Sper Scientific 850037 User Manual

Page 1
Large Display
Salinity &
Conductivity Pens
850036 & 850037
Page 2
Large Display Salinity & Conductivity Pens
Copyright ©2013 by
Sper Scientic
ALL RIGHTS RESERVED Printed in the USA
The contents of this manual may not be reproduced or transmitted in any form or by any means electronic, mechanical, or other means that do not yet exist or may be developed, including photocopying, recording, or any information storage and retrieval system without the express
permission from Sper Scientic.
Sper Scientic 8281 E. Evans Rd., Suite #103
Scottsdale, AZ 85260 (480) 948-4448
2
Page 3
TABLE OF CONTENTS
INTRODUCTION. . . . . . . . . . . . . . 4
FEATURES. . . . . . . . . . . . . . . . . 5
MATERIALS SUPPLIED . . . . . . . . . . 5
POWER SUPPLY . . . . . . . . . . . . .6
LCD DISPLAY . . . . . . . . . . . . . . . 6
APPENDICES . . . . . . . . . . . . . . 45
BATTERY REPLACEMENT . . . . . . . 52
MAINTENANCE . . . . . . . . . . . . . 53
TROUBLESHOOTING . . . . . . . . . . 54
WARRANTY . . . . . . . . . . . . . . . 56
850036 SALINITY PEN . . . . . . . . . .7
SET UP . . . . . . . . . . . . . . . . . . 8
CALIBRATION . . . . . . . . . . . . . 12
MEASUREMENT PROCEDURES . . . 16
SPECIFICATIONS . . . . . . . . . . . 19
ERROR CODES . . . . . . . . . . . . 20
850037 CONDUCTIVITY PEN . . . . . . 23
SET UP . . . . . . . . . . . . . . . . . 24
CALIBRATION . . . . . . . . . . . . . 32
MEASUREMENT PROCEDURES . . . 38
SPECIFICATIONS . . . . . . . . . . . 42
ERROR CODES . . . . . . . . . . . . 43
3
Page 4
INTRODUCTION
The Sper Scientic Large Display Salinity
(Model 850036) and Conductivity (Model
850037) Pens provide highly stable and accurate readings with a large LCD that simultaneously displays the parameter being measured and temperature in
°C or °F. The pens are designed for simple one-hand operation. The case is IP65 waterproof, and will oat if accidentally dropped into the water.
Salinity is displayed not as a percentage but as parts per thousand (ppt), which is approximately grams of salt per liter
of solution.
Conductivity is displayed as mhos per centimeter (M/cm) or siemens per centimeter (S/cm). Because a mho (or siemen) is a large unit, microsiemen or millisiemen typically is used (µS/cm,
mS/cm).
4
Page 5
FEATURES
• IP65 waterproof housing
• Dual display with ATC
• Data hold
• Pen size for portability
• Built-in probe with protective cap
• One touch, multi-point calibration
• Auto-power-off
• Low battery indicator
MATERIALS SUPPLIED
• 4 LR44 button cell batteries
• Instruction Manual
5
Page 6
POWER SUPPLY
LCD DISPLAY
The pen is powered by 4 LR44 batteries.
Check the batteries when:
• Using the unit for the rst time
• The low battery indicator appears
on the LCD
• The pen will not turn on
LCD DISPLAY
Hold data on the LCD Calibration Mode
Low battery indicator
Unit of conductivity
Measured value
Unit of TDS and Salinity
Unit of liquid temperature °C or °F
Temperature reading
6
Page 7
Large Display
Salinity Pen
850036
7
Page 8
SETUP MODE
The advanced Setup Mode allows you
to customize the pen’s preferences and
defaults:
1. Press SET to turn the pen on.
2. From Measurement Mode, press SET
for more than 2 seconds to enter Setup
Mode and change parameters.
Temperature Units (tUt)
To select the Celsius or Fahrenheit temperature scale:
1. Press SET for 2 seconds to enter setup.
2. “t” will appear on the primary display of the LCD and “P1.0” will appear on the secondary display.
3. Press SET again for less than one second to enter the unit setting.
4. Press ▲ or ▼ to select C (Celsius) or F (Fahrenheit). Press SET for less than one second to conrm the unit selection.
Press SET for more than 2 seconds to
return to P1.0 without saving.
8
Page 9
5. When in P1.0, press ▲ to advance to
SETUP MODE
the next setup function or press SET for more than 2 seconds to return to
Measurement Mode.
