Sierra Mass-Trak 810M, Mass-Trak 810C, Mass-Trak 810S-L, Mass-Trak 810 Series, Mass-Trak 810S-M Instruction Manual

Sierra 810 Series
Mass-Trak
®
Mass Flow Instruments
Part Number: IM-81
Revision H 4/15
GLOBAL SUPPORT LOCATIONS: WE ARE HERE TO HELP!
IMPORTANT CUSTOMER NOTICE: OXYGEN SERVICE
Sierra Instruments, Inc. is not liable for any damage or personal injury, whatsoever, resulting from the use of Sierra Instruments standard mass flow meters or controllers for oxygen gas. You are responsible for determining if this mass flow meter or controller is appropriate for your oxygen application. You are responsible for cleaning the mass flow meter or controller to the degree required for your oxygen flow application.
© COPYRIGHT SIERRA INSTRUMENTS 2014
No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language, in any form or by any means, electronic, mechanical, manual, or otherwise, or disclosed to third parties without the express written permission of Sierra Instruments. The information contained in this manual is subject to change without notice.
TRADEMARKS
Mass-Trak® 810 Series is a Registered Trademark of Sierra Instruments, Inc. Other product and company names listed in this manual are trademarks or trade names of their
respective manufacturers.
CORPORATE HEADQUARTERS
5 Harris Court, Building L Monterey, CA 93940 Phone (831) 373-0200 (800) 866-0200 Fax (831) 373-4402
www.sierrainstruments.com
EUROPE HEADQUARTERS
Bijlmansweid 2 1934RE Egmond aan den Hoef The Netherlands Phone +31 72 5071400 Fax +31 72 5071401
ASIA HEADQUARTERS
Second Floor Building 5, Senpu Industrial Park 25 Hangdu Road Hangtou Town Pu Dong New District, Shanghai, P.R. China Postal Code 201316 Phone: + 8621 5879 8521 Fax: +8621 5879 8586
Series 810 Table of Contents
TABLE OF CONTENTS
Chapter 1 Introduction
Introduction .................................................
1-1
Using this Manual ...................................
1-2
Safety Information ...............................
1-3
Receipt of System Components .........
1-3
Technical Assistance .............................
1-3
The 810 Series Flow Sensing Principle
1-4
Chapter 2 Installation
Installation Overview .....................................
2-1
Before Beginning an Installation.....................................
2-1
Installing the Transducer (Plumbing) ...........
2-2
Compression Fittings ...........................
2-2
VCO Fittings ........................................
2-3
VCR Fittings .........................................
2-3
1/4-inch Female NPT Fittings .............
2-3
Installing the Transducer (Electrical) ...........
2-4
D-connector PIN assignments .............
2-5
Chapter 3 Operation
Mass Flow Meter Operation .........................
3-1
Mass Flow Controller Operation ...................
3-1
Setpoint Adjustment ...............................
3-2
Setpoint Configuration............................
3-3
Over-Range Condition ...........................
3-4
Auto Shut-Off Function .......................
3-5
Manual Valve Override .......................................
... 3-5
ValvePurg e.. ... .......... ... ....... ..
............................. 3-5
Chapter 4 Maintenance
Flow Path Maintenance and Cleaning
4-1
Sensor Maintenance and Inspection .........
4-2
Valve Maintenance ..........................................
4-3
Valve Adjustment Procedure .................
4-4
Transducer Calibration .................................
4-5
Chapter 5 Troubleshooting
Transducer Troubleshooting Charts...........
5-1
Returning Equipment to the Factory ............
5-2
Table of Contents Series 810
Appendix A Conversion Formulas and Gas Tables
Appendix B Product Specifications & Mounting
Appendix C Pin Connections
Appendix D Purge and Valve Off Connections
Appendix E Warranty Policy
List of Figures
1-1. Mass-Trak Features (Typical) ...............................
1-1
1-2. Flow Paths through the Transducer ....................
1-4
1-3. Flow Measuring Principle .....................................
1-4
1-4 Sensor Temperature Distribution ..... ........ ......... ...... .... ...1-5
1-5 Linear Range of Transducer Output Signal .........................1 -5
2-1. Minimum Restriction Diameters ..................................................... 2-4
2-2. Transducer D-Connector Pin Assignments .................................
2-5
3-1 Dip Switch Settings.............................. ................ .... ..
3-3
4-1. Stainless Steel Low Flow Internals ............................................
4-2
4-2. Stainless Steel Medium Flow Internals .......................................
4-2
4-3. Nylon Transducer Internals .........................................................
