Throughout this manual, special references are made when deemed important. Three
classifi cations are used to separate these references by their order of importance.
WARNING is used in connection with a procedure or situation that may result in serious injury or death.
CAUTION is used in connection with a procedure or situation that will result in damage to the product.
NOTE is used to emphasize important information.
COPYRIGHT, TRADEMARK AND PATENT NOTICE
This instruction manual is copyrighted and all rights are reserved. It may not, in whole
or in part, be copied, photocopied, reproduced, translated, or reduced to any media
without prior written consent from SUUNTO.
SUUNTO, COBRA, Consumed Bottom Time (CBT), Oxygen Limit Fraction (OLF),
SUUNTO Reduced Gradient Bubble Model (RGBM), Continuous Decompression
and their logos are all registered or unregistered trademarks of SUUNTO. All rights
are reserved.
Patents have been issued or applied for one or several features of this product.
CE
The CE mark is used to mark conformity with the European Union EMC
directive 89/336/EEC. The SUUNTO dive instruments fulfi ll all the required EU
directives.
FIOH, Laajaniityntie 1, FIN-01620 Vantaa, Finland, notifi ed body no.0430, has
EC type-examined this type of personal protective equipment.
EN 250 Respiratory equipment - Open circuit self-contained compressed air diving
apparatus - Requirements, testing, marking.
The tank pressure gauge and the parts of this product used in measuring the tank
pressure meet the requirements set in the section of the European Standard EN 250
that concern tank pressure measurements. The instruments must be serviced by an
authorized dealer every second year or after 200 dives (whichever comes fi rst).
PrEN 13319
PrEN 13319 “Diving accessories - Depth gauges and combined depth and time
measuring devices - Functional and safety requirements, test methods” is a
European diving depth gauge standard draft. The COBRA is designed to comply
with this draft standard.
ISO 9001
SUUNTO Oyj’s Quality Assurance System is certifi ed by Det Norske Veritas to be
according to the ISO 9001 in all SUUNTO Oyj’s operations (Quality Certifi cate No.
96-HEL-AQ-220).
1
WARNINGS
SUUNTO Oyj does not assume any responsibility for losses or claims by third
parties, which may arise through the use of this device.
Due to continuous product development, the COBRA is subject to change without
notice.
WARNING!
READ THIS MANUAL! Carefully read this instruction manual in its entirety
paying close attention to all warnings listed below, including section 1.1. “Safety
Precautions”. Make sure that you fully understand the use, displays and limitations
of the dive computer because any confusion resulting from neglecting to follow
this instruction manual or from improper use of this device may cause a diver to
commit errors that may lead to serious injury or death.
WARNING!
NOT FOR PROFESSIONAL USE! Suunto dive computers are intended for
recreational use only. The demands of commercial or professional diving
may expose the diver to depths and exposures that tend to increase the risk of
decompression illness (DCI). Therefore, Suunto strongly recommends that the
device be not used for commercial or professional diving activity.
WARNING!
ONLY DIVERS TRAINED IN THE PROPER USE OF SCUBA DIVING
EQUIPMENT SHOULD USE A DIVE COMPUTER! No dive computer can
replace the need for proper dive training. Insuffi cient or improper training may
cause diver to commit errors that may lead to serious injury or death.
WARNING!
THERE IS ALWAYS A RISK OF DECOMPRESSION ILLNESS (DCI) FOR ANY
DIVE PROFILE EVEN IF YOU FOLLOW THE DIVE PLAN PRESCRIBED
BY DIVE TABLES OR A DIVE COMPUTER. NO PROCEDURE, DIVE
COMPUTER OR DIVE TABLE WILL PREVENT THE POSSIBILITY OF DCI
OR OXYGEN TOXICITY! An individual’s physiological make up can vary from
day to day. The dive computer cannot account for these variations. You are strongly
advised to remain well within the exposure limits provided by the instrument to
minimize the risk of DCI. As an added measure of safety, you should consult a
physician regarding your fi tness before diving.
2
WARNING!
SUUNTO STRONGLY RECOMMENDS THAT SPORT DIVERS LIMIT THEIR
MAXIMUM DEPTH TO 40 M [130 FT] OR TO THE DEPTH CALCULATED
BY THE COMPUTER BASED ON THE SELECTED O2% AND A MAXIMUM
PO2 OF 1.4 BAR SETTINGS!
WARNING!
DIVES WITH REQUIRED DECOMPRESSION STOPS ARE NOT
RECOMMENDED. YOU SHOULD ASCEND AND BEGIN DECOMPRESSION
IMMEDIATELY WHEN THE DIVE COMPUTER SHOWS YOU THAT A
DECOMPRESSION STOP IS REQUIRED! Note the blinking ASC TIME symbol
and the upward pointing arrow.
WARNING!
USE BACK-UP INSTRUMENTS! Make sure that you use back-up instrumentation
including a depth gauge, submersible pressure gauge, timer or watch, and have
access to decompression tables whenever diving with the dive computer.
WARNING!
PERFORM PRECHECKS! Always activate and check the device before diving
in order to ensure that all Liquid Crystal Display (LCD) segments are completely
displayed, that the device has not run out of battery power, and that the oxygen,
altitude and personal adjustments are correct. Also, exit the Data Transfer mode
before diving, as the computer does not automatically revert to Dive mode from
Data Transfer mode.
WARNING!
YOU ARE ADVISED TO AVOID FLYING ANY TIME THE COMPUTER
COUNTS DOWN THE NO-FLYING TIME. ALWAYS ACTIVATE THE
COMPUTER TO CHECK THE REMAINING NO-FLY TIME PRIOR TO
FLYING! The computer goes into the stand-by display automatically 5 minutes
after the dive has ended. The stand-by display shuts off after two hours. Flying
or traveling to a higher altitude within no-fl y time can greatly increase the risk
of DCI. Review the recommendations given by Diver’s Alert Network (DAN) in
chapter 3.6.3. “Flying After Diving”. There can never be a fl ying after diving rule
that is guaranteed to completely prevent decompression illness!
3
WARNING!
THE DIVE COMPUTER SHOULD NEVER BE TRADED OR SHARED
BETWEEN USERS WHILE IT IS IN OPERATION! Its information will not
apply to someone who has not been wearing it throughout a dive or sequence of
repetitive dives. Its dive profi les must match that of the user. If it is left on the
surface during any dive, it will give inaccurate information for subsequent dives.
No dive computer can take into account dives made without the computer. Thus
any diving activity up to four days prior to initial use of the computer may cause
misleading information and must be avoided.
WARNING!
DO NOT EXPOSE THE COBRA DIVE COMPUTER WITH IT´S STANDARD
HOSE TO ANY GAS MIX CONTAINING MORE THAN 40% OXYGEN!
Enriched air with greater oxygen content presents a risk of fi re or explosion and
serious injury or death.
WARNING!
DO NOT DIVE WITH A CYLINDER OF ENRICHED AIR IF YOU HAVE NOT
PERSONALLY VERIFIED ITS CONTENTS AND ENTERED THE ANALYZED
VALUE INTO YOUR DIVE COMPUTER! Failure to verify cylinder contents and
enter the appropriate O
% into your dive computer will result in incorrect dive
2
planning information.
WARNING!
THE DIVE COMPUTER WILL NOT ACCEPT FRACTIONAL PERCENTAGE
VALUES OF OXYGEN CONCENTRATION. DO NOT ROUND UP
FRACTIONAL PERCENTAGES! For example, 31.8% oxygen should be entered
as 31%. Rounding up will cause nitrogen percentages to be understated and will
affect decompression calculations. If there is a desire to adjust the computer to
provide more conservative calculations, use the personal adjustment feature to affect
decompression calculations or reduce the PO
setting to affect oxygen exposure.
2
WARNING!
SET THE CORRECT ALTITUDE ADJUSTMENT MODE! WHEN DIVING at
altitudes greater than 300 m [1000 ft] the Altitude Adjustment feature must be
correctly selected in order for the computer to calculate the decompression status.
THE DIVE COMPUTER is not intended for use at altitudes greater than 3000
m [10000 ft]. FAILURE TO SELECT the correct Altitude Adjustment setting or
diving above the maximum altitude limit will result in erroneous dive and planning
data.
4
WARNING!
SET THE CORRECT PERSONAL ADJUSTMENT MODE!, Whenever it
is believed that factors that tend to increase the possibility of DCI exist, it is
recommended that you use this option to make the calculations more conservative.
Failure to select the correct Personal Adjustment setting will result in erroneous
dive and planning data.
NOTE!
It is not possible to change between Air, Nitrox and Gauge modes before the
instrument has counted down the no-fl ying time.
There is one exception to this: You can change from Air to Nitrox mode even
during the no-fl ying time. When planning both air and nitrox dives during the
same dive series, you should set the instrument in Nitrox mode and modify the
gas mix accordingly.
In Gauge mode, the no-fl ying time is always 48 hours.
Congratulations on your purchase of the SUUNTO COBRA advanced dive
computer. The Cobra builds on the Suunto tradition of delivering feature-rich
dive computers. The Cobra provides many new and enhanced features that cannot
be found in any other air-integrated dive computer. Push button controls access
a wide selection of choices. The display is optimized for the dive mode chosen.
This air-integrated dive computer is a compact and sophisticated multipurpose dive
instrument, designed to give you years of trouble-free service.
CHOICE OF OPERATING MODES AND SET-UP OPTIONS
User options for the Cobra are selected using the push buttons. Pre Dive
confi guration and setup options include:
• Choice of operating mode - Air / Nitrox / Gauge
• Choice of unit - Metric / Imperial
• Maximum depth alarm
• Dive time alarm
• Backlight settings
• Clock, calendar, daily alarm
• Mix Oxygen fraction % (Nitrox mode only)
• Maximum PO
• Altitude adjustment
• Personal adjustment
(Nitrox mode only)
2
CONTINUOUS DECOMPRESSION WITH SUUNTO RGBM
The Suunto Reduced Gradient Bubble Model (RGBM) utilized in the Cobra predicts
both dissolved and free gas in blood and tissues of divers. It is a signifi cant advance
on the classic Haldane models, which do not predict free gas. The advantage of
Suunto RGBM is additional safety through its ability to adapt to a variety of
situations and dive profi les.
In order to optimize how to respond to different added risk situations an additional
category of stop, referred to as a Mandatory Safety Stop, has been introduced. Also
a countdown for the Recommended Safety Stop is included. The combination of
stop types will depend on the specifi c dive situation.
To get the most from the safety benefi ts be sure to read the summary of the Suunto
Reduced Gradient Bubble Model on page 62.
8
1.1. SAFETY PRECAUTIONS
Do not attempt to use the dive computer without reading this instruction manual
in its entirety, including all the warnings. Make sure that you fully understand the
use, displays and limitations of the instrument. If you have any questions about
the manual or the dive computer, contact your SUUNTO dealer before diving with
the dive computer.
Always remember that YOU ARE RESPONSIBLE FOR YOUR OWN
SAFETY!
When used properly, the dive computer is an outstanding tool for assisting
properly trained, certifi ed divers in planning and executing sport dives. It is NOT
A SUBSTITUTE FOR CERTIFIED SCUBA INSTRUCTION, including training
in the principles of decompression.
Diving with enriched air mixtures (nitrox) exposes the user to risks different from
those associated with diving with standard air. These risks are not obvious and
require training to understand and avoid. Risks include possible serious injury or
death.
Do not attempt to dive with any gas mix other than standard air without fi rst
receiving certifi ed training in this specialty.
1.1.1. EMERGENCY ASCENTS
In the unlikely event that the dive computer malfunctions during a dive, follow
the emergency procedures provided by your certifi ed dive training agency or,
alternatively,
STEP 1: Assess the situation calmly and then move promptly to less than 18 m
[60 ft].
STEP 2: At 18 m [60 ft], slow down your ascent rate to 10 m/min [33 ft/min] and
move to a depth between 3 and 6 meters [10 to 20 ft].
STEP 3: Stay there as long as you assess your air supply will safely allow.
After reaching the surface stay out of the water for at least 24 hours.
1.1.2. DIVE COMPUTER LIMITATIONS
While the dive computer is based on current decompression research and technology,
you must realize that the computer cannot monitor the actual physiological functions
of an individual diver. All decompression schedules currently known to the authors,
including the U.S. Navy Tables, are based on theoretical mathematical models,
which are intended to serve as a guide to reduce the probability of decompression
illness.
9
1.1.3. NITROX
Diving with nitrox provides the diver with an opportunity to reduce the risk of
decompression illness by reducing the nitrogen content in the breathing gas mix.
However, when the gas mix is altered, the oxygen content of the mix is generally
increased. This increase exposes the diver to an oxygen toxicity risk not usually
considered in recreational diving. In order to manage this risk, the dive computer
tracks the time and intensity of the oxygen exposure and provides the diver with
information to adjust the dive plan in order to maintain oxygen exposure within
reasonably safe limits.
In addition to the physiological effects of enriched air on the body there are
operational considerations to be addressed when handling altered breathing mixes.
