3-3. Service ............................................................................................................................... 23
REV 8/05
Although current at the time of publication, SKYTRON’s policy of continuous development makes this
manual subject to change without notice.
Page 2
TYPE B
EQUIPMENT
EQUIPMENT LABELS AND SPECIFICATIONS
INDICATES DANGEROUS VOLTAGE, 120 V, 60 Hz
CLASS I DEFIBRILLATION PROOF, TYPE B EQUIPMENT- IPX4 RATED.
INTERNALLY POWERED EQUIPMENT
PROTECTIVE GROUNDING.
IN ORDER TO ENSURE PROPER GROUNDING RELIABILITY,
THIS TABLE MUST BE CONNECTED TO A PROPERLY GROUNDED
HOSPITAL GRADE OUTLET.
N
IPX4
V
A
HZ
CONNECTION FOR NEUTRAL CONDUCTOR SUPPLIED
UNIT TO BE USED ONLY IN SPECIFIED ENVIRONMENTAL CONDITIONS
TEMPERATURE: 15˚ - 30˚ C (60˚ - 85˚ F)
HUMIDITY: 30% - 60% RELATIVE HUMIDITY, NON CONDENSING
AC VOLTAGE
ENCLOSURE CLASS
VOLTAGE RATING OF THE UNIT
AMPERAGE RATING OF THE UNIT
FREQUENCY OF THE UNIT
ATTENTION, CONSULT MANUAL FOR FURTHER INSTRUCTIONS.
INDICATES SPECIAL USER ATTENTION.
POWERED BY AC VOLTAGE
POWERED BY BATTERY
BATTERY TYPE:
SEALED
LEAD ACID 12V, VALVE REGULATED
16AH, 10HR (530W/10MIN)
15 AMP INTERNAL CIRCUIT BREAKER
15A
Electrical Specifications
Power requirements
Current Leakage
Power Cord
120 VAC, 60Hz, 450 Watts
Less than 100 micro amps
15 feet w/hospital grade
connector (removable)
12-1/2"
18-1/4"25"23"
14-1/2"
19-3/4"
5-3/4"
3"
82"
TOP VIEW
90˚
60˚
14-1/2"
40-3/4"
SIDE VIEW
8-3/4"
5-3/4"
19"
22"
42-1/2" MAX
24" MIN
END VIEW
ENTELA CERTIFIED
TO UL2601-1
CAN/CSA601.1, IEC 60601-2-46
SERIAL NUMBER TAG
3600B UltraSlide General Purpose Surgical Table Specifications
Page 3
Page 4
The extreme positioning capabilities of the 3600B
UltraSlide Table requires special attention for possible interference points when using multiple function positioning. As with the operation of any
surgical table, a certain amount of care should be
exercised to position the patient safely. Although
the thick pads and sheets substantially protect the
patient, pinch points, located at the joints of the top
section should always be considered. BE SURE
THAT THE ARMS, HANDS AND FINGERS OF
THE PATIENT AND THOSE OF THE OPERATING ROOM PERSONNEL ARE CLEAR OF ALL
MOVING PARTS BEFORE MOVING THE TABLE.
Proper restraints should always be used for patient
safety.
SPECIAL USER ATTENTION
NOTE
Activating any function button will activate the brake system. Using the TABLE
UP function to set the brakes provides
a visual assurance that the brakes are
locked without altering the table position, except when emergency brake is
released.
WARNING
Possible explosion hazard exists if table
is used in the presence of FLAMMABLE
ANESTHETICS.
Certain accessories such as the Uro-Drain Tray,
Armboards and X-Ray top can be damaged when
changing the position of the table top sections.
Always look first to see if a desired movement is
going to interfere with any accessories in use.
The operator has the ultimate responsibility of
preventing damage to the table and surrounding
equipment or possible injury to the patient or staff.
The operator must ensure proper positioning is
maintained to prevent compromising respiration,
nerve pathways or circulation.
In general, common sense will dictate when there
is a potential hazard.
The following precautions should be reviewed
by all personnel prior to operating the table.
A routine Instructional Program should be
implemented by the facility for proper usage
instructions for all personnel that may operate
this table.
NOTE
An equalization terminal is located under the main power panel. This is provided as an alternate pathway to reduce the risk of static shock hazards.
Always follow recommended grounding procedures to ensure patient and
staff safety.
NOTE
The table will operate correctly on battery power with the power cord connected to a wall outlet or disconnected.
NOTE
Battery Operation must be turned OFF
at the pendant control. It cannot be
turned OFF using the main power
switch.
WARNING
Indicates a possibility of personal injury.
