The model and serial numbers are located at the rear.
Record these numbers in the space provided below.
Refer to these numbers whenever you call upon your Sony
dealer regarding this product.
Model No. ____________________
Serial No. ____________________
WARNING
To reduce the risk of fire or electric shock, do not
expose this apparatus to rain or moisture.
To avoid electrical shock, do not open the
cabinet. Refer servicing to qualified personnel
only.
THIS APPARATUS MUST BE EARTHED.
To disconnect the main power, unplug the AC IN
connector.
You are cautioned that any changes or modifications not
expressly approved in this manual could void your
authority to operate this equipment.
All interface cables used to connect peripherals must be
shielded in order to comply with the limits for a digital
device pursuant to Subpart B of Part 15 of FCC Rules.
For the customers in Canada
This unit has been certified according to Standard CSA
C22.2 No.601.1.
For the customers in the U.S.A and Canada
When you use this product connected to 240 V single
phase, be sure to connect this product to a center tapped
circuit.
Important safeguards/notices for use in the
medical environments
1. All the equipments connected to this unit shall be
certified according to Standard IEC60601-1,
IEC60950-1, IEC60065 or other IEC/ISO Standards
applicable to the equipments.
This symbol indicates the equipotential
terminal which brings the various parts of a
system to the same potential.
Warning
This unit has no power switch.
When installing the unit, incorporate a readily accessible
disconnect device in the fixed wiring, or connect the power
plug to an easily accessible socket-outlet near the unit.
If a fault should occur during operation of the unit, operate
the disconnect device to switch the power supply off, or
disconnect the power plug.
For the customers in the U.S.A.
This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15
of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when
the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in
accordance with the instruction manual, may cause
harmful interference to radio communications. Operation
of this equipment in a residential area is likely to cause
harmful interference in which case the user will be
required to correct the interference at his own expense.
2. Furthermore all configurations shall comply with the
system standard IEC60601-1-1. Everybody who
connects additional equipment to the signal input part
or signal output part configures a medical system, and
is therefore, responsible that the system complies with
the requirements of the system standard IEC60601-1-1.
If in doubt, consult the qualified service personnel.
3. The leakage current could increase when connected to
other equipment.
4. For this particular equipment, all accessory equipment
connected as noted above, must be connected to mains
via an additional isolation transformer conforming with
the construction requirements of IEC60601-1 and
providing at least Basic Insulation.
5. This equipment generates, uses, and can radiate radio
frequency energy. If it is not installed and used in
accordance with the instruction manual, it may cause
interference to other equipment. If this unit causes
interference (which can be determined by unplugging
the power cord from the unit), try these measures:
Relocate the unit with respect to the susceptible
equipment. Plug this unit and the susceptible
equipment into different branch circuit.
Consult your dealer. (According to standard EN60601-1-2
and CISPR11, Class B, Group 1)
2
Important EMC notices for use in the medical environments
• The UP-DR80MD needs special precautions regarding
EMC and needs to be installed and put into service
according to the EMC information provided in this
instructions for use.
• The portable and mobile RF communications equipment
such as cellular phones can affect the UP-DR80MD.
Guidance and manufacturer’s declaration-electromagnetic emissions
The UP-DR80MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the UP-DR80MD should assure that it is used in such an environment.
The use of accessories and cables other than those
specified, with the exception of replacement parts sold by
Sony Corporation, may result in increased emissions or
decreased immunity of the UP-DR80MD.
The UP-DR80MD uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
The UP-DR80MD is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Warning
If the UP-DR80MD should be used adjacent to or stacked
with other equipment, it should be observed to verify
normal operation in the configuration in which it will be
used.
3
Guidance and manufacturer’s declaration - electromagnetic immunity
The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of the
UP-DR80MD should assure that it is used in such as environment.
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
±8 kV air
be at least 30%.
IEC 61000-4-2
Electrical fast
transient/burst
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be that of a typical commercial or
hospital environment.
