2. Starting up
2.1 Installing the scale elements
The indicator must be installed in such a way that it can be
easily disconnected from the mains.
Insert the batteries in all four load receptors. As soon as the
batteries are inserted, the load receptors start to transmit and
try to establish a radio connection to the display unit.
Attention (Fig. 1- 4 right):
Before switching on the balance, place the load receptors under
the 4 wheels of the hospital bed. Place the load receptor under
the wheels as far as it will go. Press the lever down and lift the
bed wheel. (The load receptors are designed to take wheel
diameters of 130-160 mm! Smaller diameters optional.
Due to its design, the scale may only be used on level oors.
For installation sites in Germany this is fullled if the oors
comply with the tolerances for surface-ready oors according to
DIN 18202, table 3, line 4.
For other countries, the respective applicable national standards
can be used as a basis.
2.2 Switch on / switch off
Switch the display unit on or off using the ON/OFF button.
„SOEHNLE PROFESSIONAL“ appears in the display together with a search
symbol until the radio connection to the load receptors is established. This
usually takes 20-60 seconds.
As soon as the radio connection to the load receptors is established, the current
gross weight value appears in the display.
Please ensure that the socket is freely accessible.
2.3 Radio contact
The Bluetooth radio connection works over a maximum distance of 5 m if there is no partition between
the load receptor and the display unit. After brief interruptions (e.g. replacement of a battery or loss of
contact of a battery due to vibration), the connection is automatically re-established within a few
seconds.
The following symbol appears during connection setup:
4
Page 5
2. Starting up
2.4 Switching off / ending operation safely
Press the ON/OFF key to switch the unit off. By unplugging the mains plug from
the socket, the device is disconnected from the mains on all poles.
Operation is thus safely terminated.
If the sensors are loaded or if you have called up a menu, the balance only
switches off after 5 seconds by pressing the ON/OFF key.
It is not possible to switch off the balance while it is waiting for an entry in the
editing area of the setting mode. To switch off, the editing area must be left.
The load receptors have no on/off switch, they are switched off by removing the
batteries.
OFF indicator
If the balance is switched off, but the power supply remains on, „OFF“ appears in
the display.
2.5 Charging the batteries
Open the lid of the battery charging station.
There are four charging bays for simultaneous charging of four batteries.
The four LEDs indicate the charge status of each battery.
The fth LED on the charging station is the mains indicator of the charger.
2
STARTING UP
LED status display Charger (per charging compartment):
LED lights orange: Battery is charging
LED lights green: battery charged - trickle charge
LED status display of the load receptor:
Display unit is switched on (radio connection active)
LED lights green: battery charged
LED ashes green, mostly on, only off for a short time:
Battery still has enough capacity for trouble-free operation
LED ashes green, mostly off, only on for a short time:
Battery almost empty, battery must be charged
Display unit is off or in stand-by mode
LED lights green for approx. 2 minutes and then starts to signal the standby
state by ashing (mainly off).
Charging time for empty batteries is approx. 4 hours.
Batteries should be charged and replaced in sets.
5
Page 6
2. Starting up
The batteries can remain in the charging station for any time.
Display of the connection status:
Press the INFO key + 1. The following will then appear, for example, in the
display:
Battery life:
The capacity of the batteries is sufcient for approx. 12 hours of continuous
operation.
If the balance is mainly switched off (90% OFF), the batteries can remain in the
load receptors for up to 125 hours before they have to be charged.
Attention: Even when the display unit is switched off, the Bluetooth connection
is regularly activated, which is why the batteries are also discharged in this state
and must be recharged after 125 hours at the latest.
Recommendation:
After switching off the display unit, always remove the batteries from the load
receptor and place them in the battery box or charging station. This ensures that
the balance works reliably when it is put back into operation.
If the bed scale is not used, the batteries must be removed from the load
receptors to prevent deep discharge. Eight batteries are included.
6
Page 7
3. Input and control
The device is operated via the keyboard.
The keys are assigned different functions in the different modes.
3.1 Keyboard for input and control
3
INPUT AND CONTROL
ORG key:
Call up organizational data
F key: Function key /
Cursor up /
Cursor left
TARA key /
Cursor down /
Cursor right
INFO key:
e. g. electronic nameplate
ZEROING key
ENTER key /
Conrmation key
ON/OFF key
SPECIAL CHARACTER key
ZERO key: by pressing +, - and
the space character
3.2 Alphanumeric input
Alphanumeric input can be made with the combined number/letter keypad.
