Siemens FIDAMAT 6 Operating Instructions Manual

Operating Instructions Edition 08/2004
gas analysis
1
s
FIDAMAT 6
Gas analyzer for the measurement of total hydrocarbons
Operating instructions
A5E00222135-01 Release 08/2004
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights created by the granting of patents or registration of a design are reserved. Technical data subject to change without notice.
Weitergabe sowie Vervielfältigung dieser Unterlage, Verwertung und Mitteilung ihres Inhaltes nicht gestattet, soweit nicht ausdrücklich zugestanden. Zuwiderhandlungen verpflichten zu Schadenersatz. Alle Rechte vorbehalten, insbesondere für den Fall der Patenterteilung oder GM-Eintragung. Technische Änderungen vorbehalten.
Toute communication ou reproduction de ce document, toute exploitation ou communication de son contenu sont interdites, sauf autorisation expresse. Tout manquement à cette règle est illicite et expose son auteur au versement de dommages et intérêts. Tours nos droits sont réservés pour le cas de la délivrance d'un brevet ou celui de l'enregistrement d'un modèle d'utilité. Modifications techniques sont réservées.
La divulgación y reproducción de este documento asi como el aprovechamiento de su contenido, no están autorizados, a no ser que se obtenga el consentimiento expreso, para ello. Los infractores quedan obligados a la indemnización por daños y perjucios. Se reservan todos los derechos, en particular para el caso de concesion de Patente o de Modelo de Utilidad. Salvo modificaciones ténicas.
ULTRAMAT, OXYMAT, CALOMAT, SIPAN sono marchi registrati di S
IEMENS AG.
Le denominazioni di altri prodotti menzionati in questa documentazione possono essere marchi il cui uso da parte di terzi può violare i diritti di proprietà. Conformemente alla "Legge sulle unità di misura" i dati in pollici valgono soltanto per l'esportazione.
La trasmissione a terzi e la riproduzione di questa documentazione, cosiccome lo sfruttamento del suo contenuto non è permesso, se non autorizzato per iscritto. Le infrazioni comporteranno una richiesta di danni. Tutti i diritti sono riservati, in particolare nel caso di brevetti. Modifiche tecniche possibili.
ULTRAMAT, OXYMAT, CALOMAT, SIPAN son marcas registradas de S
IEMENS AG.
Las otras designaciones que figuran en este documento puenden ser marcas cuya utilización por terceros para sus propios fines puede violar los derechos de los proprietarios de dichas marcas. Conforma a la "Ley sobre las unidades de medida", las dimensiones en pulgadas sólo son válidas para la exportación.
ULTRAMAT, OXYMAT, CALOMAT, SIPAN sont des marques déposées de S
IEMENS AG.
D'autres dénominations utilisées dans ce document peuvent également être des marques déposées dont l'utilisation par des tiers à leurs propres fins peut enfreindre les droits des propriétaires desdites marques.
ULTRAMAT, OXYMAT, CALOMAT, SIPAN sind Marken der S
IEMENS AG.
Die übrigen Bezeichnungen in diesem Handbuch können Marken sein, deren Benutzung durch Dritte für deren Zwecke die Rechte der Inhaber verletzen können. Die Angaben in Zoll (inch) gelten gemäß dem Gesetz über Einheiten im Meßwesen" nur für den Export.
ULTRAMAT, OXYMAT, CALOMAT, SIPAN are S
IEMENS registered trademarks.
A
ll other product or system names are (registered) trademarks
of their respective owners and must be treated accordingly.
A
ccording to the German law on units in measuring
technology, data in inches only apply to devices for export.
SIEMENS AG
A
utomation and Drives Process Instrumentation D-76181 Karlsruhe
Siemens Aktiengesellschaft
© Siemens AG 2004 Subject to modifications without prior notice
Order no. A5E00222135 Printed in Germany
A
G 0804 En 0.05 118 PU
Table of contents
1. INFORMATION FOR THE OWNER ..................................................................................................... 1-1
1.1
INFORMATION FOR OUR CUSTOMERS ............................................................................................. 1-2
1.2 GENERAL COMMENTS................................................................................................................... 1-2
1.3 USING THIS MANUAL ..................................................................................................................... 1-3
1.4 HAZARD INFORMATION ................................................................................................................. 1-3
1.5 USE FOR THE PURPOSE INTENDED ................................................................................................ 1-4
1.6 QUALIFIED PERSONNEL ................................................................................................................ 1-5
1.7 WARRANTY INFORMATION............................................................................................................. 1-5
1.8 DELIVERY INFORMATION ............................................................................................................... 1-6
1.9 STANDARDS AND REGULATIONS .................................................................................................... 1-6
1.10 DECLARATION OF CONFORMITY..................................................................................................... 1-7
2. TECHNICAL DESCRIPTION ............................................................................................................. 2-1
2.1
SCOPE OF APPLICATION ............................................................................................................... 2-2
2.2 DESIGN ....................................................................................................................................... 2-4
2.3 MODE OF OPERATION ................................................................................................................... 2-7
2.4 COMMUNICATIONS ..................................................................................................................... 2-10
2.5 TECHNICAL DATA........................................................................................................................ 2-14
3. INSTALLATION INSTRUCTIONS ....................................................................................................... 3-1
3.1
SAFETY INSTRUCTIONS................................................................................................................. 3-2
3.2 GENERAL INSTALLATION REQUIREMENTS ....................................................................................... 3-3
3.3 GAS CONDITIONING ...................................................................................................................... 3-5
3.4 ELECTRICAL CONNECTIONS .......................................................................................................... 3-6
3.4.1 POWER SUPPLY....................................................................................................................... 3-6
3.4.2 CONNECTING THE SIGNAL LINES ............................................................................................... 3-6
3.4.3 CIRCUIT DIAGRAMS (ELECTRICAL CONNECTIONS)....................................................................... 3-8
3.5 DIMENSION DRAWINGS ............................................................................................................... 3-11
4. START-UP.................................................................................................................................... 4-1
4.1
SAFETY INSTRUCTIONS................................................................................................................. 4-2
4.2 START-UP PREPARATIONS ............................................................................................................ 4-3
4.3 START-UP AND OPERATION........................................................................................................... 4-5
4.3.1 MEASURING RANGES ............................................................................................................... 4-5
4.3.2 CALIBRATION........................................................................................................................... 4-7
5. OPERATION ................................................................................................................................. 5-1
5.1
GENERAL COMMENTS................................................................................................................... 5-2
5.2 STATUS DIAGRAM......................................................................................................................... 5-7
5.3 OVERVIEW OF OPERATOR FUNCTIONS ........................................................................................... 5-9
5.3.1 DIAGNOSIS............................................................................................................................ 5-10
5.3.2 CALIBRATION......................................................................................................................... 5-11
5.3.3 MEASURING RANGES ............................................................................................................. 5-17
5.3.4 PARAMETERS ........................................................................................................................ 5-20
5.3.5 CONFIGURATION.................................................................................................................... 5-27
6. MAINTENANCE ............................................................................................................................. 6-1
6.1
MAINTENANCE CONCEPT .............................................................................................................. 6-2
6.1.1 PUMP MAINTENANCE................................................................................................................ 6-2
6.1.2 REPLACING THE FILTER PLATE.................................................................................................. 6-2
6.2 REPLACING THE MOTHERBOARD AND OPTION BOARD ..................................................................... 6-3
6.3 CHANGING FUSES ........................................................................................................................ 6-3
6.4 CLEANING THE UNIT ..................................................................................................................... 6-4
6.5 MAINTENANCE REQUEST AND FAULT MESSAGE .............................................................................. 6-4
6.5.1 LIST OF MAINTENANCE REQUESTS ............................................................................................ 6-5
6.5.2 FAULTS................................................................................................................................... 6-7
6.5.3 OTHER FAULTS........................................................................................................................ 6-8
7. LIST OF SPARE PARTS AND RETURNS............................................................................................ 7-1
7.1
INFORMATION ON ORDERING ......................................................................................................... 7-2
7.2 LIST OF SPARE PARTS .................................................................................................................. 7-3
7.3 RETURNS .................................................................................................................................. 7-12
7.4 ABBREVIATIONS ......................................................................................................................... 7-15
7.5 OVERVIEW OF OPERATOR FUNCTIONS ......................................................................................... 7-16
1-1
1. Information for the owner
1.1. Information for our customers
1.2. General comments
1.3. Using this manual
1.4. Hazard information
1.5. Use for the purpose intended
1.6. Qualified personnel
1.7. Warranty information
1.8. Delivery information
1.9. Standards and regulations
1.10. Declaration of conformity
Information for the owner
FIDAMAT 6 gas analyzer
1-2 Operating manual – A5E00222135-01
1.1 Information for our customers
Please read this manual carefully before beginning operation of the sensor! It will provide you with important information and data you need to ensure proper functioning of the sensor and reduce maintenance costs. Following these instructions will help you to operate the device more easily and efficiently, allowing you to achieve reliable measuring results.
1.2 General comments
Before product described in this manual left the factory, it was inspected and found to be in perfect condition as regards safety. To keep it in this state and to ensure its safe and problem-free operation the product should only be used in the manner described by the manufacturer. In addition, proper transportation, storage and installation as well as careful operation and maintenance of the product are vital for ensuring correct and safe operation.
This manual provides the information you will need for using the described product for the purpose for which it is intended. The manual is intended to be used for technically qualified personnel with special training or who have relevant knowledge in the field of automation technology (instrumentation and control technology).
Familiarity with and a technically faultless implementation of the safety information (including warnings) contained in this manual is essential for hazard-free installation and commissioning as also for safe operation and maintenance of the product described. Only qualified personnel possess the special knowledge required not only for a correct interpretation in the individual concrete case of the safety information and warnings provided in this manual for general application but also for putting them into practice.
This manual is a permanent part of the scope of supply even when ordering it separately has been permitted for logistical reasons. Due to the sheer number of technical details it is not possible to cover all versions of the product described nor every conceivable aspect of installation, operation, maintenance and use of the product as part of a system. If you require further information, or should problems be encountered which are not treated in sufficient depth in this document, contact your local Siemens representative for the information you require.
Note
Particularly before the device is used for new applications in the area of research and development, we recommend you contact us to discuss the application in question.
Information for the owner
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 1-3
1.3 Using this manual
This manual provides you with information on using, installing, operating, and maintaining this device.
Pay particular attention to all warnings and notes. Information of this type is set apart from the rest of the text and specially identified by appropriate symbols (see examples on the left). This information provides you with useful tips on how to avoid operating the analyzer incorrectly.
1.4 Hazard information
Please observe the following notes not only for your own personal safety but also to safeguard against damage to the product described as well as any devices connected to it.
Safety information and warnings are given particular emphasis in this manual by means of the various "signal” terms defined below. They apply to both users and maintenance personnel and are intended to help prevent dangers to life and limb, or to health, and also to prevent damage to property. These notes are also marked by warning symbols which reflect the meaning of the accompanying text and may therefore vary from the examples given below. Within the context of this manual and information on the product itself, the terms used are defined as follows:
Danger
means that death or severe injury and / or considerable damage to property will result if the corresponding precautions are not taken.
Warning
means that death or severe injury and / or considerable damage to property could occur if the corresponding precautions are not taken.
Caution
with a warning triangle means that minor injury may result if the corresponding precautions are not taken.
Caution
without a warning triangle means that damage to property may result if the corresponding precautions are not taken.
Information for the owner
FIDAMAT 6 gas analyzer
1-4 Operating manual – A5E00222135-01
Attention
means that an undesirable event or condition may result if the note in question is not observed.
Note
gives important information about the product, about using the product, or about the corresponding part of the manual to which special attention is to be drawn.
1.5 Use for the purpose intended
In this manual 'use for the purpose intended' means that this product may only be used for the application cases described in the catalog and in the technical description (in this regard see also "Technical description” from page 2-1 onwards) and only in conjunction with third-party devices and components recommended or approved by Siemens.
The product described in this manual has been designed, manufactured, inspected and documented in compliance with the relevant safety standards. For this reason, provided both the handling requirements described for planning, installation, operation and maintenance and also safety instructions are observed, under normal circumstances the unit will not be a source of danger with respect to personal health and safety nor with respect to damage to property. The analyzer has been designed in such a way that a safe separation between primary and secondary circuits is ensured. Extra-low voltages which are connected must also be generated by safe separation.
Warning
Once the housing or shock-hazard protection has been removed or the system cabinet opened, certain parts or components of this unit or system which could be carrying a hazardous voltage become accessible. For this reason only appropriately qualified personnel should be permitted such access to this unit. These persons must have a thorough knowledge of all sources of danger and of maintenance activities as described in this operating manual.
Information for the owner
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 1-5
1.6 Qualified personnel
Unqualified tampering with the unit or system, or failure to observe the warnings in the manual or affixed to the unit or system cabinet could result in severe bodily injury and / or damage to property. Only appropriately qualified personnel should therefore be permitted such access to this unit or system. Within the context of the safety information in this manual or on the product itself, qualified personnel is defined as persons who:
as planning and design personnel are familiar with the safety concepts applicable in automation technology, or
as operating personnel have been given instruction on working with automation technology equipment and who are familiar with the operation-related content of this manual, or
as commissioning or maintenance personnel possess the training which enables them to repair automation technology equipment of this kind or who have been authorized to operate, ground or label devices, systems and electrical circuits in accordance with accepted technical safety standards.
1.7 Warranty information
Your attention is expressly directed to the fact that the nature and quality of the product is described exclusively and conclusively in the contract of purchase. The contents of this product documentation do not constitute part of a past or existing agreement, promise, or legal relationship nor are they intended to modify these. All obligations on the part of Siemens originate from the purchase contract in question, which also includes the full and exclusively valid warranty provisions. The contractual provisions relating to liability for defects are neither extended nor in any way restricted by the information contained in the present document.
Information for the owner
FIDAMAT 6 gas analyzer
1-6 Operating manual – A5E00222135-01
1.8 Delivery information
The scope of supply corresponding to the applicable purchase contract is shown in the shipping documents enclosed with the consignment.
When opening the packaging, read and comply with the instructions provided on the packaging material. Make sure the consignment is complete and undamaged. In particular, you should also, if applicable, check the order number on the nameplates against the ordering information.
If possible, you should not discard the packaging material as this could be needed later should anything need to be returned. A returns form is provided on page 7-14 of the section entitled "Spare parts and returns.”
1.9 Standards and regulations
As far as possible the specifications and manufacture of this device were based on the harmonized European standards. Where harmonized European standards were not applied, the standards and regulations of the Federal Republic of Germany will apply (see also "Technical data” on page 2-14). If this product is used outside the area of validity of these standards and regulations, the standards and regulations valid in the owner's country should be observed.
