sensors
DPL4000
Portable low range dew point system for accurate
and reliable verification of gases using current
sensor technology.
LOW RANGE DEW POINT
MEASUREMENT
TECHNICAL FEATURES
Optimal dew point range
-112 to -22 F with +/- 3.6 F accuracy
-80 to -30 C with +/- 2 C accuracy
Operating temperature
-40 to +140 F
-40 to + 60 C
Pressure range: 0 to 725 psi
Alarm indication
Sensor threads: ‰ NPT
Sensor housing: IP65 NEMA4 (AISI316L)
o
o
o
o
oo
o
o
o
Determine very low dew points with high accuracy
Excellent long term stability
Rapid response time with fast stabilizing dew point reading
Withstands condensation
Includes hand held instrument, sensor, sensor housing, flow meter,
quick connect fittings, logging software, and protective carrying case
Typical DP Gas Measurements
Hydrogen
o
Argon
Nitrogen
OTHER FEATURES
Durable handheld
Easy to read LCD display
Units: Dew point ( F or C), Frost point,
ppm (by weight), ppb (by voume)
Process variable trending and data logging
Battery powered
Rechargeable battery included
Battery run time: 12 hours
Sensor Quick Disconnect
Includes two male quick disconnect fittings
USB Cable and Utility Software
Quickly retrieve data from handheld device
INNOVATIVE SOLUTIONS WORLDWIDE
S01-07200927-001
o
o
SuperSystem
incorporated
s
Special Features
Quick Response
Easy to Use
Sensor Purge
This function evaporates all excess
moisture molecules from the sensor to
significantly increase the response time
when the sensor goes from an ambient
condition to a dry condition. A Sensor
Purge will be initiated automatically when
the sensor detects a large wet-to-dry
environmental change or upon startup
and after every 24 hours of continuous
use.
Specifications
This device is used to measure very low dew points with high accuracy and
excellent long term stability. It can be used across a wide range of
temperatures and pressures. It measures dew point in a variety of noncorrosive gases including Nitrogen, Argon, and Hydrogen.
Total Measurement Range: -148F to +68F (-100C to +20C)
Optimal Measurement Range: -112F to -22F (-80C to -30C)
Accuracy (Within Optimal Range): +/- 3.6F (+/- 2C)
Accuracy (Outside Optimal Range): +/- 5.4F (+/- 3C)
Operating Temperature Range: -40F to +140F (-40C to +60C)
Operating Pressure Range: 0-725 psi / 0-50 bar
Sensor Housing: IP65 - NEMA4 (AISI316L)
Power: Rechargeable Battery (included) or 4 "AA" Batteries
Battery Recharging Time: Approximately 4 hours
Run Time on Battery Power: Approximately 12 hours (depends on application)
Output Units: Dew Point /Frost Point, ppm (by weight), ppb (by volume)
SuperSystems
incorporated
Utility
Software
7205 Edington Drive
Cincinnati, OH 45249
513.772.0060 phone
513.772.9466 fax
www.supersystems.com 800.666.4330