Siemens EMOTION 16-6 User Manual

Page 1
SOMATOM Emotion 6/16-slice configuration
Application Guide
Protocols Principles Helpful Hints
Software Version syngo CT 2007E
Page 2
This material does not substitute for that duty and is not intended by Siemens Medical Solutions Inc., to be used for any purpose in that regard. The drugs and doses mentioned herein are consistent with the approval labeling for uses and/or indications of the drug. The treating physician bears the sole responsibility for the diagnosis and treatment of patients, including drugs and doses prescribed in connection with such use. The Operating Instructions must always be strictly followed when operating the MR/CT System. The source for the technical data is the corresponding data sheets.
The pertaining operating instructions must always be strictly followed when operating the SOMATOM Emotion 6/16-slice configuration. The statutory source for the technical data are the corresponding data sheets.
We express our sincere gratitude to the many customers who contributed valuable input.
Special thanks to Christiane Bredenhoeller, Gabriel Haras, Ute Feuerlein, Jessica Amberg, Thomas Flohr, Rainer Raupach, Bettina Hinrichsen, Axel Barth, Kristin Pacheco and the CT-Application Team for their valuable assistance.
To improve future versions of this application guide, we would highly appreciate your questions, suggestions and comments.
Please contact us: USC-Hotline: Tel. no. +49-1803-112244 email ct-application.hotline@med.siemens.de
Editors: Wang Jian, Chen Ma Hao
Page 3
Overview
User Documentation 16
Scan and Reconstruction 18
Dose Information 42
Workflow Information 64
Contrast Medium 122
Application Information 136
Head 162
Neck 204
Shoulder 218
Thorax 226
Abdomen 262
Spine 294
Pelvis 314
Upper Extremities 332
Lower Extremities 346
Vascular 360
Specials 416
Radiation Therapy 462
3
Page 4
Overview
Respiratory Gating 484
Children 508
4
Page 5
Overview
5
Page 6
Contents
User Documentation 16
Scan and Reconstruction 18
Concept of Scan Protocols 18
Scan Set Up 19
Feed in/Feed out 19
Topo Length 20
Scan Modes 21
- Sequential Scanning 21
- Spiral Scanning 21
-Quick Scan 22
- Dynamic Multiscan 22
- Dynamic Serioscan 22
UFC detector 23
Acquisition, Slice Collimation and Slice Width
- SOMATOM Emotion 16-slice configuration
- SOMATOM Emotion 6-slice configuration 26
Increment 27
Pitch 27
Kernels 28
- Head Kernels 32
- Child Head Kernels 32
-Body Kernels 33
- Special Application Kernels 33
Extended FoV 34
Auto-FoV 35
Neuro Modes 37
Automatic Bone Correction 38
Positioning 39
Image Filters 40
24
25
Dose Information 42
CTDI
and CTDI
W
ImpactDose 44
Effective mAs 45
Vol
6
42
Page 7
Contents
CARE Dose 4D 47
- How does CARE Dose 4D work? 49
- Special Modes of CARE Dose 4D 53
- Scanning with CARE Dose 4D 54
- Adjusting the Image Noise 58
- Activating and Deactivating 61
- Conversion of Old Protocols into Protocols with CARE Dose 4D 61
- Additional Important Information 63
Workflow Information 64
WorkStream4D 64
- Recon Jobs 64
- 3D Recon 65
- 1. Sagittal/Coronal Reconstructions 71
- 2. Oblique/Double-oblique Reconstructions
- Non-square Matrix for 3D Recon 76
- Case Examples for 3D Recon and Non-Square Matrix 77
Workflow 79
- Patient Position 79
- Auto Reference Lines 79
- Navigation within the Topogram 80
-API Language 81
e - Logbook 83
- e- Logbook Configuration 83
- e- Logbook subtask card area 87
-e- Logbook Browser 88
- Study Continuation 91
- Reconstruction on the syngo CT Workplace
- Examination Job Status 93
- Auto Load in 3D and Postprocessing Presets
Scan Protocol Creation 96
- Edit/Save Scan Protocol 96
- Scan Protocol Assistant 98
- Manipulate scan protocols 100
71
92
94
7
Page 8
Contents
- Change parameters 103
- Import scan protocols from SOMATOM LifeNet/CD 117
Contrast Medium 122
Contrast Medium 122
- The Basics 122
- IV Injection 125
Bolus Tracking 126
Test Bolus using CARE Bolus 128
Test Bolus 129
- CARE Contrast 130
Application Information 136
SOMATOM LifeNet 136
- General Information 136
- Key Features 137
- SOMATOM LifeNet offline 138
- SOMATOM LifeNet online 140
Image Converter 147
Report Template Configuration 150
File Browser 151
Camtasia 155
- Key features 155
- Additional Important Information 159
Patient Protocol 160
Head 162
Overview 162
- General Hints 165
- Head Kernels 166
Scan Protocols 168
- HeadRoutine 168
- HeadNeuro 172
- HeadSeq 174
- InnerEarHR 177
8
Page 9
Contents
- InnerEarHRVol 180
- InnerEar 184
- InnerEarSeq 188
- Sinus 192
- SinusVol 196
- Orbit 198
- Dental 200
Neck 204
Overview 204
- General Hints 206
- Body Kernels 207
Scan Protocols 208
- NeckRoutine 208
- NeckThinSlice 212
- NeckVol 214
Shoulder 218
Overview 218
- General Hints 219
- Body Kernels 219
Scan Protocols 220
- Shoulder 220
- ShoulderVol 224
Thorax 226
Overview 226
- General Hints 229
- Body Kernels 231
Scan Protocols 232
- ThoraxRoutine/ ThoraxRoutine06s 232
- ThoraxCombi/ ThoraxCombi06s 235
- ThoraxVol 240
-ThoraxFast/
9
Page 10
Contents
ThoraxFast06s 244
- ThoraxHR 246
- ThoraxHRSeq 250
- ThoraxECGHRSeq 252
-LungLowDose/ LungLowDose06s 254
-LungCARE/ LungCARE06s 258
Abdomen 262
Overview 262
- General Hints 264
- Body Kernels 265
Scan Protocols 266
- AbdomenRoutine/ AbdomenRoutine06s 266
- AbdomenCombi/ AbdomenCombi06s 270
- AbdomenVol 274
- AbdomenFast/ AbdomenFast06s 278
- AbdMultiPhase/ AbdMultiPhase06s 280
- AbdomenSeq 288
- Colonography/ Colonography06s 290
Spine 294
Overview 294
- General Hints 296
- Body Kernels 297
Scan Protocols 298
- C-Spine 298
- C-SpineVol 300
- SpineRoutine 302
- SpineThinSlice 304
- SpineVol 305
10
Page 11
Contents
- SpineSeq 308
- Osteo 312
Pelvis 314
Overview 314
- General Hints 316
- Body Kernels 317
Scan Protocols 318
- Pelvis 318
- PelvisVol 322
- Hip 324
- HipVol 328
- SI_Joints 330
Upper Extremities 332
Overview 332
- General Hints 334
- Body Kernels 335
Scan Protocols 336
- WristHR 336
- ExtrRoutineHR 340
- ExtrCombi 344
Lower Extremities 346
Overview 346
- General Hints 348
- Body Kernels 349
Scan Protocols 350
- Knee 350
- Foot 352
- ExtrRoutineHR 354
- ExtrCombi 358
11
Page 12
Contents
Vascular 360
Overview 360
- General Hints 363
- Head Kernels 364
- Body Kernels 365
Scan Protocols 366
- HeadAngio/ HeadAngio06s 366
- HeadAngioVol 370
-CarotidAngio/ CarotidAngio06s 372
- CarotidAngioVol 376
- ThorAngioRoutine/ ThorAngioRoutine06s 380
- ThorAngioVol 384
- ThorAngioECG/ ThorAngioECG06s 388
- ThorAngioECGSeq 392
- Embolism/ Embolism06s 394
- BodyAngioRoutine/ BodyAngioRoutine06s 398
- BodyAngioVol 402
-BodyAngioFast/ BodyAngioFast06s 406
- AngioRunOff/ AngioRunOff06s 410
- WholeBodyAngio 414
Specials 416
Overview 416
- Trauma 416
- Interventional CT 418
- Test Bolus 420
Trauma Protocols 422
- General Information 422
- Trauma 424
- TraumaVol 425
12
Page 13
Contents
-PolyTrauma/ PolyTrauma06s 426
- HeadTrauma 430
- HeadTraumaSeq 432
- Additional Important Information 434
Interventional CT - Biopsy 436
- Biopsy 437
- Biopsy Single 438
Interventional CT - CARE Vision 439
- The Basics 439
- CAREVision 440
- CAREVisionSingle 441
- CAREVisionBone 442
- HandCARE 443
- Additional Important Information 447
General Information for Biopsy and CARE
Vision 450
- Interventional Toolbar 450
- CAREView 453
- Configuration 456
- Routine Subtask card 458
- Additional Important Information 459
TestBolus Protocol 460
- TestBolus 460
Radiation Therapy 462
Radiation Therapy Planning 462
- Benefits 465
Workflow 468
Scan Protocols 470
- Overview 470
- RT_Head 472
- RT_Thorax 474
- RT_Breast 476
- RT_Abdomen 478
- RT_Pelvis 480
- Additional Important Information 482
13
Page 14
Contents
Respiratory Gating 484
Key Features 486
- Respiratory Gating 486
- Respiration Monitoring 486
- Respiration Synchronization 487
Positioning of the respiratory sensor belt 488
Scanning Information 490
- Scan Parameters 490
- Temporal Resolution 491
- Technical Principles 491
- Respiratory Triggering 491
- Respiratory gating 492
- Prospective respiratory triggering versus retrospective respiratory gating 494
- Curve Editor 495
- Synthetic Trigger/Sync 497
Workflow 498
- Reconstruction and Post processing 498
Additional important Information 499
Scan Protocol 500
- RespSeq 500
- Resp 502
- RespModBreathRate 504
- RespLowBreathRate 506
Children 508
Overview 508
- General Hints 512
- Head Kernels 515
- Body Kernels 516
Scan Protocols 518
- HeadRoutine 518
- HeadSeq 522
- InnerEarHR 526
- InnerEar 530
- InnerEarSeq 534
- SinusOrbit 538
- NeckRoutine 542
14
Page 15
Contents
- ThoraxRoutine/ ThoraxRoutine06s 546
- ThoraxCombi/ ThoraxCombi06s 550
- ThoraxHRSeq 554
- AbdomenRoutine/ AbdomenRoutine06s 558
- Spine/ SpineRoutine 562
- SpineThinSlice 566
- ExtrRoutineHR 568
- ExtrCombi 570
- HeadAngio/ HeadAngio06s 574
-CarotidAngio/ CarotidAngio06s 578
- BodyAngioRoutine/ BodyAngioRoutine06s 582
-BodyAngioFast/ BodyAngioFast06s 586
- NeonateBody/ NeonateBody06s 587
15
Page 16

