Super System MKP Installation Manual

0831106 Rev.0
Calibration procedure
MKP PKP MKC PKC
Calibration procedure for MKP - PKP
1 GENERAL GUIDELINES FOR CALIBRATION
For an accurate calibration it is necessary to proceed as follows: a) - The instrument under calibration should be mounted in its case in
order to keep the internal temperature stable.
b) - The ambient temperature should be stable.
Avoid any drift due to air-conditioning or others. c) - The relative humidity should not exceed 70%. d) - The instrument must be in ON condition from 20 minutes at least. e) - Operate, possibly, in an environment with no electromagnetic
disturbances. f) - During calibration, connect to the instrument one input at a time. g) - Before to execute each calibration, be sure that the specific
hardware setting has been made (see "Preliminary hardware
setting" paragraph). For this calibration procedure it is necessary to use calibrators with
the following accuracy and resolution:
ACCURACY
1) For current input:
2) For voltage input : + 0.005% output + 0.001% range + 5 µV
3) For TC input: + 0.005% output + 0.001% range + 5 µV
4) For RTD input:
5) For cold junction compensation: better than 0.1 °C
RESOLUTION
1) For current input: 0.5 µA
2) For voltage input: 100 µV
3) For TC input: 1 µV
4) For RTD input: 10 m
5) For cold junction compensation: better than 0.1 °C
2 PRELIMINARY HARDWARE SETTINGS
+ 0.025% output + 0.0025% range + 0.01 µA
+ 0.02 % + 0.0025 /decade.
D
C
D
Fig. 2
2.2 MAIN INPUT SELECTION
Set J103 (see fig. 3) according to the desired input type as shown in the following table.
J103 INPUT TYPE
T/C, RTD,CJ 60 mV 5 V 10 V 20 mA 1-2 open open close open open 3-4 open open close close open 5-6 open open open open close 7-8 open open open open close 5-7 close close open close open 6-8 close close open open open
1 3 5 7
J102
2 4 6 8
1 3 5 7
J103
2 4 6 8
CPU card
2.1 HOW TO REMOVE THE INSTRUMENT FROM ITS CASE
1) Switch off the instrument.
2) Push gently the lock A on the right.
3)While the lock A is maintained out, slide out the right side of the
instrument (see fig. 1).
B
A
B
Fig. 1
4) Push gently the lock C on the left.
5)While the lock C is maintained out, slide out the instrument
(see fig. 2).
J205
ON DIP
V301
1 2 3 4
Fig. 3
2.3 AUXILIARY INPUT SELECTION (option)
Set J102 (see fig. 3) according to the desired input type as shown in the following table.
J102 INPUT TYPE
5 V 10 V 20 mA 1-2 close open open 3-4 close close open 5-6 open open close 7-8 open open close 5-7 open close open 6-8 open open open
2
Calibration procedure for MKP - PKP
2.4 IN CT / FEEDBACK SELECTION
This instrument can use the "IN CT" input or the "Feedback" input; the two inputs are not contemporarily. The current transformer input allows you to measure and display the current running in a load driven by a time proportional control output during the ON and OFF periods of the output cycle time. By this feature it is also available the "Out failure detection" function (see page 66 in the USER MANUAL). The feedback input is used when the servomotor close loop or the servomotor open loop with valve position indication outputs is required. To select the desired input type, set V301 (see fig. 3) as detailed in the following table:
Input V301.1 V301.2 V301.3 V301.4 IN CT ON OFF ON ON Feedback OFF ON OFF ON
2.5 CALIBRATION MODE SELECTION
To start the calibration procedure, the DIP SWITCH V101, mounted on CPU card, must be set as follows: V101.1 = Not care condition V101.2 = OFF V101.3 = Not care condition V101.4 = ON NOTE: during calibration procedure the serial communication interface will be disabled.
ON DIP
1 2 3 4
V101
3 CALIBRATION PROCEDURE
3.1 FOREWORD
Calibration parameters are divided in groups. Each group is comprised of two parameters (initial and full scale values) plus a specific calibration check.
Follows a complete list of the "calibrations groups".
1) ñ.I.tc = Main input calibration, TC input
2) ñ.I.CJ = Main input calibration, cold junction
3) ñ.I.rt = Main input calibration, RTD input
4) ñ.I.ñA = Main input calibration, mA input
5) ñ.I.5 = Main input calibration, 5 V input
6) ñ.I.10 = Main input calibration, 10 V input
7) A.I.ñA = Auxiliary input calibration, mA input
8) A.I.5 = Auxiliary input calibration, 5 V input
9) A.I.10 = Auxiliary input calibration, 10 V input
10) In.Ct = Current transformer input calibration
11) FEEd = Feedback input
12) 05.ñA = Out 5 calibration
13) 06.ñA = Out 6 calibration
14) CAL = Default calibration data loading NOTE: Calibration groups from group 7 to group 13 will be shown only when the specific hardware is fitted.
GENERAL NOTE ABOUT CALIBRATION PROCEDURE: During calibration procedure, when the initial or full scale value of a group is selected and the middle display shows "OFF", pushing the FUNC pushbutton the instrument will jump to the next parameter or check without to modify the previous calibration setting. In this way it is possible to recalibrate only the desired input or output. It is also possibe to make a check of one or more calibration group without to remake the specific calibration.
Fig. 4 When it is desired to exit from calibration mode proceed as follows:
1) switch off the instrument;
2) remove the instrument from its case;
3) select the desired operative mode by setting the DIP switch V101 as described in "Operative mode and hardware lock" paragraph reported at page 23 of the USER MANUAL.
3.2 CALIBRATION PROCEDURE
HOW TO PROCEED
Switch on the instrument, the upper display will show CAL while the lower display will show the firmware version. Push the FUNC pushbutton to display the first calibration group on the upper display. Depress FUNC pushbutton more times until the desired calibration group is reached.
1) "ñ.I.tc" - MAIN INPUT CALIBRATION - TC INPUT
The upper display will show "ñ.I.tc".
1.1) "Lr" - INITIAL SCALE VALUE
The lower display will show "Lr" a)Made the specific hardware setting as described at paragraph 2. b)Connect the instrument under test to the calibrator as shown in
Fig. 5.
_
3
+
1
Fig.5 c) The upper display will show "ñ.I.tc", the lower display will show
"Lr" while "OFF" will appear on the middle display. d)Set calibrator to 0.000 mV. e)Push pushbutton, the middle display will change to "On". f) After few seconds, start calibration by pushing FUNC
pushbutton. At the end of this calibration routine, the instrument
will go to the next step.
3
Loading...
+ 7 hidden pages