This chapter points out the purpose of the manual, specifies the product the
manual deals with and who the manual is intended for. Furthermore, it explains
the symbols, conventions and abbreviations used and offers other general
Purpose of This
Manual
information.
This manual describes the Balance Kit SAG 285/01and provides all information
required for the proper installation and setup of the kit as a preparatory step for
the Gravimetric Test. In addition, it provides the necessary references to the
document(s) where the Gravimetric Test is described in detail.
Target GroupThis manual is addressed to advanced users and authorized field service
engineers (FSEs) who wish to run the Gravimetric Test on instruments that are
equipped with a liquid handling arm (LiHa).
ScopeThis Operating Manual is limited to the following:
Applicable to the installation and setup of the Balance Kit for the Weighing
Module SAG 285/01 from Mettler Toledo AG.
Software for Gravimetric Test: Instrument Software, version 4.2 or later.
Symbols and
Conventions
Cross-references appear as follows: e.g. “Refer to section 1.1.1, 1-1”
–1.1.1 refers to the corresponding chapter number.
–The symbol
denotes the “page number”.
–1-1 refers to the page number, whereas the first number stands for the
chapter number (chapter 1 - page 1).
Command Sequences for functions of the Instrument Software are printed in
bold type, e.g. Select Liquid System > Balance.
Note: The symbols pertaining to safety (WARNINGS and ATTENTIONS) are
explained in Chapter 2 “Safety”,
2-1.
For Your SafetyBefore performing any work on or with the Balance Kit, first read the
Operating Manual carefully, in particular chapter 2 “Safety”.
This section provides a list of documents which are needed or may be useful in
connection with the Balance Kit.
System
Manuals
Instrument
Software
Manuals
Operating Manuals of instrument on which Balance Kit is to be installed.
Service Manual of instrument on which Balance Kit is to be installed.
Operating Manual of balance (provided by the manufacturer)
Operating Manual for Balance Kit for Mettler SAG 285/01 (Doc ID 393278)
Software Version V4.2 (Genesis Classic, Doc ID 390 791)
Software Version V4.3 (Genesis Freedom, Doc ID 392 179)
Software Version V4.5 (Freedom EVO, Genesis Freedom & Classic,
Doc ID 390 791)
Software Version V4.6 and later (Freedom EVO, EVOlyzer, Genesis Freedom
& Classic, Doc ID 392 888)
1.2 Trademarks
The following product names and any registered and unregistered trademarks
mentioned in this manual are used for identification purposes only and remain the
exclusive property of their respective owners:
Windows
SAG 285/01, AG 285 and AG 245 are trademarks of Mettler Toledo AG.
UC-232A is a trademark of Aten International Company Ltd.
EX-1331 and EX1334 are trademarks of Exsys Vertriebs GmbH.
TM
is a registered trademark of Microsoft Corporation
This chapter specifies the intended use of the Balance Kit and points out the
safety concept. It contains general rules of behavior and warnings from hazards
pertaining to the use of the kit.
Please note that the Balance Kit itself does not contain any hazardous parts.
However, it is installed on instruments where it is necessary to perform the
Gravimetric Test. Such instruments are equipped with parts that can move with
great force and at considerable speed. Moreover, these instruments often handle
toxic or other chemically and biologically hazardous substances which require
adequate safety measures.
As a consequence, the safety of users and personnel can only be ensured if the
safety instructions in this manual and the relevant hardware and software
manuals of the instruments on which the Gravimetric Test is run, are strictly
observed and followed.
Therefore, the relevant manuals must always be available to all persons
performing the tasks described herein.
2.1 User Qualification
All users (laboratory personnel, Tecan authorized field service engineers (FSEs)
etc.), must be qualified and trained to install and use the Balance Kit for the
Gravimetric Test.
In particular, users must fulfill the following qualifications:
They must have a thorough knowledge of the application run on the system.
They must be familiar with the good laboratory practice guidelines.
The must be familiar with the relevant setup and test procedures described in
the “Instrument Software Manual”.
