The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights, including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: SPR2-250.840.01.01.02 Doc. Gen. Date:09.05
Replaces: SPR2-250.898.01.07.02
0 - 2Revision
ChapterPageRevision
allall01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
ARCOSKOPSPR2-250.840.01Page 2 of 8Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
• 1 set of resolution tests (e.g. part no. 28 71 820)
• 1 set of radiation filters (e.g. part no. 97 98 596)
• Centering cross (e.g. part no. 96 60 051)
• Heat conducting paste WPS(e.g. part no. 20 48 650)
• Standard service equipment *
• 1 set of Allen keys *
• A 30 mm socket key (is required for adjusting the brake) *
• 200 N spring scale(e.g. part no. 44 15 113)
• Torque wrench 20 Nm to 100 Nm (e.g. part no. 80 86 159)
• Torque wrench 3 Nm to 20 Nm(e.g. part no. 34 24 553)
All tools, measurement and auxiliary devices with the exception
of those marked "*", are listed along with their specifications in
the STC (Service Tools Catalogue).
part no. 99 00 861)
part no. 51 38 727)
Siemens AGSPR2-250.840.01Seite 1 von 16ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
1 - 2Prerequisites
back
left
Fig. 1
left
front
right
right
front
back
Information regarding this document1
These service instructions will assist you in adjusting the ARCOSKOP for routine
operation. Not every adjustment is required each time the system is serviced.
Position designations1
The designations for the monitor trolley and the ARCOSKOP used in these instructions,
i.e. left, right, front and back, are explained in Fig.1.
ARCOSKOPSPR2-250.840.01Seite 2 von 16Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Prerequisites1 - 3
Safety Information1
WAR NING
WAR NING
Danger of physical injury up to death and damage to property!
You must comply with
- the product-specific safety notes in these instructions,
- the general safety notes in the instructions
TD00-000.860.01... and
- the general safety instructions according to ARTD Part 2.
In the case of non-compliance death or physical injury or damage
to property can arise.
Electrical voltage!
In the case of non-compliance death or physical injury or damage
to property can arise.
Parts conducting voltage can be touched after covers of the
system have been removed. To avoid this danger disconnect the
system from the power supply before removing covers.
For this purpose pull out the mains plug or switch the line voltage
connection of the system free of voltage and secure it against
being switched back on. In the case of necessary service work in
connection with electrical voltage, observe the general safety
notes in the instructions TD00-000.860.01...
WAR NING
WAR NING
Electrical voltage!
In the case of non-compliance death or physical injury or damage
to property can arise.
After service work on the primary circuit of the power-up module
(e.g. replacement of the power-up module or parts of it) measure
and record the equivalent system leakage current (see system folder or logbook).
Danger of infection with pathogens!
In the case of non-compliance death or physical injury can arise.
This product is released for operation in operating rooms and can
be contaminated with infectious blood or other body secretions.
Avoid all contact with blood or other body secretions!
Comply strictly with the preventive measures against infectious
diseases specified in ARTD-002.731.37... !
Siemens AGSPR2-250.840.01Seite 3 von 16ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
1 - 4Prerequisites
WAR NING
WAR NING
CAUTION
Electrical voltage!
In the case of non-compliance death or physical injury or damage
to property can arise.
After completion of all service work and after attaching all covers
of the system perform a protective ground wire test according to
ARTD-002.731.17... . The protective ground wire resistance must
not exceed 0.2 ohms.
X-ray radiation!
In the case of non-compliance death or physical injury can arise.
The radiation protection regulations must be complied with
during service work in which X-ray radiation must be released.
Keep the radiation activity as low as possible (low kV, mA, short
radiation time).
Keep your distance from the radiation source as large as possible. Radiation protection devices (lead partitions, etc.) and radiation protection clothing (lead apron) must be used.
Danger of burns on hot modules or components! In the case of
non-compliance slight to medium burns, primarily on hands, can
occur.
