Siemens MULTIGAS Service Manual

s

MULTIGAS and MULTIGAS+ Modules

Service Manual

EM Guidelines, 1997-04-02

E341.E536U.061.01.02.02 Replaces/Ersetzt: E341.E536U.061.01.01.02 ASK-T876-02-7600

ADVISORY

Siemens is liable for the safety of its equipment only if maintenance, repair, and modifications are performed by authorized personnel, and if components affecting the equipment's safety are replaced with Siemens spare parts.

Any modification or repair not done by Siemens personnel must be documented. Such documentation must:

be signed and dated

contain the name of the company performing the work

describe the changes made

describe any equipment performance changes.

It is the responsibility of the user to contact Siemens to determine warranty status and/or liabilities if other than an authorized Siemens Service Representative repairs or makes modifications to medical devices.

Table of Contents

Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Figure 1-1 MultiGas and MultiGas+ Modules - Rear Views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

2 Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

3 Service Strategy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

4 Preventative Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

5 Recommended Tools & Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Chapter 2: Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5

1

Introduction

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 5

 

Figure 2-1 Functional MGM Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 5

2

Overall Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 5

3

Method of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 5

 

Figure 2-2 Anesthetic Gas Subsystem Functional Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6

4

Subassemblies, Modules and Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6

5

Anesthetic Gas Subsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6

6

Main System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6

7

Agent Measurement Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 7

 

Figure 2-3 Agent Measurement Analyzer Functional Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . .

. 8

 

Figure 2-4 Agent Identification Analyzer (AIDA) Functional Block Diagram . . . . . . . . . . . . . . . . . . . . .

. 9

8

Agent Identification Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 9

9

Oxygen Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10

 

9.1

Paramagnetic O2 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10

 

9.2

Electrochemical O2 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10

10Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10

 

Figure 2-5 Pneumatics Block Diagram (excerpt from Figure 2-2.) . . . . . . . . . . . . . . . . . . . . . . . . . . . .

11

 

 

Table 2-1 Pump Flow Rates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

11

11Self-Test . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

12

12Calibration .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

12

 

12.1Factory Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

12

 

12.2Field Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

12

 

12.3Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

12

 

12.4Storage of Calibration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

13

 

12.5Calibration of Agent Measurement Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

13

 

 

12.5.1

Factory Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

13

 

 

12.5.2

Field Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

13

 

 

12.5.3

Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

13

 

 

 

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

i

mgm2.sm.fm/11-00/kaupp

MGM and MGM+ Modules

Service Manual

 

 

12.6Agent Identification Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 12.6.1 Factory Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 12.6.2 Field Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 12.6.3 Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 12.7Paramagnetic O2 Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 12.7.1 Factory Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 12.7.2 Field Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 12.7.3 Field Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 12.8Fuel Cell Type O2 Analyzer Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 12.8.1 Factory Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 12.8.2 Field Span and Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 12.8.3 Field Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 12.9Pneumatic System Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 12.9.1 Factory Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 12.9.2 Field Flow Rate Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 13Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 13.1AMA Data Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Figure 2-6 Software Functional Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 13.2AMA Signal Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 13.3AMA Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 13.4Patient Data Derivation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 13.5Host System Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 13.6Anesthetic Gas Subsystem / Host System Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 13.7AIDA Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 13.8MGM/AMA Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 13.9AIDA Data Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 13.10AIDA Signal Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 13.11AIDA Agent Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 13.12AIDA Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Chapter 3: Removing / Installing Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

21

1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

1.1 Opening Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Figure 3-1 Pump and Pneumatic Subassemblies in MGM+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

2 MGM+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

2.1 Pneumatic Tubing Subassemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

2.1.1 Removing Pneumatic Subassemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

2.1.2 Installing Pneumatic Subassemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Figure 3-2 Pump and Pneumatic Subassemblies in MGM (Slo-O2) . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

ii

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

mgm2.sm.fm/11-00/kaupp

Service Manual

MGM and MGM+ Modules

 

 

3 MGM (Slo-O2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 3.1 Replacing Pneumatic Subassemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 3.1.1 Removing Pneumatic Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 3.1.2 Installing Pneumatic Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 4 Replacing Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

4.1 Removing Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 4.2 Installing Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 5 Pump Filter Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 5.1 Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 5.2 Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 6 Nafion® or other Tubing Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 6.1 Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 6.2 Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 7 Room Air Filter Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 7.1 Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 7.2 Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 8 AMA Analyzer Head Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 8.1 Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 8.2 Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 9 AMA Sample Cell Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 9.1 Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Figure 3-3 System PC Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 9.2 Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 10Agent ID Analyzer Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 10.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 10.2Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 11MGM System Board Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 11.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 11.2Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 12Fast O2 Sensor Assembly Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 12.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Figure 3-4 MGM Module Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Figure 3-5 Pneumatic Tubing Disconnection Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Figure 3-6 Rear Panel Subassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 12.2Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 13CAN PC Board Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 13.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 13.2Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

iii

mgm2.sm.fm/11-00/kaupp

MGM and MGM+ Modules

Service Manual

 

 

14Power Supply Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 14.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 14.2Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 15Solenoid #1 & #3 Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 15.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 15.2Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 16Solenoid #2 Removal/Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 16.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 16.2Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 17NVRAM Transfer Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 17.1Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 17.2Equipment required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 17.3NVRAM Transfer procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Chapter 4: Adjustment / Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

33

1 Recommended Tools & Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 2 Leakage Check Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 3 Pump Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 4 O2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 5 Pump Flow Rate Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 6 Agent Detection and Analysis Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

Figure 4-1 Gas Calibration / Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

Chapter 5: Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 37

1

Agent Measurement Analyzer Head) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . 37

 

