Siemens MULTIMOBIL 10 Troubleshooting Guide

MULTIMOBIL 10
Troubleshooting Guide
SP
© Siemens AG 2003
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
English Print No.: SPR8-X01.840.10.02.02 Doc. Gen. Date: 10.05
Replaces: SPR8-X01.840.10.01.02
Multimobil 10
Service Manual
Service Instructions
Med
Service Instructions
Version: 4.0
The reproduction, transmission or use
of this document or its contents is not permitted without express written authority. Offenders will be liable for damages.
Siemens Ltd. Med Division India Version 4.0
Page 1 of 48
Service Manual
Service Instructions
INDEX
1 SERVICE................................................................................................................................................... 4
1.1 SAFETY PRECAUTIONS : .......................................................................................................................... 4
1.2 TROUBLESHOOTING : .............................................................................................................................. 4
1.2.1 Tools and measuring instruments required....................................................................................... 4
1.3 MEASURING WITH OSCILLOSCOPE : ......................................................................................................... 4
1.4 HANDLING PRECAUTIONS........................................................................................................................ 4
1.5 CONTROL VOLTAGES .............................................................................................................................. 5
1.6 SIGNALS & LINE VOLTAGE DISPLAY: ...................................................................................................... 5
1.7 HT CHECK : ............................................................................................................................................ 6
1.8 EXPOSURE :............................................................................................................................................. 6
1.9 TIMING :.................................................................................................................................................. 6
1.10 KVP,KVN &KV MEASUREMENT :.......................................................................................................... 6
1.10.1 kVp-meter :................................................................................................................................... 6
1.10.2 Electrical :.................................................................................................................................... 7
1.11 MA/MAS MEASUREMENT :...................................................................................................................... 7
1.11.1 mAs-meter : .................................................................................................................................. 7
1.11.2 Electrical :.................................................................................................................................... 7
1.12 CALIBRATION :........................................................................................................................................ 7
1.13 FORMING :............................................................................................................................................... 8
1.14 SAFETY NOTES AND PROTECTIVE MEASURES .......................................................................................... 8
1.15 REPLACING DAMAGED OR MISSING SCREWS ............................................................................................ 8
1.16 CLEANING............................................................................................................................................... 8
1.17 CLEANING THE UNIT............................................................................................................................... 8
1.18 DISINFECTING ......................................................................................................................................... 9
1.19 MAINTENANCE........................................................................................................................................ 9
1.20 PRODUCT DISPOSAL................................................................................................................................9
1.21 SAFETY CHECKS ................................................................................................................................... 10
1.21.1 Daily checks................................................................................................................................ 10
1.21.2 Monthly checks ........................................................................................................................... 10
1.21.3 Performance Check (Every Six months) ..................................................................................... 10
1.22 PROGRAM MODES FOR SERVICING........................................................................................................ 11
1.22.1 Program 1 : To display the DC voltage on the capacitor........................................................... 11
1.22.2 Program 2 : To check the total number of exposures taken . .................................................... 11
1.22.3 Program 3 : To check the type of codes that occurred in the unit maximum 20 ....................... 11
1.22.4 Program 4 : To Erase the errors that have occurred on the unit. .............................................. 12
1.22.5 Program 5 : To set the display value at switch ON to Default Kv mAs combination or last
selected Kv mAs combination. ....................................................................................................................... 12
1.22.6 Program 6 : To set the maximum kV and mAs value for the unit. .............................................. 12
1.22.7 Program 7 : To set the max.inverter firing frequency(REG)...................................................... 12
2 CODES..................................................................................................................................................... 14
3
TROUBLESHOOTING ......................................................................................................................... 16
3.1 REMOVING THE TOP PANEL, FRONT COVER AND X-RAY SOURCE ASSEMBLY ......................................... 16
3.1.1 Removing the top panel ................................................................................................................... 16
3.1.2 Removing the Front cover. .............................................................................................................. 17
3.1.3 Removing the X-ray source assembly..............................................................................................17
3.2 CHECKING THE LINE VOLTAGE, FUSES AND LEDS................................................................................. 18
Siemens Ltd. Med Division India Version 4.0
Page 2 of 48
Service Manual