CAUTION
Indicates a possibility of damage to
equipment.
NOTE
Indicates important facts or helpful hints.
NOTE
Turning the Main Power Switch ON will
change the table operation to 120 VAC
power.
NOTE
When the red light starts to blink (indicating low power in battery) the table
will operate for approximately 5 continuous minutes, typically long enough
to use the table for the rest of the day.
SPECIAL USER ATTENTION
Page 5
NOTE
The charging system operates ONLY
when the table is in AC120V operation
mode.
NOTE
The table can be operated on 120VAC
power while the battery is being recharged.
NOTE
If the table is stored for a period greater
than 6 months, the batteries should be
removed and stored in a dry, clean condition at a storage temperature of 68° F
(20° C). Batteries should be recharged
every 6 months of product storage.
WARNING
NOTE
If the table top is slid toward the foot
end, the back section will not go below
horizontal. An audible alarm will sound.
NOTE
To prevent damage to the kidney lift, a
safety interlock prevents the back section from going more than 45° above
horizontal if the kidney lift is not all the
way down. An audible alarm will sound.
NOTE
If the leg section is positioned more
than 45° below horizontal, the top will
not slide toward the head end. An
audible alarm will sound.
NOTE
DO NOT unlock brakes when a patient
is on the table. An uneven patient
weight load may cause instability.
NOTE
With an evenly distributed patient weight
load, all table positioning functions will
operate smoothly and quietly with a
patient weight of up to 800 pounds.
WARNING
To maximize patient safety, utilize
proper restraint methods during extreme
Trendelenburg positioning.
WARNING
To maximize patient safety, utilize
proper restraint methods during extreme
lateral tilt positioning.
If the leg section is below horizontal,
slide toward head is limited to 7-1/2"
NOTE
If the back section is positioned below
horizontal, the top will not slide toward
the foot end. An audible alarm will
sound.
CAUTION
The Leg section may hit the table base
or the floor if both the leg and elevation
systems are placed in their full down
position.
NOTE
If the top is slid toward the head end
less than 7-1/2", the leg section will only
go down 45°. An audible alarm will
sound. If the top is slid toward the head
end more than 7-1/2", the leg section
will not go below horizontal.
Page 6
SPECIAL USER ATTENTION
NOTE
If the top is slid toward the foot end, the
FLEX function will not operate. An
audible alarm will sound.
NOTE
To prevent damage to the kidney lift, a
safety interlock prevents the kidney lift
from going up if the back section is 45°
above horizontal. An audible alarm will
sound .
NOTE
Elevation and brake system functions
are not affected by the return to level
function.
NOTE
The table top must be centered or slid
toward the foot end for the Beach Chair
function to operate.
NOTE
The emergency back-up control
switches will function when the table is
operating on 120VAC power, battery
power, or turned off.
NOTE
The Emergency Brake Release Valve
must be closed and tightened (clockwise) before activating any function.
•If the Emergency Brake Release Valve
has been operated, the UNLOCK button
on the pendant control will have to be
pressed before brakes will lock again.
NOTE
To make the Back Section easier to
handle, remove the Head Section, pad
and X-ray top prior to removing the
Back Section. Remove the pad and Xray top prior to removing the Leg Section.
CAUTION
The safety interlock system is not operational when the emergency backup control switches are used.
CAUTION
The EMERGENCY BRAKE LOCK
switch does not activate the brake system timer. The switch must be held
until the brakes are completely locked,
approximately 10 seconds.
WARNING
Ensure that the Leg and Back sections
are properly engaged and secured to
pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled
for proper orientation. The Leg section
cannot be installed on the Back section
pins.
SPECIAL USER ATTENTION
Page 7
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and cardiac defibrillator monitors.
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor.
WARNING
The antistatic properties of the table
are dependent on the use of the original
pad set which was furnished with the
table or an alternate approved replacement.
CAUTION
Thoroughly read and follow the
manufacturer's directions for all cleaning fluids. DO NOT use cleaners containing phenolics.
CAUTION
When using spray cleaners DO NOT
spray fluids directly into electrical receptacles or micro switches.
CAUTION
Before replacing pads on the table, make
sure the pads and all mating surfaces
are completely dry. Moisture trapped
between the pads and mating surfaces
may cause distortion of table tops.
WARNING
Certain accessories may limit weight
capacities. Check with your SKYTRON
representative.
NOTE
Always follow current AORN Journal
Guidelines to ensure proper cleaning
and disinfection procedure.
WARNING
Always follow OSHA blood-borne pathogens standards for protective clothing,
including gloves, masks and eye protection when cleaning the surgical table.