Table 5-1 AMA Analyzer Head Problems

37

2

Agent Measurement Analyzer Optical Path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . 37

 

Table 5-2 AMA Optical Path Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . 37

3

Agent Identification Analyzer Head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 38

 

Table 5-3 AIDA Analyzer Head Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 38

4

Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 39

 

Table 5-4 Power Supply Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 39

5

O2 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 39

 

Table 5-5 O2 Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 39

6

Sample Delivery Subsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 41

 

Table 5-6 Sample Delivery Subsystem Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 41

7

Gas Flow Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 41

 

Table 5-7 Gas Flow Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 41

iv

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

mgm2.sm.fm/11-00/kaupp

Service Manual

MGM and MGM+ Modules

8

Bad MGM System Board suspected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . 41

 

Table 5-8 MGM System Board Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . 41

9

MGM System Board Test Points, Jumpers, & Connectors . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . 42

 

Table 5-9 MGM System Board Test Points . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . 42

 

Table 5-10Connector Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . 43

 

Table 5-11Jumpers, Functions, and Settings . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . 44

 

Table 5-12O2 Jumper Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . 44

Chapter 6: Functional Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

1 Pneumatics Leakage Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

2 Pump Flow Rate Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

3 Gas Identification Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

4 Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

4.1 Resistance Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

4.2 Chassis Leakage Current Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Figure 4-2 Leakage Current Tests Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Appendix A: Spare / Exchange Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

Figure A-1 MultiGas Module (Fuel Cell Type Sensor) Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Figure A-2 MultiGas+ Module (Paramagnetic Sensor) Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . 49 Figure A-3 MultilGas Module (Fuel Cell Type Sensor) Cable Interconnections . . . . . . . . . . . . . . . . . 50 Figure A-4 MultilGas+ Module (Paramagnetic Sensor) Cable Interconnections . . . . . . . . . . . . . . . . 51 Figure A-5 MGM / MGM+ Selected Component Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Figure A-6 MGM / MGM+ Cover and Mounting Feet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Figure A-7 MultiGas / MultiGas+ Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

Appendix B: Exploded-View Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

Figure B-1 MultiGas Module - Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 Figure B-2 MultiGas+ Module - Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Figure B-3 MultiGas Module Rear Panel Subassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Figure B-4 MultiGas+ Module Rear Panel Subassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Figure B-5 Agent Measurement Analyzer / Agent Identification Analyzer Subassembly . . . . . . . . . 58 Figure B-6 MGM Top Cover and Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Figure B-7 MGM System Board Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

Appendix C: Functional Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

Appendix D: Disease Prevention . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

1 Siemens Disease Prevention Policy Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

v

mgm2.sm.fm/11-00/kaupp

MGM and MGM+ Modules

Service Manual

 

 

 

2

Know the Facts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . 63

 

2.1

Types of Viruses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 63

 

2.2

Facts About ARC, AIDS, Hepatitis B, and TB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 63

3

Environmental Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 64

4

Protocol for Servicing the MGM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 65

 

4.1

Preparation for Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 65

 

4.2

Precautions During Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 65

 

4.3

After Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 65

 

4.4

Disposing of Pneumatic System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 66

5

Accidental Skin Puncture Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 66

6

Disease Prevention Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 66

Appendix E: Supplemental Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 67

1

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 67

2

Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . 67

vi

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

mgm2.sm.fm/11-00/kaupp

Chapter 1 Introduction

1 Overview

MULTIGAS™ and MULTIGAS+™ Modules (MGM) are free-standing units that perform sidestream measurements of respiratory and anesthetic gases. The modules automatically identify and measure five common anesthetic agents (Isoflurane, Halothane, Enflurane, Sevoflurane, Desflurane), and report the agent detected and its measurement data to the host device (such as an SC 9000, SC 7000, SC9000XL, or SC 8000 Patient Monitor, or KION). The modules also monitor respiratory gases CO2, N2O, and O2, and report measurements to the host as waveforms (except N2O) and parameters.

MULTIGAS and MULTIGAS+ Modules differ only in the way that they measure O2. The basic MULTIGAS Module measures O2 using a galvanic cell, and calculates average inspiratory values for O2 (labeled iO2). The MULTIGAS+ Module Incorporates a faster-acting paramagnetic sensor that provides both inspired and expired O2 measurements (iO2 and etO2). The outward appearance of the modules differs only in the rear view. The O2 galvanic cell is visible on the rear panel of the MULTIGAS Module. The paramagnetic cell is internal in the MULTIGAS+ Module. See Figure 1-1. In this service manual, the term MULTIGAS is used synonymously with MULTIGAS+ unless specifically stated otherwise.

 

Fan Filter

Exhaust Port

Grounding Stud

Fan Filter

Exhaust Port

Grounding Stud

 

(covered)

 

 

(covered)

 

 

 

 

 

Hardware Version

 

 

Hardware Version

 

 

 

Label

 

 

 

 

 

 

 

Label

 

 

 

 

 

 

 

 

 

RS232 Connector

 

 

RS232 Connector

 

 

 

 

 

 

 

 

 

Software Version

 

 

Software Version

 

 

 

Label

 

 

 

 

 

 

 

Label

 

 

 

 

 

 

O2 Cell

O2 Cell

Power Connector

 

 

Power Connector

 

 

Connector

 

 

 

 

 

 

 

MULTIGAS Module

 

 

MULTIGAS+ Module

Figure 1-1 MULTIGAS and MULTIGAS+ Modules - Rear Views

2 Hardware Installation

X12

 

CPS Connectors

 

X12

 

IDS Connectors

CAN

 

MGM connects to the host (monitor) via an INFINITY Device Communication Power Supply (CPS) or INFINITY Docking Station (IDS) or SC 8000 Patient Monitor. A cable connects the RS232 port on the rear of the MGM to the SC 8000 or to X12 on the CPS/IDS. (The IDS must have a MIB Option installed; SC 8000 requires installed Adv Com Option.) See illustrations at left. The host displays parameter and setup information, only while the module is actually connected. When the module is disconnected, all parameters, waveforms, and setup menus remain on the display until the host is powered off. If host is powered on again without MGM connected, gas parameters and waveforms do not reappear.