Types of input elds
Display elds without write option
Display/write elds with numeric input option
Display/write elds with alphanumeric input option
7
Page 8
3. Input and control
Input procedure
Numeric elds: The digits 0 to 9 can be entered directly using the KEYPAD. The entry is right-aligned. A
cursor ashes under the last digit entered. The C key deletes the digit marked with the cursor. The
NULLSTELL key cancels the entry without accepting new input values, existing values are retained. The
ENTER key accepts the entered value and switches back to the application program.
The specication of whether point, decimal point or no decimal separation is made in the setting mode
(Setting mode > Terminal > Display > Decimal separation, see page 14). The input eld can only be
described up to the specied number of characters, further input is not displayed and is not accepted.
Alphanumeric elds: The number keys are also assigned to letters (uppercase only) and characters in
multiple assignment. The o (zero) key is followed, for example, by +, - and the space character. In
addition, the SPECIAL CHARACTER button is available.
If a number key is pressed again within one second after it has been pressed, the characters assigned
to the key are activated on a rolling basis. If no further key is pressed within one second, the called
character is accepted and the cursor moves one position further.
Pure numerical entries can be made directly one after the other without waiting time, as long as a
number is not entered several times in succession. A cursor ashes under the position to be entered.
Use the F key and the TARE key to move the cursor left and right. The C key deletes the last character to
the left of the cursor or the character under which the cursor is ashing.
The ZEROING key cancels the entry without accepting new input values, existing values are retained.
The ENTER key accepts the entered value and switches back to the application program. The input eld
can only be described up to the specied number of characters, further input overwrites the last
character.
3.3 Display screen
The display shows different views depending on the situation.
It is divided into the following display elds:
Secondary displayToolbar
Display panel
Additional display
Displays e.g. BMI, gross, net. Can be switched on with the F key according to the program settings.
Toolbar
Shows information about the status of the active measurement loop.
Display panel
Displays weight value, dimension indicator.
8
Page 9
3. Input and control
3.4 Symbols / Signs
Icons in the toolbar
Alibi memory ll level 100%
Alibi memory ll level 80%
Alibi memory ll level empty
Key lock
Multirange/Multidivision 1
Multirange/Multidivision 2
Multirange/Multidivision 3
Resolution x10 (10 times)
Zero
Hold function active
Dimension sign
Signs for stability
After placing the load (bed, patient, etc.) on the scale, it takes time to settle in. As long as no stability
has been reached, the dimension sign (kg, lb) is hidden. If stability has been reached and a stable value
is available, this is indicated by the insertion of the dimension character.
3
INPUT AND CONTROL
9
Page 10
4. Setting mode
4.1 Keyboard in setting mode
ORG key:
Leaving the setting mode
T key: (TARA key)
Moving the selection down
in the list or cursor to the
right
KEYBOARD
ON/OFF key
F key:
Move the selection up in the
list or cursor to the left
PRINT/ENTER key:
Select the marked menu item
or accept an entry or change
C key:
Delete the last digit /
character
10
Page 11
4. Setting mode
4.2 Overview of menu navigation
Program settings
Terminal
Scales
Data Processing
General
Weighing + taring
Version
Display
Keypad
Date/Time
Organisation data
User password
Reset terminal
Scale parameters
Data transfer fixed values
Alibi memory
Interface 1
Hold mode
Fixed tare value
Clear fixed tare value
Auto standby time
Plausibility check sensor
Provisional tare
F key
Calibrated counter status*
approvable*
Filter setting*
Zero tracking*
Zero range at power-on*
Second unit*
Empty message*
Serial number MS*
Software version MS*
Application
Baud rate*
…*
Dialysis function
Hold mode
Second unit
BMI function
4
SETTING MODE
Print template
I/O port settings*
Barcode*
Service
* Only for service personnel authorised by Soehnle Professional
(password-protected)*
Printout request
Printout when W+T=O
Decimal separator
Special thousands separator
Hide unit
Consec. no. print key
Factory setting
11
Page 12
4. Setting mode
4.3 Entering the setting mode
Turn the unit on. After the switch-on routine, the device automatically starts the
last activated application program.
To enter the setting mode, rst press the ENTER key and then the F key
simultaneously. You are now in the Setting mode selection menu.
4.4 Input in setting mode
In most cases, you make a selection from various settings offered.
Editable values can be entered numerically or alphanumerically via the KEYPAD.
4.5 Settings in Setting Mode
You can use the parameters in the setting mode to adapt the „Weighing and taring“ application program
to your individual requirements.
