August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0vii
Page 8
Table of Contents
viiiGenesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 9
1About This Manual
1 – About This Manual
Purpose of This
Chapter
IntroductionThe structure and contents of this manual were compiled based on the results of a
ApplicabilityThe instructions given in this document are applicable for Genesis Instruments type …
Purpose of This
Document
This chapter points out the purpose of the manual, specifies the instruments the manual
deals with and for whom the manual is intended. Furthermore, it contains a list with
relevant reference documents, explains the symbols, conventions and abbreviations
used and offers general information.
worldwide survey between Tecan field service engineers. We hope the manual fits your
requirements and would be happy to get any feedback. Just contact the documentation
service group of Tecan Schweiz AG.
•Genesis RSP
•Genesis RWS Logistics
•Genesis RWS Assay
•Genesis RMP
This document serves as a reference guide for Tecan trained and authorized personnel
only. It provides all relevant information for installation and servicing of the instrument –
with exception of information covered by the respective reference documents (see
section ‘Reference Documents’ later in this chapter).
Target GroupsThis manual addresses the following target groups:
Target groupDescriptionTasks/responsibilities
Field service
engineer
System integratorA legal person (e.g. a RO within
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.01 – 1
Experienced field service
engineer who has visited a Tecan
service training and has been
authorized for servicing the
Genesis RSP, RMP and RWS
instruments.
Tecan group or a distributor) who
acts in a legal sense between
Tecan Schweiz AG
(manufacturer of sub-assembly)
and end-user.
•Site inspection
•Installation
•Commissioning
•Site acceptance
•Maintenance
•Repair
•Calibration
•Conformity assessment
and CE marking (see
Section 4.5).
Page 10
1 – About This Manual
Reference
Documents
The following documents are helpful for installation and servicing of the instrument.
They are not enclosed with this manual, nevertheless they are part of the service
documentation.
Document typeValid for …/document titleDoc ID
Operating ManualGenesis RSP and NPS
Genesis RMP
Genesis RWS
Carousel
Columbus Washer
Sunrise Reader
Spectra & Rainbow Readers
Instrument
Software Manual
Installation ManualLower DiTi eject option 2
Genesis InstrumentDoc ID 390 791
Low volume option 1: tube fastening
Access option/Signal lamp (RSP, RWS)
Doc ID 390 783
I 119 100
Doc ID 391 197
Doc ID 391 209
I109004
I137301
I139003
Doc ID 391 276
Doc ID 391 228
Doc ID 392 330
Door lock V2 (RMP)
Alarm device V2 (RMP)
Form or ChecklistDecontamination Declaration
Genesis Maintenance and Service
Logbook, including:
•Installation Qualification
•Operation Qualification
•Acceptance Protocol
•Daily/Weekly Maintenance Checklist
•Preventive Maintenance Checklist
•Service Checklist
Furthermore, you may find following Application Software Manuals useful:
Area of applicationApplication softwareDoc ID
DiagnosticsTOPS
Doc ID 391 260
Doc ID 391 255
Doc ID 390 901
Doc ID 390 924
Doc ID 391 180
Doc ID 391 182
Doc ID 391 825
Doc ID 391 193
Doc ID 391 181
Doc ID 391 183
I 117 578
Logic
Life ScienceFACTS
Gemini
1 – 2Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Doc ID 391 110
Doc ID 391 252
Doc ID 391 201
Page 11
1 – About This Manual
Symbols Used in
This Manual
Instrument
Overviews
Safety Symbols
The safety symbols are explained in Chapter 2, ‘Safety’.
Useful Notes
Useful notes appear as follows:
Note
Gives helpful information about the equipment or regarding proceedings, tentative
clarifications etc.
In Chapter 8, ‘Instruments at a Glance’ you will find useful illustrations of the
instruments giving an overview of the main components.
On the detail drawings and the text passages in this document (mainly in Chapter 6,
‘Replacement of Spare Parts, Repair’), one main part is always designated with the
identical item number – for example (8) for LiHa – as on the relevant illustration in
chapter 8. Item numbers of their components consist always of the main part item
number, a dash and the item number of the component – for example (8-04) for LiHa
backplane.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.01 – 3
Page 12
1 – About This Manual
1 – 4Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 13
2Safety
2 – Safety
Purpose of This
Chapter
Significance of
These Safety
Instructions
This chapter contains specific rules of behavior and warnings from hazards with regard
to installation, setup, maintenance and repair of the Genesis RSP, RMP or RWS
instruments.
The safety of users and personnel can only be ensured if these safety instructions and
the safety-related warnings in the individual chapters are strictly observed and followed.
Therefore, the Service Manual must always be available to all persons performing the
tasks described herein.
In addition to the safety instructions given in this Service Manual, the safety instructions
pointed out in the Operating Manuals of the Genesis RSP, RMP or RWS instruments
apply as well.
2.1User Qualification
FSE Authorization The field service engineers (FSE) are specially trained personnel. Exclusively FSEs are
entitled to perform the maintenance and service work described in this Service Manual.
The manufacturer Tecan authorizes the FSEs if they fulfill the following particular
qualifications:
WARNING
Symbols
•They must have received appropriate service and operator training from Tecan.
•They must be familiar with the good laboratory practice guidelines.
•They must have read and understood the instructions in this Service Manual.
2.2Notices and Symbols
2.2.1Warning Notices Used in This Manual
The symbols used for safety-related notices have the following significance:
WARNING
Generally, the triangular warning symbol indicates the possibility of personal injury or
even loss of life if the instructions are not followed.
Whenever possible, the symbol indicates the hazard a person is exposed to more
specifically. The symbols used in this Service Manualhave the following significance:
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.02 – 1
Page 14
2 – Safety
WARNING
Biological hazard
WARNING
Chemical hazard
WARNING
Radioactive radiation
WARNING
Electrical danger
WARNING
Laser radiation
ATTENTION
Symbols
ATTENTIONS appear as follows:
ATTENTION
With the general “STOP” symbol, ATTENTIONs indicate the possibility of equipment
damage, malfunctions or incorrect process results, if instructions are not followed.
Other symbols indicate the significance of the ATTENTION more specifically.
ATTENTION
Damage to electronics by electrostatic discharge.
Always follow ESD safety practices.
ATTENTION
Disturbance of functions by electromagnetic RF waves.
Do not use a cellular phone.
2.2.2Warning Notices Attached to the Product or Its Surroundings
Cellular Phones
Prohibited
Do not use cellular phones in the proximity of the instrument.
This symbol is attached to the safety panel of the instrument. The symbol must also be
attached to the laboratory door.
Damaged or fallen off symbols (notices or stickers) must be replaced immediately.
2 – 2Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 15
2 – Safety
2.3Use of the Product
Intended UseThe Genesis RSP, RMP or RWS instruments are to be applied and used exclusively in
the following application fields. Any other use is considered improper and is strictly
forbidden.
•The Genesis RSP is intended for liquid pipetting for generally known laboratory
methods according to common safety precautions.
•The Genesis RMP is intended for fully automated processing of 96-well microplate
based ELISA (Enzyme Linked Immuno Sorbent Assay) and ELISA-like tests,
starting from sample pipetting and ending with result reading.
All ELISA test procedures have to be validated prior to performing routine tests.
•The Genesis RWS Logistics is intended for pipetting tasks as well as storage,
identification and transfer of microplates. It is to be applied for research only.
The Genesis RWS Logistics is intended to be part of a liquid handling system and
as such installed and put into operation by a trained Tecan or Tecan authorized
system integrator.
•The Genesis RWS Assay is intended for processing of microplate based tests,
including liquid handling, incubation, wash, read and data acquisition. It is to be
applied for research only.
The Genesis RWS Assay is intended to be part of a liquid handling system and as
such installed and put into operation by a trained Tecan or Tecan authorized system
integrator.
All Genesis instruments are intended for indoor operation and storage only.
Improper UseDue to their open architecture, the Genesis RWS Assay and Logistics are not intended
for clinical and diagnostic applications. These applications would be carried out by less
qualified people, exposing them to dangerous liquids in case of instrument malfunction.
Incorrect use of the Genesis RMP may lead to false test results and may cause
exposure of the operator to potentially dangerous compounds.
Do not attempt to use liquids with undissolved particles as this could result in liquid not
being dispensed due to clogged tips.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.02 – 3
Page 16
2 – Safety
2.4Product Safety
PrincipleThe Genesis RSP, RMP or RWS instruments are designed and built in accordance with
the present state-of-the-art technology and the recognized technical safety regulations.
Nevertheless, risks to users, property, and the environment can arise when the
instrument is used carelessly or improperly.
The manufacturer has determined all residual dangers emanating from the instrument
in all life phases and from the process.
Appropriate warnings in the Operating Manuals of the Genesis RSP, RMP or RWS
instruments and in this Service Manual serve to make the user alert to these residual
dangers.
2.4.1Instrument-Related Hazards and Safety Measures
Pay attention to the following safety notices:
WARNING
Electrical shock hazard.
Switch the instrument off and disconnect from mains whenever no power is required to
perform service tasks.
ATTENTION
Crimping of the tubing between top cover and case possible.
Secure the top cover from falling down during inspection work.
ATTENTION
Damage to the electronic boards due to electrostatic discharge (ESD).
Always wear a wrist strap when handling the boards.
2.4.2Other Hazards and Safety Measures
WARNING
•Chemical, biological and radioactive hazards can be associated with the
substances used or the samples processed with the Genesis RSP, RMP or RWS
instrument.
•The same applies to waste disposal.
Always be aware of possible hazards associated with these substances.
Request a filled out and signed Decontamination Declaration prior to performing any
maintenance or repair tasks.
2 – 4Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 17
2.4.3Safety Elements
2 – Safety
Removal of
Protective
Devices
When to
Decontaminate
The protective and safety devices installed on the Genesis RSP, RMP or RWS
instrument must be neither removed nor disabled during operation.
If such elements were removed, e.g. for maintenance work, operation may only be
resumed when all protective and safety devices have been completely installed and
checked.
2.5Decontamination
Apart from regular decontamination, the user must thoroughly decontaminate the
instrument according to standard laboratory regulations in the following cases:
•Before any maintenance or service work is performed on the instrument
•Before a Tecan field service engineer (FSE) performs any in-site work on the
instrument
•Before the instrument is returned to Tecan (e.g. for repair)
•Prior to storage of the instrument
•Prior to disposal of the instrument or parts of it
•Generally before the instrument or parts of it leave the user’s site.
Decontamination
Method
Decontamination
Declaration
The decontamination method must be adapted to the respective application and the
substances associated with it. The user takes the full responsibility for the appropriate
decontamination of the entire equipment.
WARNING
Biological or chemical hazard and/or radioactive radiation.
•Contamination hazard due to parts of the instrument which are not completely
decontaminated.
•Mind that not only the parts having direct contact with chemicals or biological
material must be treated, but also the tubing system as well as the whole upstream
equipment.
Before a Tecan FSE carries out any work on the instrument, or before the instrument is
returned to Tecan, the owner of the instrument must confirm in writing that the
decontamination has been performed properly and in accordance with good laboratory
practice guidelines. For this, the owner must enclose a Decontamination Declaration
form (Doc ID 390 901), which can be provided by Tecan.
Note: Tecan will refuse any instrument or a part of it, if the decontamination form is not
filled in and duly signed.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.02 – 5
Page 18
2 – Safety
2.6General Safety Rules
Legal Regulations Legal regulations, such as local, state and federal laws which prescribe the use or
application as well as the handling of dangerous materials in connection with the
Genesis RSP, RMP or RWS instrument must be strictly followed.
Duty of
Maintenance and
Care
Appropriate
Behavior
Spare Parts to Be
Used
ModificationsModifications to the Genesis RSP, RMP or RWS instrument are only permitted with the
The user is responsible for ensuring that the Genesis RSP, RMP or RWS instrument is
operated in proper condition only, and that maintenance, service, and repair jobs are
performed with care and on schedule.
Use exclusively tools suitable for the respective work sequence. Ensure that the tools
are in sound condition.
Handling and disposing of waste has to be in accordance with all local, state and federal
environmental, health, and safety laws and regulations.
Use only genuine spare parts for maintenance and repair.
written approval of the manufacturer. Modifications and upgrades shall only be carried
out by an authorized field service engineer. The manufacturer will decline any claim
resulting from unauthorized modifications.
2 – 6Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 19
3Product Description
3 – Product Description
Purpose of This
Chapter
Chapter OverviewThis chapter consists of the following sections:
This chapter summarizes the technical data of the Genesis instruments and contains an
overview of requirements, hardware, software and firmware compatibilities.
Note
For detailed information concerning optional modules as reader, washer, carousel etc.,
refer to their respective manuals.
