Seametrics PT2X Instructions Manual

Page 1
Precision Environmental Sensors
d
C
An ON
I
CON Brand
PT2X
Pressure/Temperature Smart Sensor
and Data Logger Instructions
For PSIG
sensors, refer
to page 23
regarding
desiccant
use!
o
m
e
i
p
f
PROUDLY MADE IN THE
USA
i
t
r
e
ISO
9001:2008
a
n
y
Page 2
TABLE OF CONTENTS
General Information
General Information ........................................................................................................................................................................Page 3
Dimensions .........................................................................................................................................................................................Page 3
Specications .....................................................................................................................................................................................Page 4
How Pressure Sensors Work ........................................................................................................................................................Page 5
Initial Inspection and Handling ...................................................................................................................................................Page 6
Do’s and Don’ts .................................................................................................................................................................................Page 6
Installation
Connecting External Power ...........................................................................................................................................................Page 7
Connecting a PT2X to a Computer ............................................................................................................................................Page 7
Connecting to Sensors ...................................................................................................................................................................Page 8
Cable Wiring ......................................................................................................................................................................................Page 9
Installing Aqua4Plus2.0 ..................................................................................................................................................................Page 9
Using the PT2X without Aqua4Plus 2.0 ...................................................................................................................................Page 9
Installing the Sensor ........................................................................................................................................................................Page 10
Desiccant Use .....................................................................................................................................................................................Page 10
Grounding Issues ..............................................................................................................................................................................Page 10
Settings and Calibration
Sensor Settings ..................................................................................................................................................................................Page 11
Program Settings ..............................................................................................................................................................................Page 11
Pressure Calibration.........................................................................................................................................................................Page 12
Depth to water ..................................................................................................................................................................................Page 13
Ground Elevation ..............................................................................................................................................................................Page 14
Sta Gauge .........................................................................................................................................................................................Page 14
Removing calibration setup values ............................................................................................................................................Page 14
Operation
Real-time Data ...................................................................................................................................................................................Page 15
Data Logging......................................................................................................................................................................................Page 16
Reports .................................................................................................................................................................................................Page 18
A Word About Units ........................................................................................................................................................................Page 18
Barometric Compensation ............................................................................................................................................................Page 19
Direct Read Modbus/SDI-12
Setting Units for Direct Read .......................................................................................................................................................Page 20
Power Consideration .......................................................................................................................................................................Page 20
Reading via Modbus RTU ..............................................................................................................................................................Page 20
Reading via SDI-12...........................................................................................................................................................................Page 21
Maintenance
Removing Debris from End Cone ...............................................................................................................................................Page 23
Desiccant Tubes ................................................................................................................................................................................Page 23
Sensor/Cable/End Connections ..................................................................................................................................................Page 24
Changing Batteries ...........................................................................................................................................................................Page 24
Troubleshooting
Problems/Probable Causes/Things to Try ...............................................................................................................................Page 27
Warranty
Seametrics Limited Warranty .......................................................................................................................................................Page 28
PT2X INSTRUCTIONS
IF USING ALKALINE BATTERIES—PREVENT BATTERY LEAKAGE!
PT2X sensors are typically shipped with lithium batteries. If, however, you are using alkaline batteries, be aware that under some circumstances alkaline batteries can leak, causing damage to the sensor. To prevent leakage, the following is recommended. (Does not apply to lithium batteries.)
• Change the batteries at least every 18 months.
• If the sensor will not be deployed for 3 months or more, remove the batteries.
Seametrics • 253.872.0284 Page 2 seametrics.com
Page 3
GENERAL INFORMATION
PT2X INSTRUCTIONS
The Seametrics PT2X Smart Sensor is an integrated data logger and pressure/temperature sensor and is ideal for monitoring groundwater, well, tank, and tidal levels, as well as for pump and slug testing. This sensor networks with all of the Seametrics Smart Sensor family. IThe PT2X is a microprocessor based digital intelligent sensor designed to measure and record pressure, temperature, and time, using low power, battery operated circuitry.
Pressure is measured with an extremely rugged and stable piezo-electric media-isolated pressure element combined with a 16-bit analog-to-digital converter. This provides extremely accurate and stable pressure input into the microprocessor on the circuit board.
This industry standard digital RS485 interface device records up to 520,000 records of pressure/level, temperature, and time data, operates with low power, and features easy-to­use software with powerful features. Constructed with 316
stainless steel or titanium, PTFE, and uoropolymer, this
sensor provides high-accuracy readings in rugged and
corrosive eld conditions.
Two internal 1.5V AA batteries power the PT2X. (Auxiliary power supplies are available for data intensive applications.) The unit is programmed using Seametrics’ easy-to-use Aqua4plus 2.0 control software. Once programmed the unit will measure and collect data on a variety of time intervals.
Several PT2Xs, or a combination of PT2Xs and other Seametrics Smart Sensors, can be networked together and controlled directly from a single computer.
While most will use the PT2X with our free, easy-to-use Aqua4Plus 2.0 software, it is by no means limited to that software. You can use your own Modbus® RTU or SDI-12 software or logging equipment to read measurements, thus tying into your existing systems and databases.
Dimensions
Cableless
0.25” (0.6
cm) Shorter
0.28” (0.7 cm)
0.28” (0.7 cm)
12.18” (30.9 cm)
Battery Version
8.37” (21.3 cm)
Non-Battery Version
Diameter
0.75” (1.9 cm)
Seametrics • 253.872.0284 Page 3 seametrics.com
Page 4
GENERAL INFORMATION
PT2X INSTRUCTIONS
Specications*
Housing & Cable Weight 0.8 lb. (0.4 kg)
Body Material Acetal & 316 stainless or titanium Wire Seal
Material Cable Submersible: polyurethane, polyethylene, or ETFE (4 lb./100 ft., 1.8 kg/30 m) Desiccant 1-3 mm indicating silica gel Field Connector Standard
Temperature Operating Range Recommended: -15˚ to 55˚C (5˚ to 131˚F) Requires freeze protection kit if using pressure option in water
Storage Range Without batteries: -40˚ to 80˚C (-40˚ to 176˚F)
Power Internal Battery Two lithium ‘AA’ batteries - Expected battery life: 18 months at 15 min. polling interval (may vary do to
Auxiliary 12 Vdc - Nominal, 6-16 Vdc - range
Communication Modbus® RS485 Modbus® RTU, output=32bit IEEE oating point
SDI-12 SDI-12 (ver. 1.3) - ASCII
Logging Memory 4MB - 520,000 records
Logging Types Variable, user-dened, proled Logging Rates 8x/sec maximum, no minimum Baud Rates 9600, 19200, 38400 Software Complimentary Aqua4Plus 2.0 Networking 32 available addresses per junction (Address range: 1 to 255) File Formats .a4d and .csv
Output Channels Temperature Depth/Level¹
Element Digital IC on board Silicon strain gauge transducer, 316 stainless or Hastelloy Accuracy ±0.5°C — 0° to 55°C (32˚ to 131˚F)
Resolution 0.1˚C 0.0034% FS (typical) Units Celsius, Fahrenheit, Kelvin PSI, FtH₂O, inH₂O, mmH₂O, mH₂O, inH₂O, cmHg, mmHg, Bars, Bars,
Range -15˚ to 55˚C (5˚ to 131˚F) Gauge
Compensated --- 0˚ to 40˚C (32˚ to 104˚F) Max operating pressure 1.1 x full scale Over pressure protection 3x full scale up to 300psi - for > 300psi (650 ft or 200 m) contact factory Burst pressure 1000 psi (approx. 2000 ft or 600 m) Environmental IP68, NEMA 6P
Fluoropolymer and PTFE
below freezing.