They must know how to set up and operate the Balance that is to be used
together with this kit.
They must have read and understood the instructions in this Operating
Manual.
Only persons who meet the qualifications prescribed here are authorized to
perform the work described in this Operating Manual.
Note: Tecan recommends that users attend an appropriate training course.
•Please ask your nearest Tecan representative about the available courses.
Page 8
2 - Safety
Notices and Symbols
2.2 Notices and Symbols
2.2.1 Warning Notices Used in This Manual
The symbols used for safety-related notices have the following significance:
WARNING
Symbols
WARNINGS appear as follows:
WARNING
Generally, the triangular warning symbol indicates the possibility of personal injury
or even loss of life if the instructions are not followed.
Whenever possible, the symbol indicates the hazard a person is exposed to more
specifically. The symbols used in this Operating Manualhave the following
significance:
With the general “Read This!” symbol, ATTENTIONs indicate the possibility of
equipment damage, malfunctions or incorrect process results, if instructions are
not followed.
2.2.2 Warning Notices Attached to the Product or Its Surroundings
There are no warning notices attached to the Balance Kit.
2.3 Use of the Product
Intended UseThe Balance Kit described in this manual is to be used exclusively together with
the Mettler-Toledo balance model SAG 285/01.
Improper UseThe Balance Kit described in this manual is not suitable for other balance models
(e.g. Mettler AG 285 or AG 245).
Legal
Regulations
Duty of
Maintenance
and Care
Observe
Specific Safety
Instructions
2.4 General Safety Rules
Legal regulations, such as local, state and federal laws which prescribe the use or
application as well as the handling of dangerous materials in connection with the
Balance Kit must be strictly followed.
The user is responsible for ensuring that the Balance Kit and the balance itself are
used in proper condition only. Maintenance, service, and repair jobs on the
instrument on which the Balance Kit is to be used must be performed with care
and on schedule, and by authorized personnel only.
When setting up and using the Balance Kit pay due attention to the safety
instructions as laid out in the following manuals:
Operating Manual of the specific instrument
Service Manual of the specific instrument
Instrument Software Manual
Operating Manual of the balance
OverviewThe following figure provides a quick overview of both the Balance Kit and the
This chapter introduces the reader to the Balance Kit and its main components. It
contains technical data, requirements and performance data.
3.1 General Description
The Balance Kit is a set of components that are needed to adapt the Mettler
Toledo Weighing Module SAG 285/01 to the instrument so that it can be used for
the Gravimetric Test.
SAG 285/01 from Mettler Toledo. Brief descriptions are given in the sections after
the figure.
L
K
I
A
B
C
H
J
Fig. 3-1 Parts of Balance Kit + SAG285/01 (Weighing Cell and Control Unit)
Draft shield
BRing for evaporation compensation
CVessel
DCentering ring
EWeighing pan (Mettler)
FFoot with counternut
The following table lists the parts the Balance Kit consists of. Also refer to
Fig. 3-1, 3-1.
Tab. 3-1 Parts of the Balance Kit
Item in
Figure
A1Draft shieldProtects the vessel (C) with the liquid to be weighed
B1Evaporation
C2VesselContains the liquid to be weighed during the Gravimet-
D1Centering ringHolds the evaporation ring (B) in place
F1Adapter platePlate with two handles and positioning slots that can
G4Foot + counter-
QuantitiyDesignationExplanation
from air drafts.
This ring is placed on the Weighing Cell (H) as shown
ring
nut
in the figure. It contains a groove that can be filled with
distilled or de-ionized water. During operation, part of
this liquid evaporates and thus increases the air
humidity under the draft shield. This, in turn, has the
effect that less liquid evaporates from the vessel during the Gravimetric Test. In this way, the accuracy of
the test results is increased indirectly.
ric Test. The vessel rests on the weighing pan (E).
be placed in the appropriate grid position on the worktable. It also has four positioning holes that allow positioning the Weighing Cell (H) properly on the plate.
