Siemens SMD 18101 CS, SMD 18101 CP, SMD 18101 D Instruction Manual

SOLUTIONS
18” Gray Scale Flat Panel Display
Instruction Manual Version 11/2005
SMD 18101 CS SMD 18101 CP
display
SOLUTIONS
LC displays for medical diagnostics
18" Gray Scale Flat Panel Display SMD 18101 CS / SMD 18101 CP / SMD 18101 D
Instruction Manual
Introduction
Safety notes
Description
Application planning
Assembling
Connecting
Commissioning
1
2
3
4
5
6
7
Operation
Maintenance, service and disposal
Fault diagnostics
Technical data
Dimension drawings
Appendices
8
9
10
11
12
A
11/2005
A5E00472377A-002
Safety Guidelines
This manual contains notices you have to observe in order to ensure your personal safety, as well as to prevent damage to property. The notices referring to your personal safety are highlighted in the manual by a safety alert symbol, notices referring only to property damage have no safety alert symbol. These notices shown below are graded according to the degree of danger.
Danger
indicates that death or severe personal injury will result if proper precautions are not taken.
Warning
indicates that death or severe personal injury may result if proper precautions are not taken.
Caution
with a safety alert symbol, indicates that minor personal injury can result if proper precautions are not taken.
Caution
without a safety alert symbol, indicates that property damage can result if proper precautions are not taken.
Notice
indicates that an unintended result or situation can occur if the corresponding information is not taken into account.
If more than one degree of danger is present, the warning notice representing the highest degree of danger will be used. A notice warning of injury to persons with a safety alert symbol may also include a warning relating to property damage.
Qualified Personnel
The device/system may only be set up and used in conjunction with this documentation. Commissioning and operation of a device/system may only be performed by qualified personnel. Within the context of the safety notes in this documentation qualified persons are defined as persons who are authorized to commission, ground and label devices, systems and circuits in accordance with established safety practices and standards.
Prescribed Usage
Note the following:
Warning
This device may only be used for the applications described in the catalog or the technical description and only in connection with devices or components from other manufacturers which have been approved or recommended by Siemens. Correct, reliable operation of the product requires proper transport, storage, positioning and assembly as well as careful operation and maintenance.
Trademarks
All names identified by ® are registered trademarks of the Siemens AG. The remaining trademarks in this publication may be trademarks whose use by third parties for their own purposes could violate the rights of the owner.
Disclaimer of Liability
We have reviewed the contents of this publication to ensure consistency with the hardware and software described. Since variance cannot be precluded entirely, we cannot guarantee full consistency. However, the information in this publication is reviewed regularly and any necessary corrections are included in subsequent editions.
Siemens AG
Automation and Drives Postfach 48 48 90437 NÜRNBERG GERMANY
Order No.: A5E00472377A-002 Version 11/2005
Copyright © Siemens AG 2005. Technical data subject to change
Table of contents
1 Introduction............................................................................................................................................. 1-1
1.1 Contents of this documentation ................................................................................................. 1-1
1.2 Further documentation............................................................................................................... 1-1
2 Safety notes............................................................................................................................................ 2-1
3 Description.............................................................................................................................................. 3-1
3.1 Scope of delivery ....................................................................................................................... 3-1
3.2 Applications................................................................................................................................ 3-2
3.3 Important features...................................................................................................................... 3-2
4 Application planning................................................................................................................................ 4-1
4.1 Installation site ........................................................................................................................... 4-1
5 Assembling............................................................................................................................................. 5-1
5.1 Mounting of unit ......................................................................................................................... 5-1
6 Connecting ............................................................................................................................................. 6-1
6.1 General connection information................................................................................................. 6-1
6.2 Connection cover ....................................................................................................................... 6-2
6.3 Cable entry via stand ................................................................................................................. 6-2
6.4 Summary of signals and connections ........................................................................................ 6-3
6.5 Signal connections and DIP switches........................................................................................6-3
6.5.1 Signal connections..................................................................................................................... 6-3
6.5.2 DIP switch .................................................................................................................................. 6-4
6.6 Connecting the signal cables..................................................................................................... 6-4
6.7 Connecting the line cord ............................................................................................................ 6-6
6.8 Service interface ........................................................................................................................ 6-6
7 Commissioning ....................................................................................................................................... 7-1
7.1 Note on commissioning.............................................................................................................. 7-1
7.2 Control elements and power switches ....................................................................................... 7-1
7.2.1 Switching on the device ..................................
7.2.2 Key functions.............................................................................................................................. 7-2
........................................................................... 7-2
7.3 Description of OSD menu .......................................................................................................... 7-3
7.3.1 OSD overview ............................................................................................................................ 7-3
7.3.2 User groups and program levels................................................................................................ 7-4
7.3.3 Key functions in OSD menu....................................................................................................... 7-4
