English
Doc. Gen. Date: 02.05
2005
2Revision / Disclaimer
1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
2 _______ Changes to Previous Version _____________________________________ 17
Siemens AGSPR2-330.815.02.01.02ARCADIS Avantic
Medical Solutions
02.05CS SD 24
Page 3 of 18
4Laser Light Localizer
Safety information0
1-1Laser Light Localizer
NOTICE
NOTE
Laser
¹ This product contains two class 2 lasers. (USA: class 2
laser) Follow the safety instructions of
ARTD-002.731.03... When working with the laser light
localizer, do no look directly into the laser beam.
The eye is not in immediate danger. However, it is important not to
look directly into the beam.
ARCADIS AvanticSPR2-330.815.02.01.02Siemens AG
02.05CS SD 24
Page 4 of 18
Medical Solutions
Laser Light Localizer5
Other notes1.1
These adjustment instructions apply only to the installation of a laser light localizer in an
ARCADIS Avantic.
This document also provides adjustment instructions for service and maintenance. Store it
in a safe location together with the adjustment device.
NOTICE
Boards
¹ Switch off ARCADIS Avantic prior to disconnecting and
connecting boards
Siemens AGSPR2-330.815.02.01.02ARCADIS Avantic
Medical Solutions
02.05CS SD 24
Page 5 of 18
6Laser Light Localizer
Protective conductor resistance test1.2
After all work has been completed and all covers have been installed, perform the protective conductor test according to ARTD-002.731.17... The protective conductor resistance
may not exceed 0.2 Ohms.
ARCADIS AvanticSPR2-330.815.02.01.02Siemens AG
02.05CS SD 24
Page 6 of 18
Medical Solutions
Laser Light Localizer7
Required measuring equipment and aids1.3
(according to SPC, Spare Parts Catalog, under Tools)
Siemens AGSPR2-330.815.02.01.02ARCADIS Avantic
Medical Solutions
02.05CS SD 24
Page 7 of 18
8Laser Light Localizer
Start-up1.4
• Switch on the ARCADIS Avantic. The laser light localizer automatically shuts off after
5 minutes.
• Switch on the laser light localizer
via the
pushbutton.
- on the ARCADIS Avantic operating console -
•Position the C-arm in the preferred position for operation, i.e. horizontally with the
C-arm lowered.
ARCADIS AvanticSPR2-330.815.02.01.02Siemens AG
02.05CS SD 24
Page 8 of 18
Medical Solutions
Laser Light Localizer9
Adjusting the laser light fan beams to the I.I. grid1.5
Fig. 1:Laser_unit
Fig. 2:Laser light fan beams
•Move the excenter (1 und 2/Fig. 1 / p. 9) to the "zero position".
•Loosen the screws (3/Fig.1/p.9) until the laser diodes can be turned in their holders.
Siemens AGSPR2-330.815.02.01.02ARCADIS Avantic
Medical Solutions
02.05CS SD 24
Page 9 of 18
10Laser Light Localizer
NOTICE
Important!
¹ Do not turn the lens of the laser diodes.
•Turn the diodes in the holder until the laser light fan beams are reflected on the I.I. as
shown in (Fig.2/p.9). The laser light fan beams begin at (X/Fig.2/p.9). This setting
ensures that the laser light fan beams cross near the laser light localizer.
•Retighten the screws (3/Fig.1/p.9).
ARCADIS AvanticSPR2-330.815.02.01.02Siemens AG
02.05CS SD 24
Page 10 of 18
Medical Solutions
Laser Light Localizer11
Checking the laser light beam at the image intensifier1.6
This check is to be performed according to (1 und 2/Fig. 3 / p. 11).
The laser beam must cross the center of the fastening screws.
Fig. 3:Laser light_I.I.
Siemens AGSPR2-330.815.02.01.02ARCADIS Avantic
Medical Solutions
02.05CS SD 24
Page 11 of 18
12Laser Light Localizer
Final steps and functional test1.7
•Mount the POWERPHOS cover and secure it via the cover screws.
•Store these adjustment instructions in register 4, "Start-up" of the ARCADIS Avantic
system folder. These adjustment instructions can be used for adjustment procedures
during subsequent service and maintenance work.
•Switch on the laser light localizer. The two lasers must shut off automatically after 5
minutes.
•Perform the protective conductor test for the entire ARCADIS Avantic system.
ARCADIS AvanticSPR2-330.815.02.01.02Siemens AG
02.05CS SD 24
Page 12 of 18
Medical Solutions
Laser Light Localizer13
Checking the identification labels1.8
Fig. 4:Stickers_other_countries
Siemens AGSPR2-330.815.02.01.02ARCADIS Avantic
Medical Solutions
02.05CS SD 24
Page 13 of 18
14Laser Light Localizer
Labels for the U.S.A.1.9
•Visually inspect the identification labels, warning logo type "Laser class 2", DHHS certi-
fication label, and the laser aperture label as in (Fig.4/p.13) and (Fig.5/p.14).
Fig. 5:Sticker_laser aperture
ARCADIS AvanticSPR2-330.815.02.01.02Siemens AG
02.05CS SD 24
Page 14 of 18
Medical Solutions
Laser Light Localizer15
Warning label check0
Fig. 6:Label_not_usa
•Check the warning label according to (Fig.6/p.15).
Other countries:Check the warning label according to IEC 825 (text) in the local
language, the warning label according to IEC 825 (symbol) and
the identification label.
Great Britain:Check the warning label according to IEC 825 (text) in the local
language, the warning label according to IEC 825 (symbol) and
the identification label. Also check the laser aperture labels
(Fig.5/p.14).
Siemens AGSPR2-330.815.02.01.02ARCADIS Avantic
Medical Solutions
02.05CS SD 24
Page 15 of 18
16Laser Light Localizer
Maintenance1.10
Mechanical safety0
•Ensure that the laser light localizer is labeled as described in this document. Replace
any labels that are damaged.
•Check the I.I. ring as well as the removable grid for mechanical damage.
•Check the laser beam exit window at the tube cover for damage.
•Ensure that the laser light fan beams are set as described in this document.
Operating value inspection0
•Check the function of the "Laser light localizer on/off" button on the operating console of
the basic unit.
•Switch the laser light localizer back on.
¹ The laser diodes must shut off automatically after 5 minutes.
ARCADIS AvanticSPR2-330.815.02.01.02Siemens AG
02.05CS SD 24
Page 16 of 18
Medical Solutions
Changes to Previous Version17
2-2Changes to Previous Version
No changes; initial publication.
Siemens AGSPR2-330.815.02.01.02ARCADIS Avantic
Medical Solutions
02.05CS SD 24
Page 17 of 18
18Changes to Previous Version
ARCADIS AvanticSPR2-330.815.02.01.02Siemens AG
02.05CS SD 24
Page 18 of 18
Medical Solutions