Syntec 6515-01-556-9563 Vitreoretinal Surgical System Maint - VitMan Oxygen Blender Instruction Manual

Page 1
©
Syntec VitMan
MAINTENANCE
MANUAL
Syntec, Inc.
733 Mansion Rd. Phone: (636) 566-6500 Winfield, Mo. 63389 Fax: (636) 566-6535
Page 2

Table of Contents

System Description........................................................................................... 2
Theory of Operation.......................................................................................... 3
System Interconnect Block Diagram .............................................................. 28
Connector Pinouts .......................................................................................... 29
Pneumatic Block Diagram .............................................................................. 31
Status Line Messages .................................................................................... 32
System Test Procedures................................................................................. 5 6
System Operation Check................................................................................ 57
System Trouble Shooting Chart...................................................................... 6 1
Trouble Shooting Steps .................................................................................. 66
System Adjustment Procedures ..................................................................... 97
Adjust Pressure Regulators............................................................................ 9 8
Adjust Vacuum Regulator ............................................................................... 99
Calibrate Aspiration ...................................................................................... 101
System Assembly/Disassembly Procedures................................................ 102
Light Source Lamp Replacement ................................................................. 103
System Fan Filter Maintenance.................................................................... 1 05
System Fuse Replacement .......................................................................... 106
System Cover Removal ................................................................................ 107
System Cover Installation............................................................................. 108
Hydrophobic Filter Replacement ................................................................... 110
Pneumatic Module Removal.......................................................................... 112
Pneumatic Module Installation ....................................................................... 116
Replace Vacuum Regulator.......................................................................... 122
Analog Electronics Removal ......................................................................... 123
Analog Electronics Installation ..................................................................... 125
Front Panel Removal .................................................................................... 129
Front Panel Installation................................................................................. 133
CPU Electronics Removal ............................................................................ 139
CPU Electronics Installation ......................................................................... 140
Back Panel Removal .................................................................................... 141
Back Panel Installation ................................................................................. 143
Ultrasound Electronics Removal .................................................................. 147
Ultrasound Electronics Installation ............................................................... 148
System Fan Removal ................................................................................... 149
System Fan Installation ................................................................................ 150
Software Procedures .................................................................................... 152
Load Syntec VitMan Software ...................................................................... 15 3
Save Syntec VitMan Status Information....................................................... 155
Syntec
VitMan
ãã
ã
ãã
page 1
Contents
Page 3
System Description
page 2
Syntec
VitMan
ãã
ã
ãã
Page 4
System Description

Theory of Operation

The Syntec VitMan is a complex microprocessor controlled, electromechanical system for use in support of ophthalmic surgery. Operationally it can be broken down into surgical functions or into subsystems that can be grouped together logically because of their func­tion in the system.
The surgical functions are as follows: Vitrector drive Aspiration Irrigation Illumination Air exchange Ultrasound
Operation of the system’s menus is required to access all of the features of the system. The owner’s manual includes comprehensive information on the operation of the menus. This document will refer to a specific series of menu selections separated by a back slash (\). For example Values \ Current A/D Values signifies selecting Values from the Main menu, then selecting Current A/D Values from the Values menu. When a reference is made to a specific selection a complete listing of the menu selections will be given from the main menu, even if a shorter route from the current screen could be taken in the menu tree. Vitrector - The vitrector surgical function provides a variable rate pressure pulse used to drive a guillotine type vitrector. The vitrector surgical function LED (located on the front panel of the system) will display the current state of the vitrector surgical function. If the LED is dark, the vitrector surgical function is off. If the LED is green, the vitrector surgical function is operating within accept­able limits. If the LED is red a problem exists which may effect the operation of the vitrec­tor surgical function. If the LED is red, the front panel display will show a status line mes­sage indicating the problem. The vitrector surgical function will operate in one of two modes. The vitrector on/off button is used to change the mode of the vitrector surgical function. Each time the vitrector on/off button is pressed the mode will change. Starting with the vitrector surgical function off, if the vitrector on/off button is pressed the vitrector surgical function will come on and enter the posterior vitrectomy mode. If pressed again the vitrector surgical function will enter anterior vitrectomy mode. The vitrector up and down buttons can be used to change the vitrector cut rate setting. The vitrector cut rate setting has a range of five cuts-per-minute to twelve hundred cuts-per-minute, or it can be placed in single cut mode. If the up or
The logically grouped subsystems are as follows: Light source- see Illumination surgical function Cassette housing assembly System Pneumatics- Includes Pneumatic module Floppy disk drive LCD display CPU electronics board Analog electronics board Ultrasound electronics board Power supply Speaker System cooling Foot pedal
Syntec
VitMan
ãã
ã
ãã
page 3
Theory of operation
Page 5
System Description
down buttons are pressed and released, the vitrector cut rate setting will change by five cuts-per-minute. If the up or down buttons are pressed and held, the vitrector cut rate setting will continue to change (in five cuts-per-minute steps) until the end of the range is reached or the button is released. If the vitrector cut rate setting is at the minimum setting, and the down button is pressed, the vitrector surgical function will enter single cut mode. While in single cut mode, if the up button is pressed, the cut rate will change to the mini­mum setting. The foot pedal position is used to control when the vitrector cuts. A foot pedal side switch (left or right) is selected by the user, to enable cutting functions (ultrasound or vitrector). By specifying which switch is used to actuate reflux (Current Settings \ Aspiration Values), the other switch is used to enable cutting functions. The default cut enable switch is the right switch. Two modes are available for the cut enable switch, toggle or momentary (Current Settings \ Vitrector Values). If the user selects the toggle mode, actuating the switch will toggle the cut enable state. If cutting is enabled, it will be disabled. If cutting is disabled, it will be enabled. If the user selects momentary mode, the switch needs to be actuated to enable cutting. In addition to enabling cutting (using a side switch), the foot pedal must be depressed before cutting begins. The user is allowed to specify the foot pedal movement required before cutting begins (Current Settings \ Pedal Thresholds). When the foot pedal is depressed far enough that the vitrector threshold is reached, the vitrector will cut. Several subsystems are involved in performing the vitrector surgical function (refer to figure
1). The foot pedal is the main tool available to the surgeon for control of how the system operates the vitrector. Refer to the foot pedal subsystem for details about how it operates. The foot pedal connects to the system back panel via an integral cable. A connector at the back panel makes the transition from the external cable to an internal cable.
External Foot Pedal
Foot Pedal
System Front Panel
Output Tubing
Pneumatic Module
Signal Cable
Theory of operation
Drive Tubing
Analog-Pneumatics
Control Cable
System Pressure Sensor Tubing
System Back Panel
Internal Foot Pedal Signal Cable
Foot Pedal Isolation Electronics Board
CPU Electronics Board
CPU-Analog Board Signal Cable
Analog Electronics Board
page 4
Figure 1. Block diagram of subsystems used for the vitrector drive surgical function.
Syntec
VitMan
ãã
ã
ãã
Page 6
System Description
The internal cable terminates at a connector on the foot pedal isolation electronics board. The board optically isolates the foot pedal signals and passes them through a board to board connector to the CPU electronics board. The microprocessor on the CPU electronics board analyzes inputs from the foot pedal to determine what action to take. If the system pressure is OK and the foot pedal position is correct as defined in the system settings, then it will send a signal to the analog electronics board to activate the vitrector. A signal cable connects the CPU electronics board to the analog electronics board. Addresses and data are passed on this cable. The analog board decodes the command to operate the vitrector and applies ground to the appropriate pin on the analog-pneumatics cable. This cable constantly provides 12Vdc to the vitrector valve, which is turned on when the return wire is grounded. The vitrector valve is located in the pneumatic module. A tube connects the air output from the vitrector valve to the system front panel. A tube connects the main system air pressure from the pneumatic module to a sensor on the analog electronics board. The analog electronics board communicates the system pressure back to the microprocessor through the CPU-analog signal cable. The system monitors the system air pressure only. It can’t determine if the vitrector valve is actually operating. Nor does it measure the pressure pulses sent from the vitrector valve. Aspiration - The aspiration surgical function is used to provide vacuum to the connected surgical instrument. The aspiration surgical function LED will display the current state of the aspiration surgical function. If the LED is dark, the aspiration surgical function is off. If the LED is green, the aspiration surgical function is operating within acceptable limits. If the LED is red a prob­lem exists which may effect the operation of the aspiration surgical function. If the LED is red, the front panel display will show a status line message indicating the problem. The aspiration on/off button is used to change the state of the aspiration surgical function. The effect on the aspiration surgical function is dependent upon the state of other surgical functions as detailed below. With either the vitrector or ultrasound surgical functions on, if the aspiration surgical function is on and the aspiration on/off button is pressed, the aspira­tion function will be turned off. If the aspiration surgical function is off and the aspiration on/off button is pressed, the aspiration function will be turned on. The aspiration mode used (anterior or posterior) is the mode selected by either the vitrector or ultrasound surgi­cal function (whichever is active). With both the vitrector and ultrasound surgical functions off; the aspiration on/off button is used to place the aspiration surgical function in one of three states. Starting with the aspiration surgical function off, when the aspiration on/off button is pressed the aspiration surgical function will come on and enter posterior mode. If the aspiration on/off button is pressed again the aspiration surgical function will enter ante­rior mode. If pressed again the aspiration surgical function will be turned off. Two types of vacuum control are provided, fixed and linear. While in fixed aspiration, the user can select the desired vacuum level using the aspiration up and down buttons. The foot pedal position is monitored and when the activity threshold is reached, the aspiration valve will open and selected vacuum level is applied to the surgical instrument. The activity threshold can be specified by the user (Current Settings \ Pedal Thresholds). In linear aspiration, the vacuum applied to the surgical instrument is proportional to the posi
Syntec
VitMan
ãã
ã
ãã
page 5
Theory of operation
Page 7
System Description
tion of the foot pedal. When the activity threshold is reached, the aspiration pinch valve will open, but initially, no vacuum is applied. As the foot pedal is further depressed, the vacuum level will increase linearly, until the maximum vacuum setting is reached when the foot pedal is fully depressed. Reflux is used to apply a small positive pressure to the aspira­tion tubing. This small pressure is used to release any tissue trapped at the end of the surgical instrument. With the foot pedal up, the surgeon can activate the reflux function using a foot pedal side switch. Either the left or right side foot pedal switch can be selected for reflux function activation (Current Settings \ Aspiration Values). The aspiration up and down buttons can be used to change the maximum vacuum setting. The up button will increase the maximum vacuum setting and the down button will de­crease the maximum setting. The maximum vacuum setting has a range from zero mmHg to five hundred mmHg. If the up or down buttons are pressed and released, the maximum vacuum setting will change by five mmHg. If the up or down buttons are pressed and held, the maximum vacuum setting will continue to change (in five mmHg steps) until the end of the range is reached or the button released. While the aspiration surgical function is off, the aspiration up and down buttons will have no effect, unless a prime is in progress. The aspiration valve open/close button can be used to open and close the aspiration pinch valve and, when pressed and held, can be used to initiate a prime cycle. If the foot pedal is not active, when the aspiration valve open/close button is pressed and released, the aspi­ration valve will change state. If the valve is open, it will close. If the valve is closed, it will open. If the aspiration valve is opened, using the aspiration valve open/close button, it will close after thirty seconds or if any foot pedal activity is sensed.. If the aspiration open/close button is pressed and held for one second, a prime cycle will start. A prime cycle will apply the specified vacuum level (Current Settings \ Prime Pa-
rameters) to the 50cc chamber of the aspiration cassette, for the duration specified (Cur­rent Settings \ Prime Parameters). One of two prime methods can be selected ( Current Settings \ Prime Parameters \ Prime Mode Select). In timed prime mode, the duration of
the prime cycle is determined by the prime cycle duration for the selected surgical function. In continuous prime mode, once started the prime cycle will stop when the aspiration valve button is pressed, foot pedal activity is detected or the cassette over fills. The normal method to terminate a continuous prime cycle is to press the aspiration valve button. The aspiration cassette (refer to figure 2) is the connection between the aspiration tubing and the vacuum system. It also provides for aspirant storage. A dual chamber design is used. The larger 500cc chamber is used to store vacuum and aspirant. The smaller 50cc chamber is connected to aspiration tubing that communicates the vacuum to the surgical instrument. It’s relatively small volume allows for high speed control of it’s vacuum level. A float valve controls the opening of a channel connecting the two chambers. During use, approximately 25cc of fluid is allowed to enter the smaller chamber. Any additional fluid entering the smaller chamber causes the float valve to transfer an equivalent amount into the larger chamber. Two prisms are used to allow the system to sense the fluid level in each chamber. A third prism is used to reflect visible light towards the front of the cassette.
Theory of operation
page 6
Syntec
VitMan
ãã
ã
ãã
Page 8
Valve float
Output connection
Valve
System Description
Vacuum connections
50cc chamber
500cc chamber
Level sense prisms
Illumination prism
Figure 2. Aspiration Cassette
External Foot Pedal Signal Cable
V acuum Tubing
Vacuum Sense Tubing
Analog-Pneumatics Control Cable
System Back Panel
Internal Foot Pedal Signal Cable
Foot Pedal Isolation Electronics Board
CPU Electronics Board
Analog Electronics Board
Aspiration Tubing
Manifold

Cassette

Pinch V alves
System Front Panel
Foot Pedal
Hydrophobic Filters
Pneumatic Module
Pinch V alve Control Tubing
Figure 3. Block diagram of subsystems used for the aspiration surgical function.
CPU-Analog Board Signal Cable
Syntec
VitMan
ãã
ã
ãã
page 7
Theory of operation
Page 9
System Description
Various subsystems are used to perform the aspiration surgical function (refer to figure 3). The foot pedal is the primary method for controlling how the system applies vacuum. Refer to the foot pedal subsystem for details about how it operates. The foot pedal is connected to the system back panel with a cable that is permanently connected to the foot pedal. A connector at the back panel makes the transition from the external cable to an internal cable. The internal cable terminates at a connector on the foot pedal isolation electronics board. The board optically isolates the foot pedal signals and passes them through a board to board connector to the CPU electronics board. The microprocessor on the CPU electron­ics board analyzes inputs from the foot pedal to determine what action to take. If the foot pedal position is correct as defined in the system settings, then it will send a signal to the analog electronics board to either deliver a vacuum, change the aspiration pinch valve state, or change the reflux valve state. A signal cable connects the CPU electronics board to the analog electronics board. Addresses and data are passed on this cable. The analog board decodes the commands from the CPU electronics board. If the command is to acti­vate one of the pinch valves (aspiration pinch valve or reflux pinch valve) then the analog electronics board applies ground to the appropriate pin on the analog-pneumatics cable. This cable constantly provides 12Vdc to the valves (in the pneumatic module) that control the pinch valves. The valves turn on when their return wire is grounded. Tubing connects the control valves in the pneumatic module to the pinch valves on the system front panel. If the command is to change the vacuum level of the smaller chamber of the cassette then the gain on a multiplying D/A converter is set and the analog signal from the foot pedal is passed through the analog-pneumatics cable to a vacuum regulator in the pneumatic module. The vacuum regulator in the pneumatic module is a closed loop V to P convertor. It takes vacuum from the larger chamber of the cassette and uses it to control the level of vacuum in the smaller chamber of the cassette. Tubing connects the cassette to the pneu­matic module. The tubing includes two hydrophobic air filters to protect the system from fluid ingress. The cassette manifold provides the connection between the tubing and the cassette. Additionally the cassette manifold is connected to pressure sensors on the ana­log electronics board with two tubes. The sensors monitor the vacuum level of both cham­bers in the cassette. The analog electronics board communicates the chamber vacuums back to the microprocessor through the CPU-analog signal cable. The system monitors the vacuum level in the 50cc chamber and makes command adjustments as necessary to maintain the vacuum level requested by the foot pedal. The system monitors the 500cc chamber vacuum level and will give an error if the vacuum level measures too low. Irrigation - The irrigation pinch valve is used to control irrigation fluid flow. While in any anterior mode, the irrigation pinch valve will open just prior to the application of aspiration vacuum and close just after vacuum is removed. The foot pedal position is used to control when the pinch valve opens. The user is allowed to specify the foot pedal movement re­quired before the pinch valve opens (Current Settings \ Pedal Thresholds). While in posterior modes, the irrigation pinch valve will open (if closed) just prior to the application of aspiration vacuum and will remain open, until closed using the irrigation valve button (or by entering anterior mode). The irrigation pinch valve is to be used only in anterior modes, however, if used in posterior modes the valve will open and remain open.
Theory of operation
page 8
Syntec
VitMan
ãã
ã
ãã
Page 10
System Description
The irrigation surgical function requires multiple subsystems to operate (refer to figure 4). The foot pedal is the main means available for control of how the system operates the irrigation pinch valve. Refer to the foot pedal subsystem for details about how it operates. The foot pedal connects to the system back panel with a cable integral with the pedal. A connector at the back panel makes the transition from the external cable to an internal cable. The internal cable terminates at a connector on the foot pedal isolation electronics board. The board optically isolates the foot pedal signals and passes them through a board to board connector to the CPU electronics board. The microprocessor on the CPU electron­ics board analyzes inputs from the foot pedal to determine what action to take. If the foot pedal position is correct as defined in the system settings, then it will send a signal to the analog electronics board to activate the pinch valve. A signal cable connects the CPU electronics board to the analog electronics board. Addresses and data are passed on this cable. The analog board decodes the command to operate the pinch valve and applies
External Foot Pedal Signal Cable
System Back Panel
Internal Foot Pedal Signal Cable
Foot Pedal Isolation Electronics Board
CPU Electronics Board
Figure 4. Block diagram of subsystems used for the irrigation surgical function.
Foot Pedal
Pinch V alve
System Front Panel
Pinch V alve Control Tubing
Pneumatic Module
Analog Electronics Board
Analog-Pneumatics Control Cable
CPU-Analog Board Signal Cable
ground to the appropriate pin on the analog-pneumatics cable. This cable constantly pro­vides 12Vdc to the valve (in the pneumatic module) that controls the pinch valve. The valve is turned on when it’s return wire is grounded. Tubing connects the control valve in the pneumatic module to the pinch valve on the system front panel. Illumination - The illumination surgical function is used to provide illumination for surgery through fiber optic instruments. Three fiber optic connection ports are provided. All three ports are illuminated using the same bulb. Two bulbs are available, the primary bulb and a back up bulb. The system monitors the condition of both bulbs and switches to the back up bulb when required. The system warms each bulb prior to turning it on. The user changes the intensity of the light output by adjusting the dimming knob on the light source
Syntec
VitMan
ãã
ã
ãã
page 9
Theory of operation
Page 11
System Description
tray. The illumination on/off button can be used to turn on and off the illumination surgical function. When the illumination surgical function is turned on, the system will warm the bulb, then turn the bulb full on. The system uses several subsystems to support the illumination surgical function (see figure 5). The microprocessor on the CPU electronics board analyzes inputs from the front panel controls and the current bulb status (good or bad) to determine what action to take. When a request to turn on a bulb is made, a series of commands is sent to the analog electronics board through the CPU-Analog board signal cable. The analog board decodes the commands. If the command is to move the bulb position (move a backup bulb into position) then it applies ground to the appropriate pins (2) on the analog-pneumatics cable. This cable constantly provides 12Vdc to the valves (in the pneumatic module) that control the backup mechanism in the light source. The valves are turned on when their return wires are grounded. Tubing connects the valves in the pneumatic module to the backup mecha­nism in the light source. The backup mechanism moves to the “A” position when the valves are off and the “B” position when the valves are on. If the command is to warm a bulb then ground is applied to the appropriate pin on the appropriate analog-bulb cable. Each bulb has it’s own power resistor, which is connected in series with the bulb to warm it prior to applying full current. The power resistors are mounted to the air duct. If the command is to turn on a bulb then ground is applied to the appropriate pin on the appropriate analog-bulb cable (different pin than the warm pin). The analog-bulb cables constantly provide 24Vdc to the bulbs. The bulb is turned on when it’s return wire is grounded. The majority of the light source components are mounted on the light source tray. This allows the light source to be opened for bulb replacement. The two analog-bulb cables terminate in a drawer style connector inside the light source enclosure. The connection is broken whenever the light source tray is opened. The light source output connector (see figure 6) contains an integral dimming disk and the
CPU Electronics Board
CPU-Analog Board Signal Cable
Pneumatic Module
Backup Control Tubing
Light Source
Analog-Pneumatics
Control Cable
Analog Electronics Board
Light Source Fan Cable
Analog-Bulb Cables
Figure 5. Block diagram of subsystems used for the illumination surgical function.
mechanism for retaining the fiber optic connectors. The light source dimming mechanism is controlled by the dimming knob located on the front of the light source tray. Rotating the dimming knob will change the position of the dimming disk relative to the ends of the fiber optic cables and change the amount of light allowed to enter the fiber optic cable.
Theory of operation
page 10
Syntec
VitMan
ãã
ã
ãã
Page 12
System Description
Guide bushings
Figure 6. Illustration of light source connector
Fiber optic illuminator
Dimming disk
Connector catches
The light source enclosure has a fan to cool the light source optics (refer to figure 7). The light source fan pulls air from the inside of the system and forces it past the light source optics and out the air duct. The light source fan will run when the system power is on. Power for the light source fan is provided by the light source fan cable connected to the analog electronics board.
Figure 7. Light source enclosure, bottom back
Bulb warming power resistors
Light source
System speaker
fan
Analog-bulb cables
The light source tray contains the optics used to focus the light from the illuminated bulb to the ends of the fiber optic cables (refer to figure 8). Light passes through a collimating lens onto a hot mirror. The visible light which passes through the hot mirror then passes through a focusing lens onto the ends of the fiber optic cables. A spherical mirror reflects light from behind the bulb towards the front.
Syntec
VitMan
ãã
ã
ãã
page 11
Theory of operation
Page 13
System Description
The light source drawer is held in the light source enclosure by the light source lock knob (refer to figure 9). When the lock knob is in the unlocked position, the light source drawer may be slid to the open position. The light source drawer stop cable (see figure 8) prevents the light source drawer from being removed from the system.
Figure 8.
Drawer connector
Spherical mirror
Opened light source tray
Output connector
Dimming knob
Lock knob
Lens
Hot mirror
Stop cable
Backup mechanism
Figure 9 Light source tray bottom
Theory of operation
Tubing connectors
Drawer connector
page 12
Syntec
VitMan
ãã
ã
ãã
Page 14
System Description
Air Exchange - The air exchange surgical function is used to provide a variable flow con­stant pressure source of air. The air exchange surgical function LED will display the current state of the air exchange surgical function. If the LED is dark, the air exchange surgical function is off. If the LED is green, the air exchange surgical function is operating within acceptable limits. If the LED is red a problem exists which may effect the operation of the air exchange surgical function. If the LED is red, the front panel display will show a status line message indicating the prob­lem. The air exchange on/off button can be used to turn on and off the air exchange surgical function. The air exchange up and down buttons can be used to change the user specified pressure setting. The up button will increase the pressure setting and the down button will decrease the pressure setting. The user specified pressure setting has a range from one mmHg to one hundred mmHg. If the up or down buttons are pressed and released, the user specified pressure setting will change by one mmHg. If the up or down buttons are pressed and held, the user specified pressure setting will continue to change (in one mmHg steps) until the end of the range is reached or the button is released. While the air ex­change surgical function is off, the air exchange up and down buttons will change the user specified pressure setting. The system uses several subsystems to support the air exchange surgical function (see figure 10). The microprocessor on the CPU electronics board analyzes inputs from the front panel controls and the current function status to determine what action to take. When the air exchange surgical function is turned on, the current output port pressure is used as the desired low pressure compressor pressure set point. A command is sent to the analog electronics board through the CPU-Analog board signal cable to enable the low pressure compressor. The analog board decodes the commands. It applies ground to the appropri­ate pin on the analog-pneumatics cable, enabling the low pressure compressor. It also sets the desired pressure point for the low pressure compressor control circuit which supplies a positive drive voltage to the appropriate pin on the analog-pneumatics cable. This cable connects to the low pressure compressor on the pneumatic module. A tube connects the output of the low pressure compressor to the system front panel (after it passes through a valve). Two pressure sense tubes (one for measuring pressure at the air exchange output, called “eye”, and another for measuring the compressor output pressure, called “pump”)
Syntec
Figure 10 Block diagram of systems used for Air Exchange surgical function.
VitMan
System Front Panel
Output T ubing
Pneumatic Module
ãã
ã
ãã
Analog-Pneumatics
Control Cable
"Eye" pressure sensor tubing
"Pump" pressure sensor tubing
page 13
CPU Electronics Board
CPU-Analog Board Signal Cable
Analog Electronics Board
Theory of operation
Page 15
System Description
connect to sensors on the analog electronics board. The low pressure compressor control circuit on the analog electronics board uses the pump sensor output as feedback to vary the low pressure compressor drive voltage. The output of the pressure sensors is commu­nicated back to the CPU electronics board through the CPU-analog board signal cable. When the current compressor pressure is less than ten mmHg above the initial output port pressure, the microprocessor will send a command for the air exchange valve to open. It will also send a command to change the desired low pressure compressor pressure set point to the user specified air exchange pressure setting. The analog board decodes the commands. It applies ground to the appropriate pin on the analog-pneumatics cable that connects to the air exchange valve in the pneumatic module. This cable constantly pro­vides 12Vdc to the air exchange valve, which is turned on when the return wire is grounded. When the valve is on, output from the low pressure compressor is connected to the air exchange output on the system front panel. When the valve is off, the output from the low pressure compressor is disconnected and the air exchange output on the system front panel is blocked. Once the air exchange valve is open, if the eye sensor pressure exceeds the user specified value by fifteen mmHg, the microprocessor will command the air exchange valve to close until the pressure is reduced. If the eye sensor pressure is not within five mmHg of the user specified air exchange pressure setting, the air exchange surgical function will be in alarm (system beeps, LED goes red, and a status line message appears). The low pressure compressor control circuit monitors the pump pressure sensor. If exces­sive pressure is detected by the control circuit, it will disable the low pressure compressor. This event is communicated to the microprocessor through the CPU-analog board signal cable. If excessive pressure is reported to the microprocessor, the air exchange valve will be commanded to close. If the excessive pressure condition is removed, the microproces­sor will enable the low pressure compressor control circuit. When appropriate the air ex­change valve will be opened. Note that this excess pressure control method is separate from the normal, software dependent, method. This method is built into the control circuit and cannot be disabled by the system software. Ultrasound - The ultrasound surgical function is used to apply energy to an attached ultrasound handpiece. The energy can be delivered in either continuous or pulse mode. The relative power and pulse rate are specified by the user. The ultrasound surgical function LED will display the current state of the ultrasound surgical function. If the LED is dark, the ultrasound surgical function is off. If the LED is green, the ultrasound surgical function is operating within acceptable limits. If the LED is red a prob­lem exists which may effect the operation of the ultrasound surgical function. If the LED is red, the front panel display will show a status line message indicating the problem. The ultrasound surgical function will operate in one of four modes. The ultrasound on/off button is used to change the mode of the ultrasound surgical function. Each time the ultra­sound on/off button is pressed the ultrasound mode will change. Starting with the ultra­sound surgical function off, if the ultrasound on/off button is pressed the ultrasound surgical function will come on and enter posterior fragmentation mode. If pressed again the ultra­sound surgical function will enter anterior linear phaco. mode. If pressed again the
Theory of operation
page 14
Syntec
VitMan
ãã
ã
ãã
Page 16
System Description
ultrasound surgical function will enter anterior fixed phaco mode. If pressed again the ultrasound surgical function will enter anterior fragmentation mode. If pressed again the ultrasound surgical function will be turned off. When the ultrasound surgical function enters a new mode, the aspiration surgical function is turned on, the aspiration mode (anterior or posterior, linear or fixed) is set to be compatible with the ultrasound mode. The ultrasound up and down buttons can be used to change the ultrasound power setting. While in fixed power modes the setting will select the power level delivered to the hand piece, when energized. In linear power mode, the setting will select the power level reached upon fully depressing the foot pedal. If the ultrasound surgical function is on, the up button will increase the ultrasound power setting and the down button will decrease the ultrasound power setting. The ultrasound power setting has a range from one percent to one hundred percent. If the up or down buttons are pressed and released, the ultrasound power setting will change by one percent. If the up or down buttons are pressed and held, the ultrasound power setting will continue to change until the end of the range is reached or the button released. While the ultrasound surgical function is off, the ultrasound up and down buttons will have no effect. The pulse button is used to change the state of the ultrasound pulse mode. This is a toggle operation and requires that the ultrasound surgical function be on. If pulse mode is off and the pulse button is pressed, pulse mode will be turned on. If pulse mode is on and the pulse button is pressed, pulse mode will be turned off. While the ultrasound surgical func­tion is off, the pulse button will have no effect. The pulse up and down buttons can be used to change the pulse rate setting. If the ultra­sound surgical function is on and pulse mode is on, the up button will increase the pulse rate setting and the down button will decrease the pulse rate setting. The pulse rate setting has a range from one pulse-per-second to twenty pulses-per-second. If the up or down buttons are pressed and released, the pulse rate setting will change by one pulse-per­second. If the up or down buttons are pressed and held, the pulse rate setting will continue to change until the end of the range is reached or the button released. While pulse mode is off, the pulse up and down buttons will have no effect. Older versions of the ultrasound electronics require that the electronics be tuned in order to transfer maximum power to the needle. New versions of the ultrasound electronics con­stantly tune, so maximum power is delivered to the handpiece at all times. A tune cycle scans the entire range of frequencies looking for maximum power. The frequency of maxi­mum power is the resonant frequency of the hand piece. The tune button will initiate a tune cycle. If the ultrasound hand piece is removed from the front panel, then reconnected, a tune cycle will be required before the hand piece can be used. While the ultrasound surgi­cal function is off, the tune button will have no effect. The foot pedal position is used to control when the energy is delivered to the ultrasound hand piece. A foot pedal side switch (left or right) is selected by the user, to enable cutting functions (ultrasound or vitrector). By specifying which switch is used to actuate reflux (Current Settings \ Aspiration Values), the other switch is used to enable cutting func­tions. The default cut enable switch is the right switch. For ultrasound, the side switch is
Syntec
VitMan
ãã
ã
ãã
page 15
Theory of operation
Page 17
System Description
only used in anterior frag. and posterior frag. modes. Two modes are available for the cut enable switch, toggle or momentary (Current Settings \ Ultrasound Values). If the user selects the toggle mode, actuating the switch will toggle the cut enable state. If cutting is enabled, it will be disabled. If cutting is disabled, it will be enabled. If the user selects momentary mode, the switch needs to be actuated to enable cutting. In all ultrasound modes, the foot pedal must be depressed before energy is delivered to the ultrasound hand piece. The user is allowed to specify the foot pedal movement required before cutting begins (Current Settings \ Pedal Thresholds). When the foot pedal is depressed far enough that the ultrasound threshold is reached, power will be delivered to the handpiece. When energy is being delivered to the ultrasound hand piece, a tone is be generated. This will provide audible feedback when energy is being delivered to the hand piece. Several subsystems are involved in performing the ultrasound surgical function (refer to figure 11). The foot pedal is the main tool available to the surgeon for control of how the system delivers ultrasound power. Refer to the foot pedal subsystem for details about how it operates. The foot pedal connects to the system back panel via an integral cable. A connector at the back panel makes the transition from the external cable to an internal cable. The internal cable terminates at a connector on the foot pedal isolation electronics board. The board optically isolates the foot pedal signals and passes them through a board to board connector to the CPU electronics board. The microprocessor on the CPU electron­ics board analyzes inputs from the foot pedal to determine what action to take. If the aspi­ration surgical function is OK, the ultrasound handpiece is present, and the foot pedal position is correct as defined in the system settings, then it will send a signal to the analog electronics board to set the ultrasound power and activate power output. The CPU-analog signal cable connects the CPU electronics board to the analog electronics board. Ad
Figure 11. Block diagram of subsystems used for the ultrasound surgical function.
Chassis Ground
Theory of operation
External Foot Pedal
Foot Pedal
System Front Panel
Ultrasound Output Cable
Ultrasound Electronics Board
Signal Cable
Handpiece Present
Analog-Ultrasound Control Cable
page 16
System Back Panel
Internal Foot Pedal Signal Cable
Foot Pedal Isolation Electronics Board
CPU Electronics Board
CPU-Analog Board Signal Cable
Analog Electronics Board
Syntec
VitMan
ãã
ã
ãã
Page 18
System Description