Resetting the Pen (rSt)
When the pen is reset, all parameters
will be reset to factory default values.
1. Press SET for 2 seconds to enter setup.
2. Press ▲ to select the reset pen setting. “rst” will appear on the main display of the LCD and “P3.0” will appear on the secondary display.
3. Press SET again for less than one second to enter the reset setting.
4. Press ▲ to select Y (yes) or N (no).
9
Page 10
5. Press SET for less than one second to conrm the selection or press SET for more than 2 seconds to return to P3.0 without saving.
6. When in P3.0, press ▲ to advance to
the next setup function or press SET for more than 2 seconds to return to
Measurement Mode.
Calibration Review (CAL)
1. Press SET for 2 seconds to enter setup.
2. Press ▲ to select the calibration review setting. “CAL” will appear on the primary display of the LCD and “P4.0” will appear on the secondary display.
Range 1 Calibration Information:
3. When in P4.0, press SET for less than one second to enter P4.1 to review the last calibration concentration. If the pen is not yet calibrated, “---” will appear on the LCD.
10
Page 11
Range 2 Calibration Information:
4. When in P4.1, press ▲ to enter P4.2 to review the last calibration concentration. If range 2 is not yet calibrated, “---” will appear on the LCD.
5. When in P4.1 or P4.2, press SET for less than one second to conrm and return to P4.0.
6. When in P4.0, press ▲ to advance to
the next setup function or press SET for more than 2 seconds to return to
Measurement Mode.
Note…
P4.1 and P4.2 Modes allow review of the
calibration information but are not used
during the calibration procedure.
Selecting the Measurement Range (rAn)
There are two measurement ranges. The pen defaults to auto-ranging:
Salinity
Range 1 0 ~ 10.00 ppt
Range 2 10.1 ~ 70.0 ppt
11
Page 12
To manually select the measurement range:
1. Press SET to turn the pen on.
2. Press ▲ for more than 2 seconds to select range 1. Press ▲ for more than 2 seconds to select range 2.
CALIBRATION
Selecting the Calibration Standard Solution
For best results, select a NaCl standard near the sample value that you are
measuring. Alternatively, use a calibration solution value that is approximately ⅔ of
the full scale of the measurement range
that you plan to utilize.
Do NOT reuse the calibration solution.
Contaminants in the solution will affect the
calibration and the accuracy. Use fresh solution each time.
The previous calibration data will be replaced after re-calibrating. However,
12
Page 13
the pen will retain the calibration data for other ranges that have not yet been re-
calibrated.
Selecting the Calibration Schedule
Calibration is necessary and should be
performed regularly. Soak the probe for 15
minutes before calibration or measurement
to minimize drift. If the pen is used in the mid ranges, calibrate the unit monthly.
If the measurement is performed at extreme temperatures or within the concentration listed below, calibrate
the unit weekly.
Model Concentration
Salinity Pen < 0.10 ppt > 5.0 ppt
13
Page 14
Salinity Calibration
1. Place the probe into deionized or
distilled water for about 30 minutes to
rinse the probe.
2. Select the NaCl standard for calibration.
3. Pour 3 cm (deep) of solution into two separate containers (A & B).
4. Press SET to turn the pen on.
5. Rinse the probe in one of the containers. Gently stir the probe.
6. Dip the rinsed probe into the other container. Tap the probe on the bottom of the container to remove air bubbles. Let the probe stabilize to the solution temperature (allow 15 minutes).
7. Press CAL for more than 2 seconds to begin calibration. “CAL” and the salinity value of the solution will ash on the LCD.
8. Press ▲ or ▼ to adjust the value on the
primary display to match the value of
the standard buffer. You can adjust the
salinity reading ± 30% from the detected
value. If the detected value and the
standard value differ by more than ±
30%, clean or replace the pen.
14
Page 15
9. When “CAL” stops ashing on the
LCD, press SET for less than 1 second
to conrm the value. The pen will
return to Measurement Mode for salt
measurement. To exit Calibration Mode
without saving, press SET for more than
2 seconds. This allows you to retain the pen’s previous calibration data for the current range as you proceed. If “CAL” ashes on the LCD continually, check
that the calibration solution is stable and
that the input value (Step 8) is correct.
10. Repeat Steps 1-9 for other ranges as
needed.
Note…
If the standard value is over the measuring range or 10% less, the displayed value will be equal to the range limit or 10% of the
range limit.
