RELFILTER
RS232
MATH
REAR
4W
BUFFER
STAT
2016 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
EXIT ENTER
RANGE
AUTO
RANGE
CH1REM
STEPCH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
SCAN
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD TRIG FAST MED SLOWAUTO ERR
MATH
SHIFT
LOCAL
POWER
DCV
EX TRIG
SAVE SETUP
SOURCE
THD
ACV
HOLD
TRIG
MEAS
THD
dBm
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
RECALL
REL FILT
dB
CONT
ACI
Ω2 Ω4
TEST
GPIB
DIGITS RATE
Input Impedance = 1M and 100pF
Caution: Maximum Input = 1000V peak, 8
Maximum Common Mode = 500V peak
Crest factor
AC voltage and current accuracies are affected by the crest factor of the waveform, the ratio
of the peak value to the RMS value. 500Hz is the maximum fundamental frequency at which the
corresponding crest factor must be taken into account for accuracy calculations.
Low level considerations
For sensitive measurements, external considerations beyond the Model 2016 affect the
accuracy. Effects not noticeable when working with higher voltages are significant in microvolt
signals. The Model 2016 reads only the signal received at its input; therefore, it is important that
this signal be properly transmitted from the source. The following paragraphs indicate factors
that affect accuracy, including stray signal pick-up and thermal offsets.
×107 V•Hz
2-18Basic Measurements
Shielding
AC voltages that are extremely large compared with the DC signal to be measured may produce
an erroneous output. Therefore, to minimize AC interference, the circuit should be shielded with
the shield connected to the Model 2016 INPUT LO (particularly for low level sources). Improper
shielding can cause the Model 2016 to behave in one or more of the following ways:
•Unexpected offset voltages.
•Inconsistent readings between ranges.
•Sudden shifts in reading.
To minimize pick-up, keep the voltage source and the Model 2016 away from strong AC
magnetic sources. The voltage induced due to magnetic flux is proportional to the area of the
loop formed by the input leads. Therefore, minimize the loop area of the input leads and connect
each signal at only one point.
Thermal EMFs
Thermal EMFs (thermoelectric potentials) are generated by thermal differences between the
junctions of dissimilar metals. These can be large compared to the signal that the Model 2016
can measure. Thermal EMFs can cause the following conditions:
•Instability or zero offset is much higher than expected.
•The reading is sensitive to (and responds to) temperature changes. This effect can be
demonstrated by touching the circuit, by placing a heat source near the circuit, or by a
regular pattern of instability (corresponding to changes in sunlight or the activation of
heating and air conditioning systems).
To minimize the drift caused by thermal EMFs, use copper leads to connect the circuit to the
Model 2016. A banana plug generates a few microvolts. A clean copper conductor such as #10
bus wire is ideal for this application. The leads to the input may be shielded or unshielded, as
necessary. Refer to “Shielding”.
Widely varying temperatures within the circuit can also create thermal EMFs. Therefore,
maintain constant temperatures to minimize these thermal EMFs. A shielded enclosure around
the circuit under test also helps by minimizing air currents.
The REL control can be used to null out constant offset voltages.
Basic Measurements2-19
AC voltage offset
The Model 2016, at 5½-digits resolution, will typically display 100 counts of offset on AC
volts with the input shorted. This offset is caused by the offset of the TRMS converter. This offset
will not affect reading accuracy and should not be zeroed out using the REL feature. The
following equation expresses how this offset (V
Displayed readingVIN()2V
+=
Example: Range = 1VAC
Offset = 100 counts (1.0mV)
Input = 100mV RMS
()
) is added to the signal input (VIN):
OFFSET
OFFSET
2
Displayed reading100mV()21.0mV()
0.01V()1106–V×()+=
Displayed reading0.100005=
+=
2
The offset is seen as the last digit, which is not displayed. Therefore, the offset is negligible.
If the REL feature were used to zero the display, the 100 counts of offset would be subtracted
from V
, resulting in an error of 100 counts in the displayed reading.
IN
See Section 3 — Measurement Options for information that explain the configuration options
for DC and AC voltage measurements.
2-20Basic Measurements
Measuring current
The Model 2016 can make DCI measurements from 10nA to 3A and ACI measurements from
1µAm to 3A RMS.
NOTESee the previous discussion about crest factor in “Measuring voltage” in this section.
Connections
Assuming factory default conditions, the basic procedure is as follows:
1.Connect test leads to the AMPS and INPUT LO terminals. The front inputs must be used;
place the INPUTS button in the FRONT position.
2.Select the measurement function by pressing DCI or ACI.
3.Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE ▲ and ▼ keys to select a
measurement range consistent with the expected current.
4.Connect test leads to the source as shown in Figure 2-5.
CAUTIONDo not apply more than 3A, 250V to the input or the AMPS fuse will
open-circuit.
Figure 2-5
DC and AC current
measurements
5.Observe the display. If the “OVERFLOW” message is displayed, select a higher range
until a normal reading is displayed (or press AUTO for autoranging). Use the lowest
possible range for the best resolution.
6.Take readings from the display.
Model 2016
SENSE
INPUT
Ω 4 WIRE
350V
PEAK
INPUTS
F
FRONT/REAR
HI
1000V
!
PEAK
LO
500V
PEAK
R
3A 250V
AMPS
Current
Source
SHIFT
LOCAL
POWER
CH1REM
STEPCH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
SCAN
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD TRIG FAST MED SLOWAUTO ERR
MATH
THD
DCV
ACV
HOLD
EX TRIG
TRIG
SAVE SETUP
SOURCE
MEAS
THD
dBm
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
RECALL
REL FILT
dB
CONT
ACI
Ω2 Ω4
TEST
GPIB
DIGITS RATE
CAL
RELFILTER
RS232
MATH
REAR
4W
BUFFER
STAT
2016 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
EXIT ENTER
Caution: Maximum Input = 3A DC or RMS
RANGE
AUTO
RANGE
AMPS fuse replacement
WARNINGMake sure the instrument is disconnected from the power line and other
equipment before replacing the AMPS fuse.
1.Turn off the power and disconnect the power line and test leads.
2.From the front panel, gently push in the AMPS jack with your thumb and rotate the fuse
carrier one-quarter turn counterclockwise. Release pressure on the jack and its internal
spring will push the jack out of the socket.
3.Remove the fuse and replace it with the same type (3A, 250V, fast blow, 5 × 20mm). The
Keithley part number is FU-99-1.
CAUTIONDo not use a fuse with a higher current rating than specified or instrument
damage may occur. If the instrument repeatedly blows fuses, locate and
correct the cause of the trouble before replacing the fuse. See the
Model 2016 Service Manual for troubleshooting information.
4.Install the new fuse by reversing the procedure above.
See Section 3 — Measurement Options for information that explains the configuration
options for DC and AC current measurements.
Basic Measurements2-21
2-22Basic Measurements
Measuring resistance
The Model 2016 can make 2-wire and 4-wire resistance measurements from 100µΩ to
120MΩ.
Connections
Assuming factory default conditions, the basic procedure is as follows:
1.Connect test leads to the Model 2016 as follows:
A. For Ω2-wire, connect the test leads to INPUT HI and LO.
B. For Ω4-wire, connect the test leads to INPUT HI and LO, and SENSE Ω4 WIRE
HI and LO. Recommended Kelvin test probes include the Keithley Models 5805
and 5806. Either the front or rear inputs can be used; place the INPUTS button in
the appropriate position.
2.Select the measurement function by pressing Ω2 or Ω4.
3.Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE ▲ and ▼ keys to select a
measurement range consistent with the expected resistance.
4.Connect test leads to the resistance as shown in Figure 2-6.
CAUTIONDo not apply more than 1000V peak between INPUT HI and LO or
instrument damage may occur.
5.Observe the display. If the “OVERFLOW” message is displayed, select a higher range
until a normal reading is displayed. Use the lowest possible range for the best resolution.
resistance in a shielded enclosure and connect the shield to the INPUT LO terminal of the
instrument electrically.
options for 2-wire and 4-wire resistance measurements.
To achieve a stable reading, it helps to shield resistances greater than 100kΩ. Place the
See Section 3 — Measurement Options for information that explains the configuration
2-24Basic Measurements
Measuring frequency and period
The Model 2016 can make frequency measurements from 3Hz to 500kHz on voltage ranges
of 100mV, 1V, 10V, 100V, and 750V. Period measurements can be taken from 2µs to 333ms on
the same voltage ranges as the frequency.
The instrument uses the volts input terminals to measure frequency. The AC voltage range can
be changed with the RANGE ▲ and ▼ keys. The signal voltage must be greater than 10% of
the full-scale range.
CAUTIONThe voltage limit is subject to the 8 × 10
Trigger level
Frequency and Period use a zero-crossing trigger, meaning that a count is taken when the
frequency crosses the zero level. The Model 2016 uses a reciprocal counting technique to
measure frequency and period. This method generates constant measurement resolution for any
input frequency. The multimeter’s AC voltage measurement section performs input signal
conditioning.
Gate time
The gate time is the amount of time the Model 2016 uses to sample frequency or period
readings. The RATE key setting yields gate times as follows:
FAST = 0.01s
MEDium = 0.1s
SLOW = 1.0s
The Model 2016 completes a reading when it receives its first zero-crossing after the gate
time expires. In other words, the reading is completed 1/2 cycle after the gate time has expired.
For example, with a 1sec gate time to sample a 3Hz frequency, you may wait up to 3 seconds
before the Model 2016 returns a reading.
7
V•Hz product.
Connections
Assuming factory default conditions, the basic procedure is as follows:
1.Connect test leads to the INPUT HI and LO terminals of the Model 2016. Either the front
2.Select the FREQ or PERIOD function.
3.Connect test leads to the source as shown in Figure 2-7.
CAUTIONDo not exceed 1000V peak between INPUT HI and INPUT LO or
4.Take a reading from the display.
See Section 3 — Measurement Options for information that explains the configuration
options for frequency and period measurements.
Basic Measurements2-25
or rear inputs can be used; place the INPUTS button in the appropriate position.
instrument damage may occur.
Figure 2-7
Frequency
and period
measurements
Model 2016
SHIFT
LOCAL
POWER
CH1REM
STEPCH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
SCAN
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD TRIG FAST MED SLOWAUTO ERR
MATH
THD
DCV
ACV
HOLD
EX TRIG
TRIG
SAVE SETUP
SOURCE
MEAS
THD
dBm
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
RECALL
REL FILT
dB
CONT
ACI
Ω2 Ω4
TEST
GPIB
DIGITS RATE
CAL
RELFILTER
RS232
MATH
REAR
4W
BUFFER
STAT
2016 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
EXIT ENTER
RANGE
AUTO
RANGE
Input Impedance = 1M in parallel with <100pF
Caution: Maximum Input = 1000V peak, 8
×107 V•Hz
350V
PEAK
SENSE
Ω 4 WIRE
INPUTS
F
FRONT/REAR
INPUT
HI
1000V
!
