IMPORTANTIt is recommended that you read the “SC Series Instruction Manual”, Sen-
sotec document 008-0608-00 before using the channels described herein.
CAUTION: The operator of this instrument is advised that if the equipment
is not used in a manner not specified in this manual, the protection provided by the equipment may be impaired.
CAUTION: Only qualified, service-trained personnel who are aware of the
hazards involved should remove the cover from the instrument or connect
external wiring to the instrument.
Sensotec and the Sensotec logo are registered trademarks of Sensotec,
Inc.
Printed in U.S.A.
Sensotec continually improves its products, and thus the information herein
is subject to change without notice.
This manual will explain the setup, features and operation of additional Input and
Output channels that may be installed into Sensotec’s 3rd generation SC SEries
instruments. This series includes the models SC1000, SC2000, SC2001 and
SC3004.
This manual uses the following conventions to present information:
Introduction
[TEXT IN BRACKETS]The label of a front panel button.
DDDDIIIISSSSPPPPLLLLAAAAYYYY
->Indicates that what follows is an item from a sub-
DATA
Text that appears on the display, such as error
messages or menu items.
menu, such as
Serial communications commands sent to (or
replies from) the instrument
Chapter 1, “Introduction”, offers general information about this instruction manual.
Chapter 2, “RTD Input Channel”, explains how to wire, configure, operate and
calibrate the RTD Input channel.
Chapter 3, “Frequency Input Channel”, explains how to wire, configure, operate and calibrate the Frequency Input channel.
Chapter 4, “Setup Menu Reference”, is a list of all SETUP menus and a crossreference to related information in this instruction manual.
the text.
A stop sign highlights procedures or information
necessary to avoid damage to equipment, damage to software, loss of data, loss of calibration, or
invalid test results.
The caution symbol alerts that a specific procedure or practice which, is not followed correctly,
could cause serious personal injury.
Additional Channels for SC Series Instrumentation page 5
Page 6
1.2 Related Documents
SC Series Instruction
Manual
Customer Information
Sheet
The “SC Series Instruction Manual”, Sensotec p/n 008-0608-00, describes the wiring,
setup and operation of the chassis and channels of SC Series Instruments. The channels described in this document are:
•Strain-Gage Input channel
•AC/AC-LVDT Input channel
•High-Level Input channel
•Relay Output channel
•DAC Output channel
•Split-Display Virtual channel
•Mathematics Virtual channel
A printed copy of this document is available for order, or you may download it for free
Sensotec’s web site, www.sensotec.com.
Every instrument is shipped with a Customer Information Sheet which documents the
configuration of the instrument when it left the factory. Such information includes:
•Sensotec part number,
•date of manufacture,
•list of all installed channels and their setup information,
•customer specific SensoCode programming of Mathematics Virtual channels and
operation notes.
Communications
Guide
The “SC Series Communications Guide”, Sensotec p/n 008-0610-00, describes in
detail how to communicate with an SC Series instrument using RS-232 and RS-485.
Wiring diagrams, sample programs, and command descriptions are included.
A printed copy of this document is available for order, or you may download it for free
Sensotec’s web site, www.sensotec.com.
page 6 008-0608-01
Page 7
INTRODUCTION1
Additional Channels for SC Series Instrumentation page 7
Page 8
page 8 008-0608-01
Page 9
2.1 Features
Chapter 2
RTD Input Channel
The RTD Input channel provides a DC excitation current to and accepts millivolt
signals from a standard platinum, 100Ω Resistance Temperature Detector (RTD).
These signals are digitized, converted into engineering units, and placed into the
track, peak and valley data values of the channel.
Setup and calibration of the channel is made at the factory for operation with a
customer-supplied Pt-100 RTD with a DIN-43760 standard calibration curve. Recalibration can be made manually through the SETUP mode with the known-load
calibration method and a customer-supplied RTD calibrator.
The engineering units used to display the temperature readings are field selectable from a built-in conversion table. In addition, the engineering units used for
calibration can be independently selected.
Two rear panel control inputs can be field-configured for such functions as remote
track/hold, disabling peak/valley detection and clearing the peak/valley values. A
voltage or current digital-to-analog output is also provided.
2.2 Getting Started Quickly
As shipped from the factory, the RTD Input channel is calibrated for a standard
Pt100 (platinum, 100Ω @ 100ºC) RTD that conforms to the DIN-43760 / IEC751
standards (∝=0.00385). No further calibration is required prior to use.
Connect the RTD to the RTD Input channel according to “Wiring” on page 10.
RTD INPUT CHANNEL2
Additional Channels for SC Series Instrumentation page 9
Page 10
2.3 Wiring
2.3.1 Channel
Connector
The pin-out for the channel’s connector is shown in the following table.
Table 2-1: RTD Input Channel Pin Connections
PinLabelFunctionNOTES
1 (top)+EXC(+)Excitationreferenced to pin 10
2N/C N/CDO NOT CONNECT!
3N/C N/CDO NOT CONNECT!
4-EXC(-)Excitationreferenced to pin 10
5+SIG(+)Signalreferenced to pin 10
6-SIG(-)Signalreferenced to pin 10
7+OUTAnalog Outputreferenced to pin 8
8-OUTAnalog Return-
9N/C N/CDO NOT CONNECT!
10DGNDDigital Ground-
11AUX1Auxiliary Function 1
(connect to pin 10 to activate)
12
(bottom)
The Analog Output and Analog Return pins are electrically isolated from all other pins
on the instrument.
AUX2Auxiliary Function 2
(connect to pin 10 to activate)
referenced to pin 10
referenced to pin 10
2.3.2 4-Wire RTD
Connection
Do not connect anything to the pins labeled as “N/C”. Damage to the instru-
ment or loss of accuracy may result.
Use a 4-wire RTD connection when the highest possible accuracy is desired. The
excitation current from the instrument will be carried by two wires and the voltage drop
across the RTD will be sensed with the other two wires. Since the (+)Signal and ()Signal pins are high impedance inputs, there is little current flow through these wires
therefore the resistance of the wires doesn’t add to the resistance of the RTD.
Figure 2-1: “4-wire RTD Connection” to RTD Input Channel
page 10 008-0608-01
Page 11
2.3.3 3-Wire RTD
Connection
2.3.4 2-Wire RTD
Connection
If the cost of 4-wire cable is an issue, it is possible to use a 3-wire RTD connection that shares many of the benefits of the 4-wire connection. The wire lengths
should be equal and the wire type should be identical. This will equalize wire
resistance and thus obtain the best accuracy.
Figure 2-2: “3-Wire RTD” Connection to RTD Input Channel
The 2-wire connection scheme is not recommended unless the cable length is
short. The wire’s resistance will add to the resistance of the RTD and result in
reduced accuracy.
Figure 2-3: “2-Wire RTD Connection” to RTD Input Channel
RTD INPUT CHANNEL2
Additional Channels for SC Series Instrumentation page 11
Page 12
2.4 Calibration
A listing of all menu items for this channel is given in “Setup Menu Reference” on
page 55.
A customer-supplied RTD simulator/calibrator is required for re-calibration.
Although not recommended, field re-calibration of the channel for operation
with RTDs with outputs other than the Pt100 DIN-43760 / IEC751 standards
can be made; this may result in reduced accuracy and resolution.
Step 1: Wire the RTD simulator/calibrator to the channel’s connector.
See “Wiring” on page 10.
Step 2: Enter the
See “
CALIBRATIONDATA
CCCCAAAALLLLIB
IBRRRRAAAATTTTIIIIOOOONNNNDAT
IBIB
.
DATAAAA
DATDAT
Sub-Menu” on page17.