7.3.4 Main menu ................................................................................................................................. 7-5
7.3.5 Function description of OSD ...................................................................................................... 7-6
7.3.6 Force Mode.............................................................................................................................. 7-15
SMD 18101 CS / SMD 18101 CP / SMD 18101 D Instruction Manual, 11/2005, A5E00472377A-002
iii
Table of contents
7.3.6.1 Force Mode 1 ........................................................................................................................... 7-15
7.3.6.2 Force Mode 2 ........................................................................................................................... 7-16
7.4 Make settings ........................................................................................................................... 7-18
7.4.1 Adjusting the graphics card...................................................................................................... 7-18
7.4.2 Screen saver ............................................................................................................................ 7-18
7.4.3 Adjusting the image geometry ................................................................................................. 7-19
7.4.4 Adjusting the brightness and contrast...................................................................................... 7-19
8 Operation................................................................................................................................................ 8-1
8.1 Note for users............................................................................................................................. 8-1
8.2 Switching on the display............................................................................................................. 8-1
8.3 Control elements ........................................................................................................................ 8-2
9 Maintenance, service and disposal......................................................................................................... 9-1
9.1 Cleaning ..................................................................................................................................... 9-1
9.2 Disposal...................................................................................................................................... 9-2
10 Fault diagnostics................................................................................................................................... 10-1
11 Technical data ...................................................................................................................................... 11-1
11.1 Graphic display ........................................................................................................................ 11-1
11.2 Power supply............................................................................................................................ 11-1
11.3 Electronics................................................................................................................................ 11-2
11.4 RGB input................................................................................................................................. 11-2
11.5 Video inputs ............................................................................................................................. 11-3
11.6 Control and connection elements ............................................................................................ 11-3
11.7 Mechanical design ................................................................................................................... 11-4
11.8 Climatic conditions ................................................................................................................... 11-5
11.9 Mechanical requirements ......................................................................................................... 11-5
11.10 Safety specifications ................................................................................................................ 11-6
11.11 Electromagnetic compatibility .................................................................................................. 11-6
12 Dimension drawings ............................................................................................................................. 12-1
12.1 SMD 18101 CS/CP ...........................................
....................................................................... 12-1
12.1.1 Front view................................................................................................................................. 12-1
12.1.2 View from left ........................................................................................................................... 12-2
12.1.3 Rear view ................................................................................................................................. 12-2
12.1.4 Plan view.................................................................................................................................. 12-3
12.2 SMD 18101 D........................................................................................................................... 12-3
12.2.1 Front view................................................................................................................................. 12-3
12.2.2 View from left ........................................................................................................................... 12-4
12.2.3 Rear view ................................................................................................................................. 12-5
12.2.4 Plan view.................................................................................................................................. 12-5
12.3 Display stand............................................................................................................................ 12-6
12.3.1 Front view................................................................................................................................. 12-6
12.3.2 View from left ........................................................................................................................... 12-7
A Appendices.............................................................................................................................................A-1
A.1 Invalidity of guarantee................................................................................................................ A-1
SMD 18101 CS / SMD 18101 CP / SMD 18101 D
iv Instruction Manual, 11/2005, A5E00472377A-002
Table of contents
A.2 Warranty..................................................................................................................................... A-1
A.3 Repairs....................................................................................................................................... A-1
A.4 Environmental protection ........................................................................................................... A-1
A.5 Further devices .......................................................................................................................... A-1
A.6 Contact address.........................................................................................................................A-1
Index................................................................................................................................................ Index-1
SMD 18101 CS / SMD 18101 CP / SMD 18101 D Instruction Manual, 11/2005, A5E00472377A-002
v
Table of contents
SMD 18101 CS / SMD 18101 CP / SMD 18101 D
vi Instruction Manual, 11/2005, A5E00472377A-002