PEAK
LO
500V
PEAK
R
3A 250V
AMPS
AC Voltage
Source
F
2-26Basic Measurements
Measuring temperature
The Model 2016 measures temperature with thermocouples. The temperature measurement
ranges available depend on the type of thermocouple chosen.
Thermocouples can be connected to an external thermocouple card, such as a Model 7057A,
7402, or 7014 installed in a Model 7001 or 7002 Switch System.
Connections
igure 2-8
Thermocouple
temperature
measurements
POWER
TALK
LSTN
SRQ
SHIFT
TIMER
MATH
SHIFT
DCV
LOCAL
EX TRIG
SAVE SETUP
SOURCE
Note: This thermocouple card must be inserted into a
Keithley Model 7001 or 7002 Switch System.
Model 2016
STEPCH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
CH1REM
SCAN
HOLD TRIG FAST MED SLOWAUTO ERR
REL FILT
dBm
dB
CONT
THD
ACI
ACV
HOLD
TRIG
MEAS
THD
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
Ω2 Ω4
TEST
CAL
RECALL
RELFILTER
RS232
GPIB
DIGITS RATE
MATH
REAR
BUFFER
STAT
2016 THD MULTIMETER
PERIOD TCOUPL
FREQ
EXIT ENTER
SENSE
Ω 4 WIRE
4W
TEMP
HI
350V
PEAK
LO
INPUTS
F
R
RANGE
AUTO
FRONT/REAR
RANGE
Input
INPUT
!
3A 250V
AMPS
HI
1000V
PEAK
500V
PEAK
Input
LO
Cable
Clamp
Out HI
Out LO
H L
CH 9
H L
CH 8
Math
Basic Measurements2-27
Configuration
The following information explains the various configuration options for temperature
measurements. To select and configure the thermocouple measurement:
Press SHIFT then TCOUPL. Three choices are available using the ▲ and ▼ keys:
•UNITS — C, K, F (Centigrade, Kelvin, Fahrenheit). This parameter selects the displayed
units for temperature measurements.
•TYPE — J, K, T (thermocouple type).
•JUNC — SIM. Typically, a thermocouple card uses a single reference junction. The
Model 2016 simulates a reference junction temperature. Typical reference junction
temperatures are 0°C and 23°C.
A simulated reference temperature is the temperature of the junction where the thermocouple
voltage is sensed. It is room temperature if the thermocouple wire is terminated to banana jacks
and corrected directly to the multimeter. The accuracy of a temperature measurement depends
on the accuracy of the reference junction.
Model 2016 math operations are divided into four categories:
•mX+b and percent
•dBm and dB calculations
•Statistics of buffered readings
•Limit testing
The first two categories are discussed here; buffered reading statistics and reading limit
testing are described in Section 3 — Measurement Options.
The procedure to select and configure a math operation is summarized as follows:
1.Press SHIFT then the appropriate math key.
2.Configure the parameters for the math operation. Press ENTER when done. (Press
SHIFT then the related math function to end the calculation.)
NOTES Once enabled for a function, the mX+b and percentage calculations are in effect
across function changes.
The Model 2016 uses IEEE-754 floating point format for math calculations.
2-28Basic Measurements
MX + B
This math operation lets you manipulate normal display readings (X) mathematically
according to the following calculation:
Y= mX + b
where: X is the normal display reading
m and b are user-entered constants for scale factor and offset
Y is the displayed result
Configuration
To configure the mX+b calculation, perform the following steps:
1.Press SHIFT then MATH to display a math calculation. Use the ▲ and ▼ keys to select
MX+B.
2.Press ENTER to display the present scale factor:
M: +1.000000^
3.Enter a value and units prefix. Use the and keys to choose a numerical place and
use the ▲ and ▼ keys to increment or decrement the digits.
4.Press ENTER to confirm the M value and display the B value:
B: +00.00000 m
5.Enter a value and units prefix.
6.Press ENTER to confirm the B value and display the UNITS designation:
MXB
7.Scroll through the letters to change and press ENTER when done.
The Model 2016 then displays the result of the calculation.
Percent
Basic Measurements2-29
This item selects the percentage calculation and lets you specify a reference value. The
displayed reading will be expressed as a percent deviation from the reference value. The
percentage calculation is performed as follows:
Input - Reference
Percent
where: Input is the normal display reading.
Reference is the user entered constant.
Percent is the displayed result.
------------------------------------------
Reference
100%×=
Configuration
To configure the percent calculation, perform the following steps:
1.Press SHIFT then MATH to display a math calculation. Use the ▲ and ▼ keys to select
PERC.
2.Press ENTER to display the present value:
REF:+1.000000^
3.Enter a reference sign, value, and units prefix. Use the and keys to choose a
numerical place and use the ▲ and ▼ keys to increment or decrement the digits.
4.Press ENTER when done.
The Model 2016 will display the result of the calculation. The result is positive when the input
exceeds the reference and negative when the input is less than the reference. Engineering units
are used to show values in the range 1 nano to 1000G. Exponential notation is used above that
range.
2-30Basic Measurements
dBm calculation
dBm is defined as decibels above or below a 1mW reference. With a user-programmable
reference impedance, the Model 2016 reads 0dBm when the voltage needed to dissipate 1mW
through the reference impedance is applied. The relationship between dBm, a reference
impedance, and the voltage is defined by the following equation:
dBm = 10 log
2
/Z
V
REF
IN
--------------------------------1mW
Where: V
is the DC or AC input signal.
IN
Z
is the specified reference impedance.
REF
NOTEDo not confuse reference impedance with input impedance. The input impedance of
the instrument is not modified by the dBm parameter.
If a relative value is in effect when dBm is selected, the value is converted to dBm then REL
is applied to dBm. If REL is applied after dBm has been selected, dBm math has REL applied
to it.
Configuration
To set the reference impedance, perform the following steps:
1.After selecting dBm, the present reference impedance is displayed (1-9999Ω):
REF: 0000
2.To change the reference impedance, use the and keys to select the numeric
position. Then use the ▲ and ▼ keys to select the desired value. Be sure to press ENTER
after changing the reference impedance.
NOTES dBm is valid for positive and negative values of DC volts.
The mX+b and percent math operations are applied after the dBm or dB math. For
example, if mX+b is selected with m=10 and b=0, the display will read 10.000 MXB
for a 1VDC signal. If dBm is selected with Z
= 50Ω, the display will read 130MXB.
REF
dB calculation
Expressing DC or AC voltage in dB makes it possible to compress a large range of
measurements into a much smaller scope. The relationship between dB and voltage is defined
by the following equation:
dB= 20 log
V
IN
-----------------V
REF
Basic Measurements2-31
where: V
is the DC or AC input signal.
IN
is the specified voltage reference level.
V
REF
The instrument will read 0dB when the reference voltage level is applied to the input.
If a relative value is in effect when dB is selected, the value is converted to dB then REL is
applied to dB. If REL is applied after dB has been selected, dB has REL applied to it.
Configuration
To set the reference voltage, perform the following steps:
1.After selecting dB, the present reference voltage level is displayed:
REF: +0.000000
2.To change the reference level, use the and keys to select the numeric position.
Then use the ▲ and ▼ keys to select the desired value. Be sure to press ENTER after
changing the reference voltage.
NOTES The dB calculation takes the absolute value of the ratio V
The largest negative value of dB is -160dB. This will accommodate a ratio of
V
= 1µV and V
IN
REF
= 1000V.
/ V
IN
REF
F
C
2-32Basic Measurements
Measuring continuity
The Model 2016 uses the 1kΩ range to measure circuit continuity. After selecting continuity,
the unit prompts you for a threshold resistance level (1Ω-1000Ω). The Model 2016 alerts you
with a beep when a reading is below the set level.
To measure the continuity of a circuit, press SHIFT then CONT, set the threshold resistance
level and connect the circuit.
NOTEContinuity has a non-selectable reading rate of FAST (0.1 PLC).
Connections
Connect the circuit you want to test to the INPUT HI and INPUT LO terminals of the
Model 2016. The test current flows from the INPUT HI as shown in Figure 2-9.
You can define a threshold resistance from 1Ω to 1000Ω. The factory setting is 10Ω. Follow
these steps to define the resistance level:
1.Press SHIFT then CONT.
2.Use the and keys to choose a numerical place and use the ▲ and ▼ keys to
increment or decrement the digits. Enter a value from 1 to 1000.
3.Press ENTER to confirm your setting.
Testing diodes
With a Model 2016, you can measure the forward voltage drop of general-purpose diodes and
the zener voltage of zener diodes. To test diodes, press SHIFT then , set the test current
range, connect the diode, and take a reading from the display.
NOTEDiode test has a non-selectable reading rate of MEDium (1 PLC).
Connections
Connect the diode leads to the INPUT HI and INPUT LO terminals on the Model 2016. The
test current flows from the INPUT HI terminal as shown in Figure 2-10.
Basic Measurements2-33
Figure 2-10
Diode testing
Range
Model 2016
SENSE
INPUT
Ω 4 WIRE
PEAK
350V
INPUTS
F
FRONT/REAR
HI
1000V
!
PEAK
LO
500V
PEAK
R
3A 250V
AMPS
SHIFT
LOCAL
POWER
CH1REM
STEPCH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
SCAN
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD TRIG FAST MED SLOWAUTO ERR
MATH
DCV
HOLD
EX TRIG
TRIG
SAVE SETUP
SOURCE
MEAS
THD
REL FILT
dBm
dB
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
ACI
RECALL
CONT
Ω2 Ω4
TEST
GPIB
DIGITS RATE
THD
ACV
CAL
RELFILTER
RS232
MATH
REAR
4W
BUFFER
STAT
2016 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
EXIT ENTER
RANGE
AUTO
RANGE
Model 2016
SENSE
INPUT
Ω 4 WIRE
PEAK
350V
INPUTS
F
FRONT/REAR
HI
1000V
!
