Stryker X8000 Service Manual

APPROVALS DATE TITLE
DRAWN BY:
02/28/07
Stryker Endoscopy
5900 Optical Court, San Jose, CA 95138
X8000 Light Source
Service Guide
A. Infanger
SIZE REV. DOCUMENT NO. SHEET OF
02/28/07
A A 1000-400-933 1 71
REVISIONS REVISIONS
LTR ECN NO. DATE APPROVED LTR ECN NO. DATE APPROVED
A ECO 104650 02/28/07 A. Infanger
B
C
D
E
F
G
H
J
K
L
M
N
P
R
T
U
V
W
Y
Printing Instructions:
1. 8.5 × 11” portrait
2. Front and back
3. Color printing
4. Select binding and paper weight based on cost
Global Source:
Archive this document as a service/repair guide.
X8000 Light Source
220-200-000
Service Guide
2007/02 www.stryker.com 1000400933 A
Contents
Warnings and Cautions ..................................................................................................................4
Introduction ..........................................................................................................................................5
Service Options ....................................................................................................................................5
Required Skills, Tools, and Components .............................................................................................5
Reference Documents .........................................................................................................................5
Device Description ...........................................................................................................................6
Technical Specications .......................................................................................................................6
Device Diagrams ..................................................................................................................................7
Basic Maintenance .........................................................................................................................10
Cleaning the X8000 ............................................................................................................................10
Caring for the Bulb Module ................................................................................................................10
Replacing the Bulb Module ................................................................................................................10
Replacing the Fuses ...........................................................................................................................11
Disposing of the X8000 ......................................................................................................................11
Troubleshooting ................................................................................................................................12
Error Code Denitions ........................................................................................................................13
Diagnosis ..............................................................................................................................................14
Control Board Diagrams ....................................................................................................................18
Repair/Replacement ......................................................................................................................19
1.0 Chassis Cover ..............................................................................................................................20
2.0 Control Board ...............................................................................................................................21
3.0 Ballast ..........................................................................................................................................25
4.0 Bulb Mount Board .......................................................................................................................32
5.0 AC Inlet Board .............................................................................................................................35
6.0 Jaw Handle ..................................................................................................................................38
7.0 LED power switch .......................................................................................................................39
8.0 Display Board ..............................................................................................................................43
9.0 Front Panel Membrane .................................................................................................................45
10.0 Shutter ........................................................................................................................................47
11.0 Hot Mirror ...................................................................................................................................49
12.0 Motor ..........................................................................................................................................51
13.0 Jaw Assembly ............................................................................................................................53
14.0 Integrating Rod .........................................................................................................................55
15.0 AC Inlet Filter ..............................................................................................................................57
16.0 Ballast Fan .................................................................................................................................60
17.0 Bulb Fan .....................................................................................................................................62
Final Testing and Inspection .....................................................................................................65
Tools Needed .....................................................................................................................................65
Hi-Pot/Di-Electric Breakdown Test ....................................................................................................67
Warranty ................................................................................................................................................67
3