SectionTitlePage
3.1Technical Data and Requirements3 – 2
3.1.1Dimensions and Weights3 – 2
3.1.2Supply Ratings3 – 6
3.1.3Environmental Conditions3 – 7
3.1.4Computer and Software Requirements3 – 8
3.1.5Additional Data for System Modules3 – 8
3.2Compatibility Matrix3 – 9
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.03 – 1
Page 20
3 – Product Description
3.1Technical Data and Requirements
3.1.1Dimensions and Weights
RSP Instrument
Dimensions
830 mm
(32.7")
Genesis RSP 100: 1051 mm (41.4")
Genesis RSP 150: 1426 mm (56.1")
Genesis RSP 200: 2026 mm (79.8")
1200 mm
(47.25")
1000 mm
(39.4")
RMP Instrument
Dimensions
front view
Genesis RSP 100: 1501 mm (59.1")
Genesis RSP 150: 1876 mm (73.9")
Genesis RSP 200: 2476 mm (97.5")
top view
RMP 100: 1405 mm (55.3")
RMP 150: 1780 mm (70.1")
RMP 200: 2380 mm (93.7")
805 mm (31.7")
780 mm (30.7")
1230 mm (48.4")
left side view
840 mm
(33.1")
1000 mm (39.4")
870 mm (30.7")
1230 mm (48.4")
RMP 100: 1856 mm (73.1")
RMP 150: 2230 mm (87.8")
RMP 200: 2830 mm (111.45")
840 mm (33.1")
3 – 2Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 21
RWS Logistics
Workstation
Dimensions
830 mm
(32.7")
Logistics WS 100: 1051 mm (41.4")
Logistics WS 150: 1426 mm (56.1")
Logistics WS 200: 2026 mm (79.8")
3 – Product Description
1200 mm
(47.25")
RWS Assay
Workstation
Dimensions
830 mm
(32.7")
Front View
Logistics WS 100: 1501 mm (59.1")
Logistics WS 150: 1876 mm (73.9")
Logistics WS 200: 2476 mm (97.5")
Top View
Assay WS 150: 1796 mm (70.7")
Assay WS 200: 2396 mm (94.4")
780 mm (30.7")
840 mm
(33.1")
Left Side View
1200 mm
(47.25")
Front View
Assay WS 150: 1970 mm (77.6")
Assay WS 200: 2570 mm (101.2")
Top View
840 mm
(33.1")
780 mm (30.7")
Left Side View
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.03 – 3
Page 22
3 – Product Description
Worktable
Dimensions
Z
Y
Genesis RSP
Accessible X-range (X-travel)mm
Accessible Y-range (Y-travel)mm
Grid positions on worktablepcs.304569
Genesis RMP
Accessible X-range (X-travel)mm
X
inch
inch
inch
RSP 100RSP 150RSP 200
745
29.3
418
16.5
RMP 100RMP 150RMP 200
550
21.6
1120
44.1
418
16.5
925
36.4
1720
67.7
418
16.5
1525
60
Accessible Y-range (Y-travel)mm
inch
Grid positions on worktablepcs.243963
Genesis RWS
RWS Logistics 100 150 200
Accessible X-range (X-travel)mm
inch
Accessible Y-range (Y-travel)mm
inch
Grid positions on worktablepcs.304569
3 – 4Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
418
16.5
745
29.3
418
16.5
418
16.5
1120
44.1
418
16.5
418
16.5
1720
67.7
418
16.5
Page 23
3 – Product Description
RWS Assay 150 200
Accessible X-range (X-travel)mm
Accessible Y-range (Y-travel)mm
Grid positions on worktablepcs.4367
WeightsGenesis RSP
Platform with 4-tip/8-tip LiHa 2kg
RoMa 2kg
PosID 2 optionkg
MPO or FWOkg
inch
inch
lb.
lb.
lb.
lb.
925
36.4
418
16.5
RSP 100RSP 150RSP 200
96/103
212/227
13
18
4.4
6
8
2
115/122
253/269
6
13
11
24
2
4.4
1525
60
418
16.5
165/172
364/379
13
14
31
4.4
6
2
Packingkg
lb.
Genesis RMP
RMP 100RMP 150RMP 200
Platform with LiHa 2, RoMa 2 and
PosID 2
FWOkg
Packingkg
Genesis RWS
Logistics 100 150 200
Platform with 4-tip/8-tip LiHa 2 and
RoMa 2
kg
lb.
lb.
lb.
kg
lb.
35
77
166
366
4.4
104/111
229/245
2
47
104
188
414
4.4
43
95
123/130
271/287
156
241
531
2
218
173/180
381/397
71
2
4.4
99
PosID 2 optionkg
lb.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.03 – 5
18
8
11
24
14
31
Page 24
3 – Product Description
Logistics 100 150 200
MPO or FWOkg
lb.
Packingkg
lb.
4.4
35
77
2
4.4
47
104
2
4.4
71
156
Assay 150 200
Platform with 4-tip/8-tip LiHa 2 and
RoMa 2
PosID 2 optionkg
MPO or FWOkg
Packingkg
kg
lb.
lb.
lb.
lb.
131/138
289/304
11
24
4.4
50
110
181/188
397/414
14
31
2
4.4
74
163
2
2
3.1.2Supply Ratings
Genesis RSP
Genesis RMP
Supply ratingsRSP 100RSP 150/200
Primary voltage V AC100–240100–240
Frequency Hz50/6050/60
Power VA800 800
Module type
(for power module supplied as of September 2001)
PM 1PM 4
Supply ratingsRMP 100/150/200
Primary voltage V AC100–240
Frequency Hz50/60
Power VA1000
3 – 6Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 25
Genesis RWS
Supply ratingsRWS 100
3 – Product Description
RWS 150/200
Operating
Conditions
Primary voltage V AC100–240100–240
Frequency Hz50/6050/60
Power VA800 1200
Module type
(for power module supplied as of September 2001)
3.1.3Environmental Conditions
The Genesis RSP, RMP or RWS instruments are intended for indoor operation and
storage only.
Operating temperature °C
Operating humidity
relative (non condensing)
at 30 °C/86 °F or below
Logistics
Assay/Logistics
VAPM 1PM 2
RSPRMPRWS
°F
15–32
59–90
18–30
65–86
%30–8030–8030–80
18–30
65–86
Storage
Conditions
Pollution degree 222
Over voltage category class222
Protect the instrument against dust and debris with a cover.
Recommendation: store the instrument in its original packaging. Store all manuals and
the Service and Maintenance Logbook with the instrument.
RSPRMPRWS
Storage temperature °C
°F
Storage humidity
%30–8030–8030–80
1–60
34–140
1–60
34–140
1–60
34–140
relative (non condensing)
at 30 °C/86 °F or below
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.03 – 7
Page 26
3 – Product Description
3.1.4Computer and Software Requirements
Computer
Hardware
SoftwareIt is strongly recommended to use the latest software versions. Please contact your
Refer to the Genesis Instrument Software Manual for details on minimum computer
requirements.
nearest Tecan representative for more information.
SoftwareRSPRMPRWS
Logistics
Operating SystemX
X
Genesis Instrument SoftwareXXXX
Logic
(Clinical Diagnostics)
Gemini (Life Science)XXX
TOPS
(Operating SW)
FACTS
(Event scheduling SW)
X
X
X
X
X
X
XX
RWS
Assay
X
X
TAURUS
(Database for Clinical
Diagnostics)
Magellan
(Data reducing SW)
X = required or available for respective instrument
X
3.1.5Additional Data for System Modules
LiHa 2•4 or 8 tips (diluters) acting independently
•Variable tip spacing: 9–38 mm
•Arm movement speed: 300–500 mm/s
•Positioning precision: better than 0.4 mm
RoMa 2•Transports any rack in MP-format
X
•Arm movement speed: 400 mm/s
•Rotation: 270°
3 – 8Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 27
•Lifting force: up to 4 N (transportable mass up to 0.4 kg
•Software adjustable gripper force: 1–4 N
•Gripper space range 55 to 140 mm
PosID 2•Laser class 2
•Wave length: 670 nm
•Distinction between sample barcode/no barcode/no tube
•Reads vertically and horizontally
•Movement speed: 400 mm/s
•Barcode types:
Code 39 (length 0...32)
Code 39 full ASCII
Codabar (length 0...32)
Code 128 (length 0...32)
2/5 Interleaved (length 0...30, even only)
UPC-A (length not setable: 12)
UPC-E (length not setable: 8)
EAN-8 (length not setable: 8)
EAN13 (length not setable: 13)
3 – Product Description
3.2Compatibility Matrix
Tecan customer support provides a compatibility maxtrix which displays the most
common configurations that may be installed presently on your customer’s systems.
Please request your up-to-date copy via expertline-eu@tecan.com.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.03 – 9
Page 28
3 – Product Description
3 – 10Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 29
4Transport and Installation
4 – Transport and Installation
Purpose of This
Chapter
Chapter OverviewThis chapter consists of the following sections:
This chapter summarizes the procedures needed for installation of a Genesis RSP,
RMP or RWS instrument from site inspection to site acceptance.
SectionTitlePage
4.1Site Inspection4 – 1
4.2Transport4 – 2
4.3Installation4 – 3
4.3.1Unpacking and Positioning4 – 3
4.3.2RMP Specific Installation Tasks4 – 3
4.3.3RWS Specific Installation Tasks4 – 13
4.3.4Further Options and Modules for RSP/RWS4 – 14
4.3.5Computer, Software4 – 15
4.4Commissioning4 – 16
4.5Conformity Assessment and CE Marking for RWS4 – 18
4.6Site Acceptance4 – 18
4.1Site Inspection
Site Requirements Check the customer’s site for site requirements suitable concerning the respective
instrument:
Notice the dimensions and weights of the different instruments (see section 3.1.1).
Delivery Route
•Loading dock present?
•Load capacity and size of elevator sufficient?
•Door and entrance openings, passageway corners: big enough?
Environmental Requirements
•Check according to Section 3.1.3, ‘Environmental Conditions’.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.04 – 1
Page 30
4 – Transport and Installation
Space
•Enough space to place instrument and possible extensions?
•Enough space to open safety panels and access doors of the instrument?
•Enough walking space around the instrument?
•Space for placing system liquid/waste bottles?
•Enough space for the control computer?
Supply
Notice the power supply requirements (refer to section Section 3.1.2, ‘Supply Ratings’)
Check presence of power sockets for
•Instrument
•Computer
•Other modules
Date of Delivery•Determine the date of delivery.
4.2Transport
General Notes•The Genesis RSP, RMP and RWS are precision instruments. Handle with care. Do
not expose to excessive shock.
•Always use original packaging for shipping the instrument.
Visual Check•Visually check the crate(s) for damage sustained during transportation. If such are
found, the customer must immediately file a complaint with the transport agency.
•Tecan instrument packaging is designed to avoid damage during transportation.
Please inform Tecan about all occurrences of transportation damages. Such
information is important for further improvements.
4 – 2Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 31
4.3Installation
4.3.1Unpacking and Positioning
Transport handles
Figure 4-1 Transport handles – seen from below
•For exact positioning: always lift the instrument; do not shift it.
4 – Transport and Installation
1Unpack all components and visually
check for any damage. Leave transport
paddings and moorings in place.
2Check if the shipment is in compliance
with the packing list.
3Lift the instrument onto the assigned
work bench and place it as required.
•Because of the considerable weight of the instrument (see Section 3.1.1,
‘Dimensions and Weights’), up to four people are required for lifting the instrument.
The center of gravity is at the rear right side of the instrument. Make sure that the
two strongest persons lift the rear of the instrument.
•Use the optional transport handles (see Figure 4-1). Never lift the instrument by the
lateral worktable covers. The screws holding the covers are not designed to support
the instrument weight.
4Remove all transport paddings and moorings.
Packing material shall be stored for possible shipment of parts to be repaired by
manufacturers.
4.3.2RMP Specific Installation Tasks
Add-On ModuleOnly the RMP 200 has a separate add-on module. RMP 100 and 150 consist of one-
piece frames. Therefore, following procedure applies to RMP 200 only.
Refer to Figure 4-2, ‘RMP 200’ and Figure 4-3, ‘Add-on module assembly’.
1Place a spirit level onto the RMP worktable and level horizontally by means of the
supports below the instrument frame.
2Connect the add-on module (15) to the right side of the RMP 200.
There are two positioning pins (15-01) in the add-on module frame and two
positioning holes in the instrument frame.
While positioning the add-on module, lead the cables for the mains power switch
and the control lamp trough the upper left part of the add-on module.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.04 – 3
Page 32
4 – Transport and Installation
Control lamp cable
Figure 4-2 RMP 200
3Turn the two left supports (15-02) to their topmost positions, so that they do not
touch the work bench. The add-on module is now sitting on its two right supports
with the two positioning pins connected to the RMP 200.
Mains power switch cable
Uniport cable
Incubator/shaker cable
4Turn the two left supports down until they touch the work bench.
5Insert the module worktable (15-03).
15
15-01
15-02
15-03
Figure 4-3 Add-on module assembly
15Add-on module
15-01Positioning pin
15-02Support
15-03Module worktable
4 – 4Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 33
4 – Transport and Installation
6Insert the fixing screws and the two
distance washers (15-04) according to
Figure 4-4. Tighten the fixing screws.
7Check horizontal alignment with a spirit
M4x12
6x M4x20
15-03
4x M4x16
level; if necessary, adjust by means of
M4x6
M4x6
15-04
Figure 4-4 Module worktable
the two right supports (15-02).