environmental factors)
±2.0°C — below 0°C (32˚F)
±0.05% FSO (typical, static) ±0.1% FSO (maximum, static)
(B.F.S.L. 20˚C)
kPa
PSI: 12, 5, 7, 15, 30, 50, 100, 300
FtH₂O: 2.32, 12, 35, 69, 115, 231, 692
Absolute³
mH₂O: 0.72, 3.5, 5, 10.5, 21, 35, 70, 210
PSI: 30, 50, 100, 300
FtH₂O: 35, 81, 196, 658 mH₂O: 10, 24, 59, 200
*Specications subject to change. Please consult out web site for the most current data (seametrics.com). Modbus is a registered trademark of Schneider Electric.
1 Higher pressure ranges available upon request 2 ±0.25% accuracy FSO (max) at this range 3 Depth range for absolute sensors has 14.7 PSI subtracted to give actual depth allowed.
Seametrics • 253.872.0284 Page 4 seametrics.com
Page 5
GENERAL INFORMATION
PT2X INSTRUCTIONS
How Pressure Sensors Work
Liquids and gasses do not retain a xed shape. Both have the ability to ow and are often referred to as uids. One fundamental law for a uid is that the uid
exerts an equal pressure in all directions at a given level. Further, this pressure increases with an increasing depth of
“submergence”. If the density of a uid remains constant
(noncompressible...a generally good assumption for water at “normal” pressures and temperatures), this pressure increases linearly with the depth of “submergence”.
We are all “submerged” in the atmosphere. As we increase our elevation, the pressure exerted on our bodies decreases
as there is less of this uid above us. It should be noted
that atmospheric pressure at a given level does vary with changes in the weather. One standard atmosphere
(pressure at sea level at 20º C) is dened to be 14.7 PSI
(pounds per square inch).
There are several methods to reference a pressure measurement. Absolute pressure is measured with respect to an ideal vacuum (no pressure). Gauge pressure is the most common way we express pressure in every day life and is the pressure exerted over and above atmospheric pressure. With this in mind, gauge pressure (Pg) can be
expressed as the dierence between the absolute pressure
(Pa) and atmospheric pressure (Patm):
Water Line
P = Patm + kd
A
Patm
d
“A”
Pressure Diagram: See Detail A.
Pg = Pa - Patm.
To measure gauge pressure, atmospheric pressure is subjected to one side of the system and the pressure to be measured is subjected to the other. The result is that the
dierential (gauge pressure) is measured. A tire pressure
gauge is a common example of this type of device.
Recall that as the level of submergence increases (in a
noncompressible uid), the pressure increases linearly.
Also, recall that changes in weather cause the absolute atmospheric pressure to change. In water, the absolute pressure (Pa) at some level of depth (d) is given as follows:
Pa = Patm + kd
where k is simply a constant (i.e.: 2.307 feet of water = 1 PSI)
Seametrics’ standard gauge submersible pressure devices utilize a vent tube in the cable to allow the device to reference atmospheric pressure. The resulting
gauge pressure measurement reects only the depth of
submergence. That is, the net pressure on the diaphragm is due entirely to the depth of submergence.
Seametrics • 253.872.0284 Page 5 seametrics.com
Absolute pressure is given as Pa = Patm + kd
(where k is 2.307 feet of water)
Page 6
GENERAL INFORMATION
PT2X INSTRUCTIONS
Initial Inspection and Handling
Upon receipt of your smart sensor, inspect the shipping package for damage. If any damage is apparent, note the signs of damage on the appropriate shipping form. After opening the carton, look for concealed damage, such as a cut cable.
If concealed damage is found, immediately le a claim with the carrier.
Check the etched label on the sensor to be sure that the proper range and type were provided. Also check the label attached to the cable at the connector end for the proper cable length.
Do’s and Don’ts
Do handle sensor with care
   
Do store sensor in a dry, inside area when not in use
Do install a desiccant tube if using a gauge sensor
   
Do install sensor so the connector end is kept dry
Don’t drop into well
Lower gently to prevent damage
Seametrics • 253.872.0284 Page 6 seametrics.com
Don’t scrape cable over edge of well
May nick or fray the cable
Don’t bend cable sharply
May close o vent tube
and/or weaken internal wires
Don’t support sensor with the connector
Use a strain relief device
Page 7
INSTALLATION
PT2X INSTRUCTIONS
Connecting External Power
The PT2X comes with two 1.5V AA internal batteries.
If auxiliary power is desired, you can use a 6–16 VDC supply that can provide 15 mA. Connect to Vaux++ (pin 1 - white) and Ground (pin 5 - blue) or contact Seametrics for auxiliary power supplies.
Connecting the PT2X to a Computer
Cabled sensors are terminated with a weather-resistant connector. Cableless sensors are terminated with a weather­resistant connector that is inside a screw-cap. Connect the weather-resistant connector to your computer’s USB port as shown below.
USB to RS485
Adapter
Screw­cap
Sensor
Sensor
Cableless
Conguration
USB Port
PC or Laptop
Computer
Connecting sensor to your computer using
Seametrics’ USB to RS485 adapter.
Connecting via RS232 Serial Port
Connect the weather-resistant connector to your computer’s serial port as shown below.
Interface Cable
Screw­cap
Sensor
Cableless
Conguration
Sensor
RS485/RS232 Adapter
Serial Port
PC or Laptop
Computer
Connecting sensor to your computer using
an RS485/RS232 adapter and an interface cable.
Connecting with a USB/Serial Adapter
USB-to-Serial cables are readily available from many electronics and computer stores, as well as numerous sites on the Internet. Seametrics has tested and recommends the Keyspan USA-19HS. Install as follows:
Plug into USB port.
Install the drivers provided with the particular unit.
Once drivers are installed connect to sensor
Aqua4Plus 2.0 communicates with the sensor using the USB to RS485 adapter cable. This cable requires drivers to be installed on your computer. If you are connected to the
Internet when you rst plug in the cable, it will normally
obtain and install the correct drivers automatically.
Alternate Connection Options
Seametrics recommends connecting the sensor to your computer using the Seametrics USB cable. However, when using Aqua4Plus 2.0, the sensor can also be connected using an RS232 serial port or a USB-to-Serial cable, as described below.
