Page 1
FLUOROSPOT H/HK
Maintenance Protocol
AX
Customer:
Department: Room:
Address:
Contact person: Telephone:
Cust. specific no.: Customer no:
The Maintenance Instructions
RX41-020.101.04.01.02
are required for this protocol
Register 9 English
Print Number: RX41-020.105.04.01.02 Doc. Gen.Date: 05.97
Replaces: n. a.
© Siemens AG 1997
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _ or_ design,_are_ reserved.
Page 2
Maintenance Protocol CS No.: . . . . . . . .
SIEMENS office:
Address:
Region: Country:
Contact person: Telephone:
FSE in charge: Telephone:
Information regarding the maintenance protocol
The maintenance protocol serves as a quality record of the maintenance performed
Maintenance must be performed in the prescribed intervals.
The maintenance results must be entered in the maintenance protocol.
The page numbers in the check list at the end of the protocol refer to the corresponding
pages in the associated maintenance instructions (refer to the cover sheet).
The maintenance protocol must be completely filled out by the FSE in charge.
In other words,
• All fields must be completed. If a field does not apply to the system or if there is no
information to be entered, enter ’n.a. ’ in the field.
• The customer number must be entered in the header of every page (CS No.) so that
every page is assigned to a customer.
• In case of complaints, enter the product identification (IVK = WE) for the component as
well as the type of complaint in the ’Open items’ table provided. Record the open items in
the table with the date and signature. Correction of these open points should also be
documented in this table along with the date and signature. If there are no open points,
draw a line through the entire table and enter the date/initials.
• Enter the values measured during the maintenance in the space / table provided.
• After completing the maintenance, fill out page 3 of the protocol and sign it.
Processing and archiving the maintenance protocol:
The maintenance protocol is considered a record and must be archived. It must be filed
after completion of maintenance in the corresponding Register of the Maintenance binder.
If necessary, a copy should be distributed to the customer.
Siemens AG Register 9 RX41-020.105.04 Page 2 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 3
Maintenance Protocol CS No.: . . . . . . . .
System Status:
System: Ser. No.:
Software Version:
Maintenance contract no.:
Type of contract:
The equipment has no problems
Results of the image quality test
showed no deviations from the
required reference values.
The equipment has minor problems that do not restrict its use.
However they should be corrected as a preventive measure.
Results of the image quality test
showed no deviations from the
required reference values.
The equipment has major problems. For safety reasons, the
equipment may not be used until
the problem has been resolved.
Location
Date
Name of the FSE
Signature
Siemens AG Register 9 RX41-020.105.04 Page 3 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 4
Maintenance Protocol CS No.: . . . . . . . .
Performed tasks
Date:
Performed tasks:
OK not OKn.a.
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
❍❍❍
Siemens AG Register 9 RX41-020.105.04 Page 4 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 5
Maintenance Protocol CS No.: . . . . . . . .
Open points
IVK
(WE)
Component Open points Solved
Explanation of the acronyms:
Abbrev. Explanation
(Date)
FSE
(Date)
SI Safety Inspection
SIE Safety Inspection Electrical Safety
SIM Safety Inspection Mechanical Safety
PM Preventive Maintenance
PMP Preventive Maintenance Preventive Parts Exchange, External Inspection, etc.
PMA Preventive Maintenance Adjustments
PMF Preventive Maintenance Function, Operating-Value Check
Q Quality Check
QIQ Quality Check Image
QSQ Quality Check System
SW Software Maintenance
Siemens AG Register 9 RX41-020.105.04 Page 5 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 6
Maintenance Protocol CS No.: . . . . . . . .
Measurement devices
Please enter all measuring devices used for maintenance in the table.
Measurement Device Type Serial-No. Date of use Calibrated
until
TV Dynamics Test
Luminous intensity meter
Siemens AG Register 9 RX41-020.105.04 Page 6 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 7
Maintenance Protocol CS No.: . . . . . . . .
Image Quality Test
Indirect Dose Rate Control (IQ Chap. 4.1.2)
Full format, all control steps
Prefiltering: 2.1 mm precision radiation filter
Operating Conditions Initial Values Mainten.
+-
Control step Fluoro
Progr.
(nGy/s)
Step 1
Fluoro 1
Step 2
Step 1
Fluoro 2
Step 2
Step 1
Fluoro 3
Step 2
Zoom formats, Fluoro 1, Step 1
I.I. Format / I.I. - 1 kV
Zoom 1
Zoom 2
Zoom 3
Curve Zoom - 1 kV
- 10% mA
kV
mA
kV
mA
kV
mA
kV
mA
kV
mA
kV
mA
- 10% mA
kV
mA
kV
mA
kV
mA
+ 1 kV
+ 10% mA
Act.
