Check Components
PE202 Quick
Install Guide
PE202-0XX
Electromagnetic
Low Flow Meter
101216
Mounting Bracket
Optional
8
¼”, ¾”, ½” or ¾”
PP or PVDF MNPT
Adapter Kit
Positioning
The PE202 may be mounted vertically or horizontally.
Choose a position that will ensure a full pipe.
FLOW
FLOW
101153
X4 Bracket Screws
½” PVDF
FNPT Adapter Kit
Shielded Output Cable
Mounting Plate
It is highly recommended to use the mounting bracket
provided. The mounting bracket uses two #8 screws
(not included) on a 1.5 inch bolt pattern.
101422
8-Pin, 6 Meter
FLOW
Page 1 of 2
Pipe Mounting
Adapter Kit (Optional)
The PE202 may be supported by its piping connections if the
piping is rigid with no exion at the ttings.
Cable Connection
Power and signal connections are provided through the 8-pin
male bulkhead connector on the meter housing.
The standard NPT ttings can be used with or without adapters.
Thoroughly clean the pipe threads and nose and apply Teflon
tape (not included) to male threads. Hold adapter with wrench
while tightening the pipe to prevent damage to the meter.
OR
*Grounding Metal Pipe
For best metering accuracy, jumper the inlet and outlet pipes
together and connect to ground for best results.
*If piping is non-metallic, it is recommended that additional
metal ttings be attached at both ends and connected
together with a 12AWG or larger equalization jumper.
To complete your installation, please see ‘CONNECTIONS’ in your
PE202 instruction manual for power supply connection information.
For the most recent PE202 instruction manual please visit: http://www.seametrics.com/downloads
Page 2 of 2
LT-14231-011315
01/13/2015