If you wish to re-calibrate the instrument using a unit-of-measure other than
degrees Celsius, you can use the
CCCCAL
ALIIIIBBBBRRRRAAAATTTTIIIIOOOONU
ALAL
NUNNNNIIIITTTTSSSS
NUNU
menu item to change the
units-of-measure that the instrument uses for calibration.
Step 3: Perform the calibration.
See “
CALIBRATE
Using the
Menu Item” on page 17.
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTEEEE
menu item starts the calibration process. You will be
prompted to use your RTD simulator/calibrator to apply simulated RTD signals
to the channel as required.
page 12 008-0608-01
Page 13
2.5 Channel Specifications
Absolute Accuracy of Factory Calibra-
RTD INPUT
Excitation Current0.25mA
Factory CalibrationFor use with Pt100 RTD that use DIN-
Output voltage range5, ±5, 10 or ±10 VDC (field selectable)
Output current range (optional current
output channels)
Sourceany channel’s track, peak or valley value
Isolation500V
Resolution13 bits
Frequency Responsesame as input when driven by the same
track hold
RTD INPUT CHANNEL2
4-20 mA
channel’s tracking data
Additional Channels for SC Series Instrumentation page 13
Page 14
2.6 Channel Menu
The RTD Input channel is configured and calibrated via its channel menu. Detailed
instructions on operating the instrument in the SETUP Menu mode can be found in
the “SC Series Instruction Manual”, Sensotec p/n 008-0608-00.
2.6.1
OP
OPEEEERATION
OPOP
Sub-Menu
PPPPOOOOWWWWEEEERRRR----OOOONNNNSOURC
SOURCEEEE
SOURCSOURC
RATION
RATIONRATION
Menu
Item
This menu controls the operation of this channel when the instrument is in the RUN
mode.
This menu selects which value is displayed by the channel when first entering the
RUN mode.
The choices are:
•“
•“
•“
” means the live tracking value of the channel.
TTTTRRRRAAAACCCCKKKK
” means the highest value of the channel.
PPPPEEEEAAAAKKKK
” means the lowest value of the channel.
VVVVAAAALLLLLLLLEEEEYYYY
page 14 008-0608-01
Page 15
2.6.2
DISPLA
DISPLAYYYYUNITS
DISPLADISPLA
UNITS
UNITSUNITS
Menu Item
This menu item controls how data values are scaled into engineering units, displayed by the channel, and transmitted via serial communications. The RTD
Input channel has conversion factors for degrees Celsius, degrees Fahrenheit
and Kelvin. Note that if the
then the choices available in the
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONU
DDDDIIIISSSSPPPPLLLLAAAAYUNI
tors will change accordingly. For example, if the
then choices available for the
menu items determine what happens when the
Auxiliary Function pins (labeled as “AUX1” and “AUX2”) on the channel’s connec-
tor are activated. These pins are “activated” when they are connected to the Digital Ground (labeled as “DGND”) pin. The choices are:
•“
DDDDIIIISSSSAAAABBBBLLLLEEEEDDDD
•“
TTTTRRRRAAAACCCCKKKKHOL
” means that activating the pin does nothing.
” means that the tracking, peak and valley values will not be
HOLDDDD
HOLHOL
updated.
•“
HHHHIIIIGGGGHHHH////LLLLOOOOWWWWHOL
•“
HHHHIIIIGGGGHHHH////LLLLOOOOWWWWCLEA
” means that the peak and valley values will not be updated.
HOLDDDD
HOLHOL
” means that the peak and valley values are reset.
CLEARRRR
CLEACLEA
As the Auxiliary Function pins are not isolated, it is recommended that a push-button switch or relay is used to connect these pins to the Digital Ground pin.
RTD INPUT CHANNEL2
Additional Channels for SC Series Instrumentation page 15
Page 16
2.6.4
TYPE
TYPE
Menu Item
TYPETYPE
CALI
CALIBBBBRATION
RATION
CALICALI
RATIONRATION
The only choice is:
•“
TTTTYYYYPPPPEEEE====5P
5POOOOIIIINNNNTCA
5P5P
” means 5-Point Known Load Calibration. You are prompted to
TCALLLL
TCATCA
use your RTD simulator/calibrator to simulate the RTD output at the temperatures
entered in the “
and “
KKKKNNNNOOOOWWWWNNNNPOINT5/
POINT5/5555
POINT5/POINT5/
KKKKNNNNOW
OWNP
NPOOOOIIIINNNNTTTT1111////5555
OWOW
NPNP
” registers. This technique is used to compensate for the non-
The calibration points used for factory calibration are -200ºC, -100ºC, +0ºC, +100ºC
and +200ºC. Using these temperatures for calibration is recommended for most applications. If desired, accuracy may be improved by calibrating over a narrower temperature range which decreases the difference between the points.
Choosing calibration points where the difference between points is larger than
100ºC will result in greatly reduced accuracy.
”
page 16 008-0608-01
Page 17
2.6.5
CALIBRATI
CALIBRATIOOOONDATA
CALIBRATICALIBRATI
NDATA
NDATANDATA
Sub-Menu
CCCCAL
ALIIIIBBBBRRRRAAAATTTTIIIIOOOONNNNUNI
ALAL
KKKKNNNNOOOOWWWWNNNNPOINTx/
POINTx/yyyy
POINTx/POINTx/
Menu ItemThis menu item controls what units of measure are used for the known-load cali-
UNITTTT
UNIUNI
bration points.
Table 2-2: Temperature Conversion
Unit of Measure
degrees Celsius
degrees Fahrenheit
Kelvin
CCCCAAAALLLLIB
IBRRRRAAAATTTTIIIIOOOONNNNUNI
IBIB
Menu Item
UNITTTT
UNIUNI
ddddeeeeggggCCCC
ddddeeeeggggFFFF
KKKK
Menu ItemsThis enters the engineering unit values for the known-load calibration points.
These points must match the actual temperatures that you will apply to the instrument during calibration. The following menu items are available:
•“
KKKKNNNNOOOOWWWWNNNNPOINT1/
•“
KKKKNNNNOOOOWWWWNNNNPOINT2/
•“
KKKKNNNNOOOOWWWWNNNNPOINT3/
•“
KKKKNNNNOOOOWWWWNNNNPOINT4/
•“
KKKKNNNNOOOOWWWWNNNNPOINT5/
POINT1/5555
POINT1/POINT1/POINT2/5555
POINT2/POINT2/POINT3/5555
POINT3/POINT3/POINT4/5555
POINT4/POINT4/POINT5/5555
POINT5/POINT5/
”: point 1 of 5, usually -200ºC.
”: point 2 of 5, usually -100ºC.
”: point 3 of 5, usually 0 ºC.
”: point 4 of 5, usually +100ºC.
”: point 5 of 5, usually +200ºC.
The instrument expects the input signal applied at each known-load point
to be increasing. For example, the load applied at Known-Load Point 2/5
must cause the transducer to produce a more positive output than at
Known-Load Point 1/5.
RTD INPUT CHANNEL2
2.6.6
CALIBRAT
CALIBRATEEEE
CALIBRATCALIBRAT
Menu
Item
This menu item performs a calibration according to what was entered in the
and
BBBBRRRRAAAATTTTIIIIOOOONNNNTYP
TYPEEEE
TYPTYP
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNNDAT
Before performing a calibration, the
“
CALIBRATIONDATA
menu items.
DATAAAA
DATDAT
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOOND
must be entered (see
NDAAAATTTTAAAA
NDND
Sub-Menu” on page 17).