Elevated concentrations of oxygen present a fi re or explosion hazard. Consult with
the manufacturer of your equipment in regards to its compatibility with nitrox.
10
2. GETTING ACQUAINTED
2.1. FUNCTIONS
You can select the dive computer model between the Regular Air Dive Computer,
Nitrox Dive Computer and Depth Gauge with Timer mode.
The COBRA multipurpose dive computer is an air integrated computer featuring
three dive computer modes (AIR, NITROX, GAUGE), three main operating
modes (TIME/STAND-BY, SURFACE, DIVING), three menu based main modes
(MEMORY, SIMULATION, SET) and 16 - 18 menu based submodes (review
separate Quick Reference Guide). You can scroll through the modes using the push
buttons. The mode indicator at the left side and the mode text at the bottom of the
display indicate the selected mode.
The timekeeping display is the default display of the instrument (Fig. 2.1.). If a
button is not pressed within 5 minutes, the dive computer beeps and returns to the
timekeeping display automatically (except in Diving and Simulation modes). The
timekeeping display shuts off after two hours, but pressing the PLAN or TIME
button activates it.
2.2. CONNECTING THE COBRA TO A REGULATOR
Upon purchasing the Cobra, Suunto strongly recommends you have your dealer
attach it to the fi rst stage of your regulator.
If you decide to attach it yourself, follow the steps outlined below:
1. Remove the high pressure (HP) port plug on the fi rst stage of your
regulator using an appropriate sized wrench.
2. Thread the high pressure hose of the Cobra into the port of your regulator with your fi ngers. Tighten the hose fi tting with a 16 mm [5/8”]
wrench. DO NOT OVERTIGHTEN!
3. Attach the regulator to a charged scuba cylinder. Slowly open the valve.
Check for leaks by submerging the fi rst stage regulator in water. If leaks
are detected, consult your dealer.
11
2.3. PUSH BUTTONS
The dive computer has easy-to-use push buttons and an advisory display, which
guides the user. The SMART (MODE) button is the key to the system. The two
scroll buttons, PLAN and TIME, are used for scrolling up and down the menus
and to show the alternative displays. The dive computer is controlled with these
three push buttons as follows (see Fig. 2.2.).
Press the SMART (MODE) button
• To activate the dive computer.
• To change from the Surface Mode to the menu based modes.
• To select, confi rm or quit a submode (short press).
• To immediately exit any submode to the Surface Mode (long press).
• To activate the electroluminescent backlight (in the Surface Mode hold
down the mode button for more than two (2) seconds, during a dive for
one (1) second).
Press the arrow up scroll (PLAN) button
• To activate the timekeeping display, if the display is blank.
• To activate the Dive Planning in the Surface Mode.
• To make a special bookmark in the profi le memory during a dive.
• To scroll up the options (s, increase).
Press the arrow down scroll (TIME) button
• To activate the timekeeping display, if the display is blank.
• To activate the Time display(s) and/or alternative display(s).
• To scroll down the options (t, decrease).
The dive computer is controlled with the SMART (MODE/On/Backlight/Select/OK/Quit) and the PLAN s and TIME t push buttons and with the water contacts
as follows:
Activation press the SMART (On) button or immerse the instrument in
water for fi ve (5) seconds.
Dive Planning in the Surface Mode, press the PLAN button.
Menu Modes press the SMART (MODE) button.
The display is illuminated by holding down the SMART button for more than
two seconds.
12
2.4. WATER CONTACTS
The water contacts control the automatic activation of
the Dive Mode.
The water and data transfer contacts are located on
bottom of the case (Fig 2.3.). When submerged the water
contacts are connected to the push buttons (which are
the other pole of the water contact) by the conductivity
of the water. The “AC” text (Active Contacts, Fig. 2.4.)
will appear on display. The AC text will be shown until
the water contact deactivates or the dive computer enters
the Dive Mode automatically.
Fig. 2.1. Time keeping display.
Pressing PLAN or TIME button
activates display.
The SMART button
A
Fig 2.3. Depth sensor (A), water/data transfer contacts (B).
B
The PLAN, TIME and
SCROLL buttons
Fig. 2.2. Push buttons of the
dive computer.
Fig. 2.4. Active water contacts
are indicated by the text AC.
13
3. DIVING WITH THE COBRA
m
DIVE
ft
CEILING
S
L
O
MAXAVGPO2
W
ASC TIME
STOP
NO
SURF
O2%
TIME
AIR
OLF
C
B
T
°
C
°
F
bar
psi
OPTIONS
NO DEC TIME
DIVE
OK
SELECT
TIME
QUIT
This section contains instructions on how to operate the
dive computer and interpret its displays. You will fi nd
that this dive computer is easy to use and read. Each
display shows only the data relevant to that specifi c
diving situation.
3.1 BEFORE DIVING
Fig. 3.1. Startup I. All segments
shown.
a
OK
b
c
d
QUIT
Fig. 3.2. Startup II. Battery
power indicator.
3.1.1. ACTIVATION AND PRECHECKS
The dive computer will activate if submerged deeper
than 0.5 m (1.5 ft). However, it is necessary to turn
on the Dive Mode before diving to check the cylinder
pressure, altitude and personal adjustment settings,
battery condition, oxygen settings etc. Press the SMART
(On) button to activate the instrument.
After activation all display elements will turn on showing
mostly fi gure 8’s and graphical elements (Fig. 3.1.). A
few seconds later the battery power indicator is shown
and the backlight and the buzzer are activated (Fig. 3.2.
display a, b, c or d depending on the battery voltage). If
set to Air mode the screen will enter the Surface mode
(Fig. 3.3.). If set to Gauge mode the text GAUGE is
shown (Fig. 3.4.) and if set to Nitrox mode the essential
oxygen parameters are shown with the text NITROX
(Fig. 3.21.) before the Surface mode.
At this time, perform your precheck making sure that:
• the instrument operates in the correct mode
and provides a complete display (Air and
Nitrox modes)
• the low battery indicator is not on
• the altitude and personal adjustment settings
are correct
• the instrument displays correct units of
measurement (Metric/Imperial)
• the instrument displays correct temperature
and depth (0.0 m [0 ft])
• the buzzer beeps
14
• you have enough air for your planned dive. You should also check the
pressure reading against your back-up pressure gauge.
And if set to Nitrox mode, make sure that:
• the oxygen percentage is adjusted according to the measured Nitrox
blend in your cylinder
• the oxygen partial pressure limit is set correctly.
The dive computer is now ready for diving.
3.1.2. BATTERY POWER INDICATOR AND LOW BATTERY
WARNING
This dive computer has a unique graphic Battery Power Indicator designed to give
you an advance notice of impending need to change the battery.
The Battery Power Indicator can always be seen when the Dive Mode is activated.
The electroluminescent backlight will be on during the battery check. The following
Table and Figure show the various warning levels.
TABLE 3.1. BATTERY POWER INDICATOR
15
m
M
DIVE
bar
°
C
TIME
Fig 3.3. Startup III. Surface
mode. Depth and dive time are
zero and cylinder pressure is
300 bar [4350 psi]. Pressing
TIME button activates alternative display of temperature and
current time.
m
DIVE
bar
°
C
TIME
Fig. 3.4. Startup IV. Gauge
mode.
m
Temperature or an internal oxidation of the battery
affects the battery voltage. If the instrument is stored
for a long period, the low battery warning may be
displayed even though the battery has enough capacity.
The low battery warning may also be displayed at low
temperatures, even though the battery has enough
capacity in warmer conditions. In these cases repeat the
battery check procedure.
TIME
After battery check the Low Battery Warning is indicated
by the battery symbol (Fig. 3.5.).
If the battery symbol is displayed in the Surface mode
or if the display is faded or weak, the battery may be too
low to operate the dive computer and battery replacement
is recommended.
NOTE! For safety reasons the backlight cannot be
activated when the low battery warning is indicated
by the battery symbol.
3.1.3. DIVE PLANNING [PLAN]
To enter Planning Mode, press the PLAN button at
any time during Surface Mode. After showing the
TI
text “PLAN” (Fig. 3.6.), the display will show the nodecompression limit for the depth of 9 m [30 ft]. By
pressing the down arrow (t) TIME button, the dive
computer will calculate and show the next deeper nodecompression limits in 3 m [10 ft] increments ending
at 45 m [150 ft]. By pressing the up arrow (s) PLAN
button, the next shallower depth will be shown again.
The Planning Mode is cancelled by pressing the SMART
(QUIT) button.
NOTE! The Planning mode is disabled in Gauge
mode and in Error mode (see section 3.9. “Error
Conditions”).
DIVE
bar
TIME
Fig. 3.5. Low Battery Warning.
Battery symbol indicates that
the battery is low and battery
replacement is recommended.
16
Higher Altitude and conservative Personal Adjustment
Modes will shorten the no-decompression time
limits. These limits at different Altitude and Personal
Adjustment Mode selections are shown in Table 6.1 and
6.2 in section 6.1. “Operating Principles”.
The Planning mode also accounts for the following
information from previous dives:
• any calculated residual nitrogen
• all dive history for the past four days
• oxygen toxicity (Nitrox mode)
The no-decompression times given for different depths
will therefore be shorter than before your fi rst “fresh”
dive.
DIVE NUMBERING SHOWN DURING
DIVE PLANNING
Dives belong to the same repetitive dive series if the
instrument was still counting down the no-fl y time at
the beginning of the dive.
The surface interval must be at least 5 minutes for
a dive to be considered a repetitive dive. Otherwise,
it is considered a continuation of the same dive. The
dive number will not change and the dive time will
continue where it left off (see also section 3.6.2. “Dive
Numbering”).
DIVE
MAX
QUIT
NO DEC TIME
Fig. 3.6. Dive Planning. Planning mode is indicated by
PLAN text. No-decompression
time limit at 30.0 m [100 ft] is
14 minutes in A0/P1 mode.
3.1.4. USER DEFINABLE FUNCTIONS
AND ALARMS
This Cobra has several User Defi nable Functions and
depth and time related alarms that you can set according
to your personal preference.
The dive computer model, unit of measurement and
backlight preferences can be set in the MODE- SETSET PREF submode. The dive time and the depth alarms
can be set in the MODE- SET- SET DIVE submode and
the alarm clock can be set in the MODE- SET- SET
TIME submode. Setting of the user defi nable functions
and alarms are explained in detail in section 4.3. “Set
Modes”.
17
3.2. SAFETY STOPS
Safety stops are widely considered “good diving practice” and are an integral part of
most dive tables. Reasons to perform a safety stop include a reduction in sub clinical
DCI, microbubble reduction, ascent control, and orientation before surfacing.
The Cobra displays two different types of safety stops: Recommended Safety Stop
and Mandatory Safety Stop.
The Safety Stops are indicated by:
• STOP label, when in the depth range 3 m - 6 m [10 ft - 20 ft]
= Recommended Safety Stop Countdown
• STOP + CEILING label, when in the depth range 3 m - 6 m [10 ft - 20
ft] = Mandatory Safety Stop Time display
• STOP label, when deeper than 6 m
= Mandatory Safety Stop scheduled
3.2.1. RECOMMENDED SAFETY STOP
With every dive over 10 meters the instrument has a three minute countdown for the
recommended safety stop, to be taken in the 3 - 6 meter [10 ft - 20 ft] range. This
is shown with the STOP sign and a three-minute countdown in the center window
instead of the no-decompression time (Fig. 3.11.).
The Recommended Safety Stop, as the name implies, is recommended. If it is
ignored, there is no penalty applied to the following surface intervals and dives.
3.2.2. MANDATORY SAFETY STOP
When the ascent rate exceeds 12 meters/min [40 ft] momentarily or 10 meters/min
[33ft] continuously the micro-bubble build-up is predicted to be more than allowed
for in the decompression model. The Suunto RGBM calculation model responds
to this by adding a Mandatory Safety Stop to the dive. The time of this Mandatory
Safety Stop will depend on the severity of the ascent rate excess.
The STOP sign will appear in the display and when you reach the depth zone between
6 m to 3 m [20 ft to 10] also the CEILING label, ceiling depth and the calculated
Safety Stop time appear in the display. You should wait until the Mandatory Safety
Stop warning disappears (Fig. 3.14.).
The Mandatory Safety Stop time always includes the three minute Recommended
Safety Stop time. The total length of the Mandatory Safety Stop time depends on
the seriousness of the ascent rate violation.
You must not ascend shallower than 3 m [10 ft] with the Mandatory Safety Stop
warning on. If you ascend above the Mandatory Safety Stop ceiling, a downward
pointing arrow will appear and a continuous beeping starts (Fig. 3.15.). You should
immediately descend to, or below, the Mandatory Safety Stop ceiling depth. If
18
you correct this situation at any time during that dive, there are no affects on the
decompression calculations for future dives.
If you continue to violate the Mandatory Safety Stop, the tissue calculation model
is affected and the dive computer shortens the available no-decompression time
for your next dive. In this situation, it is recommended to prolong your surface
interval time before your next dive.