4.6 Settings in the „Program Settings“ menu
In the „Program settings“ setting mode, you can adapt the user program of your balance to your specic
requirements - when operating as a bed scale, this is always „weighing and taring“.
4.7 Settings in the „Terminal“ menu
In the „Terminal“ setting mode, you will nd the following query and setting options.
Version
Software status Display only
User software Display only
Serial number Display only
Serial number main board Display only
Display
BacklightLuminosity setting, input in percent via numeric keypad;
Contrast
Language
Decimal division
Second unit
12
0% = dark, 100% = full luminosity; factory setting: 80%.
Contrast setting, input in percent via numeric keypad; 1% is lowest
colouring, 100% is full colouring; factory setting 50%.
German/English/French
Comma, dot or none; factory setting: comma
g/t/lb/mg/oz (additional eld in the display)
Page 13
4. Setting mode
Keyboard
Keyboard lockSingle, multiple or all keys can be locked. By pressing the key to be
Key beep
Receipt beep
Date / Time
TimeInput via KEYBOARD, clock runs from conrmation of ENTER key
Date
Date format
Organizational data
Name ORG AEnter the name of the organizational data memory A; digits or text via
Name ORG B/
ORG C/ ORG D
User password
Create / ChangeInput of the password with up to 8 characters via keyboard; after
Delete
locked, the status of the key can be changed from open to locked and
vice versa.
Press the ENTER key for 3 seconds to conrm.
Selection „off“ or „on“ conrmation beep at the
Key operation. Factory setting: „off
Selection „off“ or „on“ acknowledgement level for the
Execution of a function. Factory setting: „off
Input via KEYBOARD
Selection of European or imperial date;
Factory setting: European
KEYBOARD, e.g. „Bed number“, „Room number“, max. 12 characters
Further input options for other organizational data memories, see
above.
activation, the setting mode can only be reached via the password. If
the password is forgotten, entering the word CLEAR gives access.
Deletes the password and releases access again.
Reset Terminal
Do not execute / execute
„Execute“ deletes all individual settings of the terminal and resets it to
the factory settings.
13
Page 14
5. Functions
5.1 Power On Zeroing
At each power-on, the weight value transmitted by the load receptors is set to zero as long as it is less
than 50% of the max. weighing range. This tares the empty bed and the weight display goes to zero.
5.2 Manual zeroing
Manual zeroing corrects small deviations from the zero point, e.g. due to dirt on
the balance. The following character appears in the toolbar when zero is
displayed: >0<.
Zero setting range not veriable: -1 to +3% of the weighing range.
If zeroing is not possible, the error message „Above zero range“ or „Below zero
range“ appears for two seconds.
5.3 x10 (times 10)-Display
Displays the weight value with a further decimal in 10 times higher resolution.
Press the ENTER and then the ZERO button simultaneously to switch to the x10
display.
As long as the 10 times higher display is selected, x10 appears in the toolbar.
The function is deactivated when the unit is switched off.
5.4 Weighing and taring
Display view
5.5 Weighing with determined taring
Determination and taring of a bed weight without patient.
Place the empty bed on the scale, including bed linen. Press the TARE key. The tare value is stored, the
weight display goes to zero. Now the patient can be put to bed. The weight is displayed with the note
Net.
Display value in tare memory
To display the tare memory, press the INFO key, then the TARE key.
The ZEROING key ends the display.
Manually changing the tare value afterwards
Display tare value by pressing the INFO key, then the TARE key. Enter the new value using the KEYPAD.
The ZERO key cancels the entry, existing value is retained or new tare value is saved by pressing the
TARE key. Conrm with the ENTER key and return to the application program.
14
Page 15
5. Functions
5.6 Weighing with manually entered tare
If the patient is already in bed, it can be manually tared if the bed weight is known.
Enter the bed weight (tare weight) via the KEYBOARD. The entry appears in the display with the current
weight unit. Now press the TARE key. The entry is stored in the tare memory and the tare weight is
subtracted from the weight display. The weight of the patient determined in this way is displayed with
the note Net.
Display value in tare memory
To display the tare memory, press the INFO key, then the TARE key.
The NULLSTELL key ends the display.
Manually changing the tare value afterwards
Display tare value by pressing the INFO key, then the TARE key. Enter the new value using the
KEYBOARD. The ZERO key cancels the entry, existing value is retained or new tare value is saved by
pressing the TARE key. Conrm with the ENTER key and return to the application program.
5.7 Weighing with stored tare values
Known bed weights can be stored as tare values in a power-failure-proof memory and called up as
required (100 different bed weights are possible).
