11.0 Hot Mirror ...................................................................................................................................49
12.0 Motor ..........................................................................................................................................51
Please read this manual and follow its instructions carefully.
e words warning, caution, and note carry special
meanings and should be carefully reviewed:
Warning Warnings indicate risks to the safety
of the patient or user. Failure to follow
warnings may result in injury to the
patient or user.
4. Ensure that readjustments, modications, and/
or repairs are carried out exclusively by Stryker
Endoscopy or other authorized personnel.
5. Ensure that the electrical installation of the relevant
operating room complies with the applicable IEC,
CEC, and NEC requirements.
Federal law (United States of America) restricts this device
to use by, or on order of, a physician.
Caution Cautions indicate risks to the
equipment. Failure to follow cautions
may result in product damage.
Note Notes provide special information to
clarify instructions or present additional
useful information.
An exclamation mark within a triangle
is intended to alert the user of special
warnings or important operating and
maintenance instructions in the manual.
A lightning bolt within a triangle is
intended to warn of the presence of
hazardous voltage. Refer all service to
authorized personnel.
e incorrect use of any of the required tools and/or
techniques may risk damage to the equipment or injury to
the person carrying out the procedure, subsequent operators,
or patient. Perform the repair ONLY if you have been
specically trained in the use of all pertinent equipment
and techniques. Stryker Endoscopy cannot continue to
guarantee compliance to UL, CSA, TUV, or other labeled
safety standards if service is performed by anyone other than
Stryker Endoscopy personnel.
To help avoid potential serious injury to the user and the
patient and/or damage to this device:
1. Read this manual thoroughly, especially the
warnings, and be familiar with its contents prior to
using this equipment.
2. Never sterilize any part of the X8000 console.
3. Disconnect the X8000 from the electrical outlet
when inspecting the fuses.
IMPORTANT SAFETY NOTICE
Before operating or performing any maintenance or
repairs on the X8000 Light Source, read the user guide
and service guide thoroughly and carefully. When using
any light source, such as the X8000, re and/or severe
injury may result to the patient, user or inanimate
objects, if the instructions in this manual are not
followed.
All light sources, including the X8000, can generate
signicant amounts of heat at the scope tip, the scope
light post, the light cable tip, and/or near the light cable
adapter. Higher levels of brightness from the light source
result in higher levels of heat.
To avoid the risk of burns and/or re:
• Always adjust the brightness level of the camera
and the monitor before adjusting the brightness
level of the light source.
• Adjust the brightness level of the light source to
the minimum brightness necessary to adequately
illuminate the surgical site.
• Adjust the internal shutter of the camera higher
in order to run the light source at a lower
intensity.
• Avoid touching the scope tip or the light cable
tip to the patient, and never place them on top
of the patient, as doing so may result in burns to
the patient or user.
• Never place the scope tip, the scope light post,
the light cable adapter, or the light cable tip on
the surgical drapes or other ammable material,
as doing so may result in re.
• Always place the light source in Standby mode
whenever the scope is removed from the light
cable or the device is unattended. e scope
tip, scope light post, light cable adapter, and
light cable tip will take several minutes to cool
o aer being placed in Standby mode, and
therefore may still result in re or burns to the
patient, user, or inanimate objects.
4
Introduction
Required Skills, Tools, and Components
is manual is intended to be used as a service guide for
Stryker repair technicians in the installation, maintenance,
and repair of the Stryker X8000 Light Source. It is meant
to be used in conjunction with the X8000 Light Source
User Manual (1000-400-885) and does not replace existing
documentation.
Service Options
Factory Service
Due to the complexity of the X8000 Light Source, Stryker
recommends that any malfunctioning system be returned
to Stryker Endoscopy for repair or replacement, where
specialized equipment and technicians are available to
perform repairs while maintaining full product quality and
safety. If service is needed either during or aer the warranty
period:
1. Contact Stryker Endoscopy at 1-800-624-4422,
or contact your local Stryker Endoscopy sales
representative.
2. Package all the components carefully in the original
shipping container, if possible.
3.