Information for the owner
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 1-7
1.10 Declaration of conformity
Product name, model
FIDAMAT 6E 7MB2421-xxxxx-xxxx
7MB2427-xxxxx-xxxx
is in compliance with the following standard(s) or documents:
- Low-Voltage Equipment Directives (73/23/EEC and 93/68/EEC)
- EMC Directives (89/336/EEC, 91/263/EEC, 92/31/EEC, 93/68/EEC and 93/97/EEC)
- Harmonized standards applied to
all devices EN 61326
EN 61010
In accordance with the aforementioned EC directives, the EC Declarations of Conformity are kept available for the relevant authorities by
Siemens Aktiengesellschaft Automation and Drives A&D PI 2 D-76181 Karlsruhe
If this product is used outside the European Union, the standards and regulations valid in the owner's country must be observed.
Information for the owner
FIDAMAT 6 gas analyzer
1-8 Operating manual – A5E00222135-01
2. Technical description
2.1. Scope of application
2.2. Design
2.3. Mode of operation
2.4. Communications
2.5. Technical data
Technical description
FIDAMAT 6 gas analyzer
2-2 Operating manual – A5E00222135-01
2.1 Scope of application
The FIDAMAT 6 gas analyzer is used for the quantitative determination of hydrocarbons. The analyzer is also suitable for measurements in corrosive and condensing gas mixtures.
The measuring principle is based on flame ionization detection (FID) with a minimum detection level of about 0.1 ppm. During combustion of hydrocarbons in a hydrogen flame, ions are produced which are in turn converted into a current flow by application of an electrical field. The current intensity arising from this is a measure of the number of carbon atoms.
Benefits
The FIDAMAT 6 gas analyzer stands out due to its very wide scope of application. Accordingly, measurement is possible:
With H
2
O vapor concentrations up to
100%
With ultrapure gas applications
With high-boiling components (up to
200°C)
When corrosive gases are present
Unlike other comparable devices, the FIDAMAT 6 has:
A very low transverse gas sensitivity compared to unwanted gasses
Low consumption of combustion air
Low susceptibility to being affected by
oxygen
The device is also equipped with warning and fault messages, for example:
Failure of combustion gas supply
Detector flame extinguished
Malfunctions of pump and filter
and so on
Examples of applications
Environmental protection
Wastewater (in combination with a
stripping unit, determination of hydrocarbon content of liquids)
Measurement in flue gases in accordance with Emissions Prevention Directives and Clean Air Guidelines (13. BlmSchV /
17. BlmSchV and TA-Luft) for these types of fuel: oil, coals, gases and refuse (with TÜV [Technical Inspectorate] approval)
Monitoring at workplaces with respect to maximum allowable concentration of pollutants
Quality surveillance
Process 14
Ultrapure-gas measurement in media
such as O2, CO2, inert gases and cold measurement gases
Measurement of corrosive and condensing gases
Process optimization
Automotive industry (engine
development, vehicle assembly development and certification)
Measurement of emissions (environmental protection)
Locations
Chemical plants
Gas manufacturers (quality surveillance
for ultrapure gases)
Research and development
Cement industry
Paint shops and chemical cleaning
facilities
Refineries (tank farms, wastewater)
Drying plants
Solvent recovery facilities
Pharmaceutical industry
Technical description
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 2-3
Main features
Four measuring ranges freely parametrizable, also with suppressed zero point, all ranges linear
Hydrogen required: H2
Extremely short measuring spans
Electrically isolated measured-value
output 0/ 2/ 4 to 20 mA (negated also)
Automatic or manual range changeover selectable; remote changeover is also possible
Measured values can be retained during calibration
Time constants selectable within wide limits (static / dynamic noise suppression); in other words, the Unit's response time can be adjusted to suit the measurement task on hand
Simple to use due to menu-controlled operation
User operation based on NAMUR recommendations
Short response time
Low long-term drift
Three operator control levels each with
own authorization code to prevent inadvertent or unauthorized operator actions
External pressure pick-ups can be connected to correct process gas pressure fluctuations
Units easy to swap over since electrical connections only need to be disconnected from unit
Wear-free, corrosion-resistant filter
No clogging of measuring-gas capillary
since a quartz capillary is used
Low consumption of combustion air
Digital display of gas concentration, of
measuring gas, combustion air and hydrogen pressure
Response factors meet the minimum requirements of the Clean Air Guidelines and of the German Automotive Industry Working Group
Simplicity of operation due to a numerical membrane keyboard and operator guidance
Parametrizable automatic measuring range calibration
PROFIBUS (DP/ -PA) / AK
Menu and interface are compatible with
the other Series 6 devices (CALOMAT 6, OXYMAT 6, ULTRAMAT 6, OXYMAT 61)
Customer-specific versions of the device, such as:
- Customer acceptance
- TAG plates
- Drift recording
Technical description
FIDAMAT 6 gas analyzer
2-4 Operating manual – A5E00222135-01
2.2 Design
Design of the overall unit
The FIDAMAT 6 consists of two main modules.
a) Analysis module b) Electronics
a) Analyzer section
This consists of a oven in which the following components are installed:
- Measuring gas filter (heated)
- Flame ionization detector (FID)
- Various restrictors
- Two pressure regulators
- Two pressure sensors
- Solenoid valves
- Bulkhead connections 6 mm or ¼"?
The unit has been designed in such a way that interior parts are easily accessible for maintenance. Access is possible both from the top (for maintenance of individual parts) and from the rear (for replacing the measuring-gas filter).
The measuring-gas pump is easily accessible from above. Both the gas connections and the electrical connections are located on the rear of the unit.
b) Electronics module
This consists of:
- Operator control board with display
- Motherboard
- Adapter board
- Option boards: PROFIBUS (DP/ PA), AK,
AUTOCAL
The operator control board is integrated into the front panel. The adapter board accommodates the preamplifier for measured data acquisition and for the control unit.
Design of the FIDAMAT 6 housing
19" rack module with 4 HU for installation in swing frame
19" rack module with 4 HU for installation in cabinets, with telescopic or support rails
Front panel for maintenance access, swings downwards (for example, for laptop connection, RS 485)
Gas path Material
Piping Gas entry Seals Measuring-gas restrictor Reactant-gas restrictors Pump diaphragm Pump head
Stainless steel 1.4571 Stainless steel 1.4571 Graphite
Quartz
Stainless steel 1.4571 PTFE Stainless steel 1.4571
Detector
Nozzle FID housing
Quartz Stainless steel 1.4571
Gas connections for measuring-gas entry (sample) and measuring-gas outlet (exhaust) and also hydrogen and combustion air; pipe diameter 6 mm or ¼" (Swagelok)
Technical description
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 2-5
Display and control panel
Large LCD panel for simultaneous display of:
- Measured value (digital and analog
display)
- Measuring ranges
- Status bar
LCD panel contrast adjustable via menu
Permanent LCD backlighting
Five-digit measured-value display
(decimal point counts as a digit)
Wipe-clean diaphragm keyboard and front panel
Menu-guided operation for parametrization, test functions and calibration
User guidance in plain text
Graphic display of concentration curve;
time intervals can be parametrized
User software in two languages in each case: German/English; English/Spanish; French/English; Spanish/English; Italian/English
Inputs and outputs
Two analog inputs can be configured
One analog output 0/ 2/ 4 - 20 mA
Six binary inputs can be configured as
required for example, for measuring range changeover, processing of external signals from sample preparation
Six relay outputs can be configured as required for example, for faults, maintenance request, limit value alarm, external solenoid valves
Further eight binary inputs and relay outputs can be added in each case for automatic calibration with a maximum of four calibration gases
Communications
ELAN (RS 485) contained in base unit (connection at rear)
Options
AK interface for the automotive industry with expanded functions
Converter to RS 232 (for ELAN)
Converter to TCP/IP Ethernet (for ELAN)
Incorporation in networks via
PROFIBUS DP/PA interface
SIPROM GA software as service and maintenance utility (for ELAN)
Response factors
Substance Response factors
n-butane 1,00
n-propane 1,00
n-heptane 1,00 Cyclohexane 1,08 Isopropanol 0,81 Toluene 1,06 Acetone 0,92 Ethyl acetate 0,76 Isobutyl acetate 0,83 Methane 1,06 Ethane 0,99 n-hexane 1,01 Isooctane 1,04 Acetylene 0,91 Propene 0,84 Methanol 0,87 Ethanol 0,83 Acetic acid 1,13 Methyl acetate 0,67 Benzole 1,01 Ethyl benzole 0,96 p-xylene 1,03 Dichloromethane 1,13 Trichloroethene 1,01 Tetrachloroethene 1,07 Chloroform 0,72 Chlorobenzene 1,15
Technical description
FIDAMAT 6 gas analyzer
2-6 Operating manual –A5E00222135-01
Fig. 2-1: Membrane keyboard and graphics display
Fig. 2-1: Membrane keyboard and graphics display
Technical description
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 2-7
Fig. 2-2: FIDAMAT 6 – Mode of operation
2.3 Mode of operation
The FIDAMAT 6 measures compound-class-specifically not component-specifically. It measures the total of all organic carbons in a measuring gas but giving different weighting to the hydrocarbon molecules.
As a first approximation, the value displayed is proportional to the number of C atoms in the molecule in question. In practice, however, there are deviations from this. The reading deviation for the molecule in question is expressed by the response factor.
The measuring gas is supplied to the FIDAMAT 6 under overpressure or drawn in by the built-in diaphragm pump (optionally via a heated line and an additional filter) and then routed to the flame ionization detector via a non-clogging fused-silica restrictor. Within the detector the hydrocarbons contained in the measuring gas are combusted in an oxy-hydrogen flame. During the combustion process the organic hydrocarbons are ionized.
Technical description
The ions which are released are converted into an ion current by the polarization voltage between two electrodes and then measured using a highly sensitive amplifier.
The current measured is proportional to the number of organic C atoms of the hydrocarbons in the measuring gas.
A pressure regulator keeps the pressure of the hydrogen constant. A mutually adjusted system of pump, capillaries and combustion-air pressure regulator ensures a constant measuring-gas pressure.
After switching on the analyzer, the flame will be ignited as soon as a flame temperature of 165°C has been reached. The pump is started automatically once the flame reaches a temperature of 220°C.
The FIDAMAT 6 sends out various signals in the form of floating contacts:
Maintenance request Measuring-gas flow rate (filter / pump) Fan failure (advance warning for measurement accuracy) The unit continues to measure as normal and the measured value is not affected.
Fault Hydrogen, combustion air and measuring gas pressure, temperature, physical section and pump, fault in the electronics (temperature). The measured value can be affected; under certain circumstances the unit will switch into a different state (the combustion-gas valve is closed). See also "List of maintenance requests” on page 6-5.
Note
Sample gases must be free of dust when they enter the analyzers. Condensate should be avoided. For this reason, it will in most application cases be necessary to use a gas conditioning device suitable for the measurement task in question.
2-8 Operating manual – A5E00222135-01
FIDAMAT 6 gas analyzer
Technical description
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 2-9
Fig. 2-3: Gas circuit (unit with pump)
Technical description
FIDAMAT 6 gas analyzer
2-10 Operating manual – A5E00222135-01
2.4 Communications
All of the gas analyzers in Series 6, ULTRAMAT 6, OXYMAT 6/ 61, CALOMAT 6, ULTRAMAT 23 and also FIDAMAT 6 provide the following communication methods:
ELAN interface (also with SIPROM GA)
PROFIBUS DP / PA
AK interface
ELAN interface
ELAN is a serial interface (RS 485) built in as standard and which allows communication between several analyzers. A maximum of 12 analyzers with no more than four components may be networked together. Connection can be made even without a PC. The following diagram shows how this works.
Fig. 2-4: Typical arrangement for an ELAN network
(RS 232)
No. Key
1 Computer
2
RS-485 / RS-232 converter with RS 485 and RS 232 connecting leads
3 RS 485 bus plug with jumper 4 Analyzer 5 RS 485 cable 6 RS 485 bus plugs 7 RS 485 network 8 9-pole SUB D connector 9 Optional: RS-485 repeater
Interface parameters
Level RS 485 Baud rate 9600 Data bit 8 Stop bit 1 Start bit 1 Parity None No information feedback
Ordering information Order no.
Interface description
C79000-B5200-C176
RS 485-RS 232 converter
C79451-Z1589-U1
SIMATIC cable / bus cable
6XV1 830-OEH10
SIMATIC bus plugs
6ES7 972-OBB11-OXAO
9-pole SUB D connector
6ES7 972-OBB11-OXAO
Repeater
6ES7 972-OAA01-OXAO
For further information, see the ELAN interface description. Order number:
C79000-B5200-C176 German C79000-B5274-C176 English
Technical description
FIDAMAT 6 gas analyzer
Operating manual –A5E00222135-01
2-11
SIPROM GA
SIPROM GA is a software utility especially for service and maintenance tasks. It allows remote operation and monitoring of all functions of the analyzers, either as individual units or networked together.
Functions:
Display and storage of all device data
Remote operation of all device functions
Parameter and configuration settings
Comprehensive diagnostics information
Remote calibration
On-line help
Cyclic saving of measured values
Status on hard disk and exportation to
user programs available on the market
Downloading of new software
Hardware requirements:
PC or laptop, Pentium 133 MHz, RAM 32 MB, CD drive
At least 10 MB free on hard disk
VGA graphics card supported by
Windows
Printer supported by Windows
Unused COM port: RS 232 or for
connection to ELAN network RS 485 / RS 232
For connecting the Ethernet / RS 485 interface converter, a standard 10 Mbit or 100 Mbit network (RJ 45 connection) with TCP/ IP is required. With an RS 485 network the distance must not be greater than 500 m or a repeater will need to be installed.
Software requirements (optional):
Windows 95
Windows 98
Windows NT 4.0
Windows 2000
Windows XP
Ordering information Order no.
SIPROM GA software, German or English selectable at time of installation, consisting of 1 CD with installation instructions, software, product certificate and registration form
S79610-B4014-A1
Firmware retrofit kit for older analyzers:
FIDAMAT 6
- German
- English
- French
- Spanish
- Italian
A5E00223093 A5E00223146 A5E00223149 A5E00223152 A5E00223155
PROFIBUS DP / PA
PROFIBUS DP / PA is the field-bus market leader. All Siemens gas analyzers when equipped with an optional plug-in board (can be retrofitted) have Profibus capability and meet the mandatory "device profile for analysis devices” defined by the PNO (PROFIBUS Nutzer Organisation) (Profibus users organization). Centralized access to the analyzers installed in the system is possible using the SIMATIC PDM operator control utility.