Each memory location has:
• Three-digit consecutive number
• Name (text 18 characters)
• Tare value
5
FUNCTIONS
Enter tare weights in xed tare value memory
• Open the input menu
(Setting mode > Program settings > General > Fixed tare values)
• Press the ENTER key and F key simultaneously to enter the setting mode.
• The arrow points to „Program Settings“ and conrm with the ENTER key.
• Conrm the „General“ display with the ENTER key.
• Clock to „Fixed tare values“ with the TARE key and conrm with the ENTER key.
Entering the values
• If no tare memory is yet occupied, the rst free, otherwise the last selected xed tare memory is
displayed.
• Press the ORG key and then the ENTER key, the input eld for the designation of a new tare memory
is displayed.
• Enter the name using the KEYBOARD and conrm with the ENTER key.
• The arrow points to „Tare“. Press the ENTER key to display the input eld for the Tare value is
displayed.
• Enter the tare value and conrm with the ENTER key.
• Use the ZEROING key to return to tare memory selection.
15
Page 16
5. Functions
• If additional tare memories are to be used, select the next free memory location by pressing the
ORG key and repeat the procedure described above.
• To return to the weighing mode, press the ZERO key and then the ORG key.
Transfer of xed tare values from a PC to the display device
• For this you need a special software.
For more information, please contact the Soehnle Professional service staff.
Changing or overwriting the xed tare value memory
• Open the input menu
(Setting mode > Program settings > General > Fixed tare values)
• Press the ENTER key and F key simultaneously to enter the setting mode.
• The arrow points to „Program Settings“ and conrm with the ENTER key.
• Conrm the „General“ display with the ENTER key.
• Clock to „Fixed tare values“ with the TARE key and conrm with the ENTER key.
Correction of values
• The last selected xed tare value memory is displayed. Use the F or TARE key to select the tare
memory to be changed or enter the memory location number and conrm with the ENTER key.
• Press the ENTER key, the input eld for the name of the tare memory is displayed.
• Change the name using the KEYBOARD and conrm with the ENTER key.
• The arrow points to „Tare“.
• Press the ENTER key, the input eld for the tare value is displayed.
• Change the tare value via the KEYBOARD and conrm with the ENTER key.
• Use the ZEROING key to return to tare memory selection.
• If you want to change other tare memories, use the F or TARE key to select the next memory location
to be changed and repeat the procedure described above.
• To return to the weighing mode, press the ZEROING key and then the ORG key.
Calling up the xed tare value memory list for taring
The tare value from the xed tare value memory can be activated when the balance is unloaded (i.e.
before weighing) or when the balance is loaded (during weighing).
• The display shows a weight value ≥ 0.000 kg.
• First press the ENTER key and then the TARE key simultaneously. The display shows the last xed
tare value memory used.
• Use the F or TARE key to move to the desired xed tare value memory or enter the memory location
number and conrm with the ENTER key.
• Activate the xed tare value by pressing the ENTER key. The display shows the gross value minus
the tare value (net).
• Further tare values from the memory are activated by carrying out the procedure described above.
• ENTER-Taste und F-Taste gleichzeitig drücken, um in den Einstellmodus zu gelangen.
• Der Pfeil zeigt auf „Programmeinstellungen“, mit ENTER-Taste bestätigen.
• Anzeige „Allgemein“ mit ENTER-Taste bestätigen.
• Mit TARA-Taste zu „Tarafestwerte“ takten und mit ENTER-Taste übernehmen.
Deleting a tare value
• Der zuletzt angewählte Tarafestwertspeicher wird angezeigt. Mit der F- oder TARA-Taste zum
gewünschten Tarafestwertspeicher takten oder die Speicherplatz-Nr. eingeben und mit der
ENTER-Taste bestätigen.
• Mit der C-Taste den angezeigten Tarafestwertspeicher löschen.
• Die entstandene Lücke in der fortlaufenden Nummerierung wird geschlossen. Die nachfolgenden
Tarafestwertspeicher erhalten eine um 1 reduzierte Nummer.
ATTENTION: Under Setting Mode > Program Settings > General > Delete Fixed Tare Values all xed
values are deleted from the memory!
5.8 Intermediate tare / dialysis functionn
Setting and using the operating mode
• Setting in setting mode
(Setting mode > Program settings > Weighing + taring > F key > Dialysis function)
• Press the ENTER key and F key simultaneously to enter the setting mode.
• The arrow points to „Program Settings“. Conrm with the ENTER key.
• Scroll down to „Weighing + taring“ with the TARE key and conrm with the ENTER key.
