Page 2
0 - 2Revision
ChapterPageRevision
allall02
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Opdima SPB7-230.814.03Page 2 of 4Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
BACK COVER CPL
LOWER BACK COVER
UPPER BACK COVER
Holding Plate left
Holding Plate right
CABLE SOCKET CPL
Grip
LOCTITE_243
Lever axle
Lever Bushing
Spring 1x9x30 DIN 2098
EMC Filter
Screws M4x12 DIN 912-12.9 0000
Washer 4 SN70093-FST
Washer A8.4 DIN 125-ST 6980
Transformer 215 VA RONDOLL
Screw MC6S6x65 DIN 912
Washer A6.4 DIN 125-ST 6980
Nut M6NY6 DIN 985
PROM kit Master VA00D
Software kit V2.1
1.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
2.000
2.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
Siemens AGSPB7-230.814.03Page 1 of 2Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
CABLE BRACKET
Cable tie SN70764
CABLE Z801
Cable socket 6.3 - 2.5 DIN 46247 - BRONZE
Heat shrink tubing 4.8 SW
Label 8 x 17 Paper yellow
PROM KIT MASTER PLD VA00D
1.000
3.000
1.000
0.070
7.000
1.000
Phantom / Auxiliary1
ItemRemarksMaterial NumberWith System
RMI 15688 81 265no
One 0.1 mm, two 0.2 mm and
one 0.5 mm sheets of aluminium
99 % (1100) alloy, 4 x 4 cm
Mammography line pair phantom2 - 10 Ip/mm88 81 299no
4.2 cm plexiSize of the
detector
Centering cross96 60 051no
88 81 273no
74 47 720no
Opdima SPB7-230.814.03Page 2 of 2Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 7
Important Notes2
Symbols2
This symbol indicates exposure of X-ray radiation.
"This symbol indicates that the measured values have to be entered.
Protective measures2
System Power Supply2
It is very important that any intervention in the equipment will start with disconnecting it
from the power supply with the main circuit-breaker. Before removing or inserting any of
the printed circuit boards, switch off the equipment.
2 - 1
DR
DR
system is only switched off at the
.
WARNING
WARNING
If the MAMMOMAT Novation
control panel or with the circuit breaker CB S700, there is still
voltage present within the generator. See the wiring diagram for
MAMMOMAT Novation
Life-threatening electric shock hazard exists.
Disconnect mains cable and comply with the information on this
page.
After shut-down of the system, there may still be 380 V DC
present on the intermediate circuit of the MAMMOMAT generator.
Life-threatening electric shock hazard exists.
A bleader resistor is connected over the output DC capacitors.
The DC voltage will discharge below 60 Volt in 3 minutes.
The voltage level is indicated by LEDs H3 and H4.
PC Boards2
CAUTION
The p.c. boards contain electrostatic highly sensitive components.
If disregarded, the components could be damaged.
Use ESD equipment, ground prior to making contact and place
the components on a conductive surface.
Siemens AGSPB7-230.814.03Page 1 of 4Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Page 8
2 - 2Important Notes
Metal Curtain2
WARNING
The edges of the metal curtain of the stand are very sharp.
They may cause severe injury.
These strips must always be applied onto the edges of the metal
curtain during service and maintenance work.
Opdima SPB7-230.814.03Page 2 of 4Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 9
Important Notes2 - 3
Apply the protective strips after removing the covers from the stand. Remove the protective strips only when the covers are to be mounted or when vertical adjustment of the
swivel arm system is necessary.
1
2
Fig. 1 Protective strips for the metal curtain
Holders (1/Fig.1 and 2/Fig. 1) for storing the protective strips when not in use are provided
on both sides of the curtain.
Measurements with the oscilloscope2
WARNING
The existing ground conductor in the mains cable must under no
circumstances be disconnected when operating the oscilloscope.
A life-threatening hazard of electric shock exists.
For measurements in which a resulting ground loop may falsify
the measuring result, use the differential amplifier (difference
measurement).
Swivel-arm rotation 2
WARNING
When rotating the swivel arm with the detector wing in parking
position, make sure that the main cable harness does not get
damaged. Stop rotation immediately, if the cable harness appears
to be extensively pulled.
Service Software2
The MAMMOMAT
Novation
• Acquisition Workstation
The Acquisition Workstation service software is web-based and can be started from the
Options application menu.