Page 8
SECTION I INTRODUCTION
SIDE
RAIL
HEAD SECTION
LOCKING KNOB
ACCESS COVER
HEAD
SECTION
PENDANT CONTROL
STORAGE BRACKET
SERVICE
POWER
CORD
REMOVABLE
BACK SECTION
PENDANT
CONTROL
SEAT
SECTION
REMOVABLE
LEG SECTION
LEG SECTION
RELEASE LEVER
(RECESSED)
FLOOR/LOCK
BRAKE (4)
MAIN POWER
SWITCH
Figure 1-1. 3600B UltraSlide
1-1. General
SKYTRON’s 3600B UltraSlide Surgical Table is an
electro-hydraulically operated, general purpose surgical table. See figure 1-1.
The electro-hydraulic positioning functions operated by the hand-held, push button, pendant control unit are: Trendelenburg, lateral tilt, back section, elevation, leg section, top slide, flex/reflex,
kidney lift, return to level, beach chair and the floor
lock/brake system.
Manual controls are provided for head section
positioning, emergency brake release, back section removal and leg section removal.
EMERGENCY
BRAKE RELEASE
main power ON/OFF switch is located on the
electrical panel on the front edge of the table base.
See figure 1-2.
The battery charging indicator and foot control
connector are also located on the electrical panel.
1-2. Power Requirements
The 3600B UltraSlide Surgical Table requires a
120VAC, 60 Hz electrical power supply. The table
is equipped with a removable 15 foot long power
cord with a three prong, hospital grade plug. The
POWER
CORD
MAIN POWER
SWITCH
BATTERY
INDICATOR
Figure 1-2. Electrical Panel
FOOT CONTROL
CONNECTOR
1-3. Pendant Control Unit
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
TABLE UP
(BRAKE LOCK)
LATERAL
TILT LEFT
BACK UP
LEG UP
SLIDE HEAD
FLEX
KIDNEY
LIFT UP
RETURN
TO LEVEL
BATTERY
LATERAL
TILT RIGHT
BACK DOWN
TABLE DOWN
LEG DOWN
SLIDE FOOT
REFLEX
KIDNEY
LIFT DOWN
BEACH
CHAIR
BRAKE
UNLOCK
REVERSE
TRENDELENBURG
TRENDELENBURG
Page 9
The hand-held pendant control unit (figure 1-3)
has a non-slip rubber cover which assures a positive grip during use. A spring clip hanger is located
on the back of the control for storage. When the
Pendant Control is not in use, it should be stored on
a convenient side or end rail. A bracket is located
under the table top next to the pendant control
connector for storage of the Pendant Control when
the table is not in use and during cleaning. See
figure 1-4.
POWER
SIDE RAIL
CLIP
INDICATOR
FUNCTION
BUTTONS
The function push buttons are identified with abbreviated descriptions for all functions. See figure
1-5. When illuminated the Trendelenburg and
table up buttons are red, the remaining buttons are
all green.
PENDANT CONTROL
STORAGE BRACKET
Figure 1-3. Pendant Control Unit
BACK
SECTION
Figure 1-4. Pendant Control Storage Bracket
SEAT
SECTION
Figure 1-5. Function Buttons
1-4. Floor Lock/Brake System
The floor lock/brake system consists of four selfleveling, hydraulic brake cylinders which raise and
support the table base off from the casters. Press
the TABLE UP button on the pendant control to set
the table’s brakes. An electronic timer will activate
the brake system until the brakes are completely
set, approximately 8-10 seconds.
NOTE
Activating any function button will activate the brake system. Using the TABLE
UP function to set the brakes provides
a visual assurance that the brakes are
locked without altering the table position, except when emergency brake is
released.
Page 10
SECTION II OPERATION
2-1. Electrical Power
The 3600B table will operate on either 120 VAC or
battery power.
WARNING
Prior to operating the table, observe
all table caution labels and review the
SPECIAL USER ATTENTION section
in the front of this manual.
WARNING
Possible explosion hazard exists if table
is used in the presence of FLAMMABLE
ANESTHETICS.
POWER
CORD
MAIN POWER
SWITCH
BATTERY
INDICATOR
TWIST TO
LOCK
FOOT CONTROL
EQUALIZATION
GROUNDING TERMINAL
Figure 2-1. Electrical Panel
CONNECTOR
NOTE
An equalization terminal is located under the main power panel. This is provided as an alternate pathway to reduce the risk of static shock hazards.
Always follow recommended grounding procedures to ensure patient and
staff safety.