• Conrm the „F key“ display by pressing the ENTER key.
• Press the TARE key (down) or F key (up) to select „Dialysis function“ and conrm with the ENTER key.
An acoustic acknowledgement is made.
• Press the ORG key to exit the setting mode.
The gross value is displayed.
Use of the dialysis function
• Press the F key (when the scale is loaded).
Message: „ Put on tare“. Put on the additional tare. Press the TARE key to conrm.
The additional tare is stored in the tare memory, the net value remains unchanged.
• Repeat the procedure for additional tare. The weight is displayed with the message „Net“.
If the balance is unloaded, it shows the tare weight with a negative sign.
• Delete tare by pressing the ZEROING key.
Display value in tare memory
To display the tare memory, press the INFO key, then the TARE key.
The ZEROING key ends the display.
Manually changing the tare value afterwards
Display tare value by pressing the INFO key, then the TARE key. Enter the new value using the
KEYBOARD. The ZEROING key cancels the entry, existing value is retained or new tare value is saved by
pressing the TARE key. Conrm with the ENTER key and return to the application program.
5
FUNCTIONS
17
Page 18
5. Functions
5.9 Error message when taring
„Error 06 Taring not possible“
Possible causes:
• Empty scale
• Balance in overload or underload
• Manual tare input with tare value above maximum load
5.10 Display of the tare memory
To display the tare memory, press the INFO key, then the TARE key.
The ZEROING key ends the display.
5.11 BMI-Function
The Body Mass Index (BMI) is used to assess whether people are normal or overweight.
The BMI is calculated according to the following formula:
body weight (kg)
BMI =
body size (m) x body size (m)
The unit of the BMI is therefore kg/m2.
BMI classication by gender
Classicationmw
Underweight
Normal weight
Overweight
Adiposites
massive Adiposites
18
< 20
20 - 2519 - 24
25 - 3024 - 30
30 -4030 - 40
> 40
< 19
> 40
Page 19
5. Functions
Setting and using the operating mode
• Setting in setting mode
(Setting mode > Program settings > Weighing + taring > F key > BMI function)
• Press the ENTER key and F key simultaneously to enter the setting mode.
• The arrow points to „Program Settings“. Conrm with the ENTER key.
• Scroll down to „Weighing + taring“ with the TARE key and conrm with the ENTER key.
• Conrm the „F key“ display by pressing the ENTER key.
• The arrow points to the last activated function.
• Clock to „BMI function“ with the TARE key and conrm with the ENTER key.
A beep conrms the takeover.
• Press the ORG key to exit the setting mode. The weight value is displayed.
Using the BMI function
• Press the F key, the message „Enter body size“ appears.
Enter your body height in meters.
• Press the ENTER button to conrm.
Now the BMI value is displayed in parallel to the person weight in the secondary display.
• The BMI display can be switched off by pressing the F key again.
5.12 Hold-Function
• Setting in Setting Mode
(Setting mode > Program settings > General > Hold mode)
The hold function „Hold mode“ freezes the measured value depending on the selected setting in the
display. The following settings are possible:
5
FUNCTIONS
not active: Holdmode is off.
Silence/key: Value freezes, hold mode is cancelled by pressing the ON/OFF key.
Silence/empty: Value freezes, Holdmode is canceled by unscaling the balance.
Max/key: Max. Value (at stability) freezes, hold mode is cancelled by pressing the
ON/OFF key.
Max/empty: Max. Value (at stability) freezes, hold mode is canceled by unloading the scale.
Drag/key:Max. Value (on movement) freezes, hold mode is cancelled by pressing the ON/OFF key.
Drag/empty: Max. Value (on movement) freezes, hold mode is canceled by unloading the scale.
19
Page 20
5. Functions
5.13 Printing
The 3020 program terminal can be connected to a printer via the serial interface to print out weighing
slips. A maximum of 15 images can be stored.
The Soehnle Professional Service Centres will provide service software for the 3020/3030 for designing
printed images with individual content.
The parameters listed below in the standard print images can be used to create individual print images.
In addition, organizational data (identiers) and individual parameters such as company names can be
recorded. Please contact your Soehnle Professional service partner for print images. The print is
triggered either by the ENTER key or by another print-triggering condition, e.g. F key.
Conguring the interface for the printer
• Setting in Setting Mode
(Setting mode > EDP > Interface 1 > Use > 2795.11)
• Press the ENTER key and F key simultaneously to enter the setting mode.
• The arrow points to „Program Settings“. Clock to „EDP“ with the TARE key.
Press the ENTER key.
• The arrow points to „Data transmission“.
