Scott PORTABLE GAS DETECTION INSTRUMENT,MINI-SA User Manual

MINI-SA PORTABLE
GAS DETECTION
INSTRUMENT
1
INTRINSICALLY SAFE GAS
LEVEL DETECTOR
R
USC
39NO
CERTIFIED BY UNDERWRITERS LABORATORIES INC.
ONLY AS TO INTRINSIC SAFETY
FOR USE IN HAZARDOUS LOCATIONS
CLASS I, DIV, 1, GROUPS A, B, C, & D.
CERTIFIED TO CANADIAN CSA STANDARD C22.2 NO. 157-92
WARNING:
SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
USE ONLY WITH APPROVED BATTERIES.
TEMPERATURE CODE T3C.
CAUTION:
FOR SAFETY REASONS, THIS EQUIPMENT MUST BE
OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY.
READ AND UNDERSTAND MANUAL BEFORE OPERATING AND SERVICING
CHANGE BATTERIES IN NON-HAZARDOUS AREA
SCOTT INSTRUMENTS
2
SCOTT INSTRUMENTS
MINI-SA PORTABLE GAS
DETECTION INSTRUMENT
Contents
1.0 INTRODUCTION ................................................................. 4
2.0 OPERATIONAL ENVIRONMENT ........................................ 4
3.0 GENERAL INSTRUCTIONS ............................................... 5
3.1 UNPACKING .......................................................... 5
3.2 REGISTRATION ..................................................... 5
3.3 BATTERY INSTALLATION ..................................... 5
4.0 INSTRUMENT DESCRIPTION ............................................ 6
4.1 ALARM LEGENDS ................................................. 7
4.2 AUDIBLE ALARMS ................................................ 8
4.3 VIBRATION ALARM ............................................... 9
4.4 ALARM LED LIGHTS ............................................. 9
4.5 STATUS LIGHT ...................................................... 9
5.0 SWITCH FUNCTIONS AND DISPLAYS ........................... 10
6.0 OPERATING INSTRUCTIONS ........................................... 11
6.1 INSTRUMENT TURN ON / OFF............................ 12
6.2 MENU SELECTION .............................................. 14
6.3 ZERO CALIBRATION MENU ................................ 17
6.4 SPAN CALIBRATION MENU ................................ 18
6.5 SENSOR REPLACEMENT MESSAGE ................. 22
6.6 PEAK / HOLD MENU............................................ 23
6.6.1 PEAK ............................................................. 23
6.6.2 HOLD ............................................................. 25
6.7 ALARM SETPOINT MENU ................................... 26
6.8 TWA / STEL MENU .............................................. 29
6.9 CONFIGURATION MENU ..................................... 32
6.9.1 BATTERY SAVER MODE ............................... 32
6.9.2 PASSWORD PROTECTION .......................... 33
7.0 ERROR MESSAGES ........................................................ 37
8.0 CALIBRATION GASES .................................................... 38
9.0 SPECIFICATIONS ............................................................ 39
3
SCOTT INSTRUMENTS MINI-SA PORTABLE GAS DETECTOR
1.0 INTRODUCTION
The SCOTT INSTRUMENTS MINI-SA Portable Gas Detector is a self contained instrument designed to detect and provide audible, visual, and tactile indication of the presence of oxygen or toxic gas in the air. The toxic gas concentration is monitored and shown in the graphic display in parts per million (ppm) of the atmosphere. The oxygen concentration is shown in the graphic display as a percent (%). When the gas level exceeds the preset level, alarms actuate including a pulsing audible alarm tone, flashing LED, and a vibration alarm.
The alarm functions for toxic gases include a Time Weighted Average (TWA) Alarm and a Short Term Exposure Level (STEL) Alarm.
Other features include AUTO CALIBRATION and self diagnostic capabili­ties, a HOLD function that displays the lowest level detected, and user adjustable high and low ALARM SETPOINTS. A battery life indicator con­stantly monitors remaining battery capacity and a backlight provides visibil­ity in low light situations.
