Siemens Opdima Installation manual [gb]

Page 1
Opdimar
Installation
Installation and Start-Up
SP
On MAMMOMAT Novation
English Print No.: SPB7-230.814.03.02.02 Doc. Gen. Date: 01.05
Replaces: SPB7-230.814.03.01.02
DR
Serial No < 1100
© Siemens AG 2004
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
Page 2
0 - 2 Revision
Chapter Page Revision
all all 02
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revi­sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho­rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Opdima SPB7-230.814.03 Page 2 of 4 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 3
Contents 0 - 3
Page
1 _______Prerequisites __________________________________________________1 - 1
Documentation required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Auxiliary Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Content of the Upgrade Kit Part No 6655984 . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Phantom / Auxiliary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
2 _______Important Notes________________________________________________ 2 - 1
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
Protective measures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
System Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
PC Boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Metal Curtain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Measurements with the oscilloscope . . . . . . . . . . . . . . . . . . . . . . . .2 - 3
Swivel-arm rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 3
Service Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Delay times between two exposures . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 4
3 _______Settings ______________________________________________________3 - 1
Presettings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
4 _______Worksteps procedures __________________________________________4 - 1
Covers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 1
Lift the generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 2
Cable Socket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
EMC filter (Z801) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
Cable Z801 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
Cable Bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 5
Flying Wing lever . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 7
Software Update V1.0 to 2.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 10
Parameter settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 11
Checks, adjustments and calibrations. . . . . . . . . . . . . . . . . . . . . . . . . 4 - 12
Rotation Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 12
Checking the OPCOMP function when Opdima wing selected . . . . . . . . . . . . 4 - 13
Collimator checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 13
Collimator wing difference Calibration. . . . . . . . . . . . . . . . . . . . . . . . . 4 - 13
Disable the film exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 15
Checking the OPDOSE for Opdima wing . . . . . . . . . . . . . . . . . . . . . . . 4 - 15
Grid Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 15
Calibration of detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 16
QC Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 16
Saving the MAMMOMAT Parameters . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 16
Siemens AG SPB7-230.814.03 Page 3 of 4 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 4
0 - 4 Contents
Page
5 ______ Final work steps________________________________________________5 - 1
Without OPDIMA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
6 ______ Changes to previous version _____________________________________6 - 1
Opdima SPB7-230.814.03 Page 4 of 4 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 5

Prerequisites 1

Documentation required 1

MAMMOMAT Novation
MAMMOMAT Novation
SPB7-250.815.03...
Instruction for use
QC Manual
MAMMOMAT Novation
Software Description V2.1. SPB7-250.816.03.

Auxiliary Materials 1

Drill 7 mm
Drill Machine
Center punch
Seeger circlip ring nipper
DR
Wiring Diagram SPB7-250.844.01.03..
DR
Startup with WH AWS, from System Serial No. 1101
DR
Replacements of parts SPB7-250.841.01.01..
1 - 1
PLCC Extractor
Service PC see the CS Intranet (CS/For Service/Common Services/Service Laptop for
CSEs) for details.
Power ground-wire tester 44 15 899
Serial Service Cable (Stand) Part No. 99 00 440

Content of the Upgrade Kit Part No 6655984 1

0010 0020 0025 0030 0040 0050 0060 0070 0080 0090 0100 0110 0120 0130 0140 0150 0160 0170 0180 0240 0245
06646678 06622927 06622919 06636661 06636679 06655869 06192236 05507038 06248046 06421494 09026634 06383066 03433919 03416666 03342532 06298637 06023980 03342292 06034581 06659747 06659762
BACK COVER CPL LOWER BACK COVER UPPER BACK COVER Holding Plate left Holding Plate right CABLE SOCKET CPL Grip LOCTITE_243 Lever axle Lever Bushing Spring 1x9x30 DIN 2098 EMC Filter Screws M4x12 DIN 912-12.9 0000 Washer 4 SN70093-FST Washer A8.4 DIN 125-ST 6980 Transformer 215 VA RONDOLL Screw MC6S6x65 DIN 912 Washer A6.4 DIN 125-ST 6980 Nut M6NY6 DIN 985 PROM kit Master VA00D Software kit V2.1
1.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
2.000
2.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
Siemens AG SPB7-230.814.03 Page 1 of 2 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 6
1 - 2 Prerequisites
0190 0195 0200 0210 0220 0230 0240
06658863 06501712 06655950 03343829 06582407 08102415 06659747
CABLE BRACKET Cable tie SN70764 CABLE Z801 Cable socket 6.3 - 2.5 DIN 46247 - BRONZE Heat shrink tubing 4.8 SW Label 8 x 17 Paper yellow PROM KIT MASTER PLD VA00D
1.000
3.000
1.000
0.070
7.000
1.000