dresses and data are passed on this cable. The analog board decodes the command to operate ultrasound and applies +5Vdc to the appropriate pin on the analog-ultrasound control cable. It also sets the desired power level by applying a 0.5-5Vdc signal to the appropriate pin on the analog-ultrasound control cable. The ultrasound electronics board receives the command and outputs power to the ultrasound output cable. This cable con­nects to the ultrasound output connector on the system front panel. The ultrasound hand­piece connects to the front panel connector. The ultrasound handpiece has a multiple crystal piezoelectric motor. A single conductor supplies drive voltage to the motor. The return path is comprised of multiple conductors and is tied to chassis ground at the front panel. The ultrasound electronics board monitors the ground path for the handpiece. If it senses a loss of ground, it applies 5Vdc to the appropriate pin of the analog-ultrasound control cable. The analog electronics board communicates this to the CPU electronics board over the CPU-analog signal cable. The microprocessor inhibits ultrasound power output and the surgical function LED goes red. An additional connection for the ultrasound output cable connects from the ultrasound output connector on the system front panel to the CPU electronics board. It provides the handpiece present signal to the microprocessor. The microprocessor inhibits ultrasound power output and the surgical function LED will go red if the handpiece isn’t connected when the surgical function is on. Cassette Housing Assembly- Physical alignment of the various components that support the cassette is provided by the cassette housing assembly (refer to figures 12 and 13). Located on the bottom front of the housing is the lock knob. It is used to align the cassette front to back in the housing and to compress the top ports of the cassette to the seals in the cassette manifold. This is accomplished through a 270 degree rotation of the lock knob. Stops are provided at both extremes of knob rotation. The cassette manifold is

Figure 12. Cassette housing
Vacuum connections
Figure 13. View of cassette manifold seals
Cassette manifold
Sensors
Seals
Analog electronics board
Lock knob
Syntec
VitMan
ãã
ã
ãã
page 17
Theory of operation
Page 19
System Description

located towards the top front of the housing. It provides connections to the cassette, vacuum tubing, and sensors. The seals to the cassette are accomplished through O-rings mounted under the manifold. Both the manifold and the lock knob bracket are connected to the front panel providing it mechanical support. The left wall of the housing is used to mount the housing to the right side of the light source enclosure. It also contains a leaf spring that is used to hold the cassette against the right side of the housing. The rear wall also contains springs. They are used to eject the cassette after it is released by the lock knob. The right side of the housing holds the analog electronics board. The side has holes which allow LED’s and optical sensors access to the side of the cassette. A series of opti­cal sensors on the analog electronics board to monitor the prisms on the side of the cas­sette. Fluid in the cassette causes light to not be reflected back by the prism in the cas­sette. This allows the system to measure the amount of fluid in the cassette. It also allows the system to sense when a cassette is installed into the system. The front cross rail of the system mounts to the right side of the housing. System Pneumatics - The VitMan air system affects every surgical function of the system. The Pneumatic Module is designed to be removed from the system as a unit for mainte­nance or replacement. It contains every the pneumatic component in the system that doesn’t need to be elsewhere because of it’s function. Refer to figure 14 for a schematic of the system pneumatics and figures 15 and 16 for a picture of the pneumatic module. Pneu­matic components that are not mounted on the pneumatic module are drawn with a red outline. The various pressure sensors are all mounted on the analog electronics board. They contain the same references on the schematic as is used on the board. The three pinch valves and two function outputs (vitrector and air exchange) are mounted on the system front panel. The air cylinder is part of the light source backup mechanism. The cassette manifold is part of the cassette housing assembly, and is the connection to the aspiration cassette for vacuum control. The two filters are located to the rear of the cas­sette housing assembly. The pressure and vacuum input connections are mounted on the system back panel. The tubing that connects the different components is shown color coded for the size and type of tubing used. The schematic is a physical connection style schematic, the connections on the components are positioned similar to the physical con­nections used in the system. The bubble encapsulated numbers represent various fittings used to accomplish connections. Of special note is bubble 6. They are the quick discon­nects that must be disconnected to remove the pneumatic module. Beside them is the reference used to mark the tubes for easy reassembly. If a marker is missing or illegible, refer to these references to aid assembly. The pneumatic module contains seven valves, low pressure compressor, and the vacuum regulator that all require electrical control signals. The signals are provided by the analog­pneumatics control cable connected to the analog electronics board. Six of the valves (all except the vitrector valve) have wire connectors at the valve and an LED to indicate when the valve is on. The vitrector valve has permanently connected wiring and no indication of when it is on. All seven valves can be manually activated by pressing the button on their sides. The low pressure compressor has permanently connected wiring. The vacuum regulator wiring is connected to screw tightened terminals under the cover of the regulator. The cover can be removed by removing the two screws securing the cover to the regulator

Theory of operation
page 18
Syntec
VitMan
ãã
ã
ãã
Page 20
System Description
Light Source Backup
16
Air Cylinder
16
8
Low Pressure
Compressor
P
V
11
25mm 0.3 micron GMF
Filter
Intake
Accumulator
4
12
12
2-Way Normally Closed
8 P
Air Ex
A 8
12
6
Air Exchange Out
7 on for A bulb
0 on for B bulb
"0"
6
Ref. Pinch Valve
8
Irr. Pinch Valve
8
Asp. Pinch Valve
6
"3"
"2"
6
"9"
"7"
6
"6"
6
"5"
6
"1"
6
0.005 Orifice
5
5
Pilot
4
"Pump"
Sensor
X103
0.005 Orifice
Sensor
X102
"Eye"
0.005 Orifice
8
8
8
8
A
8
8
P
1
Out Vac
1
Bleed
8
7
8
8
8
8
15
Vacuum
Regulator
E
9
3-way Normally Open
Light Y
A
E 8
Ref.
Pinch
3-way Normally Closed
Irr.
Pinch
8
Asp.
Pinch
8
Light X
8
60mm
1 micron
Filter
9
9
10
10
Cassette
Manifold
Exhaust
Air flow
Intake
Exhaust
Intake
Air flow
Air Input
Connector
Filter
50mm
1 micron
5
12
0.005 Orifice
Bleed
Open
4
Relief
Valve
13
13
Vacuum
Pump
Red
White
17
"500CC"
Sensor
X101
5
"50CC"
1/8 O.D Tube
1/4 O.D. Tube
3/16 O.D. Tube
1/4 O.D. Silicone Tube
5/32 O.D. Silicone Tube
2 Filter
17
2
5
4
Cylinder
Regulator
13
3
Manifold
4
3
2
3
Vitrector
Regulator
2
13
Vac Input Connector
Sensor
X100
14
15
16
17
10-32 to 1/8 Barb (Large)
1/4 Quick Disconnect "T"
10-32 to 1/8 Quick Disconnect
1/4 NPT Plug
6
13
13
Vac. Reg.
Regulator
3
4
13
10
11
12
13
1/8 NPT to 1/8 Quick Disconnect
1/8 NPT to 3/16 Barb
1/8 Barb "T"
1/8 NPT Plug
Main Compressor
B
14
1
A
"4"
13
7
P A
15
"8"
6
7
8
9 1/8 NPT to 1/4 Quick Disconnect
10-32 to 1/4 Barb
10-32 to 1/8 Barb
C
D 1
Sensor
X104
"Drive"
3-way Normally Closed
15
15
PE
E
Vit
Vitrector Out
1
234
5
6
1/8 Union Quick Disconnect
1/8 NPT to 1/4 Barb
1/8 NPT to 1/8 Barb
1/8 Barb to 3/16 Barb
1/8 NPT Nipple
1/4 NPT to 1/4 Barb "L"
Figure 14. Schematic of system Pneumatics
ãã
ã
Syntec
VitMan
ãã
page 19
Theory of operation
Page 21
System Description

body. Power for the main compressor and vacuum pump are provided by the analog­pneumatics power cable. Both sets of wires are permanently attached to the compressor and pump. System Pressure- The VitMan requires air pressure to operate. Two methods of providing pressure are provided. A connection for an external air pressure supply is accessible on the system back panel. The system will error if the system pressure is below 42psi. The system will vent any pressure above 50psi. Any external air supply should maintain be­tween these two pressures. It should also be clean, dry, and oil vapor free. An internal pressure compressor is contained in the pneumatic module. The compressor will maintain the appropriate pressure under all normal operating conditions. The intake of the compres­sor is connected to a length of tubing, which acts as a muffler. The output of the compres­sor is connected via a “T” to a coalescing air filter. The other leg of the “T” is connected to the external air input connector. The external air input connector has a built in shut-off valve that is opened when a hose is connected to it on the outside of the system. This is of inter­est if the system is relying on it’s internal compressor and a hose is connected to the exter­nal air input connector. If the hose isn’t shut off at the distal end or if it is connected to something that represents too great of a load, then the internal compressor will be unable to provide the necessary air to the system. The valves in the compressor do not put an additional load on the external air supply. The coalescing air filter will remove any water from the air, but will not remove water vapor. Water collected is plumbed to a spot where it will not do harm. It is normal for the system in humid environments to “make” water as it operates. Water that collects as a result of decompressing the air will eventually go to one of the exhausts and evaporate. If the vitrector is operated without connecting to the output port a small amount of water may spit from the port. This is normal. The output of the coalescing air filter is connected to system pressure manifold. The manifold distributes the air to various regulators that supply the rest of the air system. Also connected to the mani­fold are the system pressure sensor (X104) and the pressure relief valve. The system pressure sensor measures the current operating pressure of the system. This measure­ment is used to ensure that the system doesn’t try to operate certain functions if the supply pressure is too low i.e. vitrector. The pressure relief valve will vent any pressure above 50 psi. It is adjustable. Operating the relief valve at greater pressure will result in a system pressure too high message on the front panel display. It also will put excess stress on the air system components. Vitrector Drive- The vitrector pressure regulator connects to the system pressure mani­fold. It is adjusted to 42 psi. This pressure results in the correct pressure pulse being deliv­ered to the vitrector. The vitrector valve is connected to the regulator. It is a 3-way, normally closed, solenoid operated, air piloted valve. Two exhausts are routed to the valve manifold. The output pressure pulses are connected to the output connector on the system front panel. Timing, pressure, and air flow have all been optimized to provide the high speed cut rate that the VitMan possesses. Kinks in either the pressure or exhaust tubes can cause performance degradation. Air Cylinder Control- The cylinder pressure regulator connects to the system pressure manifold. It is adjusted to 30 psi. The regulated pressure output is connected to the valve manifold. This manifold directly supplies pressure to four 3-way, normally closed, solenoid

Theory of operation
page 20
Syntec
VitMan
ãã
ã
ãã
Page 22
Vac. reg. regulator
Cylinder regulator
Relief valve
System Description
Figure 15. Pneumatic module
Low pressure compressor intake filter
Main compressor
Vac. reg. zero and span adjust
Vacuum regulator
Vitrector regulator
Vibration mounts
Filter drain tube
Coalescing filter
Cylinder valves
Low pressure compressor
Figure 16. Pneumatic module, bottom
ãã
ã
Syntec
VitMan
ãã
Hydrophobic filters
page 21
Vacuum pump
Theory of operation
Page 23
System Description

operated, air piloted valves. Three of the valves are used to control air cylinders which drive the pinch valves on the system front panel. The fourth is connected to the air cylinder that is used in the light source backup mechanism. The exhausts on three of the valves is plumbed back to the valve manifold (separate from the pressure channel). The exhaust of the valve manifold is connected to a length of tubing that acts as a muffler. The pressure from the valve manifold is connected to an additional valve (Light Y). It is a 3-way, normally open, solenoid operated, air piloted valve. The valve is connected to the air cylinder in the light source backup mechanism. It works in concert with the normally closed valve (Light X), mounted on the valve manifold, to control the position of the air cylinder. Connections to the air cylinder from both valves contain an orifice. The orifices act to slow the cylinder motion. Vacuum Regulator Pilot Pressure- The vac. reg. pressure regulator connects to the system pressure manifold. It is adjusted to 20 psi. The regulated pressure is connected to the vacuum regulator. The vacuum regulator uses this pressure to drive the internal regula­tion valve. System Vacuum- The VitMan requires vacuum to operate. Two methods of providing vacuum are provided. A connection for an external vacuum supply is accessible on the system back panel. The system requires an minimum of 550 mmHg vacuum to operate correctly. An internal vacuum pump is contained in the pneumatic module. The pump will maintain the appropriate vacuum under all normal operating conditions. The exhaust of the pump is connected to a length of tubing, which acts as a muffler. The intake of the pump is connected via a “T” to the cassette manifold. The connection is protected with a hydropho­bic filter. The other leg of the “T” is connected to the external vacuum input connector. The external vacuum input connector has a built in shut-off valve that is opened when a hose is connected to it on the outside of the system. This is of interest if the system is relying on it’s internal pump and a hose is connected to the external vacuum input connector. If the hose isn’t shut off at the distal end or if it is connected to something that represents too great of a load, then the internal pump will be unable to provide the necessary vacuum to the system. The valves in the pump do not put an additional load on the external vacuum supply. The cassette manifold provides the connection to the aspiration cassette (see aspiration surgical function for explanation on how the cassette operates). The cassette manifold contains a channel that connects the source vacuum to the 500cc chamber of the cassette, which acts as an accumulator. Vacuum Output Control- The vacuum stored in the 500cc chamber of the aspiration cassette is used to control the vacuum level of the 50cc chamber of the aspiration cassette. Vacuum is routed from the cassette manifold to the input of the vacuum regulator. The vacuum regulator responds to a voltage control signal. It uses the pilot pressure to control the vacuum output. It contains an internal pressure sensor for feedback. The output of the vacuum regulator is protected with a hydrophobic filter and is then connected to the cas­sette manifold. The cassette manifold contains a channel that connects the output of the vacuum regulator to the 50cc chamber of the cassette. Two sensors (X101 and X100) are connected to the cassette manifold. They monitor the vacuum level in both chambers of the cassette. The system is able to modify the voltage command to the vacuum regulator to fine tune the output vacuum. The tubing connecting X100 to the cassette manifold in

Theory of operation
page 22
Syntec
VitMan
ãã
ã
ãã
Page 24
System Description
cludes a “T” that is connected to an orifice. The orifice ensures that there is always a slight load on the vacuum system, even when the output of the aspiration cassette is plugged. This helps stabilize the control loop and improves response time. Air Exchange- The components that comprise the air exchange portion of the system pneumatics are separated from the rest of the pneumatics. The air pressure source is a small rotary vane compressor. The intake of the compressor is protected by a filter. The output of the compressor is connected to an accumulator. The accumulator dampens pressure pulses. The other side of the accumulator is connected to a sensor, an orifice and a valve. The sensor monitors the output pressure of the compressor, regardless of whether the output of the air exchange surgical function is enabled or not. The orifice ensures that a slight load is always placed on the compressor. The valve is a 2-way, solenoid operated, air piloted valve. It is used to isolate the system air exchange output from the pressure source. It allows the pressure source to start and stabilize before the output is connected. It also allows the system to disconnect the output in case the pressure source exceeds allowable limits. The output of the valve is connected to a sensor and to the air exchange output connector. The sensor monitors the output pressure of the system. It allows the to compare the output to the pressure source pressure and make decisions accordingly. Floppy Disk Drive - The floppy disk drive can be used to perform system software up­dates, load diagnostics, configure the system to user preferences and to record system function during surgery. Once a minute, the floppy disk drive is checked for the presence of a diskette. If a disk is present (and not locked) the system function for the past minute will be recorded. The data will be appended to the PC compatible file VITMAN.LOG. If the file does not exist, it will be created. If the data cannot be saved, it is discarded. Several menu mode commands are available to perform actions on a diskette. The direc­tory of a diskette may be displayed. The system software and user preferences may be saved or loaded. The file VITMAN.LOG may be deleted. A diskette may be formatted. The event log or system data may also be saved. The floppy disk drive is mounted to the side of the system. It is controlled by the CPU electronics board. The CPU-floppy signal cable routes control signals to the floppy disk drive. The analog-floppy power cable is connected to the analog electronics board. It pro­vides power to the floppy disk drive. LCD Display - The LCD display is used to display the current settings and operational state of the surgical functions. Two modes are available, normal and menu mode. While in normal mode the state of the surgical functions is displayed. The display area is divided among the various surgical functions. The bottom line of the display area is the status line. Any problem detected will result in a message being displayed in the status line. If more than one problem exists a message will display for two seconds before changing to the next message. Menu mode allows the user to change all aspects of the system configuration. The button is used to enter and exit menu mode. While in normal mode, if the
Display
pressed and held for one second, menu mode is entered. While in menu mode, pressing
Display
button is
Syntec
VitMan
ãã
ã
ãã
page 23
Theory of operation
Page 25
System Description
the
Display
active. Only the LCD display is effected, however, values modified may effect the surgical functions. The entire menu may be traversed and all information can be displayed without using the While in normal mode, the user can adjust the contrast and back light intensity. The left and right arrow buttons can be used to change the LCD display contrast level. The left arrow will make the contrast darker. The right arrow will make the contrast lighter. The up and down arrow buttons can be used to change the LCD display back light intensity. The up arrow will increase the back light intensity. The down arrow will decrease the back light intensity. The LCD electronics board is mounted to the CPU electronics board. It receives power and control signals from the CPU electronics board via a board to board connector. Power Supply- The VitMan receives electrical power through a detachable cord that con­nects to the power entry module on the system back panel. The main power switch for the
button will exit menu mode. While in menu mode, all surgical functions remain
Enter
button. The
Enter
button is required to save any change.
Figure 17. Block diagram of power distribution
GND A
Power Entry Module
CPU electronics board
Ultrasound electronics board
Floppy disk drive
Power Filter
Chassis
Analog electronics board
Power Supply L1 +24 L2/N
GNDD
+S
-S
+12
+S
-S
-
-
system and the system fuse/s are contained in the power entry module. Care should be taken that the proper rating and number of fuses are installed. The power entry module is connected to the power filter, which is mounted on top of the power supply. The filter re­duces conducted electrical noise. The output of the filter is connected to the AC inputs of the power supply. It is a dual output 600W “universal input” switching power supply. The power supply provides an AC OK indication which is connected to the CPU electronics board with the AC OK cable. If AC OK is lost a message is displayed in the status line, and the system attempts to do an emergency shutdown. The power supply is mounted to the system crossrails under the air duct. It provides +12 and +24 volts DC to the analog
Theory of operation
page 24
Syntec
VitMan
ãã
ã
ãã
Page 26
System Description
electronics board via the analog-power supply cable. The analog electronics board serves as a power conditioning and distribution point. Unregulated +12 and +24 volts, regulated +12 and -12 volts, +5 volts, VCC, -22 volts, and various grounds are distributed from here. Speaker - The speaker is used to generate warning and error tones at the request of the surgical functions. If enabled (Audio Services \ Key Click), a key click tone will be gener­ated when a valid keyboard button is depressed. When energy is being delivered to the ultrasound hand piece, a tone will be generated. The audio volume can be adjusted (Audio Services \ Audio Amplitude). The speaker can be disabled ( Audio Services). Several other options are available through menu mode. The speaker is mounted to the bottom of the air duct behind the light source fan (refer to figure 7). The analog-speaker signal cable is connected to the analog electronics board on one end and permanently attached to the speaker at the other. A D/A channel on the ana­log electronics board is used by the system to generate tones. The output of the D/A is connected to an audio amplifier. The output of the audio amplifier is connected to the speaker. The gain of the audio amplifier is connected to another D/A channel. System Cooling- The internal temperature of the system is maintained at an acceptable level by the flowing room temperature air through the system. The system fan (mounted on the system back panel) pulls in air from outside the system and forces it past the main compressor. Air circulates within the system. The light source fan (mounted on the back of the light source enclosure) draws air from inside the system and blows it into the light source. Air passes through the light source and exits the back of the system through the air duct. Directing air flow in this manner protects the rest of the system from the high tem­peratures developed by the light source. Both fans operate when the system power is on. The analog-fan power cables provide power to the fans. The system fan’s power cable has connectors to detach from the fan. The light source fan’s power cable is permanently attached to the fan. The system fan has a filter on the intake side of the fan. The filter should be monitored regularly and cleaned when it becomes dirty. A temperature sensor is located on the analog electronics board. The system will detect and report excesses in the system temperature. If the system temperature exceeds reasonable limits, the system attempts to reduce it by turning off high current draw items. If the current system tempera­ture is 55 °C or higher, the message “Device disabled” will be displayed in the status line and an audible indication will be given. Ten seconds after this error occurs the illumination bulbs will be turned off and the pressure and vacuum pumps will be shut off. Keyboard - The keyboard allows the user to communicate with the system. It can be used to turn on and off the surgical functions, as well as modify their operating characteristics. The keyboard is an integral portion of the front panel overlay. It is comprised of a matrix of membrane switches. A ribbon cable exits the keyboard near the top of the system and connects to the CPU electronics board. Foot Pedal - The foot pedal allows the surgeon communicate control commands to the system. This is accomplished through rotation of the foot pedal position. An analog vertical rotation signal is produced with a linear potentiometer coupled to a gear. Refer to figure 18 for a schematic of the foot pedal wiring and figure 21 for a view inside the foot pedal.
Syntec
VitMan
ãã
ã
ãã
page 25
Theory of operation
Page 27
System Description
Horizontal rotation is monitored with two switches. A cam mechanism causes horizontal rotation to “stop” in three positions. Either left or right rotation causes a switch to activate. The center position is neutral. The left and right switches are used to request reflux and enable the vitrector and ultrasound surgical functions. Vertical rotation is used to change the vacuum level in linear aspiration modes, and specify the power level for linear ultra­sound. An additional switch indicates when the foot pedal is all the way up (rest position). The foot pedal up switch is used as an redundant indication that aspiration, irrigation, and
Top
7
Blue (Position)
6
Green (Up)
5
Yellow (Left)
4
Orange (Right)
3
Red (Present)
2
Brown (+5VDC)
1
Black (Gnd)
9
Drain (Chassis GND)
8
Shield (Chassis GND)
Bottom
NO
C
Left
C
Right
C
Up
NC
NO
NC
NC
NO
Figure 18. Foot pedal schematic
Left switch is located on the right side of the pedal
cutting (vitrector or ultrasound) should stop. The top of the foot pedal includes a pair of adjustable foot guides (refer to figure 19). By pressing on the lock release, located on the left front of the pedal, the guides are allowed to move. This feature allows the foot pedal to accommodate a wide range of foot sizes with­out causing smaller feet to slide around in between the guides. The foot guides pivot near the back of the foot pedal top. A guide rod, at the front of the foot guides, controls their position. Underneath the cover for the foot pedal top is the locking mechanism for the guide rods (refer to figure 20). When the lock release lever is depressed, the guide rods should move freely in the lock mechanism. The top of the foot pedal may be removed from the base by depressing the pivot pin located on either side of the pedal.
Theory of operation
page 26
Syntec
VitMan
ãã
ã
ãã
Page 28
Pivot pin
System Description
Figure 19. VitMan foot pedal
Foot guide
Lock release lever
Cable
Figure 20. Foot guide lock mechanism
Potentiometer
Cam
Left switch
Guide rod
Position return springs
Syntec
VitMan
Up switch
ãã
ã
ãã
Right switch
page 27
Figure 21. Foot pedal internal mechanisms
Theory of operation
Page 29
Keyboard
System Description