Additional features include a PASSWORD PROTECT menu and a BAT­TERY SAVER mode.
2.0 OPERATIONAL ENVIRONMENT
The SCOTT INSTRUMENTS MINI-SA Portable Gas Detector is designed to be “INTRINSICALLY SAFE” suitable for operation in Class I, Division 1, Groups A, B, C & D Hazardous Locations.
4
3.0 GENERAL INSTRUCTIONS
3.1 UNPACKING
Remove the SCOTT INSTRUMENTS MINI-SA instrument from its packaging and examine it for external damage. If any damage is found, notify your SCOTT INSTRUMENTS distributor, sales representative, or the SCOTT IN­STRUMENTS Service Department.
3.2 REGISTRATION
Fill out and mail the postage paid Registration Card provided. Registering your SCOTT INSTRUMENTS MINI-SA instrument assures that you will re­ceive notification of any future enhancements and/or upgrades that become available. You may also register your instrument on SCOTT INSTRUMENTS web site “www.scottinstruments.com”.
3.3 BATTERY INSTALLATION
The SCOTT INSTRUMENTS MINI-SA instrument is powered by two “AAA” Alkaline batteries. Use only alkaline (Eveready Energizer Industrial EN 92, Duracell PC2400, or Ray-O-Vac AL-AAA) batteries. Use of any other batter­ies except those listed will void the intrinsic safety listing of the SCOTT INSTRUMENTS MINI-SA Portable Gas Detector. Use of rechargeable batter­ies is prohibited. Two of the specified “AAA” Alkaline batteries will provide approximately one hundred (100) hours of continuous non-alarming opera­tion before requiring replacement.
Install the batteries as follows:
1) Remove the battery cover by loosening the #1-phillips-head captive cover screw counterclockwise, then lifting the cover out of the housing. Then lift the two batteries from the housing, using a small coin or tool.
2) Install two (2) AAA Alkaline batteries of the specified type. Note polarity orientation (+/-) as shown inside the battery box and push the batteries securely in between the contacts.
3) Replace the battery cover by pushing it into the bottom detents and secure by tightening the cover screw.
NOTE
The metal belt clip is designed to be removed from the instrument. After removing the battery cover, remove the cover screw completely and slide the clip out of the cover. Store the clip for future use. Replace the cover screw and reattach battery cover to the instrument.
5
4.0 INSTRUMENT DESCRIPTION
THE SCOTT INSTRUMENTS MINI-SA INSTRUMENT is designed for use as a personal monitor. The MINI-SA is equipped with a snap mounting clip for ease of use. A break-away neck strap and rubber boot are also available.
All operations of the instrument are controlled by two push-button actuators which are large enough to accommodate use with gloved hands. In each menu screen, “PROMPTS” over each push-button define their use.
The SCOTT INSTRUMENTS MINI-SA INSTRUMENT case is constructed of high impact resistant, flame retardant plastic. The push-button actuators and the graphic display overlay are UV and chemical resistant.
The Graphic Liquid Crystal Display (LCD) provides continuous readings of gas concentration as well as battery life remaining and other information.
Alarm conditions are signaled by an alarm legend on the graphic display, an audible tone, a flashing light, and an internal vibration. Audible tones for certain alarms can be suspended for thirty (30) second intervals while oth­ers will continue until the unit is reset in an atmosphere that corresponds to the preset safe values. Alarms can also be configured to be latching or nonlatching. Latching alarms operate until acknowledged. Nonlatching alarms clear automatically when the detected gas is on the safe side of the alarm setting.
SENSOR
GRAPHIC
PUSHBUTTONS
(w/ H2S Sensor shown for reference)
BATTERY
MENU
LED
(GREEN OR RED
DEPENDING
FIGURE 1
MINI-SA INSTRUMENT
6
4.1 ALARM LEGENDS
The following alarm legends appear in the graphic display:
“ALARM” flashing legend indicates one of the following:
a) Gas concentration which exceeds the preset level
b) A Time Weighted Average (TWA) Alarm
c) A Short Term Exposure Level (STEL) Alarm.