15-03Module worktable
15-04Distance washer
Set the mains power switch to off and disconnect the mains power connection of
the instrument before performing any further work!
Observe precautions for handling electrostatic discharge sensitive devices.
Wear a wrist strap.
#3
#4
8Open the right access door (6) and
install the mains power switch (33-02)
A
33-07
into the add-on module.
Note the correct orientation of the
switch according to Figure 4-5!
9Connect the mains power cable (33-07)
to the power switch (33-02) according
to the installation diagram (33-06) on
33-02
correct
P2P1
view A
33-06
33-05
the housing.
33-08
33-06
Washer
Figure 4-5 Electric connections for RMP
33-07
33-05
10 Fix the earth cable (33-05) to the add-
on module frame as shown in Figure
4-5.
11 Connect the control lamp cable (33-08)
to the power control lamp in the add-on
module.
12 Connect the Uniport cable coming from
the Optibo to the Uniport (see Figure
4-6, ‘Cable connections on Supervisor
and Uniport board’).
The Uniport is located in the rear part of
the add-on module.
33-02Power switch
33-05Earth cable
33-06Installation diagram
33-07Mains power cable
33-08Control lamp cable
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.04 – 5
Page 34
4 – Transport and Installation
Alarm device J25
2nd RT incubator
J16, J17, J12
RT incubator
Balance
OEM
Washer
Reader
from Optibo
Figure 4-6 Cable connections on Supervisor and Uniport board
Supervisor board
Waste (LICOS)
System Liquid (LICOS)
J5 Niveau (H-sensor)
Uniport board
33-09
Figure 4-7 Connector sheet
13 Fix the connector sheet (33-09).
The connector sheet guarantees a
proper electromagnetic compatibility
(EMC).
14 Insert and fix the module cover (15-05).
15-05
Figure 4-8 Module cover
4 – 6Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 35
Reader (Option)
19
19-02
16-06
15-07
4 – Transport and Installation
1Remove the frame cover (15-07) from
the add-on module in order to get
easier access.
2For RMP 200 only:
Lock the adjusting bracket (16-06) in
designated place (positioning pins)
onto the module worktable (15-03) and
fix it with the knurled screws.
3Place the reader (19) onto the module
worktable.
4Set selection switch (19-02) into correct
position, according to the label (19-01)
on the adjusting bracket:
•TOPS V3.0 and earlier: set switch
to “RMP mode”.
•TOPS V4.0 and later: set switch to
“Stand-alone mode”.
15-06
15-03
Figure 4-9 Reader assembly
19-03
3x M4x12
19-01
5Put a spirit level in X-direction onto the
reader housing. If necessary, level
horizontally by means of the adjusting
wheel (15-06).
6Connect the reader cable (19-03) to the
connection cable on the Uniport (TOPS
V3.0 and earlier) or to the second PCcom port (TOPS V4.0 and later).
15-03Module worktable
15-06Adjusting wheel
15-07Frame cover
16-06Adjusting bracket
19Reader
19-01Label
19-02Selection switch
19-03Reader cable
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.04 – 7
Page 36
4 – Transport and Installation
Washer (Option)
16-01
16
1If not yet done: remove the frame cover
(Figure 4-9, 15-07) from the add-on
module in order to get easier access.
2For RMP 200 only:
Screw the positioning rack (16-01) to
the adjusting bracket (16-06).
16-02
16-05
Figure 4-10 Washer assembly
16-06
16-04
3x M4x10/
washer
16-03
3Place the washer (16) on the
positioning rack.
4Check horizontal alignment with a spirit
level. If necessary, adjust by means of
the adjusting wheel (16-02).
16Washer
16-01Positioning rack
16-02Adjusting wheel
16-03Waste tubing positioning rack
16-04Waste tubing washer
16-05Washer cable
16-06Adjusting bracket
5Attach the washer cable (16-05) –
which is connected to the Uniport – to
the washer.
6Lead the waste tubing (16-03, 16-04)
through the module frame.
4 – 8Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 37
Heated Incubator
(Option)
4 – Transport and Installation
Installation of One Heated Incubator
1Connect the incubator cable (20-01) to
the incubator 1 (20-A) and the Optibo.
2Place the incubator 1 (20-A) into the
respective recesses on the module
worktable.
20-01
20-A
Figure 4-11 Heated incubator 1
4
5
3
6
2
7
1
8
0
9
F
A
E
B
D
C
20-A
3
2
1
0
F
E
20-B
20-01
20-02
3Check horizontal alignment with a spirit
level. If necessary, adjust by means of
the supplied special key (see 17-05 in
Figure 4-13).
20-AHeated incubator 1
20-01Incubator cable
Installation of Two Heated Incubators
4
5
6
7
8
9
A
B
D
C
1Connect the incubator cable (20-01) to
the incubator 1 (20-A) and the Optibo.
2Join the two incubators with the
connection cable (20-02).
3Change address switch of incubator 2
(20-B) from #2 (= standard setting) to
#3.
4Place the incubator 1 (20-A) into the
respective recesses on the module
worktable.
20-A
20-B
20-03
Figure 4-12 Installation of two heated incubators
5Insert adapter plate (20-03) beneath
incubator 1 onto the worktable.
6Place the incubator 2 (20-B) into the
recesses on the adapter plate.
7Check horizontal alignment with a spirit
level. If necessary, adjust by means of
the supplied special key (see 17-05 in
Figure 4-13).
20-AHeated incubator 1 (address #2)
20-BHeated incubator 2, with shaker
(address #3)
20-01Incubator cable
20-02Connection cable
20-03Adapter plate
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.04 – 9
Page 38
4 – Transport and Installation
Room
Temperature
Incubator
Installation of One RT Incubator
1Fix the mounting bracket (17-01) with
the knurled nuts to the module
worktable.
2Place the RT incubator 1 into the
recesses on the module worktable and
screw it to the mounting bracket.
17-02
17-01
17-03
17-03
17-02
17-A
17-04
17-01
17-A
17-B
M4x8/
washer
17-04
3Check horizontal alignment with a spirit
level. If necessary, adjust by means of
the supplied special key (17-05).
4Attach the incubator cable (17-02) –
which is connected to the Supervisor
board – to the incubator 1 (17-A).
17-ARoom temperature incubator 1
17-BRoom temperature incubator 2
17-01Mounting bracket
17-02Incubator cable RT1
17-03Incubator cable RT2
17-04Adapter plate
17-05Special key
Installation of Two RT Incubators
(Option)
1Place the adapter plate (17-04) onto
the worktable.
2Fix the mounting bracket (17-01) with
the knurled nuts to the module
worktable and the adapter plate.
17-B17-A
3Place the RT incubator 1 into the
recesses on the module worktable and
screw it to the mounting bracket.
4Place the RT incubator 2 into the
recesses on the adapter plate and
screw it to the mounting bracket.
5Check horizontal alignment with a spirit
level. If necessary, adjust by means of
the supplied special key (17-05).
6Attach the incubator cables (17-02,
17-05
17-03) – which are connected to the
Supervisor board – to the incubators.
Figure 4-13 Room temperature incubators
4 – 10Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 39
Wash Bottle Rack
18-01
18
4 – Transport and Installation
Recommendation: As the fixing screws
(18-01) might get into contact with spilled
wash liquid, grease them to prevent from
stucking.
1For RMP 200 only:
•Place the wash bottle rack (18) on
the module worktable and insert the
fixing screws (18-01), but do not
tighten yet.
•Mount the add-on module safety
panel.
•Make sure the wash bottle rack is
aligned parallel to the closed safety
panel.
•Tighten the fixing screws (18-01).
18
18-05
18-02
16
18-04
18-03
•Place the bottles (18-02) in the
wash bottle rack.
16Washer
18Wash bottle rack
18-01Fixing screws
18-02Bottle
18-03Wash liquid tubing
18-04Tubing clip
18-05Cable
2Connect the wash liquid tubing (18-03)
to the bottles and their respective
channels at the washer (16).
Make sure for each tubing that the
number printed on the wash bottle rack
corresponds with the one printed on the
rear of the washer.
3Fix the tubing into tubing clip (18-04).
4Attach the cable (18-05) marked
“Niveau” – which is connected to the
Supervisor board – to the wash bottle
rack.
5Install the frame cover (15-07) on the
15-07
Figure 4-14 Installation of wash bottle rack
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.04 – 11
add-on module.
Page 40
4 – Transport and Installation
LICOS
15-08
1For RMP 200 only:
•Insert the print cover (15-08) and fix
it with the knurled nuts.
•Place the system liquid container
(15-09) in the compartment and
connect the system liquid tubing
(white connector, leading to FWO)
and the LICOS tubing (blue
connector, leading to Supervisor
board) to the container.
•Lead the fill tubing into the system
liquid container.
LICOS
15-09
Waste
liquid
System
liquid
2Connect waste liquid tubing, leading to
the Supervisor board, to LICOS sensor
rod. Place LICOS sensor rod into the
waste container.
15-08Print cover
15-09System liquid container
Figure 4-15 Installation of LICOS
Alarm DevicePlease refer to Installation Instructions for Alarm Device V2, Doc ID 391 255.
4 – 12Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 41
Workstation
Extension
(RWS Assay)
4.3.3RWS Specific Installation Tasks
16
4 – Transport and Installation
The workstation extension is an option for
RWS Assay instruments only. It is used as
support for reader and washer.
19
24
12
Figure 4-16 RWS Assay workstation extension
25
25-01
1Install the workstation extension (25) at
the right side of the cut out worktable
(12).
2Check horizontal alignment with a spirit
level; if necessary, adjust by means of
the two right supports (25-01).
12Cut out work table
16Washer
19Reader
24Mounting assembly for washer
25Workstation extension
25-01Support
Variable
Extension Deck
(RWS Logistics)
The variable extension deck is an option for RWS Logistics instruments only. It is used
as support for Ultra reader and further external devices.
The variable extension deck can either be placed on the right or on the left side of the
instrument.
Reader (Option)1For RWS Assay:
•Place the reader (19) onto the workstation extension (25).
For RWS Logistics:
•Install the positioning plate on the variable extension deck.
•Place the Ultra reader onto the positioning plate.
2Connect the reader to the power supply and to PC-com port.
Washer (Option)For RWS Assay only.
1Install the mounting assembly for washer (24) onto the workstation extension (25).
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.04 – 13
Page 42
4 – Transport and Installation
2Place the washer (16) on top of the mounting assembly.
3Check horizontal alignment with a spirit level. If necessary, adjust by means of the
mounting assembly legs.
4Connect the washer to the power supply and to PC-com port.
5Connect wash liquid tubing and waste tubing to the washer and respective liquid
containers.
Heated Incubator
(Option)
4.3.4Further Options and Modules for RSP/RWS
Installation of One Heated Incubator
1Install adapter plate onto worktable.
2Place incubator into respective recesses on the adapter plate.
3Connect the incubator cable to the Optibo/Optibo Power.
4Check horizontal alignment with a spirit level. If necessary, adjust by means of the
supplied special key (see 17-05 in Figure 4-13).
Installation of Up To Four Heated Incubators
Corresponding to the workstation configuration, up to four incubators can be placed at
the rear of the worktable.
1Install each incubator as described under Installation of One Heated Incubator.
It depends on the application software, if it is also possible to connect maximal two
incubators in series.
2Assure proper address setting for each incubator in compliance with possibly other
options (Refer to Section 6.2.2 Jumper and Address Settings Overview).
Standard setting for one incubator is address #2.
Connecting
Options and
Modules
LICOS1Connect LICOS sensor tubing, leading to MPO board, to the LICOS sensor rods.
4 – 14Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
1Make sure all options and modules are compatible (Refer to Section 3.2
Compatibility Matrix).
2Assure proper address settings (Refer to Section 6.2.2 Jumper and Address
Settings Overview).
3Prepare all options and modules according to manufacturers instructions and
connect them to Optibo/Optibo Power or Uniport or CANDI board.
2Place LICOS sensor rods into respective containers:
•Tube 1: system liquid container
•Tube 2: waste container
3Place system liquid container on worktable level to avoid pressure differences within
tubing system.
Page 43
4 – Transport and Installation
Access Option/
Signal Lamp
4.3.5Computer, Software
Please refer to Installation Manual ID 392 330
1Place the computer on the left side of the instrument. Install the computer according
to manufacturer’s instructions.
2Make sure the instrument is switched off. Install communication cable between
computer (COM 2) and instrument (Optibo or Optibo Power).
3Install Genesis instrument software and – if necessary – perform firmware download
(refer to Genesis Instrument Software Manual Doc ID 390 791).
4Install respective application software.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.04 – 15
Page 44
4 – Transport and Installation
4.4Commissioning
Mechanical
Checks
Quick Jumper
Settings Test
Retightening
Screws
Manually check smooth movement of the LiHa 2 and – if installed – RoMa 2 and
PosID 2.
If there is any resistance in the movement, check for dirt residue and clean thoroughly.
Control CAN-bus resistance to check correct jumper settings (see Section 6.2.3, ‘CAN-
Bus Resistance Test’).
1Tighten all power cable screws on the
Optibo/Optibo Power (left access door).