System Interconnect Block Diagram

Pneumatic moduleFront panel Rear panel
Foot Pedal
Ultrasound Out
Valve
Control
Analog Data
Pump Power
Main Fan Power
AC
Connections
CPU electronics board
Floppy
Data
CPU Power
Floppy Power
Speaker
Drive
Analog electronics
board
"A"
Bulb
Light Source
Fan Power
Power
Floppy disk drive Light source
"B" Bulb
Power
Ultrasound Power
Ultrasound Data
AC OK Indication
System DC Power
Ultrasound electronics
board
Power supply
Earth ground
Speaker

System interconnect block diagram

page 28
Ground
ChassisSystem covers
Syntec
VitMan
ãã
ã
ãã
Page 30
System Description

Connector Pinouts

CPU – Analog Signal Cable
niPnoitcnuF
1)BSL(0tibsuBataDlanoitceriD-iB 21tibsuBataDlanoitceriD-iB 32tibsuBataDlanoitceriD-iB 43tibsuBataDlanoitceriD-iB 54tibsuBataDlanoitceriD-iB 65tibsuBataDlanoitceriD-iB 76tibsuBataDlanoitceriD-iB 87tibsuBataDlanoitceriD-iB 98tibsuBataDlanoitceriD-iB
019tibsuBataDlanoitceriD-iB 1101tibsuBataDlanoitceriD-iB 2111tibsuBataDlanoitceriD-iB 3121tibsuBataDlanoitceriD-iB 4131tibsuBataDlanoitceriD-iB 5141tibsuBataDlanoitceriD-iB
61)BSM(51tibsuBataDlanoitceriD-iB 71dnuorG
81)woLevitcA(etirWrossecorporciM 91dnuorG
02)woLevitcA(daeRrossecorporciM 12dnuorG 221tibsuBsserddA 322tibsuBsserddA 423tibsuBsserddA 524tibsuBsserddA 625tibsuBsserddA 726tibsuBsserddA
82)woLevitcA(1tceleSpihCrossecorporciM 92desUtoN 03kcolCmorPEE
13)woLevitcA(2tceleSpihCrossecorporciM 23ataDmorPEElanoitceriD-iB 33dnuorG 43TOPlanretnisladeptoofehtmorfegatloV 53egatloverapmocDELlevelettessaC
63)woLevitcA(TESERrosneserutarepmeT
73)WOLevitcA(langisesnesdellatsnieceipdnahnoitatnemgarF 83kcolCrosneserutarepmeT 93dnuorG 04ataDlanoitceriD-iBrosneserutarepmeT 14dnuorG 24langistrevnoCfodnEretrevnocD/A 34dnuorG 44langiskcolCretrevnocD/A 54dnuorG
64)woLevitcA(tceleSpihCretrevnocD/A 74egatlovylppusTOPladeP 84langisNIDretrevnocD/A
94)woLevitcA(TESERrossecorporciM 05langisTUODretrevnocD/A
AC OK Cable
niPnoitcnuF 1+KOCAylppusrewopoT 2oTKOCAylppusrewop-
CPU – Analog Power Cable
niPnoitcnuF
1)CCV(CDV5+
2)CCVgolanA(CDV5+
3)DNG(dnuorG
4)A-CDV21-golanA(CDV21-
5)DNGgolanA(dnuorG
6)A-CDV21+golanA(CDV21+
7)ylppustsartnocDCL(CDV22­8desUtoN
CPU – Floppy Signal Cable
niPnoitcnuFnoitceriD
1)woLevitcA(teseRegnahCksiDoT
2)woLevitcA(egnahCksiDksiDmorF 3dnuorG
4)woLevitcA(esUnIksiDoT 5dnuorG 6desUtoN 7dnuorG 87tibsuBataDlanoitceriD-iB 9dnuorG
019tibsuBataDlanoitceriD-iB 11dnuorG 2111tibsuBataDlanoitceriD-iB 31dnuorG 41desUtoN 51dnuorG
61)BSM(51tibsuBataDlanoitceriD-iB 71dnuorG
81)woLevitcA(noitceriDpetSksiDoT 91dnuorG
02)woLevitcA(dnammoCpetSksiDoT 12dnuorG 22ataDetirWksiDksiDoT 32dnuorG 42dnammoCetirWksiDksiDoT 52dnuorG 62esnes0kcarTksiDmorF 72dnuorG
82)woLevitcA(1tceleSpihCrossecorporciMksiDoT 92dnuorG 03ataDdaeRksiDksiDmorF 13dnuorG
23)woLevitcA(tceleSdaeHksiDksiDoT 33dnuorG 43]nipetanretlA[)woLevitcA(egnahCksiDksiDmorF
Analog - Floppy Power Cable
niPnoitcnuF 1CDV5+ 2dnuorG 3dnuorG 4CDV21+
Syntec
VitMan
ãã
ã
ãã
page 29
Connector pinouts
Page 31
System Description
Analog - Bulb Cable
niPnoitcnuFnoitcennoC
1blubotCDV42+601FaivCDV42+ 2nruterMRAWbluBrotsiserrewopaivblubrofnruteR 3nruterbluBdnuorgotdehctiwS
Analog - Pneumatics Power Cable
niPnoitcnuFnoitcennoC
1pmuPmuucaVotrewoPCDV21+101FaivCDV21+ 2nruteRpmuPmuucaVdnuorgotdehctiwS
3
4nruteRrosserpmoCniaMdnuorgotdehctiwS
niaMotrewoPCDV21+
rosserpmoC
201FaivCDV21+
Foot Pedal Signal Cable (internal)
niPnoitcnuF
1dnuorG 22FesufaivylppusladepCDV5+ 3C/N 4*dellatsnIladeP 5*thgiRladeP 6C/N 7*tfeLladeP
8)desutoN(*nwoDladeP 9*PUladeP
01noitisoPladeP
* This signal is active when the line is near 0Vdc
Analog - Pneumatics Control Cable
niPnoitcnuFnoitcennoC
1
)CDV5 2CDV21+rotalugeRmuucaVotrewoP 3dnuorGrotalugeRmuucaVotnruteR 4noitcennoCoN 5nruteRevlaVrotcertiV evlavrotcertivehtetavitcaotdnuorgotdehctiwS 6rewoPevlaVrotcertiVevlavehtotrewop001FaivCDV21+
7
nruteR
8
rewoP
9nruteRevlaVxulfeRevlavhcnipehtetavitcaotdnuorgotdehctiwS
01rewoPevlaVxulfeRevlavehtotrewop001FaivCDV21+
11nruteRevlaVnoitagirrI evlavhcnipehtetavitcaotdnuorgotdehctiwS
21rewoPevlaVnoitagirrIevlavehtotrewop001FaivCDV21+
31nruteRevlaVegnahcxEriA
41rewoPevlaVegnahcxEriAevlavehtotrewop001FaivCDV21+
51
nruteR
61
rewoP
bluBB/ApukcaBevlaV
71
nruteR
bluBB/ApukcaBevlaV
81
rewoP
91nruteRrotoMegnahcxEriA
02rewoProtoMegnahcxEriA .derisederusserpehtnognidnepedCDV22+ot0
-0(langiSlortnoCmuucaV
evlaVhcniPnoitaripsA
evlaVhcniPnoitaripsA
evlaVpukcaBbluBB/A
evlaVpukcaBbluBB/A
rotalugeRmuucaVoT
.tuptuoria
.71nipotsuoenatlumiS
.pukcabecruosthgil
rotomegnahcxEriAehtevird
evlavhcnipehtetavitcaotdnuorgotdehctiwS
evlavehtotrewop001FaivCDV21+
selbanE.evlavehtetavitcaotdnuorgotdehctiwS
.evlavehtetavitcaotdnuorgotdehctiwS
evlavehtotrewop001FaivCDV21+
sevoM.evlavehtetavitcaotdnuorgotdehctiwS
evlavehtotrewop001FaivCDV21+
otsedoidrenezV9.3owtaivdnuorgotdehctiwS
Analog - Fan Power Cable
niPnoitcnuF
1401FaivCDV21+
2dnuorG
Analog - Speaker Cable
niPnoitcnuF
1+rekaepsoT
2-rekaepsoT
Analog - Ultrasound Power Cable
niP
1detcennoCtoN1detcennoCtoN 2detcennoCtoN2detcennoCtoN 3detcennoCtoN3detcennoCtoN 4detcennoCtoN4detcennoCtoN 5detcennoCtoN5detcennoCtoN 6detcennoCtoN6detcennoCtoN 7detcennoCtoN7)DNG(dnuorG
8)DNG(dnuorG8)DNG(dnuorG
9)901Faiv(CDV42+9detcennoCtoN
draoBgolanA
rotcennoC
niP
01detcennoCtoN 11detcennoCtoN 21detcennoCtoN
31)901Faiv(CDV42+
41)901Faiv(CDV42+
rotcennoC
draoBdnuosartlU
Analog – Power Supply Cable
Analog - Ultrasound Control Cable
niPnoitcnuFsesuFsdeeF 1CDV21+311F,501F,401F 2CDV21+001F 3CDV42+901F,601F,301F 4dnuorG 5dnuorG 6dnuorG 7CDV21+101F,201F 8CDV21+ 9dnuorG
dnuosartlUniP
noitcnuF
.nnoCdraoB
1detcennoCtoN 2dnuorG31 3detcennoCtoN 4detcennoCtoN 5detcennoCtoN 6detcennoCtoN 7dednuorg=wol,tluaFdnuorG8 8detcennoCtoN 9detcennoCtoN
01detcennoCtoN
11)CDV0.5-5.0+(lortnoCrewoP41 21detcennoCtoN 31hgihevitca,elbanEtuptuOrewoP6 41detcennoCtoN 51detcennoCtoN 61detcennoCtoN
golanAniP
.nnoCdraoB
Connector pinouts
page 30
Syntec
VitMan
ãã
ã
ãã
Page 32
System Description

Pneumatic Block Diagram

Vitrector Out
Vitrector
Control
Valve
Regulator
Aspiration
Pinch Valve
Fluid Control
Control
Control
Valve
Regulator
Valve
Reflux
Pinch Valve
Control
Valve
Irrigation
Pinch Valve
Air Cylinder
Light Source Backup
Control
Valve
Control
Valve
Sensor
Sensor
Control
Valve
Air Exchange Out
Accumulator
Sensor
Air Supply
Main Compressor
Manifold
Filter
Relief
Valve
Aspiration
Regulator
Vacuum
Regulator
Vacuum Pump
Sensor
Aspiration Out
Controlled
Vacuum
Sensor
Vacuum Accumulator
and Aspirant Storage
Air Exchange
Low Pressure
Compressor
Syntec
VitMan
ãã
ã
ãã
page 31

Pneumatic block diagram

Page 33

Status Line Messages

Status Line Messages
The status line of the LCD display will show information concerning the current state of the device. All error and warning conditions will generate a status line message. In addition to error and warning messages, informational messages are displayed as well. This list contains all types of status line messages, sorted in alphabetical order. Following each message is an explanation of, and in some cases a possible solution to, the condition being reported.

dd-mmm-yy hh:mm:ss

The date and time will display when the date & time is selected for the status line default and no other messages need to be displayed. The status line default may be selected using the “Status Line Default” command in the “Display” menu. The date and time can be modified using the “Date & Time” menu, found in the main menu. The date and time are used to time stamp any files created.

Dr. Cutright

The user name will display when the user name is selected for the status line default and no other messages need to be displayed. The status line default may be selected using the “Status Line Default” command in the “Display” menu. The user name can be modified using the “Edit User Name” command, found in the “Select User” menu.

+12 volt fuse problem

The +12 volt fuse on the analog board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
This message indicates that the +12 volt power is not present. In addition to the fuse opening, this message could be caused by a problem with the DC-DC converter that gener­ates the +12 volt power.

-12 volt fuse problem

The -12 volt fuse on the analog board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
This message indicates that the -12 volt power is not present. In addition to the fuse open­ing, this message could be caused by a problem with the DC-DC converter that generates the -12 volt power.

-22 volt fuse problem

The -22 volt fuse on the analog board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
This message indicates that the -22 volt power is not present. The -22 volt power is used to generate the contrast for the LCD display. If the -22 volt power is not present it is un­likely that this message will be legible. In addition to the fuse opening, this message could
Status line messages
page 32
Syntec
VitMan
ãã
ã
ãã
Page 34
Status Line Messages
be caused by a problem with the DC-DC converter that generates the -22 volt power or the absence of the +12Vdc-Main voltage. The +12Vdc-Main voltage is fused and monitored.

500cc sensor problem

The vacuum sensor used to monitor the vacuum pump pressure is indicating a problem. If the aspiration cassette is present and the vacuum level is zero or if the aspiration cassette is missing and pressure is present, this message will be displayed in the status line. This message may appear along with the “Low vacuum level”, “Check cassette tubing” and “Please lock cassette” messages, when the aspiration cassette is present but not locked.
This message indicates that the 500cc sensors output voltage appears to be low. The 500cc sensors output voltage is read using an A/D converter. This voltage is then reduced by the voltage present when the sensors port is at atmosphere. The result represents the vacuum level present on the sensors port. If the voltage read is less than the previously recorded voltage present at atmosphere, then pressure may be present on the sensor, the A/D converter may have a problem or the sensor may have a problem. A leak in the vacuum system may also cause this message.

50cc sensor problem

The vacuum sensor used to monitor the small chamber of the aspiration cassette is indicat­ing a problem. If the aspiration cassette is present and the vacuum level is zero (when vacuum should be present) or if the aspiration cassette is missing and pressure is present, this message will be displayed in the status line.
This message indicates that the 50cc sensors output voltage appears to be low. The 50cc sensors output voltage is read using an A/D converter. This voltage is then reduced by the voltage present when the sensors port is at atmosphere. The result represents the vacuum level present on the sensors port. If the voltage read is less than the previously recorded voltage present at atmosphere, then pressure may be present on the sensor, the A/D converter may have a problem or the sensor may have a problem. A leak in the aspiration system may also cause this message.

Air exchange xxx mmHg

While the desired air exchange pressure is modified, the current eye pressure will be dis­played in the status line, where “xxx” is the value of the eye pressure.

Air exchange excess

When the air exchange pressure exceeds the warning pressure limit for a time longer than the warning time limit, this message is displayed in the status line. The warning pressure limit and the warning time limit may be adjusted using the “Air Exchange Values” menu in the “Current Settings” menu.

Air exchange problem

If the air exchange valve is open, the pressure difference between the eye pressure sensor and the air exchange pump pressure sensor should be less than 15 mmHg (for low flow rate conditions the pressure difference will be close to zero mmHg). If a pressure differ
Syntec
VitMan
ãã
ã
ãã
page 33
Status line messages
Page 35
Status Line Messages
ence of 15 mmHg or greater is detected, this message will be displayed. This message will appear along with the “Service required” message.

Air exchange set: xxx

While the air exchange surgical function is on, the pressure setting is displayed in the status line, where “xxx” is the pressure setting value in mmHg.

Air pump fuse problem

The air exchange pump fuse on the analog board has opened. This fuse is not resettable and will need to be replaced in order to clear the problem.
This message indicates that the +24 volt power is not present on the air exchange pump control circuitry. This fuse also protects the audio circuitry. The speaker will not operate if the fuse has failed. In addition to the fuse opening, this message could be caused by a problem with the +24 volt output of the power supply.

Air pump hours excess

The run time on the air exchange pump has exceeded the maximum limit. The pneumatics module needs to be serviced. The “Display Pump Hours” command, found in the “Service Pneumatics” menu of the “Service Menu”, can be used to display the current air pump hours.
If the air exchange pump hours exceed 2000, this message will be displayed. While the air exchange surgical function is on this message is disabled.

Air pump problem

The test of the air exchange pump indicates that it may not be capable of generating pres­sure. This message may be accompanied by the “Air pump fuse problem” or the “Service required” messages.
At power-up, a test is made of the ability of the air exchange pump to generate pressure. If pressure is not detected at power-up, the test will be run each time the air exchange surgi­cal function is enabled. If the test fails, this message will be displayed.

Air sensor problem

The pressure sensor used to monitor the air pump used in the air exchange surgical func­tion is indicating a problem. If the air exchange surgical function is off and vacuum is present or if the air exchange surgical function is on and no pressure is present, this mes­sage will be displayed in the status line.
This message indicates that the air exchange pump sensors output voltage appears to be low. The air exchange pump sensors output voltage is read using an A/D converter. This voltage is then reduced by the voltage present when the sensors port is at atmosphere. The result represents the pressure level present on the sensors port. If the voltage read is less than the previously recorded voltage present at atmosphere, then vacuum may be
Status line messages
page 34
Syntec
VitMan
ãã
ã
ãã
Page 36
Status Line Messages
present on the sensor, the A/D converter may have a problem or the sensor may have a problem. A leak in the air exchange system may also cause this message.

Alarm cut off

While not in menu mode, the ENTER button will allow alarm cut off. Pressing and releas­ing the ENTER button will mute any currently playing alarm or warning tone. This is nor­mally only required when the error or warning tones are in continuous mode. The error and warning tone modes may be adjusted using the “Audio Services” menu.

Anterior Aspiration

This message indicates that the aspiration surgical function has entered anterior aspiration mode, using the aspiration on/off button.

Anterior Fixed Phaco

This message indicates that the ultrasound surgical function has entered the anterior fixed phaco mode, using the ultrasound on/off button.

Anterior Frag

This message indicates that the ultrasound surgical function has entered the anterior frag mode, using the ultrasound on/off button.

Anterior Linear Phaco

This message indicates that the ultrasound surgical function has entered the anterior linear phaco mode, using the ultrasound on/off button.

Anterior Linear Rate

This message indicates that the vitrector surgical function has entered anterior linear cut rate mode, using the vitrector on/off button.

Anterior Vitrectomy

This message indicates that the vitrector surgical function has entered the anterior vitrec­tomy mode, using the vitrector on/off button.
Bulb “A” FET problem
If bulb “A” is present and enabled but is not illuminated, this message will be displayed in the status line. A field effect transistor (FET) is used to sink the bulb current and illuminate the bulb. This message indicates a hardware problem that requires servicing.
A field effect transistor (FET) is used to sink the bulb current and illuminate the bulb. The bulb detection logic uses the voltage across the FET to determine the status of the bulb. The voltage across the FET is read using an A/D converter. If the voltage across the FET is greater than 2.5 volts while the bulb is enabled, this message will be displayed. If the A/ D converter has a problem and reports a high voltage while the bulb is enabled, this mes­sage will be displayed.
Syntec
VitMan
ãã
ã
ãã
page 35
Status line messages
Page 37
Status Line Messages
Bulb “A” hours xxx
This is the in use time for bulb “A”, where “xxx” is in hours. This message will be displayed when the illumination surgical function is turned on, and the current bulb is bulb “A”.
Bulb “A” large hours
When the VitMan is powered up or if illumination is turned on while bulb “A” is selected, this message will be displayed if the bulb “A” hours exceed 40 hours.
Bulb “A” missing
If bulb “A” is not detected when the VitMan is powered up, this message will be displayed. A field effect transistor (FET) is used to sink the bulb current and illuminate the bulb. The
bulb detection logic uses the voltage across the FET to determine the presence of the bulb. The voltage across the FET is read using an A/D converter. If the voltage across the FET is less than 2.5 volts, if the A/D converter has a problem and reports a low voltage, if the bulb power fuse is open or if the +24 volt output of the power supply is not present, the bulb will not be detected.
Bulb “A” open
When bulb “A” is not detected, this message will be displayed in the status line. The most likely solution is to replace the bulb.
A field effect transistor (FET) is used to sink the bulb current and illuminate the bulb. The bulb detection logic uses the voltage across the FET to determine the presence of the bulb. The voltage across the FET is read using an A/D converter. If the voltage across the FET is less than .03 volts, this message will be displayed. If the A/D converter has a problem and reports a low voltage, this message will be displayed. The power to the bulbs is fused and if the fuse is open this message will be displayed. In addition to the fuse opening, this message could be caused by a problem with the +24 volt output of the power supply.
Bulb “A” problem
If the bulb detection logic indicates bulb “A” is on, while the illumination surgical function is off, this message will be displayed in the status line.
A field effect transistor (FET) is used to sink the bulb current and illuminate the bulb. The bulb detection logic uses the voltage across the FET to determine the presence of the bulb. The voltage across the FET is read using an A/D converter. If illumination is off and the voltage across the FET is greater than .03 volts, but less than 2.5 volts, or if the A/D con­verter has a problem and reports a voltage within this range, this message will be dis­played. If illumination is on this message will not be displayed.
Bulb “B” FET problem
If bulb “B” is present and enabled but is not illuminated, this message will be displayed in the status line. A field effect transistor (FET) is used to sink the bulb current and illuminate the bulb. This message indicates a hardware problem that requires servicing.
Status line messages
page 36
Syntec
VitMan
ãã
ã
ãã
Page 38
Status Line Messages
A field effect transistor (FET) is used to sink the bulb current and illuminate the bulb. The bulb detection logic uses the voltage across the FET to determine the status of the bulb. The voltage across the FET is read using an A/D converter. If the voltage across the FET is greater than 2.5 volts while the bulb is enabled, this message will be displayed. If the A/ D converter has a problem and reports a high voltage while the bulb is enabled, this mes­sage will be displayed.
Bulb “B” hours xxx
This is the in use time for bulb “B”, where “xxx” is in hours. This message will be displayed when the illumination surgical function is turned on, and the current bulb is bulb “B”.
Bulb “B” large hours
When the VitMan is powered up or if illumination is turned on while bulb “B” is selected, this message will be displayed if the bulb “B” hours exceed 40 hours.
Bulb “B” missing
If bulb “B” is not detected when the VitMan is powered up, this message will be displayed. A field effect transistor (FET) is used to sink the bulb current and illuminate the bulb. The
bulb detection logic uses the voltage across the FET to determine the presence of the bulb. The voltage across the FET is read using an A/D converter. If the voltage across the FET is less than 2.5 volts, if the A/D converter has a problem and reports a low voltage, if the bulb power fuse is open or if the +24 volt output of the power supply is not present, the bulb will not be detected.
Bulb “B” open
When bulb “B” is not detected this message will be displayed in the status line. The most likely solution is to replace the bulb.
A field effect transistor (FET) is used to sink the bulb current and illuminate the bulb. The bulb detection logic uses the voltage across the FET to determine the presence of the bulb. The voltage across the FET is read using an A/D converter. If the voltage across the FET is less than .03 volts, this message will be displayed. If the A/D converter has a problem and reports a low voltage, this message will be displayed. The power to the bulbs is fused and if the fuse is open this message will be displayed. In addition to the fuse opening, this message could be caused by a problem with the +24 volt output of the power supply.
Bulb “B” problem
If the bulb detection logic indicates bulb “B” is on, while the illumination surgical function is off, this message will be displayed in the status line.
A field effect transistor (FET) is used to sink the bulb current and illuminate the bulb. The bulb detection logic uses the voltage across the FET to determine the presence of the bulb. The voltage across the FET is read using an A/D converter. If illumination is off and
ãã
ã
Syntec
VitMan
ãã
page 37
Status line messages
Page 39
Status Line Messages
the voltage across the FET is greater than .03 volts, but less than 2.5 volts, or if the A/D converter has a problem and reports a voltage within this range, this message will be dis­played. If illumination is on this message will not be displayed.

Bulb fuse problem

The illumination fuse on the analog board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
This message indicates that the bulb power is not present. In addition to the fuse opening, this message could be caused by a problem with the +24 volt output of the power supply.

Calibrate air pump

If the tables used to regulate the air exchange operation are invalid, this message will be displayed in the status line. If this message is displayed, it is recommended that the air exchange be calibrated before use. The air exchange can be calibrated using the “Calibra­tion Routines” found in the “Service Menu”.

Calibrate aspiration

If the tables used to regulate the aspiration operation are invalid, this message will be displayed in the status line. If this message is displayed, it is recommended that the aspi­ration surgical function be calibrated before use. Aspiration can be calibrated using the “Calibration Routines” found in the “Service Menu”.

Calibrate foot pedal

If the values used to regulate the foot pedal operation are invalid, this message will be displayed in the status line. If this message is displayed, it is recommended that the foot pedal be calibrated before use. The foot pedal can be calibrated using the “Calibration Routines” found in the “Service Menu”.

Cassette full

If the cassette level detection logic determines that the aspiration cassette is full, this mes­sage will be displayed in the status line. This is the second level of warning and occurs after about 300 cc of fluid are aspirated.

Cassette over full

If the cassette level detection logic determines that the aspiration cassette is over full, this message will be displayed in the status line. This is the last level of warning and occurs after about 325 cc of fluid are aspirated. When this condition is detected the vacuum pump will stop and aspiration is no longer possible. Replacement the aspiration cassette is recommended.

Cassette position bad

If the cassette level detection logic determines that the aspiration cassette is not properly aligned, this message will be displayed in the status line. This generally indicates that the cassette is pushed into the cassette chamber prior to rotating the knob into the lock posi­tion. To correct the problem, rotate the knob to the unlocked position then back to the
Status line messages
page 38
Syntec
VitMan
ãã
ã
ãã
Page 40
Status Line Messages
locked position without pressing on the cassette. The “Re-position cassette” and “Use locking knob only” messages will also be displayed.

Cassette problem

If a high fluid level is detected in the small chamber of the aspiration cassette, this mes­sage will be displayed in the status line. This condition will stop the vacuum pump and prevent the use of aspiration. If this condition is suspected to have been detected in error, remove the aspiration cassette, wait for one second, and reinsert the aspiration cassette. If the condition persists, replace the aspiration cassette. The “Replace cassette” message will also be displayed.
This message could be caused by fluid drops on the 50cc chamber prism. It could also be caused by a failure of the cassette valve. If the valve doesn’t open, then the fluid level will eventually cover the prism. Mechanical alignment of the analog electronics board could cause this message to display when using functional cassettes. The analog electronics board should be aligned directly against the inside of the front panel of the system enclo­sure.

Check cassette tubing

If the foot pedal is in the rest position and the vacuum level in the 500cc chamber falls below 400 mmHg, this message will be displayed in the status line. This may be the result of using the aspiration surgical function or requesting a prime cycle, with no tubing con­nected to the aspiration cassette. This message may appear along with the “Low vacuum level” and “Please lock cassette” messages.
When the aspiration surgical function is on and the foot pedal is in the rest position, an attempt is made to keep the 50cc chamber vacuum level at 1 mmHg. If tubing is not present, the vacuum level of the 500cc chamber may be reduced to below an acceptable value. If trying to keep the 50cc chamber at 1 mmHg causes a reduction in the 500cc chamber vacuum level, the “Check cassette tubing” status line message will be displayed. If the vacuum level is reduced during normal use, this status line message should not be displayed.

Check external source

This message will be displayed if a problem is detected with the system pressure or vacuum level and the external connections are selected for use. If the pressure compres­sor mode is set to external pressure or auto select and the “Low system pressure” or “High system pressure” status line messages are displayed, this message will also be displayed. If the vacuum pump mode is set to external vacuum or auto select and the “Low vacuum level” status line message is displayed, this message will also be displayed. The most likely cause of this message is that the hose(s) are connected to the VitMan, but not to the external source of pressure (or vacuum). The user should check the external connections, and if connected, then check the external source pressure levels.
Syntec
VitMan
ãã
ã
ãã
page 39
Status line messages
Page 41
Status Line Messages

Contrast adjust

While not in menu mode, the LEFT ARROW and RIGHT ARROW can be used to adjust the contrast of the LCD display. This message will be displayed as the contrast is adjusted.

Currently in stage 1

When the stage button is pressed and released, the current stage will be displayed. If the current stage is stage one, this message will be displayed in the status line.

Currently in stage 2

When the stage button is pressed and released, the current stage will be displayed. If the current stage is stage two, this message will be displayed in the status line.

Currently in stage 3

When the stage button is pressed and released, the current stage will be displayed. If the current stage is stage three, this message will be displayed in the status line.

Currently in stage 4

When the stage button is pressed and released, the current stage will be displayed. If the current stage is stage four, this message will be displayed in the status line.

Currently in stage 5

When the stage button is pressed and released, the current stage will be displayed. If the current stage is stage five, this message will be displayed in the status line.

Currently in stage 6

When the stage button is pressed and released, the current stage will be displayed. If the current stage is stage six, this message will be displayed in the status line.