Clock to „Interface 1“ with the TARE key.
Press the ENTER key.
• The arrow points to „Use“. Press ENTER key.
• The arrow points to „2795.11“. Use the TARE or F key to select the printer to be connected.
• Conrm with the ENTER key.
Connection options (factory setting):
none -
2795.11 (Soehnle Professional Printer)
2795.12 (Soehnle Professional Printer)
EDV (PC)
Barcode (Barcode reader)
Large display S302 (Screen)
About the printer
The Soehnle Professional printers are included as standard, but other printers can also be created with
the 3020/3035 service software. The other parameters baud rate, data bits, parity and XON/XOFF can
be set according to the information from the
printer manual. For this purpose, please contact Soehnle Professional at
authorized service personnel. The parameters for Soehnle Professional printers are set at the factory.
20
Page 21
5. Functions
Standard print images, factory setting
Print images can be assigned to the print-triggering keys using the service software. The following print
image is already stored and assigned in the factory setting, so it can be printed out. This standard print
image is an example that shows which parameters are available.
Date
Time
Scale no.
Gross
Tar e
Print-triggering conditions
The service software is used to assign print images to print triggering conditions.
The following conditions can be assigned to print images:
• Print key (ENTER key)
• Auto imprint/data record once immediately
• Auto imprint/data record once after weight change at stability
• Auto printing/data record after weight change at stability
• Autoprint/data record after weight change at stability above the empty signal
• Autoprint/data record after weight change at stability above the empty signal if the empty signal
was active before.
• Send current
• ENTER/Capture key when entering ORG A
• ENTER/Capture key when entering ORG B
• ENTER/Capture key when entering ORG C
• ENTER/Capture key when entering ORG D
• Key sequence INFO, F and ENTER (depending on setting)
• ENTER/Capture key for manual input
• Print alibi memory
5
FUNCTIONS
5.14 Assignment of the F key
You can use the F key to call the functions. This takes on different functions depending on the
occupancy. The current function is selected in the setting mode.
You can enter the setting mode by simultaneously pressing the ENTER and then the
F key.
In the setting mode, you can nd the assignment options for the F key via Program Settings > Weighing
+ Taring > F key. Select the desired function with the F or TARE key and conrm with the ENTER key.
The following functions are available:
Intermediate tare
Dialysis function
Hold fashion
Second Unit
BMI function
Exit the setting mode by pressing the ORG key.
21
Page 22
6. Error messages
Error no.CauseRemedy / Info
05Zeroing not possibleDeviation from zero point too large
06Taring not possible
07Pressure not allowed
08No unit switching
09Print character buffer full
10
11Measuring point signal
12
13Converter does not work
15Measuring point write-protected
16No measuring point available
18Measurement point is not set to zero
20Print format empty
23Print failed
24Corrupt print image memory
25No printed image(s)
26Printer update aborted
30Alibi memory full
33Content Alibi memory corrupt
34Alibi memory empty
35Alibi memory missing or defective
36Cannot write alibi
37Cannot read alibi
40General data format errors
41Input value less than permitted
42Input value greater than permitted
44Switching off not allowed
50Value below zero setting limit
51Value above zero setting limit
52OverloadMeasured value over weighing range
53UnterloadMeasured value below zero
55Place weight
75Replacing the backup battery
76Charging the batteries
80Cancel external command
81Timeout external command
82Number already existsThe ID assigned to a xed value already exists
90ROM occupied / fullA total of 100 memory locations are available
98Barcode wrong
Overow Load cell signal
Check sensor position
Empty scale, scale in underload or overload,
manual tare input with tare value above maximum load
Incorrect lever position with at least one load receptor or no weight
on at least one load receptor (plausibility check active)
Zeroing not possible, because measured value is outside zeroing limit
Zeroing not possible, because measured value is outside zeroing limit
The balance/scale is empty or not yet unloaded according to the
set unload factor
22
Page 23
7. Technical Information
7.1 Technical Data
Weighing range: see nameplate
Division: see nameplate
Min. / max. load: see nameplate
Max. bed weight: 250 kg
Ambient temperature for operation: –10°C to +40°C, rapid temperature changes,
e.g. due to cold drafts, can impair the measurement result
Storage temperature: –40°C to +70°C
Air pressure: 950 to 1050 hPa
Humidity: 20% to 85% relative humidity, non-condensing
Power supply of the display unit: Power supply unit/voltage 100 to 240 V / 50 to 60 Hz
Power supply of the load receptors: Via 6V lead-gel battery, battery capacity: 1200 mA
Continuous operation approx. 12 hours
Transport conditions: Transport the device only in its original packaging,
maintain storage temperature, shake and avoid shocks.