CAUTION
WAR NING
After covers are removed modules or components (e.g. power
modules, heat sinks, magnetic brakes) that can have temperatures > 50° Celsius in operation can be touched. To avoid burns
switch the system free of voltage and wait for a cooling down time
of at least 5 minutes before touching modules or components.
Electrostatic voltage!
There can be damage to property in the case of non-compliance.
Comply with the ESD protection regulations in service work.
Electrical voltage!
In the case of non-compliance death or physical injury or damage
to property can arise.
The Elko batteries of the generator and the Elkos on the generator
board remain charged for a long time even after the system is
switched off. The Elkos must be discharged before working on the
Elko batteries and before working on the generator. Switch the
switch D21.S2 into the UZ_OFF position and wait until the LEDs
D21.X22 and D21.X23 no longer light up. Then check the voltage
between the test points X109.UZ_IST and X109.ANA_GND.
The voltage must be < 0.2 V (corresponds to Uz < 20 V).
ARCOSKOPSPR2-250.840.01Seite 4 von 16Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Prerequisites1 - 5
CAUTION
CAUTION
Danger of injury when dismantling / assembling mechanical
parts! Non-compliance can lead to minor to medium severe injuries, especially to the hands!
If parts that are under mechanical stress have to be dismantled or
assembled, as for example the cover over the horizontal carriage
of the basic unit, inattention can cause injuries due to crushing,
cutting or grazing the skin, especially to the hands.
Perform the corresponding work with special care and attentiveness.
Wear working gloves if necessary.
Danger of injury on mechanical parts!
Non-compliance can lead to minor up to moderately severe injuries, especially to the hands!
With covers removed, you can come into contact with parts such
as flat plugs, threaded bolts, cut off cable ties or edges of components on which, in the case of inattention, injuries can arise
due to crushing, cutting or grazing the skin, especially to the
hands.
Perform the corresponding work with special care and attentiveness.
Siemens AGSPR2-250.840.01Seite 5 von 16ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
1 - 6Prerequisites
Information on the equivalent leakage current measurement1
Regulations and scope for the subsidiaries1
Within the scope of DIN VDE0751 part 1, the equivalent leakage current test must be performed.
Outside the scope of DIN VDE0751, the subsidiaries must observe the following:
(see ARTD-002.731.731.17, Safety regulations for installation and maintenance).
The legal national regulations apply to the subsidiaries. In the event that there are no existing regulations, observe the following rules in the interest of safety for customers, patients, employees and third parties, as well as the company.
Initial measurement1
The measurement was performed at the factory and the value measured was recorded in
test protocol 1b. The measurement was performed with the line voltage and line frequency
specified in the test protocol. During start-up of the ARCOSKOP on-site, the values were
checked and remeasured and recorded if they deviated from the line voltage. The protocol
is filed in Register 9 of the System Manual and System Owner Manual,
Register 6.
Repeat measurements1
When performing service or repair work in the primary circuit of the power supply, the
equivalent leakage current must be measured again and compared to the initial value
measured. The measurement setup must match the measurement setup for the equivalent leakage current / protocol. During measurement, the ARCOSKOP must be switched
on.
When using the Bender safety tester, the tester must be set to manual measurement.
The values measured during repeat tests must not exceed the original value by more than
50%. In addition, the limit of 2 mA must not be exceeded. ARCOSKOP systems which
exceed this threshold must be serviced. The values measured are to be recorded in the
equivalent leakage current / protocol (see System Manual, Register 9).
Hand over a copy of the protocol to the customer for filing in the System Owner Manual.
ARCOSKOPSPR2-250.840.01Seite 6 von 16Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Prerequisites1 - 7
Fig. 2Fig. 3
Fig. 4Fig. 5
Covers1
WAR NING
NOTE
Installing/removing the covers on the support1
• To loosen the fastening screws for the end caps on the support arms, the rubber at the
ends must be pulled out from the slot, refer to Fig. 2 to Fig. 5. After removing the Phillips
screws, press the internally located retainer a little to the inside with a thin screwdriver
and simultaneously pull the end cap out a little. After the end cap has been loosened on
both sides it can be removed.
