Tektronix 2000 Reference manual

www.keithley.com
Model 2000 Digital Multimeter
Quick Start Guide
2000-903-01 Rev. C / October 2007
A GREATER MEASURE OF CONFIDENCE
Model 2000 Digital Multimeter
Quick Start Guide
©2007, Keithley Instruments, Inc.
Document Number:
Cleveland, Ohio, U.S.A.
2000-903-01 Rev. C / October 2007
Manual Print History Model 2000 Digital Multimeter Quick Start Guide

Manual Print History

The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision Level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the man ual. Addenda are numbered sequ entially. When a new Revision is created, all Addenda associated with the previous Revision of the manual are incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page.
Revision A (Document Number 2000-903-01)..........................................................April 1995
Revision B (Document Number 2000-903-01)..................................................Feburary 2000
Revision C (Document Number 2000-903-01) ....................................................October 2007
All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective holders.
2000-903-01 Rev. C / October 2007
The following safety precautions should be observed before using this product an d any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired. The types of product users are: Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equipment is
operated within its specifications and operating limits, and for ensuring that operators are adequately trained. Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instrument.
They must be protected from electric shock and contact with hazardous live circuits. Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or
replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.

Safety Precautions

Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained service
personnel may perform installation and service procedures. Keithley Instruments products are designed for use with electrical signals that are rate d Measurement Category I and Measurement
Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data I/O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with high transient over-voltages. Measurement Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
08/07
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power
!
disconnect device must be provided in close proximity to the equipment and within easy reach of the operator. For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test.
AL W AYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions, or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections. If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a lid
interlock. If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation. The symbol on an instrument indicates that the user should refer to the operating instructions located in the user documentation. The symbol on an instrument shows that it can source or measure 1000V or more, including the combined effect of normal and
common mode voltages. Use standard safety precautions to avoid personal contact with these voltages. The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns. The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and all test cables. T o maintain protection from electric shock and fire, replacement components in mains circuits - including the power transformer, test leads,
and input jacks - must be purchased from Keithley Instruments. Standard fuses with applicable national safety approvals may be used if the rating and type are the same. Other components that are not safety-related may be purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.

Table of Contents

Section T opic Page
1 Introduction................................................................................................. 1-2
Operation Summary.................................................................................... 1-2
Power-on defaults................................................................................ 1-2
Remote Operation Summary ............................................................. 1-13
Table of Contents Model 2000 Digital Multimeter Quick Start Guide
This page left blank intentionally.
2 2000-903-01 Rev. C / October 2007

List of Figures

Section Figure Title Page
1 Figure 1-1 Basic measurement connections................................................... 1-3
Figure 1-2 Standard event status.................................................................. 1-13
Figure 1-3 Operation event status................................................................. 1-14
Figure 1-4 Measurement event status .......................................................... 1-14
Figure 1-5 Questionable event status ........................................................... 1-15
Figure 1-6 Status byte and service request (SRQ) ....................................... 1-15
List of Figures Model 2000 Digital Multimeter Quick Start Guide
This page left blank intentionally.
4 2000-903-01 Rev. C / October 2007

List of Tables

Section Figure Title Page
1 Table 1-1 Commands to select sense mode ................................................. 1-3
Table 1-2 Factory defaults ............................................................................. 1-7
Table 1-3 Error and status messages............................................................ 1-9
Table 1-4 IEEE-488.2 common commands and queries............................. 1-16
Table 1-5 Signal oriented measurement commands ................................... 1-17
Table 1-6 CALCulate command summary................................................... 1-18
Table 1-7 DISPlay command summary....................................................... 1-19
Table 1-8 FORMat command summary....................................................... 1-19
Table 1-9 ROUTe command summary........................................................ 1-20
Table 1-10 SENSe command summary ........................................................ 1-20
Table 1-11 STATus command summary........................................................ 1-28
Table 1-12 SYSTem command summary...................................................... 1-29
Table 1-13 TRACe command summary ........................................................ 1-29
Table 1-14 Trigger command summary......................................................... 1-30
Table 1-15 TEMPerature command summary............................................... 1-31
Table 1-16 Models 196/199 device-dependent command summary............. 1-32
Table 1-17 Models 8840A/8842A device-dependent comma nd summary.... 1-36
List of Tables Series 2600 System SourceMeter® Instruments User’s Manual
This page left blank intentionally.
4 2600S-900-01 Rev. B / September 2007
In this section:
Topic Page
Introduction................................................................................1-1
Operation Summary.................................................................1-2
Section 1
Introduction
Power-on defaults..............................................................1-2
Remote Operation Summary.........................................1-13
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide

Introduction

This quick reference guide includes summary information on front panel and remote operation for the Model 2000 Multimeter. For detailed operating information, consult the Model 2000 User’s Manual.Front Panel