Currently in stage 7

When the stage button is pressed and released, the current stage will be displayed. If the current stage is stage seven, this message will be displayed in the status line.

Device disabled

If the internal device temperature is 55 °C or higher, this message is displayed in the status line. This message indicates an error condition and after ten seconds the surgical functions will be affected. The illumination surgical function will be turned off. The system pressure compressor and vacuum pump will be turned off, effectively disabling aspiration, irrigation, the vitrector and the ultrasound. The air exchange surgical function will remain unaffected.
The temperature sensor monitors the internal temperature of the system and resides on the analog board. The temperature sensor is a serial device that provides a properly for­matted response to a number of commands. If proper communication is not established, any values read are not used. If the temperature sensor has a problem and reports a high temperature condition, this message may be displayed.
Status line messages
page 40
Syntec
VitMan
ãã
ã
ãã
Page 42
Status Line Messages

Disk available hh:mm

This is the minimum time remaining to record data on the floppy disk, where “hh” is the hours and “mm” is the minutes. After the data from the previous minute is saved on the disk, this message is displayed in the status line.

Disk space warning

If the minimum time remaining is less than 30 minutes, this message will be displayed in the status line.

Disk write protected

The write protect status of the floppy disk is checked when an attempt is made to save data on the disk. If the floppy disk is write protected, this message will be displayed in the status line.

Diskette is full

If no free space remains on the floppy disk, this message is displayed in the status line.

Diskette read problem

If the boot block, file allocation tables or root directory of the floppy disk cannot be read, this message is displayed in the status line. If this condition persists, the disk may need to be formatted.
If the data cable between the CPU board and the disk drive is faulty or if power is not present on the disk drive, this message will be displayed. The disk drive has a jumper on the drive select signal. If the jumper is missing or selecting drive 0, this message will be displayed.

Drive fuse problem

The main compressor (vitrector drive pressure) fuse on the analog board has opened. This fuse is not resettable and will need to be replaced in order to clear the problem.
This message indicates that the main compressor power is not present. In addition to the fuse opening, this message could be caused by a problem with the +12 volt output of the power supply.

Drive sensor problem

The pressure sensor used to monitor the main compressor (vitrector drive pressure) is indicating a problem. The sensor is indicating that the system pressure is zero.
This message indicates that the system pressure sensors output voltage appears to be low. The system pressure sensors output voltage is read using an A/D converter. This voltage is then reduced by the voltage present when the sensors port is at atmosphere. The result represents the pressure level present on the sensors port. If the voltage read is less than the previously recorded voltage present at atmosphere, then vacuum may be
Syntec
VitMan
ãã
ã
ãã
page 41
Status line messages
Page 43
Status Line Messages
present on the sensor, the A/D converter may have a problem or the sensor may have a problem. A leak in the system pressure may also cause this message.

Eye pressure alarm

If the air exchange surgical function is on and the current eye pressure is not within (5 mmHg of the current set point, this message is displayed in the status line.

Eye sensor problem

The pressure sensor used to monitor the eye pressure, for the air exchange surgical func­tion, is indicating a problem. If the air exchange surgical function is off and vacuum is present or if the air exchange surgical function is on and no pressure is present, this mes­sage will be displayed in the status line.
This message indicates that the eye pressure sensors output voltage appears to be low. The eye pressure sensors output voltage is read using an A/D converter. This voltage is then reduced by the voltage present when the sensors port is at atmosphere. The result represents the pressure level present on the sensors port. If the voltage read is less than the previously recorded voltage present at atmosphere, then vacuum may be present on the sensor, the A/D converter may have a problem or the sensor may have a problem. A leak in the air exchange system may also cause this message.

Fan fuse problem

The fan fuse on the analog board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
This message indicates that the fan power is not present. In addition to the fuse opening, this message could be caused by a problem with the +12 volt output of the power supply.

Floppy fuse problem

The disk drive fuse on the analog board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
This message indicates that the disk drive power is not present. In addition to the fuse opening, this message could be caused by a problem with the +12 volt output of the power supply or the disk drives +5 volt regulator.

Foot pedal is active

When the foot pedal is active, several actions are prohibited. The user may not change the mode of the aspiration, vitrector or ultrasound surgical functions. The ultrasound hand­piece may not be tuned. The aspiration valve button may not be used to open or close the aspiration valve or start a prime cycle. While in anterior mode, the irrigation valve button may not be used to open or close the irrigation valve. The stage button may not be used to change mode of the surgical functions.
Status line messages
page 42
Syntec
VitMan
ãã
ã
ãã
Page 44
Status Line Messages

Foot pedal problem

If the foot pedal down switch indicates that the foot pedal is up, but the foot pedal position indicates depression of 10% or more, this message will be displayed in the status line. This indicates a problem with the foot pedal down switch (internal to the foot pedal), the foot pedal cable or connector. The hardware that senses the foot pedal down switch may also have a problem.
The condition of the foot pedal, as measured by the system software, can be viewed using the “Calibrate Foot Pedal” command under “Calibration Routines” under the “Service Menu”.

Ground fault detected

If a ground fault is detected by the ultrasound hardware, this message will be displayed in the status line. The ground fault indicates that the ultrasound handpiece is not properly grounded. No energy will be delivered to an ungrounded ultrasound handpiece. This problem may be corrected by disconnecting and reconnecting the ultrasound handpiece.
An interruption of the ground path anywhere can cause this message to display. Check the ground connections of the ultrasound output cable at both the front and rear panels of the system. Also, the ultrasound electronics board uses the board mounting hardware to sense ground. If the mounting hardware isn’t grounded then a ground fault will be sensed. Note that the entire system chassis is part of the ground path, so the front and rear panels of the system are ground.

High cassette level

If the cassette level detection logic determines that the aspiration cassette level is high, this message will be displayed in the status line. This is the first level of warning and occurs after about 250 cc of fluid are aspirated.

High pump pressure

If the air exchange surgical function is on and the air exchange pump pressure exceeds the current set point by more than 8 mmHg, this message is displayed in the status line.

High system pressure

If the main pressure pump pressure exceeds the upper limit, this message is displayed in the status line. This is only a warning, no surgical functions will be disabled.
The nominal value for the main pump pressure is 2585 mmHg (50 psi). This message will be displayed if the main pump pressure is above 2844 mmHg (55 psi).

High temperature xx (C

If the internal device temperature is 50 °C or higher, this message is displayed in the status line. The current temperature is “xx”. This message is only a warning and no surgical functions are effected. The temperature messages may be disabled in the “Display” menu.
Syntec
VitMan
ãã
ã
ãã
page 43
Status line messages
Page 45
Status Line Messages
The temperature sensor monitors the internal temperature of the system and resides on the analog board. The temperature sensor is a serial device that provides a properly for­matted response to a number of commands. If proper communication is not established, any values read are not used. If the temperature sensor has a problem and reports a high temperature condition, this message will be displayed. Under normal operating conditions, the internal temperature of the system won’t reach this limit. Check for air flow restrictions around the system. Especially near the main system fan and the air duct at the rear of the system.

High vacuum level

If the aspiration surgical function is on and the 50cc vacuum level exceeds the desired level by more than 20 mmHg, this message is displayed in the status line.

Hold for menu mode

While not in menu mode, holding the DISPLAY button for one second will enter menu mode. If the button is released before entering menu mode, this message is displayed in the status line.

Hold to change bulbs

When the illumination surgical function is turned off, this message will be displayed in the status line. If the illumination surgical function is off, pressing and holding the illumination on/off button can be used to switch to the backup bulb.

Hold to enter stage 1

When the stage button is pressed and held, if stage one will be entered when the stage button is released this message will be displayed in the status line.

Hold to enter stage 2

When the stage button is pressed and held, if stage two will be entered when the stage button is released this message will be displayed in the status line.

Hold to enter stage 3

When the stage button is pressed and held, if stage three will be entered when the stage button is released this message will be displayed in the status line.

Hold to enter stage 4

When the stage button is pressed and held, if stage four will be entered when the stage button is released this message will be displayed in the status line.

Hold to enter stage 5

When the stage button is pressed and held, if stage five will be entered when the stage button is released this message will be displayed in the status line.

Hold to enter stage 6

When the stage button is pressed and held, if stage six will be entered when the stage button is released this message will be displayed in the status line.
Status line messages
page 44
Syntec
VitMan
ãã
ã
ãã
Page 46
Status Line Messages

Hold to enter stage 7

When the stage button is pressed and held, if stage seven will be entered when the stage button is released this message will be displayed in the status line.

Hold to save stage 1

When the stage button is pressed and held, this message will be displayed in the status line, prior to saving stage one. If the stage button is continued to be held, the current system state will be saved.

Hold to save stage 2

When the stage button is pressed and held, this message will be displayed in the status line, prior to saving stage two. If the stage button is continued to be held, the current sys­tem state will be saved.

Hold to save stage 3

When the stage button is pressed and held, this message will be displayed in the status line, prior to saving stage three. If the stage button is continued to be held, the current system state will be saved.

Hold to save stage 4

When the stage button is pressed and held, this message will be displayed in the status line, prior to saving stage four. If the stage button is continued to be held, the current system state will be saved.

Hold to save stage 5

When the stage button is pressed and held, this message will be displayed in the status line, prior to saving stage five. If the stage button is continued to be held, the current sys­tem state will be saved.

Hold to save stage 6

When the stage button is pressed and held, this message will be displayed in the status line, prior to saving stage six. If the stage button is continued to be held, the current sys­tem state will be saved.

Hold to save stage 7

When the stage button is pressed and held, this message will be displayed in the status line, prior to saving stage seven. If the stage button is continued to be held, the current system state will be saved.

Hold to start prime

When the aspiration valve button is pressed and released (and not held long enough to start the prime cycle), this message will be displayed in the status line.
Syntec
VitMan
ãã
ã
ãã
page 45
Status line messages
Page 47
Status Line Messages

LED/backlight adjust

While not in menu mode, the UP ARROW and DOWN ARROW can be used to adjust the intensity of the LCD display backlight and the surgical function LEDs. This message will be displayed as the intensity is adjusted.

Lift pedal to prime

To start a prime cycle the aspiration pinch valve button must be pressed and held for 1/2 second. If the foot pedal is down when a prime cycle is started, this message will be dis­played in the status line.

Lift pedal to reflux

If the foot pedal is depressed to a level that allows aspiration, reflux is disabled. If reflux is attempted while aspirating, this message will be displayed in the status line.

Low air pump pressure

If the air exchange surgical function is on and the air exchange pump pressure is more than 8 mmHg below the current set point, this message is displayed in the status line.

Low system pressure

If the main pressure pump pressure is below acceptable levels, this message is displayed in the status line. When low system pressure is detected the vitrector, ultrasound and aspiration surgical functions will be disabled.
When the vitrector is used at high cut rates the volume of air consumed may cause this message to be displayed. The nominal value for the main pump pressure is 2585 mmHg (50 psi). This message will be displayed if the main pump pressure is below 2172 mmHg (42 psi). When the system is operating properly, the system pressure won’t drop this low. Improper adjustment of the vitrector pressure regulator (too high), air leaks, clogged air filter, and poor compressor performance could all cause this problem.

Low temperature (xx (C

If the internal device temperature is 10 °C or lower, this message is displayed in the status line. The current temperature is “xx” and when below zero will be preceded by a minus sign (-). This message is only a warning and no surgical functions are effected. The tem­perature messages may be disabled in the “Display” menu.
The temperature sensor monitors the internal temperature of the system and resides on the analog board. The temperature sensor is a serial device that provides a properly for­matted response to a number of commands. If proper communication is not established, any values read are not used. If the temperature sensor has a problem and reports a low temperature condition, this message may be displayed.

Low vacuum level

If the vacuum pump vacuum is below acceptable levels, this message is displayed in the status line. When this condition occurs aspiration is not disabled, however, not all levels of
Status line messages
page 46
Syntec
VitMan
ãã
ã
ãã
Page 48
Status Line Messages
vacuum may be obtainable. This message may appear along with the “Check cassette tubing” and “Please lock cassette” messages.
If the 500cc vacuum level is below the aspiration surgical function’s maximum vacuum setting or is below 400 mmHg, then this message is displayed. This message indicates that the 500cc sensors output voltage appears to be low. The 500cc sensors output volt­age is read using an A/D converter. This voltage is then reduced by the voltage present when the sensors port is at atmosphere. The result represents the vacuum level present on the sensors port. If the voltage read indicates a low vacuum level the A/D converter may have a problem or the sensor may have a problem. A leak in the vacuum system may also cause this message. When the aspiration surgical function is on and the foot pedal is in the rest position, an attempt is made to keep the 50cc chamber vacuum level at 1 mmHg. If tubing is not present and connected properly, the vacuum level of the 500cc chamber may be reduced to below an acceptable value.

Main +12 fuse problem

The Main +12 fuse on the analog board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
This message indicates that the +12Vdc-Main power is not present. In addition to the fuse opening, this message could be caused by a problem with the +12 volt output of the power supply.

Minimum cut rate:xxxx

When the vitrector surgical function enters posterior or anterior linear rate mode this mes­sage will display the minimum cut rate, where “xxxx” is the cut rate.

No cassette present

If the aspiration cassette is not detected, this message will be displayed in the status line. While the aspiration cassette is missing, the vacuum pump is disabled and the aspiration cassette illumination is reduced.
The system uses the fluid level sensor array on the analog electronics board to detect the presence of a cassette. The 500cc chamber prism causes light to be reflected back to the sensor array. If the cassette is extremely full of fluid, the sensor array will fail to “see” the prism.

No diskette present

Once a minute the trace data is written to the floppy disk. If no floppy disk is detected in the disk drive, this message is displayed in the status line.

No foot pedal present

If the foot pedal cannot be detected, this message is displayed in the status line. Without the foot pedal, many surgical functions are inoperable. Only the air exchange and illumina­tion will function without the foot pedal.
ãã
ã
Syntec
VitMan
ãã
page 47
Status line messages
Page 49
Status Line Messages
The foot pedal presence is sensed through the cable connected to the foot pedal. Pin 1 is shorted to pin 4 internal to the foot pedal. The condition of the foot pedal, as measured by the system software, can be viewed using the “Calibrate Foot Pedal” command under “Calibration Routines” under the “Service Menu”

No handpiece present

If the ultrasound surgical function is on and the handpiece is not present, this message will be displayed in the status line.
The handpiece presence is sensed through the cable connector. 2 pins are shorted internal to the cable connector. A connection is made between the ultrasound output connector on the front panel and the CPU electronics board to enable the system to measure for the presence of the handpiece.

No keyboard heartbeat

If the keyboard encoder does not answer a heartbeat request, this message will be dis­played in the status line. The heartbeat request is made once every second. If the heart­beat request is unanswered, the keyboard encoder is reset.

No stage selected

When the stage button is pressed and released, the current stage will be displayed. If no stage was selected, this message will be displayed in the status line.

No stages enabled

If no stages (other than the current stage) are enabled, and the stage button is pressed and held, this message will be displayed in the status line. This indicates that no stage is available as the next stage

No ultrasound present

If the ultrasound option is not present when the ultrasound on/off button is pressed, this message will be displayed in the status line.
The ultrasound electronics has a serial EEPROM that holds the serial number and other parameters associated with the ultrasound surgical function. If the ultrasound data cable is disconnected, if power is not present on the ultrasound board or if the serial EEPROM has a problem, the serial EEPROM will not be detected. If the serial EEPROM is not detected or if the contents are invalid, this message will be displayed.

Now saving stage 1

When the stage button is pressed and held, this message will be displayed in the status line, when the current machine state is saved in stage one.

Now saving stage 2

When the stage button is pressed and held, this message will be displayed in the status line, when the current machine state is saved in stage two.
Status line messages
page 48
Syntec
VitMan
ãã
ã
ãã
Page 50
Status Line Messages

Now saving stage 3

When the stage button is pressed and held, this message will be displayed in the status line, when the current machine state is saved in stage three.

Now saving stage 4

When the stage button is pressed and held, this message will be displayed in the status line, when the current machine state is saved in stage four.

Now saving stage 5

When the stage button is pressed and held, this message will be displayed in the status line, when the current machine state is saved in stage five.

Now saving stage 6

When the stage button is pressed and held, this message will be displayed in the status line, when the current machine state is saved in stage six.

Now saving stage 7

When the stage button is pressed and held, this message will be displayed in the status line, when the current machine state is saved in stage seven.

Over temperature

If the internal device temperature is 53 °C or higher, this message is displayed in the status line. This message is only a warning and no surgical functions are effected.
The temperature sensor monitors the internal temperature of the system and resides on the analog board. The temperature sensor is a serial device that provides a properly for­matted response to a number of commands. If proper communication is not established, any values read are not used. If the temperature sensor has a problem and reports a high temperature condition, this message may be displayed. Under normal operating conditions, the internal temperature of the system won’t reach this limit. Check for airflow restrictions around the system. Especially near the main system fan and the air duct at the rear of the system.

Overpressure detected

If the current eye pressure exceeds the warning pressure limit, this message will be dis­played in the status line. The warning pressure level may be adjusted using the “Air Ex­change Values” menu in the “Current Settings” menu.

Pedal fuse problem

The foot pedal fuse on the CPU board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
This message indicates that the foot pedal power is not present. The foot pedal power is a fused version the CPU boards +5 volt power.
ãã
ã
Syntec
VitMan
ãã
page 49
Status line messages
Page 51
Status Line Messages

Please lock cassette

If the aspiration cassette is present and the vacuum level is below acceptable limits, this message will be displayed in the status line. This message may appear along with the “Low vacuum level” and “Check cassette tubing” messages.
This message will be displayed if the 500cc vacuum level is below 50 mmHg, while the cassette is present. A leak in the vacuum system, a problem with the 500cc vacuum sen­sor or a problem with the A/D converter, may also cause this message.

Please tune handpiece

If the ultrasound surgical function is on and the handpiece is not in tune, this message will be displayed in the status line. If the handpiece is removed and replaced, it must be tuned. If the power detection logic determines that the handpiece is out of tune, the probe should be retuned.

Posterior Aspiration

This message indicates that the aspiration surgical function has entered posterior aspira­tion mode, using the aspiration on/off button.

Posterior Frag

This message indicates that the ultrasound surgical function has entered posterior frag mode, using the ultrasound on/off button.

Posterior Linear Rate

This message indicates that the vitrector surgical function has entered posterior linear cut rate mode, using the vitrector on/off button.

Posterior Vitrectomy

This message indicates that the vitrector surgical function has entered posterior vitrectomy mode, using the vitrector on/off button.

Power fail detected

If the main power supply detects a problem with the mains input, this message will be displayed in the status line. When this condition is detected, disk activity is inhibited for five seconds.
The system power supply provides an AC LOW indication. The AC LOW indication is used to generate a non-maskable interrupt (NMI) on the CPU.

Pressure pump hours

The run time on the main pressure pump has exceeded the maximum limit. The pneumat­ics module needs to be serviced. The “Display Pump Hours” command, found in the “Ser­vice Pneumatics” menu of the “Service Menu”, can be used to display the current pressure pump hours.
Status line messages
page 50
Syntec
VitMan
ãã
ã
ãã
Page 52
Status Line Messages
If the pressure pump hours exceed 2000, this message will be displayed. While the vitrec­tor surgical function is on this message is disabled.

Prime cycle aborted

This message indicates that the aspiration prime cycle was aborted. Any foot pedal activ­ity, or pressing the aspiration valve button, will abort a prime cycle.

Prime level xxx mmHg

While a prime cycle is in progress, the aspiration up and down buttons can be used to change the aspiration level. This message will be displayed when a change is made to the prime aspiration level. The aspiration prime level is “xxx”.

Re-position cassette

If the cassette level detection logic determines that the aspiration cassette is not properly aligned, this message will be displayed in the status line. This generally indicates that the cassette is pushed into the cassette chamber prior to rotating the knob into the lock posi­tion. To correct the problem, rotate the knob to the unlocked position then back to the locked position without pressing on the cassette. The “Cassette position bad” and “Use locking knob only” messages will also be displayed.

Replace cassette

If a high fluid level is detected in the small chamber of the aspiration cassette, this mes­sage will be displayed in the status line. If this condition is suspected to have been de­tected in error, remove the aspiration cassette, wait for one second, and reinsert the aspira­tion cassette. If the condition persists, replace the aspiration cassette. The “Cassette problem” message will also be displayed.

Reverse flow detected

If the air exchange surgical function is on and the pressure on the air exchange port ex­ceeds the desired pressure level then this message will be displayed in the status line. This generally indicates that the three way valve is incorrectly positioned to apply the saline solution to the air exchange port rather than to the infusion cannula. When this condition is detected, the air exchange valve is closed in an attempt to reduce fluid flow into the VitMan.

Service pneumatics

If the time on the pneumatics module has exceeded the maximum limit, when the VitMan is powered up this message will be displayed in the status line. The “Display Pump Hours” command, found in the “Service Pneumatics” menu of the “Service Menu”, can be used to display the current pressure pump hours.
If the pressure pump hours exceed 1000, this message will be displayed.
ãã
ã
Syntec
VitMan
ãã
page 51
Status line messages
Page 53
Status Line Messages

Service required

This message indicates that a serious problem was detected and repair of the Syntec VitMan is required. The message will be accompanied by other status line messages that will indicate the problem area.
Air exchange problem - If the air exchange valve is open, the pressure difference between the eye pressure sensor and the air exchange pump pressure sensor should be less than 15 mmHg (for low flow rate conditions the pressure difference will be close to zero mmHg). If a pressure difference of 15 mmHg or greater is detected, this message will be displayed.
Air pump problem - At power-up, a test is made of the ability of the air exchange pump to generate pressure. If pressure was not detected, this message will be displayed when the air exchange surgical function is enabled.

Shutdown at 55°C

If the internal device temperature is 53 °C or higher, but less than 55 °C, this message is displayed in the status line.
The temperature sensor monitors the internal temperature of the system and resides on the analog board. The temperature sensor is a serial device that provides a properly for­matted response to a number of commands. If proper communication is not established, any values read are not used. If the temperature sensor has a problem and reports a high temperature condition, this message may be displayed. Under normal operating conditions, the internal temperature of the system won’t reach this limit. Check for airflow restrictions around the system. Especially near the main system fan and the air duct at the rear of the system.

Stabilizing...

This message indicates that a hardware problem was detected while waiting for the pres­sure sensors to thermally stabilize. If this message is present, the VitMan will not function. If this message should appear, power cycle the device.

Stage 1 selected

When the stage button is pressed and held, if stage one will be entered when the stage button is released the message “Hold to enter stage 1” will be displayed. Releasing the stage button will cause stage one to be entered and this message to be displayed in the status line.

Stage 2 selected

When the stage button is pressed and held, if stage two will be entered when the stage button is released the message “Hold to enter stage 2” will be displayed. Releasing the stage button will cause stage two to be entered and this message to be displayed in the status line.
Status line messages
page 52
Syntec
VitMan
ãã
ã
ãã
Page 54
Status Line Messages

Stage 3 selected

When the stage button is pressed and held, if stage three will be entered when the stage button is released the message “Hold to enter stage 3” will be displayed. Releasing the stage button will cause stage three to be entered and this message to be displayed in the status line.

Stage 4 selected

When the stage button is pressed and held, if stage four will be entered when the stage button is released the message “Hold to enter stage 4” will be displayed. Releasing the stage button will cause stage four to be entered and this message to be displayed in the status line.

Stage 5 selected

When the stage button is pressed and held, if stage five will be entered when the stage button is released the message “Hold to enter stage 5” will be displayed. Releasing the stage button will cause stage five to be entered and this message to be displayed in the status line.

Stage 6 selected

When the stage button is pressed and held, if stage six will be entered when the stage button is released the message “Hold to enter stage 6” will be displayed. Releasing the stage button will cause stage six to be entered and this message to be displayed in the status line.

Stage 7 selected

When the stage button is pressed and held, if stage seven will be entered when the stage button is released the message “Hold to enter stage 7” will be displayed. Releasing the stage button will cause stage seven to be entered and this message to be displayed in the status line.

Temp sensor problem

If the temperature sensor is undetectable, this message is displayed in the status line. Powering the VitMan off and then on may clear the problem.
The temperature sensor monitors the internal temperature of the system and resides on the analog board. The temperature sensor is a serial device that provides a properly for­matted response to a number of commands. If proper communication is not established, any values read are not used. If the temperature sensor has a problem, this message will be displayed.

Threshold is xx mmHg

The aspiration threshold allows the user to specify the vacuum level required in the 50cc chamber, before the aspiration valve is opened. If aspiration is prevented by the current threshold setting, this message is displayed in the status line, where “xx” is the current threshold value. The aspiration threshold may be adjusted using the “Aspiration Threshold” command, found in the “Aspiration Values” menu of the “Current Settings” menu.
Syntec
VitMan
ãã
ã
ãã
page 53
Status line messages
Page 55
Status Line Messages

Tuning aborted

If the ultrasound handpiece tuning cycle is terminated by turning off the ultrasound surgical function or by foot pedal activity, this message is displayed in the status line.

Tuning handpiece

While the ultrasound handpiece is being tuned, this message will be displayed in the status line.

Tuning not allowed

If aspiration is off or in alarm, tuning of the ultrasound handpiece is not allowed. If an attempt to tune the ultrasound handpiece is made, this message will be displayed in the status line.

Tuning unsuccessful

If the ultrasound handpiece tuning cycle is unsuccessful, this message will be displayed in the status line.

Ultrasound fuse open

The ultrasound fuse on the analog board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
This message indicates that the +24 volt power is not present on the ultrasound electron­ics. In addition to the fuse opening, this message could be caused by a problem with the +24 volt output of the power supply.

Urgent system problem

If a communications problem is detected between the CPU board and the analog board, this message will be displayed in the status line. This is a serious problem that requires attention before use. Powering the VitMan off and then on may clear the problem.
The A/D converter, present on the analog board, provides several constant values that are used to determine the quality of information and integrity of the communications path. If these values are found to be in error, this message will be displayed. The A/D converter is the primary source of information on the status of the system. The quality of the informa­tion provided is vital to the proper operation of the system.

Use locking knob only

If the cassette level detection logic determines that the aspiration cassette is not properly aligned, this message will be displayed in the status line. This generally indicates that the cassette is pushed into the cassette chamber prior to rotating the knob into the lock posi­tion. To correct the problem, rotate the knob to the unlocked position then back to the locked position without pressing on the cassette. The “Cassette position bad” and “Re­position cassette” messages will also be displayed.
Status line messages
page 54
Syntec
VitMan
ãã
ã
ãã
Page 56
Status Line Messages

Vacuum fuse problem

The vacuum pump fuse on the analog board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
This message indicates that the vacuum pump power is not present. In addition to the fuse opening, this message could be caused by a problem with the +12 volt power.

Vacuum pump hours

The run time on the vacuum pump has exceeded the maximum limit. The pneumatics module needs to be serviced. The “Display Pump Hours” command, found in the “Service Pneumatics” menu of the “Service Menu”, can be used to display the current vacuum pump hours.