Warnings and Cautions

Please read this manual and follow its instructions carefully. e words warning, caution, and note carry special meanings and should be carefully reviewed:
Warning Warnings indicate risks to the safety
of the patient or user. Failure to follow warnings may result in injury to the patient or user.
4. Ensure that readjustments, modications, and/ or repairs are carried out exclusively by Stryker Endoscopy or other authorized personnel.
5. Ensure that the electrical installation of the relevant operating room complies with the applicable IEC, CEC, and NEC requirements.
Federal law (United States of America) restricts this device to use by, or on order of, a physician.
Caution Cautions indicate risks to the
equipment. Failure to follow cautions may result in product damage.
Note Notes provide special information to
clarify instructions or present additional useful information.
An exclamation mark within a triangle is intended to alert the user of special warnings or important operating and maintenance instructions in the manual.
A lightning bolt within a triangle is
intended to warn of the presence of hazardous voltage. Refer all service to authorized personnel.
e incorrect use of any of the required tools and/or techniques may risk damage to the equipment or injury to the person carrying out the procedure, subsequent operators, or patient. Perform the repair ONLY if you have been specically trained in the use of all pertinent equipment and techniques. Stryker Endoscopy cannot continue to guarantee compliance to UL, CSA, TUV, or other labeled safety standards if service is performed by anyone other than Stryker Endoscopy personnel.
To help avoid potential serious injury to the user and the patient and/or damage to this device:
1. Read this manual thoroughly, especially the warnings, and be familiar with its contents prior to using this equipment.
2. Never sterilize any part of the X8000 console.
3. Disconnect the X8000 from the electrical outlet when inspecting the fuses.
IMPORTANT SAFETY NOTICE
Before operating or performing any maintenance or repairs on the X8000 Light Source, read the user guide and service guide thoroughly and carefully. When using any light source, such as the X8000, re and/or severe injury may result to the patient, user or inanimate objects, if the instructions in this manual are not followed.
All light sources, including the X8000, can generate signicant amounts of heat at the scope tip, the scope light post, the light cable tip, and/or near the light cable adapter. Higher levels of brightness from the light source result in higher levels of heat.
To avoid the risk of burns and/or re:
Always adjust the brightness level of the camera and the monitor before adjusting the brightness level of the light source.
Adjust the brightness level of the light source to the minimum brightness necessary to adequately illuminate the surgical site.
Adjust the internal shutter of the camera higher in order to run the light source at a lower intensity.
Avoid touching the scope tip or the light cable tip to the patient, and never place them on top of the patient, as doing so may result in burns to the patient or user.
Never place the scope tip, the scope light post, the light cable adapter, or the light cable tip on the surgical drapes or other ammable material, as doing so may result in re.
Always place the light source in Standby mode whenever the scope is removed from the light cable or the device is unattended. e scope tip, scope light post, light cable adapter, and light cable tip will take several minutes to cool o aer being placed in Standby mode, and therefore may still result in re or burns to the patient, user, or inanimate objects.
4

Introduction

Required Skills, Tools, and Components

is manual is intended to be used as a service guide for Stryker repair technicians in the installation, maintenance, and repair of the Stryker X8000 Light Source. It is meant to be used in conjunction with the X8000 Light Source User Manual (1000-400-885) and does not replace existing documentation.

Service Options

Factory Service
Due to the complexity of the X8000 Light Source, Stryker recommends that any malfunctioning system be returned to Stryker Endoscopy for repair or replacement, where specialized equipment and technicians are available to perform repairs while maintaining full product quality and safety. If service is needed either during or aer the warranty period:
1. Contact Stryker Endoscopy at 1-800-624-4422, or contact your local Stryker Endoscopy sales representative.
2. Package all the components carefully in the original shipping container, if possible.
3.
Ship the X8000 Light Source, prepaid and insured, to:
Stryker Endoscopy Customer Service Attention: Repair Department 5900 Optical Court San Jose, California 95138
You may request a loaner unit during the repair period.
On-Site Service
Stryker Endoscopy accepts responsibility for the eects on safety, reliability, and performance of the equipment only if readjustments, modications, and repairs have been carried out exclusively by a person specically authorized by Stryker Endoscopy to do so.
On-site repair should be carried out only by qualied technicians with the proper test equipment listed in this manual, so that the safety of operators and patients is not compromised.
In no event shall Stryker Endoscopy be liable for incidental or consequential damages in connection with or arising from the performance or use of its products aer unauthorized modication or repair.
e repair procedures described in this manual require a basic set of skills, tools, and replacement components.
Skills
Stryker recommends diagnostic and repair procedures be performed by authorized, qualied technicians with training or experience in the following:
• Basic electronics techniques
• Multimeter operation
• Oscilloscope operation
• Vectorscope operation
• BioTek Safety Analyzer operation
To o l s
Most of the procedures described in this manual can be performed using a basic tool kit that includes the items listed below.
Basic tools Advanced tools
• Phillips screwdriver
• Flathead screwdriver
• 8˝ adjustable wrench
• Needlenose pliers
• Color video monitor
• Glass fuse puller
• Multimeter
• Oscilloscope (20 Mhz or higher)
• NTSC/Pal Vectorscope
• BioTek Model 601 PRO Safety Analyzer or equivalent current leakage tester
Components
Replacement parts are identied in the “Device Diagrams” section of this manual. Certain components are for in-house use only and will not be available to any non-Stryker entity.
Before any parts are purchased, an Indemnication Letter must be signed and submitted to Stryker Endoscopy, available from Customer Service at 1-800-624-4422.
Stryker reserves the right to incorporate improvements to the X8000 without notice and will inform customers of any signicant upgrades. All updated parts will be fully interchangeable with older versions and will oer at least the same level of quality and performance.
For up-to-date information on upgrades to the product or to this manual, contact your Stryker representative.