2-2. AC 120V Operation
Use the following procedures to operate the table
on 120 VAC power.
a. Make sure the Power cord is securely attached to the table. To install the power cord, align
the cord connector with the base connector, insert
the cord and twist clockwise to lock the cord into the
connector. See figure 2-1. Plug the cord into a
properly grounded, Hospital Grade, 120 VAC outlet. Make sure the power cord is routed to the outlet
to prevent it from being in the way of operating
personnel.
The pendant control buttons and the green AC
120V, POWER indicator light located in the upper
right corner of the pendant control will illuminate.
See figure 2-2.
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
Figure 2-2. Pendant Control
b. Activate the POWER SWITCH located on the
electrical panel. The switch will illuminate.
c. The table is now ready for 120VAC operation.
Page 11
BATTERY
INDICATOR
2-3. Battery Operation
a. Make sure the Main Power Switch indicator
light, on the electrical panel, are OFF. See figure
2-2. If the indicator light is ON, turn AC120V
operation OFF with the main power switch.
NOTE
The table will operate correctly on battery power with the power cord connected to a wall outlet or disconnected.
b. Press the BATT button on the hand-held
pendant control. The pendant control buttons, the
red BATTERY indicator light, located in the upper
right corner of the pendant control and the Battery
Indicator on the electrical panel will illuminate.
c. The table is now ready for BATTERY operation.
d. To extend the battery charge life, turn the
BATTERY power OFF with the pendant control
when the table is not going to be used.
NOTE
Battery Operation must be turned OFF
at the pendant control. It cannot be
turned OFF using the main power
switch.
2-5. Charging the Battery
Batteries should be charged:
• When the table is placed into initial
service
• As indicated by Battery Indicator
• Every week under normal service
conditions
a. Battery Indicator The Battery Indicator
consists of ten lighted bars, 3 red, 4 yellow and 3
green. See figure 2-3. Each bar represents a
percentage of the battery charge condition. When
all ten bars are illuminated, the batteries are fully
charged. The following list shows the battery
charge level as indicated by the lighted bars;
3 green100% -Fully charged
2 green89%
1 green78%
4 yellow67%
3 yellow56%
2 yellow45% -Needs Charging (BATT
indicator on pendant will flash)
1 yellow34% -Needs Charging
3 red23% -Needs Charging
During charging, the bars will light in sequence to
the respective charge level, turn off and light in
sequence again.
2-4. Automatic Shut-Off
a. To prevent unnecessary discharge of the
battery, a timer is built into the battery circuit. This
timer will automatically shut the battery power OFF
after 1½ hours of table inactivity.
b. To turn the table ON again, press the BATT
button on the pendant control, the pendant control
buttons and the red indicator light will illuminate.
NOTE
Turning the Main Power Switch ON will
change the table operation to 120 VAC
power.
Figure 2-3. Battery Indicator
Page 12
b. If the battery needs to be charged when
operating the table on battery power, the red indicator light on the pendant control will begin to blink.
2-6. Positioning Functions
The hand-held pendant control (figure 2-4) activates the following table functions:
NOTE
When the red light starts to blink (indicating low power in battery) the table
will operate for approximately 5 continuous minutes, typically long enough
to use the table for the rest of the day.
NOTE
The charging system operates ONLY
when the table is in AC120V operation
mode.
c. To recharge the battery, make sure the power
cord is connected, plugged into a 120VAC wall
outlet and the main POWER SWITCH - ON.
NOTE
The table can be operated on 120VAC
power while the battery is being recharged.
d. A full battery charge will last approximately 2
weeks under normal operating conditions. However, it is recommended to charge the batteries at
the end of each week to establish a normal routine
protocol. Lead acid batteries last longer if they are
not permitted to fully discharge. The table features
(2) 12 volt, sealed, lead acid batteries which require no manual maintenance. Lead acid gel batteries, under a proper charging program, feature
an approximate normal life of 4 years.
TRENDELENBURG
REVERSE
LATERAL
TILT LEFT
BACK UP
TABLE UP
(BRAKE LOCK)
SLIDE HEAD
LEG UP
FLEX
KIDNEY
LIFT UP
RETURN
TO LEVEL
BATTERY
AC120V POWER ON
INDICATOR LIGHT
(GREEN)
TRENDELENBURG
LATERAL
TILT RIGHT
BACK DOWN
TABLE DOWN
SLIDE FOOT
LEG DOWN
REFLEX
KIDNEY
LIFT DOWN
BEACH
CHAIR
BRAKE
UNLOCK
Figure 2-4. Pendant Control Function Buttons
a. Floor Lock/Brake System. To activate the
brakes without affecting table positioning, press
the TABLE UP button. See figure 2-5. The
elevation cylinder will not function until the brakes
are completely extended.