Phantom / Auxiliary 1

Item Remarks Material Number With System
RMI 156 88 81 265 no One 0.1 mm, two 0.2 mm and
one 0.5 mm sheets of aluminium 99 % (1100) alloy, 4 x 4 cm
Mammography line pair phantom 2 - 10 Ip/mm 88 81 299 no
4.2 cm plexi Size of the detector
Centering cross 96 60 051 no
88 81 273 no
74 47 720 no
Opdima SPB7-230.814.03 Page 2 of 2 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 7

Important Notes 2

Symbols 2

This symbol indicates exposure of X-ray radiation.
" This symbol indicates that the measured values have to be entered.

Protective measures 2

System Power Supply 2

It is very important that any intervention in the equipment will start with disconnecting it from the power supply with the main circuit-breaker. Before removing or inserting any of the printed circuit boards, switch off the equipment.
2 - 1
DR
DR
system is only switched off at the
.
WARNING
WARNING
If the MAMMOMAT Novation control panel or with the circuit breaker CB S700, there is still voltage present within the generator. See the wiring diagram for MAMMOMAT Novation
Life-threatening electric shock hazard exists. Disconnect mains cable and comply with the information on this
page.
After shut-down of the system, there may still be 380 V DC present on the intermediate circuit of the MAMMOMAT generator.
Life-threatening electric shock hazard exists. A bleader resistor is connected over the output DC capacitors. The DC voltage will discharge below 60 Volt in 3 minutes. The voltage level is indicated by LEDs H3 and H4.

PC Boards 2

CAUTION
The p.c. boards contain electrostatic highly sensitive compo­nents.
If disregarded, the components could be damaged. Use ESD equipment, ground prior to making contact and place
the components on a conductive surface.
Siemens AG SPB7-230.814.03 Page 1 of 4 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 8
2 - 2 Important Notes

Metal Curtain 2

WARNING
The edges of the metal curtain of the stand are very sharp. They may cause severe injury. These strips must always be applied onto the edges of the metal
curtain during service and maintenance work.
Opdima SPB7-230.814.03 Page 2 of 4 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 9
Important Notes 2 - 3
Apply the protective strips after removing the covers from the stand. Remove the protec­tive strips only when the covers are to be mounted or when vertical adjustment of the swivel arm system is necessary.
1
2
Fig. 1 Protective strips for the metal curtain
Holders (1/Fig.1 and 2/Fig. 1) for storing the protective strips when not in use are provided on both sides of the curtain.

Measurements with the oscilloscope 2

WARNING
The existing ground conductor in the mains cable must under no circumstances be disconnected when operating the oscilloscope.
A life-threatening hazard of electric shock exists. For measurements in which a resulting ground loop may falsify
the measuring result, use the differential amplifier (difference measurement).

Swivel-arm rotation 2

WARNING
When rotating the swivel arm with the detector wing in parking position, make sure that the main cable harness does not get damaged. Stop rotation immediately, if the cable harness appears to be extensively pulled.

Service Software 2

The MAMMOMAT
Novation
Acquisition Workstation
The Acquisition Workstation service software is web-based and can be started from the Options application menu.
BRICK
The BRICK service software can be accessed via the Acquisition Workstation.
Siemens AG SPB7-230.814.03 Page 3 of 4 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
DR
has various service software packages:
Page 10
2 - 4 Important Notes

Delay times between two exposures 2

The delay times listed below must be followed to prevent the tube from overheating.
Exposure mAs value Delay time between two exposures (seconds)
max. 100 min. 15 max. 200 min. 30 max. 300 min. 45 max. 400 min. 60 max. 500 min. 75
Tab. 1 Delay times between two exposures
Opdima SPB7-230.814.03 Page 4 of 4 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 11