• To remove the two-part cover of the support on the false ceiling, remove the two white
rubber plugs and unscrew the Allen screws, refer to the figure at the right above. The two
parts must be pulled apart. Take care in this case that the two internally located retaining
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the covers, switch off the line voltage supply to
the system and secure against accidental switch-on.
Switch off the compressed air brakes, turn off the compressed air
and let it escape.
Siemens AGSPR2-250.840.01Seite 7 von 16ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
1 - 8Prerequisites
rods remain in one half of the two covers in order to keep the space required for pulling
the halves apart small.
Installing/removing the covers on the C-arm mounting1
• Remove the cover screws and remove both covers towards the side (Fig. 5).
• After completing the procedure, reinstall both covers in the reverse order.
ARCOSKOPSPR2-250.840.01Seite 8 von 16Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Prerequisites1 - 9
Cover over the horizontal carriage1
Fig. 6Fig. 7
Removing the cover1
NOTE
Fig. 6 and Fig. 7 are examples and show the cover of the SIREMOBIL Iso-C.
The cover of the ARCOSKOP horizontal carriage is open to the
front in the area of the verical lifting column and has no emergency stop button and only 3 brake buttons.
However, the cover of the ARCOSKOP can be removed and attached in practically the same way.
• Loosen the 6 mounting screws (1/Fig. 6).
CAUTION
Danger of injury when dismantling / assembling mechanical
parts! Non-compliance can lead to minor up to moderately severe
injuries, especially to the hands!
The cover over the horizontal carriage is bent U-shaped and must
be pulled on the side to the outside when it is removed.
The cover is then under mechanical stress.
We recommend that you wear light working gloves to avoid injuries.
• Expand the cover at the rear end in the direction of the arrow (see Fig. 2) and remove it
downwards.
• If needed disconnect the protective ground wire, the cable of the keypad (brake buttons)
and the cable of the emergency stop button.
Siemens AGSPR2-250.840.01Seite 9 von 16ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
1 - 10Prerequisites
Attaching the cover1
• Re-fasten or connect the protective ground wire, the cable of the keypad (brake buttons)
and the cable of the emergency stop button.
CAUTION
Danger of injury when dismantling / assembling mechanical
parts! Non-compliance can lead to minor up to moderately severe
injuries, especially to the hands!
When the cover is attached and pressed down, it is under mechanical stress and snaps into place over the bottom edge of the
horizontal carriage.
Expand the cover at the rear end (see Fig. 6) so far that as shown
in Fig. 7 it lies obliquely over the horizontal carriage in the edge 1/
Fig. 7.
Press on the surface on the top of the cover, without grasping the
sheet metal edges, until this slides over the lower edge of the
horizontal slide.
• Expand the cover at the rear end (see Fig. 6) so far that as shown in Fig. 7 it lies obliquely
over the horizontal carriage in the edge 1/Fig. 7.
• Press on the top surface of the cover, without grasping the sheet metal edges, until this
slides over the lower edge of the horizontal slide. Here make sure that the cover does not
slide out from the edge (1 / Fig. 7).
• Check the operation of the emergency stop button and of the brake buttons.
• Fasten the cover with the 6 Allen screws (1/ Fig. 6).
ARCOSKOPSPR2-250.840.01Seite 10 von 16Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Prerequisites1 - 11
Fig. 8
Installing/removing the cover at the C-arm/I.I. side1
CAUTION
Danger of injury!
In the case of non-compliance death or physical injury or damage
to property can arise.
The C-arm cover contains counterbalance weights and weighs
> 10 kg.
Hold the cover firmly when installing or removing it.
