21919 30th Drive SE
Bothell, WA 98021-3904
USA
Telephone: 1-888-482-9449 or 1-425-951-1200
Fax: 1-425-951-1201
SonoSite Ltd
Alexander House
40A Wilbury Way
Hitchin, Herts
SG4 OAP UK
T: +44-1462-444800
F: +44-1462-444801
Caution:
M-Turbo, SiteLink, SonoCalc, SonoHD, SonoMB, and SonoSite are registered trademarks or trademarks of SonoSite, Inc.
DICOM is the registered trademark of the National Electrical Manufacturers Association for its standards publications relating to digital communications
of medical information.
Non-SonoSite product names may be trademarks or registered trademarks of their respective owners.
Index ........................................................................................................................ 51
iv
Chapter 1: Introduction
Before servicing the M-Turbo ultrasound system, please read this manual. The information applies only to the
SonoSite M-Turbo ultrasound system product manufactured after December 5, 2007.
The ultrasound system has multiple configurations and feature sets. All are described in this service manual but
not every option may apply to your system. System features depend on your system configuration, transducer, and
exam type.
Refer to the M-Turbo Ultrasound System User Guide for additional information regarding safety, system controls,
operation, capabilities, and specifications.
Audience
The intended audience of this manual is properly trained field and in-house service personnel.
Conventions
These conventions are used in this service manual:
•A WAR NIN G describes precautions necessary to prevent injury or loss of life.
•A Caution describes precautions necessary to protect the products.
• Numbered steps must be performed in a specific order.
• Bulleted lists present information in list format but do not imply a sequence.
Labeling symbols are in the user guide.
Contact Information
Questions and comments are encouraged. SonoSite is interested in your feedback regarding the service manual.
If you encounter difficulty with the system, use the information in this manual to help correct the problem. If the
problem is not covered here, contact SonoSite Technical Support as follows:
Technical Support (USA, Canada)1-877-657-8118
Technical Support fax:1-425-951-6700
Technical Support e-mail:service@sonosite.com
SonoSite website:www.sonosite.com (Select Resources > Support & Service)
International Technical Support:Contact your local representative or call (USA) +425-951-1330
European Service Center+44-(0)1462-444-800
Japan Service Center+81-3-5304-5337
e-mail: uk.service@sonosite.com
Chapter 1: Introduction1
2Chapter 1: Introduction
Chapter 2: System Overview
About the System
The SonoSite M-Turbo high-resolution ultrasound system is a portable, full featured, general purpose, software
controlled, diagnostic ultrasound system using all digital architecture. The system is used to acquire and display
high-resolution, real-time ultrasound data in 2D, M Mode, Pulsed Wave (PW) Doppler, Continuous Wave (CW)
Doppler, Color Power Doppler (CPD), and color Doppler (Color) or in a combination of these modes.
The system has an electrocardiography (ECG) display feature and supports a 3-lead ECG cable assembly to collect
data for M Mode and Doppler measurements. The system provides measurement capabilities for anatomical
structures and fetal biometry that provide information used for clinical diagnostic purposes. The system has a PW
and CW Doppler audio output feature and cine review, image zoom, labeling, biopsy, measurements and
calculations, image storage and review, printing, and recording capabilities.
The system includes the ability to measure the intima-media thickness (IMT) of the carotid artery using digital
ultrasound images. The IMT measurement of the carotid artery may be used adjunctively with other medical data
obtained by a physician to help assess the cardiovascular health of a patient.
The system includes Digital Imaging and Communications (DICOM) capabilities as well as general computer
communication capabilities to provide the acceptance, transfer, display, storage, and digital processing of
ultrasound images and loops. Security support is also provided to facilitate HIPAA compliance.
The system/transducer is capable of exceeding a TI or an MI of 1.0 in certain operating modes or mode
combinations. The system displays the current output level in terms of one of two bioeffects indices (“Mechanical
Index [MI]” and “Thermal Index [TI]”) in accordance with the AIUM/NEMA Standard for Real Time Display of
Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment.
