INDEX .......................................................... 111
C4 Manual Rev 7.5.2
7
8
C4 Manual Rev 7.5.2
1. INTRODUCTION
This manual describes the operating procedures for the Sigma Systems Models C4
& CC-3.5 Controllers, microprocessor based controllers and control communications
interfaces for the family of Sigma Systems temperature chambers and thermal
platforms.
1.1 Models C4 & CC-3.5 Explained
The models C4 & CC-3.5 controllers are successors to the model CC-3.
The model C4 is a completely redesigned controller that uses a
completely different and more modern set of internal components with
a new processor and completely new firmware. Model C4 controllers
are only available as new products from Sigma Systems.
The model CC-3.5 is a hybrid upgrade controller that uses only the
digital circuitry of the model C4. It is made by substituting the C4
digital p.c. board (known within Sigma Systems as the “CPU board”, or
the “A” board) for the CC-3's “A” board. The upgrade from CC-3 to
CC-3.5 also includes a few small modifications to other internal
components. The power supply, power switching, analog, and front
panel components of the CC-3 remain. Model CC-3.5 controllers are only
available as the product of upgrading a CC-3 controller.
From a functional perspective, models C4 & CC-3.5 are identical because all of the
functionality of the controllers is defined by the processor, bus interface components,
and firmware... all of which are integral with the C4 “A” board. The C4 analog and
front panel components that remain unique to the C4 (not included in the CC-3
upgrade from CC-3) provide slightly better accuracy and substantially better noise
and static immunity.
For the balance of this manual, the term C4 will mean to
include both the model C4 controller and the model CC-3
.5
controller. In the event that there is a difference between
the two models, that difference will be explicitly detailed.
C4 Manual Rev 7.5.2
.5
9
1.2 General Description
Using the model C4, temperature control is available manually from the front
panel, by use of user entered programs, or via remote control via either a EIA-232
or IEEE-488 GPIB. The controller has a precision temperature reading capability
with a digital read-out. Two temperature probes can be connected to the
controller allowing either probe or both probes to be the control probe(s) while
either probe can be used to take measurements.
Two additional controlled device ports are available. They are intended for on/off
control of a refrigeration compressor and an external load such as a device under
test, or a main chamber or platform power relay. These ports normally are
supplied as TTL level (low voltage) ports but are optionally available with solid
state relays to control line voltage as in the case of units with mechanical
refrigeration. The compressor control port may be toggled from the front panel at
any time in the manual (LOCAL) mode. The compressor is designed to not cycle
on and off with the temperature control function.
The controller operates in each of four modes:
Local
Mode
Program
Mode
Remote
Mode
Setup
Mode
Single Setpoint control from the front panel. Simple
Start/Stop functionality.
Programmed control using programs entered, stored, and
run from the front panel. 100 Temperature/Duration
Program steps available. Multiple programs may be stored
and called as needed.
Control via EIA-232 or IEEE-488 GPIB. The IEEE-488
interface implementation is a TALKER/LISTENER with
serial poll. Extended addressing and parallel poll
capabilities are not supported. The EIA-232 interface is
fully configurable for baud rate, data bits, stop bits and
parity.
Used to define and store operation and environment
variables that control how the C-4 behaves.
10
C4 Manual Rev 7.5.2
1.3 Custom Features / Interchangeability WARNING
Each Sigma Systems C4 Controller has been custom configured for the chamber or
platform with which it was supplied or for which it was specified. Many units
include special wiring for custom control applications, precision fail-safe additions,
non-standard voltages, external unit power control, etc. Units that may appear
to be identical may be internally quite different. Do not interchange
controllers between controlled devices (chambers and/or platforms)
unless you are certain that the controllers have been identically
constructed.
Failure to heed this warning voids your warranty, may cause unpredictable
controlled device behavior that could cause damage to persons or property, pose a
risk of fire, or cause other problems. If you must move controllers between
controlled devices, please contact the Sigma technical support department for
assistance and advice.
1.4 Release 7.5.2 Firmware
This manual is specifically written to cover the features of Release 7.5.2 firmware
for the Sigma Systems Model C4 controller. The features of this release firmware
are largely backward compatible with all CC-3 and earlier CC-3.5 & C4 firmware.