The four feet are standard knurled screws with a plastic head. They are screwed to the bottom of the Weighing Cell (H) and can be used to level out the Weighing
Cell. The counternuts serves for securing the feet.
Ordering the
Balance Kit
The Balance Kit (or certain parts of it) can be ordered from Tecan or via a local
Tecan representative. For more details see section
7 “Spare Parts and Accessories”, 7-1.
Note: The Balance Kit does not comprise the Weighing Module (weighing cell +
control unit), nor are the connection cable and other accessories part of the kit.
These items must be ordered directly from Mettler Toledo AG (see next section).
The SAG 285/01 is a product of Mettler Toledo AG and has been available since
January 2004. This SAG 285/01 consists of a Weighing Cell (H) and a separate
Control Unit (J) that are interconnected with a cable. Thanks to this separation a
higher precision and stability can be attained.
The SAG 285/01 is a substitute for the predecessor model AG 285 which
incorporated the weighing cell, the display and the operating and control elements
in one unit. Note, that this model is no longer available.
The following two tables list the parts that must be ordered directly from Mettler
Toledo:
Tab. 3-2, 3-3 lists the parts the SAG 285/01consists of.
Tab. 3-3, 3-3 lists the accessories that must be ordered separately.
See also Fig. 3-1 , 3-1.
Tab. 3-2 Weighing Module SAG 285/01
Item in
figure
H1Weighing CellIs placed on the adapter plate (F) and actually weighs
J1Control UnitThe control unit is separated from the weighing cell
I1Connection
QuantityDesignationExplanation
the liquid in the vessel.
(H). It contains a digital display and the necessary controls that allow operating the whole Weighing Module.
Connects the Weighing Cell to the Control Unit. Please
cable 1.5 m
note that a 5 m cable is available as an option (see
below)
Ordering the
SAG 285/01 and
Accessories
L1Power supply This power supply unit is connected to the control unit.
Tab. 3-3 Accessories for the SAG 285/01
Item in
Figure
E1Weighing pan,
QuantityDesignationExplanation
Carries the vessel with the liquid to be weighed.
∅ 50 mm
K1Cable LC-RS9Cable with local CAN bus to RS-232 adapter that con-
nects the Control Unit to the computer.
I1Connection
cable 5 m
--1Transport caseOptional accessory that allows transporting the
Optional. Can be used instead of the 1.5 m cable to
connect the Weighing Cell to the Control Unit.
SAG 285/01.
See section 7 “Spare Parts and Accessories”, 7-1.
InstrumentThe Balance Kit together with the SAG 285/01can be used on the following
instruments types: Freedom EVO instruments, Genesis Freedom and Genesis
Classic instruments.
ComputerHardware requirements as specified in the Instrument Software Manual that
corresponds to the installed version of the Instrument Software.
Possible interfaces for connecting the instrument and the SAG 285/01 see below.
Interfaces, Adapters
There is more than one possibility to connect the instrument and the SAG 285/01
to the computer.
Two R S-23 2
Ports
USB PortsTo connect the instrument and/or the SAG 285/01 to the computer via USB ports,
If the computer has two free RS-232 ports, both the SAG 285/01 and the
instrument can be connected directly.
Note: Depending on the connectors at the instrument cable and the computer,
adequate 9 to 25-pole RS-232 adapters and / or gender changers may be
needed.
an appropriate USB to Serial converter is needed.
Recommended USB to Serial converters:
Aten UC-232A USB to RS232 Converter. Has one RS-232 connector and
plugs into a USB port on the computer.
Exsys EX-1331 USB to RS232 Converter. One RS-232 and 1 one USB
connector.
Exsys EX-1334 USB to RS-323 Converter. Has four RS-232 connectors and
plugs into a single USB port on the computer.
The above converters feature the following:
Data rate: Up to 230 Kb per second
RS-232 connectors: Male DB9
Compliant with USB Specifications V1.1
Suitable for Windows 2000 and XP
Note: A mixed connection (instrument to RS-232, SAG 285/01 via USB) is also
possible.