Operation Summary

Power-on defaults
Power-on defaults are those settings the instrument assumes when it is turned on. The Model 2000 offers two choices for the settings: user or factory. With user, the instrument will power-on to the last configuration that you saved. With factory, the instrument will power on to the factory default settings.
To store a user configuration in memory:
1. Configure the instrument as desired for USER default.
2. Press SHIFT and then SAVE.
3. Use the ▲ and ▼ keys to select YES or NO. Note: Select NO if you do not wish to change the last configuration that you saved.
4. Press ENTER.
To restore factory or user settings:
1. Press SHIFT and then SETUP.
2. Use the ▲ and ▼ keys to select FACTory or USER.
3. Press ENTER.
NOTE Factory default settings are listed in Table 2.
Instrument connections
Basic connections for Model 2000 measurements are shown in Figure 1. Th ese drawings show connections to the front panel terminals (INPUTS switch in the FRONT position). If using the rear panel terminals, place the INPUTS switch in the REAR position. Note that current (DCI and ACI) measurements cannot be made from the rear panel terminals.
1-2 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
!
F
500V PEAK
FRONT/REAR
2A 250V
AMPS
HI
INPUT
LO
SENSE
Ω 4 WIRE
INPUTS
350V PEAK
1100V PEAK
R
Model 2000
Measure DCV, ACV, Ω2, FREQ (PERIOD) or TEMP *
* Temperature measurements are typically performed through a thermocouple scanner card, such as the Model 2001-TCSCAN. See the User's Manual for details.
!
F
500V PEAK
FRONT/REAR
2A 250V
AMPS
HI
INPUT
LO
SENSE
Ω 4 WIRE
INPUTS
350V PEAK
1100V PEAK
R
Model 2000
Measure
DCI or ACI
!
F
500V PEAK
FRONT/REAR
2A 250V
AMPS
HI
INPUT
LO
SENSE
Ω 4 WIRE
INPUTS
350V PEAK
1100V PEAK
R
Model 2000
Measure
Ω4
R
Figure 1-1 Basic measurement connections
For resistance measurements (¾2 and ¾4) greater than 100k¾, more stable readings can be achieved by using shielding. Place the resistance in a shielded enclosure and connect the shield to INPUT LO. Shielded cable should be used such that the shield (INPUT LO) encircles the other cable conductor(s).
Maximum inputs for the Model 2000 are summarized in Table 1. Table 1-1
Commands to select sense mode
Function Maximum input
DCV 1000V peak ACV 750V rms, 1000V peak, 8×107V•Hz DCI 3A dc, 250V ACI 3A rms, 250V FREQ (PERIOD) 1000V peak, 8× 107V•Hz
Basic measurement procedure
1. Connect the instrument as explained in “Instrument connection”.
2. From the function keys, select the desired measurement function.
3. Use the RANGE keys to select autoranging or a manual range. Pressing the AUTO range key toggles autoranging. You can select a different range with the and RANGE keys.
4. Take a reading from the display.
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-3
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Storing readings (2 to 1024)
1. Set up the instrument for the desired configuration.
2. Press the STORE key.
3. Use the , , ▲ and ▼ keys to select the number of readings to stor e in th e bu ffer.
4. Press ENTER to start the storage process. The asterisk (*) annunciator turns on to indi cate a data storage operation. It turns off when the storage is finished.
Recalling readings
1. Press the RECALL key. The BUFFER annunciator indicates that stored readings are being displayed. The arrow annunciator indicates that more dat a can be viewed with the , , and keys.
2. Use the , , ▲ and ▼ keys to navigate through the reading numbe rs, reading values, and statistics. For any of the buffer st atistics ( maximum, minim um, average, st and ard devia tion) the STA T annunciator is on.
3. Use the EXIT key to return to the normal display.
Changing speed, resolution and filter
RATE — Measurement speed (reading RATE) can be changed for any measurement function
except frequency, period, continuity and diode test. Use the RATE key to select FAST, MEDium or SLOW. Use FAST (0.1 PLC) if speed is of primary importance (at the expense of increased noise and fewer usable digits of resolution). SLOW (10 PLC) provides better noise performance at the expense of speed. MEDium (1 PLC) provides a compromise between noise and speed.
DIGITS — Display resolution can be changed for any measurement function. Use the DIGITS key to select the desired number of digits (3 to 6 ).
FILTER — Filter state (on/off) and configuration can be changed for any measurement function, except frequency, period, continuity and diode test. The FILTER key toggles between filter on (FILT annunciator on) and filter off. After pressing FILTER to enable the filter, you can then enter the filter count (1 to 100), and select the filter type (moving or repeating). (If the FILT annunciator is already on, press FILTER twice before entering the filter count.)
Changing temperature measurement configuration
Press SHIFT and then TCOUPL. Three choices are available using the and keys:
UNITS — C, K or F. Specify units.
TYPE — J, K or T. Specify thermocouple type.
JUNC — SIM or CH1. Simulate a reference junction temperature (for direct thermocouple connection to the INPUT), or use the reference junction (channel 1) provided on a switching card. In order to keep the reference calculations update d and accurate, Channel 1 needs to be read periodically.
To assign a value to a parameter, use the and keys to scroll to the desired parameter. Select the
key, and use the and keys to scroll through and choose the p referred value. Select the
ENTER key to save any changes.
Relative (rel)
Rel can be used to null offsets or subtract a baseline reading from present and future read ings. When rel is enabled by pressing REL (REL annunciator on), the instrument uses the present reading as the relative value. Subsequent readings will be the difference between the actual input value and the rel value. You can define a rel value for each function. Pressing REL a second time disables rel.
1-4 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Percent
Input - Reference
Input
--------------------------------------------- - 100%¥=
dBm 10 log
V
2
IN
Z
REF
§
⎝⎠
⎛⎞
1mW
---------------------------------- -=
V
IN
V
REF
----------------=
Math
To enable and configure a math oper ation, press SHIFT and then the desir ed math key (MX+B, %, dBm or dB). Use the
, , and keys to configure the math operation and press ENTER when done. Pressing SHIFT and then the related math key a second time disables the math operation.
MX+B
Y = mX + b Where: X is the normal display reading.
m and b are the user entered constants. Y is the displayed result.
Percent (%)
Where: Input is the normal display reading.
Reference is the user entered constant. Percent is the displayed result.
dBm
Where: VIN is the DC or AC input signal.
Z
is the specified reference impedance.
REF
dB
Where:VIN is the DC or AC input signal.
V
is the specified voltage reference impedance.
REF
Measuring continuity
With this feature, the instrument alerts you with a beep when a resist ance re ading is below the set level. To measure continuity, press SHIFT and then CONT. Use the
, , and keys to set the
resistance threshold level, and press ENTER. Connect the test leads to INPUT HI and LO.
T esting diodes
With this feature, the instrument measures the forward voltage drop of general-purpose diodes and the zener voltage of zener diodes. To test diodes, press SHIFT and then ( keys to select the test current.
). Use the ▲ and
Limit operations
With this feature, you can check to see of readings are within (“IN” displayed), below (“LO” displayed) or above (“HI” displayed) specified low and high reading limits. You can also set the instrument to beep when readings are inside or outside of the limit range.