Reference Documents

None of the in-house Manufacturing Assembly Procedures (MAPs), Quality Inspection Procedures (QIPs), specialty tools, jigs, or xtures listed in this manual are available for purchase.
5

Device Description

Technical Specifications
e Stryker X8000 Light Source is a light-generating unit designed to illuminate surgical sites during endoscopic applications. e X8000 uses a 300-watt xenon bulb to generate light, which it delivers to the surgical site via a
beroptic light cable.
e X8000 is compatible with all Stryker light cables, and,
with the proper light cable and adapters, can connect to any
exible or rigid endoscope.
e X8000 is equipped with Electronic Scope Sensing
Technology (ESST), a special safety feature that helps prevent accidental burns caused by a light cable that is not connected to the scope. When operated with an ESST light cable, the X8000 senses when the scope and the light cable are separated and places the light source in Standby mode. In Standby mode, the X8000 will reduce light output to a minimum, preventing the light cable from generating excessive heat.
Electrical
Primary: 100 – 120 VAC, 50/60 Hz, 450 W 220 – 240VAC, 50/60 Hz, 450 W Fuses (2): 5.0A, 250V
Dimensions
Height: 4.75˝ (12.1 cm) Width: 12.5˝ (31.8 cm) Depth: 16.8˝ (42.7 cm) Weight: 16.0 lbs (7.3 kg)
Fiberoptic Cable Range
2 – 6.5 mm diameter
Bulb
Type: 300 Watt Xenon (Elliptical) Life: Approximately 500 hours
Operating Conditions
10 – 40°C 30 – 75% Relative Humidity
Transportation and Storage
-20 – 50°C 10 – 75% Relative Humidity 700 – 1060 hPa
Classications and Approvals
Complies with medical safety standards:
• IEC 60601-1:2005
• CAN/CSA C22.2 No.601.1-M90
• UL 60601-1: 2003
Complies with medical EMC standard:
• IEC 60601-1-2:2001
Class 1 Equipment
Type CF applied parts
Water Ingress Protection, IPX0 — Ordinary Equipment
Continuous Operation
Patent Protection
U.S. #5,850,496; 6,110,107; and 6,689,050. Other patents pending.
6

Device Diagrams

e individual components of the X8000, which are referred to throughout this manual, are identied in the following diagrams.
To p Chassis Assembly
7
Bottom Chassis Assembly
8
9

Basic Maintenance

Replacing the Bulb Module

Cleaning the X8000

Caution Unplug the X8000 before cleaning the
unit.
1. As needed, clean the external surfaces of the X8000 using a cloth or sponge dampened with a mild detergent or disinfectant.
2. Clean and maintain the light cable according to the manufacturer’s instructions.
Caution Do not use any abrasive cleaners. Do not
allow any liquid to drip into the unit.
Caution Do not sterilize or immerse the X8000.