3.2.1 Checking the line voltage ................................................................................................................ 18
3.2.2 Checking the fuses........................................................................................................................... 19
3.2.3 Checking the LEDs on D801 ........................................................................................................... 19
3.3 CHECKING THE CONTROL VOLTAGES ................................................................................................... 19
3.4 CHECKING THE INTERMEDIATE CIRCUIT VOLTAGE................................................................................ 20
3.5 CHECKING THE MAXIMUM MAIN INVERTER FREQUENCY ...................................................................... 20
3.6 FILAMENT CURRENT MEASUREMENT .................................................................................................... 20
3.7 CHECKING THE HIGH VOLTAGE KVSOLL AND KVIST............................................................................. 20
3.8 SETTING THE MAXIMUM FILAMENT FREQUENCY .................................................................................. 21
3.9 CHECKING THE TUBE CURRENT ............................................................................................................. 21
3.10 CHECKING THE KV AND TUBE CURRENT (JR)........................................................................................ 21
3.11 CHECKING THE MAS VALUES ................................................................................................................ 21
3.12 ADJUSTING THE MAS ............................................................................................................................ 22
3.13 SWITCHING THE COLLIMATOR LIGHT ON............................................................................................. 22
3.14 ALIGNING THE LIGHT FIELD TO THE RADIATION FIELD .......................................................................... 22
3.14.1 Operating sequence .................................................................................................................... 22
3.14.2 Evaluation: ................................................................................................................................. 22
3.15 CHECKING AND READJUSTING THE COUNTERWEIGHT ........................................................................... 23
3.16 ADJUSTMENT INSTRUCTIONS : .............................................................................................................. 23
3.17 REPLACING THE SINGLE TANK............................................................................................................... 23
3.18 REPLACING THE COLLIMATOR............................................................................................................... 24
3.19 REPLACING THE COLLIMATOR LAMP .................................................................................................... 24
Service Instructions
4 SPECIFIC CODE HANDLING ............................................................................................................ 25
4.1 UNIT NOT SWITCHING ON ..................................................................................................................... 25
4.2 NO RADIOGRAPHY ................................................................................................................................27
4.3 NO STANDBY ........................................................................................................................................ 29
4.4 INITIALISATION CODES.......................................................................................................................... 30
4.4.1 Code 90 (EPROM CHECKSUM FAILURE)................................................................................... 30
4.4.2 Code 96 (KV SOLL FAILURE) ......................................................................................................30
4.4.3 Code 97 ( mA FAILURE) ................................................................................................................ 30
4.5 STANDBY CODES .................................................................................................................................. 31
4.5.1 Code 02 : +15 V Supply Code......................................................................................................... 31
4.5.2 Code 03 & 04 : Iheiz < Istby & Iheiz > Istby.................................................................................. 32
4.5.3 Code 05 : kVist <> 0....................................................................................................................... 33
4.5.4 Code 06 : JR <> 0........................................................................................................................... 34
4.5.5 Code 33 : Main Inverter Short Circuit............................................................................................ 35
4.6 EXPOSURE CODES ...........................................................................................................................36
4.6.1 Code 11 : Main Inverter Short Circuit............................................................................................ 36
4.6.2 Code 12 : kVist > kVmax................................................................................................................. 37
4.6.3 Code 13 : Iheiz > Imax OR JR > Jrmax......................................................................................... 38
4.6.4 Code 14 : kVist < kVsoll.................................................................................................................. 39
4.6.5 Code 15 : JR < JRS......................................................................................................................... 40
4.6.6 Code 17 : Exposure terminated by Backup Timer........................................................................... 41
4.6.7 Code 18 : Premature termination of Exposure................................................................................ 42
4.6.8 Code 21 : Iheiz > Iheiz maximum.................................................................................................... 43
4.6.9 Code 22 : Maximum Preparation Time........................................................................................... 44
4.6.10 Code 10 : Rotating Anode not OK.............................................................................................. 45
4.6.11 Code 31 : No charging ............................................................................................................... 46
4.6.12 Code 01 : Err signal on D950A active ....................................................................................... 48
Siemens Ltd. Med Division India Version 4.0
Page 3 of 48
Service Manual
Service Instructions
1 Service
1.1 Safety Precautions :
Before working on the unit ensure that the Mains is switched OFF. For completely safe working remove mains plug.
The unit must be OFF before removing or putting back the assemblies.
The protective earthing of subassemblies must not be removed in any case.