User Documentation

For further information about the basic operation, please refer to the corresponding syngo CT Operator Manual:
syngo CT Operator Manual Volume 1:
syngo Security Package Siemens Virus Protection Basics SOMATOM LifeNet
syngo Patient Browser syngo Data Set Conversion
Camtasia SaveLog E-Logbook
syngo Viewing syngo Filming
syngo CT Operator Manual Volume 2:
Preparations Examination MPPS HeartView CT Respiratory Gating CT CARE Bolus CT CARE Vision CT
syngo CT Operator Manual Volume 3:
syngo 3D syngo Dental CT syngo Osteo CT
16
Page 17
User Documentation
syngo CT Operator Manual Volume 4:
syngo LungCARE CT syngo Pulmo CT syngo Neuro Perfusion CT syngo Body Perfusion CT
syngo CT Operator Manual Volume 5:
syngo Calcium Scoring syngo Circulation syngo Volume Calculation syngo Dynamic Evaluation syngo Neuro DSA CT
syngo CT Operator Manual Volume 6:
syngo InSpace 4D CT syngo Colonography
17
Page 18

Scan and Reconstruction

Concept of Scan Protocols

The scan protocols for adult and children are defined according to body regions - Head, Neck, Shoulder,
Thorax, Abdomen, Pelvis, Spine, Upper Extremities, Lower Extremities, Vascular, RT, Specials and
optional Cardiac, PET, SPECT and Private. The protocols for special applications are defined in the
Application Guide “Clinical Applications” or in the case of a Heart View examination, in the Application Guide “Cardiac CT“.
The general concept is as follows: All protocols without a suffix are standard spiral modes. For example, “Sinus” means the spiral mode for the sinus.
The suffixes of the protocol name are follows: “Routine“: for routine studies “Seq”: for sequence studies “Fast“: use a higher pitch for fast acquisition “ThinSlice“: use a thinner slice collimation “Combi“: use a thinner and a thicker slice collimation “05s”: use the rotation time of 0.5 seconds “ECG“: use a ECG-gated or triggered mode “Neuro“: for neurologicial examinations with a special
mode “Vol“: use the 3D Recon workflow “HR“: use a thin slice width for High Resolution studies A prefix of the protocol name is as follows: “RT”: for radio therapy studies The availability of scan protocols depends on the sys-
tem configuration. “Resp”: for respiratory gated studies The availability of scan protocols depends on the sys-
tem configuration.
18
Page 19
Scan and Reconstruction

Scan Set Up

Scans can be simply set up by selecting a predefined examination protocol. To repeat any mode, just click the chronicle with the right mouse button for repeat. To delete it, select cut. Each range name in the chron­icle can be easily changed before load.
Multiple ranges can be run either automatically with auto range, which is denoted by a bracket connecting the two ranges, or separately with a pause in between.

Feed in/Feed out

The performance of the different buttons (soft but­tons, gantry buttons, control box buttons) is standard­ized as follows:
•in NOT loaded modes
1 mm
•in loaded Biopsy mode:
Feed In/Out = slice width x No. slice positions per scan
2
19
Page 20
Scan and Reconstruction

Topo Length

SOMATOM Emotion 16
Length [mm] 128, 256, 512, 768, 1024,
1500 Slice width [mm] 4x0.6 Angle Top, Bottom, Lateral
SOMATOM Emotion 6
Length [mm] 128, 256, 512, 768, 1024,
1500, 1536*, 2000**,
2048*** Slice width [mm] 3x1 Angle AP, PA, Lateral
* only in combination with PET and SPECT, option
** only in combination with SPECT, option *** only in combination with PET, option
20
Page 21
Scan and Reconstruction

Scan Modes

Sequential Scanning

This is an incremental, slice-by-slice imaging mode in which there is no table movement during data acquisi­tion. A minimum interscan delay in between each acquisition is required to move the table to the next slice position.

Spiral Scanning

Spiral scanning is a continuous volume imaging mode. The data acquisition and table movements are per­formed simultaneously for the entire scan duration. There is no inter-scan delay and a typical range can be acquired in a single breath hold.
Each acquisition provides a complete volume data set, from which images with overlapping can be recon­structed at any arbitrary slice position. Unlike the sequence mode, spiral scanning does not require addi­tional radiation to obtain overlapping slices.
21
Page 22
Scan and Reconstruction

Quick Scan

The data is usually acquired during a full 360° rotation – this is a Full scan. Data acquisition not using a full 360° rotation is called a “Quick scan”. Quick scans are employed to reduce motion artifacts and improve the temporal resolution.

Dynamic Multiscan

Multiple continuous rotations at the same table posi­tion are performed for data acquisition. Normally, it is applied for fast dynamic contrast studies, such as syngo Neuro Perfusion CT.

Dynamic Serioscan

Dynamic serial scanning mode without table feed. Dynamic serio can still be used for dynamic evaluation such as Test Bolus. The image order can be defined on the Recon subtask card.
22
Page 23
Scan and Reconstruction

UFC detector

Siemens’ proprietary, high-speed Ultra Fast Ceramic (UFC) detector enables a virtually simultaneous read­out of two projections for each detector element.
The detector configuration with the routine acquisition of the Emotion 6/16-slice configuration:
SOMATOM Emotion 16-slice configuration:
SOMATOM Emotion 6-slice configuration:
23
Page 24
Scan and Reconstruction