Caring for the Bulb Module

e X8000 uses a Xenon bulb, which has a guaranteed life of 500 hours when used properly. Always follow these guidelines to ensure maximum bulb life:
1. If the bulb is touched, clean the bulb face with alcohol and a cotton swab. Dirt or oil on the bulb face will cause the bulb to heat unevenly and fail.
2. Do not power on and o the bulb in rapid succession. Allow the bulb to run for at least ve minutes once it has been powered on. Failure to do so can cause the bulb to rapidly darken and fail.
Replace the bulb module when the LCD indicates 500 hours or when the bulb no longer suciently illuminates the surgical site. If possible, replace the bulb module between surgical procedures.
Warning During operation, the bulb and the
housing around the bulb may be hot. Wa it at least three minutes for the bulb to cool before handling it.
To replace the bulb module,
1. Power down the X8000 and wait at least three minutes for the bulb to cool.
2. Open the bulb door.
3. Rotate the bulb module handle to the right, grasp it, and withdraw the bulb.
Warning Do not reach inside the bulb door for
any reason other than replacing the bulb module. Touching parts other than the bulb module may cause burns or product damage.
Warning Do not touch any part of the bulb module
except for the handle. e module may be very hot and cause burns.
Caution e X8000 has been designed to start the
bulb under most conditions, even when it is hot. However, if the bulb does not start within 10 seconds aer the X8000 has powered on, turn the unit o and wait at least ve minutes for the bulb to cool before restarting. Further attempts to start the bulb can damage the bulb and possibly the internal circuitry.
10
4. Insert the new bulb module (Stryker part number 220-201-000) along the guide rails until it is fully seated on the mating connectors.
5. Rotate the handle to the le to secure the bulb module within the light source.
6. Close and latch the bulb door.
7. Power on the unit.
8. Power on the bulb (if not already on) and verify that the LCD displays the bulb hours as “0.”
Note Using a non-Stryker replacement bulb
module will result in no light output.
Caution Do not operate the X8000 with a burned-
out bulb or with no bulb installed.

Replacing the Fuses

1. Unplug the light source from the AC outlet and remove the power cord from the rear of the unit.
2. Unlatch the fuse holder and remove the fuse(s).
3. Replace the fuse(s) with fuse(s) of the same value and rating.
Warning To help avoid the risk of re, use only
5.0A 250V fuses.
4. Reinstall the fuse holder.

Disposing of the X8000

is product is considered electronic equipment. It must not be disposed of as unsorted municipal waste and must be collected separately. Please contact the manufacturer or other authorized disposal company to decommission your equipment.
11