Ship the X8000 Light Source, prepaid and insured, to:
Stryker Endoscopy Customer Service
Attention: Repair Department
5900 Optical Court
San Jose, California 95138
You may request a loaner unit during the repair period.
On-Site Service
Stryker Endoscopy accepts responsibility for the eects on
safety, reliability, and performance of the equipment only if
readjustments, modications, and repairs have been carried
out exclusively by a person specically authorized by Stryker
Endoscopy to do so.
On-site repair should be carried out only by qualied
technicians with the proper test equipment listed in this
manual, so that the safety of operators and patients is not
compromised.
In no event shall Stryker Endoscopy be liable for incidental
or consequential damages in connection with or arising from
the performance or use of its products aer unauthorized
modication or repair.
e repair procedures described in this manual require a
basic set of skills, tools, and replacement components.
Skills
Stryker recommends diagnostic and repair procedures be
performed by authorized, qualied technicians with training
or experience in the following:
• Basic electronics
techniques
• Multimeter operation
• Oscilloscope operation
• Vectorscope operation
• BioTek Safety Analyzer
operation
To o l s
Most of the procedures described in this manual can be
performed using a basic tool kit that includes the items listed
below.
Basic toolsAdvanced tools
• Phillips screwdriver
• Flathead screwdriver
• 8˝ adjustable wrench
• Needlenose pliers
• Color video monitor
• Glass fuse puller
• Multimeter
• Oscilloscope
(20 Mhz or higher)
• NTSC/Pal Vectorscope
• BioTek Model 601
PRO Safety Analyzer
or equivalent current
leakage tester
Components
Replacement parts are identied in the “Device Diagrams”
section of this manual. Certain components are for in-house
use only and will not be available to any non-Stryker entity.
Before any parts are purchased, an Indemnication Letter
must be signed and submitted to Stryker Endoscopy,
available from Customer Service at 1-800-624-4422.
Stryker reserves the right to incorporate improvements
to the X8000 without notice and will inform customers of
any signicant upgrades. All updated parts will be fully
interchangeable with older versions and will oer at least the
same level of quality and performance.
For up-to-date information on upgrades to the product or to
this manual, contact your Stryker representative.
Reference Documents
None of the in-house Manufacturing Assembly Procedures
(MAPs), Quality Inspection Procedures (QIPs), specialty
tools, jigs, or xtures listed in this manual are available for
purchase.
5
Device Description
Technical Specifications
e Stryker X8000 Light Source is a light-generating unit
designed to illuminate surgical sites during endoscopic
applications. e X8000 uses a 300-watt xenon bulb to
generate light, which it delivers to the surgical site via a
beroptic light cable.
e X8000 is compatible with all Stryker light cables, and,
with the proper light cable and adapters, can connect to any
exible or rigid endoscope.
e X8000 is equipped with Electronic Scope Sensing
Technology (ESST), a special safety feature that helps
prevent accidental burns caused by a light cable that is
not connected to the scope. When operated with an ESST
light cable, the X8000 senses when the scope and the light
cable are separated and places the light source in Standby
mode. In Standby mode, the X8000 will reduce light output
to a minimum, preventing the light cable from generating
excessive heat.
Electrical
Primary: 100 – 120 VAC, 50/60 Hz, 450 W
220 – 240VAC, 50/60 Hz, 450 W
Fuses (2): 5.0A, 250V
Water Ingress Protection, IPX0 — Ordinary
Equipment
Continuous Operation
Patent Protection
U.S. #5,850,496; 6,110,107; and 6,689,050.
Other patents pending.
6
Device Diagrams
e individual components of the X8000, which are referred to throughout this manual, are identied in the
following diagrams.
To p Chassis Assembly
7
Bottom Chassis Assembly
8
9
Basic Maintenance
Replacing the Bulb Module
Cleaning the X8000
Caution Unplug the X8000 before cleaning the
unit.
1. As needed, clean the external surfaces of the X8000
using a cloth or sponge dampened with a mild
detergent or disinfectant.
2. Clean and maintain the light cable according to the
manufacturer’s instructions.
Caution Do not use any abrasive cleaners. Do not
allow any liquid to drip into the unit.
