Sper Scientific 850038 User Manual

Conductivity, TDS,
Salinity Meter
850038
Instruction Manual
SPER SCIENTIFIC
INDEX
Introduction..............................................1
MaterialSupplied....................................1
Ports&Probe ............................................1
Specifications .........................................2
LCDDisplay ..................................................2
KeypadOperation ........................................4
Auto Shut Off .................................. 6
RecallMode ..............................................7
ReviewMode.................................................7
Peparation For Calibration.............. 8
ConductivityCalibration .........................11
TDS Calibration.......................................14
SaltCalibration .......................................16
ConductivityMeasurement....................18
TDS Measurement ................................20
SalinityMeasurement.............................21
ParametersSetting.................................23
Maintenance& Storage..........................35
Trouble Shooting..................................... 36
Warranty ......................................................... 37
IrDATransmission..................................38
RS232Output ........................................39
AppendixA..............................................40
AppendixB .............................................41
AppendixC..............................................41
AppendixD..............................................42
INTRODUCTION
This meter measures conductivity value in uS and mS, as well as total dissolved solids (TDS) and Salinity. Simultaneously displays readings, time, date, and temper­ature in °C or °F.
Features a detachable probe, ATC, 99 memory points, min-max-average, hold, RS232 port, large LCD, auto shut off, and indicators for low battery and over range.
MATERIAL SUPPLIED
This package contains:
Meter Probe
Batteries AAA x 4pcs Instruction manual Hard carrying case
Meter
PORTS & PROBE
Top view
Conductivity probe
connector port
Probe
Connector to meter
~ 87 cm
Right side view
IrDA port
3x10mm sensing area on each side
Standard probe
*ABS body, graphite cell. *Highest durability with
Temp.
reliable measurements.
sensor
*Integral temperature.
inside
*Temperature steady time <10 seconds.
1
DC 9V
RS232
DISPLAY
Themeterwill briefly displayall LCDsegmentswhen it is first turnedon.
The primary display shows: Themeasuredconductivityvalue (CON) in uS or mS per cm.
TDS in ppm or ppt Salinity in ppt
The secondary display below the primary display shows:
The temperature of the reading. Year/Month/Date and
Hour/Min/Sec (displayed
at the bottom of the screen).
HLD(Hold),Max,Min,Avg,
ReadyandATC are shownon theleftside of the screen.
REC (Recall)isshownatthe left
bottom left.
The top right number ("88")
indicates the records in memory.
3
KEYPAD OPERATION
1. SET (Power) key: Turn on the meter by pressing the SET key. The entire display
flashes. Once the m,eter is on,
press this key for more than 2 seconds to enter the
settings m,ode.
Manually turn the meter off by pressing and releasing this key.
2. ▲ MODE key: Press the Mode key to switch between the displayed measurements for conductivity (CON), TDS, or Salinity (SALT).
4
3. CAL/Esc key:
ATC
CON
H:M:SouS
ATC
Y.M.D
CON
o
uS
In normal mode, press this key > 2 seconds to enter calibration mode. While the meter is in calibration, setting or recall mode, press this key (Esc) to return to previous mode.
4. MEM :
In normal mode, press this key to store the current reading with RTC (real time clock), the primary display will flash 3 times. The occupied memory number (88 on the top right corner) changes from xx to yy, eg: 00 to 01 .(Fig. A& B)
In setting mode, press this key to change the setting content. Under recall mode, press this key to read memories.
C
Fig.A Fig.B
5-1.HLD/REC :
In normal mode, press the key < 1 sec. To hold the current reading, then
quickly press the key again to unlock it. In normal mode, press the key more than 2 seconds to enter recall mode. (Fig. C, page 6)
5
Press HLD+ENTER for the backlight. The backlight will turn off automatically after 10 seconds.
CON
uS
Fig. C
REC
Y-M-D
o
C
6-1.
MI/MX/AV key: In normal and recall mode, press this key to view Max./Min./Avg. readings.