The term "field bus” designates a digital communication system with which decentrally installed field devices in a system are networked together using a single cable and at the same time connected to automation devices or to a process control system.
Due to its high transmission speeds coupled with relatively small data volumes for each device, the PROFIBUS DP variant is used widely in production automation, while PROFIBUS PA takes particular account of properties demanded in process control, such as large quantities of data and deployment in explosion-hazard areas.
The benefits to the user lie is considerable potential for savings in all parts of the system, from engineering and commissioning, to operation and maintenance, and later on to installation expansion.
Technical description
FIDAMAT 6 gas analyzer
2-12 Operating manual – A5E00222135-01
The gas analyzers can be operated from a control system or a separate PC using the SIMATIC PDM (Process Device Manager) operator control utility, a program which runs under Windows XP or Windows 2000 and which can also be incorporated into the SIMATIC PCS 7 process control system. This means that is possible to represent in a clear and logical manner not only the incorporation of the devices in the system but also the complex parametrical structure of the analyzers and operating the units becomes simply a matter of clicking. The PROFIBUS users organization (PNO) is an independent international body representing the interests of a large number of manufacturers and users. Alongside services such as giving advice, training and the certification of devices, its primary duty is the further development, standardization and promotion of PROFIBUS technology. Definition of a mandatory functionality for a class of devices in a profile is a necessary condition for devices from different manufactures behaving in a uniform manner – what is known as "interoperability”. At the end of 1999 a mandatory profile for analysis devices was established down thereby ensuring that all devices in a system with PROFIBUS capability will work together.
This profile defines the functionalities of the analysis devices in a block model. For example, the device block (physical block) describes the measurement procedure, the analyzer and manufacturer names, serial number and operational state (in operation, under maintenance). Different functional blocks contain the execution of specific functions such as measured-value processing and alarm response. The transducer blocks describe the functionality of the measurement procedure itself as well as control of the procedure – for example, preprocessing of a measured value, cross-corrections, curves, measuring ranges and also switching and control operations. Data transmission between bus users is defined in protocols.
Here a distinction is drawn between cyclic and acyclic services. Time-critical data, such as measured values and states, are transmitted using cyclic services. Acyclic services, on the other hand, permit device parameters to be interrogated or modified during operation.
All of the gas analyzers in Series 6 (ULTRAMAT 6, OXYMAT 6/ 61, CALOMAT 6, ULTRAMAT 6 and FIDAMAT 6 and also ULTRAMAT 23) have PROFIBUS capability once fitted with an optional plug-in board (which can also be retrofitted) (see also the electronics spare parts from page 7-3 onward of the spare parts list).
Fig. 2-5: Typical layout of a PROFIBUS system
Technical description
FIDAMAT 6 gas analyzer
Operating manual –A5E00222135-01
2-13
AK interface
AK is an interface which is not a standard integrated part of Series 6 gas analyzers but which can only be used when an additional board is fitted. It can be ordered under this order number: C79451-A3474-B61
Unlike PROFIBUS and ELAN, communication with AK is only possible between a device and a PC and takes place on the master-slave principle. The devices transfers data only upon being so requested by a command telegram, and here it can only process and reply to one command at a time.
Fig. 2-6: Basic structure of an AK interface
The AK menu can be accessed via function 88 and the parameters set.
The benefit to the user lies in its numerous functions which are primarily required in the automotive industry in order to carry out relinearization,
The principle behind the connection technology corresponds to an RS 232 and is shown in the diagram below.
For further information, see the AK interface description.
Order number: C79000-B5200-C188 German
C79000-B5276-C188 English
Technical description
FIDAMAT 6 gas analyzer
2-14 Operating manual – A5E00222135-01
2.5 Technical data
General comments
Measuring ranges
Autoranging
4, can be switched over internally or externally; manual and automatic measuring range changeover possible
Hysteresis, selectable
Smallest possible measuring span
0 - 10 ppm
Threshold of detection 0.1 ppm
Largest possible measuring span
Concentration units (ppm)
99,999 ppm
C
1
, C3, C6, mg C/m3
Measured-value display
Digital concentration reading (5 positions with floating point)
Resolution of the digital display
Position of normal use
0.1% of the measured value
Front panel vertical
Conformity
Oven temperature
CE symbols to EN 50081-1, EN 50082-2
Adjustable between 100-200 °C (depending on application)
Design, housing
Degree of protection IP 20 as per EN 60529
Dimensions
See "Dimensional drawings” diagram
Weight
approx. 23 kg
Electrical features
Auxiliary energy
100 - 120 V AC (nominal consumption range 90 -132 V), 48 - 63 Hz, or
200 - 240 V AC (nominal consumption range 180 - 264 V), 48 - 63 Hz
Power consumption
approx. 150 VA in operation approx. 350 VA during warming-up
phase EMC – resistance to interference (electromagnetic compatibility)
Complies with standard requirements of
NAMUR NE21 (August 1998)
Electrical safety To EN 61010-1,
overvoltage category II
Fuse-protection values 100... 120 V: 4.0 T/ 250
200... 240 V: 2.5 T/ 250
Gas entry conditions
Permitted measuring­gas pressure
- With pump
installed
Atmospheric pressure
Measuring-gas flow rate 18 ... 60 l/h (0.3 to 1 l/min)
Measuring-gas temperature
0 - 200°C
Measuring-gas humidity
< 90% RH (RH = relative humidity)
Consumption figures for the gases
Combustion air
approx. 350 ml/min
Hydrogen approx. 20 ml/min
Sample approx. 1 l/min
Neutral gas / calibration gas
approx. 2 l/min
Time behavior
Warming-up time
At room temperature, approx. 2 - 3 h
Display delay (T90) 2 - 3 s
Attenuation (electrical time constant)
0 -100 s, parametrizable
Dead time (time for flushing out gas path inside unit at 1 l/min)
With filter 2 - 3 s
Time for signal
rocessing within unit
p
< 1 s
Measurement behavior
Output signal fluctuations < 0.75% of smallest measuring
range possible according to nameplate with electronic attenuation constant of 1 s (this corresponds to +/- 0.25% at 2 σ)
Zero-point drift
< 0.5% per month of smallest measuring range possible according
o nameplate
t
Measured-value drift
< 1% per week of measuring span concerned
R epeatability < 1% of measuring span concerned
Linearity deviation
< 1% of measuring span concerned
Technical description
FIDAMAT 6 gas analyzer
Operating manual –A5E00222135-01
2-15
Influencing variables
Ambient temperature
Ambient pressure Sample pressure
< 1% / 10 K with respect to the smallest measuring range possible according to nameplate
Auxiliary energy
< 1% / 50 hPa
Frequency
< 2% of measuring span / 1% pressure change
Influence due to position < 1% with change of output signal
span of ±10%
< 1%
< 1% with inclination <15°
Electrical inputs and outputs
Analog output 0/ 2/ 4 to 20 mA, floating, output
load max. 750
Relay outputs 6, with changeover contacts,
parametrizable, for example for measuring range identification; output loading capability: AC / DC 24 V / 1 A floating
Analog inputs 2, designed for 0/ 2/ 4 to 20 mA for
external pressure pick-up and accompanying gas correction (transverse gas correction)
Binary inputs 6, designed for 24 V , floating, freely
paramerizable, e. g. for autoranging)
Serial interface ELAN (RS 485)
Options AUTOCAL functions with each
8 additional binary inputs and relay outputs, also for PROFIBUS PA or PROFIBUS DP
Climatic conditions
Permitted ambient temperature
+5 to +45 °C in operation
-30 °C to +70 °C in storage or during transportation
Permitted humidity <90 RH (RH = relative humidity)
annual mean, during storage and transportation (no falling below dew point)
Unit with pump
Gases
Input pressure
hPa (rel.)
Operating pressure
hPa (rel.)
Flow rate
through
FID
ml/ min.
Flow rate
through
bypass
ml/ min.
Without pump
start-up
With pump
start-up
Hydrogen 2.000 – 4.000 1 ± 20 ~ 25 ---
Combustion air 2.000 – 4.000 420 ± 20 500 ~ 320 ~ 500
Sample atmosphere --- 500 ± 2 ~ 3 ~ 1000
Zero gas 2.500 – 3.000 --- 500 ± 2 ~ 3 ~ 1000
Turning gas 2.500 – 3.000 --- 500 ± 2 ~ 3 ~ 1000
FIDAMAT 6 gas analyzer
Operating manual – A5E00222135-01
2-16
3. Installation instructions
3. Installation instructions
3.1. Safety instructions
3.2. General installation requirements
3.3. Gas conditioning
3.4. Electrical connections
3.4.1. Power supply
3.4.2. Connecting the signal lines
3.4.3. Circuit diagrams (electrical connections)
3.4.3.1. Pin assignments on the motherboard (default)
3.4.3.2. Pin assignments on the optional plug-in board and PROFIBUS
3.4.3.3. ELAN interface line
3.5. Dimension drawings
Installation instructions
FIDAMAT 6 gas analyzer
3-2 Operating manual –A5E00222135-01
3.1 Safety instructions
Warning
It is essential that you observe the following information and instructions!
Hazardous voltages
Certain parts of this device are carrying hazardous voltages. Before the unit is switched on, the housing must be closed and grounded. Failure to comply may result in death, physical injury or damage to property. In this regard you should also read and comply with the section entitled "Electrical connection” on page 3-5.
Poisoning hazard
When toxic or corrosive gases are being analyzed, the exhaust gases must be taken off in such a way that persons present are not exposed to any danger. Exhaust gas lines must be made of material which is resistant to the sample (see also "Design of housing and FID” on page 2-
4). Failure to comply may result in serious symptoms of poisoning or even death.
Explosion hazard!
Use of the FIDAMAT 6 gas analyzer for measuring operationally explosive mixtures is not permitted. If the unit is to be used for measuring flammable gases which in combination with air or oxygen could form an explosive mixture, special explosion-protection measures will need to be taken.
Combustion hazard
During operation the interior of the unit's oven reaches a temperature of around 200°C. With the aid of function 2 not only the actual temperature of the oven but also any other temperature setting may be accessed. Once the unit has been switched off, temperatures fall only at a very slow rate. This is why the oven is not accessible to operating personnel during normal operation. Protective gloves must always be worn during maintenance work. Note too that hydrogen burns colorlessly. Failure to observe this may result in serious burns.
Flushing the housing
In accordance with details in the technical data starting on page2-14, there may be what is to be regarded as a limited escape of flammable components from leaks in the sample path. With the FIDAMAT 6 there is no need to flush the housing provided it has been ensured that there is a natural exchange of air from the environment.
Note
However, this does not apply to toxic or corrosive gases! In this regard the maximum allowable workplace concentration figures must be observed.
Installation instructions
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 3-3
3.2 General installation requirements
Installation conditions A location as free as possible of vibration should be selected for
installation.
Make sure that the permissible ambient temperature of 5°C – 45°C is observed during operation (see also "Technical data” on page 2-14). You should also ensure that the unit is not exposed to direct sun light.
If the FIDAMAT 6 is installed in a cabinet or desktop casing, it must be accommodated on support or telescopic rails. It can also be fitted into a swing frame. Mounting the unit by its front end is not satisfactory since the unit's own weight will place too great a load on the chassis.
Due to the unit's great weight the mountings will need to be sized correctly. Use all four screws to fix the housing in place.
Ventilation
Although the FIDAMAT 6 has an internal fan you still need to ensure
adequate ventilation. If several devices are installed in a single cabinet, the FIDAMAT 6 should be fitted as the highest-positioned device due to the great amount of heat it gives off. If this is not possible, a space of about 4 - 5 cm (1 height unit) must be left above the unit.
Gas connections Bulkhead connectors with an external pipe diameter of 6 mm or ¼"
(Swagelok) are provided as gas connections. Gas connection assignments are given in both the gas circuit diagrams and the identification labeling on the rear panel of the unit.
Electrical connections All connecting cables (apart from the mains power line) must be shielded.
The connected cables are shielded via the plug-in connectors. The core cross-section should be 0.5 mm². Cables of the type JE- LiYCY...BD are suggested. The line length of the analog outputs depends on the load.
Caution
When installing the wiring please observe that the device is emitting heat. Therefore there must be a minimum distance of >2 cm between wiring and device.
Maintenance In order to ensure that the unit is not only electrically safe but also is
functionally reliable, the FIDAMAT 6 should be given an annual servicing. In particular, the gas path inside the unit (containment system) must be checked for leaks. After each servicing of the containment system, a leak test must be carried out (in this regard see "Checking for leaks” on page 4-3).
Installation instructions
FIDAMAT 6 gas analyzer
3-4 Operating manual –A5E00222135-01
Sample gas line For gas entry and exit there is a directly heated entry (sample) and a
directly heated outlet. Separate heated sample supply lines each require their own energy supply and temperature regulator. The hexagon of the bulkhead connectors is seated in an anti-rotation lock. When tightening up the union nuts on the connectors make absolutely certain that they are properly locked using a suitable wrench. Failure to do so will mean the risk of the gas line becoming leaky.
The sample is drawn in by a diaphragm pump (approx. 1 l/min) and if at all possible should be made available at atmospheric pressure. Notwithstanding this, the analyzer can also be operated at a constant over- or underpressure of up to 100 hPa.
Exhaust gas line
The greater part of the sample delivered by the pump is routed via a so-
called bypass to the exhaust gas outlet. If the exhaust gases are flammable, toxic or otherwise injurious to health, it is essential that they are taken off using the appropriate measures.
If the exhaust gas is to discharge into a collecting main the following points should be noted:
The flow resistance in the collecting main should be kept low by means of either the shortest length of pipe possible or a transition to a larger diameter.
The exhaust gas line must not be subject to rapid pressure fluctuations. If this is not the case, you will need to put in a separate exhaust gas line or fit a damping vessel with throttle value (>1 l) between the analyzer and the exhaust gas line (pneumatic low-pass filter).
To prevent blockages of the exhaust gas line (for example, due to condensed water) and thereby damage to the unit due to water formation, all parts of the line must have a downward gradient. To this end, it is advisable to install a siphon as a water trap.
Fig. 3-1: FIDAMAT 6, gas and electrical connections
Installation instructions
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 3-5
Sample path within the unit (containment system)
Sample-wetted materials must be resistant to the sample. Materials used
for the seals are quartz, stainless steel 1.4571, graphite and / or PTFE.
Regular inspections for leaks must be carried out to check the integrity of the containment system. If the result is negative, the cause of the leaks must be corrected before you can use the analyzer. A possible remedy in this case is to fit new seals.
After each servicing of the containment system, a leak test must be carried out (in this regard see "Checking for leaks” on page 4-3).