This release fixes a number of bugs found in the CC-3.5 interim release versions
6.8.6, 6.9.0, and 7.0.0. It also adds a number of new features. See Section 1.5.
Note: This release implements probe temperature correction on a per
probe basis. Interim CC-3.5 releases implemented this feature on a global
basis.
Sigma highly recommends that all users update to the latest firmware release.
Contact Sigma Technical Support. See Section 9.10.
Note: This release is not available as an upload file for serial port firmware
updating of your controller. Due to non-backward compatible change in the way
some data is stored internally, this version must be installed using a PROM
obtained from Sigma Systems.
C4 Manual Rev 7.5.2
11
1.5 C4 vs. CC-3 Differences (What’s New)
1.5.1 Hardware & Stability Improvements
The new C4 introduces a number of new improvements to make the controller
faster and more reliable than its predecessor.
The C4 has a completely new digital processing board. The new board has a much
faster processor, more memory, and a much faster IEEE-488 GPIB controller.
Interrupts have been completely restructured to improve stability. The multilayer
design is far more tolerant of both static and power line interference. There is
now a watchdog timer to detect system lockups in the event that something does
interfere with the system. The battery backed RAM has been replaced by an
EEPROM. The Vactrol type isolation device in the failsafe circuit has been
replaced. The C4 contains no components that have a time based failure mode.
The new firmware also monitors the state and integrity of internal memory.
Critical system information is stored in multiple places so that minor errors due to
uncontrollable transients or other causes can be repaired on the fly with no
disruption in process control. Likewise, both the front panel display and the GPIB
interface are monitored constantly to assure that their operation has not been
compromised by static discharge or line transient. In the event of a problem,
either device can be reset on the fly to allow operations to continue normally.
System integrity is further enhanced by a series of successive shutdown processes
that monitor the integrity of the data coming from the sensor probes. In the event
that any probe reports a temperature more than 20
/C beyond the limits set in the
controller, the controller will shut down all heating and cooling and display a
warning message. Likewise, in the event that a sensor probe reports an extreme
temperature, either hot or cold, the system will assume that a probe has become
compromised by an open or shorted circuit and stop applying heat and cooling,
shut down the system and display an appropriate warning message..
1.5.2 Hardware Change (EEPROM replaces BBSRAM)
All C4 controllers and CC-3.5 controllers converted after January 1998 have the
battery backed static RAM (BBSRAM) replaced with an EEPROM. This change
was implemented to reduce the possibility that the controller will require service.
Although changing the BBSRAM when the battery died (about every 5-12 years)
was a fairly simple matter, the necessity for doing so was found to be a nuisance
as was diagnosing the need for the change. The BBSRAM or EEPROM provides
the non volatile memory where the C4 stores both the setup parameter
information and the user programs.
12
C4 Manual Rev 7.5.2
1.5.3 Firmware Uploads
When new firmware is available for your C4, you can easily upload it into the
controller using the controller’s serial port. The procedure requires only a diskette
bootable PC and a serial cable and takes only about ten minutes. Firmware
updates, when available, may be obtained on diskette for a fee from Sigma
Systems or for free by download from www.SigmaSystems.Com or
ftp.SigmaSystems.Com. See Section 3.2.2
1.5.4 Forced Start from PROM Firmware
The controller can be started from the original firmware version that is stored in
the PROM. Uploaded versions are stored in flash memory. If a firmware upload
session should go astray somehow, this feature allows the controller to still
operate. See Section 9.7.8.
1.5.5 Front Panel Information Display at Startup
The controller model is displayed at startup - See Section 3.1.1
The firmware version number is displayed at startup - See Section 3.1.1
The operating range may be displayed at startup - See Section 3.1.2
The controller serial number may be displayed - See Section 3.1.3
1.5.6 Temperature Out of Range Shutdown
The controller now stores the operational limits for itself, the controlled device
(Sigma chamber or platform), and the unit under test (UUT). The operating
temperature is checked against these limits, if it is too far outside these limits,.
the system is shutdown with an appropriate error message displayed. See
Section 3.5.