Chapter 2: System Overview3
Theory of Operation
AQ BusRF Bus
Acquisition
subsystem
Processing
subsystem
Transduce r
Display
subsystem
Control Bus
Control
subsystem
User
interface
Battery
pack
assembly
Pulser voltage
Video
External video to monitor,,
printer
Power
subsystem
Power
adapter
External power
IrDA
Serial Bus
Logic power
Display power
The M-Turbo ultrasound systemhas seven (7) major functional groups:
•Transducer
•Acquisition Subsystem
• Processing Subsystem
•Display Subsystem
• Control Subsystem
• User Interface Subsystem
•Power Subsystem
Figure 2.1 is a system block diagram that shows the relationship of the functional groups.
Figure 2.1 SonoSite High-Resolution Ultrasound System (M-Turbo) Block Diagram
The Transducer elements convert the pulser voltage to acoustic energy during the transmit portion of the
ultrasound acquisition cycle. The elements convert the acoustic echo to voltage in the receive portion of the
acquisition. The voltage developed on the transducer elements is sensed by the acquisition subsystem. The system
transducers have 64 to 192 elements.
The Acquisition Subsystem consists of the beamformer and interface to the transducer. The beamformer controls
the timing of the transmit pulses to focus the acoustic beam. The beamformer amplifies the low-level received
echos and controls the receive focusing. The system beamformer transmits on up to 128 elements and receives on
64 elements.
The Processing Subsystem includes capabilities for interfacing with the beamformer and performing high speed
processing. The processing subsystem demodulates, filters, detects, and compresses the signal supplied by the
beamformer into display information.
4Chapter 2: System Overview
The Display Subsystem converts the detected ultrasound data into picture elements (pixels). The software user
interface graphics are combined with the ultrasound information and converted to a video stream. The external
video port supports NTSC and PAL format.
The Control Subsystem consists of the central processing unit, program and video memory, permanent image
storage and retrieval memory, external communication interface ports, and connection to the user interface keys.
The control software includes the acoustic power and intensity software subsystem, power group monitors, and a
beamformer monitor. This software guarantees a level of patient safety by ensuring the system is operating within
acoustic power and intensity limits.
The User Interface Subsystem represents the software interface and form factor. The software interface is the
interaction between the user and the screen layout components. The form factor is the type of physical buttons,
location, and grouping of the buttons and the device size, shape, and weight. Dedicated controls are for high usage
activities and grouped according to the user workflow.
The Power Subsystem provides the system power and protects the hardware from destructive and/or unsafe
conditions by detecting failures in the system through hardware and software monitors. Detection of a fault results
in disabling of the pulser supply, and signaling of an error to the Control Group. The power subsystem includes the
battery pack and battery charging electronics.
Description of Operating Modes
2D Mode2D mode is a two dimensional image of the amplitude of the echo signal. It is used for
location and measurement of anatomical structures and for spatial orientation during
operation of other modes. In 2D, a two-dimensional cross-section of a 3-dimensional soft
tissue structure such as the heart is displayed in real time. Ultrasound echoes of different
intensities are mapped to different gray scale or color values in the display. The outline of the
2D cross-section may be a rectangle, parallelogram, trapezoid, sector, or a full circle,
depending on the particular transducer used. 2D mode can be used in combination with any
other modes.
MModeM Mode is also known as “T-M mode” or “time-motion” mode. It is used primarily for cardiac
measurements such as valve timing and septal wall thickness when accurate timing
information is required.
Ultrasound echoes of different intensities are mapped to different gray scale values in a
scrolling display. M Mode displays time motion information of the ultrasound data derived
from a stationary beam. Depth is arranged along the vertical axis with time along the
horizontal axis. M Mode can be used alone but is normally used in conjunction with a 2D
image for spatial reference. The 2D image has a graphical line (M-line) superimposed on the
2D image indicating where the M Mode beam is located.
Chapter 2: System Overview5
Color
Doppler
(Color)
In color Doppler, a real-time, two-dimensional cross-section of blood flow is displayed. The
2D cross-section may be presented as a rectangle, parallelogram, trapezoid, sector, or a full
circle, depending on the particular transducer used.
The 2D cross-section is presented as a full color display, with various colors being used to
represent the velocity, both positive and negative, of the blood flow echoes. Often, to
provide spatial orientation, the full color blood flow cross-section is overlaid on top of the
gray scale cross-section of soft tissue structure (2D echo). For each pixel in the overlay, the
decision of whether to display VCD, gray scale (echo) information or a blended combination
is based on the relative strength of echoes from the soft-tissue structures and from the red
blood cells.