Pressure sensors The FIDAMAT 6 has two internal pressure sensors for measuring the
sample pressure (= combustion air pressure) and the hydrogen pressure.
Maintenance Both the electrical safety and the serviceability of all unit components
must be checked regularly. For this reason the FIDAMAT 6 needs an annual servicing.
The servicing frequency can be reduced in individual cases on the basis of the owner's judgment if no negative effects on the sample-wetted seals are to be expected from chemical attack.
The operator panel (viewing window and keyboard) should only be cleaned by wiping with a damp cloth.
3.3 Gas conditioning
Sample The sample is drawn in through a sample filter heated to 200°C in order to
refilter the sample gases. The filter is made of sintered metal (CrNi steel), has a pore size of 3 µm and a surface area of 3 cm².
Note
Sample gases must be free of dust when they enter the analyzers. Condensate should be avoided. For this reason, it will in most application cases be necessary to use a gas conditioning device suitable for the measurement task in question.
Supply gas
The supply gases (hydrogen, combustion air) must have a purity level of
5.0 if correct measurement is to be assured. The purity level will need to be increased in the case of very low concentrations of hydrocarbons (<1 ppm). The supply gases must be made available with a input pressure between 2000 and 4000 hPa rel. The pressure of zero gas and test gas is around 3000 hPa in each case. Higher pressures will result in increased gas consumption while lower pressure could on the other hand lead to incorrect calibration (calibration) since sample could also be drawn in as well.
Installation instructions
FIDAMAT 6 gas analyzer
3-6 Operating manual –A5E00222135-01
3.4 Electrical connections
Warning
When carrying out electrical installation, you should observe the provisions of VDE 0100 or an equivalent international standard relating to high-voltage systems with nominal voltages below 1000 V.
Failure to comply with these provisions may result in death, physical injury or damage to property.
3.4.1 Power supply
An inlet connector for non-heating apparatus is supplied with the unit and should only be connected to the mains supply line by qualified personnel (see "Qualified personnel” on page 1-5). The mains supply line must satisfy the regulations and requirements applicable at the place of installation and also be provided with a protective conductor at housing potential. The cross-section of each core must be 1 mm². The phase-carrying connection cable must be connected at the marked location in the plug.
The power cable must be routed separately from the signal lines.
A power isolator device must be provided in the immediate vicinity of
the unit (load capability shown on nameplate). It must be easily accessible and have identification labeling.
A check should be made to ensure that the mains voltage present is the same as specified on the nameplate.
Caution
Due to the heat radiation a minimum distance of one inch from the housing must be observed when wiring the device.
3.4.2 Connecting the signal lines
Warning
Signal lines should only be connected to units which are safely electrically isolated from their auxiliary power.
Installation instructions
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 3-7
With the FIDAMAT 6 the signal lines are connected to the SUB D connectors at the rear of the unit.
The connection cables to the relay outputs, the binary inputs and the analog inputs and outputs must be shielded. They should be connected to the corresponding trapezoidal plug (SUB-D plug) as shown in the allocation lists (see "Pin assignments” on page 3-8).
The reference ground of the analog inputs is the housing potential.
The analog output is floating.
The shielding of the interface line must be connected to housing
potential. The cable shielding must have a large-area connection with the shielding of the SUB D plug. The cross-section of the cores should be 0.5 mm². The interface line should not exceed 500 m in length.
All connecting cables (apart from the mains power line) must be shielded.
The cable shielding of the connected lines must be applied over a large area and without any gaps at the respective threaded connec­tions. The core cross-section should be 0.5 mm². Cables of the type JE- LiYCY...BD are suggested. The line length of the analog outputs depends on the load.
Installation instructions
FIDAMAT 6 gas analyzer
3-8 Operating manual –A5E00222135-01
3.4.3 Circuit diagrams (electrical connections)
3.4.3.1 Pin assignments on the motherboard (default)
Fig. 3-2: FIDAMAT 6, 19" rack module, pin assignments
Installation instructions
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 3-9
3.4.3.2 Pin assignments on the optional plug-in board and PROFIBUS
Fig. 3-3: FIDAMAT 6, 19" rack module, pin assignments of AUTOCAL board and PROFIBUS connector
.
Installation instructions
FIDAMAT 6 gas analyzer
3-10 Operating manual –A5E00222135-01
3.4.3.3 ELAN interface line
Specifications of the interface cable
Characteristic impedance
100 to 300 at a measuring frequency of
> 100 kHz Cable capacitance
Typically < 60 pF per meter
Core cross-section
>0.22 mm
2
, corresponds to AWG 23
Cable type twisted pair, 1 x 2 conductors
Signal attenuation max. 9 dB over entire length of line section
Shielding Copper braiding, or braided shield and foil shield
Bus termination resistors
Before bus termination resistors can be connected, in the first and last
connectors of a bus cable pin 3 has to be jumpered with pin 7, and pin 8 with pin 9 (see Fig. 3-4 "Pin assignments”).
Note
If the bus cable is longer than 500 m or there is a high level of interference it is advisable to install a repeater. For further information see function 73 in "ELAN configuration” on page 5-33 or in the description of the ELAN interface.
Order number: C79000-B5200-C176 German
C79000-B5276-C176 English
Fig. 3-4: Pin assignments
Installation instructions
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 3-11
3.5 Dimension drawings
Fig. 3-5: Dimension drawing of the FIDAMAT 6 (dimensions in mm)
Installation instructions
FIDAMAT 6 gas analyzer
3-12 Operating manual –A5E00222135-01
4. Start-up
4. Start-up
4.1. Safety instructions
4.2. Start-up preparations
4.3. Start-up and operation
4.3.1. Measuring ranges
4.3.2. Calibration
Start-up
FIDAMAT 6 gas analyzer
4-2 Operating manual –A5E00222135-01
4.1 Safety instructions
Warning
It is essential that you observe the following information and instructions!
Explosion hazard!
Use of the FIDAMAT 6 gas analyzer for measuring operationally explosive mixtures is not permitted. If the unit is to be used for measuring flammable gases which in combination with air or oxygen could form an explosive mixture, special explosion-protection measures will need to be taken.
Electrical safety
Certain parts of this device are carrying hazardous voltages. Before the FIDAMAT 6 is switched on, the housing must be closed and grounded. Failure to comply may result in death, physical injury or damage to property. For this, please also note "
Electrical connection” on page 3-5.
Flushing the housing
In accordance with details in the technical data, there may be what is to be regarded as a limited escape of flammable components from leaks in the sample path. With the FIDAMAT 6 there is no need to flush the housing provided it has been ensured that there is a natural exchange of air from the environment. Your attention is particularly drawn to this in the case where the analyzer is installed in a cabinet.
Warning
However, this does not apply to toxic or corrosive gases! When working with corrosive or toxic gases the device must be flushed
Start-up
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 4-3
4.2 Start-up preparations
Position of unit The FIDAMAT 6 may only be operated in a vertical position.
Gas conditioning Gas preparation takes place via a heated sample gas filter
(approx. 200°C) with a pore size of 3 µm and with a very limited capacity. For this reason, samples with a high degree of contamination must be prefiltered before entering the analyzer.
Check for leaks A leakage inspection should be conducted after any maintenance activity
affecting the flame ionization detector (FID) or the gas path. This is done as follows:
FID
1. Connect
the sample gas outlet (exhaust gas) to a relative
pressure manometer
(0 - 1000 hPa, resolution 0.1 hPa)
2. Apply an overpressure of about 950 hPa at the sample entry and close off.
3. Wait two minutes or so, until pressure equalization has occurred and then make a note of the pressure.
4.
Wait another minute and then read off the pressure The gas path is satisfactory as far as leaks are concerned when the pressure drop within a minute does not exceed 10 + 2 hPa.
Note
The temperature of the gas path, including the FID, must remain constant while this measurement is being taken.
Test conditions: The pressures given are relative pressures.
Test medium: Synthetic air (sample path must not become contaminated)
Test pressure: Operating pressure x 1.5 (operating pressure = 500 hPa)
Unit electrically switched off:
Test pressure [hPa]
Permissible pressure drop p [hPa]
Sample path
750 + 10 (750 + 10 hPa)
p 10+2 (10 + 2 hPa)
Supply gas inlets
3,000 +100 (3,000 + 100 hPa)
p 12+5 (12 + 5 hPa)
Start-up
FIDAMAT 6 gas analyzer
4-4 Operating manual –A5E00222135-01
Pressure equalization times: time sequence of test
After application of test pressure, observe start and end of measurement for determination of ∆p.
Measurement starts x min after
application of test pressure
Measurement ends y min after
application of test pressure
Sample path
2 3
Supply gas inlets
1 2
Operation Before connecting up and switching on the unit, the operator must have
familiarized himself with how the unit is operated (see on page 5-2).
Interfaces
Before the unit is used, the interfaces must be correctly assigned and
parametrized (see "Connecting the signal lines” on page 3-6).
Noise damping The output signal fluctuations caused by noise in the measurement signal
can be reduced by means of function 50. Among other things, this function allows you to parametrize a low-pas filter which can be assigned a time constant of up to 300 s.
Temperature influence Make sure that the permissible ambient temperature of 5°C – 45°C is
observed during operation (see also "Technical data” on page 2-14).
Pressure influence If measurement is carried out externally, the unit will carry out an
automatic pressure correction. If there are pressure fluctuations within the unit, pressure compensation will be carried out automatically.
Start-up
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 4-5
4.3 Start-up and operation
Check Before switching the unit on, check to see that hydrogen, combustion air, zero
gas and test gas are connected and the required pressures are present.
Switching on mains power Shortly after switch-on, the measured value display appears on the LCD
screen. Above this, on the top line, is the status display (Section 5.1).
Warming up
With the FIDAMAT 6 it is necessary to wait for about one hour for the
warming-up phase to finish. During this time, the oven heats up to 200°C (accessible via function 2 "Diagnostics values”). In contrast, the measuring cell and FID are warmed up with a delay whereby their temperature derives from the setpoint temperature of the oven. Next, the unit automatically checks in turn to see whether the corresponding pressure of the sample and combustion air obtains and whether the hydrogen supply is adequate. If this is the case, the flame is ignited automatically. The flame burns when the flame temperature is 20°C higher than the oven temperature.
Not until the unit has detected that the flame is burning will the pump be switched on.
So as not to block the automatic heating-up procedure, the unit must be in its measurement state. Here the current operational status can be read off from the bottom line.
4.3.1 Measuring ranges
Measuring ranges and measuring spans
The measuring spans desired (initial value of measuring range and final
value of measuring range) should be defined using function 41. Analog current values 0/2/4 mA to 20 mA are assigned to the initial and final values. If the same values are assigned for a measuring range's initial and starting values this measuring range will be deemed non-existent. The smallest measuring span (MS) is to be assigned to measuring range 1, and so on. In general, the following arrangement will apply: MS1 < MS2 < MS3 < MS4
Setpoint value of zero point The setpoint value for the zero point is entered under function 22. It
applies to all measuring ranges.
Calibration of zero point Zero calibration is carried out with the aid of function 20. Here both the
setpoint value and the actual value are displayed.
Start-up
FIDAMAT 6 gas analyzer
4-6 Operating manual –A5E00222135-01
Electronically suppressed zero point If the measuring range initial value (ppm,% vol., etc.) is not at
concentration zero, we speak of measuring ranges with suppressed zero point (for example, 200 - 300 ppm CO). Here 200 ppm is the initial value, 300 ppm the final value, and 100 ppm the measuring span.
Channels with electronic zero-point suppression are physically identical to non-suppressed zero points. Where they differ is solely in the parametri­zation of the measuring ranges and of the setpoint value for the suppressed zero point (200 ppm CO, for example).
The characteristic
curve
is stored from zero to the characteristic final value (as stated on
the nameplate). Units with non-suppressed zero points can be repara-
metrized subsequently to suppressed zero points by modifying this para­meter (functions 22 and 41). It should however be noted that influences such as noise, temperature and pressure errors are magnified by the factor
F =
Final value of smallest measuring range
Final value of smallest measuring range – initial value
Here F should not exceed a value of 7. In general it is recommended to increase the smallest measuring span by about 30%.
Span setpoint value The setpoint value for span should be checked under function 22.
The setpoint value displayed must be the same as the test gas value. If this is not the case, the values must be made to coincide with the aid of function 22.
With a total calibration, the leading measuring range should be selected.
The setpoint values must be as far away as possible from the zero point (at least 60% of the corresponding full-scale value). The corresponding test gases should be made available for the span calibration. The setpoint value is input under function 22.
Changing measurement ranges
For the largest measuring range (according to the nameplate) the
linearized curve is saved in the memory. If the largest measuring range is changed (function 41) this final value must not be exceeded.
Dropping below the smallest measuring range (according to the nameplate) is not permitted as in this case noise and the influence of temperature on the measured value would increase relative to the measuring range and reproducibility and drift behavior would deteriorate.
Start-up
FIDAMAT 6 gas analyzer Operating manual – A5E00222135-01 4-7
4.3.2 Calibration
Zero calibration Zero calibration is carried out with the aid of function 20. Here both the
setpoint value and the actual value are displayed.
Single and total calibration Single calibration means that each measuring range is adjusted with its
own test gas. This is to be recommended when the changeover ratio of the measuring spans exceeds 1:10.
With total calibration only the leading measuring range (function 22) is adjusted; the other measuring ranges are "pulled in” via the changeover ratio.
Total calibration can be activated or deactivated with the aid of function
23.
Next, select function 21 and carry out the calibration desired. In this regard, note that this function cannot be selected unless AUTOCAL is disabled.
Total calibration therefore means that all measuring ranges are adjusted. With single calibration, the procedure described above has to be repeated separately for each measuring range with the corresponding test gases. (Under function 22 a test gas must be specified for each measuring range!)
Calibration of zero point, span Apply zero gas or test gas to the unit at 30 – 90 l/ h (0.5 – 1.5 l/ min).
Note here also the input pressures on page 2-14. Zero point calibration is carried out with the aid of function 20, span calibration with function 21.
Examples of calibration
a) Detection of hydrocarbon in air
Measuring range: 0 - 20 ppm C
1
Test gas: 15 ppm methane in air Zero gas: Synthetic air (free of hydrocarbons)
Procedure Function no. Input Comments
Selection of measuring range initial and final values
41 0 – 20 ppm
Inputting setpoint value for zero points and span
22
Zero point: 0 ppm Recommended: 15 ppm
Note input pressures (see "Technical data” on page 2-14)
Calibration of zero point 20
Span calibration 21
Start-up
FIDAMAT 6 gas analyzer
4-8 Operating manual –A5E00222135-01
b) Detection of hydrocarbon in air
Measuring range: 0 - 50 ppm C
3
Test gas: 90 ppm methane Zero gas: Nitrogen (free of hydrocarbons)
Procedure
Function
no.