Troubleshooting

Problem Possible Solution
No light output
• Ensure the AC power cord is properly connected to a hospital-grade power outlet and the inlet on the rear console panel.
• Ensure the power switch on the front panel is powered on. (It will illuminate when powered on.)
• Ensure all fuses are operating.
• Ensure the bulb is properly seated in the bulb housing. e bulb handle should be turned all the way to the le within the light source.
• Ensure the bulb is in operating condition. Replace the bulb if necessary.
• Ensure the light cable is correctly engaged with the cable port. As a safety feature, the X8000 will provide no light output unless a beroptic light cable is properly seated in the cable port.
• Ensure the bulb access door is completely shut.
• Check for error codes E-1, E-2, E-3, E-4 or E-5. See the Error Code Denitions section of this manual for details.
• Check that vents are not obstructed.
• If the safety shuto has been activated, please return the X8000 for service. Please see the Component Replacement section of this manual for additional information.
To o much or too little light output
Excessive glare in the video
• Ensure the light cable is correctly engaged with the cable port.
• Ensure the bulb has adequate bulb life remaining. e bulb has a warranty of 500 hours.
• Ensure the X8000 is in Run mode. (e Run LED should be illuminated.) If necessary, press the Mode button to switch from Standby to Run. If the unit remains in Standby:
1. Ensure the light cable is correctly engaged with the cable port.
2. If an ESST cable is connected to the X8000, ensure the cable is attached to the scope using an ESST scope adapter.
• Use the up/down buttons to adjust the brightness.
• Ensure the beroptic cable is transmitting light properly. Hold the light-source end of the cable to an overhead room light and look into the scope end of the light cable. If the pattern contains any black spots, the light cable may be worn out and may require replacement.
• Ensure the light cable is of an adequate size for the application. e cable diameter may be too small to provide adequate light transmission for the medical video camera in the endoscopic application.
• Ensure the electronic shutter on the camera is operating properly to control the video signal brightness. If further light reduction is required, decrease the light source brightness with the down button.
12
Error Code Definitions
e X8000 displays error codes when one of the following conditions occurs. Follow the recommended action to correct the error.
Code Definition Recommended Action
E-1
E-2
E-3
E-4
E-5
Blinking 500
All conditions are met for the bulb to illuminate, yet it remains o.
All conditions are not met for the bulb to illuminate, yet it remains on.
e bulb has higher than expected current or voltage applied to it.
e bulb is kept o because the ballast fan is not working properly.
e bulb is kept o because the bulb fan is not working properly.
e bulb has exceeded its recommended lifetime of 500 hours.
Consult the Diagnosis section of this manual.
Consult the Diagnosis section of this manual.
Consult the Diagnosis section of this manual.
Consult the Diagnosis section of this manual.
Consult the Diagnosis section of this manual.
Install a new Stryker bulb.
Replace Bulb
e bulb has reached 1000 hours and has exceeded its recommended lifetime of 500 hours.
Install a new Stryker bulb.
13