Valve fuse problem

The valve fuse on the analog board has opened. This fuse is resettable, powering the VitMan off and then on may clear the problem.
Syntec
VitMan
ãã
ã
ãã
page 55
Status line messages
Page 57

System Test Procedures

page 56
Syntec
VitMan
ãã
ã
ãã
Page 58
System Test Procedures

System Operation Check

The
Syntec
functions of a vitrectomy system. This procedure confirms that the system is assembled correctly, and that no major system malfunctions exist. It does not verify calibration of the internal pressure sensors or of internal operating pressures. Various circumstances require verification of correct system operation:
In order to evaluate the system, certain items are required:
VitMan includes various subsystems that operate together to provide the
System installation Complaint that system isn’t operating properly After maintenance on the system After upgrade of system hardware or software
VitMan system Power cord System foot pedal System owner’s manual Aspiration cassette Vitrector Air Exchange tubing w/ air exchange filter Infusion cannula Endoilluminator Ultrasound handpiece (if appropriate) Ultrasound needle (need not be new) Beaker or bowl of water, at least 350cc
An assumption is made in this procedure that the operator of the system, is familiar with system operation and setup. Refer to the owner’s manual for information on how to operate the system. Operation of the system’s menus is required to verify system operation. The owner’s manual includes comprehensive information on the operation of the menus. This procedure will refer to a specific series of menu selections separated by a \. For example Values \ Current A/D Values signifies selecting Values from the Main menu, then selecting Current A/D Values from the Values menu. A complete listing of the menu selections will be given from the main menu, even if a shorter route from the current screen could be taken in the menu tree. This allows the operator to access the correct menu screen without follow­ing this procedure step by step. Perform the following steps. If the system does not react as described, refer to the troubleshooting procedure for a remedy.
1. Connect the power cord and foot pedal to the system. Turn system on.
2. Install the aspiration cassette. Connect the vitrector (both pressure and aspiration tubes, route the asp. tube correctly), endoilluminator, air exchange tubing, and infu­sion cannula..
3. Check the system light source. a. Ensure that all other functions are off. b. Remove both lamps from the light source. Close the light source drawer. Don’t
touch the lamps with fingers. Use a clean cloth or gloves. Refer to the light
Syntec
VitMan
ãã
ã
ãã
page 57
System operation check
Page 59
System Test Procedures
source lamp replacement procedure for more information. c. Turn the light source dimming knob fully clockwise. Turn the light source on. The
system display should indicate that both bulbs have failed. Turn the light source off.
d. Install a lamp in the “B” socket (right socket). Close the light source drawer. Turn
the light source on. Light should radiate from the endoilluminator after a few seconds for the bulb warm-up. Turn the light source off. The system display should indicate that the “A” bulb has failed.
e. Wait approximately one minute for the lamp to cool. If the light source was turned
on for more than just a few seconds it will take longer for it to cool. Remove the lamp from the “B” socket and install it in the “A” socket . (Note that the system will automatically move the lamp sockets into the “A” position when the light source drawer is opened. There could be a delay of a couple of seconds for this to happen after the drawer is opened.) Close the light source drawer. Turn the light source on. Light should radiate from the endoilluminator after a few seconds for the bulb warm-up. Turn the light source off. The system display should indi­cate that the “B” bulb has failed.
f. Wait approximately one minute for the lamp to cool. Install the other lamp into
the “B” socket. Close the light source drawer.
g. Turn the light source on. The system display should indicate that the “A” bulb is
on. It should not indicate that either lamp has failed. Observe the light radiating from the endoilluminator by projecting onto a surface from approximately 5cm. It should appear very bright white. Turn the dimming knob counterclockwise. The endoilluminator output should start to dim after some rotation of the knob. The output should stay substantially unchanged, except for intensity, through most of the rotation of the dimmer knob. Repeat this test for the other two connector outputs. (Don’t forget to turn the dimmer back up.)
h. Turn the light source off. Press and hold the light source on/off button (This will
initiate a change in the illuminated lamp from A to B. Release the button after the light source output has started to radiate light then has turned off again.). Ob­serve the light radiating from the endoilluminator by projecting onto a surface from approximately 5cm. It should appear very bright white. Turn the dimming knob counterclockwise. The endoilluminator output should start to dim after some rotation of the knob. The output should stay substantially unchanged, except for intensity, through most of the rotation of the dimmer knob. Repeat this test for the other two connector outputs.
i. Turn the light source off.
4. Check aspiration function. Note that whenever an aspiration cassette is used to aspi­rate fluids (especially a cassette that may have been abused, as is usually the case with test components), it is advisable to ensure that the float valve in the smaller chamber is operating correctly. Fluid should start transferring from the smaller cham­ber to the larger chamber when the fluid level reaches approximately half way up the float. If the fluid goes more than 3/4 up the float, then the cassette should be consid­ered faulty. Use of a faulty cassette could result in aspiration of fluids into the system. a. Immerse the vitrector tip in water. b. Turn aspiration function on. Enter anterior aspiration mode. The status LED
System operation check
page 58
Syntec
VitMan
ãã
ã
ãã
Page 60
System Test Procedures
should turn green. c. Adjust the maximum aspiration to 500 (or the system max). d. Enter into the system menu, go to Values \ Current A/D Values. e. The 500cc value should read greater than 550. The 50cc value should read 5
maximum.
f. Depress the foot pedal until the Pedal value is slightly more than the activity
threshold. The activity threshold default is 10%. It can be found under Current Settings \ Pedal Thresholds \ Activity Threshold. The irrigation pinch valve should open. Release the foot pedal. The irrigation pinch valve should close.
g. Depress the foot pedal until the Pedal value is slightly more than the irrigate to
aspirate threshold. The irrigate to aspirate threshold default is 10%. It can be found under Current Settings \ Pedal Thresholds \ Irrigate To Aspirate. The aspiration pinch valve should open. Release the foot pedal. The aspiration pinch valve should close. Note that if the activity threshold is at 10% and the irrigate to aspirate threshold is at 10% then the aspiration pinch valve should open after 20%. To calculate the position that the aspiration pinch valve opens, the two thresholds are added together in anterior surgical modes . Note also that the aspiration threshold setting under Current Settings \ Aspiration Values can also affect the opening of the aspiration pinch valve. The default is 0 mmHg. The aspiration pinch valve will not open until the vacuum level in the 50cc (smaller) chamber of the cassette has reached the aspiration threshold setting, regardless of threshold adjustments for the foot pedal.
h. Rotate the foot pedal to activate reflux. The direction (left or right) for reflux is
determined under Current Settings \ Aspiration Values. The reflux valve should respond based on the setting under Current Settings \ Aspiration Values. Continuous Reflux mode causes the reflux valve to remain closed for the entire time that the foot pedal is rotated. Timed Reflux causes the reflux valve to close (after the foot pedal is rotated) for the time determined under Current Settings \ Aspiration Values \ Reflux Hold Time.
i. Rapidly depress the foot pedal fully. The 50cc value should quickly (less than 1
second) reach the maximum aspiration adjustment. It is normal for the value to over shoot up to 10 on very high maximum aspiration settings. The value should stabilize to within 2 of the maximum aspiration setting in a couple of seconds. The aspiration function should not go into alarm, as would be indicated by the status LED turning red.
j. Adjust the maximum aspiration setting to 250 (or 1/2 of system max). Repeat
step 4.i.
k. Adjust the maximum aspiration setting to 50 (or 1/10 of system max). Repeat
step 4.i. The value over shoot should be less than 5. l. Return to the system operating display (By pushing the display button). m. Adjust the maximum aspiration setting to 250 (or 1/2 of system max). Aspirate
fluid into the aspiration cassette. If the cassette icon is enabled in the Display
menu, then it should graphically indicate an increase in fluid by “filling up” as the
fluid level raises in the larger chamber of the aspiration cassette. At approxi-
mately 250cc of fluid in the larger chamber of the aspiration cassette an alarm
should sound (If the speaker is enabled under the Audio Services menu.), and
Syntec
VitMan
ãã
ã
ãã
page 59
System operation check
Page 61
System Test Procedures
a status line message should indicate High cassette level . At approximately 300cc of
fluid in the larger chamber of the aspiration cassette an alarm should sound, and a status line message should indicate Cassette full . At approximately 325cc of fluid in the larger chamber of the aspiration cassette an alarm should sound, and a status line message should indicate Cassette over full. Aspiration should fail to function any further until the over full cassette condition is removed.
n. Remove and empty fluid from the aspiration cassette. Replace the aspiration
cassette.
5. Check vitrector function. a. Turn the vitrector function on. Turn the aspiration function off. b. Adjust the cut rate to 1000 cpm c. Fully depress the foot pedal. Rotate the foot pedal to activate the vitrector. The
direction (left or right) for cutting is determined under Current Settings \ Aspira- tion Values, and is opposite the side reflux is set to. For example if reflux is on the left then vitrector will be on the right. Refer to the owner’s manual under the description of the menu item Current Settings \ Vitrector Values for more information on vitrector function options. The system should be able to drive the vitrector indefinitely at 1000 cpm without going into alarm (the status LED turns red if the function goes into alarm). Observation of the cutting port on the vitrec­tor should indicate that the internal needle is fully closing off the port as the vitrector operates.
d. Turn the vitrector function off.
6. Check ultrasound function (if the ultrasound option is present). a. Turn the ultrasound function on. The status LED should turn red. The display
status line should indicate No handpiece present.
b. Connect the ultrasound handpiece. Install the ultrasound needle. The status LED
should remain red. The display status line should now indicate Please tune
handpiece. c. Tune the handpiece. The tune should be successful. d. Set the ultrasound power to 50%. e. Fully depress the foot pedal. Rotate the foot pedal to activate ultrasound. The
direction (left or right) for cutting is the same as for the vitrector. Power should be
applied to the ultrasound needle. Apply the ultrasound needle tip to a test sur-
face (raw macaroni is acceptable). Noticeable needle activity should be observed
against the test surface. f. Turn the ultrasound function off.
7. Check air exchange function. a. Turn the air exchange function on. b. Adjust the air exchange output to 100. The function should be able to maintain at
least 80 with the infusion cannula open to atmosphere.
c. Adjust the output to 50. The output should stabilize to 50 quickly. Pinch the tube
connecting the infusion cannula. The output should over shoot the setting mo­mentarily, then settle back to 50. Release the tube. The output should under shoot the setting momentarily then settle back to 50.
System operation check
page 60
Syntec
VitMan
ãã
ã
ãã
Page 62
System Test Procedures

System Trouble Shooting Chart

:motpmyS:melborPelbissoP:otrefeR
launaMgnitoohSelbuorTsihtotsreferST
launaMsrenwOnaMtiVcetnySehtotsreferMO
PUrewoPtnoWnepoesufrewopniaMa.2ST
notonhctiwsrewoP2retpahCMO
nideggulptontinU2retpahCMO
drocrogulpevitcefeD2retpahCMO
ylppusrewopniamevitcefeDg.1-a.1ST
)C55sdeecxeerutarepmetlanretni(tohootsitinUc.71-a.71ST
segatloVlanretnIevitcefeDh.3-a.3ST
draobscinortceleUPCevitcefeDd.4-a.4ST
erehttubnursnaF
nielbisivgnihtonsi
yalpsideht
)eruliafretrevnocCD
krad/thgilootdetsujdatsartnoC3retpahCMO
ehtneewtebnekorb/gnissimssenrahrewoP
UPCehtdnadraobscinortcelegolanA
)CDV22-(nepo011FesuFc.3ST
-CDCDV22-(draobscinortcelegolanAevitcefeD
.eruliafylppusrewopdraobscinortcelegolanA
)gnissimCDV21ro,CDV21+,CDV5+(
draobscinortceleUPCnoyalpsidDCLevitcefeDa.4ST
draobscinortceleUPCevitcefeDb.4-a.4ST
ST
h.3-a.3ST
b.3ST
Syntec
OGOLcetnyS
yalpsidniyltnatsnoc
sutatsDERruofllA
nosDEL
rosserpmocniaM
tratstonseod
ãã
ã
VitMan
ãã
nekorb/ylluf
nekorb/ylluf
nepoesufrosserpmocniaMb.7-a.7ST
page 61
detaeston/gnissimelbaclortnocgolanAUPC
)X-CDV5+(gnissimecnereferD/Aa.5ST
)evitcefedD/A(draobscinortcelegolanAevitcefeDa.6ST
)MARdekcabyrettabevitcefeD(UPCevitcefeD
detaeston/gnissimelbaclortnocgolanAUPC
)X-CDV5+(gnissimecnereferD/Aa.5ST
)evitcefedD/A(draobscinortcelegolanAevitcefeDa.6ST
)MARdekcabyrettabevitcefeD(UPCevitcefeD
Symptom troubleshooting chart
b.4-a.4ST
onsierehT
.sihtrofkcehc
b.4-a.4ST
onsierehT
.sihtrofkcehc
Page 63
System Test Procedures
:motpmyS:melborPelbissoP:otrefeR
/nideggulptoneludomcitamuenPotelbacrewoP
nekorb/yllufdetaeston
ssenrahrewopCDmetsysevitcefeDc.7ST
TEFevird(draobscinortcelegolanAevitcefeD
)nepo
elbacnoitacinummocUPCgolanAevitcefeDa.6ST
rosserpmocevitcefeDd.7ST
e.7ST
d.7ST
pmuPmuucaV
tratstonseod
llufettessaCb.8ST
nekorb/yllufdetaeston
)nepo
pmupevitcefeDf.8ST
seodecruosthgiL
srebifetanimulliton
midyllufotdenrutlortnocgnimmiD41erugiFMO
srebifelbitapmocnI95erugiFMO
gnissim/tuodenruberasblubhtoBk.9&a.9ST
yltcerrocdellatsniton/gnissimettessaCb.8-a.8ST
/nideggulptoneludomcitamuenPotelbacrewoP
g.8-d.8ST
)neercssutatsaivkcehc(nepo101FesuFMO
TEFevird(draobscinortcelegolanAevitcefeD
f.8ST
elbacnoitacinummocUPCgolanAevitcefeDh.8&a.6ST
sissahcotnidellatsniylluftonecruosthgiLg.9ST
rotcennocrebifelpirtotnidetresniylluftonsrebiF41erugiFMO
ylppusrewoprociVmorferuliafCDV42+l.9ST
)neercssutatsaivkcehc(nepo601FesuFl.9ST
UPCmorfdeggulpnuelbacdraobyeKc.4ST
)evirdTEF(draobscinortcelegolanAevitcefeDm.9-l.9ST
tucotsliaFrotcertiVdetcelestonedomrotcertiVb.01-a.01ST
Symptom troubleshooting chart
rotcertiVevitcefeD
page 62
)DER(mralanisiedomrotcertiVb.01ST
edomtucelgnisnisirotcertiV
.egnargnittucotnidesserpedro
dehcnip/tcerrocnignibuT
)dehctiwssitucfitfel(thgirotdetatortonladeptooF
31
Syntec
:YMOTCERTIV:YMOTCERTIV
:YMOTCERTIV:YMOTCERTIV
:YMOTCERTIV
&roiretsoP
roiretnA
LADEPTOOFLADEPTOOF
LADEPTOOFLADEPTOOF
LADEPTOOF
SLORTNOC
YMOTCERTIVYMOTCERTIV
YMOTCERTIVYMOTCERTIV
YMOTCERTIV
RETTUC
-11serugiF
VitMan
2retpahCMO
2retpahCMO
4retpahCMO
2retpahCMO
ãã
ã
ãã
Page 64
System Test Procedures
:motpmyS:melborPelbissoP:otrefeR
2retpahCMO
LENAPKCABLENAPKCAB
LENAPKCABLENAPKCAB
evitcefed/detcennoctonladeptooF
.sevlavotrewopon/nepo001FesuFa.51ST
.ladepotrewopon/nepo2FesuFa.11ST
eriwnekorb/kcutsevlavtiVe.01-a.01ST
.nekorb/yllufdetaestonelbaclanretniladeptooFc.11-a.11ST
)evirdTEF(draobscinortcelegolanAevitcefeDe.01ST
)noitalosiladeP(draobscinortceleUPCevitcefeDc.11-a.11ST
ladeP(.elbaclortnocgolanAUPCevitcefeD
)draobscinortcelegolanAgnihcaertonegatlov
LENAPKCAB
&SNOITCENNOC
SLORTNOC
h.8&a.6ST
otsliaFdnuosartlU
tuc
.delbanetonnoitcnufdnuosartlUb.21-a.21ST
)DER(mralanisiedomdnuosartlUb.21ST
oreztaroraentesrewoP
.egnargnittucotnidesserpedro
eceipdnahevitcefeD4retpahCMO
.stibelbanero,D/A
draobscinortceleUPCevitcefeDf.21ST
2noitceSMO
ROIRETSOPROIRETSOP
ROIRETSOPROIRETSOP
ROIRETSOP
:DNUOSARTLU
SNEL
-OITATNEMGARF
N
SLORTNOC
2retpahCMO
LADEPTOOFLADEPTOOF
LADEPTOOFLADEPTOOF
LADEPTOOF
)dehctiwssitucfitfel(thgirotdetatortonladeptooF
elbaclangisdnuosartlU/detaesyllufton/evitcefeDe.21ST
elbaclortnocdnuosartlU/detaesyllufton/evitcefeDf.21ST
elbacrewopdnuosartlU/detaesyllufton/evitcefeDh.21ST
CD-CD(draobscinortcelednuosartlUevitcefeD
).cte,csO,pmA(,)MORPEElaireS(,)retrevnoc
,A/DgniliaF(.draobscinortcelegolanAevitcefeD
h.21ST
h.21-e.21ST
Syntec
VitMan
sliaFnoitaripsAdelbanetonedomnoitaripsA
)DER(mralaniedomnoitaripsAa.31ST
orezottesnoitaripsA
evitcefed/detcennoctonladeptooF
.neercs
yltcerrocdellatsniton/gnissimettessaC
ãã
ã
ãã
page 63
Symptom troubleshooting chart
2retpahCMO
NOITARIPSA
2retpahCMO
ROIRETSOPROIRETSOP
ROIRETSOPROIRETSOP
ROIRETSOP
NOITARIPSA
2retpahCMO
LENAPKCABLENAPKCAB
LENAPKCABLENAPKCAB
LENAPKCAB
&SNOITCENNOC
SLORTNOC
sutatskcehC.ladepotrewopon/nepo2FesuF
a.11ST
2retpahCMO
NOITARIPSANOITARIPSA
NOITARIPSANOITARIPSA
NOITARIPSA
Page 65
System Test Procedures
:motpmyS:melborPelbissoP:otrefeR
deggulp/deknik/detuoryltcerrocnignibuT
egnarnoitaripsaotnidesserpedtonladeptooF
morfpmupmuucavgnitneverpnepo101FesuF
neercssutatskcehC.gninnur
c.7ST
enihcamotlanretnikaelriAb.31ST
eriwnekorb/kcutsrotalugermuucaVh.31ST
rotcennocscitamuenptatneserptonA-CDV21+i.31ST
nekorb/yllufdetaestonelbaclanretniladeptooFc.11-a.11ST
diulfhtiwdellif)s(retlifcibohpordyHg.31ST
.draobscinortcelegolanAevitcefeDi.31ST
elbaclortnoCgolanAUPCevitcefeDh.8&a.6ST
sutatskcehC.sevlavotrewopon/nepo001F
neercs
a.51ST
eriwnekorb/evlavlortnochcnipevitcefeD
evlavhcnipevitcefeD
2retpahCMO
NOITARIPSANOITARIPSA
NOITARIPSANOITARIPSA
NOITARIPSA
2retpahCMO
LADEPTOOFLADEPTOOF
LADEPTOOFLADEPTOOF
LADEPTOOF
SLORTNOC
f.31-e.31STf.31-e.31ST
f.31-e.31STf.31-e.31ST
f.31-e.31ST
sliaFegnahcxEriAdelbanetonnoitcnuFegnahcxEriA
2retpahCMO
EGNAHCXERIAEGNAHCXERIA
EGNAHCXERIAEGNAHCXERIA
EGNAHCXERIA
)DER(mralanisinoitcnufegnahcxEriAa.41ST
orezottesegnahcxEriA
deknikgnibuT
riagnikcolbevlavpotsyaw-3
2retpahCMO
EGNAHCXERIAEGNAHCXERIA
EGNAHCXERIAEGNAHCXERIA
EGNAHCXERIA
2retpahCMO
EGNAHCXERIAEGNAHCXERIA
EGNAHCXERIAEGNAHCXERIA
EGNAHCXERIA
2retpahCMO
EGNAHCXERIAEGNAHCXERIA
EGNAHCXERIAEGNAHCXERIA
EGNAHCXERIA
neercssutatskcehC.deneposah301FesuFe.41ST
eriwnekorb/rosserpmocegnahcxeriAevitcefeDb.41ST
sutatskcehC.sevlavotrewopon/nepo001F
neercs
a.51ST
eriwnekorb/evlavlortnochcnipevitcefeDc.41ST
enihcamotlanretnikaelriAe.41ST
.draobscinortcelegolanAevitcefeDb.41ST
sliaFksiD).eerfrorreebTSUM(.etteksidevitcefeDe.61-d.61ST
detcetorpetirWc.61ST
sutatskcehC)yppolfotCDV5+(.nepo211FesuF
neercs
neewtebelbaclortnoc/detaesyllufton/evitcefeD
draobscinortceleUPCdnayppolf
b.61ST f.61ST
elbacrewop/detaesyllufton/evitcefeDf.61ST
scinortcelegolanAnorotalugerCDV5+evitcefeD
draob
a.61ST
evirDyppolFevitcefeDf.61ST
Symptom troubleshooting chart
page 64
Syntec
VitMan
ãã
ã
ãã
Page 66
System Test Procedures
:motpmyS:melborPelbissoP:otrefeR
UPCevitcefeDf.61ST
2retpahCMO
LENAPKCABLENAPKCAB
LENAPKCABLENAPKCAB
sliaFladePtinuotnideggulptonladePtooF
sutatskcehC).ladepotCDV5+(.nepo2FesuF
.neercs
elbac/ladeptoofevitcefeDc.11-a.11ST
toHooT c.71-a.71ST
ysioN e.81-a.81ST
LENAPKCAB
&SNOITCENNOC
SLORTNOC
a.11ST
Syntec
VitMan
ãã
ã
ãã
page 65
Symptom troubleshooting chart
Page 67
System Test Procedures

