Register 3English
Print No.: R71-020.033.04.05.02 Doc. Gen. Date:08.97
Replaces: R71-020.033.04.04.02
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
OPTILIX tube assemblies without the designation ”L” (without LOADIX) are
equipped with 3-phase stators and can only be operated on generators with
starters that have been developed specifical ly for this.
These tube assemblies are shown with (3
The technical data of the tube assembly are supplied with the assembly.
Siemens AGRegister 3R71-020.033.04Page 1 of 2BIANGULIX / OPTILIX
Medical EngineeringRev. 0412.96TD SD 31
1 - 2Prerequisites
Mounting the x-ray tube assembly1
The installation of the x-ray tube assembly to the unit is described in the corresponding
unit installation instr uctions.
Please observe especially the notes in the TI 376 (R71-020.038.03..., tube assembly
flange adjusted at the factory).
Do not loosen
screws secured
with varnish
BIANGULIX / OPTILIXRegister 3R71-020.033.04Page 2 of 2Siemens AG
Rev. 0412.96TD SD 31Medical Engineering
CAUTION
Installation2
Fig. 1Fig. 2
2 - 1
Connection to the tube assembly (on the stator side)2
2-phase stator (Fig.1)
II, O, IConnection for rotating anode cable (4/Fig.1), see stator connection.
(insulate shielding - do not connect!)
Abweichende Angaben in den Generator Montageanleitungen sind zu
beachten.
3-phase stator (Fig.2)
Connection of the double-shielded rotati ng-anode cable on the tube assembly with
3-phase stator U
Controls
1, 2:LOADIX measured value acquisition (only for tube assemblies with
designation ”L”)
Jumper 1-2- (1/Fig.1) is removed for LOADIX mode.
The measuring resistor (2/Fig.1) is installed in the LOADIX
(see LOADIX installation and setting instructions R71-020.033.01...).
1, 2, 4: Connection of the oil pressure switch (5/Fig.1)
- Allocation with POLYDOROS: 2+1
- Allocation with PANDOROS: 1+4
= II, V1 = I, W1 = 0 (designation in the cover)
1
If the soldering connections for the oil pressure switch are
overheated, the plastic assembly can melt, which can adversely
affect the function of the switch.
Recommended soldering temperature: 350° C
Siemens AGRegister 3R71-020.033.04Page 1 of 10BIANGULIX / OPTILIX
Medical EngineeringRev. 0508.97TD SD 31
for max. 3 sec.
2 - 2Installation
CAUTION
NOTICE
Ground wire
Connection for
wire connection on the high-voltage generator.
≥ 6 mm
Stator connection2
Use a cable 3 x 0,75 mm2 provided with shielding as stator cable.
2
ground wire (3/Fig.1) to the central ground
During the start-up phase, voltages o f
reason, use only shielded cables!
The shielded braiding of the stator cable is not
tube assembly side, except when the double-shielded rotatinganode cable is used, or in the case of other information specified
in the generator installation instructions.
Run cable so that it can neither be kinked nor strained with the
unit in motion!
≥ 1800 V can result; for this
connected on the
• Secure the stator to the x-ray tube assembly us ing a compression clip and clamp.
Secure the double-shielded rotati ng anode cable to the x-ray tube assembly using t he
clamp but not the compression clip (Fi g.2) (braid to ground).
Stator frequency 50 Hz or 150 Hz
• Connect the stator cable to the terminal bl ock of the x-ray tube assembly (Fig . 1)
according to the provided starter (stator frequency 50/60 Hz or 150/180 Hz) .
- Observe notes on the inner side of the anode-s ide protective cover.
-With 3
~ stator s, the stator windings are not changed over.
• Observe notes in the generator instru ctions and wiring diagrams. Connect starter s N80,
N81 and N90 generally for 150 Hz.
Stator frequency 300 Hz (17000 rpm)
Only x-ray tube assemblies which are intended for a stator frequency of 300 Hz may be
connected to a 300 Hz starter. For reasons of safety, the stator in the 300 Hz tubehousing ”117 GL” is switched so that the high speed (17.000 rpm) is only obtained with
this tube assembly.
• For this purpose, run 2 separate shielded st ator cables to the x-ray tube assembly. Se e
also the corresponding documents refe rring to the 300 Hz starter.
The connection of 150 Hz tube housings (100,102 etc.) to the
300 Hz starter results in a failure of the power safety circuit
breaker in the starter!
Protective ea rthing2
• Run ground wire ≥ 6 mm
on the high-voltage generator.
BIANGULIX / OPTILIXRegister 3R71-020.033.04Page 2 of 10Siemens AG
Rev. 0508.97TD SD 31Medical Engineering
2
from the tube assembly to the central ground wire c onnection
Installation2 - 3
CAUTION
LOADIX connection2
Not for three-phase stators!
Terminals 1 and 2 are connected with the LOADIX measuring cell.