Interface
Screw­cap
Sensor
Cableless
Conguration
Cable
Adapter
RS232/RS485 Adapter
Sensor
Connecting sensor to your computer using
a USB to Serial adapter and an interface cable.
USB PortUSB-to-Serial
PC or Laptop
Computer
Seametrics • 253.872.0284 Page 7 seametrics.com
Page 8
INSTALLATION
PT2X INSTRUCTIONS
Connecting to Sensors
Aqua4Plus 2.0 is designed to automatically detect your communication cable and scan for sensors. It is recommended you connect your USB/RS485 cable to your PC and have the sensor connected before opening Aqua4Plus 2.0.
If your cable and sensor were not connected before opening Aqua4Plus 2.0 simply connect and click Rescan. While
scanning is active you’ll see a green dot ash in the upper right corner of the program. Scanning is complete when this dot stops ashing.
If your sensor still won’t connect you can expand the Modbus address range under program settings here:
Simply drag the Address Range slider higher up to increase the maximum Modbus address scanned. If you’ve scanned all the way up through address 255 and still have no connection click Troubleshooting for further troubleshooting or contact Seametrics Tech Support for assistance.
Click at any time to refresh sensor information.
Seametrics • 253.872.0284 Page 8 seametrics.com
Page 9
INSTALLATION
PT2X INSTRUCTIONS
Cable Wiring
If you buy your cabled sensor with a connector installed
(the normal conguration), no further wiring is needed. For reference purposes, the rst three diagrams below
show the pinout from the connector for various scenarios.
The nal diagram shows the pinout if you bought your
sensor without a connector for use with SDI-12.
White Purple Yellow Brown Blue Shield
1 2 3 4 5
5-Pin Connector
12 VDC+ (Vaux) Modbus D­Modbus D+ Digital I/O (Not used) 12 VDC- (GND)
For Modbus® with rmware lower than 2.0
— with 5-pin connector
White Purple Yellow Brown Blue Shield
1 2 3 4 5
5-Pin Connector
12 VDC+ (Vaux) Modbus D­Modbus D+ SDI-12 (Not used) 12 VDC- (GND)
For Modbus® with rmware 2.0 or higher
— with 5-pin connector
Installing Aqua4Plus 2.0
The PT2X comes with the Aqua4Plus 2.0 host software that is installed on your PC or laptop. Use this software to program the datalogger, to retrieve data from the logger,
to view collected data, and to export data to external les
for use with spreadsheets or databases.
Refer to the software manuals for details on installing and using Aqua4Plus 2.0.
Using the PT2X Without Aqua4Plus 2.0
Most users will use the PT2X with Seametrics’ Aqua4Plus
2.0 software. However, the PT2X is quite versatile, communicating via either Modbus® or SDI-12 interfaces, allowing you to do the following:
®
Read a PT2X via Modbus
using your own software.
Read a PT2X via SDI-12 protocol.
Display readings from a PT2X on a panel meter.
If you want to use one of these methods, see page 20.
White Purple Yellow Brown Blue Shield
1 2 3 4 5
5-Pin Connector
12 VDC+ (Vaux) Modbus D- (Not used) Modbus D+ (Not used) SDI-12 Signal 12 VDC- (GND)
For SDI-12 with rmware 2.0 or higher
— with 5-pin connector
White Purple Yellow Brown Blue Shield
12 VDC+ (Vaux) Modbus D- (Not used) Modbus D+ (Not used) SDI-12 Signal 12 VDC- (GND) Earth ground
For SDI-12 with rmware 2.0 or higher
— without connector
Seametrics • 253.872.0284 Page 9 seametrics.com
Page 10
INSTALLATION
PT2X INSTRUCTIONS
Installing the Sensor
The PT2X measures pressure. The most common application is measuring liquid levels in wells and tanks. In order to do this, the sensor must be installed below the water level at a
xed depth. The installation depth depends on the range
of the sensor. One (1) PSI is equal to approximately 2.31 feet of water. If you have a 5 PSI sensor, the range is 11.55 feet of water and the sensor should not be installed at a depth below 11.55 feet. If the sensor is installed below its maximum range, damage may result to the sensor and the output reading will not be correct.
Note: If you are using an absolute sensor and you want to enter a depth-to-water reference after data is collected, then see page 19 before proceeding.
Lower the sensor to the desired depth.
Fasten the cable to the well head using a weather proof strain-relief system. When securing a vented cable, make sure not to pinch the cable too tightly or the vent tube inside the cable jacket may be
sealed o.
Take a measurement to insure the sensor is not installed below its maximum range.
Be sure the supplied cap is securely placed on the weather­resistant connector at the top of the cable. Do not install such that the connector might become submerged with changing weather conditions. The connector can withstand incidental splashing but is not designed to be submerged.
For vented sensors, install the sensor so that the desiccant
tube will not ood or lie in water.
The sensor can be installed in any position; however, when it leaves the factory it is tested in the vertical position. Strapping the sensor body with tie wraps or tape will not hurt it. Seametrics can provide an optional 1/4” NPT input adapter which is interchangeable with the standard end cone for those applications where it is necessary to directly attach the sensor to a pipe, tank, or other pipe port. If the
sensor is being installed in a uid environment other than water, be sure to check the compatibility of the uid with
the wetted parts of the sensor.
Desiccant Use
On vented sensors a desiccant tube prevents moisture in the air from being sucked into the vent tube, which can cause erratic readings and sensor damage.
The desiccant tube is lled with blue silica gel beads. A locking barb and a hydrophobic water lter are attached to the end of the desiccant tube. This lter prolongs the life of
the desiccant as much as three times over a desiccant tube
without the lter.
Install the sensor so that the desiccant tube and cable
connector will not ood or lie in water.
The desiccant is a bright blue color when active and dry. See Maintenance section for care and changing of desiccant.
Grounding Issues
It is commonly known that when using electronic equipment, both personnel and equipment need to be protected from high power spikes that may be caused by lightning, power line surges, or faulty equipment. Without
a proper grounding system, a power spike will nd the path
of least resistance to earth ground—whether that path is through sensitive electronic equipment or the person operating the equipment. In order to ensure safety and prevent equipment damage, a grounding system must be used to provide a low resistance path to ground.
When using several pieces of interconnected equipment, each of which may have its own ground, problems with noise, signal interference, and erroneous readings may be noted. This is caused by a condition known as a Ground Loop. Because of natural resistance in the earth between
the grounding points, current can ow between the points, creating an unexpected voltage dierence and resulting
erroneous readings.
The single most important step in minimizing a ground loop is to tie all equipment (sensors, dataloggers, external power sources, and any other associated equipment) to a single common grounding point. Seametrics recommends
connecting the shield to ground at the connector end.
Seametrics • 253.872.0284 Page 10 seametrics.com
Page 11
SETTINGS AND CALIBRATION
PT2X INSTRUCTIONS
Sensor Settings
Once connected you’ll see the Sensor screen appear and display the connected sensor(s) details. Mousing over--- icons will provide tool-tips, mouse over to view sensor
rmware and serial number details.