2Tighten all ILID plug fixing screws.
Make Liquid
System Ready
for Use
Figure 4-17 Retightening screws
1Tighten tubing connections and the
syringe screw on diluters.
For these tasks, a better result is
achieved when the valves with syringes
are removed from the diluters.
2Tighten tubing connections on
distributors (top cover).
3Mount tips.
4Open worktable front cover (10),
remove sealing strip. Place wash
station on the worktable. Cut the
sealing strip at the appropriate length
and reinsert.
Figure 4-18 Diluter
Power-on1Connect the instrument to the mains.
2Set mains power switch to on.
4 – 16Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 45
4 – Transport and Installation
Check Readiness
for Operation
CleaningCheck the whole system for dirt residue resulting from transport and installation.
1According to the document Installation Qualification, Doc ID 391 180, perform Setup
& Service software module.
2Perform calibration process according to document Operation Qualification, Doc ID
391 182.
Observe appropriate sections in the Operating Manual.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.04 – 17
Page 46
4 – Transport and Installation
4.5Conformity Assessment and CE Marking for RWS
Legal SituationGenesis RWS is an open and flexible platform providing our clients with many
configuration possibilities. A Genesis RWS instrument is considered as sub-assembly
that must be integrated into a complete system. Each system has to be assessed and
investigated to comply with particular local requirements.
The openness of the Genesis RWS system makes it impossible to declare it as CE
conform and affix the corresponding label.
ResponsibilityA system integrator is a legal person (e.g. a RO within Tecan group or a distributor) who
acts in legal sense between Tecan Schweiz AG (manufacturer of sub-assembly) and
the end-user of the system. The system integrator takes the responsibility of a remanufacturer for the whole system and must fulfill the following points:
1Description of complete system (product)
2Draw up (technical file)
3Assure conformity to local regulations
4Declare Conformity
5Affix CE-marking (within EU-region only)
Tecan Schweiz AG can provide support to help system integrators with points 2, 3, 4
and 5 above.
Safety StandardsPlease note that Genesis RWS Instruments meet the following:
•Safety requirements for laboratory equipment
EN 61010-1 / UL 3101-1 / CSA C22.2 No 1010-1
except Part 5 and 7, which are met only in part due to the open nature of the
instrument
•EMC requirements for laboratory equipment
EN 50081-1, part 1
EN 50082-1, part 1
The above is stated in the manufacturer’s declaration.
4.6Site Acceptance
Completion1In case of a RWS instrument: make a conformity declaration.
2Fill out the Acceptance Protocol (Doc ID 391 825) and have it signed by the
customer.
4 – 18Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 47
5Maintenance
5 – Maintenance
Purpose of This
Chapter
This chapter summarizes the preventive maintenance activities intended to retain the
Genesis RSP, RMP or RWS instrument in a state in which it maintains the required or
specified performance.
5.1Concerning Your Safety
DecontaminationDepending on the application, the instrument may be contaminated with compounds
hazardous to your health.
WARNING
The instrument may be contaminated. Before performing any maintenance or repair
tasks:
•Make sure that the instrument has been decontaminated by an expert according to
standard laboratory regulations.
•Request a filled out and signed Decontamination Declaration (Doc ID 390 901)
Unless otherwise noted, always switch off power and disconnect from mains before
carrying out any maintenance tasks.
5.2Consumables
Cleaning Agents
ATTENTION
Strong detergents may dissolve carrier and worktable surface coatings.
Use only cleaning agents that are recommended by Tecan and according to table
Cleaning in Section 5.3.2.
Alcohol
Use ethyl-alcohol or 2-Propanol (Isopropanol).
Water
Use distilled or de-ionized water.
Bleach
Use sodium hypochlorite solution, max. 6 %.
Detergent
Use a weak detergent like
•RoboScrub
®
, order through Tecan US (order number 70-736 for 16 oz. bottle) or
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.05 – 1
Page 48
5 – Maintenance
•CLEAN SYSTEM, order through Tecan-D:
Daily System Clear = to clean the system daily = order number 30000938 (250 ml),
Setup Clean = to clean contaminated systems = order number, 30000937 (500 ml),
Protolyse = to clean the system from proteins = order number 30000939 (1000 ml),
Carrystop = against carryover = order number 30000940 (1000 ml).
Cleaning MaterialUse lint-free tissue only.
5.3Maintenance Schedule
Spare PartsFor spare parts list with part numbers refer to Chapter 7, ‘Check Lists Spare Parts’.
•Use original Tecan spare parts only.
General
Note
Spare parts must correspond to the technical requirements laid down by the
manufacturer.
5.3.1Daily/Weekly Maintenance
For daily and weekly maintenance please follow the descriptions in the document Daily/
Weekly Maintenance Checklist, Doc ID 391 193. Check off the appropriate boxes as
you go through the maintenance tasks.
5.3.2Half-Yearly Maintenance
Refer to document Preventive Maintenance Checklist, Doc ID 391 181 and check off
the appropriate boxes as you go through the maintenance tasks.
Task
Print out system information
Run counter read out
5 – 2Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.05 – 3
Page 50
5 – Maintenance
5.3.3Yearly Maintenance
Fill/flush system
Disposable tips
Liquid detection test
LICOS test
PosID test
Washer: prime each washer channel
Reader: QC test
Door lock test
Refer to document Preventive Maintenance Checklist, Doc ID 391 181 and check off
the appropriate boxes as you go through the maintenance tasks.
GeneralPerform the half-yearly maintenance tasks as listed in Section 5.3.2.
Parts to be
Replaced
Test to Be
Performed
PartInterval: 12 month or …Refer to …
Syringesafter 1 million movesSection 6.4.3
3-way valve of diluterafter 1 million movesSection 6.4.3
Tips
Aspirating tubingSection 6.4.2
Interconnecting tubingSection 6.4.2
Pipetting tubingSection 6.4.2
Waste tubingSection 6.4.2
Perform every 12 month:
Test
Precision test
5 – 4Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 51
6 – Replacement of Spare Parts, Repair
6Replacement of Spare Parts, Repair
Purpose of This
Chapter
Chapter OverviewThis chapter consists of the following sections:
This chapter describes the activities carried out after a failure has occurred, intended to
restore an item to a state in which it can perform its required function.
SectionTitlePage
6.1General Notes on Repair6 – 4
6.1.1Concerning Your Safety6 – 4
6.1.2General Repair Tasks6 – 5
6.1.3Operating Tests after Repair6 – 6
6.2Abstract of Important Data6 – 7
6.2.1Special Tools6 – 7
6.2.2Jumper and Address Settings Overview6 – 8
6.2.3CAN-Bus Resistance Test6 – 10
6.2.4Software Error Messages6 – 11
6.3Instrument – Disassembly and Assembly Procedures6 – 22
6.3.1Worktable6 – 22
6.3.2Door Locks6 – 24
6.3.3X-Drive Assembly6 – 25
6.3.4Power Modules6 – 34
6.3.5Electronic Boards6 – 37
6.4Liquid System – Disassembly and Assembly Procedures6 – 46
6.4.1Overview6 – 46
6.4.2Tubing System6 – 47
6.4.3Diluter and Dilback6 – 50
6.4.4MPO/FWO6 – 54
6.4.5Low Volume Option6 – 62
6.4.66-Way Valve Option6 – 69
6.5LiHa 1 – Disassembly and Assembly Procedures6 – 73
6.5.1Overview6 – 73
6.5.2Complete LiHa 1 Assembly6 – 75
6.5.3X-Flex Cable6 – 76
6.5.4ILID Cable6 – 76
6.5.5ILID Flat Cable6 – 76
6.5.6Electronic Boards for LiHa 16 – 78
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 1
Page 52
6 – Replacement of Spare Parts, Repair
SectionTitlePage
6.5.7Y-Belt and Y-Spreading Belt6 – 79
6.5.8Y-Motor and Y-Spreading Motor6 – 80
6.5.9Tip Adapter6 – 81
6.5.10Ensure Operating Readiness6 – 81
6.6LiHa 2 – Disassembly and Assembly Procedures6 – 83
6.6.1Overview6 – 83
6.6.2Complete LiHa 2 Assembly6 – 85
6.6.3X-Flex Cable6 – 88
6.6.4ILID Cable6 – 89
6.6.5ILID Flat Cable6 – 90
6.6.6Electronic Boards for LiHa 26 – 91
6.6.7Y-Belt and Y-Spreading Belt6 – 95
6.6.8Y-Motor and Y-Spreading Motor6 – 97
6.6.9Tip Adapter6 – 98
6.6.10Lower DiTi Eject Option6 – 99
6.6.11Ensure Operating Readiness6 – 104
6.7RoMa 1 – Disassembly and Assembly Procedures6 – 106
6.7.1Overview6 – 106
6.7.2Complete RoMa 1 Assembly6 – 109
6.7.3Gripper Fingers6 – 109
6.7.4Gripper Module Head, Gripper Board6 – 109
6.7.5Rotator Motor6 – 110
6.7.6X-Flex Cable6 – 110
6.7.7Gripper/Rotator Flex Cables6 – 110
6.7.8Y/R- and Z/G-DC-Servo Board6 – 110
6.7.9RoMa 1 Backplane6 – 111
6.7.10Y-Belt6 – 111
6.7.11Y- and Z-Motor6 – 111
6.7.12Z-Brake6 – 112
6.7.13Ensuring Operating Readiness6 – 113
6.8RoMa 2 – Disassembly and Assembly Procedures6 – 114
6.8.1Overview6 – 114
6.8.2Complete RoMa 2 Assembly6 – 117
6.8.3Mechanical Adjustment After Reinstallation6 – 118
6.8.4Gripper Fingers6 – 121
6.8.5Gripper Module Head, Gripper Board6 – 122
6.8.6Rotator Motor6 – 125
6 – 2Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 53
6 – Replacement of Spare Parts, Repair
SectionTitlePage
6.8.7X-Flex Cable6 – 126
6.8.8Gripper/Rotator Flex Cables6 – 129
6.8.9Y/R- and Z/G-DC-Servo Board6 – 131
6.8.10RoMa 2 Backplane6 – 132
6.8.11Y-Belt6 – 133
6.8.12Y-Motor6 – 135
6.8.13Z-Motor6 – 136
6.8.14Z-Brake6 – 137
6.8.15Ensuring Operating Readiness6 – 138
6.9PosID 1 – Disassembly and Assembly Procedures6 – 139
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 3
Page 54
6 – Replacement of Spare Parts, Repair
6.1General Notes on Repair
Useful References •Refer to Chapter 8, ‘Instruments at a Glance’ for an overview concerning the
location of the most common parts.
•For spare parts list with part numbers refer to Chapter 7, ‘Check Lists Spare Parts’.
•As Chapter 7 is structured in the sequence of disassembly of an item, it can also be
used by experienced FSEs as brief instructions.
•For electrical diagrams refer to Chapter 9, ‘Diagrams’.
Spare Parts•Use original Tecan spare parts only.
Note
Spare parts must correspond to the technical requirements laid down by the
manufacturer.
Disposal of Waste •Dispose used items and consumables according to the relevant national
environmental, health and safety laws and regulations.
6.1.1Concerning Your Safety
Chemical,
Biological and
Radioactive
Hazards
DecontaminationFor your own and any other affected person’s safety, make sure the instrument or parts
Depending on the applications, parts of the instrument may have been in contact with
biohazardous, poisonous or even radioactive materials. A potential risk may arise from
the liquids that have been handled on the instrument.
•Strictly apply appropriate safety precautions according to general laboratory and
applicable local, state and federal regulations.
WARNING
Potential Biohazard. The instrument might be contaminated! For servicing, use
appropriate personal protective equipment!
have been thoroughly decontaminated prior to carrying out any maintenance and repair
task on the instrument or before sending it or parts of it for repair:
•Thoroughly clean and decontaminate all relevant parts.
•Provide a filled-out and signed Decontamination Declaration (Doc ID 390 901).
6 – 4Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 55
Electrical Shock
Hazard
6 – Replacement of Spare Parts, Repair
WARNING
Unless otherwise noted, always switch off power and disconnect from mains before
carrying out any tasks described in this chapter.
Electrostatic
Sensitive Devices
Discharge static electricity from your body by wearing a wrist strap to protect yourself
and sensitive control electronics installed:
•Wear a wrist strap during all installation and maintenance tasks, e.g. when
exchanging components.
ATTENTION
Observe precautions for handling electrostatic discharge sensitive devices.
6.1.2General Repair Tasks
Loctite® 638•Several set screws are glued with Loctite 638. Use a hot-air apparatus to soften the
adhesive before unscrewing these screws.
Disconnecting
Cables
•When disconnecting cables, always mark designation (Jxx) on connector using a
water resistant pen.
If due to insufficient space, designations
are not marked on the PCB use electrical
diagrams instead.
Figure 6-1 Connector designation
When replacing
any PCB …
Cleaning•Wipe off grease residue and thoroughly clean affected instrument parts after any
•Make sure the jumpers for CAN-bus resistance are set identically to the replaced
PCB (see also Section 6.2.2, ‘Jumper and Address Settings Overview’).