Diagnosis

To diagnose system errors, follow the strategies provided below. For steps that involve test points on the control board, refer to the control board diagram at the end of the Diagnosis section.
Diagnosis strategies are grouped by symptom:
A. Bulb does not turn on
B. Bulb does not turn o
C. Bulb ignites immediately upon system startup
D. LCD display is blue or completely dark
E. Fan does not run
F. Light source does not switch from STANDBY to
RUN mode
G. Brightness cannot be adjusted
H. E-1 error appears on the LCD display
I. E-2 error appears on the LCD display
J. E-3 error appears on the LCD display
K. E-4 error appears on the LCD display
L. E-5 error appears on the LCD display
M. Bulb hours value does not display
N. Shutter makes noise or does not move properly.
A. Bulb does not turn on
1. Check the voltages on the control board test points TP22 and TP23.
• If the test points read 12V and 0V respectively, then the ballast is faulty. Replace the ballast.
• If the voltages are in the vicinity of 5.5V and 7V respectively, proceed to step 2.
2. Visually check if the bulb door is properly closed with the cover or a stand-alone magnet. If necessary, close the door.
3. Visually check if the light cable is properly inserted into the jaw. If necessary, reinsert the cable.
4. Visually check if the bulb has the one-wire memory chip. If it does, ensure the chip makes contact with the bulb mount board connector. If necessary, adjust contact between the chip and the connector.
5. Using a multimeter, electronically check the following test points on the control board. Each test point should read 3.3 volts.
• PD1—reads 3.3 volts if the one wire chip is detected by the control board. If PD1 reads 0 volts (and PD2 reads 3.3 volts), try a dierent bulb. Push the bulb rmly towards the plastic separator panel to ensure contact.
• PD2—reads 3.3 volts if the door reed switches are closed. If PD2 reads 0 volts,
- Check if the reed switches on the bulb mount boards are soldered properly. If necessary, repair the soldering.
- Check for shorts using the magnet on the reed switch. Repair/replace as necessary.
- Check if the cable wire colors are correctly oriented. Adjust as necessary.
• PD3—reads 3.3 volts if the fans are functioning properly
• PD4—reads 3.3 volts if a regular light cable is inserted. If PD4 reads 0 volts, remove the front panel and check if the beam sensor is properly positioned and connected to the display board. Adjust as necessary.
• PD5—reads 3.3 volts if an ESST light cable is inserted. If PD5 reads 0 volts, remove the front panel and check if the beam sensor is properly positioned and connected to the display board. Adjust as necessary.
6. If none of these help, contact Technical Support at +1.800.624.4422.
B. Bulb does not turn o
If the bulb turns on immediately aer AC power is applied to the light source, proceed to diagnosis strategy C, “Bulb ignites immediately upon system startup.”
If the bulb ignites aer the regular startup sequence but then will not turn o, perform the following steps until the source of the problem is identied.
1. Remove the light cable from the jaw.
2. Open the bulb door or remove the magnet from the reed switches.
3. Using a multimeter, check the following test points on the control board:
• PD2—If PD2 reads 3.3 volts, the reed switches are faulty. Replace the reed switches.
• PD4 and PD5—If either PD4 or PD5 reads 3.3 volts, the light cable sensor (beam sensor) is faulty. Replace the light cable sensor.
4. If PD4 or PD5 reads 0 volts, measure the voltage dierence between TP22 and TP23.
• If the voltage dierence is around -1.3 volts, replace the ballast.
• If the voltage dierence is around 12 volts, proceed to step 5.
5. Measure the voltage level for TP32.
• If TP32 reads 0 volts, replace the control board. (See the Repair/Replacement Procedures section of this manual.)
• If TP32 reads 3.3 volts, proceed to step 6.
14
6. Measure the voltage for TP34.
• If TP34 reads 0 or 1.6 volts, replace the control board. (See the Repair/Replacement section of this
manual.)
• If TP34 reads otherwise, contact Technical Support at +1.800.624.4422.
C. Bulb ignites immediately upon system startup
1. Check if the jumpers on the ballast (JP300) are properly installed. When the ballast is viewed from the bulb side, the jumpers should read 0 I I 0 I 0 0. Adjust as necessary.
2. Check the voltage dierence between TP22 and TP23 on the control board:
• Turn o system power
• Connect the positive multimeter probe to TP22 and the negative probe to TP23
• Turn on system power
• If the voltage dierence is around 12V, proceed to step 3
• If the voltage dierence is around -1.3 volts, replace the ballast (See the Repair/Replacement section of this manual.).
3. Check the voltage of TP32. If it does not read 3.3 volts, replace the control board (See the Repair/ Replacement section of this manual.).
4. Make sure that the 6-pin power cable from the ballast to the control board is connected.
• If the power button lights up, perform steps 1 – 3.
• If the power button still does not light up, replace the control board (See the Repair/ Replacement section of this manual.).
E. Fan does not run
1. If one or both fans are not working, an E-4 (ballast) or E-5 (bulb) error should display on the LCD screen. Conrm this.
2. If both fans do not work, measure the voltage on the
10.2V test point on the control board (underneath lamp status connector).
• If it does not read around 10.2 volts, replace the control board (See the Repair/Replacement section of this manual.).
3. If only one fan works, switch the fan connectors between the working and non-working fans.
• If non-working fan still fails to run, then it is faulty. Replace it. (See the Repair/Replacement section of this manual.)
• If the non-working fan begins to run, but the working fan stops running, then the control board is faulty. Replace it. (See the Repair/ Replacement section of this manual.).
D. LCD display is blue or completely dark
Power on the light source:
• If you hear the fans running, or if the power button is lighted, perform steps 1 – 3.
• If you don’t hear the fans running, and the power button is not lighted, perform step 4.
1. Check if the cable between the control board and front display board is connected properly. Adjust as necessary.