Acquisition, Slice Collimation and Slice Width

Slice collimation is the slice thickness resulting from the effect of the tube-side collimator and the adaptive detector array design. In Multislice CT, the Z-coverage per rotation is given by the product of the number of active detector slices and the collimation (e.g. 6 x
1.0mm for the SOMATOM Emotion 6-slice configura­tion or 16 x 0.6mm for the SOMATOM Emotion 16­slice configuration ).
Slice width is the FWHM (full width at half maximum) of the reconstructed image.
With the SOMATOM Emotion 6/16-slice configuration, you select the slice collimation together with the slice width desired. The slice width is independent of pitch, i.e. what you select is always what you get. Actually, you do not need to care about the algorithm any more; the software does it for you.
If Metrorecon/Fastrecon is not selected you will rou­tinely get “Real Time” images. The Recon icon on the chronicle will be labeled with “RT”. After the scan the Real Time displayed image series has to be recon­structed.
In some cases – this depends also on Scan pitch and Reconstruction increment – the Recon icon on the chronicle will be labeled with “RT”. This indicates the Real Time display of images during scanning. The Real Time displayed image series has to be reconstructed after completion of spiral.
The Acq (Acquisition) is displayed on the Examination task card. The Acquisition is simply "number of slices acquired per rotation" x "width of one slice".
24
Page 25
Scan and Reconstruction
SOMATOM Emotion 16-slice configu­ration
Spiral Mode
Collimation/ Acquisition
16 x 0.6 mm 0.75, 1.0, 1.5, 2.0, 3.0, 4.0, 5.0 mm
16 x 1.2 mm 1.5, 2.0, 3.0, 4.0, 5.0, 6.0, 8.0,
HR/Neuro Spiral Mode
Collimation/ Acquisition
4 x 0.6 mm 0.6, 0.75, 1.0, 1.5, 2.0, 3.0, 4.0,
Sequence Mode
Collimation/ Acquisition
4 x 0.6 mm 0.6, 1.2, 2.4 mm
12 x 0.6 mm 0.6, 2.4, 7.2 mm
16 x 0.6 mm 1.2, 2.4, 4.8, 9.6 mm
Slice width
10.0 mm
Slice width
5.0 mm
Slice width
2 x 5 mm 5.0, 10.0 mm
12 x 1.2mm 1.2, 3.6, 4.8 mm
2 x 8 mm 8.0, 16.0 mm
16 x 1.2 mm 2.4, 4.8, 9.6, 19.2 mm
25
Page 26
Scan and Reconstruction
SOMATOM Emotion 6-slice configura­tion
Spiral Mode
Collimation Slice width
1 mm 1,1.25,2,2.5,3,4,5,6,8,10mm
2 mm 2.5, 3, 4, 5, 6, 8, 10 mm
3 mm 4, 5, 6, 8, 10 mm
Sequence Mode
Collimation Slice width
1 mm 1, 2, 3 mm
2 mm 2, 4, 6, 12 mm
3 mm 3, 6, 9, 18 mm
5 mm 5, 10 mm
HR Spiral Mode
Collimation Slice width
0.5 mm 0.63, 0.75, 1, 1.25, 2, 2.5, 3, 4, 5mm
HR Sequence Mode
Collimation Slice width
1 mm 1 mm
26
Page 27
Scan and Reconstruction

Increment

The increment is the distance between the recon­structed images in Z direction. When the chosen incre­ment is smaller than the slice thickness, the images are created with an overlap. This technique is useful for reducing partial volume effect, giving you better detail of the anatomy and high quality 2D and 3D postpro­cessing.
The increment can be freely adapted from 0.1 - 10 mm.

Pitch

Pitch = feed per rotation z-coverage
z-coverage = detector rows x collimated slice width
Feed/Rotation = table movement per rotation
The Pitch Factor can be freely adapted from 0.45 – 2.0, in Cardio, there is a fixed pitch down to 0.1.
With the SOMATOM Emotion 6/16-slice configuration, you select the slice collimation together with the slice width desired.
The slice width is independent of pitch, i.e. what you select is always what you get. Actually, you do not need to be concerned about the algorithm any more; the software does it for you.
Pitch values with a step width of 0.05 can be selected for all modes.
We recommend to use a Pitch Factor of 0.45 for MPR reconstructions.
27
Page 28
Scan and Reconstruction

Kernels

There are 4 different types of kernels: “H“ stands for Head, “B“ stands for Body, “C“ stands for ChildHead and ”S” stands for Special Application, e.g. syngo Osteo CT.
The image sharpness is defined by the numbers – the higher the number, the sharper the image; the lower the number, the smoother the image.
Head Kernels:
Kernel description
H10s very smooth
H19s very smooth H20s smooth H21s smooth + H22s smooth FR + H29s smooth + H30s medium smooth H31s medium smooth + H32s medium smooth FR + H37s medium smooth (Emotion 16-slice
configuration only) H39s medium smooth H40s medium H41s medium + H42s medium FR H45s medium H47s medium smooth (Emotion 16-slice
configuration only) H48s medium smooth (Emotion 16-slice
configuration only)
H50s sharp H60s medium H70s very sharp H80s inner ear H90s inner ear
28
Page 29
Scan and Reconstruction
Body Kernels:
Kernel description
B08s very smooth B10s very smooth
B19s very smooth B20s smooth B29s smooth B30s medium smooth B31s medium smooth + B35s HeartView medium B39f HeartView medium B40s medium B41s medium+
B46s medium
B50s medium sharp
B60s sharp
B65s sharp
B70s very sharp
B75s very sharp (Emotion 16-slice configu-
ration only) B80s ultra sharp B90s ultra sharp
Child Head Kernels:
Kernel description
C20s smooth
C30s medium smooth C60s sharp
29
Page 30
Scan and Reconstruction
Topogram Kernels:
Kernel description
T10s smooth
T20s standard
T21s standard T80s sharp T81s sharp T90s ultra sharp
Special Application:
Kernel description
S30s Shepp-Logan S80s Shepp-Logan with notch filter S90s Shepp-Logan without notch filter U90s specification kernel
30
Page 31
Scan and Reconstruction
PET-Kernel:
Kernel PET
B19s smooth B29s medium smooth B39s medium H19s smooth H29s medium smooth H39s medium
SPECT-Kernel:
Kernel SPECT
H08s very smooth
B08s very smooth
31
Page 32
Scan and Reconstruction

Head Kernels

For soft tissue head studies, the standard kernel is H40s; softer images are obtained with H30s or H20s, H10s, sharper images with H50s. The kernels H21s, H31s, H41s yield the same visual sharpness as H20s, H30s or H40s, respectively. The image appearance, however, is more acceptable due to a "fine-grained" noise structure; quite often, the low contrast detect­ability is improved by using H31s, H41s instead of H30s, H40s.
In emergency examinations, kernels H22s, H32s, and H42s can be used because they allow fast reconstruc­tion (FR) and easy patient positioning (50 cm FoV). To ensure best performance, special online bone correc­tion (PFO) is not used.
High Resolution head studies should be performed with H50f, H60f (for example, for dental and sinuses). It is essential to position the area of interest in the cen­ter of the scan field.
For a better gray-white brain tissue differentiation use the H37s, H38s or H47s kernel (Emotion 16-slice con­figuration only).

Child Head Kernels

For head scans of small children, the kernels C20s, C30s (for example for soft tissue studies) and C60s (for example, provided for sinuses) should be chosen instead of the "adult" head kernels H20s, H30s and H60s.
32
Page 33
Scan and Reconstruction

Body Kernels

As standard kernels for body tissue studies B30s or B40s are recommended; softer images are obtained with B20s or B10s (extremely soft). The kernels B31s or B41s have about the same visual sharpness as B30s respectively, B40s, the image appearance, however, is more acceptable due to a "fine-grained" noise struc­ture; quite often, the low contrast detectability is improved by using B31s, B41s instead of B30s, B40s.
For higher sharpness, as is required for example, in patient protocols for cervical spine, shoulder, extremi­ties, thorax, the kernels B50s, B60s, B70s, B80s are available.

Special Application Kernels

The special kernels are mostly used for "physical" mea­surements with phantoms, e.g. for adjustment proce­dures (S80s), for constancy and acceptance tests (S80s, S90s), or for specification purposes (S90s). For special patient protocols, S80s and S90s are chosen, e.g. for osteo (S80s).
Note:
In case of 3D study only, use kernel B10s and at least 50% overlapping for image reconstruction.
Do not use different kernels for body parts other than what they are designed for.
33
Page 34
Scan and Reconstruction

Extended FoV

SOMATOM Emotion 16/6-slice configuration offers the extended field of view. The range can be individually adapted by the user from 50 cm up to 70 cm.
To use this feature you have to select the extended FoV checkbox on the Recon subtask card. The default setting is 65 cm, but can be modified.
Extended FoV can be used with each scan protocol. The extended FoV value should be adapted carefully
to the exact patient size in order to achieve best possi­ble image quality outside the standard scan field.
34
Page 35
Scan and Reconstruction

Auto-FoV

After scanning a topogram the available ranges are dis­played in the topo segment. They can be automatically adapted according to the patient contours. When mov­ing the scan range over the topogram and press the "ctrl" key simultaneous, the adaptation will be done automatically. Please make sure, that the whole object is covered within the default FoV.
In case the FoV is too small, please press the "ctrl" key and move the scan range over the object once, and it will be adapted automatically.
The Auto-FoV will also work with the snap function, when an examination has two or more ranges. The snap function will also cover the Auto-FoV and there­fore you have the possibility to merge different ranges. To be able to use the snap function, it is necessary to have the same FoV and the same x and y coordinates for all available ranges.
Do not use Auto-FoV for asymmetric objects (e.g. only one arm within the scan field).
35
Page 36
Scan and Reconstruction
Hints
• When positioning the arms along the body, the Auto­FoV will also cover the arms.
• When scanning two extremities at the same time, the Auto-FoV will also cover both extremities.
36
Page 37
Scan and Reconstruction