19
m
MAX
NO DEC TIME
C
B
T
bar
DIVE
TIME
Fig. 3.7. Dive has just begun
and no air time is shown. First
estimate of remaining air time
will be shown after 30-60
seconds.
m
MAX
MAX
AIR
TIME
C
NO DEC TIME
B
T
bar
DIVE TIME
Fig. 3.8. Diving display. Present
depth is 19.3 m [63 ft] and nodecompression stop time limit
is 23 minutes in A0/P1 mode.
Maximum depth during this
dive was 19.8 m [65 ft], elapsed
dive time is 16 minutes.
m
MAX
MAX
3.3. DIVING IN AIR MODE
3.3.1. BASIC DIVE DATA
The dive computer will remain in the Surface mode at
depths less than 1.2 m [4 feet]. At depths greater than
1.2 m the instrument will go into the Diving mode
(Fig. 3.7.).
All information on the display is labeled (Fig. 3.7.
and 3.8.). During a no-decompression stop dive, the
following information will be displayed:
• your present depth in meters [ft]
• the Remaining Air time
• the Altitude Adjustment setting on the left
side of the center window with a wave and
mountain symbols (A0, A1, or A2) (see Table
3.3.)
• the Personal Adjustment setting on the left
side of the center window with a diver symbol
and + signs (P0, P1, or P2) (see Table 3.4.)
• the maximum depth during this dive in meters
[ft], indicated as MAX
• the cylinder pressure in bar [or psi] displayed
in the lower left corner
• the elapsed dive time in minutes, shown as
DIVE TIME in the lower right corner
• the available no-decompression time in
minutes in the center window as NO DEC
TIME and as a bar graph on the left side of
the display. It is calculated based on the fi ve
factors described in Section 6.1. “Operating
Principles”.
AIR
TIME
C
°
C
NO DEC TIME
B
T
bar
DIVE TIME
TIME
Fig. 3.9. Diving display. Present cylinder pressure is 210 bar [3045 psi]
and remaining air time 41 minutes. Alternative display of current time
and temperature is shown for 5 seconds after pressing TIME button.
20
Alternative displays by pressing the TIME button (Fig.
3.9.):
• the current time, shown as TIME
• the water temperature followed by °C for
Centigrade [or °F for Fahrenheit].
NOTE! In Dive Mode the TIME / TEMPERATURE
display automatically changes back to DIVE TIME
/ CYLINDER PRESSURE after 5 seconds.
3.3.2. BOOKMARK
It is possible to make special marks in the profi le memory
during a dive. These Bookmarks will be shown as a dive
log symbol when scrolling the profi le memory on the
computer display. The Bookmarks will also be shown as
annotations in the PC-software, Suunto Dive Manager.
To make a bookmark on the profi le memory during a
dive press the PLAN button (Fig. 3.10.).
3.3.3. CYLINDER PRESSURE DATA
The air pressure of your scuba cylinder in bars [psi] will
be shown digitally in the lower left corner of the display.
Anytime you enter into a dive, the remaining air time
calculation begins. After 30 - 60 seconds (sometimes
more, depending on your air consumption), the fi rst
estimation of remaining air time will be shown in the left
center window of the display. The calculation is always
based on the actual pressure drop in your cylinder and
will automatically adapt to your cylinder size and current
air consumption.
The change in your air consumption will be based on
constant one second interval pressure measurements over
30 - 60 second periods. An increase in air consumption
will infl uence the remaining air time rapidly, while a
drop in air consumption will increase the air time slowly.
Thus a too optimistic air time estimation, caused by a
temporary drop in air consumption, is avoided.
The remaining air time calculation includes a 35 bar [500
psi] safety reserve. This means that when the instrument
shows the air time to be zero, there is still about 35 bar
[500 psi] pressure left in your cylinder depending on your
m
MAX
AIR
TIME
C
NO DEC TIME
B
T
bar
DIVE TIME
Fig. 3.10. Bookmark activation.
An annotation, Bookmark, is
placed in the profi le memory
during a dive by pressing PLAN
button. Note Logbook symbol.
m
MAX
STOP
AIR
TIME
C
B
T
bar
DIVE TIME
Fig. 3.11. A three minute Recommended Safety Stop.
m
MAX
AIR
TIME
C
NO DEC TIME
B
T
bar
DIVE TIME
Fig. 3.12. Cylinder pressure
warnings. Pressure has dropped
below 50 bar [725 psi]. Pressure
display is blinking and there is
an audible alarm.
21
m
AIR
TIME
C
B
T
bar
STOP
NO DEC TIME
DIVE TIME
S
L
O
MAX
W
Fig. 3.13. Ascent Rate Indicator. Blinking depth reading,
SLOW and four segments are
shown together with an audible
alarm: ascent rate is more than
10 m/min [33 ft/min]. This is
a caution to slow down! STOP
sign means that you are advised
to make a Mandatory Safety
Stop when you reach a depth of
6 m [20 ft].
m
CEILING
STOP
AIR
TIME
C
B
T
bar
DIVE TIME
air consumption rate. With a high consumption rate
the limit will be close to 50 bar [725 psi] and with a
low rate close to 35 bar [500 psi].
NOTE! Filling your BC will affect the air time
calculation, due to the temporary increase in air
consumption.
NOTE! A change of temperature will affect the
cylinder pressure and consequently the air time
calculation.
LOW AIR PRESSURE WARNINGS
The dive computer will warn you with three audible
double beeps and a blinking pressure display when
the cylinder pressure reaches 50 bar [725 psi] (Fig.
3.12.). The next three double beeps are heard when
the cylinder pressure reaches 35 bar [500 psi] and
when the remaining air time reaches zero.
MAX
TIME
°
C
Fig. 3.14. A Mandatory Safety
Stop. You are advised to make
a Mandatory Safety Stop in the
depth zone between 6 m and
3 m [20 ft and 10 ft]. Pressing
TIME button shows alternative
display.
m
CEILING
STOP
AIR
TIME
C
B
T
bar
DIVE TIME
22
Fig. 3.15. Violated Mandatory Safety Stop. Downward
pointing arrow and an audible alarm indicate you should
descend to ceiling zone.
3.3.4. CONSUMED BOTTOM TIME (CBT)
The available no-decompression stop time is also shown visually in the multifunction bar graph on the left side of the display (Fig. 3.7., 3.8. and 3.9.). When
your available no-decompression time decreases below 200 minutes, the fi rst
(lowest) bar graph segment appears. As your body absorbs more nitrogen, more
segments start to appear.
Green Zone - As a safety precaution Suunto recommends you should maintain
the no-decompression bar graph within the green zone. Segments start to appear
when the available no-decompression time decreases below 100, 80, 60, 50, 40,
30 and 20 minutes.
Yellow Zone - As the bars reach the yellow zone, your no-decompression stop time
is less than 10 or 5 minutes and you are getting very close to no-decompression
limits. At this point, you should start your ascent towards the surface.
Red Zone - As all of the bars appear (red zone), your no-decompression stop time
has become zero and your dive has become a decompression stop dive (for more
information see section 3.3.6. “Decompression dives”).
3.3.5. ASCENT RATE INDICATOR
The ascent rate is shown graphically along the right side of the display as
follows:
TABLE 3.2. ASCENT RATE INDICATOR
23
When the maximum allowed ascent rate is exceeded, the fi fth SLOW warning
segment and the STOP sign appear and the depth reading starts to blink, indicating
that the maximum ascent rate has been exceeded continuously or that the current
ascent rate is signifi cantly above the allowed rate.
Whenever the SLOW warning segment and the STOP sign appear (Fig. 3.13.),
you should immediately slow down your ascent. When you reach the depth zone
between 6 m to 3 m [20 ft to 10 ft] the STOP and CEILING depth labels will
advise you to make a Mandatory Safety Stop. Wait until the warning disappears
(Fig. 3.14.). You should not ascend shallower than 3 m [10 ft] with the Mandatory
Safety Stop warning on.
WARNING!
DO NOT EXCEED THE MAXIMUM ASCENT RATE! Rapid ascents increase
the risk of injury. You should always make the Mandatory and Recommended
Safety Stops after you have exceeded the maximum recommended ascent rate.
If this Mandatory Safety Stop is not completed the decompression model will
penalize your next dive(s).
3.3.6. DECOMPRESSION DIVES
When your NO DEC TIME becomes zero, your dive becomes a decompression
stop dive, i.e. you must perform one or several decompression stops on your way
to the surface. The NO DEC TIME on your display will be replaced by an ASC
TIME notation and the maximum depth will be replaced by a CEILING notation
and an upward pointing arrow (Fig. 3.16.).
If you exceed the no-decompression limits on a dive, the dive computer will provide
decompression information required for ascent. After this, the instrument will
continue to provide subsequent interval and repetitive dive information.
Rather than requiring you to make stops at fi xed depths, the dive computer lets you
to decompress within a range of depths (Continuous Decompression).
24
The ascent time (ASC TIME) is the minimum amount of time needed to reach the
surface in a decompression dive. It includes:
• the time needed to ascend to the ceiling at an ascent rate of 10 m/min
[33 ft/min] plus
• the time needed at the ceiling. The ceiling is the shallowest depth to
which you should ascend plus
• the time needed at the Mandatory Safety Stop (if any) plus
• the 3 minute Recommended Safety Stop plus
• the time needed to reach the surface after the ceiling and safety stops
have been completed.
WARNING!
YOUR ACTUAL ASCENT TIME MAY BE LONGER THAN DISPLAYED
BY THE INSTRUMENT! The ascent time will increase if you:
- remain at depth
- ascend slower than 10 m/min [33 ft/min] or
- make your decompression stop deeper than at the ceiling.
These factors will also increase the amount of air required to reach the
surface.
CEILING, CEILING ZONE, FLOOR AND DECOMPRESSION
RANGE
When in decompression, it is important that you understand the meaning of ceiling,
fl oor, and decompression range (Fig. 3.20.):
• The ceiling is the shallowest depth to which you should ascend when in
decompression. At this depth, or below, you must perform all stops.
• The ceiling zone is the optimum decompression stop zone. It is the zone
between the minimum ceiling and 1.8 m [6 ft] below the minimum
ceiling.
• The fl oor is the deepest depth at which the decompression stop time will
not increase. Decompression will start when you pass this depth during
your ascent.
• The decompression range is the depth range between the ceiling and
fl oor. Within this range, decompression takes place. However, it is
important to remember that the decompression will be very slow at, or
close to, the fl oor.
25
m
CEILING
ASC TIME
AIR
AIR
AIR
AIR
AIR
TIME
TIME
TIME
TIME
TIME
C
B
T
bar
DIVE TIME
Fig. 3.16. Decompression dive,
below fl oor. Upward pointing
arrow, blinking ASC TIME
label and an audible alarm tell
you to ascend. Minimum total
ascent time including safety
stop is 7 minutes. Ceiling is at 3
m [10 ft].
m
CEILING
The depth of the ceiling and fl oor will depend on your
dive profi le. The ceiling depth will be fairly shallow when
you enter the decompression mode, but if you remain at
depth, it will move downward and the ascent time will
increase. Likewise, the fl oor and ceiling may change
upwards while you are decompressing.
When conditions are rough, it may be difficult to
maintain a constant depth near the surface. In this case
it will be more manageable to maintain an additional
distance below the ceiling, to make sure that the waves
do not lift you above the ceiling. Suunto recommends
that decompression takes place deeper than 4 m [13 ft],
even if the indicated ceiling is shallower.
NOTE! It will take more time and more air to
decompress below the ceiling than at the ceiling.
ASC TIME
AIR
TIME
C
B
T
bar
DIVE TIME
Fig. 3.17. Decompression dive,
above fl oor. Upward pointing
arrow has disappeared and ASC
TIME label has stopped blinking, which means that you are in
the decompression range.
m
CEILING
ASC TIME
STOP
AIR
TIME
C
B
T
bar
DIVE TIME
°C
26
WARNING!
NEVER ASCEND ABOVE THE CEILING! You
must not ascend above the ceiling during your
decompression. In order to avoid doing so by accident,
you should stay somewhat below the ceiling.
DISPLAY BELOW THE FLOOR
The blinking ASC TIME and an upward pointing arrow
indicate that you are below the fl oor (Fig. 3.16.). You
should start your ascent immediately. The ceiling depth
is shown on the right top corner and the minimum total
ascent time on the right side of the center window.
MAX
Fig. 3.18. Decompression dive, at ceiling zone. Two arrows
point at each other (“hour glass”). You are in the optimum
TIME
ceiling zone at 3.5 m [11 ft] and your minimum ascent time is
5 minutes. Pressing TIME button activates alternative display.
DISPLAY ABOVE THE FLOOR
When you ascend above the fl oor, the ASC TIME display
stops blinking and the upward pointing arrow disappears
(Fig. 3.17.). Decompression will now begin, but is very
slow. You should therefore continue your ascent.
DISPLAY AT THE CEILING ZONE
m
CEILING
ASC TIME
STOP
AIR
TIME
C
B
T
bar
DIVE TIME
When you reach the ceiling zone, the display will show
you two arrows pointing at each other (the “hour glass”
icon, Fig 3.18.). Do not ascend above this zone.