For toxic gas, the ppm, TWA, or STEL level detected and the “ALARM” legend will flash alternately.
A graphic battery level indicator appears continuously on the display screen. The indicator shows battery capacity as ten (10) lines inside a picture of a battery. Each line represents approximately 10% of battery time remaining.
The battery level indicator flashes when the batteries are low. This will occur when approximately 10% of battery life remains.
EACH LINE REPRESENTS
APPROXIMATELY 10%
OF BATTERY TIME REMAINING
FIGURE 2
BATTERY LEVEL INDICATOR
An exhausted battery alarm will occur after the last bar disappears. See Figure 2.
“HOLD ACTIVATED” flashing legend indicates that the HOLD feature has been selected. The instrument will display the highest reading of a toxic gas or the lowest oxygen reading. This reading will continue to be displayed until the HOLD function is reset.
7
4.2 AUDIBLE ALARMS
The SCOTT INSTRUMENTS MINI-SA has an audible alarm which is acti­vated when any of the following conditions occurs:
- The preset gas alarm level is exceeded (ALARM)
- The Time Weighted Average (TWA) is exceeded
- The Short Term Exposure Limit (STEL) is exceeded
- The battery is exhausted
- The sensor is missing or incompatible
In addition, each time the instrument is turned ON, the audible alarm will pulse three times.
The following describes the function of the Audible Alarm during the alarm conditions listed above:
GAS ALARM – The Audible Alarm sounds when a gas concentration is de­tected that exceeds the chosen alarm level, or when a TWA or STEL alarm value is exceeded. A repetitive pulsing tone will occur at one (1) second intervals while the alarm condition exists.
In the LATCHING alarm mode, selected in the Alarm Setpoint Menu, after an alarm condition has cleared, the alarm indication will continue until the alarm is acknowledged by depressing a pushbutton. In the NON-LATCHING alarm mode, all alarm indications clear when the alarm condition clears. In either mode, the instrument gives an “all clear” signal with three audible beeps, three flashes of the Red Alarm Light, and three vibration pulses when the alarm indication is cleared.
When an Audible Alarm sounds, depressing either pushbutton on the instru­ment will silence the Audible Alarm for thirty (30) seconds. If the original alarm condition still exists after the thirty (30) second silence period, the Audible Alarm will sound again. The Audible Alarm can be silenced repeat­edly by depressing a pushbutton.
LOW BATTERY – An Audible Alarm tone will pulse once every fifteen (15) seconds to indicate when less than 10% of battery life remains.
MISSING SENSOR – The Audible Alarm for a missing or incompatible sen­sor is identical to the pulsing one (1) second tone that occurs during an ALARM condition. Depressing a pushbutton to acknowledge the condition turns the unit OFF.
8
4.3 VIBRATION ALARM
An internal vibrating motor activates whenever the Audible Alarm sounds. The Vibration Alarm operates as a secondary alarm feature in high noise environ­ments where the audible alarm may be difficult to hear.
The Vibration Alarm is reset the same as the Audible Alarm as described in Section 4.2.
The Vibration Alarm will also pulse briefly whenever the MINI-SA instrument is turned ON or OFF.
WARNING THE VIBRATION ALARM DOES NOT FUNCTION FOR THE LOW BAT­TERY INDICATION.
4.4 ALARM LIGHT
The SCOTT INSTRUMENTS MINI-SA instrument contains a Red Light Alarm LED (Light Emitting Diode) which flashes whenever any Alarm Condition oc­curs. It will continue to flash when the Audible Alarm is silenced and will flash as long as the alarm condition is present. In the LATCHING alarm mode, selected in the Alarm Setpoint Menu, after an alarm condition has cleared, the alarm indication will continue until the alarm is acknowledged by de­pressing a pushbutton. In the NON-LATCHING alarm mode, all alarm indica­tions clear when the alarm condition clears. In either mode, the instrument gives an “all clear” signal with three audible beeps, three flashes of the Red Alarm Light, and three vibration pulses when the alarm indication is cleared.