For maximum accuracy, allow at least five minutes of warm-up before cal-
have applied this simulated temperature, press [ENTER].
•The display will read
of
KKKKNNNNOOOOWWWWNNNNPOINT2
POINT2////5555
POINT2POINT2
and
, then
WWWWOOOORRRRKKKKIIIINNNNGGGG
CA
CALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNNUNI
CACA
AAAAPPPPPPPPLLLLYYYY-100.00d
-100.00deg
-100.00d-100.00d
are). When you have applied this simu-
UNITTTT
UNIUNI
(or whatever the settings
egCCCC
egeg
Additional Channels for SC Series Instrumentation page 17
CCCCAAAALLLLIIII----
(or
Page 18
lated temperature, press [ENTER].
•The display will read
KKKKNNNNOOOOWWWWNNNNPOINT3
POINT3/5
POINT3POINT3
/5
/5/5
and
WWWWOOOORRRRKKKKIIIINNNNGGGG
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNNUNI
temperature, press [ENTER].
•The display will read
KKKKNNNNOOOOWWWWNNNNPOINT4
POINT4/5
POINT4POINT4
/5
/5/5
and
WWWWOOOORRRRKKKKIIIINNNNGGGG
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNNUNI
temperature, press [ENTER].
•The display will read
KKKKNNNNOOOOWWWWNNNNPOINT5
POINT5/5
POINT5POINT5
/5
/5/5
and
WWWWOOOORRRRKKKKIIIINNNNGGGG
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNNUNI
temperature, press [ENTER].
•The display will indicate
DDDDOOOONNNNEEEE
, then
AAAAPPPPPPPPLLLLYYYY000.00d
000.00deeeeggggCCCC
000.00d000.00d
are). When you have applied this simulated
UNITTTT
UNIUNI
, then
AAAAPPPPPPPPLLLLYYYY100.00d
100.00deg
100.00d100.00d
are). When you have applied this simulated
UNITTTT
UNIUNI
, then
AAAAPPPPPPPPLLLLYYYY200.000d
200.000deeeeggggCCCC
200.000d200.000d
are). When you have applied this simulated
UNITTTT
UNIUNI
(or whatever the settings of
(or whatever the settings of
egCCCC
egeg
(or whatever the settings of
and the instrument will return to the RUN mode.
page 18 008-0608-01
Page 19
2.6.7
DDDDAAAACCCC....CHANNE
CHANNELLLL
CHANNECHANNE
DDDDAAAACCCC....SOURC
SOURCEEEE
SOURCSOURC
DA
DACCCCSETUP
SETUP
DADA
SETUPSETUP
Sub-Menu
This sub-menu contains four items that control the Digital-to-Analog Converter
(DAC) output of the channel.
Menu ItemThis chooses which channel will drive the DAC output. Normally, the DAC located
on a particular channel will be driven by that channel, but that need not be the
case. For example, if a Mathematics Virtual channel is installed on the instrument, the output of this Virtual channel could be used to drive this channel’s DAC
output.
If the DAC is set-up to be driven by its own channel’s tracking value, the Analog
Output will have the frequency response specified by “Channel Specifications” on
page 13. Otherwise, the Analog Output will operate more slowly.
Menu ItemThis designates the data source of the channel monitored by the analog output.
Each channel has three data sources: the live tracking value (TRACK), its highest
value (PEAK), and its lowest value (VALLEY).
The options for this menu item are:
•“
•“
” means the live tracking value of the channel.
TTTTRRRRAAAACCCCKKKK
” means the highest value of the channel since the peak/valley detector
PPPPEEEEAAAAKKKK
was last cleared.
•“
” means the lowest value of the channel since the peak/valley detector
VVVVAAAALLLLLLLLEEEEYYYY
was last cleared.
If the DAC is set-up to be driven by its own channel’s tracking value, the Analog
Output will have the frequency response specified by “Channel Specifications” on
page 13. Otherwise, the Analog Output will operate more slowly.
DA
DACCCC....ZERO-SCAL
ZERO-SCALEEEE
DADA
ZERO-SCALZERO-SCAL
DA
DACCCC....FULL-SCAL
FULL-SCALEEEE
DADA
FULL-SCALFULL-SCAL
Menu ItemThis specifies what value, in engineering units, corresponds to zero output on the
Analog Output.
“Zero output” might be 0 Volts, 2.5 Volts, 5 Volts, 4 mA or 12 mA depending on if
the channel has a voltage or current output and how it is configured. See the
“Digital-to-Analog Output” section of this chapter for details.
Menu ItemSpecifies what value, in engineering units, corresponds to full output on the Ana-
log Output.
“Full output” might be 5 Volts, 10 Volts or 20 mA depending on if the channel has
a voltage or current output and how it is configured. See the “Digital-to-Analog
Output” section of this chapter for details.
RTD INPUT CHANNEL2
Additional Channels for SC Series Instrumentation page 19
Page 20
2.6.8
DIAG
DIAGNNNNOSTICS
OSTICS
DIAGDIAG
OSTICSOSTICS
Sub-Menu
DDDDAAAACCCCFFFFUUUULLLLLS
LSCCCCAAAALLLLEEEE
LSLS
DDDDAAAACCCCZZZZEEEERRRROS
OSCCCCAAAALLLLEEEE
OSOS
VVVVEEEERRRRSSSSIIIIOOOONNNNINF
LLLLIIIINNNNEEEEAAAARRRRIIIIZZZZAAAATTTTIIIION
INFOOOO
INFINF
DDDDIIIISSSSPPPPLLLLAAAAYYYYAD
ADCCCC
ADAD
ON
ONON
Menu ItemWhen this menu item is selected, the Analog Output of the channel is forced to its full-
scale output, then
DDDDAAAACCCCUUUUPPPPDDDDAAAATTTTEEEEDDDD
is displayed. This is useful when calibrating or trimming
the readout connected to the Analog Output.
Menu ItemWhen this menu item is selected, the Analog Output of the channel is forced to its
zero-scale output, then
DDDDAAAACUP
CUPDDDDAAAATTTTEEEEDDDD
CUPCUP
is displayed. This is useful when calibrating or trim-
ming the readout connected to the Analog Output.
Menu ItemThis menu item displays the part number and revision level of the firmware used by
this channels microprocessor.
Menu ItemThe options for this menu item are:
•“
•“
” will allow the [VALUE] button to cycle through “
OOOOFFFFFFFF
value) and “
” will allow the “
OOOONNNN
” (valley value). This is the recommended option.
LLLLOOOO
” (percentage of the Analog-to-Digital converter’s full-scale
AAAADDDD
” (tracking value), “
TTTTKKKK
digitizing capability) display source to be available along with “
“
” (peak value) and “
HHHHIIII
” (valley value) when the [VALUE] button is pressed on
LLLLOOOO
” (peak
HHHHIIII
” (tracking value),
TTTTKKKK
the front panel.
The Analog-to-Digital converter counts are displayed as a percentage from -100.00%
to 100.00%.
This can be used to verify the correct operation of the input circuitry of the channel.
Menu ItemThe options for this menu item are:
•“
” will allow the linearization data obtained from the 5-Point Known Load Cali-
OOOONNNN
bration to affect a channel’s scaled values. This is the recommended option.
•“
” will not allow linearization data to effect the channel’s scaled values. This is
OOOOFFFFFFFF
useful when diagnosing problems in a calibration procedure.
” will bypass the INITIALIZE and RUN modes of the channel. The track, peak
OOOONNNN
and valley values of the channel are forced to 0. This option is helpful when a
channel is installed but not being used.
•“
” will allow normal operation of the channel.
