Taylor 9306N User Manual

Infrared Lens
Infrared
Reading
Probe
White Light
Model 9306N Thermometer
Operating Instructions
Probe
Reading
Scan
(infrared)
Key
Probe Key
9306N
HACCP Check LED
Scan
Mode
Probe
NSF
Mode Key
In Non-contact Infrared Thermometer function (IRT Mode)
# The white light targeting feature will automatically turn on while the Scan button is pressed.
Distance:Spot (FOV)= 3:1 Emissivity = 0.1~1 Step.01 Wave Length = 8um-14um
Simply aim the thermometer at the target to be measured with the “Infrared Lens” and press the Scan (infrared) key to display the surface temperature. The distance to target ratio is 3:1; therefore the thermometer should be positioned as close to the target as possible.
While scanning, the newest temperature will be updated on the LCD and the measurement will continue as long as the Scan (infrared) key is depressed. When the Scan (infrared) key is released, the “Hold” icon will appear on the display and the last measurement will remain visible for 15 seconds before the display goes blank.
Mode Selection MIN
Minimum or Maximum Mode
Min/Max mode: single press Scan=Max; 2 press Scan = Min
The thermometer will display the minimum or maximum reading during the measurement period only until the Mode key is pressed. To utilize the maximum mode, please press the Scan (infrared) key > Mode key > Scan (infrared) key. Keep pressing the Scan (infrared) key for measurements. To utilize the minimum mode, please press the Scan (infrared) key > Mode key *twice > Scan (infrared) key. Keep pressing the Scan (infrared) key for measurements.
Lock Mode
The lock mode is particularly useful for continuous monitoring of temperatures.To utilize the lock mode, please press the Scan (infrared) key > Mode key *three times > Probe (infrared) key. The thermometer will continuously display the temperature for up to 60 minutes or until the Probe (infrared) key button is pressed. Othewise, when the Scan (infrared) key is released, the “Hold” icon will appear on the display and the last measurement will remain visible for 15 seconds before the display goes blank.
ºC or ºF Mode
To change the 'ºC' or 'ºF' mode, please press the Scan (infrared) key > Mode key *four times > Scan (infrared) key. The same steps can be taken when switching from ºF to ºC.
Emissivity
The infrared thermometer is supplied with a default emissivity of 0.95. The emissivity can be changed from 0.10 (10E) to 1 (100E).Note: non-contact infrared thermometers are not recommended for use in measuring the temperature of shiny or polished metals.To change the emissivity, please Scan (infrared) key > Mode key *five times > Scan (infrared) key for each 0.01 (1E) adjustment > Mode key.
Rubber Gasket (Screw’s location)
Battery Cover
o
C/ oF > EMIS
Leading the Way in Accuracy®
The thermometer is a non-contact infrared thermometer with a probe thermometer. You can select only one mode at a time but can change the mode at will. Please remember to keep away from children and do not use for safety related applications.
* Special Features:
1. Food Code Zone Display;
2. splash proof (IP65)
In-Contact Thermocouple Probe function (COT Mode)
Insert the probe into the item to be measured and press the Probe key to continuously display the temperature for up to 4 minutes. After that the device will automatically shut off to extend the battery life. Pressing the Probe key will interrupt the scanning to display the last temperature with a 'Hold' icon. If scan mode is active during the 4 minutes while using the probe, the auto-off will activate for both the scan and probe after 15 seconds.
1. Do not twist the probe or rotate the probe in wrong direction.
2. Placing stress on the probe may cause breakage.
3. After measuring high temperatures, the probe may remain HOT for a while. Allow the probe to cool before
touching.
4. Use caution when the probe is in an open position to avoid injury. Remember to fold the probe back to the closed position when not in use.
The probe may be damaged if used to measure a temperature that exceeds the specification of measurement temperature range.
HACCP check
The “HACCP CHECK” feature is incorporated in our thermometer temperature to graphically indicate critical temperature zone. The icons and LED indicators located above the display indicate if a food product stays in a safe or unsafe HACCP “ Danger Zone” temperature.
A Green LED appears with the ” ” icon to indicate a safe cool or frozen condition below 4ºC(40ºF) or appears with the ” ” icon to indicate a safe holding temperature above 60ºC(140ºF).