Trouble Shooting Steps

1. Main Power supply checkout
petSerudecorPelanoitaRsetoN
ytirojamehtrofdesusiylppusCDV21+ehT
CDehT.naMtiVehtnisnoitcnufehtfo
ehtmorfyltceridCDV21+ehtsekatssenrah
potehtevomeR
.srevocmottobdna stinotinuehtecalP
ehthtiw(edistfel
)hgihecruosthgil
a.1
b.1
c.1
golanAeht
otretemtlovaesU CDV21+ehtyfirev
notneserpegatlov
ehtfotuptuoeht
ylppusrewopniam
tinuehthtiw
.puderewop
siCDV21+ehtfI
,CDV5.11+woleb
niamehttcennocsid
morfssenrahCD
.draobscinortcele
ehterusaeM
.niagaCDV21+
ehterusaeM
egatlovCDV42+
ehtnotneserp
niamehtfotuptuo
.ylppusrewop
.scinortcele
.tuptuoevitcefed
.ylppusrewop
.sesufetarapes
.tuptuoevitcefed
golanAehtotylppusrewopniam
aivdetubirtsidsitierehwdraobscinortcele
dna,spmup,snafehtotsesufetarapes
yrt,noitacificepsfotuosiCDV21+ehtfI
ehtmorfssenrahCDehtgnitcennocsid
wollalliwsihT.draobscinortcelegolanA
narehtieotmelborpehtfonoitalosi
aroylppusrewopehtnodaolevissecxe
ehtkcehc,tnesbasiCDV21+ehtfI
sioottifI.tneserpsitifieesotCDV42+
deliafsahrehtieylppusrewopeht,tnesba
.tupniCAgniviecertonsiro
gniylppusfoelbapacsiylppusrewopehtfI
ssenrahCDehthtiwCDV21+tcerroca
ylppusrewopniamehtrehtie,detcennocsid
CDehtnitrohsasierehtrodeliafsah
lavomeRylppuSrewoPehtesU.ssenrah
CDehtotsseccaniagoterudecorp
egamadrofylhguorohttikcehC.ssenrah
niamehtrossenrahehtecalperrehtiedna
dnaW002tadetarsiylppusCDV42+ehT
sitI.naMtiVehtnisnoitcnufruofrofdesusi
,)W051(noitanimullirofecruosrewopeht
dna,)W3(egnahcxeria,)W54(dnuosartlu
ssenrahCDehT.)W1(rekaepslanretnieht
golanAehtotyltceridCDV42+ehtsekat
aivdetubirtsidsitierehwdraobscinortcele
yrt,noitacificepsfotuosiCDV42+ehtfI
dnuosartlUgolanAehtgnitcennocsid
scinortcelegolanAehtmorfelbacrewop
lliwsihT.ssenrahCDehtro/dnadraob
narehtieotmelborpehtfonoitalosiwolla
aroylppusrewopehtnodaolevissecxe
CDV21+ehT
erusaemdluohs
5.11neewteb .CDV5.21dna
CDV42+ehT
erusaemdluohs
dna32neewteb
.CDV52
Troubleshooting steps
page 66
Syntec
VitMan
ãã
ã
ãã
Page 68
System Test Procedures
petSerudecorPelanoitaRsetoN
siCDV42+ehtfI ,CDV32+woleb
niamehttcennocsid
d.1
morfssenrahCD
golanAeht
.draobscinortcele
ehterusaeM
.niagaCDV42+
.ylppusrewop
tupnitahtyfireV
tatneserpsirewop
slanimrettuptuoeht
yrtnerewopehtfo
ylluferacybeludom
raerehtgnisaeler
e.1
erusaemdnalenap
ssorcaegatloveht
yrtnerewopeht
ehthtiwslanimret
sihT.nodenruttinu
dluohsegatlov
.tupnienilehthctam
petssihtmrofreP
onsierehtfiYLNO
ondnaCDV21+
ehtmorfCDV42+
rewopmetsys
CDehthtiwylppus
ssenrahrewop
morfdetcennocsid
golanAeht
.draobscinortcele
f.1
CAnaesU
yfirevotretemtlov
yltcerrocsiCAtaht
ehtotdetcennoc
sihT.ylppusrewop
.krowotdeificeps
ehttahteriuqerlliw
ylppusrewop
ebylbmessa
.devomer
CDehtotsseccaniagoterudecorp
.retlifenilrewopehtnineporo
gniylppusfoelbapacsiylppusrewopehtfI
ssenrahCDehthtiwCDV42+tcerroca
ylppusrewopniamehtrehtie,detcennocsid
CDehtnitrohsasierehtrodeliafsah
lavomeRylppuSrewoPehtesU.ssenrah
egamadrofylhguorohttikcehC.ssenrah
niamehtrossenrahehtecalperrehtiedna
nisnoitcennoclanretnilareveseraerehT
erehtylekilnusitI.eludomyrtnerewopeht
ehtfokcehcatubnepolanretninaeblliw
lliwslanimrettuptuoehtnotneserpegatlov
.stsixemelborpyrtnerewoponyfirev
htapCAehtniretlifenilrewopasierehT
dnahctiws/esufyrtnerewopehtneewteb
rewopfokcehcehT.ylppusrewopeht
rewopehtfoslanimrettupniehttatneserp
eriwnepoonsierehttahtyfirevlliwylppus
021001rehtieseriuqerylppusrewopehT
CA.zH06/05CAV042002roCAV
tonerasegnaresehtedistuosegatlov
ehttaCAehT
ylppusrewop
slanimrettupni
hctamdluohs
CAnwonkeht
.enil
Syntec
VitMan
ãã
ã
ãã
page 67
Troubleshooting steps
Page 69
System Test Procedures
petSerudecorPelanoitaRsetoN
dnaCDV21+htobfI
eraCDV42+
tahtyfirev,tneserp
erasegatlovlla
golanAehtgnihcaer
g.1
draobscinortcele
gnitcennocsidyb
ssenrahrewopeht
golanAehtmorf
draobscinortcele
ehtgniyfirevdna
.tneserpsegatlov
2. Main fuse check
petSerudecorPelanoitaRsetoN
tonmetsysafoesuacylekiltsomehT
esufehtnepoyrP
a.2
ehtkcehcdnarevoc
.esuf
nometsysehtgninnurnehtdna)A3.6(
esufsihT.esufyrtneniamehteblliwgninnur
rotnerruchsurnievissecxenowolbnac
noitarepoCAV052rofdesuftinuehtgnivah
.)dedeenA01(CAV021
nipehT
sitnemngissa
nidetsil
:AxidneppA
rociVgolanA
.elbaCrewoP
3. Internal DC Voltage Check
petSerudecorPelanoitaRsetoN
potehtevomeR
.srevocmottobdna
ehttcennocsiD
gninnurelbacrewop
ehtneewteb
a.3
ehtdnadnuosartlU
scinortcelegolanA
ylluferaC.draob
llaerusaem
.elbac
yfirevotsegatlov
-CDV21+setarenegretrevnocCD-CDehT
nip ehtnoesurofCDV5+dna,A-CDV21-,A scinortcelednuosartlUdna,golanA,UPC
erasegatlovesehtfoeerhtllA.sdraob
rewopdnuosartlUgolanAehtnotneserp
golanA
rotcennocehT
stnemngissa
nidetsilera
:AxidneppA
dnuosartlU
.elbaCrewoP
.tneserperayeht
ehttcennocsiD
gninnurelbacrewop
UPCehtneewteb
golanAehtdna
.draobscinortcele
b.3
erusaemylluferaC
otCDV22eht
sititahtyfirev
ehtnotneserp
siylppussihT.knalbeblliwyalpsid
ehtsetaercretrevnocCD-CDlanoitiddanA
ehtotsaibedivorpotdedeenCDV22
eht,gnissimsiylppussihtfI.yalpsidDCL
ehtfo7nipaivUPCehtotdetcennoc
tneserptonsiCDV22ehtfI.elbacrewop
.evitcefedsidraobscinortcelegolanAeht
scinortcelegolanA
.lanimretdraob
Troubleshooting steps
page 68
Syntec
VitMan
ãã
ã
ãã
Page 70
System Test Procedures
petSerudecorPelanoitaRsetoN
ehttcennoceR
dnaelbacrewop
llaehtyfirev
erasegatlov
UPCehttatneserp
.draobscinortcele
ehterusaeM
ehtnosegatlov
ehtfoedisetisoppo
c.3
d.3
e.3
f.3
ehtmorfUPC
roF.rotcennoc
seriweht,ecnerefer
ehtevahsroloc
:gninaemgniwollof
dnuorG:kcalB
CDV5+:deR
CDV21+:teloiV
CDV21-:yerG
CDV22-:wolleY
siCDV22ehtfI
ylluferac,gnissim
rewopehtevomer
rotcennocssenrah
UPCehtmorf
draobscinortcele
erusaemdna
wolleyehtneewteb
.seriwkcalbdna
ehtrofkcehC
foecneserp
ehtnoCDV21+
.311Ffosedishtob
.draob
siCDV21+fI
yamereht,gnissim
ehtnimelborpaeb
.ssenrahrewop
ylekilsisegatlovehtfoenofoeruliaF
rewopehtnieriwnekorbaoteud
erewsesufylppusehtecnisssenrah
ehtfI.a.3petsnidekcehcyltceridni
sawtinehwgnissimwonsiCDV22
arehtiesiereht,b.3petsnitneserp
roeriwwolleynekorbahtiwmelborp gnidaolsidraobscinortceleUPCeht
.ylppusehtnwod
ehtfienimreteddluohskcehcsihT
ehtfodneUPCehttastsixeCDV22
eht,tneserpsiCDV22ehtfI.elbac
evitcefedsiUPC
ro,A-CDV21-,A-CDV21+ehtfI
aelbissopsitignissimeraCDV5+
otretemtlovaesU.nwolbsahesuf
noCDV21+foecneserpehtrofkcehc
retrevnocCD-CDehtfoedisecruoseht
dluowesufsihtnoneponA.311Fesuf
scinortcelegolanAevitcefedaetacidni
rewopniamehtmorfnekatsiCDV21+
golanAehtotdetuordnaylppus
etarapes3gnoladraobscinortcele
niamehtotdeilppussiCDV21+.shtap
snipaivpmupmuucavdnarosserpmoc
.rotcennoc401PJehtfo8&7
ehtotecruosasadeilppussiCDV21+
ehT.2nip401PJaivsevlavlortnoc
ehtgnidulcni,sdaolCDV21+gniniamer
rofrewopsevirdtahtretrevnocCD-CD
.1nip401PJaivdeilppusera,UPCeht
golanA
dnuosartlU
.devomer
ybdevloser
golanAeht
.evitcefed
ehtsiCDV22ehT
hcihwegatlovylno
detsettonsaw
ehtnotneserp
ehtnehwrotcennoc
UPCgolanA
.tneserpsawelbac
ehtno011FesuF
scinortcelegolanA
sinepolliwdraob
notrohsasiereht
ehtnoCDV22eht
scinortceleUPC
siesufsihT.draob
lliwdnaelbatteser
sirewopnehwteser ytilibissopdnocesA
nieruliafaebdluoc
ylppusCDV22eht
golanAehtno
.draobscinortcele
tonsimelborpehtfI
UPCehtgnignahc
,draobscinortcele
sidraobscinortcele
Syntec
VitMan
ãã
ã
ãã
page 69
Troubleshooting steps
Page 71
System Test Procedures
petSerudecorPelanoitaRsetoN
ehtrofkcehC
foecneserp
-,A-CDV21+
g.3
dnaA-CDV21
htobnoCDV5+
sesuffosedis
dna,801F,701F
111F
.detcennocsid
CDV21+rofkcehC
h.3
htobnotneserp
501Fesuffosedis
ehtnosesufeerhtoslaeraerehT
.retrevnocCD-CDehtfoedistuptuo dluowsesufesehtfoenonineponA
golanAehthtiwmelborpaetacidni
ehthtobnehwdraobscinortcele
dnaelbacrewopdnuosartlU-golanA
eraelbacrewopUPCgolanAeht
snurretrevnocCDCDCDV22ehT
aivylppusniamCDV21+ehtmorf
rewopseilppusosla501F.501Fesuf
rotalugerraenilksidyppolfehtot
-otpOeerhtehtsevirddna)451U( follA.301OSI101OSIsrotalosi
golanAehtnoerasdaoleseht
ehtnieruliafA.draobscinortcele
aetacidnidluowniamCDV21+
.draobscinortcelegolanAevitcefed
4. CPU electronics board check
petSerudecorPelanoitaRsetoN
yameruliafehT
ehtnoebyllautca
potehtevomeR
.srevocmottobdna
rewopehtkcehC
a.4
UPCehtgniretne
golanAehtaiv
.elbacrewopUPC
ehtsedulcnisihT
.CDV22
.evitcefed
yameriwelbacrewoptnettimretninA
.UPCgninoitcnufnonanitirplucehteb
snipUPCehttarewopehtfokcehcA
sirewopehttahteetnarauglliw
nisretcarahconeraerehtfI.tneserp
,CDV5+elihwyalpsidDCLeht
,tneserperaCDV22dnaA-CDV21+
sidraobscinortceleUPCeht
eludomDCL
UPCehtotdehcatta
.draobscinortcele
ehtgnitsujdayrtoslA
ybtsartnoc
thgirehtgnisserped
ekamotnottubeulb
rorethgilyalpsideht otnottubeulbtfeleht
yalpsidehtekam
.rekrad
Troubleshooting steps
page 70
Syntec
VitMan
ãã
ã
ãã
Page 72
System Test Procedures
petSerudecorPelanoitaRsetoN
ehtpu-rewopnO
rossecorporcim
llaffonrutdluohs
llufrofkcehC
egamad/gnitaes
b.4
UPCehtno
lortnocgolanA
.elbac
).yalpsid
yalpsidehtnosiogoLcetnySehtfI
,DERerasDELruofehtdnayltnatsnoc
nimelborpaebyamereht
dnaUPCehtneewtebsnoitacinummoc
UPCehT.sdraobscinortcelegolanAeht
ehtstcennocelbaclangisgolanA
snipnoUPCehtotretrevnocD/AgolanA
tonnacUPCehtfI.05dna,84,64,44,24
golanAehtnoD/Aehthtiwetacinummoc
ehtniniamerlliwti,draobscinortcele
,tcatnisielbacehtfI.etatsogoLcetnyS
ehtroUPCehtnoebnacmelborpeht
golanAehT.sdraobscinortcelegolanA
evitcefedaevahdluocdraobscinortcele
sseccaotgniliafebdluocUPCehtroD/A
ehtemussa,esacsihtnI.yltcerrocD/Aeht
.evitcefedsidraobscinortcelegolanA
ylriafgnikrowebtsumUPCehtecniS(
ehtotnoogolcetnySehtdaolotllew
elihwogol
dabasah
).draob
lenaptnorfruof
nehtdnasDEL
cetnySehtyalpsid
.nurerascitsongaid
snoitacinummocfI
dehsilbatsetonera
UPCehtneewteb
golanAdna
,draobscinortcele
edocemit-nureht
ehtro,muskcehc sitpurretnism01 ehtmorfgnissim
kcolcemitlaer
erasDELeht,)61U(
ehT(.DERottes
dnaedocemit-nur
kcolcemitlaereht
nodetacolhtobera
scinortceleUPCeht
ehttahtkcehC
liatgipdraobyek
c.4
Syntec
VitMan
yllufsielbac
ehtnodetaes
scinortceleUPC
.rotcennocdraob
ãã
ã
ãã
.evitcefedylekil
page 71
ehtotdnopserotmetsysehtfoeruliaF
ehthtiwmelborpaotebdluocdraobyek
simelborponfI.elbaclangisdraobyek
tsomsidraobscinortceleUPCeht,dnuof
Troubleshooting steps
tahtsemussasihT
semocmetsyseht
lamronstihtiwpu
htiW.neercslaitini
draobyekon
.gnitareposnoitcnuf
Page 73
System Test Procedures
petSerudecorPelanoitaRsetoN
eraslangisehT
ssorcadessap
rotalosiotpona
scinortcele
sihcihwdraob
ehtnodetnuom
UPCehtforaer
scinortcele
eruliafA.draob
ehtfoenofo
srotalosiotpo
ehttneverplliw
morfladeptoof gnitarepo .yltcerroc
noitarepoehT
ladeptoofehtfo
nideyalpsidsi
ETARBILAC
LADEPTOOF
sihT.unem
ehtsetacidni
ehtfoetats
sehctiwseerht
rabasahdna
othparg
ehtetacidni
ladeptoof
.noitisop
toofehtkcehC
d.4
otnoitcennocladep
scinortceleUPCeht
.draob
aivCDV5+htiwderewopsiladeptoofehT
denrutersinoitisopladeptoofehT.2Fesuf
dellatsnitopK5.2aaivladeptoofehtmorf
nehwdenrutersiCDV0~.ladeptoofehtni
denrutersiCDV5~dnapusiladepeht
eerht,noitiddanI.nwodyllufsitinehw
CDV5~ylppusotdeiterasehctiwsorcim
nehwCDV0~dnaetatsevitcaniehtninehw
ehtetacidnisehctiwseerhtesehT.evitca
eht
.tfelro,thgir,pusiladep
neercs
5. A/D Reference check
petSerudecorPelanoitaRsetoN
potehtevomeR
.srevocmottobdna
otMVDaesU
ehterusaem
a.5
ehtnoCDV5+
scinortcelegolanA
neewtebdraob
dna)dnG(101PT
4nip001X
Troubleshooting steps
page 72
draobscinortcelegolanAehtnoD/AehT
otecnerefer)%5(CDV5+asdeen
oslasiecnereferD/AehT.ylreporpetarepo
.srosneserusserpevifllaotdef
siCDV21+nehwCDV5+foecnesba
111Fesuffosedishtobnotneserp
raenilCDV5+ehtnieruliafasetacidni
eht.e.I.tnenopmocdetrohsarorotaluger
.evitcefedsidraobscinortcelegolanA
Syntec
CDV5+
.D/A
lliwUPCehT
ehttagnah
fiogoLcetnyS
otsliafD/Aeht
tcerroctroper
aoteudsedoc
evitcefed
ehttaecnerefer
ãã
ã
VitMan
ãã
Page 74
System Test Procedures
6. Analog electronics board check
petSerudecorPelanoitaRsetoN
amrofreP
daerMORPEE
golanAehtmorf
draobscinortcele
DAERehtgnisu
DRAOBGOLANA
ehtniunembus
a.6
.UNEMECIVRES
,sliafdaersihtfI
amrofrep
daerMORPEE
dnuosartlUehtmorf
draobscinortcele
daerehtgnizilitu
draobdnuosartlu
.dnammoc
ytpmenallatsnI
yfirevdnaettessac
b.6
ettessaCehttaht
setacidninocileveL
siettessaceht
.yltcerroc
.ytpme
gnibutatcennoC
noitaripsaehtottes
norotcennocreul
.ettessaceht
noitaripsaelbanE
toofehtesudna
c.6
retawwardotladep
.ettessacehtotni
.yltcerroc
ettessacehthctaW
yfirevdnanocilevel
ehtsetacidnititaht
ehtfognillif
.ettessac
draobscinortcelegolanAotUPCeht
erasdaerhtobfI.evitcefedsielbac
ebtsumsDELlevelettessacehT
ebtsumsDELlevelettessacehT
rossecorporcimehttahtseifirevnoitcasihT
fonoitroplartnecehtnoetacinummocnac
ehtrehtie,liafsdaerhtobfI.elbaclortnoc
ehtrodeliafsahdraobscinortceleUPC
lortnocdraobscinortceleUPCotgolanA
scinortceleUPCotgolanAeht,lufsseccus
ylnofoeruliafA.KOsielbaclortnocdraob
gniliafehtsetacidnisdraobowtehtfoeno
.decalperebtsumdnaevitcefedsieludom
noitcnufotnoitaripsaotylreporpgninoitcnuf
noitcnufotnoitaripsaotylreporpgninoitcnuf
MORPEEehtfI
ehtsliaf
laireSegassem
MORPEE
eblliweruliaf
.deyalpsid
Syntec
VitMan
ãã
ã
ãã
page 73
Troubleshooting steps
Page 75
System Test Procedures
7. Main compressor check
petSerudecorPelanoitaRsetoN
fonoitcayrevehT(
sutatsehtotgnitteg
ehttahtseilpmineercs
rewopniamCDV21+
niamehttahtyfireV
dnarosserpmoc
a.7
erapmupmuucav
ninurotdelbane
..neercssutatseht
potehtevomeR
.srevocmottobdna
yfirevotMVDaesU
b.7
si)%01(CDV21+
neewtebtneserp
pilcesufrewoleht
dna201Fesuffo
.)dnG(001PT
siCDV21+fI
rewolehtnotnesba
dnapilcesuf201F
,doogsiesufeht
niamehttcennocsid
ssenrahrewop
niamehtneewteb
c.7
dnaylppusrewop
golanAeht
.draobscinortcele
yfirevotMVDaesU
tneserpsiCDV21+
8/7snipneewteb 9dna)CDV21+( .401PJfo)dnG(
ehtronrosserpmocniamehtrehtieN
nidelbasidfinurlliwpmupmuucav
ehttahtylekilnusitI.erawtfoseht
sititubffomehtdenrutsaherawtfos
kcehcehthtrow
eht
dnaesufehtfoedisdaolehtsisihT llatatneserpCDV21+evahdluohs
yfirev,tnesbasiCDV21+fI.semit
gnivomerybneposiesufehttaht
.VMDahtiwgnitsetdna
tnesbanasahtahtesufdoogA
melborpaetacidnidluowCDV21+
noitcennocCDV21+ehthtiw
dnaylppusrewopniamehtneewteb
.draobscinortcelegolanAeht
niamehtmorfnekatsiCDV21+
ehtotdetuordnaylppusrewop
3gnoladraobscinortcelegolanA
deilppussiCDV21+.shtapetarapes
muucavdnarosserpmocniamehtot
401PJehtfo8&7snipaivpmup
asadeilppussiCDV21+.rotcennoc
aivsevlavlortnocehtotecruos
gniniamerehT.2nip401PJ
CD-CDehtgnidulcni,sdaolCDV21+
ehtrofrewopsevirdtahtretrevnoc
.1nip401PJaivdeilppusera,UPC
CDV21+eht
.)snoitcennoc
NONRUTNONRUT
NONRUTNONRUT
NONRUT
unembus
niamehtro
.seriweht
gninoitcnufsiylppus
rewopUPCehtecnis
foenomorfdeviredsi
ehtnisineercssihT
rednuuneMecivres
SPMUPFFONRUT/SPMUP
lliwesufnwolbA
ylniatrectsomla
nimelborpaetacidni
scitamuenpeht
erehtrehtiE.eludom
gniriwehtnitrohsasi
sahrosserpmoc
asA.detrohs/dellats
,revewoh,yrttsrif
nitsujesufehtecalper
esufevitcefedaesac
.eruliafehtdesuac
lacisyhprofkooL
A.snipehtotegamad
ssenrahevitcefed
gnicraesuacdluoc
rotcennocehtedisni
foeruliafotdaeldna
Troubleshooting steps
page 74
Syntec
VitMan
ãã
ã
ãã
Page 76
System Test Procedures
petSerudecorPelanoitaRsetoN
siCDV21+fI
ehtnotneserp
esuf201Frewol
ehtkcehc,pilc
ybnruterpmup
gnisuYLLUFERAC
d.7
erusaemotMVDa
ehtneewteb
batlatemdesopxe
001PTdna101Qfo
egatlovsihT.)dnG(
.evitcefed
raenebdluohs
.orez
setacidnid.7petsfI
,egatlovorezraena
301Pgulpevomer
e.7
golanAehtmorf
draobscinortcele
snipehttcepsnidna
.seriwdna
noitarbilacehtnuR
f.7
riAehtrofenituor
egnahcxE
.rosserpmoC
niamehtnineponarehtiesi ehtrogniriwrosserpmoc .deziessahrosserpmoc
sidraobscinortcelegolanA
niaMehtelbaneotdedeen
dnaCDV0raensi101QfobatehT
erehtseilpmigninnurtonspmupeht
eht,101QnotneserpsiCDV21+fI
esuacdluocnip)gnicra(evitcefedA
scinortcelegolanAehtotegamad
agnisuacot,noitiddani,draob
rewopscitamuenpehtnieruliaf
scitamuenpehtecalpeR.ssenrah
ehtotegamadsierehtfieludom
golanAehtecalpeR.sniprossenrah
egamadsierehtfidraobscinortcele
.rotcennocdetnuomBCPehtot
ehttahtyfirevoterudecorpsihtnuR
neewtebgnitareposisubataderitne ehtdnadraobscinortceleUPCeht riAehT.draobscinortcelegolanA
sezilituenituornoitarbilacegnahcxE
slangissubataddnasserddaemas
erasaegnahcxEriAehtetarepoot
osladluocnepoehT
evitcefedaeb
rehtiE(.niprotcennoc
golanAehtno
nirodraobscinortcele
rewopcitamuenpeht
).ssenrah
etomeryrevA
aebdluowytilibissop
noitacinummoc
ehtneewtebmelborp
golanAdnaUPC
.sdraobscinortcele
.ylekilnuYREVsisihT
8. Vacuum pump check
petSerudecorPelanoitaRsetoN
niamehttahtyfireV
dnarosserpmoc
a.8
erapmupmuucav
ninurotdelbane
neercssutatseht
ãã
ã
Syntec
VitMan
ãã
rosserpmocniamehtrehtieN
nurlliwpmupmuucavehtron
tI.erawtfosehtnidelbasidfi
erawtfosehttahtylekilnusi
sititubffomehtdenrutsah
.kcehcehthtrow
page 75
CDV21+ehtfoeno
SPMUP
unembus
Troubleshooting steps
rewopniamCDV21+eht
morfdeviredsirewopUPC
FFONRUT/SPMUPNONRUTFFONRUT/SPMUPNONRUT
FFONRUT/SPMUPNONRUTFFONRUT/SPMUPNONRUT
FFONRUT/SPMUPNONRUT
otgnittegfonoitcayrevehT
tahtseilpmineercssutatseht
ehtecnisgninoitcnufsiylppus
sineercssihT.)snoitcennoc
ehtrednuuneMecivresehtni
Page 77
System Test Procedures
petSerudecorPelanoitaRsetoN
fonmulocasiereh
ehtnosDELderarfni
scinortcelegolanA
erahcihwdraob
ytpmenallatsnI
yfirevdnaettessac
ehtfoenontaht
b.8
segassemgniwollof
:deyalpsidera
llufrevoettessaCllufrevoettessaC
llufrevoettessaCllufrevoettessaC
llufrevoettessaC
tneserpettessacoN
ettessacnoitisop-eR
ehttahtkcehC
c.8
esufpmupmuucav
esufehtniKOsi
neercssutats
potehtevomeR
.srevocmottobdna
ehttcennocsiD
rewopcitamuenp
d.8
MVDaesU.elbac
(CDV21+yfirevot
tneserpsi)%01
fo3nipneewteb
001PTdna301PJ
.draob
.)dnG(
ehttahtetacidnisegassemesehT
siettessacehtdenimretedsahUPC
elbaneotnoitisoptcerrocehtniton
.pmupmuucaveht
llufrevodna
.nurotpmup
ylniatrectsomlalliwesufnwolbA
scitamuenpehtnimelborpaetacidni
ehtnitrohsasierehtrehtiE.eludom
sahpmupmuucavehtrogniriw
sutatsesufehT.detrohs/dellats
rednuunemecivresehtnisineercs
.unembussesufecivedgolanaeht
esufdeliafa
.noitacidni
dnaesufehtfoedisdaolehtsisihT
llatatneserpCDV21+evahdluohs
asetacidnitnesbaCDV21+.semit
CDV21+ehthtiwmelborp
rewopniamehtneewtebnoitcennoc
scinortcelegolanAehtdnaylppus
nomsirpatademia
yllamroN.ettessaceht
sisDELfonmulocsiht
ehttcetedotdesu
ehtnidiulffolevel
llufehtdnaettessac
ehtfI.snoitidnoc
denoitisopsiettessac
UPCeht,yltcerrocni
ehtelbanetonlliw
pmupmuucavehT
citamotuanasiesuf sidnaesufgnitteser
ehT.elbaecalperton
nehwflestisteseresuf
detcennocsidsirewop
eruliafesufehT.timorf
ehtsezilitutiucric
ehtfoecneserp
ehtsaCDV21+
esufonfonoitacidni
deliafA.eruliaf
niamehtotnoitcennoc
eviglliwylppusrewop
Troubleshooting steps
page 76
Syntec
VitMan
ãã
ã
ãã
Page 78
System Test Procedures
petSerudecorPelanoitaRsetoN
siCDV21+fI
301PJhtiwtnesba
yrt,detcennocsid
ehtgnitcennocsid
rewopniam
neewtebssenrah
rewopniameht
e.8
ehtdnaylppus
scinortcelegolanA
MVDaesU.draob
siCDV21+yfirevot
neewtebtneserp
8/7snip
9dna)CDV21+( .401PJfo)dnG(
siCDV21+fI
,301PJnotneserp
ehttcennocer
rewopcitamuenp
ehtkcehC.ssenrah
ybnruterpmup
f.8
gnisuYLLUFERAC erusaemotMVDa
ehtneewteb
batlatemdesopxe
001PTdna501Qfo
.evitcefed
egatlovsihT.)dnG(
raenebdluohs
.orez
setacidnif.8petsfI
,egatlovorezraena
301Pgulpevomer
g.8
golanAehtmorf
draobscinortcele
snipehttcepsnidna
.seriwdna
noitarbilacehtnuR
h.8
riAehtrofenituor
egnahcxE
niamehtmorfnekatsiCDV21+
ehtotdetuordnaylppusrewop
3gnoladraobscinortcelegolanA
deilppussiCDV21+.shtapetarapes
muucavdnarosserpmocniamehtot
401PJehtfo8&7snipaivpmup
asadeilppussiCDV21+.rotcennoc
aivsevlavlortnocehtotecruos
gniniamerehT.2nip401PJ
CD-CDehtgnidulcni,sdaolCDV21+
ehtrofrewopsevirdtahtretrevnoc
.1nip401PJaivdeilppusera,UPC
dnaCDV0raensi501QfobatehT
sierehtseilpmigninnurtonpmupeht
pmupmuucavehtnineponarehtie
.deziessahrosserpmocehtrogniriw
eht,501QnotneserpsiCDV21+fI
sidraobscinortcelegolanA
esuacdluocnip)gnicra(evitcefedA
scinortcelegolanAehtotegamad
agnisuacot,noitiddani,draob
rewopscitamuenpehtnieruliaf
scitamuenpehtecalpeR.ssenrah
ehtotegamadsierehtfieludom
golanAehtecalpeR.sniprossenrah
egamadsierehtfidraobscinortcele
.rotcennocdetnuomBCPehtot
ehttahtyfirevoterudecorpsihtnuR
neewtebgnitareposisubataderitne ehtdnadraobscinortceleUPCeht riAehT.draobscinortcelegolanA
sezilituenituornoitarbilacegnahcxE
slangissubataddnasserddaemas
erasaegnahcxEriAehtetarepoot
.pmuPmuucaVehtelbaneotdedeen
lacisyhprofkooL
A.snipehtotegamad
ssenrahevitcefed
gnicraesuacdluoc
rotcennocehtedisni
foeruliafotdaeldna
.seriweht
osladluocnepoehT
evitcefedaeb
rehtiE(.niprotcennoc
golanAehtno
nirodraobscinortcele
rewopcitamuenpeht
).ssenrah
etomeryrevA
aebdluowytilibissop
noitacinummoc
ehtneewtebmelborp
golanAdnaUPC
.sdraobscinortcele
.ylekilnuYREVsisihT
Syntec
VitMan
ãã
ã
ãã
page 77
Troubleshooting steps
Page 79
System Test Procedures
9. Light source check
petSerudecorPelanoitaRsetoN
dednelbenosierehT
ehT.ytilibissoperuliaf
sesuyartecruosthgil
evomoterusserpria
sblubBdnaAeht
owtehtfI.noitisopotni
dessorcerewsenilria ehtro,riaperagnirud ecruosthgillaudividni
ehtnosrotcennoc
scinortcelegolanA
dessorcerewdraob eht,riaperagnirud pukcabehtniblub
tubthgillliwnoitisop
ehtthgilotliaflliw
tonlliwtiecnissrebif
.htapcitpoehtnieb
erawtfoslortnocehT
ehtedilssyawlalliw
ehtsatfelehtotsblub
sitI.deneposireward
yfitnediottnatropmi
pmaltcerrocehttaht
rewardehtsatilsaw
.denepotsrifsi
.pugnithgil
thgilehttahtyfireV
a.9
nodenrut
pusthgilecruos
sinoitcnufehtnehw
}f.9
}m.9
ytisnetniehtkcehC
ehtnobonklortnoc
thgilehtfotnorf
ehtfI.ecruos
tatessiytisnetni
ehtdnamumixam
situptuonoitanimulli
b.9
,etauqedaton
rebifehtkcehc
rofelbaccitpo
rehtollafI.sknik
erasnoitacidni
ehtecalper,lamron
citporebif
.rotanimulliodne
ehttahtyfireV
ehtnisiblubtcerroc
ybhtapcitpo
thgilehtgninepo
c.9
htiwrewardecruos
ehT.nonoitanimulli
dluohsblubgniwolg
ehtfotnorfnieb
rewardehtsarorrim
.sneposi
shtapgnireffidowteraerehT
erasblubehtrehtehwnognidneped
melborpehtneht,thgilsierehtfI)A
fotuoblubthgiL(.htapcitpoehtnisi
,msinahcemgnimmid,tnemngila
c.9spetsesU{).cte,srettuhsrebif
ehtneht,thgilonsierehtfI)B
h.9spetsesU{.lacirtcelesimelborp
gnihctiwsblubehtnieruliafA
asiblubehtevaeldluocmsinahcem
siblubtcerrocniehterehwetats
ybdesuacebnacsihT.detanimulli
egnahcxeblubehtnikaelriana
blubehtnoeruliafybromsinahcem
.sevlavegnahcxe
Troubleshooting steps
page 78
Syntec
VitMan
ãã
ã
ãã
Page 80
System Test Procedures
petSerudecorPelanoitaRsetoN
siblubtcerrocniehtfI
ekam,detanimulli
sblubhtobtahterus
-erdnadoogera ecruosthgilehtllatsni
.tinuehtotnireward dnapotehtevomeR
d.9
nruT.srevocmottob
dnanoitanimullino
ehtesuneht
FFO/NOnoitanimulli
ehttcelesotnottub
tahtyfireV.blubB lortnocblubhtob riehtevahsevlav
.tilsDEL
sDELevlavhtobehtfI
nrut,etanimullitonod
edisstinometsyseht
ehttcennocsiddna
lortnocscitamuenp
ehterusaeM.elbac
e.9
neewtebecnatsiser
tI.61dna51snip
~erusaemdluohs
erusaeM.smho052
ecnatsisereht
dna71snipneewteb
osladluohstI.81
smho052~erusaem
nisiblubtcerrocehtfI
,htapscitpoeht
f.9
ehtnepoylluferac
rewardecruosthgil
ehttcepsnidna
.htapscitpo
ehtniatrecekaM
g.9
yllufsirewardscitpo
ehtotnidellatsni
.metsys
golanAeht,doogerastiucric
.rewardscitpoehtecalper
.sblubehtotrewoptcennoc
dellortnocsiedilsnoitisopthgilehT
citamuenpehtnosevlavforiapayb
ebdluohssevlavhtoBeludom
pukcabehtecrofotdezigrene
.blubBehtottniopotmsinahcem
asetacidnitiucricrehtienineponA
htobfI.eludomscitamuenpevitcefed
.evitcefedsidraobscinortcele
dna,retlifder-arfnieht,sesnelhtoB
dlehebdluohsrorrimlacirehpseht
gniniaterriehthtiwecalpniylmrif
gnimmideht,noitiddanI.sgnir
.yleerfetatordluohsmsinahcem
otmsinahcemrettuhsehttcepsnI
esoolnekorbtonsahtierusekam
ynafI.noitacoldetnemecstimorf
,htapscitpoehtnidnuofsimelborp
noitisopnidlehsirewardscitpoehT erugiFMOeeS(hctalgnitatorahtiw
yllufrewardehtevahoteruliafA.)41
oteruliafanitluserlliwdellatsni
Syntec
VitMan
ãã
ã
ãã
page 79
Troubleshooting steps
Page 81
System Test Procedures
petSerudecorPelanoitaRsetoN
noitanimulliehtyalpsidlliwtinuehT
scinortcelegolanAehtnehwnocI
ehtnrutotdednammocsidraob
thgilehttahtyfireV
h.9
detanimulliecruos
BCLehtnisinocI
.yalpsid
noitanimullifI
tonotsraeppa
i.9
,ylreporpnoitcnuf
elosnocehtkcehc
gninrawrofyalpsid
.segassem
,liafsnottubllafI
ehttahtkcehc
ecafretnidraobyek
j.9
detaesyllufsielbac
snipgnitamehtotno
UPCehtfo
.draobscinortcele
nimelborpaebyamereht
aotetalosiotdetavitcaebnac
eht,eruliafasahnottubelgnis
.evitcefedsidraobyek
.sregniferabhtiwblubeht
noitcennocdetaesyltcerrocA
.draobscinortceleUPC
,raeppatonseodnocIehtfI.nopmal
.nottubnoitanimulliehtgnizingocer
snoitcnuflacigrusrehtotahtkcehC
afI.eruliafasahtahtnottubcificeps
eB.yrassecenfi)s(blubehtecalpeR tohtonsiblubehttahtluferacyrev hcuottonoD.devomersitierofeb
UPCehtdnadraobyekehtneewteb
ehtnisieruliafehtetacidnidluow
Troubleshooting steps
page 80
Syntec
VitMan
ãã
ã
ãã
Page 82
System Test Procedures
petSerudecorPelanoitaRsetoN
setacidniegassemgnissimXblubehT
ehteestonseodmetsysehttaht
ro,tuodenrubebyamtI.blubdeificeps
htapCDV42+ehtnikaerbasiereht
ehtotkcabdna)s(blubehthguorht
erasblubehT.draobscinortcelegolanA
niamehtmorfCDV42+htiwdeilppus
ehT.601Fesufaivylppusrewop
lliwdnadesnessi601Ffonoitidnoc
.nepositifieruliafesufblubayalpsid
teserlliwtI.esufelbatteserasi601F
asA.devomersirewopnehwflesti
ylekiltsomeruliafesufbluba,tluser
ylppusblubehtfoenonitrohsasetacidni
ecruoSthgiLehtedisnitrohsaroselbac
eruliafesufblubehT
rucconacnoitidnoc
blubafinoitarepo tuotrohsoterew
.sliaftisayllanretni
esufpirtlliwtrohsehT
metsysehT.601F
ehtothctiwslliw
lliwdna,revewoh
sanoitanimullierotser
sah601Fsanoos
enontahtyfireV
gniwollofehtfo
erasegassem
DCLehtni
:enilsutats
k.9
gnissimAbluB)1gnissimAbluB)1
gnissimAbluB)1gnissimAbluB)1
gnissimAbluB)1
gnissimBbluB)2
eruliafAbluB)3
eruliafBbluB)4
eruliafesufbluB)5
.rewarDecruoS
.draobscinortcele
ecruoSthgiLehtesaeleR.rewarD
.tinuehtfotuoyawtraptillupdnarewarD
,spotsegassemeruliafesufbluBehtfI
lamrongnirud
seriwehtfoenonitrohsasiereht
bluBehtfI.stekcospmalehtotgnidael eht,stsisrepegassemeruliafesuf ebotdeenlliweludomscitamuenp
CDV42+EGNAROehttahtosdevomer
yehtsadetcepsniebnacseriwecruos
dnadraobscinortcelegolanAehtevael
,blubpukcab
thgiLehtfoediskcabehtotdetuorera
setacidniegassemeruliafXblubehT
nrutersblubehtgnivirdtonsimetsyseht
.derevocer
ssalcegassemsihT.dnuorgotylluf
golanAehtnoeruliafasetacidni
Syntec
VitMan
retfastsisrep)2ro)1egassemrehtiefI
nwonkahtiwdecalperneebsahblubeht
/eriwnekorbaylekilsiereht,blubdoog
golanAehtneewtebelbacdetaesnu
ecruoSthgiLehtdnadraobscinortcele
ehtotdetcennocsiblubAehT.rewarD
hguorhtdraobscinortcelegolanA
siblubBehT.701PJrotcennoc
htoB.601PJrotcennocaivdetcennoc
tarotcennocnip-4nommocaerahssblub
.rewarDecruoSthgiLehtforaereht
ãã
ã
ãã
page 81
Troubleshooting steps
Page 83
System Test Procedures
petSerudecorPelanoitaRsetoN
eruliafesufblubehT
llatahtyfireV
ehtfoeerht
gniwollof
erasegassem
ehtnitonni
l.9
eruliaf
nepo
:enilsutatsDCL
eruliafesufbluB)1eruliafesufbluB)1
eruliafesufbluB)1eruliafesufbluB)1
eruliafesufbluB)1
esufpmupriA)2
esufdnuosartlU)3
tahtyfireV
ehtforehtien
gniwollof
m.9
erasegassem
:deyalpsid
melborpAbluB)1melborpAbluB)1
melborpAbluB)1melborpAbluB)1
melborpAbluB)1
melborpBbluB)2
.draobscinortcele
ehtrofkcehC
CDV42+sreviledylppusrewopniamehT
eerhtllA.sesuftnereffideerhthguorht
aylekiltsomsinepognitacidnisesuf
ehtroylppusniamCDV42+ehtnieruliaf
rewopniamehtneewtebssenrahelbac
.draobscinortcelegolanAehtdnaylppus
ehT.ssenrah
siCDV42+
egats-owtaninodenrutsiblubehT
blubehtsmrawegatstsrifehT.ssecorp
lliwsihT.rotsiserW5seiresagnisuyb
dnocesehT.wolgotblubehtesuac
tuognitrohsybyllufblubehtsthgilegats
blubehtthgiloteruliafA.rotsiserW5eht
golanAehtnieruliafasetacidniylluf
rucconacnoitidnoc
CDV42+foecneserp
rewopniamehtta
tneserpsitifI.ylppus
,ylppusrewopehtta
nekorbarofkcehc
ninoitcennoc/eriw
rewopCDniameht
3nipotdetcennoc
.eriwEGNAROnaaiv
10. Vitrector check
petSerudecorPelanoitaRsetoN
ehttahtyfireV
a.01