Settings 3

Presettings 3

3 - 1
NOTE
Switch ON the MAMMOMAT Novation
All the orientations left and right is always seen from the front of the MAMMOMAT System.
DR
System.
Connect the Service PC to the stand.
With the service software revision level V1.0 available at site make a backup of all
components. (i.e. of the stand and of the generator. The generator data are stored within the control panel.)
You may either "copy installation area to disc", or "copy installation area to floppy".
Label the floppy with "backup with SSW V1.0 / current date".
If the copy was made to the (hard) disc of the service PC, note down the location where
to the data were stored. (This backup may only be needed if some irregularity would occur.)
Switch OFF the MAMMOMAT System and switch OFF the main power line.
Siemens AG SPB7-230.814.03 Page 1 of 2 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 12
3 - 2 Settings
This page intentionally left blank.
Opdima SPB7-230.814.03 Page 2 of 2 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 13

Worksteps procedures 4

Covers 4

Remove the back cover.
Remove the back cover service.
Left front cover.
4 - 1
NOTE
Remove lower and back cover from the detector wing.
Remove the tube head cover.
Remove the covers below the bellow.
Attach metal curtain protection.

Lift the generator 4

NOTE
Loosen the four screws (One turn counter clockwise Fig. 2) fixing the generator to the
stand.
Remove the bottom hinge with the two screws.
NOTE
Place the washer 8.4 x16 x1.6 (Pos 140) on the lower hinge axle according to (1/Fig. 1).
Lift the generator door up and mount the hinge back to its original position.
Remove the four screws and swing the generator out.
The generator has to be lifted to make room for the EMC filter.
The generator will slide down a little.
Siemens AG SPB7-230.814.03 Page 1 of 16 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 14
4 - 2 Worksteps procedures
Screws for securing the generator
1
Fig. 1 Fig. 2

Transformer 4

Drill a hole (φ 7 mm) for mounting of the transformer according to Fig. 3 (spring site).
See to it that no remnants of the drilling are left inside the stand. For marking the place of the hole a drill template (No: 1) may be used.
Mount the transformer (Pos 150) inside the stand with a MC6S M6x65 (Pos 160), one
washer 6.4 x 12 (Pos 170) and a nut M6M (Pos 180) according to Fig. 3 and Fig. 4. The cable connections pointing downwards. Fasten the label T803 (use one off the yellow labels and write T803 on it) and stick it to the inside off the stand above the transformer.
Opdima SPB7-230.814.03 Page 2 of 16 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 15
Worksteps procedures 4 - 3
70
123
Fig. 3 Fig. 4

Cable Socket 4

Dismount the foot pedals strain relieves, the mains cable strain relief and the mains
supply connection from the Cable Socket CPL.
Remove the old Cable Socket CPL.
Mount the left and right foot pedal strain relieves on the inner positions on the new Cable
Socket CPL (Pos 50) according to (1/Fig. 5)
Install the new Cable Socket CPL.
1
Fig. 5
Siemens AG SPB7-230.814.03 Page 3 of 16 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 16
4 - 4 Worksteps procedures

EMC filter (Z801) 4

Mount the EMC filter (Pos. 110) upside down with two MC6S 4 x 12 screws (Pos. 120)
and two contact washers (Pos. 130), and fasten the labels according to Fig 6.
Label 1 to 4, Z801, Z801-1, Z801-2 and Z801 GND.
Fig. 6

Cable Z801 4

Dismount the D-sub plate.
Thread the Z801 cable (Pos 200) through the conduit entry according to Fig. 7.
Restore the D-sub plate. Attach the AMP connector to the transformer inside the stand.
Fig. 7
Attach the connectors to the EMC filter and to the connector X803 according to markings
on the connectors, according to Fig. 8.
NOTE
Opdima SPB7-230.814.03 Page 4 of 16 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Use the numbering of terminal contacts on the lower side of terminal row Z 803.
Page 17
Worksteps procedures 4 - 5
If you don't connect an Opdimar mains cable isolate the outputs 1 and 2 of the filter Z 801 by means of the supplied cable socket and heat shrink tubing in order to prevent electro­cution.
Fig. 8