SC 8000 ADV COM Option

Refer to the User Guide for the software version installed in the monitor, for applicable Technical Data, and for procedures to access the MGM menu structure in the monitor.

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

1

mgm2.sm.fm/11-00/kaupp

MGM and MGM+ Modules

Service Manual

 

 

3 Service Strategy

In light of the state-of-the-art technology used in the manufacture of Siemens' equipment, proprietary nature of the software, and specialized equipment required for replacement of most individual parts, Siemens’ policy is for the MGM to be serviced only to the field-replaceable subassembly level, after expiration of the warranty period. While in the warranty period, an MGM found to be malfunctioning should be returned to the factory for repair or replacement. After expiration of the warranty period, replacement of components other than those listed in Spare / Exchange Parts should be performed only at Siemens service depots.

4 Preventative Maintenance

Siemens recommends that the following preventative maintenance procedures be performed annually.

Warning

All parts of a MultiGas/+ module that come in contact with the patient’s airway (such as all internal and external tubing, water trap and water trap manifold, and filters) may be contaminated. Handle according to the hospital’s procedures and guidelines for handling infectious substances. Also, see Disease Prevention.

Before initiating preventative maintenance procedures, do the following:

With MGM running with host, verify that the reported revision of the software and hardware is up to date in accordance with the Software Compatibility Chart for the INFININTY NETWORK (or with the installed software in the host if the host is operating in standalone mode). If not, the unit can be updated later in this procedure.

Verify status that no errors are flagged. If any errors are flagged, troubleshoot and repair the MGM before completing the following procedure.

1.Turn off power to MGM.

2.Unscrew top cover, and gently remove cover.

3.Inspect and replace the following, if necessary (expected replacement rate of these parts is once per year):

Internal Nafion® Tubing Assy (qty=2)

Room air filter

Pump filter

Internal Bacterial filter (qty=2)

Water trap seals (qty=2)

Fan filter

Water trap

4.Clean and remove any excess dust, etc.

5.If necessary, update software and/or hardware.

6.Power up MGM.

2

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

mgm2.sm.fm/11-00/kaupp

Service Manual

MGM and MGM+ Modules

 

 

7.Perform Leakage Check Procedure.

8.Perform Pump Flow Rate Verification Procedure.

9.Perform Span Verification Procedure.

10.Turn unit off and replace top cover.

11.Power up unit and verify status is okay.

5Recommended Tools & Test Equipment

SC 9000, SC 7000 / SC 9000XL Patient Monitor with CPS or IDS (with installed MIB Option + CAN), or SC 8000 (with installed Adv Com Option), or KION

Appropriate communication cables (from host to MGM).

Siemens Calibration Kit - SVC TOOL MGM/MGM+ CAL KIT, Art. No. 52 07 415 E536U, containing the following:

Calibration gas - contains 3.00% Isoflurane, 5.00% CO2, 40.00% N2O, 52% Oxygen (with a 1% gas concentration accuracy), Siemens Art. No. 57 36 322 E536U.

Gas Regulator

Tubing w/ Luer-lock fittings T-Piece w/ Luer-lock fittings Two-way valve w/ Luer-lock fittings Gas collection bag

Flow meter with a range of minimum 0 - 350 ml/min, accuracy ±5% or better, (Sierra Flow Control Model 822-13-OV1-PV1-V1 calibrated for “standard - l/min” is recommended

Pressure Gauge

Recommended: Setra Digital Pressure Gauge, Model 370 or equiv.

Note: Pressure gauge required only if verifying and/or calibrating the pressure channel. The hospital and/or a local weather station or airport may be able to provide a reading.

Exhaust system (for exhausting calibration gas).

Digital Voltmeter w/ 3½ digit resolution (minimum)

Oscilloscope (optional)

Hand tools:

Medium sized Phillips screwdriver

Medium sized flat head screw driver

Wire cutters

Non - serrated needle nose pliers

Loctite adhesive or equivalent

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

3

mgm2.sm.fm/11-00/kaupp

MGM and MGM+ Modules

Service Manual

 

 

This page intentionally left blank

4

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

mgm2.sm.fm/11-00/kaupp

Chapter 2 Functional Description

1 Introduction

Both THE MULTIGAS and MULTIGAS+ Modules provide a non-dispersive infrared measurement of respiratory and anesthetic gases. The O2 analyzer subsystem of the MULTIGAS module uses an electrochemical fuel cell, and the MULTIGAS+ module uses a paramagnetic cell. Both modules are designed to work with the host monitor through a serial digital interface. The MGM is intended for measuring airway gases of ventilated patients, within the anesthesia workplace, during the induction and maintenance of, and emergence from, general anesthesia.