• BRICK
The BRICK service software can be accessed via the Acquisition Workstation.
Siemens AGSPB7-230.814.03Page 3 of 4Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
DR
has various service software packages:
Page 10
2 - 4Important Notes
Delay times between two exposures2
The delay times listed below must be followed to prevent the tube from overheating.
Exposure mAs valueDelay time between two exposures (seconds)
Opdima SPB7-230.814.03Page 4 of 4Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 11
Settings3
Presettings3
3 - 1
NOTE
• Switch ON the MAMMOMAT Novation
All the orientations left and right is always seen from the front of
the MAMMOMAT System.
DR
System.
• Connect the Service PC to the stand.
• With the service software revision level V1.0 available at site make a backup of all
components. (i.e. of the stand and of the generator. The generator data are stored
within the control panel.)
• You may either "copy installation area to disc", or "copy installation area to floppy".
• Label the floppy with "backup with SSW V1.0 / current date".
• If the copy was made to the (hard) disc of the service PC, note down the location where
to the data were stored. (This backup may only be needed if some irregularity would
occur.)
• Switch OFF the MAMMOMAT System and switch OFF the main power line.
Siemens AGSPB7-230.814.03Page 1 of 2Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Page 12
3 - 2Settings
This page intentionally left blank.
Opdima SPB7-230.814.03Page 2 of 2Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 13
Worksteps procedures4
Covers4
• Remove the back cover.
• Remove the back cover service.
• Left front cover.
4 - 1
NOTE
• Remove lower and back cover from the detector wing.
• Remove the tube head cover.
• Remove the covers below the bellow.
Attach metal curtain protection.
Lift the generator4
NOTE
• Loosen the four screws (One turn counter clockwise Fig. 2) fixing the generator to the
stand.
• Remove the bottom hinge with the two screws.
NOTE
• Place the washer 8.4 x16 x1.6 (Pos 140) on the lower hinge axle according to (1/Fig. 1).
• Lift the generator door up and mount the hinge back to its original position.
• Remove the four screws and swing the generator out.
The generator has to be lifted to make room for the EMC filter.
The generator will slide down a little.
Siemens AGSPB7-230.814.03Page 1 of 16Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Page 14
4 - 2Worksteps procedures
Screws for
securing the
generator
1
Fig. 1Fig. 2
Transformer4
• Drill a hole (φ 7 mm) for mounting of the transformer according to Fig. 3 (spring site).
See to it that no remnants of the drilling are left inside the stand. For marking the place of
the hole a drill template (No: 1) may be used.
• Mount the transformer (Pos 150) inside the stand with a MC6S M6x65 (Pos 160), one
washer 6.4 x 12 (Pos 170) and a nut M6M (Pos 180) according to Fig. 3 and Fig. 4.
The cable connections pointing downwards. Fasten the label T803 (use one off the
yellow labels and write T803 on it) and stick it to the inside off the stand above the
transformer.
OpdimaSPB7-230.814.03Page 2 of 16Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 15
Worksteps procedures4 - 3
70
123
Fig. 3Fig. 4
Cable Socket4
• Dismount the foot pedals strain relieves, the mains cable strain relief and the mains
supply connection from the Cable Socket CPL.
• Remove the old Cable Socket CPL.
• Mount the left and right foot pedal strain relieves on the inner positions on the new Cable
Socket CPL (Pos 50) according to (1/Fig. 5)
• Install the new Cable Socket CPL.
1
Fig. 5
Siemens AGSPB7-230.814.03Page 3 of 16Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Page 16
4 - 4Worksteps procedures
EMC filter (Z801)4
• Mount the EMC filter (Pos. 110) upside down with two MC6S 4 x 12 screws (Pos. 120)
and two contact washers (Pos. 130), and fasten the labels according to Fig 6.
• Label 1 to 4, Z801, Z801-1, Z801-2 and Z801 GND.
Fig. 6
Cable Z8014
• Dismount the D-sub plate.
• Thread the Z801 cable (Pos 200) through the conduit entry according to Fig. 7.
• Restore the D-sub plate. Attach the AMP connector to the transformer inside the stand.
Fig. 7
• Attach the connectors to the EMC filter and to the connector X803 according to markings
on the connectors, according to Fig. 8.