C4 Manual Rev 7.5.2
13
1.5.7 Internal Error Shutdown Conditions
The C4 monitors system health by keeping track of three additional areas;
they are:
Processor healthTracked by watchdog timer
Memory conditionChecked at startup
Setup parameter tableChecked continuously as used
Some detected internal errors can be repaired on the fly. If this is possible, the C4
will recover from the error and you will not know the error existed. If, however,
the error is not repairable and the system must be shut down, an error message
will be displayed to help you understand what happened and how to prevent or
cope with it. See Section 3.7.
1.5.8 Fahrenheit Temperature Scale Supported
The controller will now operate in either Celsius /C or Fahrenheit /F mode. See
Section 3.4.
1.5.9 Temperature Probe Correction (Calibration)
The C4 will allow you to enter data via the Setup mode that will correct anomalies
in the temperature readings and control. Such adjustments are sometimes
necessary to optimize accuracy at a particular temperature, or to compensate for
differences between raw probe temperature data and actual temperatures. These
differences can be the result of probe placement, effects of the unit under test on
the temperature data, or other causes. See Section 3.8
1.5.10 “Bumpless” Temperature Control
(Not in this release. Available in next release, without charge. Check the
Sigma Systems FTP or WWW site for downloadable file.)
When the setpoint is changed, the PID control algorithm begins a new “search” for
the right amount of heat and/or cooling to maintain the new setpoint. Normally,
for each new setpoint, the PID routines begin the search anew... behaving as if the
controller was just turned on. The controller will quickly determine that heat or
cooling is called for, and while the chamber or platform advances toward the new
Setpoint, the PID routine adjusts for the response to heat and cool and
methodically settles the chamber or platform in on the new setpoint. For most
setpoint changes, where the new and old setpoints are quite different, this is a
fast, accurate and appropriate method of control.
14
C4 Manual Rev 7.5.2
However, when the change in the Setpoint is very small, this “start from the
beginning” search routine can search over such a wide range that it will introduce
a “bump” in the platform or chamber temperature that can exceed the amount of
the Setpoint change. The C4 includes an intelligent PID routine that constrains
the search appropriately for the change in Setpoint and thus eliminates the “PID
bump”.
1.5.11 Intelligent 2 Probe Control (Probe Averaging)
(Not in this release. Available in next release, without charge. Check the
Sigma Systems FTP or WWW site for downloadable file.)
Intelligent 2 Probe Control allows the internal temperature of the unit under test
(UUT) to be used in the temperature control algorithm. Both the primary probe,
located in the chamber airstream or platform, and the secondary probe, typically
located inside the UUT, are used to provide a chamber or platform response that
can accelerate testing while respecting the absolute and relative limits of all the
affected components.
Common single probe control strives to maintain the Setpoint temperature in the
chamber airstream, or at the platform surface. If the UUT is massive, or is a poor
thermal conductor, the internal temperature of the UUT can lag the chamber or
platform temperature considerably. Conversely, using a second probe, buried
inside the UUT, to control the temperature may achieve better UUT interior
temperature control, but it will do so at the risk of extreme temperatures in the
chamber or on the platform. If not carefully monitored, second probe only control
can result substantial damage to the chamber or platform and UUT and risk
operator injury.
Intelligent 2 Probe Control is designed to achieve the Setpoint temperature inside
the UUT (probe 2) either as quickly as possible, or at a controlled ramp rate, while
always respecting the limits of the controller, chamber or platform, and UUT. The
user may specify the absolute limits of the exterior of the UUT as well as limit
themal shock by specifying a dynamically changing “sliding scale” maximum
temperature differential for the UUT skin to core temperature. Intelligent 2
Probe Control will maximize speed in achieving internal UUT Setpoint
temperatures, while, at the same time, controlling the thermal stress on the UUT.
An in depth discussion of this feature can be found in Section 4.
1.5.12 Default Setup Parameters Restore
There is a procedure for erasing the current setup parameter table data and
restoring it basic default values. See Section 3.3
C4 Manual Rev 7.5.2
15
1.5.13 Program Mode Step Insert & Delete
Program steps may now be deleted from or inserted into programs.
See Sections 6.5 & 6.6.