2. Insert the bulb, close the bulb door, and measure the voltages of PD1 and PD2 on the control board.
• If they do not read 3.3 volts, replace the control board. (See the Repair/Replacement section of this manual.)
• If they do read 3.3 volts, proceed to step 3.
3. Measure the voltage across the C47 and C48 capacitors (directly beneath the display board cable on the control board side).
• If they read 3.3 volts and 12.3 volts respectively, replace the display board (See the Repair/ Replacement section of this manual.).
• If not, replace the control board (See the Repair/Replacement section of this manual.).
F. Light source does not switch from STANDBY to RUN mode
1. Check if the bulb is lit. If it is lit, go to step 2. If it is not lit, then go to diagnosis strategy A.
2. Press the brightness up and brightness down buttons on the membrane. Check if the mini brightness meter in the top right corner of the display increases or not.
• If the meter does not change, remove the front panel and check if the front panel membrane connector is fully connected to the display board. Adjust as necessary.
• If the meter does change and the membrane is properly connected, cycle the light source power.
- As the LCD powers up, check the soware version numbers. ey should read at least SW REV P 1. 2 and SW REV P.1.2 consecutively. (e revision numbers may be higher if newer soware versions are installed.)
- If the version numbers are not at least
1.2, contact Technical Support at +1.800.624.4422 for soware updates.
- If the version numbers are adequate, contact Technical Support.
15
G. Brightness cannot be adjusted
1. Switch the light source between STANDBY and RUN modes by pressing on the membrane switch. (Make sure to use a non-ESST cable, or an ESST cable that has the ESST adaptor at the tip.)
• If the light source does not switch between modes, replace the control board (See the Repair/Replacement section of this manual.)
• If the light source switches between modes, but brightness cannot be increased or decreased, proceed to step 2.
2. Remove the jaw handle and the plastic front panel. Ensure that the front panel membrane is properly connected to the front LCD display board. Adjust as necessary.
3. If the membrane is properly connected, but brightness still cannot be adjusted, replace the membrane.
4. If the new membrane does not work, check for proper connection between the control board and the display board. Adjust as necessary.
5. If the membrane works aer checking the connections, replace the front LCD display board (See the Repair/Replacement section of this manual.)
H. E-1 error appears on the LCD display
1. Make sure that the light cable is inserted, the bulb door is closed with a bulb inside, and the lamp is not lit.
2. Check the voltages on TP22 and TP23.
• If the voltage dierence between TP22 and TP23 is not 12V, replace the control board.
• If the voltage dierence between these two test points is 12V, ensure the J3 cable is connected properly to the ballast and that the individual wires are not crossed.
- Ensure that L3 and L4 ferrites, located underneath the J3 connector, are soldered properly to the control board.
- If these ferrites are missing, broken, or not soldered properly, repair or replace the control board.
3. If none of the above steps resolves the problem, replace the ballast.
I. E-2 error appears on the LCD display
1. Make sure that the bulb is lit and the light is on.
2. Repeat diagnosis strategies B and C, respectively.
J. E-3 error appears on the LCD display
1. Probe TP13 with a multimeter.
• If it reads 3.3 volts, and E-3 still appears on the display, replace the control board.
• If the voltage on TP13 is low, replace the ballast.
K. E-4 error appears on the LCD display
1. Verify that the ballast fan is not running.
2. Ensure the ballast fan is connected to the control board.
3. If the fan seems to be working, visually compare the speed of the ballast fan to the bulb fan.
• If the ballast fan seems to be running slower, replace the fan and verify that the E-4 error disappears.
• If the E-4 error does not disappear with the new fan, replace the control board.
L. E-5 error appears on the LCD display
1. Check to verify that the bulb fan is not running.
2. Ensure the ballast fan is connected to the control board.
3. If the fan seems to be working, visually compare the speed of the bulb fan to the ballast fan.
• If the ballast fan seems to be running slower, replace the fan and verify that the E-5 error disappears.
• If the E-5 error does not disappear with the new fan, replace the control board.
M. Bulb Hours value does not display
1. Connect a regular light cable to the light source.
2. Make sure there is a bulb in the light source, close the door, and verify that it is lit.
3. Verify that there is a one-wire memory properly attached.
4. Verify that the one-wire memory is making contact with the connectors on the bulb mount board.
5. If the bulb does not light, perform diagnosis strategy A.
6. If you are unsure about the proper contact, insert a new bulb to see if the hours will display.
7. If the hours still do not display, verify that the cable between the bulb mount board and the control board is properly connected.
8. If the cables are properly connected, but the hours still do not display, replace the control board.
16
N. Shutter makes noise or does not move properly.
1. Check the pinouts of the motor cable and the bulb mount board/control board cable. Make sure the wire colors match the drawing.
2. Make sure the connectors J1 and J2 on the bulb mount board are not soldered in the wrong direction.
3. Make sure the minimum shutter switch is soldered properly. Make sure the switch lever is properly connected to the switch itself.
4. Make sure the shutter wheel is properly connected to the motor sha.
5. Make sure the shutter wheel is not rubbing against the hot mirror holder.
6. If the motor is rotating the shutter in the reverse direction when power is applied to the light source, replace the motor.
7. If the motor is hesitantly rotating the shutter back and forward, then replace the motor.
8. If the shutter’s mini tab is catching the bottom of the minimum shutter switch, replace the bulb mount board.
9. If none of the steps above solves the problem, contact Technical Support at +1.800.624.4422.
17