15
Page 16
MEASUREMENT PROCEDURES
1. Remove the probe cover from the pen.
2.
Press SET to turn the pen on.
3. The pen defaults to auto-ranging mode.
The auto-ranging function determines and selects a range that will provide the
greatest resolution and accuracy.
4. To manually select the measurement range: Press SET to turn the pen on. Press ▲ for more than 2 seconds to select range 1. Press ▲ for more than 2 seconds to select range 2.
Note...
Accuracy is a percent of full-scale; use your pen in the lowest range to achieve
the greatest accuracy. Error codes
E02/E03 will appear on the LCD if the measured value is below or above the
limit value. Please re-select the range.
16
Page 17
5. Rinse the probe tip with de-ionized or
distilled water before use to remove
any impurities adhering to the probe. If
the probe tip is dehydrated, soak it for 30 minutes to prevent contaminants from slowing down the measurement
and accuracy.
6. Immerse the probe tip completely into the sample. Ensure that there are no air bubbles trapped in the probe slot.
Gently stir the probe to remove air bubbles, ensuring that the probe tip
is submerged while stirring.
7. Stir the probe gently to create a uniform sample. Wait about 15 minutes for the reading to stabilize.
8. The unit of measure will ash on the
LCD to indicate that the pen is in Measurement Mode. When the reading is stable, the unit of measure
will stop ashing.
Hold Function
1. Press HLD to freeze the current readings on the display. “Hold” will appear on the LCD.
17
Page 18
2. Press HLD again to release the hold function.
Turning the Pen Off
1. Press SET to turn the pen off from Measurement Mode.
2. To turn the pen off from Setup Mode,
press SET for more than 2 seconds to return to Measurement Mode, then press SET again.
Auto Power Off
1. To save battery life, the pen will
automatically turn off after 20 minutes
of inactivity. To disable this function:
2. Press SET to turn the pen off.
3. Press SET and ▼ simultaneously until “n” appears on the LCD. Release the keys to return to Measurement Mode. Auto-power-off is now disabled.
Note…
The auto-power-off function will be enabled each time the pen is turned off.
18
Page 19
SPECIFICATIONS
±1% Full Scale
Accuracy
Range
Resolution 0.01 ppt or 0.1 ppt
Calibration One point per range
TDS Factor
Temperature
Coefcient
Reference Temperature
Operating Temperature
Battery Life >80 hours continuous use
Weight 2.5 oz (68 g)
Dimensions
± 1 digit (0.0~10.0 ppt)
or ±2% Full Scale
±1 digit (10.1~70.0 ppt)
0.00~10.00 ppt (NaCl) or 10.1~70.0 ppt (NaCl)
Built-in NaCl conductivity to TDS factor
Built-in NaCl temperature
coefcient
Fixed at 25°C
0~50°C (32~122°F)
6½" x 1½" x 1½" (165 x 38 x 38 mm)
19
Page 20
ERROR CODES
Salinity
- - - The pen is in manual ranging 1 but
the salinity measurement is higher
than 10.0 ppt; Press ▲ for more
than 2 seconds to change to manual
ranging 2 or auto-ranging mode.
E03 The salt value is over the range limit
(70.0 ppt) or the pen is damaged;
Put the pen in standard buffer solution (the buffer concentration
must be lower than the range limit).
If E03 still appears, send the pen to
Sper Scientic for repair.
E04 Error in measuring original data
(damaged temperature sensor,
temperature out of specications or salinity error). E04 will disappear when the original error is resolved.
20
Page 21
Temperature
E01 The temperature circuit is
damaged; Send the pen to
Sper Scientic for repair.
E02 The temperature value is below the
range limit (0°C) or the temperature circuit is damaged; Place the pen in room temperature for 5 minutes, if E02 still appears, send the pen
to Sper Scientic for repair.
E03 The temperature value is above
the range limit (50°C) or the temperature circuit is damaged; Place the pen in room temperature for 5 minutes, if E03 still appears,
send the pen to Sper Scientic for repair.
21
Page 22
22
Page 23
Large Display
Conductivity Pen
850037
23
Page 24
SETUP MODE
The advanced Setup Mode allows you to customize the pen’s preferences and defaults:
1. Press SET to turn the pen on.
2. From Measurement Mode, press SET
for more than 2 seconds to enter Setup
Mode and change parameters.
24
Page 25
Temperature Units (tUt)
To select the Celsius or Fahrenheit temperature scale:
1. Press SET for 2 seconds to enter setup.
2. “t” will appear on the primary display of the LCD and “P1.0” will appear on the secondary display.
3. Press SET again for less than one second to enter the unit setting.
4. Press ▲ or ▼ to select C (Celsius) or F (Fahrenheit). Press SET for less than one second to conrm the unit selection.
Press SET for more than 2 seconds to
return to P1.0 without saving.
5. After conrming the temperature unit
selection, you may continue to set other temperature related parameters by pressing SET for more than 2 seconds
to return to P1.0.
6. When in P1.0, press SET for more
than 2 seconds to return to
Measurement Mode.
25
Page 26
press " " more than 2 sec to enter setup mode. Press " "or " " or " " to select momentarily again to enter unit setting.
Press " " or " " to select C /F. Press " " momentarily to confirm the unit or press " " more than 2 sec to return to P1.0 without saving.
After confirming the unit, you can continuously set another temp. related parameters (except 8371) or press " " more than 2 seconds to return to P1.0.
While in P1.0, press " " for more than 2 seconds to return to measure- ment mode.
P1.0, then press " "
Reference (Normalization) Temperature (tnr)
After saving the temperature unit setting, the pen will automatically enter the
reference temperature setup.
1. Press ▲ to select 20°C or 25°C.
2. Press SET for less than one second to conrm the value or press SET for more than 2 seconds to return to P1.0 without saving.
or
Long pres s
Long pres s
26
Shor t pres s
or
Page 27
Temperature Coefcient (tCo)
Press " "or " "
o
o
lue.
P1.1 P 1. 2 P1 .3
1. Press ▲ or ▼ to change the temperature coefcient from 0.0 to 4.0.
2. Press SET for less than one second to conrm the value or press SET for more than 2 seconds to return to P1.0 without conrming the temperature coefcient value.
27
Page 28
Resetting the Pen (rSt)
NOTE: Pay attentio n to the dif
fe- rence betwee n short or long press " " after setting the tds value.
When the pen is reset, all parameters (including calibration information) will be
reset to factory default values.
1. Press SET for 2 seconds to enter setup.
2. Press ▲ to select the reset pen setting. “rst” will appear on the main display of the LCD and “P3.0” will appear on the secondary display.
3. Press SET again for less than one second to enter the reset setting.
4. Press ▲ to select Y (yes) or N (no).
5. Press SET for less than one second to conrm the selection or press SET for more than 2 seconds to return to P3.0 without saving.
28
Page 29
Calibration Review (CAL)
1. Press SET for 2 seconds to enter setup.
2. Press ▲ to select the calibration review setting. “CAL” will appear on the primary display of the LCD and “P4.0” will appear on the secondary display.
Range 1 Calibration Information:
3. When in P4.0, press SET for less than one second to enter P4.1 to review the last calibration concentration. If the pen is not yet calibrated, “---” will appear on the LCD.
Range 2 Calibration Information:
4. When in P4.1, press ▲ to enter P4.2
to review the last calibration concentration. If range 2 is not yet calibrated, “---” will appear on the LCD.
Note…
When in P4.1 or P4.2, press SET for less than one second to conrm and return to P4.0.
29
Page 30
P4.1 and P4.2 Modes allow review of the
calibration information but are not used
during the calibration procedure.
When in Temperature Parameter (P1.0), Resetting the Pen (P3.0) or Calibration Review (P4.0) Setup Modes, press SET
for more than 2 seconds to return to
Measurement Mode.
30
Page 31
Selecting the Measurement Range (rAn)
2.Pr ess " " or " " mo re than 2 seco nds to se lect th e ran ge .
To manua l selec t the m ea sur ement rang e:
1. Turn o n the met er and st ay at nor mal meas ureme nt mode .
Pres s" " > 2 seco nd
Pres s" " > 2 seco nd
There are two measurement ranges. The pen defaults to auto-ranging
Conductivity
Range 1
Range 2
0 ~ 1999 uS
0 ~ 19.99 mS
To manually select the measurement range:
1. Press SET to turn the pen on.
2. Press ▲ for more than 2 seconds to select range 1. Press ▲for more than 2 seconds to select range 2.
(Def ault)
31
Page 32
CALIBRATION
Selecting the Calibration Standard Solution
For best results, select a conductivity standard near the sample value that
you are measuring. Alternatively, use
a calibration solution value that is approximately 2/3 of the full scale of the measurement range that you plan
to utilize.
For example, in the 0 to 1999 uS range,
use 1413 uS solution for calibration. Do NOT reuse the calibration solution.
Contaminants in the solution will affect the
calibration and the accuracy. Use fresh solution each time.
Refer to the following table. For best
results, use the recommended solution for various conductivity ranges
32
Page 33
Conductivity Measuring Range
1 0 ~ 1999 uS 600 ~ 1700 uS
2 0 ~ 19.99 mS 6.0 ~ 17.00 ms
Recommended Cal.
Solution Range
The previous calibration data will be
replaced after re-calibrating. For example,
if you previously calibrated the conductivity pen at 1413 uS in the 0 ~ 1999 uS range, when you re-calibrate at 1500 uS again (also in the 0 ~ 1999 range), the previous 1413 uS calibration point will be replaced
in this range (0 ~ 1999 uS). However,
the pen will retain the calibration data for other ranges that have not yet been re-
calibrated.
Note…
The temperature coefcient of the pen defaults to 2.1% per °C and provides good results for most applications. To reset the coefcient, see page 27. You may
also refer to Appendix B on page 47 to
calculate the temperature coefcient.
33
Page 34
The default value of the reference
temperature is set at 25°C. To normalize to a different value, see page 33. Determine
the calibration standard value for your reference temperature before resetting
the reference temperature. Contact the
solution distributor for the calibration
standard value of the solution.
Selecting the Calibration Schedule
Calibration is necessary and should be
performed regularly. Soak the probe
for 15 minutes before calibration or
measurement to minimize drift. If the pen
is used in the mid ranges, calibrate the
unit monthly.
If the measurement is performed at extreme temperatures or within the concentration listed below, calibrate
the unit weekly.
Model Concentration
Conductivity Pen < 100 uS > 2 mS
34
Page 35
Conductivity Calibration
CALIBRATION
3. Pour 3 cm (deep) of buffer solution into two separate containers (A & B).
4. Press SET to turn the meter on. Press CAL for more than 2 seconds to enter Calibration Mode.
5. Rinse the probe in one of the containers. Gently stir the probe.
6. Dip the rinsed probe into the other container. Tap the probe on the bottom of the container to remove air bubbles. Let the probe stabilize to the solution temperature (allow 15 minutes).
1. Place the probe into deionized or
2. Select the conductivity standard for
3. Pour 3 cm (deep) of calibration solution
4. Press SET to turn the pen on. Press
5. Rinse the probe in one of the
6. Dip the rinsed probe into the other
distilled water for about 30 minutes to
rinse the probe.
calibration. (Refer to page 26)
into two separate containers (A & B).
CAL for more than 2 seconds to enter
Calibration Mode.
containers. Gently stir the probe.
container. Tap the probe on the bottom of the container to remove air bubbles. Let the probe stabilize to the solution temperature (allow 15 minutes).
A (Step 5) B (Step 6)
35
Page 36
7. Press CAL for more than 2 seconds to begin calibration. The conductivity value of the solution will ash on the LCD.
8. Press ▲ or ▼ to adjust the value on
the primary display to match the value
of the standard buffer. You can adjust
the conductivity reading ± 30% from the
detected value. If the detected value
and the standard value differ by more
than ± 30%, clean or replace the pen.
Example: Standard: 10 uS; Detected value: 19 uS Adjustable range: ± 5.7 uS (19*30%) Under the above situation, the values
differ by over 30%.
Note…
If the standard value is over the measuring range or 10% less, the displayed value will be equal to the range limit or 10% of the
range limit.
Example 1: Standard: 22 uS; Detected value: 19 uS Adjustable range: ± 5.7 uS (19*30%)
36
Page 37
Although the values differ less than 30%,
the 22 uS is still over the range limit. The maximum input value is 19.99 uS.
Example 2:
Standard: 1.6 mS; Detected value: 2.1 mS Adjustable range: ± 0.63 mS (2.1*30%)
Although the measured value differs less
than 30%, the 1.6 mS is still less than the 10% range limit (1.99). The maximum input value is 1.99 mS.
9. When “CAL” stops ashing on the LCD,
press SET for less than 1 second to
conrm the value. The pen will return to Measurement Mode. To exit Calibration
Mode without saving, press SET for
more than 2 seconds. This allows you to retain the pen’s previous calibration
data for the current range as you
proceed. If “CAL” ashes on the LCD
continually, check that the calibration solution is stable and that the input
value (Step 8) is correct.
10. Repeat Steps 1-9 for other ranges
as needed.
37
Page 38
Note…
Remove the cover from the bottom of meter to expose the electrode.
Elect rode
Remove the cover from the bottom of meter to expose the electrode.
Electrode
3.
The meter is default in aut o-ranging status once powered on. Auto-ranging determi nes and selects
a range which gives you the gr eatest resolution and accuracy. Altern atively, you can manually select th e ranges. For example, if you prefer the meter to display a reading as 0.50 mS instead of 500 µS, you just need to press m
select 19.99 mS range when you
are in normal measurement mode.
ore than 2 seconds
to
See page 12 for the detail
When the pen is switched from Measurement Mode to Calibration Mode,
it will display the factory default value.
Therefore, if the pen was previously
calibrated, the display may appear to jump
to the factory default value when entering Calibration Mode
MEASUREMENT PROCEDURES
1. Remove the probe cover from the pen.
Remove the cover from the bottom of pen to
expose the electrode.
38
Page 39
Remove the cover from the bottom of meter to expose the electrode.
Electrode
3.
The meter is default in aut o-ranging status once powered on. Auto-ranging determi nes and selects a range which gives you the gr eatest resolution and accuracy. Altern atively,
you can manually select th e ranges. For example, if you prefer the meter to display a reading as 0.50 mS instead of 500 µS, you just need to press m
select 19.99 mS range when you
are in normal measurement mode.
ore than 2 seconds
to
2. Press SET to turn the pen on.
3. The pen defaults to auto-ranging mode.
The auto-ranging function determines and selects a range that will provide the greatest resolution
and accuracy.
4. The pen also provides a manual range option. For example, to view a reading of 0.50 mS rather than 500 uS: From Measurement Mode, press ▲ for more than 2 seconds to select the 19.99 mS range. See page 31 for instructions on manually selecting the range.
See page 12 for the detail
Note...
Accuracy is a percent of full-scale; use your pen in the lowest range to achieve
the greatest accuracy. Error codes
E02/E03 will appear on the LCD if the measured value is below or above the
limit value. Please re-select the range.
39
Page 40
5. Set the temperature coefcient to the correct value. The pen defaults to 2.1%
per °C, which will typically provide
good results. See page 27 to reset the temperature coefcient.
6. Select the reference temperature. The pen defaults to 25°C. To more than 2 seconds to select the 19.99 mS range. See page 26 for instructions on manually selecting the range. normalize to 20°C instead, see page 26.
7. Rinse the probe tip with de-ionized or
distilled water before use to remove any impurities adhering to the probe. If the probe tip is dehydrated, soak it for 30 minutes to prevent contaminants from slowing down the measurement
and accuracy.
8. Immerse the probe tip completely into the sample. Ensure that there are no air bubbles trapped in the probe slot.
Gently stir the probe to remove air bubbles, ensuring that the probe tip is
submerged while stirring.
40
Page 41
9. Stir the probe gently to create a uniform
8351
8351
sample. Wait about 15 minutes for the reading to stabilize.
10. The unit of measure
will ash on the LCD to
indicate that the pen is in
Measurement Mode. When
the reading is stable, the unit of measure will stop
ashing.
Hold Function
1. Press HLD to freeze the
current readings on the
display. “Hold” will appear on the LCD.
2. Press HLD again to release the hold function.
41
Page 42
SPECIFICATIONS
Accuracy
Range
Resolution 1 uS or 0.01 mS
Temperature Accuracy
Temperature Resolution
Calibration One point per range
Temperature
Coefcient
Reference Temperature
Operating Temperature
Battery Life >80 hours continuous use
Weight 2.5 oz (68 g)
Dimensions
± 1% Full Scale ± 1 digit
0 ~ 1999 uS or
0 ~ 19.99 mS
± 2°F ± 1°C
0.1°C/°F
0 ~ 4.0%/°C
20°C or 25°C
0~50°C (32~122°F)
6½" x 1½" x 1½" (165 x 38 x 38 mm)
42
Page 43
ERROR CODES
Conductivity
- - - The pen is in manual ranging 1 but
the conductivity measurement is
higher than 1999 uS; Press ▲ for
more than 2 seconds to change to manual ranging 2 or auto-ranging
mode.
E03 The conductivity value is over
the range limit (19.99 mS) or the
pen is damaged; Put the pen in standard buffer solution (the buffer concentration must be lower than
the range limit). If E03 still appears, send the pen to Sper Scientic for repair.