A high pass filter (wall filter) is used to remove the signals from stationary or slowly moving
structures. Tissue motion is discriminated from blood flow by assuming that blood is moving
faster than the surrounding tissue, although additional parameters may also be used to
enhance the discrimination. The remaining signal after wall filtering may be averaged over
time (persistence) to present a steady state image of blood flow distribution. Variance
information may also be displayed to provide information when large variance is observed in
the velocity information.
Color Power
Doppler
(CPD)
Continuous
Wave (CW)
Doppler
In CPD, a real-time two-dimensional cross-section of blood flow is displayed. The 2D
cross-section may be presented as a rectangle, parallelogram, trapezoid, sector, or a full
circle, depending on the particular transducer used.
The 2D cross-section is presented as a full color display, with various colors being used to
represent the power in blood flow echoes. Often, to provide spatial orientation, the full color
blood flow cross-section is overlaid on top of the gray scale cross-section of soft tissue
structure (2D echo). For each pixel in the overlay, the decision of whether to display CPD, gray
scale (echo) information or a blended combination is based on the relative strength of
echoes from the soft-tissue structures and from the red blood cells.
A high pass filter (wall filter) is used to remove the signals from stationary or slowly moving
structures. Tissue motion is discriminated from blood flow by assuming that blood is moving
faster than the surrounding tissue, although additional parameters may also be used to
enhance the discrimination. The power in the remaining signal after wall filtering may be
averaged over time (persistence) to present a steady state image of blood flow distribution.
CW provides a real-time representation of blood flow and is displayed as a
velocity-versus-time sweeping output. Velocity (or frequency) is presented as the vertical
axis with time along the horizontal axis. The magnitude of the detected signal is represented
as different gray scale values.
CW Doppler mode provides the clinician with the ability to obtain blood flow velocities
focused about a user specified focal region. A continuous transmit waveform of ultrasound
energy with a known frequency is transmitted and focused by the system; on the receive
side, the transducer receive echoes are continuously amplified, focused about the focal
region and converted to a base band quadrature signal. The signal is analyzed by a
quadrature phase detector that establishes two receive channels to allow detection of flow
direction. These two channels are then analyzed by a fast complex Fourier transform (FFT)
circuit to establish the spectrum of frequencies present in the echoes. The data are displayed
as spectrum frequencies with respect to time.
CW can be used alone but is normally used in conjunction with a 2D image for spatial
reference. The 2D image has a graphical line (D-line) superimposed on the 2D image
indicating where the M-mode beam is located.
6Chapter 2: System Overview
Pulsed Wave
(PW) Doppler
PW provides a real-time representation of blood flow and is displayed as a
velocity-versus-time sweeping output. Velocity (or frequency) is presented as the vertical
axis with time along the horizontal axis. The magnitude of the detected signal is represented
as different gray scale values. The ultrasound data is derived from a single area, the sample
volume, on a stationary beam.
PW Doppler mode provides the clinician with the ability to obtain blood flow velocities
about a spatial sample volume. A burst of ultrasound with a known spectrum is transmitted
by the system; on the receive side, the transducer receive echoes are amplified and range
gated at the appropriate depth. The signal is analyzed by a quadrature phase detector that
establishes two receive channels to allow detection of flow direction. These two channels are
then analyzed by a fast complex Fourier transform (FFT) circuit to establish the spectrum of
frequencies present in the echoes. The data are displayed as spectrum frequencies with
respect to time.
PW can be used alone but is normally used in conjunction with a 2D image for spatial
reference. The 2D image has a graphical line (D-line) superimposed on the 2D image
indicating where the M-mode beam is located. The sample volume position (depth) and size
are also indicated on the D-Line.
Additional System Feature Performances
Broadband ImagingThis ultrasound acquisition system uses high resolution broadband technology in
the transmit pulsers, transducer, and receivers. The receive path can capture and
process signals over a wide spectrum, from below 2.0 MHz to beyond 10 MHz. For
each application, the transmit pulse is designed to produce an appropriate
bandwidth. For example, in 2D grayscale imaging, a wide band pulse is used to
support good axial resolution. For Doppler modes, a narrower band pulse is used,
which improves the spectral resolution of the detected Doppler signal.
In addition to transmit pulse control, programmable digital signal processing is used
in the receive path to further refine the bandwidth used to produce the final image.