To secure this measuring cell, connections 1 and 2 are provided with a jumper.
• The jumper and the adaptation resistor must be removed only on connection to the
LOADIX. See LOADIX installation instruc tions.
Temperature limit value switch2
BWhen the temperature limit value is ex ceeded, the oi l pres sure switch r esponds an d the
radiation release is inibit ed in the generator if the swit ch in the generator is connected e.g.
in series with the door contact. Depending on the type of generator, an error message is
output.
If the ”LOADIX” is connected, the tube assembly symbol lights up and a warning signal
sounds.
Fan mode
- Basically, all tube assemblies can be operated with one or two fans.
- Exception: STEREOLIX
- The installation or connection of th e fans is described in the fan instal lation
instructions.
Siemens AGRegister 3R71-020.033.04Page 3 of 10BIANGULIX / OPTILIX
Medical EngineeringRev. 0508.97TD SD 31
2 - 4Installation
I
L
II
S
0C
3-polig
tripolar
NOTICE
NOTICE
0
C
IIS
L
4-polig
quadripolar
Fig. 3Fi g . 4
High-voltage connection2
The tube assembly is provided with ”0” plug connectors. This implies high-voltage cables
with ”0” sealing ends.
I
G
For installation of the high voltage cables, see the Instal lation
Instructions, RX0-000.031.01... in the generator binder, Reg.3.
High-voltage cables with 3-pin ”0”plugs can be connected both on the anode and on the
cathode side (Fig.3). With tube assemblies with final designations GL a 4-core high-voltage cable is necessary on the cathode side (Fig. 4).
If a 4-pole
→ 3-pole transition piece is used (supplied wit h part-no. 97 46 223 X1751), this
may be inserted only on the high-voltage transformer side.
In order to prevent overheating of the filaments, a check must be
performed with the cathotest tube phantom before insertion of
the cathode-side high-voltage cable in the x-ray tube assembly
See the generator setting instructions and the instructions ”
cathode tube phantom” (TI 300 RA0-000.073.06...)
BIANGULIX / OPTILIXRegister 3R71-020.033.04Page 4 of 10Siemens AG
Rev. 0508.97TD SD 31Medical Engineering
Installation2 - 5
Pre-filtering of the x-ray tube assembly2
With supply with additional filter, inherent filtering of the x-ray tube assembly is at least
2.5 mm Al equivalent. Additional filtering comprises a filter cone (0.5 mm Al) and one or
two filter disks (0.5 mm Al each).
• Remove the unnecessary filter dis ks according to the attached collimat or*.
* Collimators with mirror, note on ”Al equivalent 1 mm” attached.
Replacement of the additional filter2
Replacement of the additional filters on the BIANGULIX, OPTILIX- and OPTITOP-tube
assembly
(not OPTILIX-HSG)
(not Optilix-HSG)
Attention!
With 150/180 Hz rotation, the
protective cone must not be omitted
Insert filter disks and protective
cone according to the unit documents
Lead cone, see I.I. adaptation
Unscrew ring nut
protective cone
filter disk
with supplied
tools and reinsert
after check of
pre-filtering
Siemens AGRegister 3R71-020.033.04Page 5 of 10BIANGULIX / OPTILIX
Medical EngineeringRev. 0508.97TD SD 31
2 - 6Installation
CAUTION
Replacement of the additional filters on the ”OPTILIX 12/50 HSG” tube assembly
For reasons of safety, the filter assembly for these tube assemblies is composed in a
different way, replacement is also possible.
Unscrew ring nut with
supplied tools and reinsert
after check of pre-filtering
Protective cone
Insert filter disks according
to the unit documents
Neither the protective cone nor the lead cone or the beryllium
disk may be omitted!
During installation, ensure correct positioning of the seal!
Filter disks
Rubber flat packing
Lead cone
Beryllium disk
BIANGULIX / OPTILIXRegister 3R71-020.033.04Page 6 of 10Siemens AG
Rev. 0508.97TD SD 31Medical Engineering
Installation2 - 7
I.I.-cone adaptation2
Cone diameter (openings) ”d” and I.I . sizes with 80 cm source- I.I. distan ce are listed in the
following table.
Cone opening d (mm)
RBV
BIANGULIX
1)
OPTILIX
1)
OPTILIX
and
OPTITOP
2)
1712(10)231412(11)
27161414
33191620
40231922
For distances A
(cm) deviating from 80 cm, t he necessary cone open ings dx (mm) can be
x
calculated.
d
(mm) =
x
80 cm
A
cm
x
x d (mm)
”Diameters” not contained in the cone sets can be obtained by boring open.
1)Cone set, par t no. 81 92 353 X1122:
Ø 3, 12, 14, 16, 19, 23 mm must be ordered
separately for BIANGULIX or OPTILIX (without Opti 40/73 C).