1.5.14 Program Mode Any Step Points to Step 100
Any program step may now point to step 100 (program end) as the next step to
execute.
1.5.15 Program Mode Safer Program Clear
Some deliberate delays have added to the key sequence to completely clear the
program memory to lessen the likelihood that all of program memory will be
erased by accident.
1.5.16 Program Mode Run Time Program Pre-check
When a program is run in Program Mode, the C4 pre-reads the program to look
for run-time errors that it can report to you before starting. By pre-checking your
program, errors are dealt with immediately rather than after the program has
partially completed. The following items are checked:
Calls for probe 2 when only one probe is defined for the system
Calls for setpoints that are not within the system operating range
Loop counter numbers not within the range of 1 to 999 integer
See Section 6.7.2 for a full explanation.
1.5.17 Remote Mode EIA-232 Baud Rate Improvement
EIA-232 communications are now supported at 19,200 and 38,400 bps.
See Section 8.
16
C4 Manual Rev 7.5.2
1.5.18 Remote Mode EIA-232 Port Initialization
The CC-3 required that to use the EIA-232 port, the port had to be initialized by
switching the mode switch to SETUP before switching to REMOTE mode. The C4
eliminates this requirement. The EIA-232 port is initialized each time the remote
switch is switched to REMOTE mode.
1.5.19 Remote Mode Fault Tolerant Parser
The C4 uses a very fault tolerant parser. Command strings received over either
the EIA-232 or GPIB ports are converted to upper case, extra spaces and tabs are
removed, commas are converted to spaces and line terminators are corrected if
necessary. For this reason, programs that work properly with the C4 and not with
a CC-3, that use only CC-3 commands, probably have syntax errors that the C4
parser corrects. See Section 9.7.7
1.5.20 Remote Mode System Information Queries
QVQuery Firmware Version ..................... See Section 7.4.1
QNQuery Serial Number ........................ See Section 7.4.2
QRQuery Temperature Range .................... See Section 7.4.3
QSQuery Setpoint & Control Probe Number ........ See Section 7.4.4
QFQuery Setup Parameter Data (Binary Query)..... See Section 7.4.5
QFA Query Setup Parameter Data (ASCII Query) ..... See Section 7.4.5
1.5.21 Remote Mode Operation Information Queries
RSARequest Status Byte (ASCII Query) ............. See Section 7.5.1
REA Request Error Byte (ASCII Query) ............. See Section 7.5.2
QEQuery Error/Status String (Binary Query) ....... See Section 7.5.3
QEA Query Error/Status String (ASCII Query) ........ See Section 7.5.3
QCQuery Last Command........................ See Section 7.5.4
C4 Manual Rev 7.5.2
17
1.5.22 Remote Mode Setup Parameter Commands
SCSet Correction (Calibration) for Probe ........... See Section 7.6.1
WPSet PID Constants ........................... See Section 7.6.2
BFTurn Blowers Off ............................ See Section 7.6.3
BOTurn Blowers On ............................ See Section 7.6.3
SLSet UUT Temperature Limits .................. See Section 7.6.4
SDSet UUT Temperature Differential Limits........ See Section 7.6.5
UPWrite SRAM Parameters to EEPROM ........... See Section 7.6.6
1.5.23 Remote Mode IEEE-488 (GPIB) Monitoring
The IEEE-488 bus controller is automatically reset if a problem is detected. No
message to the user is generated as no data is typically lost in the process.
1.5.24 Setup Mode Easier Parameter Access
Setup parameters may now be accessed directly by number rather than having to
scroll the entire list. If scrolling is used, there is now a backup key so the list can
be scrolled in either direction. Likewise, for parameters that are chosen from a
list, the list can be scrolled in either direction or the parameters can be keyed in
directly without scrolling the list. See Section 8.1.
18
C4 Manual Rev 7.5.2
2. PHYSICAL DESCRIPTION
The SIGMA SYSTEMS Model C4 Controller fits into an area 3.5" x 5.5" x 9.5"
long. Connection to the chamber or platform is made through an umbilical cable
with a 12 pin connector that is standard for all SIGMA chambers and thermal
platforms. Controllers destined for use with SIGMA thermal platforms have an
additional 6 pin connector used for the sensor probe circuits. All C4 controllers
have a 3 pin “pigtail” connector at the back of the controller for control of
refrigeration via solid state relay. This feature was optional on CC-3 controllers
and is thus not a consistent feature of CC-3.5 controllers.
