1
WD1160 - WD4060 - WD4190 Series
- PROGRAM TABLE -
WD1160 Series: instrument washers, 60cm, without
drying system.
WD4060 Series: instrument washers, 60cm, equipped with forced
air drying system.
WD4190 series: instrument washers, 90cm, with
forced-air drying system and side cabinet to take
detergent cans.
WARNING
This document is an appendix to the appliance's user manual and must be consulted together with the manual for proper, safe use of the device.
It contains up-to-date information concerning the characteristics of the washing programs installed in the appliance. Due to changes made in response to improvements in products,
the programs described here may differ from those described in the manual: in this case, only refer to the information provided in this document.
Programs may vary depending on the firmware version installed in the appliance. If in doubt, have authorised staff check on the appliance that the installed version is the same as the
version stated in the table at the back of this document.
ABBREVIATIONS USED IN THE TABLE OF PROGRAMS
DW, Demi: demineralised water, the type of water used in a specific phase of the program.
CW, Cold: cold mains water, the type of water used in a specific phase of the program.
No fill: this abbreviation refers to specific phases in which the water in the tank is not pumped out and fresh water is not taken in; instead, the water already in the well from the previous
phase is reused.
N.C.: "NOT CONTROLLED" - refers to the relatively short heating phases during which the heating elements are on but no target temperature value is set.
P1, ... P3: the code refers to the activation of the relative peristaltic pump, with dispensing of the liquid detergent associated to it.
Each pump has a default dispensing rate set in the factory.
P2a: the code refers to dispensing by pump P2 at a rate different from the default rate, used in special cases, as specified in the table.
Pump P3 is always optional. The programs which use it are not available if the pump is not installed.
TD: thermal disinfection - phase during which the target temperature is kept above a target value for a specified time. The efficacy of the thermal disinfection process is indicated by
means of parameter A0. In Standard Original programs, the target value for a thermal disinfection process is always at least 80°C.
Temp. T [°C]: Target temperature of the current phase. The temperature inside the tank is kept constant around the specified value for the time shown in the table. The phase during
which the temperature is held at the target value is called the EXTENSION phase. When no target temperature is indicated or the indication is (“0”), the heating elements are not
activated (e.g. during a cold prewash or for cold rinses).
2
VARIOUS NOTES:
WATER INTAKE: water intake for each phase is 9 litres. For program “16 Service” only, water intake is 6 litres for phases 1 and 2 and 9 litres for the rinse phase 3. Program “16 Service”
only comprises 3 phases and is generally used during installation, to prime the detergent intake lines correct and allow use of the device.
PHASE NUMBERING: a record of the parameters for the various cycle phases is maintained in the appliance’s memory. The first phase is numbered “PHASE 1”, the second “PHASE 2”, etc.
CYCLE TIME: the stated time for the complete cycle refers to standard test conditions: intake water pressure 3bar, cold water temperature 25°C. The single-phase connection performs
the activation of a single heating element instead of three, the cycle times are significantly longer.
PARAMETER A0
Parameter A0 is used by product standard EN ISO 15883-1 to assign a numerical value to the thermal disinfection process.
It is calculated mathematically using the formula:
z = 10°C - t = duration of time interval considered in seconds - T = load temperature in °C.
When calculating the parameter, only the time intervals during which the temperature is above 65°C are considered.
For thermal disinfection programs, the calculation is simplified by only including the “extension” phase, when the temperature is kept constant with reference to the target value set.
PROCESSABLE LOAD: never exceed a total load of 23 kg to ensure good washing results.
The items in the load must be positioned correctly, with no overlapping.
SELF DISINFECTION CYCLE: after 24 hours of inactivity, the device automatically proposes a self-disinfection cycle 93°C. The display shows "Pdis". It is advisable to run the program
without load, no load inside the washing chamber. Press the "Start / Stop" button to start the program. Press the Cycle Selection buttons, "Left selection" or "Right selection" to ignore
the suggestion.
WD DEFAULT DOSAGE
[ml/litre]
DOSAGE FOR SERVICE PROGRAM 16
ONLY [ml/litre]