1.2 Troubleshooting :
1.2.1 Tools and measuring instruments required
Usual service tools
Digital Multimeter:
2 channel storage oscilloscope with 2 probes and one current probe.
mAs meter
Protective ground wire and leakage current tester: Bender safety tester
Radiation detector
During Oscilloscope operation the protective ground wire connection
in the power cable must not be interrupted under any circumstances. For measurements where ground loops that may be present could impair the measuring result, use the Tek amplifier and the trigger attachment.
1.3 Measuring with oscilloscope :
For normal working, the protective earthing of mains must not be removed. In case system noise is getting coupled to the scope, use the scope in differential mode. To measure high voltage, use DIFFERENTIAL probe isolated from ground.
1.4 Handling precautions
Strictly observe ESD precautions while handling PCBs and ESD sensitive devices.
After the unit is switched ON, the capacitors C1~C12 on D970A and D971 charge to 355V DC. After switching the unit OFF, it takes about 08 minutes for capacitors to discharge. The presence of DC voltage is indicated by Green LED V24 on D950A and also V1 ~ V12 LED’s on D970A / D971.
Siemens Ltd. Med Division India Version 4.0
Page 4 of 48
Service Manual
Service Instructions
1.5 Control Voltages
Voltage Measuring Point Range
+5 V D915. X15.4 4.8v to 5.2v
+15 V D915. X15.5 13.5v to 16.5v
-15 V D915. X15.7 -13.5v to -16.5v
Collimator Supply D801 . X6 12V ac to 16V ac
1.6 Signals & Line voltage display:
Signal Measuring Point Range Adjustment
kVsoll D915.TP.KVS
2.0 ± 0.1 V for 60kV
D915.P7
during Exposure
mAsoll D915.TP.JRS
3.2 V ± 0.1 V for 160mA
D915.P6
during Exposure
D915.TP. IH during
1.5V± 0.1V
Standby
IH
REG: Max
Main Inv. req.
CAL :Max
Fil.Inv. Freq
D915.TP. IH during Preparation
D915.TP.REG point during exposure
D915 TP CAL during start up
2V ±0.3V (for critically damped mA)
15 kHz ± 0.5 kHz
15KHz ± 0.5 kHz
D915.P5
D915.P4
D915 .P3
The actual mains Line voltage, at 110 and 230 V supply, is being monitored for display on the 7 segment display during first step of hand release switch for radiography. Ensure that display value mathes with actual line voltage with allowed tolerance ± 5 %. Using P1 potetiometer on D801 PCB, set the voltage accordingly.
This is applicalbe while replacing D915 and D801 PCB also.
If the Line voltage is less than 205 volts on on 230 V line, microcontroller switches “ON“ the ’LN’ contactor which will support extra voltage through auto transformer for rotating anode to run at higher speed. In preperation and exposure mode contactor retains its position. After 5 second of each termination of radiography Line voltage is being monitored once again.
When unit is connected to 110 V line LN contactor does not function even though there is a variation in the supply voltage.
Siemens Ltd. Med Division India Version 4.0
Page 5 of 48
Service Manual
Service Instructions
1.7 HT Check :
Presence of High Tension
Observe Safety Precautions.
Use radiation protection
1.8 Exposure :
Switch the generator ON. Set exposure parameters as 60Kv, 16 mAs. Pull the recoilable exposure release switch and release an exposure. The yellow LED on top panel lights up for a short duration equal to exposure time. Audible indication by a Buzzer from the control and muffled sound from inverter confirms the working of the inverter. The LED’s on D915 light in the following sequence: S27 - STEP 1(half press) : V22 Exposure state : V23 In case of code : V24 Code corresponding to the code is displayed on the seven-segment display. Sharps knocking sound from inverter indicates inverter short circuit, resulting in code display on the top panel.
1.9 Timing :
Set 60 kV 16mAs Release an exposure. The radiation indication LED will light for 100msec ± 5msec.
1.10 KVp,KVn &KV Measurement :
The kVp, KVn & KV measurement can be carried out either by kVp-meter or by electrical measurement.
1.10.1 kVp-meter :
Switch the unit ON. Insert the kV sensor into collimator channel and connect the cable to the meter. Open collimator flaps such that the sensor is well covered by the field. Set the kVp meter at 12pulse and 55-85kV ranges. Set exposure parameters as 60kV 20mAs. Release an exposure. The radiation LED will light up during exposure. The meter will read 57-63kV. Repeat with meter setting 77-150kV and exposure parameter setting, as 90kV 20mAs.The meter will read 85.5-94.5 kV.