Introduction

1.1 Contents of this documentation

This document explains the functionality and the intended application of the SMD 18101 Gray Scale Flat Panel Display.
We also point out that the contents of this document are not part of a previous or existing agreement, commitment or statutory right and do not change the latter.

1.2 Further documentation

SMfit ACT® Basics
The Instruction Manual for the "SMfit ACT® Basics" software is provided on the CD supplied, along with the program itself.
1
SMD 18101 CS / SMD 18101 CP / SMD 18101 D Instruction Manual, 11/2005, A5E00472377A-002
1-1
Introduction
1.2 Further documentation
SMD 18101 CS / SMD 18101 CP / SMD 18101 D
1-2 Instruction Manual, 11/2005, A5E00472377A-002

Safety notes

2.1 Safety notes
Perfect, safe and reliable operation of the display assumes that it has been professionally transported, stored, mounted and installed as well as careful operator control and service. The unit must only be used for applications for which displays are normally used. The technical specifications of the product must be observed.
For the sake of safety, the following precautions must be observed:
Danger There is a danger to life if the warning information is not observed. Severe personal injury or
damage to property may occur. Do not open the unit yourself Certain components inside the unit are at high-voltage, i.e. touching these components
presents a danger to life! Only use a perfect line cord A damaged line cord may result in a fire or electric shock. When disconnecting the line cord,
always do so by holding the plug. Do not insert any objects into the housing Objects inserted into the housing (e.g. through the ventilation slots) may result in damage to
the unit or personal injury.
2
Do not place any objects on top of the units Penetrating liquids may result in a fire or electric shock. Connecting No contact to a patient must be possible when handling the connection cables.
SMD 18101 CS / SMD 18101 CP / SMD 18101 D Instruction Manual, 11/2005, A5E00472377A-002
2-1
Safety notes
Caution Incorrect installation may result in extensive damage to property.
Installation must be carried out by specialists:
1. When installing your medical electrical system with our products in an environment with patients, please observe the safety requirements of EN 60601-1-1 (IEC 601-1-1) "Specifications for the safety of medical electrical systems" in order to prevent injury to patients and users of your systems.
2. Use appropriate measures to ensure that the leakage currents in particular remain below the necessary limits. Appropriate measures are:
Separators for signal input or signal output unit Use of a safety isolating transformer Use of the additional protective conductor connection
3. Only use video cables with BNC connectors and interface cables specified by the system manufacturer for the installation.
4. The serial interface cable must have a female Sub-D connector at the computer end.
5. Use line cords with contactor contact. Only insert into sockets with contactor contact.
6. For certain applications, the video ground can be connected to the protective ground via the additional protective conductor in the plug panel. Observe EN 60 601-1-1 (IEC 601-1-
1).
7. Close the plug panel using the provided cover, and lock with the screw.
8. Mounting information: The stability of the display must be ensured following mounting of the stand/holder. The mounting screws must only have the defined length and the defined strength. All these requirements are satisfied when using the original stand. These requirements must be observed when using customer-specific mounting solutions. In addition, the country-specific regulations regarding steadiness must be observed.
9. Note for users: The closed connection panel must not be opened by users.
10. Servicing information: If housing components have to be removed for servicing, this must not be carried out in the presence of patients, users, or other persons not involved with servicing. Displays with a VESA mounting surface must only be connected to the power supply with the VESA board screwed on.
11. The following applies to installations in the USA and Canada: Molded power supply connectors must comply with the requirements for "hospital grade attachments" CSA Std. C22.2 No. 21 and UL 498.
SMD 18101 CS / SMD 18101 CP / SMD 18101 D
2-2 Instruction Manual, 11/2005, A5E00472377A-002
Safety notes
See also
Caution Failure to observe the warnings may result in substantial damage to property.
Do not subject unit to impermissible shocks Take care when transporting! Use the original packaging! The panel in particular should be protected against shocks.
Notice When touching the panel surface, the mechanical contact or an electrostatic discharge may
cause a brief disturbance in the picture quality.
Mounting of unit (Page 5-1) Cleaning (Page 9-1) Installation site (Page 4-1) General connection information (Page 6-1)
SMD 18101 CS / SMD 18101 CP / SMD 18101 D Instruction Manual, 11/2005, A5E00472377A-002
2-3
Safety notes
SMD 18101 CS / SMD 18101 CP / SMD 18101 D
2-4 Instruction Manual, 11/2005, A5E00472377A-002