Input Comments
Selection of measuring range initial and final values
41 0 – 50 ppm
Inputting setpoint value for zero points and span
22
Zero point: 0 ppm C
3
Recom.: 30 ppm C3
90 ppm C
1
gives 30 ppm C3
Calibration of zero point 20
Span calibration 21
A detailed description of operator and input possibilities for the above­mentioned functions may be found in Chapter 5 "Operation” starting on page 5-1.
5. Operation
5.1. General comments
5.2. Overview of operator functions
5.2.1. Diagnosis
5.2.2. Calibration
5.2.3. Measuring ranges
5.2.4. Parameters
5.2.5. Configuration
Operation
FIDAMAT 6 gas analyzer
5-2 Operating manual –A5E00222135-01
5.1 General comments
Fig. 5-1: Display and control panel
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-3
Switches / keys and their significance
Key Meaning
CLEAR Clears a numerical input which has been started
ENTER
Every numerical input (apart from fast selection of a function) must be confirmed with ENTER
ESC Goes back one step in the operator interface structure. Changes are accepted.
INFO Information about the current menu
MEAS
Return from any position in the operator interface structure in the display mode (if necessary, the interrogation is done before accepting the entered data).
Pressing the MEAS key again will lock the analyzer
in other words, changing back again to operator mode is only possible after inputting the code.
Soft key
Variable significance; possibilities here include:
- Menu item selection in menu tree
- Selection function
Entering inputs The values shown in menus depicted in this section are for illustrative
purposes only.
An active input field is shown with colons (for example: :10:) as
delimiters. Here the cursor appears as a flashing line beneath the number to be input (for example: : 2
3.45 :).
Pressing the ENTER key ends inputting and the value is saved. If
several input fields are present in a menu window, the cursor will simultaneously be positioned over the next input field.
Every value entered has to be confirmed with ENTER before leaving
the menu. This includes the last of multiple values in a menu.
The CLEAR key clears a numerical input which has been started The
cursor then jumps back to the first position of the input field.
Graphic symbols
Switching function (ON state)Switching function (OFF state; also status display on the status
line)
Entry into a subsequent menu
Initiation of a function (for example, Start calibration…)
Operation
FIDAMAT 6 gas analyzer
5-4 Operating manual –A5E00222135-01
Fig. 5-2: Display and operator panel
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-5
Attention
To prevent static charges the keyboard should only be used for maintenance or operating purposes.
Operational sequence
Measured value display (measurement mode)
Entry into the main menu
The measurement component, furnished with a rightward pointing arrow (), appears on the right-hand side of the screen. A soft key is assigned to this component. Pressing this soft key calls the main menu.
The main menu consists of the following sub-sections (sub-menus). To their right is shown the corresponding code level:
Diagnosis Not coded Calibration Code level 1 Measuring ranges Code level 1 Parameters Code level 1 Configuration Code level 2
The code for level 1 is factory-set to "111” and that for level 2 is set to "222.”
Entry into a sub-menu
Once a sub-menu has been selected, you will be asked for the code for
the operator level (exception: the "Diagnosis” submenu is accessible without restrictions).
Decoding level 2 also decodes level 1. With decoding, signaling to the outside can take place via a relay contact provided a corresponding relay has been configured via function 71 with CTRL. The warming-up and calibration phases of the unit are also signaled via this relay contact.
With decoding the measured value memory is also activated, provided it was set under function 77. The coding of the unit is to be seen on the display (Fig.:
Measured value display) as the symbol CODE,
decoding with
CODE.
Operation
FIDAMAT 6 gas analyzer
5-6 Operating manual –A5E00222135-01
Return to measuring mode
From any location in the menu structure, pressing the MEAS key will take you back immediately to measuring mode. Any input you may have started will be aborted.
Before the jump, the question shown on the left will appear.
Pressing the YES or NO soft keys takes you back to measuring mode. YES causes the changes to be saved in the main memory of the parameter memory, NO discards them.
Pressing the ESC key takes you back to the last function screen. Changes are accepted without checking with you first.
Coding the unit
Following a return via ESC or MEAS to measuring mode, the unit can be coded again by pressing MEAS again
(code) and thus returned to measuring mode. All states brought into existence with decoding are thereby cancelled.
Quick selection of functions In the case of frequent operator activity a "power user” feature has been
implemented to allow the user to go directly to the desired function screen from the "Measured value display” screen. This allows direct access to the desired settings by inputting functions. This simply bypasses menu levels. However these functions can only be launched from measuring mode and include the following operator steps:
In measuring mode ("Measured value display” screen), input the number of the desired function using the number key.
Press the soft key with the arrow.
If the desired function is protected by a code, you will be
asked to input the code (see "Entry into the main menu”).
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-7
5.2 Status diagram
Fig. 5-3: Status diagram
Operation
FIDAMAT 6 gas analyzer
5-8 Operating manual –A5E00222135-01
Change of state following error/maintenance request
Fault / maintenance request State before State after
S1: Parameter storage - - ­S2: Pump does not start warm-up / stand by stand-by S3: Flame does not ignite warm-up pause S4: External fault - - ­S5: Oven temperature warm-up / stand-by / measure / pause pause without heating S7: Flame temperature warm-up / stand-by / measure / pause pause S8: Combustion air/gas pressure warm-up / stand-by / measure / pause pause S9: Heating switched off warm-up pause without heating S10: 24h RAM/flash check - - ­S12: Mains power supply warm-up / stand-by / measure / pause pause S13: Hardware / mains frequency warm-up / stand-by / measure / pause pause S14: Measured value > full-scale value - - ­S15: Calibration cancelled - - -
W1: Calibration tolerance exceeded - - ­W4: Set clock - - ­W5: Combustion air/gas pressure stand-by / measure warm-up W6: Temp. sensor LCD display - - ­W7: Electronics / analyzer temperature - - ­W8: Flame out Stand-by / measure warm-up W9: External maintenance request - - ­W10: Autocal check deviation - - -
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-9
5.3 Overview of operator functions
The unit functions are shown below in an overview. This listing corresponds to version 4 of the software.
Subsection of main menu
Function
no.
Name of function
5.2.1. Diagnosis
1 2 3 4
Factory data Diagnostic values Logbook Display measuring ranges
5.2.2. Calibration (Code 1)
20 21 22 23 24
Calibration of zero point Span calibration Setpoint values of zero / span Calibration settings (single and total calibration) AUTOCAL / check (only with option card)
5.2.3 Measuring ranges
5.2.3 (Code 1)
40 41
Select measuring ranges Define measuring ranges
5.2.4. Parameters (Code 1)
50 51 52 53 54 55 56 57 58 59 60 61
Electrical time constants Limits In and out functions Status messages Graphic representation of measured values Measured-value display LCD contrast Ignite flame Date and time Measurement point changeover Logbook setting Switch internal valves
5.2.5 Configuration (Code 2)
70 71 72 73 74 75 76 77 78 79 80 81 83 85 86 87 88
90 91 92 93 99
Analog output Relay assignments Binary inputs ELAN configuration Reset Save or load data Suppression of brief fault signals Analog output memory Calibration tolerances Change codes Unit test Language selection Accompanying gas correction Switch valves Linear temperature compensation Fault on/off AK (only with optional board)
PROFIBUS configuration (only with optional board) Start-up state Pressure values Units Factory settings
Table 5-1: Overview of operator functions
Operation
FIDAMAT 6 gas analyzer
5-10 Operating manual –A5E00222135-01
5.3.1 Diagnosis
When the diagnostic functions are selected in the main menu by pressing the first soft key ("Diagnosis”) the adjacemt display appears.
Access to diagnostic functions is not restricted. For this reason you will not be asked for the code.
Every unit provides the following diagnostic functions:
1 Factory data When this function is selected, important manufacturer's data for the unit
are displayed:
Motherboard:
Firmware no.: Order number of the software held in the EPROM for the motherboard
Order no.: Information about ordering data for the unit
Serial no.: Information about date of manufacture and unit serial number
Article status: Information about hardware configuration of the unit
Software status and date Information about the unit's functional scope at the motherboard
2 Diagnostic values The most important diagnostic values are listed under function 2. They
may allow conclusions to be drawn for evaluating setting activities or errors.
3 Logbook In the logbook will be found a list of all errors which resulted in a
maintenance request (W) or a fault message (S). (see also on page 6-5 "List of maintenance requests”) Limit alarm (LIM) and function check (CTRL) are also recorded. These, however, do not trigger a maintenance request or fault message.
The logbook contains a maximum of eight pages, each of which can hold four messages. It works on the principle of the cyclic buffer – in other words, when all eight pages (all 32 slots) are occupied, the oldest message will be overwritten.
Logbook entries can be deleted or locked (function 60) and even switched off individually (function 87).
Note
If an error occurs and its error message has been switched off with function 87, there will be no reaction at an interface which may have been configured. This applies to the ELAN interface and also to the analog output and relay output.
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-11
In the case of maintenance, function 110 Diagnostic service should also be taken into consideration. This menu is however only accessible to maintenance personnel.
4 Display measurement ranges
The measuring ranges defined under function 41 are listed under
function 4. However, they cannot be modified in this menu.
5.3.2 Calibration
If at all possible, calibration should be carried out using test gases which very closely resemble the sample both in composition and concentration. If the measuring task is to measure hydrocarbons in nitrogen, a hydro­carbons in nitrogen test gas should also be used. If, for example, propane in synthetic air were to be used here, the measurement result would be falsified by the response factor of the methane compared with the propane and by the oxygen error.
The FIDAMAT 6 provides both manual and automatic calibration (AUTOCAL: function 24). The latter is only possible with the aid of an optional board which in addition contains eight binary inputs and eight relay outputs.
Setpoint values for zero-point and span calibration must be set under function 22.
20 Zero-point calibration
Calibration of the zero point is carried out for all measuring ranges simultaneously even if the span of the measuring ranges is being adjusted individually.
The calibration procedure should not therefore be initiated until the measured value (actual value) has stabilized following addition of the zero gas.
If the measured value fails to settle, the time constant (function 50) should be increased before carrying out calibration.
Operation
FIDAMAT 6 gas analyzer
5-12 Operating manual –A5E00222135-01
21 Span calibration of
Single calibration:
The display shows as many measuring ranges as was previously defined via function 41. The adjacemt display therefore represents an example of single calibration of four measuring ranges.
If measuring range 3, for example, is to be adjusted, the corresponding softkey should be pressed.
The setpoint value and the current (ACTUAL) value for measuring range 3 appear on the screen.
Once the actual value has stabilized, the calibration procedure can be started by pressing the fourth soft key. The actual value is now brought into agreement with the setpoint value.
In the case of an inadvertently incorrect calibration (after using the wrong calibration gas, for example), pressing the "ABORT" soft key will reload the original calibration.
Total calibration:
Total calibration means that all measuring ranges are adjusted simultaneously. The "leading” measuring range is defined under function 22 (see also menu 23 "General and single calibration”). It makes sense to select the largest measuring range as the leading range. The setpoint value and the current value for the "leading” measuring range appear on the screen.
Once the actual value has stabilized, the calibration procedure can be started by pressing the fourth soft key. The actual value is now brought into agreement with the setpoint value.
In the case of an inadvertently incorrect calibration (after using the wrong calibration gas, for example), pressing the "ABORT" soft key will reload the original actual value.
22 Setting setpoint values
The adjacent example to the left shows setpoint value input in the case of total calibration. The fourth measuring range has been chosen as the leading measuring range.
In the case of single calibration, it is not possible to specify the leading measuring range.
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-13
23 Total and single calibration:
This function is used for selecting general or single calibration of the measuring ranges.
With total calibration a "leading measuring range” is adjusted and all of the other measuring ranges are "pulled in” via the changeover ratio.
If this function is not activated (see illustration), each measuring range will be adjusted individually.
24 AUTOCAL
Automatic calibration (AUTOCAL) cannot be carried out unless the unit has an additional electronic module (optional board).
If it does not and an AUTOCAL parameter is selected, the corresponding message will be shown on the display.
AUTOCAL / check operating mode
This subfunction allows you to parameterize various AUTOCAL operating modes.
AUTOCAL / Check on / off
When the "AUTOCAL off” state is active (shown as
) it will not be
possible to operate the "AUTOCAL Start by cycle” (parameter) or "AUTOCAL Start by binary input” switches.
Even "AUTOCAL run once” is switched off. The cycle time continues to run. Nevertheless no automatic calibration is initiated.
AUTOCAL / check Start by cycle AUTOCAL can be activated to use a regularly repeating cycle provided
the "Time between two AUTOCALs” has been set. "AUTOCAL – check” is used for checking the calibrations. As with "AUTOCAL”, the sequence parameterized in the "AUTOCAL sequence” menu is carried out. In contrast to "AUTOCAL” no new calibrations are however performed, but deviations from selectable calibration tolerances are checked.
Operation
FIDAMAT 6 gas analyzer
5-14 Operating manual –A5E00222135-01
AUTOCAL / check Start by binary input AUTOCAL can be activated via a binary input provided one has been
configured with the aid of function 72.
AUTOCAL run once In the "AUTOCAL on” state, an AUTOCAL sequence can be started at
any time by means of the "AUTOCAL run once” soft key – provided the unit is in the measuring state. A run initiated in this way does not have any influence on the time cycle of an AUTOCAL run – in other words, the cycle time continues to run independently of this.
After AUTOCAL initiation, this item disappears until the process has completed.
AUTOCAL abort An automatic calibration run which is in progress can be stopped at any
time by means of the "AUTOCAL abort” soft key. This results in all calibration data so far obtained being discarded and the calibration data valid before AUTOCAL started (zero point and span) being used as previously.
Aborting in this way has no effect on the time cycle. All valid calibration events are retained.
AUTOCAL sequence This subfunction allows you to group several calibration phases into one
AUTOCAL sequence.
The sequence used for automatic calibration can be defined as you see fit. It is possible to put together a sequence consisting of as many as twelve different phases.
In addition to inputting a zero gas and up to four test gases for each component, you can also program in flushing with sample gas, and a sample gas intermediate operation period as well as a signaling contact. This signaling contact will be available provided it has already been assigned to a relay output by function 71.
Sample gas intermediate operation A sample gas intermediate operation period may become necessary
if the system is only permitted to leave measurement mode for a very specific time. If the necessary flushing times then add up to a period greater than the permitted outage time, there must be a return to measurement mode between the individual calibration runs.
Signaling contact The signaling contact can, for example, be used for starting an
automatic calibration run in a second unit or for signaling the start or end of an AUTOCAL run.