Siemens Ltd. Med Division India Version 4.0
Page 6 of 48
Service Manual
Service Instructions
1.10.2 Electrical :
Connect oscilloscope at D915.TP.kV. Set the exposure parameters as 60kV 16mAs. Release an exposure. The recorded waveform will read 2.9V - 3.1V. Repeat with 100kV 10mAs. The recorded waveform will read 5V +/-0.25V.
Lower kV indicates code in kVsoll / kV sensing / kV-regulator OR high mA. Higher kV indicates code in kVsoll / kV sensing / kV regulator.
1.11 mA/mAs Measurement :
mAs measurement could be carried out either by mAs-meter or by electrical measurement.
1.11.1 mAs-meter :
Remove mAs link on D800 PCB on the Single tank. Insert mAs-meter leads in the banana sockets mA+ and mA- on D800. Set meter at, 200mAs and in the mAs mode. Set exposure parameters as 60kV 32mAs. Release an exposure. The meter will read 31 – 33 mAs. Repeat for exposure parameters 100kV 40mAs. The meter will read 39 to 41mAs.
1.11.2 Electrical :
Connect oscilloscope at D915.TP.JR. Set exposure parameters 60kV 16mAs. Release an exposure. The recorded waveform will read 3 – 3. 2V. The exposure time will be 95 - 105mS.
Lower mA indicates code in mAsoll / mA sensing / mA regulator circuit. Higher mA indicates code in mAsoll / mA sensing / mA regulator or Low kV. Incorrect mAs indicate code in timer circuitry.
1.12 Calibration :
It is advisable to re-calibrate the unit in case of undershoot or overshoot observed in the mA wave shape. Recalibration is required when the single tank is replaced.
Siemens Ltd. Med Division India Version 4.0
Page 7 of 48
Service Manual
Service Instructions
1.13 Forming :
If the unit is being installed after six months from the date of despatch, the capacitors C1~C12 on D970A/D971 need forming. Refer Installation instructions for forming procedure.
1.14 Safety notes and protective measures
Refer to the General safety note in Chapter on Installation.
Before working on the open MULTIMOBIL 10, switch off the unit with the power switch at the control panel, the MCB in off position and remove the Mains cable from socket. Before inserting or removing PC boards, switch off the generator and observe the ESD regulations.
1.15 Replacing damaged or missing screws
Damaged or missing screws should only be replaced by steel screws having the specified tensile strength with Nickel Plating (6492) specified in the Surface Finish Documents.
1.16 Cleaning
The unit must always be switched OFF or disconnected before cleaning. Use only water or a Luke warm mixture of a household cleaner diluted with water to clean the unit. Do not use an abrasive or organic solvents or cleaning agents containing solvents such as gasoline used for cleaning purposes, alcohol or stain remover. Do not spray water on the unit.
1.17 Cleaning the Unit
Switch OFF the system prior to cleaning, disinfecting or sterilizing.
Clean the stand with a damp cloth or cotton pad. Dampen the cloth or pad with water or lukewarm, diluted commercially available liquid cleaning solution.
Note: Do not use scouring powder; organic solvents (may damage materials) or solvent based cleaning fluids (benzene, alcohol, spot remove).
Siemens Ltd. Med Division India Version 4.0
Page 8 of 48
Service Manual
Service Instructions
Do not spray any fluid on the unit. The cleaning fluid must not seep into the generator.
1.18 Disinfecting
As is commonly known, the ingredients in disinfectants are hazardous to your health. Ensure that the room is well ventilated when using any disinfectant. Follow manufacturer’s instructions for the use of the product.
To disinfect surfaces, we recommend common liquid solutions of aldehyde-based or ampholytic surfactant -based disinfectant.
Substituted phenol based or chlorine-releasing disinfectant can weaken materials and are not recommended. The same restrictions apply to undiluted solutions with a high alcohol content (e.g. for disinfecting hands). Disinfectant sprays should generally not be used, the spray can seep into the system, and safety features can no longer be guaranteed (possible damage to electrical components, formation off flammable mixtures of air and the solution vapor.