2)Cone set, part no. 89 52 699 X1953:
Ø 3, 14, 20, 22 mm must be ord ered separately
for tube assembly OPTILIX 150/40/73C-100...and OPTITOP 150/40/80 HC-100...
Siemens AGRegister 3R71-020.033.04Page 7 of 10BIANGULIX / OPTILIX
Medical EngineeringRev. 0508.97TD SD 31
2 - 8Installation
Primary leaves2
Basically, the primary leaves
must be positioned parallel to
the tube axis.
Exception: STEREOLIX
Fig. 5
Fig. 6
In order to facilitate the adjustment
of older collimator types, the adjustment
ring (2) has been created for attachment
to the tube assembly with radiation
output flange adjusted at the factory.
Prerequisite:
Mirror and light localizer lamp must
be adjusted correctly.
The recesses must be positioned above
the clamping slides (1) (adjustment ring
has a very firm seat).
BIANGULIX / OPTILIXRegister 3R71-020.033.04Page 8 of 10Siemens AG
Rev. 0508.97TD SD 31Medical Engineering
Installation2 - 9
Useful radiation f ield2
Depending on the source-image dis tance and the anode inclination angle, cer tai n film formats are exposed. See diagram.
SID [cm]
140
120
100
80
60
40
20
0
0
10
20
6°
30
8°
40
10°
16°
50
Field size [cm]
Field size [cm] on radiation receptor in the longitudinal axis of tube:
Siemens AGRegister 3R71-020.033.04Page 9 of 10BIANGULIX / OPTILIX
Medical EngineeringRev. 0508.97TD SD 31
2 - 10Installation
This page intentionally left blank.
BIANGULIX / OPTILIXRegister 3R71-020.033.04Page 10 of 10Siemens AG
Rev. 0508.97TD SD 31Medical Engineering
Startup3
Startup of the tube assembly with high voltage3
Referring to ”Setting the tube pre-heating”:
• Perform setting of the tube pre- heating according to the corresponding gen erator
setting instructions.
Depending on the generator, the tube assembly can be warmed up before setting the
tube pre-heating.
Pre-heating must be performed.
With generators without this funct ion, the tube assembly must be warmed up as follows
after setting the pre-heating:
- Switch on fluoroscopy at lowest kV stage
- Adjust up to 110 kV, 4.1 mA within approx. 1 min
- Leave fluoroscopy switched on for 10 min
Monitor kV and mA on the oscilloscope
In the case of malfunctions (ignit ion of the tube), fluoroscopy must be perfo rmed for a
longer time at a reduced kV value. Only then may t he kV value be increased. If
fluoroscopy is working without any mal functions, begin adjustment of the ex posure
heating.
If operation is not possible without any malfunctions in spite of longer fluoros copy, the
tube assembly must be replaced.
3 - 1
Referring to ”Setting the exposure heating”:
• Perform setting of the exposure heat ing according to the corresponding generator
setting instructions.
Siemens AGRegister 3R71-020.033.04Page 1 of 2BIANGULIX / OPTILIX
Medical EngineeringRev. 0412.96TD SD 31
3 - 2Startup
Service notes3
Determination of the stator winding resistance
In the case of malfunctions in the rotating anode run, the stator can be checked as
follows.
Connect ohmmeter to the anode side and compare resistance values as follows.
X-ray tube assembly with 1-phase stator
BIANGULIX
(50 Hz)
(2 800...3 500
1
/
)
min
BIANGULIX-Rapid,
OPTITOP and OPTILIX
(8 500...10 000
Test points
0 - I0 - III - II0 - I0 - III - II0 - III/I
Winding
resistance
Ω)
(
70-8013-1683-9618-2013-1631-3614-1618-20
X-ray tube assembly with three-phase stator
OPTITOP and OPTILIX... -100
(150 Hz)
1
/
min
)
Test points
(8 500...10 000
0 - I0 - III - II
Winding
resistance
Ω)
(
2,0-2,62,0-2,62,0-2,6
(150 Hz)
OPTILIX - HSG
(300 Hz)
1
/
1
/
)
min
(17 000
min
B
- 1
)
A
BIANGULIX / OPTILIXRegister 3R71-020.033.04Page 2 of 2Siemens AG
Rev. 0412.96TD SD 31Medical Engineering
Changes to previous version4
Chap. 0Revision level and Table of Contents updated.
Chap. 2
Page 1"Caution" new included
Chap. 4Updated
4 - 1
Siemens AGRegister 3R71-020.033.04Page 1 of 2BIANGULIX / OPTILIX
Medical EngineeringRev. 0308.97TD SD 31
4 - 2Changes to previous version
TD PS 21 / Link
TD SD 31 / Kern R., Tropia
BIANGULIX / OPTILIXRegister 3R71-020.033.04Page 2 of 2Siemens AG
Rev. 0308.97TD SD 31Medical Engineering
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.