During the decompression stop, ASC TIME will count
down towards zero. When the ceiling moves upwards,
you can ascend to the new ceiling. You may surface
only after the ASC TIME and CEILING labels have
disappeared, which means that the decompression stop
and any Mandatory Safety Stop has been completed. You
are advised, however, to stay until the STOP sign has also
gone. This indicates that the three minute Recommended
Safety Stop has also been completed.
DISPLAY ABOVE THE CEILING
If you ascend above the ceiling during a decompression
stop, a downward pointing arrow will appear and a
continuous beeping starts (Fig. 3.19.). In addition, an
error warning Er reminds you that you have only three
minutes to correct the situation. You must immediately
descend to or below the ceiling.
If you continue to violate the decompression, the dive
computer goes into a permanent Error Mode. In this
mode the instrument can only be used as a depth gauge
and timer. You must not dive again for at least 48 hours
(see also section 3.9. “Error Conditions”).
Fig. 3.19. Decompression dive,
above ceiling. Note downward
pointing arrow, Er warning and
an audible alarm. You should
immediately (within 3 minutes)
descend to or below ceiling.
CEILING
3m / 10ft
6m / 18ft
FLOOR
Fig. 3.20. Ceiling and Floor
zone. Recommended and
Mandatory Safety Stop zone
between 6 m and 3 m [20 ft and
10 ft].
27
m
O2%
OLF
Fig. 3.21. Nitrox display. Maximum depth based on set O2%
(21%) and PO2 (1.4 bar) is 54.1
m [177 ft].
O2%
OLF
Fig. 3.22. Diving in Nitrox
mode. The O
PO2
m
NO DEC TIME
bar
DIVE
TIME
% is set to 32%.
2
MAX
3.4. DIVING IN NITROX MODE
3.4.1. BEFORE DIVING
This dive computer can be set for diving with standard air
only (Air mode) or it can be set for diving with Enriched
Air Nitrox (EANx) (Nitrox mode).
If set to the Nitrox mode, the correct oxygen percentage
of the gas in your cylinder must always be entered into
the computer to ensure correct nitrogen and oxygen
calculations. The dive computer adjusts its mathematical
nitrogen and oxygen calculation models according to
the entered O
on Nitrox result in longer no-decompression times and
shallower maximum depths. When the dive computer
is set in Nitrox mode then the Dive Planning and Dive
Simulation modes both calculate with the O
values that are currently in the computer.
DEFAULT NITROX SETTINGS
In the Nitrox mode, the default setting is for standard
air (21% O
is adjusted to any other percentage of oxygen (22%
- 50%).
If unused, the computer will retain the manually entered
value for the selected oxygen percentage for about two
hours, after which it will revert to the default setting of
21% O
The default setting for maximum oxygen partial pressure
is 1.4 bar, however you are able to set it in the range of
1.2 - 1.6 bar.
% and PO2 values. Calculations based
2
). It remains in this setting until the O2%
2
.
2
% and PO2
2
28
3.4.2. OXYGEN DISPLAYS
If set to Nitrox mode the NITROX display, with all
labeled oxygen information and the label NITROX, is
shown after activation. In Dive Planning mode the nitrox
display shows (Fig. 3.21.):
• the oxygen percentage, labeled with O
shown in the left side of the center window
• the set oxygen partial pressure limit, labeled
with PO
, is shown in the upper right display
2
• the maximum allowed depth based on the set
oxygen percentage and partial pressure limit
• the current oxygen toxicity exposure shown
with an Oxygen Limit Fraction (OLF) bar
graph along the left side of the display
(instead of the CBT).
%, is
2
m
MAX
AIR
TIME
OLF
Fig. 3.23. Diving in Nitrox
mode. After air time is less than
30 minutes O2% is substituted
by Air Time display.
NO DEC TIME
bar
DIVE TIME
In the Dive modes, the oxygen percentage labeled with
O
% and the current oxygen toxicity exposure shown
2
with an Oxygen Limit Fraction (OLF) bar graph are
shown (Fig. 3.22. and Fig. 3.23.). The O
% is shown until
2
the remaining air time is less than 30 minutes. After this,
the remaining air time is displayed in its place. During
a dive, the oxygen partial pressure, labeled with PO
, is
2
also shown instead of the maximum depth in the upper
right display, if the partial pressure is greater than 1.4
bar or the set value (Fig. 3.24.).
By pressing the TIME button during a nitrox dive,
the alternative display appears, which includes (Fig.
3.25.):
• current time
• temperature
• Consumed Bottom Time
• maximum depth (during decompression stop
dive).
• Oxygen percentage when remaining air time
less than 30 min
After fi ve seconds the display will automatically revert
to the original display.
m
O2%
OLF
Fig. 3.24. Oxygen partial pressure and OLF displays. There
is an audible alarm as oxygen
partial pressure is greater than
1.4 bar or set value, and/or the
OLF has reached 80% limit.
PO2
NO DEC TIME
bar
DIVE
TIME
29
m
O2%
C
B
T
°
C
MAX
NO DEC TIME
TIME
Fig. 3.25. Alternative display.
Pressing TIME button displays
current time, maximum depth,
temperature, CBT and O2%,
if Air Time is less than 30
minutes.
m
MAX
NO DEC TIME
AIR
TIME
OLF
bar
DIVE TIME
Fig. 3.26. The lowest bar graph
blinks to indicate that the OLF
value shown relates to OTU.
3.4.3. OXYGEN LIMIT FRACTION (OLF)
In addition to tracking the diver’s exposure to nitrogen,
the instrument tracks the exposure to oxygen, if set to
Nitrox mode. These calculations are treated as entirely
separate functions.
The dive computer calculates separately for Central
Nervous System oxygen toxicity (CNS) and Pulmonary
Oxygen toxicity, the latter measured by the addition of
Oxygen Toxicity Units (OTU). Both fractions are scaled
so that the maximum tolerated exposure for each is
expressed as 100%.
The Oxygen Limit Fraction (OLF) has 11 segments, each
representing 10%. The OLF bar graph displays only the
value of the higher of the two calculations. When the
OTU value meets and exceeds the CNS value then in
addition to displaying its percentage the lowest segment
blinks to indicate that the value shown relates to OTU.
The oxygen toxicity calculations are based on the factors
listed in section 6.3. “Oxygen Exposure”.
30
3.5. DIVING IN GAUGE MODE
If set to Gauge mode, the dive computer can be used
for diving with technical diving mixed gases. If you are
trained for technical diving and you plan to use Gauge
mode on a regular basis, it may be preferable to set the
instrument permanently to its Gauge mode (see section
4.3. “Set Modes”).
If set to Gauge mode the text GAUGE is shown after
activation (Fig. 3.4.).
In the Gauge mode the present depth, maximum depth,
dive time, cylinder pressure, remaining air time and
ascent rate indicator are displayed during the dive (Fig.
3.27.). In addition, temperature and current time are
shown in the alternative display.
NOTE! If you dive with the Gauge mode, it is not
possible to change between the modes within 48
hours.
m
MAX
AIR
TIME
bar
DIVE TIME
Fig. 3.27. Diving in Gauge
mode.
31
m
NO
bar
Fig. 3.28. Surface display.
You have surfaced from a 18
minute dive, which maximum
depth was 20.0 m [66 ft]. The
present depth is 0.0 m [0 ft].
Airplane symbol indicates that
you should not fl y and diver
attention symbol indicates that
you should prolong your surface
interval time because of excess
micro-bubbles.
NO
SURF
TIME
°
C
MAX
DIVE
TIME
m
MAX
TIME
3.6. AT THE SURFACE
3.6.1. SURFACE INTERVAL
An ascent to any depth shallower than 1.2 m [4 ft]
will cause the DIVING display to be replaced by the
SURFACE display, giving the following information
(Fig. 3.28.):
• maximum depth of last dive in meters [ft]
• present depth in meters [ft]
• no-fl ying warning indicated by an airplane
icon
• Altitude Adjustment setting
• Personal Adjustment setting
• Diver attention symbol indicates if you should
prolong your surface interval time
• STOP label for 5 min, if the Mandatory Safety
Stop was violated
• Er, if the decompression ceiling was violated
(= Error Mode) (Fig. 3.31.)
• Cylinder pressure in bars [psi]
• dive time of last dive in minutes, shown as
DIVE TIME.
Fig 3.29. Surface interval, Surface
time display. Pressing TIME button
once will show surface time display.
32
Or when the TIME button is pressed once or twice:
• the current time, shown as TIME instead of the
DIVE TIME
• the current temperature with °C for Centigrade
[or °F for Fahrenheit]
• the surface time in hours and minutes
(separated by a colon), telling the duration of
the present surface interval (Fig. 3.29.)
• the desaturation/no-fl ying time in hours and
minutes is shown next to the airplane in the
center window of the display (Fig. 3.30.).
If set to Nitrox mode, the following information will
also be shown:
• the oxygen percentage labeled with O
% is
2
shown on the left side of the center window
• the current oxygen toxicity exposure shown
with an Oxygen Limit Fraction (OLF) bar
graph along the left side of the display.
m
NO
°
C
MAX
TIME
3.6.2. DIVE NUMBERING
Several repetitive dives are considered to belong to the
same repetitive dive series when the dive computer has
not counted the no-fl ying time to zero. Within each
series, the dives are given individual numbers. The
fi rst dive of the series will be numbered as DIVE 1, the
second as DIVE 2, the third as DIVE 3, etc.
If you start a new dive with less than 5 minutes of
surface interval time, the dive computer interprets this
as a continuation of the previous dive and the dives are
considered to be the same. The diving display will return,
the dive number will remain unchanged, and the dive
time will begin where it left off. After 5 minutes on the
surface, subsequent dives are, by defi nition, repetitive.
The dive counter displayed in the Planning Mode will
increment to the next higher number if another dive is
made.
Fig. 3.30. Surface interval,
no-fl ying time. Pressing TIME
button twice will show no-fl ying
time, indicated by an airplane
symbol.
m
NO
°
C
MAX
DIVE
TIME
Fig. 3.31. Surface Mode after
a violated decompression dive.
Er symbol indicates that you
have violated the ceiling for
more than three minutes. You
must not dive again for at least
48 hours.
NO
Diver Attention Symbol
Violated Decompression
Ceiling
NO
Do Not Fly Symbol
33
3.6.3. FLYING AFTER DIVING
The no-fl ying time is shown in the center window next to the airplane image. Flying
or travelling to a higher altitude should be avoided at any time the computer counts
down the no-fl ying time.
NOTE! The airplane symbol is not shown on the stand-by display. You should
always activate the dive computer and check that the airplane symbol is not
displayed prior to fl ying.
The no-fl ying time is always at least 12 hours or equivalent to the so-called
desaturation time (if longer than 12 hours).
In the permanent Error mode and Gauge mode the no-fl ying time is 48 hours.
Divers Alert Network (DAN) recommends the following on no-fl ying times:
• A minimum surface interval of 12 hours would be required in order to
be reasonably assured a diver will remain symptom free upon ascent to
altitude in a commercial jetliner (altitude up to 2400 m [8000 ft]).
• Divers who plan to make daily, multiple dives for several days, or make
dives that require decompression stops, should take special precautions
and wait for an extended interval beyond 12 hours before fl ight. Further,
the Undersea and Hyperbaric Medical Society (UHMS) suggests divers
using standard air tanks and exhibiting no symptoms of decompression
illness wait 24 hours after their last dive to fl y in an aircraft with
cabin pressure up to 2400 m [8000 ft]. The only two exceptions to this
recommendation are:
• If a diver had less than 2 hours total accumulated dive time in the last 48
hours, then a 12 hour surface interval before fl ying is recommended.
• Following any dive that required a decompression stop, fl ying should be
delayed for at least 24 hours, and if possible, for 48 hours.
• Suunto recommends that fl ying is avoided until all the DAN and UHMS
guidelines and the dive computer wait to fl y conditions are satisfi ed.
34
3.7. AUDIBLE AND VISUAL ALARMS
The dive computer features audible and visual alarms to advise when important
limits are approached or to acknowledge preset alarms.
A short single beep occurs, when:
• the dive computer is activated.
• when the dive computer automatically returns to the TIME mode.
Three double beeps occur, when:
• the cylinder pressure reaches 50 bar [725 psi]. The cylinder pressure
display will start to blink (Fig. 3.12.).
• the cylinder pressure reaches 35 bar [500 psi].
• the calculated remaining air time reaches zero.
Three single beeps with a two second interval and the backlight activated for
5 seconds occur, when:
• the no-decompression dive turns into a decompression stop dive. An
arrow pointing upwards and the blinking ascent warning ASC TIME
will appear (Fig. 3.16.).
Continuous beeps and the backlight activated for 5 seconds occur, when:
• the maximum allowed ascent rate, 10 m/min [33 ft/min], is exceeded.
SLOW and STOP warnings will appear (Fig. 3.13.).
• the Mandatory Safety Stop ceiling is exceeded. A downward pointing
arrow will appear (Fig. 3.15.).