If the computer has only one free RS-232 port and no USB port, the SAG 285/01
may be connected via the Uniport board, provided the following conditions are
fulfilled:
Instrument Software V4.5 or earlier is installed on the computer.
The instrument is equipped with a Uniport board.
Note:
•Apart from Genesis RMP or instruments used in customized solutions, Tecan
instruments are not normally equipped with a Uniport board.
•If possible, a separate Uniport board may be installed for the Gravimetric
Test.
•The connection via the Uniport is not supported by the Instrument Software
version V4.6 or later.
3.3.2 Software
Version V4.2 or later is required.
Note: The connection of the SAG 285/01 via Uniport is no longer supported from
Instrument Software Version 4.6 or later.
According to specifications of installed version of Instrument Software.
Drivers for
USB / RS-232
Converters
The delivery of the recommended USB to RS-232 converters includes drivers that
are suitable for Windows 2000 and Windows XP.
To carry out the Balance Setup function and the Gravimetric Test with the
Instrument Software you need the following access rights:
Version V4.6x or earlier: No password required.
Version V4.7 or later: From version V4.7 onwards, all users need a user
account, together with a user name and a password.
–There are several predefined user groups with different authorization
levels.
–The Balance Setup and the Gravimetric Test can be carried out by users
with SnS_Customer access rights.
Note: For detailed information about user accounts, passwords and access rights
refer to the Instrument Software Manual (for software version V4.7 or later),
section “Tecan User Administration System”.
3.3.4 Syringes and Liquid for Gravimetric Test
Syringes: 250 µl, 500 µl, 1 ml.
System liquid: Demineralized or de-ionized and degassed water. Water
temperature should correspond to ambient temperature.
Pipetting liquid: Tap water. Water temperature should correspond to ambient
This chapter describes how the Balance Kit is installed and gives instructions on
initial operation.
4.1 Setting up the Balance Kit and the SAG 285/01
4.1.1 Overview
The following figure provides an overview of the Balance Kit and the SAG 285/01
Weighing Module. For a description and the detailed specifications see section
3.1, 3-1.
L
K
I
H
A
B
C
D
E
F
J
G
Fig. 4-1 Parts of Balance Kit + SAG285/01 (Weighing Cell and Control Unit)
Draft shield
BRing for evaporation compensation
CVessel
DCentering ring
EWeighing pan (Mettler)
FFoot with counternut
4.1.5 Connecting the Weighing Cell to the Control Unit
To connect the SAG 285/01 electrically to the other parts of the system proceed
as follows (also see Fig. 4-1 , 4-1):
Weighing Cell
to Control Unit
1Place the Control Unit (J) on a table beside the instrument.
2Lead the connection cable (I) from the Weighing Cell (H) to the Control Unit
(J). A possible way to lay the cable is shown in Fig. 4-1 , 4-1.
3Plug the connectors at both ends of the cable into the corresponding sockets
on the Weighing Cell and the Control Unit and fix the connectors with their
fixing screws.
Power Supply4Connect the power supply unit (L) to the Control Unit (J) as shown in
Fig. 4-1 , 4-1.
4.1.6 Connecting the SAG 285/01 and the Instrument to the Computer
Depending on the computer model and its interfaces there is more than one
possible way to connect the SAG 285/01 and the instrument to the computer.
Note: The Optibo board shown in the following figures is located behind the left
access door of the instrument.
Computer with
2 RS-232 Ports
If two RS-232 ports are available on the computer, both the instrument and the
Control Unit can be connected directly as shown in the following figure. This is the
simplest and least problematic arrangement.
Optibo-Board
L
RS-232
M
I
J
H
Computer
RS-232
RS-232
Fig. 4-2 Direct connection to RS-232 ports
HWeighing Cell
ICable Weighing Cell to Control Unit
JControl Unit
Modern computers, particularly notebooks, often have no RS-232 ports at all, but
are equipped with two or more USB-ports. In such a case the instrument and the
Control Unit must be connected via appropriate USB to Serial converters
according to the following figure.