•Control if address switch is set correctly (see Section 6.2.2, ‘Jumper and Address
Settings Overview’).
•Check CAN-bus resistance (see Section 6.2.3, ‘CAN-Bus Resistance Test’).
repair task.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 5
Page 56
6 – Replacement of Spare Parts, Repair
6.1.3Operating Tests after Repair
Service Checklist•After replacing or removing and reinstalling modules or spare parts, carry out the
appropriate tests according to Service Checklist Doc ID 391 183 (included in the
Genesis Maintenance and Service Logbook).
6 – 6Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 57
6.2Abstract of Important Data
6.2.1Special Tools
LiHa reference tipPart No. 612 503
384 well tip adjustment toolPart No. 613 103
6 – Replacement of Spare Parts, Repair
RoMa calibration toolPart No. 612 622
RoMa teach platePart No. 613 101
PosID service rackPart No. 613 100
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 7
Page 58
6 – Replacement of Spare Parts, Repair
6.2.2Jumper and Address Settings Overview
The following communication overview shows …
•the default address settings on the electronic boards of the different modules,
•on which modules the jumpers have to be set/switched on, or removed/switched off
according to the configuration of the instrument:
Communication
Overview
RS 232
RS 232
PC
RS 232 / RS 232
Multiport 8x RS232
Washer
Reader
Z/R
Liconic Carousel
Stepper Drive / Scanner
Option-Can
O7
Te-Shake #6
Option-Can
Te-Mags #5
O6
3
Option-Can
Te-Vacs #4
O5
3
O4
Option-Can
Te-Sonic #3
Option-Can
O3
Incubator #2
32
SMIO #1
O2
Option-Can
6
MPO #0
O1
Option-Can
#0
Can
DC-
ST5
Local-
Servo
5
Option-Can
T1..T6
TEMPO #8..D
7
O2
Option-Can
Te-Flow #1
5
V1
Option-Can
LOW-VOLUME
O7
V2
O2
U1..4
Q1...8 S1..8
...4
T1
72
4
RMP
4
1
2
RSP
9
Option-Can
Option-Can
Option-Can
Option-Can
Option-Can
Local-
Local-
Option-Can
Te-Stack #0...F
WRC (2) #6
6-way valve
Supervisor #1
Uniport
position)
(depending on
TeMo #0
Can
DC-
Servo
Can
DC-
Servo
#1
#0
Local-
T3TR6
Can
2
The Device default number and name, i.e. O5 as device is identified in FW
The Device default adress, i.e. #4. Setting for the adress switch
Some configurations may require different adress settings.
#0
DC-
Servo
Option-Can
WRC (1) #6
= always open (Switch off or Jumper not set)
= always terminated if the module is installed (Switch on or Jumper set)
O5
The termination of the option-CAN and the system-CAN is priorized
depending on configuration:
Option-Can
The two installed modules with the lowest numbers / letters within the box
have to be terminated:
{
Vacs #4
= priorized termination of Option CANA= priorized termination of System CAN
1
CU
RS-232
M1
Can
DC-
System-
Optibo
MO2
D9
System-Can Option-Can
#5
#7
Can
D8
Diluter
System-
#6
Can
D7
Diluter
System-
#5
Can
D6
Diluter
System-
#4
Can
D5
Diluter
System-Can
Diluter Backplane
System-
#3
Can
D4
Diluter
System-
#2
Can
D3
Diluter
System-
#1
Can
D2
Diluter
System-
#0
Can
D1
Diluter
System-
C
E9
Servo
8
Option-Can
P1
CU PosID2
#0
Can
DC-
MO1
Local-
Servo
#4
Can
DC-
MO5
Servo
System-
#3
Can
DC-
MO4
Servo
System-
B
System-Can
A1
Backplane (LiHa)
Can
System-
Can
System-
Can
System-
DC-
DC-
DC-
#2
MO3
Servo
#1
MO2
Servo
#0
MO1
Servo
Scanner
Y/B PosID
Local-Can / RS232
#2
Can
DC-
MO2
Local-
Servo
#7
Can
DC-
MO1
Servo
RoMa
Backplane
System-
Can
System-
DC-
#6
MO2
Servo
A
R1
System-Can
6 – 8Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 59
Jumper Allocation
6 – Replacement of Spare Parts, Repair
PCPSystem
CAN-bus
Optibo, Optibo PowerJ12J13—
CU boardJ1J22
Option
CAN-bus
Local
CAN-bus
MPO board V2.0
MPO board V3.1/V4.0
LiHa backplaneJ14——
DilbackJ10——
RoMa 1 backplane
RoMa 2 backplane
PosID 1/PosID 2 CU board—J8J6
—
—
J2
J1
J4
J2
—
—
—
—
—
—
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 9
Page 60
6 – Replacement of Spare Parts, Repair
6.2.3CAN-Bus Resistance Test
Purpose of this test is to verify whether bus end jumper settings on control electronics
are correct.
Will Be Applicable •When commissioning a new instrument.
•After exchange or new installation of a LiHa 1/LiHa 2, RoMa 1/RoMa 2, PosID 1/
PosID 2 or other options as e.g. Te-VacS etc.
•After exchange or supplemental installation of PCBs.
Required Special
Tools
Test Procedure1Switch off the instrument.
•Multimeter
2On the Optibo or Optibo Power (left access door) check the CAN-bus resistance
between the respective check points:
Figure 6-2 Check points for System CAN-bus
•System CAN
Measure resistance between CAN_L
and CAN_H.
•Option CAN
Measure resistance between CANO_L
and CANO_H.
Figure 6-3 Check points for Option CAN-bus
3Correct CAN-bus resistance:
50–65 Ω:correct number of jumpers
< 50 Ω:too many jumpers are set
~ 40 Ω:1 jumper too much
> 65 Ω:too few jumpers are set
~ 120 Ω:1 jumper too few
If CAN-bus resistance is not between 50 an 65 Ω: compare jumper settings and
CAN-bus connections with ‘Communication Overview’in Section 6.2.2, ‘Jumper and
Address Settings Overview’ and correct.
6 – 10Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 61
6.2.4Software Error Messages
Toolbox General
Errors
6 – Replacement of Spare Parts, Repair
Toolbox
error code
0OK (no error)
1Toolbox not initialized (from V2.1 on: not opened)
2Toolbox already initialized (from V2.1 on: already open)
3Toolbox init error (from V2.1 on: error in opening)
4Communication driver error
5Incorrect answer string format (cannot extract)
6Illegal command parameter value
7Error list contains at least one error
8No such worktable
Explanation
9Open rack database failed
10Worktable access error (carrier/rack)
11Evaluation access (not a single position defined)
12Evaluation access (not a single position is accessible due to Y)
13Evaluation access (position out of range)
14Accessing all selected positions not possible
15No carrier/rack defined
16Worktable access error (tube)
17Open comm log file failed
18No carrier on specified grid position
19Worktable object creation error
20Open communication driver failed
21Close communication driver failed
22Evaluation access (not a single position is accessible due to X)
23Genesis DB not open
24Communication driver not open
25Plate stuck on gripper
26Balance initialization error
27Balance isn't initialized
28Balance isn't correct tared
29Measurements not enough linear. Preparing/weighing not possible.
30Wrong machine type. No balances available.
31Balance not prepared. Weighing not possible.
32Error while getting a balance value. Weighing not possible.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 11
Page 62
6 – Replacement of Spare Parts, Repair
Toolbox
error code
33Wrong machine type. No low volume support.
34This machine has no pinch valves
35No installable easy option available
36Easy option slot occupied
37Defined Easy Option module not available
38The initialization of the serial communication port failed
39Sending through the serial communication port failed
40Receiving through the serial communication port failed
41Command not implemented (for the PosID 1)
42Command not implemented (for the PosID 2)
43Difference too big between “prepare weigh” and “weigh” in
44Worktable View: nothing selected
45Worktable View: index out of range
46Worktable View: …
Explanation
evaporation
Toolbox User
Interactions
Toolbox Data
Errors
47Worktable View: object not created
Toolbox
error code
90Abort from user
91Retry from user
92User (calling application) handles error
93Ignore from user
94“Go to ZBottom” from user
Toolbox
error code
101Incorrect tip number
102Data pointer is NULL
103Individual tip data set (no global data available)
Explanation
Explanation
104Incorrect liquid index
105Data set not defined
106Selector out of bounds
107Not enough space (for complete report)
6 – 12Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 63
6 – Replacement of Spare Parts, Repair
Toolbox
Communication
Errors
Toolbox Device
Errors (Ranges)
Toolbox
Explanation
error code
108Invalid diluter number
109Parameter value(s) out of range
110Invalid arm ID
111Invalid number of tips
112Invalid number of diluters
Toolbox
Explanation
error code
201COM unknown error
202COM device still busy
203COM device time-out
.
Toolbox error
code range
Device
Address
Explanation
Common Errors
for All Devices
300Common
400MCU
500ALiHa
527ALiHa (tip mounted)
600OMPO, Incubator, Supervisor
700DDiluter
800PPosID
900RRoMa
1000VLow Volume Option
1100UUniport
10,000PPosID extended errors
Toolbox error code = Toolbox error code range + Communication driver error code
Toolbox error code = Toolbox error code range (10,000 for PosID) + Communication
driver error code
Communication
driver error code
x5010,x50Unknown axis mnemonic
x5110,x51Invalid axis identifier
Toolbox
error code
Explanation
x5210,x52# of sub-axis is out of limit
x5510,x55Collision detection is active
x5810,x58Time-out calculation error
6 – 16Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 67
6 – Replacement of Spare Parts, Repair
Communication
driver error code
x5910,x59Dynamic read calculation error
x6110,x61Initialization: target not reachable
x6210,x62Target out-of-range
x6310,x63Target not reached
x6510,x65Axis movement time-out
1x2011,x20CAN over run
1x2111,x21No load
1x2211,x22No liquid detected
1x2311,x23Break point not reached
1x2411,x24Overload detected
1x2511,x25Delayed action pending (ready to execute)
1x2611,x26Delayed action pending (ready to execute)
1x2711,x27Command not yet implemented
1x2911,x29DC-Servo multiple error
Toolbox
error code
Explanation
PosID 1
Miscellaneous
Errors
Toolbox error code = Toolbox error code range (10,000 for PosID) + Communication
driver error code
BCS = Barcode Scanner; HOST = PosID CU
Communication
driver error code
40110,401BCS initialization: no ID REQUEST from BCS
40510,405Unexpected response from BCS after START
41110,411No decoded barcode from BCS
41210,412Barcode string length out-of-range
41310,413NR (No Read) message from BCS
42110,421CMNDNAK (CoMmaND Not Acknowledge)
42210,422No ACK or NAK respond from BCS
42510,425No respond from BCS
42610,426Checksum error detected by BCS
Toolbox
error code
Explanation
after ATTENTION from HOST
DECODE
respond from BCS
42710,427Unexpected respond from BCS
43510,435No respond from BCS
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 17
Page 68
6 – Replacement of Spare Parts, Repair
Communication
driver error code
43610,436Checksum error detected by HOST
43710,437Unexpected respond from BCS
43810,438No complete respond from BCS
44110,441HOST received message from wrong
45110,451BC type out-of-range
45210,452Verify mode out-of-range
45510,455Unexpected character in BC string
45610,456BC check digit error
46010,460Invalid test number
47010,470Parameter out-of-range
50110,501Invalid field id
50210,502Invalid parameter
51010,510Invalid sector id
51110,511Sector already active
Toolbox
error code
Explanation
message source
51210,512Sector not yet active
51310,513Sector parameter out-of-range
51410,514Sector limit reached
52010,520Invalid fixed field id
52110,521Fixed field already active
52210,522Fixed field not yet active
52310,523Fixed field parameter out-of-range
53010,530Unknown carrier state
53110,531Invalid carrier code
53210,532Invalid carrier code string length
53310,533Invalid carrier code type
53410,534Invalid carrier placement
53510,535Invalid carrier id
53610,536Carrier id already active
53710,537Carrier id not yet active
53810,538Invalid carrier code format
54510,545Invalid carrier type
54610,546Carrier type already active
54710,547Carrier type not yet active
54810,548Invalid carrier data id
54910,549Carrier is at remove position
6 – 18Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 69
6 – Replacement of Spare Parts, Repair
Communication
driver error code
55010,550Unknown rack/sample state
55510,555Invalid rack/sample type
55610,556Rack/sample type already active
55710,557Rack/sample type not yet active
55810,558Invalid rack/sample data id
56010,560Move right carrier locked
60110,601Invalid port number
60210,602Invalid bit number
70110,701BC scan task not started
71010,710No carrier present
71110,711Expected carrier not found
72010,720Calibration Y max read
74110,741Reload of same carrier
74210,742Comparison of 1st rack/sample impossible
74310,743Illegal removal of carrier detected
Toolbox
error code
Explanation
74410,744Illegal movement of carrier detected
75010,750Previous BC read was equal
PosID 2 ErrorsToolbox error code = Toolbox error code range (= 800 for PosID) + Communication
driver error code
Communication
driver error code
9 809No liquid detected by DC-Servo
10810Drive with no load detected
11811Break point reached
12812No access to parameter block
13813Collision between LiHa and PosID avoided
14814Extended error
16816Power fail circuit failure
17817Not locked to grid
Toolbox
error code
Explanation
(error code currently not used)
18818Sector not defined
20820Barcode scanner failure (communication
problem, e.g. wrong answer to ESC
sequence)
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 19
Page 70
6 – Replacement of Spare Parts, Repair
Communication
driver error code
21821Barcode scanner communication error
22822Carrier not movable
23823Break point not reached
24824Carrier not defined
25825Rack/sample not defined
26 826 Carrier not loaded
27827Invalid carrier id barcode
28828Carrier not present
29829Illegal BCS movement
30 830 BCS failed during initialization
Toolbox
error code
Explanation
(transmission errors upon OV, FE, PT, VRC
cmds)
(error code currently not used)
(error code currently not used)
Low Volume
Option Errors
IncubatorErrorsToolbox error code = Toolbox error code range (= 600 for incubator)+ Communication
Toolbox error code = Toolbox error code range (= 1000 for Low Volume Option) +
Communication driver error code
Communication
driver error code
91009Selected valve not installed
101010Endless mode not possible
111011Wrong node defined
driver error code
Communication
driver error coder
9609Shaker not installed
10610Shaker position initialization error
11611Close door solenoid short or open circuit
12612EEPROM checksum error
Toolbox
error code
Toolbox
error code
Explanation
Explanation
13613EEPROM access error
16616Slot temperature overrun
17617Temperature sensor short- or open circuit, or
suspect
18618Shaker fail
6 – 20Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 71
6 – Replacement of Spare Parts, Repair
SupervisorErrorsToolbox error = Toolbox error range (= 600 for Supervisor)+ Communication driver error
Communication
driver error coder
12612LICOS not calibrated
13613EEPROM access error
16616Door already open
17617Hall sensors not calibrated
UniportErrorsSpectra/Sunrise Errors
Toolbox error = Toolbox error range (= 1100 for Uniport)+ Communication driver error
Communication
driver error coder
161116Lamp low
171117Lamp high
181118Transport error
191119Filter error
201120Checksum error
Toolbox
error code
Toolbox
error code
Explanation
Explanation
211121No data uploaded by “D” or “D2” cmd.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 21
Page 72
6 – Replacement of Spare Parts, Repair
6.3Instrument – Disassembly and Assembly Procedures
This section describes mechanical and electronic exchange routines on instrument
level.