To change general sensor settings click in the sensor screen. This allows you to change the following:
Program Settings
To view/change Aqua4Plus 2.0 settings click in the blue side-bar menu
Under the General Settings tab you can change the default data folder location. This is where your Reports are saved to on your PC.
The Zoom Factor slider can be used to adjust the font size within Aqua4plus 2.0.
Uncheck the Allow app to collect anonymous usage statistics box if you would like to opt out. This information
is used to track Aqua4plus 2.0 reliability across dierent system congurations.
Click to rename the sensor
To change Modbus address and/or Baud Rate simply select the desired address and/or Baud Rate from the drop down menus. Sensor will automatically reconnect at new address and/or Baud Rate
To change the Direct Read output units (for direct Modbus or SDI12 integration) simply select the desired output units from the drop down menus. To change Aqua4plus
2.0 display units scaling see Program Settings.
Sensor Clock can be synced with your PC time or set manually if desired. To set manually enter your desired date/time and click Set Time.
When batteries are changed out make sure to reset the battery information here, simply check the I have just put in new batteries box and select the battery type that was installed from the drop down menu.
Under the Display Units tab you can select your desired display units for the supported channels. These may be changed at any time and associated Real-Time readings and Reports will rescale to the currently selected display unit. To change Direct Read units scaling see Sensor Settings.
Seametrics • 253.872.0284 Page 11 seametrics.com
Page 12
SETTINGS AND CALIBRATION
PT2X INSTRUCTIONS
Pressure Calibration
To perform a calibration setup on the pressure channel
rst connect to the sensor and ensure all data has been
uploaded and erased from the sensor. Next select the calibration button (example images can be found under conductivity calibration page 12).
Next select the calibration setup you’d like to perform:
Comrm desired measurement units and click Continue.
Submergence:
Under the Communications tab you can change your Modbus communication settings. Typically you will only need to change the address range to connect to sensors outside of Modbus address 1-10. In certain cases we may need to change the Retry and Timeout settings to overcome communication issues on very long, or corroded cabling. See Troubleshooting section or contact Seametrics Tech Support for details.
To restore factory default settings click
One Point/Zero Point Calibration:
To zero pressure output to atmospheric pressure position the sensor in air in its desired installation position (typically vertical, if sensor will be installed horizontally position as such during 0 point calibration). Select 1 point Calibration under Calibration Type.
For PSIG sensors use 0 as the reference value and click Measure.
For PSIA sensors enter current barometric pressure from a known accurate barometer set to matching units. Enter your reference value in the Ref. Point box and click Measure.
Seametrics • 253.872.0284 Page 12 seametrics.com
Page 13
SETTINGS AND CALIBRATION
PT2X INSTRUCTIONS
Aqua4Plus 2.0 will take 10 readings and display the average. Watch for stability while Aqua4Plus 2.0 is measuring to ensure an accurate calibration.
Click Ok to accept the reading and Aqua4Plus 2.0 will
calculate a new pressure oset. Click Apply to conrm the new oset value and Aqua4Plus 2.0 will provide a real time
reading to verify calibration was successful:
Select 2 point Calibration under Calibration Type
Perform rst point calibration as listed above and click Next
Enter known pressure value in matching units in the Ref. Point box for second point measurement
2 Point submergence calibration ONLY RECOMMENDED IF YOU HAVE AN ACCURATE PRESSURE REFERENCE. Our Smart Sensors rarely change slope during normal use, however if you have an accurate pressure source it is possible to perform a 2 point calibration on the pressure channel.
Aqua4Plus 2.0 will take 10 measurements and display the average. Watch for stability while Aqua4Plus 2.0 is measuring to ensure an accurate calibration.
Click Ok and Aqua4Plus 2.0 will calculate the new slope
and oset values. Click Apply to conrm the new slope and oset values and Aqua4Plus 2.0 will provide a real time
reading to verify calibration was successful.
Depth to water (for PSIG sensors only, see Barometric Compensation Utility for PSIA sensors)
Position the sensor in its desired location and ensure all data has been uploaded and erased from the sensor before proceeding with calibration.
Once positioned connect to sensor and select the calibration button, followed by selecting the Depth to Water option. Double check measurement units selection before proceeding.
Enter your current depth to water (Typically obtained from a water level indicator) value in the Ref. Point box, making sure to match measurement units.
Seametrics • 253.872.0284 Page 13 seametrics.com
Page 14
SETTINGS AND CALIBRATION
PT2X INSTRUCTIONS
Click Measure and Aqua4Plus 2.0 will take 10 readings and display the average. Watch for stability while Aqua4Plus 2.0 is measuring to ensure an accurate calibration.
Click Accept and Aqua4Plus 2.0 will calculate a new slope
and oset, click Apply to accept the new slope and oset
and Aqua4Plus 2.0 will provide a real time reading to verify calibration was successful.
Groundwater Elevation
Position the sensor in its desired location and ensure all data has been uploaded and erased from the sensor before proceeding with calibration.
Once positioned connect to sensor and select the calibration button, followed by selecting the Groundwater Elevation option. Double check measurement units selection before proceeding.
Enter your current Groundwater Elevation reading in the Ref. Point box, making sure to match measurement units.
Click Measure and Aqua4Plus 2.0 will take 10 readings and display the average. Watch for stability while Aqua4Plus 2.0 is measuring to ensure an accurate calibration.
Click Accept and Aqua4Plus 2.0 will calculate a new oset, click Apply to accept the new oset and Aqua4Plus 2.0
will provide a real time reading to verify calibration was successful.
Removing calibration setup values
To return to factory default simply enter an oset of 0 and slope of 1 in the Oset and Slope boxes:
Click Accept and Aqua4Plus 2.0 will calculate a new oset, click Apply to accept the new oset and Aqua4Plus 2.0
will provide a real time reading to verify calibration was successful.
Sta Gauge
Position the sensor in its desired location and ensure all data has been uploaded and erased from the sensor before proceeding with calibration.
Once positioned connect to sensor and select the
calibration button, followed by selecting the Sta Gauge
option. Double check measurement units selection before proceeding.
Enter your current Sta Gauge reading in the Ref. Point
box, making sure to match measurement units.
Click Measure and Aqua4Plus 2.0 will take 10 readings and display the average. Watch for stability while Aqua4Plus 2.0 is measuring to ensure an accurate calibration.
Click Apply to conrm settings and Aqua4Plus 2.0 will provide a real time reading to conrm.
Adjusting for specic gravity
You man enter the specic gravity of your uid in the Slope eld to adjust for specic gravity when needed. Click Apply to conrm settings and Aqua4Plus 2.0 will provide a real time reading to conrm.