TABLE UP
(BRAKE LOCK)
NOTE
If the table is stored for a period greater
than 6 months, the batteries should be
removed and stored in a dry, clean condition at a storage temperature of 68° F
(20° C). Batteries should be recharged
every 6 months of product storage.
BRAKE
UNLOCK
Figure 2-5. Brake System Activation
Press the BRAKE UNLOCK button on the pendant
30˚
30˚
REVERSE
TRENDELENBURG
TRENDELENBURG
control to release the four self-leveling brake feet in
order to move the table. See figure 2-5. The brake
delay circuit automatically retracts the brake system. It takes approximately 7-8 seconds to totally
release the system.
WARNING
DO NOT unlock brakes when a patient
is on the table. An uneven patient
weight load may cause instability.
NOTE
Page 13
Figure 2-6. Trendelenburg Positioning
With an evenly distributed patient weight
load, all table positioning functions will
operate smoothly and quietly with a
patient weight of up to 800 pounds.
b. Trendelenburg. To place the table in a
Trendelenburg (head down) position, press the
TREND button (figure 2-6). To place the table in a
reverse Trendelenburg (head up) position, press
the REV TREND button. Trendelenburg positioning of up to 30° may be obtained.
WARNING
To maximize patient safety, utilize
proper restraint methods during extreme
Trendelenburg positioning.
c. Lateral Tilt. To achieve lateral tilt right (as
viewed from the head end of the table), press the
TILT RIGHT button (figure 2-7). To achieve lateral tilt
left, press the TILT LEFT button. Tilt of up to 30° may
be obtained.
WARNING
To maximize patient safety, utilize
proper restraint methods during extreme
lateral tilt positioning.
LATERAL
TILT LEFT
LATERAL
TILT RIGHT
30˚30˚
Figure 2-7. Lateral Tilt Positioning
Page 14
d. Back Section. To raise the back section,
press the BACK UP button (figure 2-8). The back
section will raise up to 90° above horizontal. To
lower the back section, press the BACK DOWN
button. The back section will go down to 40° below
horizontal.
NOTE
If the table top is slid toward the foot
end, the back section will not go below
horizontal. An audible alarm will sound.
NOTE
42-1/2"
24"
TABLE
UP
TABLE
DOWN
To prevent damage to the kidney lift, a
safety interlock prevents the back section from going more than 45° above
horizontal if the kidney lift is not all the
way down. An audible alarm will sound.
BACK UP
90˚
40˚
BACK DOWN
Figure 2-9. Elevation Function
f. Top Slide. To move the table top toward the
head end, press the SLIDE HEAD button. From
center position, the top will slide up to 9-1/2". See
figure 2-10.
NOTE
If the leg section is positioned more
than 45° below horizontal, the top will
not slide toward the head end. An
audible alarm will sound.
NOTE
If the leg section is below horizontal,
slide toward head is limited to 7-1/2"
Figure 2-8. Back Section Positioning
e. Elevation. To raise table top, press the
TABLE UP button (figure 2-9). The table will lift a
patient weight of 1,000 pounds up to a maximum
height of 42-1/2" (46-1/2" with X-Ray top and 2"
pad). To lower the table top, press the TABLE
DOWN button. The table top will go down to a
minimum height of 24" (minus pad).
To move the table top toward the foot end, press
the SLIDE FOOT button. From center position, the
top will slide up to 13-1/2". Slide function will stop
and RETURN CTR Indicator will illuminate when
table is centered.
NOTE
If the back section is positioned below
horizontal, the top will not slide toward
the foot end. An audible alarm will
sound.
Page 15
SLIDE HEAD
9-1/2"
SLIDE FOOT
13-1/2"
Figure 2-10. Top Slide
g. Leg Section. To lower the leg section, press
the LEG DOWN button (figure 2-11). The leg
section will go down to 100° below horizontal. To
raise the leg section, press the LEG UP button.
The leg section will go up to 15° above horizontal.
LEG UP
LEG DOWN
15˚
100˚
Figure 2-11. Leg Section Positioning
h. Flex Positioning. To place the table top in
a flex position from horizontal, press the FLEX
button (figure 2-12). To return the table top to a
horizontal position or into a reflex position, press
the LEVEL or REFLEX button.
CAUTION
The Leg section may hit the table base
or the floor if both the leg and elevation
systems are placed in their full down
position.
NOTE
If the top is slid toward the head end less
than 7-1/2", the leg section will only go
down 45°. An audible alarm will sound.
If the top is slid toward the head end
more than 7-1/2", the leg section will not
go below horizontal.
