Model 952 15
The display is reset to zero. 0.0 appears
in the display. The note “NET” appears.
– Now put the patient and the additional
weight on the scale.
The scales calculate the patient’s
weight without the additional
weight.You can now perform as many
weighing operations as you like.
– Hold the key down again to deactivate
the Tare function; the scale is then back
in normal weighing mode and the tare
value has been erased.
– The tare value is likewise erased by the
scale being switched off.
5. Cleaning
Clean the seat and the housing as required using a domestic cleaning agent or commercially-available disinfectant. Follow the manufacturer’s instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like for
cleaning. Such substances can damage the high-quality surfaces.
6. What do I do if…
… no weight display comes on?
– Is the scale switched on?
– Check the batteries.
… 0.0 does not appear before weighing?
– Switch off the scale with the Start key and start the scale again – there must not be
any load on the scale – and only its feet should be in contact with the floor.
… one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.
… the display shows ?
– Battery voltage is running low. Change the batteries in the next few days.
… batt appears in the display?
– Batteries are empty. Put in new batteries.
… St0p appears in the display?
– Maximum load has been exceeded.
… the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-
tected a steady weight value. The display will then stop flashing.
kg
0.0