Relay outputs If relay outputs have been defined for sample, zero gas, test gases
and / or measurement or calibration (function 71), the corresponding external solenoid valves will be operated via these outputs. The same applies to the "AUTOCAL" signaling contact as well – this is closed for about one second when the command is being executed.
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-15
Example: The following sequence is to be programmed:
1. Zero gas calibration, with zero gas flushing after 15 minutes
2. Calibration with test gas 1, with flushing with test gas after 10 minutes
3. Flushing with sample: 8 minutes
4. Sample gas intermediate operation: 30 minutes
5. Calibration with test gas 2, with flushing after 8 minutes
6. Calibration with test gas 3, with flushing after 8 minutes
7. Calibration with test gas 4, with flushing after 10 minutes
8. Flushing with sample: 8 minutes
9. Momentary signaling contact so as to be able to start "AUTOCAL" in another unit or channel.
Note
The FIDAMAT 6 only has one connection for test gas. Additional test gases will need to be connected up externally and switched using external valves.
The AUTOCAL sequence specified is shown in the screen to the left.
List for the AUTOCAL sequence:
Step Component
Zero gas 1 Component 1 Zero gas 2 – " – Test gas 1 – " – Test gas 2 – " –
Test gas 3
– " – Test gas 4 – " – Flush sample
Sample intermediate operation
Signaling contact Zero gas 1b Component 2 Test gas 1b – " – Test gas 2b – " – Test gas 3b – " – Test gas 4b – " –
Operation
FIDAMAT 6 gas analyzer
5-16 Operating manual –A5E00222135-01
Note
Zero gas 2 is not used with the FIDAMAT 6!
AUTOCAL / check Cycle parameters
With this subfunction various time constants for activating a cyclically repeating AUTOCAL run can be parameterized.
Time between two AUTOCAL cycles. Any setting between 0 and 1 (hour) will be accepted by the unit. Here a "0” does not correspond to any AUTOCAL cycle.
Time until the first AUTOCAL run (from time when setting made). If a "0” is input here and AUTOCAL is switched on (see "AUTOCAL on / off”) the analyzer will start the AUTOCAL sequence immediately. If AUTOCAL is switched off, the unit will not start an AUTOCAL sequence unless AUTOCAL is switched on within a minute of the "0” being input. If this does not occur, from time the "0” is input the entire time between two AUTOCAL cycles will be cancelled.
This internal clock will therefore run even when AUTOCAL is switched off! It starts the first time the unit is switched on, and needs to be reset to the current time via function 58.
Number of cycles before test-gas calibration is performed.
With every AUTOCAL run the zero point is calibrated. If, in order to save on test gas, the span is not to be adjusted as well every time the zero point is calibrated, a value greater than 1 must be entered in the line "Perform test gas calibration every : : cycle.”
Note
As long as AUTOCAL is activated (AUTOCAL
), access to functions
20 and 21 will also be blocked.
Should you still attempt to access the
functions, a corresponding message will appear in the display.
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-17
Sequence of AUTOCAL check
1. Input the desired calibration tolerances in the "AUTOCAL check”
menu. If necessary, select the relay output and binary input for "AUTOCAL check”.
2. Start the "AUTOCAL check” by the key in the "AUTOCAL check” menu or via binary input.
3. The unit now performs a run as parameterized in the "AUTOCAL sequence”.
4. Should an calibration limit be crossed, maintenance request W10 will be issued and also, if parametried, the relay "Acal Chk Dif.” set.
5. Following a fault-free AUTOCAL run they will both be reset. W10 will be retained in the logbook.
5.3.3 Measuring ranges
When the measuring range functions are selected in the main menu by pressing the third soft key ("Measuring ranges”) the adjacemt display appears.
40 Select measuring ranges
It is possible to select a fixed measuring range or to switch over to automatic measuring range changeover. All possible selections are subject to mutual interlocking.
Automatic measuring range changeover is possible under the following conditions:
At least two measuring ranges must be available. A measuring range is deemed present when:
measuring range initial value full-scale value
The measuring spans must be made larger
The measuring range must adjoin each other or overlap
This yields the following permitted measuring range arrangements which are shown in the following diagrams:
Operation
FIDAMAT 6 gas analyzer
5-18 Operating manual –A5E00222135-01
Only two measuring range types are distinguished:
type A: The full-scale value must be less than the full-scale value which follows it. This
means that the upper measuring range limit is higher with each successive measuring range.
smr : start of measuring range emr: end of measuring range LP: lower changeover point: select
smaller measuring range
UP: upper changeover point: select
larger measuring range
The following applies to measuring range changeover:
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-19
If the upper changeover point (OU) is crossed, the next largest available measuring range will be selected. If the lower changeover point (UU) goes below that of the next smallest available measuring range, this will be selected. The UU is located at 80% (HystA) of the measuring range. The OU is located at 90% (HystE) of the measuring range.
type B: The full-scale value must be greater than or equal to the full-scale value
which follows it. Since the measuring spans must at the same time become larger, the measuring range initial values of the following measuring ranges must always be smaller.
smr : start of measuring range emr: end of measuring range LP: lower changeover point: select
smaller measuring range
UP: upper changeover point: select
larger measuring range
The following applies to measuring range changeover:
If the upper changeover point (OU) is crossed downwardly, the next largest available measuring range will be selected. If the lower changeover point (UU) of the next smallest available measuring range is exceeded, this will be selected.
The UU is located at 10% (100% - HystE) of the measuring range. The OU is located at the start of the measuring range (mba).
41 Define measuring ranges
A maximum of four measuring ranges can be defined whose initial values are assigned to the lower value (0/ 2/ 4 mA) and whose final values are assigned to the upper value (20 mA) of the analog output.
If the message "Measuring ranges implausible” appears, this means that automatic measuring range changeover is not possible. If the initial or final value is 0, the measuring range will be switched off.
Operation
FIDAMAT 6 gas analyzer
5-20 Operating manual –A5E00222135-01
5.3.4 Parameters
Following selection of parameter functions in the main menu by pressing the fourth soft key ("Parameters”) the adjacemt display appears, showing parameter functions 50 to 53. If you press the fifth soft key ("…more”), this will display parameter items 54 to 61 instead.
50 Electrical time constants With the aid of this function various time constants which reduce noise
can be set. The reduction feature corresponds more or less to a low-pass filter with the corresponding time constant. The display delay is shown in 90% time.
The time constant ti is effective within an action interval which is to be parameterized and which is defined as a percentage of the smallest measuring span. On the one hand it damps minor changes in measured value (noise, for example) but has no effect immediately the measured value exceeds the action interval. In this case the measured value is damped by the external time constant ta.
Values up to 100% can be parameterized for the action interval and values up to 300 s for the time constants ti and ta. A low display delay (90% time) can still be achieved by a skilful combination of these three parameters despite a high level of noise suppression.
The effect of the damping parameters set can be observed on the bottom line where the measured value is shown "live.”
51 Limits For each measurement component the unit can monitor up to four limits
which can be assigned to the measuring range as desired. Any relay can be assigned to each limit (see function 71).
Only positive limit values up to 100% can be parameterized.
Furthermore, you can specify whether a limit alarm should be activated when the limit value entered is exceeded or fallen below.
The limit value is assigned to the measuring range by repeatedly pressing the third soft key. Here pointers travel above the boxed measuring range identification numbers and indicate the measuring ranges where limit monitoring is to be activate (in the example to the left this is measuring range 3).
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-21
Limit monitoring can be switched off individually for any limit value (see also function 52).
Tripping of a limit relay is recorded in the logbook (function 3). As soon as the cause of this disappears, the limit relay will reset automatically. If you press the fifth soft key ("…more”), the program will jump to the next limit screen.
52 On / off functions
Unit-specific on / off functions can only be accessed via the first component. This menu allows you to switch off in a simple manner the functions which are listed in the adjacemt display.
This simplified operator action means that the longer routes through the various menu levels are not needed for these functions. However, for the sake of clarity, the function numbers are still quoted.
Activated functions are indicated by , inactivated by □. If you press
the fifth soft key ("more”), the program will jump to the next screen.
Operation
FIDAMAT 6 gas analyzer
5-22 Operating manual –A5E00222135-01
The following functions are switched on or off with the aid of function 52:
Key F. no. Comments
Total calibration 23
Autoranging 40
Limit monitoring 1 51
Limit monitoring 2 51
Limit monitoring 3 51
Limit monitoring 4 51
Suppress negative measured values for display 55
Block logbook 60
Suppress negative measured values 70
Fault / Maint. Req. / CTRL / NAMUR 72
Analog output memory 77
Signal tolerance exceeded 78
Temperature re-compensation of zero point 86
Temperature re-compensation of span 86
Temperature compensation 102 Accessible for maintenance only
Pressure compensation 104 Accessible for maintenance only
TC precompensation 111 Accessible for maintenance only
Linearization 113 Accessible for maintenance only
Table: 5-2: These functions can be selected via function 52
In addition to the functions listed in Table 5-2, additional maintenance functions can be accessed with the aid of function 52. These are reserved for the use of maintenance personnel only and are hidden until the maintenance code (code level 3) is input.
53 Status messages
With the aid of this function a maximum of three different states can be shown in the measured values display.
Functional check shows the current state of the unit (for example, warming-up phase, calibration, and so on) or which actions the operator should carry out (for example, decoding the unit). After quitting a menu, press the MEAS key twice. If you do not do so, the analog output will remain with the fault current.
Note
If functional check is active, this could result in falsification of the measured value. Since the functional check is no longer displayed, the measured value will have to be checked.
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-23
Status Displayed on screen depending on functions 52 and 53
Ft. 53
Ft. 52 Ft. 53
Ft. 52 Ft. 53
Calibration: CAL
None CAL
CAL
CAL Calibration in progress (also in
AUTOCAL)
Measured-value memory: STO
None STO
STO
STO Analog output connected to memory
(see function 77 on page
5-36)
Limit: LIM
None LIM
LIM
LIM Limit exceeded or fallen below (see
function 51 on page
5-20)
Automatic measuring range changeover: AR
None AR
AR
AR Measuring range changeover (active
possible) (see function 40 on page
5-17)
Functional check: CTRL
None CTRL
CTRL
CTRL Unit is not in measurement mode /
warming
-up phase /
Calibration in progress
Table 5-3: Status message The status type "Code” always appears on the status line. Should a
fault occur during operation, the message "Maintenance request”, "Fault” and "Measurement protection” will appear on the status line, depending on the weighting given to the fault. This message is displayed in alternation with the status messages.
In addition, with communication via the ELAN interface – also alternating – the message "Remote” will appear, together with the unit state.
54 Graphic measured value representation
This function allows you to track on screen the measured value time curves for the last 10 minutes or 24 hours.
Following selection of a time axis (time period) the measured value is plotted against the time axis. The very latest measured value is located at the extreme left of the time axis at t = 0.
Operation
FIDAMAT 6 gas analyzer
5-24 Operating manual –A5E00222135-01
Under "Parameters” a specific measuring range can be assigned to the measured value axis. In addition there is also the possibility of an "optimum measured value display.” This means that when this parameter is activated the program automatically carries out rescaling of the measured value axis. Here the scale is adjusted to suit the measured value scatter.
55 Measured-value display
This function allows you to suppress output of negative measured values.
It is also possible to specify the total number of positions and the number of positions after the decimal point.
It should be noted that no more than five positions can be displayed (the decimal point counts as one position).
56 LCD contrast
With this function you can adjust the contrast of the display to a brighter or darker setting.
If the contrast is set wrong, by pressing the third soft key ("Default setting”) you can restore the contrast set in the factory.
It is also possible to carry out an LCD test by pressing the fourth soft key ("test”). A succession of test images is displayed.
If the LCD was grossly incorrect and the unit is in measured value display, if you press the key sequence ENTER, you can restore the default setting.
8 8 8 8
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-25
57 Ignite flame
To get the FID running the hydrogen flame must be ignited. This is done via function 57.
Ignition is activated or deactivated  by pressing the second hot key. In addition, the temperature of the flame and of the sample stream are shown on the display. When the flame is burning, "Flame burning” appears on the display. If the flame goes out, this message disappears.
58 Date and time
With this function you can set the time and date precisely. When the function is called, the presets shown in the screen shot on the left appear in the display.
The cursor will already be standing on the first position of the date display which can be changed in the order day, month, year. The same procedure applies to the time (hours, minutes).
The unit has a system clock which is not protected against power failure (not a real-time clock).
Attention
When switched off, the clock remains at a standstill and is not updated.
The main importance of the settings is for trouble-shooting. With the aid of the time and date it is easier to classify faults which have occurred and which are always recorded in the logbook (function 60).
By pressing the third soft key ("Set clock”) the data which have been set are taken over for use. They will then appear at the bottom edge of the screen.
Operation
FIDAMAT 6 gas analyzer
5-26 Operating manual –A5E00222135-01
59 Measurement point changeover
This function allows you to assign a maximum of six measurement points and to cycle through them automatically.
What this requires is that the measurement point relays have already been parameterized via function 71 ("Relay assignment”) and these will then control the corresponding solenoid valves.
A time duration is also assigned to each measurement point relay and is to be input into the corresponding editing field. For this, values between 0 and 60 (minutes) are possible.
If you press the fifth soft key you can switch measurement point changeover on and off.
It is also possible to assign a signal relay to each measurement point relay. This will make possible a measurement point identification which is separate from the measurement point relay. These signal relays will also need to be configured via function 71.
60 Logbook settings
With the aid of this function, logbook entries can be deleted or locked (see also function 3). The logbook entries can also be deleted by the key sequence ENTER. This is, however, only possible in the measured value display.
5 5 55
This cannot be used for suppressing status messages such as maintenance requests or fault messages which will still be displayed even if the logbook is locked.
61 Internal valves
With the aid of this function, the internal valves can be operated to control the gases.
Pressing the corresponding soft key activates or de-
activates the valves.
The internal valves open or close the hydrogen, combustion air, zero gas and test gas streams.
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-27
5.3.5 Configuration
All of the following functions in this block are only accessible via the level 2 code.
Entry menu
When the configuration functions have been selected in the main menu, pressing the fifth soft key ("…more”) takes you to more configuration functions.
70 Analog output
With the aid of this function you can determine the lower value of the measuring range 0 - 20 mA/ 2 - 20 mA/ 4 - 20 mA/ NAMUR = 4 - 20 mA with limit at 3.8 mA. See also the table below.
The desired value is selected by pressing the soft key assigned to it and simultaneously the two other values are reset.
In addition, the analog output can be negated; for example, 0 … 100 ppm THC 0 ... 20 mA or 0 ... 100 ppm THC 20 ... 0 mA.
Note
The second analog output is not used.
