Sper Scientific 850052 User Manual

BATTERY REPLACEMENT
Replace all 4 LR44 batteries when the low battery icon is flashing, or when troubleshooting (below) fails.
Unscrew the cover and install the batteries with the flat positive (+) side of the batteries facing the spring.
Do not remove the O-ring.
Remove the batteries when the unit is
not in use for an extended period.
TROUBLESHOOTING
If there is no display, or you have erratic readings after having soaked the electrode in distilled/de-ionized water for at least 1 hour prior to calibration, check the continuity of the batteries by removing them for 1 minute, then reinsert.
Slow Response: Soak the electrode in tap water for 10-15 minutes, then rinse with distilled/de-ionized water.
"- - -" displays when the pH sample is too acidic or too alkaline.
"H" or "L" displays when the temperature is too hot (H) or too cold (L).
No Display: Be sure to press for at least one second. If the problem persists, see Battery Replacement (above).
If your readings are still erratic after calibration, soak the electrode for 10 to 30 minutes in distilled water.
Note: It is normal for the pH value to fluctuate quickly when the electrode is exposed to air and not immersed.
SPECIFICATIONS
pH Range 0~14 pH pH Resolution 850051 850052 Accuracy 850051 850052 Adjustable Calibration Points Temp.
0.1 pH
0.01 pH
±0.2 pH ±0.05 pH pH 4 pH 7 pH 10 0~60°C
0.5 °F/°C
0.1°F/°C
±1°C ±0.5°C
3.5~4.5
6.5~7.5
9.5~10.5
Range Operating Environ.
31~122°F 0~50°C
<80 RH
Weight 2.4 oz, 68g with battery Dimension 6½" x 1½" x 1"
165 x 35 x 32 mm
WARRANTY: Sper Scientific warrants this product against defects in materials a n d w o r k m a n s h i p f o r a period of 1 year from the date of purchase, and agrees to repair or replace any defective unit without charge. If your model has since been discontinued, an equivalent Sper Scientific product will be substituted if available. This warranty does not cover probes, batteries, or damage resulting from accident, misuse, or abuse. In or­der to obtain warranty service, ship the unit postage prepaid to:
SPER SCIENTIFIC
7720 E Redfield Rd, Ste 7
Scottsdale, AZ 85260
Email: info@sperscientific.com
Include a description of the problem and your return address along with the re­t u r n e d i t e m . R e t u r n t h e warranty card within 10 days or register at: www.sperscientific.com.
LARGE DISPLAY
ATC PH PENS
850051 and 850052
Loading...