PEAK
LO
500V
PEAK
R
3A 250V
AMPS
SHIFT
LOCAL
POWER
STEPCH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
SCAN
CH1REM
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD TRIG FAST MED SLOWAUTO ERR
MATH
DCV
HOLD
EX TRIG
TRIG
SAVE SETUP
SOURCE
MEAS
THD
REL FILT
dBm
dB
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
ACI
RECALL
CONT
Ω2 Ω4
TEST
GPIB
DIGITS RATE
THD
ACV
CAL
RELFILTER
RS232
MATH
REAR
4W
BUFFER
STAT
2016 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
EXIT ENTER
RANGE
AUTO
RANGE
Note: Source current flows from the INPUT HI to INPUT LO terminals.
General-Purpose
Diode
Zener
Diode
You can set the test current range from the front panel. The choices are 1mA, 100µA, and
10µA. The factory test current setting is 1mA. To set the test current, do the following:
1.Press SHIFT then .
2.Use the ▲ and ▼ keys to scroll through the three test current selections.
The diode test measures voltages on the 3V range for the 1mA test current and the 10V range
for the 100µA and 10µA ranges. If a reading is more than 10V, the Model 2016 displays the
“OVERFLOW” status message.
2-34Basic Measurements
Measuring distortion
The Model 2016 can make distortion measurements from 0.002% to 100% (-94dB to 0dB).
The 2016 uses a digital signal processor (DSP) to perform a fast Fourier transform on the signal
applied to the front or rear voltage inputs. It then analyzes the levels of the harmonics present in
the signal to calculate THD, THD+noise, and SINAD. A direct digital synthesis module
included in the distortion circuitry provides a programmable sine source. The source has a
second output that can provide the inverse of the sine output (shifted 180˚), or output 0-5V logic
level pulses in phase with the main output. Distortion measurement types are explained below:
•THD — Total harmonic distortion is the default distortion measurement type. It is
expressed in percent or dB. The measurement is calculated as follows:
is magnitude of the ith harmonic and f is the magnitude of the fundamental
i
2
2
h
h
4
3
f
waveform.
•THD+n — This is total harmonic distortion plus noise. It is what conventional (analog)
THD meters display. A conventional THD meter has a notch filter that removes the
fundamental frequency from the signal, and measures THD based on what remains. This
includes all of the harmonics, but also includes any random noise in the signal. Since the
Model 2016 uses a DSP to perform a Fourier transform on the signal, noise can be
eliminated from the distortion measurement, thus providing a true reading. This
measurement is expressed in percent or dB, and is calculated as follows:
where h is the harmonic, n is the noise, and f is the magnitude of the fundamental waveform.
•SINAD — This is another way of expressing THD+noise. It is the RMS magnitude of
the signal divided by the RMS magnitude of the difference between the signal and the
fundamental. This measurement is expressed only in dB. The calculation is as follows:
where h is the harmonic, n is the noise, and f is the magnitude of the fundamental waveform.
The digital averaging filter (moving or repeating) can be used with distortion measurements.
The RATE key has no effect on distortion measurements, because there is no NPLC setting.
NOTEThe Rel key is not supported for distortion measurements.
Configuration
These configuration options are presented in the order commonly used to make a distortion
measurement. The SCPI commands are generic; actual syntax depends on the test programming
language used. Factory defaults are assumed.
Select the function
From the front panel, select the measurement function by pressing SHIFT then THD. For
remote operation, send the command:
:SENSe:FUNCtion 'DISTortion';select distortion measurement function
Set the distortion measurement type
1.Press THD-MEAS and the display shows “TYPE: THD”.
2.Use the right cursor to highlight the type selection, then use the ▲ and ▼ keys to select
THD, THD+N, or SINAD.
3.Press ENTER, then press EXIT.
For remote operation, send the command:
:SENSe:DISTortion:TYPE THD|THDN|SINAD;select THD, THD+n, or SINAD
Set the distortion frequency acquisition
Basic Measurements2-35
The Model 2016 must know the fundamental frequency of the input waveform. Even a
difference of a few hertz can cause large errors. The choices are AUTO, SET, and ACQUIRE.
AUTO acquires the frequency before each distortion reading. This slows down the reading rate,
but it is useful if the source frequency is unknown or unstable. ACQUIRE takes a frequency
measurement once and uses it for distortion measurements. SET allows you to select a frequency
between 20 and 20kHz (60Hz is default). AUTO or ACQUIRE are recommended instead of SET,
as the programmed frequency of the external source may not be accurate.
1.Press THD-MEAS.
2.Press ENTER once and the display shows “FREQ: AUTO”.
3.Use the cursor keys to select AUTO, SET, or ACQUIRE. (The source must be connected
and turned on for ACQUIRE.)
4.Press ENTER. If SET was selected, you are prompted for the frequency to be used. Use
the cursor keys and press ENTER to select it.
5.Press EXIT.
The following commands set the distortion frequency acquisition:
:SENSe:DISTortion:FREQuency:ACQuire;acquire the frequency once
:SENSe:DISTortion:FREQuency:AUTO ON|OFF ;turn AUTO on or off
:SENSe:DISTortion:FREQuency xxxxx.xxx;set frequency in hertz
2-36Basic Measurements
Set number of highest harmonic
This option sets the number (n) of the highest harmonic included in the distortion calculation,
where “n” is between 2 and 64 (2 is default). For example, the harmonic with a frequency twice
that of the fundamental is the second harmonic (n=2). This function acts as a brick-wall, lowpass filter. It has an upper limit of 50kHz or 64 times the fundamental frequency, whichever is
lower. This option has no effect in the THD+n or SINAD modes.
1.Press THD-MEAS.
2.Press ENTER until the display shows “UPPR HARM: 02”.
3.Set the number of the highest harmonic using the cursor keys.
4.Press ENTER, then press EXIT.
Use this remote command:
:SENSe:DISTortion:HARMonic xx;set number of highest harmonic
Set measurement units
For the THD and THD+n modes, the measurement units can be percent or dB. SINAD is
always displayed as dB.
1.Press THD-MEAS.
2.Press ENTER until the display shows “UNITS: PERC”.
3.Select PERC or dB using the cursor keys.
4.Press ENTER, then EXIT.
For remote operation, use this command:
:UNIT:DISTortion PERCent|DB;select percent or dB for units
Select shaping filter
The Model 2016 has digital shaping filters to simulate having the sample signal pass through
various types of telephone lines. The filter is turned off by default (NONE).
1.Press THD-MEAS.
2.Press ENTER until the display shows “SFIL: NONE”.
3.Select NONE, C (C message weighting), CCITT, CCIRARM, A (A weighting), or CCIR
using the cursor keys.
The Model 2016 defaults to autoranging for the voltage input range. If you prefer manual
ranging, use the lowest possible range for the signal level. Not using the appropriate range
causes inaccurate readings, or the display may show “OVRFLW” or “UDRFLW.”
Basic Measurements2-37
While the 2016 is measuring distortion, press the manual range keys. The display shows the
new range briefly. Press AUTO for autoranging. For remote operation, use these commands:
:SENSe:DISTortion:RANGe xxx.xx;set range based on parameter
:SENSe:DISTortion:RANGe:AUTO ON|OFF;turn autoranging on or off
Configure the internal function generator
The Model 2016 has a 10Hz-20kHz (60Hz default) two-channel function generator. It can be
used to evaluate amplifiers, filters or other devices, or can be connected to the 2016 inputs. The
function generator has selectable output impedances of 50Ω, 600Ω, and HIZ (50Ω default), and
is unbalanced (coaxial). Its amplitude is 0-2Vrms for 50Ω and 600Ω (0.5Vrms default), and
0-9.5Vrms for HIZ (1.0Vrms default). A second output provides an inverted sine (opposite in
phase to the main output), or a 0-5V logic level pulse in phase with the main output and having
the same frequency.
NOTEThe output impedance needs to be set before the amplitude. The impedance of the
second output reflects the impedance of the main output.
The only difference between 50Ω and HIZ is that the requested output voltage is corrected
for the actual load. For example, if 1.5V is requested:
1.With 50Ω selected and a 50Ω load, the result is 1.5V (correct).
2.With HIZ selected and a high impedance load, the result is 1.5V (correct).
3.With HIZ selected and a 50Ω load, the result is 0.75V (half of expected).
4.With 50Ω selected and a high impedance load, the result is 3.0V (twice expected).
5.With 50Ω selected and a 25Ω load, the result is 1.0V (incorrect for either 50Ω or high
impedance).
From the front panel:
1.Press THD-SOURCE and the display shows “SINE OUT: OFF”.
2.Use the cursor keys to select ON, then press ENTER.
3.When the display shows “FREQ: 00.0600k”, use the cursor keys to select the frequency
(.01 to 20k), then press ENTER.
4.When the display shows “IMPEDANCE: 50”, use the cursor keys to select 50, HIZ, or
600, then press ENTER.
5.When the display shows “AMPL: 0.5000V”, use the cursor keys to set the amplitude
(0 to 2V for 50Ω and 600Ω, 0 to 9.5V for HIZ), then press ENTER.
6.When the display shows “CHAN2: ISINE”, use the cursor keys to select ISINE (inverted
sine) or PULSE (square wave), then press ENTER.
For remote operation:
:OUTPut ON|OFF;turn output on or off
:OUTPut:FREQuency xxxxx.xxx;set frequency of source
:OUTPut:IMPedance OHM50|OHM600|HIZ;set output impedance
:OUTPut:AMPLitude x.xxx;set amplitude of source
:OUTPut:CHANnel2:SHAPe ISINE|PULSE;set Channel 2 waveform
2-38Basic Measurements
Retrieve magnitude of individual harmonic
(remote operation only)
The 2016 can return the levels of individual harmonics (relative to the level of the
fundamental, in dB). The parameters for this command are the starting and ending harmonics.
Specify 2,3 for the second and third harmonics, or 2,2 for the second harmonic. The harmonic
levels returned correspond to the last triggered reading, and the unit has to be set for one-shot
readings (:INIT:CONT OFF). For remote operation, send this command:
Retrieve RMS volts, THD+n, or THD for an acquired reading
(remote operation only)
Once a single reading has been triggered, the corresponding RMS volts value, THD+noise
value, or THD value can be read for the same set of data, regardless of what distortion mode is
set. SINAD can be calculated from the THD+noise reading. Note that the following commands
only work if the unit is set to trigger one reading at a time (:INIT:CONT OFF). Errors occur if
the unit is continuously updating.
:SENSe:DISTortion:RMS?;return the calculated RMS volts value for the last
:SENSe:DISTortion:THD?;return the THD+noise reading for the last triggered
:SENSe:DISTortion:THDN?;return the THD reading (number of harmonics depends
triggered reading
reading
on the last setting of harmonic number)
Querying the RMS volts value in the distortion mode with :sens:dist:rms? may yield a reading
slightly different from a reading in AC volts mode. This is because of the differences in how the
two modes make the measurement.