Initial Values Mainten.
+-
+ 1 kV
Act
+ 10% mA
Remarks: __________________________________________________________
______________________________________________________________________
Siemens AG Register 9 RX41-020.105.04 Page 7 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 8
Maintenance Protocol CS No.: . . . . . . . .
Indirect Dose Control (IQ Chap. 8.1.) DR
Dose kV F/s Scene Focus
100 70 1 10s
I.I. For-
mat (cm)
Zoom
Dose fac.
Initial Values Qg/n Mainten.
- Act +
- 15% + 15%
DSA
Dose kV F/s Scene Focus
500 70 0.5 10s
I.I. For-
mat (cm)
Remarks: __________________________________________________________
_______________________________________________________________________
Monitor Adjustment (IQ Chap.8.3)
DFR Test Images
Zoom
Dose Fac.
Initial Values Qg/n Mainten.
- Act +
- 15% + 15%
Monitor Initial Values
Luminous intensity
Live
Examination
room
Ref
5% / 95% fields visible?
Live
Control room
Ref
5% / 95% fields visible?
On SIMOMED monitors, switch off the ambient light sensor for
the measurement!
2
(cd/m
-
black 0.7 0.9
white 240 280
black 0.7 0.9
white
240
)
Act +
280
yes
no
black 0.7 0.9
white 240 280
black 0.7 0.9
white 240 280
yes
no
Mainten.
yes
no
yes
no
Remarks: __________________________________________________________
_______________________________________________________________________
Siemens AG Register 9 RX41-020.105.04 Page 8 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 9
Maintenance Protocol CS No.: . . . . . . . .
Dynamics Test with Fluoroscopy (IQ Chap. 5.8.1) with VIDEOMED SX
B/BA-Signal Amplitudes
Focus-I.I. distance (SID) = _______ cm - with grid
I.I. format 22 cm; 23 cm; 27 cm; 28 cm - Dynamics test (close to I.I.)
- without heart contour collimator
- without capillary test
- B-Signal without bias light
Fluoro curve _______ = _______
B-/BA-Signal (mV) Field No. Initial Values (mV) Mainten.
µGy/s
-
B-Signal
Input
TV CCU
Fluoro dyn.
BA-Signal
Output
TV CCU
Fluoro kV __ __
2L __ __
16 5 9 5
5R __ __
2L
13 0 0 4 0 0
5R
≥ 1300
__ __
Act
*1 Procedure with SIREGRAPH D2/D3/CF
+
__
Dynamics factor (f) = level/field 2L: level/field 1
Remarks: __________________________________________________________
_______________________________________________________________________
Siemens AG Register 9 RX41-020.105.04 Page 9 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 10
Maintenance Protocol CS No.: . . . . . . . .
Dynamics Test with DR and DSA (IQ Chap.8.5)
Do not select any additional filtering in the collimator
DR: Dose 100, single image Focus-I.I. distance (SID) = _______ cm
Iris corr. values_______ - I.I. format 22 cm; 23 cm; 27 cm
or 28 cm
- Dynamics test (close to I.I.)
- with heart contour collimator
DSA: Dose 200, F/s = 6, scene 5s - with capillary test
Iris corr. values_______ - with grid
upper window value: W1 = 511
lower window value: W2 = 256
- Measure BA-Signal at DFR output
Amplitudes
Field No. Initial Values (mV) Mainten.
Act
70
70
DR
BA-Signal
(mV)
DSA *
BA-Signal
(mV)
(kV)
U
a
U
(V)
Iris
-
2L 520 620
12 0 0 2 5 0
5R __ __
2L
11 9 0 2 6 0
5R
DR __
DSA __
DR __ __
DSA __ __
__
__ __
Reference values applicable for 70 kV
* Evaluate Native image (next to last image);
If 2L field would be > 650 mV, field 1 < 190 mV is admissible.
+
< 650
__
__
Remarks: __________________________________________________________
_______________________________________________________________________
Siemens AG Register 9 RX41-020.105.04 Page 10 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 11
Maintenance Protocol CS No.: . . . . . . . .
Edge Enhancement (IQ Chap.8.5.2)
DR exposure
Function configured
yes
no
Remarks: __________________________________________________________
_______________________________________________________________________
Mean Value Calculation (GGM) (IQ Chap. 8.5.3)
Dig. Fluoro
Function configured
yes
no
Remarks: __________________________________________________________
Maintenance
Maintenance
_______________________________________________________________________
Siemens AG Register 9 RX41-020.105.04 Page 11 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 12
Maintenance Protocol CS No.: . . . . . . . .