Digital filters are applied to the digitized received signal to limit and shape the
spectral bandwidth used to generate the displayed output.
Tissue Specific
Imaging
Biopsy GuidanceThe system can display a pair of biopsy guidelines that represent the anticipated
Measurement and
Calculation
Capabilities
In this feature, parameters for signal and image processing are optimized to
maximize the image quality or to obtain the best compromise of resolution and
penetration for different specific clinical applications. These parameters include: the
order of received filters, the bandwidth, the dynamic range, the compression curve,
the gain setting and parameters for compounding frequency band, etc. For
example, different system parameter setups are used for abdominal or peritoneal
scanning. This feature is for ease of use for the operator by automatically setting up
system control parameters rather than manually adjusting settings for best
performance.
path of the biopsy needle. The image of an anatomical target, biopsy guidelines, a
scan plane marker, and a biopsy needle are displayed to assist in guiding the biopsy
needle to the target. The system also provides needle guidance for vascular access
procedures. For additional information, see the biopsy user guides.
The system offers a variety of measurements and calculations, specific to exam type
and transducer. A list of them , and author references, are in the system user guide.
Measurement accuracy is also discussed.
Chapter 2: System Overview7
Continuous Wave
Doppler Audio
Output
The system provides for audio output of the CW velocity information. This can be
presented as stereo information, with flow moving towards the transducer on one
channel and flow away on the other, or as a mono output with the single audio
output representing the summation of the flow directions.
Pulsed Wave Doppler
Audio Output
Electrocardiograph
(ECG) Display
ECG Module
The ECG module allows a representation of the heart electrical activity to be displayed in real time with ultrasound
images acquired and displayed on the system video display.
The ECG module interfaces to the patient through three (3) ECG leads: Right Arm ECG lead (RA), Left Arm ECG lead
(LA), and Left Leg ECG lead (LL). The ECG received signal from the ECG electrodes are isolated, amplified, and
filtered by the ECG module before it is sent to the system for further processing and display.
The ECG module and cable are an integrated assembly. The module receives power from the system. Patient
isolation is provided by the ECG module, allowing the connection and signals to the system to be system-ground
referenced. The isolation between the patient and the system meets the requirements of IEC 601-1 for Type BF
equipment.
The system provides for audio output of the PW velocity information. This can be
presented as stereo information, with flow moving towards the transducer on one
channel and flow away on the other, or as a mono output with the single audio
output representing the summation of the flow directions.
ECG is provided to measure the electrical signal generated by the heart. A three lead
interface: Right Arm (RA), Left Arm (LA) and Left Leg (LL), is provided on the system.
The ECG signal is displayed as an amplitude-versus-time sweeping output.
Amplitude is presented on the vertical axis with time along the horizontal axis.
DICOM
The system features Digital Imaging and Communications (DICOM) capability to provide the acceptance, transfer,
display, storage, and digital processing of single ultrasound images as well as loops of ultrasound images.
IMT
The system includes the ability to measure the intima-media thickness (IMT) of the carotid artery using digital
ultrasound images. The intima is that region of the arterial wall from and including the endothelial surface at the
lumen to the luminal margin of the media. The media layer extends from the intima to the adventitia of the vessel
wall. The adventitia is normally quite echogenic on ultrasound images when compared to the media. The IMT
measurement of the carotid artery may be used adjunctively with other medical data obtained by a physician to
help assess the cardiovascular health of a patient.
System Specifications
This section contains system and accessory specifications and agency approvals. The specifications for
recommended peripherals can be found in the manufacturers’ instructions. See the applicable SonoSite accessory
user guide for information on the accessories.