Basic Measurements2-39
Configuring and using the internal sweep
(remote operation only)
The Model 2016's internal source can be set to sweep up to 200 frequencies and then return
the distortion and/or RMS volts for each frequency. Sweep results can be returned using the
SREAL data format (fastest—IEEE754 single precision), DREAL data format (IEEE754
double precision), or in ASCII (default). The sweep uses the present distortion measurement
settings and must be allowed to complete before requesting the data (otherwise the data will be
incomplete). Sweep end can be detected by reading bit 3 of the Operation Event register
(:stat:oper?) which is set to 1 at the end of a sweep. Bit 3 of the Operation Event register can
also be used to trigger a service request when the sweep is completed.
The output sweep can only be performed in distortion mode. When performing a sweep, use
SET mode for the distortion frequency acquisitions mode for maximum speed (remote
command is :sens:dist:freq:auto off).
NOTEDo not place the unit in autorange mode while sweeping. A value of +812 (not
permitted in autorange) will be returned if this command is issued while in autorange
mode. Refer to Appendix B for a complete list of status and error messages.
:OUTPut:LIST:MODE LIST;Sets single frequency or sweep mode
:OUTPut:LIST <ampl1>, <freq1>, <ampl2>, <freq2>… ;Creates a sweep list (50 points
max)
:OUTPut:LIST:APPend: <ampl#x>, <freq#x> ;Adds additional points to the existing
sweep list (change trigger count to match
number of points)
:OUTPut:LIST:ELEMents DIST,AMPL;Selects data to be returned (distortion,
RMS volts, or both)
:OUTPut:LIST:DELay xxx.xxx;Sets the source delay time (the time
between setting the source and taking the
measurement) in seconds
:TRIGger:COUNt xx;Sets the number of measurements to make
during the sweep; must equal the number
of sweep points
:OUTPut ON;Output must be on before starting sweep
:INIT;Starts the sweep
:OUTPut:LIST:DATA?;Returns the list of sweep data in the format
<dist1>, <ampl1>, <dist2>,… (dependent
on data elements selected)
2-40Basic Measurements
Configuring high and low cutoff filters
(remote operation only)
The Model 2016 includes low and high cutoff filters used to limit the range of frequencies
used in distortion measurements. The filters can be set in the range of 20Hz to 50kHz. Use the
low cutoff to limit noise frequencies below the fundamental for THD+noise and SINAD
measurements (the low filter does not have an effect in THD mode). When setting the low cutoff,
set the value lower than the fundamental signal frequency. Similarly, the high cutoff filter limits
noise frequencies above the specified frequency for THD+noise and SINAD measurements. In
THD mode, the limiting frequency is equal to the lower of the high cutoff filter or the harmonic
value.
:SENSe:DISTortion:LCO xxxx;Sets the low cutoff frequency
:SENSe:DISTortion:LCO:STATe ON|OFF;Turns the low cutoff on or off
:SENSe:DISTortion:HCO xxxx;Sets the high cutoff frequency
:SENSe:DISTortion:HCO:STATe ON|OFF;Turns the high cutoff on or off
The digital averaging filter (moving or repeat) can be used for THD. The RATE key sets the
gate time (FAST = 0.01s, MEDium = 0.1s, and SLOW = 1.0s).
Auto or Acquire frequency modes are recommended instead of Set mode, as the programmed
frequency of an external source may not be accurate.
Querying the RMS volts value in the distortion mode (:sens:dist:rms?) may yield a reading
slightly different from a reading in AC volts modes due to a difference in the way each mode
makes measurements.
The output impedance needs to be set prior to setting the amplitude of starting the sweep. The
impedance of the second output reflects the impedance of the main output.
:LCO <NRf>;Sets the low cutoff frequency of the Bandpass filter.
:LCO?;Querys the low cutoff frequency of the Bandpass filter.
:HCO <NRf>;Sets the high cutoff frequency of the Bandpass filter.
:HCO?;Queries the high cutoff frequency of the Bandpass filter.
Connections
Distortion connections
Assuming factory default conditions, the basic procedure is as follows:
1.Connect test leads to the INPUT HI and LO terminals. Either the front or rear inputs can
NOTEIf the internal source is connected directly to the inputs, be sure the outer conductor
:STATe <b>;Sets LCO state on or off.
:STATe?;Queries LCO state.
:STATe <b>;Sets HCO state on or off.
:STATe?;Queries HCO state.
be used; place the INPUTS button in the appropriate position.
is connected to INPUT LO.
Fi
11
D
gure 2-
istortion
measurements
Basic Measurements2-41
2.Select the measurement function by pressing SHIFT then THD.
3.Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE ▲ and ▼ keys to select a
measurement range consistent with the expected voltage.
NOTEPressing AUTO to turn off autoranging for distortion readings displays a message
indicating the present range.
4.Connect test leads to the source as shown in Figure 2-11.
Model 2016
SENSE
INPUT
Ω 4 WIRE
350V
PEAK
INPUTS
F
FRONT/REAR
HI
!
LO
R
3A 250V
AMPS
7
V•Hz
1000V
PEAK
500V
PEAK
AC Voltage
Source
SHIFT
LOCAL
POWER
STEPCH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
CH1REM
SCAN
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD TRIG FAST MED SLOWAUTO ERR
MATH
THD
DCV
ACV
HOLD
EX TRIG
TRIG
SAVE SETUP
SOURCE
MEAS
THD
dBm
DCI
LIMITS ON/OFFDELAY
STORE
CONFIG HALT
STEP SCAN
RECALL
REL FILT
dB
CONT
ACI
Ω2 Ω4
TEST
GPIB
DIGITS RATE
CAL
RELFILTER
RS232
MATH
REAR
4W
BUFFER
STAT
2016 THD MULTIMETER
PERIOD TCOUPL
FREQ
TEMP
EXIT ENTER
RANGE
AUTO
RANGE
Input Impedance = 1M and <100pF
Caution: Maximum Input = 1000V peak, 8 × 10
CAUTIONDo not apply more than 1000V peak to the input or instrument damage
may occur. The voltage limit is subject to the 8×10
7
V•Hz product.
5.Observe the display. If the “UDRFLW %THD” or “UDRFLW dB” message is displayed
(2% of range), select a lower range until a normal reading is displayed (or press AUTO
for autoranging). Use the lowest possible range for the best resolution.
NOTEPressing the AUTO key to turn off autorange for distortion readings will display a
short message indicating the present range.
When the fundamental frequency is outside the 20Hz to 20kHz range, an underflow or
overflow condition exists. When a traping filter is enabled, the range can be much
narrower, depending on the type of filter enabled.
6.Take readings from the display.
NOTEWhen measuring distortion on the 100mV range or autoranging, open input leads
cause the distortion modes to display apparently valid readings. This is noise pickup
at the power line frequency. The amplitude of the signal depends on the type and
length of input connections. If the issue cannot be resolved by selecting a higher
range, place a resistor across the input leads creating an underflow condition when
the leads are open. The resistor value depends on the amount of noise present, but
typically is 50kΩ to 150kΩ.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Fi
12
F
2-42Basic Measurements
Function generator connections
Figure 2-12 shows typical connections for the function generator. In this example,
connections are made to the SOURCE OUTPUT BNC jack using a coax cable. Connections to
the INV/PULSE SOURCE OUTPUT jack are similar. See “Configure the internal function
generator” earlier in this section for details on setting up the function generator.
WARNINGMaximum voltage between SOURCE OUTPUT or INV/PULSE SOURCE
OUTPUT low (BNC jack shell) and chassis ground is 42V peak. Exceeding
this value will result in a shock hazard.
gure 2-
unction generator
connections
Measurement examples
Model 2016
Coax
Cable
350V
PEAK
SOURCE
OUTPUT
HI
1000V
TRIGGER
PEAK
!
500V
PEAK
LO
SENSE
INPUT
Ω 4W
42V PEAK
INV/PULSE
SOURCE
OUTPUT
35
1
46
2
!
!
LINE
FUSE
100 VAC
500 mAT
120 VAC
(SB)
220 VAC
250 mAT
240 VAC
(SB)
LINK
!
VMCEXT TRIG
LINE RATING
50, 60
25 VA MAX
MADE IN
U.S.A.
RS232
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
120
Warning: Maximum voltage between
SOURCE OUTPUT or INV/PULSE
SOURCE OUTPUT and chassis
ground is 42V peak.
Input
Device Under Test
Measuring distortion
NOTEThe Rel key is not supported for distortion measurements.
Turn on the Model 2016 and connect the function generator output (SOURCE OUTPUT) to
the front or rear voltage inputs (INPUT HI/LO). Set the INPUTS button accordingly.
1.Press SHIFT then ACV(THD) to put the 2016 into distortion measurement mode.
2.Press MEAS (the THD measure key).
3.When the display shows “TYPE” select THD and press ENTER.
4.When the display shows “FREQ:” select AUTO and press ENTER.
5.When the display shows “UPPR HARM:” select 02 and press ENTER.
6.When the display shows “UNITS:” select PERC and press ENTER.
7.When the display shows “SFIL:” select NONE and press ENTER.
8.Press SOURCE (the THD source key).
9.When the display shows “SINE OUT:” select ON and press ENTER.
10.When the display shows “FREQ:” select 01.0000kHz and press ENTER.
Basic Measurements2-43
11.When the display shows “IMPEDANCE: 50”, select HIZ and press ENTER. (Setting is
ignored in this example.)
12.When the display shows “AMPL:” select 1.0000V and press ENTER.
13.When the display shows “CHAN2:” select ISINE and press ENTER.
The display should read approximately 0.2 %THD.
To illustrate why the 2016 needs to know the exact source frequency, perform the following:
1.Press THD-MEAS and press ENTER once.
2.When the display shows “FREQ:”, select SET and press ENTER.
3.For the frequency, enter 01.0002kHz and press ENTER four times.
The display should read approximately 0.02 %THD and will be unstable. Note that the setting
for the function generator did not change.
Using the steps listed previously, return to the THD-MEAS menu and select ACQUIRE for
the frequency mode. The readings will return to approximately 0.02 %THD.