• the decompression ceiling depth is exceeded. An error warning Er and a
downward pointing arrow appear. You should immediately descend to,
or below, the ceiling. The instrument will otherwise enter a permanent
Error Mode within three minutes, indicated by a permanent Er (Fig.
3.19.).
35
You are able to preset alarms before the actual dive. The user programmable alarms
can be set for maximum depth, dive time and time. The alarms activate when:
• The preset maximum depth is reached
- continuous beep series for 24 seconds or until any button is
pressed.
- the maximum depth blinks as long as the present depth value
exceeds the adjusted value.
• The preset dive time is reached
- continuous beep series for 24 seconds or until any button is
pressed.
- the dive time blinks for one minute, if no button is pressed.
• The preset alarm time is reached
- the current time is shown.
- continuous beep series for 24 seconds or until any button is
pressed.
- the current time blinks for one minute, if no button is pressed.
OXYGEN ALARMS IN NITROX MODE
Three double beeps and the backlight activated for 5 seconds occur, when:
• the OLF bar graph reaches 80%. The segments exceeding the 80% limit
start to blink (Fig. 3.24.)
• the OLF bar graph reaches 100%.
The blinking of the segments exceeding 80% will stop, when the OLF is not loading
anymore. At that point the PO
Continuous beeps for 3 minutes and the backlight activated for 5 seconds
occur, when:
• the set oxygen partial pressure limit is exceeded. The maximum depth
is replaced with a current blinking PO
ascend above the PO
NOTE! When the backlight is turned OFF, it does not illuminate when an
alarm is activated.
is less than 0.5 bar.
2
value. You should immediately
depth limit (Fig. 3.24.).
2
2
36
WARNING!
WHEN THE OXYGEN LIMIT FRACTION INDICATES THAT THE
MAXIMUM LIMIT IS REACHED, YOU MUST IMMEDIATELY ASCEND
UNTIL THE WARNING STOPS BLINKING! Failure to take action to reduce
oxygen exposure after the warning is given can rapidly increase the risk of
oxygen toxicity and the risk of injury or death.
3.8. HIGH ALTITUDE DIVES AND PERSONAL
ADJUSTMENT
The dive computer can be adjusted both for diving at altitude and also to increase
the conservatism of the mathematical nitrogen model.
3.8.1. ALTITUDE ADJUSTMENT
When programming the instrument for the correct altitude, you need to select the
correct Altitude Mode according to Table 3.3. The dive computer will adjust its
mathematical model according to the entered altitude mode, giving shorter nodecompression times at higher altitudes (see Section 6.1. “Operating Principles”,
Table 6.1. and 6.2.).
TABLE 3.3. ALTITUDE ADJUSTMENT RANGES
The entered Altitude Adjustment Mode is indicated by mountain symbols (A0, A1
= one mountain, or A2 = two mountains). Section 4.3.1.1. “Altitude Adjustment
and Personal Adjustment Setting” describes how the Altitude Mode is adjusted.
Traveling to a higher elevation can temporarily cause a change in the equilibrium
of dissolved nitrogen in the body. It is recommended that you acclimate to the new
altitude by waiting at least three hours before making a dive.
37
3.8.2. PERSONAL ADJUSTMENT
There are adverse personal factors for DCI which divers can predict in advance
and input into the decompression model. Factors that may affect susceptibility to
decompression illness vary between divers and also for the same diver from one
day to another. The three-step Personal Adjustment Mode is available, if a more
conservative dive plan is desired.
The personal factors which tend to increase the possibility of DCI include, but
are not limited to:
• cold exposure - water temperature less than 20 °C [68 °F]
• the diver is below average physical fi tness level
• diver fatigue
• diver dehydration
• previous history of DCI
• stress
• obesity
The Personal Adjustment Mode is indicated by a diver symbol and plus signs (P0
= a diver, P1 = diver +, or P2 = diver ++). Section 4.3.1.1. “Altitude Adjustment
and Personal Adjustment Setting” describes how the Personal Mode is adjusted.
This feature should be used to adjust the computer to be more conservative,
according to personal preference, by entering the suitable Personal Adjustment
Mode with the help of Table 3.4. In ideal conditions, retain the default setting, P0.
If conditions are more diffi cult or other mentioned factors which tend to increase
the possibility of DCI exist, select P1 or even the most conservative P2. As a
result the dive computer adjusts its mathematical model according to the entered
Personal Adjustment Mode, giving shorter no-decompression times (see section
6.1. “Operating Principles”, Table 6.1 and 6.2).
TABLE 3.4. PERSONAL ADJUSTMENT RANGES
38
3.9. ERROR CONDITIONS
The dive computer has warning indicators that alert the user to react to certain
situations that would signifi cantly increased risk thew of DCI. If you do not respond
to its warnings, the dive computer will enter an Error Mode, indicating that the
risk of DCI has greatly increased. If you understand and operate the dive computer
sensibly, it is very unlikely you will ever put the instrument into the Error Mode.
OMITTED DECOMPRESSION
The Error Mode results from omitted decompression, i.e. when you stay above
the ceiling for more than three minutes. During this three-minute period the Er
warning is shown and the audible alarm beeps. After this, the dive computer will
enter a permanent Error Mode. The instrument will continue to function normally
if you descend below the ceiling within this three-minute period.
Once in the permanent Error Mode only the ER warning is shown in the center
window. The dive computer will not show times for ascent or stops. However, all
the other displays will function as before to provide information for ascent. You
should immediately ascend to a depth of 3 to 6 m [10 to 20 ft] and remain at this
depth until air supply limitations require you to surface.
After surfacing, you should not dive for a minimum of 48 hours. During the
permanent Error Mode, the Er text will be displayed in the center window and the
Planning Mode will be disabled.
39
OPTIONS
4. MENU BASED MODES
To make yourself familiar with the menu based
QUIT
functions, please use your Quick Reference Guide
supplied with the Cobra together with the information
in this chapter.
The main menu based functions are grouped under 1)
memory, 2) dive simulation and 3) setting modes.
Fig. 4.1. Main menu based
Mode options. [3 MODE].
SELECTSELECTSELECT
Fig. 4.2. Memory option. [1
MEMORY].
SELECT
Fig. 4.3. Simulation option. [2
SIMUL].
THE USE OF THE MENU BASED
FUNCTIONS
1. Activate the menu based modes by pressing once
the SMART (MODE) button in the Dive Mode
(Fig. 4.1.).
2. Scroll the mode options by pressing the arrow up/
down buttons. When scrolling the options, the
label and an equivalent number are shown on the
display (Fig. 4.2. - 4.4.).
3. Press the SMART (Select) button once to select
the desired option.
4. Scroll the submode options by pressing the arrow
up/down buttons. When scrolling the options, the
label and an equivalent number are shown on the
display.
5. Select the desired option by pressing once the
SMART (Select) button. Repeat the procedure, if
there are more submodes.
6. Depending on the mode, you are now able to have
a look at the memories, simulate a dive, or make
desired settings (use the the arrow up/down
buttons). The SMART button is used to Quit or to
confi rm the settings (OK).
If you do not press any of the buttons for 5 minutes
while in a Menu based mode, the instrument beeps
and returns to the timekeeping display. In the
Simulation Mode, however, the equivalent time is
60 minutes.
40
EXIT / QUIT!
By pressing the SMART button for more than 1 second,
any menu based function or submode can be quit and the
dive computer will return directly to the Dive Mode.
THE LIST OF THE MENU BASED MODES
1. MEMORIES AND DATA TRANSFER [1
MEMORY]
1. Logbook and Dive Profi le Memory [1 LOGBOOK]
2. Dive History Memory [2 HISTORY]
3. Data Transfer and PC-Interface [3 TR-PC]
2. SIMULATION MODE [2 SIMUL]
1. Dive Simulator [1 SIMDIVE]
2. Dive Planning Simulator [2 SIMPLAN]
3. SET MODES [3 SET]
1. Dive Parameter Settings [1 SETDIVE]
1. Altitude Adjustment and Personal Adjustment Settings [1 AdJ MODE]
2. Dive Time Alarm Setting [2 d ALARM]
3. Maximum Depth Alarm Setting [3 MAX DPTH]
4. Nitrox/Oxygen Settings [4 NITROX]
2. Setting Time [2 SET TIME]
1. Adjusting Time [1 AdJ TIME]
2. Adjusting Date [2 AdJ DATE]
3. Adjusting Daily Alarm [3 T ALARM]
3. Setting Preferences [3 SET PREF]
1. Backlight Setting [1 LIGHT]
2. Dive Computer Units Setting (Metric/Imperial) [2 UNITS]
3. Dive Computer Model Settings
(Air/Nitrox/Gauge) [3 MODEL]
SELECT
Fig. 4.4. Set option. [3 SET].
QUIT
OPTIONS
Fig. 4.5. Memory options. [3
MEMORY].
SELECT
NOTE! The menu based modes cannot be activated
until 5 minutes after the dive.
Fig. 4.6. Logbook option. [1
LOGBOOK].
41
DIVE
TIME
page 1
SELECT
Fig. 4.7. Logbook, page I. Scroll
different pages of specifi c dive.
TIME
page 2
S
L
O
MAX
W
QUIT
DIVE
ASC TIME
STOP
O2%
OLF
°
C
DIVE
Fig. 4.8. Logbook, page II.
Main dive related data.
DIVEDIVE
AVG
SURF
TIME
bar
page 3
QUIT
Fig. 4.9. Logbook, page III. Surface
interval time, average depth and
consumed air indicated by
∆P.
42
4.1. MEMORIES AND DATA
TRANSFER [1 MEMORY]
The memory options (Fig. 4.5.) for this dive computer
include the combined Logbook and Dive Profile
Memory (Fig. 4.6. - 4.12.), Dive History Memory (Fig.
4.13. - 4.14.) and the Data Transfer and PC-Interface
functions (Fig. 4.15.).
The dive entry time and date is registered in the
Logbook memory. Always check before diving that
the time and date are correctly set, especially after
travelling between different time zones.
4.1.1. LOGBOOK AND DIVE PROFILE
MEMORY [1 LOGBOOK]
This instrument has a very sophisticated high capacity
Logbook and Profi le Memory, with data being recorded
every 20 seconds. Dives shorter than the recording
interval are not registered. You have the possibility to
set the recording interval to 10, 20, 30 or 60 seconds
with the optional PC-interface unit and software.
To enter the Logbook Memory Mode select MODE- 1
MEMORY- 1 LOGBOOK.
For each dive there are four pages of logbook dive
information. Use the scroll buttons to scroll the display
between Logbook pages I, II, III and IV. The data of
the most recent dive is shown fi rst.
The logbook dives can be scrolled through displaying
only the fi rst page of each dive, or scrolling through
the 4 different pages of a dive.
When at the fi rst page of a dive use the MODE button
to change the scroll sequence. When the arrow icon is
next to the Mode button the scroll button will scroll
through the fi rst page only of each dive.
When Select is displayed next to the mode button the
scroll buttons will scroll through the four pages of the
selected dive.
The END text is displayed between the oldest and most
recent dive. (Fig. 4.12.)
Note that chronological sequence in the logbook is
determined by the date, not by the dive number.
The following information will be shown on four
pages:
DIVE
QUIT
Page I, main display (Fig. 4.7.)
• dive number in the dive series
• dive entry time and date.
Page II (Fig. 4.8.)
• dive number in the dive series
• maximum depth
(NOTE! Due to lower resolution, the reading
may differ from the maximum depth reading
of the Dive History up to 0.3 m [1 ft].)
• total dive time
• temperature at the maximum depth
• Altitude Adjustment setting (not displayed in
Gauge mode)
• Personal Adjustment setting (not displayed in
Gauge mode)
• SLOW label, if the diver has violated the
maximum ascent rate
• STOP label, if the Mandatory Safety Stop was
violated
• ASC TIME label, if the dive was a
decompression stop dive
• Diver attention symbol, if the dive was started
while symbol was displayed
• downward pointing arrow, if the ceiling was
violated
• oxygen percentage (only in Nitrox mode)
• maximum OLF during the dive (only in Nitrox
mode).
TIMEDIVE
page 4
Fig. 4.10. Logbook, page IV.
Profi le of specifi c dive.
DIVE
TIME
Fig. 4.11. Logbook, page I.
Press SMART (Select) button
once to be able to scroll between
different dives.
SELECT
QUIT
Fig. 4.12. Logbook, end of the
memory. END text is displayed
between oldest and the most
recent dive.
43
Page III (Fig. 4.9.)
• dive number in the dive series
• average depth
• surface interval time before the dive
• ∆P describing Cylinder pressure drop during the dive.
Page IV (Fig. 4.10.)
• dive number in the dive series
• the profi le of the dive, automatic scrolling during which:
• logbook symbol blinks when the user has pressed the PLAN button for
Bookmark
• blinking SLOW label when registered
• blinking ASC TIME label when the dive became a decompression dive.
Press the SMART (Select) button once to change the scroll buttons to scroll the
different dives forward and backward (Fig. 4.11.). Press the SMART (>Select)
button again to change function of the scroll buttons back to scroll the different
pages of the selected dive. When searching the dives, only the page 1 is shown. The
END text is displayed between the oldest and the most recent dive (Fig. 4.12.).
