Step 1 – Install Configurator 2.0 on the Local Computer .......................................................................................... 5
Step 2 – Connect the SSi DAQ to a Network or Local Computer ................................................................................ 7
Step 3 – Configure Configurator 2.0 on the local computer ....................................................................................... 8
Step 4 – Complete Configurator Range Setups Menu Option ..................................................................................... 9
Step 5 – Complete Configurator Input Offsets Menu Option.................................................................................... 10
Analog Inputs ........................................................................................................................................................ 11
Adding a Jumper to an Input ............................................................................................................................ 11
Voltage connections ........................................................................................................................................ 11
4 – 20 mA. Current Loop connections ............................................................................................................... 12
Setting the DIP switches to Assign Board Numbers ............................................................................................ 12
Configurator Menu ................................................................................................................................................. 13
Aux Instruments ................................................................................................................................................. 14
Communication Setup ......................................................................................................................................... 15
Range Setups ..................................................................................................................................................... 18
Aux Instrument Setup ......................................................................................................................................... 20
DAQ Data Sheets ................................................................................................................................................... 25
Appendix 1 – Input Number Table ........................................................................................................................... 27
Revision History ..................................................................................................................................................... 29
4576 DAQ Manual Revision A Page 2 Super Systems Inc
OOvveerrvviieeww
The SSi DAQ data acquisition module is a simple, powerful and expandable product for measuring a vast array of signals
in the process control environment. The DAQ can be used with as few as 3 inputs and can be expanded to as many
as 48 per DAQ depending upon your needs. Each SSi fully-isolated 3 input analog module can be configured on an inputby-input basis for a range of voltage inputs from as little as 20 mV to as much as 10V or for a 4 - 20 mA current signal.
Communications to the DAQ are available via Modbus serial protocol or through the DAQ's Ethernet port via
ModbusTCP. Calibration and configuration are available through SSi's Configurator software (provided) and the DAQ
also features a limited web server for simple interfacing from any PC in your plant (Ethernet communications required).
4576 DAQ Manual Revision A Page 3 Super Systems Inc
SSSSii DDAAQQ WWiirriinngg DDiiaaggrraamm
RS232
SSi
CINCINNATI, OHIO
RS485
HOST
-
S
+
+
24V
DC
REMOTE
SSi
CINCINNATI, OHIO
+
24V
DC
REMOTE
RS485
RS485
HOST
-
S
+
ETHERNET
3
2
-
S
1
+
-
6
+
-
5
+
-
4
+
12
3
2
1
-
6
+
-
5
+
-
4
+
12
+
+
+
-
9
+
-
8
+
-
7
+
+
+
+
3
SSi
CINCINNATI, OHIO
+
24V
DC
REMOTE
RS485
3
2
-
S
1
+
-
6
+
-
5
+
-
4
+
12
+
+
+
4576 DAQ Manual Revision A Page 4 Super Systems Inc
SSSSii DDAAQQ SSeettuupp
This section will explain how to set the SSi DAQ up. The steps required to set up the SSi DAQ through a local computer
are:
1. Install Configurator 2.0 on the local computer
2. Connect the SSi DAQ to a network or local computer
3. Configure Configurator 2.0 on the local computer
4. Complete Configurator
5. Complete Configurator
Range Setups
Input Offsets
menu option
menu option
Step 1 – Install Configurator 2.0 on the Local Computer
Configurator 2.0 is a configuration utility developed by SSi
that will allow the user to interface with an SSi instrument
directly or over a network. The installation file,
ConfiguratorSetup.msi, should be included with the
installation CD
provided by SSi. If this file is not on the CD, contact Super
Systems at 513-772-0060.
Double-click on the installation file to begin the installation
process.
The first page displayed is just for information purposes.
Click on the Next > button to move to the next page, or
press the Cancel button to cancel the installation.
The second page is a warning about Configurator 2.0. Since
Configurator 2.0 is a Microsoft .Net 2.0 product, the local
computer will have to have the .Net 2.0 framework installed
before Configurator 2.0 can be used. Click on the Next >
button to continue or the Cancel button to cancel the
installation.
Page 3 will allow the user to select the location of the
installation. The default location is “C:\SSi\”. To change this
location, click on the Browse button and select a new location
from the dialog box that is displayed. The Disk Usage button
is a utility that will display the available hard drive space on the
local computer. Click on the Next > button to move to the
next page.
4576 DAQ Manual Revision A Page 5 Super Systems Inc
Page 4 will allow the user to review the installation
settings, if necessary. Click on the Install button to install
the software.
Page 5 will display a progress bar as the installation proceeds.
Note: The installation should only take a few minutes
4576 DAQ Manual Revision A Page 6 Super Systems Inc
.
Page 6 is the finishing screen, which is displayed after the
software has been installed. Click on the Next > button to
continue.
Page 7 is the informational screen about the makers of the
installation software. Click on the Finish button to close out
this screen.