OOOOFFFFFFFF
Page 21
2.7 Analog Output Configuration
2.7.1 Identifying the
Output Type
2.7.2 Channel Menu
Items
2.7.3 Output Selection
A RTD Input channel is available with one of two types of digital-to-analog (DAC)
outputs: voltage or current. You can determine which type of output a channel
has by one of three ways:
•Consulting the instrument’s Customer Information Sheet
•Examining the
SSSSYYYYSSSSTTTTEEEEMMMMMEN
nn is the number of the channel. If the channel’s type is
age output. If the channel’s type is
Additional Channels for SC Series Instrumentation page 21
Page 22
2.8 Troubleshooting
2.8.1 Error
Messages
See the “Error Messages” chapter of the “SC Series Instruction Manual”, Sensotec p/
n 008-0608-00, for information relating to error messages.
2.8.2 Common
Problems and
Solutions
Erratic DisplayMake sure that the temperature on the RTD is constant. A RTD simulator/calibrator
may be used.
The RTD Input channel is calibrated at the factory for Pt100 RTD’s with the DIN stan-
dard output curve. See “Channel Specifications” on pag e13.
It is recommended that the RTD Input channel be calibrated at -200ºC, -100ºC, 0ºC,
+100ºC and +200ºC for rated accuracy.
+OVLD or -OVLD on
Display
Analog Output
Incorrect
Check “Wiring” on page 10 and make certain that nothing is connected to the pins
labeled as “N/C”.
The RTD Input channel is calibrated at the factory for Pt100 RTD’s with the DIN stan-
dard output curve. See “Channel Specifications” on pag e13.
Make certain of the type of Analog Output (voltage or current) that the channel is
equipped with; see “Identifying the Output Type” on page21.
Use the
DDDDIIIIAAAAGGGGNO
NOSSSSTTTTIIIICCCCSSSS
NONO
->
DDDDAC
ACFFFFUL
ACAC
ULLLLLSC
SCAAAALLLLEEEE
ULUL
SCSC
and
DDDDIIIIAAAAGGGGNNNNOOOOSSSSTTTTIIIICCCCSSSS
->
DDDDAAAACZ
CZEEEERRRROS
CZCZ
OSCCCCAAAALLLLEEEE
OSOS
menu items to
force the Analog Output to a known output. Then, adjust your readout device, panel
meter, PLC or data acquisition system to match.
Auxiliary Function
Pins Not Operating
Check the output selection jumpers; see “Output Selection” on page 21.
Make sure that the
if they are set to
DDDDIIIISSSSAAAABBBBLLLLEEEEDDDD
AAAAUX
UX1111FFFFUUUUNNNNCCCCTTTTIIIIOOOONNNN
UXUX
then they will not operate.
or
AAAAUUUUXXXX2F
2FUUUUNNNNCCCCTTTTIIIIOOOONNNN
2F2F
SETUP menu items are set correctly;
The Auxiliary Function (“AUX1” and “AUX2”) pins must be connected to pin 10, not pin
8, to activate them.
page 22 008-0608-01
Page 23
2.9 Communication Commands
2.9.1 Introduction
2.9.2 Descriptions
This section only documents the communication commands that affect the operation
of the RTD Input channel. Refer to document p/n 008-0610-00, “SC Series Communications Guide” for details regarding serial communications. This document can be
downloaded from Sensotec’s web site, www.sensotec.com.
F0Transmit Track Data
PurposeTo transmit the channel’s tracking data value
Usage"
Example"
Reply"
RemarksThe input channel continuously reads the RTD. This function
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵ is the ‘carriage return’ character (ASCII decimal 13).
#0001F0
0037.5
transmits the most recent reading.
↵"
F0
↵"
” (typical)
F9Transmit Peak Data
PurposeTo transmit the channel’s peak data value
Usage"
Example"
Reply"
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵ is the ‘carriage return’ character (ASCII decimal 13).
#0001F9
0162.5
↵"
F9
↵"
” (typical) or "
N/A
”
RTD INPUT CHANNEL2
Additional Channels for SC Series Instrumentation page 23
Page 24
FATransmit Valley Data
PurposeTo transmit the channel’s valley data value
Usage"
Example"
Reply"
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵ is the ‘carriage return’ character (ASCII decimal 13).
#0001FA
-0012.5
↵"
FA
↵"
” (typical) or "
N/A
”
FBClear Peak and Valley Data
PurposeTo reset the channel’s peak and valley data values to the track
value.
Usage"
Example"
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵ is the ‘carriage return’ character (ASCII decimal 13).
#0001FB
FB
↵"
↵"
Reply"
RemarksThis command is not available on the Model SC1000.
OK
” or "
N/A
”
FFTransmit Analog-to-Digital Converter
Reading
PurposeTo transmit the Analog-to-Digital converter reading as a per-
centage from -100% to +100% of the A/D converters’s full
scale.
Usage"
Example"
Reply"
Remarksnone
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵ is the ‘carriage return’ character (ASCII decimal 13).
#0001FF
50.000
↵"
FF
↵"
” (typical)
page 24 008-0608-01
Page 25
FHWrite DAC Control Value
PurposeTo control the channel’s Digital-to-Analog Converter (DAC)
manually or automatically.
Usage"
aaccFHn
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
n
is the argument defined below.
↵"
↵ is the ‘carriage return’ character (ASCII decimal 13).
Argument"
” returns the DAC to its normal automatic operation
AUTO
(power-on default). That is, it is controlled by the "
"
DDDDAAAACCCC.SO
.SOUUUURRRRCCCCEEEE
.SO.SO
”, "
DDDDAAAACCCC.Z
.ZEEEERRRROOOO----SSSSCCCCAAAALLLLEEEE
.Z.Z
” and "
DDDDAAAACCCC.F
.FUUUULLLLLLLL----SSSSCCCCAAAALLLLEEEE
.F.F
DDDDAAAACCCC....CCCCHHHHAAAANNNNNNNNEEEELLLL
”,
” menu items.
If the argument is a numeric value between -1 and +1, the
DAC is forced to manual control and the value is used to drive
the DAC from -100% to +100% of its output.
Example"
#0001FH.5
↵" will force channel 01’s DAC to +50% of full-
scale.
Reply"
OK
” or "
ERROR
”
RemarksYou might wish to force a channel’s DAC to a certain output to
help calibrate the attached data acquisition system or other
device.
R6Read Display Units
PurposeReads the channel’s "
Usage"
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵" to read.
R6
DDDDIS
ISPPPPLLLLAAAAYYYYUNIT
ISIS
↵ is the ‘carriage return’ character (ASCII decimal 13).
Example"
#0001R6
↵”
ReplyWhen reading: a four-character string.
” menu item
UNITSSSS
UNITUNIT
RTD INPUT CHANNEL2
Additional Channels for SC Series Instrumentation page 25
Page 26
RK/WKRead/Write Known-Load Calibration
Point
PurposeReads or writes the channel’s "
Usage"
aaccRKpp
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
pp
is the parameter defined below.
n
is the known-load calibration point in engineering units.
↵" to read, "
aaccWKppn
#
↵ is the ‘carriage return’ character (ASCII decimal 13).
Parameter"
Example"
ReplyWhen writing: "OK
” accesses the "
00
"
” accesses the "
01
"
” accesses the "
02
"
” accesses the "
03
"
” accesses the "
04
#0001RK01
↵”
KKKKNNNNOOOOWWWWNNNNPOINT1
POINT1////5555
POINT1POINT1
KKKKNNNNOOOOWWWWNNNNPOINT2
POINT2////5555
POINT2POINT2
KKKKNNNNOOOOWWWWNNNNPOINT3
POINT3////5555
POINT3POINT3
KKKKNNNNOOOOWWWWNNNNPOINT4
POINT4////5555
POINT4POINT4
KKKKNNNNOOOOWWWWNNNNPOINT5
POINT5////5555
POINT5POINT5
" or "
ERROR
”.