The memory will retain approximately the last 36 hours of dive time. After that,
when new dives are added, the oldest dives are deleted. The contents of the memory
will remain when the battery is changed (assuming that the replacement has been
done according to the instructions).
DIVE PROFILE MEMORY [PROF]
The scrolling of the profi le will start automatically when the Logbook page IV
(PROF) is entered.
With the default setting, the dive profi le is recorded and displayed in 20-second
increments with each display being shown for about three seconds. The depths
shown are the maximum values of each interval.
Pressing any button stops the scrolling of the profi le.
NOTE! Several repetitive dives are considered to belong to the same repetitive
dive series if the no-fl ying time has not ended. See “Dive Numbering” in section
3.6.2. for further information.
44
4.1.2. DIVE HISTORY MEMORY
[2 HISTORY]
The Dive History is a summary of all the dives recorded
by the dive computer. To enter the Dive History Memory
Mode select MODE- 1 MEMORY- 2 HISTORY (Fig.
4.13.).
The following information will be shown on the display
(Fig. 4.14.):
• the maximum depth ever reached
• the total accumulated dive time in hours
• the total number of dives.
The Dive History Memory can hold a maximum of
999 dives and 999 diving hours. When these maximum
values are reached, the counters will start again from
zero.
NOTE! The maximum depth can be reset to 0.0 m [0
ft] using the optional PC-Interface unit and Suunto
Dive Manager software.
4.1.3. DATA TRANSFER AND PCINTERFACE [3 TR-PC]
The instrument can be connected to an IBM compatible
personal computer (PC), using the optional PC-Interface
and software. With the PC-Interface unit, dive data
from the dive computer can be downloaded to a PC.
The PC software can be used for educational and
demonstration purposes, for planning dives, and for
keeping a complete record of your dive history with
the instrument. Additional Logbook data can also be
added. Paper copies of your dive log and profi les can
easily be printed.
The data transfer is carried out using the connector on
the bottom of the unit.
SELECT
Fig. 4.13. Dive History Memory
Mode. [2 HISTORY].
DIVE
MAX
QUIT
TIME
DIVE
Fig. 4.14. Dive History information. Total number of dives, dive
hours and maximum depth.
SELECT
Fig. 4.15. Data Transfer mode.
[3 TR-PC].
45
The following data is transferred to the PC:
• depth profi le of the dive
• dive time
• preceding surface interval time
• dive number
• Altitude and Personal adjustment settings
• oxygen percentage setting and maximum OLF (in Nitrox mode)
• tissue calculation data
• temperature at the beginning of the dive, maximum depth and end of the
dive
• cylinder pressure at the beginning and at the end of the dive (i.e. pressure
drop ∆P during dive)
• surface air consumption.
Using the PC software, you are able to enter setup options such as:
• change the sample rate for profi le records/logbook from the default 20
second interval to 10, 30 or 60 seconds
• input a personal, 30 character fi eld into the Cobra (i.e. your name)
• reset the Dive History’s maximum depth to zero.
It is also possible to manually add comments and other personal information to
the PC based dive data fi les. The PC-Interface package comes complete with the
interface unit, the software and an installation guide.
To enter the Data Transfer mode select MODE- 1 MEMORY- 3 TR - PC (Fig.
4.15.).
NOTE! While in the Data Transfer mode, the connector/water contacts are
used only for the data transfer. The Dive Mode IS NOT automatically activated
if the contacts are submerged.
After you have fi nished the data transfer, press the SMART (Quit) to exit Data
Transfer mode. If a button is not pressed or no data is transferred within 5 minutes,
the instrument beeps and returns to the timekeeping display automatically.
46
4.2. SIMULATION MODE [2 SIMUL]
The Simulation mode can be used to acquaint yourself
with the features and displays of the instrument before
diving, to plan dives in advance, for demonstration or
education purposes, or just for pleasure.
The dive computer has two simulation modes (Fig.
4.16.):
• The DIVE SIMULATOR (Fig. 4.17.)
• The DIVE PLANNING SIMULATOR (Fig.
4.19.).
In Simulation Mode, time moves four times faster than
real time (when diving), i.e., 15s = 1min.
4.2.1. DIVE SIMULATOR [1 SIMDIVE]
The Dive Simulator mode is an excellent tool for
becoming familiar with the dive computer and
planning your dives. Suunto recommends using the
dive simulator to run through different dive scenarios.
The Dive Simulator will allow you to “perform” dive
profi les of your choice and see what the display would
look like during an actual dive. This includes basic dive
information, as well as audible and visual warnings.
Cylinder pressure drop is simulated with a constant,
depth dependant consumption.
To enter the Dive Simulator mode select MODE- 2
SIMUL- 1 SIMDIVE (Fig. 4.17. and 4.18.).
QUIT
OPTIONS
Fig. 4.16. Dive Simulation options. [2 SIMUL].
SELECT
Fig. 4.17. Dive Simulator mode.
[1 SIMDIVE].
m
QUIT
DIVE
bar
Fig. 4.18. Selected Dive Simulator mode. Descend by
pressing the arrow down (TIME) button and ascend by
pressing the arrow up (PLAN) button.
Fig. 4.20. Selected Dive Planning Simulator mode. Add
desired surface interval increment (to present shown surface
interval) by pressing TIME and
PLAN buttons. This display is
only shown for repetitive dives.
4.2.2. DIVE PLANNING SIMULATOR [2
SIMPLAN]
The Dive Planning Simulator mode shows you the
present no-decompression limits. In this mode, you are
also able to add the desired surface interval increment to
present surface interval time, which allows you to plan
your dives in advance.
This mode is also used to add the desired surface
intervals for dive simulations. Add the desired surface
interval increment to present surface interval by
pressing the arrow down (TIME) and arrow up (PLAN)
buttons.
To enter the Dive Planning Simulator mode select
MODE- 2 SIMUL- 2 SIMPLAN (Fig. 4.19.).
NOTE! This display is only shown for repetitive
dives.
NOTE! The Dive Planning Simulator mode is
disabled in Gauge mode and in Error mode (see
section 3.9. “Error Conditions”).
DIVE
NO DEC TIME
Fig. 4.21. Dive Planning
Simulation.
48
MAX
QUIT
4.3. SET MODES [3 SET]
The Set Modes (Fig. 4.22.) are divided into three
submodes for setting the dive-related parameters, timerelated parameters and your personal preferences.
QUIT
4.3.1. DIVE PARAMETER SETTINGS [1
SET DIVE]
To enter the Dive Parameter Setting Mode select MODE3 SET- 1 SET DIVE. The Dive Parameter Setting mode
has two to four options depending on the dive computer
mode. In Gauge mode there are two options, in Air mode
three options and in Nitrox mode four options.
4.3.1.1. ALTITUDE ADJUSTMENT AND
PERSONAL ADJUSTMENT SETTINGS
[1 ADJ MODE]
The current Altitude and Personal Adjustment modes are
shown when diving and at the surface. If the mode does
not match the altitude or personal conditions (see section
3.8. “High Altitude Dives and Personal Adjustment”), it
is imperative that you enter the correct selection before
diving. Use the Altitude Adjustment to select the correct
altitude mode. Use the Personal Adjustment to add an
extra level of conservatism.
To enter the Altitude Adjustment and Personal
Adjustment Setting mode, select MODE- 3 SET- 1
SET DIVE- 1 AdJ MODE. Now you are able to select
from the three altitude modes (Fig. 4.23.) and the three
personal modes (Fig. 4.24.).
Fig. 4.24. Setting Personal
adjustment. Press scroll buttons
to change personal mode.
49
OK
TIMEDIVE
Fig. 4.25. Setting Dive Time
Alarm. Press scroll buttons to
change alarm on/off and to set
dive time value.
m
MAX
OK
Fig. 4.26. Setting Maximum
Depth Alarm. Press scroll buttons to change alarm on/off and
to set maximum depth value.
m
O2%
PO2
OK
4.3.1.2. DIVE TIME ALARM SETTING
[2 D ALARM]
The instrument has one Dive Time Alarm Setting,
which can be used for several purposes to add to your
diving safety. The alarm can be set, for example, to your
planned bottom time.
To enter the Dive Time Alarm Setting Mode select
MODE- 3 SET- 1 SET DIVE- 2 d ALARM. The Dive
Time alarm can be set on or off and the time from 1 to
999 minutes.
4.3.1.3. MAXIMUM DEPTH ALARM
SETTING [3 MAX DPTH]
You can set one depth alarm in the dive computer.
To enter the Maximum Depth Alarm Setting select
MODE- 3 SET- 1 SET DIVE- 3 MAX DEPTH .
The depth alarm is set to 40 m [131 ft] at the factory,
but you are able to adjust it according your personal
preference or switch it off. The depth range can be set
from 3.0 m to 100 m [9 ft to 328 ft] (Fig. 4.26.).
4.3.1.4. NITROX/OXYGEN SETTINGS
[4 NITROX]
If set to the Nitrox mode, the correct oxygen percentage
of the gas in your cylinder must always be entered into
the computer to ensure correct nitrogen and oxygen
calculations. Also, in Nitrox mode, the oxygen partial
pressure limit must be set. When in the Nitrox Setting
mode the equivalent allowed maximum depth based on
the chosen setting will also be displayed.
To enter the Nitrox/Oxygen Setting Mode select MODE3 SET- 1 SET DIVE- 4 NITROX. The default oxygen
percentage (O
pressure (PO
%) setting is 21% (air) and oxygen partial
2
) setting 1.4 bar (Fig. 4.27.).
2
Fig. 4.27. Setting Oxygen Percentage and Partial Pressure. The equivalent
maximum depth is displayed as 32.8 m [107 ft]. Press scroll buttons to change
oxygen percentage and to set oxygen partial setting value.
50
4.3.2. SETTING TIME [2 SET TIME]
To enter the Time Setting Mode select MODE- 3 SET- 2
SET TIME. The Time Setting mode has three options:
1 Time, 2 Date and 3 Daily Time Alarm.
4.3.2.1. ADJUSTING TIME [1 ADJ TIME]
To enter the Time Setting Mode select MODE- 3 SET- 2
SET TIME- 1 AdJ TIME. After entering this mode you
are able to select between 12 h and 24 hour time formats
and set the correct time by using the SMART (MODE)
and scroll buttons (Fig. 4.28.).
4.3.2.2. ADJUSTING DATE [2 ADJ DATE]
To enter the Date Setting Mode select MODE- 3 SET- 2
SET TIME- 2 AdJ DATE. After entering this mode you
are able to set the correct year, month and day in this
order (Fig. 4.29.).
NOTE! The day of the week is automatically
calculated in accordance with the date. The date
can be set within the range of Jan 1, 1990 to Dec.
31, 2089.
OK
Fig. 4.28. Adjusting Time.
OK
4.3.2.3. ADJUSTING DAILY ALARM
[3 T ALARM]
You can set one daily alarm in the dive computer. When
the daily alarm activates, the time symbol blinks for 1
minute and the alarm sounds for 24 seconds. The alarm
is given at the preset time each day. Press any button to
stop the audible alarm, after it has activated.
To enter the Daily Alarm Setting Mode to adjust the
alarm select MODE- 3 SET- 2 SET TIME- 3 T ALARM.
After entering this mode you are able to set the desired
alarm time (Fig. 4.30.).
Fig. 4.29. Adjusting Date.
OK
OK
Fig. 4.30. Adjusting Daily
Alarm.
51
OK
TIME
Fig. 4.31. Setting Backlight On
Time. Press scroll buttons to
change backlight on/off and to
set time value.
4.3.3. SETTING PREFERENCES
[3 SET PREF]
To enter the Setting Preferences Mode select MODE- 3
SET- 3 SET PREF. The Setting Preferences mode has
three options: 1 Light, 2 Units and 3 Model.
4.3.3.1. BACKLIGHT SETTING [1 LIGHT]
In the Backlight Setting mode the backlight can be
turned off or on and the on time can be set to be 5 to
30 seconds (Fig. 4.31.).
To enter the Backlight Setting Mode select MODE- 3
SET- 3 SET PREF- 1 LIGHT.
NOTE! When the backlight turned OFF, it does not
illuminate when an alarm is given.
m
OK
°
C
°
F
Fig. 4.32. Setting Metric/Imperial units.
OK
Fig. 4.33. Setting Dive Computer model.
52
ft
4.3.3.2. DIVE COMPUTER UNITS SETTING
[2 UNITS]
To enter the Dive Computer Units Setting mode, select
MODE- 3 SET- 3 SET PREF- 2 UNITS. This will
enable you to choose between metric and imperial
units (Fig. 4.32.).
4.3.3.3. DIVE COMPUTER MODEL
SETTING [3 MODEL] - AIR/NITROX/
GAUGE
In the Dive Computer Model Setting mode the dive
computer can be set to be an AIR computer, NITROX
computer or a depth GAUGE with a timer (Fig.
4.33.).
To enter the Dive Computer Model Setting mode select
MODE- 3 SET- 3 SET PREF- 3 MODEL.