Note: The connection of the computer to the instrument via an USB to Serial
converter will reduce the communication speed between the instrument and the
computer. As a consequence, the execution time for the Gravimetric Test may
increase by 10 to 20 %, compared to the time it takes when the instrument is
connected to the computer directly.
Optibo-Board
L
J
K
RS-232
M
I
H
Instrument
RS-232
RS-232
Cnv
Notebook
Fig. 4-4 Connection via USB to Serial converter
HWeighing Cell
ICable Weighing Cell to Control Unit
JControl Unit
KLC-RS9 (Control Unit to computer)
Recommended converters: See sections 3.3.1, 3-6 and 7.2.2, 7-4.
USB
LPower supply for SAG 285/01
MCable instrument to computer
Cnv USB to serial converter
As has been mentioned earlier (see section 3.3.1, 3-6) the SAG 285/01 may be
connected via the Uniport board under the following conditions:
Uniport board must be installed and connected to the Optibo board as shown
in the following figure.
The connection via the Uniport is only supported by Instrument Software
versions V4.5 or earlier.
Note: This configuration requires a gender changer between the LC-RS9 cable
(K) and the Uniport board. A special intermediate cable (N) that serves as a
gender changer is available from Tecan (see section 7.1.2,
Uniport
L
K
J
RS-232
7-2).
Optibo-Board
RS-232
N
M
I
H
Computer
RS-232
Fig. 4-5 Connection via the Uniport
HWeighing Cell
ICable Weighing Cell to Control Unit
JControl Unit
KLC-RS9 (Control Unit to Uniport)
LPower supply for SAG 285/01
MCable instrument to computer
NIntermediate cable between LC-RS9
Instrument
and Uniport (gender changer)
ProcedureTo connect the SAG 285/01 and the instrument to the computer:
1Decide which of the following connection methods is suitable in your case:
–Direct connection to two RS-232 ports (see Fig. 4-2 , 4-3).
–Mixed connection, instrument to RS-232 and SAG 285/01 via a USB to
Serial converter (see Fig. 4-3 , 4-4).
–Connection via USB to Serial converter (see Fig. 4-4 , 4-5)
–Connection of SAG 285/01 via Uniport (see Fig. 4-5 , 4-6).
2Connect the units together.
3If you use a USB to Serial converter you must also install the appropriate
driver software on the computer. Follow the instructions provided by the
supplier of the converter.
4.1.7 Set Communication Protocol and Balance Parameters
Next you must set up the following:
Communications protocol
Balance parameters
Communication
Protocol
The communication protocol can be adapted by means of the three rotary
switches located inside the protocol converter of the LC-RS9 cable (over which
the Control Unit is connected to the computer).
Weighing process adapter1Suitable for small quantities
ReproducibilityStdStandard reproducibility
Weighing unit 1unit 1 mgBoth weighing units must be mg
Weighing unit 2unit 2 mg
Automatic zero correctionaz. offNo automatic zero correction
Automatic offaoff -No automatic switch off
Start modequ startQuick start. No display check at
startup.
ATTENTION
Do not reset the parameters unnecessarily, otherwise you will have to set them
again!
4.1.8 Balance Setup with the Instrument Software
PurposeThe purpose of this setup is to establish the proper communication with the
SAG 285/01.
NomenclatureThe Instrument Software uses the term “Balance” for the Weighing Module,
regardless of whether the Weighing Cell and the Control Unit are separated or
incorporated in the same device.
Required for
the Setup
To run the setup procedure you need the following:
Instrument Software Manual that corresponds to the installed version.
–A PDF-file of the manual is stored on the installation CD of the Instrument
Software.
–Operating Manual of the SAG 285/01 Weighing Module.
If version V4.7 or later is installed on the computer you need a user account
with a user name and a password.
–Required access rights: SnS_Customer access level or higher.
–If necessary turn to the system administrator for an appropriate user
account and the necessary access rights.