Neuro Modes

In addition to the standard collimations, the SOMA­TOM Emotion 16-slice configuration provides a special mode which is optimized for Neuro applications. Excel­lent low contrast and detail resolution are achieved.
For spiral scans 4 x 0.6 acquisition mode is provided in the range of the cerebrum. This approach shows a min­imized partial volume effect, i.e. low level of artifacts in the base of the skull or near vertebral bodies, as
0.6 mm detector rows are used and the narrow colli­mation reduces scattered radiation.
One scan protocol is predefined for adults: – HeadNeuro using an acquisition 16 x 0.6 mm in the
base and an acquisition of 4 x 0.6 mm in the cere­brum
We recommend using this special protocol for dedi­cated Neuro examinations.
For fast standard examinations such as rule out of hemorrhage or ischemia, the "Routine" protocol should be used.
37
Page 38
Scan and Reconstruction

Automatic Bone Correction

The head protocols provide significant improvements regarding image quality for heads. An automatic bone correction algorithm has been included in the standard image reconstruction. Using a new iterative technique, typical artifacts arising from the beam-hardening effect, for example, Hounsfield bar, are minimized without additional post-processing. This advanced algorithm produces excellent images of the posterior fossa, but also improves head image quality in general. Bone correction is activated automatically for body region “Head”. The reconstruction algorithm for “Head” also employs special adaptive convolution kernels which help to improve the sharpness-to-noise ratio. More precisely, anatomic contours are clearly dis­played while noise is suppressed at the same time without causing a blurring of edges.
Head image without correction.
38
Head image with corrections.
Page 39
Scan and Reconstruction

Positioning

In order to optimize image quality versus radiation dose, scans in body regions “Head” and “AngioHead” are provided within a maximum scan field of 300 mm with respect to the iso-center. No recon job with a field of view exceeding those limits will be possible. There­fore, patient positioning has to be performed accu­rately to ensure a centered location of the skull.
correct positioning wrong positioning of the head of the head
For trauma examinations of the head we provide two protocols, to be found in the specials folder: – HeadTrauma – HeadTraumaSeq.
The scan protocols enable you to utilize the full 50 cm FoV, resulting in easier patient positioning for trauma examinations and to ensure the highest performance, the dedicated PFO head filter is not used.
39
Page 40
Scan and Reconstruction

Image Filters

If you use kernels, the images are reconstructed again with the selected kernel value. If you use image filters, the images are not reconstructed again and the result is much quicker.
Three different filters are available: LCE: The Low-contrast enhancement filter enhances
low-contrast detectability. It reduces the image noise.
• Similar to reconstruction with a smoother kernel
• Reduces noise
• Enhances low-contrast detectability
• Adjustable in four steps
• Automatic post-processing
Image taken without the LCE filter
40
Image taken with the LCE filter
Page 41
Scan and Reconstruction
"HCE": The High-contrast enhancement (HCE) filter
enhances high-contrast detectability. It increases the image sharpness, similar to reconstruction with a sharper kernel.
• Increases sharpness
• Faster than raw-data reconstruction
• Enhances high-contrast detectability
• Automatic post-processing
Image taken without the HCE filter
"ASA": The Advanced Smoothing Algorithm (ASA) filter reduces noise in soft tissues while edges with
high contrast are preserved.
• Reduces noise without blurring of edges
• Enhances low-contrast detectability
• Individually adaptable
• Automatic post-processing
Image taken with the HCE filter
41
Page 42

Dose Information

CTDIW and CTDI
Vol
The average dose in the scan plane is best described by the CTDIW for the selected scan parameters. The CTDIW is measured in dedicated plastic phantoms – 16 cm diameter for head and 32 cm diameter for body (as defined in IEC 60601 – 2 – 44). For scan modes with z­Sharp the CTDI100 is calculated using the single num­ber of tomographic sections (not doubled by z-Sharp) to remain within the terms of IEC 60601-2-44. The z­coverage with and without z-Sharp is the same and so is the dose. This dose index gives a good estimation of the average dose applied in the scanned volume, as long as the patient size is similar to the size of the respective dose phantoms.
Since the body size can be smaller or larger than 32 cm, the CTDI
value displayed can deviate from the
W
dose in the scanned volume. The CTDIW definition and measurement are based on
single axial scan modes. For clinical scanning, i.e.scan­ning of entire volumes in patients, the average dose will also depend on the table feed between axial scans or the feed per rotation in spiral scanning. The dose, expressed as the CTDI
, must therefore be corrected by
W
the pitch factor of the spiral scan or an axial scan series to describe the average dose in the scanned volume.
For this purpose the IEC defined the term “CTDI
“ in
Vol
September 2002:
CTDI
Pitch factor
=
Vol
CTDI
w
This dose number is displayed on the user interface for the selected scan parameters.
42
Page 43
Dose Information
Note: Previously the dose display on the user interface
was labeled “CTDIW“. This displayed CTDIW was also cor­rected for the pitch and was therefore identical to the current CTDI
The CTDI tion of the radiation risk associated with CT examina­tion. For this purpose, the concept of the “Effective Dose“ was introduced by ICRP (International Commis­sion on Radiation Protection). The effective dose is expressed as a weighted sum of the dose applied not only to the organs in the scanned range, but also to the rest of the body. It could be measured in whole body phantoms (Alderson phantom) or simulated with Monte Carlo techniques.
The calculation of the effective dose is rather compli­cated and has to be performed by sophisticated pro­grams. These have to take into account the scan parameters, the system design of the individual scan­ner, such as X-ray filtration and gantry geometry, the scan range, the organs involved in the scanned range and the organs affected by scattered radiation. For each organ, the respective dose delivered during the CT scanning has to be calculated and then multiplied by its radiation risk factor. Finally, the weighted organ dose numbers are added up to get the effective dose.
The concept of effective dose allows the comparison of radiation risk associated with different CT or X-ray exams, i.e. different exams associated with the same effective dose would have the same radiation risk for the patient. It also allows a comparison of the applied X-ray exposure to the natural background radiation, for example, 2 – 3 mSv per year in Germany.
.
Vol
value does not provide the entire informa-
w
43
Page 44
Dose Information

ImpactDose

For most of the scan protocols, the effective dose num­bers for standard male* and female* are calculated, and listed the result in the description of each scan pro­tocol.
The calculation was performed using the commercially available program "ImpactDose" (Wellhoefer Dosime­try).
For pediatric protocols, the ImpactDose calculation and the correction factors published in "Radiation Exposure in Computed Tomography"** are used. These only include conversion factors for ages 8 weeks and 7 years.
*The Calculation of Dose from External Photon Expo-
sures Using Reference Human Phantoms and Monte Carlo Methods. M. Zankl et al. GSF report 30/91
**Radiation Exposure in Computed Tomography,
edited by Hans Dieter Nagel, published by COCIR c/o ZVEI, Stresemannallee 19, D-60596, Frankfurt, Ger­many.
44
Page 45
Dose Information

Effective mAs

In sequential scanning, the dose (D
) applied to the
seq
patient is the product of the tube current-time (mAs)
CTDIw
per mAs:
w
x mAs
spiral
) is
and the CTDI
D
= D
seq
In spiral scanning, however, the applied dose (D influenced by the conventional mAs (mA x Rot Time) and additionally by the pitch factor. For example, if a Multislice CT scanner is used, the actual dose applied to the patient in spiral scanning will be decreased when the pitch factor is greater than 1, and increased when the pitch factor is less than 1 (for constant mA). Therefore, the dose in spiral scanning has to be cor­rected by the pitch factor:
D
spiral
= (D
x mA x Rot Time)
CTDIw
Pitch Factor
To simplify this task, the concept of the “effective“ mAs was introduced with the SOMATOM Multislice scan­ners.
The effective mAs takes into account the influence of pitch on both the image quality and dose:
Effective mAs = mAs
Pitch Factor
To calculate the dose, you simply multiply the CTDI
w
per mAs with the effective mAs of the scan:
D
spiral
= D
x effective mAs
w
CTDI
45
Page 46
Dose Information
For spiral scan protocols, the indicated mAs is the effective mAs per image. The correlation between tube current and effective mAs of spiral scans on a Multi­slice CT scanner is expressed by the following formula:
Effective mAs = mA x RotTime Pitch Factor
Pitch Factor = Feed per Rotation nrow x Slice collimation
mA = effective mAs x Pitch Factor RotTime
where Slice collimation refers to the collimation of one detector row, and nrow is the number of used detector rows.
46
Page 47
Dose Information

CARE Dose 4D

CARE Dose 4D is an automated exposure control, which ensures constant diagnostic image quality over all body regions at the lowest possible dose.
CARE Dose 4D combines three different adaptation methods to optimize image quality at the lowest dose level:
• Automatic adaptation of the tube current to the patient size
• Automatic adaptation of the tube current to the attenuation of the patient’s long axis, the so-called z­axis.
• Automatic adaptation of the tube current to the angular attenuation profile measured online for each single tube rotation, the so-called angle modu­lation.
Based on a single a.p. or lateral topogram, CARE Dose 4D determines the adequate mAs level for every sec­tion of the patient. Based on these levels, CARE Dose 4D modulates the tube current automatically during each tube rotation according to the patient’s angular attenuation profile. Thus, the best distribution of dose along the patient’s long axis and for every viewing angle can be achieved.
47
Page 48
Dose Information
Based on a user defined Image Quality Reference mAs, CARE Dose 4D automatically adapts the (eff.)
mAs to the patient size and attenuation changes within the scan region. With the setting of the Image Quality Reference mAs you can adjust image quality (image noise) to the diagnostic requirements and the individual preference of the radiologist.
Note: The Image Quality Reference mAs should not be adjusted to the individual patient size!
48
Page 49
Dose Information

How does CARE Dose 4D work?