When reading: an ASCII-floating-point value.
Changing this value has no effect until the channel is re-calibrated to the RTD with the Known-Load Calibration method.
KKKKNNNNOOOOWWWWNNNNPOINTx
POINTx////yyyy
POINTxPOINTx
↵" to write.
” menu item.
” menu item.
” menu item.
” menu item.
” menu item.
” menu items
page 26 008-0608-01
Page 27
RM/WMRead/Write DAC Channel & Value
PurposeReads or writes the channel’s "
DDDDAAAACCCC....CHANNE
CHANNELLLL
CHANNECHANNE
” and "
DDDDAC
AC....SOURC
SOURCEEEE
ACAC
SOURCSOURC
menu items
Usage"
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number.
n
is the argument defined below.
↵" to read, "
RM
aaccWMn
#
↵" to write.
↵ is the ‘carriage return’ character (ASCII decimal 13).
ArgumentA number which selects a channel’s value. This number is
created by adding together the values of the desired options
as shown.
ExampleIt is desired to disable the [TARE] button for channel 02 when
the Protection jumper is installed. Sending "
#0002WT1
↵" will
accomplish this.
ReplyWhen writing: "
OK
" or "
ERROR
”.
When reading: an ASCII-floating-point value described above.
page 30 008-0608-01
RemarksIf a disabled front panel button is pressed, the message "
” will appear on the display.
TTTTEEEECCCCTTTTEEEEDDDD
PPPPRRRROOOO----
Page 31
RTD INPUT CHANNEL2
Additional Channels for SC Series Instrumentation page 31
Page 32
page 32 008-0608-01
Page 33
3.1 Features
Chapter 3
Frequency Input Channel
The Frequency Input channel accepts square wave or sine wave input signals
from such transducers as magnetic pickups, devices with TTL outputs, or devices
with PNP or NPN transistor outputs. These signals are AC coupled into a zerocrossing detector, digitized, converted into engineering units, and placed into the
track, peak and valley data values of the channel. A 5V excitation is available in
case it is required by the device.
Setup and calibration of the channel is made at the factory. Re-calibration can be
made manually through the SETUP mode with the known-load calibration method
and a customer-supplied frequency calibrator.
The engineering units used to display the frequency are Hz. A Mathematics Virtual channel can be factory programmed for specific applications that required
conversion into units of RPM or horsepower.
Two rear panel control inputs can be field-configured for such functions as remote
track/hold, disabling peak/valley detection and clearing the peak/valley values. A
voltage or current digital-to-analog output is also provided.
3.2 Getting Started Quickly
As shipped from the factory, the Frequency Input channel is calibrated and ready
for use.
Connect the output of your transdcuer to the Frequency Input channel according
to “Wiring” on pag e34.
FREQUENCY INPUT CHANNEL3
Additional Channels for SC Series Instrumentation page 33
Page 34
3.3 Wiring
3.3.1 Channel
Connector
The pin-out for the channel’s connector is shown in the following table.
Table 3-1: Frequency Input Channel Pin Connections
PinLabelFunctionNOTES
1 (top)+EXC(+)Excitationreferenced to pin 10
2N/C N/CDO NOT CONNECT!
3N/C N/CDO NOT CONNECT!
4-EXC(-)Excitationreferenced to pin 10
5+SIG(+)Signalreferenced to pin 10
6-SIG(-)Signalreferenced to pin 10
7+OUTAnalog Outputreferenced to pin 8
8-OUTAnalog Return-
9N/C N/CDO NOT CONNECT!
10DGNDDigital Ground-
11AUX1Auxiliary Function 1
(connect to pin 10 to activate)
12
(bottom)
The Analog Output and Analog Return pins are electrically isolated from all other pins
on the instrument.
AUX2Auxiliary Function 2
(connect to pin 10 to activate)
referenced to pin 10
referenced to pin 10
3.3.2 NPN OpenCollector
Connection
Do not connect anything to the pins labeled as “N/C”. Damage to the instru-
ment or loss of accuracy may result.
.
Figure 3-1: NPN Open-Collector Connection to Frequency Input Channel
page 34 008-0608-01
Page 35
3.3.3 PNP OpenCollector
3.3.4 TTL/CMOS Logic
Connection
Figure 3-2: PNP Open-Collector Connection to Frequency Input Channel
3.3.5 Magnetic Pickup
Connection
Figure 3-3: TTL or CMOS Logic Connection to Frequency Input Channel
FREQUENCY INPUT CHANNEL3
Figure 3-4: Magnetic Pickup Connection to Frequency Input Channel
Additional Channels for SC Series Instrumentation page 35
Page 36
3.4 Calibration
A listing of all menu items for this channel is given in “Setup Menu Reference” on
page 55.
A customer-supplied signal generator is required for re-calibration. It should be set up
to produce a square wave output at 1V peak-to-peak amplitude.
Step 1: Wire the signal generator to the channel’s connector.
See “Wiring” on page 34.
Step 2: Enter the
See “
CALIBRATIONDATA
CCCCAAAALLLLIB
IBRRRRAAAATTTTIIIIOOOONNNNDAT
IBIB
.
DATAAAA
DATDAT
Sub-Menu” on page40.
Step 3: Perform the calibration.
See “
CALIBRATE
Using the
Menu Item” on page 40.
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTEEEE
menu item starts the calibration process. You will be
prompted to use your signal generator to apply signals at various frequencies to
the channel as required.
Output voltage range5, ±5, 10 or ±10 VDC (field selectable)
Output current range (optional current
output channels)
Sourceany channel’s track, peak or valley value
Isolation500V
Resolution13 bits
Frequency Responsesame as input when driven by the same
track hold
FREQUENCY INPUT CHANNEL3
4-20 mA
channel’s tracking data
Additional Channels for SC Series Instrumentation page 37
Page 38
3.6 Channel Menu
The Frequency Input channel is configured and calibrated via its channel menu.
Detailed instructions on operating the instrument in the SETUP Menu mode can be
found in the “SC Series Instruction Manual”, Sensotec p/n 008-0608-00.
3.6.1
Sub-Menu
PPPPOOOOWWWWEEEERRRR----OOOONNNNSOURC
3.6.2
SOURCEEEE
SOURCSOURC
AUXnF
AUXnFUUUUNCTION
AUXnFAUXnF
Menu Items
OP
OPEEEERATION
RATION
OPOP
RATIONRATION
Menu
Item
NCTION
NCTIONNCTION
This menu controls the operation of this channel when the instrument is in the RUN
mode.
This menu selects which value is displayed by the channel when first entering the
RUN mode.
The choices are:
•“
•“
•“
The
” means the live tracking value of the channel.
TTTTRRRRAAAACCCCKKKK
” means the highest value of the channel.
PPPPEEEEAAAAKKKK
” means the lowest value of the channel.
VVVVAAAALLLLLLLLEEEEYYYY
AAAAUUUUXXXX1F
1FUN
1F1F
UNCCCCTTTTIIIIOOOONNNN
UNUN
and
AAAAUUUUXXXX2F
2FUUUUNNNNCCCCTTTTIIIIOOOONNNN
2F2F
menu items determine what happens when the
Auxiliary Function pins (labeled as “AUX1” and “AUX2”) on the channel’s connector
are activated. These pins are “activated” when they are connected to the Digital
Ground (labeled as “DGND”) pin. The choices are:
•“
DDDDIIIISSSSAAAABBBBLLLLEEEEDDDD
•“
TTTTRRRRAAAACCCCKKKKHOL
•“
HHHHIIIIGGGGHHHH////LLLLOOOOWWWWHOL
•“
HHHHIIIIGGGGHHHH////LLLLOOOOWWWWCLEA
” means that activating the pin does nothing.