4-2
Series 810 Instruction Manual Chapter 1 Introduction
Chapter 1 Introduction
Sierra's Mass-Trak' mass flow meters and controllers are designed to accurately measure and control flows of process gases. This instruction manual covers the installation, operation and maintenance of the Mass-Trak product line, which includes the following Sierra models:
Model 810M Mass-Trak Flow Meter Model 810C Mass-Trak Flow Controller Model 810S-L Mass-Trak Flow Controller for flows up to 15
SLPM
Model 810S-M Mass-Trak Flow Controller for flows from 15
to 100 SLPM All of these models can be ordered with or without a digital display.
Sierra Instruments' Series 810 Mass-Trak products can be calibrated to measure and control the mass flow rate of gases in several ranges from 0-10 standard cubic centimeters per minute (SCCM) to 0-100 standard liters per minute (SLPM). Accuracy is ±1.5% of full scale over a wide temperature and pressure range, and time response is 5 seconds to within 2% of set point. Detailed product specifications may be found in Appendix B.
Mass -Trak Design (typical)
Figure 1-1. Mass-Trak Design (typical)
Mass-Trak is ideally suited to a wide range of gas flow applications including general process control, laboratories, instrument OEM's, gas panels, and flow calibration.
FLOW
ZERO
SET POINT
IM-81 1-1
Chapter 1 Introduction Series 810 Instruction Manual
Mass-Trak requires a regulated 24 VDC external power source. The 0-5 VDC and 4-20 mA output signals, which are linearly proportional to gas mass flow rate, are provided for recording, data logging, or control. A 15-pin "D" connector is provided for power input, output signal and set point control. Mass-Trak is available in a variety of configurations with either a nylon or stainless steel flow body. A 24VDC power supply is available as an option.
Using This Manual
This manual is organized into five chapters:
Chapter 1 includes the introduction and theory of operation
Chapter 2 provides installation, plumbing and wiring
instructions
Chapter 3 describes operation and features
Chapter 4 covers routine maintenance
Chapter 5 contains the troubleshooting guide and service
recommendations
There are also 5 Appendices:
Appendix A lists conversion formula and gas tables
Appendix B includes the product specifications,
dimensional drawings and mounting instructions
Appendix C describes the PIN connections
Appendix D describes the purge and valve off connections
Appendix E details the warranty policy for the product
Throughout this manual, we use the word
transducer
as a generic term to represent all models of Sierra's 810 Series Mass-Trak instruments.
1-2 IM-81
Series 810 Instruction Manual Chapter 1 Introduction
Safety Information
Caution and warning statements are used throughout this book to draw your attention to important information.
Warning! Caution!
This statement appears with information that This statement appears with information that is is important to protect people and equipment important for protecting your equipment and from damage. Pay very close attention to all performance. Read and follow all cautions that warnings that apply to your application. apply to your application.
Receipt of System Components
When receiving a Sierra transducer, carefully check the outside packing carton for damage incurred in shipment. If the carton is damaged, notify the local carrier and submit a report to the factory or distributor. Remove the packing slip and check that all ordered components are present and match your specifications (as ordered). Make sure any spare parts or accessories are not discarded with the packing material. Do not return any equipment to the factory without first contacting Sierra Customer Service.
Technical Assistance
If you encounter a problem with your transducer, review the configuration information for each step of the installation, operation and set up procedures as explained in this manual. Verify that your settings and adjustments are consistent with factory recommendations. Refer to Chapter 5, Troubleshooting, for specific information and recommendations.
If the problem persists after following the troubleshooting procedures outlined in Chapter 5, contact Sierra Instruments by fax or by E-mail (see inside front cover). For urgent phone support you may call (800) 866-0200 or (831) 373-0200 between 8:00 a.m. and 5:00 p.m. PST. In Europe contact Sierra Instruments BV at +31 72 5071400. When contacting Technical Support, make sure to include this information:
The flow range, serial number, Sierra order number and
model number (all marked on the transducer nameplate)
The problem you are encountering and any corrective action
taken
Application information (gas, pressure, temperature, pipe
and fitting configuration)
IM-81 1-3
Chapter 1 Introduction Series 810 Instruction Manual
The 810 Mass -Trak Flow Sensing Principle
The operating principle of Mass-Trak transducers is based on heat transfer and the first law of thermodynamics. During operation process gas enters the instrument's flow body, where it encounters the laminar flow bypass. The laminar flow bypass generates a pressure drop, P1~P2, forcing a small fraction of the total flow to pass through the sensor tube (m1). The remainder of the flow continues through the body (m2). (Figure 1-2)
Figure 1-2. Flow Paths through the Transducer
Two resistance temperature detector (RTD) coils around the sensor tube direct a constant amount of heat (H) into the gas stream. In actual operation, the gas mass flow carries heat from the upstream coil to the downstream coil. The resulting temperature difference (OT) is detected by the RTD coils and gives the output signal. Since the molecules of the gas carry away the heat, the output signal is linearly proportional to gas mass flow. (Figure 1-3)
Figure 1-3. Flow Measuring Principle
Figure 1-3 shows the mass flow through the sensor tube as inversely proportional to the temperature difference of the coils.