To perform this example remotely, send these commands:
*RST;start from defaults
:sens:func ‘dist’;select distortion function
:sens:dist:type thd;select THD type
:sens:dist:harm 2;set highest harmonic to 2
:unit:dist:perc;select percent distortion
:sens:dist:sfil none;no shaping filter
:sens:dist:rang:auto on;turn on autoranging
:outp:freq 1000;set frequency to 1kHz
:outp:imp HIZ;set high impedance source
:outp:ampl 1;set one volt
:outp:chan2 isine;select inverted sine
:outp on;turn on source
:read?;trigger one reading, the distortion measurement can be read from
the bus
:sens:dist:rms?;return the RMS volts value corresponding to the above
measurement
2-44Basic Measurements
Measuring AC volts or frequency
To measure AC volts or frequency using the Model 2016, follow this procedure. Note that the
function generator does not work outside the distortion function.
1.Select ACV or frequency.
Front panel:
• Press ACV or FREQ.
Remote:
:sens:func ‘volt:ac’;select ACV
:sens:func ‘freq’;select frequency
2.Set voltage range.
Front panel:
• Press RANGE ▲ and ▼. The display briefly shows the new range. Autorange is
Remote:
:sens:volt:ac:rang xxx.xxx;select ACV range based on number
:sens:volt:ac:rang:auto on|off;turn autorange on or off
:sens:freq:thr:volt:rang xxx.xxx;select frequency range
3.Set integration rate for ACV.
Front panel:
• Press RATE. An annunciator indicates FAST (0.1 NPLC), MED (1 NPLC), or SLOW
Remote:
:sens:volt:ac:nplc xx.xx;set number of power line cycles to integrate over
4.Set number of displayed digits.
Front panel:
• Press DIGITS to cycle through the number of digits.
Remote:
:sens:volt:ac:dig x;select number of display digits in ACV
:sens:freq:dig x;select number of display digits in frequency
available for ACV but not frequency.
(10 NPLC).
Basic Measurements2-45
5.Set digital filter for ACV.
The 2016 has a digital averaging filter to stabilize readings. The repeating filter takes the
set number of readings, averages, then updates the display. The moving filter throws out
the oldest reading, takes a new reading, then updates the display. The moving filter
updates the display faster, but takes longer for readings to stabilize.
Front panel:
• Press FILTER.
• When the display shows “010 RDGS”, select the number of readings to average (1 to
100) and press ENTER.
• When the display shows “TYPE:MOVING AV”, select MOVING AV or REPEAT
and press ENTER.
• The display will have the FILT annunciator lit to indicate the filter is on. Press
FILTER again to disable the filter.
Remote:
:sens:volt:ac:aver:tcon mov|rep;select moving or repeating filter for ACV
:sens:volt:ac:aver:coun xxx;set number of averaged readings
:sens:volt:ac:aver:stat on|off;turn filter on or off
2-46Basic Measurements
Distortion and RMS volts sweep example
This is an example of the bus commands which should be sent to the 2016 to configure1 start,
and receive the data for a 10-point sweep. This also includes the use of the high and low cutoff
filters.
*RST;Resets the 2016 to default conditions
*CLS;Clears the status registers
:STAT:OPER:ENAB 8;This will cause the Operation Summary Bit to set
*SRE 128;Enables the Operation Summary Bit mask to cause
:SENS:FUNC 'DIST';Selects Distortion mode
:SENS:DIST:RANG:AUTO OFF;Disables auto range
:SENS:DIST:FREQ:AUTO OFF;Turns off the AUTO frequency mode
:SENS:DIST:TYPE THDN;Selects THD+noise mode
:SENS:DIST:LCO 500;Configures the low cutoff to filter out noise below
:SENS:DIST:LCO:STAT ON;Turns on the low cutoff filter
:SENS:DIST:HCO 10000;Configures the high cutoff to filter out noise above
:SENS:DIST:HCO: STAT ON;Tums on the high cutoff filter
:OUTP:LIST:DEL .1;Sets a source delay of 0.1 seconds
:OUTP:LIST:ELEM DIST,AMPL;Selects distortion and amplitude as the data
:TRIG:COUN 10;The 2016 will take 10 triggered measurements
OUTP ON;Tums the output on
:INIT;Begins the sweep (Wait for SRQ to be asserted)*
:OUTP:LIST:DATA?;Queries the sweep data (Read data from the 2016)*
when the sweep has been completed, so that the
sweep end can be detected
the SRQ line to be asserted when the sweep is
completed
500 Hz
10kHz
;This will set 10 sweep points, from 1000 Hz to 1900
Hz in 100Hz steps, all at one volt amplitude
elements to be returned
The actual syntax for these steps depends on the type of GPIB interface and control software
used. For example, the WaitSRQ function can be used with a National Instruments interface to
wait for the SRQ, and the IBRD function can be used to read the data.
Running this same sweep again only requires sending *cls and init over the bus. Sending the
*cls again is necessary to reset the sweep done bit.
Measure-
ment Options
3
Measurement
Options
3-2Measurement Options
Introduction
This section describes the front panel features of the Model 2016. For those measurement
options accessible only by a remote interface, refer to Sections 4 and 5. This section is organized
as follows:
•Measurement configuration — Describes ranging, filtering, relative readings, digits of
resolution, and measurement rate.
•Trigger operations — Uses a trigger model to explain trigger modes and sources.
•Buffer operations — Discusses the reading storage buffer and buffer statistics.
•Limit operations — Defines how to set reading limits.
•Scan operations — Explains the external scanning capabilities.
•System operations — Gives details on setup saving and restoring, selecting a remote
interface, and accessing test and calibration.
Measurement configuration
The following paragraphs discuss configuring the multimeter for making measurements. See
the end of Appendix A for information about optimizing readings for speed or accuracy.
Range
The selected measurement range affects both the ultimate digits and accuracy of the measurements as well as the maximum signal that can be measured. The range setting (fixed or auto) for
each measurement function is saved when changing functions.
Maximum readings
The full scale readings for every range on each function are 20% overrange except for the
1000VDC, 750VAC, 3ADC, 3AAC, diode test, and distortion ranges.
Input values more than the maximum readings cause the “OVERFLOW” messages to be
displayed.
Manual ranging
To select a range, simply press the RANGE ▲ or ▼ key. The instrument changes one range
per keypress. The selected range is displayed for one second.
If the instrument displays the “OVERFLOW” message on a particular range, select a higher
range until an on-range reading is displayed. Use the lowest range possible without causing an
overflow to ensure best accuracy and resolution.
NOTE Temperature and continuity functions have just one range.
Filter
Measurement Options3-3
Autoranging
To enable autoranging, press the AUTO key. The AUTO annunciator turns on when
autoranging is selected. While autoranging is selected, the instrument automatically chooses the
best range to measure the applied signal. Autoranging should not be used when optimum speed
is required.
Note that up-ranging occurs at 120% of range, while down-ranging occurs at 10% of nominal
range, except for distortion measurements. Up-ranging for the distortion function is 106-112%
of range, while down-ranging is approximately 8.5% of range.
To cancel autoranging, press AUTO or the RANGE ▲ or ▼ key. Pressing AUTO to cancel
autoranging leaves the instrument on the present range.
The AUTO key has no effect on the temperature, continuity, and diode test functions.
FILTER lets you set the filter response to stabilize noisy measurements. The Model 2016 uses
a digital filter, which is based on reading conversions. The displayed, stored, or transmitted reading is simply an average of a number of reading conversions (from 1 to 100).
To select a filter:
1.Press FILTER once if the FILT annunciator is off; press twice if FILT is on.
2.Enter the number of readings.
3.Select the type of filter you want (moving average or repeating), then press ENTER.
The FILT annunciator turns on. When a filter is enabled, the selected filter configuration for
that measurement function is in effect.
Pressing FILTER once disables the filter.
NOTEThe filter can be set for any measurement function except frequency, period,
continuity, and diode test.
F
M
3-4Measurement Options
Filter types
The moving average filter (Figure 3-1) uses a first-in, first-out stack. When the stack becomes
full, the measurement conversions are averaged, yielding a reading. For each subsequent conversion placed into the stack, the oldest conversion is discarded, and the stack is re-averaged,
yielding a new reading.
For the repeating filter (Figure 3-1), the stack is filled and the conversions are averaged to
yield a reading. The stack is then cleared and the process starts over. Choose this filter for scanning so readings from other channels are not averaged with the present channel.
igure 3-1
oving average and
repeating filters
Response time
Conversion #10
#9
Conversion #1
Conversion #10
#9
Conversion #1
#8
#7
#6
#5
#4
#3
#2
A. Type - Moving Average, Readings = 10
#8
#7
#6
#5
#4
#3
#2
B. Type - Repeating, Readings = 10
Reading
#1
Reading
#1
Conversion #11
#10
Conversion #2
Conversion #20
#19
Conversion #11
#9
#8
#7
#6
#5
#4
#3
#18
#17
#16
#15
#14
#13
#12
Conversion #12
#11
#8
Reading
#2
Conversion #3
Reading
#2
#10
#9
#7
#6
#5
#4
Conversion #30
#29
#26
Conversion #21
#28
#27
#25
#24
#23
#22
Reading
#3
Reading
#3
The filter parameters have speed and accuracy tradeoffs for the time needed to display, store,
or output a filtered reading. These affect the number of reading conversions for speed versus
accuracy and response to input signal changes.
Relative
Measurement Options3-5
The rel (relative) function can be used to null offsets or subtract a baseline reading from
present and future readings. When rel is enabled, the instrument uses the present reading as a
relative value. Subsequent readings will be the difference between the actual input value and the
rel value.
NOTEThe rel function is not supported for distortion measurements.
You can define a rel value for each function. Once a rel value is established for a measurement
function, the value is the same for all ranges. For example, if 50V is set as a rel value on the
100V range, the rel is also 50V on the 1000V, 10V, 1V, and 100mV ranges.
Thus, when you perform a zero correction for DCV, Ω2, and Ω4 measurements by enabling
REL, the displayed offset becomes the reference value. Subtracting the offset from the actual
input zeroes the display, as follows:
Actual Input – Reference = Displayed Reading
A rel value can be as large as the highest range.
Selecting a range that cannot accommodate the rel value does not cause an overflow
condition, but it also does not increase the maximum allowable input for that range. For
example, on the 10V range, the Model 2016 still overflows for a 12V input.
Digits
To set a rel (relative) value, press REL key when the display shows the value you want as the
relative value. The REL annunciator turns on. Pressing REL a second time disables rel.
You can input a REL value manually using the mX+b function. Set M for 1 and B for any
value you want. Pressing REL enables that value to be the relative value. See Section 2 for more
information on the mX+b function.
The display resolution of a Model 2016 reading depends on the DIGITS setting. It has no
effect on the remote reading format. The number of displayed digits does not affect accuracy or
speed. Those parameters are controlled by the RATE setting.