In normal mode, press key > 2 sec. to clear the value. The entire LCD will flash for 2 seconds.
In setting mode, press this key to enter content setting.
7. SET + : When the meter is off, pressing these two keys simultaneously for more than one second disables auto-off mode.
AUTO POWER OFF
The meter will auto-powered off in 20 minutes after last key press. This function saves battery power.
To override this function, press POWER and UP keys at the same time until "n" is displayed. Release the keys and meter will enter normal mode.
6
RECALL MODE
In the normal mode, press more
HLDHLD
RecRec
than 2 seconds to enter RECALL (REC) mode. Press MIN/MAX/AV to view Maximum, Minimum and Average of the memorized data ( Fig. D). You could also press or to review
MODEMODE
MEMMEM
the memories one by one.
"REC" is flashing in recall mode. To exit recall mode, press more than 2 sec. Quickly press to return to normal mode. LCD will show the last memory
CALCAL
Es cEs c
HLDHLD
RecRec
count saved before you entered recall mode.
CON
MIN
REC
uS
o
C
Fig. D
Press MX/MN/AV key w
hen REC is
flashing, so that the MAX value of the all memories will diaplay. Press MX/MN/AV key again to view MIN. Press again to view Average. The top right corner record number will not be displayed while viewing MX/MN/AV.
REVIEW MODE
NN
/AV/AV
XX
In normal mode, press key < 1sec.
MM
to view MAX, MIN, AVG. These values are cleared when the meter is turned off.
You can press the key > 2 second
NN
/AV/AV
XX
MM
to clear these values (LCD will flash for 2 seconds). Note: Clear the values when
the measurement is stable.
7
PREPARATION FOR CALIBRATION
Two issues should be considered before operation:
What is the right calibration standard? When should you calibrate?
Selecting a calibration standard
For best results, select a conductivity, TDS or NaCl standard near the sample value you are measuring.
Alternatively, use a calibration solution
value which is approximately 2/3 of the
full scale of the measurement range you plan to use.
For example, in the 0 to 1999 uS range, use 1413 uS solution for calibration.
DO NOT reuse the calibration solution. Contaminants in the solution will affect the calibration and the accuracy. Be sure to use fresh solution each time.
Refer to below table. Use the recommended solution for different conductivity and TDS ranges
for best results.
Conductivity
measuring range
0~19.99uS
1
0~199.9uS
2
0~1999uS
3 4
0~19.99mS 0~199.9mS
5
Recommended cal.
solution range
6.00~17.00uS
60.0~170.0uS 600~1700uS
6.00~17.00mS
60.0~170.0mS
TDS measuring
range (factor=0.5)
0.00~9.99ppm
1
0.0~99.9ppm
2
0~999ppm
3
0.00~9.99ppt
4
0.0~199.9ppt
5
Recommended cal.
solution range
3.00~8.50ppm
30.0~85.0ppm 300~850ppm
3.00~8.50ppt
30.0~85.0ppt
8
This meter has a built-in algorithm to linearize the measurement of Sodium Chloride concentration. For SALT calibration, you only need to perform one-point calibration.
The previous calibration data will be replaced after re-calibrating. For example, if you previously calibrated the conductivity meter at 1413 uS in the 0 to 1999 uS range, when you re-calibrate it at 1500 uS again (also in the 0 to 1999uS range), the previous 1413uS will be replaced in this range (0~1999uS). However, the meter will retain the calibration data for other ranges which are not yet re-calibrated.
If you use solution to calibrate one range and then manually input the cell constant (P5.2), the cell constant of range 1 to 5 will be changed simultaneously. .
NOTE: The temperature coefficient of the meter defaults to 2.1% per C (Fig. E) and
o
provides good results for most applications. See Program P4.1 on page 27 to reset the coefficient.
Alternatively, refer to Appendix D to calculate the temperature coefficient and determine the appropriate temperature coefficient for solution.
o
C
Fig. E
NOTE:
The default normalization temperature value is 25°C.