Control Board Diagrams

18

Repair/Replacement

e X8000 Light Source is a precision instrument that has been engineered and manufactured with great care to ensure the safety of operators and patients. In order to maintain the high level of safety and reliability required in devices of this nature, it is important to fully understand and comply with all required procedures set forth herein.
If some part of a procedure is omitted or adequate equipment is not used, the safety and performance of the devices may be unknowingly compromised. If any of the procedures described in this manual are beyond the scope of the technician’s training, consult the “Service Options” section of this manual for information on how to obtain service from Stryker Endoscopy.
As is the case with all AC powered devices, dangerous voltages are present. If adequate safety precautions are not taken, results may include damage to the equipment, personal injury, or death. It is imperative that these procedures are approached only by trained technicians with proper equipment aer fully reading and understanding the steps involved.
Note For all steps that involve the application of
Loctite, apply Loctite to the threaded holes rather than to the threads on the screws unless otherwise specied.
Note When using a power screwdriver, set
torque setting to 3 high unless otherwise specied.
Each of the following tables provides repair/replacement instructions for specic components of the X8000. For each instruction, the required tools, equipment, and replacement part numbers are provided.
Caution ESD Protection is required to perform
this assembly. Failure to follow this procedure can cause undetectable damage to electrical components.
19

1.0 Chassis Cover

Seq. Ta s k Description Figure To o l s Needed Parts
10 Unscrew Chassis Cover
from Chassis.
Torque setting of 5.
• Power Screwdriver (Phillips)
• 6-32 x 0.25 PH
• X8000 Cover
• X8000 Chassis
Ext. Sems. (2) (198)
(407)
(343)
20 Remove Chassis Cover
from Chassis.
• X8000 Cover (407)
• X8000 Chassis (343)
20

2.0 Control Board

Seq. Ta s k Description Figure To o l s Needed Parts
30 Remove Chassis Cover
from Chassis (procedure
1.0).
40 Remove all Cables
connected to the Control Board. Unscrew S-Bracket from Chassis as shown.
• 4mm Nut Driver
• 2-56 Hex Nut
• S-Bracket
• X8000 Chassis
(866)
(585)
(343)
50 Unscrew U-Bracket from
Chassis as shown.
60 Unscrew Control Board
from lower Control Board mounts as shown.
Do not bend Control
Board!
70 Unscrew Control Board
from Fan Support. Remove Control Board from Unit.
• 4mm Nut Driver
• Power Screwdriver (Phillips)
• Power Screwdriver (Phillips)
• 2-56 Hex Nut (2) (866)
• U-Bracket (584)
• X8000 Chassis (343)
• 6-32 x 5/16 PH Screw (4) (102)
• X8000 Control Board (873)
• X8000 Chassis (343)
• 6-32 x 5/16 PH Screw (2) (102)
• Fan Support (223)
• X8000 Control Board (873)
21
Loading...
+ 49 hidden pages