dnuosartlU
TONsinoitcnuf
.delbane
yllautumsinoitcnufdnuosartlUehT
ehtfI.rotcertivehthtiwevisulcxe
der(detanimullisiDELdnuosartlU
sinoitcnufdnuosartlUeht,)neergfo FFO/NOdnuosartlUehT.detceles elpitlumdesserpedebtsumnottub
.dehsiugnitxesiDELehtlitnusemit
ehtfonoisserpedhcaE
nottubFFO/NOdnuosartlU
ehthguorhtselcyc
ehtfosedomlanoitarepo
ehT.eceipdnahdnuosartlu
.ffosiedomtsal
Troubleshooting steps
page 82
Syntec
VitMan
ãã
ã
ãã
Page 84
System Test Procedures
petSerudecorPelanoitaRsetoN
sDELehtfoytisnetniehT
DELehtnidellortnocsi
ehtfounembusYTISNETNI
ekaM.unemYALPSID
pusiytisnetniehttahterus
ehtfosutatsehteesotylluf .moortilylthgirbanisDEL
nacsrorrelanretnineveS
morfrotcertivehtpeek
rotcertivehtfI.gnitarepo
rofkcehc,DERsiDEL
gniwollofehtfoeno
eruliafrosnvesevirD)3
erusserpevirdhgiH)4
erusserpevirdwoL)5
tneserpladeptoofoN)6
,seslupriaoneraerehtfI fonoitarepoehtrofnetsil
dnaevlavrotcertiveht
fonoitarepoehtevresbo
DELevlavrotcertiveht
tonseodDELevlavehtfI
aebyamereht,etanimulli
yllufton/eriwnekorb
/rotcennocdetaes
scitamuenpehtneewteb
golanAehtdnaeludom
.draobscinortcele
ehttahtyfireV
noitcnufrotcertiv
b.01
nodelbanesi
tahtdnatinueht
rotcertiveht
.neergsiDEL
.dehsiugnitxe
ehttcennocsiD
ehtmorfrotcertiv
reullenaptnorf
dnarotcennoc
ehtsserped
c.01
.ladeptoof
riatahtkcehC
eraseslup
ehtmorftuptuo
rotcertiv
.rotcennoc
.metsys
oneraerehtfI
,seslupria
potehtevomer
mottobdna
.srevoc
d.01
ehtrofnetsiL
ehtfonoitarepo
,evlavrotcertiv
ehtevresbodna
ehtfonoitarepo
evlavrotcertiv
.DEL
ybedameraskcehclanretnilareveS
eht,stsixerorrenafI.metsyseht
delbasideblliwnoitcnufrotcertiv
ehtfI.dereblliwDELehtdna
rotcertiveht,KOeraskcehclanretni
.neergeblliwDELsutats
rotcertivehtfidraobyekehttcepsuS
onsierehtdnathgiltonlliwDEL
.yekehtgnisserpedmorfkcilcyek
nidelbasidebdluockcilcyekehT(
)unembussecivreSoiduAeht
yalpsidehtfiUPCehttcepsuS
sahedomrotcertivehttahtsetacidni
:segassem
delbasideciveD)1delbasideciveD)1
delbasideciveD)1delbasideciveD)1
delbasideciveD)1
eruliafesufevirD)2
llitssiDELehttubderetneneeb
eruliafesufevlaV)7
niruccodluocnoitcnuflamrotcertivA
.lacirtceleebrohtapcitamuenpeht
eht,tneserpseslupriaeraerehtfI
aebdluowsmelborpylekiltsom
yltcerrocninarorotcertivevitcefed
ehtnorotalugererusserpdetsujda
.eludomscitamuenp
naebdluowytilibissopetomerA
neewtebebutdetcennocyltcerrocni
ehtdnaeludomcitamuenpeht
gniynapmoccanasahevlavhcaE
ehtnehwdetanimullisitahtDELder
ehtfoesacehtnI.detavitcasievlav
detavitcasievlaveht,evlavrotcertiv
ehttuoesluperusserpaecalpot
.rotcennocrotcertiv
siriatubgnihsalfsiDELevlavehtfI
aylekiltsomsiereht,tnesba
eht,htapriaehtninoitcurtsbo
rotcertivehtrofrotalugererusserp
sievlavrotcertivehtro,deliafsah
otgniliaftubrewopgniviecer
tahteriuqeresehtfoynA.etarepo
.decalperebeludomscitamuenpeht
Syntec
VitMan
ãã
ã
ãã
page 83
Troubleshooting steps
Page 85
System Test Procedures
petSerudecorPelanoitaRsetoN
yllufton/eriwnekorbaebyamerehT
DELevlavehtfI
,etanimullitonseod
nometsysehtnrut
.draobscinortcele
dnaedissti
ehttcennocsid
e.01
lortnocscitamuenp
ehterusaeM.elbac
neewtebecnatsiser
tI.6dna5snip
~erusaemdluohs
.smho052
.decalper
rotcertivehtfI
siDELnoitcnuf
ehttahtyfirev,neerg
siladeptoof
f.01
ehtgnidnammoc
ehtfI.tucotrotcertiv
seodnocienitolliug
ees,raeppaton
ehtnimelborp
).unembusFFO
.ecafretniladeptoof
ehtneewteb/rotcennocdetaes
golanAehtdnaeludomscitamuenp
,tiucricneponasetacidniecnatsiserehtfI
,elbacevlavehtnikaerbarehtiesiereht
,esacrehtienI.evitcefedsievlavehtro
.evitcefedsieludomscitamuenpeht
siereht,desolcsitisetacidnitiucricehtfI
.scinortceleevirdehthtiwmelborpaylekil
ebdluohsdraobscinortcelegolanAehT
nocienitolliugawohslliwyalpsidehT
ehtotdetatorsiladeptoofehtnehw
yllamronsisihT.noitisopgnittucelbane
ehT(.ladepehtfonoitisopthgireht
ehtmorfdelbasidebdluocnocienitolliug
SINOCIRETTUC/NOSINOCIRETTUC
rotcertivehT
sievlav
otdetcennoc
ehtfo6&5snip
.elbaclortnoc
noitisoptucehT
dehctiwsebnac
ehthtiwtfelehtot
NOSIXULFER
XULFER/TFEL
THGIRNOSI
.dnammocunem
11. Foot pedal interface check
petSerudecorPelanoitaRsetoN
eerhtsniatnocladeptoofehT
nehwevitcaeratahtsehctiws
stimorfdevomsiladepeht
ehtnieruliafA.noitisopeldi
dluocelbacstiroladeptoof
scinortceleUPCehtgnihcaer
ehtedisnislangisehtsetuor
lenapraerehtmorfmetsys
nekorbA.draobscinortcele
/detaesyllufton/eriw
ehtesuacdluocrotcennoc
esehtfoynanieruliafA
melborpasetacidnisegatlov
noeruliafA
ladeptoofeht
oslalliwesuf ehtpeek sehctiws
gniebmorf
nA.detceted
esufnepo
anitluserlliw
esufladeP
eruliaf
ninoitacidni
yalpsideht
.draobybab
.detcennocsiladepeht
a.11
.esiwrehtoCDV5~
.esiwrehtoCDV5~
.esiwrehto
.desserpedylluf
rotalosiotpoehtno1Pfo01dna
.CDV5~taebdluohs2niP)1
tadnathgirdevomsiladepeht
nehwCDV0~aebdluohs7niP)4
tadnatfeldevomsiladepeht
CDV0~tadnapusiladepeht
nehwCDV0~taebdluohs4niP)2 nehwCDV0~taebdluohs7niP)3
nehwCDV5~taebdluohs9niP)5
segatlovehterusaemotMVDaesU
,9,7,5,4,2snipdna1nipneewteb
morflangisehttneverp
eht,noitiddanI.draob
elbaclortnocladeptoof
UPCehtotrotcennoc
.melborp
nehwCDV0~taebdluohs01niP)6
sitinehwCDV5~dnapusiladepeht
ladeptoofehtni
Troubleshooting steps
page 84
Syntec
VitMan
ãã
ã
ãã
Page 86
System Test Procedures
petSerudecorPelanoitaRsetoN
dnapotehtevomeR
.srevocmottob
ladeptoofehttcennocsiD
otretemtlovaesudna
ehtfoytivitcennocyfirev
lortnocladeptoofretni
b.11
ehtgnimrofrepybelbac
raerehttastsetgniwollof
ladeptooflenap
:rotcennoc
notneserpsiCDV5+~
.1dna2snipneewteb
lanretniehttcennocsiD
elbaclortnocladeptoof
rotalosiotpoehtmorf
aesU.draobscinortcele
c.11
ehterusaemotMVD
ehtneewtebecnatsiser
,1sniprotcennocenilni
ehtdna01,9,7,5,4,2
rotcennoclenapraer
7,6,5,4,3,2,1snip
ehtroladeptoofehtnitrohsA
ehtwolbnacelbacladeptoof
ehtfI.esufladeptoofelbatteser
nehwsraelceruliafesufladeP
,detcennocsidsiladeptoofeht
evitcefedsiladeptoofeht
xisehtfoenoynanikaerbA
nieruliafasetacidnisrotcudnoc
tooF,naMtiV,yssAelbaCeht
.A4650001NPlanretnIladeP
morf)CDV5+(2niP
rotalosiotpoeht
sidraobscinortcele
ehtni2nipotdetuor
.rotcennocladeptoof
morf)dnuorG(1niP
rotalosiotpoeht
sidraobscinortcele
ehtni1nipotdetuor
.rotcennocladeptoof
12. Ultrasound check
petSerudecorPelanoitaRsetoN
yllautumsinoitcnufrotcertiVeh
fI.dnuosartlUehthtiwevisulcxe
ehtfonoisserpedhcaE
nottubFFO/NOrotcertiv
ehthguorhtselcyc
ehtfosedomlanoitarepo
edomtsalehT.rotcertiv
ehttahtyfireV
a.21
sinoitcnufrotcertiV
.delbaneTON
detanimullisiDELrotcertiVeht
rotcertiveht,)neergfoder(
ehT.detcelessinoitcnuf
ebtsumnottubFFO/NOrotcertiV
litnusemitelpitlumdesserped
.dehsiugnitxesiDELeht
.ffosi
Syntec
VitMan
ãã
ã
ãã
page 85
Troubleshooting steps
Page 87
System Test Procedures
petSerudecorPelanoitaRsetoN
sisDELehtfoytisnetniehT
DELehtnidellortnoc
ehtfounembusYTISNETNI
erusekaM.unemYALPSID
yllufpusiytisnetniehttaht
ehtfosutatsehteesot
.moortilylthgirbanisDEL
nacsrorrelanretnineetniN
morfdnuosartlUehtpeek
dnuosartlUehtfI.gnitarepo enorofkcehc,DERsiDEL :segassemgniwollofehtfo
eceipdnahenutesaelP)11
ylnobonkgnikcolesU)81
ebnacnoitisoptucehT
ehthtiwtfelehtotdehctiws
NOSIXULFER/TFEL
.dnammocunemTHGIR
ehtevahoterusekaM
nehwdemirpdnuosartlu
ehthtiwstsetgnimrofrep
ehT.eceipdnahdnuosartlu
emoceblliweceipdnah
onsierehtnehwmrawetiuq
ehthguorhttnaripsafowolf
ehttahtyfireV
sinoitcnufdnuosartlU
b.21
tinuehtnodelbane
dnuosartlUehttahtdna
siDELdnuosartlU
.neerg
dnuosartlUehtfI
,neergsiDELnoitcnuf
ladeptoofehttahtyfirev
ehtgnidnammocsi
c.21
ehtfI.tucotdnuosartlU
tonseodnocievaw
.11ees,raeppa
ecafretnIladeptooF
kcehC
seodnocievawehtfI
ehtkcehc,raeppa
niegatnecreprewop
d.21
rewopehT.yalpsideht
elbatsujdasidereviled
ni%001ot%1morf
htiwstnemercni%1
noyllufgnieb%001
eraskcehclanretnilareveS
nafI.metsysehtybedam
dnuosartlUeht,stsixerorre
dnadelbasideblliwnoitcnuf
ehtfI.dereblliwDELeht
eht,KOeraskcehclanretni
lliwDELsutatsdnuosartlU
.neergeb
ehtfidraobyekehttcepsuS
thgiltonlliwDELdnuosartlU
morfkcilcyekonsierehtdna
ehT(.yekehtgnisserped
nidelbasidebdluockcilcyek
secivreSoiduAeht
)unembus
ehtfiUPCehttcepsuS
ehttahtsetacidniyalpsid
neebsahedomdnuosartlU
llitssiDELehttubderetne
.dehsiugnitxe
evawawohslliwyalpsidehT
siladeptoofehtnehwnoci
elbaneehtotdetator
sisihT.noitisopgnittuc
fonoitisopthgirehtyllamron nocievawehT(.ladepeht ehtmorfdelbasidebdluoc
SINOCIRETTUC
FFOSINOCIRETTUC/NO
).unembus
ebyamlevelrewop%1A
ynamrofrepotllamsoot
.gnittucelbaeciton
llufrevoettessaC)1llufrevoettessaC)1
llufrevoettessaC)1llufrevoettessaC)1
llufrevoettessaC)1
dabnoitisopettessaC)2
eruliafevlavettessaC)3
delbasideciveD)4
eruliafesufevirD)5
levelmuucavhgiH)6
tneserpettessaCoN)7
tneserpladeptoofoN)8 ettessackcolesaelP)01 ettessacnoitisop-eR)21
!detrobagninuT)31
!deliafgninuT)41
eceipdnahgninuT)51
dewollatongninuT)61
eruliafesufmuucaV)91
eruliafesufevlaV)02
NOSIXULFER
.eldnah
tneserpeceipdnahoN)9
nepoesufdnuosartlU)71
Troubleshooting steps
page 86
Syntec
VitMan
ãã
ã
ãã
Page 88
System Test Procedures
petSerudecorPelanoitaRsetoN
seilppushcihwelbaclanretniehT
tnorfehtotrewopdnuosartlu
ehttahtyfireV
tcennocretnidnuosartlU
e.21
ylmrifsielbac
ehtotdetcennoc
scinortcelednuosartlu
.draob
tonsieceipdnahehtfI
ehttahtyfirev,detceted
elbaclortnocdnuosartlu ehtneewtebsnurhcihw
scinortcelegolana
f.21
ehtdnadraob
scinortcelednuosartlu
tadetaesyllufsidraob
erehttahtdnasdnehtob
ehtotegamadonsi
.elbac
sieceipdnahehtfI
tonlliwtubdetceted
g.21
ehtrofkcehc,enut
sutatsesufdnuosartlU
.egassem
sieceipdnahehtfI
tonlliwtubdetceted
rewopehtkcehc,enut
ehtotdereviled
scinortcelednuosartlu
ehtesaeleR.draob
ylluferacdnalenapraer
h.21
rewopehttcennocsid
ehtmorfssenrah
scinortcelednuosartlu
otMVDaesU.draob
ehttahterusaem
reptcerrocerasegatlov
dnuosartlUgolanAeht
.tuonipelbaCrewoP
ehT.ygreneehtseilppus
.eceipdnah
ehtfognisnestcerroceht
.eceipdnah
dnuosartlueht,tneserpsi
evitcefed
evitcefedsidraob
.evitcefedsi
rednudetuorsirotcennoclenap
dnaylppusrewopmetsyseht
ehtnosrotcennocowtotno
.draobscinortcelednuosartlu
thgirreppuehtnirotcennocehT
ehtsesneswolebtsujrotcennoc
yletamitlusieceipdnahehT
UPCehttatneserpsadesnes
nekorbA.draobscinortcele
tneverplliwelbacrehtienilangis
egassemesufdnuosartlUehtfI
.evitcefedsidraobscinortcele
draobscinortcelednuosartluehT
ehtmorfetarapesyllanoitcnufsi
scinortceleBCPdnagolanA
ehtnieruliafasierehtfI.sdraob
eruliafeht,scinortcelednuosartlu
gniniamerehtpeektonlliw
ehtfI.gnitarepomorfsnoitcnuf
tcerrocehtsahelbacrewop
smetigniwollofeht,segatlov
:melborpehtgnisuacebdluoc
sieceipdnahdnuosartluehT.1
scinortcelednuosartluehT.2
draobscinortcelegolanaehT.3
snoisarbarofkcehC
sitI.elbacehtno
tahttnatropmiYREV
otruccoegamadon
.elbacsiht
ytilibissopylekilnunA
oteruliafehttahtsi
eceipdnahehtesnes
asitneserpsa
UPCevitcefed
.draobscinortcele
onsierehterusekaM
golanAehtotegamad
rewoPdnuosartlU
gnicalpererofebelbac
dnuosartlueht
.draobscinortcele
noitalusnieriwehT
neebevahsroloc
ehtevahotnesohc
:gninaemgniwollof
dnuorG:kcalB
CDV5+:deR
CDV21+teloiV
CDV21-:yerG
CDV42+:egnarO
Syntec
VitMan
ãã
ã
ãã
page 87
Troubleshooting steps
Page 89
System Test Procedures
13. Aspiration check
petSerudecorPelanoitaRsetoN
DELehtnidellortnoc
skcehclanretnilareveS
.metsysehtybedamera
eht,stsixerorrenafI
lliwnoitcnufnoitaripsA
dnaettessacallatsnI
a.31
ehttahtdnatinu
siDELnoitaripsAehtfI
siereht,desserpedsi
kaelrianaylekil
ehthtiwdetaicossa
b.31
.srevocmottobdna
gnibutehttcepsnI
golanAehtfosrosnes
oslA.draobscinortcele
gnioggnibutehttcepsni
rotalugermuucavehtot
.retlif
noitaripsAehttahtyfirev
ehtnodelbanesinoitcnuf .neergsiDELnoitaripsA
ladeptoofehtlitnuneerg
noitaripsaehtfonoitarepo
potehtevomeR.noitcnuf
muucavehtotdetcennoc
cibohpordyhegralehtdna
DELehtdnadelbasideb
lanretniehtfI.dereblliw
eht,KOeraskcehc
DELsutatsnoitaripsA
.neergeblliw
fidraobyekehttcepsuS
lliwDELnoitaripsAeht
onsierehtdnathgilton
gnisserpedmorfkcilcyek
kcilcyekehT(.yekeht
ehtnidelbasidebdluoc
secivreSoiduA
)unembus
ehtfiUPCehttcepsuS
ehttahtsetacidniyalpsid
sahedomnoitaripsA
ehttubderetneneeb
.dehsiugnitxellitssiDEL
nacmetsysmuucavehT
otskaeletarelotton
kaelynA.erehpsomta
stceffeesrevdaevahlliw
noitaripsaehtno
.noitcnuf
:segassem
.gnibuteht
.moortilylthgirba
eruliafrosnescc005)1eruliafrosnescc005)1
eruliafrosnescc005)1eruliafrosnescc005)1
eruliafrosnescc005)1
eruliafrosnescc05)2
lluFettessaC)3
llufrevoettessac)4
dabnoitisopettessaC)5
eruliafevlavettessaC)6
delbasideciveD)7
melborpladeptooF)8
eruliafesufevirD)9
levelmuucavhgiH)01
erusserpevirdwoL)11
levelmuucavwoL)21
tneserpettessaCoN)31
tneserpladeptoofoN)41
eruliafesufladeP)51
ettessackcolesaelP)61 ettessacnoitisop-eR)71
ylnobonkgnikcolesU)81
eruliafesufmuucaV)91
eruliafesufevlaV)02
gniwollofehtfoeromro
ehtraensessapgnibut
sisDELehtfoytisnetniehT
ehtfounembusYTISNETNI
erusekaM.unemYALPSID
otyllufpusiytisnetniehttaht
nisDELehtfosutatsehtees
nacsrorrelanretniytnewT
morfnoitaripsAehtpeek
noitaripsAehtfI.gnitarepo
enorofkcehc,DERsiDEL
ebdluohsnoitcepsniehT
ehterehwhguorohtyrev
pmupsihT.pmupmuucav
dnastnuomstinoevomnac
tsniaganoisarbaesuacnac
Troubleshooting steps
page 88
Syntec
VitMan
ãã
ã
ãã
Page 90
System Test Procedures
petSerudecorPelanoitaRsetoN
ot0morfelbatsujdasinoitaripsaehT
DELnoitaripsAehtfI
tahtkcehc,neergsi
gnittesnoitaripsaeht
rehtoeulavaottessi
c.31
ehttahtdnaoreznaht
evlavhcnipnoitaripsa
ehtnehwsnepo
yllufsiladeptoof
.desserped
evlav
hcnipnoitaripsaehtfI
,nepootsliafevlav
d.31
fognittesehtkcehc
NOITARIPSAeht
ehtniDLOHSERHT
unemsgnittestnerruc
ehtfognittesehtfI
NOITARIPSA
siDLOHSERHT
potehtevomer,dilav
.srevocmottobdna
tcerrocehtrofkcehC
ehtfonoitarepo
e.31
hcnipnoitaripsa
evlavlortnocsevlav
ehtgnivresboyb
lortnocehtnoDEL
DELsihT.evlav
.evlavlortnoc ehtnehwtilebdluohs ebdluohsevlavhcnip
.nepo
ehT.stnemercnigHmm5nigHmm005
ehtotevitalermetsysehtfonoitisop
,yllaedI.noitaripsastceffeoslatneitap
emasehttadecalpeblliwmetsyseht
.elttobnoitagirriehtsathgieh
ehtnahtrehgihdetacolsimetsysehtfI
evissaptegotelbissopsiti,elttob
ehtotpusiladepehtnehw.e.I.xulfer
raensinoitaripsaehttahttniop
wolfesreveresuaclliwytivarg,gHmm0
othguonehgihdesiarsiladepehtlitnu
.evlavhcnipnoitaripsaehtesolc
ehtnahtrewoldetacolsimetsysehtfI
evissaptegotelbissopsiti,elttob
otpusiladepehtnehw.e.I.noitaripsa
raensinoitaripsaehttahttniopeht
esuacoteunitnoclliwytivarg,gHmm0
hgihdesiarsiladepehtlitnuwolf
hcnipnoitaripsaehtesolcothguone
ehtseificepsdlohserhtnoitaripsaehT
ebtsumhcihwlevelmuucavmuminim
hcnipnoitaripsaehterofebtneserp
fI.nepootdednammoceblliwevlav
lliwevlaveht,hgihoottessieulavsiht
.nepootdednammocebreven
gniebsirewoptahtsetacidniDELehT
noitaripsaehT.evlavehtotdeilppa
erusserpriaybdetareposievlavhcnip
riaehT.eludomscitamuenpehtmorf
.evlavlortnocehthguorhtdeilppussi
nehwdesolcnevirdsievlavhcnipehT
sidnatneserperusserpriaonsiereht
ehthguorhtriagniylppusybdenepo
sinoitaripsA
ehtybdellortnoc
.noitisopladeptoof
levelnoitaripsaehT
morfraenilsi
ehtotgHmm0
gnittesmumixam
ehtnidetceles
:sedomgniwollof
noitaripsAroiretsoP
roiretsoP
ymotcertiV
roiretsoP
dnuosartlU
noitaripsAroiretnA
ymotcertiVroiretnA
dnuosartlUroiretnA
cetnySehteeS
leunaMsrenwO
.CxidneppA
DELehtfoeruliafA
etacidninacthgilot
ehtnieriwnekorba
lortnoccitamuenp
evitcefeda,elbac
melborparo,evlav
golanAehthtiw
.draobscinortcele
Syntec
VitMan
ãã
ã
ãã
page 89
Troubleshooting steps
Page 91
System Test Procedures
petSerudecorPelanoitaRsetoN
DELevlavehtfI
,etanimullitonseod
nometsysehtnrut
dnaedissti
ehttcennocsid
f.31
lortnocscitamuenp
ehterusaeM.elbac
neewtebecnatsiser
tI.8dna7snip
.evitcefed
~erusaemdluohs
.smho521
noitaripsaehtfI
snepoevlavhcnip
tub,dluohstisa
,muucavonsiereht
g.31
dnapotehtevomer
.srevocmottob
egralehttahtyfireV
siretlifcibohpordyh .diulfhtiwdellifton revocehtevomeR
muucavehtmorf
aesudnarotaluger
tahtyfirevotMVD
tneserpsiCDV21+ 1slanimretneewteb
ehtedisni3dna
h.31
ehttsujdA.rotaluger
otlevelnoitaripsa
dnamumixam
ladepehtsserped
tahtyfireV.ylluf
tneserpsiCDV5~
2slanimretneewteb
.langis
ehtedisni3dna
rotaluger
yllufton/eriwnekorbaebyamerehT
ehtneewteb/rotcennocdetaes
golanAehtdnaeludomscitamuenp
.draobscinortcele
neponasetacidniecnatsiserehtfI
evlavehtnikaerbarehtiesiereht,tiucric
rehtienI.evitcefedsievlavehtro,elbac
sieludomscitamuenpeht,esac
siereht,desolcsitisetacidnitiucricehtfI
evirdehthtiwmelborpaylekil
scinortcelegolanAehT.scinortcele
.decalperebdluohsdraob
ehtstcetorpretlifcibohpordyhegralehT
tahttnaripsaynamorfrotalugermuucav
forebmahccc05ehtmorfepacsethgim
ettessacllufafI.ettessacnoitaripsaeht
denrutsitiehtelihwmetsysehtnitfelsi
cc05ehtnitnaripsayna,dneno
muucavehtotniniardlliwrebmahc
ehtnithguacebdnagnibutrotaluger
.retlifcibohpordyh
htiwmelborpasierehttahtelbissopsitI
sirotalugerehT.rotalugermuucaveht
golanAehtmorfCDV21+htiwderewop
scitamuenpehtaivdraobscinortcele
dellortnocsimuucavehT.elbaclortnoc
CDV0~.langisCDV5otCDV0ahtiw
CDV5~elihwGHmm0setacidni
,ladeptoofehT.GHmm055setacidni
noitaripsamumixamehthtiwdenibmoc
ehtfoedutilpmaehtslortnoc,gnittes
noitaripsaehT
detcennocsievlav
fo8&7snipot
.elbaclortnoceht
dnuofsitnaripsafI
cibohpordyhehtni
ebnacti,retlif
tuohtiwdecalper
ehtgnicalper
scitamuenperitne
.eludom
eraerehT
elbatsujda
fosdlohserht
levartladeptoof
ebtsumtaht
etaitiniotdessap
ehT.noitaripsa
erasdlohserht
ehtnidellortnoc
LADEP
SDLOHSERHT ehtfounembus sgnittestnerruc
.unem
Troubleshooting steps
page 90
Syntec
VitMan
ãã
ã
ãã
Page 92
System Test Procedures
petSerudecorPelanoitaRsetoN
edisnisegatlovehtfI
erarotalugereht
ehtnrut,tcerrocni
edisstinometsys
ehttcennocsiddna
lortnoccitamuenp
ehtmorfelbac
i.31
scinortcelegolanA
MVDaesU.draob
YLLUFERACot
ehterusaem
nosegatlov
101PJrotcennoc
htiw2&1snip
3nipottcepser
14. Air exchange check
golanAevitcefedaromelborp
ehthtiwmelborpasierehtneht
.melborpladeptoof
ehtsieriwnekorbatahtelbissopsitI
neewtebegatlovehtfI.draobscinortcele
egatlovehtdnaCDV21+si3dna2snip
,noitisopladeptoofehtswollof1nipno erasegatlovehtfI.eludomscitamuenp
sidraobscinortcelegolanAeht,tnesba
ehthtiwmelborpasierehtroevitcefed
tnesbaCDV21+.ecafretniladeptoof
sidraobscinortcelegolanAehtsetacidni
ton1nipnoegatlovehT.evitcefed
aebdluocladeptoofehtgniwollof
noegatlovehtfI
tonseod1nip
ehtwollof
,noitisopladeptoof
ehtmrofrep
nideniltuoskcehc
ladeptooF.11
fI.kcehCecafretnI
eht,ssapeseht
scinortcelegolanA
.evitcefedsidraob
petSerudecorPelanoitaRsetoN
ehtfoytisnetniehT
nidellortnocsisDEL
YTISNETNIDELeht
retlifcibohpordyhallatsnI
otnokcocpotsyaw-3dna
elamegnahcxeriaeht
ehtesolC.rotcennocreul
a.41
riAehtelbane,kcocpots
dna,noitcnufegnahcxE
riAehttahtyfirev
snrutDELegnahcxE
neerg
eraskcehclanretnilareveS
nafI.metsysehtybedam
egnahcxEriAeht,stsixerorre
dnadelbasideblliwnoitcnuf
ehtfI.dereblliwDELeht
riAeht,KOeraskcehclanretni
eblliwDELsutatsegnahcxE
.neerg
riAehtfidraobyekehttcepsuS thgiltonlliwDELegnahcxE morfkcilcyekonsierehtdna
ehT(.yekehtgnisserped
nidelbasidebdluockcilcyek
)unembussecivreSoiduAeht
yalpsidehtfiUPCehttcepsuS
egnahcxEriAehttahtsetacidni
tubderetneneebsahedom
.dehsiugnitxellitssiDELeht
ehtfounembus
ekaM.unemYALPSID
siytisnetniehttahterus
sutatsehteesotyllufpu
anisDELehtfo
eniN.moortilylthgirb
peeknacsrorrelanretni
morfegnahcxEriAeht
riAehtfI.gnitarepo
,DERsiDELegnahcxE
eromroenorofkcehc
gniwollofehtfo
:segassem
ssecxeegnahcxeriA)1ssecxeegnahcxeriA)1
ssecxeegnahcxeriA)1ssecxeegnahcxeriA)1
ssecxeegnahcxeriA)1
eruliafpmupriA)2
eruliafesufpmupriA)3
eruliafrosnesriA)4
mralaerusserpeyE)5
eruliafrosneseyE)6
erusserppmuphgiH)7
erusserppmupriawoL)8
melborpmetsystnegrU)9
Syntec
VitMan
ãã
ã
ãã
page 91
Troubleshooting steps
Page 93
System Test Procedures
petSerudecorPelanoitaRsetoN
mralaeruliafrosneseyEehT
ehtfIylnopetssihtmrofreP
sieruliafrosneseyE
potehtevomeR.deyalpsid
c.41
dnasrevocmottobdna
riAehtnoDELehttcepsni
DELsihT.evlavegnahcxe
.detanimulliebdluohs
tonseodDELevlavehtfI
metsysehtnrut,etanimulli
tcennocsiddnaedisstino
d.41
lortnocscitamuenpeht
ehterusaeM.elbac
snipneewtebecnatsiser
dluohstI.41dna31
.smho052~erusaem
ehtfiylnopetssihtmrofreP
sieruliafesufpmupriA
ehtecalpeR.deyalpsid
e.41
elihW.eludomscitamuenp
,devomersieludomeht
ahtiw301Fesufecalper
.esufA2/1wen
tsujda,neergsiDELehtfI erusserpegnahcxEriAeht
yfireV.gHmm05otlevel
setacidniyalpsidehttaht
f.41
derevileddnatniopteseht
tifI.GHmm05sierusserp
kaelaylekilsiereht,tonsi
snoitcennocrognibutehtni
.metsysehtedisni
onsierehttahtsetacidni
seyeehtnodetcetederusserp
.evlavegnahcxEriAehtfoedis
dluocrosneseyeehthguohT
elbissoposlasiti,deliafevah
nonepootdeliafevlavehttaht
siDELehtfI.dnammoc
golanAehtemussa,detanimulli
fI.deliafsahdraobscinortcele
scinortcelegolanAehtgnignahc
ehttcerroctonseoddraob
eludomscitamuenpeht,melborp
.deliafyllautca
ton/eriwnekorbaebyamerehT
neewteb/rotcennocdetaesylluf
ehtdnaeludomscitamuenpeht
ehtfI.draobscinortcelegolanA
neponasetacidniecnatsiser
nikaerbarehtiesiereht,tiucric
sievlavehtro,elbacevlaveht
eht,esacrehtienI.evitcefed
.evitcefedsieludomscitamuenp
,desolcsitisetacidnitiucricehtfI
ehthtiwmelborpaylekilsiereht
golanAehT.scinortceleevird
ebdluohsdraobscinortcele
.decalper
sahpmupegnahcxEriAehtfI
pmupriaehtesuaclliwti,dezies
tonsiesufsihT.nepootesuf
.decalperebtsumdnaelbatteser
wolasipmupegnahcxEriAehT
lliwwolfssecxE.ecivedemulov
ehtwolebpordottiesuac
.erusserpdeificeps
erudecorp
nonepolliw
.dnammoc
.eruliaf
siereht
etomerasierehT
ehttahtytilibissop
tileblliwDELevlav
liaflliwevlavehttub
sihT.nepoot
gnitoohselbuort
evlavehtsemussa
egnahcxeriaehT
detcennocsievlav
fo41&31snipot
.elbaclortnoceht
etomerasierehT
atahtytilibissop
golanAehtnotrohs
draobscinortcele
dnaruccodluoc
ehtnitluserosla
esufpmupriA
wolfonrednU
teseht,snoitidnoc
lautcadnatniop
ebdluohserusserp
nehW.lacitnedi
,wolfelbaredisnoc
tonlliwpmupeht
.pupeekotelbaeb
Troubleshooting steps
page 92
Syntec
VitMan
ãã
ã
ãã
Page 94
System Test Procedures
petSerudecorPelanoitaRsetoN
egnahcxEriAehT
krowotdengisedsi
A.alunnacaotni
nahtgnineporegral
etelpedlliwag02
foesaelertxenehT
erawtfoslortnoceht
ehtegnahclliw
niebottuodaer
rotomehT.stlov
ebdluohsegatlov
02nahtretaerg
g.41
swolfriatahtyfirevdna
gninepoehtmorf
metsysehtnrut,etanimulli
h.41
lortnocscitamuenpeht
ehterusaeM.elbac
snipneewtebecnatsiser
dluohstI.41dna31
.smho052~erusaem
15. Valve fuse failure checkout
kcocpotsyaw-3ehtnepO
tonseodDELevlavehtfI
tcennocsiddnaedisstino
ehttahtevorplliwnoitcasihT
wollofnacyrtiucriclortnoc
tahtdnaetarwolfnisegnahc
niegakcolblaitraponsiereht
.htapriaeht
stroperyalpsidseulavD/AehT
niegatlovlautcaeht
egatlovrotomehT.lamicedaxeh
fI.000Dnahtretaergebdluohs
,llamsootsiedocdenrutereht
sidraobscinortcelegolanAeht
ehtnisiedocehtfI.evitcefed
scitamuenpeht,deificepsegnar
.evitcefedsieludom
elbaliavaeht
.erusserp
.stlov
petSerudecorPelanoitaRsetoN
ylppusotdesusiesufevlavehT
nosevlavlortnocriaehtotrewop
scitamuenpehttcennocsiD
ehtmorfelbaclortnoc
a.51
.draobscinortcelegolanA
esufevlaVehttahtyfireV
.spotsegassemeruliaf
llA.eludomscitamuenpeht
htiwdeilppuserasevlav
foynafotrohsadnaCDV21+
esufehtnitluserlliwseriweht
spotsegassemehtfI.gninepo
sielbaclortnocehtnehw
ehtfosutatsehT
oslanacesufevlav
gnisudekcehceb
unemSEULAVeht
ESUFehtdna
unembusETATS
ehtnosimelborpeht,devomer
ehtfI.ylbmessascitamuenp
melborpeht,stsisrepegassem
.draobscinortcelegolanAehtno
16. Floppy disk drive checkout
petSerudecorPelanoitaRsetoN
ehtfoenontahtkcehC
nierasegassemgniwollof
:yalpsideht
a.61
detcetorpetirwksiD)1
llufsietteksiD)2
eruliafdaeretteksiD)3
eruliafesufyppolF)4
ãã
ã
Syntec
VitMan
ãã
page 93
Troubleshooting steps
Page 95
System Test Procedures
petSerudecorPelanoitaRsetoN
eruliafesufyppolFehtfI
siereht,sruccoegassem
ehtnimelborparehtie
niroelbacrewopyppolf
b.61
.flestievirdyppolfeht
dnapotehtevomeR
tcepsnidna.srevocmottob
rewopevirdyppolfeht
.elbac
detcetorpetirwksiDehtfI
yfirev,sraeppaegassem
tonsiksidyppolfehttaht
c.61
etirwsitifI.detcetorpetirw
citsalpehtevom,detcetorp
wodniwehttahtosrettuhs
atcelesrodesolcsi
.yppolftnereffid
rollufsietteksiDehtfI
eruliafdaeretteksiD
atceles,ruccosegassem
d.61
nacstnetnocesohwyppolf
yppolfehttamroF.tsoleb
KSIDTAMROFehtgnizilitu
ksidehtniunembus
.unemsecivres
otsliafetteksidehtfI
asierehtrehtie,tamrof
ehthtiwmelborp
e.61
asierehtroscinortcele
yrT.yppolfehtnitcefed
yppolfdnocesagnittamrof
.gniunitnocerofeb
dnapotehtevomeR
tahtyfireV.srevocmottob
lortnocevirdyppolfeht
ehtnidetaesylmrifsielbac
f.61
ehtnodnaevirdyppolf
.draobscinortceleUPC
onsierehttahtoslayfireV
rewopehtotegamad
.elbac
ehtotegamadsierehtsselnU
eht,elbacrewopevirdyppolf
ebotylekilsievirdetteksid
.evitcefed
citsalpgnidilsllamsasierehT
ehtstcetorpetirwhcihwrettuhs
nepoehtnisitinehwksid
.noitisop
sezilitunaMtiVcetnySehT
ehT.tamrofMBIniseippolf
yppolfehtgnittamroffossecorp
dnaaidemehttahtyfirevlliw
gninoitcnuferaevirdetteksid
.yltcerroc
seriuqernaMtiVcetnySehT
dnocesehT.setteksideerf-rorre
siyppolftnereffidahtiwtpmetta
tahtytilibissopehtetanimileot
.melborpylnoehtsawaidemeht
ereht,tamroftonlliwsetteksidfI
ehthtiwmelborparehtiesi
eht,elbaclortnocevirdyppolf
UPCehtro,evirdyppolf
lortnocehtfI.draobscinortcele
eht,ylreporpdetaessielbac
yppolfehtylekiltsomsimelborp .evirdyppolfehtecalpeR.evird
ecalper,sniamermelborpehtfI
.draobscinortceleUPCeht
.titeser
tamrofehT
.tsoleb
siesufyppolfehT
oN.elbatteser
spetslanoitidda
otyrassecenera
lliwnoitarepo
noatadllayortsed
ekaM.yppolfeht
yppolfehterus
onsniatnocnesohc
tonnactahtatad
Troubleshooting steps
page 94
Syntec
VitMan
ãã
ã
ãã
Page 96
System Test Procedures
17. Excess heat check
petSerudecorPelanoitaRsetoN
ehtrofhtapriayramirpehT
retneotriarofsinaMtiVcetnyS
siretlifnafehttahtyfireV
a.71
dnadeggolcrodekcolbton
sitsuahxeecruosthgileht
.dekcolbton
erasnafhtobtahtyfireV
ehtgnivomerybgnitarepo
niamehttcepsnI.revocpot
b.71
ehtnepO.noitatorrofnaf
dnarewardecruosthgil
rofnafblubehttcepsni
.noitator
c.71
esufnaFehttahtyfireV
.yalpsidehtnitonsieruliaf
.taehrevo
eb,retlifnafehthguorht
dnatinuehthguorhtdetalucric
ecruosthgilehttuotsuahxeneht
roegakcolbynA.tcud
ottinuehtesuaclliwnoitcurtsbo
ebtsumti,deliafsahnafrehtiefI
sinafrehtienfI.decalper
esufnaFarofkcehc,gnitarepo
.yalpsidehtnieruliaf
rehtie,deliafsahesufnafehtfI
gniriwnafehtnitrohsasiereht
nafniamehtronafblubehtro
nafniamehT.deziessah
golanAehtotstcennoc
rotcennocaivdraobscinortcele
nafblubehtelihw111PJ
.901PJrotcennocaivstcennoc
ehtwollasrotcennocesehT
ehtsahhcihwnafehtotnoitalosi
ehtecalpeR.melborp
.nafevitcefedetairporppa
siretlifnafehtfI
tievomer,deggolc
.naelcdna
18. Noise Check
petSerudecorPelanoitaRsetoN
otdedivorperaskcolgnippihsehT
mottobdnapotehtevomeR
a.81
owtehttcepsnI.srevoc
mottobehtnoskcolgnippihs
.tinuehtfo
morfeludomscitamuenpehttneverp
eraswercsehtfI.rafootgnivom
eht,ylesoolootroylthgitootdetsujda
etarbivlliweludomscitamuenp
.sissahcehttsniaga
fodetcurtsnocerastnuomrotomehT
ehtnoruoferaerehT.rebburenocilis
stnuomrotomeulbehttcepsnI
b.81
eramehtfoenonerusekamot
.nekorb
,eludomscitamuenpehtfomottob
niamehttnuomotdesueraeerht
,tekcarbtroppusehtotrosserpmoc
ehttnuomotdesueraruofdna
.tekcarbtroppusehtotpmupmuucav
ãã
ã
Syntec
VitMan
ãã
page 95
Troubleshooting steps
Page 97
System Test Procedures
petSerudecorPelanoitaRsetoN
elbacehtfognituorehttcepsnI
ehtraendnarevoseilbmessa
nA.draobscinortcelegolanA
c.81
.elttaranitluser
d.81
e.81
nacelbacnoitacol-fo-tuo
ehtfognitfihsaesuac
dnaylbmessascitamuenp
.erusserpmetsysehtevresbO
D/ATNERRUCehtesU
ehtniyalpsidSEULAV
ehtyalpsidotunemSEULAV
.eulavevirdtnerruc
.ISP55
stlobtnemhcattallatahtyfireV
sihT.denethgitylerucesera
raerdnatnorfehtsedulcni
.swercsgnitnuomlenap
nacevlavfeilerISP05ehtfoeruliafA
atagninnurmetsysehtnitluser
sihT.dedeennahterusserprehgih
eromesuaclliwerusserplanoitidda
tisarosserpmocniamehtninoitom lanoitiddanitluserlliwdnasetarepo
deyalpsideulavehT.tinuehtniesion
dna54neewtebebdluohsevirDsa
Troubleshooting steps
page 96
Syntec
VitMan
ãã
ã
ãã
Page 98

