Sony UP-DR80MD User Manual

4-118-614-11(1)
Digital Color Printer
Instructions for Use
UP-DR80MD
© 2008 Sony Corporation
Owner’s Record
The model and serial numbers are located at the rear. Record these numbers in the space provided below. Refer to these numbers whenever you call upon your Sony dealer regarding this product.
Model No. ____________________ Serial No. ____________________
WARNING
To reduce the risk of fire or electric shock, do not expose this apparatus to rain or moisture.
To avoid electrical shock, do not open the cabinet. Refer servicing to qualified personnel only.
THIS APPARATUS MUST BE EARTHED.
To disconnect the main power, unplug the AC IN connector.
You are cautioned that any changes or modifications not expressly approved in this manual could void your authority to operate this equipment.
All interface cables used to connect peripherals must be shielded in order to comply with the limits for a digital device pursuant to Subpart B of Part 15 of FCC Rules.
For the customers in Canada
This unit has been certified according to Standard CSA C22.2 No.601.1.
For the customers in the U.S.A and Canada
When you use this product connected to 240 V single phase, be sure to connect this product to a center tapped circuit.
Important safeguards/notices for use in the medical environments
1. All the equipments connected to this unit shall be certified according to Standard IEC60601-1, IEC60950-1, IEC60065 or other IEC/ISO Standards applicable to the equipments.
This symbol indicates the equipotential terminal which brings the various parts of a system to the same potential.
Warning
This unit has no power switch. When installing the unit, incorporate a readily accessible disconnect device in the fixed wiring, or connect the power plug to an easily accessible socket-outlet near the unit. If a fault should occur during operation of the unit, operate the disconnect device to switch the power supply off, or disconnect the power plug.
For the customers in the U.S.A.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
2. Furthermore all configurations shall comply with the system standard IEC60601-1-1. Everybody who connects additional equipment to the signal input part or signal output part configures a medical system, and is therefore, responsible that the system complies with the requirements of the system standard IEC60601-1-1. If in doubt, consult the qualified service personnel.
3. The leakage current could increase when connected to other equipment.
4. For this particular equipment, all accessory equipment connected as noted above, must be connected to mains via an additional isolation transformer conforming with the construction requirements of IEC60601-1 and providing at least Basic Insulation.
5. This equipment generates, uses, and can radiate radio frequency energy. If it is not installed and used in accordance with the instruction manual, it may cause interference to other equipment. If this unit causes interference (which can be determined by unplugging the power cord from the unit), try these measures: Relocate the unit with respect to the susceptible equipment. Plug this unit and the susceptible equipment into different branch circuit.
Consult your dealer. (According to standard EN60601-1-2 and CISPR11, Class B, Group 1)
2
Important EMC notices for use in the medical environments
• The UP-DR80MD needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this instructions for use.
• The portable and mobile RF communications equipment such as cellular phones can affect the UP-DR80MD.
Guidance and manufacturer’s declaration-electromagnetic emissions
The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of the UP-DR80MD should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Group 1
Class B
Class A
Complies
Warning
The use of accessories and cables other than those specified, with the exception of replacement parts sold by Sony Corporation, may result in increased emissions or decreased immunity of the UP-DR80MD.
The UP-DR80MD uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The UP-DR80MD is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Warning
If the UP-DR80MD should be used adjacent to or stacked with other equipment, it should be observed to verify normal operation in the configuration in which it will be used.
3
Guidance and manufacturer’s declaration - electromagnetic immunity
The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of the UP-DR80MD should assure that it is used in such as environment.
Immunity test
Electrostatic discharge (ESD)
IEC 60601 test
level
±6 kV contact
±8 kV air
Compliance level Electromagnetic environment-guidance
±6 kV contact
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should
±8 kV air
be at least 30%.
IEC 61000-4-2
Electrical fast transient/burst
±2 kV for power supply lines
±2 kV for power supply lines
Mains power quality should be that of a typical commercial or hospital environment.
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
±1 kV for input/ output lines
±1 kV differential mode
±2 kV common mode
< 5% U
T
(> 95% dip in U for 0.5 cycle
40% UT (60% dip in U for 5 cycles
70% UT (30% dip in U for 25 cycles
< 5% U
T
(> 95% dip in U for 5 sec
±1 kV for input/ output lines
±1 kV differential mode
±2 kV common mode
< 5% U (> 95% dip in U
)
T
for 0.5 cycle
40% UT (60% dip in U
)
T
for 5 cycles
70% UT (30% dip in U
)
T
for 25 cycles
< 5% U
)
(> 95% dip in U
T
for 5 sec
Mains power quality should be that of a typical commercial or hospital environment.
T
Mains power quality should be that of a typical commercial or hospital environment. If the user of the UP-DR80MD requires
)
T
continued operation during power mains interruptions, it is recommended that the UP-DR80MD be powered from an uninterruptible power supply or a battery.
)
T
)
T
T
)
T
3 A/m 3 A/m Power frequency magnetic fields should be at least
characteristic of a typical location in a typical commercial or hospital environment.
4
Guidance and manufacturer’s declaration - electromagnetic immunity
The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of the UP-DR80MD should assure that it is used in such as environment.
Immunity test
IEC 60601 test
level
Compliance level Electromagnetic environment-guidance
Portable and mobile RF communications equipment should be used no closer to any part of the UP-DR80MD, including cables, than the recommended separation distance calculated from the equation appliance to the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the UP-DR80MD is used exceeds the applicable RF compliance level above, the UP-DR80MD should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the UP-DR80MD.
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
d = 1.2 P
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b
Interference may occur in the vicinity of equipment marked with following symbol:
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
5
Recommended separation distances between portable and mobile RF communications equipment and the
UP-DR80MD
The UP-DR80MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the UP-DR80MD can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (Transmitters) and the UP-DR80MD as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2P
80 MHz to 800 MHz
m
800 MHz to 2.5 GHz
d = 1.2 P
d = 2.3 P
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
Caution
When you dispose of the unit or accessories, you must obey the law in the relative area or country and the regulation in the relative hospital.
Warning on power connection
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
2. Use the Power Cord (3-core mains lead) / Appliance
If you have questions on the use of the above Power Cord / Appliance Connector / Plug, please consult a qualified service personnel.
reflection from structures, objects and people.
Appliance Connector / Plug with earthing-contacts that conforms to the safety regulations of each country if applicable.
Connector / Plug conforming to the proper ratings (Voltage, Ampere).
Warning on power connection for medical use
Please use the following power supply cord. With connectors (plug or female) and cord types other than those indicated in this table, use the power supply cord that is approved for use in your area.
United States and Canada
Plug Type HOSPITAL GRADE*
Cord Type Min.Type SJT
Minimum Rating for Plug and Appliance Couplers
Safety Approval UL Listed and CSA
*Note: Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked ‘Hospital Only’ or ‘Hospi­tal Grade’.
Min.18 AWG
10A/125V
For the customers in Europe
The manufacturer of this product is Sony Corporation, 1­7-1 Konan, Minato-ku, Tokyo, Japan. The Authorized Representative for EMC, medical devices, and product safety is Sony Deutschland GmbH, Hedelfinger Strasse 61, 70327 Stuttgart, Germany. For any service or guarantee matters, please refer to the addresses given in separate service or guarantee documents.
6