Negative measured values: Should negative measured values prove to be unsuitable for further processing, by activating this function you can limit the negative measured values at the analog output to 0 (or 2/ 4/ 3.8 (NAMUR)) mA. The display continues to show the correct measured value.
Measured-value limitation during normal operation
Measured-value limitation during fault / CTRL
Defined analog output / mA
Initial value [mA] Final value [mA] Initial value [mA] Final value [mA]
0 - 20 -1 21 0 21
2 - 20 1 21 2 21
4 - 20 2 21 4 21
4 -20 (NAMUR) 3.8 20.5 3 21.5
Table 5-4: Measuring ranges
Note
A defective electronics module may cause the analog output to remain fixed at approx. -1 mA or approx. +24 mA.
Operation
FIDAMAT 6 gas analyzer
5-28 Operating manual –A5E00222135-01
71 Relay assignment
In the basic unit, six programmable relays are available. Their reversible output contacts (max. 24 V / 1 A) can be used for signaling, controlling valves, and so on. Should these six relays be insufficient it is possible to obtain a further eight relays by installing an additional electronics module (option).
One of the functions listed in Table 5.4 can be assigned to each relay but each function may only be assigned once. What this means is that a fault signal, for example, cannot be sent to two relays.
Pin assignments for the individual relays in the non-energized state can be seen in the assignments diagram "Electrical connections” on page 3-5. The unit is shipped with the relays preset.
Up to four inputs can be configured in a menu screen. Switching to further menu screens – and thus to more binary inputs – is always done by pressing the fifth soft key ("…more”).
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-29
Function
Relay is passive
when
Relay is carrying a
current
Comment
Unassigned
Relay is permanently passive (at zero current)
Fault Fault
Maintenance request
Maintenance request
Also appears on the display (measuring mode) (see Section 6.5)
Calibration Calibration in progress For identification Measuring range
1 (...4)
Measuring range 1 (...4) on
Measuring range detection
Limit 1 (...4)
Limit 1 (...4) has responded
Limit signaling
Functional check (CTRL)
Functional check on
Decoding, warming-up phase, AUTOCAL in progress Calibration in progress
Signaling when:
Unit is decoded
Warming-up phase (max. 2 h)
Calibration in progress
(AUTOCAL)
Sample Supply of sample
Zero gas 1 Supply of zero gas
Test gas 1 (…4) Supply of test gas
Control of the valves
Measurement point 1 (...6)
Measurement point 1 (…6) selected
For gas extraction via solenoid valves at different measurement points
Signal – measurement point 1 (...6)
Measurement point 1 (…6) selected
For measurement-point identification (runs in parallel with measurement point)
Signaling contact
During signaling the relay briefly carries a current
For example, with AUTOCAL: control of a second unit
Heating – gas path
Heating ready for operation
For identification
AUTOCAL / check
AUTOCAL difference too great (function 24)
Hydrogen Hydrogen valve is open Control of the valves
Combustion air
Combustion air valve is open
Control of the valves
Table 5-5: Relay assignments
Operation
FIDAMAT 6 gas analyzer
5-30 Operating manual –A5E00222135-01
72 Binary inputs
In the basic version of the unit, six floating binary inputs are available ['0' = 0 V (0...4.5 V); '1' = 24 V (13 ... 33 V)] which can be configured as required. Should these six inputs be insufficient, you will need to install an additional electronics module with a further eight binary inputs (optional).
This is where the manner of functioning of the binary inputs is defined.
With the operating mode NAMUR () the binary inputs behave as shown in Table 5.6 by 'N'.
If operating mode NAMUR is not activated (), the binary inputs will
behave compatibly with the software edition states of the older version V4.3.0 (marked with 'X' in Table 5.6).
One of the control functions listed below can be assigned to each input as required but each function may only be assigned once.
Pin assignments for the individual inputs can be seen in the assignments diagram 'Electrical connections' on page 3-6. The unit is shipped with none of the binary inputs preset. Up to four inputs can be configured in a menu screen. Switching to further menu screens – and thus to more binary inputs – is always done by pressing the fifth soft key ("…more”).
Attention
It is essential that any changes made to the configuration of the binary inputs are saved to the user data memory by means of function 75. Failure to do so runs the risk of an earlier (and unwanted) configuration being accessed when 'Load user data' (function 75) is executed.
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-31
Driving functions / NAMUR
Function Driving voltage required Comment / effects 0 V 24 V 24 V (min.1 s)
Unassigned No effect when addressed External fault 1, 2, ..., 7 N X
External maintenance request 1, 2, ..., 7
N X
e.g.: Signaling by a gas preparation
unit: condensate overflow, gas cooler defective or similar
Deletion of logbook entries N, X
Manual deletion of logbook entries is possible. But if the cause of a fault or a maintenance request has not yet been corrected the corresponding message will reappear in the logbook.
Functional check (CTRL) 1 … 4 N X
Relays under function 71 must be configured for a functional check if, for example, the function is to be checked in the case of a second unit.
AUTOCAL N, X
AUTOCAL must be parameterized (function
24)
Measuring range 1 (... 4) N, X
For remote measuring range changeover, autoranging (function 52) must be deactivated.
Zero gas N, X
Relay must be configured for zero gas and the corresponding valves connected. See also example 1, below
Test gas N, X
Relay must be configured for sample and the corresponding valves connected. See also example 1, below
Sample N, X
Relay must be configured for sample and the corresponding valves connected. Before the sample gas valve is opened, the other valves must have been connected. The unit then switches to measurement mode and all other processes which had previously been activated (such as
calibration) are aborted. Zero-point calibration Span calibration
N, X
Start calibration
See also example 1, below Autorange N, X Automatic measuring range changeover
AUTOCAL / check N, X Start AUTOCAL check (function 24)
Measurement protection N, X
A "Measurement protection” binary input
can be defined which has the following
effect:
if the unit is in the "Measurement'” state (no
functional check), it will remain in this state;
in other words:
- the unit can no longer be decoded
- the unit can no longer be set to "Remote”
- The message "Measurement protection
switched on” appears on the status line of the measurement screen.
Table 5-6: Driving functions
The meaning of "N" and "X" in the "Driving voltage” columns is explained in function 72 "Binary inputs.”
Operation
FIDAMAT 6 gas analyzer
5-32 Operating manual –A5E00222135-01
Example 1:
The following procedure allows you to adjust the measuring range via the binary inputs.
1. Zero gas:
The "Zero gas'” relay is opened and thus switches into calibration mode.
2.
Test gas: Relay 1, 2, 3 or 4 of the test gas is opened. Distinguishing which relay is concerned will be made clear by the following differentiation:
With total calibration ON: The relay corresponds to the measuring range number which has already been selected under function 22 (on page 5-12).
With total calibration OFF: The relay corresponds to the measuring range number of the measuring range currently selected.
3.
Zero-point calibration is carried out. This will not take place until the previous steps have been carried out and the setting "Common zero-point calibration ON” has been enabled. This means that it is necessary for the unit to be in calibration mode.
4.
Span calibration is carried out. Here too it should be noted that the unit must be in measurement mode for the calibration actually to be carried out. The measuring range currently selected is adjusted.
The following diagrams should make the procedure clear:
Fig. 5-3: Range calibration (zero gas)
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-33
Fig. 5-4: Range calibration (calibration gas)
Note
In this regard you should also note the explanatory material under function 40 "Selecting measuring ranges” on page 5-17.
73 ELAN configuration
With the aid of this dialog you can set the parameters for an ELAN network.
Channel address: This is where the channel address can be specified. Addresses from 1 to 12 can be set. In an ELAN network each address may only be used once.
Measured-value telegrams (on / off): The autonomous transmission of measured values cyclically every 500 ms can be switched on or off here.
When putting together your own communication control system, there is a simple possibility to examine an ELAN telegram. However, to prevent unnecessary load on the unit and the ELAN network this function should only be enabled when actually needed.
Operation
FIDAMAT 6 gas analyzer
5-34 Operating manual –A5E00222135-01
Note
Further information about ELAN may be found in the ELAN interface description (C79000-B5200-C176 German) (C79000-B5276-C176 English).
74 Reset
This function is used for restarting (warm start) (see ”Warming up” on page 4-5). Launching this function puts the unit in the warm-up phase. Only when this phase has ended is the unit fully ready for operation.
75 Save or load data
With this function you can save user-specific data in the user data memory.
This should happen, for example, following successful start-up of a system. All individual settings are then saved and can if necessary be reloaded ("Load user data”).
This is important when repair or maintenance work is to be carried out on a unit, for example, or when a new parameterization needs to be tried out.
If necessary, the state as shipped from the factory can be restored with the aid of the function "Load factory data”.
In addition, this menu incorporates a safety check question. To really load the data into the memory, you must first answer "Yes” to this question. Answering "No” aborts.
Attention
It is essential that any changes made to the configuration of the relay outputs are saved to the user data memory by means of function 75. Failure to do so runs the risk of an earlier (and unwanted) configuration being accessed when "Load user data” (function 75) is executed.
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-35
The next diagram gives an overview of the interactions between the various memory modules.
Fig. 5-5: Memory modules
76 Suppression of short-duration fault signals
This function is used for getting rid of unwanted steep-edged faults which exceed a configurable threshold.
Steep-edged faults ("spikes”) are caused by electromagnetic interference or occasional mechanical jolts. These faults can be suppressed by inputting a "blanking-out time” of 0…5 s. The time input causes the last measured value before a spike occurred to be output and the measurement result is thus no longer affected.
Input can be made in steps of 0.1 s.
If a change in concentration follows immediately upon a fault, this change may under certain circumstances be displayed after a delay. When activating this function, the settings for function 50 ("Electrical time constants”) must be taken into account since this function is executed first.
Operation
FIDAMAT 6 gas analyzer
5-36 Operating manual –A5E00222135-01
77 Memory - Analog output
In the event of a fault (S) or CTRL (when the unit is not in measurement
mode),
- the last measured value registered
- or 0/ 2/ 4 mA
- or 21 mA
is output at the analog output.
78 Calibration tolerances
The calibration tolerance at the zero point defines in percentage form the maximum deviation with respect to the smallest measuring span. The calibration tolerance for span defines in percentage form the maximum deviation with respect to the current measuring span. With the aid of this function it is possible to signal changes in the zero point or span with respect to the last calibration as "maintenance request” provided a relay output has been configured for "maintenance request” under function 71.
For this function to be effective the unit must also have been set to "Total calibration” (function 22).
The calibration tolerance, which can be set between 0 – 99%, relates in the case of the zero point to the smallest measuring range span, and in the case of the span to the measuring range span in which total calibration is carried out.
Let this be made clearer with the aid of an example: Measuring range 1: 5 ... 100 ppm THC Measuring range 2: 20 ... 200 ppm THC Smallest measuring span: 95 ppm THC Measuring range in which calibration is carried out: Measuring range 2
Response threshold for zero point 95 ppm THC x 10% = 9.5 ppm THC
Response threshold for span 180 ppm THC x 10% = 18 ppm THC
If the zero point (or the span) now deviates with respect to the last calibration performed by more than the configured value, the corresponding relay with signal a maintenance request.
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-37
79 Change codes
This function allows you to replace the factory codes ("111” for level 1, "222” for level 2) to be replaced by your own. If the value "000” is now entered for a code, code locking is cancelled and access to the corresponding operator control level is now unrestricted.
80 Selftest
The following three tests will allow you to test the functional serviceability of the unit.
Keyboard test:
With the keyboard test all of the keys on the control panel can be tested. With the five soft keys on the right-hand edge the corresponding item can be made to appear or disappear. If a number key or sign key is pressed, the corresponding number will be displayed in the editing field on the bottom line of the screen. Pressing the "INFO" key will issue a plain-text message; here the "MEAS" and "ESC" retain their return functions.
Relay and binary-input test
The first screen shot shows the status of the six relay and binary inputs in a standard unit. If an option board is fitted, on a second screen page a further eight relays or binary inputs will be found.With the relay test individual relays can be activated. This is done via the input field. With a "1” the relay picks up, with a "0” it returns to its release state. The input field will not allow you to input anything except "0” and "0.”
After you quit function 80 the relays will have the status which they had before the relay and binary inputs test was selected.
The current state of the binary inputs is shown in the "Binary” column.
Analog test
With the analog test the analog output can be set for testing purposes to a constant current of between 0 and 24000 µA.
The analog input shows the current input currents in µA.
Operation
FIDAMAT 6 gas analyzer
5-38 Operating manual –A5E00222135-01
81 Language selection
With the aid of this function the unit can be switched over to a second
dialog language.
The unit is always shipped with the language ordered. As a rule English is provided as the second language (if English is the first language, Spanish will be set as the second).
83 Accompanying gas correction
With the FIDAMAT 6, accompanying gas correction is not usually
necessary. Should an accompanying gas correction still be required in a particular case, it will first be necessary to decide what type of accompanying gas influence is involved. By pressing the first soft key you can ascertain the type of accompanying gas influence. The following distinctions are possible:
No accompanying gas influence
Accompanying gas correction with constant accompanying gas
influence
Accompanying gas correction with variable accompanying gas influence via analog inputs
Accompanying gas correction with variable accompanying gas influence via ELAN
Accompanying gas correction will be disabled while an calibration (zero point or span) is in progress. Once calibration is completed and the unit has returned to measurement mode, it will be reactivated.
Accompanying gas correction with constant accompanying gas influence
The unit must be informed of the value of the zero point offset – here referred to as the accompanying gas deflection. In addition, you can specify whether accompaning gas correction is only to apply to certain measuring ranges.
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-39
With a variable accompanying gas the situation is different. What is concerned here is a variable accompanying gas influence which can be picked up by a suitable separate gas analysis device and which is then fed as an analog current or digital signal (via ELAN) into the device for accompanying gas offsetting.
If accompanying gas correction is to use the RS 485 serial interface (ELAN), the same inputs should be made as with accompanying gas correction via analog input.
The following are also required: Channel number and component number of the accompanying gas analysis device. The display will show the gas type assigned to the channel and the component, and also the device status (see also function 82 "Pressure compensation”).
In addition, the parameter "Measured-value telegrams” (function 73) must be set to "on” for the device which is supplying the correction data.
85 Switch valves
With this function you can switch up to six valves by hand. This is done via the relays assigned to the individual valves and which are available on the motherboard and option board.
A requirement is that the corresponding relays must have already been configured under function 71 ("Relay assignment”). The function "Switch valves” applies only for the relay configurations "Zero gas”, "Test gas 1…4” and "Sample”.
Only one valve can be switched in each case since the valves are interlocked.
86 Linear temperature compensation
The FIDAMAT 6 is temperature-compensated both at the zero point and also for the span. Should an additional temperature fault occur during operation it can be compensated with this function.