Perform the following steps to set digits for a measurement function:
1.Press the desired function.
2.Press the DIGITS key until the desired number of digits is displayed (3½ to 6½).
NOTEFrequency and period can be displayed with four to seven digits.
3-6Measurement Options
Rate
The RATE operation sets the integration time of the A/D converter, the period of time the
input signal is measured (also known as aperture). The integration time affects the usable digits,
the amount of reading noise, as well as the ultimate reading rate of the instrument. The
integration time is specified in parameters based on a number of power line cycles (NPLC),
where 1 PLC for 60Hz is 16.67msec and 1 PLC for 50Hz and 400Hz is 20msec.
In general, the fastest integration time [FAST (0.1 PLC) from the front panel, 0.01 PLC from
the bus] results in increased reading noise and fewer usable digits, while the slowest integration
time (10 PLC) provides the best common-mode and normal-mode rejection. In-between settings
are a compromise between speed and noise.
The RATE parameters are explained as follows:
•FAST sets integration time to 0.1 PLC. Use FAST if speed is of primary importance (at
the expense of increased reading noise and fewer usable digits).
•MEDium sets integration time to 1 PLC. Use MEDium when a compromise between
noise performance and speed is acceptable.
•SLOW sets integration time to 10 PLC. SLOW provides better noise performance at the
expense of speed.
NOTEThe integration time can be set for any measurement function except frequency,
period, continuity (FAST), diode test (MEDium) and distortion. For frequency and
period, this value is gate time or aperture.
For the AC functions, MEDium and SLOW have no effect on the number of power line
cycles. See the discussion on “Bandwidth” that follows.
Measurement Options3-7
Bandwidth
The rate setting for AC voltage and current measurements determines the bandwidth setting:
•Slow — 3Hz to 300kHz.
•Medium — 30Hz to 300kHz.
•Fast — 300Hz to 300kHz.
Bandwidth is used to specify the lowest frequency of interest. When the Slow bandwidth
(3Hz to 300kHz) is chosen, the signal goes through an analog RMS converter. The output of the
RMS converter goes to a fast (1kHz) sampling A/D and the RMS value is calculated from 1200
digitized samples (1.2s).
When the Medium bandwidth (30Hz to 300kHz) is chosen, the same circuit is used. However,
only 120 samples (120ms) are needed for an accurate calculation because the analog RMS
converter has turned most of the signal to DC.
In the Fast bandwidth (300Hz to 300kHz), the output of the analog RMS converter (nearly
pure DC at these frequencies) is simply measured at 1 PLC (16.6ms).
Table 3-1 lists the rate settings for the various measurement functions. The FAST, MED, and
SLOW annunciators are only lit when conditions in the table are met. In other cases, the
annunciators are turned off.
Table 3-1
Rate settings for the measurement functions
Function
DCV, DCI
ACV, ACI
Ω2W, Ω4W
FREQ, PERIOD
dB, dBm (ACV)
dB, dBm (DCV)
Continuity
Diode test
Distortion
Notes:
NPLC = number of power line cycles.
BW = lower limit of bandwidth (in Hz).
APER = aperture in seconds.
N/A = not available.
X = setting ignored.
The following paragraphs discuss front panel triggering, the programmable trigger delay, the
reading hold feature, and external triggering.
Trigger model
The flowchart of Figure 3-2 summarizes triggering as viewed from the front panel. It is called
a trigger model because it is modeled after the SCPI commands used to control triggering. Note
that for stepping and scanning, the trigger model has additional control blocks. These are
described in “Scan operations” later in this section.
igure 3-2
Idle
ront panel triggering
without stepping/scanning
Control
Source
Immediate
External
Event
Detection
Delay
Device
Action
Output
Trigger
Idle
The instrument is considered to be in the idle state whenever it is not performing any
measurements or scanning functions. From the front panel, the unit is considered idle at the end
of a step or scan operation when the reading for the last channel remains displayed. To restore
triggers, use the SHIFT-HALT keys.
Once the Model 2016 is taken out of idle, operation proceeds through the flowchart.
Control source and event detection
The control source holds up operation until the programmed event occurs and is detected. The
control sources are described as follows:
•Immediate — With this control source, event detection is immediately satisfied allowing
operation to continue.
•External — Event detection is satisfied for any of three conditions:
• An input trigger via the Trigger Link line EXT TRIG is received.
• A bus trigger (GET or *TRG) is received.
• The front panel TRIG key is pressed. (The Model 2016 must be taken out of remote
before it will respond to the TRIG key. Use the LOCAL key or send LOCAL 716 over
the bus.)
Measurement Options3-9
Delay
A programmable delay is available after event detection. It can be set manually or an auto
delay can be used. With auto delay, the Model 2016 selects a delay based on the function and
range. The AUTO settings are listed in Table 3-2.
Table 3-2
Auto delay settings
FunctionRange and delay
DCV
ACV, Distortion
FREQ
DCI
ACI
Ω2W, Ω4W
Continuity
Diode testing
The delay function is accessed by pressing the SHIFT-DELAY keys. The present delay
setting (AUTO or MANual) is displayed. Use the ▲ and ▼ keys to select the type of delay. If
MANual is chosen, also enter the duration of the delay. The maximum is shown following:
Press ENTER to accept the delay or EXIT for no change.
100mV
1ms
100mV
400ms
100mV
1ms
10mA
2ms
100Ω
3ms
1V
1ms
1V
400ms
1V
1ms
100mA
2ms
1kΩ
3ms
1kΩ
3ms
1mA
1ms
10V
1ms
10V
400ms
10V
1ms
1A
2ms
1A
400ms
10kΩ
13ms
100µA
1ms
99H:99M:99.999S
100V
5ms
100V
400ms
100V
1ms
3A
2ms
3A
400ms
100kΩ
25ms
10µA
1ms
1000V
5ms
750V
400ms
750V
1ms
1MΩ
100ms
10MΩ
150ms
100MΩ
250ms
Changing the delay to MANual on one function changes the delays on all functions to MANual.
3-10Measurement Options
Device actions
The primary device action is a measurement. However, the device action block could include
the following additional actions:
•Source and Delay (while in LIST mode) — If MODE is set to LIST (rather than FIXED),
readings will be taken for each listed amplitude and frequency pair, up to the maximum
number of pairs allowed (see :OUTPut:LIST command in Section 5). This sweep
functionality is tied into Device Action so all trigger model features such as
SAMP:COUN, TRIG:COUN, and also the different trigger sources are functional while
in LIST mode.
•Filtering — If the repeating filter is enabled, the instrument samples the specified
number of reading conversions to yield single filtered reading. Only one reading
conversion is performed if the filter is disabled, or after the specified number of reading
conversions for a moving average filter is reached. The output of filter feeds hold.
•Hold — With hold enabled, the first processed reading becomes the “seed” reading and
operation loops back within the device action block. After the next reading is processed,
it is checked to see if it is within the selected window (0.01%, 0.1%, 1%, 10%) of the
“seed” reading. If the reading is within the window, operation again loops back within
the device action block. This looping continues until the specified number (2 to 100)
consecutive readings are within the window. If one of the readings is not within the
window, the instrument acquires a new “seed” reading and the hold process continues.
•Channel closure — When stepping or scanning, the last device action is to open the
previous channel (if closed) and close the next channel. Using the hold feature provides
an auto settling time for the scanner relays. Each open/close transition will restart the
hold process and a reading for each channel will not occur until the relay settles.
Output trigger
After the device action, an output trigger occurs and is available at the rear panel Trigger Link
connector. This trigger can be used to trigger another instrument to perform an operation (e.g.,
select the next channel for an external scan).
Counters
The trigger model for stepping and scanning contains additional blocks for counting samples
(the number of channels to scan) and counting triggers. These counters are explained in the
paragraph “Scan operations” later in this section.
Reading hold (autosettle)
F
R
When a hold reading is acquired as described in “Device actions”, an audible beep is sounded
(if enabled) and the reading is considered a “true measurement”. The reading is held on the
display until an “out of window” reading occurs to restart the hold process.
When operating remotely or scanning, the hold process seeks a new “seed” once it has been
satisfied and the reading has been released. When operating from the front panel, the hold
process does not seek a new “seed” until the held condition is removed.
Hold example
1.Enable HOLD, select a window percentage and enter a count.
2.Apply test probes to a signal. Once the signal becomes stable enough to satisfy the hold
condition, the reading is released, and the beeper sounds (if enabled).
3.Remove the hold condition by lifting the probes. Hold will then seek a new “seed”.
External triggering
The EXT TRIG key selects triggering from two external sources: trigger link and the TRIG
key. When EXT TRIG is pressed, the TRIG annunciator lights and dashes are displayed to
indicate that instrument is waiting for an external trigger. From the front panel, you can press
the TRIG key to trigger a single reading. Pressing the EXT TRIG key again toggles you back to
continuous triggers.
Measurement Options3-11
The Model 2016 uses two lines of the Trigger Link rear panel connector as External Trigger
(EXT TRIG) input and Voltmeter Complete (VMC) output. The EXT TRIG line allows the
Model 2016 to be triggered by other instruments. The VMC line allows the Model 2016 to
trigger other instruments.
At the factory, line 1 is configured as VMC and line 2 as EXT TRIG. (Changing this
configuration is described in the Model 2016 Service Manual.) A connector pinout is shown in
Figure 3-3.
igure 3-3
ear panel pinout
6
8
7
5
#2
EXTERNAL
TRIGGER
INPUT
34
2
1
#1
VOLT METER
COMPLETE
OUTPUT
Pin Number Description
1 Voltmeter Complete Output
2 External Trigger Input
3 no connection*
4 no connection*
5 no connection*
6 no connection*
7 Signal Ground
8 Signal Ground
* Either pin 3 or 5 may be configured as an output
instead of pin 1. Either pin 4 or 6 may be
configured as an inputinstead of pin 2. See the
Model 2016 Service Manual for details.