” means that the tracking, peak and valley values will not be updated.
HOLDDDD
HOLHOL
” means that the peak and valley values will not be updated.
HOLDDDD
HOLHOL
” means that the peak and valley values are reset.
CLEARRRR
CLEACLEA
As the Auxiliary Function pins are not isolated, it is recommended that a push-button
switch or relay is used to connect these pins to the Digital Ground pin.
page 38 008-0608-01
Page 39
3.6.3
CCCCALIBRATI
ALIBRATIOOOONTYPE
ALIBRATIALIBRATI
NTYPE
NTYPENTYPE
Menu Item
The only choice is:
•“
TTTTYYYYPPPPEEEE=5
=5PPPPOOOOIIIINNNNTC
=5=5
” means 5-Point Known Load Calibration. You are prompted
TCAAAALLLL
TCTC
to use your singal generator to apply the frequencies that were entered in the
“
KKKKNNNNOOOOWWWWNPO
NPOIIIINNNNTTTT1111////5555
NPONPO
PPPPOOOOIIIINNNNTTTT5/
5/5555
5/5/
”, “
KKKKNNNNOOOOWWWWNP
NPOOOOIIIINNNNTTTT2222////5555
NPNP
”, “
KKKKNNNNOOOOWWWWNNNNPPPPOOOOIIIINNNNT3/
T3/5555
T3/T3/
”, “
KKKKNNNNOOOOWN
WNPPPPOOOOIIIINNNNT4/
T4/5555
WNWN
T4/T4/
” and “
KKKKNNNNOOOOWWWWNNNN
” registers. This technique is used to compensate for the non-linear-
ity inherent in the channel’s frequency-to-voltage converter.
The calibration points used for factory calibration are 0 Hz, 35kHz, 45kHz,
47.5kHz and 50kHz.
Choosing calibration points other than these will result in greatly reduced
accuracy.
FREQUENCY INPUT CHANNEL3
Additional Channels for SC Series Instrumentation page 39
Page 40
3.6.4
DA
DATTTTAAAA
DADA
CALI
CALIBBBBRATION
RATION
CALICALI
RATIONRATION
Sub-Menu
KKKKNNNNOOOOWWWWNNNNPOINTx/
POINTx/yyyy
POINTx/POINTx/
3.6.5
CA
CALLLLIBRATE
CACA
Menu Item
Menu
Items
IBRATE
IBRATEIBRATE
This enters the engineering unit values for the known-load calibration points. These
points must match the actual frequencies that you will apply to the instrument during
calibration. The following menu items are available:
•“
KKKKNNNNOOOOWWWWNNNNPOINT1/
•“
KKKKNNNNOOOOWWWWNNNNPOINT2/
•“
KKKKNNNNOOOOWWWWNNNNPOINT3/
•“
KKKKNNNNOOOOWWWWNNNNPOINT4/
•“
KKKKNNNNOOOOWWWWNNNNPOINT5/
POINT1/5555
POINT1/POINT1/POINT2/5555
POINT2/POINT2/POINT3/5555
POINT3/POINT3/POINT4/5555
POINT4/POINT4/POINT5/5555
POINT5/POINT5/
”: point 1 of 5, must be 0Hz.
”: point 2 of 5, must be 35kHz.
”: point 3 of 5, must be 45kHz.
”: point 4 of 5, must be 47.5kHz.
”: point 5 of 5, must be 50kHz.
The instrument expects the frequency applied at each known-load point to be
increasing. For example, the frequency applied at Known-Load Point 2/5
must be greater than the frequency applied at Known-Load Point 1/5.
This menu item performs a calibration according to what was entered in the
and
TTTTIIIIOOOONNNNTYP
TYPEEEE
TYPTYP
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNNDAT
Before performing a calibration, the
“
CALIBRATIONDATA
menu items.
DATAAAA
DATDAT
CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNNDAT
must be entered (see
DATAAAA
DATDAT
Sub-Menu” on page40).
CCCCAAAALLLLIIIIBBBBRRRRAAAA----
For maximum accuracy, allow at least five minutes of warm-up before calibra-
This sub-menu contains four items that control the Digital-to-Analog Converter
(DAC) output of the channel.
Menu ItemThis chooses which channel will drive the DAC output. Normally, the DAC located
on a particular channel will be driven by that channel, but that need not be the
case. For example, if a Mathematics Virtual channel is installed on the instrument, the output of this Virtual channel could be used to drive this channel’s DAC
output.
If the DAC is set-up to be driven by its own channel’s tracking value, the Analog
Output will have the frequency response specified by “Channel Specifications” on
page 37. Otherwise, the Analog Output will operate more slowly.
Menu ItemThis designates the data source of the channel monitored by the analog output.
Each channel has three data sources: the live tracking value (TRACK), its highest
value (PEAK), and its lowest value (VALLEY).
The options for this menu item are:
•“
•“
” means the live tracking value of the channel.
TTTTRRRRAAAACCCCKKKK
” means the highest value of the channel since the peak/valley detector
PPPPEEEEAAAAKKKK
was last cleared.
•“
” means the lowest value of the channel since the peak/valley detector
VVVVAAAALLLLLLLLEEEEYYYY
was last cleared.
If the DAC is set-up to be driven by its own channel’s tracking value, the Analog
Output will have the frequency response specified by “Channel Specifications” on
page 37. Otherwise, the Analog Output will operate more slowly.
DA
DACCCC....ZERO-SCAL
ZERO-SCALEEEE
DADA
ZERO-SCALZERO-SCAL
DA
DACCCC....FULL-SCAL
FULL-SCALEEEE
DADA
FULL-SCALFULL-SCAL
Menu ItemThis specifies what value, in engineering units, corresponds to zero output on the
Analog Output.
“Zero output” might be 0 Volts, 2.5 Volts, 5 Volts, 4 mA or 12 mA depending on if
the channel has a voltage or current output and how it is configured. See the
“Digital-to-Analog Output” section of this chapter for details.
Menu ItemSpecifies what value, in engineering units, corresponds to full output on the Ana-
log Output.
“Full output” might be 5 Volts, 10 Volts or 20 mA depending on if the channel has
a voltage or current output and how it is configured. See the “Digital-to-Analog
Output” section of this chapter for details.
FREQUENCY INPUT CHANNEL3
Additional Channels for SC Series Instrumentation page 41
Page 42
3.6.7
DIAG
DIAGNNNNOSTICS
OSTICS
DIAGDIAG
OSTICSOSTICS
Sub-Menu
DDDDAAAACCCCFFFFUUUULLLLLS
LSCCCCAAAALLLLEEEE
LSLS
DDDDAAAACCCCZZZZEEEERRRROS
OSCCCCAAAALLLLEEEE
OSOS
VVVVEEEERRRRSSSSIIIIOOOONNNNINF
LLLLIIIINNNNEEEEAAAARRRRIIIIZZZZAAAATTTTIIIION
INFOOOO
INFINF
DDDDIIIISSSSPPPPLLLLAAAAYYYYAD
ADCCCC
ADAD
ON
ONON
Menu ItemWhen this menu item is selected, the Analog Output of the channel is forced to its full-
scale output, then
DDDDAAAACCCCUUUUPPPPDDDDAAAATTTTEEEEDDDD
is displayed. This is useful when calibrating or trimming
the readout connected to the Analog Output.