Breathing

 

 

Circuit

 

 

 

 

 

Main System &

Central

 

 

Subsystem Hardware

Processing

 

 

 

(Firmware)

 

 

Pneumatic System

Host Unit

 

 

Communications

 

Airway

Agent Analyzer

Data Processing

 

Oxygen Analyzer

Flow Control

 

Adapter

Agent Identification

 

 

 

Sample Gas

 

 

 

Exhaust

 

Figure 2-1

Functional MGM Block Diagram

 

2 Overall Functionality

The MGM pulls the sample gas off the endotracheal tube of a ventilated patient and leads the sample gas through three analyzer subsystems: the Agent Measurement Analyzer (AMA), the Oxygen (O2) Analyzer, and the Agent Identification Analyzer (AIDA). The computational processing unit in the MGM derives waveform data for CO2, anesthetic agents (one out of Halothane, Enflurane, Isoflurane, Sevoflurane, and Desflurane), and O2, together with airway respiration rate and inspired and end-tidal values for the gases, and also including N2O. The derived data is transmitted to the host system which derives alarms from the received data, displays all the alarms and data, and communicates them to other functional modules in the monitoring system.

3 Method of Operation

The airway gases measurement technique used in the AMA subsystem and the AIDA subsystem are based on the non-dispersive infrared absorption of light by molecular gases.

The airway gases measurement technique used in the oxygen analyzer subsystem of the Anesthetic Gas Subsystem is dependent on the type of O2 transducer used.

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

5

mgm2.sm.fm/11-00/kaupp

 

MGM and MGM+ Modules

Service Manual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 2-2 Anesthetic Gas Subsystem Functional Block Diagram

4 Subassemblies, Modules and Components

This document describes the subassemblies, modules, and components of the MGM, how they are controlled by the central processing unit, how the CPU processes the data received from the analyzer subsystems, and how communication between the MGM and the host system (SC 9000, SC 7000, SC 9000XL, or SC 8000 Patient Monitor) works.

5 Anesthetic Gas Subsystem

Figure 2-2 shows a functional block diagram of the Anesthetic Gas Subsystem, which houses the system board and the following major components:

Agent Measurement Analyzer (AMA)

Agent Identification Analyzer (AIDA)

Oxygen (O2) Analyzer

Pneumatic System

Power Supply

These components are typically built into a metal box whose dimensions, weight, and additional features meet the unique requirements of the SC9000, SC 7000, SC 9000XL, SC 8000 or similar host. Typically it includes a power switch, a power connector, an RS-232 connector, a gas inlet, and an exhaust tube.

6 Main System

The Pneumatic System (consisting of the pump, tubing system, solenoid valves, and flow control components) pulls the gas from the gas inlet through the analyzer subsystems at a well-defined flow rate. The second solenoid valve is used when both the Oxygen (O2) Analyzer and the Agent Identification Analyzer are installed.

6

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

mgm2.sm.fm/11-00/kaupp

Service Manual

MGM and MGM+ Modules

 

 

The Agent Measurement Analyzer determines the concentration of CO2, N2O and one anesthetic agent in the gas sample. The AMA is plumbed “first in line,” so that CO2 data is not distorted and capnographic waveforms can accurately be displayed by the host monitoring system.

The O2 Analyzer determines the oxygen concentration in the gas sample.

The Pressure Transducer measures the differential pressure of the gas contained in the pneumatic system. During a Zero calibration this equals the ambient environmental pressure. This pressure transducer is physically housed in the AMA, but plumbed after the O2 Analyzer.

The Agent Identification Analyzer determines which anesthetic agents, if any, are contained in the gas sample.

The Power Supply provides the Anesthetic Gas Subsystem and all of its components with the power necessary to keep the system working. It operates at an input voltage range of 100 - 240 Vac, and is certified to be in compliance with the applicable requirements of UL544 (Patient Care Equipment), CSA 22.2 No. 234 (Level 3), IEC 601-1 (1988), EN60601, and VDE 0750/5.82.

The Electronics Subsystem, with memory (ROM and RAM), multiplexers, A-D converter, and power line supervision, is responsible for the following functions:

Acquisition and processing of data from, and control of, the AMA

Acquisition and processing of data from the Oxygen Analyzer

Controlling the Pneumatic System

Controlling the communications between the Anesthetic Gas Subsystem and the host monitoring system

Controlling the communications between the Anesthetic Gas Subsystem and the Agent Identification Subsystem.

The MGM/AMA Electronics Subsystem has two communications channels -- one connected to an external RS-232 port and the other connected to the AIDA Electronics Subsystem.

The AIDA Electronics Subsystem, with memory (ROM and RAM), multiplexers, A-D converter, and power line supervision, acquires and processes data from agent identification and controls the AIDA. The only communications channel in the AIDA Electronics Subsystem is the one connected to the MGM/AMA Electronics Subsystem.

Full functionality of the Anesthetic Gas Subsystem is controlled by its firmware.

7 Agent Measurement Analyzer

The proven, known, and widely used technology of non-dispersive infrared gas analysis is used by the AMA in the Anesthetic Gas Subsystem.Figure 2-3 on page 8 is a functional block diagram of this analyzer subsystem.

The infrared light source is constructed of tungsten powder metal which is embedded in an Al2O3 ceramic. This source is electrically heated to an operating temperature of 600°C. Infrared emission from this source is distributed as a blackbody radiator.