1-4 IM-81
Series 810 Instruction Manual Chapter 1 Introduction
Figure 1-4. Sensor Temperature Distribution
The RTD coils are legs of a bridge circuit with an output voltage in direct proportion to the difference in the coils' resistance; the result is the temperature difference (OT), Figure 1-4. Two other parameters, heat input (H) and coefficient of specific heat (Cp) are both constant. The resulting output is nearly linear over the transducer's normal operating range (Figure 1-5).
Figure 1-5. Linear Range of the Transducer's Output Signal
For mass flow
controllers
, once the gas flows through the monitoring section, it is then controlled by the built-in servo-control valve. All models of the Mass-Trak utilize Sierra's proprietary high­efficiency Fast-Trak Electromagnetic Valve. The normally closed Fast-Trak valve is similar to an on/off solenoid valve, except that the current to the valve coil, and hence the magnetic field, is modulated so that the ferromagnetic valve armature, or valve plug, assumes the exact height above the valve's orifice required to maintain the valve's command flow. The result is nearly infinite resolution.
IM-81 1-5
Series 810 Instruction Manual Chapter 2 Installation
Chapter 2 Installation
Installation Overview
Mass-Trak- transducers are supplied with female NPT, compression, VCO, or VCR process connections. To ensure a successful installation, inlet and outlet tubing should be in a clean state prior to plumbing the transducer into the system. The shipping caps covering the inlet/outlet fittings should not be removed until immediately prior to installation.
Before Beginning the Installation
Before installing the transducer, verify the following:
1. Ensure that the installation site conforms to the specific operating
parameters recorded on the transducer's nameplate.
This is critical as each transducer is factory-configured for a specific gas and flow range, pressure differential, temperature range and mounting orientation. The line pressure should not exceed the upper limit for the model in hand: 150 psig (Models 810M, 810C), 500 psig (Model 810S-M) or 1000 psig (Model 810S-L). The temperature should not exceed 122°F (50°C).
2. Double-check to be sure that the transducer o-ring material is compatible
with the gas to be measured.
3. If the gas contains any particulate matter, install an in-line filter
upstream of the transducer. Recommended filter size: 15 micron for flows of 10 to 30 slpm, 30 micron for flows above 30 slpm.
4. Do not locate the transducer in areas subject to sudden temperature
changes, excessive moisture, and frequent drafts or near equipment radiating significant amounts of heat. Be sure to allow adequate space for cable connectors and wiring.
5. Ensure the installation location provides a minimum of two inches of
straight pipe upstream and downstream from the process connections.
6. For controllers, use a properly sized pressure regulator and verify that
the controller orifice size is the smallest in the system. There can be no restrictions (such as valves, tubing or pipe internal diameters, reducers, etc.) upstream or downstream of the controller with a dimension that is less than the valve orifice diameter.
7. Output Signals: The Mass-Trak has 0-5 VDC (0-10 VDC optional) and 4-20 mA output signals that are linearly proportional to the gas mass flow rate. The instrument was calibrated using either voltage or current output as specified on the data tag. Note that there may be a slight difference between the voltage and the current output signals (up to
IM-81 2-1
Chapter 2 Installation Series 810 Instruction Manual
Caution!
Only qualified
personnel should
install the transducer.
1 %). Using the output that is recorded on the data tag will optimize instrument performance.
8. Integral Display: The 3 1/2 digit LCD display reads directly in
engineering units or percent of full scale. The full-scale range and gas are shown on the instrument data tag. The decimal point for the flow rate is set at the factory and will show automatically (e.g., "5.54" SLM or "76.4" %).
9. Over range conditions are indicated by the display and/or output going
to a high level, approximately 150% of the full-scale range. After the over-range condition has been removed, it may take several minutes for the Mass-Trak to recover and resume normal operation.
10. The transducer has stringent power supply requirements. Because the valve on flow controllers is operated in a control loop, power supply variations cannot be tolerated. See Appendix B for a complete listing of power requirements.
Installing the Transducer-Plumbing
Follow the installation instructions that are applicable to your transducer's process connection. Before use, all plumbing should be checked carefully for leaks and the transducer should be purged with dry nitrogen. Ensure that the tubing is free from burrs, or sharp rims that may result from cutting.