Table of Contents

Introduction
Features ...................................................................8
System Configuration .........................................8
Location and Function of Parts and Controls .....8
Front ....................................................................8
Rear .....................................................................9
Preparation
Supplied Accessories ............................................10
Connections ..........................................................10
USB Port Connection .......................................10
Installing the Printer Driver ..............................11
Operation
Before Printing .....................................................12
Notes on the Handling of the Ink Ribbon and
Paper ...............................................................12
Loading the Ink Ribbon and Paper ...................13
Replacing the Ink Ribbon and Paper ................16
Printing .................................................................17
Miscellaneous
Precautions ...........................................................19
Safety ................................................................19
Installation ........................................................19
On Transportation .............................................19
Cleaning ............................................................19
Cleaning the internals of the printer .................19
Ink Ribbon and Paper .........................................20
Specifications ........................................................21
About the Self-Laminating Color Printing Pack
(UPC-R80MD/R81MD) .................................21
Troubleshooting ....................................................22
Explanation of Indicators ....................................23
If a paper jam occurs ...........................................24
7

Introduction

Introduction

Features

The UP-DR80MD Digital Color Printer is a dye sublimation thermal transfer printer providing high quality and high resolution printing of computer image data on A4 or letter size paper in full color (256 gradations process and 16.7 million colors).

System Configuration

Computer Controls the printer operations.

Location and Function of Parts and Controls

Front

USB cable
UP-DR80MD Digital Color Printer
A 1 ON/STANDBY switch/indicator
When this switch is pressed, the indicator lights, and the printer enters the ready state. When it is pressed again, the indicator flashes and then goes out, and the printer enters the standby state.
B Paper cover
Printouts stack here.
C Paper outlet
Printouts are ejected here.
D Vent openings
E Media tray (page 13)
The ink ribbon and paper are loaded here.
F Stopper (page 17)
Flip this stopper up in order to prevent the ejected printouts from falling.
8
Features / Location and Function of Parts and Controls
Loading...
+ 17 hidden pages