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-5
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Setting limits — To set limits (absolute values), press SHIFT and then LIMITS to display the high limit. Use the limit value and again press ENTER.
Enabling limits — Press SHIFT and then ON/OFF to display the beeper status. Use the and keys to change the beeper status (NEVER, OUTSIDE or INSIDE) and press ENTER to enable limits. Pressing SHIFT and then ON/OFF a second time disables limit operations.
, , and keys to enter the desired value, and press ENTER. Enter the low
Reading hold
With this feature, an audible beep is sounded when a valid, settled reading is acquired. A valid reading is determined by the specified window and count.
1. Press SHIFT and then HOLD. Select a window percentage and enter a count.
2. Apply the test probes to the signal and wait for the beeper to sound to indicate a valid read­ing.
3. Remove the hold condition by lifting the probe s.
Pressing SHIFT and then HOLD a second time disables Hold.
Triggers
Factory defaults place the instrument in a continuous (immediate) measurement mode. The instrument can be placed in an external trigger mode where a programmed event is required to cause a measurement (device action).
The instrument is placed in the external trigger mode by pressing th e EXT TRIG key. In this mode, a device action (measurement) will occur when the TRIG key is pressed, a bus trigger (*TRG or GET) is received, or an input trigger via the TRIGGER LINK connector is received. After the device action (measurement), an output trigger occurs and is available at the TRIGGER LINK connector. Refer to the User’s Manual for details on using the trigger link. Pressing EXT TRIG a second time returns the instrument to continuous operation.
Delay — A programmable delay before the device action (measureme nt) occurs is available. It can be set manually or an auto delay can be used. To set a delay, press SHIFT and then DELAY. Use the and keys to select AUTO or MANual. If MANual is chosen, enter the duration of the delay and press ENTER. The AUTO delay times are listed in Table 3-2 of the User’s Manual.
Scan operations
The Model 2000 can be used with an internal scanner car d (Model 2000 SCAN or 2001-TCSCAN), or with external scanner cards installed in a switching main frame (i.e. Model 7001 and 7002).
Step/Scan configuration — A “walk-thr ough” menu is provided to configure stepp ing or scanning. The main menu is accessed by pressing SHIFT and then CONFIG. Use the ENTER keys to select scan type (INTernal or EXTernal), first channel in the scan (MINimum CHANnel), last channel in the scan (MAXimum CHANnel), time between scans, and reading co unt (RDG CNT).
and keys — These keys can be used to manually scan through channels on the internal scanner card. Press to manually increment channels or to manually decrement channels. Hold down either key to scan continuously.
, , , and
OPEN and CLOSE keys — Use these keys to open and close channels on the internal scanner card. After pressing CLOSE, use the and press ENTER. Selecting a different channel will cause the previously closed channel to open. For the ¾4 function, CLOSE will close the selected channel and the paired channel. Paired channels for the Model 2000 SCAN are 1&6, 2&7, 3&8, 4&9, and 5&10. Channel pairs are similar for the Model 2001-TCSCAN except that channels 1&6 are not paired.
1-6 Return to Section Topics 2000-903-01 Rev. C / October 2007
, , and keys to display the desired channel (1 to 10)
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
STEP key — Pressing this key starts a stepping operation of consecutive channels, where output triggers are sent after every channel closure.
SCAN key — Pressing this key starts a scanning operation of consecutive channels, wh ere an output trigger is sent at the end of the scan list.
HALT — Pressing SHIFT and then HALT stops stepping or scanning.
Factory default conditions
Table 1-2
Factory defaults
Setting Factory default
Autozero Buffer Continuity Beeper Digits Rate Threshold Current (AC and DC) Digits (AC) Digits (DC) Filter Count Mode Range Relative Value Rate (AC) Rate (DC) Diode test Digits Range Rate Frequency and Period Digits Range Relative Value Rate Function GPIB Address Language Limits Beeper High limit Low limit mX+b Scale factor Offset Percent References
On No effect
On 4 Fast (0.1 PLC) 10¾
5 6 On 10 Moving average Auto Off
0.0 Medium* Medium (1 PLC)
6 1mA Medium (1 PLC)
6 10V Off
0.0 Slow (1 sec) DCV No effect (16 at factory) (SCPI at factory) Off Never +1
-1 Off
1.0
0.0 Off
1.0
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-7
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-2
Factory defaults (cont.)
Setting Factory default
Resistance (2-wire and 4-wire) Digits Filter Count Mode Range Relative Value Rate RS-232 Baud Flow Tx term Scanning Channels
6 On 10 Moving average Auto Off
0.0 Medium (1 PLC) Off No effect No effect No effect Off 1-10
Internal Mode Temperature Digits Filter Count Mode Junction Temperature Relative Value Rate Thermocouple
5
On
10
Moving average
Simulated
23°C
Off
0.0
Medium (1 PLC)
J
°C Units Triggers Continuous Delay
On
Auto
Immediate Source Voltage (AC and DC) dB reference dBm reference Digits (AC) Digits (DC) Filter Count Mode Range Relative Value Rate (AC)
No effect
75¾
5
6
On
10
Moving average
Auto
Off
0.0
Medium*
Medium (1 PLC) Rate (DC)
*DETector:BANDwidth 30
1-8 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Error and status messages
Table 1-3
Error and status messages
Number Description Event
-440
-430
-420
-410
-363
-350
-330
-314
-315
-285
-284
-282
-281
-260
-241
-230
-225
-224
-223
-222
-221
-220
-215
-214
-213
-212
-211
-210
-202
-201
-200
-178
-171
-170
-168
-161
-160
-158
-154
-151
-150
-148
Query unterminated after indefinite response Query deadlocked Query unterminated Query interrupted Input buffer overrun Queue overflow Self-test failed Save/recall memory lost Configuration memory lost Program syntax error Program currently running Illegal program name Cannot create program Expression error Hardware missing Data corrupt or stale Out of memory Illegal parameter value Too much data Parameter data out of range Settings conflict Parameter error Arm deadlock Trigger deadlock Init ignored Arm ignored Trigger ignored Trigger error Settings lost due to rtl Invalid while in local Execution error Expression data not allowed Invalid expression Expression error Block data not allowed Invalid block data Block data error String data not allowed String too long Invalid string data String data error Character data not allowed
EE EE
EE EE SYS SYS EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-9
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-3
Error and status messages (cont.)
Number Description Event
-144
-141
-140
-128
-124
-123
-121
-120
-114
-113
-112
-111
-110
-109
-108
-105
-104
-103
-102
-101
-100
Character data too long Invalid character data Character data error Numeric data not allowed Too many digits Exponent too large Invalid character in number Numeric data error Header suffix out of range Undefined header Program mnemonic too long Header separator error Command header error Missing parameter Parameter not allowed GET not allowed Data type error Invalid separator Syntax error Invalid character Command error
EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE EE
EE +000 No error SE +101