2.1 Front Panel
The front panel consists of four major parts: the LED digit display, the LED
indicators, the mode switch and the keypad. The numbers in circles on the front
panel guide the user through the steps for changing the temperature setting in
the LOCAL (manual) mode of operation. See Section 5.
2.1.1 Digital LED Display
The Digital LED Display is a 7 digit display arranged in the following
configuration:
***-****
The display is used for showing probe temperatures, setpoints, program steps,
fail-safe status and setup information.
2.1.2 LED Indicators
Below the Digital LED Display are four discrete LED Indicators. These indicators
are arranged in the following order:
REFERCOOL
HEAT
RUN
REFER indicator lights if power is applied to the refrigeration compressor
control port (not all units are equipped with compressors).
C4 Manual Rev 7.5.2
19
COOL indicator lights when cooling is active, typically during the time the
cryogenic valve is open on cryogenically cooled units.
HEAT indicator is pulsed along with the on and off function of the heaters.
These indicators are active in all modes of operation.
RUN indicator shows whether the chamber temperature is being controlled.
2.1.3 Mode Switch
The Mode Switch is used to apply power to the chamber and to select the mode of
operation. It is a 5 position rotary switch arranged as follows:
The OFF position removes the power from heating and cooling and compressor
control circuits of both the controller and the chamber or thermal platform.
CAUTION: Placing the mode switch in the OFF position
does not remove all power from either the controller or the
chamber or thermal platform. Only the heating, cooling,
and compressor control circuits are turned off. Full line
voltage potential is still available in many places in both the
controller and the chamber or platform. See servicing
warnings and instructions in the appendix of this manual.
Moving the rotary switch changes the mode of operation of the chamber. The
different modes are described in detail in separate sections of this manual. When
the rotary mode switch is moved to a new position, other than OFF, the
temperature control loop, if running, is turned off, heating and cooling are
disabled but the chamber blowers will continue to run unless disabled via setup
20
C4 Manual Rev 7.5.2
parameter F15. Likewise, at the end of a local or remotely controlled program,
chamber blowers will continue to run, but heating and cooling will be disabled.
Note that a chamber in this condition, with blowers running, will exhibit some
heating due to blower air friction. This effect is exaggerated in units equipped
with high velocity blowers. Moving the rotary switch between OFF and LOCAL
will maintain the last used setpoint for the next operation of the controller. See
6.3 for description of auto-start function.
2.1.4 Keyboard
The keyboard consists of 20 momentary contact keys. Certain keys are functional
only in some of the modes of operation. With some keys, such as the Display
Control, the function is different depending on the mode of operation. The
Keyboard is arranged in the following configuration:
123
456
789
.
CLEAR PROGClear program from memory
START STOPStart/Stop temperature control
REFERToggle refrigeration compressor on or off *
ENTEREnter/finalize current keyboard entry
ADVAdvance to next program step
DISP TEMPDisplay temperature
CLEAR ENTRYClear current keyboard/display entry
DISP CNTLDisplay control setpoint
0
+
-
CLEAR
PROG
REFERENTER
ADV
CLEAR
ENTRY
START
STOP
DISP
TEMP
DISP
CNTL
* = LOCAL MODE ONLY.
C4 Manual Rev 7.5.2
21
2.2 Rear Panel Connections
At the rear panel of the controller, a cable is provided to connect to the power,
cooling solenoid, mechanical refrigeration and heaters of the chamber. In
addition, a six lug screw terminal block (J1) is provided for eyelet terminal
connection of the temperature probes and for connection of one additional optional
device such as the Sigma PFS-2 Precision Fail Safe.
J1 TOP
Probe 1 (Black lead)
ì
Probe 2 (Black lead)
í
Chassis Ground (Probe shield)
î
(Chamber or platform ground, etc.)
Probe 1 & 2 return (Red or White lead)
ï
Auxiliary device ground
ð
12 VDC for Fail-Safe
ñ
Please observe all standard anti-static procedures when making connections to
these points!
There is also an IEEE-488 (GPIB) connector, series 57 (metric threads), and an
EIA-232 connector, female DB-25, for the remote modes of operation. Two TTL
outputs and one TTL input are also available on J6. (See Section 9.5)
22
C4 Manual Rev 7.5.2
3. GENERAL OPERATION & ERROR CONDITIONS
3.1 Startup Displays
3.1.1 Model Number and Firmware Version Number Display
The C4 identifies itself upon power up. It will display the model number for 1
second, then the firmware version number for 2½ seconds as follows:
c3-5rel 7.5.2
c4rel 7.5.2
3.1.2 Temperature Range Display
Each C4 controller is set at the factory for use with a specific device (chamber or
platform). Because the controlled device was made to specific thermal limit
specifications, and because constraining the operating range of the controller
improves it’s accuracy, the controller to be used with each device is set to operate
only within the range appropriate for that device. The setting of an operating
range involves a number of internal adjustments and calibrations and the loading
of specific control tables for that range. This setting can only be changed by
Sigma Systems service personnel. We recommend that each controller be kept
with the device for which it was originally configured.
However, if you find that it is necessary to move a controller to another device, it
is important that you check to be sure that the controller that is moved is
configured to operate in the appropriate range for the device it is to control. As of
November, 1998, controllers are supplied for five ranges as follows:
/C to 200/C-148/F to 392/F-100/C to 300/C-148/F to 572/F
-100
-100
/C to 350/C-148/F to 662/F-175/C to 400/C-283/F to 752/F
-195/C to 300/C-319/F to 572/F
To check the internal range setting of your C4 controller, turn the mode switch
from OFF to any other position. During the 2½ seconds that the firmware release
number is shown on the display, rapidly press <DISP/TEMP> three times. The
internal range setting will be displayed for 5 seconds following the version display.
An example:
200 - 100c392 - 148f
(Celsius mode)(Fahrenheit mode)
C4 Manual Rev 7.5.2
23
3.1.3 Serial Number Display
The controller serial number can be displayed at startup by pressing
<DISP/CTRL> while the model number (
number will be displayed in the format:
c3-5 or c4) is displayed. The serial
4-0 3276
The first digit (either a 3 or 4) indicates whether the controller is a CC-3
and the 5 digits to the right of the dash is the sequential part of the number.
There is no significance to the leftmost of the 5 sequential digits being separated
from the other 4 digits. The separation is a limitation of the display. When
recording or reporting serial numbers, please always use the 7 digit string,
including the leftmost digit and the dash in this format: 4-03276.
When the serial number is displayed, the controller is not controlling. The
number will remain on the display until the controller is turned off.
3.2 Updating Firmware
There are two methods of updating the firmware in the C4.
.5
or C4,
1. Physically replace the EEPROM that contains the firmware.
(There is usually a charge for firmware EEPROMs)
.
2. Use a PC to upload the firmware through the C4 serial port.
(Firmware and upload software are free from Sigma
System’s internet sites, www.SigmaSystems.com or
ftp.SigmaSystems.com)
Note: All firmware updates are available on EEPROM chips for physical
installation. Not all updates are available as downloads from the Sigma Systems
web site. The reason for this is that some firmware updates make changes in the
basic data structures inside the C4. Because firmware loaded into the C4 through
the serial port is stored separately, and in addition to, the EEPROM firmware,
and because the C4 can be started from either resident version, the two versions
must use compatible data structures. Thus, versions that will modify the data
structures in the C4 may only be installed by changing the firmware EEPROM.
24
C4 Manual Rev 7.5.2
3.2.1 Upgrading Firmware by PROM Replacement
Before starting, be certain that you have a properly grounded antistatic
surface and a grounding strap to prevent damaging the C4 components
during disassembly and reassembly.
Remove the C4 from its cabinet or rack, then remove the EEPROM chip that
contains the firmware and replace it with a new one containing the updated
firmware. Firmware
EEPROMs are available
from Sigma Systems.
There is a charge for
firmware EEPROMs.
The firmware EEPROM
is located on the digital or
“A” board, the same board
that has the EIA-232 and
GPIB connectors on the
end. The location of the
firmware EEPROM is
shown by the triangle
pointer. The EEPROM
socket is labeled U13. Make certain that the end of the EEPROM with the notch
or notch mark is toward the inside, or center, of the board.
Note that it is possible to remove and replace the firmware EEPROM without
disassembling the C4. If you lay the controller upside down, and look at the
digital board from the bottom, the firmware EEPROM is the first large chip from
the front of the controller. You will not be able to see the U13 label, but you can
see the C32 label on the small capacitor located next to the firmware EEPROM
socket. Using a small flat bladed screwdriver or chip removal tool, carefully pry
the old EEPROM from it’s socket. Try to keep the chip as flat as possible while
removing it to prevent damage to the pins or socket. When replacing, keep the
replacement EEPROM flat to the socket, make sure all pins are started in their
respective receivers in the socket, then press the chip firmly into place.
3.2.2 Upgrading Firmware by Uploading through C4 Serial Port
Note: CC-3.5 controllers with version 7.0.0 firmware or earlier have battery
backed RAMS for non volatile storage. These units MUST have the BBRAM
replaced with an EEPROM before newer firmware is installed.
C4 Manual Rev 7.5.2
25
Use a DOS bootable diskette and a PC type computer to load the new firmware
through the C4's serial port. You can obtain the required diskette from Sigma
Systems for a fee, or you can provide your own DOS bootable diskette and obtain
the necessary files for the upload process from the Sigma Systems site on the
Internet. The Internet address for this site is:
www.SigmaSystems.com
The download file, SSUPxxxx.EXE, may contain specific instructions that replace
the instructions in the next paragraph. After downloading the file, copy it to an
otherwise empty DOS bootable diskette and run SSUPxxxx.EXE. This file is a
self extracting ZIP archive that will install all the files you need onto the diskette.
Check for the presence of a README.TXT file on the diskette for instructions that
supplement or supercede these instructions.
Making the physical connection for upload
To upload firmware to your C4, you will need to connect the C4 to a DOS bootable
PC compatible computer using a straight-through, or modem type serial cable.
The C4 is configured as DCE (Data Communications Equipment). Therefore, do
NOT use a “null modem” type serial cable that has pins 2 & 3 crossed. Connect
the cable to either COM1 or COM2 on the PC and to the female DB25 connector
on the back of the C4.
Running the upload software on the PC
After you run the SSUPxxxx.EXE file on your DOS bootable diskette you are
ready to run the software. Place the floppy in the “A” drive of the PC and turn on
or reset the PC to allow it to boot from the floppy. Follow the instructions on the
screen. Note that the menu will allow you to do successive uploads without
restarting and will allow you to change serial ports as you do so. Users who will
be uploading to a number of controllers in one session will find that this feature
saves some time.
If the uploads are being done on a bench (as opposed to leaving the controller in
it’s chamber or platform housing), be certain to follow all precautions about
removing all power before removing the controller and about taking care to use
proper anti-static procedures when handling the controller. Be certain that the
bench has a properly grounded anti-static surface.
If you are going to upload firmware into many controllers successively in a bench
environment, you can use a PC with two serial ports (COM1 and COM2) and two
serial cables. You can then have the upload program alternate serial ports for the
26
C4 Manual Rev 7.5.2
uploads so that you can upload to one controller while connecting and
disconnecting the other.
Starting the upload on the C4
When the PC is ready, turn the C4 controller on. During the 1 second display of
the model number, rapidly press <CLEAR/PROG> three times. The display will
read:
rs load
The firmware upload should begin immediately. The PC will indicate when the
load is complete.
If you have difficulties, see the trouble shooting information in Section 9.7.8.
3.3 Restoring Setup Parameters to Default Values
Note: The default values are very generic. They are not the values that were
likely in the controller when it was shipped from the factory. After completing
this procedure, please check each value to make certain that it is appropriate.
To restore the default values to the setup parameter table, turn the controller off,
then back on. During the 1 second in which the controller model is displayed,
press <CLEAR/ENTRY> very quickly 3 times. The display should then read
su res
Press <ENTER> to confirm that you want restore the default values. (Any other
key will abort the process). The display should then read
ee busyand thensu done
Turn the controller off, then on again to resume operation with the new values.
The first time the controller is turned back on after restoring the default setup
parameters, the display will show
turned to SETUP. (You may see an ee busy display and model number
and/or firmware version number at this time.) The controller forces you to SETUP
mode before it will function as a reminder that the default values have been
loaded and that the parameters needed for your operation have not yet been set.
all res . The mode switch must then be
C4 Manual Rev 7.5.2
27
When you have set the necessary parameters for your needs, you are ready to go
to LOCAL, PROGRAM, or REMOTE mode.
28
C4 Manual Rev 7.5.2
3.4 Fahrenheit Operation
The C4 can use either Celsius or Fahrenheit temperature scales. Celsius is the
default mode. When the controller is operating in Celsius mode it’s behavior is
identical to the model CC-3 controller. When the controller is operating in
Fahrenheit mode, both the displays and the bus communications are different.
These differences are intentional and designed to prevent errors that might arise
from a user using setpoints in one scale while the controller is operating in the
other scale. Note, however, that there is no difference in the Program mode
programming operation of the C4 when operating in either Fahrenheit or Celsius
mode. It is incumbent upon the user to notice the differences in the temperature
displays and program accordingly.
To change the temperature scale, use SETUP mode (See Section 8) to access setup
parameter F16. Select either of these field values:
0Celsius
1Fahrenheit
When in Fahrenheit mode, temperature displays have an “F” following the probe
identifier on the left side of the display. Example:
p1f 102.4
Likewise, temperature inquires over the bus (PT command - See Section 7.5.7)
will return a string with an “F” as the third byte in the string. The string will
therefore be 1 byte longer than the string returned in Celsius mode. This
difference was intentional as it requires a modification of the parsing routine that
will assure that Celsius and Fahrenheit data are recognized properly. Thus, if the
current probe 1 temperature is 86.7/F, the result of a Fahrenheit mode get
temperature query for probe 1 (query command = PT1) would be
T1F 86.7<CR><LF>
If the temperature at probe 1 was 55.4/C, the same command in Celsius mode
would return
T1 55.4<CR><LF>
During Remote mode operation, the C4 may be queried to determine the current
temperature scale by using the QS, Query Setpoint, command. See Section 7.4.4.
C4 Manual Rev 7.5.2
29
3.5 System Operating (Temperature) Range
The C4 controller operates within the limits of the system devices. There are 3
pairs (low-high) of temperature limits that constrain the range of operations.
1.The range of the C4 as it is set at the factory
2The range of the controlled device (chamber or platform) as
described by setup parameters F25 & F26.
3The range of the UUT (unit under test) as described by
setup parameters F27 & F28.
Each of these ranges is characterized by a low limit and a high limit. The highest
of the 3 low limits is the low temperature System Operation Limit. The lowest
of the 3 high limits is the high temperature System Operation Limit. In other
words, the controller will respect the most restrictive limits that are described by
the combination of the 3 ranges listed above.
The low and high temperature System Operating Limits define the SystemOperating Range. The controller will not accept setpoints outside the System
Operating Range and will report error conditions and stop controlling if the
controlling probe(s) report a temperature too far outside that range.
Note: Probe 1 is always in the chamber airstream or platform and is always
monitored for conformance with the System Operating Range. Probe 2 is
monitored for conformance only if it is being used for control.
3.6 Probe Out of Range Shutdown
When the C4 is actively controlling the temperature of a chamber or platform, it
frequently checks to be certain that the control temperature, as reported by any
active control probe, is within the System Operating Range.
In the event that the temperature reported by the active control probe is more
than 20/C, and less than 50/C, outside the controller’s System Operating Range,
then the controller will presume that a “run away” condition exists. The controller
will turn off all heating and cooling and stop controlling. Blowers will be left in
the condition determined by setup parameter F15 (Blower shut-off mode). The
controller display will show
p1 lop1 hi
30
C4 Manual Rev 7.5.2
Loading...
+ 90 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.