When temperature is between 4ºC and 60ºC (40 and 140ºF), the red LED with the ” ” icon will appear and indicate that the temperature has fallen within the HACCP “Danger Zone” from 4ºC to 60ºC (40~140ºF).
Lcd Error Messages
The thermometer incorporates visual diagnostic messages as follows:
」 「
」「
For all other error messages it is necessary to reset the thermometer. To reset it, turn the instrument off, remove the battery and wait for a minimum of one minute, reinsert the battery
and turn on. If the error message remains, please contact Customer Service for further assistance.
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the range of the instrument, ‘Hi’ when higher than +330oC (626
o
than -55
C (-67 oF).
‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. ‘Er3’ is displayed when the ambient temperature exceeds
o
C (32oF) or +50 oC (122 oF). The thermometer should be allowed plenty of time
0 (minimum 30 minutes) to stabilize to the working/room temperature.
o
F) and ‘Lo’ when lower
Batteries
The thermometer incorporates visual low battery indication as follows:
‘Battery OK’: measurements
are possible
> > > > >
CAUTION: Since the thermometer is splash proof, please make sure the battery cover and rubber stopper are on tight.
>
1. Please pick out the rubber stopper on the battery cover by carefully using a small, pointed tool, such as a screwdriver. Then remove the screw that is under the rubber stopper.
2. Remove the battery compartment cover.
3. Replace the 2 AAA 1.5V batteries.
4. Replace the battery compartment cover.
5. Replace the screw and the rubber stopper.
When the 'Low Battery' icon indicates the battery is low, or the readings grow dim or irregular, the batteries should be replaced immediately. Please note: It is important to turn the instrument off before replacing the batteries, otherwise the thermometer may malfunction. Dispose of used batteries properly and keep away from children. Do not dispose of batteries in fire, as they may explode or leak.
Specification
Measurement Range -67°F to 482°F/-55°C to 250°C
Operating Range 0 to 50°C (32 to122°F)
Accuracy
‘Battery Low’: batteries needs to be
replaced, measurements are possible
Infrared Scan function
(IRT Mode)
-67°F to 32°F +/- 1.8°F + (-0.18 x (reading – 32))
-55°C to 0°C +/- 1°C + (-0.1 x reading) 32°F to 149°F/0°C to 65°C +/-2°F(1°C) 149°F to 482°F/65°C to 250°C +/-1.5% of reading
‘Battery Exhausted’:
measurements are not possible
Thermocouple Probe (K type, Grounded) (COT Mode)
-67°F to 626°F/-55°C to 330°C
Below 14°F /-10°C +/-2°F(1°C) 14°F to 149°F/-10°C to 65°C +/-1°F(0.5°C) Above 149°F/65°C +/-1% of reading
EMC/RFI
Readings may be affected if the unit is operated within a radio frequency electromagnetic field strength of approximately 3 volts per meter, but the performance of the instrument will not be permanently affected.
Two Year Limited Warranty
This product is warranted against defects in materials or workmanship for two (2) years from date of original purchase for the original purchaser. It does not cover damages or wear resulting from accident, misuse, abuse, or unauthorized adjustment and/or repair. Should this product require service (or replacement at our option) while under warranty, please pack the item carefully and return it prepaid, along with receipt showing date of purchase and a note explaining reason for return to:
Taylor Precision Products
2220 Entrada Del Sol
Las Cruces, New Mexico 88001, Suite A
Customer Service Phone: 1-800-225-4834
Customer Service Fax: 1-575-526-4347
www.taylorusa.com
There are no expressed warranties except as listed above. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.
For additional product information, or warranty information in Canada or elsewhere outside the USA, please contact us through www.taylorusa.com.
©2011 Taylor Precision Products and its affiliated companies, all rights reserved. Taylor® and Leading the Way in Accuracy® are trademarks of Taylor Precision Products and its affiliated companies. All rights reserved.
Made to our exact specifications in China.
NOT LEGAL FOR TRADE
Resolution (-9.9 to 199.9°C) 0.2°C/0.5°F 0.2°C/0.5°F
Dimension 22.18*38*190.09mm
Weight (with battery) 98.1g
Battery Life Typically 40 hours, minimum 30 hours continuous use
Auto Shut-Off
Thermocouple – Infrared – After 15 seconds if no change in temperature
After 4 minutes if no change in temperature
9306N 4.11TT
Loading...
+ 4 hidden pages