SMEG WD3060 - WD5090
- WASHER DISINFECTOR -
1 INSTRUCTION SHEET
THIS DOCUMENT
Brief instruction sheet for basic procedures.
For all details relating to correct, safe use of the appliances concerned, the updated table of programs, interpretation of a
messages and installation instructions, please refer to the full version of the operator’s manual and program table.
SELECTING AND STARTING A WASHING PROGRAM
Power up the device by turning the master switch from “0” to “I”.
For models with side cabinet, the switch is access
ible by opening the cabinet door.
The latest program selected appears on the display after the manufacturer’s logo.
ESC” to change the preselected program.
UP” (or “DOWN”) to choose the program required.
ENTER” to select the program required.
START” to start the selected program.
RESET CYCLE
“RESET” button: starts the RESET cycle to return the device to its initial status.
The reset cycle should be used when alarms appear on the display.
Press “STOP” to halt a program. The door cannot be opened. Two possibilities:
-
If the “START” button is pressed within 60 seconds: the cycle is restarted.
- If the “RESET” button is pressed: the reset cycle is performed.
ENTER” to activate opening of the door. Do not force the door open by hand; a safety device is fitted.
ALARMS
The display and the buzzer alert the user to any alarms. When an alarm occurs, the device will attempt to deal with
the problem itself. When an alarm occurs, the user may: start a RESET cycle. If an alarm persists, contact the after
sales service.
There is no need to start a RESET cycle in the event of an alert
such as “No salt” or “Lack of detergent”; respond as
necessary to enable the device to function correctly.
11 Surgical instruments 93°C – 10’ [ + Drying ]
12 Thermal disinfection 93°C - 3’ [ + Drying ]
3 Plastics 70°C 5’ (P3) [ + Drying ]
13 Thermal disinfection 93°C – 10’ [ + Drying ]
4 Short 60°C - 5’ (P3) [ + Drying ]
14 Intensive thermal disinfection 93°C – 10’ [ + Drying ]
5 Clogs 80°C - 7’ (P3) [ + Drying ]
15 Intensive plastics 70°C - 5’ (P3, P5) [ + Drying ]
6 Feeding bottles 93°C - 5’ [ + Drying ]
16 Intensive clogs 80°C - 5’ (P3, P5) [ + Drying ]
7 Surgical instruments 80°C – 10’ [ + Drying ]
17 Lenses 80°C – 10’ (P5) [ + Drying ]
8 Short surgical instruments 90°C - 1’ [ + Drying ]
18 Surgical instruments 80°C – 10’ (P5) [ + Drying ]
9 Surgical instruments 90°C - 5’ [ + Drying ]
19 Surgical instruments 93°C – 10’ (P5) [ + Drying ]
10 Surgical instruments 93°C - 5’ [ + Drying ]
20 Intensive prewash (P5) [+ Drying]
[+ Drying]: end-of-cycle drying phase, on models with Drying System only.