Before the C-arm cover is loosened, the C-arm must be moved to
its middle orbital position.
If the C-arm covers are loosened, the C-arm must not be moved to
its orbital end positions (I.I. or POWERPHOS in the orbital end
position).
• Remove the 4 fastening screws (Fig. 8). Hold the covers firmly while removing them.
• Install them in the reverse order.
Siemens AGSPR2-250.840.01Seite 11 von 16ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
1 - 12Prerequisites
Monitor trolley1
Removing the covers
• Remove the upper back cover plate of the log book compartment.
• Remove the protective ground wire.
• Remove the lateral Allen screws from the lower back cover.
• Remove the protective ground wire from the cover.
Installing the covers
• Begin with the lower back cover.
• Reattach the protective ground wire for the lower cover.
• Install the cover and tighten the screws.
• Reattach the protective ground wire for the log book compartment.
• Install the cover and tighten the screws.
Installing/ removing the covers on the monitor support system1
Fig. 9 Monitor support system cover 1-2
• Remove the screw fastening of the cover halves on the upper joint of the cover,
see arrow Fig. 9.
• Loosen the remaining cover screws and remove the two cover halves to the side.
• After taking the necessary action install the covers again in the reverse order.
ARCOSKOPSPR2-250.840.01Seite 12 von 16Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Prerequisites1 - 13
Service position for the board access of the electronics box1
Fig. 10
• The cover latch must be closed before loosening the screws (see Fig. 10).
• Remove the screws from the sides of the cover (see Fig. 10).
• Be aware of the weight added by the components attached to the rear side when opening
the latch and the service cover (see 2/Fig. 11).
Fig. 11
• You can now lower the generator unit. Use an extension to loosen the two screws on the
upper sides (see 3/Fig. 11).
Siemens AGSPR2-250.840.01Seite 13 von 16ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
1 - 14Prerequisites
Fig. 12
• Cables D20, X7, X8, and X9 should be fastened to the cable clamps with a torque of
4.8 Nm. Loop the cables and attach the shield as shown in Fig. 12. Ensure that the
cables do not become crimped or torn when moving the generator.
• When routing cables in the E-box, ensure that the cover can be opened and closed
without crimping or stretching the cables or loosening the plug connection.
ARCOSKOPSPR2-250.840.01Seite 14 von 16Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Prerequisites1 - 15
Fig. 13Fig. 14
Fig. 15
I.I. service position1
Moving the I.I. into the service position1
• Take the threaded rods, spacers and nuts out of the accessory pack ("Service Pack I.I./
Camera" part no. 71 43 519 / Fig. 15).
• Move the C-arm into the vertical position, with the I.I. below and the POWERPHOS
above.
• Remove both attachment screws for the I.I. and replace them with the threaded rods
from the service pack.
• Insert the threaded rods from the service pack at least 5 turns into the threaded holes of
the I.I.
• Then mount the spacers and nuts (see Fig. 14).
• Tighten the nuts to fasten the I.I. again.
Siemens AGSPR2-250.840.01Seite 15 von 16ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
1 - 16Prerequisites
• Turn the C-arm so that the I.I. is above and the POWERPHOS is below (angulation).
• Lower the I.I. by gradually loosening the nuts alternately. Be careful not to damage the
cables. Fig. 13 shows the I.I. in the service position.
Installing the I.I. 1
• Lift the I. I. a little alternately on each side and retighten the nuts.
Repeat this until the I.I. is fastened again. Be careful not to damage the cable.
• Tighten the nuts to fasten the I.I. again.
• Move the C-arm so that the POWERPHOS is above and the I.I. is below (angulation).
• Remove the two threaded rods (service pack) again from the threaded holes of the I.I.
and reinstall the I.I. with the two attachment screws.
ARCOSKOPSPR2-250.840.01Seite 16 von 16Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Power supply adaptation2
2 - 1
WAR NING
Electrical voltage!
See Chapter 1, Safety Information.