NOTE
OpdimaSPB7-230.814.03Page 4 of 16Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Use the numbering of terminal contacts on the lower side of terminal
row Z 803.
Page 17
Worksteps procedures4 - 5
If you don't connect an Opdimar mains cable isolate the outputs 1 and 2 of the filter Z 801
by means of the supplied cable socket and heat shrink tubing in order to prevent electrocution.
Fig. 8
Cable Bracket4
• Remove the Upper Back Cover and Lower Back Cover from the detector.
• Remove the detector (Info in Replacement of parts SPB7-250.841.01.01..).
• Remove the screws holding the bellow. Remove the strain relief from underneath the
Flying Wing (detector side). Lift the bellow and replace the old Cable Bracket with the
new one (POS 190) according to Fig. 9.
Fig. 9
Siemens AGSPB7-230.814.03Page 5 of 16Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Page 18
4 - 6Worksteps procedures
• Mount two cable ties (Pos 195) according to Fig. 9 and one cable tie according to Fig. 10.
Fig. 10
• Replace the strain relief underneath the Flying Wing on the detector side according to
(1/Fig. 11).
1
Fig. 11
• Mount the detector base blade.
OpdimaSPB7-230.814.03Page 6 of 16Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 19
Worksteps procedures4 - 7
Flying Wing lever4
• Dismount the Clamp Rings (2 spare in the upgrade kit), the Pin and the Torsion Spring
according to (1/Fig. 12) and take out the Catch.
1
Fig. 12
Fig. 13
• Assemble the Lever (Pos 60), Bushing (Pos 90), Axle (Pos 80), Catch and Spring
(Pos 70) according to Fig. 13 and Fig. 14 (apply Loctite 243 (Pos 70) on the Axle threads
before turning it in to the lever handle.
Fig. 14
Siemens AGSPB7-230.814.03Page 7 of 16Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Page 20
4 - 8Worksteps procedures
• Mount the Flying Wing Lever together with the Pin, Torsion Spring and Clamp Rings
according to Fig. 15.
Fig. 15
• Reinstalling the Detector
• Remove the existing rails left and right from the Detector wing.
• Remove all the assemblies from the existing rails and reinstall it by the new one
(1/Fig. 16).
• Install the new Spring plate according to the (2/Fig. 16).
1
2
2
1
Fig. 16
OpdimaSPB7-230.814.03Page 8 of 16Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 21
Worksteps procedures4 - 9
• Reinstall, i.e. attach both rails back to the Detector wing and reconnect the cables.
• Move the Cover Bracket and the Service Cover from the old Lower Back Cover to the
new Lower Back Cover (Pos 20) according to Fig. 17.
Fig. 17
• Mount the Upper Back Cover and the Lower Back Cover. Notice how the cables run
through the new opening on the Lower Back Cover according to Fig 18.
Fig. 18
Siemens AGSPB7-230.814.03Page 9 of 16Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Page 22
4 - 10Worksteps procedures
Software Update V1.0 to 2.14
OPDIMA can only be connected to NovationDR with the FW/SSW revision level V2.1 or
higher !!
The actual installation of OPDIMA to a NovationDR can follow only when the NovationDR
was already adapted, or if the NovationDR has had a serial number higher than 1100 !
The adaptation of NovationDR includes the Upgrade of the Firmware/ServiceSoftware
from revision level V1.0 to revision level V2.1 !
Thereby following has to be exchanged:
On the PC Board D750:
• exchange the IC 62 (MatNo.: 066 59 770), see 3 / Fig. 19.
• exchange the IC 45 (MatNo.: 066 59 796), see 1 / Fig. 19.
• exchange the IC 53 (MatNo.: 066 59 747), see 4 / Fig. 19.
• The IC 52 (MatNo.: 066 59 440), see 2 / Fig. 19 - will remain, as it is.
Fig. 19 PCB D750
OpdimaSPB7-230.814.03Page 10 of 16Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 23
Worksteps procedures4 - 11
On the PC Board D801:
• Exchange the IC 9 (MatNo.: 066 59 788).
On the PC Board D740:
• Exchange the IC 10 (MatNo.: 066 59 804), see Fig. 20.
Fig. 20 PCB D740
• Service PC
• Install the Service Software V2.1, delivered on a 3,5 inch floppy under the
MatNo.. 066 59 812, on the service PC.
NOTE
For further service use only the delivered service software V2.1 .