Caution Do not sterilize or immerse the X8000.
Caring for the Bulb Module
e X8000 uses a Xenon bulb, which has a guaranteed
life of 500 hours when used properly. Always follow these
guidelines to ensure maximum bulb life:
1. If the bulb is touched, clean the bulb face with
alcohol and a cotton swab. Dirt or oil on the bulb
face will cause the bulb to heat unevenly and fail.
2. Do not power on and o the bulb in rapid
succession. Allow the bulb to run for at least ve
minutes once it has been powered on. Failure to do
so can cause the bulb to rapidly darken and fail.
Replace the bulb module when the LCD indicates 500 hours
or when the bulb no longer suciently illuminates the
surgical site. If possible, replace the bulb module between
surgical procedures.
WarningDuring operation, the bulb and the
housing around the bulb may be hot.
Wa it at least three minutes for the bulb
to cool before handling it.
To replace the bulb module,
1. Power down the X8000 and wait at least three
minutes for the bulb to cool.
2. Open the bulb door.
3. Rotate the bulb module handle to the right, grasp it,
and withdraw the bulb.
WarningDo not reach inside the bulb door for
any reason other than replacing the bulb
module. Touching parts other than the
bulb module may cause burns or product
damage.
WarningDo not touch any part of the bulb module
except for the handle. e module may be
very hot and cause burns.
Caution e X8000 has been designed to start the
bulb under most conditions, even when
it is hot. However, if the bulb does not
start within 10 seconds aer the X8000
has powered on, turn the unit o and
wait at least ve minutes for the bulb to
cool before restarting. Further attempts
to start the bulb can damage the bulb and
possibly the internal circuitry.
10
4. Insert the new bulb module (Stryker part number
220-201-000) along the guide rails until it is fully
seated on the mating connectors.
5. Rotate the handle to the le to secure the bulb
module within the light source.
6. Close and latch the bulb door.
7. Power on the unit.
8. Power on the bulb (if not already on) and verify that
the LCD displays the bulb hours as “0.”
Note Using a non-Stryker replacement bulb
module will result in no light output.
Caution Do not operate the X8000 with a burned-
out bulb or with no bulb installed.
Replacing the Fuses
1. Unplug the light source from the AC outlet and
remove the power cord from the rear of the unit.
2. Unlatch the fuse holder and remove the fuse(s).
3. Replace the fuse(s) with fuse(s) of the same
value and rating.
WarningTo help avoid the risk of re, use only
5.0A 250V fuses.
4. Reinstall the fuse holder.
Disposing of the X8000
is product is considered electronic
equipment. It must not be disposed of
as unsorted municipal waste and must
be collected separately. Please contact
the manufacturer or other authorized
disposal company to decommission
your equipment.
11
Troubleshooting
ProblemPossible Solution
No light output
• Ensure the AC power cord is properly connected to a hospital-grade power outlet and
the inlet on the rear console panel.
• Ensure the power switch on the front panel is powered on. (It will illuminate when
powered on.)
• Ensure all fuses are operating.
• Ensure the bulb is properly seated in the bulb housing. e bulb handle should be
turned all the way to the le within the light source.
• Ensure the bulb is in operating condition. Replace the bulb if necessary.
• Ensure the light cable is correctly engaged with the cable port. As a safety feature, the
X8000 will provide no light output unless a beroptic light cable is properly seated in
the cable port.
• Ensure the bulb access door is completely shut.
• Check for error codes E-1, E-2, E-3, E-4 or E-5. See the Error Code Denitions section
of this manual for details.
• Check that vents are not obstructed.
• If the safety shuto has been activated, please return the X8000 for service. Please see
the Component Replacement section of this manual for additional information.
To o much or too little
light output
Excessive glare in the
video
• Ensure the light cable is correctly engaged with the cable port.
• Ensure the bulb has adequate bulb life remaining. e bulb has a warranty of 500 hours.
• Ensure the X8000 is in Run mode. (e Run LED should be illuminated.) If necessary,
press the Mode button to switch from Standby to Run. If the unit remains in Standby:
1. Ensure the light cable is correctly engaged with the cable port.
2. If an ESST cable is connected to the X8000, ensure the cable is attached to the scope
using an ESST scope adapter.