Disconnect the ARCOSKOP power plug.
• Open the back cover of the monitor trolley.
• Measure the on-site line voltage and line frequency.
• Adjust transformer T1 to the required line voltage and line frequency. Refer to the label
on the switch-on assembly.
• Adjust transformer T2 to the required line voltage and line frequency. Refer to the label
on the switch-on assembly.
• Check fuses F1 and F2 to see as to whether they correspond with the values on the label
of the switch-on assembly and replace them, if necessary.
• Mark the correct line voltage and line frequency on the line voltage label at the back of
the monitor trolley.
Checking the line voltages on the monitor trolley2
• Plug X10 for the basic unit must be disconnected.
• The monitor, memory device and other devices, if present, such as multiformat camera,
video printer or video recorder must be disconnected from the line voltage supply.
• Check the secondary voltages for line voltage transformer T1 according to table 1.
From test pointto test pointVoltageComment
Transformer T1.31Transformer T1.32190 V ~ to
205 V ~
Transformer T1.31Transformer T1.33230 V ~ to
246 V ~
Power cable
Memoskop, N
Power cable
Monitor 1, N
Option: Power cable
Monitor 2, N
Option: Power cable
Multispot, N
Option: Power cable
Video printer, N
Option: Power cable
Video recorder, N
Option
Power cable optical
video separation, N
Power cable
Memoskop, L
Power cable
Monitor 1, L
Power cable
Monitor 2, L
Power cable
Multispot, L
Power cable
Video printer, L
Power cable
Video recorder, L
Power cable optical
video separation, L
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
if Video printer present, not
not adjustable
not adjustable
not adjustable
not adjustable
not adjustable
if Multispot present, not
adjustable
adjustable
if Video recorder present,
not adjustable
if optical video separation
present, not adjustable
Tab. 1
Siemens AGSPR2-250.840.01Seite 1 von 2ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
2 - 2Power supply adaptation
Diese Seite wurde bewusst leer gelassen.
ARCOSKOPSPR2-250.840.01Seite 2 von 2Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Checking the operating voltages3
Fig. 1Fig. 2
3 - 1
Fig. 3Fig. 4
Siemens AGSPR2-250.840.01Seite 1 von 4ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
3 - 2Checking the operating voltages
Low voltages3
• Check the operating voltages of the ARCOSKOP according to tables 1 and 2 (Fig.1 - 4).
From test pointTo test pointVoltageAdjuster
D1.X37 (0V)D1.X30 (+5 V)+5.1 V to 5.2 VM14. +5 V/Adj.
D1.X39 (0V)D1.X32 (+15 V)+4.9 V to 15.1 VM14. +15 V/Adj.
D1.X39 (0V)D1.X33 (-15 V)- 4.8 V to -15.2 VM14. -15 V/Adj.
D30.X1.8 (0V)D30.X1.9 (+24 V)+24.0 V to 29.6 Vnot adjustable
D30.X21.1 (0V)D30.X21.10 (+24 V)+22.8 V to +29.5 Vnot adjustable
D30.X20.1 (0V)D30.X20.3 (+27 V)+27.3 V to +27.8 VD30 R115
D42.MP1 (0V)D42.MP2 (+5 V)+ 5,0 V bis +5,2 VD40.P1
D40.D17.A (0V)TP (+5 V)+5,0 V to +5,2 VD40.P1 *1
M13.S- (0V)M13.S+ (+13 V)+13.1 V to 13.3 VM13.TR1
Ta b . 1
*1
With existing "operation free" option, i.e. the control console is detachable, you must measure and check on board D40 that the voltage does not exceed 5.2 V. Should the voltage
on D40 be too high, the voltage on D42 can be reduced to 5.0 V.
Image intensifier voltages3
• Take the voltages E1 / E2 / E3 and A from the test protocol of the image intensifier and
check or adjust them at the test points of the control unit listed in Table 2.