Description

3.1 Scope of delivery

SMD 18101 CS/CP:
SMD 18101 D:
Note It is recommendable to keep the packaging material for subsequent transport of the display.
Display without stand
CD ROM with this Instruction Manual and SMfit ACT® Basics
SMfit ACT-cable
Display with stand
3
CD ROM with this Instruction Manual and SMfit ACT® Basics
European power cord
US power cord
DVI video cable
Video cable 15-contact Sub-D to 5x BNC
SMfit ACT-cable
SMD 18101 CS / SMD 18101 CP / SMD 18101 D Instruction Manual, 11/2005, A5E00472377A-002
3-1
Description
3.2 Applications

3.2 Applications

The gray scale flat panel display
SMD 18101 CS Order no.: 6GF6221-2CD13 SMD 18101 CP Order no.: 6GF6221-2CD19
SMD 18101 D Order no.: 6GF6221-2CD14 is approved for medical diagnostics according to IEC 60601-1. The display is available in the following versions: ceiling suspension (CS), ceiling suspension
with integrated protective glass (CP = CS + protective glass) and desktop (D). It complies with the safety standards UL 2601 and EN 60601-1.

3.3 Important features

The display has the following features, which permit a wide range of applications:
Compact design
Low weight and short overall depth are important advantages of the flat panel display compared to conventional CRT monitors.
Perfect picture reproduction thanks to LCD technology
Distortions in the image geometry are not present because of the use of LCD technology. The flat panel display provides a flicker-free picture even with low refresh rates (60 Hz). The
display thus fulfills even the strictest ergonomic requirements.
Modification of gray scale setting
In medical applications (e.g. for x-rays, scanners and nuclear spin tomography), where pictures are usually displayed in shades of gray, the gray values can be adapted to the user's eye using an on-screen display (OSD) menu. In the delivered state, several adjustment curves can be selected. Further curves can be set using the SMfit ACT® calibration software and a Siemens photometer (Siemens Serial Spot Meter or Siemens Universal Serial Luminance Meter).
Screen resolution
The display is equipped with an active 18" TFT display module, which provides a very wide viewing angle. The picture resolution is 1280 x 1024 pixels.
RGB input
The display can be connected to the computer system using BNC sockets. The display is adapted to the signal using an OSD menu.
SMD 18101 CS / SMD 18101 CP / SMD 18101 D
3-2 Instruction Manual, 11/2005, A5E00472377A-002
Description
3.3 Important features
DVI input
The display is connected to the computer system via the DVI-I input. Because of the digital signal transmission, no adaptation using the OSD is required.
Video inputs
The display has two additional analog video inputs and can thus be operated with standard analog video signals (PAL/NTSC). The RGB/DVI input and the two video inputs can also be simultaneously connected to signal sources.
Protective glass
The SMD 18101 CP has anti-glare protective glass fitted over the top of the LCD panel to protect the surface of the panel against bumps and scratches. The space between the protective glass and the panel is sealed against dust so that the internal surfaces cannot get dirty.
SMD 18101 CS / SMD 18101 CP / SMD 18101 D Instruction Manual, 11/2005, A5E00472377A-002
3-3
Description
3.3 Important features
SMD 18101 CS / SMD 18101 CP / SMD 18101 D
3-4 Instruction Manual, 11/2005, A5E00472377A-002