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

7

mgm2.sm.fm/11-00/kaupp

MGM and MGM+ Modules

Service Manual

 

 

Metal Cavity

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Infrared

 

Filter Wheel

 

 

 

 

 

Detector

Wideband

 

 

 

 

 

With TE

With IR Optical

Sample Cell

 

 

 

Sample Cell

 

 

 

Cooler

Light Beam

 

Thermistor

Bandpass Filters

Window

 

Window

 

 

Containing

 

 

 

 

 

 

 

 

Infrared

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Preamp

 

 

 

 

 

 

 

 

Circuits

 

 

 

 

Gas

 

 

 

 

Infrared

 

 

Gas

Sample

Gas

 

 

 

Mot or

Cel l

 

 

 

Light

 

Input

 

Output

 

 

 

 

 

 

 

 

Source

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NVRAM

 

Filter Wheel

 

 

 

 

 

 

Memory

 

Cavity Heater

Thermistor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AMA Preamp

 

 

 

 

 

 

 

 

Board Assembly

Figure 2-3 Agent Measurement Analyzer Functional Block Diagram

 

The sample cell is constructed of a stainless steel tube with a reflective

 

inside surface which serves as a light pipe. The sample cell length has been

 

designed to provide an adequate absorption length to obtain the desired

 

signal-to-noise ratio for the weakest anticipated absorption. Sapphire serves

N2O

as the sample cell window material for the two ends of the sample cell.

 

Agent

Reference

CO2

Blank

Filter Wheel

The gas sample to be analyzed enters the sample cell through the gas inlet and leaves it through the gas outlet. While in the cell, the gas sample is penetrated by light from the infrared light (IR) source. This light is filtered by coated optical bandpass filters mounted on the filter wheel (see illustration at left). The attached brushless DC motor spins the filter wheel so that the appropriate filter for each gas type (CO2, N2O, agent) comes into place one after the other. The filter wheel cavity heater maintains the metal cavity at 65°C under control of a thermistor. The wavelengths used are --

4.3µ for CO2

3.6µ for N2O

3.3µ for anesthetic agents

The thermistor attached to the sample cell wall provides a measure for the sample cell temperature. Knowledge of sample pressure and sample temperature is vital to accurately determine gas concentrations in the gas sample. Sample pressure is provided by a pressure transducer housed in the AMA but actually plumbed behind the O2 Analyzer. It is therefore its own gas connection.

The photoresistive lead selenide (PbSe) infrared detector, mounted on the preamplifier board assembly, converts the IR radiation not absorbed by the gas sample to an electrical signal. The transmittance of IR radiation is a measure of the total number of molecules of a given gas in the sample cell. The detector’s output signals are preamplified and consist of a pulse stream, one pulse for each IR filter, corresponding to the fraction of this gas type in the sample. The IR detector temperature is kept at 2°C by a thermoelectric cooler to enhance signal-to-noise ratio.

A calibration mechanism guarantees long-term stable measurements and eliminates filter variations.

8

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

mgm2.sm.fm/11-00/kaupp

Service Manual

MGM and MGM+ Modules

 

 

Sample Cell

Thermopile

Window

Infrared

 

Detectors

 

(4 of 7 illustrated)

 

 

 

 

Specific

 

 

Chopped

 

IR Wavelengths

 

 

 

 

Wideband

Light

Wideband

 

 

Infrared

Sample Cell

 

Light Beam

Chopper

 

Light

Window

 

Containing

 

 

 

 

 

 

Infrared

 

 

 

 

 

 

 

Gas

 

 

 

 

Sample

 

 

 

 

Cell

 

Infrared

 

Motor

 

 

Light

 

 

 

 

 

 

 

Source

 

 

 

 

 

 

 

Gas

IR Optical

 

 

 

Input

Bandpass

 

 

 

 

Filters

 

 

 

Gas

(4 of 7 illustrated)

 

 

 

Output

 

Thermistor

Preamp

Circuits

NVRAM

Memory

AIDA Preamp

Board Assembly

Figure 2-4 Agent Identification Analyzer (AIDA) Functional Block Diagram

8 Agent Identification Analyzer

The agent identification function identifies which of the following anesthetic agents is being used:

one agent out of Isoflurane, Halothane, Enflurane, or

one agent out of Isoflurane, Halothane, Sevoflurane, or

one agent out of Isoflurane, Halothane, Desflurane.

Like the AMA, the AIDA in the Anesthetic Gas Subsystem uses the technology of non-dispersive infrared gas analysis. Figure 2-4 shows a functional block diagram of this analyzer subsystem.

Infrared light from the IR light source (which is identical to the AMA IR light source) is modulated using a rotating chopper wheel driven by a stepping motor which is speed controlled by the Electronics Subsystem.

Narrow band filtering and demodulation techniques greatly enhance the quality of the signal generated in the infrared absorption process.

The sample cell is made of thermoplastic, and has a conical shape and nonreflective walls. The cell window material is silicon.

Seven thermopile IR detectors which do not require cooling, each output an analog signal whose magnitude is inversely proportional to the infrared light absorption at the corresponding frequency. These frequencies are determined by the bandpass filters (4 of 7 illustrated in Figure 2-4) operating in the wavelength region from 10µ to 13µ. The thermistor output is used to compensate for the effect of IR filter temperature changes. The analog signals are directly related to the anesthetic agent gas concentrations in the sample cell.

The IR detector outputs are measured during both chopper wheel phases. Measurements taken when the IR light beam is interrupted provide the dark level reference needed by the signal processing software.

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

9

mgm2.sm.fm/11-00/kaupp

MGM and MGM+ Modules

Service Manual

 

 

9 Oxygen Analyzer

9.1ParamagneticMeasurement O2

9.2Electrochemical O2 Measurement

10Pneumatic System

These signals are amplified, filtered, and digitized by the pre-amplifier on the pre-amplifier board assembly. The digitized waveform is then demodulated by the electronics subsystem to obtain a transmission value for each detector.