Temperature compensation at the zero point:
Starting from a mean physical temperature, two different correction variables can be defined for areas of higher temperature and of lower temperature.
Temperature compensation in the measured value (sensitivity):
The procedure is the same for the zero point, but the change is with respect to the measured value.
Operation
FIDAMAT 6 gas analyzer
5-40 Operating manual –A5E00222135-01
Note
Should the zero point deviate towards minus during a temperature change, will be positively signed and vice versa. The same applies to the measured value.
87 Fault on / off
Signaling of maintenance requests and faults (see Tables 6.5.1 and 6.5.2) can be switched off separately with this function so that there will be no logbook entry, no status message and no outward-directed signaling.
88 AK configuration
This function cannot be accessed unless the unit has an additional AK electronics module.
The following parameters of the serial interface can be set:
Baud rate: 300, 600, 1200, 2400, 4800 9600 *)
Transmission format: 7 data bits, no parity bit 2 stop bits
7 data bits, even parity 1 stop bit 7 data bits, odd parity 1 stop bit 8 data bits, no parity bit 1 stop bit * 7 data bits, even parity 2 stop bits 7 data bits, odd parity 2 stop bits 8 data bits, even parity 1 stop bit 8 data bits, odd parity 1 stop bit 8 data bits, no parity 2 stop bits
* Default setting
Start character: All characters from 1 to 255 are possible but must
differ from the end character! Default setting: 2 STX
End character: All characters from 1 to 255 are possible but must
differ from the start character! Default setting: 3 ETX
Don’t-care character: All characters from 1 to 255 are possible but must
differ from the start and end characters!
Default setting: 10 LF (Line Feed)
Operation
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 5-41
90 PROFIBUS Configuration
This function cannot be accessed unless the unit has an additional PROFIBUS electronics module.
This function allows you to set the PROFIBUS station address. The address can be set between 0…126.
The configuration behavior of the unit can be set with the "ID number”. Parameters 0, 1 and 3 can be selected and have the following significance:
0: Only the PROFILE identification number is acknowledged
positively.
1: Only the device-specific identification number is acknowledged
positively.
3: Only the PROFILE identification number for multivariable devices
(complex analysis devices) is acknowledged positively.
91 Start-up state
With this function you can put the unit into various start-up states. With this function there is also the possibility of a standby setting. This is used for relieving the load on the pump without interrupting operations.
Standby: In this setting the pump is switched off.
Quitting this mode results in the pump being started and the unit going into measurement mode.
Time-out This setting allows you to reduce gas
consumption; the pump is switched off. This mode is quit via the warm-up phase.
Time-out without heating This setting allows you to reduce gas
consumption and output. The pump is switched off. The heating is also switched off. This mode is quit via the warm-up phase.
92 Pressure values
In this menu various values relating to the pressure are output and can in part be modified.
In the case of the hydrogen pressure the following values are displayed:
Set pressure: Displayed automatically
Max. deviation: If this deviation from the setpoint pressure is
exceeded the unit will quit the measurement state. This value can be changed.
Actual pressure: Displayed automatically
Operation
FIDAMAT 6 gas analyzer
5-42 Operating manual –A5E00222135-01
In the case of combustion air pressure there is a further value which can be input in addition to the functions listed above.
Min. increase at pump start-up:
When the pump starts the combustion air pressure must rise by this value for the unit to go into measurement mode. This value can be changed.
93 Units
The measurement unit can be changed in this menu. Its name cannot on the other hand be changed.
The following settings for the units are possible:
ppm C
1
, ppm C3, ppm C6, mg C/m³
99 Factory settings With the aid of this menu the operator can access the factory functions
level. These settings are intended for the exclusive use of maintenance personnel and are therefore protected by a further code.
When you select this menu a message will appear asking you to input the code.
6. Maintenance
6.1. Maintenance concept
6.1.1. Pump maintenance
6.1.2. Replacing the filter plate
6.2. Replacing the motherboard and option board
6.3. Changing fuses
6.4. Cleaning the unit
6.5. Maintenance request and fault message
6.5.1. List of maintenance requests
6.5.2. Faults
6.5.3. Other faults
Maintenance
FIDAMAT 6 gas analyzer
6-2 Operating manual –A5E00222135-01
6.1 Maintenance concept
Warning
Before opening the unit shut off both the gas and power supply.
Use the appropriate tools for setting work so as to prevent short circuits occurring on the electronic boards.
Incorrect assembly, installation or calibration can under certain circumstance result in dangerous gas escaping. This represents endangerment of the health of personnel (symptoms of poisoning, chemical burns) and the unit can also suffer damage from corrosive attack.
In the case of the FIDAMAT 6, only the pump needs regular maintenance. It is recommended that the pump diaphragm be changed every six months.
6.1.1 Pump maintenance
First of all, disconnect the analyzer from the power supply.
Remove the unit cover and open the oven. The pump can now be taken out of the unit. Care should be taken that the gas connections at the pump do not get damaged.
A pump defect may be noticed by reduced pump performance and also increased noise in the measurement signal. Check the filter disk as well – it could be clogged.
Note
Please observe also Section 'Check for leaks' on page 4-3.
6.1.2 Replacing the filter plate
The filter plate must be replaced at longer or shorter intervals, depending on the application. The seal will also need to be changed at the same time in order to ensure the filter does not leak. After the inserted insulation material has been removed, the filter is accessible from the rear of the unit.
Note
Please observe also Section 'Check for leaks' on page 4-3.
Maintenance
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 6-3
6.2 Replacing the motherboard and option board
Replacing or installing the motherboard or option board is a simple matter.
Removal and installation of the motherboard To remove the board, follow the following procedure:
Disconnect the analyzer from the power supply
Unscrew and remove the housing cover
Remove the data connector from the housing back panel
Undo the 3 screws which are located between the SUB D connectors
at the rear of the unit
Pull out the ribbon-cable plug-in connectors from the motherboard
Carefully remove the motherboard
Removal of option board / adapter board The procedure is the same as for the motherboard. Unlike the
motherboard, only two screws are used in each case to fix the option board and the adapter board to the unit back panel.
The reverse order of operations applies to installing the two boards.
6.3 Changing fuses
Warning
Always disconnect the unit from the power supply before changing fuses!
Apart from this, the requirements applicable to operator and maintenance personnel described in Section 1.6 will apply.
The unit is protected against various influences (for example, mains voltage, heating) by several fuses. However, should even just one mains fuse be defective, nevertheless replace both of them!
To replace the temperature fuse, proceed as follows:
1. Disconnect the analyzer from the power supply
2. Remove the housing cover panel
Maintenance
FIDAMAT 6 gas analyzer
6-4 Operating manual –A5E00222135-01
Warning
During operation the interior of the unit's oven reaches a temperature of 200°C. The actual temperature of the oven may be ascertained via the diagnostic values. Once the unit has been switched off, temperatures fall only at a very slow rate. Protective gloves must always be worn during maintenance work. Failure to observe this may result in serious burns.
3. Open the oven.
4. Remove the temperature fuse by undoing the fastener clip.
5. Undo the clamp used to connect the temperature fuse and remove it.
The reverse order of operations applies to installing the new temperature fuse.
6.4 Cleaning the unit
Surface The front panel and control panel can be wiped clean. In explosion-hazard
areas they should only be cleaned by wiping with a damp cloth. As cleaning agent we recommend water with conventional commercial detergent added.
Only slight pressure should be applied to the surface of the screen when cleaning it as otherwise the thin film could be damaged.
It is absolutely essential to ensure that no water gets into the unit while it is being cleaned.
Interior When the unit has been opened, the interior can, if necessary, be
carefully cleaned with a compressed air gun.
6.5 Maintenance request and fault message
The FIDAMAT 6 is capable of detecting functional irregularities. These appear on the status line as "maintenance request” or ”fault”. At the same time they are registered in the logbook and can be viewed there. By pressing the soft key beside an entry, the corresponding message can be acknowledged. The messages will reappear if the cause of them has not been corrected.
When a new message occurs, the record saved in the logbook is shifted by one memory location. There are 32 memory locations in all. Once all 32 memory locations are occupied, the arrival of a further record (no. 33) causes the oldest (no. 1) to be overwritten.
Function 60 allows you to lock the logbook and even to delete all of the messages it contains. This may be useful during trial operation in particular. Under function 87 all messages can be switched off separately. During normal operation use of this possibility is not recommended.
Maintenance
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 6-5
Maintenance request
Should there be indications of changes to parameters within the unit,
"maintenance request” will appear on the status line of the display. At the time of their occurrence, changes of this kind will not affect the measuring capability of the unit. But to guarantee long-term measuring capability, remedial measures may need to be taken.
If the relay output of the unit has been configured correspondingly (see also "Calibration” on page 5-11), signaling to the outside may also take place.
Fault Defects in the hardware or changes in unit parameters which render the
unit unserviceable for measurement purposes will cause a fault message to be generated. "Fault” then appears on the status line, provided the unit is in measurement mode. The measured value flashes and remedial measures must be taken without fail.
As with the maintenance request, there can be signaling to the outside via the relay output (function 71). In addition the analog output can be linked to the output current range which was set using function 77 (measured value memory).
Other messages In addition to maintenance requests and fault messages, other important
messages are recorded in the logbook.
LIM 1-4 Î Limit values (measuring range 1-4) have been
exceeded / fallen below
CTRL Î Functional check (see "Diagnosis" on page 5-10)
6.5.1 List of maintenance requests
With the FIDAMAT 6 the following messages entail a maintenance request (screen display) and are signaled to the outside provided a corresponding relay has been configured via function 71.
Each maintenance request can be switched off (deactivated) with the aid of function 87.
Maintenance
FIDAMAT 6 gas analyzer
6-6 Operating manual –A5E00222135-01
No. Message Possible causes Remedy Comments
Test gas has been changed
Repeat calibration
W1
Calibration tolerance exceeded
Drift caused by contaminated gas lines
Check if the drift is normal check pipes carrying gas for contamination, replacing if necessary
Calibration tolerance – see also function 78; drift of channel as per technical data: zero point: 1% of full­scale value per week 1% of full-scale value per week
W4
Set clock Unit has been switched off
Fresh input of time and date
Function 58
W5
Combustion air (sample) or hydrogen pressure too high or too low
Combustion air pressure or sample pressure is too high or too low
Hydrogen input pressure is too high or too low
Limits have been set too tight.
Malfunctions of sample supply (pump, filter, lines) Limits have been set too tight.
Closure of exhaust gas restrictor (throttle valve 4), piping, pressure sensor, readjust with pressure regulator
see above
see above
Pump diaphragm defective, filter clogged, exhaust gas restrictor clogged, lines or connections leaking, external sample supply defective or clogged
W6
Temperature LCD display out of tolerance
Ambient temperature is outside the 5 – 45°C tolerance range specified in the technical data
Ensure that the ambient temperature is within the 5 – 45°C range replace the display, senor is defective
W7
Electronic / physical temperature out of tolerance
Temperature sensor defective, ambient temperature outside tolerance
Ensure that the ambient temperature is within the 5 – 45°C range
W8
Flame has gone out
Hydrogen cylinder is empty, combustion air cylinder is empty, lines to the FID have blockage
Check hydrogen pressure, check combustion air pressure
W9
External maintenance request
Signaling from outside Check
Function 72 must have been configured correspondingly
W10
AUTOCAL check deviation
Table 6-1: Causes of maintenance requests
Maintenance
FIDAMAT 6 gas analyzer Operating manual –A5E00222135-01 6-7
6.5.2 Faults
The following faults result in a fault message (screen display) and are signaled to the outside provided a corresponding relay has been configured via function 71. Immediate action should be taken without fail by qualified maintenance personnel to remedy the problem. There is a defect in the unit and the measured values are not correct.
Each fault can be switched off (deactivated) with the aid of function 87.
No. Fault message Possible causes Remedy
S1
Parameters memory test failure
EEPROM holding incorrect or incomplete data in the main memory
1.
Carry out a RESET or switch the unit off and back on again If fault message S1 reappears:
2. Load user data (function 75)
3. Notify Service department
Leave unit operating to make it easier for servicing personnel to diagnose the problem.
S2
Pump does not start
Combustion air pressure (sample) too high, pump defective
Check combustion air pressure repair or if necessary replace pump
S3
Flame will not ignite
No hydrogen or no combustion air, ignition cable defective,
no ignition spark
ignition spark energy inadequate
Check hydrogen / combustion air,
replace cable check high voltage (cable) Notify Service department
S4
External fault message Signaling from outside
Check Function 72 must have been configured correspondingly
S5
Temperature of oven out of tolerance
Temperature sensor defective Check the temperature sensor
S6
Temperature of the catalysator out of tolerance (applies only with ANM or ANMP versions)
Temperature sensor defective Check the temperature sensor
S7
Temperature of flame out of tolerance
Temperature sensor defective Check the temperature sensor
S8
Combustion air pressure (sample) or hydrogen pressure out of tolerance
Pressure sensor defective
Pressures incorrectly set
Check pressure sensor, check pressure and if necessary adjust via the pressure regulators
S9
The heating has switched off
The heating temperature set has not been reached
Check heating elements,
check temperature sensor in the oven / catalysator
S10
24 h RAM / flash memory check
RAM or flash memory defective Replace motherboard
Maintenance
FIDAMAT 6 gas analyzer
6-8 Operating manual –A5E00222135-01
No. Fault message Possible causes Remedy
S12
Mains power supply
Mains power supply out of tolerance
Mains voltage must be within tolerance limits shown on nameplate
S13
Mains frequency / hardware
Mains frequency out of tolerance Internal power supply out of tolerance
Check the diagnostic values, replace motherboard or adapter board
S14
Measured value greater than characteristic final value (+ 5%)
Incorrect calibration, wrong test gas,
sample concentration too high
Repeat calibration, check test gas, check measuring range
S15
Calibration aborted
During an AUTOCAL run the process was aborted
Table 6-2: Causes of fault messages
6.5.3 Other faults
Within an appropriate period of time according to the drift information (see also "Technical data” on page 2-14) the unit should be calibrated at the zero point (function 20) and at deflection (function 21) using the corresponding test gases. Attention should be given to clean gas preparation. In most cases a greater zero point drift is an indication of matter or dust particles being deposited in the analysis chamber (see "Cleaning” on page 6-4).
Fault Possible cause and remedy
Large zero point drift towards +
Check gas preparation (filter), clean analysis chamber (see "Cleaning” on page 6-4)
Large span drift Detector leaking, FID nozzle defective
Sensitivity markedly dependent on flow rate
Exhaust gas line restricted, for example by condensed water
Table 6-3: Causes of unstable measured-value readings
Loading...