8Chapter 2: System Overview
System Dimensions
Length: 11.8 in. (29.97 cm)
Width: 10.8 in. (27.43 cm)
Height: 3.1 in. (7.87 cm)
Weight: 8.5 lbs. (3.9 kg) with the C60x transducer and battery installed
Display Dimensions
Length: 8.4 in. (21.34 cm)
Height: 6.3 in. (16 cm)
Diagonal: 10.4 in. (26.4 cm)
Transducers
C11x/5-2 MHz 11 mm curved array (6 ft./1.8 m)
C60x/5-2 MHz 60 mm curved array (5.5 ft./1.7 m)
HFL38x/13-6 MHz 25 mm linear array (5.6 ft./1.7 m)
ICTx/8-5 MHz 11 mm intracavitary array (5.5 ft./1.7 m)
L25x/13-6 MHz 25 mm linear array (7.5 ft./2.3 m)
L38x/10-5 MHz 38 mm linear array (5.5 ft./1.7 m)
P21x/5-1 MHz 21 mm phased array (6 ft./1.8 m)
Imaging Modes
2D (256 gray shades)
Color power Doppler (CPD) (256 colors)
Color Doppler (Color) (256 colors)
Continuous Wave (CW) Doppler
MMode
Pulsed wave (PW) Doppler
Tissue Doppler Imaging (TDI)
Tissue Harmonic Imaging (THI)
Image and Clips Storage
The number of images and clips you can save varies with imaging mode and file format.
Accessories
Hardware, Software, and Documentation
Barcode Scanner
Battery
Biopsy Guide
Carry case
Chapter 2: System Overview9
ECG Cable (6 ft/1.8m)
External display
Footswitch
Kensington Security Cable
Mini-Dock
Mobile Docking System Lite II (MDS Lite II)
Mobile Docking System M Series (MDSm)
Needle Guide
Power supply
Quick Reference Guide
SiteLink Image Manager 4.0
SonoCalc IMT
System User Guide
System AC PowerCcord (10 ft / 3.1 m)
Triple Transducer Connect
Video and printer cables
Cables
See the M-Turbo Ultrasound System User Guide, MDSm User Guide, and the MDS Lite II User Guide for information on
cables.
Peripherals
Peripherals include the following medical grade (conforming to the requirements of EN60601-1) and non-medical
grade (commercial) products. Manufacturer’s instructions accompany each peripheral. System setup instructions
are in the M-Turbo Ultrasound System User Guide. Instructions for using peripherals with the system are in the
applicable SonoSite accessory user guide.
Medical Grade
Black-and-white printer
Recommended sources for printer paper: Contact CIVCO at 1-800-445-6741 or www.civco.com to order
supplies or to find the local distributor.
Color printer
DVD recorder
15” External monitor
Non-Medical Grade
USB Memory Stick
10Chapter 2: System Overview
Temperature, Pressure, and Humidity Limits
Note: The temperature, pressure, and humidity limits apply only to the ultrasound system and transducers.
Operating Limits: System
• 10–40°C (50–104°F), 15–95% R.H.
• 700 to 1060hPa (0.7 to 1.05 ATM)
Operating Limits: Battery
• 10–40°C (50–104°F), 15–95% R.H.
• 700 to 1060hPa (0.7 to 1.05 ATM)
Operating Limits: Transducer
10–40°C (50–104°F), 15–95% R.H.
Shipping/Storage Limits: System without Battery
• -35–65°C (-31–149°F), 15–95% R.H.
• 500 to 1060hPa (0.5 to 1.05 ATM)
Shipping/Storage Limits: Battery
• -20–60°C (-4–140°F), 0–95% R.H.*
• 500 to 1060hPa (0.5 to 1.05 ATM)
* For storage longer than 30 days, store at or below room temperature.
• 10–40°C (50–104°F), 15–95% R.H.
Shipping/Storage Limits: Transducer
• -35–65°C (-31–149°F), 15–95% R.H.
Electrical
Power Supply Input: 100-240 VAC, 50/60 Hz, 2.0 A Max @ 100 VAC.
Power Supply Output 115 VDC, 5.0A Max (system)
Power Supply Output 212 VDC, 2.3A Max (battery)
Combined output not exceeding 75W.
Battery
6-cell, 11.2 VDC, 5.2 amp-hours, rechargeable lithium ion battery pack.
Run time is up to 2 hours, depending on imaging mode and display brightness.
Chapter 2: System Overview11
Electromechanical Safety Standards
EN 60601-1:1997, European Norm, Medical Electrical Equipment–Part 1. General Requirements for Safety.
EN 60601-1-1:2001, European Norm, Medical Electrical Equipment–Part 1. General Requirements for
Safety–Section 1-1. Collateral Standard. Safety Requirements for Medical Electrical Systems.
EN 60601]2]37:2001 + Amendment A1:2005, European Norm, Particular requirements for the safety of ultrasonic
medical diagnostic and monitoring equipment.