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common
mode
< 5% U
T
(> 95% dip in U
for 0.5 cycle
40% UT
(60% dip in U
for 5 cycles
70% UT
(30% dip in U
for 25 cycles
< 5% U
T
(> 95% dip in U
for 5 sec
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common
mode
< 5% U
(> 95% dip in U
)
T
for 0.5 cycle
40% UT
(60% dip in U
)
T
for 5 cycles
70% UT
(30% dip in U
)
T
for 25 cycles
< 5% U
)
(> 95% dip in U
T
for 5 sec
Mains power quality should be that of a typical commercial or
hospital environment.
T
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the UP-DR80MD requires
)
T
continued operation during power mains interruptions, it is
recommended that the UP-DR80MD be powered from an
uninterruptible power supply or a battery.
)
T
)
T
T
)
T
3 A/m3 A/mPower frequency magnetic fields should be at least
characteristic of a typical location in a typical commercial or
hospital environment.
4
Guidance and manufacturer’s declaration - electromagnetic immunity
The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of the
UP-DR80MD should assure that it is used in such as environment.
Portable and mobile RF communications equipment should
be used no closer to any part of the UP-DR80MD, including
cables, than the recommended separation distance
calculated from the equation appliance to the frequency of
the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the UP-DR80MD is used exceeds the applicable RF
compliance level above, the UP-DR80MD should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the UP-DR80MD.
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance
in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey, a should be less than the
compliance level in each frequency range. b
Interference may occur in the vicinity of equipment marked
with following symbol:
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
5
Recommended separation distances between portable and mobile RF communications equipment and the
UP-DR80MD
The UP-DR80MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the UP-DR80MD can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (Transmitters) and the UP-DR80MD as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
m
800 MHz to 2.5 GHz
d = 1.2 √P
d = 2.3 √P
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
Caution
When you dispose of the unit or accessories, you must
obey the law in the relative area or country and the
regulation in the relative hospital.
Warning on power connection
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
2. Use the Power Cord (3-core mains lead) / Appliance
If you have questions on the use of the above Power Cord
/ Appliance Connector / Plug, please consult a qualified
service personnel.
reflection from structures, objects and people.
Appliance Connector / Plug with earthing-contacts that
conforms to the safety regulations of each country if
applicable.
Connector / Plug conforming to the proper ratings
(Voltage, Ampere).
Warning on power connection for medical
use
Please use the following power supply cord.
With connectors (plug or female) and cord types other than
those indicated in this table, use the power supply cord that
is approved for use in your area.
United States and Canada
Plug TypeHOSPITAL GRADE*
Cord TypeMin.Type SJT
Minimum Rating for Plug
and Appliance Couplers
Safety ApprovalUL Listed and CSA
*Note: Grounding reliability can only be achieved when the equipment is
connected to an equivalent receptacle marked ‘Hospital Only’ or ‘Hospital Grade’.
Min.18 AWG
10A/125V
For the customers in Europe
The manufacturer of this product is Sony Corporation, 17-1 Konan, Minato-ku, Tokyo, Japan.
The Authorized Representative for EMC, medical devices,
and product safety is Sony Deutschland GmbH,
Hedelfinger Strasse 61, 70327 Stuttgart, Germany.
For any service or guarantee matters, please refer to the
addresses given in separate service or guarantee
documents.
6
Table of Contents
Introduction
Features ...................................................................8
System Configuration .........................................8
Location and Function of Parts and Controls .....8
Front ....................................................................8
Explanation of Indicators ....................................23
If a paper jam occurs ...........................................24
7
Introduction
Introduction
Features
The UP-DR80MD Digital Color Printer is a dye
sublimation thermal transfer printer providing high quality
and high resolution printing of computer image data on A4
or letter size paper in full color (256 gradations process
and 16.7 million colors).
System Configuration
Computer
Controls the printer
operations.
Location and Function of
Parts and Controls
Front
USB cable
UP-DR80MD Digital Color Printer
A 1 ON/STANDBY switch/indicator
When this switch is pressed, the indicator lights, and
the printer enters the ready state. When it is pressed
again, the indicator flashes and then goes out, and the
printer enters the standby state.
B Paper cover
Printouts stack here.
C Paper outlet
Printouts are ejected here.
D Vent openings
E Media tray (page 13)
The ink ribbon and paper are loaded here.
F Stopper (page 17)
Flip this stopper up in order to prevent the ejected
printouts from falling.
8
Features / Location and Function of Parts and Controls
Loading...
+ 17 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.