Parameter settings4
• Switch ON the main power line.
• Switch ON the MAMMOMAT Novation
acquisition workstation doesn't need to be switched ON, yet.
• The control console may display the ERROR 821, meaning: "The stand EEPROM has
the wrong version number (old or corrupt)". Disregard the message and acknowledge/
purge the same on the control console in the usual manner.
• The service PC is still connected to the stand; reconnect if disconnected.
Siemens AGSPB7-230.814.03Page 11 of 16Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
DR
System, during these work steps the
Page 24
4 - 12Worksteps procedures
• Perform "Backup" of both, the stand and the panel (control console) /generator, i.e.
"copy installation area to disc" or "copy installation area to floppy". Note down where to
on the hard disc this backup data were stored, or attach corresponding label to the
floppy.
• Switch OFF the MAMMOMAT Novation DR system from the control console, wait for half
a minute.
• Switch the system ON again, the ERR821 shouldn't appear this time.
Checks, adjustments and calibrations4
NOTE
In the last few working steps the firmware / software V 1.0 of the
MAMMOMAT NovationDR system was changed to the revision
level V 2.1. All parameters, which have been previously configured, entered and stored on pertaining places ("installation
areas") by the FW/SW V1.0 are present within the system in an
unchanged format. EEPROMs, containing such settings, aren't
supposed to be replaced in course of this upgrade. Since, however, in course of some work steps the detector has been
removed and the connection of the OPDIMA option was made
available, following activities have to be performed additionally:
1) Rotation test/check.
2) Calibration/check of the optimized compression for OPDIMA;
(or, setting of that parameter to the same value as for the DR).
3) Calibration of the collimator for the radiation beam and the light
field.
4) Calibration of the wing difference; (or , setting of that parameter
to the same value as for Mo large focus).
5) Disabling film exposure, i.e. the setting of the "Allow film exposure" software switch to OFF.
6) OPDOSE - Configuration of "Auto limits" for both, DR (Digital
Radiography) and for the OPDIMA.
7) Check of the grid lines.
8) Calibration of the detector gain.
Rotation Test4
• Move the wing in different positions and watch to the cable.
• The rotation of the swivel arm has to stop on position 0°, + 45°, + 90° and + 120°, when
OPDIMAr wing is selected.
• The rotation of the swivel arm has to stop on position 0°, + 45°, + 90° and + 135°/ - 180°,
when Detector wing is selected.
• Installing the wing cover and the bellow.
OpdimaSPB7-230.814.03Page 12 of 16Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 25
Worksteps procedures4 - 13
Checking the OPCOMP function when Opdima wing selected4
• Decompress.
• Set the compression force to 20 kg by using the presetting knob and compress.
The compression movement must be interrupted by OPCOMP before 20 kg is reached,
i.e. at 5-15 kg depending on the material used.
• Adapting OPCOMP.
NOTE
NOTE
If the factory-set OPCOMP is not to the customer's satisfaction,
an adjustment can be carried out with the tautness factors in the
"Optimized compression dr" table, see Service Instructions (Software, SPB7-250.816.03...). A lower tautness value will increase the
compression force and vice versa. Increase/decrease the tautness factor by e.g. 10 the first time. Let the customer use this setting for at least two weeks before further changes are made.
OPCOMP has been thoroughly tested by Siemens. If any factors
are changed, OPCOMP will no longer be an optimization of compression force and image quality according to Siemens' clinical
test.
Collimator checks4
• Checks have to be performed according to the chapter Collimator Adjustments of the
corresponding Startup with WH AWS, from System Serial No. 1101 SPB7-250.815.03...
Collimator wing difference Calibration4
Select the OPDIMAr wing by moving the detector in the parking position. OP is then displayed on the control console.
• Connect the compression plate simulator and the collimator-mounted plexi phantom.
• Start the service software on the service PC and select:
Configuration -> Collimator -> Calibration of wing difference
NOTE
If no precise wing difference calculation is performed, or if you have to start from scratch,
you can start with the "large focus/molly" values that have been adjusted with the
"Collimator beam calibration".
Siemens AGSPB7-230.814.03Page 13 of 16Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Configuration -> Collimator -> Calibration of wing difference.