6.3.1Worktable
Spare Parts
Worktable
Required Special
Tool
Removal of
Worktable
For spare parts list with part numbers refer to Section 7.1.1, ‘Spare Parts Worktable’.
Following items can be replaced as spare parts:
•Worktable
•Positioning pins
•Sealing strip
•Reference tip for LiHa
Will Be Applicable …
•If a new or another type of worktable has to be installed.
•If you need to access the PosID/PosID 2 or an older version of PosID/PosID 2
CU board (located underneath the worktable).
Removal
For more information refer to Section 4.3, ‘Installation’, inverse order of installation
instructions.
1Empty the worktable completely.
2For RMP: detach the module worktable.
3Open the front worktable cover (10).
•For RMP: remove the door locks (21).
•For RWS: remove magnet holders.
4Remove all screw caps (tip: use a watch maker screwdriver) and all fixing screws
from the worktable.
5Remove the worktable by carefully lifting and pulling forward.
Installation of
Worktable
6 – 22Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
1Carefully install the worktable.
2RSP and RWS: insert the 4 countersunk screws on the rear right and rear left but do
not tighten them yet.
RMP: insert the 2 countersunk screw on the rear left but do not tighten them yet.
Page 73
6 – Replacement of Spare Parts, Repair
3Insert but only slightly tighten the oval head screws.
4Manually align the worktable, then tighten the front left screw.
This screw will act as a pivot point during the following alignment.
5Install the reference tip onto the LiHa:
•4-Tip LiHa: mount reference tip on position 1 (first tip from behind).
•8-Tip LiHa: mount reference tip on position 3 (third tip from behind).
6Manually position the reference tip at the edge of the foremost positioning pin in grid
position 1 (most left grid). Check the gap between pin and reference tip.
7Manually move the LiHa to the
rightmost grid position and check the
gap between pin and reference tip.
The gap must be the same on either
side of the worktable.
Ensure Operating
Readiness
Figure 6-4 Gap between reference tip and positioning pin
8When the worktable is positioned: tighten the fixing screws and insert the screw
caps.
9Reinstall all previously removed parts. For detailed information about installing …
•RMP door locks: refer to Section 6.3.2.
•RMP 200 add-on module: refer to Section 4.3, ‘Installation’.
•Reader, washer and incubators: refer to Section 4.3, ‘Installation’.
10 Ensure Operating Readiness as described later in this section.
For detailed information refer to the Genesis Instrument Software Manual.
After replacing or reinstalling the worktable or replacing the positioning pins, control –
and if necessary carry out – following settings:
StepTest or Setup
1LiHa
•Check reference positions
•Adjust scale factors
Only necessary if check failed
•Adjust reference positions
Only necessary if check failed
•Set absolute Z
•Set individual Z
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 23
Page 74
6 – Replacement of Spare Parts, Repair
2PosID
•Test
•PosID adjustment
Only necessary if test failed
3RoMa
•Check reference positions
•RoMa setup
Only necessary if check failed
6.3.2Door Locks
RMPFor the description of disassembly and assembly procedure for the Genesis RMP door
locks, please refer to the document Doc ID 391 260.
RSP, RWSThe door lock is part of the access option for RSP and RWS instruments. Please refer
to document Doc ID 392 330 for further information.
6 – 24Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 75
6 – Replacement of Spare Parts, Repair
6.3.3X-Drive Assembly
The X-drive assembly is installed in the X-bay (5) of the instrument. It consists of one Xmotor and a X-belt which drives the LiHa. If the instrument is equipped with a RoMa, a
second X-drive assembly is installed which drives the RoMa.
The X-DC-Servo board – which controls the X-motors of the LiHa and RoMa – is
described in Section 6.3.5, ‘Electronic Boards’.
Spare Parts For spare parts list with part numbers refer to Section 7.1.3, ‘Spare Parts X-Drive
Assembly’.
Following items can be replaced as spare parts:
•Belt tensioner
•X-belt (5.5 m)
•X-DC-Servo motor with cable and bracket
Overview X-Drive
Assembly for LiHa
5-01
5-10
2x M4x10
5-02
5-035-04
5-11
5-06
2x M4x10
5-07
5-08
M4x10
5-015-035-07
5-09
3x M4x8
5-05
M4x60
Figure 6-5 X-drive assembly for LiHa
5-01X-motor LiHa5-06Fixing screw/washer for belt take-up
5-02X-motor cable LiHa (J13)5-07X-slide LiHa
5-03X-belt LiHa5-08Fixing screw for X-belt LiHa
5-04Belt take-up LiHa5-09Fixing screw for X-motor LiHa
5-05Belt tensioner (belt tensioning screw/5-10Fixing screw for X-motor LiHa
washer/pressure spring)5-11Driver LiHa
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 25
5-09Fixing screw for X-motor LiHa5-18Fixing screw/washer for belt take-up
5-10Fixing screw for X-motor LiHa5-19X-slide RoMa
5-12X-motor RoMa5-20Fixing screw for X-motor RoMa
5-13X-motor cable RoMa (J09)5-21Fixing screw for belt lock
5-14X-belt RoMa5-22Fixing screw for X-motor RoMa
5-15Fixing screw for X-belt RoMa5-23Belt lock RoMa
5-16Belt take-up RoMa5-24Driver RoMa
Removal of X-Belt For LiHa and RoMa X-Belt
The item numbers mentioned refer to Figure 6-5, ‘X-drive assembly for LiHa’ and Figure
6-6, ‘X-drive assembly for RoMa’.
Note
In case the instrument is not yet equipped with the new belt tensioner (5-05 or 5-17), we
recommend to upgrade the instrument when removing the X-belt.
6 – 26Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 77
6 – Replacement of Spare Parts, Repair
1Remove the arm of the relevant X-drive.
For LiHa 1 see Section 6.5.2, ‘Complete LiHa 1 Assembly’,
for LiHa 2 see Section 6.6.2, ‘Complete LiHa 2 Assembly’,
for RoMa 1 see Section 6.7.2, ‘Complete RoMa 1 Assembly’
for RoMa 2 see Section 6.8.2, ‘Complete RoMa 2 Assembly’.
2Unscrew and remove the X-bay covers, open right access door (6).
3For RMP instruments only:
•Remove system liquid container.
•Unscrew the steel plate supporting the Supervisor/Uniport board (41, 37) in
order to get access to the belt tensioning screw (5-17).
4From below the X-bay loosen the respective belt take-up fixing screws (5-06 or
5-18).
5Loosen the respective belt tensioning screw (5-05 or 5-17) to slacken the X-belt
•Open belt lock (5-23) and X-belt fixing screws (5-15), remove X-belt (5-14).
For LiHa X-Belt
5-04
5-03
5-11
5-08
M4x10
2x M4x10
5-07
5-05
M4x60
5-06
Figure 6-7 Installing X-belt LiHa
5-03X-belt LiHa5-06Fixing screw/washer for belt take-up
5-04Belt take-up LiHa5-07X-slide LiHa
5-05Belt tensioner (belt tensioning screw/5-08Fixing screw for X-belt LiHa
washer/pressure spring)5-11Driver LiHa
1Cut the new belt to required size according to the replaced one.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 27
Page 78
6 – Replacement of Spare Parts, Repair
2If installed: remove belt tensioning screw (5-05).
3Fix – but do not tighten yet – the belt take-up fixing screws (5-06) which should be
positioned more likely in the left part of the slotted hole.
5-03
5-04
5-06
5-08
5-03
5-11
Figure 6-8 X-belt LiHa
4Lead the X-belt (5-03) over the pulleys
5-03X-belt LiHa
5-04Belt take-up LiHa
5-06Fixing screw for belt take-up
5-08Fixing screw for X-belt
5-11Driver LiHa
5Loosen the fixing screws of the driver (5-11).
.
6Screw the belt tensioning screw (5-05)
6.5 mm
5-04
5-05
as shown in Figure 6-7 and fasten it
(fixing screw 5-08) in the driver (5-11).
Make sure the belt is placed correctly
over the pulleys (Figure 6-8).
with washer and pressure spring into
the belt take-up (5-04) until the distance
between inner face of the washer and
the frame of the instrument is 6.5 mm
(this corresponds to a tension of 9 N).
5-04Belt take-up LiHa
5-05Belt tensioning screw
5-06Fixing screw for belt take-up
7Check the position of the belt take-up
fixing screws (5-06) in the slotted hole;
it should bein the left third of the slot.
5-06
1/3 2/3
Figure 6-9 Tensioning the X-belt LiHa
Otherwise you need to shorten the Xbelt.
8Tighten the belt take-up fixing screws (5-06) and glue them with Tree Bond 1342.
9Reinstall LiHa (see Section 6.5.2, ‘Complete LiHa 1 Assembly’ or Section 6.6.2,
‘Complete LiHa 2 Assembly’) and ensure operating readiness as described in
Section 6.5.10, ‘Ensure Operating Readiness’ (for LiHa 1) or Section 6.6.11,
‘Ensure Operating Readiness’ (for LiHa 2).
6 – 28Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 79
For RoMa X-Belt
6 – Replacement of Spare Parts, Repair
5-15
5-16
5-14
5-17
5-24
5-23
5-18
2x M4x10
5-19
Figure 6-10 Installing X-belt RoMa
5-14X-belt RoMa5-18Fixing screw/washer for belt take-up
5-15Fixing screw for X-belt RoMa5-19X-slide RoMa
5-16Belt take-up RoMa5-23Belt lock RoMa
5-17Belt tensioner (belt tensioning screw/5-24Driver RoMa
washer/pressure spring)
M4x60
1Cut the new belt to required size according to the replaced one.
2If installed: remove belt tensioning screw (5-17).
3Fix – but do not tighten yet – the belt take-up fixing screws (5-18) which should be
positioned more likely in the left part of the slotted hole.
4Fix one end of the X-belt (5-14) in the belt lock (5-23), then lead it over the pulleys
as shown in Figure 6-10. Insert the other belt end in the belt take-up (5-16) and
fasten the fixing screws (5-15).
Make sure the belt is placed correctly over the pulleys as shown in the upper
illustration of Figure 6-8.
5Loosen the fixing screws of the driver (5-24).
6Screw the belt tensioning screw (5-17) with washer and pressure spring into the belt
take-up (5-16) until the distance between inner face of the washer and the frame of
the instrument is 6.5 mm (this corresponds to a tension of 9 N). (See Figure 6-11,
‘Tensioning the X-belt RoMa’.)
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 29
Page 80
6 – Replacement of Spare Parts, Repair
.
6.5 mm
5-16
5-17
5-16Belt take-up RoMa
5-17Belt tensioning screw
5-18Fixing screw for belt take-up
7Check the position of the belt take-up
fixing screws (5-18) in the slotted hole;
5-18
1/3 2/3
Figure 6-11 Tensioning the X-belt RoMa
it should bein the left third of the slot.
