Read this manual before installing, operating, or servicing this equipment. Do not install
substitute parts, or perform any unauthorized modification of the product. Return the
product to Sigma Instruments for service and repair to ensure that safety features are
maintained.
Safety Symbols
WARNING: Calls attention to a procedure, practice, or condition, that could possibly
cause bodily injury or death.
CAUTION: Calls attention to a procedure, practice, or condition, that could possibly cause
damage to equipment or permanent loss of data.
Refer to manual for specific Warning or Caution information to avoid personal
injury or equipment damage.
Hazardous voltages may be present.
Earth ground symbol.
Chassis ground symbol.
Warranty Information
This Sigma Instruments product is warranted against defects in material and
workmanship for a period of 1 year from the date of shipment, when used in accordance
with the instructions in this manual. During the warranty period, Sigma Instruments will,
at its option, either repair or replace products that prove to be defective.
Limitation of Warranty
Defects from, or repairs necessitated by, misuse or alteration of the product, or any
cause other than defective materials or workmanship are not covered by this warranty.
NO OTHER WARRANTIES ARE EXPRESSED OR IMPLIED, INCLUDING BUT NOT
LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE. UNDER NO CIRCUMSTANCES SHALL SIGMA
INSTRUMENTS BE LIABLE FOR CONSEQUENTIAL OR OTHER DAMAGES
RESULTING FROM A BREACH OF THIS LIMITED WARRANTY, OR OTHERWISE.
Return Policy
The purchaser may return this product in like new condition within 30 days after
shipment for any reason. In case of return, purchaser is liable and responsible for all
freight charges in both directions.
A. Material Parameters
B. Specifications
C. I/O Connections
D. Declaration of Conformity
Chapter 1Quick Start
1.0 Introduction
Congratulations on your purchase of the SQM-160 Deposition Rate/Thickness Monitor.
The SQM-160 is an easy-to-use instrument for measuring many types of thin film
coatings. This chapter will help to get you up and running quickly. Please review the
entire manual for detailed operational, programming, and safety information.
1.1 Installation
Refer to Sections 1.3 and 1.4 for detailed system hookup information.
Rack
Installation
Power
Connection
Sensor
Connections
Digital I/O
Connections
Computer
Connection
The SQM-160 occupies a 3.5” high, half-rack space. An
optional kit (900-008) is available to adapt to a full rack. Install
the unit in a 19” rack with the supplied hardware.
WARNING: The SQM-160 is shipped with a ½ A Fuse for use
with 90–120 VAC. If your line voltage is 180–240 VAC, replace
the fuse with the ¼A fuse included with your SQM-160.
WARNING: Verify that the Voltage Selector Switch located next
to the power switch matches your line voltage.
Plug the SQM-160 into your power source with the included
power cord.
Connect the BNC cables and oscillator from your vacuum
chamber feedthrough to the desired SQM-160 Sensor Input(s).
Refer to Appendix C for details on wiring digital I/O to the
SQM-160 Relay I/O connector.
If you would like to use the supplied Windows software to
collect data or program the SQM-160, attach a 9 pin straightthru cable from the SQM-160 RS-232 connector to your
computer’s serial port.
Option
Connections
Move the rear panel power switch to the On (|) position. The SQM-160 will briefly
display its software version, then go to normal operating mode.
If you purchased the Analog Output Option (502-021), connect
a BNC cable form the Rate or Thickness connector to your
chart recorder.
1-1
Chapter 1Quick Start
1.2 Front Panel
Display1 Display2Crystal Status LEDs
Control Knob
Rate A/sThickness kA
Zero
Xtal Life
Open
Closed
Shutter
ControlConfiguration
Σ
Control Section
Time
Sigma
instruments
SQM-160
Rate/Thickness Monitor
Time SP
Thk SP
Final Thk
1
2
3
4
5
Crystal Status
6
NextPrevClearProgram
Clear
Configuration SectionSetpoint LEDs
Front Panel Controls
Display 1Displays rate and thickness in normal operation.
Displays setup parameter name in program mode.
Display 2Displays deposition time in normal operation.
Displays setup parameter value in program mode.
Control
Section
Zero thickness reading. Toggle Display 1 between Crystal Life
and Rate/Thickness readings. Toggle shutter relay on/off.
Configuration
Section
Program enters setup mode. Next/Prev moves thru menus.
Clear cancels change and returns value to its original setting.
Setpoint LEDsIlluminate when the indicated setpoint is reached.
Crystal Status
LEDs
Illuminate when a crystal is active and operating.
Flashes when an active crystal fails.
Control KnobUsed to adjust values or scroll though menu selections.
Pushing the control knob stores the current setting.
1-2
Chapter 1Quick Start
1.3 Rear Panel
Option A Card
3456
Sensor 1 Sensor 2
Manufactured By
Sigma
Σ
instruments
RateThick
RS-232
Model No. SQM-160
Ser No. 80901
Option B Card
Relay I/O
Ground
110
Line Fuse
Fuse 1/2 A SB at 120VAC
Fuse 1/4 A SB at 240VAC
220
10
Rear Panel Connections
Sensor 1 & 2Connection to quartz crystal sensor remote oscillator. See
Section 1.4 for detailed hookup.
RS-232Connection to computer serial port for programming and data
acquisition.