The following transmission data is used to obtain the gas concentration values used by the agent identification routine.

The seven preamplified IR detector outputs

The thermistor output

The Zero calibration constants

Factory characterization constants

Gas concentration algorithms

Primary agent ID thresholds

Secondary agent ID thresholds

Primary to secondary agent ID crosstalk factors

A calibration mechanism guarantees long-term stable measurements and eliminates filter variations.

In the MULTIGAS+ Module, the Paramagnetic Oxygen Transducer provides “fast” O2 measurement. In the MULTIGAS Module, the electrochemical type sensor provides O2 measurement with a slower response time. Both the paramagnetic sensor and the electrochemical cell deliver an analog signal linearly proportional to the oxygen concentration in the sample gas.

O2 is paramagnetic, which means that a magnetic field induced in O2 will be in the same direction as, and in greater strength than, the magnetizing field. In the paramagnetic oxygen transducer, O2 is placed in two sealed spheres of a dumb-bell assembly, which is suspended on a spring device in a symmetrical non-uniform magnetic field. The assembly assumes a position away from the most intense part of the field.

Sample gas surrounds the dumb-bell assembly, and when the surrounding gas contains O2, the dumb-bell spheres are pushed further out of the field by the relatively stronger paramagnetic O2. The strength of the torque acting on the dumb-bell is proportional to the paramagnetism of the surrounding gas, and is converted into an analog voltage which is likewise proportional to the oxygen concentration.

The electrochemical O2 analyzer operates like a battery. O2 in the gas sample, in contact with an electrolyte, generates a voltage proportional to the concentration of O2.

The Anesthetic Gas Subsystem includes a gas sampling system which accurately controls the flow rate of gas through the analyzer system. Nafion® tubing, a hygroscopic material made from Teflon and polypropylenesulfonic acid copolymer, is added to the sampling line inside the Anesthetic Gas Subsystem to eliminate residual water. Anesthetic agents, N2O, and CO2 are impermeable to the tubing.

10

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

mgm2.sm.fm/11-00/kaupp

Service Manual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MGM and MGM+ Modules

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 2-5 Pneumatics Block Diagram (excerpt from Figure 2-2.)

As illustrated in Figure 2-5, pneumatic solenoid valves are incorporated in the gas stream to switch between the patient gas stream (normal operation) and room air (during Zero calibration). The selected gas (patient or room air) is directed to the Agent Measurement Analyzer, O2 analyzer, and Agent Identification Analyzer.

A servo controlled pump is attached to the exhaust of the analyzer. The pump generates the flow through the system and pulls the gas from the airway adapter through the analyzers to the exhaust outlet. It also delivers the Zero calibration gas to the sample cells of the analyzer subsystems for the periodic zero procedures, and exhausts the patient’s sample gas, zero calibration and field calibration gases. The pump can be operated at four different flow rates, which are hardware-adjusted during factory calibration of the MGM. See Table 2-1.

Table 2-1 Pump Flow Rates

Flow

Flow

Description

Type

Rate

 

 

 

 

 

 

 

 

 

 

Idle

No Flow

Pump switched off

 

 

 

 

 

Low

120

With Paramagnetic O2

With Electrochemical O2

 

ml/min

analyzer

analyzer

 

 

 

 

 

 

Used for analysis of

Optionally used for

 

 

patient gas samples

analysis of patient gas

 

 

 

samples.

 

 

 

 

High

200

With Paramagnetic O2

With Electrochemical O2

 

ml/min

analyzer

analyzer

 

 

 

 

 

 

Used for purging

Used for normal analysis

 

 

Agent Measurement

of patient gas samples

 

 

and paramagnetic O2

 

 

 

analyzers before and

 

 

 

after zero calibration

 

 

 

 

 

Purge

350 ml/

With Paramagnetic O2

With Electrochemical O2

 

min

analyzer

analyzer

 

 

 

 

 

 

Used for purging the

Used for purging the

 

 

Agent Identification

Agent Measurement and

 

 

Analyzer before and

Agent Identification

 

 

after Zero calibration

Analyzers before and

 

 

 

after Zero calibration

 

 

 

 

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

11

mgm2.sm.fm/11-00/kaupp

MGM and MGM+ Modules

Service Manual

 

 

A flow sensor, consisting of a differential pressure transducer, a dampener, and a flow restrictor, is used to determine, stabilize and limit the flow rate of the sampled gas. The output from the pressure transducer is used in a servo system to control the drive power to the pump. The dampener (a 15cc container) isolates the sample cells of the analyzer subsystems from pulsations, enabling a smooth flow through the system. The flow rate control logic works as hard as necessary to maintain the selected flow rate. A partial occlusion, or an inefficient pump, result in the pump being driven harder. A serious occlusion results in the pump being driven at or near its maximum drive. A sense circuit is then triggered to report an occlusion.

11Self-Test

A power-up self-test is performed to validate the contents of firmware memory (ROM), read/write memory (RAM) and non-volatile memory (NVRAM), and to verify errorless access to these storage devices for read and write operations.

12Calibration

12.1Factory Calibration

12.2Field Span Calibration

12.3 Zero Calibration

In order to guarantee long-time stable measurement performance the MGM must be enabled to cope with three types of conditions that can lead to measurement errors:

Small differences among the components of a subsystem (e.g., caused by limitations in manufacturing precision)

Changes of the physical properties of some components over time (e.g., caused by aging or pollution)

Limitations in the compensation for certain effects (e.g., changes in cell temperature/pressure or cross-gas interference)

Each of these conditions can be handled by an appropriate calibration process performed either during original manufacture, as part of normal preventive maintenance, or during normal use.