• Use the up/down buttons to adjust the brightness.
• Ensure the beroptic cable is transmitting light properly. Hold the light-source end of
the cable to an overhead room light and look into the scope end of the light cable. If
the pattern contains any black spots, the light cable may be worn out and may require
replacement.
• Ensure the light cable is of an adequate size for the application. e cable diameter may
be too small to provide adequate light transmission for the medical video camera in the
endoscopic application.
• Ensure the electronic shutter on the camera is operating properly to control the video
signal brightness. If further light reduction is required, decrease the light source
brightness with the down button.
12
Error Code Definitions
e X8000 displays error codes when one of the following conditions occurs. Follow the recommended action to correct the
error.
CodeDefinitionRecommended Action
E-1
E-2
E-3
E-4
E-5
Blinking
500
All conditions are met for the bulb to illuminate, yet it
remains o.
All conditions are not met for the bulb to illuminate, yet it
remains on.
e bulb has higher than expected current or voltage
applied to it.
e bulb is kept o because the ballast fan is not working
properly.
e bulb is kept o because the bulb fan is not working
properly.
e bulb has exceeded its recommended lifetime of 500
hours.
Consult the Diagnosis section of this manual.
Consult the Diagnosis section of this manual.
Consult the Diagnosis section of this manual.
Consult the Diagnosis section of this manual.
Consult the Diagnosis section of this manual.
Install a new Stryker bulb.
Replace
Bulb
e bulb has reached 1000 hours and has exceeded its
recommended lifetime of 500 hours.
Install a new Stryker bulb.
13
Diagnosis
To diagnose system errors, follow the strategies provided
below. For steps that involve test points on the control
board, refer to the control board diagram at the end of the
Diagnosis section.
Diagnosis strategies are grouped by symptom:
A. Bulb does not turn on
B. Bulb does not turn o
C. Bulb ignites immediately upon system startup
D. LCD display is blue or completely dark
E. Fan does not run
F. Light source does not switch from STANDBY to
RUN mode
G. Brightness cannot be adjusted
H. E-1 error appears on the LCD display
I. E-2 error appears on the LCD display
J. E-3 error appears on the LCD display
K. E-4 error appears on the LCD display
L. E-5 error appears on the LCD display
M. Bulb hours value does not display
N. Shutter makes noise or does not move properly.
A. Bulb does not turn on
1. Check the voltages on the control board test points
TP22 and TP23.
• If the test points read 12V and 0V respectively,
then the ballast is faulty. Replace the ballast.
• If the voltages are in the vicinity of 5.5V and
7V respectively, proceed to step 2.
2. Visually check if the bulb door is properly
closed with the cover or a stand-alone magnet. If
necessary, close the door.
3. Visually check if the light cable is properly inserted
into the jaw. If necessary, reinsert the cable.
4. Visually check if the bulb has the one-wire memory
chip. If it does, ensure the chip makes contact with
the bulb mount board connector. If necessary,
adjust contact between the chip and the connector.
5. Using a multimeter, electronically check the
following test points on the control board. Each test
point should read 3.3 volts.
• PD1—reads 3.3 volts if the one wire chip is
detected by the control board. If PD1 reads 0
volts (and PD2 reads 3.3 volts), try a dierent
bulb. Push the bulb rmly towards the plastic
separator panel to ensure contact.
• PD2—reads 3.3 volts if the door reed switches
are closed. If PD2 reads 0 volts,
- Check if the reed switches on the bulb
mount boards are soldered properly. If
necessary, repair the soldering.
- Check for shorts using the magnet on the
reed switch. Repair/replace as necessary.
- Check if the cable wire colors are
correctly oriented. Adjust as necessary.
• PD3—reads 3.3 volts if the fans are functioning
properly
• PD4—reads 3.3 volts if a regular light cable is
inserted. If PD4 reads 0 volts, remove the front
panel and check if the beam sensor is properly
positioned and connected to the display board.
Adjust as necessary.