5. CARE AND MAINTENANCE
This Suunto dive computer is a sophisticated precision instrument. Though designed
to withstand the rigors of scuba diving, you must treat it with proper care and
caution as any other precision instrument.
5.1. IMPORTANT INFORMATION
DEPTH SENSOR
The built-in depth sensor is a sophisticated precision component. Keep the area
around the depth sensor clear from dirt, sand, dust or other substances. Rinse the
device thoroughly with fresh water and dry it with a soft towel. Never use pins,
needles or other objects to clean the depth sensor (Fig. 2.3.).
WATER CONTACTS AND PUSH BUTTONS
Contamination or dirt on the water contacts/connector or push buttons may prevent
the automatic activation of the Dive Mode and cause problems during data transfer.
Therefore, it is important that the water contacts and push buttons are kept clean.
If the water contacts are active (AC-text remains on display) or the Dive Mode
activates on its own, there is probably contamination or invisible marine growth
that may create an unwanted electric current between the contacts. It is important
that the dive computer is carefully washed in fresh water after the day’s diving
is completed. The contacts can be cleaned with fresh water and, if necessary, a
mild detergent and a soft brush. Sometimes it might be necessary to remove the
instrument from the protective boot for cleaning.
53
5.2. CARE OF YOUR DIVE COMPUTER
• NEVER try to open the case of the dive computer.
• Service your dive computer every two years or after 200 dives (whichever
comes fi rst) by an authorized dealer or distributor. This service will include
a general operational check, replacement of the battery, and water resistance
check. The service requires special tools and training. Therefore, it is
advisable to contact an authorized Suunto dealer or distributor for biennial
service. Do not attempt to do any service that you are not sure about how to
do.
• Should moisture appear inside the case or battery compartment, have the
instrument checked immediately by your Suunto dealer or distributor.
• Should you detect scratches, cracks or other such fl aws on the display that
may impair its durability, have it replaced immediately by your Suunto
dealer or distributor.
• Wash and rinse the unit in fresh water after every use.
• Protect the unit from shock, extreme heat, direct sunlight, and chemical
attack. The dive computer cannot withstand the impact of heavy objects
like scuba cylinders, nor chemicals like gasoline, cleaning solvents, aerosol
sprays, adhesive agents, paint, acetone, alcohol etc. Chemical reactions with
such agents will damage seals, case and fi nish.
• Store your dive computer in a dry place when you are not using it.
• The dive computer will display a battery symbol as a warning when the
power gets too low. When this happens, the instrument should not be used
until the battery has been replaced (see also section 3.1.1. “Activation and
Prechecks”).
• Periodically check the hose for cracks or other signs of deterioration.
Replace hose if any damage is found.
54
5.3. MAINTENANCE
The instrument should be thoroughly rinsed with fresh water, then dried with a
soft towel after each dive. Make sure that all salt crystals and sand particles have
been fl ushed out.
Check the display and the transparent battery compartment cover for possible
moisture or water. DO NOT use the dive computer, if you detect any moisture or
water inside.
CAUTION!
- Do not use compressed air to blow water off the unit.
- Do not use solvents or other cleaning fl uids that might cause damage.
- Do not test or use the dive computer in pressurized air.
5.4. WATER RESISTANCE INSPECTION
Ensure the water resistance of the device. Always check the water resistance of
the battery compartment when replacing the battery. Moisture inside the device
or battery compartment will seriously damage the unit.
Check the transparent battery compartment cover and the display for any sign of
leaks. If you fi nd moisture inside your dive computer, there is a leak. A leak must
be corrected without delay, as moisture will seriously damage the unit, even beyond
repair. Suunto does not take any responsibility for damage caused by moisture in
the dive computer, if the instructions of this manual are not carefully followed.
In case of a leak, immediately take the dive computer to an authorized Suunto
dealer or distributor. Only an authorized Suunto dealer or distributor should do
service activities.
5.5. BATTERY REPLACEMENT
NOTE! It is imperative that the change is made in a proper manner to avoid any
leakage of water into the battery compartment or computer. When in doubt, it is
advisable to contact an authorized Suunto dealer for battery replacement.
NOTE! When the battery is changed, all nitrogen and oxygen uptake data is lost.
Therefore, before changing the battery, the no-fl ying time shown by the computer
should have reached zero. Otherwise, wait for 48 hours or preferably even 100
hours before you dive again.
55
CAUTION!
Defects caused by improper battery installation are not covered by the
warranty.
All history and profi le data, as well as the altitude, personal and alarm settings,
will remain in the dive computer memory after the battery change. However, the
clock time and time alarm setting is lost. In the Nitrox mode, the nitrox settings
revert back to default settings (21% O
When working with the battery compartment, cleanliness is extremely important.
Even the smallest dirt particles may cause a leakage when you dive.
, 1.4 bar PO2).
2
BATTERY KIT
The battery kit includes a 3.0 V coin type lithium cell battery and a lubricated Oring. When handling the battery do not make contact with both of the poles at the
same time. Do not touch the surfaces of the battery with your bare fi ngers.
TOOLS REQUIRED
• A Philips screwdriver.
• Soft cloth for cleaning.
• Needlenose pliers or small screwdriver for turning securing ring.
BATTERY REPLACEMENT
The battery and the buzzer are located in the back of the instrument in a separate
compartment. Console and battery compartment parts are shown in Fig. 5.1. and
Fig. 5.2. To change the battery, follow the procedure below:
1. Unscrew the four screws at the back of the elastomer console and remove
the back cover of the console.
2. Thoroughly rinse and dry the computer.
3. Open the securing ring of the battery compartment lid by pushing it down
and rotating it clockwise. You may use a pointed nose pliers or a small
screwdriver as an aid to rotate. Put the pliers ends into the holes of the
securing ring or the screwdriver onto the side of the right tooth on the ring
(Fig. 5.3.) and turn the ring clockwise. Care should be taken not to damage
any of the parts.
4. Remove the ring.
56
5. Carefully remove the lid with the beeper attached to it. The lid can be
removed by pressing with your fi nger on the outermost edge of the lid
while at the same time pulling with your nail at the opposite side. Do not
use sharp metal objects as they might damage the O-ring or the sealing
surfaces.
6. Remove the O-ring and the battery retainer.
7. Carefully remove the battery. Do not damage the electrical contacts or the
sealing surface.
Check for any traces of fl ooding, particularly between the beeper and the lid, or for
any other damage. In case of a leak or any other damage, bring the dive computer
to an authorized Suunto dealer or distributor for check and repair.
8. Check the condition of the O-ring; a defective O-ring may indicate sealing
or other problems. Dispose the old O-ring, even if it seems to be in good
condition.
9. Check that the battery compartment, battery holder and lid are clean. Clean
with soft cloth if necessary.
10. Check the polarity of the battery: the “-” mark should point toward the
bottom of the compartment and the “+” mark upwards. Gently insert the
new battery in the battery compartment
11. Reinstall the battery retainer in its correct position.
12. Check that the new lubricated O-ring is in good condition. Put it in the right
position in the battery compartment. Be very careful not to get any dirt on
the O-ring or its sealing surfaces.
13. Carefully press the lid onto the battery compartment with your thumb,
while making sure that the O-ring is not at any point protruding out at the
edge.
14. Put your other thumb through the locking ring. Press this thumb fi rmly
against the lid and release the other one. Make sure that the lid is pressed
completely down!
15. Note the correct orientation of the locking ring. Turn the locking
ring counterclockwise with your free thumb and fi ngers until it snaps into
its locked position.
57
16. The dive computer should now activate it’s timekeeping mode and show
time 18:00 [6:00 PM] and date SA 01,01. Activate the instrument. Check
that:
• All display segments work.
• The low battery warning is off.
• The buzzer beeps and backlight works.
• All the settings are correct. Reset current time, date and other settings, if
necessary.
17. Reassemble computer into the console. The instrument is now ready for
use.
CAUTION!
Check after the fi rst dives for possible moisture under the transparent battery
compartment lid, indicating a leak.
Cobra console,
end piece
K5606
Cobra display
shield V5850
Cobra console,
front K5607
CB-71 with
SK-7 compass
5517
Cobra case
Console screw
(3,5 x 16mm)
K5609
Cobra console,
back K5608
Fig. 5.1. Console parts. The code after the name stands for
spare part order number.
Securing Ring
V5844
Battery
Compartment Lid
with buzzer
V5843
O-Ring
K5664
Battery
Retainer
V5842
Battery
K5597
Fig. 5.2. Battery compartment parts. The code
after the name stands for
spare part order number.
58
5.6. DISPLAY SHIELD REPLACEMENT
Unscrew the four screws at the back of the console and remove the back cover.
Remove the computer and the display shield. Install new display shield and reinstall
computer and console back.
5.7. COMPASS ATTACHMENT
Unscrew the four screws at the back of the console and remove the back cover.
Replace the console end part with compass. Reassemble console.
Fig. 5.3. Opening of securing ring.
59
6. TECHNICAL DESCRIPTION
6.1. OPERATING PRINCIPLES
NO-DECOMPRESSION LIMITS
The no-decompression limits displayed by the dive computer for the fi rst dive to
a single depth (see Table 6.1. and Table 6.2.), are slightly more conservative than
those permitted by the U.S. Navy tables.
TABLE 6.1. NO-DECOMPRESSION TIME LIMITS (MIN) FOR
VARIOUS DEPTHS (M) FOR THE FIRST DIVE OF A SERIES
The atmospheric pressure is lower at high altitudes than at sea level. After traveling
to a higher altitude, the diver will have additional nitrogen in his body, compared
to the equilibrium situation at the original altitude. This “additional” nitrogen is
released gradually in time and equilibrium is reached. It is recommended that
you acclimate to the new altitude by waiting at least three hours before making
a dive.
Before high altitude diving, the instrument must be set to the Altitude Adjustment
mode to adjust the calculations for the new altitude. The maximum partial pressures
of nitrogen allowed by the mathematical model of the dive computer are reduced
according to the lower ambient pressure.
As a result, the allowed no-decompression stop limits are considerably reduced.
SURFACE INTERVALS
The dive computer requires a minimum surface interval of 5 minutes between
dives. If a surface interval is shorter than 5 minutes, the next dive is treated as a
continuation of the previous dive.
61
6.2. REDUCED GRADIENT BUBBLE MODEL,
SUUNTO RGBM
The Suunto Reduced Gradient Bubble Model (RGBM) is a modern algorithm for
predicting both dissolved and free gas in the tissues and blood of divers. It was
developed in a co-operation between Suunto and Bruce R. Wienke BSc, MSc.
PhD. It is based on both laboratory experiments and diving data, including data
from DAN.
It is a signifi cant advance on the classical Haldane models, which do not predict free
gas (microbubbles). The advantage of Suunto RGBM is additional safety through its
ability to adapt to a wide variety of situations. Suunto RGBM addresses a number
of diving circumstances outside the range of just dissolved gas models by:
• Monitoring continuous multiday diving
• Computing closely spaced repetitive diving
• Reacting to a dive deeper than the previous dive
• Adapting for rapid ascents which produce high micro-bubble (silentbubble) build-up
• Incorporating consistency with real physical laws for gas kinetics.
SUUNTO RGBM ADAPTIVE DECOMPRESSION
The SUUNTO RGBM algorithm adapts its predictions of both the effects of microbubble build-up and adverse dive profi les in the current dive series. It will also
change these calculations according to the personal adjustment you select.
The pattern and speed of decompression at the surface is adjusted according to
micro-bubble infl uence.
Also on repetitive dives adjustment may be applied to the maximum allowable
nitrogen overpressure in each theoretical tissue group.
Depending on circumstances Suunto RGBM will adapt the decompression
obligations by doing any or all of the following:
• Reducing no-decompression stop dive times
• Adding Mandatory Safety Stops
• Increasing decompression stop times
• Advising an extended surface interval (Diver Attention symbol).
62
DIVER ATTENTION SYMBOL - ADVICE TO EXTEND
SURFACE INTERVAL
Some patterns of diving cumulatively add a higher risk of DCI, e.g. dives with
short surface intervals, repetitive dives deeper than earlier ones, multiple ascents,
substantial multiday diving. When this is detected in addition to adapting the
decompression algorithm Suunto RGBM will in some circumstances also advise,
with the Diver Attention Symbol (review page 33) that the diver extend the surface
interval
6.3. OXYGEN EXPOSURE
The oxygen exposure calculations are based on currently accepted exposure time
limit tables and principles. In addition to this the dive computer uses several methods
to conservatively estimate the oxygen exposure. These include for example:
• the displayed oxygen exposure calculations are rounded to the next
higher whole percentage value
• for recreational scuba diving, the recommended upper limit of 1.4 bar
is used as a default
PO
2
• the CNS% limits up to 1.4 bar are based on 1991 NOAA Diving Manual
limits, but the limits higher than 1.4 bar are signifi cantly shortened
• the OTU monitoring is based on the long-term daily tolerance level and
the recovery rate is reduced.