During factory calibration, the individual performance of each subsystem unit is measured. Polynomial coefficients are then calculated from these individual response curves and stored in the unit itself. These coefficients are later used to compensate for possible unit-to-unit component differences.

During field Span calibration, accurately known concentrations of each gas of interest are introduced into the AMA and O2 Analyzer sample cells and measured. Differences between the known and the measured values are used to calculate the appropriate coefficients for compensation of these differences.

Field Span calibration of the anesthetic agent, CO2 and N2O channels is typically part of preventive maintenance.

The paramagnetic O2 analyzer typically does not require Span calibration.

When an electrochemical O2 sensor is in use, the O2 channel must periodically be Span calibrated.

During Zero calibration, the analyzer subsystems are purged with room air or nitrogen to eliminate any gas of interest (concentration of these gases is “zero”). Oxygen or nitrogen are convenient “Zero calibration gases” since they do not absorb infrared radiation in the wavelengths used by the AMA and AIDA. Since atmospheric air is composed primarily of oxygen and

12

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

mgm2.sm.fm/11-00/kaupp

Service Manual

MGM and MGM+ Modules

 

 

nitrogen, with small amounts of water and CO2, normal room air is used for zeroing the MGM system. Optionally, the MGM can be Zero calibrated with 100% dry nitrogen.

Since changes of physical properties of subsystem components have the same effect on room air measurement performance, this is used as a reference and therefore to calculate new coefficients which help compensate for such changes. Zero calibration of the MGM analyzer subsystem is requested by the MGM as follows:

12.4Storage of Calibration Data

12.5Calibration of Agent Measurement Analyzer

after certain time intervals

after certain changes in operation

after certain operational failures have been detected.

The Zero calibration process measures the infrared signal strength (transmittance) when no IR absorbing gases are in the sample cell. AMA and AIDA field calibration software compensates for the small absorption from atmospheric CO2.

With one exception, all Zero calibration data is stored in and used from volatile RAM memory. The Zero calibration data calculated by the first successful Zero calibration after is stored in non-volatile NVRAM memory for use after subsequent resets.

All Span calibration data is stored in non-volatile NVRAM memory. Critical data is replicated into a second location in both RAM memory (for immediate use) and NVRAM memory.

Block checksums are used to confirm continued validity of NVRAM and RAM data. Power cycling does not affect this data

Both the AMA and AIDA subsystems have their own NVRAM memory for storing their own calibration data, enabling interchangeability of these subsystems with system boards.

Calibration data for the oxygen analyzer is stored in AMA NVRAM memory.

12.5.1

Factory Calibration

Factory calibration compensates for small differences among the following

 

 

components: pressure transducer, infrared (IR) light source, sample cell

 

 

thermistor, filter heater element, filter cavity thermistor, each of the IR

 

 

 

bandpass filters, IR detector, and thermo-electric cooler. Near the end of

 

 

the manufacturing process, binary gases are used to characterize each

 

 

 

AMA. The characterization process also analyzes individual cross-gas

 

 

 

interference. The last function performed during characterization is to

 

 

 

verify performance by sampling cocktail gases. Each unit ends up with is

 

 

own unique set of response curves, and the ability to accurately report gas

 

 

concentrations based on its individual parts and characteristics.

 

12.5.2

Field Span Calibration

The AMA should be calibrated by trained service personnel once every 12

 

 

months using precision calibration gases. The resulting Span calibration data

 

 

is stored in NVRAM memory. The host system can replace field calibration

 

 

data with the original factory calibration data via software command.

 

12.5.3

Zero Calibration

To maintain the highest gas concentration measurement accuracy

 

 

 

possible, the MGM requests that the host command Zero calibration at the

 

 

following time intervals The Zero calibration is performed automatically,

 

 

requiring no user intervention.

 

 

 

 

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

13

mgm2.sm.fm/11-00/kaupp

MGM and MGM+ Modules Service Manual

:

Zero Calibration

Time Interval

 

 

 

 

1st

8 minutes after power-up or reset

 

 

2nd

15 minutes after power-up or reset

 

 

3rd

30 minutes after power-up or reset

 

 

4th

45 minutes after power-up or reset

 

 

5th

90 minutes after power-up or reset

 

 

next

8 hours after previous calibration

 

 

12.6Agent Identification Analyzer

12.6.1

Factory Calibration

Factory calibration compensates for small differences among the following

 

 

components: infrared (IR) light source, detector thermistor, each of the

 

 

seven IR bandpass filters, and each of the seven IR detectors. Near the end

 

 

of the manufacturing process, binary gases are used to characterize each

 

 

AIDA. The last function performed during characterization is to verify per-

 

 

formance by sampling cocktail gases. Each Analyzer ends up with its own

 

 

unique set of response curves, and the ability to accurately measure and

 

 

identify anesthetic agents based on its individual parts and characteristics.

12.6.2

Field Span Calibration

The function of the AIDA is to measure accurately very low gas

 

 

concentrations, most critically in the range of 0.0% to 0.5% where

 

 

identification thresholds are set. Since field Span calibration would not

 

 

influence the performance of the Analyzer in this very narrow range, none

 

 

is required.