NOTE
If the top is slid toward the foot end, the
FLEX function will not operate. An
audible alarm will sound.
FLEX
REFLEX
Figure 2-12. Flex/Reflex Positioning
Page 16
i. Kidney Lift. To raise the built-in kidney lift,
press the KIDNEY UP button (figure 2-13). Up to
5-3/4" of lift can be achieved. Press the KIDNEY
DOWN button to lower the kidney lift.
NOTE
To prevent damage to the kidney lift, a
safety interlock prevents the kidney lift
from going up if the back section is 45°
above horizontal. An audible alarm will
sound .
KIDNEY UP
5-3/4"
KIDNEY DOWN
RETURN
TO LEVEL
Figure 2-14. Return To Level
k. Beach Chair. To place the top in the beach
chair position from a level position, press the
BEACH CHAIR button (figure 2-15). The back
section will raise, the leg section will lower and the
Trendelenburg positioning will function simultaneously. The functions will stop when Trendelenburg reaches it limit.
NOTE
The table top must be centered or slid
toward the foot end for the Beach Chair
function to operate.
Figure 2-13. Kidney Lift Positioning
j. Return To Level. To return the table top to a
level position, press the LEVEL button (figure 2-14).
NOTE
Elevation and brake system functions
are not affected by the return to level
function.
BEACH
CHAIR
Figure 2-15. Beach Chair Positioning
2-7. Emergency Back-up Controls
a. The emergency back-up control switches are
located under the access door on the service
access cover in the table base. See figure 2-16.
Page 17
CAUTION
The EMERGENCY BRAKE LOCK
switch does not activate the brake system timer. The switch must be held
until the brakes are completely locked,
approximately 10 seconds.
NOTE
FUNCTION CONTROL
ACCESS DOOR
Figure 2-16. Emergency Controls Location
b. In the event of either a power failure or a
problem with the hand-held pendant control, the
table can be operated using the emergency backup switches. Simply push the desired emergency
switch in the appropriate direction to operate the
table functions. See figure 2-17.
CAUTION
The safety interlock system is not operational when the emergency backup control switches are used.
The emergency back-up control
switches will function when the table is
operating on 120VAC power, battery
power, or turned off.
c. Switches are provided for Trendelenburg,
lateral tilt, back section, elevation, leg section,
kidney down and brake lock. These switches are
spring-loaded so they return to the neutral or
center position when released.
REV
TILT
TREND
TRENDTILT
BACKUPTABLEUPLEGUPBRAKE
RIGHT
BACK
DOWN
TABLE
DOWN
LEFT
LEG
DOWN
LOCK
KIDNEY
DOWN
Figure 2-17. Emergency Back-Up Controls
Page 18
2-8. Emergency Brake Release.
In case of a power failure or an electrical problem
within the table, the emergency brake release
system can be used to move the table. The control
knob for this function is located on the side of the
table base and is identified by an EMERGENCY
BRAKE RELEASE label. Turn the knob counterclockwise to release the brakes. See figure 2-18.
SERVICE ACCESS
COVER
BRAKE (4)
POWER
CORD
EMERGENCY
BRAKE RELEASE
2-9. Head Section
a. A quick release positioning bar located under
and to the front of the head section (figure 2-19) is
used to raise or lower the head section. Pull the
release bar toward the head end to allow the
section to pivot up or down. Positioning from 60°
above horizontal to 90° below horizontal in 15°
increments is available. Release the bar to lock the
head section in position.
HEAD SECTION
RELEASE BAR
Figure 2-18. Emergency Brake Release
NOTE
The Emergency Brake Release Valve
must be closed and tightened (clockwise) before activating any function.
•If the Emergency Brake Release Valve
has been operated, the UNLOCK button
on the pendant control will have to be
pressed before brakes will lock again.
Figure 2-19. Head Section Adjustment
b. By loosening two locking knobs beneath the
back section, an additional 2" of longitudinal adjustment can be achieved. If desired, the head
section may be removed by loosening the locking
knobs and pulling it straight out of the back section.
3600B UltraSlide Table has the capability of attaching the head section to the leg section
for use
as a foot extension ONLY. Do Not reverse the
patient on the table without first consulting with
SKYTRON.
Two locking knobs are located on the inside of the
leg section for securing the head section. See
figure 2-20.
FOOT/LEG
SECTION
LOCKING
KNOB
HEAD
SECTION
LOCKING
KNOB
Figure 2-20. Repositioning Head Section
(for use as a Foot Extension)
2-10. Leg and Back Section Removal.
a. The Leg Section and the Back Section on the
3600B UltraSlide table are both removable. See
figure 2-21 (Leg Section shown).