Resolution (IQ Chap. 8.7) and Minimum Contrast (IQ Chap. 8.8)
- 17 µm Cu strips
- With grid,
- Resolution test, Type 41 turned 45°
and collimate to the test size
- Insert 1.2 mm Cu in the collimator
- E (%) = 0; temp. filter = 6%
- Fluoro 1 at fluoro ( 70 kV)
- Dose step 100 with DR ( 70 kV)
- Upper window value: W1 = 511; lower window value: W2 = 256
2
Format
Initial Values (cd/m
(LP/mm)
)
Mainten.
Min. contrast visible?
Full format __
Dig.
Fluoro
1
Min. contrast visible?
Full format __
DR
exposure
Min. contrast visible?
Full format __
Hardcopy
-
__
Zoom 1 __
Zoom 2 __
Zoom 3 __
__
Zoom 1 __
Zoom 2 __
Zoom 3 __
__
Zoom 1 __
Zoom 2 __
Zoom 3 __
Act
yes
yes
yes
+
Remarks: __________________________________________________________
_______________________________________________________________________
Siemens AG Register 9 RX41-020.105.04 Page 12 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 13
Maintenance Protocol CS No.: . . . . . . . .
Check of the DSA Device (IQ Chap. 8.9)
Focus-I.I. distance (SID)= _______ cm
I.I. format 22 cm; 23 cm; 27 cm; 28 cm
Window values for DSA image: brightness W1 = 80; contrast: W2 = 220
- With grid
- E (%) = 0
- Dynamics test with heart contour collimator and capillary test 37 90 180
- Trigger a DSA scene: 6 F/s, Dose 200, scene 5s, 70 kV
3 mm
2 mm
1 mm
Contrast Sensitivity
16 %
4 %
0.75 %
10 %
3 %
2 %
7.5 %
4 %
2.5 %
Initial Values
Maintenance *1
2L 1 5R 2L 1 5R
Tolerance
n+2
nn - 3
-1
LOG amplifier *2
3 mm 16 %
Subtraction *3
Tol . 0
± 1
*1 mark capillaries not visible
*2 Enter gray step of capillary
*3 Enter basic gray value of test field
Remarks: __________________________________________________________
_______________________________________________________________________
Siemens AG Register 9 RX41-020.105.04 Page 13 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 14
Maintenance Protocol CS No.: . . . . . . . .
Image Disturbances (Artifacts) (Chap. 8.10)
DR Mode (Chap. 8.10.1)
Images from
MP
8.2
8.5.1 POST PROCESSING
8.5.1 Ghost images
8.2 / 8.5.1 Contouring
*2 Horizontal jitter
8.2 / 8.5.1 Island formation
8.4.3 Vignetting
Other disturbances: ____________________________________
DSA Mode (IQ Chap. 8.10.2)
Images from
MP
8.2
8.5.1 POST PROCESSING
Type of disturbance, Artifact Initial Values
RAM
Pixel errors/
Type of Disturbance, Artifact Initial Values
RAM
Pixel errors/
Mainten. *1
Mainten. *1
8.9 Artifacts caused by exposure
8.9 Logarithming errors
8.9 / *2 Convergence errors
8.9 / *2 Inhomogeneousness (Hum, Line
Errors)
8.9 / *2 Microphony
8.9 / 8.2 Island formation
8.9 Noise patterns
Other disturbances: ____________________________________
*1 Enter No. 1, 2, 3
1 = No disturbances, artifacts
2 = Slight disturbances, artifacts
3 = Intolerable disturbances, artifacts
*2 Soldering wire cross on middle of I.I.; 0.6 mm Cu prefilter
Siemens AG Register 9 RX41-020.105.04 Page 14 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 15
Maintenance Protocol CS No.: . . . . . . . .
Digital Tomography (IQ Chap. 8.11)
- Tomo test object on tabletop
- Tomo height approx. 11.5 cm
(depending on tomo test object)
- , 0.6 mm Cu
- I.I. format 17 cm (20 cm)
- With grid
- Focus-I.I. distance (SID) = ___________ cm
- Exposure voltage = __________ kV; Dose 100
- Tomo profile
- max. tomo angle: ___________ °
- min. tomo time: ______________ s
Initial Values
Mainten.
* ∆ SHa = SHa - SH (mm)
∆ S
HL
= S
- SH (mm)
HL
Rg (LP/mm)
Shape and
sweep
/ o.k.