CARE Dose 4D combines two types of tube current modulation:
1) Axial tube current modulation:
Based on a single Topogram (a.p. or lateral) the atten­uation profile along the patient’s long axis is measured in direction of the projection and estimated for the per­pendicular direction by a sophisticated algorithm.
Lateral
Attenuation (log)
Scan Range
Example of lateral and a.p. attenuation profile evalu­ated from an a.p. Topogram.
Based on these attenuation profiles, axial tube current profiles (lateral and a.p.) and the resulting eff. mAs for every table position are calculated. The correlation between attenuation and tube current is defined by an analytical function which results in an optimum dose and image noise in every slice of the scan.
49
Page 50
Dose Information
2) Angular tube current modulation: Based on the above described axial eff. mAs profiles, the tube current is modulated during each tube rota­tion. Therefore the angular attenuation profile is mea­sured automatically during the scan and the tube cur­rent is modulated accordingly in real time to achieve an optimum distribution of the X-ray intensity for every viewing angle.
e
s
rel. tube current
e
s
a
e
e
r
crease
s
c
e
a
e
e
d
r
k d
c
a
e
e
g
d
a
we
r
g
e
n
v
o
a
r
t
s
reference attention
i
o
n
e
g
a
m
i
t
n
a
t
s
n
o
c
o
b
e
se
r
t
s
e
v
a
W
constant dose
s
l
i
m
rel. attenuation
e
s
a
e
r
c
n
i
g
n
o
r
e
s
a
e
r
nc
i
e
g
a
n
i
k
a
e
reference tube current
e
as
e
r
c
Image Quality
Relation between relative attenuation and relative tube current. The adaptation strength may be adjusted by user separately for the left branch (slim) and the right branch (obese) of the curve. This adjust­ment effects all examinations. The gray lines here indicates the theoretical limits of the adaptation (con­stant dose resp. constant image noise). The absolute (eff.) mAs value is scaled with the Image Reference mAs value, which may be adjusted in the Scan Card by the user.
50
Page 51
Dose Information
Scan with constant mA
x-ray dose
Reduced dose level based on topogram
Real-time angular dose modulation
slice position
51
Page 52
Dose Information
Principle of automatic tube current adaptation by CARE Dose 4D for a spiral scan from shoulder to pelvis (very high table feed for demonstration): High tube current and strong modulation in shoulder and pelvis, lower tube current and low modulation in abdomen and thorax. The dotted lines represent the min. and max. tube current at the corresponding table position and result from the attenuation profile of the Topo­gram.
The mAs value displayed in the user interface and in the patient protocol is the mean (eff.) mAs value for the scan range.
The mAs value recorded in the images is the local (eff.) mAs value.
52
Page 53
Dose Information

Special Modes of CARE Dose 4D

For certain examination protocols CARE Dose 4D uses modified tube current modulation, to meet specific conditions, for example:
• for Adult Head protocols the tube current is adapted to the variation along the patient’s long axis and not to the angular attenuation profile.
• for Extremities, CARE Vision, syngo Neuro Perfusion CT, syngo Body Perfusion CT and other special proto­cols (indicated as CARE Dose), only angular tube cur­rent modulation is supported.
• for Osteo and Cardio protocols the mAs setting is adjusted to the patient size and not modulated dur­ing the scan, except if ECG pulsing is switched on.
53
Page 54
Dose Information

Scanning with CARE Dose 4D

If the settings of Image Quality Reference mAs are cor­rectly predefined*, no further adjustment of the tube current is required to perform a scan.
CARE Dose 4D automatically adapts the tube current to different patient sizes and anatomic shapes, but it widely ignores metal implants.
Note: Otherwise the magnification of the topogram would be distorted which would lead to an under­estimation or overestimation of the required eff. mAs.
For an accurate mAs adaptation to the patient’s size and body shape with CARE Dose 4D, the patient should be carefully centered in the scan field.
When using protocols with CARE Dose 4D for body regions other than those they are designed for, the image quality should be carefully evaluated.
As CARE Dose 4D determines the (eff.) mAs for every slice of the topogram, a topogram must be obtained for use of CARE Dose 4D.
*For Siemens scan protocols of SW version syngo CT
2007E, the settings of CARE Dose 4D are already pre­defined but may be changed to meet the customer’s preference of image quality (image noise).
54
Page 55
Dose Information
Outside the topogram range, CARE Dose 4D will con­tinue the scan with the last available topogram infor­mation. Without a topogram, CARE Dose 4D cannot be switched on. Repositioning of the patient on the table and excessive motion of the patient must be avoided between the topogram and the scan. If two topograms of the same projection exist for one scan range, the last acquired will be used for determining the (eff.) mAs. If a lateral and a.p. topogram exist for one scan range, both will be used for determining the (eff.) mAs.
55
Page 56
Dose Information
After the topogram has been scanned, the (eff.) mAs value in the Routine tab card displays the mean (eff.) mAs estimated by CARE Dose 4D based on the topo­gram*. After the scan has been performed this value is updated to the mean (eff.) mAs that was applied. The values may differ slightly due to the online modulation according to the patient’s angular attenuation profile.
*When tuning the CARE Dose 4D parameter setting to
the individual preference for image quality, we rec­ommend keeping track of this value and comparing it with the values used without CARE Dose 4D.
56
Page 57
Dose Information
The Quality reference mAs value is displayed on the Scan tab card. This defines the overall image quality of
the scan protocol currently being used. This value can be adapted for each protocol according to the user’s individual requirements of image quality. Here you can also view the effective mAs value that the system is going to use for the current scan range.
You can also deselect CARE Dose4D on this tab card.
57
Page 58
Dose Information

Adjusting the Image Noise

The correlation between attenuation and tube current is defined by the analytical function described above. This function may be adjusted to adapt the image qual­ity (image noise) according to the diagnostic require­ments and the individual preference of the radiologist. – To adapt the image noise for a scan protocol the
Image Quality Reference mAs value in the Scan tab card may be adjusted. This value can be adapted for each protocol according to the user’s individual pref­erences of image quality, and reflects the mean (eff.) mAs value that the system will use for a reference patient with that protocol and the corresponding body region. The reference patient is defined as a typical adult, 70 kg to 80 kg or 155 to 180 lbs (for adult protocols), or as a typical child, 5 years, appr. 20 kg or 45 lbs (for child protocols). Based on that value, CARE Dose 4D adapts the tube current (or the mean (eff.) mAs value) to the individual patient size or body region.
Note: Do not adapt the Image Quality Reference mAs for an individual patient’s size. Only change this value if you want to adjust the image quality.
58
Page 59
Dose Information
If you change the quality ref. mAs, a pop-up window is displayed.
• To change the configuration of CARE Dose 4D, please open the Examination Configuration dialog box under Options > Configuration. In the window that then appears, please double-click the Examina- tion icon to display the configuration window. The adaptation strength of CARE Dose 4D may be influ­enced for slim, obese patients, or body parts of a patient by changing the CARE Dose 4D settings in the Patient tab card. This may be desirable:
– if the automatic dose increase for obese patients (or
patient sections) has to be stronger than the preset (choose obese: strong increase), resulting in less image noise and a higher dose for those images.
59
Page 60
Dose Information
– if the automatic dose increase for obese patients (or
patient sections) has to be more moderate than the preset (choose obese: weak increase), resulting in more image noise and a lower dose for those images.
– if the automatic dose decrease for slim patients (or
patient sections) has to be stronger than the preset (choose slim: strong decrease), resulting in more image noise and a lower dose for those images.
– if the automatic dose decrease for slim patients (or
patient sections) has to be more moderate than the preset (choose slim: weak decrease), resulting in less image noise and a higher dose for those images.
On the Patient tab card you can adjust the image qual­ity (for more information see chapter How does CARE
Dose 4D work). Note: Changing this adaptation strength effects all
protocols!
60
Page 61
Dose Information