Page 19
WARNING
Ensure that the Leg and Back sections
are properly engaged and secured to
pins before use to prevent injury.
b. To remove either section, level the table top,
simultaneously depress both release levers and
pull the section out. Press the LEG DOWN or
BACK DOWN button on the pendant control to
position the attachment pins down and out of the
way.
NOTE
To make the Back Section easier to
handle, remove the Head Section, pad
and X-ray top prior to removing the
Back Section. Remove the pad and Xray top prior to removing the Leg Section.
FOOT/LEG SECTION
LOCKING
LEVERS
NOTE
The Leg and Back sections are labeled
for proper orientation. The Leg section
cannot be installed on the Back section
pins.
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and cardiac defibrillator monitors.
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor.
Figure 2-21. Leg Section Removal
c. To install the Leg Section, press and hold the
LEG UP button until the attachment pins completely stop. Install the section on the pins. Level
the table top and pull out on the section to make
sure the release levers are completely locked.
d. To install the Back Section if the attachment
pins are not aligned, make sure the table top is
centered or toward the head end, press and hold
the BACK DOWN button until the attachment pins
completely stop (40° down). Press REV TREND to
bring the pins up to a level position and install the
section on the pins. Level the table top and pull out
on the section to make sure the release levers are
completely locked.
WARNING
The antistatic properties of the table
are dependent on the use of the original
pad set which was furnished with the
table or an alternate approved replacement.
2-11. Positioning
The use of certain optional accessories available
from SKYTRON further extend the positioning
capabilities of the 3600B UltraSlide Table. Refer to
the following "Positioning Guidelines" or contact
your SKYTRON representative for further details.
WARNING
Certain accessories may limit weight
capacities. Check with your SKYTRON
representative.
Endovascular (with optional 48” Carbon
Fiber Leg Extension)
Page 22
SECTION III MAINTENANCE
3-1. Preventive Maintenance
The following preventive maintenance checks and
services are recommended to ensure the serviceability and proper operation of your SKYTRON
Surgical Table, and should only be performed by
qualified SKYTRON trained personnel.
a. During normal cleaning, a general visual
examination should be made checking for leaks,
loose bolts or parts, and cracked, chipped, or
missing paint. Any necessary repairs should be
made.
b. Semi-annually the following checks and services should be performed:
1. Check all hydraulic fittings, mini-valves and
slave cylinders for proper operation and
any signs of leaks.
2. Check the hydraulic speed controls and
adjust if necessary.
3-2. Cleaning Recommendations
NOTE
Always follow current AORN Journal
Guidelines to ensure proper cleaning
and disinfection procedure.
The following procedures should be followed when
cleaning the surgical table between cases.
Place table top in level position prior to starting
cleaning procedure.
WARNING
Always follow OSHA blood-borne pathogens standards for protective clothing,
including gloves, masks and eye protection when cleaning the surgical table.
3. Pressure check (with a gauge) the pressure relief valve.
4. Check all mechanical adjustments and adjust as necessary.
5. Check hydraulic fluid level.
6. Lubricate the slider assembly.
7. Check function of back and foot leg release
levers, lubricate as necessary.
Remove major contaminants from the table with
disposable materials following appropriate biohazard waste disposal procedures.
Remove all table pads and place them on a flat
surface for cleaning.
CAUTION
Thoroughly read and follow the
manufacturer's directions for all cleaning fluids. DO NOT use cleaners containing phenolics.
Apply cleaning fluid liberally to top and sides of
each pad and wipe with a clean lint-free cloth.
Using a clean, damp, lint-free cloth, wipe the pads
to remove the cleaning fluid.
Using a clean, dry, lint-free cloth, wipe the pads to
remove all moisture.
Repeat the steps to clean the bottom of the each
pad.
Page 23
3-3. Service
Table maintenance can be performed by trained
maintenance personnel using SKYTRON authorized replacement parts and service techniques.
Service instructions and parts are available from
SKYTRON.
CAUTION
When using spray cleaners DO NOT
spray fluids directly into electrical receptacles or micro switches.
Repeat cleaning procedure for all table surfaces
including the top, sides, elevation column, base
and all accessories.
CAUTION
Before replacing pads on the table, make
sure the pads and all mating surfaces
are completely dry. Moisture trapped
between the pads and mating surfaces
may cause distortion of table tops.
When the cleaning procedure is complete, replace
all pads and accessories as applicable.
Remove pendant control from table side rail and
apply cleaning solution to the pendant control and
cord.
Use a clean cloth dampened with water to remove
cleaning solution.