Seametrics • 253.872.0284 Page 14 seametrics.com
Page 15
OPERATION
Real-time Data
Connect to sensor and select the Real-time data tab
To start real-time readings click Start, readings default to table view
PT2X INSTRUCTIONS
Real-time readings default to a 1 second interval for 1 minute, to adjust enter your desired settings here:
Note: Currently this data is not saved and is for viewing current conditions only. To save the data to sensor memory see Data Logging section. You can run Real-time Data while logging is active.
To switch to Real-time graphing view click the graph icon
Seametrics • 253.872.0284 Page 15 seametrics.com
Page 16
OPERATION
Data Logging
Select Set Up Logging from the sensor screen
PT2X INSTRUCTIONS
If there are no les currently on the sensor you’ll see the
Set Up Logging button active under the Data Files tab as
well as in the upper menu. Once les have been started/
logged on the sensor they will be displayed under the Data Files tab.
Setting Up Logging Window
Here you will name your data le and set up the recording
interval and duration of each logging phase. Select your desired recording interval and duration for each phase, Aqua4Plus will display the available memory at the bottom of the window.
Click to switch between interval and continuous data recording (PT2X & CT2X only) Select your continuous rate from the drop down box here.
Duration can be set by either number of records
Or by setting a duration time
When set by number of records the time of the recording phase will be displayed detailing how long that phase will run. When set by time, the total number of records for that phase will be displayed.
If you need to check settings or perform a calibration click
before proceeding with logging setup to switch to the
Settings and Calibration screen.
You may sync the sensor clock with the PC clock when starting logging by clicking the slider here.
Check the Delayed Start box and enter the desired date/ time you would like logging to start. This is useful for syncing data when setting up multiple sensors on a site. Data will start logging at the set date/time rather than immediately when Start is pressed.
Seametrics • 253.872.0284 Page 16 seametrics.com
Page 17
OPERATION
PT2X INSTRUCTIONS
Data le name defaults to Test File # and may be re-named
here.
The 3 previous Logging Schedules that were programmed to a sensor will be listed under the Select Template drop
down menu. There you will also nd pre-programmed
logging schedules such as 24 hour pump test, along with any custom logging schedules saved by the user.
To save a logging schedule as a template enter desired settings and click This will add your custom schedule to the Select Template menu.
Once all the desired settings are made simply click Start to begin logging.
Data les already downloaded will show in the Reports column, clicking here will bring you to the reports screen to view the data. See Reports section for details.
You may only have 1 active data le recording on each sensor, however you can store multiple les in memory if
desired.
Starting a new le will automatically terminate the active
logging and begin the new logging schedule. Real-time data is available during active logging.
This will return you to the Sensor screen and your status
will change to Active with the data le displayed under the Data Files tab. Mouse over an active le to pause, terminate,
download, or view logging setup details.
Seametrics • 253.872.0284 Page 17 seametrics.com
To delete les from memory make sure they have all been
downloaded to Reports. Files are removed from memory all at once rather than individually.
Once conrmed les are permanently deleted from the
sensor memory.
Page 18
OPERATION
Reports
Data downloaded from your sensor is stored in the Reports
section of Aqua4Plus 2.0 for viewing and editing. The les
will be saved to default data folder on your PC as well. See Program Settings for default data folder location.
In the main view you’ll see a list of reports sorted by date,
size, or le name as selected here
PT2X INSTRUCTIONS
Click to view data as a table
Click to view data statistics
The Information tab is a new feature allowing users to add
metadata to their reports such as site location, eld notes,
or comments.
You can also search reports by keyword using the search box
Click on a report to bring up the report details
Reports are displayed in graphing view by default. You can
zoom to specic sections by selecting a section with you
mouse or by adjusting the slider below the graph, shown below.
Click to switch to full screen graphing view
Graph saving and export options are available here
The Schedule tab will display the logging setup details for the report
Click Export to export the report as a .csv le or .a4d le for
distribution or use in 3rd party software.
Click Delete to delete the report from Aqua4Plus 2.0
You can also import .a4d les from compatible sensors into Aqua4Plus 2.0 by clicking at the top of the Reports screen.
A Word About Units
Readings from the PT2X Smart Sensor can be displayed in various units, such as PSI, Ft. H2O, or mm H2O for pressure, or degrees Celsius or degrees Fahrenheit for temperature. Select the units you want from the Options | Display Units
menu or from the Congure Menu | Program Conguration
| Set Computer Display Units.
Seametrics • 253.872.0284 Page 18 seametrics.com
Page 19
OPERATION
Barometric Compensation
PT2X INSTRUCTIONS
For PSIA sensors we’ve built a new barometric compensation utility into the Reports section. Click on a report to compensate the data for barometric pressure.
Corresponding barometric les are ltered by date/time
and displayed to the left. Select the barometric le you
would like to use to compensate the report, select either the Submergence or Depth To Water tab, then click Continue.
Aqua4Plus 2.0 will perform the barometric compensation and create a new compensated report. Original reports are retained as uploaded.
Compensated report can then be viewed and exported as needed.
If compensating for Depth to Water enter your depth to water reference measurement and the date/time the measurement was taken (typically taken with a water level indicator before data is uploaded from the sensor) before clicking Continue.
Seametrics • 253.872.0284 Page 19 seametrics.com
Page 20
DIRECT READ MODBUS/SDI-12
Direct Read (Modbus® or SDI-12)
PT2X INSTRUCTIONS
While the PT2X comes with Seametrics’ easy to use Aqua4Plus 2.0, you can also use standard Modbus® RTU or SDI-12 equipment to easily take readings, so as to tie into your existing equipment or networks.
You may need to use Aqua4Plus 2.0 to make a few settings prior to directly reading the PT2X with your equipment. These might include the units for the returned values and/or the Modbus baud rate. These are described in the following sections.
For Modbus direct read, you must have PT2X rmware 1.5 or higher. For SDI-12, you must have rmware 2.0 or higher.
Setting Units for Direct Read
By default, the PT2X uses the following units:
Temperature Degrees Celsius Pressure PSI
If you want to change to dierent units, for example,
degrees Fahrenheit for temperature or feet of water for pressure, set these units using Aqua4Plus 2.0. See sensor settings.
Power Consideration
change the baud rate for Aqua4Plus 2.0, as follows:
Click the Options menu, and then select Baud Rate.
On the popup box, click the down-arrow, select the baud rate you need, and then click OK.
The current Aqua4plus 2.0 baud rate is displayed in the lower right corner of the main Aqua4plus 2.0 window.
Taking Measurements
Reading Registers
Read measurements using Modbus function 03 – Read Holding Registers. Readings are located in two registers each, starting at address 62592. (PT2X register addressing is zero based, i.e., starts at zero. If your equipment uses one based addressing, you will need to add one to the register addresses.)
Addresses for PT2Xs with rmware lower than 2.0
Zero based One based Pressure 62592 62593 Temperature 62594 62595
Addresses for PT2Xs with rmware 2.0 or higher
Zero based One based Temperature 62592 62593 Pressure 62594 62595
If your sensor does not have internal batteries and is not powered continuously by an auxiliary power supply, then you must turn power on to the sensor at least two seconds before a reading is to be taken to allow the sensor to warm up.