Activating and Deactivating

CARE Dose 4D may be activated or deactivated for the current scan in the Scan tab card. If CARE Dose 4D is activated as default, the Image Quality Reference mAs value is set to the default value of the protocol. After deactivating CARE Dose 4D, the Image Quality Reference mAs is dimmed and the (eff.) mAs value has to be adjusted to the individual patient’s size! If CARE Dose 4D is switched on again, the Image Quality Reference mAs is reactivated. Note that the last setting of the Image Quality Reference mAs or the (eff.) mAs will be restored when you switch from and back to CARE Dose 4D usage. The default activation state of CARE Dose 4D may be set in the Scan Protocol Man­ager. CARE Dose 4D must be selected (column CARE Dose type). The corresponding column for activating CARE Dose 4D is called CARE Dose (4D), with possible default on or off.
Conversion of Old Protocols into Proto­cols with CARE Dose 4D
Protocols of SW versions VA70, VA47 and VA45 may be converted to CARE Dose 4D in the Scan Protocol Man­ager.
Prior to activating CARE Dose 4D an Image Quality Ref­erence mAs value has to be set in the corresponding column.
61
Page 62
Dose Information
If you are unsure about the correct Image Quality Ref­erence mAs value, follow this simple procedure:
• Enter the (eff.) mAs value used for that type of pro­tocol without CARE Dose 4D.
• There is a simple way of ascertaining what eff. mAs CARE Dose 4D will use along the scan range: When the topogram is complete shrink the scan range to it's minimum. As you move this small box over the topogram you can see how the eff. mAs displayed in the Routine and Scan tab card varies along the patient's body. To achieve a certain eff. mAs at a patient's particular body region you can move the small scan range to this position and then adjust the Quality reference mAs so that the displayed eff. mAs value is as desired. After resizing the scan range to the range for the examination, carefully observe the displayed mean eff. mAs. After the subsequent scan is com­pleted inspect the image quality to ensure that the chosen Quality reference mAs is the right value.
• With that setting perform the first scan and carefully inspect the image quality. In that first step the dose may not be lower than without CARE Dose 4D but will be well adapted to the patient’s attenuation, resulting in improved image quality.
• Starting from that setting, reduce the Image Quality Reference mAs step by step to meet the necessary image quality level.
• Store the scan protocol with the adapted image qual­ity reference mAs.
62
Page 63
Dose Information

Additional Important Information

For ideal dose application it is very important to posi­tion the patient in the isocenter of the gantry.
Example for an a.p. topogram:
X-ray tube
Patient
(centered)
Detector
Patient is positioned in the isocenter – optimal dose and image quality
X-ray tube
Patient
(not centered)
Detector
Patient is positioned too high – increased mAs
X-ray tube
Patient
(not centered)
Detector
Patient is positioned too low – reduced mAs and increased noise
63
Page 64

Workflow Information

WorkStream4D

Recon Jobs

In the Recon card, you can define up to eight recon­struction jobs for each range with different parameters either before or after you acquire the data. When you click on Recon, these jobs are performed automatically in the background. If you want to add more than eight recon jobs, simply click the icon for an already completed recon job in the chronicle with the right mouse button and select delete recon job. Another recon job will now become available on the Recon tab card.
Note: What you delete is just the job from the display, not the images that have been reconstructed. Once reconstructed, these completed recon jobs stay in the browser, until deleted from the local database.
You can also reconstruct images for all scans per­formed by not selecting any range in the chronicle, prior to clicking Recon.
Another entry you will find in the right mouse menu is copy/replace recon parameters. This function is available for spiral scans only.
The main goal is to support the transfer of volume parameters between oblique recon jobs of ranges which cover mainly the same area, e.g., two spiral scans with/without contrast media.
64
Page 65
Workflow Information

3D Recon

3D Recon allows you to perform oblique and/ or double oblique reconstructions in any user-defined direction directly after scanning.
No further post-processing or data loading is needed. The high-quality SPO (spiral oblique) images are calcu­lated by using the system’s raw data.
Key Features
• Reconstruction of axial, sagittal coronal and oblique/ double oblique images
• 3 planning images in the 3 standard orientations (coronal, axial, sagittal)
• Image types for planning MPR Thick (10 mm), MIP Thin (3 mm)
• Field of view and reference image definition possible in each planning segment
• Asynchronous reconstruction (several reconstruc­tion jobs are possible in the background, axial and non-axial)
• Workstream 4D performs reconstructions on the basis of CT raw data
• If the raw data are saved you can start the 3D recon­struction on your syngo CT Workplace.
• It is also possible to perform the reconstruction with non-square matrix.
65
Page 66
Workflow Information
Workflow Description
WorkStream 4D improves your workflow whenever non-axial images of a CT scan are required, for example examinations of the spine.
3D reconstructions are possible: – spiral scan is needed – as soon as one scan range is finished and at least one
axial reconstruction job has been performed (RTD or RTR images).
Select a new recon job and mark Recon Job Type – 3D on the Recon card. The first recon job that is suitable for the 3D reconstruction is used as Available plan- ning volumes.
66
Page 67
Workflow Information
Additional Important Information
Pitch factor for 3D Recon
• For reconstruction of 3D recon jobs the maximum pitch factor is 1.5. If the pitch factor is > 1.5 a message window informs you that this 3D recon job cannot be started and may be deleted. In this case use the standard 3D task card with an axial image series for reconstruction.
67
Page 68
Workflow Information
Three planning segments in perpendicular orienta­tions will appear in the upper screen area. You can choose between MPR Thick (3 mm) and MIP Thin (10 mm) as the image type for your planning volume using the relevant buttons.
In each segment you will find a pink rectangle which represents the boundary of the result images. The image with the right down marker represents the field of view (FoV) of the result images (viewing direction).
Right down marker
Reference lines
68
Page 69
Workflow Information
The rectangle with the grid represents the reference image (topogram) which is added to the Topogram series including the reference lines after reconstruc­tion.
Topographics indicator
Reference lines
Recon area
69
Page 70
Workflow Information
Preview Image
A preview of the actual FoV is now available.
• After pressing the button Preview Image the actual
FoV to be reconstructed will be displayed.
• Clicking again on the button deactivates the preview image and displays the whole reference image again.
• Double clicking into the FoV image activates or deac­tivates the Preview Image function as well.
If the Preview Image function is active and you move or rotate the box, or change the recon begin and end position, the Preview image in the FoV segment will be updated accordingly.
70
Page 71
Workflow Information
Depending on the desired resultant images, choose coronal, sagittal or oblique recon axis.

1. Sagittal/Coronal Reconstructions

• Adjust the field of view size to your needs.
• It is only possible to reconstruct images with a squared matrix.

2. Oblique/Double-oblique Reconstructions

If you want to define the orientation of the result images independent of the patient’s axis:
• Enable the Free V iew Mode and rotate the reference lines in the three segments until the desired image orientation is displayed. The vertical and horizontal line are always perpendicular to each other. With the default orientation button you can reset the image orientation at any time.
• It is only possible to reconstruct images with a squared matrix.
• Set the field of view to the active segment by clicking the Set FoV button. The result images will then be orientated as in the FoV segment. You can adjust the extension perpendicular to the field of view in the same way in the other two segments.
71
Page 72
Workflow Information
To define the reference image (topogram) to the active segment, click on the Set Reference Segment button. This defines the orientation of the reference image which will be added to the result images.
Once you have finished the adjustment, start calcula­tion of the result images by clicking on the Recon but­ton. You can start a recon job at any time, indepen­dently of other ongoing jobs (asynchronous reconstruction). After starting the recon job the layout of the Examination task card changes back to the standard layout. If "auto recon" is selected, all defined recon jobs start automatically after scanning. The progress of reconstruction is displayed by the slider in the tomo segment.
72
Page 73
Workflow Information
Additional Information
As soon as you define a new recon range, all recon ranges will be shown in the topo segment. The two numbers on the right-hand side at the beginning of each recon range indicate the recon job the range belongs to. The first number stands for the scan range, the second number stands for the recon job to which the range belongs. If no recon job is pending, only the scan ranges are shown in the topo segment. Only one number on the right-hand side at the beginning of each scan range indicates which scan the range belongs to.
• If the first recon job is saved as an Oblique recon job, RTD images are displayed after scanning and the
Examination task card is automatically switched to 3D reconstruction
• Patient Browser: for each double oblique recon job, one series is added in the Patient Browser.
•If Auto Reference Lines is selected the correspond- ing reference image is added to the 3D recon series.
• All reconstructions are performed in the background
• Do not use high resolution images
• Do not use extended FoV
• If no entry is selected in the chronicle, all open reconstructions are automatically reconstructed.
•If Autorecon is selected on the Recon tab card, this recon job (axial and oblique) will be automatically reconstructed after scanning.
73
Page 74
Workflow Information
Recon Planning
During planning of a 3D recon range, the image dis­played in the FoV segment will be updated to the new position of the recon start and end position.
The corresponding reference line displayed in both planning segments is the reference line to the actual image displayed in the FoV segment.
One click on the start or end position of the recon range displays either the reference image to the start position of the recon range or the reference image to the end position of the recon range in the FoV seg­ment.
Case Examples Some scan protocols are supplied with predefined
oblique reconstructions. These protocols are marked with the suffix “VOL”.
•Coronal and sagittal reconstruction of the spine:
– Scan a topogram – Plan your axial spiral scan range – Reconstruction of the spiral images (RT images) – Select Recon job Type sagittal/coronal – Select the axial image segment –Press button Set FoV Segment – Adjust the FoV to your needs – Define your desired reconstruction parameters
(for example, image type SPO) – Start reconstruction – Repeat the reconstruction steps for the other
orientation (sagittal/coronal)
74
Page 75
Workflow Information
Oblique reconstruction of the sinuses:
– Scan a topogram – Plan your axial spiral scan range – Reconstruction of the spiral images
(RT images)
–Select Recon job Type oblique – Select the sagittal image segment – Enable Free Mode – Rotate the reference lines until the best view of the
sinuses is displayed in one of the other segments
– Select this segment and press the Set FoV Seg-
ment button
– Adjust the FoV to your needs – Define your desired reconstruction parameters
(e.g., image type SPO)
– Start reconstruction
Oblique reconstruction of the vascular tree:
– Scan a topogram – Plan your spiral scan range – Axial reconstruction of the spiral images
(RTD images)
–Select Recon job Type oblique – Select button MIP Thin as image type for the
planning volume on the toolbar
– Enable Free Mode – Rotate the reference lines until the best view
of the entire vascular tree is displayed in one of the other segments
– Select the coronal image segment – Select this segment and press the Set FoV Seg-
ment button
– Adjust the FoV to your needs – Define your desired reconstruction parameters
(e.g., Type MIP Thin)
– Start reconstruction
75
Page 76
Workflow Information