Cable Bracket 4

Remove the Upper Back Cover and Lower Back Cover from the detector.
Remove the detector (Info in Replacement of parts SPB7-250.841.01.01..).
Remove the screws holding the bellow. Remove the strain relief from underneath the
Flying Wing (detector side). Lift the bellow and replace the old Cable Bracket with the new one (POS 190) according to Fig. 9.
Fig. 9
Siemens AG SPB7-230.814.03 Page 5 of 16 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 18
4 - 6 Worksteps procedures
Mount two cable ties (Pos 195) according to Fig. 9 and one cable tie according to Fig. 10.
Fig. 10
Replace the strain relief underneath the Flying Wing on the detector side according to
(1/Fig. 11).
1
Fig. 11
Mount the detector base blade.
Opdima SPB7-230.814.03 Page 6 of 16 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 19
Worksteps procedures 4 - 7

Flying Wing lever 4

Dismount the Clamp Rings (2 spare in the upgrade kit), the Pin and the Torsion Spring
according to (1/Fig. 12) and take out the Catch.
1
Fig. 12
Fig. 13
Assemble the Lever (Pos 60), Bushing (Pos 90), Axle (Pos 80), Catch and Spring
(Pos 70) according to Fig. 13 and Fig. 14 (apply Loctite 243 (Pos 70) on the Axle threads before turning it in to the lever handle.
Fig. 14
Siemens AG SPB7-230.814.03 Page 7 of 16 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 20
4 - 8 Worksteps procedures
Mount the Flying Wing Lever together with the Pin, Torsion Spring and Clamp Rings
according to Fig. 15.
Fig. 15
Reinstalling the Detector
Remove the existing rails left and right from the Detector wing.
Remove all the assemblies from the existing rails and reinstall it by the new one
(1/Fig. 16).
Install the new Spring plate according to the (2/Fig. 16).
1
2
2
1
Fig. 16
Opdima SPB7-230.814.03 Page 8 of 16 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 21
Worksteps procedures 4 - 9
Reinstall, i.e. attach both rails back to the Detector wing and reconnect the cables.
Move the Cover Bracket and the Service Cover from the old Lower Back Cover to the
new Lower Back Cover (Pos 20) according to Fig. 17.
Fig. 17
Mount the Upper Back Cover and the Lower Back Cover. Notice how the cables run
through the new opening on the Lower Back Cover according to Fig 18.
Fig. 18
Siemens AG SPB7-230.814.03 Page 9 of 16 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 22
4 - 10 Worksteps procedures

Software Update V1.0 to 2.1 4

OPDIMA can only be connected to NovationDR with the FW/SSW revision level V2.1 or higher !!
The actual installation of OPDIMA to a NovationDR can follow only when the NovationDR was already adapted, or if the NovationDR has had a serial number higher than 1100 !
The adaptation of NovationDR includes the Upgrade of the Firmware/ServiceSoftware from revision level V1.0 to revision level V2.1 !
Thereby following has to be exchanged:
On the PC Board D750:
exchange the IC 62 (MatNo.: 066 59 770), see 3 / Fig. 19.
exchange the IC 45 (MatNo.: 066 59 796), see 1 / Fig. 19.
exchange the IC 53 (MatNo.: 066 59 747), see 4 / Fig. 19.
The IC 52 (MatNo.: 066 59 440), see 2 / Fig. 19 - will remain, as it is.
Fig. 19 PCB D750
Opdima SPB7-230.814.03 Page 10 of 16 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 23
Worksteps procedures 4 - 11
On the PC Board D801:
Exchange the IC 9 (MatNo.: 066 59 788).
On the PC Board D740:
Exchange the IC 10 (MatNo.: 066 59 804), see Fig. 20.
Fig. 20 PCB D740
Service PC
Install the Service Software V2.1, delivered on a 3,5 inch floppy under the
MatNo.. 066 59 812, on the service PC.
NOTE
For further service use only the delivered service software V2.1 .