• PD5—reads 3.3 volts if an ESST light cable is
inserted. If PD5 reads 0 volts, remove the front
panel and check if the beam sensor is properly
positioned and connected to the display board.
Adjust as necessary.
6. If none of these help, contact Technical Support at
+1.800.624.4422.
B. Bulb does not turn o
If the bulb turns on immediately aer AC power is applied
to the light source, proceed to diagnosis strategy C, “Bulb
ignites immediately upon system startup.”
If the bulb ignites aer the regular startup sequence but then
will not turn o, perform the following steps until the source
of the problem is identied.
1. Remove the light cable from the jaw.
2. Open the bulb door or remove the magnet from the
reed switches.
3. Using a multimeter, check the following test points
on the control board:
• PD2—If PD2 reads 3.3 volts, the reed switches
are faulty. Replace the reed switches.
• PD4 and PD5—If either PD4 or PD5 reads 3.3
volts, the light cable sensor (beam sensor) is
faulty. Replace the light cable sensor.
4. If PD4 or PD5 reads 0 volts, measure the voltage
dierence between TP22 and TP23.
• If the voltage dierence is around -1.3 volts,
replace the ballast.
• If the voltage dierence is around 12 volts,
proceed to step 5.
5. Measure the voltage level for TP32.
• If TP32 reads 0 volts, replace the control board.
(See the Repair/Replacement Procedures
section of this manual.)
• If TP32 reads 3.3 volts, proceed to step 6.
14
6. Measure the voltage for TP34.
• If TP34 reads 0 or 1.6 volts, replace the control board.
(See the Repair/Replacement section of this
manual.)
• If TP34 reads otherwise, contact Technical
Support at +1.800.624.4422.
C. Bulb ignites immediately upon system
startup
1. Check if the jumpers on the ballast (JP300) are
properly installed. When the ballast is viewed from
the bulb side, the jumpers should read 0 I I 0 I 0 0.
Adjust as necessary.
2. Check the voltage dierence between TP22 and
TP23 on the control board:
• Turn o system power
• Connect the positive multimeter probe to TP22
and the negative probe to TP23
• Turn on system power
• If the voltage dierence is around 12V, proceed
to step 3
• If the voltage dierence is around -1.3 volts,
replace the ballast (See the Repair/Replacement
section of this manual.).
3. Check the voltage of TP32. If it does not read 3.3
volts, replace the control board (See the Repair/
Replacement section of this manual.).
4. Make sure that the 6-pin power cable from the
ballast to the control board is connected.
• If the power button lights up, perform steps
1 – 3.
• If the power button still does not light up,
replace the control board (See the Repair/
Replacement section of this manual.).
E. Fan does not run
1. If one or both fans are not working, an E-4 (ballast)
or E-5 (bulb) error should display on the LCD
screen. Conrm this.
2. If both fans do not work, measure the voltage on the
10.2V test point on the control board (underneath
lamp status connector).
• If it does not read around 10.2 volts, replace
the control board (See the Repair/Replacement
section of this manual.).
3. If only one fan works, switch the fan connectors
between the working and non-working fans.
• If non-working fan still fails to run, then it is
faulty. Replace it. (See the Repair/Replacement
section of this manual.)
• If the non-working fan begins to run, but the
working fan stops running, then the control
board is faulty. Replace it. (See the Repair/
Replacement section of this manual.).
D. LCD display is blue or completely dark
Power on the light source:
• If you hear the fans running, or if the power
button is lighted, perform steps 1 – 3.
• If you don’t hear the fans running, and the
power button is not lighted, perform step 4.
1. Check if the cable between the control board and
front display board is connected properly. Adjust as
necessary.
2. Insert the bulb, close the bulb door, and measure the
voltages of PD1 and PD2 on the control board.
• If they do not read 3.3 volts, replace the control
board. (See the Repair/Replacement section of
this manual.)
• If they do read 3.3 volts, proceed to step 3.
3. Measure the voltage across the C47 and C48
capacitors (directly beneath the display board cable
on the control board side).
• If they read 3.3 volts and 12.3 volts respectively,
replace the display board (See the Repair/
Replacement section of this manual.).