1.19 Maintenance
All parts, modules of this product must be tested, inspected, for performance and safety aspects every 12 months to ensure that the product functions properly and is safe for patients, operating personnel and other persons. Every 12 months, trained technical personnel should inspect and if necessary replace system components of the generator, which may produce hazardous, conditions due to excessive wear and tear. These instructions should be included in the annual maintenance performed by Siemens Customer Service.
If more frequent inspections and maintenance are required by federal or local regulations, ensure compliance with them.
1.20 Product Disposal
Legal regulations may contain special prescriptions concerning the disposal of this product. In order to avoid environmental damage and/or personal injury, please inform SIEMENS Customer Service if you want to put the unit out of operation and dispose it. Radioprotection Material Lead in X-ray tube assembly and in Collimator 3 kg.
Siemens Ltd. Med Division India Version 4.0
Page 9 of 48
Service Manual
Service Instructions
Transformer Oil
Oil in X-ray tube assembly, approx. 7 kg.
Plastics
Epoxy resin on PC boards, PVC of cables approx. 10 kg.
Electrolytic Capacitors
These capacitors must be emptied. Please inform the Siemens Customer Service. Capacitors approx. 15 kg Remaining Electrolytic Capacitors approx. 0.5 kg.
1.21 Safety Checks
The user shall follow the following safety checks:
1.21.1 Daily checks
Check the power cord. Do not use the unit if the power cord is damaged.
Check the exposure indicator and audible signal during exposure.
Check that the Single tank assembly and the arm system remain in the desired position.
1.21.2 Monthly checks
Check the lifting and lowering movements of the support arm.
Check function of locking lever and cap screw on the X-ray tube
Check the legibility of the labels according to those in section Location of Labels. If any labels require to be replaced, please notify Siemens customer service.
1.21.3 Performance Check (Every Six months)
Check counterbalancing of the support arm at each position.
Check the locking mechanism for the support arm.
Check brake for proper functioning .
Take a sample exposure at 60 kV, 16 mAs.
Check for Radiation indication LED and the Audio indication in form of Buzzer.
Siemens Ltd. Med Division India Version 4.0
Page 10 of 48
Service Manual
Service Instructions
Note : Annual Maintenance must be performed as per the guidelines described in
the Maintenance section. Notify Siemens Customer Service if you do not have a maintenance contract.
1.22 Program Modes for servicing.
To put the unit into service mode, Short the ST link on the D915 and put ON the unit.Use the Kv+ and Kv- switches to select the required program from 1 to 7.
To check the parameters in the program,press the DL-serv switch once.
1.22.1 Program 1 : To display the DC voltage on the capacitor.
Select the program one in service mode and the display will read as below
Pr 1
Press the switch DL-serv on the D936 once.
The display will read as below
Cap XXX
XXX corresponds to the DC voltage on the capacitors.
1.22.2 Program 2 : To check the total number of exposures taken .
Select the program two in service mode and the display will read as below
Pr 2
Press the DL-serv switch and enter into the program.The display will be
XXX XXX
XXX corresponds to the total number of exposures taken on the unit.
1.22.3 Program 3 : To check the type of codes that occurred in the unit maximum 20 .
Select the program three in service mode and the display will read as below
Pr 3
Press the switch DL-serv on the D936 once.The display will read as below.
AA XX
AA displays the serial no of the errors occurred in chronological order. XX corresponds to the type of code.
Siemens Ltd. Med Division India Version 4.0
Page 11 of 48
Service Manual
Service Instructions
Press the kV+ switch to check all the codes stored. Maximum 20 codes can be stored.
1.22.4 Program 4 : To Erase the errors that have occurred on the unit.
Select the program four in service mode and the display will read as below
Pr 4
Press the switch DL-serv on the D936 once.The display will read as below.
Era Err
To erase the errors keep the mAs + switch pressed till the display reads
0 Err
1.22.5 Program 5 : To set the display value at switch ON to Default Kv mAs combination or last selected Kv mAs combination.
Select the program five in service mode and the display will read as below
Pr 5
Press the switch DL-serv on the D936 once.The display will read as below.
las Val
The default value can be set by selecting the desired kV, mAs combination. Last selected combination can be stored by pressing the DL-serv switch twice.
1.22.6 Program 6 : To set the maximum kV and mAs value for the unit.
Select the program six in service mode and the display will read as below
Pr 6
Press the switch DL-serv on the D936 once.The display will read as below.