Non-square Matrix for 3D Recon

If you perfrom a 3D reconstruction of your spiral scan you have the possibility to choose between three dif­ferent FoV matrices: 512 square, 512 non-square, 256 non-square. In some cases it is already saved to the scan protocol (Spine, CarotidAngio) set up a new scan protocol or want to modify an existing one you can save the non-square matrix together with the recon parameters.
• 512 square: the FoV stays quadratic with a 512x512 matrix size.
• 512 non-square: the FoV can be adjusted as a rectan­gle to your needs, for example spine reconstruction. Its max. side ratio is 1:4.
• 256 non-square: the FoV can be adjusted as a rectan­gle to your needs but with a lower matrix size and a lower resolution for example RunOff , Cardiac recon­structions. The maximum side ratio is then 1:8.
If you use the non-square matrix and you extend the side length of your FoV more then the max. ratio then the shorter side will be stretched to fit into the ratio again.
You will find the FoV displayed in the image text for the non-square matrix. It will be displayed like this: FoV X x FoV Y.
76
Page 77
Workflow Information
Case Examples for 3D Recon and Non­Square Matrix
Some scan protocols are delivered with predefined oblique and non-square matrix reconstructions. These protocols are marked with the suffix “VOL”
•Coronal and sagittal reconstruction of the spine:
– Scan a topogram – Plan your axial spiral scan range – Reconstruction of the spiral images (RTR/RTD
images)
–Select Recon job Type sagittal/coronal – Select the axial image segment –Press button Set FoV Segment – Select the Matrix size for example, non-square 512
and adjust the FoV to your needs.
– Define your desired reconstruction parameters
(e.g. image type SPO) – Start reconstruction – Repeat the reconstruction steps for the other
orientation (sagittal/coronal)
Oblique reconstruction of the carotid:
– Scan a topogram – Plan your spiral scan range – Axial reconstruction of the spiral images
(RTR/RTD images) –Select Recon job Type oblique – Select the coronal image segment – Enable Free Mode
77
Page 78
Workflow Information
– Rotate the reference lines until the best view on
the sinuses is displayed in one of the other seg­ments
– Select this segment and press button Set FoV Seg-
ment button
– Select the Matrix size for example, non-square 512
and adjust the FoV to your needs
– Define your desired reconstruction parameters
(e.g. image type SPO)
– Start reconstruction
Double-oblique reconstructions of the heart
For detailed information on heart reconstructions please refer to your "Cardiac CT" Application Guide.
78
Page 79
Workflow Information

Workflow

Patient Position

A default patient position can be linked and stored to each scan protocol. The SIEMENS default protocols are already linked to a default patient position. (Head first - supine)
If a scan protocol is selected and confirmed in the Patient Model Dialog, the linked patient position stays active until the user changes it, even if a scan pro­tocol with different patient position is selected.

Auto Reference Lines

The Auto Reference lines settings defined in the Patient Model Dialog can be linked and saved to each scan protocol.
If a scan protocol is selected and confirmed in the
Patient Model Dialog, the linked Auto Reference lines settings stay active until the user changes them, even if a scan protocol with different Auto Reference lines settings is selected.
79
Page 80
Workflow Information

Navigation within the Topogram

Navigation within the topogram helps you to plan a reconstruction range. The minimum conditions for its use are a scanned range and the availability of RTD (Real time display) images. After scanning, an orange line is displayed within the topogram. This line corre­sponds to the axial image in the tomo segment.
• If you scroll through the axial image stack, the orange line in the topogram is displayed as a refer­ence line to the currently displayed axial image in the tomo segment.
• If you change the reconstruction begin or end, the orange reference line automatically jumps to this new position and the axial image in the tomo seg­ment will be updated accordingly to the newly selected position.
• If you move the whole recon box in the topogram, the orange reference line automatically jumps to this new position and the axial image in the tomo seg­ment will be updated accordingly to the newly selected position.
80
Page 81
Workflow Information

API Language

The API language can now be selected directly in the Patient Model Dialog.
When the API language is selected, only the relevant, language specific API entries can be selected in the Scan subtask card. This way it is much easier to select the correct patient instruction.
81
Page 82
Workflow Information
Before recording a new API text, first define the API lan­guage in the API setup dialog under Setup > API/Com- ment Setup in the main menu.
82
Page 83
Workflow Information

e - Logbook

The goal of e-Logbook is to offer an effective and effi­cient functionality to process examination informa­tion.
The e-Logbook consists of three components:
•The e-Logbook Configuration
•The e-Logbook subtask card area
•The e-Logbook Browser, where all examinations
can be listed for viewing, sorting, searching and printing

e- Logbook Configuration

You will find the e- Logbook Configuration u nd er Options >Configuration >e- Logbook Configuration.
The configuration is divided into three tab cards:
•General
• System Entries
• Manual Entries Under General you can activate and deactivate the e-
Logbook, as default the e-Logbook is activated. If the e-Logbook is deactivated, no patient information is
recorded. If you do not want to have the e-Logbook displayed in
the subtask area you can switch it off, even though the system entries will be recorded.
Additionally you can select a Default printer from a drop down menu.
83
Page 84
Workflow Information
Over Default Time period you can determine how the recordings should be listed inside the e-Logbook Browser:
– Today (which is the default setting) – This week – This month –Yesterday – Last week –Last month
Any changes can be saved by selecting "Apply"
84
Page 85
Workflow Information
On the Manual Entries tab card you can configure the System Entries and Manual Entries, that should be displayed in the e-Logbook.
System Entries are automatically filled out by the sys-
tem and displayed in the e-Logbook as read-only if they are configured.
Default settings are:
• Date of Examination
• Patient Name
• Patient ID
•Date of birth
• Scan Protocol Name
•Total mAs The Continuous Number field is an incremental num-
ber to mark each recorded study within a defined time range. In addition the Star t No. can be set to ensure for example an ongoing numbering after a software update.
85
Page 86
Workflow Information
Furthermore the Continuous Number can be set to:
Daily Monthly Yearly
If you set Continuous Number to Daily, the continu­ous number starts with one each day.
86
Page 87
Workflow Information
Additionally the user can define Manual Entries which will also be displayed in the e-Logbook. These infor­mation can be pre-configured and then selected over a drop down menu in the e-Logbook.
To configure new entries of the drop down menu for each Manual Entry, just type the desired information inside and click on add.
To remove already existing entries, just select the entry and click on delete.
Additionally you can customize up to five Manual Entries fields. If you want to rename the customized entry fields type select Rename.