12.6.3

Zero Calibration

As with the AMA, regular Zero calibration is required. The Zero calibration

 

 

process is exactly the same as for the AMA, but the time intervals are

 

 

slightly different. The first Zero calibration is performed automatically

 

 

(without host involvement) 2 minutes after power-up. Other Zero cali-

 

 

brations are requested of the host system as described in Section 12.5.3.

12.7Paramagnetic O2 Analyzer

12.7.1

Factory Calibration

The paramagnetic O2 analyzer is calibrated with potentiometers at the 0%

 

 

and 100% point of its measurement range.

12.7.2

Field Span Calibration

The paramagnetic O2 analyzer may be Span calibrated in the field via Span

 

 

calibration commands using appropriate precision calibration gases.

 

 

The resulting Span calibration data is stored in NVRAM memory. The host

 

 

can replace field calibration data with the original factory calibration data via

 

 

software command.

12.7.3

Field Zero Calibration

Zero calibration of the paramagnetic O2 analyzer is performed with room

 

 

air. This is done every time the AMA is Zero calibrated.

12.8Fuel Cell Type O2 Analyzer Calibration

12.8.1 Factory Calibration

The fuel cell type O2 analyzer does not require “characterization” or factory

 

 

calibration. In the case where a fuel cell type O2 analyzer is installed prior

 

 

to original equipment shipment, field Span and Zero calibration are

 

 

performed.

 

 

 

 

 

 

14

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

 

mgm2.sm.fm/11-00/kaupp

Service Manual

MGM and MGM+ Modules

 

 

12.8.2Field Span and Zero Calibration

12.8.3Field Zero Calibration

The fuel cell type O2 analyzer must be periodically Span calibrated since its output is continually degrading during use. Fuel cell type O2 analyzer Span and Zero calibration is performed using the MGM 2-point Span calibration command as discussed in Section 12.7.2. The resulting O2 calibration data is stored in NVRAM memory.

Separate Zero calibration of the fuel cell type O2 analyzer is not required.

12.9Pneumatic System Calibration

12.9.1

Factory Calibration

The pneumatic system of the Agent Analyzer Subsystem is factory-

 

 

calibrated by performing range adjustments. A flow meter is used to adjust

 

 

the four possible flow rates in which the pump can operate. These values

 

 

are set by appropriately adjusting three potentiometers.

 

 

The specific flow rates listed below are representative of one MGM

 

 

configuration. Other configurations may use slightly different flow rates.

 

 

 

 

 

 

 

 

 

Flow Type

Flow Rate

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Idle

No flow

 

 

 

 

 

 

 

 

 

 

Low

120 ml/min

 

 

 

 

 

 

 

 

 

 

High

200 ml/min.

 

 

 

 

 

 

 

 

 

 

Purge

350 ml/min.

 

 

 

 

 

 

 

12.9.2

Field Flow Rate Calibration

Trained service personnel may perform a field calibration of the pneumatic

 

 

system. A field calibration consists of the same range adjustments done

 

 

during the factory calibration.

 

 

13Software

13.1AMA Data Acquisition

Figure 2-6 on page 16 shows a functional block diagram of MGM software. Each bubble indicates a submodule of that software and represents a functional task that is described in more detail below. Each box indicates a hardware part controlled by the firmware. Figure 2-6 also shows that the AIDA Control and Data Processing submodules run on its own AIDA Electronics Subsystem while all remaining submodules are executed by the MGM/AMA Electronics Subsystem, both shown in Figure 2-6.

The AMA Data Acquisition submodule, physically located on the AMA, acquires IR detector output signal pulses, sample cell pressure, sample cell temperature, filter wheel cavity temperature, and IR detector temperature data from the AMA. Additional data acquired by the submodule includes ambient temperature, pump flow rate, four MGM system board voltage measurements and the output of the O2 analyzer.

These analog signals are digitized by an A-D converter in the MGM/AMA Electronics Subassembly. The submodule stores this raw data in shared RAM memory so that the AMA Signal Processing and the Control submodules can access and further process them.

ASK-T876-02-7600

Siemens Medical Systems, EM-PCS Danvers

15

mgm2.sm.fm/11-00/kaupp

Siemens MULTIGAS Service Manual

MGM and MGM+ Modules

Service Manual

 

 

Figure 2-6

Software Functional Block Diagram

13.2AMA Signal Processing

The AMA Data Acquisition and Control submodules send control signals to the AMA to control its IR source, TE cooler temperature, filter wheel motor speed, and filter wheel heater temperature.

Additional data acquired by the submodule includes ambient temperature, pump flow rate, four MGM system board voltage measurements and the output of the O2 analyzer.

These analog signals are digitized by an A-D converter in the MGM/AMA Electronics Subassembly. The submodule stores this raw data in shared RAM memory so that the AMA Signal Processing and the Control submodules can access and further process them.

The AMA Data Acquisition and Control submodules send control signals to the AMA to control its IR source, TE cooler temperature, filter wheel motor speed, and filter wheel heater temperature.

The Data Acquisition submodule checks the digitized data against A-D converter boundary conditions and issues an A-D limit error to the MGM/ AMA Control submodule if necessary.

The AMA Signal Processing submodule reads the AMA IR detector output data from RAM memory, takes an average of four samples and normalizes this average by multiplying the zero constant offset determined during the last Zero calibration of the AMA. This zero constant is read from the AMA RAM memory. The normalized data is stored in RAM memory from where it is read and further processed by the AMA Gas Concentration Algorithms submodule.

16

Siemens Medical Systems, EM-PCS, Danvers

ASK-T876-02-7600

mgm2.sm.fm/11-00/kaupp

Loading...
+ 54 hidden pages