Relay I/OConnects 4 relays and 4 digital inputs to external equipment for
process control. See Appendix C for connections.
Option A CardProvides an additional four sensor measurement channels.
Option B CardProvides two 0-5V analog outputs for rate and thickness
readings.
Ground Terminal
Measurement ground terminal.
Voltage SelectorWARNING: Select 115V for 90-120VAC, 230V for 180-240VAC.
Line FuseWARNING: Replace line fuse only with a fuse of the specified
type and rating.
Power ConnectorWARNING: Use removable power cords only of the specified
type and rating.
1-3
Chapter 1Quick Start
1.4 System Connections
The diagram and table below identifies typical vacuum system components and their
function.
Sensor
In-Vac
Cable
Feedthrough
6" BNC Cable
Oscillator
Source
Shutter
10' BNC Cable
Ground Wire
Option A Card
Sensor 1 Sensor 2
Manufactured By
Sigma
Σ
instruments
RS-232
Model No. SQM-160
Ser No. 80901
Option B Card
Relay I/O
Ground
Line Fuse
Fuse 1/2 A SB at 120VAC
Fuse 1/4 A SB at 240VAC
220
110
System Components
SensorHolds the quartz crystal used to measure rate and thickness.
Crystals must be replaced occasionally.
In-Vac CableA coax cable, typically 30” long, that connects the sensor to the
feedthrough.
10
FeedthroughProvides isolation between vacuum and atmosphere for
electrical and cooling lines.
6” BNC CableProvides a flexible connection from the feedthrough to the
oscillator. Keep this cable as short as possible.
OscillatorContains the electronics to operate the quartz crystal. Total
cable length to the crystal should be under 40”.
10’ BNC CableConnects the oscillator to the SQM-160. Lengths up to 100’
are acceptable.
Ground WireA wire, typically braided, that connects the vacuum system to
the SQM160 ground terminal.
1-4
Chapter 1Quick Start
1.5 Film Setup
This section will help you setup the SQM-160 to measure a film. Refer to Chapter 2 for
detailed film programming instructions.
Note: User actions with front panel controls are indicated by a Box. Results shown on
displays, are indicated by a Dashed Box.
Enter Program
Mode
Select a FilmTurn the Control Knob to select one of the nine possible films,
Set Film
Parameters
Set System
Parameters
If the Crystal Life display is shown, press Xtal Life to return to
Rate/Thickness mode. Press Program to enter the film setup
menu. Pressing the program switch while in program mode
returns the SQM-160 to normal mode.
then press the Control Knob to enter the film parameters
menu.
Turn the Control Knob to set the first film parameter (density).
The parameter value is shown in Display 2. Press the Control
Knob to save the value and move to the next film parameter. If
you press Clear, the film parameter returns to its original value.
Continue to set each parameter. When the last film parameter
is entered, the SQM-160 returns to normal mode. See Chapter
2 for detailed instructions on setting Film Parameters.
To Enter the system menu, press Program, then Prev. Set
system parameters by turning, then pushing, the Control Knob
as described above. See Chapter 2 for detailed instructions on
setting System Parameters.
If the sensor(s) you selected during film setup are connected to the SQM-160, the
Crystal Status LEDs should be lit. If not, return to the Film Parameters menu and set
the Sensor Average parameter to the desired sensor(s).
If the Crystal Status LED is flashing, it is most likely that sensor is not properly
connected. Sigma supplies a small test crystal with each oscillator module that can be
used to test sensor connections external to the vacuum chamber. Referring to Section
1.4, disconnect the oscillator from its 6” BNC cable. Attach the test crystal to the
oscillator’s feedthrough connector. The Crystal Status LED will remain lit if the external
sensor connections are correct.
Refer to the Sensor Selection, Frequency Display and Sensor Min/Max sections of
Chapter 2 for information that can assist in troubleshooting sensor problems.
1-5
Chapter 1Quick Start
1.6 Depositing a Film
If you have followed this Quick Start chapter, you are ready to deposit a film. Follow the
procedure below to begin deposition.
Verify Sensor
Operation
Display
Rate/Thickness
Zero
Thickness
Start
Deposition
The Rate and Thickness displays should begin to move from zero. Please take time to
review the remainder of this manual for detailed operational, programming, and safety
information.
Verify that the Crystal Status LED for the measuring sensor is
lit, and not flashing.
Display 1 should be displaying Rate on the left and Thickness
on the right. If the Crystal Life display mode is active, press
the Xtal Life switch to return to Rate/Thickness mode. If the
Program Mode is active, press Program to return to normal
mode.
If needed, press the Zero switch to zero the thickness reading.
Apply power to your source evaporation supply. If the SQM160 shutter relay is connected, press the Shutter switch to
begin deposition.
If the displays remain at zero, check your system setup to assure that you are actually
evaporating. Also check that the deposited material is reaching the sensor.
If the display is erratic or noisy, first check your sensor connections. Refer to the
Timebase, Rate Filter and Rate Resolution sections of Chapter 2 for information that
can help in troubleshooting noisy reading problems.
If the rate and thickness readings do not match your expectations, refer to the Film
Parameter (Density, Z-Factor, Tooling) and Sensor Tooling sections of Chapter 2.
1-6
Loading...
+ 24 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.