+121 +122 +123 +124 +125 +126 +161 +171 +174 +301 +302 +303 +304 +305 +306 +307 +308 +309 +310 +311
Operation complete Device calibrating Device settling Device ranging Device sweeping Device measuring Device calculating Program running Waiting in trigger layer Re-entering the idle layer Reading overflow Low limit 1 event High limit 1 event Low limit 2 event High limit 2 event Reading available Voltmeter complete Buffer available Buffer half full Buffer full Buffer overflow
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
1-10 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-3
Error and status messages (cont.)
Number Description Event
+400 +401 +402 +403 +404 +405 +406 +407 +408 +409 +410 +411 +412 +413 +414 +415 +416 +417 +418 +419 +420 +421 +422 +423 +424 +425 +438 +439 +450 +451 +452 +453 +454 +455 +456 +457 +458 +459 +460 +461 +462 +463 +464 +465 +466
+467 +468 +469 +470 +471 +472 +473
Calibration messages: 10 vdc zero error 100 vdc zero error 10 vdc full scale error
-10 vdc full scale error 100 vdc full scale error
-100 vdc full scale error 1k 2-w zero error 10k 2-w zero error 100k 2-w zero error 10M 2-w zero error 10M 2-w full scale error 10M 2-w open error 1k 4-w zero error 10k 4-w zero error 100k 4-w zero error 10M 4-w sense lo zero error 1k 4-w full scale error 10k 4-w full scale error 100k 4-w full scale error 1M 4-w full scale error 10M 4-w full scale error 10m adc zero error 100m adc zero error 10m adc full scale error 100m adc full scale error 1 adc full scale error Date of calibration not set Next date of calibration not set 100m vac dac error 1 vac dac error 10 vac dac error 100 vac dac error 100m vac zero error 100m vac full scale error 1 vac zero error 1 vac full scale error 1 vac noise error 10 vac zero error 10 vac full scale error 10 vac noise error 100 vac zero error 100 vac full scale error 750 vac zero error 750 vac full scale error 750 vac noise error
Post filter offset error 1 aac zero error 1 aac full scale error 3 aac zero error 3 aac full scale error Input time constant error Frequency gain error
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-11
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-3
Error and status messages (cont.)
Number Description Event
+500 +510 +511 +512 +513 +514 +515 +522 +610 +611 +800 +802 +803 +805 +806 +807 +808 +900
Calibration data invalid Reading buffer data lost GPIB address lost Power-on state lost AC calibration data lost DC calibration data lost Calibration dates lost GPIB communication language lost Questionable Calibration Questionable Temperature RS-232 Framing Error detected RS-232 Overrun detected RS-232 Break detected Invalid system communication RS-232 Settings Lost RS-232 OFLO: Characters Lost ASCII only with RS-232 Internal System Error
DDC Status Model: +950 +951 +952 +953 +954 +955 +956 +957 +958 +959 +960 +961
DDC Trigger Overrun Error
DDC Interval Overrun Error
DDC Big String Error
DDC Uncalibrated Error
DDC No Scanner Error
DDC Maximum Channel is 4
DDC Maximum Channel is 8
DDC Calibration Locked
DDC Conflict Error
DDC No Remote Error
DDC Mode IDDC Error
DDC Mode IDDCO Error
Keithley 199 Serial Poll Byte +962 +963 +964 +965 +966
Events:
DDC Ready
DDC Reading Done
DDC Buffer Half Full
DDC Buffer Full
DDC Reading overflow
Fluke 8842 serial poll byte +970 +971 +972
events:
Fluke Error
Data Available
Overrange EE = error event
SE = status event SYS = system error event
EE EE EE EE EE EE EE EE SE SE EE EE EE EE EE EE EE EE
EE EE EE EE EE EE EE EE EE EE EE EE
SE SE SE SE SE
SE SE SE
1-12 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
* ESR ?
PON
(B7)
URQ
(B6)
CME
(B5)
EXE
(B4)
DDE
(B3)
QYE
(B2) (B1) (B0)
OR
Standard Event
Status Register
Standard Event
Status Enable
Register
PON = Power On
URQ = User Request
CME = Command Error
EXE = Execution Error
DDE = Device-Dependent Error
QYE = Query Error
OPC = Operation Complete
& = Logical AND
OR = Logical OR
&
&
&
&
&
OPC
&
&
PON
(B7)
URQ
(B6)
CME
(B5)
EXE
(B4)
DDE
(B3)
QYE
(B2) (B1) (B0)
OPC
* ESE
* ESE ?
To Event
Summary
Bit (ESB) of
Status Byte
Register (See
Figure 6).
(B15 - B8)
(B15 - B8)
Remote Operation Summary
At the factory, the instrument is set for GPIB bus operation at primary address 16 using the SCPI programming language.
GPIB bus — Use the GPIB bus configuration menu to enable or disable GPIB bus operation (ON or OFF), and to check and/or change the primary address (0 to 30) and language (SCPI, 199, or
8842). The configuration menu is accessed by pressing SHIFT and then GPIB. Commands for the SCPI language are provided in Tables 4 through 15. The DDC commands for the 199 and 8842 languages are provided in Tables 16 and 17.
RS-232 interface — Use the RS-232 interface menu to enable or disable RS-2 32 interface operation (ON or OFF), and to check and/or change the baud rate (300 to 19.2k), flow control (NONE or XonXoFF) and terminator (LF, CR or LFCR). Note that only the SCPI language commands can be used over the RS-232 interface.
Status structure
Figure 1-2
Standard event status
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-13
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
OR
Idle = Idle state of the 2000
Trig = Triggering
Meas = Measuring
& = Logical AND
OR = Logical OR
(B15 - B11)
Idle
(B10) (B9) (B8) (B7) (B6)
Trig
(B5)
Meas
(B4) (B3) (B2) (B1) (B0)
Operation Event
Register
(B15 - B11)
Idle
(B10) (B9) (B8) (B7) (B6)
Trig
(B5)
Meas
(B4) (B3) (B2) (B1) (B0)
Operation Event
Enable Register
&
To Operation
Summary Bit
(OSB) of Status
Byte Register.
(See Figure 6).
&
&
(B15 - B11)
Idle
(B10) (B9) (B8) (B7) (B6)
Trig
(B5)
Meas
(B4) (B3) (B2) (B1) (B0)
Operation
Condition Register
OR
BFL = Buffer Full
BHF = Buffer Half Full
BAV = Buffer Available
& = Logical AND
OR = Logical OR
(B15 - B12) (B10) (B9)
BHF
(B8)
BAV
(B7) (B6)
RAV
(B5) (B4) (B3)
(B2)
LL
(B1) (B0)
Measurement Event
Register
(B15 - B12)
Measurement Event
Enable
Register
To Measurement
Summary Bit
(MSB) of Status
Byte Register.
(See Figure 6).
&
&
(B15 - B12) (B10) (B9) (B8) (B7) (B6)
RAV
(B5) (B4) (B3)
(B2)
LL
(B1) (B0)
Measurement
Condition Register
BFL
HL
ROF
BFL
ROF
&
&
&
BHF BAV
HL
&
&
(B10) (B9)
BHF
(B8)
BAV
(B7) (B6)
RAV
(B5)
(B4) (B3)
(B2)
LL
(B1) (B0)
BFL ROFHL
RAV = Reading Available
HL = High Limit
LL = Low Limit
ROF = Reading Overflow
(B11)
(B11)
(B11)
Figure 1-3
Operation event status
Figure 1-4
Measurement event status
1-14 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
(B13 - B9)(B14) (B8)(B15)
OR
Questionable
Condition Register
Questionable Event
Enable Register
Warn = Command Warning
Cal = Calibration Summary
Temp = Temperature Summary
& = Logical AND
OR = Logical OR
&
&
&
0
(B13 - B9)(B14)
(B8)
(B15)
0
Warn Cal
Questionable Event
Register
(B13 - B9)(B14) (B8)(B15)
To Questionable
Summary Bit (QSB)
of Status
Byte Register
(See Figure 6).
(B7 - B5)
(B7 - B5)
(B7 - B5)
Warn Cal
0
Warn Cal
(B4)
Temp
(B4)
Temp
(B3 - B0)
(B3 - B0)
(B3 - B0)(B4)
Temp
&
Status Summary Messages
* STB?
Serial Poll
OSB
(B7)
RQS
(B6)
MSS
ESB
(B5)
MAV
(B4)
QSB
(B3)
EAV
(B2)
(B1) (B0)
OR
* SRE
* SRE?
Status Byte
Register
Service
Request
Enable
Register
OSB = Operation Summary Bit
MSS = Master Summary Status
RQS = Request for Service
ESB = Event Summary Bit
MAV = Message Available
QSB = Questionable Summary Bit
EAV = Error Available
MSB = Measurement Summary Bit
& = Logical AND
OR = Logical OR
OSB
(B7) (B6)
ESB
(B5)
MAV
(B4)
QSB
(B3)
EAV
(B2)
(B1) (B0)
&
&
&
&
&
MSB
MSB
&
Service
Request
Generation
Read by Serial Poll
Read by *STB?
Figure 1-5
Questionable event status
Figure 1-6
Status byte and service request (SRQ)