e- Logbook subtask card area

If you close the current patient examination you will get an e-Logbook subtask area which shows you all the information that will be saved in your system. Here you can edit the manual entries and save these as well by clicking on "ok".
87
Page 88
Workflow Information

e- Logbook Browser

You will find the e- Logbook Browser in the main menu under Patient > e- Logbook browser or you can
use F12 key on your keyboard.
You can list the e-Logbook recordings by date. Select your desired timeframe in the calendar and click List now.
If you want to list the e-Logbook recordings from today, click on Tod ay and the recordings will be dis­played immediately, no confirmation is needed.
A shortcut to yesterday’s recordings is accessible over the black arrow on the right side of the Today button.
The system behaves the same if you want to list the recordings from This Week/Last Week and This Month/Last Month.
88
Page 89
Workflow Information
Additional to the dates, certain criteria can be selected to have a more specific search. Search criteria can be defined for all entries recorded inside the e-Logbook.
For example, the entr y Number of images is recorded. A search for datasets which have a certain amount of images can be defined.
Additional conditions can be defined in this case:
is greater thangreater or equalis less thanless or equalequals
The conditions vary with the selected search criteria. You will find under the only within drop-down menus
only the System and Manual Entries you have config- ured before.
The list can be printed:
• Select from the main menu File> Print. The whole
list will be printed at the Default printer, which is configured under Options> Configuration> e-Log- book> General.
The list can be exported:
• Select from the main menu File> Export.
•A Save As dialog pops up.
• The list will be automatically exported to H:\Site-
Data\e-Logbook.
• A file name can be given. The number of columns inside the displayed list
depend on the configurations under Options> Config- uration> e-Logbook> System/Manual Entries.
89
Page 90
Workflow Information
Hints for the Record List:
1. Calling up Patient Browser
When you double click on any record in the e-Logbook Browser, the patient data of the Patient Browser will be opened, if still available and the customer has the chance to edit the patient information. This change will also be updated inside the e-Logbook.
2. Updating the Recon List
The e-Logbook is updated automatically when the examination data has changed within the Patient Browser with Correct and Rearrange. Patient name, date of birth, patient ID and study ID will be updated automatically.
3. Sorting data within the Recon List
You can decide if the data in the Record List should be listed in ascending or descending order. The default sorting order is ascending. Just click on the column head and a small arrow will appear, clicking on it will change the sorting order.
4. Resizing and reorder the columns
It is possible to reorder the table columns by drag and drop the column head. For resizing the columns you just have to move the vertical column lines together.
5. Modify Manual Entries
Make a right mouse click into the cell and modify the information to your needs. The cell background color will be changed to green as an indication. Click on the enabled "Apply" button, then the changes will be applied to the database and the cell background color will be white again.
90
Page 91
Workflow Information

Study Continuation

An existing study can be continued at a later time. To load an existing study:
• Select the desired study in the Patient Browser.
•Select Register from the Patient drop-down menu.
• The patient data is loaded in the Registration dialog
box.
The previous scan protocol is already preselected, but it is also possible to select any desired scan protocol.
After the patient has been registered, the patient is loaded into the Examination card.
The ranges already scanned are listed. The following chronicle entry is shown between the ranges already scanned and the new ranges: Exam Continue <Patient Position>.
If you want to continue a contrast media study, the sys­tem asks you if the next scan should be continued as a non-contrast scan instead.
• If you want to continue as a non-contrast scan, the chronicle entries for the new scan range is indicated as a non-contrast scan. (No injector symbol is shown.)
If you continue a study as a contrast study, the chroni­cle entries of the new scan range are indicated as a contrast scan. (An injector symbol is shown.)
91
Page 92
Workflow Information
Reconstruction on the syngo CT Work- place
It is possible to start all reconstructions from your sat­ellite console.
• Raw data has to be available in the local database
• Select the raw data series of the patient in the
Patient Browser and load it into the Recon card
• Plan your recon jobs as usual
92
Page 93
Workflow Information

Examination Job Status

You can get an overview of all recon jobs by clicking on the recon task symbol in the status bar or selecting Transfer – Examination Job status in the patient main menu in the Patient Browser.
The Examination Job Status dialog box will appear where all recon jobs (completed, queued and in work) are listed. You can stop, restart and delete each job by clicking the according button. To give a selected job a higher priority click urgent.
The column Type shows you which kind of reconstruc­tion is queued.
Two types are displayed: – Recon
all recon jobs from the Recon card, either on the syngo Acquisition Workplace or syngo CT Workplace.
Auto 3D
all 3D reconstructions which you have send via Auto postprocessing automatically into the 3D Card. These jobs will be deleted from the job list as soon as the patient is closed in the 3D card.
93
Page 94
Workflow Information

Auto Load in 3D and Postprocessing Presets

You can activate the Auto load in 3D function on the Examination task card/Auto Tasking and link it to a
recon job, for example, the 2 slice width in some of the examination protocols. If the post-processing type is chosen from the pull-down menu, the reconstructed images will be loaded auto­matically into the 3D task card on the syngo Acquisi- tion Workplace with the corresponding post-process­ing type.
On the 3D task card you can create parallel and radial ranges for Multi-Planar-Reconstruction (MPR) and Thin Maximum-Intensity-Projection (MIP Thin), which can be linked to a special series.
For example, if you always perform sagittal MPRs for a spine examination, as soon as you load a spine exami­nation into the 3D task card, select the image type (MPR), orientation, and open the Range Parallel func­tion. Adapt the range settings (image thickness, dis­tance between the images etc.) and click the link but­ton and save your settings. You now have a predefined post-processing protocol linked to the series descrip­tion of a spine examination.
nd
recon job with thinner
94
Page 95
Workflow Information
The same can be done for VRT presets. In the main menu under Type > VRT Definition, you can link VRT presets with a series description.
Some of the scan protocols, primarily for Angio exami­nations, are already preset in the protocol with Auto load in 3D. If you prefer not to have this preset, dese­lect the Auto load in 3D and save your scan protocol.
Some of the scan protocols are preset in the protocol with links to a post-processing protocol. If you prefer not to have this preset, please delete the Range Parallel preset or overwrite them with your own settings.
95
Page 96
Workflow Information

Scan Protocol Creation

You can modify or create your scan protocols in two different ways:
• by editing/saving scan protocols
• via scan protocol manager.

Edit/Save Scan Protocol

If you want to modify an existing protocol or create a new one, for example, you want to have two Abdo- menRoutine Protocols with different slice widths, we recommend you do this directly on the Examination task card.
96
Page 97
Workflow Information
User-specific scan protocols can be saved with the fol­lowing basic procedure:
• Register a patient, you can choose any patient posi­tion in the Patient Model Dialog.
• Select an existing scan protocol in the Patient Model Dialog.
• Modify the scan protocol, change parameters, add new ranges etc. to adapt the new protocol to your requirements.
• Scan your patient as usual.
• Check that all parameters are as you desire.
•Select Edit/Save Scan Protocol in the main menu.
• Select the folder where you want the new protocol to appear and the scan protocol name in the pop-up dialog box.
• You can either use the same name to overwrite the existing scan protocol or enter a new name, which will create a new protocol name and will not alter any of the existing protocols already stored.
• If you want to save an "old" protocol again, you may have to modify the protocol name. The old protocol (with the old name) must be cancelled explicitly.
97
Page 98
Workflow Information
Additional important information:
• You can save your scan protocol at any time during the examination.
• It is recommended that you save your own scan pro­tocol under a new name in order to avoid overwrit­ing the default scan protocol.
• Do not use special characters or blanks. Allowed are all numbers from 0 to 9, all characters from A to Z and a to z and explicitly the _ (under-score), but no country-specific characters, for example, à, ê, å, ç, ñ.
• Do not rename scan protocol files at Windows level. This will lead to inconsistencies.
• You can now save your own scan protocols in any pre-defined folder. The organ characteristics will belong to the scan protocol, not to the region.
•In the Patient Model Dialog, the modified scan pro­tocols are marked by a dot in front of the protocol.

Scan Protocol Assistant

If you want to modify special parameters for several existing scan protocols or you want to modify the folder structure, we recommend doing this in the "Scan Protocol Assistant".
You will find the Scan Protocol Assistant under Options > Configuration in the main menu.
98
Page 99
Workflow Information
Step 1 - What do you want to do? Four different operation types are possible:
• Manipulate Scan protocols (cut, copy, paste and delete)
• Change parameters
• Import scan protocols from SOMATOM LifeNet/CD
• Restore protocols to Siemens default
Additional important information
• Each workflow consists of up to five steps, indicated by the footprints.
• Depending on the workflow step you can list all scan protocols or all selected scan protocols. The <Export> button exports the listed scan protocols in MS Excel recognizable format. The export file is always created in H:\SiteData\proto­cols folder. The folder selection cannot be changed.
• The section "Change Parameters" is preselected as default.
• Depending on the selected workflow in Step 1, the Scan protocol Assistant leads you through the cor­responding steps.
• From Step 3 an <Undo> and <Redo> button are added. Both buttons only affect the last operation.
• You can navigate through the selected workflow via the footprints or with the "Back" and "Next" buttons. With the "Quit" Button you can leave the Scan Proto- col Assistant at any time without any changes saved.
99
Page 100
Workflow Information

Manipulate scan protocols

This workflow consists of four steps.
Step 2 - Manipulate scan protocols (cut/copy/paste/ delete)
To be able to modify existing scan protocols you just have to click with the right mouse button onto the pro­tocol you want to cut, copy or delete, there a menu will be displayed where you can choose what you want to do.
Cutting a scan protocol out of one of the folders and putting it into another one can be done with cut and then click again on the right mouse button on the desired folder and you will get the option to paste this scan protocol there. If you want to leave the scan pro­tocol in the old folder and save it into another one you just use the copy/paste function. Removing a scan pro­tocol will delete it. Adult and Child protocols are man­aged separately.
100
Loading...