Page 26
4 - 14Worksteps procedures
Fig. 21
• Enter the numbers in the table below:
• Narrow the X-ray field so that the edges of the detector will not be exposed to the X-ray
and white lines are visible in the image at the edges of the detector. To do this, decrease
the YC value and increase the YS, XR and XL values by 5 numbers.
• Make sure a patient is registered and start an exposure.
• Use a magnification of 4 in the application software and go to the edges of the detector.
White areas should be visible at all edges of the image. This shows you that the X-ray
does not cover the entire detector and shows where the beam really ends.
• Measure the white line thickness at each image edge with the application graphical user
interface and adjust the corresponding XL, XR, YC and YS values with the service
software.
• Start an exposure.
• Check if the white lines at the edges of the image are gone using a magnification of 4 in
the application graphical user interface.
• Save the calibration and Exit from the Service Software.
NOTE
If you see a white thin line on the chest wall when using the magnification.
table, it might be the magnification table itself. Move the magnification table out by approximately 1 mm and try again.
OpdimaSPB7-230.814.03Page 14 of 16Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 27
Worksteps procedures4 - 15
Disable the film exposure4
• Connect the Service PC to the MAMMOMAT Stand and start the service software
program.
• Go to Configuration -> Miscellaneous -> Allow film exposure and make sure that the
Switch is set to OFF Fig. 22.
Fig. 22
Checking the OPDOSE for Opdima wing4
• Select the Opdimar wing. As a result OP is displayed on the control console.
• Mount the spot 1.0 compression plate.
• Select a thickness of calibration plexi (part no. 65 61 240) that is within the thickness
interval of the program to be tested.
• Select AUTO on the control panel.
• Place the calibration plexi on the object table and compress to 6 kg or more.
• Check that the correct program is blinking on the control panel. Select the program by
pressing the corresponding program button on the control panel.
• Repeat the procedure described in steps 3-6 for the other programs used.
Grid Lines4
• Checks have to be performed according to the chapter Image Quality / Checking grid
lines of the corresponding Startup with WH AWS, from System Serial No. 1101
SPB7-250.815.03...
Siemens AGSPB7-230.814.03Page 15 of 16Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Page 28
4 - 16Worksteps procedures
Calibration of detector 4
• The Detector Gain calibration has to be performed according to Startup with WH AWS,
from System Serial No. 1101 SPB7-250.815.03...
QC Test4
NOTE
The following Quality Control Manual tests have to be carried out.
• Chest wall missed tissue; (No: 2).
• Collimator assessment; (No: 3).
• Compression plate overlap on the chest wall side; (No: 4).
• Spatial resolution; (No: 5).
• Phantom image quality; (No: 6).
• Detector uniformity; (No: 8).
• Mean glandular dose; (No: 12).
• AEC Image stability and reproducibility, signal-to-noise ratio; (No: 13).
• Select Backup > Copy installation Main menu: area to floppy A:.
• Leave the program with <F10> when the backup is ready.
• Keep the backup floppy with other documentation for the MAMMOMAT Novation
question.
DR
in
OpdimaSPB7-230.814.03Page 16 of 16Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 29
Final work steps5
Without OPDIMA5
• Fix the Generator to the stand.
• Close all the covers
• Check the grounding of the systems . Using the protective ground wire tester ,
measure the resistance between the protective ground terminal in the stand
(Mains input terminal) and all metallic parts of the equipment which the patient/
operator may come in contact with, incl. all object tab les. The protective ground
resistance must not exceed 0.1 Ω
• Perform a test exposure with collimator mounted phantom and when everything is
working fine turn the system over to the customer.
.
5 - 1
Siemens AGSPB7-230.814.03Page 1 of 2Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Page 30
5 - 2Final work steps
This page intentionally left blank.
Opdima SPB7-230.814.03Page 2 of 2Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 31
Changes to previous version6
Chapter 3Page 1Text added
Chapter 4Page 10Procedure described more in detail
6 - 1
Siemens AGSPB7-230.814.03Page 1 of 2Opdima
Medical SolutionsRev. 0201.05CS PS 24Installation
Page 32
6 - 2Changes to previous version
This page intentionally left blank.
Opdima SPB7-230.814.03Page 2 of 2Siemens AG
InstallationRev. 0201.05CS PS 24Medical Solutions
Page 33
Attention!
Print paper A4 size
Template No. 1
70 mm
125 mm
fold paper at dotted line
fold paper at dotted line