Reading Via Modbus® RTU
Setting Baud Rate
Your PT2X comes congured to communicate at 38,400
baud, with 8 data bits, one stop bit, and no parity. The sensor can also be set to 19,200 or 9600 baud, if needed for your application. You must use Aqua4Plus 2.0
If needed, set your PT2X to the desired baud rate as follows:
Click on the Congure menu, and then select Advanced.
From the yout menu, select Sensor Baud Rate. (You may be asked for a password. Enter admin.)
On the popup box, click the down-arrow, select the baud rate you need, and then click OK.
Once you have changed the baud rate on the PT2X, you will not be able to talk to it with Aqua4plus 2.0 until you
Measurement Timing
When you request a reading via Modbus, the sensor wakes up, returns the current values in the registers, and then starts taking new readings and updating the registers. After approximately 10 seconds, if no more readings have been requested, the sensor goes back to sleep.
Because of this, the rst reading you get will be old. If you
are taking readings at intervals of less than 10 seconds,
simply ignore the rst reading — all remaining readings will
be current. On the other hand, if you are taking readings at intervals of greater than 10 seconds, take a reading, ignore it, wait one second, take another reading. Record this second reading.
Data Format
The data is returned as 32-bit IEEE oating-point values, highword rst, also referred to as big-endian or oat
inverse.
For further information and detailed Modbus examples, see Seametrics’ Technical Bulletin, available on www. seametrics.com.
Seametrics • 253.872.0284 Page 20 seametrics.com
Page 21
DIRECT READ MODBUS/SDI-12
PT2X INSTRUCTIONS
Reading Via SDI-12
Note: The default units setting for pressure is PSI. The default units setting for temperature is Celsius. To change these, use the Direct Read Units in the Aqua4Plus 2.0 Control Software.
Addressing
Default SDI-12 Address: 0
SDI-12 Command Nomenclature
<a> = Sensor address {crc} = SDI-12 compatible 3-character CRC <cr> = ASCII carriage return character <lf> = ASCII line feed character
highlighted values indicate variable data
All SDI-12 requests consist of a command followed by a request for values. Some software or equipment may combine these, making the second one unnecessary. Refer to your software or equipment documentation for details.
Request Response
<a>M1! <a>0021<cr><lf>
Address
Command
Request Response
Address
Time until
(in seconds)
response
<a>D0! <a>+7.15863<cr><lf>
Address
Request for values read
Address
Returned value(s)
SDI-12 Commands
Sensor Identication
<a>I! <a>13 INWUSA PT2X 213ssssssssss<cr><lf> Note: 213 will change to reect current rmware version.
Acknowledge Active, Address Query
<a>! <a><cr><lf> ?! <a><cr><lf>
ssssssssss = device serial number
Carriage Return
Linefeed
# of values to
be returned
Carriage Return
Linefeed
Change Address
<a>A<b>! <b><cr><lf> Change address to <b>
Request Measurement
<a>M! <a>0022<cr><lf> Request all measurements <a>D0! <a>+22.0512+12.0512<cr><lf> Read temperature and pressure
<a>M1! <a>0021<cr><lf> Request temperature measurement only <a>D0! <a>+22.0512<cr><lf> Read temperature
<a>M2! <a>0021<cr><lf> Request pressure measurement only <a>D0! <a>12.0512<cr><lf> Read pressure
Seametrics • 253.872.0284 Page 21 seametrics.com
Page 22
DIRECT READ MODBUS/SDI-12
Request Measurement with CRC
<a>MC! <a>0022<cr><lf> Request all measurements with CRC <a>D0! <a>+22.0512+12.0512{crc}<cr><lf> Read temperature and pressure
<a>MC1! <a>0021<cr><lf> Request temperature measurement only with CRC <a>D0! <a>+22.0512{crc} <cr><lf> Read temperature
<a>MC2! <a>0021<cr><lf> Request pressure measurement only with CRC <a>D0! <a>12.0512{crc} <cr><lf> Read pressure
Concurrent Measurement
<a>C! <a>0022<cr><lf> Request all measurements <a>D0! <a>+22.0512+12.0512<cr><lf> Read temperature and pressure
<a>C1! <a>0021<cr><lf> Request temperature measurement only <a>D0! <a>+22.0512<cr><lf> Read temperature
PT2X INSTRUCTIONS
<a>C2! <a>0021<cr><lf> Request pressure measurement only <a>D0! <a>12.0512<cr><lf> Read pressure
Concurrent Measurement with CRC
<a>CC! <a>0022<cr><lf> Request all measurements with CRC <a>D0! <a>+22.0512+12.0512{crc}<cr><lf> Read temperature and pressure
<a>CC1! <a>0021<cr><lf> Request temperature measurement only with CRC <a>D0! <a>+22.0512{crc} <cr><lf> Read temperature
<a>CC2! <a>0021<cr><lf> Request pressure measurement only with CRC <a>D0! <a>12.0512{crc} <cr><lf> Read pressure
For further information and SDI-12 examples, see Seametrics application note, “PT2X SDI-12 Interface Specication” available from our web site at www.seametrics.com.
Seametrics • 253.872.0284 Page 22 seametrics.com
Page 23
MAINTENANCE
PT2X INSTRUCTIONS
Removing Debris from End Cone
At times mud, silt, or other debris may foul the water inlets to the pressure element. The end cone can be removed to clean out the debris.
Twist Open Housing
1. Gently twist o end cone portion only - do not twist
o pressure element!
2. Remove debris. Do not poke anything into the
sensor. This can damage the sensor element and
void the warranty.
3. Replace and retighten the end cone.
Pressure element
Water inlet
End cone
Gently twist o the end cone and carefully remove debris
Set Screw Housing
1. Remove the two set screws at the bottom of the
housing tube, using a 1/16” allen wrench.
2. Gently remove the end cone.
3. Remove debris. Do not poke anything into the
sensor. This can damage the sensor element and void the warranty.
4. Replace the end cone and secure with set screws.
Desiccant Tubes
On vented sensors, inspect the desiccant tube at least once every two months. The desiccant tube prevents moisture in the air from being sucked into the vent tube, which can cause erratic readings and sensor damage.
The desiccant tube is lled with blue silica gel beads. A locking barb and a hydrophobic water lter are attached to the end of the desiccant tube. This lter prolongs the life of
the desiccant as much as three times over a desiccant tube
without the lter.
Install the sensor so that the desiccant tube and cable
connector will not ood or lie in water.
The desiccant is a bright blue color when active and dry. As moisture is absorbed the color will begin to fade, becoming a light pink, which indicates full saturation and time to replace. Replacement desiccant and hydrophobic
lters can be purchased from Seametrics.
To Change the Desiccant:
Pulling gently remove the black tube tting from the clear desiccant tube.