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-15
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
IEEE-488.2 common commands and queries
Table 1-4
IEEE-488.2 common commands and queries
Mnemonic Name Description
*CLS *ESE <NRf> *ESE? *ESR?
*IDN?
*OPC
*OPC?
*OPT?
Clear status Event enable command Event enable query Event status register query
Identification query
Operation complete command
Operation complete query
Option identification query
Clears all event registers and Error Queue. Program the Standard Event Enable Register. Read the Standard Event Enable Register. Read the Standard Event Enable Register and
clear it. Returns the manufacturer, model number,
serial number and firmware revision levels of the unit.
Set the Operation Complete bit in the Standard Event Status Register after all pending commands have been executed.
Places an ASCII “1” into the output queue when all pending selected device operations have been completed.
Returns an ID code that indicates which memory option is installed and whether or not the optional scanner card is installed.
*RCL <NRf>
*RST
*SAV <NRf>
*SRE <NRf>
*SRE? *STB? *TRG *TST?
*WAI
1-16 Return to Section Topics 2000-903-01 Rev. C / October 2007
Recall command
Reset command
Save command
Service request enable command
Service request enable query Read status byte query Trigger command Self-test query
Wait-to-continue command
Returns the Model 2000 to the setup configuration stored in the specified memory location.
Returns the Model 2000 to the *RST default conditions.
Saves the current setup to the specified memory location.
Programs the Service Request Enable Register.
Reads the Service Request Enable Register. Reads the Status Byte Register. Sends a bus trigger to the 2000. Performs a checksum test on ROM and returns
the result. Wait until all previous commands are executed.
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
SCPI command subsystems
NOTE 1. Brackets ([]) are used to denote optional character sets. These
optional characters do not have to be included in the program message. Do not use brackets in the program message.
2. Angle brackets (<>) are used to indicate parameter type. Do not use angle brackets in the program message.
3. Upper case characters indicate the short-from version for each command word.
Table 1-5
Signal oriented measurement commands
Command Description
:CONFigure:<funct ion>
:FETCh? :READ?
MEASure[:<functio n>]?
Places the Model 2000 in a “one-shot” measurement mode for the specified function.
Requests the latest reading. Performs an :ABORt,
:INITiate, and a :FETCh?. Performs an :ABORt,
:CONFigure:<function>, and a :READ?.
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-17
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-6
CALCulate command summary
Command Description
:CALCulate[1] :FORMat <name>
:FORMat? :KMATh :MMFactor <NRf>
:MMFactor? :MBFactor <NRf>
:MBFactor? :MUNits <name>
:MUNits? :PERCent <NRf>
:ACQuire :PERCent? :ST ATe <b> :STATe? :DATA?
:CALCulate2 :FORMat <name>
:FORMat? :STATe <b> :STATe? :IMMediate :IMMediate? :DATA?
:CALCulate3 :LIMit[1] :UPPer [:DATA] <n> [:DATA]? :LOWer [:DATA] <n> [:DATA]? :STATe <b> :STATe? :FAIL? :CLEAR [:IMMediate] :AUTO <b> :AUTO? :IMMediate
Subsystem to control CALC 1: Select math format (NONE, MXB, PERCent). Query math format. Path to configure math calculations: Set “m” factor for mx+b (-100e6 to 100e6). Query “m” factor. Set “b” factor for mx+b (-100e6 to 100e6). Query “b” factor. Specify units for mx+b reading (three characters ‘A’ through ‘Z’). Query “mx+b” units. Set target value for PERCent calculation (-100e6 to 100e6). Use input signal as target value. Query percent. Enable or disable kmath calculation. Query state of kmath function. Read result of kmath calculation.
Subsystem to control CALC 2: Select math format: (MEAN, SDEViation, MAXimum, MINimum, or NONE). Query math format. Enable or disable calculation. Query state of math function. Recalculate raw input data in buffer. Perform calculation and read result. Read math result of CALC 2.
Subsystem to control CALC 3 (limit test): Path to control LIMIT 1 test: Path to configure upper limit: Set upper limit (-100e6 to 100e6). Query upper limit. Path to configure lower limit: Set lower limit (-100e6 to 100e6). Query lower limit. Enable or disable limit test. Query state of limit test. Query test result (1 = pass, 0 = fail). Path to clear failed test: Clear failed test indication. Enable or disable auto clear. Query auto clear. Re-perform limit tests.
1-18 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-7
DISPlay command summary
Command Description
:DISPlay [:WINDow[1]] :TEXT :DATA <a>
:DATA? :STATe <b> :STATe?
:ENABle <b> :ENABle?
Path to control user text messages. Define ASCII message “a” (up to 12 characters). Query text message. Enable or disable message mode. Query text message state.
Enable or disable the front panel display. Query state of the display.
Table 1-8
FORMat command summary
Command Description
:FORMat [:DATA] <type>[,<length>]
[:DATA]? :ELEMents <item list>
:ELEMents? :BORDer <name>
:BORDer?
Select data format: (ASCii, SREal or DREal). Query data format. Specify data elements: (READing, CHANnel, and UNITs). Query data elements. Select binary byte order: (NORMal or SWAPped). Query byte order.
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-19
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-9
ROUTe command summary
Command Description
:ROUTe :CLOSe <chan num>
:STATe? :OPEN:ALL :MULTiple :CLOSe <list>
:STATe? :OPEN <list> :SCAN [:INTernal] <list>
[:INTernal]? :EXTernal <list>
:EXTernal? :LSELect <name>
:LSELect?
Table 1-10
SENSe command summary
Commands to control scanner card: Close specified channel (1 to
10) or channel pair (1 to 5). Query closed channel (or channel pair). Open all input channels (1 through 10). Path to close and open multiple channels: Close specified channels (1 to
11). Query closed channel. Open specified channels (1 to
11). Path to scan channels. Specify internal scan list (2 to 10 channels). Query internal scan list. Specify external scan list (2 to 800 channels). Query external scan list. Select scan operation (INTernal, EXTernal, or NONE). Query scan operation.
Command Description
[:SENSe[1]] :FUNCtion <name>
:FUNCtion? :DATA? :FRESh? :HOLD :WINDow <NRf> :WINDow? :COUNt <NRf> :COUNt? :ST ATe <NRf> :STATe?
Select measurement function: ‘VOLTage:AC’, VOLTage :DC’, RESistance’, ‘FRESistance’, ‘CURRent:AC’, ‘CURRent: DC’ , ‘FREQuency’,‘TEMPerature’, ‘PERiod’, ‘DIODe’, “CONTinuity’. Query function. Return the last instrument reading. Returns a new (fresh) reading. Path to control Hold feature: Set Hold window (%); 0.01 to
20. Query Hold window. Set Hold count; 2 to 100. Query Hold count. Enable or disable Hold. Query state of Hold.
1-20 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-10
SENSe command summary (cont.)
Command Description
:CURRent:AC :NPLCycles <n>
:NPLCycles? :RANGe
[:UPPer] <n> [:UPPer]? :AUTO <b> :AUTO? :REFerence <n> :STATe <b> :STATe? :ACQuire :REFerence? :DIGits <n>
:DIGits? :AVERage
:TCONtrol <name> :TCONtrol?
:COUNt <n> :COUNt? :STATe <b> :STATe?
:CURRent:AC :DETector :BANDwidth <NRf>
:BANDwidth?
Path to configure AC current. Set integration rate (line cycles; 0.01 to 10). Query line cycle integration rate. Path to configure measurement range: Select range (0 to 3.1). Query range. Enable or disable auto range. Query auto range. Specify reference (-3.1 to