Menu ItemWhen this menu item is selected, the Analog Output of the channel is forced to its
zero-scale output, then
DDDDAAAACUP
CUPDDDDAAAATTTTEEEEDDDD
CUPCUP
is displayed. This is useful when calibrating or trim-
ming the readout connected to the Analog Output.
Menu ItemThis menu item displays the part number and revision level of the firmware used by
this channels microprocessor.
Menu ItemThe options for this menu item are:
•“
•“
” will allow the [VALUE] button to cycle through “
OOOOFFFFFFFF
value) and “
” will allow the “
OOOONNNN
” (valley value). This is the recommended option.
LLLLOOOO
” (percentage of the Analog-to-Digital converter’s full-scale
AAAADDDD
” (tracking value), “
TTTTKKKK
digitizing capability) display source to be available along with “
“
” (peak value) and “
HHHHIIII
” (valley value) when the [VALUE] button is pressed on
LLLLOOOO
” (peak
HHHHIIII
” (tracking value),
TTTTKKKK
the front panel.
The Analog-to-Digital converter counts are displayed as a percentage from -100.00%
to 100.00%.
This can be used to verify the correct operation of the input circuitry of the channel.
Menu ItemThe options for this menu item are:
•“
” will allow the linearization data obtained from the 5-Point Known Load Cali-
OOOONNNN
bration to affect a channel’s scaled values. This is the recommended option.
•“
” will not allow linearization data to effect the channel’s scaled values. This is
OOOOFFFFFFFF
useful when diagnosing problems in a calibration procedure.
” will bypass the INITIALIZE and RUN modes of the channel. The track, peak
OOOONNNN
and valley values of the channel are forced to 0. This option is helpful when a
channel is installed but not being used.
•“
” will allow normal operation of the channel.
OOOOFFFFFFFF
Page 43
3.7 Analog Output Configuration
3.7.1 Identifying the
Output Type
3.7.2 Channel Menu
Items
3.7.3 Output Selection
A Frequency Input channel is available with one of two types of digital-to-analog
(DAC) outputs: voltage or current. You can determine which type of output a
channel has by one of three ways:
•Consulting the instrument’s Customer Information Sheet
•Examining the
SSSSYYYYSSSSTTTTEEEEMMMMMEN
nn is the number of the channel. If the channel’s type is
voltage output. If the channel’s type is
Additional Channels for SC Series Instrumentation page 43
Page 44
3.8 Troubleshooting
3.8.1 Error
Messages
See the “Error Messages” chapter of the “SC Series Instruction Manual”, Sensotec p/
n 008-0608-00, for information relating to error messages.
3.8.2 Common
Problems and
Solutions
Erratic DisplayCheck “Wiring” on page 34 and make certain that nothing is connected to the pins
labeled as “N/C”.
Incorrect display
readings over 50kHz
Analog Output
Incorrect
Auxiliary Function
Pins Not Operating
The Frequency Input channel can only digitize signals with the minimum pulse specified in “Channel Specifications” on pag e37.
Make certain of the type of Analog Output (voltage or current) that the channel is
equipped with; see “Identifying the Output Type” on page43.
Use the
DDDDIIIIAAAAGGGGNO
NOSSSSTTTTIIIICCCCSSSS
NONO
->
DDDDAC
ACFFFFUL
ACAC
ULLLLLSC
SCAAAALLLLEEEE
ULUL
SCSC
and
DDDDIIIIAAAAGGGGNNNNOOOOSSSSTTTTIIIICCCCSSSS
->
DDDDAAAACZ
CZEEEERRRROS
CZCZ
OSCCCCAAAALLLLEEEE
OSOS
menu items to
force the Analog Output to a known output. Then, adjust your readout device, panel
meter, PLC or data acquisition system to match.
Check the output selection jumpers; see “Output Selection” on page 43.
Make sure that the
if they are set to
DDDDIIIISSSSAAAABBBBLLLLEEEEDDDD
AAAAUX
UX1111FFFFUUUUNNNNCCCCTTTTIIIIOOOONNNN
UXUX
then they will not operate.
or
AAAAUUUUXXXX2F
2FUUUUNNNNCCCCTTTTIIIIOOOONNNN
2F2F
SETUP menu items are set correctly;
The Auxiliary Function (“AUX1” and “AUX2”) pins must be connected to pin 10, not pin
8, to activate them.
page 44 008-0608-01
Page 45
3.9 Communication Commands
3.9.1 Introduction
3.9.2 Descriptions
This section only documents the communication commands that affect the operation of the Frequency Input channel. Refer to document p/n 008-0610-00, “SC
Series Communications Guide” for details regarding serial communications. This
document can be downloaded from Sensotec’s web site, www.sensotec.com.
F0Transmit Track Data
PurposeTo transmit the channel’s tracking data value
Usage"
Example"
Reply"
RemarksThe input channel continuously reads the RTD. This function
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵ is the ‘carriage return’ character (ASCII decimal 13).
#0001F0
0037.5
transmits the most recent reading.
↵"
F0
↵"
” (typical)
F9Transmit Peak Data
PurposeTo transmit the channel’s peak data value
Usage"
Example"
Reply"
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵ is the ‘carriage return’ character (ASCII decimal 13).
#0001F9
0162.5
↵"
F9
↵"
” (typical) or "
N/A
”
FREQUENCY INPUT CHANNEL3
Additional Channels for SC Series Instrumentation page 45
Page 46
FATransmit Valley Data
PurposeTo transmit the channel’s valley data value
Usage"
Example"
Reply"
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵ is the ‘carriage return’ character (ASCII decimal 13).
#0001FA
-0012.5
↵"
FA
↵"
” (typical) or "
N/A
”
FBClear Peak and Valley Data
PurposeTo reset the channel’s peak and valley data values to the track
value.
Usage"
Example"
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵ is the ‘carriage return’ character (ASCII decimal 13).
#0001FB
FB
↵"
↵"
Reply"
RemarksThis command is not available on the Model SC1000.
OK
” or "
N/A
”
FFTransmit Analog-to-Digital Converter
Reading
PurposeTo transmit the Analog-to-Digital converter reading as a per-
centage from -100% to +100% of the A/D converters’s full
scale.
Usage"
Example"
Reply"
Remarksnone
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵ is the ‘carriage return’ character (ASCII decimal 13).
#0001FF
50.000
↵"
FF
↵"
” (typical)
page 46 008-0608-01
Page 47
FHWrite DAC Control Value
PurposeTo control the channel’s Digital-to-Analog Converter (DAC)
manually or automatically.
Usage"
aaccFHn
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
n
is the argument defined below.
↵"
↵ is the ‘carriage return’ character (ASCII decimal 13).
Argument"
” returns the DAC to its normal automatic operation
AUTO
(power-on default). That is, it is controlled by the "
"
DDDDAAAACCCC.SO
.SOUUUURRRRCCCCEEEE
.SO.SO
”, "
DDDDAAAACCCC.Z
.ZEEEERRRROOOO----SSSSCCCCAAAALLLLEEEE
.Z.Z
” and "
DDDDAAAACCCC.F
.FUUUULLLLLLLL----SSSSCCCCAAAALLLLEEEE
.F.F
DDDDAAAACCCC....CCCCHHHHAAAANNNNNNNNEEEELLLL
”,
” menu items.
If the argument is a numeric value between -1 and +1, the
DAC is forced to manual control and the value is used to drive
the DAC from -100% to +100% of its output.
Example"
#0001FH.5
↵" will force channel 01’s DAC to +50% of full-
scale.