Otherwise you need to shorten the Xbelt.
8Tighten the belt take-up fixing screws (5-18) and glue them with Tree Bond 1342.
9For RMP instruments only:
•Install the steel plate supporting the Supervisor/Uniport board (41, 37).
•Install system liquid container.
10 Install the X-bay covers.
11 Reinstall RoMa (see Section 6.7.2, ‘Complete RoMa 1 Assembly’ or Section 6.8.2,
‘Complete RoMa 2 Assembly’) and ensure operating readiness as described in
Section 6.7.13, ‘Ensuring Operating Readiness’ (for RoMa 1) or Section 6.8.15,
‘Ensuring Operating Readiness’ (for RoMa 2).
Removal of
X-motor
For Instruments without RoMa
The item numbers mentioned refer to Figure 6-5, ‘X-drive assembly for LiHa’.
1Release tension of the X-belt as described earlier in this section (see ‘Removal of X-
Belt’, step 1 to 4).
2Remove tubing cove
3which is located in the left service
compartment – to access the X-motor
fixing screws 5-10 (see Figure 6-12).
4Carefully disconnect the X-motor cable
from the CU board (see Figure 6-22,
‘CU board cable connections’).
5Open fixing screws (5-09, 5-10),
remove X-motor (5-01).
Do not attempt to remove motor cable
and bracket.
Figure 6-12 Tubing cover
6 – 30Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 81
6 – Replacement of Spare Parts, Repair
For Instruments with RoMa
The item numbers mentioned refer to Figure 6-6, ‘X-drive assembly for RoMa’.
1Release tension of both X-belts as described earlier in this section (see ‘Removal of
X-Belt’, step 1 to 4), irrespective of which X-motor has to be replaced.
2Take the RoMa X-belt (5-14) out of the belt lock (5-23).
3Remove tubing cover – which is located in the left service compartment – to
access the X-motor fixing screws (5-10) (see Figure 6-12, ‘Tubing cover’).
4Carefully disconnect both X-motor cables from the CU board (see Figure 6-22, ‘CU
6Unscrew fixing screw (5-22) to separate the two X-motors (5-01 and 5-12).
Do not attempt to remove motor cable and bracket.
Installation of
X-motor
For Instruments without RoMa
The replacement X-motor is delivered with cable and bracket.
.
5-01
5-01
5-25
5-03
Figure 6-13 Installing X-motor LiHa
5-01X-motor LiHa5-09Fixing screw for X-motor LiHa
5-02X-motor cable LiHa (J13)5-10Fixing screw for X-motor LiHa
5-03X-belt LiHa5-25Motor bracket
5-02
5-10
2x M4x10
5-25
5-09
3x M4x8
1Install the X-motor (5-01) into the X-bay and tighten the fixing screws (5-09, 5-10).
2Lead the X-motor cable trough the slot in the X-bay (see arrow in Figure 6-13) and
connect it to the CU board (J13) (see Figure 6-22, ‘CU board cable connections’).
3Install the X-belt (5-03) as described earlier in this section (see ‘Installation of X-
Belt’, ‘For LiHa X-Belt’).
4Install the tubing cover in the left service compartment (see Figure 6-12, ‘Tubing
cover’).
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 31
Page 82
6 – Replacement of Spare Parts, Repair
5Reinstall LiHa (see Section 6.5.2, ‘Complete LiHa 1 Assembly’ or Section 6.6.2,
‘Complete LiHa 2 Assembly’) and ensure operating readiness as described in
Section 6.5.10, ‘Ensure Operating Readiness’ (for LiHa 1) or Section 6.6.11,
‘Ensure Operating Readiness’ (for LiHa 2).
For Instruments with RoMa
The replacement X-motor is delivered with cable and bracket.
.
5-01
5-01
5-25
5-03
5-14
5-12
5-03
5-13
5-12
5-02
5-01
5-22
5-10
2x M4x10
Figure 6-14 Installing the X-motors
5-01X-motor LiHa5-12X-motor RoMa
5-02X-motor cable LiHa (J13)5-13X-motor cable RoMa (J09)
5-03X-belt LiHa5-14X-belt RoMa
5-09Fixing screw for X-motor LiHa5-20Fixing screw for X-motor RoMa
5-10Fixing screw for X-motor LiHa5-22Fixing screw for X-motor RoMa
5-09, 5-20
5x M4x8
1Place the two X-motors (5-01) and (5-12) on a plane surface and tighten the fixing
screw (5-22).
2Install the X-motor assembly into the X-bay; first tighten the fixing screws (5-09) and
(5-20), then the fixing screws (5-10).
3Lead both X-motor cables trough the slot in the X-bay (see arrows in Figure 6-14)
and connect them to the CU board (J13 = LiHa, J09 = RoMa) (see Figure 6-22, ‘CU
board cable connections’).
4Install the X-belts (5-03, 5-14) as described earlier in this section (see ‘Installation of
X-Belt’).
5Install the tubing cover in the left service compartment (see Figure 6-12, ‘Tubing
cover’).
6Reinstall LiHa (see Section 6.5.2, ‘Complete LiHa 1 Assembly’ or Section 6.6.2,
‘Complete LiHa 2 Assembly’).
7Reinstall RoMa (see Section 6.7.2, ‘Complete RoMa 1 Assembly’ or Section 6.8.2,
‘Complete RoMa 2 Assembly’.
6 – 32Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 83
6 – Replacement of Spare Parts, Repair
8Ensure operating readiness as described in
•Section 6.5.10, ‘Ensure Operating Readiness’ (for LiHa 1) or
•Section 6.6.11, ‘Ensure Operating Readiness’ (for LiHa 2) and
•Section 6.7.13, ‘Ensuring Operating Readiness’ (for RoMa 1) or
•Section 6.8.15, ‘Ensuring Operating Readiness’ (for RoMa 2).
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 33
Page 84
6 – Replacement of Spare Parts, Repair
6.3.4Power Modules
Note
•The old generation of power modules for Genesis RSP and RWS are replaced by
the power modules PM 1, PM 2 and PM 4 as of September 2001.
•For supply ratings and allocation to the different Genesis instruments refer to
Section 3.1.2, ‘Supply Ratings’.
•For more information about wiring of the power module see the appropriate wiring
diagram in Section 9.2.1, ‘Power Modules’.
•Check power rating printed on the type plate of the instrument for correct
identification.
Spare Parts
Power Modules
For spare parts list with part numbers refer to Section 7.1.4, ‘Spare Parts Power
Modules’.
Following item can be replaced as spare parts:
•Power module
.
The power module as a whole part is a spare part.
For safety reasons no replacement of items is allowed. Do not attempt to perform
repairs within this part. Faulty wiring will seriously endanger user and
instrument.
Following descriptions are valid for the old and the new types of power modules.
Switch off the mains power and disconnect the mains power connection of the
instrument before performing any work!
RemovalThe item numbers mentioned refer to Figure 6-15.
1Make sure the instrument is not connected to the mains and the power switch is off.
2Move arm(s) to the middle of the instrument, open top cover (2) and right access
door (6).
3Unscrew possibly present panel (33-04) between diluter compartment and power
module (33).
4Remove the fixing screws (33-03), the fixing screws and sealing washers (33-01).
6 – 34Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 85
6 – Replacement of Spare Parts, Repair
5x M4x10
33-01
33-03
4x M4x10
Figure 6-15 Power module (old generation) on Genesis RSP instrument
3333-0433-02
2
6
2Top cover
6Right access door
33Power module (old generation)
33-01Fixing screws with sealing washers
33-02Power switch
33-03Fixing screws
33-04Panel
5For RMP:
•Unscrew the connector sheet (see
Figure 6-16).
•Unscrew the earth cable (33-05) from
instrument frame and disconnect the
mains power cable from the power
switch (see Figure 6-18, ‘Power switch
for RMP’).
Figure 6-16 Connector shield on RMP instrument
new
or
old
J11
red
black
Figure 6-17 Connection to Optibo
•Disconnect the control lamp cable from
the control lamp (access from the right
access door).
6Disconnect the red and black power
cables from the Optibo/Optibo Power
(see Figure 6-17, ‘Connection to
Optibo’).
7For old generation of power modules:
disconnect the power fail cable from
the Optibo/Optibo Power.
Note: for the new power modules the
jumper J11 must be set on Optibo/
Optibo Power.
8Remove the power module by carefully
pulling forward.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 35
Page 86
6 – Replacement of Spare Parts, Repair
Installation
Switch off the mains power and disconnect the mains power connection of the
instrument before performing any work!
1Install in reverse order.
In doing so, observe the directions given hereafter:
#3
#4
•For RMP: when installing the power
switch, note the installation diagram
A
33-07
(33-06) and the correct orientation of
the switch according to Figure 6-18.
33-02
correct
Figure 6-18 Power switch for RMP
P2P1
view A
33-06
33-05
33-02Power switch
33-05Earth cable
33-06Installation diagram
33-07Mains power cable
•When fixing the power module to its place: tighten the fixing screws (33-03) first,
hereafter the fixing screws (33-01).
Do not forget the sealing washers on all top screws (33-01).
(The item numbers mentioned refer to Figure 6-15).
•Tip for leading the power cables easily through the cable channel: fix the cables
together with a strip of adhesive tape.
Refer to Figure 6-17, ‘Connection to Optibo’ for correct cable connection.
•Before connecting the instrument to the mains, make sure the power switch is
switched off.
2Ensure operating readiness: control and carry out following settings:
(For detailed information refer to the Genesis Instrument Software Manual.)
StepTest or Setup
1Random move
Attention: never perform a random move test with a Genesis RMP!
2Liquid detection
6 – 36Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 87
6.3.5Electronic Boards
ATTENTION
Observe precautions for handling electrostatic discharge sensitive devices.
Wear a wrist strap.
6 – Replacement of Spare Parts, Repair
Spare Parts
Covered in This
Section
For spare parts list with part numbers refer to Section 7.1.5, ‘Spare Parts Electronic
Boards’.
Following items can be replaced as spare parts:
•Optibo
•Optibo Power
•Fuses
•CU board
•X-DC-Servo board
•Supervisor board
•Uniport board
•CANDI board
•SMIO/SAFY
The PosID 1 CU board and the PosID 2 CU board are described in Section 6.9.3,
‘Electronic Boards for PosID 1’ and Section 6.10.3, ‘Electronic Boards for PosID 2’.
Optibo,
Optibo Power
Optibo: for all RSP, all RMP and for RWS 100
Optibo Power: for RWS 150/200
The Optibo is located in the left service
compartment. On its rear side it holds the
CU board and, plugged into latter, the X-
39
DC-Servo board which controls the Xmotor(s) for LiHa and RoMa.
39Optibo/Optibo Power
Figure 6-19 Optibo or Optibo Power location
Removal
1Make sure the instrument is switched off and the mains power connection is
disconnected.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 37
Page 88
6 – Replacement of Spare Parts, Repair
2Disconnect all cables from Optibo/Optibo Power.
Do not forget to disconnect the Diluter cable on the solder side of the PCB.
3Disconnect all cables from the CU board.
4Remove Optibo/Optibo Power fixing screws, take care not to damage the PCBs
connected to the Optibo/Optibo Power rear side.
5Disconnect CU board from Optibo/Optibo Power as described later in this section.
Installation
1Make sure the instrument is switched off and the mains power connection is
disconnected.
2Make sure the CU board and X-DC-Servo board (address switch must be set to #5)
are connected to the Optibo/Optibo Power rear side.
3Control if the jumpers J12 and J13 on Optibo/Optibo Power have been set correctly
according to the configuration of the instrument (see Section 6.2.2, ‘Jumper and
Address Settings Overview’).
4Install in reverse order.
For correct cable connection see Figure 6-20, ‘Optibo cable connections’ and
Figure 6-22, ‘CU board cable connections’.
Power fail cable
(if present)
Power cable (red)
Power cable (black)
39
RoMa flex cable
LiHa flex cable
CAN-bus cable for options
or CANDI, Uniport
Figure 6-20 Optibo cable connections
J12 J13
PC cable (RS 232)
CAN-bus cables
for options
40-01
40
Diluter cable
39
39Optibo
40CU board
40-01X-DC-Servo board (LiHa/RoMa)
5Ensure operating readiness: control and carry out following tests:
(For detailed information refer to the Genesis Instrument Software Manual.)
6 – 38Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 89
6 – Replacement of Spare Parts, Repair
StepTest or Setup
1CAN-bus resistance (see Section 6.2.3)
2RoMa test
3FWO pump test
4PosID test
5Incubator test
6Random move
Attention: never perform a random move test with a Genesis RMP!
FusesThe fuses are located on the Optibo or Optibo Power.
See Figure 6-21, ‘Fuses on Optibo and Optibo Power’ for correct allocation.
39-01
Option 1Option 1
39-02
Option 5
Figure 6-21 Fuses on Optibo and Optibo Power
39-01Fuse 6.3 AT
39-02Fuse 8 AT, sand filled
Option 2
Option 3
Option 4
CU board
Option 1 to 4
Diluter
LiHa
RoMa
CANDI and Uniport
or Option 5
39-01
Option 1
Option 2
39-02
Option 3
Option 4
Option 5
1Check if all the green LEDs near the fuses are on; otherwise the corresponding fuse
is blown.