125 100
Set the required limit for kV and mAs using the kV /mAs increment and decrement switches.
1.22.7 Program 7 : To set the max.inverter firing frequency(REG).
Select the program seven in service mode and the display will read as below
Pr 7
On activation of this program by pressing the DL-serv switch , the display indicates
Adj F
Siemens Ltd. Med Division India Version 4.0
Page 12 of 48
Service Manual
Service Instructions
Press the Radiographic exposure switch and monitor REG on Digital storage Oscilloscope.
The display indicates
F On
Adjust REG for 15 kHz ±0.5kHz by using P4 on D915 card.
Siemens Ltd. Med Division India Version 4.0
Page 13 of 48
Service Instructions
2 Codes
Initialisation Codes
90
EPROM Checksum Failure
96
KVsoll Failure
97
JRS Failure
99 Last Reset by Watchdog
Timer
02
+15V Supply Code
03
Iheiz < Istby
04
Iheiz > Istby The maximum value of Standby current permitted is 3.3 A.
05
KVist <> 0
06
JR <> 0
33
Main Inverter Short Circuit
34
Filament Inverter Short Circuit
07
Rot 0
08
Braking Failure
01
Err signal on D950A active
31
No Charging
10
No Rotation of Rotating Anode
11
Main Inverter Short Circuit
12
KVist > kVmax
13
Iheiz > Imax OR JR > JRmax
The EPROM Checksum is stored at 7FFEh and 7FFFh as a 16 bit word. During self Diagnostics the software calculates the checksum of the EPROM and compares with the stored checksum.
During Self Diagnostics the software outputs 7Fh to the D/A converter (B). The 2.5V at the output of the D/A converter is Read by the µC through Analog Port 4. The value read should be greater than 7Ah and less than 86h. (I.e. between 2.39V and 2.62V)
During Self Diagnostics the software outputs 7Fh to the D/A converter (A). The 2.5V at the output of the D/A converter is Read by the µC through Analog Port 3. The value read should be greater than 7Ah and less than 86h. (I.e. between 2.39V and 2.62V)
The built-in Watchdog timer (WDT) is reset by the software every 25 mSec. If due to some failure the software doesn’t reset the WDT, the WDT in turn will reset the µC after 65 mSec
Standby Codes
The +15V supply from SMPS is polled by the µC through Analog port 0. The +15V supply should be between +12V to +18V.
The Filament Standby current is 3A. Iheiz read by the µC through Analog port 2 should be greater than 2.75A. i.e. 1.35V (Iheiz ratio : 1V = 2A)
kVist is read by the µC through Analog port 7. During Standby the value of kVist read should be Zero. (kVist ratio : 1 V = 20kV)
JR is read by the µC through Analog port 1. During Standby the value of JR read should be Zero. (JR ratio : 1 V = 50mA)
If the Main Inverter Driver (Cable) is disconnected this Code gets activated. Also if IGBT Drivers on D960 Card detects Short Circuit of IGBT’s this Code gets activated.
This Code gets activated if IGBT Drivers on D801 Card detects Short Circuit of Filament Inverter IGBT’s
The current flowing through both the windings of motor is monitored separately. During standby Rotor should be off.RT1, RT2=0.
After exposure is terminated the Unit returns to Standby state and braking of Rotor is started by allowing a DC current to flow through the KI2.During this state which is for 8 sec after exp, RT1=0 and RT2=1.
Temperature on the heat sink of D950A exceeds it’s limit or Voltage exceeds 365v across capacitors C1~C12 or IGBT V24 on D950A is short circuit.
Voltage on capacitors C1~ C12 does not increase within 3.5sec of CS picking up or it is less than 335 V DC
Exposure Codes
Anode Rotation in the Single Tank is not taking place. This Code is displayed if any one of the RT1and Rt2 signal goes low during preparation.
This Code is displayed, when the drivers of the Main Inverter detect short circuit.
The PkV and NKV is monitored for Max.68kV separately. If the actual value of kV is greater than this i.e. if the voltages at J12.8 and J12.10 on D915 cross
4.5V, Code is displayed.
Maximum value of Iheiz above which Code will be displayed is 7 A (3.5V).
Service Manual
Siemens Ltd. Med Division India Version 4.0
Page 14 of 48
Loading...
+ 34 hidden pages