Tecfluid LD61R User Manual

Page 1
LD61R
Level switch
Instructions manual
R-MI-LD61R Rev.: 1 english version
1
Page 2
2 3
Page 3
TABLE OF CONTENTS
1. INTRODUCTION .................................................................................... 4
3. INSTALLATION ...................................................................................... 4
4. ELECTRICAL CONNECTION ................................................................ 5
4.1. Power supply connection ......................................................... 6
4.2. Relay output connection ........................................................... 6
5. OPERATION ........................................................................................... 7
6. MAINTENANCE ...................................................................................... 7
6.1. Cleaning …………………………..……………………………........ 7
6.2. Fuse ……....…...…………………..……………………………........ 7
7. TECHNICAL CHARACTERISTICS ........................................................ 8
7.1. Materials ..................................................................................... 8
7.2 Process connection ........................................................................ 8
7.3 Power supply .............................................................................. 8
7.4 Relay characteristics ................................................................. 8
7.5 General characteristics ............................................................. 8
9. TROUBLESHOOTING ……..................................................................... 11
Page 4

1. INTRODUCTION

The LD61R level switch is based on the variation of the natural resonant frequency of a vibrating fork, when it is wetted by a liquid.
This variation is detected by the internal electronics and is used to determine the state of the output .

2. RECEPTION

The LD61R level switches are supplied ready for installation and operation.
They are supplied packaged for their protection during storage and haulage.
To handle the detectors, they should always be held by the head, never by the vibrating fork.
The vibrating fork should not be modified or bent, as this can damage the detector beyond repair.

3. INSTALLATION

The LD61R level switch can be mounted in any position. The more usual insertion points are shown in the figure 1.
In positions 1 the LD61R acts as a maximum level switch. In positions 2 the LD61R acts as a minimum level switch and in position 3 it acts as an empty pipe detector (for example, to protect a pump).
If the LD61R is installed in a horizontal position it is recommended to install it with the tines vertical to avoid accumulation of substances, especially in the case of high viscosity liquids.
To indicate the position of the tines the LD-61 has a round mark on two of the flats of the nut. These marks should be in a vertical position when the detector is installed in a horizontal position.
Punto de
Positioning mark
posicionamiento
Figure 1
4
Page 5
In the same way, when the switch is installed where there is flow, the position must be taken into account. The flat part of the tines must be aligned parallel with the flow direction as shown in figure 1.
If the viscosity is high, the tines must be kept away from other objects (such as the wall of the tank). In these cases it is preferable to install a longer detector (figure 2).
It is recommended that the cables go downwards from the cable glands .
Figure 2

4. ELECTRICAL CONNECTION

For the electrical connection, the LD61R has a screw terminal strip.
For the electrical installation it is recommended to use multiple conductor cables with individual cable sections in the order of 0.25 to 0.5 mm connect. It is better to maintain the cables with mains voltage (power supply) separated from the cables with low level signals.
Before starting the installation, check that the cable glands are the right size for the cables to be used, this will guarantee the instrument will stay watertight. The PG11 cable glands used are for cables with outside diameters between 6 mm and 10 mm.
Peel the outside insulation to free the inner cables. It is recommended to tin the ends of the wires to avoid loose ends. Pass the cables through the cable glands and screw down in the corresponding positions of the terminal strip. Once the wiring is finished make sure that the cables are well gripped by the cable glands to maintain the degree of protection.
IMPORTANT NOTE: In order to comply with the electrical safety requirements as per EN­61010-1 (IEC 1010-1), the installation of the equipment must take into account the following:
 A mains switch must be provided to disconnect the equipment. This switch must
be marked as the disconnecting device for the equipment and be within easy reach of the operator.
 The mains supply must have an earth line.
 The housing must not be opened when the instrument has mains supply
connected.
2
in order to make it easier to
5
Page 6
Before installing the equipment, check that the supply voltage available in the installation is compatible with the marked on the label of the instrument.
To help in the connecting of the equipment, the description of the terminals is marked on the printed circuit next to the terminal strip.

4.1 Power supply connection

Terminal Power supply AC Power supply DC
Earth Earth
N Neutral (-)
L Live (+)
IMPORTANT NOTE: Power supply terminals are different from the ones in the previous
version. Be sure to connect properly the power supply as in the following image:

4.2 Relay output connection

Terminal
NO Normally open
NC Normally closed
C Common
6
Page 7

5. OPERATION

The LD61R can be installed as a detector of minimum or maximum level. To do that, it has a switch inside the cover. Furthermore, the instrument has a LED that indicates the relay ststus.
Change of polarity switch (maximum level)
Change of polarity switch (minimum level)
With the switch in the maximum level position, the common contacts of the relays are in “normally open” when the detector is not in contact with the liquid (level below the detector), and change to “normally closed” when the detector is submerged. The LED changes from green to red.
With the switch in the minimum level position, the common contacts of the relays are in “normally open” when the detector is in contact with the liquid (level above the detector), and change to “normally closed” when the liquid falls below the detector. The LED changes from green to red.