VoltageTest pointGround
point (0V)
E1UE1P10P111:1
E2UE2P6P71:1
E3UI 15P2P31:10000
30 kV, anodeUI 30P1P11:10000
Ta b . 2
Potentiome-
ter for
full format
Potentiome-
ter for
zoom format
Volt ag e d ivi -
der ratios
ARCOSKOPSPR2-250.840.01Seite 2 von 4Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Checking the operating voltages3 - 3
D41/D42 Programming3
• Make sure that the jumpers on boards D41 and D42 are installed as follows:
BoardJumperPinsonoffComponents
D41211 - 2XE-Box
D41221 - 2XE-Box
D41232 - 3XE-Box
D4124XE-Box
D4125XE-Box
D4126XE-Box
D42212 - 3XPanel
D42222 - 3XPanel
D42231 - 2XPanel
D4224XPanel
D4225XPanel
D4226XPanel
Tab. 3
• Board D41 for the E-box and board D42 for the control panel have to be clearly identified
with an X in the respective field.
Siemens AGSPR2-250.840.01Seite 3 von 4ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
3 - 4Checking the operating voltages
Diese Seite wurde bewusst leer gelassen.
ARCOSKOPSPR2-250.840.01Seite 4 von 4Siemens AG
System ManualRev. 0109.05CS PS 24Medical Solutions
Operating software4
Loading the operating software4
• Connect the service PC to the serial service interface on ARCOSKOP.
• Insert the "Parameter" disk in drive A.
• Start the service program.
• Select the Download menu, select the relevant file (Host, Gen....) and start Download.
• After successfully downloading, wait for the system to boot (1 minute).
• The ARCOSKOP is operational.
4 - 1
NOTE
If you cannot run Download (e.g. power failure during Download):
- Switch the ARCOSKOP off and back on
- Restart Download
Backing up the system parameters4
NOTE
Prerequisite:
The service PC is already connected, the service program is already started.
• Select the "Data" menu.
• Select the "Backup" submenu.
• Click the "Parameter" check box.
• If required select the back-up of further data by selecting the corresponding check box.
After changing settings / parameters, back these up on the parameter disk.
Note that the data are stored correctly only after ending the service program
Therefore always exit the service program with "Logoff" and
"Quit".
• Click the "Backup" button. Wait until the data are backed up temporarily on disk.
• Then back to the main menu.
• Select the "System" menu.
• Select the "Logoff" submenu. Wait until the data are stored on disk.
• You can now exit the service program by selecting the "System" and "Quit" menus.
Siemens AGSPR2-250.840.01Seite 1 von 4ARCOSKOP
Medical SolutionsRev. 0109.05CS PS 24System Manual
4 - 2Operating software
Reading and storing the load counter data4
NOTE
The procedure is divided in 4 steps:
Before the replacement of the POWERPHOS, all load counter
values have to be stored on disk. After storing the load counter
values, the load counter has to be reset. All values are set to 0.
After the replacement and all adjustments, send back the disk
with the defective POWERPHOS for repair.
• Store the actual load counter values on disk.
• Release the resetting of the load counters.
• Reset the load counter values to 0.
• Return the defective POWERPHOS, including the load counter data disk.
Storing the actual load counter values4
• Connect the service PC to the serial interface of the unit.
• Start the service program.
• Select the "Diagnostics / Monitoring" menu.
• Select "LOAD COUNTERS" in the "Group 1:" combo box.
• Click on "Update Start" and wait until the actual values are displayed.
• Click on "Update Stop".
• Click on the "Save" button.
• Change the data disk of the service program in drive A against a formatted empty one.
• Select the drive A for storage.
• For the file name, enter the current date, the part no. and the serial no. of the
POWERPHOS.
Example:
The current date is:
Use the date format DD-MM-YY:
The POWERPHOS part no. is:
The POWERPHOS serial no. is:
Enter the file name: