Sper Scientific 840046 User Manual

Page 1
   
Microwave Meter
840046
Instruction Manual
Page 2
ALL RIGHTS RESERVED
Printed in the USA
The contents of this manual may not be reproduced or transmitted in any form or by any means electronic, mechanical, or other means that do not yet exist or may be developed, including photocopying, recording, or any information storage and retrieval system without the express permission from Sper Scientific.
- 2 -
Page 3
TABLE OF CONTENTS
INTRODUCTION. . . . . . . . . . . . . . .
FEATURES . . . . . . . . . . . . . . . . . . .
MATERIALS SUPPLIED. . . . . . . . .
KEYPAD. . . . . . . . . . . . . . . . . . . . . .
MEASUREMENT PROCEDURES. .
BATTERY REPLACEMENT. . . . . . .
MAINTENANCE. . . . . . . . . . . . . . . .
OPTIONAL ACCESSORIES. . . . . .
SPECIFICATIONS. . . . . . . . . . . . . .
WARRANTY . . . . . . . . . . . . . . . . . .
4
5
6
7
8
16
16
17
18
20
- 3 -
Page 4
INTRODUCTION
Measure hazardous microwave oven leakage
With use microwave oven door seals may wear enabling hazardous microwaves to leak out. Sper Scientific Microwave Meter is calibrated at
2.45 GHZ, the precise frequency of microwave ovens to accurately measure the intensity of such micro­wave leakage. Features maximum and minimum hold, an audible alarm, and a zero adjustment to eliminate background EMF. The display also indicates overload and low battery. Comes with a soft carrying case and standard 9V battery, which provides approximately 100 hours of use.
- 4 -
Page 5
FEATURES
 Frequency selection of normal
or microwave range
 Tripod mounting socket
 Maximum and minimum
 Hold function
 Auto-power-off
 Built-in alarm
 Zero offset adjustment
 Overload indicator
 Low battery indicator
- 5 -
Page 6
MATERIALS SUPPLIED
 Meter
 9 Volt Battery
 Instruction Manual
 Soft Carrying Case
- 6 -
Page 7
KEYPAD
1. Sensor position
2. LCD
3. Zero button
4. Power on/off
5. Maximum/minimum button
6. Frequency range selection button
7. Data hold button
8. Tripod mounting socket
9. Battery cover
- 7 -
Page 8
MEASUREMENT PROCEDURES
1. Press POWER to turn the meter on.
2. Press FREQ. RANGE to select
microwave (2.45 GHZ) or RF (50 MHZ~3.5 GHZ). When 2.45 GHZ is selected, “2450 MHZ” will appear on the lower right corner of the LCD. When normal range is selected, “2450 MHZ” will disappear from the lower right corner of the display.
3. For general information on microwave oven leakage and instructions on preparing for microwave measurement, see page 14.
4. Position the top of the meter to measure the electromagnetic waves. Adjust the measurement angle or position until the highest reading is obtained.
- 8 -
Page 9
MEASUREMENT PROCEDURES
Measurement Position
Figure 1 shows the correct measurement position of RF electromagnetic field
strength.
Figure 2 shows the incorrect measurement positions of RF electromagnetic field
strength.
Figure 1
Figure 2
- 9 -
Page 10
MEASUREMENT PROCEDURES
Note…
Due to background environmental magnetic fields, this meter may display a reading value under 0.002 mW/cm prior to measuring.
Maximum, Minimum and Hold
1. Press MAX/MIN to retain the maxi-
mum value. “MAX” will appear on the LCD. The meter will continue measuring, holding the maximum value on the LCD until exceeded.
2. Press MAX/MIN again to retain the
minimum value. “MIN” will appear on the LCD. The meter will continue measuring, holding the minimum value on the LCD until exceeded.
3. Press MAX/MIN again to retain the
maximum and minimum values simultaneously. “MAX MIN” will
- 10 -
2
Page 11
MEASUREMENT PROCEDURES
flash on the LCD. The meter will continue measuring, holding the maximum and minimum values on the LCD until exceeded.
4. Press and hold MAX/MIN for more
than one second to exit and return to Normal Mode.
5. Press HOLD to freeze the current
value on the display. “DH” (Data Hold) will appear on the LCD.
Press HOLD again to release
the data hold.
Auto Power Off
1. To save battery life, the meter
will automatically turn off
after 30 minutes of inactivity. To disable this function:
- 11 -
Page 12
MEASUREMENT PROCEDURES
2. Press POWER to turn the meter off. Press and hold MAX/MIN while simultaneously pressing POWER to turn the meter on. Auto-power-off is
now disabled.
3. To restore auto-power-off: Press
POWER to turn the meter off and then press POWER again to turn the meter back on. will appear on the
LCD to indicate that auto-power-off is enabled.
Alarm
The meter features a built-in alarm that will sound when the measurement exceeds 1 mW/cm2. The alarm will stop sounding when the measurement falls under 1 mW/cm
cannot be changed.
2
. This threshold value
- 12 -
Page 13
MEASUREMENT PROCEDURES
Zero Offset Adjustment
If the meter shows a measurement other than 0 due to background EMF, zero adjusting will eliminate this electromagnetic interference.
Press ZERO to adjust and obtain a
0 measurement. “ZERO” will appear on the LCD.
Overload Indicator
When the reading exceeds 4 mW/cm2, “OL” will appear on the LCD to indicate overload.
Low Battery Indicator
The symbol will appear on the
LCD to indicate a low battery. Replace the battery (see page 16).
- 13 -
Page 14
MEASUREMENT PROCEDURES
Microwave Oven Leakage
With time, microwave leakage around door seals, hinges and metal fittings is not uncommon. Therefore, it is important that all microwave ovens be periodically checked to ensure that no damage or malfunction has occurred.
Microwaves may easily pass through porcelain, glass, ceramics and most plastics. Highly potent microwave energy generated by microwave ovens is also capable of penetrating living tissue to a depth of approximately
1.18” (30 mm).
Do NOT stand or look directly into any area where microwave leakage may be present.
Remove any rotating plates, etc. before testing.
- 14 -
Page 15
MEASUREMENT PROCEDURES
1. Because microwave ovens should not be operated while empty, fill a microwave safe container with 275 ml of water and place it in the center of the oven during testing.
2. Start the microwave at maximum power.
3. Follow the measurement procedures on page 8, concentrating measurements on door seams, window seals, welds and rivets.
- 15 -
Page 16
BATTERY REPLACEMENT
1. Press POWER to turn the meter off.
2. Remove the battery cover on the back of the meter.
3. Remove the old battery and replace with 1 new 9V battery, ensuring correct polarity.
4. Replace the battery cover.
MAINTENANCE
Use a soft, dry cloth to clean the meter. Do NOT use a wet cloth, solvents, water or any liquid to clean the meter.
- 16 -
Page 17
OPTIONAL ACCESSORIES

840090 Water Resistant Instrument Pouch
 840092 Bench-Top Tripod
 840093 Field Tripod
 850000 Rubber Boot
- 17 -
Page 18
SPECIFICATIONS
Accuracy
Resolution
Display
RF Power
Density
RF Frequency
Microwave Frequency
Axis
Sample Rate
Operating
Temperature &
Humidity
± 2 dB at 2.45 GHZ
± 50 MHZ
0.001 mW/cm2
3 3/4 digits LCD;
maximum reading
3999 (4 mW/cm2)
0.003 ~ 2.700 mW/cm2
50 MHZ ~ 3.5 GHZ
2450 MHZ
Single
2.5 times per second
5°C ~ 40°C
˂80% RH
- 18 -
Page 19
SPECIFICATIONS
Storage
Temperature &
-10°C ~ 60°C ˂70% RH
Humidity
Operating
2000 M
Altitude
Operating
Environment
Battery
Indoor use;
pollution degree 2*
9V NEDA 1604, IEC
6F22 or JIS 006P
Battery Life
Approximately
100 hours
Dimensions
5 ¼” x 2 ¼” X 1 ¼”
(130 x 56 x 38 mm)
Weight
*Pollution degree is a classification of the amount of dry pollution and condensation present in an environment. This classification affects clearance distances required for product safety. Office and laboratory areas are considered pollution degree 2 environments.
6 oz (170 g)
- 19 -
Page 20
WARRANTY
Sper Scientific warrants this product against defects in materials and workmanship for a
period of five (5) years from the date of
purchase, and agrees to repair or replace any defective unit without charge. If your model has since been discontinued, an equivalent Sper Scientific product will be substituted if available. This warranty does not cover probes, batteries, battery leakage, or damage resulting from accident, tampering, misuse, or abuse of the product. Opening the meter to expose its electronics will void the warranty.
To obtain warranty service, ship the unit postage prepaid to:
SPER SCIENTIFIC LTD
8281 E. Evans Rd., Suite #103
Scottsdale, AZ 85260
The defective unit must be accompanied by a description of the problem and your return address. Register your product online at www.sperscientific.com, or return your warranty card within 10 days of purchase.
Revised 6/21/2013
- 20 -
Loading...