HYDROGEN SENSOR
(PART NUMBER 20624)
Please read, understand, and follow these instructions before operating this equipment. Super Systems Inc. is not
responsible for damages incurred due to a failure to comply with these instructions. If at any time there are questions
regarding the proper use of this device, please contact us at (800) 666-4330 for assistance.
7205 Edington Drive
Cincinnati, OH 45249, USA
Tel. (513) 772-0060
Fax (513) 772-9466
Manual #4583A Hydrogen Sensor Operations Manual Page #1 of 10
Hydrogen Sensor (P/N 20624) Operations Manual
Description This sensor is designed to accurately measure Hydrogen through thermal conductivity technology. It is capable of providing additional computations based on the Hydrogen measurement, and it has multiple methods of digital and analog communications capabilities.
Features
Measures H2 from 0 to 100% with 0.01% resolution
Calculates NH
, DA, and Kn for nitrider applications
3
Two isolated 4 to 20ma analog outputs
One RS-232 port with Modbus RTU or a simple ASCII protocol
One RS485 port with Modbus RTU protocol
Wide power supply input range (9 to 30VDC)
Small physical size
Specifications
Hydrogen measurement: Range: 0 to 100% Accuracy: +/- 0.01% Repeatability: +/- 0.01% Resolution: +/- 0.01%
Calculated Variables for simple nitrider applications: %NH3 0 to 100% %DA (dissociated ammonia) 0 to 100% Kn nitriding potential
Analog Outputs: Two 4-20mA outputs common supply and isolated from other electronics Output variables: %H2, %NH3, %DA, and Kn on either output Adjustable range of PV: zero and span Resolution: 0.005 mA Accuracy: +/- 0.01% of range Linearity: +/- 0.01% Minimum load resistance: 0 Ohms Maximum load resistance: 500 Ohms
Sample flow rate:
1.5-2 cfh
RS-232 Serial Communications: Protocols: Modbus RTU or a simple ASCII Baud rates: 9600, 19200, or 38400 Format: 8 bits No parity, 1 stop bit, No handshaking Connection: DB-9P
RS-485 Serial Communications: Protocol: Modbus RTU Baud rates: 9600, 19200, or 38400. Format: 8 bits No parity, 1 stop bit, No handshaking
Manual #4583A Hydrogen Sensor Operations Manual Page #2 of 10
Power requirement: 9 to 30 volts DC @ 2 watts
Temperature and Humidity Electronics Operating: 0 to 50 °C, RH 0 to 90% non-condensing Sample gas: 0 to 70 °C, RH 0 to 90% non-condensing Storage: -20 to 70 °C, RH 0 to 90% non-condensing
Dimensions: See figure 1
Weight: 350g
Material: Stainless steel sensor housing