9
To normalize to another value, see
K=0.966
P4.2 on page 28 (Fig. F). Before
resetting this value,
the calibration
standard value of the normalized temp. must be known. (Refer to the datasheet enclosed with your solution.)
o
C
Fig .F
When should you do the calibration?
For first useand for best results, use solution to calibrate or manually input the cell constant
if
solution is not readily available.
(refer to the label on the probe). NOTE: The cell constant has been programmed into meter. To completely calibrate the meter, clear all previous calibration data, (to erase all calibration data, see P7.1 on page 32 (Fig. G)).
K=
0.966
Fig .G (8306)
Cell Constant If the conductivity of measured solutions are < 100 S, or TDS are < 50 ppm, calibrate the meter at least once a week for accuracy.
If the meter is used in the mid ranges, it is needed to do calibration at least once a month.
If the measurement is performed at extreme temperatures, calibrate at least once a week.
10
CONDUCTIVITY CALIBRATION
Follow the below steps for conductivity calibration:
Insert the probe into demineralized
1.
water or distilled water for about 30 minutes to rinse the probe.
Select the conductivity standard for
2.
calibration. ( See page 8)
Pour 4 cm (deep) of the solution into
3.
two separate clean containers(A&B).
Turn on the meter, the entire display
4.
flashes three times and then returns to the normal measurement mode.
5.Rinse the probe in one of the containers. Gently stir the probe.
6.Dip the rinsed probe into the other container. Tap the probe on the bottom of container to remove air bubbles. Let the probe stabilize to the solution temperature.
66
K=0.966
K=0.9
4cm
A(Step 5)
4cm
B( Step 6)
7.Press CAL key more than2 seconds to enter calibration mode. The probe auomatically detects the conductivity value of solution and blinks the value on the LCD (Fig. H)
CON
uS
READY
H:M:S
o
C
Fig .H
11
8. Press the UP/DOWN keys to change the value on the primary display to match the value to the standard ( normalization temp. is 25°C). Adjust the conductivity reading up to +20% from the detected value. However, if the detected value and standard value differ by more than +20%, clean or replace the probe. Example: Standard: 10uS; Detected value:19uS Adjustable range: +3.8us (19*20%) However, under above situation, the values already differed over 20%.
NOTE: * When the calibration is stable, "Ready" will display. If not, check that the calibration solutions aand the input value (step 8) are correct.
* Cell constant may degrade with time and usage. Use this feature as a reminder for changing to a new probe.
* The meter will automatically detect the solution, If the standard value is over the measuring limit or less than 10% of measuring
value will equal the range limit or 10% of range limit. Under this situation, go to parameter settings first to manually select a suitable range. (P1.0, see page 23).
Example 1: Standard: 22uS; Detected value:19uS Adjustable range: +3.8us (19*20%) Although the values differ less than
limit, the displayed
12
20% but the 22uS is already over range limit so the maximum value could be input is 19.99uS only. To exactly adjust the value to 22uS, please manual select the range as 0~199.9 in P1.3
Example 2: Standard: 1.6uS; Detected value:2.1uS Adjustable range: +0.42us (2.1*20%) Although the values differ less than 20% but the 1.6uS is already less than 10% range limit (1.99) so the max. value that could be input is 1.99uS.
9.After "Ready" is displayed, press "ENTER" to confirm the calibration. The LCD will stop flashing and the meter will switch back to normal measurement mode.
10.Repeat 1~9 for other ranges. NOTE:
* When switching the meter from measurment to calibration mode, the meter will auto defect the solution value based on the previously selected cell constant (0.1,1.0 or 10), so, sometimes the primary display may seem to jump to the auto defected value after entering calibration. This means that even in the same solution, it is normal for the displayed values in measurment and calibration mode s to be different.
* To exit conductivity calibration mode without confirming calibration, DO NOT press the ENTER key in step 9. Press Esc to retain the meter's previous calibration data for the current range.
13
Loading...
+ 31 hidden pages