Membrane keyboard: 19 keys, Function key freely assignable
Organizational data memory: 4 pieces, alphanumeric input via numeric keyboard
Display: Dot matrix 240 x 64 Pixel, dimmable backlit
Memory capacity: Fixed memory for 100 xed values, cross-program, power-failure-proof, optional alibi memory for 2 million
entries
Interfaces: 1 x RS 232, more possible, Bluetooth Qualication
Material: Program terminal: Housing made of plastic;
Load receptor: Aluminium
Wheels: ø 130–160 mm, width max. 71 mm
7
TECHNICAL INFORMATION
7.2 Further technical information
Circuit diagrams, spare parts lists and adjustment instructions can be requested from the manufacturer‘s customer service.
7.3 Electronic nameplate
The device has an electronic nameplate.
Display after pressing the INFO button.
Exit the display by pressing the INFO key again.
Serial number of the balance (balance type, year of
manufacture, counting number)
EU conformity marking
Serial number of the load receptors
Product number of the load receptors
Bluetooth address
Protection class
Power supply
Current draw
7.5 Explanation of the symbols on the packaging
Attention fragile
Observe storage direction during transport
Protect from moisture and wetness
Maintain storage temperature
24
Page 25
8. General Informationen
8.1 Intended use
The device is used to determine the weight of lying persons and may only be used for this purpose. Any
other use is prohibited. This balance is intended for use in legal metrology. It complies with the
applicable requirements of EC Directives 2004/108/EC, 2006/95/EC and 2009/23/EC.
The manufacturer‘s product advisor (address see address) must be notied in writing of any errors that
could endanger the patient and of errors that could falsify the measurement results. These operating
instructions and the operating instructions for the 3020 display terminal are part of the device. They
must be kept close to the equipment. The exact observance of these instructions is a prerequisite for
the proper use and correct operation of the device.
If the device does not work as intended, there is a suspicion of damage. The device must then be sent to
a service station authorised by Soehnle Professional. Only original spare parts may be used for repairs
by an authorized service station. These original parts are listed in the service documentation (spare
parts list: 470.014.037 balance / 470.630.273 evaluation unit / 470.014.038 load receptor) with order.
No. described.
Soehnle Professional will only assume responsibility for the safety of the device if the instructions are
observed and the device is operated in accordance with the operating instructions. The device may only
be used by persons who, due to their training, have knowledge that ensures proper handling. Before
using the device, the user must ensure the functional safety and the proper condition of the device. The
operator must be familiar with the operation of the device. The device is not intended for hazardous
areas. Potentially explosive areas may be created by the use of ammable anesthetics, skin cleansing
agents and skin disinfectants.
8
GENERAL INFORMATION
8.2 CE labeling
Das Produkt trägt das CE- Zeichen
nach folgenden Richtlinien:
EMV-Richtlinie: 2004/108/EC
Niederspannungsrichtlinie: 2006/95/EC
Waagenrichtlinie: 2009/23/EC
8.3 Classication
Electrical protection class II (insulated, with double insulation and functional earth). Water
protection according to EN60529: IPX2 for the whole device.
Application part
The balance has no application part. No part is intended for direct contact by the patient. However, the
leakage currents can be measured on the load shoes according to the classication of Type B application parts.
25
Page 26
8. General Information
8.4 Safety instructions
This device is designed for professional use only. Please read the information in the operating instructions carefully before using the device. They contain important information for installation, proper use
and maintenance of the device. The manufacturer is not liable if the following instructions are not observed: When using electrical components under increased safety requirements, the corresponding
regulations must be observed.
Never carry out work on the device while it is live.
There are no special requirements for the personnel, the display unit can be used everywhere. It is not
intended for use in critical environments (CT, debrillator). The product cannot be used in potentially
explosive environments. The mains-operated indicator must be installed in such a way that it can be
easily disconnected from the mains and the display (make sure that the socket is accessible). The
display unit must not get wet.
This device is interference suppressed in accordance with the applicable EC Directive 2004/108/EC.
However, under extreme electrostatic and electromagnetic inuences, e.g. when operating a radio or
mobile phone in the immediate vicinity of the device, the display value may be affected. At the end of the
disturbance, the product can be used again as intended; it may have to be switched on again. In case of
permanent electrostatic interference, please contact the responsible service partner.
The instrument is a measuring instrument. Drafts, vibrations, rapid temperature changes
and solar radiation can inuence the weighing result.