F
p
F
3-12Measurement Options
External trigger
The EXT TRIG input requires a falling-edge, TTL-compatible pulse with the specifications
shown in Figure 3-4. In general, external triggers can be used to control measure operations. For
the Model 2016 to respond to external triggers, the trigger model must be configured for it.
igure 3-4
Trigger link input
ulse specifications
(EXT TRIG)
Voltmeter complete
The VMC output provides a TTL-compatible output pulse that can be used to trigger other
instruments. The specifications for this trigger pulse are shown in Figure 3-5. Typically, you
would want the Model 2016 to output a trigger after the settling time of each measurement.
igure 3-5
Trigger link output pulse
specifications (VMC)
TTL High
(2V-5V)
TTL Low
(≤0.8V)
TTL High
(3.4V Typical)
Triggers on
Leading Edge
2µs
Minimum
Meter
Complete
TTL Low
(0.25V Typical)
External triggering example
In a typical test system, you may want to close a channel and then measure the DUT
connected to the channel with a multimeter. Such a test system is shown in Figure 3-6, which
uses a Model 2016 to measure ten DUTs switched by a Model 7011 multiplexer card in a
10µs
Minimum
Model 7001/7002 Switch System.
Fi
gure 3-
6
D
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPEAND RATING.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
The Trigger Link connections for this test system are shown in Figure 3-7. Trigger Link of
the Model 2016 is connected to Trigger Link (either IN or OUT) of the Model 7001/7002. Note
that with the default trigger settings on the Model 7001/7002, line #1 is an input and line #2 is
an output. This complements the trigger lines on the Model 2016.
HI
1000V
PEAK
!
500V
PEAK
LO
INPUT
!
INV/PULSE
SOURCE
FUSE
OUTPUT
500 mAT
(SB)
250 mAT
(SB)
2016 Multimeter
MADE IN
U.S.A.
TRIGGER
LINK
RS232
!
35
1
VMC
46
2
EXT TRIG
!
LINE
LINE RATING
100 VAC
50, 60
120 VAC
25 VA MAX
220 VAC240 VAC
Trigger
Link
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
120
7001 or 7002 Switch System
MADE IN USA
IN
OUT
Trigger
Link
350V
PEAK
SENSE
SOURCE
OUTPUT
Trigger
Link Cable
(8501)
Ω 4W
42V PEAK
For this example, the Model 2016 and 7001/7002 are configured as follows:
Model 2016:
Factory defaults restored (accessed from SHIFT-SETUP)
External scanning, channels 1 - 10, no timer, 10 readings (accessed from SHIFT-CONFIG)
External triggers (accessed from EXT TRIG)
Model 7001 or 7002:
Factory defaults restored
Scan list = 1!1-1!10,
Number of scans = 1
Channel spacing = TrigLink
To run the test and store readings in the Model 2016 with the unit set for external triggers,
press STEP or SCAN. The Model 2016 waits (with the asterisk annunciator lit) for an external
trigger from the Model 7001/7002.
F
f
3-14Measurement Options
Press STEP on the Model 7001/7002 to take it out of idle and start the scan. The scanner’s
output pulse triggers the Model 2016 to take a reading, store it, and send a trigger pulse. The
following explanation on operation is referenced to the operation model shown in Figure 3-8.
igure 3-8
Operation model
or triggering
example
7001or 7002
Press STEP to start scan
Idle
Bypass
Wait for
Trigger Link
Trigger
2016
Arm
B
Trigger Link
A
Wait for
Trigger
No
C
D
Scan
Channel
Output
Trigger
Scanned
10
Channels
?
Trigger
Trigger
Make
Measurement
Output
Trigger
Made
10
Measurements
?
E
F
No
Yes
Yes
Measurement Options3-15
A
Pressing EXT TRIG then STEP or SCAN on the multimeter places it at point A in the
flowchart, where it is waiting for an external trigger.
B
Pressing STEP takes the Model 7001/7002 out of the idle state and places operation at
point B in the flowchart.
C
For the first pass through the model, the scanner does not wait at point B for a trigger.
Instead, it closes the first channel.
D
After the relay settles, the Model 7001/7002 outputs a Channel Ready pulse. Since the
instrument is programmed to scan ten channels, operation loops back up to point B, where it
waits for an input trigger.
EF
and Remember that the Model 2016 operation is at point A waiting for a trigger.
The output Channel Ready pulse from the Model 7001/7002 triggers the multimeter to measure
DUT #1 (point E). After the measurement is complete, the Model 2016 outputs a completion
pulse (point F) and then loops back to point A, where it waits for another input trigger.
The trigger applied to the Model 7001/7002 from the Model 2016 closes the next channel in
the scan. This triggers the multimeter to measure the next DUT. The process continues until all
ten channels are scanned and measured.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
F
D
3-16Measurement Options
External triggering with BNC connections
An adapter cable is available to connect the micro-DIN Trigger Link of the Model 2016 to
instruments with BNC trigger connections. The Model 8503 DIN to BNC Trigger Cable has a
micro-DIN connector at one end and two BNC connectors at the other end. The BNC cables are
labeled VMC (trigger line 1) and EXT TRIG (trigger line 2).
Figure 3-9 shows how a Keithley Model 706 Scanner can be connected to the Trigger Link
of the Model 2016 using the adapter cable. With this adapter, a Model 706 could be substituted
for the Model 7001/7002 in the previous example. With the Model 706 set for External
Triggering, the test would start when the single scan mode is selected and initiated.
If the Model 2016 trigger line configuration has been changed from the factory setting, the
Model 8502 Trigger Link Adapter must be used to interface with instruments having BNC
trigger connections. It has two micro-DIN connectors and six BNC connectors, one for each
trigger line.
igure 3-9
IN to BNC trigger cable
Model 8503 DIN to BNC Trigger Cable
350V
PEAK
SOURCE
OUTPUT
HI
1000V
TRIGGER
PEAK
!
500V
PEAK
LO
SENSE
INPUT
Ω 4W
42V PEAK
INV/PULSE
SOURCE
OUTPUT
35
1
46
2
!
!
FUSELINE
100 VAC
500 mAT
120 VAC
(SB)
220 VAC
250 mAT
240 VAC
(SB)
Model 2016
LINK
!
VMC
EXT TRIG
LINE RATING
50, 60
25 VA MAX
MADE IN
U.S.A.
RS232
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
Channel
Ready
External
120
Trigger
706 Scanner
Buffer operations
The Model 2016 has a buffer to store from two to 1024 readings and units. It also stores the
channel number for scanned readings and overflow readings. In addition, recalled data includes
statistical information, such as minimum, maximum, average, and standard deviation.
The buffer fills with the requested number of readings and stops. Readings are placed in the
buffer after any math operations are performed. Buffered data is overwritten each time the
storage operation is selected. The data is volatile; it is not saved through a power cycle.
The following paragraphs discuss storing and recalling buffered data.
Storing readings
F
B
Use the following procedure to store readings:
1.Set up the instrument for the desired configuration.
2.Press the STORE key.
3.Using the , , ▲, and ▼ keys, select the number of readings desired.
4.Press ENTER. The asterisk (*) annunciator turns on to indicate a data storage operation.
It will turn off when the storage is finished.
Recalling readings
Use the following steps to view stored readings and buffer statistics:
1.Press RECALL. The BUFFER annunciator indicates that stored readings are being
displayed. The arrow annunciator indicates that more data can be viewed with the ,
, ▲, and ▼ keys.
2.As shown in Figure 3-10, use the cursor keys to navigate through the reading numbers,
reading values, and statistics. For any of the buffer statistics (maximum, minimum,
average, standard deviation), the STAT annunciator is on.
3.Use the EXIT key to return to the normal display.
Measurement Options3-17
igure 3-10
uffer locations
RANGE
RANGE
RDGNO.10Reading Value
RDGNO.9Reading Value
RDGNO.8Reading Value
RDGNO.7Reading Value
RDGNO.6Reading Value
RDGNO.5Reading Value
RDGNO.4Reading Value
RDGNO.3Reading Value
RDGNO.2Reading Value
RDGNO.1Reading Value
STDDEVStandard Deviation Value
AverageAverage Value
MinAtXXMinimum Value
MaxAtXXMaximum Value
3-18Measurement Options
Buffer statistics
The MAX AT and MIN AT values are the maximum and minimum values in the buffer. The
AVERAGE value is the mean of the buffered readings. The equation used to calculate the mean
is:
where: xi is a stored reading
The STD DEV value is the standard deviation of the buffered readings. The equation used to
calculate the standard deviation is:
n
X
∑
i
i1=
-----------------=
y
n
n is the number of stored readings
where: xi is a stored reading
n is the number of stored readings
NOTEThe Model 2016 uses IEEE-754 floating point format for math calculations.
Limit operations
Limit operations set and control the values that determine the HI / IN / LO status of
subsequent measurements. Limits can be applied to all measurement functions except
continuity. The limit test is performed after mX+b and percent math operations. Unit prefixes
are applied before the limit test, for example:
•Low limit = -1.0, High limit = 1.0
A 150mV reading equals 0.15V (IN).
•Low limit = -1.0, High limit = 1.0
A 0.6kΩ reading equals 600Ω (HI).
You can configure the multimeter to beep or not when readings are inside or outside of the
Use the following steps to enter high and low limit values:
1.Press the SHIFT-LIMITS keys to view the present HI1 limit value:
Measurement Options3-19
HI1:+1.000000
This value represents the absolute value of that function.
2.Use the or keys to move to the number field. Use the , , ▲, and ▼ keys to
enter the desired value. Move the cursor to the rightmost position (^) and use the ▲ and ▼ keys to move the decimal point.
3.Press ENTER to view the present LO1 limit value:
LO1:-1.000000
This value represents the absolute value of that function.
4.Enter the desired value for this low limit.
5.Press ENTER to view the present HI2 limits value:
HI2: +2.000000^
This value represents the absolute value of that function.
6.Enter the desired value for this high limit.
7.Press ENTER to view the present LO2 limit value:
LO2: -2.000000^
This value represents the absolute value of that function.
8.Enter the desired value for the low limit. Pressing ENTER returns to the normal display.
^
^
F
3-20Measurement Options
Enabling limits
Use the following procedure to turn on the limits operation:
1.Press the SHIFT-ON/OFF keys to view the present beeper status:
BEEP: NEVER
2.Use the ▲ and ▼ keys to change the beeper status (NEVER, OUTSIDE, INSIDE). Press
ENTER when done.
When the multimeter returns to the normal display, the HI/IN/LO status is displayed along
with the reading. To disable limit testing, press SHIFT-ON/OFF again. An example of using
limits to sort resistors is shown in Figure 3-11.
igure 3-11
Using limit test to sort
100Ω, 10% resistors
subsystem. From the front panel, the same menu is used to control the beeping state and
conditions (inside or outside) for both limits. Since there is only one beeper, there are two
distinct tones used for the two limits, but limit set 1 will take precedence.
Set the beeper to beep inside. Then, apply 0.9 volts. The beep will be higher in pitch. When the
voltage is increased past 1V, the input is no longer inside limit set 1 but is still inside limit set 2.