Oxygen related information displayed by the dive computer is also designed to
ensure all warnings and displays in the appropriate phases of a dive. E.g. the
following information will be shown before and during a dive, when the computer
is set in Nitrox mode:
• the selected O
• the color-coded OLF% bar graph for either CNS% or OTU%
• audible alarms are given and the OLF bar graph starts to blink when the
80% and 100% limits are exceeded
• the blinking of the bar graph stops, when the PO
• audible alarms are given and the actual PO2 value blinks when it exceeds
the preset limit
• in dive planning the maximum depth according to the O
maximum PO
• Recording interval: 20 seconds (the recording interval can be set to 10s,
30s or 60s with the optional PC-interface unit and software), records the
maximum depth of each interval.
• Memory capacity: approximately 36 hours of diving with 20 seconds
recording interval.
• Depth accuracy: 0.3 m [1 ft].
Operating Conditions
• Normal altitude range: 0 to 3000 m [10000 ft] above sea level.
• Operating temperature: 0°C to 40°C [32°F to 104°F].
• Storage temperature: -20°C to +50°C [-4°F to +122°F].
It is recommended that the instrument be stored in a dry place at room
temperature.
NOTE! Do not leave the dive computer in direct sunlight!
Tissue Calculation Model
• Suunto RGBM algorithm (developed by SUUNTO and Bruce R.
Wienke, BS, MS and PhD).
• 9 tissue compartments.
• Tissue compartment halftimes: 2.5, 5, 10, 20, 40, 80, 120, 240 and 480
minutes (on gassing). The off gassing halftimes are slowed down.
• Reduced gradient (variable) “M” values based on diving habit and dive
violations. The “M” values are tracked up to 100 hours after a dive.
• The EAN and oxygen exposure calculations are based on
recommendations by R.W. Hamilton, PhD and currently accepted
exposure time limit tables and principles.
65
Battery
• One 3 V lithium battery: CR 2450 (K5597) and O-ring 1,78 mm x 31,47
mm 70 ShA (K5664).
• Battery storage time (shelf life): Up to three years.
• Replacement: Every two years or more often depending on dive activity.
• Life expectancy at 20°C [68°F]:
- 0 dives/y -> 3 years
- 100 dives/y -> 2 years
- 400 dives/y -> 1 year
The following conditions have an affect on the expected battery lifetime:
• The length of the dives.
• The condition in which the unit is operated and stored (e.g. temperature/
cold conditions). Below 10°C [50°F] the expected battery lifetime is
about 50-75% of that at 20°C [68°F].
• The use of the backlight and audible alarms.
• The quality of the battery (some lithium batteries may exhaust
unexpectedly, which cannot be tested in advance).
• The time the dive computer has been stored until it gets to the customer.
The battery is installed into the unit at the factory.
NOTE! Low temperature or an internal oxidation of the battery may cause the
battery warning even though the battery has enough capacity. In this case, the
warning usually disappears when the Dive Mode is activated again.
66
7. WARRANTY
NOTE! The warranty arrangements are different in different countries. The dive
computer packaging contains information regarding the warranty benefi ts and
requirements applicable to your purchase.
This Suunto dive computer is warranted against defects in workmanship and
materials for a period of two years after purchase to the original owner subject to
and in accordance with the terms and conditions set forth below:
The dive computer should be serviced or repaired only by an authorized Suunto
dealer or distributor.
This warranty does not cover damage to the product resulting from improper
usage, improper maintenance, neglect of care, alteration or unauthorized repair.
This warranty will automatically become void if proper preventive maintenance
procedures have not been followed as outlined in the use and care instructions for
this product.
If a claim under this or any other warranty appears to be necessary, return
the product, freight prepaid, to your Suunto dealer or distributor or qualifi ed
repair facility. Include your name and address, proof of purchase and/or service
registration card, as required in your country. The claim will be honored and the
product repaired or replaced at no charge and returned in what your Suunto dealer
or distributor determines a reasonable amount of time, provided all necessary
parts are in stock. All repairs made, not covered under the terms of this warranty,
will be made at the owner’s expense. This warranty is non-transferable from the
original owner.
All implied warranties, including but not limited to the implied warranties of
merchantability and fi tness for a particular purpose, are limited from date of
purchase and in scope to the warranties expressed herein. Suunto shall not be
liable for loss of use of the product or other incidental or consequential costs,
expenses or damage incurred by the purchase. All warranties not stated herein are
expressly disclaimed.
Some states do not allow the exclusion or limitation of implied warranties of
consequential damages, so the above exclusions or limitations may not apply to
you. This warranty gives you specifi c legal rights, and you may also have other
rights that vary from state to state.
This warranty does not cover any representation or warranty made by dealers or
representatives beyond the provisions of this warranty. No dealer or representation
is authorized to make any modifi cations to this warranty or to make any additional
warranty.
Battery replacement is not covered by this warranty.
This Instruction Manual should be kept with your dive computer.
67
8. GLOSSARY
Air Time The amount of dive time remaining, based on a calculation
of cylinder pressure and ambient pressure and present air
consumption.
Altitude diveA dive made at an elevation greater than 300 m [1000 ft] above
sea level.
Ascent rate The speed at which the diver ascends toward the surface.
ASC RATE Abbreviation for ascent rate.
Ascent time The minimum amount of time needed to reach the surface on
a decompression stop dive.
ASC TIME Abbreviation for ascent time.
Ceiling On a decompression stop dive the shallowest depth to which a
diver may ascend based on computer nitrogen load.
Ceiling ZoneOn a decompression stop dive the zone between the ceiling and
the ceiling plus 1.8 m [6 ft]. This depth range is displayed with
the two arrows pointing toward each other (the “hour glass”
icon).
CNS Abbreviation for Central Nervous System toxicity.
Central Nervous System Toxicity
Toxicity caused by oxygen. Can cause a variety of neurological
symptoms. The most important of which is an epileptic-like
convulsion which can cause a diver to drown.
CNS%Central Nervous System toxicity limit fraction. Also note
Oxygen Limit Fraction
Compartment See “Tissue group”.
DAN Divers Alert Network.
DCI Abbreviation for decompression illness.
Decompression Time spent at a decompression stop or range before surfacing,
to allow absorbed nitrogen to escape naturally from tissues.
Decompression range On a decompression stop dive the depth range, between the fl oor
and ceiling, within which a diver must stop for some time
during ascent.
Decompression illness
Any of a variety of maladies resulting either directly or indirectly
from the formation of nitrogen bubbles in tissues or body
fl uids, as a result of inadequately controlled decompression.
Commonly called “bends” or “DCI”.
68
Dive seriesA group of repetitive dives between which the dive computer
indicates some nitrogen loading is present. When nitrogen
loading reaches zero the dive computer deactivates.
Dive timeElapsed time between leaving the surface to descend, and
returning to the surface at the end of a dive.
∆P Delta P, describing the Cylinder Pressure drop during the
dive; difference in tank pressure between beginning and end
of dive.
EAD Abbreviation for equivalent air depth.
EAN Abbreviation for enriched air nitrox.
Enriched Air Nitrox Also called Nitrox or Enriched Air = EANx. Air that has some
oxygen added. Standard mixes are EAN32 (NOAA Nitrox I =
NN I) and EAN36 (NOAA Nitrox II = NN II).
Equivalent Air Depth Nitrogen partial pressure equivalent table.
Floor The deepest depth during a decompression stop dive at which
decompression takes place.
Half-timeAfter a change in ambient pressure, the amount of time required
for the partial pressure of nitrogen in a theoretical compartment
to go half-way from its previous value to saturation at the new
ambient pressure.
HP Abbreviation for high pressure (= cylinder pressure).
Multi-level dive A single or repetitive dive that includes time spent at various
depths and whose no-decompression limits are therefore not
determined solely by the maximum depth attained.
NITROXIn sports diving refers to any mix with a higher fraction of
oxygen than standard air.
NOAAUnited States National Oceanic and Atmospheric
Administration.
No-decompression time
The maximum amount of time a diver may remain at a particular
depth without having to make decompression stops during the
subsequent ascent.
No-decompression dive
Any dive which permits a direct, uninterrupted ascent to the
surface at any time.
NO DEC TIME Abbreviation for no-decompression time limit.
OEA = EAN = EANx Abbreviation for Oxygen Enriched Air Nitrox.
OLF Abbreviation for Oxygen Limit Fraction.
69
Oxygen Tolerance Unit
Is used to measure the Whole-Body-Toxicity.
Oxygen Limit Fraction
A term used by SUUNTO for the values displayed in the
Oxygen Toxicity bargraph. The value is either the CNS% or
the OTU%.
O
% Oxygen percentage or oxygen fraction in the breathing gas.
2
Standard air has 21% oxygen.
Oxygen partial pressure
Limits the maximum depth to which the used Nitrox mixture
can safely be used. The maximum partial pressure limit for
enriched air diving is 1.4 bar. The contingency partial pressure
limit is 1.6 bar. Dives beyond this limit have the risk for
immediate oxygen toxicity.
PO
Abbreviation for oxygen partial pressure.
2
RGBM Abbreviation for Reduced Gradient Bubble Model.
Reduced Gradient Bubble Model
Modern algorithm for tracking both dissolved and free gas in
divers.
Repetitive diveAny dive whose decompression time limits are affected by
residual nitrogen absorbed during previous dives.
Residual nitrogen The amount of excess nitrogen remaining in a diver after one
or more dives.
SURF TIME Abbreviation for surface interval time.
Surface interval time Elapsed time between surfacing from a dive and beginning a
descent for the subsequent repetitive dive.
Tissue groupTheoretical concept used to model bodily tissues for the
construction of decompression tables or calculations.
Whole-Body Toxicity Another form of oxygen toxicity, which is caused by prolonged
exposure to high oxygen partial pressures. The most common
symptoms are irritation in the lungs, a burning sensation in the
chest, coughing and reduction of the vital capacity. Also called
Pulmonary Oxygen Toxicity. See also OTU.
70
71
DISPOSAL OF THE DEVICE
Please dispose of the device in an appropriate way, treating it
as electronic waste. Do not throw it in the garbage. If you wish,
you may return the device to your nearest Suunto dealer.
72
MENU BASED
MAIN MODE
MENU BASED
SUBMODE 1
MENU BASED
SUBMODE 2
MODES AND
OPERATIONS
Backlight
MAX
MAX
NO DEC TIME
m
TIME
AIR
DIVING MODE
1.2 M
DIVE MODE
MODE
Backlight (>2 s)
m
LCD &
battery
check
ON
TIME-KEEPING AND STAND-BY MODESURFACE MODE
Time & Alternative Display(s)
DIVE TIME
bar
T
B
C
Bookmark in
the profile memory
TIME
DIVE
Time & Alternative Display(s)
Plan
QUIT
display
3 MODE OPTIONS
bar
nitrox /
gauge
Activate the timekeeping display
QUIT
3 SET1 MEMORY
SELECTSELECT
3 MEMORY OPTIONS
SELECT
2 SIMUL
2 SIMULation OPTIONS
3 SET OPTIONS
QUIT
SELECT
QUIT
SELECT
2 HISTORY
SELECT
3 TR - PC1 LOGBOOK
dive simulation:
surface mode
QUIT
Add surface
interval minutes
Add surface
interval hours
2 SIMPLAN1 SIMDIVE
SELECTSELECT
DIVE
1 SET
QUIT
TIME
2 SET
SELECTSELECT
3 SET
PREFerences
page 1
Dive n-1
page 1
Dive n-2
Dive n
page 1
QUIT
END
SELECT
QUIT
page 2
QUIT
page 3
Dive n
page 1
QUIT
profile
page 4
history
data
transfer
QUIT
simulate diving
ascenddescend
OK
OK
QUIT
SIMulate PLANning
increase depthdecrease depth
SELECT
Altitude
Personal
P0, P1, P2
A0, A1, A2
OK
1 Adj MODE
SELECT
2 d ALARM
QUIT
3 / 4 SET DIVE OPTIONS
3 MAX DPTH
SELECTSELECT
OK
PO2
O2%
4 NITROX
OK
SELECT
hours
24h / AM / PM
OK
year
QUIT
SELECT
OK
2 Adj DATE
3 SET TIME OPTIONS
hours
on / off
3 T ALARM1 Adj TIME1 Adj TIME
SELECTSELECT
OK
OK
OK
SELECT
on time
backlight
1 LIGHT
QUIT
SELECT
2 UNITS
2 / 3 SET PREF OPTIONS
AIR
SELECT
NITROX
3 MODEL
OK
on / off
on / off
min
OK
month
OK
OK
minutes
OK
METRIC
IMPERIAL
GAUGE
OK
DT minutes
alarm depth
OKOK
OK
OK
day
OK
OK
COPYRIGHT
This publication and its contents are proprietary to Suunto Oy.
Suunto, Wristop Computer, Suunto Cobra, Replacing Luck and their logos
are registered or unregistered trademarks of Suunto Oy. All rights reserved.
While we have taken great care to ensure that information contained
in this documentation is both comprehensive and accurate,
no warranty of accuracy is expressed or implied.
Its content is subject to change at any time without notice.
Printed in Finland 09.2006 / Esa Print Oy / 161253