Application planning

4.1 Installation site

Provide adequate ventilation
Ventilation slots are located on the housing base, the rear panel and the side panels. The distance at the rear must be at least 10 cm from a wall or 15 cm from other units.
Observe the permissible ambient temperature range
The unit must not be operated outside the permissible ambient temperature range.
Installation on a desk
The unit must be screwed onto a stand and installed on a hard horizontal surface (unit stand for SMD 18101 CS/CP can be ordered separately). Only the stands recommended by the manufacturer may be used.
4
Avoid reflections on the screen
The display has an anti-glare surface which is only effective with a clean, grease-free screen. If the screen surface is dirty, clean it using a suitable microfiber cloth.
The display should be positioned so that reflections from lights, windows, furniture with shiny surfaces or light-colored walls do not appear on the screen.
In order to reduce mirroring on the unit, only use non-dazzling reflector lights for the ceiling lighting.
Change of environment
If the unit is brought into a warm environment from a cold one, condensation may form in the unit. The unit should not be switched on until all the condensed water has evaporated, including that inside the unit. This may take several hours, depending on the conditions.
See also
Safety notes (Page 2-1) Climatic conditions (Page 11-5) Cleaning (Page 9-1)
SMD 18101 CS / SMD 18101 CP / SMD 18101 D Instruction Manual, 11/2005, A5E00472377A-002
4-1
Application planning
4.1 Installation site
SMD 18101 CS / SMD 18101 CP / SMD 18101 D
4-2 Instruction Manual, 11/2005, A5E00472377A-002

Assembling

5.1 Mounting of unit

Ceiling or stand mounting
Several displays can be butt-mounted horizontally and vertically. The following must be observed when mounting (100 mm hole spacing according to VESA
standard):
Information about the fastening screws Number 4 Thread M4 Strength 8.8 Insertion depth Min. 8 mm/max. 10 mm Torque Max. 3 Nm
5
SMD 18101 CS / SMD 18101 CP / SMD 18101 D Instruction Manual, 11/2005, A5E00472377A-002
5-1
Assembling
5.1 Mounting of unit
SMD 18101 CS / SMD 18101 CP / SMD 18101 D
5-2 Instruction Manual, 11/2005, A5E00472377A-002

Connecting

6.1 General connection information

Caution All information and warnings for this product must be observed in order to guarantee
operation free of danger.
The unit is designed for a point-to-point connection to a graphics card.
Caution Observe shielding measures
All shielding precautions contained in the corresponding EMC guidelines must be observed. If these guidelines are not observed, interference signals could penetrate and disturb the display operation.
6
See also
Information on cable installation
Only shielded cables are permitted for all signal connections.
All cable connections must be screwed or clamped.
Signal and power cables must not be routed in the same duct.
The display must not share a power supply with motors or valves (glitches!).
Safety notes (Page 2-1) Electromagnetic compatibility (Page 11-6)
SMD 18101 CS / SMD 18101 CP / SMD 18101 D Instruction Manual, 11/2005, A5E00472377A-002
6-1