Parameter settings 4

Switch ON the main power line.
Switch ON the MAMMOMAT Novation
acquisition workstation doesn't need to be switched ON, yet.
The control console may display the ERROR 821, meaning: "The stand EEPROM has
the wrong version number (old or corrupt)". Disregard the message and acknowledge/ purge the same on the control console in the usual manner.
The service PC is still connected to the stand; reconnect if disconnected.
Siemens AG SPB7-230.814.03 Page 11 of 16 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
DR
System, during these work steps the
Page 24
4 - 12 Worksteps procedures
Perform "Backup" of both, the stand and the panel (control console) /generator, i.e.
"copy installation area to disc" or "copy installation area to floppy". Note down where to on the hard disc this backup data were stored, or attach corresponding label to the floppy.
Switch OFF the MAMMOMAT Novation DR system from the control console, wait for half
a minute.
Switch the system ON again, the ERR821 shouldn't appear this time.

Checks, adjustments and calibrations 4

NOTE
In the last few working steps the firmware / software V 1.0 of the MAMMOMAT NovationDR system was changed to the revision level V 2.1. All parameters, which have been previously config­ured, entered and stored on pertaining places ("installation areas") by the FW/SW V1.0 are present within the system in an unchanged format. EEPROMs, containing such settings, aren't supposed to be replaced in course of this upgrade. Since, how­ever, in course of some work steps the detector has been removed and the connection of the OPDIMA option was made available, following activities have to be performed additionally:
1) Rotation test/check.
2) Calibration/check of the optimized compression for OPDIMA; (or, setting of that parameter to the same value as for the DR).
3) Calibration of the collimator for the radiation beam and the light field.
4) Calibration of the wing difference; (or , setting of that parameter to the same value as for Mo large focus).
5) Disabling film exposure, i.e. the setting of the "Allow film expo­sure" software switch to OFF.
6) OPDOSE - Configuration of "Auto limits" for both, DR (Digital Radiography) and for the OPDIMA.
7) Check of the grid lines.
8) Calibration of the detector gain.

Rotation Test 4

Move the wing in different positions and watch to the cable.
The rotation of the swivel arm has to stop on position 0°, + 45°, + 90° and + 120°, when
OPDIMAr wing is selected.
The rotation of the swivel arm has to stop on position 0°, + 45°, + 90° and + 135°/ - 180°,
when Detector wing is selected.
Installing the wing cover and the bellow.
Opdima SPB7-230.814.03 Page 12 of 16 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 25
Worksteps procedures 4 - 13

Checking the OPCOMP function when Opdima wing selected 4

Decompress.
Set the compression force to 20 kg by using the presetting knob and compress.
The compression movement must be interrupted by OPCOMP before 20 kg is reached, i.e. at 5-15 kg depending on the material used.
Adapting OPCOMP.
NOTE
NOTE
If the factory-set OPCOMP is not to the customer's satisfaction, an adjustment can be carried out with the tautness factors in the "Optimized compression dr" table, see Service Instructions (Soft­ware, SPB7-250.816.03...). A lower tautness value will increase the compression force and vice versa. Increase/decrease the taut­ness factor by e.g. 10 the first time. Let the customer use this set­ting for at least two weeks before further changes are made.
OPCOMP has been thoroughly tested by Siemens. If any factors are changed, OPCOMP will no longer be an optimization of com­pression force and image quality according to Siemens' clinical test.

Collimator checks 4

Checks have to be performed according to the chapter Collimator Adjustments of the
corresponding Startup with WH AWS, from System Serial No. 1101 SPB7-250.815.03...

Collimator wing difference Calibration 4

Select the OPDIMAr wing by moving the detector in the parking position. OP is then dis­played on the control console.
Connect the compression plate simulator and the collimator-mounted plexi phantom.
Start the service software on the service PC and select:
Configuration -> Collimator -> Calibration of wing difference
NOTE
If no precise wing difference calculation is performed, or if you have to start from scratch, you can start with the "large focus/molly" values that have been adjusted with the "Collimator beam calibration".
Siemens AG SPB7-230.814.03 Page 13 of 16 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Configuration -> Collimator -> Calibration of wing difference.
Page 26
4 - 14 Worksteps procedures
Fig. 21
Enter the numbers in the table below:
Narrow the X-ray field so that the edges of the detector will not be exposed to the X-ray
and white lines are visible in the image at the edges of the detector. To do this, decrease the YC value and increase the YS, XR and XL values by 5 numbers.
Make sure a patient is registered and start an exposure.
Use a magnification of 4 in the application software and go to the edges of the detector.
White areas should be visible at all edges of the image. This shows you that the X-ray does not cover the entire detector and shows where the beam really ends.
Measure the white line thickness at each image edge with the application graphical user
interface and adjust the corresponding XL, XR, YC and YS values with the service software.
Start an exposure.
Check if the white lines at the edges of the image are gone using a magnification of 4 in
the application graphical user interface.
Save the calibration and Exit from the Service Software.
NOTE
If you see a white thin line on the chest wall when using the mag­nification.
table, it might be the magnification table itself. Move the magnifi­cation table out by approximately 1 mm and try again.
Opdima SPB7-230.814.03 Page 14 of 16 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 27
Worksteps procedures 4 - 15