CAUTION: All instruments are leak-tested prior to shipping. To check your installation, test the fittings only. Do not use liquid leak detectors such as Snoop® to search for leaks inside or outside the Mass-Trak. Instead, monitor pressure decay.
Compression Fittings
1. Position the transducer with the flow direction arrow pointing in the direction of flow.
2. Verify the position of the front and back
ferrule as shown above. Insert the tubing into the fitting. Be sure that the tubing rests firmly on the shoulder of the fitting and that the nut is finger-tight. Scribe the nut at the six o'clock position.
3. While holding the fitting body steady with a backup wrench , tighten
the nut 1-1/4 turns, watching the scribe mark make one complete revolution and continue to the nine o'clock position. For 1/16-inch, 1/8-inch and 3/16-inch (2, 3 and 4 mm) sizes, tighten only 3/4 turns from finger-tight. Do not over-tighten!
2-2 IM-81
Series 810 Instruction Manual Chapter 2 Installation
4. Check the system's entire flow path thoroughly for leaks. (Do not use liquid leak detectors. Instead, monitor pressure decay.
Exposing the transducer to leak detector fluid may damage the unit.)
VCO Fittings
1. Position the transducer with the flow direction arrow pointing in the
direction of flow.
2. Tighten the nut finger-tight, and then 1/4 turn tighter with a wrench.
Do not over-tighten!
3. Check the system's entire flow path thoroughly for leaks. (Do not
use liquid leak detectors. Instead, monitor pressure decay. Exposing the transducer to leak detector fluid may damage the unit.)
VCR Fittings
1. Position the transducer with the flow direction arrow pointing in the
direction of flow.
2. Install new washers that are compatible with the gas to be used.
3. Tighten the nut finger-tight, and then 1/4 turn tighter with a wrench. Do
not over-tighten!
4. Check the system's entire flow path thoroughly for leaks. (Do not
use liquid leak detectors. Instead, monitor pressure decay. Exposing the transducer to leak detector fluid may damage the unit.)
1/4 inch Female NPT
(standard on nylon flow bodies
)
1. Position the transducer with the flow direction arrow pointing in the
direction of flow.
2. Use a good quality paste pipe thread sealant. Apply to the inlet and
outlet fittings.
3. Tighten the fittings by hand. Then, tighten no more than 1 turn.
Caution! Do not over-tighten. Damage to the instrument may result from over-tightening fittings.
IM-81 2-3
Chapter 2 Installation Series 810 Instruction Manual
IMPORTANTStraight-run requirement (critical for nylon flow body instruments
)
Install a section of straight pipe at least five pipe diameters in length upstream of the transducer. DO NOT use reducers. In the case of a flow meter, at least two pipe diameters downstream are required for optimum performance. In the case of a flow controller, there can be no restrictions (such as valves, tubing or pipe internal diameters, reducers, etc.) upstream or downstream of the MFC with a dimension that is less than the valve orifice diameter. Failure to comply with this requirement will result in severely impaired performance and possible oscillations in flow controllers. Refer to Table 2-1 below for minimum restriction diameters upstream or downstream of the flow controller.
Flow Ranges
Relative to N2
Valve Orifice Diameter
(Typical-in inches)
0-10 to 0-2000 SCCM
0.020
0-2 to 0-11 SLM
0.040
0-11 to 0-22 SLM
0.055
0-15 to 0-28 SLM
0.063
0-20 to 0-50 SLM
0.073
0-51 to 0-73 SLM
0.094
0-61 to 0-100 SLM
0.110
Table 2-1 Typical Minimum Restriction Diameters (Under Standard
Pressure Conditions of 30 PSIG Inlet and Ambient Outlet)
Installing the TransducerMechanical Mounting
Please refer to Appendix B for mechanical mounting instructions.
Installing the TransducerElectrical Connections
Caution!
The transducer is not a loop-powered device! Do NOT apply power to the 4-20 mA output or input connections.
The Mass-Trak transducers require a 24VDC power supply. If you are using the Power-Pak power supply, connect it to the 15-pin "D" connector on the side of the transducer. If you are supplying your own power source, it must be a regulated 24 VDC with ripple not to exceed 100 mV peak to­peak and capable of producing at least 340 mA (8 watts).
Mass-Trak is provided with a 15-pin "D" connector located on the side of the enclosure. The pin numbers and assignments for the Mass-Trak "D" connector are shown in Figure 2-2. The connections for input power, output signal and input signal (controllers only) are supplied via the "D" connector. Additional pin assignments can be found in Appendix C.
2-4 IM-81
Loading...
+ 28 hidden pages