Step 2 – Connect the SSi DAQ to a Network or Local Computer
To connect the instrument to the network, through a wall port or switch, use a regular Ethernet cable. To connect the
instrument directly to a computer, use an Ethernet crossover cable. Contact your IT Department for the necessary
cables. Once the DAQ is connected to a network, the Configurator 2.0 software will be able to find it during any
searches. Connecting the instrument to your network or directly to a PC is accomplished using the Ethernet port on the
instrument. If you are connecting the instrument to your network, you will need an Ethernet cable. The cable is plugged
into the instrument Ethernet plug and then other end should be plugged into a network hub. If the IP Address of the
instrument needs to be changed, this can be done through the Configurator software (see Step 3 below). If you are not
putting the instrument on the network, you
should use an Ethernet crossover cable.
Ethernet crossover cables are most often
used when connecting two Ethernet
computers without a hub. An Ethernet
crossover cable has its send and receive
wires crossed. When using a hub or switch,
this is automatically done for you. With a
crossover cable, you are forming a network
between the computer that you are directly
plugged into and the SSi DAQ. There will be
some network settings on the computer that
you will have to configure for the 2 devices to
communicate. The DAQ will have the network
setting already setup with a default IP address – normally supplied by
the customer. This can be modified through the Configurator software.
Network settings can be found through the
Windows. By selecting
list of the current available connection types. Using the crossover
cable will require the “Local
Area Connection” as seen in the diagram to be modified. The
Properties can be changed by highlighting the connection and using
the right mouse button to click and select the
highlighting the connection and clicking on Change setting of this
connection. Once the
displayed, highlight the Internet Protocol (TCP/IP) option. Click the
Properties button to display Internet Protocol
(TCP/IP) Properties. On the Internet Protocol (TCP/IP) Properties tab, you will need to select the option for Use The
Following IP Address.
4576 DAQ Manual Revision A Page 7 Super Systems Inc
Network Setting
Local Area Connection Properties
Control Panel
, the operator will be given a
Properties
in Microsoft
tab or by
screen is
Enter in the following data on these fields:
IP Address: 192.168.0.209
Subnet Mask: 255.255.255.0
Default Gateway: 192.168.1.1
Note: These fields are suggestions. Contact your IT department to get
a valid IP address, Net Mask, and Gateway for the local computer.
To change the network settings on your computer you may need
addition information so please refer to the computer manual.
Step 3 – Configure Configurator 2.0 on the local computer
When Configurator starts up for the first time, the user will see the main screen, which will be blank because no
instruments have been set up yet. The first step is to set up an instrument in Configurator.
First, the user will need to log in with
administrative rights.
be logged in with at least administrative rights;
Supervisor rights will not allow the user to add an
instrument
Configurator are: operator, supervisor,
administrator, SSi Special. The lock on the
toolbar will let the user know what level is
currently logged in. Operators are blue,
supervisors are gold, administrators are green,
and the SSi Special, which is used for
configuration purposes before the unit is shipped,
is red. Click on the lock and log in with the
following information: username = administrator,
password = 2.
. The four levels of rights in
administrator passwords can be changed on the
Furnace Setup
be green. Click on the
Settings
Management
Click on the Add button to display the rest of the screen. First, give the instrument a name. The name can be anything
the user wants, but it is suggested that the user makes the name descriptive. Next, select the model from the drop-down
list. For the DAQ, the model number is “9021”. Next, enter the IP address in the “Ethernet” section and make sure the
“Ethernet” option is selected.
screen.
Note: The SSi DAQ is shipped with an IP address supplied by the customer
menu page
. This will display the
the local computer is hooked up directly to the instrument through a crossover cable, the IP address will still need to be
correct
the “Ethernet” IP address box. This will set up Configurator for Ethernet communications. To set it up for serial or
SuperData communications, the proper option will need to be selected and filled out.
. The user can also scan the network to find all available SSi instruments by clicking on the search button next to
Note: The user will need to
Note: The supervisor and
. The lock should now
Options
menu, then select
System
.
Note: even if
4576 DAQ Manual Revision A Page 8 Super Systems Inc
The DAQ has RS-232 host communications and RS-485 host communications capability (shown below). To set the DAQ
up for serial communications, the user will need to know which
COM port will be used. Select that COM port (COM 1 – COM
20 is available). The Baud rate can be left at the default –
19200. The address should be changed to 1. If the user is
using RS-232 communications, or if the DAQ is the only device
on RS-485 communications, then the address of 250 will be
sufficient.
Click on the Save button to save the information. Click on the
Done button to close down this screen.
The device can now be selected from the drop-down list on the
Configurator main screen.
Step 4 – Complete
Configurator Range
Setups Menu Option
Note: This menu item is also located
later in the manual under the
Configurator Menu
Range Setups
The
allow the user to view/modify the
selected input ranges for each of the
three inputs on a board. The user
will also be able to view/modify the
input linearization for each input per
board. There are a maximum of
sixteen boards.
The Input Type can be one of the
following:
B T
C 2.5 V
E 1.25 V
J 160 mV
K 80 mV
N 40 mV
NNM 20 mV
4576 DAQ Manual Revision A Page 9 Super Systems Inc
section
menu option will
.
Loading...
+ 20 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.