Using needle-nose pliers, remove the dark gray foam plug. Do not discard the plug.
Dump out the old desiccant beads and rell with new desiccant beads – tapping desiccant tube
frequently during relling to ensure that the beads
are fully seated in tube.
Push the foam plug back into the tube.
Reinsert the black tting.
Pressure element
Water inlet
Set screw
End cone
Remove end cone by removing set screws.
Gently clean out debris.
Seametrics • 253.872.0284 Page 23 seametrics.com
Page 24
MAINTENANCE
PT2X INSTRUCTIONS
Sensor
There are no user-serviceable parts, other than the batteries. If problems develop with sensor stability or accuracy, contact Seametrics. If the transducers have been exposed to hazardous materials, do not return them
without notication and authorization.
Cable
Cable can be damaged by abrasion, sharp objects, twisting, crimping, crushing, or pulling. Take care during installation and use to avoid cable damage. If a section of cable is damaged, it is recommended that you send your sensor back to replace the cable harness assembly.
End Connections
The contact areas (pins & sockets) of the connectors will wear out with extensive use. If your application requires repeated connections other types of connectors can be provided. The connectors used by Seametrics are not submersible, but are designed to be splash-resistant.
Changing Batteries Battery Type: Two 1.5V AA batteries—Lithium or Alkaline
(lithium recommended)
IMPORTANT!
Because changing the batteries involves opening the water-tight seal, this must be done in a clean, dry environment to avoid contamination or moisture damage to the circuitry.
IF USING ALKALINE BATTERIES —
PREVENT BATTERY LEAKAGE!
PT2X sensors are typically shipped with lithium batteries. If, however, you are using alkaline batteries, be aware that under some circumstances alkaline batteries can leak, causing damage to the sensor. To prevent leakage, the following is recommended. (Does
not apply to lithium batteries.)
• Change the batteries at least every 18 months.
• If the sensor will not be deployed for 3 months or more, remove the batteries.
Battery Life Calculator (Firmware 2.13 or higher)
When changing batteries, it is important to reset the Battery Life Calculator. If the battery life calculator is not reset, the remaining life information will be incorrect.
Access the Battery Life Calculator from the Congure
Menu - Battery Information and Reset. If you have put in new batteries, checkmark the box “I have just put in fresh batteries.” Click Save and Close. See sensor settings.
Tips
Never place a tool on the sensor body, it is very thin and will deform causing leaks at o-ring seals and potentially crushing the circuit board!
Always twist the sensor body o the top cap assembly rather than twisting the top cap assembly
o of the sensor body.
For cabled sensors, always clamp the sensor on the swaged area when applicable, the shoulder above it will allow you to press down without the worry of the sensor slipping out of the clamping device.
If the sensor body is slippery or you are unable to grip it hard enough to twist, try a piece of rubber cabinet liner for additional friction.
There is a black, compressible square ring near the top of the sensor. This ring acts as a spring to lock the cable in. This needs to be compressed in order to allow removal of the top cap. Once this ring is compressed, a gentle counterclockwise twist is all that is needed to remove the cable from the sensor body. Compressing the black square ring does take force, twisting does not.
HousingTop cap
Swage Knurling Black square ring
Cabled Sensor
Top cap
Housing
Seametrics • 253.872.0284 Page 24 seametrics.com
Knurling Black square ring
Cableless Sensor
Page 25
MAINTENANCE
PT2X INSTRUCTIONS
Care must be taken to compress the black square ring before attempting to twist the housing. Forceful twisting
of the housing can permanently damage the sensor.
Securing the Sensor
In order to compress the black square ring, the sensor must be secured so that you can apply downward pressure to compress the ring. This can be done by holding in your hand, using a vise, or using pliers, as follows.
By Handcabled version only
1. Tightly grasp the top cap in one hand.
2. Brace your hand against something such as a table or the ground. (Do not allow the cable to be pinched against the brace.)
Continue to Removing the Housing on the next page.
With Viserecommended method
Cabled Sensor
1. If possible, use a set of soft jaws as shown to prevent marring the surfaces of the top cap assembly.
2. Place the sensor in a vise clamping gently on the swaged area. You do not need to clamp the vise very hard.
Continue to Removing the Housing on the next page.
Continue to Removing the Housing on the next page.
Cableless Sensor—gripping on knurled area
With Pliers or Vise Grips—good for eld use
Cabled Sensor
1. Grasp the pliers on the swaged area (do not grab the knurled diameter).
2. Find a hard edge and place the tips or side of the jaws of the pliers onto this edge as shown. This will allow you to press down with your weight to compress the square ring.
Continue to Removing the Housing in on the next page.
Cabled Sensor—gripping on swage
Cableless Sensor
1. If possible, use a set of soft jaws as shown to prevent marring the surfaces of the top cap assembly.
2. Remove the cableless top cap.
3. Place the sensor in a vise clamping gently on the knurled area. You do not need to clamp the vise very hard.
Seametrics • 253.872.0284 Page 25 seametrics.com
Cabled Sensor
Cableless Sensor
1. Leave the cableless cap on in order to protect the pins inside.
2. Grasp the pliers on the knurled area tightly being careful to avoid grabbing the knurled cap.
3. Find a hard surface and place the cableless cap down onto it. This will allow you to press down with your weight to compress the square ring.
Page 26
MAINTENANCE
PT2X INSTRUCTIONS
Continue to Removing the Housing below.
Cableless Sensor
Removing the Housing
1. With your free hand, grasp the sensor body. Press down to compress the square ring. Twist gently. Once the body begins to twist, you can stop the compression action.
2. Finish gently twisting until the sensor body is removed.
3. Carefully disconnect the wiring connector inside from the circuit board in the top cap.
Replacing Batteries and Resealing Sensor
4. Hold the top cap assembly at 90° to the housing opening as shown. Depress the spring with your
ngertip and tuck the wiring into the cutaway on
the circuit board with your thumb to protect it while being installed back into the housing.
Wires tucked into slot and spring tucked into housing
5. Rotate the top cap assembly into the opening in the housing being very careful not to nick or pinch any wires.
Push top cap in before twisting and locking
1. Gently pull wiring to one side in order to allow batteries to fall out. Shake gently if needed.
2. Replace batteries with button (+) facing open end.
3. Reinstall wiring connector — it only goes in one way, so make sure not to force it.
Pull wires gently to the side to allow battery removal
6. Gently press down until the assembly stops and then twist it into place. It will click in and decompress the gasket when it is fully engaged.
Properly completed—black ring uncompressed
Connector connected properly
Seametrics • 253.872.0284 Page 26 seametrics.com
Page 27
TROUBLESHOOTING
Problem Probable Causes Things to try…
PT2X INSTRUCTIONS
Software will not
Loose cable Make sure all cable connections are secure communicate with sensor
Contacts in connector loose Be sure all wires are securely fastened inside the
USB driver not installed See Connecting the PT2X to a Computer in
Incorrect USB or COM port selected If using Aqua4Plus, be sure USB is selected in
SEE ALSO ERRATIC READINGS BELOW
Erratic readings Poor connection due to moisture between
contacts in connector
Loose or broken wires in connector Repair or return for evaluation and repair
Damaged cable, cracked or fraying Replace cable
round connector
the Installation section or see the USB/RS485 Adapter Installation application note on our web site.
the dropdown box on the tool bar or the correct COM port if using an alternate connection method. (Aqua4Plus Lite automatically uses only the USB connection.)