E04 Error in measuring original data
(damaged temperature sensor or
temperature out of specications)
results in this conductivity value error; Refer to the temperature
error codes (page 44). E04 will
disappear when the temperature
error is resolved.
43
Page 44
Temperature
E01 The temperature circuit is
damaged; Send the pen to Sper
Scientic for repair.
E02 The temperature value is below the
range limit (0°C) or the temperature circuit is damaged; Place the pen in room temperature for 5 minutes, if E02 still appears, send the pen to
Sper Scientic for repair.
E03 The temperature value is above
the range limit (50°C) or the temperature circuit is damaged; Place the pen in room temperature for 5 minutes, if E03 still appears,
send the pen to Sper Scientic for repair.
44
Page 45
Appendices
45
Page 46
APPENDIX A:
Program Function Default Note
P1.1 Select °C/°F °C Temp unit
P1.2 Reference
Temperature
P1.3 Temperature
Coefcient
P3.1 Factory
Default
P4.1 Viewing
previous calibration data
P4.2 Viewing
previous calibration data
25°C Selectable
25°C or 20°C
2.1% /°C
NO Retain
---- No cal
---- No cal
Adjustable
0.4 ~ 10%
your current settings
data for 1st range
data for 2nd range
46
Page 47
APPENDIX B:
Temperature Effect
Conductivity measurements are temperature dependent; if the temperature
increases, conductivity increases. For
example, the conductivity measured in a
0.01 M KCl solution at 20°C is 1.273 mS/
cm, whereas at 25°C, it is 1.409 mS/cm.
The concept of reference temperature
(normalization temperature) was
introduced to allow the comparison of conductivity results obtained at
different temperatures. The reference temperature is usually 20°C or 25°C. The
pen measures the actual conductivity and temperature, then converts it to the reference temperature using a temperature correction function and displays the conductivity at the
reference temperature. It is mandatory to
associate the temperature together with
a conductivity result. If no temperature
correction is applied, the conductivity
47
Page 48
exam ple, if t he T DS valu e is ppm, t he cond uctiv ity valu e must be i n uS; if th e TD S value i s in ppt, t he cond uctiv ity val ue must b e in mS.
Chec k this nu mber by m ultip lying t he cond uctiv ity read ing by th e facto r in the ab
ove fo rmula a nd the
resu lt is the TDS i n ppm.
Appendix D: Temperature Effect
Cond uctiv ity mea surem ents ar e tempe ratur e depe ndent , if the te mpera ture in creas es, con ducti vity incr eases . eg: the c onduc tivit y measu red in a 0. 01
0
M KCl so lutio n at 20 C is 1 .273 mS /cm, wh ereas , at
0
25 C, it is 1 .409 mS /cm.
The co ncept o f refer ence te mpera ture (N ormal izati on temp eratu re) was i ntrod uced to a llow th e compa rison of con ducti vity re sults o btain ed at diffe rent te mpera ture.
0 0
The re feren ce temp eratu re is usu ally 20 C or 25 C. The co nduct ivity m eter me asure s the act ual CON D. and te mpera ture an
d then c onver ts it to th e refer ence temp eratu re usin g a tempe ratur e corre ction f uncti on and di splay s the con ducti vity at t he refe rence t emp.. It is ma ndato ry to alw ays ass ociat e the tem perat ure toge ther wi th a cond uctiv ity res ult. If n o tempe ratur e corr ectio n is appl ied, th e condu ctivi ty is the v alue take n at meas ureme nt temp eratu re. The 8 30x met er used l inear t emper ature c orrec tion.
Line ar temp eratu re corr ectio n:
In mod erate ly and hi ghly co nduct ive sol ution s, temp eratu re corr ectio n can be ba sed on a li near equa tion in volvi ng a temp era
ture c oeffici ent ( ). The
coeffi cient i s usual ly expr essed a s a condu ctivi ty
0
vari ation i n %/ C. Line ar temp eratu re corr ectio n is used , e.g. fo r salin e , acid s and lea ching s oluti ons.
is the value taken at the measurement
temperature. The Large Display Salinity
(Model 850036) and Conductivity (Model
850037) Pens automatically use linear
temperature correction.
Linear Temperature Correction
In moderately and highly conductive solutions, temperature correction can be based on a linear equation involving a
temperature coefcient (θ). The coefcient
is usually expressed as a conductivity
variation in %/°C. Linear temperature
correction is used for saline, acids, and
leaching solutions.