• If not, replace the control board (See the
Repair/Replacement section of this manual.).
F. Light source does not switch from
STANDBY to RUN mode
1. Check if the bulb is lit. If it is lit, go to step 2. If it is
not lit, then go to diagnosis strategy A.
2. Press the brightness up and brightness down buttons
on the membrane. Check if the mini brightness
meter in the top right corner of the display increases
or not.
• If the meter does not change, remove the front
panel and check if the front panel membrane
connector is fully connected to the display
board. Adjust as necessary.
• If the meter does change and the membrane
is properly connected, cycle the light source
power.
- As the LCD powers up, check the soware
version numbers. ey should read at
least SW REV P 1. 2 and SW REV P.1.2
consecutively. (e revision numbers may
be higher if newer soware versions are
installed.)
- If the version numbers are not at least
1.2, contact Technical Support at
+1.800.624.4422 for soware updates.
- If the version numbers are adequate,
contact Technical Support.
15
G. Brightness cannot be adjusted
1. Switch the light source between STANDBY and
RUN modes by pressing on the membrane switch.
(Make sure to use a non-ESST cable, or an ESST
cable that has the ESST adaptor at the tip.)
• If the light source does not switch between
modes, replace the control board (See the
Repair/Replacement section of this manual.)
• If the light source switches between modes, but
brightness cannot be increased or decreased,
proceed to step 2.
2. Remove the jaw handle and the plastic front panel.
Ensure that the front panel membrane is properly
connected to the front LCD display board. Adjust as
necessary.
3. If the membrane is properly connected, but
brightness still cannot be adjusted, replace the
membrane.
4. If the new membrane does not work, check for
proper connection between the control board and
the display board. Adjust as necessary.
5. If the membrane works aer checking the
connections, replace the front LCD display board
(See the Repair/Replacement section of this
manual.)
H. E-1 error appears on the LCD display
1. Make sure that the light cable is inserted, the bulb
door is closed with a bulb inside, and the lamp is
not lit.
2. Check the voltages on TP22 and TP23.
• If the voltage dierence between TP22 and
TP23 is not 12V, replace the control board.
• If the voltage dierence between these two test
points is 12V, ensure the J3 cable is connected
properly to the ballast and that the individual
wires are not crossed.
- Ensure that L3 and L4 ferrites, located
underneath the J3 connector, are
soldered properly to the control board.
- If these ferrites are missing, broken, or
not soldered properly, repair or replace
the control board.
3. If none of the above steps resolves the problem,
replace the ballast.
I. E-2 error appears on the LCD display
1. Make sure that the bulb is lit and the light is on.
2. Repeat diagnosis strategies B and C, respectively.
J. E-3 error appears on the LCD display
1. Probe TP13 with a multimeter.
• If it reads 3.3 volts, and E-3 still appears on the
display, replace the control board.
• If the voltage on TP13 is low, replace the
ballast.
K. E-4 error appears on the LCD display
1. Verify that the ballast fan is not running.
2. Ensure the ballast fan is connected to the control
board.
3. If the fan seems to be working, visually compare the
speed of the ballast fan to the bulb fan.
• If the ballast fan seems to be running slower,
replace the fan and verify that the E-4 error
disappears.
• If the E-4 error does not disappear with the
new fan, replace the control board.
L. E-5 error appears on the LCD display
1. Check to verify that the bulb fan is not running.
2. Ensure the ballast fan is connected to the control
board.
3. If the fan seems to be working, visually compare the
speed of the bulb fan to the ballast fan.
• If the ballast fan seems to be running slower,
replace the fan and verify that the E-5 error
disappears.
• If the E-5 error does not disappear with the
new fan, replace the control board.
M. Bulb Hours value does not display
1. Connect a regular light cable to the light source.
2. Make sure there is a bulb in the light source, close
the door, and verify that it is lit.
3. Verify that there is a one-wire memory properly
attached.
4. Verify that the one-wire memory is making contact
with the connectors on the bulb mount board.
5. If the bulb does not light, perform diagnosis
strategy A.
6. If you are unsure about the proper contact, insert a
new bulb to see if the hours will display.