Disable the film exposure 4

Connect the Service PC to the MAMMOMAT Stand and start the service software
program.
Go to Configuration -> Miscellaneous -> Allow film exposure and make sure that the
Switch is set to OFF Fig. 22.
Fig. 22

Checking the OPDOSE for Opdima wing 4

Select the Opdimar wing. As a result OP is displayed on the control console.
Mount the spot 1.0 compression plate.
Select a thickness of calibration plexi (part no. 65 61 240) that is within the thickness
interval of the program to be tested.
Select AUTO on the control panel.
Place the calibration plexi on the object table and compress to 6 kg or more.
Check that the correct program is blinking on the control panel. Select the program by
pressing the corresponding program button on the control panel.
Repeat the procedure described in steps 3-6 for the other programs used.

Grid Lines 4

Checks have to be performed according to the chapter Image Quality / Checking grid
lines of the corresponding Startup with WH AWS, from System Serial No. 1101 SPB7-250.815.03...
Siemens AG SPB7-230.814.03 Page 15 of 16 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 28
4 - 16 Worksteps procedures

Calibration of detector 4

The Detector Gain calibration has to be performed according to Startup with WH AWS,
from System Serial No. 1101 SPB7-250.815.03...

QC Test 4

NOTE
The following Quality Control Manual tests have to be carried out.
Chest wall missed tissue; (No: 2).
Collimator assessment; (No: 3).
Compression plate overlap on the chest wall side; (No: 4).
Spatial resolution; (No: 5).
Phantom image quality; (No: 6).
Detector uniformity; (No: 8).
Mean glandular dose; (No: 12).
AEC Image stability and reproducibility, signal-to-noise ratio; (No: 13).
Ghost image evaluation; (No: 14).
SNR and CNR measurements; (No: 15).
AEC testing / AEC Thickness Tracking Test; (No: 16).

Saving the MAMMOMAT Parameters 4

Start the service software.
Select Backup > Copy installation Main menu: area to floppy A:.
Leave the program with <F10> when the backup is ready.
Keep the backup floppy with other documentation for the MAMMOMAT Novation
question.
DR
in
Opdima SPB7-230.814.03 Page 16 of 16 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 29

Final work steps 5

Without OPDIMA 5

Fix the Generator to the stand.
Close all the covers
Check the grounding of the systems . Using the protective ground wire tester ,
measure the resistance between the protective ground terminal in the stand (Mains input terminal) and all metallic parts of the equipment which the patient/ operator may come in contact with, incl. all object tab les. The protective ground resistance must not exceed 0.1
Perform a test exposure with collimator mounted phantom and when everything is
working fine turn the system over to the customer.
.
5 - 1
Siemens AG SPB7-230.814.03 Page 1 of 2 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 30
5 - 2 Final work steps
This page intentionally left blank.
Opdima SPB7-230.814.03 Page 2 of 2 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 31

Changes to previous version 6

Chapter 3 Page 1 Text added Chapter 4 Page 10 Procedure described more in detail
6 - 1
Siemens AG SPB7-230.814.03 Page 1 of 2 Opdima Medical Solutions Rev. 02 01.05 CS PS 24 Installation
Page 32
6 - 2 Changes to previous version
This page intentionally left blank.
Opdima SPB7-230.814.03 Page 2 of 2 Siemens AG Installation Rev. 02 01.05 CS PS 24 Medical Solutions
Page 33