Dry thoroughly. Be sure desiccant is fresh (see Maintenance section).
Oscillating readings over time (usually
0.5 to 1.5 feet of water)
Zero readings when pressurized
Moisture in the unit Return for evaluation and repair
Damaged transmitter Return for evaluation and repair
Plugged vent tube (if using a vented unit) Be sure desiccant tube is installed. Test by
gently applying a small amount of pressure to the end of the desiccant tube and seeing
if this aect the transmitter reading. If it does
not, then the vent tube is plugged. Return for evaluation and repair.
Actual water level changes in the aquifer
itself in response to barometric pressure
changes. This eect can occur in tight
You will need to record barometric pressure as well as the water level pressure and compensate
the data formations where the transmitter will immediately pick up barometric changes but the the aquifer will not.
Poor connection due to moisture between contacts in connector
Dry thoroughly. Be sure desiccant is fresh (see
Maintenance section).
Loose or broken wires in connector Repair or return for evaluation and repair
Damaged cable, broken, cracked, or fraying Replace cable
No apparent damage upon visual inspection Return for evaluation and repair
Seametrics • 253.872.0284 Page 27 seametrics.com
Page 28
SEAMETRICS LIMITED WARRANTY
The limited warranty set forth below is given by Seametrics, with respect to Seametrics branded purchases in the United States of America.
Seametrics warrants that products manufactured by Seametrics, when delivered to you in new condition in their original containers and properly installed, shall be free from defects in material and workmanship. Seametrics products are warranted against defects for a
period of two (2) years from date of installation, with proof of install date. If no proof of install date can be provided, warranty
period will be two (2) years from date of shipment from Seametrics, as dened on Seametrics’ invoice. Seametrics’ obligation under this warranty shall be limited to replacing or repairing the part or parts, or, at Seametrics’ option, the products, which prove defective in material or workmanship. The following are the terms of Seametrics’ limited warranty:
a. Buyer must give Seametrics prompt notice of any defect or failure and satisfactory proof thereof. b. Any defective part or parts must be returned to Seametrics’ factory or to an authorized service center for inspection. c. Buyer will prepay all freight charges to return any products to Seametrics’ factory, or another repair facility. as designated by
Seametrics.
d. Defective products, or parts thereof, which are returned to Seametrics and proved to be defective upon inspection, will be repaired
to factory specications.
e. Seametrics will deliver repaired products or replacements for defective products to the buyer (ground freight prepaid) to the
destination provided in the original order.
f. Products returned to Seametrics for which Seametrics provides replacement under this warranty shall become the property of
Seametrics.
g. This limited warranty covers all defects encountered in normal use of Seametrics products, and does not apply to the following
cases: i. Loss of or damage to Seametrics product due to abuse, mishandling, or improper packaging by buyer ii. Failure to follow operating, maintenance, or environmental instructions prescribed in Seametrics’ instruction manual iii. Products not used for their intended purpose iv. Alterations to the product, purposeful or accidental v. Electrical current uctuations vi. Corrosion due to aggressive materials not approved for your specic product vii. Mishandling, or misapplication of Seametrics products viii. Products or parts that are typically consumed during normal operation ix. Use of parts or supplies (other than those sold by Seametrics) which cause damage to the products, or cause abnormally
frequent service calls or service problems
h. A new warranty period shall not be established for repaired or replaced material, products, or supplied. Such items shall remain
under warranty only for the remainder of the warranty period on the original materials, products, or supplies.
i. In the event that equipment is altered or repaired by the buyer without prior written approval by Seametrics, all warranties are void.
Damage caused by equipment or accessories not manufactured by Seametrics may void the product’s warranty.
j. SOFTWARE: The Seller grants the user a non-exclusive license to use Seametrics’ software, according to the following limitations
and conditions: i. The user may install the software on one or more desktop or laptop computers. ii. All title and intellectual rights to the software are owned by Seametrics. iii. No copies may be made or distributed except as described above. iv. The user may not modify or reverse-engineer the software.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, WHETHER ORAL, WRITTEN, EXPRESSED, IMPLIED OR STATUTORY. NO IMPLIED WARRANTY, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, APPLIED TO THE PRODUCTS AFTER THE APPLICABLE PERIOD OF THE EXPRESS LIMITED WARRANTY STATED ABOVE, AND NO OTHER EXPRESS WARRANTY OR GUARANTY, EXCEPT AS MENTIONED ABOVE, GIVEN BY ANY PERSON OR ENTITY WITH RESPECT TO THE PRODUCTS, SHALL BIND SEAMETRICS. SEAMETRICS SHALL NOT BE LIABLE FOR LOSS OF REVENUES, OR PROFITS, OR INCONVENIENCES, EXPENSE FOR SUBSTITUTE EQUIPMENT OR SERVICE, STORAGE CHARGES, LOSS OF DATA, OR ANY OTHER SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGE CAUSED BY THE USE OR MISUSE OF, OR INABILITY TO USE THE PRODUCTS, REGARDLESS OF THE LEGAL THEORY ON WHICH THE CLAIM IS BASED, AND EVEN IF SEAMETRICS HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES. IN NO EVENT SHALL RECOVERY OF ANY KIND AGAINST SEAMETRICS BE GREATER IN AMOUNT THAN THE PURCHASE PRICE OF THE PRODUCT SOLD BY SEAMETRICS AND CAUSING THE ALLEGED DAMAGE. WITHOUT LIMITING THE FOREGOING, YOU ASSUME ALL RISK OF LIABILITY FOR LOSS, DAMAGE, OR INJURY TO YOU AND YOUR PROPERTY AND TO OTHERS AND THEIR PROPERTY ARISING OUT OF USE OR MISUSE OF, OR INABILITY TO USE THE PRODUCTS NOT CAUSED DIRECTLY BY THE NEGLIGENCE OF SEAMETRICS.
SOME STATES DO NOT ALLOW LIMITATIONS ON THE DURATION OF AN IMPLIED WARRANTY, SO THE ABOVE LIMITATIONS MAY NOT APPLY TO YOU. SIMILARLY, SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATIONS OF CONSEQUENTIAL DAMAGE, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU. THIS LIMITED WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS; HOWEVER, YOU MAY ALSO HAVE OTHER RIGHTS WHICH MAY VARY FROM STATE TO STATE.
Seametrics • 19026 72nd Avenue South • Kent, Washington 98032 • USA
(P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • seametrics.com
LT-14310r23 20180226
2/26/18
Loading...