Preventive Maintenance contracts are available
through your local SKYTRON representative.
To obtain service instructions, replacement parts,
factory service or preventive maintenance contracts, contact the SKYTRON representative listed
below.
Or contact:
SKYTRON
5000 36th Street S.E.
Grand Rapids, MI 49512
1-800-SKYTRON (1-800-759-8766)
Fax. 1-616-957-5053
Use another clean damp cloth to remove any
remaining residue.
Install pendant control on side rail for storage when
cleaning procedure is complete.
SKYTRON SURGICAL L IGHTS AND TABLES LIMITED WARRANTY
Limited Warranty. SKYTRON, a Division of the KMW Group, Inc., warrants only to the original purchaser (the “Customer”) that the products that it
sells, either directly or through its distributors or other authorized representatives, shall not be “defective” for the applicable time provided for, and
subject to the conditions and exceptions stated in, this warranty. A product shall be considered “defective” if SKYTRON determines that it is
defective in materials or workmanship and that the defect materially impairs the value of the product under normal use, maintenance and service.
This warranty does not apply to (i) any product that is modified or altered, (ii) minor product adjustments necessitated by ordinary wear and tear or
(iii) damage to the products caused by accident, use of replacement parts other than original OEM parts, unauthorized repairs, improper installation
or use, abuse, lack of user training, or failure to comply with SKYTRON’s instructions. THIS WARRANTY IS IN LIEU OF ALL OTHER
WARRANTIES EXPRESSED OR IMPLIED, INCLUDING, BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR ANY PARTICULAR PURPOSE.
Warranty Period. The warranty on replacement parts, spares, bulbs (for surgical lights), pads and accessory items (for surgical tables) shall extend
for 90 days from the date of receipt by the Customer. The warranty on all SKYTRON Surgical Lights and Tables shall extend for two years from (i)
the initial installation of the product by Customer or (ii) the first anniversary of the date of shipment to the Customer, whichever is earlier. If
SKYTRON supplies a replacement part or product under this warranty, then the warranty on such replacement part or product shall terminate upon
(i) 90 days or (ii) the end of the remaining period of the original warranty, whichever is of greater benefit to the Customer.
Remedies. If a product proves defective during the applicable warranty period, and if the Customer gives SKYTRON and the distributor from whom
the Customer purchased the product written notice of the defect within such period and complies with all other terms of this warranty, then
SKYTRON or its authorized service agent shall repair or replace the defective product or part thereof, at SKYTRON’s option. If necessary, this
warranty includes the cost of service labor and travel time to the site of product use for two years; provided, however, that delays caused by the
Customer in accessing the product for repair will be chargeable to the Customer at SKYTRON’s authorized service representative’s normal hourly
rate. If SKYTRON (or its authorized representative) fails to repair or replace any defective product or part thereof within a reasonable time, then
SKYTRON shall be liable for the reasonable cost of repair by a third party, but the Customer shall not obtain repair by a third party without giving
SKYTRON at least 15 days prior written notice, during which time SKYTRON (or its authorized representative) may repair or replace the defective
product or part thereof. This is the sole and exclusive remedy for any defect.
Limitations. SKYTRON SHALL NOT BE LIABLE TO THE CUSTOMER OR TO ANYONE ELSE FOR CONSEQUENTIAL, INCIDENTAL, SPECIAL,
EXEMPLARY, INDIRECT, OR PUNITIVE DAMAGES ARISING FROM ANY DEFECT, DELAY, NONDELIVERY, RECALL, OR OTHER BREACH
BY SKYTRON, INCLUDING, BUT NOT LIMITED TO, PERSONAL INJURY, DEATH, PROPERTY DAMAGE, LOST PROFITS, OR OTHER
ECONOMIC INJURY. SKYTRON shall not be liable to the Customer or anyone else in tort for any negligent design or manufacture of any products,
or for the omission of any warning with respect thereto.
Miscellaneous. No dealer, employee, or other representative of SKYTRON may extend or enlarge this warranty, waive any of the limitations, or
make any different or additional warranties with respect to the products. Any statements to the contrary are null and void unless made in writing
signed by an authorized officer of SKYTRON. This warranty shall be governed by Michigan law. Any action arising out of or relating to this warranty
shall be brought in any federal or state court located in Kent County, Michigan and having jurisdiction of the subject matter, and any such court shall
have personal jurisdiction over the parties and the parties waive any objection that the court is an inconvenient forum.
5000 36th Street S.E., Grand Rapids, MI 49512
1-800-SKYTRON or 1-616-957-0500 • FAX 1-616-957-5053