/ not o.k.
/ o.k.
Blur
/ not o.k.
*) Explanations
S
S
S
= Displayed tomo height ( ≤ 5 mm)
Ha
= Actual tomo height
H
= Light line height ( ≤ 5 mm)
HL
Rg = Measured resolution ( ≥ 2.0)
Remarks:
_______________________________________________________________________
__________________________________________________________
Siemens AG Register 9 RX41-020.105.04 Page 15 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 16
Maintenance Protocol CS No.: . . . . . . . .
This page intentionally left blank.
Siemens AG Register 9 RX41-020.105.04 Page 16 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
Page 17
Maintenance Protocol CS No.: . . . . . . . .
OK notOKn.a.
Page
1 Requirements 1 - 1
1.1. General Remarks 1 - 1
1.2. Required Documents 1 - 1
1.3. Required Test Equipment and Aids 1 - 2
1.4. Explanation of Symbols 1 - 2
2 VIDEOMED SX 2 - 1
2.1. Safety Inspection 2 - 1
SI Checking the Camera Mount. ❍ ❍ ❍ 2 - 1
SI Checking the Camera Housing
SI Checking the Ground Wire
❍ ❍ ❍ 2 - 1
❍ ❍ ❍ 2 - 1
2.2. Maintenance 2 - 1
PM Camera Optics ❍ ❍ ❍ 2 - 1
PM CCU
❍ ❍ ❍ 2 - 1
3 FLUOROSPOT H/HK 3 - 1
3.1. External Visual Inspection 3 - 1
PM Check the Control Console for Damage. ❍ ❍ ❍ 3 - 1
3.2. Maintenance 3 - 1
PM Opening the Unit ❍ ❍ ❍ 3 - 1
PM Remove dust that has collected on the inside of the unit
3.3. Inspection of the internal head vents 3 - 1
PM Check the fan for proper function. ❍ ❍ ❍ 3 - 1
3.4. Monitors 3 - 1
PM Cleaning the Monitor Screens ❍ ❍ ❍ 3 - 1
3.5. Function Inspection 3 - 2
PM Checking / Read Out the Error Log ❍ ❍ ❍ 3 - 2
❍ ❍ ❍ 3 - 1
4 Image Quality Test 4 - 1
4.1. General Remarks 4 - 1
Q Indirect Dose Rate Control (IQ Chap. 4.1.2) ❍ ❍ ❍ 4 - 2
Q Indirect Dose Control (IQ Chap. 8.1.2)
Q Monitor Adjustment (IQ Chap. 8.3)
Q Dynamics Test (Fluoroscopy) (IQ Chap. 5H, 5.8)
Q Dynamics Test (DFR) (IQ Chap. 8.5).
Q Edge Enhancement (IQ Chap. 8.5.2)
Q Mean Value Evaluation (GGM) (IQ Chap. 8.5.3)
Q Resolution (IQ Chap. 8.7) and Minimum Contrast (IQ Chap. 8.8)
Q Checking the DSA Device (IQ Chap. 8.9)
Q Image Disturbances (Artifacts) (Chap. 8.10)
Q Digital Tomo IQ Chap. 8.11)
Q Check of the Tomo Height
Q Checking the Resolution
4.2. Final Work Steps 4 - 24
Siemens AG Register 9 RX41-020.105.04 Page 17 of 18 FLUOROSPOT H/HK
Medical Engineering Rev. 01 05.97 TD AX 5 Register 9
❍ ❍ ❍ 4 - 3
❍ ❍ ❍ 4 - 5
❍ ❍ ❍ 4 - 6
❍ ❍ ❍ 4 - 9
❍ ❍ ❍ 4 - 12
❍ ❍ ❍ 4 - 12
❍ ❍ ❍ 4 - 13
❍ ❍ ❍ 4 - 14
❍ ❍ ❍ 4 - 17
❍ ❍ ❍ 4 - 20
❍ ❍ ❍ 4 - 21
❍ ❍ ❍ 4 - 23
Page 18
CS No.: . . . . . . . . Maintenance Protocol
OK notOKn.a.
4.2.1 Generator Fluoro Curves 4 - 24
4.2.2 SIMOMED H 4 - 24
4.2.3 Function Checks 4 - 24
Page
5 Changes to Previous Version 5 - 1
FLUOROSPOT H/HK Register 9 RX41-020.105.04 Page 18 of 18 Siemens AG
Register 9 Rev. 01 05.97 TD AX 5 Medical Engineering