The scale complies with protection class IPX2 and must not be exposed to high humidity, vapours,
aggressive liquids or heavy soiling.
Improper installation will void the warranty. Electrical connection conditions must correspond to the
values printed on the power supply unit. The devices are designed for operation in buildings. Observe the
permissible ambient temperatures for operation (see Technical Data). The device meets the require-
ments for electromagnetic compatibility. Exceeding the maximum values specied in the standards
must be avoided. If used as directed, there is no risk of other devices being affected by this product. In
case of problems, contact your service partner.
26
Page 27
8. General Information
8.5 Cleaning
Before each cleaning, disconnect the device from the mains by unplugging the mains plug. The unit may
only be cleaned with a damp cloth. Under no circumstances may water penetrate into the device.
Disinfectant may only be used on the membrane keypad of the display.
The following disinfectants are permitted:
Methylated spirits; Isopropanol; 2% Cohrsoline; 1% Aqueous Sokrena solution; 5% Sagrotan; 5% Gigasept.
Spraying of the device and the connection plug is not permitted.
8.6 Maintenance and service
Repairs may only be carried out by a body authorised by Soehnle Professional using original spare parts.
The balance is equipped with a non-removable power cord. The mains cable must not be replaced by
maintenance personnel.
If the device does not work as intended, there is a suspicion of damage. The device must then be sent to
a service station authorised by Soehnle Professional. Only original spare parts may be used for repairs
by an authorized service station. These original parts are described in the service documentation with
order number.
8.7 Warranty / Guarantee / Liability
If the delivered goods are defective for which the manufacturer is responsible, the manufacturer is entitled
either to remedy the defect or to deliver a replacement. Replaced parts become the property of the manufacturer.
If the remedy of the defect in the replacement delivery fails, the statutory provisions shall apply.
The warranty period is 24 months and begins on the day the product is rst purchased. Please keep the
invoice as proof. In case of service please contact your dealer or the manufacturer customer service.
In particular, no warranty is assumed for damages arising from the following reasons:
Unsuitable, improper storage or use, faulty assembly or commissioning by the customer or by third parties, natural wear and tear, changes or interventions, faulty or negligent treatment, in particular excessive
stress, chemical, electrochemical, electrical inuences or moisture, unless these can be attributed to the
manufacturer. Should operational, climatic or other inuences lead to a signicant change in the condi-
tions or material condition, the guarantee for the perfect overall function of the devices shall lapse. The
warranty period for wearing parts (e.g. batteries) is 6 months.
8
GENERAL INFORMATION
Keep the original packaging for possible return transport!
27
Page 28
8. General Information
8.8 Disposal of batteries and accumulators
Batteries and rechargeable batteries containing harmful substances are marked with
the symbol of a crossed-out dustbin and may not be disposed of with household waste. As a consumer you are legally obliged to return used batteries and accumulators.
You can dispose of your old batteries and rechargeable batteries as hazardous waste
at public collection points in your community or wherever batteries of this type are
sold. These signs can be found on batteries containing harmful substances: Pb = battery contains lead, Cd = battery contains cadmium, Hg = battery contains mercury.
Lead-gel batteries are used to operate the load receptors. You can hand them in at public collection points
or return them to the dealer when you make a new purchase.
8.9 Disposal of the scale
According to the current state of knowledge, the device does not contain any substances that are particularly
hazardous to the environment. This product is not to be treated as normal waste, but must be returned to a
collection point for the recycling of electrical and electronic equipment. For more information, contact your
local authority, municipal waste management company or the company from which you purchased the product.
8.10 Service interface
Can only be used for service purposes.
28
Page 29
9. Warnings
In case of malfunction, contact the dealer or the manufacturer. Unauthorized modications or repairs
may damage your scale and void the manufacturer‘s warranty.
Indicators must not get wet. Liquids (e.g. water) can damage the displays. Use a dry cloth, e.g. a
towel, to dry the display.
Disconnect the power supply to this unit before installation, cleaning or maintenance. Doing so could
damage the unit.
If the display is not to be used for a long period of time, remove the power supply or switch off the
display when the battery is running. Recharge the batteries after 4 weeks at the latest to compensate
for self-discharge.
Avoid stacking materials on the display or placing weights on the display. This can lead to damage.
Place the balance on a rm, stable and level surface to ensure accurate measurement results. With
carpets or with inclination the measurement results change.
Do not connect the display to unstable power sources.
Use only the original equipment. The use of other brands can lead to damage to the display.
9
WARNINGS
29
Page 30
Notes