Where:
K
= Conductivity at Tref
Tref
KT = Conductivity at T
T
= Reference temperature
ref
T = Sample temperature
100
KTr ef=
100+
( )
T
ref
T-
KT
θ = Temperature coefcient
48
Page 49
Note…
APPENDICES
The correction is accurate only within a limited temperature range around T1 and
T2. The greater the difference between T and Tref, the higher the risk of error.
Calculating Temperature Coefcients (θ)
By measuring the conductivity of a sample at temperature T1 close to Tref and another temperature T2, you can
calculate the temperature coefcient by
using the following equation:
T2 should be selected as a typical sample temperature and should be approximately
10°C different from T1. The temperature coefcients of the following electrolytes
generally fall into the following ranges:
49
Page 50
Acids: 1.0 - 1.6%/°C Bases: 1.8 - 2.2%/°C Salts: 2.2 - 3.0%/°C Drinking water: 2.0%/°C Ultra-pure water: 5.2%/°C
Average temperature coefcients of
standard electrolyte solutions are expressed as %/°C of the conductivity
value at 25°C.
Temp Range °C
15 ~25 1.725 1.863 1.882 1.981
15 ~ 25 ~35
25 ~ 35
KCl
1M
1.730 (15 ~ 27°C)
1.762 (25 ~ 27°C)
KCl
0.1 M
1.906
1.978
50
KCl
0.01 M
1.937
(15 ~ 34°C)
1.997 (25 ~ 34°C)
Satu­rated NaCl
2.041
2.101
Page 51
Battery Replacement,
Maintenance,
& Troubleshooting
51
Page 52
BATTERY REPLACEMENT
BATTERY REPLACEMENT
1. Press SET to turn the pen off.
2. Remove the battery cover located
on the top of the unit by rotating
counterclockwise. Do NOT discard the black washer.
3. Remove the old batteries and replace
with 4 new LR44 batteries, ensuring
correct polarity.
4. Replace the battery cover by rotating in the clockwise direction.
If the pen will not be in
use for a month or longer,
remove the batteries and
replace when ready to
use the pen.
52
Note…
Page 53
MAINTENANCE
Ensure that the probe is clean before use
and storage. To clean:
1. Remove the probe cap.
2. Rinse the probe carefully in deionized
water, replace the cap, and store dry in
an environment of 0 ~ 50°C.
Note…
If the probe has been exposed to a solvent that is immiscible with water, clean the probe with a solvent that is miscible with
water (i.e., ethanol or acetone), then carefully rinse with water.
53
Page 54
TROUBLESHOOTING
Pen does not turn on:
1. Press SET for more than 1 second.
2. Check that the batteries are in place,
making good contact, and placed with
correct polarity.
3. Replace with new batteries and try again.
4. Remove the batteries for one minute and then replace the batteries.
Display disappears:
If the low battery icon appears on the LCD before the display disappears, replace with
new batteries.
Air bubbles adhering to the probe:
It is typical for air bubbles to adhere to the
large slot in the probe. These air bubbles
will greatly affect the accuracy and should
be removed.
54
Page 55
Many a ir bu bbles adhe red !
To remove the air bubbles:
1. Stir the probe.
2. Dip the probe into the solution at
an oblique angle (at a slant, not perpendicular to solution). Dipping the probe perpendicularly will cause
air bubbles to form.
3. Soak the probe in solution for 15-30 minutes. Inspect the probe carefully for air bubbles.
4. If air bubbles remain on the probe, tap
the bottom of the container gently and
stir the probe.
5. If the preceding methods fail, remove
the probe from the solution and blow on
the probe to remove the air bubbles.
55
Page 56
WARRANTY
Sper Scientic warrants this product against
defects in materials and workmanship for a period of one (1) year from the date of purchase, and agrees to repair or replace
any defective unit without charge. If your
model has since been discontinued, an
equivalent Sper Scientic product will be substituted if available. This warranty
does not cover probes, batteries, battery leakage, or damage resulting from accident,
tampering, misuse, or abuse of the product.
Opening the pen to expose its electronics
will void the warranty.
To obtain warranty service, ship the unit postage prepaid to:
SPER SCIENTIFIC, LTD. 8281 E. EVANS RD., SUITE #103
SCOTTSDALE, AZ 85260
Be sure to include your name, address, phone number, and a detailed explanation
of why you are returning the item. Mail your
warranty card within 10 days or register
online at: www.sperscientic.com.
Revised 03/15/13
56
Loading...