Angiography, Fluoroscopic and Radiographic Systems
Siemensstrasse 1
D-91301 Forchheim
Germany
Telephone: +49 9191 18-0
www.siemens.com/medical
FLUOROSPOT Compact Imaging System
FLUOROSPOT Compact
Operator Manual
Imaging System for AXIOM Iconos R200
Software VE22 or higher
www.siemens.com/medical
Usability
BEFORE USING THIS SYSTEM
This system should be used only by employees adequately trained in the use of this equipment.
Before using this system, the operator should be thoroughly acquainted with the instructions for use and safety
recommendations provided in this manual.
Failure to follow the instructions for use and safety recommendations provided in this manual can cause serious
injury to the patient, to the operator or to other persons.
Please observe the system’s Safety Operator Manual.
Important information is given there.
Siemens AG
Wittelsbacherplatz 2
D-80333 Muenchen
Germany
For general notes on safety see the Safety part of the system Operator Manual.
Please observe the safety notes in the following chapters.
Note
Do not use the spacer on the rear side of the computer housing as a handle or use
for carrying.
Image failuresIf no image appears on the monitor or if image quality has deteriorated noticeably,
this can be due to a defect in the system. To check this:
Shut down the system and turn it on again.
– The system performs a self-test when turned on.
Pay attention to error messages during restart.
If the fault remains:
Shut down the system and call Siemens Service.
Caution
If image display quality deteriorated due to a defect.
Risk of error in diagnosis caused by poor image quality
Shut down the system and call Siemens Service.
Image archivingThe FLUOROSPOT Compact offers image archiving via network or CD-R.
Transferring/Archiving images to DICOM on page 159
Caution
The hard disk of the imaging system is not suitable for image archiving.
Risk of data loss due to system failure (e.g. defective hard disk)
Always archive patient and image data on suitable archive media.
Third-party softwareOnly software authorized by Siemens for use with this product may be used.
Caution
Use of non-approved or manipulated software or hardware components
Risk of malfunctions posing a hazard to patients and equipment
Only use software or hardware components approved by Siemens.
Modifications shall be performed only with our expressed written approval.
FLUOROSPOT CompactAX41-060.621.23.01.029 / 236
FLUOROSPOT Compact Imaging System
Safety
Computer virusesThe FLUOROSPOT Compact imaging system must be operated within a ’secured
network’, i.e. the network to which the imaging system is connected must be protected from computer viruses, worms, etc.
Caution
Computer virus infections, computer worms etc.
Risk that the system is not operational anymore
Risk of data loss
Always operate the imaging system within a secured network.
Virus scannerTo protect the FLUOROSPOT Compact imaging system from common computer vi-
ruses, a virus scanner is installed.
If a virus is found, a window pops up.
Note
In this case it is mandatory to call the Siemens Service!
Don’t try to act by yourself!
OptionsThis operator manual describes all system features of all system variants.
Note
The complete system is described with all options and components that have been
released. Possible options have not been specially marked.
Particular options or components may not be available for specific systems.
The quotation text of your order is the sole reference for the functional scope of
your system.
If your system does not have a specific feature, please contact your local sales
representative.
10 / 236AX41-060.621.23.01.02Operator Manual
FLUOROSPOT Compact Imaging System
Safety
Value statementsAll technical data are typical values unless specific tolerances are stated.
Values shown in pictures of the software user interface have no clinical meaning.
Please only set the values preset in the organ programs provided or the values
recommended by experienced application specialists.
IllustrationsAll illustrations of equipment and of the program user interface shown in this Op-
erator Manual are examples only. The functions actually available depend on the
type of system, the installed options, and the current configuration.
Other differences in detail may occur in your system due to constant development
and improvement of the system.
Reproduction of images can cause loss of detail. Pictures in this Operator Manual
do not therefore provide any indication of image quality.
NamesAll names of patients shown in figures are purely fictional. Any similarities with ex-
isting persons are entirely coincidental.
Warnings and hintsIn some cases, the imaging system has to inform you about certain facts. Then, an
information window is displayed.
There are different types:
Question!
The system asks you to decide something.
Information!
The system informs you about a certain fact.
Warning!
The system warns you about possible consequences.
Ignoring the warning may lead to severe consequences.
Operation hint
You tried to perform an action which is not possible in the current state of the
system.
FLUOROSPOT CompactAX41-060.621.23.01.0211 / 236
FLUOROSPOT Compact Imaging System
Safety
12 / 236AX41-060.621.23.01.02Operator Manual
PartFLUOROSPOT Compact Imaging System
ChapterOverview / Operating Elements and Displays
Introduction
The FLUOROSPOT Compact (FLC) you have purchased is a digital imaging system
which opens up new dimensions in digital fluororadiography.
The basic system of the FLUOROSPOT Compact is based on a PC with the
Windows XP® operating system.
With only a basic knowledge of the operation of PCs and Windows® you can quickly learn how to operate the imaging system.
CapabilitiesThe imaging system provides simple use and numerous functions for radiological
diagnostics, such as:
Study administration
Organ program management
Integrated generator control console
Image acquisition and display
1
–DFR
Image postprocessing
– Annotation functions
–Graphic functions
–DSA functions
– Functions for checking the image quality
Documentation (film preview, filming, paper printing, medical report and
DICOM send)
Archiving via network or on CD-R
Acquisition systemThis version of the FLUOROSPOT Compact is suitable for image acquisition and re-
porting with the following acquisition systems:
AXIOM Iconos R200
ApplicationThe FLUOROSPOT Compact imaging system can be used for many types of radio-
logical examinations.
Please also refer to the operator manual of the system.
With the FLUOROSPOT Compact imaging system, digital radiographic images can
be acquired, processed, displayed and managed.
images or series from the image intensifier and the CCD camera
1
DFR = Digital Fluoro Radiography
FLUOROSPOT CompactAX41-060.621.23.01.0213 / 236
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
Overview
Basic system
The FLUOROSPOT Compact digital imaging system consists of the following components:
Computer and camera electronics
Image monitors (TFT LCDs) in the examination room and in the control room
CD drive
Keyboard and mouse
At the camera system:
CCD camera
Using the appropriate inherent software you control the imaging system and perform image analysis.
Imaging system configuration
The FLUOROSPOT Compact imaging system provides the following features:
Patient registration DICOM Worklist (HIS/RIS) for receiving patient data from a clinical network,
MPPS (included in DICOM Worklist)
Barcode reader for entering patient data
Examination Fluoro loop
DSA and ROADMAP
PERI Stepping
Reference image monitors (TFT LCDs) in the examination room and/or in the
control room
Additional disk capacity for storing a larger number of images
CAREPROFILE
CAREPOSITION
CAREVISION
SUPERVISION
Postprocessing Harmonization (DDO), also online
Stenosis quantification
14 / 236AX41-060.621.23.01.02Operator Manual
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
Documentation CD-R recording for archiving your images in various formats
Interface for storing fluoroscopy images and acquisitions onto a video recorder
DICOM Send for transferring images into a clinical network
DICOM Print for transferring images to a hardcopy camera via a clinical net-
work
DICOM Query/Retrieve for retrieving images from a clinical network
Paper printer
Transfer operations are carried out in the background.
HIPAAThe HIPAA (Health Insurance Portability and Accountability Act) feature provides
mechanisms to secure the FLUOROSPOT Compact imaging system, patient data
and images from misuse by unauthorized persons.
Operating elements and displays
Places of operation
The imaging system functions can be operated from the control room as well as
from the examination room:
Via keyboard and mouse on the console from the control room
Via infrared remote control from the examination or control room
Infrared remote control on page 18
FLUOROSPOT CompactAX41-060.621.23.01.0215 / 236
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
The console
FLC console
(1)
(2)
(4)
(3)
(1) Image monitor (LCD)
(2) Keyboard and mouse
(3) Computer with camera electronics
(4) CD-R drive
Keyboard
The imaging system is equipped with a standard PC keyboard.
(1) Function keys
(2) Alphanumeric keys
(3) Cursor control keys
Function overview and key assignment on page 26
(4) Numeric/syngo® keypad
16 / 236AX41-060.621.23.01.02Operator Manual
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
Note
If a syngo® keyboard is installed, the symbols on the keys do not apply.
The numeric/syngo® keypad is without function except the Enter key.
The Enter key has the Return key function.
Mouse
A mouse with two buttons and one wheel is connected to the PC.
(1) With the left mouse button, you execute most actions.
(2) With the right mouse button, you can mark images, for example.
1
(3) By rotating the wheel, you can scroll in the study lists and in the image direc-
tories.
TermsIn this operator manual, various terms are used in connection with the mouse:
Click: Move the mouse in such a way that the mouse pointer on the screen is
pointing to the object in question. Now press the mouse button
1
and release
it again.
Double-click: Like clicking, but you press the mouse button twice in quick suc-
cession.
1
Drag (and drop): Like clicking, but you keep the mouse button pressed while
moving the mouse.
1
1
Where click appears in the text, we always mean the left mouse button
unless the right mouse button is mentioned explicitly.
FLUOROSPOT CompactAX41-060.621.23.01.0217 / 236
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
Infrared remote control
With the infrared remote control, you can control certain functions of the imaging
system remotely in the examination room:
Note
The function of the keys on the infrared remote control depends on the system.
Some keys may not work on your system. Some keys may have been labelled differently. Depending on configuration the following functions are possible:
IR keyShort description
Org - / Org +
CarepositionCAREPOSITION on/off
Paging Show previous image (of a series)
*
Switch to previous/next organ program
Selection of remote organ programs 1 to 7
†
Select previous mask in remask function
PagingShow next image (of a series)
†
Select next mask in remask function
Mon. SplitToggle study / series overview, fullscreen display (discards
LIH or not stored fluoro loop; no function with fluoro loop
replay)
Mark ImageMark /unmark an image
‡
Mon 1 2Copy image to reference image monitor
Store ImageStore a LIH or a fluoro image of a fluoro loop
Sub. / NativeAlternate between subtracted and native image display
RemaskSelect function to change mask image
‡
**
**
RoadmapSelect/deselect ROADMAP
RPreselect label R on the left side of the image
LPreselect label L on the right side of the image
Store ref.Store the current image as a reference image
Recall ref.Recall the current reference image
Replay LoopReplay series: start/stop loop
††
‡‡
††
Store FluoroStore the current fluoro loop
F/s Change frame rate before and during series acquisition
18 / 236AX41-060.621.23.01.02Operator Manual
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
IR keyShort description
Fluoro - /
Fluoro +
Switches to next lower/higher fluoro mode
On systems with SUPERVISION:
Switch between standard fluoroscopy and supervision
On systems with CAREVISION (pulsed fluoroscopy):
Change the pulse frequency for fluoroscopy (including
selection of standard fluoroscopy)
*
not with SIRESKOP CX, ICONOS R200
†
no function with LIH or with fluoro loop replay
‡
also during fluoroscopy
**
only with stored DSA series
††
no function with LIH, or non stored image, or with fluoro loop replay
‡‡
no function with single LIHs or images
Note
During acquisition only the key F/s is enabled.
During fluoroscopy only the keys Fluoro -/+, Mark Image and Store Image are enabled.
Note
If you press a button on the infrared remote control, which function is not applicable in the current operating state of the imaging system, then the symbol
shown is displayed on the upper right-hand corner of the screen. The function will
not be executed.
FLUOROSPOT CompactAX41-060.621.23.01.0219 / 236
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
The application program
Languages
The FLUOROSPOT Compact system can be configured by the Siemens Customer
Service in one of the following languages (user interface and keyboard layout):
English
German
French
Spanish
Russian
Swedish keyboard (for use with the English user interface)
Portuguese (with Spanish keyboard)
Italian
Chinese
The user interface for HIPAA administration is available in English only.
Imaging system user interface
On the image monitor all acquired images are displayed. You can perform all functions required for examination, data entry, image acquisition, postprocessing and
documentation.
20 / 236AX41-060.621.23.01.02Operator Manual
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
(1) Display area for images and lists
(2) Control area
(3) Tabs of the task cards for mode selection
(4) Patient information
(5) System information
(6) Resource information
(7) Display control
(8) Subtask cards for function selection
(9) Indication of radiation parameters
FLUOROSPOT CompactAX41-060.621.23.01.0221 / 236
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
Basic modes / task cards
To simplify operation, the application program has four modes. Each mode is
available on a task card.
Patient mode: study lists, input of patient data and archiving; basic settings
Examination mode: fluoroscopy and acquisition, settings for organ and fluoro
programs
Postprocessing mode: image postprocessing
Documentation mode: hardcopy and DICOM send
To select a mode:
Click on the corresponding task card tab with the mouse.
System and resource information
On the top right-hand side, the following system information is displayed
(example):
(1) Marked images of current study
(2) Images in current study
(3) Images in current series
(4) Free space in study / on disk
(5) Acquisition mode and frame rates for acquisition and fluoroscopy
(6) Selected organ program, fluoro mode and I.I. symbol with zoom stage
22 / 236AX41-060.621.23.01.02Operator Manual
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
Bottom rightOn the bottom right-hand side, the following system information is displayed
(example):
Prior to fluoroscopy/
acquisition
During fluoroscopyAfter fluoroscopyAfter acquisition
(1) Fluoroscopy progress bar
– During fluoroscopy: current to maximum fluoroscopy scene length
– After fluoroscopy: current position in fluoroscopy scene
(2) Prefiltration and dose data
1
(3) Fluoroscopy and acquisition data
(4) Reset fluoroscopy time
Note
A flashing value indicates that the actual value is different from the setting in the
organ program.
Text information in images
In the four corners of an image (in Examination or Postprocessing mode) and on
the film sheet the following information is displayed:
1
not in Patient mode
FLUOROSPOT CompactAX41-060.621.23.01.0223 / 236
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
LeftRight
Top Patient name (last, first name)
Patient ID
Patient’s date of birth
Ward
Request ID
Accession No.
†
Mark symbol
: solid white circle or
none
BottomImage content data:
R/L label or mirror/flip indication
Acquisition data:
kV
mAs
Dose value
I.I. format
Image number Ö page 24
Institution name Ö page 196
Product name: FLUOROSPOT
Referring physician
*
Date of acquisition
Time of acquisition
R/L label
†
Image type
Ö page 24
Image settings:
Tomo height or Peri subseries
F: Filter (edge enhancement)
H: Harmonization
‡
LM: landmark (anatomic back-
ground)
‡
C: Contrast
*
only on film sheet (configurable)
†
not on film sheet
‡
depending on image type (native or subtracted)
Note
You can switch off and on image text by pressing the spacebar.
Image numbering and image types
In the following table you can see the different image types and how they will be
numbered:
B: Brightness
‡
WW:
window width
(only with automatic contrast)
X:/Y:
‡
pixelshift values
24 / 236AX41-060.621.23.01.02Operator Manual
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
Image
number
(left)
#nonesingle image13
#single image of a tomography
#noneimage of a fluoro loop (not stored)
S#SERseries or DSA seriesS5
S#PERIPeristepping or Phlebostepping seriesS5
S#_#SERimage of a series (not subtracted)
SL#_#SERimage of a stored fluoro loop
S#_#-#DSAimage of a DSA series (subtracted)
Image
type
(right)
Short descriptionExample
for image number
single image no. 13
series no. 5
S5_12
series number _ image number
series number _ image number
series number _ image number mask number
series no. 5, image no. 12
SL6_13
stored fluoro loop no. 6, image no. 13
S5_12-4
series no. 5, image no. 12
current mask is image no. 4
noneOPACmax./min. opacification image
(not stored)
PO_S#nonestored max./min. opacification image
PO (for peak opacification)
series number
noneSUMaveraged image (not stored)-
SU_S#nonestored averaged image
SU (for sum) series number
noneLIHLIH image or fluoro loop (not stored)-
LIH#nonestored LIH imageLIH3
SL#nonestored fluoro loopSL4
SL#_#noneimage out of a stored fluoro loopSL4_22
noneREF#reference imageREF6
-
PO_S5
peak opacification image of
series no. 5
SU_S6
averaged image of series no. 6
3rd stored LIH image
4th stored fluoro loop
4th stored fluoro loop, image no. 22
6th reference image
SM_#noneStore Monitor image
or image read from CD
FLUOROSPOT CompactAX41-060.621.23.01.0225 / 236
SM_3
3rd Store Monitor image
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
Image
number
(left)
TW_#nonestored twinview imageTW_2
SUB_#nonestored subtraction imageSUB_1
Image
type
(right)
Short descriptionExample
for image number
2nd twinview image
1st subtraction image
Function overview and key assignment
On the following pages, a short overview is given of the basic functions of the
FLUOROSPOT Compact imaging system.
The function keys (F1 to F12 and other keys on the keyboard) are assigned certain
functions. Depending on the mode of the imaging system, different functions are
assigned to the keys.
General functions
IconKeyShort description
*
Enter
EscCancel selected function
left mouse click and ShiftSelect several objects in sequence
left mouse click and CtrlSelect several objects
*
Enter always operates the default action. If none is set Enter has no function.
Depending on the selected function the default action may be set to OK or Cancel.
OK; accept selected function
Patient mode (study lists)
General functions
IconKeyShort description
-F1Call up the operator manual
-Ctrl + F1Call up the operator manual of the system
*
*
-F5No function
26 / 236AX41-060.621.23.01.02Operator Manual
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
IconKeyShort description
ExaminationF6Switch to Examination mode
PostprocessingF7Switch to Postprocessing mode
DocumentationF8Switch to Documentation mode
-F10No function
F12Emergency registration
+
-Read study from CD /
Write study to CD-R
-Left click: Copy screen contents to reference
image monitor
Right click: Clear reference image monitor
-Retrieve study from archive
(only with Archiving Information list)
-Query and retrieve study from archive
(only with Archiving Information list)
mouse wheel/Scroll columns of study list
-Select previous study
-Select next study
-HomeSelect first entry of study list
-EndSelect last entry of study list
Ctrl ASelect all study list entries
-Ctrl RReset width and order of all study list columns
-Prt ScPrint study list data of selected study or
study list of all entries on paper
-PauseShow status of paper printer
*
in Patient mode only. The operator manual must have been installed by Siemens Service.
*
FLUOROSPOT CompactAX41-060.621.23.01.0227 / 236
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
Patient subtask card
IconKeyShort description
F2Call up the input form to create a new study
with a new patient
F3Add a study to selected patient
F4Modify patient data
-Delete selected study/studies
F9Protect / unprotect study
-Apply filter to study list
-Get worklist / Update HIS/RIS
-F11Update HIS/RIS
-Update storage commitment status
-Close MPPS study
Settings subtask card
IconKeyShort description
-Change general settings
-Enter/leave service mode
-Enable/disable access for remote diagnosis (on
service login dialog)
28 / 236AX41-060.621.23.01.02Operator Manual
FLUOROSPOT Compact Imaging System
Overview / Operating Elements and Displays
IconKeyShort description
Show status dialog
-Display test images and switch to
Postprocessing mode
-Enable IQ (image quality) subtask card and
switch to Postprocessing mode using a selected study
-Create a study statistics
Ctrl F2Logout
*
-Call up audit trail dialog*† for saving the audit
trail to a CD-R or saving/restoring users and
groups to/from a CD-R
-User and groups administration
*
HIPAA
†
only available when logged-in as administrator
Examination mode
General functions
IconKeyShort description
-F1No function
*†
PatientF5Switch to Patient mode
-F6No function
PostprocessingF7Switch to Postprocessing mode
DocumentationF8Switch to Documentation mode
-Fluoroscopy parameters
-Acquisition parameters
-Reset fluoroscopy signal
FLUOROSPOT CompactAX41-060.621.23.01.0229 / 236
Loading...
+ 207 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.