System Adjustment Procedures

Syntec
VitMan
ãã
ã
ãã
page 97
Page 99
System Adjustment Procedures

Adjust Pressure Regulators

1. Remove the top system cover.
2. Turn the VitMan on. Install a cassette. Plug the aspiration port on the cassette.
3. Enter menu mode by pressing the display button. Select Values from the Main Menu. Select Current A/D Values from the Values Menu (see figure 1).
Note: The 3 pressure regulators and 1 pressure relied valve all have locking adjustment
knobs. In order to adjust them, it is necessary to pull the knob away from the regulator body. It is usually necessary to disconnect and reconnect the pressure gage to the regulator output several times (or cause a change in flow load in some other manner) in order to receive a true pressure reading after the regulator has been adjusted. The internal mechanism needs to accommodate to the new adjustment and several read­justments may be needed before the regulator will work at the desired pressure.
4. Adjust the relief valve to obtain a reading of 2585 (2585 mmHg = 50 psi) on the drive sensor. This reading can be taken from the A/D Values screen on the front panel display. The adjustment can be checked by disconnecting and reconnecting the drive sensor tube labeled “4” at the quick disconnect. It should return to approximately the correct value after the tube has been reconnected.
5. Measure the vacuum regulator pilot pressure. It is most easily accessed at the output port of the vac. reg. regulator. Adjust the vac. reg. regulator to 20 psi.
6. Measure the cylinder operating pressure. It is most easily accessed at the tube exiting the pneumatic module labeled “1”. Connect the pressure gage to the tube at the quick disconnect. By pressing the aspiration pinch valve button, on the system front panel, pressure will be alternately supplied to this tube. Adjust the cylinder regulator to 30 psi. Simply causing the system to open and close the aspiration pinch valve (causing pressure to be routed to then vented away from the pressure gage) will change the load enough on the regulator to allow for a stable adjustment.
7. Measure the vitrector operating pressure. It is most easily accessed at the vitrector output on the system front panel. By pressing the manual valve actuation button, located on the side of the vitrector valve, pressure will be supplied to the vitrector output. Adjust the vitrector regulator to 42 psi. Opening and closing the vitrector valve (causing pressure to be routed to then vented away from the pressure gage) with the manual actuation button will change the load enough on the regulator to allow for a stable adjustment.
8. Replace the top system cover.
cylinder regulator
vac. reg. regulator
relief Valve
Adjust Pressure Regulators
vitrector regulator
page 98
Figure 1
Syntec
“Drive” sensor reading
VitMan
ãã
ã
ãã
Page 100
System Adjustment Procedures

Adjust Vacuum Regulator

1. Remove the top system cover.
2. Loosen 2 screws holding the black cover to the vacuum regulator. Remove the vacuum regulator cover. Note the position of the vibration pad on the cover for later replacement.
3. Turn the VitMan on. Install a cassette. Plug the aspiration port on the cassette. Turn the aspiration surgical function on. Adjust maximum aspiration to 500.
4. Enter menu mode by pressing the display button. Select Values from the Main Menu. Select Current A/D Values from the Values Menu (see figure 1).
5. Measure the vacuum regulator pilot pressure. It is most easily accessed at the output port of the vac. reg. regulator located on the regulator manifold in the pneumatic module. The pressure should be set to 20 psi. If the pressure in incorrect, adjust the regulator by turning the knob. Note it should be necessary to unlock the knob by pulling away from the regulator body. It is usually necessary to disconnect and recon­nect the pressure gage to the regulator output several times in order to receive a true pressure reading after the regulator has been adjusted. The internal mechanism needs to accommodate to the new adjustment and several readjustments may be needed before the regulator will work at the desired pressure.
6. Measure the power voltage to the vacuum regulator. The voltage is most easily ac­cessed at the wire connector on the vacuum regulator. Measure between pins 1 and
3. The voltage should be 12Vdc. The 12Vdc supplied to the vacuum regulator is the same as that used for a lot of the sensitive analog circuitry on the analog electronics board. So if the rest of the VitMan seems to be operating, and the 12Vdc is low or missing, suspect a problem with the wires or the connector pins going to the analog electronics board.
7. Fully depress the foot pedal (it helps to put a weight on the foot pedal). The vacuum level in the 50cc chamber will eventually reach 500 (mmHg). Even if the vacuum regulator adjustment is incorrect the system software will try to compensate. The software will slowly adjust the command (voltage) given to the vacuum regulator until the 50cc chamber reads 500. If the system fails to reach 500, first calibrate the foot pedal. This is accomplished by entering the Service Menu from the Main Menu. Select Calibration Routines from the Service Menu. Then select Calibrate Foot Pedal from the Calibration Routines menu. Follow the directions on the screen. If the system still fails to reach 500 the vacuum regulator may be grossly out of adjust­ment, the software may not be able to adjust the voltage high enough to reach 500. If this is the case, adjust the potentiometer marked “S” on the vacuum regulator circuit board clockwise (as viewed from the end with the adjustment) in 1/4 turn increments until the system software is able to reach 500.
8. Measure the control voltage to the vacuum regulator between pins 2 and 3 on the wire connector. The reading should be 3.87Vdc. If the voltage is high, turn the potentiom­eter on the vacuum regulator circuit board marked “S” clockwise. Wait for the system software to return the vacuum to 500. Likewise, if the voltage is low then adjust the “S” pot counter clockwise. After the system has returned to 500, take a new voltage read­ing and adjust as necessary.
Syntec
VitMan
ãã
ã
ãã
page 99
Adjust Vacuum Regulator
Loading...