Sunrise Medical 900-0880 User Manual

Installation & Instruction Manual
Guardian
®
Easytrack System
Bath Base Kit Assembly
Bath Base Kit
2
Introduction
TABLE OF CONTENTS
Introduction .......................................................................................... 2
Important Safeguards.............................................................................. 3
Requirements ......................................................................................... 4
Kit Contents & Part Numbers .................................................................... 5
Assembly of Bath Base Kit ....................................................................... 5
Bath Base Kit Specifications..................................................................... 8
Maintenance .......................................................................................... 8
Easytrack Products & Accessories .............................................................. 10
Warranty ............................................................................................... 11
This manual contains important safety and maintenance instructions. Please read it carefully before using your Easytrack System and refer to it as often as needed for safe and efficient use.
If you have questions regarding the safe use and/or assembly, maintenance or specifications of your Easytrack System, you should call Customer Service at 800-333-4000, 303-218-4600 or from Canada at 800-263-3390.
For service and repair, remember your authorized Sunrise supplier is able to provide the assistance you need.
Save this manual for future reference.
SUNRISE LISTENS
Thank you for choosing a Guardian Easytrack System. We want to hear your questions or comments about this manual, the safety and reliability of your Easytrack System and the service you receive from your Sunrise Medical Supplier. Please feel free to write or call us at the address and telephone number below:
Sunrise Medical
Customer Service Department
7477 East Dry Creek Parkway
Longmont, Colorado 80503
(303) 218-4600 or (800) 333-4000
FOR ANSWERS TO YOUR QUESTIONS
Your authorized supplier knows your Easytrack System best and can answer most of your questions about its safety, use and maintenance. For future reference, fill in the following:
Supplier: ___________________________________________________________________
Address:____________________________________________________________________
Telephone: __________________________________________________________________
Serial #: ____________________________________________________________________
Date Purchased: _____________________________________________________________
Bath Base Kit
IMPORTANT SAFEGUARDS
Do not operate this equipment without first reading and understanding this manual.
Should you be unable to understand the warnings, cautions and instructions, contact our technical support or your local dealer before attempting to use this equipment– otherwise injury or damage may result.
Important information is highlighted by these terms:
WARNING– Notices as used in the manual apply to hazards or unsafe practices which could result in serious bodily harm.
CAUTION– Notices as used in this manual apply to hazards or unsafe practices, which could result in minor personal injury or property
damage.
NOTES Highlight procedures and contain information which assist the operator in understanding the information contained in this
manual.
NOTE– The information contained in this document is subject to change without notice.
SAVE THESE INSTRUCTIONS AND KEEP WITH EASYTRACK AT ALL TIMES.
The Easytrack System and its accessories are intended to be used as an assistive device for transferring a person between the posts. The Easytrack system is only to be used with the Guardian Voyager Portable lift.
USE ONLY AS INTENDED - This product may cause serious injury if not installed and operated according to the instruction manual.
Failure to observe warnings listed below could result in serious injury.
Before You Install the Easytrack
• Carefully inspect product upon arrival. If any pieces are missing or appear damaged, do not install. Contact your local supplier
for further instructions.
• If you are unsure of the strength or stability of your floor and/or ceiling, do not install the Easytrack System - consult you
local supplier.
• Easytrack can not be used with any type of suspended ceiling, including suspended drywall, acoustic tile, etc.
• Installation on a plaster ceiling may cause cracking. Sunrise Medical cannot be held responsible for damage.
• Do not install on a ceiling with water damage.
• Do not over-tighten ceiling plate as this may cause ceiling to crack.
• If flooring is not according to local building code standards, flooring may be damaged.
• Do not install on a sloped floor.
• Do not drop product; may cause breakage.
• Do not completely submerge in water.
• Do not attempt to over-extend the rail past the locking point.
• Do not install product if neoprene on top or bottom plate appears to be damaged or absent.
When the Easytrack is in Use
• Do not hit posts - this might cause unit to become unstable. If posts are hit, reassemble Easytrack System to ensure the
posts are straight.
• If you use a power wheel chair, use extreme care. Hitting posts may cause unit to become unstable. If posts are hit, reinstall
the Easytrack System to ensure the posts are straight.
• Do not lean against posts.
• Do not use the Easytrack System as a swing.
• Refrain from swinging user sideways of rail.
Important Safeguards
3
Bath Base Kit
4
Requirements
REQUIREMENTS
Failure to meet requirements listed below could result in serious bodily harm.
Ceiling Requirements
• Install to a rigid ceiling (must have a framework above it).
• Must have a maximum of 24" (610mm) between the trusses in the ceiling.
• Ceiling must be made of wood or 1/2"(12.7mm) sheet rock or thicker.
• Ensure ceiling and floor are clear of any dust, grease, water or any foreign substance.
A build-up of nicotine, grease or water on the ceiling will affect the integrity of the Easytrack System.
• Installing a post directly under a joist or truss is not required, but preferable.
Floor Requirements
• Ensure flooring meets local building code standards. (Termites or water damage may affect the structure of the floor).
• Ensure floor is clear of any dust, grease, water or any foreign substance.
• Floor must not have a slope of more than 0.5 °. See drawing below to determine the floor slope. Use a level or contact a contractor if unsure of floor slope.
Bath Requirements
• Ensure your tub is made of porcelain or metal and has sufficient structure underneath the tub to withstand the weight required (see drawing below). If you are unsure of the structural integrity of the tub, contact the manu­facturer or a contractor to evaluate the structure.
• Ensure you have 3" of ledge space available on each end of the tub (see drawing). The round "foot" on the bath base must entirely be in contact with the tub ledge.
• The tub length must be between 53”–65” (1350-1650 mm).
• Any protrusion from the wall next to the tub must be removed to provide space for the post.
• A molded plastic tub with a protruding wall may obstruct the post or base if there is not a straight line between the corners of the tub.
• Tub must be clean of any soap or dust and must be dry before placing the tub base onto the tub. DO NOT ATTEMPT TO INSTALL THE BATH BASE WITHOUT THOROUGHLY CLEANING AND DRYING THE SURFACE OF THE TUB.
• Floor beneath the tub must be according to local building code standards. (Water damage or termites may affects the integrity of the floor or wall structure.)
• If you have any doubt with respect to the ceiling or bath structure, consider installing the Easytrack System with a wall mount. The wall mount must be installed by a licensed contractor.
• Ensure posts are straight using the level provided; failure to do so may cause injury.
• Every month a maintenance check should be done to ensure all posts are level. (See "Maintenance" section in this manual).
• Make sure the padded surface of the top plate and foot plate are kept clean and are intact.
• Protect the Easytrack System and its accessories during transport.
24" max
1/2" min.
Joists
Joists
Suspended wire or rods
Sheet rock or tile
Sheet rock
X 12" (300mm) 24" (600mm) 36" (900mm) 48" (1200mm)
Y 0.10" (2.6mm) 0.21" (5.2mm) 0.31" (7.9mm) 0.42" (10.5mm)
Wall
3" (75mm)
Supporting Structure
1.5" min
Ledge of Tub
(1350 – 1650mm)
53" – 65"
Loading...