ProcedureThe procedure varies slightly, depending on the version of the installed Instrument
Software: Always follow the instructions provided in the Instrument Software
Manual and/or the SAG 285 Operating Manual.
To perform the Balance Setup:
1Switch on the computer and the SAG 285/01.
2Tare the Weighing Cell (see SAG 285 Operating Manual).
4From the Windows task bar start the Setup & Service software (which is part
of the Instrument Software package) with
Start > Programs > Instrument Software > Setup & Service
5If the Instrument Software V4.7 or later is installed:
–You will be prompted to log in with your user name and password.
–If you log in for the first time you must change the password.
6Call up the function Liquid System > Balance. Follow the instructions in the
Instrument Software Manual to set up and test the communication with the
balance.
4.1.9 Loading the Worktable Map
Setting up the Balance Kit and the SAG 285/01
4 - Putting into Operation
General
Procedure
In order to perform the Gravimetric Test you must load a suitable worktable map.
Refer to the Instrument Software Manual for details.
1Call up the Setup & Service software function System > Load Worktable.
2Choose the appropriate worktable map for the Gravimetric Test (see
remarks below).
3If necessary, store the worktable map under a new name and use the
Worktable Complete Editor to adapt it to your needs.
Remarks
Instrument
Software V4.7
or later
Earlier Versions
of Instrument
Software
The Instrument Software V4.7 or later includes a worktable map that corresponds
to the adapter plate for the SAG 285/01 Weighing Cell. If necessary, you can
modify this map according to your needs with the Worktable Complete Editor.
If version V4.6 or earlier is used, you will probably not find a worktable map that
corresponds to the adapter plate for the SAG 285/01, but one that was suitable for
the predecessor model (for AG 285/245 and other models).
This previous adapter plate was 10 grid positions wide, whereas the SAG 285/01
plate occupies 6 grid positions. Moreover, the SAG 285/01 adapter plate (with
Weighing Cell and draft shield) is approx. 5 mm higher than the previous model.
Workaround:However, you can still use the old worktable map if you place the SAG 285/01
adapter plate into a grid position that corresponds to the middle of the old map.
Example: If the left grid position for the old adapter plate is 2, then place the
SAG 285/01 adapter plate in grid position 4.
Generally speaking: Left grid position of SAG 285/01 adapter plate = Left grid
position of AG 285/245 plate + 2.
Professional
Solution
A better solution is to store the existing worktable map under a new name and to
use the Worktable Complete Editor to adapt the following parameters:
Width of adapter plate: 6 grid positions
Overall height: 130 mm (corresponds to height of adapter plate with installed
Weighing Cell and draft shield, also see section 3.2.2, 3-5).
This chapter briefly explains the operating elements and gives instructions on how
to carry out the Gravimetric Test.
5.1 Operating and Display Elements
A
B
E
Operating
Elements
C
Fig. 5-1 Operating elements for Gravimetric Test
Computer display
BSAG 285 Control Unit
CKeyboard (not directly visible)
The Balance Kit itself has no operating and display elements of its own. Most of
the operating and display elements needed for the Gravimetric Test are shown in
D
DMouse
EPower switch of instrument (not
directly visible)
the above figure.
Power SwitchesThe location of the power switches of the computer and the instrument depend on
This chapter helps to resume operation after a minor problem has occurred with
the Balance Kit or the SAG 285/01 when used for Gravimetric Test and its
preparation. It lists possible problems, their probable cause and suggests how to
remedy the problems.
6.1 Troubleshooting Table
The following table lists problems and errors and gives hints on how to eliminate
them.
Tab. 6-1 Troubleshooting table
Problem / errorPossible causeCorrective measures
Weighing Cell wobblesFeet not adjusted
properly
No display on Control
Unit
Control Unit does not
react to changes in
weight on Weighing Cell
Control Unit not connected to mains
Control Unit not connected to Weighing
Cell
Level out the Weighing Cell by adjusting the
feet and secure with counternuts.