Reply"
OK
” or "
ERROR
”
RemarksYou might wish to force a channel’s DAC to a certain output to
help calibrate the attached data acquisition system or other
device.
R6Read Display Units
PurposeReads the channel’s "
Usage"
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
↵" to read.
R6
DDDDIS
ISPPPPLLLLAAAAYYYYUNIT
ISIS
↵ is the ‘carriage return’ character (ASCII decimal 13).
Example"
#0001R6
↵”
ReplyWhen reading: a four-character string.
” menu item
UNITSSSS
UNITUNIT
FREQUENCY INPUT CHANNEL3
Additional Channels for SC Series Instrumentation page 47
Page 48
RK/WKRead/Write Known-Load Calibration
Point
PurposeReads or writes the channel’s "
Usage"
aaccRKpp
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number
pp
is the parameter defined below.
n
is the known-load calibration point in engineering units.
↵" to read, "
aaccWKppn
#
↵ is the ‘carriage return’ character (ASCII decimal 13).
Parameter"
Example"
ReplyWhen writing: "OK
” accesses the "
00
"
” accesses the "
01
"
” accesses the "
02
"
” accesses the "
03
"
” accesses the "
04
#0001RK01
↵”
KKKKNNNNOOOOWWWWNNNNPOINT1
POINT1////5555
POINT1POINT1
KKKKNNNNOOOOWWWWNNNNPOINT2
POINT2////5555
POINT2POINT2
KKKKNNNNOOOOWWWWNNNNPOINT3
POINT3////5555
POINT3POINT3
KKKKNNNNOOOOWWWWNNNNPOINT4
POINT4////5555
POINT4POINT4
KKKKNNNNOOOOWWWWNNNNPOINT5
POINT5////5555
POINT5POINT5
" or "
ERROR
”.
When reading: an ASCII-floating-point value.
Changing this value has no effect until the channel is re-calibrated to the RTD with the Known-Load Calibration method.
KKKKNNNNOOOOWWWWNNNNPOINTx
POINTx////yyyy
POINTxPOINTx
↵" to write.
” menu item.
” menu item.
” menu item.
” menu item.
” menu item.
” menu items
page 48 008-0608-01
Page 49
RM/WMRead/Write DAC Channel & Value
PurposeReads or writes the channel’s "
DDDDAAAACCCC....CHANNE
CHANNELLLL
CHANNECHANNE
” and "
DDDDAC
AC....SOURC
SOURCEEEE
ACAC
SOURCSOURC
menu items
Usage"
aacc
#
#
is the ‘pound’ or ‘hash’ character (ASCII decimal 35).
aa
is the two-character instrument address.
cc
is the two-character channel number.
n
is the argument defined below.
↵" to read, "
RM
aaccWMn
#
↵" to write.
↵ is the ‘carriage return’ character (ASCII decimal 13).
ArgumentA number which selects a channel’s value. This number is
created by adding together the values of the desired options
as shown.
Additional Channels for SC Series Instrumentation page 57
Page 58
page 58 008-0608-01
Page 59
Index
C
Communications Guide 6
Customer Information Sheet 6
F
Frequency Input channel
Analog Output 41–44
Auxilliary Function pins 38, 44
calibration 36
communication commands 45
features 33
frequency response 37
Getting Started Quickly 33
remote control of Track/Hold, Peak/Valley clear,
Peak/Valley hold 38
SETUP menu 38
software revision 42
specifications 37
step response 37
troubleshooting 44
wiring 34
R
RTD Input channel
Analog Output 19, 21–22
Auxilliary Function pins 15, 22
calibration 12
communication commands 23
display
message 22
OVLD
units of measure 15
features 9
frequency response 13
getting started quickly 9
remote control of Track/Hold, Peak/Valley clear,
Peak/Valley hold 15
SETUP Menu 14
software revision 20
specifications 13
step response 13
troubleshooting 22
wiring 10
S
Serial Communications Guide 6
INDEX
Additional Channels for SC Series Instrumentation page 59
Page 60
page 60 008-0608-01
Page 61
Warranty/Repair Policy
Limitation of Remedy and
Disclaimer of Warranty
Obtaining Service Under
Warranty
Any of our products which, under normal operating conditions, proves defective in
material or in workmanship within one year from the date of shipment by Sensotec, Inc., will be repaired or replaced free of charge, provided that the buyer (1)
promptly notifies Sensotec, Inc. of any such defect; (2) provides Sensotec, Inc.
with satisfactory proof of the defect and that the product was properly installed,
maintained, and operated within the limits of rated and normal usage; and (3)
obtains from Sensotec, Inc. authorization to return the product. Any such product
shall be returned with transportation charges prepaid. The replacement product
will be shipped F.O.B. our plant.
The remedy set forth herein does not extend to any product or part thereof which,
under normal usage, has an inherently shorter useful life than one year. The remedy set forth herein does not apply to damage or to defects in any product caused
by the buyer’s misuse or neglect, nor does it apply to any product which has been
repaired or disassembled which, in the sole judgement of Sensotec, Inc. affects
the performance of the product.
The remedy set forth herein is the buyer's exclusive remedy, and will satisfy all
obligations of Sensotec, Inc. whether based on contract, negligence, or otherwise.
Sensotec, Inc. is not responsible for any incidental or consequential loss or damage which might result from a failure of any Sensotec, Inc. product.
THIS EXPRESS WARRANTY IS MADE IN LIEU OF ANY AND ALL OTHER
WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING IMPLIED WARRANTY
OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE.
Advanced authorization is required prior to the return to Sensotec, Inc. Before
returning the items, either write to the Repair Department c/o Sensotec, Inc., 2080
Arlingate Lane, Columbus, Ohio 43228, or call (800) 848-6564 with: 1) a part
number; 2) a serial number of the defective product; 3) a technical description* of
the defect; 4) a no-charge purchase order number (so products can be returned to
you correctly); and 5) ship and bill addresses. Shipment to Sensotec, Inc. shall be
at Buyer's expense and repaired or replacement items will be shipped F.O.B. our
plant in Columbus, Ohio. Non-verified problems or defects may be subject to an
evaluation charge. Please return the original calibration data with the unit.
WARRANTY/REPAIR POLICY
Obtaining Non-warranty
Service
Repair WarrantyAll repairs of Sensotec, Inc. products are warranted for a period of 90 days from
Advance authorization is required prior to the return to Sensotec, Inc. Before
returning the item, either write to the Repair Department c/o Sensotec, Inc., 2080
Arlingate Lane, Columbus, Ohio 43228, or call (800) 848-6564 with: 1) a part
number; 2) a serial number of the defective product; 3) a technical description* of
the malfunction; 4) a purchase order number to cover Sensotec, Inc.'s repair cost;
and 5) ship and bill addresses. After the product is evaluated by Sensotec, Inc.,
we will contact you to provide the estimated repair costs before proceeding. The
minimum evaluation charge is $95. Shipment to Sensotec, Inc. shall be at Buyer's
expense and repaired items will be shipped to you F.O.B., our plant in Columbus,
Ohio. Please return the original calibration data with the unit.
date of shipment. This warranty applies only to those items which were found
defective and repaired. It does not apply to products in which no defect was found
and returned as is or merely re-calibrated. Out of warranty products may not be
capable of being returned to the exact original specifications or dimensions.
*Technical description of the defect: In order to properly repair a product, it is necessary for
Sensotec, Inc. to receive information specifying the reason the product is being returned.
Specific test data, written observations on the failure and specific corrective action you
require is needed.
Additional Channels for SC Series Instrumentation page 61