6. MAINTENANCE

The life of the vibrating fork depends basically on the abrasive characteristics of the product used.
The LD-61 level switch only needs maintenance when the product, which can adhere to the tines, will not let the detector change to a non detection state. In these cases the tines must be cleaned.

6.1 Cleaning

To clean the tines, a brush suitable for removing the incrusted product, must be used . You should not try to knock off the adhered product from the tines.
During the cleaning process, be careful not to apply force against the tines as this can bend them and damage the detector beyond repair.

6.2 Fuse

In the event that the fuse blows, this should be replaced with a slow blow “T” fuse, size Ø5 x 20 mm and 250 mA rating.
7
Page 8

7. TECHNICAL CHARACTERISTICS

7.1 Materials:
Fork: EN 1.4404 (AISI 316L). Others on demand. Housing: Polycarbonate.
7.2 Process connection:

7.3 Power supply

24 ... 250 VAC / VDC Power consumption: ≤ 1 W

7.4 Relay characteristics

Number of contacts and type: Double pole, double throw Maximum switching current: 3 A Maximum switching voltage: 220 VDC, 250 VAC Maximum switching power: 60 W, 125 VA Mechanical resistance: 15.0 x 10

7.5 General Characteristics

Viscosity: up to 10.000 cSt Density: > 0,6 kg / l Switching time: ≈ 1s Hysteresis
: ± 2 mm with H2O
Process temperature (liquid): -30 ... 115ºC Ambient temperature: -5 ... 40ºC Standard pressure: PN25 Degree of Protection: IP67
Threads G1 (BSP), 1” NPT, DIN 11851, Clamp ISO 2852 fittings.
6
operations
Conforms with the 2006/95/CE Directive
Conforms with the 2002/96/CE Directive
Conforms with the 2004/108/CE Directive
Conforms with the 97/23/CE Directive
This equipment is considered as being a pressure accessory and NOT a safety accessory as defined in the 97/23/CE directive, Article 1, paragraph 2.1.3.
8
Page 9

8 DIMENSIONS

G1 or 1” NPT thread
9
Page 10
LD61TF/AISI316L SANITARY CONNECTION TF DIN405 Rd52 x 1/6
CLAMP-ISO2852 DN33,7/50,5 and 38/50,5
10
Page 11
LD61R ML

9 TROUBLESHOOTING

Problem Probable cause Remedy
The instrument is always detecting product
The detector does not vibrate and the LED inside the cover is off
The detector vibrates but does not change its state
The detector changes its state without presence of fluid
Rest of product adhered on the fork tines.
Power supply not correct.
Fused fuse. Change the fuse (250 mA T).
Damaged instrument. Contact with technical service
Interferences in the installation Contact with technical service
11
Clean softly the fork with a brush, avoiding to press on the tines.
Verify that the power supply cables are well connected to the terminal strip and that the voltage between them is correct.
Page 12
WARRANTY
Tecfluid S.A. guarantees all the products for a period of 24 months from their sale, against all faulty materials, manufacturing or performance. This warranty does not cover failures which might be imputed to misuse, use in an application different to that specified in the order, the result of service or modification carried out by personnel not authorized by Tecfluid S.A., wrong handling or accident.
This warranty is limited to cover the replacement or repair of the defective parts which have not damaged due to misuse, being excluded all responsibility due to any other damage or the effects of wear caused by the normal use of the devices.
Any consignment of devices for repair must observe a procedure which can be consulted in the website www.tecfluid.com, “After-Sales” section.
All materials sent to our factory must be correctly packaged, clean and completely exempt of any liquid, grease or toxic substances.
The devices sent for repair must enclose the corresponding form, which can be filled in via website from the same “After-Sales” section.
Warranty for repaired or replaced components applies 6 months from repair or replacement date. Anyway, the warranty period will last at least until the initial supply warranty period is over.
TRANSPORTATION
All consignments from the Buyer to the Seller´s installations for their credit, repair or replacement must always be done at freight cost paid unless previous agreement.
The Seller will not accept any responsibility for possible damages caused on the devices during transportation.
12
Loading...