3.1). Enable or disable reference. Query state of reference. Use input signal as reference. Query reference value. Specify measurement resolution (4 to 7). Query resolution. Path to configure and control the filter. Select filter type: (MOVing or REPeat). Query filter type. Specify filter count (1 to 100). Query filter count. Enable or disable filter. Query state of digital filter.
Path to configure bandwidth: Specify bandwidth (3 to 300e3). Query bandwidth.
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-21
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-10
SENSe command summary (cont.)
Command Description
:CURRent:DC :NPLCycles <n>
:NPLCycles? :RANGe
[:UPPer] <n> [:UPPer]? :AUTO <b> :AUTO? :REFerence <n> :STATe <b> :STATe? :ACQuire :REFerence? :DIGits <n>
:DIGits? :AVERage
:TCONtrol <name> :TCONtrol?
:COUNt <n> :COUNt? :STATe <b> :STATe?
:VOLTage:AC :NPLCycles <n>
:NPLCycles? :RANGe
[:UPPer] <n> [:UPPer]? :AUTO <b> :AUTO? :REFerence <n>
Path to configure DC current. Set integration rate (line cycles; 0.01 to 10). Query line cycle integration rate. Path to configure measurement range: Select range (0 to 3.1). Query range. Enable or disable auto range. Query auto range. Specify reference (-3.1 to
3.1). Enable or disable reference. Query state of reference. Use input signal as reference. Query reference value. Specify measurement resolution (4 to 7). Query resolution. Path to configure and control the filter. Select filter type: (MOVing or REPeat). Query filter type. Specify filter count (1 to 100). Query filter count. Enable or disable filter. Query state of digital filter.
Path to configure AC voltage. Set integration rate (line cycles; 0.01 to 10). Query line cycle integration rate. Path to configure measurement range: Select range (0 to 757.5). Query range. Enable or disable auto range. Query auto range. Specify reference (-757.5 to
757.5).
1-22 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-10
SENSe command summary (cont.)
Command Description
:VOLTage:AC :REFerence <n> :STATe <b> :STATe? :ACQuire :REFerence? :DIGits <n>
Enable or disable reference. Query state of reference. Use input signal as reference. Query reference value.
Specify measurement :DIGits? :AVERage
resolution (4 to 7).
Query resolution.
Path to configure and control :TCONtrol <name>
the filter.
Select filter type: (MOVing or :TCONtrol? :COUNt <n> :COUNt? :STATe <b> :STATe? :DETector :BANDwidth <NRf>
REPeat).
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
Path to configure bandwidth:
Specify bandwidth (3 to :BANDwidth?
300e3).
Query bandwidth. :VOLTage:DC
:NPLCycles <n>
Path to configure DC voltage:
Set integration rate (line
cycles; (0.01 to 10). :NPLCycles? :RANGe
Query line cycle integration
rate.
Path to configure [:UPPer] <n> [:UPPer]? :AUTO <b> :AUTO? :REFerence <n>
measurement range:
Select range (0 to 1010).
Query range.
Enable or disable auto range.
Query auto range.
Specify reference (-1010 to :STATe <b> :STATe?
+1010).
Enable or disable reference.
Query state of reference (0 or :ACQuire :REFerence? :DIGits <n>
1).
Use input signal as
reference.
Query reference value. :DIGits?
Specify measurement
resolution (4 to 7).
Query resolution. :VOLTage:DC
:AVERage
Path to configure and control
the filter. :TCONtrol <name>
Select filter type: (MOVing or
REPeat). :TCONtrol? :COUNt <n> :COUNt? :STATe <b> :STATe?
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-23
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-10
SENSe command summary (cont.)
Command Description
:RESistance :NPLCycles <n>
:NPLCycles? :RANGe
[:UPPer] <n> [:UPPer]? :AUTO <b> :AUTO? :REFerence <n> :STATe <b> :STATe? :ACQuire :REFerence? :DIGits <n>
:DIGits? :AVERage
:TCONtrol <name> :TCONtrol?
:COUNt <n> :COUNt? :STATe <b> :STATe?
Path to configure resistance:
Set integration rate (line
cycles; 0.01 to 10).
Query line cycle integration
rate.
Path to configure
measurement range:
Select range (0 to 120e6).
Query range.
Enable or disable auto range.
Query auto range.
Specify reference (0 to
120e6).
Enable or disable reference.
Query state of reference.
Use input signal as
reference.
Query reference value.
Specify measurement
resolution (4 to 7).
Query resolution.
Path to configure and control
filter.
Select filter type: (MOVing or
REPeat).
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
1-24 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-10
SENSe command summary (cont.)
Command Description
:FRESistance :NPLCycles <n> :NPLCycles?
:RANGe [:UPPer] <n>
[:UPPer]? :AUTO <b> :AUTO? :REFerence <n> :STATe <b> :STATe? :ACQuire :REFerence? :DIGits <n>
:DIGits? :AVERage
:TCONtrol <name> :TCONtrol?
:COUNt <n> :COUNt? :STATe <b> :STATe?
:TEMPerature :NPLCycles <n>
:NPLCycles? :REFerence <n> :STATe <b> :STATe? :ACQuire :REFerence? :DIGits <n>
:DIGits? :AVERage
Path to configure four-wire
resistance:
Set integration rate (line
cycles; 0.01 to 10).
Query line cycle integration
rate.
Path to configure
measurement range:
Select range (0 to 101e6).
Query range.
Enable or disable auto range.
Query auto range.
Specify reference (0 to
+101e6).
Enable or disable reference.
Query state of reference.
Use input signal as
reference.
Query reference value.
Specify measurement
resolution (4 to 7).
Query resolution.
Path to configure and control
filter.
Select filter type: (MOVing or
REPeat).
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
Path to configure
temperature:
Set integration rate (line
cycles; 0.01 to 10).
Query line cycle integration
rate.
Specify reference; -200 to
1372.
Enable or disable reference.
Query state of reference.
Use input signal as
reference.
Query reference value.
Specify measurement
resolution (4 to 7).
Query resolution.
Path to configure and control
the filter.
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-25
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-10
SENSe command summary (cont.)
Command Description
:TEMPerature :AVERage :TCONtrol <name>
Select filter type: (MOVing or
REPeat). :TCONtrol? :COUNt <n> :COUNt? :STATe <b> :STATe? :TCouple :TYPE <name>
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
Path to configure
thermocouple:
Select thermocouple type (J, :TYPE? :RJUNction
K, or T).
Query thermocouple type.
Path to configure reference :RSELect <name>
junction:
Select reference type :RSELect? :SIMulated <n>
(SIMulated or REAL).
Query reference type.
Specify simulated :SIMulated?
temperature in °C (0 to 50).
Query simulated :REAL
temperature.
Path to configure real :TCOefficient <n>
reference junction:
Specify temp coefficient :TCOefficient? :OFFSET <n>
(-0.09999 to 0.09999).
Query temp coefficient.
Specify voltage offset at 0°C
(-0.09999 to 0.09999). :OFFSet?
:FREQuency :APERture
Query voltage offset.
Path to configure frequency.
Sets gate time for period
measurements (0.01-1.0s). :APERture? :THReshold
Query period gate time.
Path to select the threshold
voltage range: :VOLTage :RANGe <n>
Select threshold range (0 to
1010). :RANGe? :REFerence <n> :ST ATe <b> :STATe? :ACQuire :REFerence? :DIGits <n>
Query threshold range. Specify reference (0 to
1.5e7)
Enable or disable reference. Query state of reference. Use input signal as reference. Query reference value.
:DIGits?
Specify measurement resolution (4 to 7). Query resolution.
1-26 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-10
SENSe command summary (cont.)
Command Description
:PERiod :APERture
:APERture? :THReshold
:VOLTage :RANGe <n>
:RANGe? :REFerence <n> :ST ATe <b> :STATe? :ACQuire :REFerence? :DIGits <n>
:DIGits?
:DIODe :CURRent :RANGe [:UPPer] <NRf> [:UPPer]?
:CONTinuity :THReshold <NRf>
:THReshold?
Path to configure period. Sets gate time for period measurements (0.01-1.0s). Query period gate time. Path to select the threshold voltage range:
Select threshold range (0 to
1010).
Query threshold range. Specify reference (0 to 1). Enable or disable reference. Query state of reference. Use input signal as reference. Query reference value. Specify measurement resolution (4 to 7). Query resolution.
Paths to configure diode test: Path to select range.
Select range (0 to 1e-3) Query range.
Path to configure continuity test: Set threshold resistance (1 to
1000).
Query threshold resistance.
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-27
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-11
STATus command summary
Command Description
:STATus :MEASurement
Path to control measurement
event registers: [:EVENt]? :ENABle <NRf> :ENABle? :CONDition? :OPERation
Read the event register.
Program the enable register.
Read the enable register.
Read the condition register.
Path to control operation status
registers: [:EVENt]? :ENABle <NRf> :ENABle? :CONDition? :QUEStionable
Read the event register.
Program the enable register.
Read the enable register.
Read the condition register.
Path to control questionable
status registers: [:EVENt]? :ENABle <NRf> :ENABle? :CONDition? :PRESet :QUEue [:NEXT]? :ENABle <list>
Read the event register.
Program the enable register.
Read the enable register.
Read the condition register.
Return status registers to
default states.
Path to access error queue:
Read the most recent error
message. :ENABle? :DISable <list>
Specify error and status
messages for queue.
Read the enabled messages. :DISable? :CLEar
Specify messages not to be
placed in queue.
Read the disabled messages.
Clears all messages from Error
Queue.
1-28 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-12
SYSTem command summary
Command Description
:SYSTem :PRESet :POSetup <name>
:POSetup? :FRSWitch?
:VERSion? :ERRor? :AZERo :STATe <b> :STATe? :KEY <NRf>
:KEY? :CLEar
:BEEPer [:STATe] <b> [:STATe]? :LOCal
:REMote :RWLock
:KCLick <b> :KCLick :LFRequency?
Return to :SYST:PRES defaults. Select power-on setup: (RST, PRESet or SAV0). Query power-on setup. Query INPUTS switch (0=rear, 1=front). Query rev level of SCPI standard. Query (read) Error Queue. Path to set up autozero. Enable or disable autozero. Query autozero. Simulate key-press (1 to 31; see Figure 5-10 of the User’s Manual). Query the last “pressed” key. Clears messages from the Error Queue. Path to control beeper. Enable or disable beeper. Query state of beeper. Take 2000 out of remote and restore operation of front panel controls (RS-232 only). Place 2000 in remote (RS-232 only). Lockout front panel controls (RS-232 only). Turn the keyclick on/off. Query the keyclick status. Query power line frequency.
Table 1-13
TRACe command summary
Command Description
:TRACe|:DATA :CLEar
:FREE? :POINts <NRf>
:POINts? :FEED <name>
:CONTrol <name> :CONTrol?
:FEED? :DATA?
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-29
Use :TRACe or :DATA as root command. Clear readings from buffer. Query bytes available and bytes in use. Specify size of buffer (2 to
1024). Query buffer size. Select source of readings (SENSe[1], CALCulate[1], NONE. Select buffer control mode (NEVer or NEXT) Query buffer control mode. Query source of readings for buffer. Read all readings in the buffer.
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-14
Trigger command summary
Command Description
:INITiate [:IMMediate] :CONTinuous <b>
:CONTinuous? :ABORt :TRIGger[:SEQuen ce[1]] :COUNt <n>
:COUNt? :DELay <n> :AUTO <b> :AUTO? :DELay? :SOURce <name>
:SOURce? :TIMer <n>
:TIMer? :SIGNal
:SAMPle :COUNt <NRf> :COUNt?
Subsystem command path: Initiate one trigger cycle. Enable or disable continuous initiation. Query continuous initiation. Reset trigger system. Path to program Trigger Layer: Set measure count (1 to 9999, or INF). Query measure count. Set delay (0 to 999999.999 sec) Enable or disable auto delay. Query state of delay. Query delay. Select control source (IMMediate, TIMer, MANual, BUS, or EXTernal). Query control source. Set timer interval (0 to
999999.999 sec). Request the programmed timer interval. Loop around control source.
Specify sample count (1 to
1024). Query sample count.
1-30 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-15
TEMPerature command summary
Command Description
:UNIT :TEMPerature <name>
Select temperature measurement units (C, F, or
K). :TEMPerature? :VOLTage :AC <name>
Query temperature units.
Path to configure voltage
units.
Select ACV measurement :DB
units (V, DB or DBM).
Path to set DB reference :REFerence <n>
:REFerence? :DBM
voltage.
Specify reference in volts
-7
to 1000).
(1e
Query DB reference.
Path to set DBM reference :IMPedance <n>
impedance.
Specify reference :IMPedance?
impedance (1 to 9999).
Query DBM reference :AC? [:DC] <name>
impedance.
Query ACV units.
Select DCV measurement :DB
units (V, DB, or DBM)
Path to set DB reference :REFerence <n>
voltage:
Specify reference in volts (0 :REFerence? :DBM
to 1000).
Query reference.
Path to set DBM refernece :IMPedance <n>
:IMPedance? :DC?
impedance:
Specify reference
impedance (1e
-7
to 9999). Query reference impedance. Query DCV units.
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-31
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Models 196/199 and 8840A/8842A commands
Table 1-16
Models 196/199 device-dependent command summary
Mode Command Description
Execute X Execute other device-
dependent commands.
Function F0
Range
R0 R1 R2 R3 R4 R5 R6 R7
Zero (rel) Z0
Filter P0
Rate S0
Trigger mode
F1 F2 F3 F4 F5 F6 F7 F8 F9
DCV ACV DCA ACA Ohms* ACV dB Freq
Auto Auto Auto Auto Auto Auto — 1V 1V 100mA 1A 1k¾ 1V 0.1V 10V 10V 3A 3A 10k¾ 10V 1V 100V 100V 3A 3A 100k¾ 100V 10V 1000V 750V 3A 3A 1M¾ 750V 100V 1000V 750V 3A 3A 10M¾ 750V 750V 1000V 750V 3A 3A 100M¾ 750V — 1000V 750V 3A 3A 100M¾ 750V
*2-wire and 4-wire ohms
Z1 Z2
P1 P2
S1
S2
T0 T1 T2 T3 T4 T5 T6
DC volts AC volts 2-wire ohms DC current AC current ACV dB Not valid Frequency Temperature 4-wire ohms
Zero disabled Zero enabled Zero enabled using a zero value (V)
Filter disabled Moving filter (count = 10) Repeat filter (count = 10)
0.1 PLC integration Line cycle integration (16.67msec, 60Hz; 20msec, 50Hz) 10 PLC (166.67msec integration, 60Hz; 200msec integration, 50Hz)
Continuous on Talk One-shot on Talk Continuous on GET One-shot on GET Continuous on X One-shot on X Continuous on External Trigger
T7
Reading mode
1-32 Return to Section Topics 2000-903-01 Rev. C / October 2007
B0 B1
B2
One-shot on External Trigger Readings from A/D converter
Individual readings from data store All readings from data store (buffer dump)
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-16
Models 196/199 device-dependent command summary (cont.)
Mode Command Description
Data store size
I0 In
Interval Q0
Qn
Value V±nn.nnnn or
V±n.nnnnnnE+n
Default
L0
conditions
L1
Data format G0
G1 G2
G3 G4 G5 G6 G7
SRQ M0
M1 M2 M4 M8 M16 M32
EOI and bus
K0
hold-off
K1 K2 K3
Terminator Y0
Y1 Y2 Y3
Disable data store Data store of n (n=1 to 500), fill and stop
Default interval, 175msec (SELECT OFF) n=interval in milliseconds (15msec to 999999msec)
Zero value, simulated reference junction temperature
Restore factory default conditions and save (L1) Save present machine states as default conditions
Reading with prefix. Reading without prefix. Reading and buffer location with prefix. Reading and buffer location without prefix. Reading and channel with prefix. Reading and channel without prefix. Reading, buffer location, and channel with prefix. Reading, buffer location, and channel without prefix.