CAN/CSA C22.2, No. 601.1]M90, Canadian Standards Association, Medical ElectricalEquipment.Part 1. General
Requirements for Safety (including CSA 601.1 Supplement 1:1994 and CSA 601.1 Amendment 2:1998)
.CEI/IEC 61157:1992, International Electrotechnical Commission, Requirements for the Declaration of the Acoustic
Output of Medical Diagnostic Ultrasonic Equipment.
UL 60601]1 (1st Edition), Underwriters Laboratories, Medical Electrical Equipment] Part 1: General Requirements
for Safety.
EMC Standards Classification
EN 60601-1-2:2001, European Norm, Medical Electrical Equipment. General Requirements for Safety-Collateral
Standard. Electromagnetic Compatibility. Requirements and Tests.
CISPR11:2004, International Electrotechnical Commission, International Special Committee on Radio Interference.
Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment Electromagnetic Disturbance
Characteristics-Limits and Methods of Measurement.
The Classification for the SonoSite system, SiteStand, accessories, and peripherals when configured together is:
Group 1, Class A.
Airborne Equipment Standards
RTCA/DO]160E:2004, Radio Technical Commission for Aeronautics, Environmental Conditions and Test Procedures
for Airborne Equipment, Section 21.0 Emission of Radio Frequency Energy, Category B.
DICOM Standard
NEMA PS 3.15: 2000, Digital Imaging and Communications in Medicine (DICOM)-Part 15: Security Profiles.
HIPAA Standard
The Health Insurance and Portability and Accountability Act, Pub.L. No. 104-191 (1996).
45 CFR 160, General Administrative Requirements.
45 CFR 164, Security and Privacy.
12Chapter 2: System Overview
Chapter 3: Troubleshooting
This chapter contains information to help you correct problems with system operation and provides instructions
on the proper care of the system, transducer, and accessories.
Periodic Maintenance
There is no recommended periodic or preventive maintenance required for the system, transducers, or accessories.
There are no internal adjustments or alignments required. There are no functions that require periodic testing or
calibration. Performance tests are described in Chapter 5, “Performance Testing” of this manual. Performing
maintenance activities not described in this manual may void the product warranty.
Local regulations may require electrical safety testing.
Contact SonoSite Technical Support for any maintenance questions.
System and Subsystem Diagnosis
This section covers basic diagnostic and troubleshooting procedures you may follow if the system does not
operate properly. To diagnose system failures, consult the referenced diagnostic figures that follow or SonoSite
Technical Suppor t.
Table 3.1: Troubleshooting Subassemblies and Diagnostic Figures
SubassembliesDiagnostic Figures or Table
DICOMTab le 3. 2
DipslayTBA
BatteryTBA
Control PanelTBA
System Repair
The system is repairable through subassembly replacement or through replacement of parts as recommended by
SonoSite in Chapter 4, “Replacement Procedures.” Component level repair of Printed Circuit Board Assemblies is
performed only at the SonoSite repair facility. Replacement of board level components by unauthorized service
facilities voids the SonoSite warranty.
Test Equipment
Test equipment is not required for this troubleshooting section. Troubleshooting test aids include an external
monitor and a spare battery.
Chapter 3: Troubleshooting13
Failure (Assert) Codes
Assert code
The system displays an “assert screen” for hardware and software issues related to main PCBA failures. Main PCBA
failures typically result in “assert codes” that are output to the display. If an assert screen appears, note the assert
code and contact SonoSite Technical Support to clarify the failure. Figure 3.1 shows an assert screen. The assert
code is the bracketed number on the line labeled “C:”.
Figure 3.1 Assert Screen
Verifying a System Assert Code
System asserts are caused by hardware and/or software faults. Hardware asserts typically require main PCBA
replacement. Software asserts can be reset and the system may recover. A simple method to identify the cause of
the assert is identified here:
Assert Cause1Record the assert code.
2Press and release the Power button to power the system down.
3Press the Power button again to power on the system.
• If the system powers on normally, it has recovered from the fault (software assert) and
you may use the system.
• If the assert condition remains, corrective action must be taken; usually replacement of
the main PCBA is required. Contact SonoSite Technical Support for assistance and to
obtain repair parts.
If the Power button is not functional, all sources of power must be removed to allow the
system to power down. I.e., disconnect AC power and remove the battery.
14Chapter 3: Troubleshooting
Loading...
+ 40 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.