.
Use sand filled 8 AT type fuses only – except for the fuse for the CU board, which is
6.3 AT and not sand filled.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 39
Page 90
6 – Replacement of Spare Parts, Repair
CU BoardThe CU board is located in the left service compartment. As it is connected to the solder
side of the Optibo or Optibo Power, it can only be accessed by removing the Optibo/
Optibo Power.
The X-DC-Servo board, which controls the X-motors of the LiHa and RoMa, is plugged
into the CU board.
Special Tools
•IC release tool
Removal
1Switch off instrument.
2Remove Optibo/Optibo Power as described earlier in this section (see Optibo,
Optibo Power).
3Unscrew the CU board fixing screws. With a pair of pliers press the barbs on the two
board clips, carefully remove the CU board from the Optibo/Optibo Power.
4Remove X-DC-Servo board.
5Use IC release tool to remove the EEPROM (IC6) if you want to install it on the new
board (see Figure 6-22, ‘CU board cable connections’).
Use of EEPROM from the replaced CU board ! an error might be carried forward.
Use of new EEPROM ! you loose all settings.
Installation
Spare CU boards may be delivered with an older firmware version. In this case,
upgrading of the firmware is necessary.
LEDs
J1
IC6
Door lock cable left
(RMP)
RS-485 cable
I/O cable
40-01
X-motor cable RoMa
X-motor cable LiHa
Door lock cable right
(RMP)
J22
Figure 6-22 CU board cable connections
40CU board40-01X-DC-Servo board
6 – 40Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
40
Page 91
6 – Replacement of Spare Parts, Repair
1Control if the jumpers J1 and J22 on the CU board have been set correctly
according to the configuration of the instrument (see Section 6.2.2, ‘Jumper and
Address Settings Overview’).
2X-DC-Servo board: check if address switch is set to #5.
3Install in reverse order.
Refer to Figure 6-22, ‘CU board cable connections’ for correct cable connections.
4Ensure operating readiness: control and carry out following tests:
(For detailed information refer to the Genesis Instrument Software Manual.)
StepTest or Setup
1CAN-bus resistance (see Section 6.2.3)
2‘Check Error States’ as described later in this section
3Check reference positions
– LiHa
– RoMa
4Liquid detection test
5Disposable tips (DiTi eject test)
6Precision test
7RoMa
8FWO pump test
9LICOS test
10PosID test
11Door lock test (RMP only)
12Random move
Attention: never perform a random move test with a Genesis RMP!
Check Error States
•Check the LEDs on the CU board according to the following table:
Green LEDRed LEDError
blinkingoffNormal operation
stays on or offstays on or offFatal hardware error
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 41
Page 92
6 – Replacement of Spare Parts, Repair
X-DC-Servo Board The X-DC-Servo board – which controls the X-motors of the LiHa and RoMa – is
plugged into the CU board that in turn is connected to the solder side of the Optibo/
Optibo Power.
It is not necessary to remove the Optibo/Optibo Power to access the X-DC-Servo
board.
Removal
•From inside of the instrument remove the X-DC-Servo board from the CU board.
Installation
4
5
3
6
2
7
1
8
0
9
F
A
E
B
D
C
1Set address switch of the X-DC-Servo
board to #5.
2From within the instrument, connect the
40-01
X-DC-Servo board to the CU board.
40
39Optibo/Optibo Power
40CU board
39
40-01X-DC-Servo board
Figure 6-23 X-DC-Servo board, connected to the CU board
3Ensure operating readiness: control and carry out following test:
(For detailed information refer to the Genesis Instrument Software Manual.)
StepTest or Setup
1Random move
Attention: never perform a random move test with a Genesis RMP!
6 – 42Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 93
6 – Replacement of Spare Parts, Repair
Supervisor Board,
Uniport Board,
CANDI Board
For RMP Instruments
41,
37,
38
15
Figure 6-24 PCB location for RMP instruments
For RSP/RWS Instruments
39,
37,
38
The Supervisor and the Uniport boards are
located in the right service compartment of
the add-on module. The CANDI board may
be installed either in the left service
compartment (directly connected to the
Optibo/Optibo Power) or in the right service
compartment (connected to the Uniport).
15Add-on module
37Uniport board
38CANDI board
41Supervisor board
There is no Supervisor board necessary.
The Uniport and CANDI boards may be
installed either in the left service
compartment (directly connected to the
Optibo/Optibo Power) or in the right service
compartment (connected via cable to the
Optibo/Optibo Power).
1Left access door
37Uniport board
1
Figure 6-25 PCB location for RSP/RWS instruments
38CANDI board
39Optibo/Optibo Power
Special Tools
•IC release tool
Removal of Supervisor, Uniport, CANDI Board
The Firmware version on the new Supervisor board must be identical with the version
on the now installed Supervisor board. Therefore check the Firmware version prior to
replacing this PCB.
Default Firmware version on all new Supervisor PCBs is V1.21. However, an EPROM
with the Supervisor Firmware V1.11 is enclosed to the delivery of a new board.
1For Supervisor board:
•If the Firmware version of the installed PCB is different to the one on the new
PCB, remove EPROM (IC7) from the installed PCB using an IC release tool.
•Disconnect LICOS and waste tubing from Supervisor board.
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 43
Page 94
6 – Replacement of Spare Parts, Repair
2Disconnect all cables from the PCB to be removed.
3Remove appropriate board fixing screws.
4Carefully disconnect the board from the other connected board and remove it.
Installation
RSP, RWS
39
Alarm
device J25
2nd RT incubator
J16, J17, J12
RT incubator
37
Reader
Washer
OEM
Balance
Balance
OEM
Washer
Reader
38
RMP
41
IC 7 EPROM
Waste
(LICOS)
System
(LICOS)
J5 Niveau
CAN-bus
cable
Figure 6-26 Cable connections on Supervisor, Uniport and CANDI boards
37Uniport board39Optibo/Optibo Power
38CANDI board41Supervisor board
6 – 44Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 95
6 – Replacement of Spare Parts, Repair
1For Supervisor board:
•If necessary: install the EPROM (IC7) with the correct Firmware version on the
PCB.
•Control if address switch of the Supervisor board is set to #1.
2Install in reverse order.
For correct cable connection refer to the respective diagram in Section 9.2.2,
‘Electronic Boards’ or to Section Figure 6-26, ‘Cable connections on Supervisor,
Uniport and CANDI boards’.
3For Uniport:
•Ensure operating readiness: control and carry out following tests:
(For detailed information refer to the Genesis Instrument Software Manual.)
StepTest or Setup
1CAN-bus resistance (see Section 6.2.3)
2Washer test
3Reader test
SMIO/SAFY
The SMIO/SAFY is located in the right
service compartment (connected via cable
to the Optibo/Optibo Power).
The SMIO/SAFY is part of the access
option for RSP and RWS instruments.
Please refer to Document Doc ID 392 330
for further information.
Figure 6-27 SMIO/SAFY
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 45
Page 96
6 – Replacement of Spare Parts, Repair
6.4Liquid System – Disassembly and Assembly Procedures
6.4.1Overview
Definition of
Liquid System
Standard Liquid
System for RSP
and RWS
The term “liquid system” refers to all instrument modules and parts which contain or
directly influence liquid. These main components are:
•LiHa 1 and LiHa 2(see Section 6.5 and Section 6.6)
•Tips(refer to respective Instrument Operating Manual)
•Tubing system(see Section 6.4.2)
•Diluter/Dilback(see Section 6.4.3)
•MPO/FWO*(see Section 6.4.4)
•Low volume option*(see Section 6.4.5)
•6-way valve option*(see Section 6.4.6)
Components marked with an asterisk (*) are optional and therefore not available on all
instruments.
The following figure shows the main components of a standard liquid system for RSP or
RWS instruments with 8 tips and optional FWO/MPO*.
30-03
30-01
35
Figure 6-28 Overview of standard liquid system for RSP/RWS with 8 tips and MPO/FWO
4Diluter30-03 Pipetting tubing
8LiHa1, LiHa 230-04 Wash station
9Tips30-05Waste tubing
30-01Aspiration tubing32Dilback
30-02Interconnecting tubing35MPO/FWO
30-02
8
9
4, 32
30-04
30-05
6 – 46Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Page 97
6 – Replacement of Spare Parts, Repair
Standard Liquid
System for RMP
The following figure shows the main components of a standard liquid system for RMP
instruments with 8 tips and optional MPO/FWO*.
30-01
35
Figure 6-29 Overview of standard liquid system for RMP with 8 tips and MPO/FWO
30-02
4, 32
30-05
30-03
8
9
30-04
4Diluter30-03 Pipetting tubing
8LiHa1, LiHa 230-04Wash station
9Tips30-05Waste tubing
30-01Aspiration tubing32Dilback
30-02Interconnecting tubing35MPO/FWO
6.4.2Tubing System
This section describes the customary tubing system which is available in three material
variants: standard, high resistant type A, high resistant type B.
For the MPO/FWO tubing system refer to Section 6.4.4.
For the Low volume option tubing system refer to Section 6.4.5.
For 6-way option tubing system refer to Section 6.4.6.
Spare PartsFor spare parts list with part numbers refer to Section 7.2.1, ‘Spare Parts Tubing
Systems’.
Following items can be replaced as spare parts:
•Tubing and fittings for Standard Tubing
•Tubing and fittings for High Resistant Tubing Type A (FEP/PVDF)
•Tubing and fittings for High Resistant Tubing Type B (FEP/PP)
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 47
Page 98
6 – Replacement of Spare Parts, Repair
Standard Tubing
4 Tips
30-08
8 Tips
30-03
High Resistant
Tubing Type A and
Type B
30-02
30-01
30-01
30-06
Figure 6-30 Standard tubing
30-01Aspiration tubing30-05Waste tubing
30-02Interconnecting tubing30-06I-connector
30-03Pipetting tubing30-07 Y-connector
30-04Wash station30-08Distributor 1 to 4
4 Tips
30-07
30-10
30-08
30-07
30-08
8 Tips
30-10
30-07
30-02
30-04
30-05
30-09
30-01
Figure 6-31 High resistant tubing type A and B
30-01Aspiration tubing30-08Distributor 1 to 4
30-02Interconnecting tubing30-09Screw plug
30-07Distributor 1 to 230-10/11 Tubing to distributor 1 to 4
6 – 48Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
30-02
30-11
30-01
30-08
30-02
Page 99
Replacement
6 – Replacement of Spare Parts, Repair
WARNING
The liquid system may contain compounds hazardous to your health.
Make sure the complete liquid system has been properly decontaminated before you
perform any service tasks.
..
1Empty the liquid system.
2Switch off the instrument.
3Replace relevant tubing.
For detailed descriptions of the
exchange routines please refer to the
Genesis Maintenance and Service
Logbook Doc ID 390 924.
•When replacing the pipetting
#1
800 mm
#1#8
#8
tubing: diluter #1 (on the far left) is
connected to tip #1 (the closest to
the back of the instrument), diluter
#2 to tip #2 etc.
•Adjust all pipetting tubing length to
800 mm.
Do not cut the remaining tubing, but
pull it back and place it in the liquid
pan of the instrument.
4Ensure Operating Readiness:
Ensure Operating
Readiness
•Tightness check
•FaWa pump test
•Precision test
… as described later in this section.
Figure 6-32 Installing the pipetting tubing
StepTest or Setup
1– Tightness check: fill/flush system and check all tubing connections
– FaWa pump test
Necessary after replacement of
•Pipetting tubing
•Interconnecting tubing
•Aspiration tubing
2– Precision test
Necessary after replacement of
•Pipetting tubing
•Interconnecting tubing
August 2002Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.06 – 49
Page 100
6 – Replacement of Spare Parts, Repair
6.4.3Diluter and Dilback
Spare PartsFor spare parts list with part numbers refer to Section 7.2.2, ‘Spare Parts Diluter/
Dilback’.
Following items can be replaced as spare parts:
•Diluters
XP 3000
XP 3000 plus
•Diluter spare part
3-way valve
Syringe with flat caps, Syringe with conical caps
Syringe caps, flat, Syringe caps, conical
•Dilback-8
Diluter XP 3000,
XP 3000 Plus
Diluter type XP 3000 is installed on RSP, RMP and RWS instruments up to serial
number 4999.
Diluter type XP 3000 plus is installed on RSP, RMP and RWS instruments with serial
numbers ≥ 5000.
Overview
30-03
4
30-02
4-01
4-02
4-03
4Diluter XP 3000 or XP 3000 plus
4-013-way valve
4-02Syringe cap
4-03Syringe
30-02Interconnecting tubing (in)
30-03Pipetting tubing (out)
Figure 6-33 Diluter – principal components
Removal
1Empty the liquid system.
2Switch off the instrument.
3Open the top cover (2) of the instrument.
6 – 50Genesis RSP, RWS and RMP Service Manual – Doc ID 391895 V1.0August 2002
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.