Connect the control unit via the power
supply unit.
Check connection between units and connect properly.
Values shown on computer screen do not correspond to those shown on
display of Control Unit.
Control Unit not connected to computer
Control Unit not set up
properly
Wrong communication
protocol
Uniport: Connection
via Uniport not correct
Connection via USB:
RS-232 to USB adapters to slow
USB adapters not
installed properly or
drivers not compatible with operating system
Check connection to computer.
Repeat setup.
Check the settings of the protocol converter
in the LC-RS9 cable.
Check the following connections:
• SAG 285/01 Control Unit to Uniport
• Uniport to Optibo board
Use recommended USB adapters.
Install adapters correctly according to
instructions provided by the supplier. Make
sure the correct drivers are installed.
This chapter lists the parts and accessories used in connection with the Balance
Kit and the Weighing Module SAG 285/01, including the ordering information.
Identify the parts from the following figure and look up the ordering information in
the tables later in this chapter.
Spare parts must be ordered from the corresponding manufacturer or from its
nearest representative:
Balance Kit: From Tecan. Addresses see section 8 “Customer Support”, 8-1
Parts and accessories for the SAG 285/01: Directly from Mettler Toledo.
USB to Serial Converters: From manufacturer or specialized distributor.
Note: This chapter only contains parts that can be replaced by the operator him/
herself. If you want to order Tecan parts that are not listed here, please contact
the Tecan Customer Support.
A
L
K
B
I
C
H
J
Fig. 7-1 Parts of Balance Kit + SAG285/01 (Weighing Cell and Control Unit))
Draft shield
BRing for evaporation compensation
CVessel
DCentering ring
EWeighing pan (Mettler)
FFoot with counternut
This chapter contains a glossary to explain terms and expressions used in this
Operating Manual.
–Instrument for determining the weight of something, scales.
–Setup function of the Instrument Software in which the communication of
the balance with the instrument is established so that the balance can be
used for the Gravimetric Test.
Balance Setup
Steps to be performed to adapt the balance parameters to the requirements of
an application (range, units, precision, etc.).
Communication Protocol
A standard set of rules that define how computers communicate with other
computers or devices. In order that connected computers and devices
“understand” each other they must use the same protocol.
Controller Area Network (CAN)
Network used to exchange information between connected devices.
Information is sent/received in the form of “messages” via a high-speed bus,
the CAN-bus. Each “message” consists of a “label” that is unique throughout
the network and a “data part”. From the label, the receiving device knows what
kind of information the data part contains.
Gravimetric Test
Test function implemented in the Instrument Software with which the precision
and the accuracy of the liquid transfer can be checked with the aid of an
external precision balance.
Instrument Software
Software package consisting of several interacting software programs (Setup
& Service, NPS Quick Test, Worktable Editor and Communication driver).
Protocol
See Communication Protocol
SAG 285/01
Precision balance, manufactured by Mettler Toledo, consisting of a Weighing
Cell and a separate Control unit.
Setup & Service Software
Part of the Instrument Software package with which the devices and options
installed on or connected to the instrument can be set up and tested.
Uniport
Interface board (used in Genesis RMP) with 4 RS-232 ports over which up to
4 external devices can be connected to the CAN bus of the instrument. Note
that the connection via Uniport is no longer supported by Instrument Software
V4.6 or later.
Universal Serial Bus (USB)
An external bus standard that supports high data transfer rates. A single USB
port can be used to connect up to 127peripherals of various sorts (e.g. mice,
modems, keyboards etc.). USB also supports Plug and Play installation.
The surface of the instrument on which various objects (carriers, tube racks,
wash stations etc.) are arranged.
Worktable Complete Editor
Graphical editor that allows creating, modifying and deleting worktable maps.
Worktable Map
A graphical representation of the worktable and the objects arranged on it.
The parameters of these objects (e.g. grid position, width, height) are needed
by the software which must know where they are placed. They also serve to
avoid collisions with moving arms, e.g. RoMa or LiHa.