At that point, the tone of the beep will change, indicating that you are still inside limit set 2.
NOTELimit 1 takes priority over Limit 2 when beeper is set to outside. No change in tone
LOINHI
90Ω
LO Limit
110Ω
HI Limit
The CALC3:LIMit2 subsystem has all the same commands available as the CALC3:LIMit[1]
Example: Power up with default limits (HLIM1 = +1, LLIM1 = -1,HLIM2 = +2, LLIM2 = -2).
will be detected.
Scan operations
The Model 2016 can be used with external scanner card installed in switching mainframes
such as the Models 707, 7001, and 7002. The following paragraphs discuss various aspects of
using scanning with the Model 2016.
Scanning overview
A scanner lets you switch among a number of input signals to the Model 2016 for measurement.
Refer to the documentation supplied with the scanner card for specific connection information.
When using external channels, the switching mainframe controls the opening and closing of
individual channels. To synchronize Model 2016 measurements with external channel closures,
connect the Trigger Link lines of the multimeter and switching mainframe. Refer to “Trigger
operations” earlier in this section for details and an example on using external triggering.
Front panel scanner controls
In addition to the trigger keys discussed previously, front panel keys that affect scanner card
operation include:
Measurement Options3-21
•SHIFT-CONFIG — Selects scan list, time between scans, and reading count.
•STEP — Starts a stepping operation of consecutive channels, where output triggers are
sent after every channel closure.
•SCAN — Starts a scanning operation of consecutive channels, where an output trigger
is sent at the end of the scan list.
•SHIFT-HALT — Stops stepping or scanning and restores the trigger model to a nonscanning mode.
Stepping and scanning trigger model additions
The trigger model presented in “Trigger operations” earlier in this section has some addi-
tional capabilities when stepping or scanning. These are outlined below:
•Timer — With this control source, event detection is immediately satisfied on the initial
pass. Each subsequent detection is satisfied when the programmed timer interval (up to
99H:99M:99.99S) elapses.
•Reading counter — For both stepping and scanning, the reading count can be entered
from SHIFT-CONFIG. (This is referred to as the trigger counter over the bus.) The reading counter can bypass the idle state. Operation will wait until the programmed control
source event occurs.
•Channel counter — For scanning, the scan list length (maximum channel less minimum
channel) is used to bypass the control source allowing a specified number of device
actions to occur. (This counter is referred to as the sample counter over the bus.)
F
F
Fi
13
F
3-22Measurement Options
These additional blocks are shown in the trigger models of Figure 3-12 and Figure 3-13. Uses
of the timer control source, reading counter, and channel counter are shown in the scanning
examples later in this section.
igure 3-12
ront panel triggering
with stepping
gure 3-
ront panel triggering
with scanning
Control
Source
Immediate
External
Timer
Control
Source
Immediate
External
Timer
Idle
Event
Detection
Delay
Idle
Event
Detection
Device
Action
Yes
Yes
Yes
No
More
Readings
?
Output
Trigger
No
More
Readings
?
Output
Trigger
No
More
Channels
?
Reading
Count
(Trigger counter)
Reading
Count
(Trigger counter)
Scan List
Length
(Sample counter)
Delay
Device
Action
Measurement Options3-23
Using SHIFT-CONFIG to configure stepping and scanning
From the SHIFT-CONFIG key combination, the minimum and maximum channels in the
scan list, the time between scans, and the reading count.
1.To configure stepping or scanning, perform the following:
2.Select the desired measurement function.
3.Press the SHIFT-CONFIG keys to access the step/scan configuration.
4.Select the first channel in the scan list (MINimum CHANnel) by using the , , ▲,
and ▼ keys and pressing ENTER.
5.Select the last channel in the scan list (MAXimum CHANnel) and press ENTER to
confirm.
6.The next selection is for timed scans. (This is the Timer control source in the trigger
model.) It sets a user-specified interval for starting scans. If you choose timed scans, the
Model 2016 prompts for a time interval:
00H:00M:00.000S
Use the , , ▲, and ▼ keys to select a time interval and press ENTER to confirm.
7.Next, you are prompted for a reading count (RDG CNT). This can be less than, equal to,
or greater than the scan list length (up to 1024). It is the number of readings that will be
stored in the buffer. The effects of these choices are further described in the scanning
examples.
8.Press ENTER when done to return to the normal display. Note that scanned readings are
always stored in the buffer, up to the setting for RDG CNT.
3-24Measurement Options
Scanning examples
The following example demonstrates the use of external scanning.
External scanning
The example of Figure 3-14 shows the front panel operations to configure an external scan.
The trigger and signal connections were shown previously in “Trigger operations”. Both
instrument setups assume factory defaults. Set the Model 2016 for the desired measurement
function.
1
On the Model 7001 Switch System, enter a scan list of channels 1 to 10 on card 1.
2
Also on the Model 7001, configure the instrument for Trigger Link triggers and one
scan of ten channels.
3
On the Model 2016 Multimeter, configure an external scan of the first ten channels.
4
Set the Model 2016 for external triggers by pressing EXT TRIG. The display will be
dashes.
5
Press STEP or SCAN on the Model 2016. The asterisk and STEP or SCAN
annunciator will light.
6
Press STEP on the Model 7001 to start channel closures.
7
After the scan, you can recall ten readings from the Model 2016 buffer.
NOTEWhen using an external thermocouple scanner card and channel 1 as a reference, the
Model 2016 recognizes only channel 1 when a step or scan is performed. If using a
Model 7001 or 7002 to close channel 1 manually, the Model 2016 will not interpret
that channel as the reference junction without a step or scan operation.
Fi
14
E
M
Measurement Options3-25
gure 3-
xternal scanning
example with
odel 7001
1
CONFIGURE SCAN
2
Model 7001
(from “reset setup”)
SCAN CHANNELS 1!1-1!10
CHAN-CONTROL
CHANNEL-SPACING
TRIGLINK
ASYNCHRONOUS
CHAN-COUNT
10
SCAN-CONTROL
SCAN-COUNT
1
3
Model 2016
(from “factory setup”)
SHIFT-CONFIG
TYPE:EXT
MIN CHAN: 001
MAX CHAN: 010
TIMER? OFF
RDG CNT: 0010
ENTER
EXT TRIG
4
5
STEP or SCAN
6
STEP
RECALL (10 readings)
7
▲
, , ▲ , ▼
▲
EXIT
3-26Measurement Options
System operations
The Model 2016 has other front panel operations. Saving and restoring setup information is
described in Section 2 — Basic Measurements. Selecting the remote interface and language is
covered in Section 4 — Remote Operation.
Calibration
The CAL selections are used to view the calibration date and next due date, to perform
calibration, and to view the number of times calibration has been performed. Some of the items
are password-protected to prevent unintended changing of calibration constants.
To view the calibration dates, press SHIFT-CAL. Press ENTER at the DATES prompt. The
first date is the last time calibration was performed. The NDUE date is the calibration due date.
Running calibration is password-protected. Refer to the Model 2016 Service Manual for
details.
To view the calibration count, press ENTER at the COUNT prompt.
4
Remote
Operation
4
Remote
Operation
4-2Remote Operation
Introduction
This section includes the following information:
•Selecting an interface
•RS-232 operation
•GPIB bus operation and reference
•Status structure
•Trigger model (GPIB operation)
•Programming syntax
•Common commands
Selecting an interface
The Model 2016 THD Multimeter supports two built-in remote interfaces:
•GPIB bus
•RS-232 interface
You can use only one interface at a time. The factory interface selection is the GPIB bus. You
can select the interface only from the front panel. The interface selection is stored in non-volatile
memory; it does not change when power has been off or after a remote interface reset.
NOTEChanging the interface (GPIB to RS-232 or RS-232 to GPIB) will clear the data
buffer.
RS-232
Remote Operation4-3
You can connect a controller to the RS-232 interface. Some considerations for selecting the
RS-232 interface are the following:
•You must define the baud rate, enable or disable software handshake XON/XOF.
•You can only use the SCPI programming language with the RS-232 interface.
To select RS-232 as the remote interface, do the following:
1.Access the RS-232 configuration by pressing SHIFT then RS232.
You see:
2.Move to the on/off selection by pressing the key.
You see
3.Turn on the RS-232 interface by toggling the selection to ON using the ▼ or ▲ key and
press ENTER.
You can exit the configuration menu by pressing EXIT.
For more information about the RS-232 interface, see section RS-232 operation.
RS232: OFF
OFF selection blinking.
GPIB bus
THD Multimeter. The address is displayed when the multimeter is turned on. At the factory, the
address is set to 16.
the GPIB interface if you have been previously using the RS-232 remote programming interface:
interface.
The GPIB bus is the IEEE-488 interface. You must select a unique address for the Model 2016
Since GPIB is the interface selection defined by the factory, only follow these steps to select
1.Select the GPIB option by pressing SHIFT then GPIB.
You see: GPIB: OFF
2.Move to the on/off selection by pressing the key.
You see
3.Turn on the GPIB interface by toggling the selection to ON using the ▼ or ▲ key and
press ENTER.
Turning off the RS-232 interface automatically selects GPIB as the remote programming
OFF selection blinking slowly.
4-4Remote Operation
RS-232 operation
Sending and receiving data
The RS-232 interface transfers data using 8 data bits, 1 stop bit, and no parity. Make sure the
controller you connect to the multimeter also uses these settings.
You can break data transmissions by sending a ^C or ^X character string to the multimeter.
This clears any pending operation and discards any pending output.
Selecting baud rate
The baud rate is the rate at which the Model 2016 THD Multimeter and the programming
terminal communicate. Choose one these available rates:
•19.2k
•9600
•4800
•2400
•1200
•600
•300
The factory selected baud rate is 4800.
When you choose a baud rate, make sure that the programming terminal that you are
connecting to the Model 2016 THD Multimeter can support the baud rate you selected. Both the
multimeter and the other device must be configured for the same baud rate. To select a baud rate,
follow these steps:
1.Access the RS-232 configuration by pressing SHIFT then RS232.
You see: RS232: ON (assuming you have already selected the RS-232 interface).
2.Go to the baud rate field by pressing the ▼ key.
You see
3.Access the baud rate list by pressing the key. You see the rate selection blinking.
4.Scroll through the available rates by pressing the ▼ and ▲ key until you find the rate you
want.
5.Confirm your selection by pressing ENTER. The multimeter prompts you to define
signal handshaking. Continue for information about handshaking. You can return to
measurement mode by pressing EXIT.
BAUD:<rate>.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.