Disable Reading overflow Data store full Data store half full Reading done Ready Error
Enable EOI and bus hold-off on X Disable EOI, enable bus hold­off on X Enable EOI, disable bus hold­off on X Disable both EOI and bus hold­off on X
CR LF LF CR CR LF
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-33
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-16
Models 196/199 device-dependent command summary (cont.)
Mode Command Description
Status U0
Send machine status word (199 format only)
U1
Send error conditions (only supports no scanner, IDDC, IDDCO)
U2
Send Translator word list (since Translator is not supported, replies with one
space character) U3 U4
Send buffer size
Send current value of “V” (199
format, equivalent to U7 for
196)
U5
Send input switch status (front
/rear) (199 format, equivalent
to U8 for 196) U6
Send simulated temperature
(set by H0)
Multiplex A0
A1
Auto/Cal multiplex disabled
Auto/Cal multiplex enabled
Delay Wn n=delay period in milliseconds,
(0msec to 999999msec)
Display Da
Display up to 12-character
message (a=character) D
Cancel display mode
1-34 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-16
Models 196/199 device-dependent command summary (cont.)
Mode Command Description
Scanning N0
N1 N2 N3 N4 N5 N6 N7 N8 N9 N10
N11 N12 N13 N14 N15 N16 N17 N18 N19
N20 N21
N22 N23 N24 N25 N26 N27 N28 N29
Thermocou ple
J0 J1 J2
Open all - stop scanning or
stepping if applicable
Close channel 1
Close channel 2
Close channel 1
Close channel 4
Close channel 5
Close channel 6
Close channel 7
Close channel 8
Close channel 9
Close channel 10
Step mode, max channel is 2
Step mode, max channel is 3
Step mode, max channel is 4
Step mode, max channel is 5
Step mode, max channel is 6
Step mode, max channel is 7
Step mode, max channel is 8
Step mode, max channel is 9
Step mode, max channel is 10
Open all - stop scanning or
stepping if applicable
Scan mode, max channel is 2
Scan mode, max channel is 3
Scan mode, max channel is 4
Scan mode, max channel is 5
Scan mode, max channel is 6
Scan mode, max channel is 7
Scan mode, max channel is 8
Scan mode, max channel is 9
Scan mode, max channel is 10
Type J thermocouple
Type K thermocouple
Type T thermocouple O0
Simulated reference junction
(for temperature function) O1
Real reference junction (for
temperature function) H0
Set simulated reference
junction temperature using “V”
command; 0 to 50 (°C).
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-35
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-17
Models 8840A/8842A device-dependent command summary
Mode Command Description
Function F1
F2 F3 F4 F5 F6 F7 F8
Range
VDC VAC k¾* mA DC mA AC Freq
R0 R1 R2 R3 R4 R5 R6 R7 R8
** ** ** ** ** — 1V 1V 1k¾ 10mA 3A 1V 10V 1 0V 10k¾ 10mA 3A 10V 100V 100V 100k¾ 100mA 3A 100V 1000V 750V 1M¾ 1A 1A 750V 1000V 750V 10M¾ 3A 3A 750V 1000V 750V 100M¾ 3A 3A 750V *** *** *** *** ***
100mV 750V 100¾ 3A 3A 750V *2-wire and 4-wire ohms **Auto on ***Auto off
Reading rate S0
S1 S2
Trigger mode
T0 (default) T1 T2 T3 T4
Offet (Rel) B0
B1
Display D0
D1
Suffix Y0
Y1
Terminator W0
W1 W2 W3 W4 W5 W6 W7
Clear *
X0
VDC (default) VAC 2-wire k¾ 4-wire k¾ mA DC mA AC Freq Temp
Slow, 10 PLC Medium, 1 PLC, 6 -digit resolution (default) Fast, 0.1 PLC, 5 -digit resolution
Trigger Rear panel Auto mode trigger delay
Internal Disabled — External Enabled On External Disabled On External Enabled Off External Disabled Off Note: Delay is enabled by entering EXT TRIG mode while in local.
Offset off (default) Offset on
Normal display (default) Blank display
Disable output suffix (default) Enable output suffix
Enable CR LF EOI (default) Enable CR LF only Enable CR EOI only Enable CR only Enable LF EOI only Enable LF only Enable EOI only Disable all output terminators
Device clear (resets 8842A to default conditions) Clear error register
1-36 Return to Section Topics 2000-903-01 Rev. C / October 2007
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Table 1-17
Models 8840A/8842A device-dependent command summary (cont.)
Mode Command Description
Single-trigger ? Trigger measurement GET G0
Get instrument configuration (function, range, speed,
trigger) G1 G2
Get SRQ mask
Get calibration input prompt
(generates error message
51)
G3
Get user-defined message
(replies with 16 space
characters) G4
Get calibration status (replies
with “1000”) G5
Get IAB status (input F/R,
autorange on/off, offset on/
off) G6
Get YW status (suffix
enabled/disabled, terminator
selection) G7 G8
Get error status
Get instrument ID
(replies with
“FLUKE,8842A,0,EMUL”) G9
Get status of JKM commands
(temp. units; TC type and
junction)
String = 1jkm
GET (cont.) G10
Get closed channel number
String = 10nn
Where: nn =
00 (all open)
01 through 10 (closed
channel) G11
Get simulated reference
junction temperature
String = xx.xxx (in °C)
Note: G2 valid only in
calibration mode.
PUT P0
Put instrument configuration
(function, range, speed,
trigger) P1 P4
Put SRQ mask
Set simulated junction
temperature; 0 to 50 (°C)
Default temperature = 23°C
PUT format N <value> P0
N <value> P1 N <value> P4
Thermocoupl e type
Temperature unit
K0 K1
K2 J0
J1 J2
Type J thermocouple
(default)
Type K thermocouple
Type T thermocouple
°C (default)
K
°F
2000-903-01 Rev. C / October 2007 Return to Section T opics 1-37
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Table 1-17
Models 8840A/8842A device-dependent command summary (cont.)
Mode Command Description
Thermocoupl e reference junction
Scanner channel
SRQ mask values
M0
Select simulated junction
(default) M1
Select real junction (channel
1 for 2001-TCSCAN card) Ln Control specified channel of
internal scan card
where: n =
0 (Open all channels)
1 through 10 (close specified
channel)
Note: For 4-wire ohms, only
channels 1 through 5 are
valid due to automatic
channel pairing. 00
01 16 32
SRQ disabled (default)
SRQ on overrange
SRQ on data available
SRQ on any error
Note: Add SRQ mask values
for combinations.
Example: 33 for SRQ on
overrange or any error.
1-38 Return to Section Topics 2000-903-01 Rev. C / October 2007
Service Form
Model No. Serial No. Date Name and Telephone No.
Company
List all control settings, describe problem and check boxes that apply to problem.
Intermittent Analog output follows display Particular range or function bad; specify
IEEE failure Obvious problem on power-up Batteries and fuses are OKFront panel operational All ranges or functions are bad Checked all cables
Display or output (check one)
Drifts Unable to zeroUnstable Will not read applied inputOverload
Calibration only Certificate of calibration requiredData required
(attach any additional sheets as necessary) Show a block diagram of your measurement system including all instruments connected (whether power is turned on
or not). Also, describe signal source.
Where is the measurement being performed? (factory, controlled laboratory, out-of-doors, etc.)
What power line voltage is used? Ambient temperature?°F
Relative humidity? Other?
Any additional information. (If special modifications have been made by the user, please describe.)
Be sure to include your name and phone number on this service form.
12/06
12/06
Specifications are subject to change without notice.
All Keithley trademarks and trade names are the property of Keithley Instruments, Inc.
All other trademarks and trade names are the property of their respective companies.
A GREATER MEASURE OF CONFIDENCE
Keithley Instruments, Inc.
Corporate Headquarters • 28775 Aurora Road • Cleveland, Ohio 44139 • 440-248-0400 • Fax: 440-248-6168 • 1-888-KEITHLEY • www.keithley.com
Loading...