7. If the hours still do not display, verify that the cable
between the bulb mount board and the control
board is properly connected.
8. If the cables are properly connected, but the hours
still do not display, replace the control board.
16
N. Shutter makes noise or does not move
properly.
1. Check the pinouts of the motor cable and the bulb
mount board/control board cable. Make sure the
wire colors match the drawing.
2. Make sure the connectors J1 and J2 on the bulb
mount board are not soldered in the wrong
direction.
3. Make sure the minimum shutter switch is soldered
properly. Make sure the switch lever is properly
connected to the switch itself.
4. Make sure the shutter wheel is properly connected
to the motor sha.
5. Make sure the shutter wheel is not rubbing against
the hot mirror holder.
6. If the motor is rotating the shutter in the reverse
direction when power is applied to the light source,
replace the motor.
7. If the motor is hesitantly rotating the shutter back
and forward, then replace the motor.
8. If the shutter’s mini tab is catching the bottom
of the minimum shutter switch, replace the bulb
mount board.
9. If none of the steps above solves the problem,
contact Technical Support at +1.800.624.4422.
17
Control Board Diagrams
18
Repair/Replacement
e X8000 Light Source is a precision instrument that has
been engineered and manufactured with great care to ensure
the safety of operators and patients. In order to maintain the
high level of safety and reliability required in devices of this
nature, it is important to fully understand and comply with
all required procedures set forth herein.
If some part of a procedure is omitted or adequate equipment
is not used, the safety and performance of the devices may
be unknowingly compromised. If any of the procedures
described in this manual are beyond the scope of the
technician’s training, consult the “Service Options” section of
this manual for information on how to obtain service from
Stryker Endoscopy.
As is the case with all AC powered
devices, dangerous voltages are present.
If adequate safety precautions are not
taken, results may include damage to the
equipment, personal injury, or death. It
is imperative that these procedures are
approached only by trained technicians
with proper equipment aer fully reading
and understanding the steps involved.
Note For all steps that involve the application of
Loctite, apply Loctite to the threaded holes
rather than to the threads on the screws
unless otherwise specied.
Note When using a power screwdriver, set
torque setting to 3 high unless otherwise
specied.
Each of the following tables provides repair/replacement
instructions for specic components of the X8000. For each
instruction, the required tools, equipment, and replacement
part numbers are provided.
Caution ESD Protection is required to perform
this assembly. Failure to follow this
procedure can cause undetectable
damage to electrical components.
19
1.0 Chassis Cover
Seq.Ta s k DescriptionFigureTo o l s NeededParts
10Unscrew Chassis Cover
from Chassis.
Torque setting of 5.
• Power
Screwdriver
(Phillips)
• 6-32 x 0.25 PH
• X8000 Cover
• X8000 Chassis
Ext. Sems. (2)
(198)
(407)
(343)
20Remove Chassis Cover
from Chassis.
• X8000 Cover
(407)
• X8000 Chassis
(343)
20
2.0 Control Board
Seq.Ta s k DescriptionFigureTo o l s NeededParts
30Remove Chassis Cover
from Chassis (procedure
1.0).
40Remove all Cables
connected to the Control
Board. Unscrew S-Bracket
from Chassis as shown.
• 4mm Nut
Driver
• 2-56 Hex Nut
• S-Bracket
• X8000 Chassis
(866)
(585)
(343)
50Unscrew U-Bracket from
Chassis as shown.
60Unscrew Control Board
from lower Control Board
mounts as shown.
Do not bend Control
Board!
70Unscrew Control Board
from Fan Support. Remove
Control Board from Unit.
• 4mm Nut
Driver
• Power
Screwdriver
(Phillips)
• Power
Screwdriver
(Phillips)
• 2-56 Hex Nut
(2) (866)
• U-Bracket
(584)
• X8000 Chassis
(343)
• 6-32 x 5/16
PH Screw (4)
(102)
• X8000 Control
Board (873)
• X8000 Chassis
(343)
• 6-32 x 5/16
PH Screw (2)
(102)
• Fan Support
(223)
• X8000 Control
Board (873)
21
Loading...
+ 49 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.