4.5 STATUS LIGHT
The instrument also has a Green Status LED that flashes periodically, indi­cating continuing instrument operation.
9
5.0 SWITCH FUNCTIONS AND DISPLAYS
All operations on the MINI-SA Portable Detector are controlled by two push­buttons located on the bottom of the unit. The operations vary depending on the operating mode and the graphic display appearing above the two switches. The unit sounds a brief audible “beep” whenever either pushbutton is de­pressed while the instrument is on. Whenever the instrument is turned ON or OFF, three successive audible tones will sound and the Red Alarm LED will flash. The Green Status LED will flash once every fifteen (15) seconds during instrument operation. When in the BATTERY SAVER mode, the Green Sta­tus LED flash rate increases to once every five (5) seconds.
SENSOR
LED
(GREEN OR RED
DEPENDING
GRAPHIC
PUSHBUTTONS (2)
FIGURE 3
COMPONENT LOCATIONS
When the unit is in the operating mode, depressing the Display Backlight Pushbutton will illuminate the graphic display for fifteen (15) seconds. A light bulb icon for the Backlight is located above the pushbutton on the operat­ing mode screen. Depressing the button again extinguishes the Backlight. The Backlight is sustained ON or OFF while the MENU operations are in progress, except for calibration. In an alarm condition, the Backlight flashes for the duration of the alarm indication.
When in the BATTERY SAVER mode, the instrument graphic display will be blank. Depressing either pushbutton will turn on the graphic display. The display remains on for thirty (30) seconds after the last pushbutton opera­tion.
USE OF THE DISPLAY BACKLIGHT FOR EXTENDED PERIODS OF TIME WILL REDUCE BATTERY SERVICE LIFE.
WARNING
10
6.0 OPERATING INSTRUCTIONS
Before using this instrument, read and understand ALL instructions for ALL operating modes described in this Section. Prior to use, the instrument should be checked for proper calibration as described in this Section.
WARNING
BEFORE EACH DAY’S USAGE, THE SENSITIVITY OF THE GAS SENSOR MUST BE TESTED ON A KNOWN CONCENTRATION OF CALIBRATION GAS.
WARNING
ACTUAL CALIBRATION SHOULD ONLY BE PERFORMED BY QUALI­FIED INSTRUMENT MAINTENANCE PERSONNEL.
WARNING
ANY INDICATION OR READING DENOTING A TOXIC GAS OR A LACK OF OXYGEN SIGNIFIES THE NECESSITY TO FOLLOW ESTAB­LISHED COMPANY PROCEDURES FOR THE INDICATED CONCEN­TRATIONS OF THE PARTICULAR GAS.
WARNING
USE IN OXYGEN ENRICHED ATMOSPHERES (ABOVE 25% O2) MAY BE HAZARDOUS.
WARNING
CONFIRM THE INSTRUMENT’S BASIC CALIBRATION IS APPRO­PRIATE FOR THE INTENDED USE.
11
6.1 INSTRUMENT TURN ON / OFF
To turn the instrument ON, depress either pushbutton. Three successive “beeps” will sound. The Red LED will flash and the vibration alarm will pulse for each “beep.” The graphic display will show one of the following screens while the instrument initializes. The sensor installed in the unit defines which display is shown:
SCOTT INSTRUMENTS
For Toxic Gases MINI-SA VERSION X.X
<gas> SENSOR
INITIALIZING
All toxic gas screens appear similar, therefore wherever the symbol <gas> appears on any of the graphic display screens in this manual, it will be substi­tuted with the applicable toxic sensor (either CO, Cl NO2, NH
3, PH3
or SO2) installed in the instrument at the time. For some toxic
, ClO
2
HCN, H2S, NH3,
2,
gases, the screen display will show 0.0 or 0.00 resolution.
The following toxic gas sensors are available in the MINI-SA:
CO = Carbon Monoxide
Cl2 = Chlorine
ClO
= Chlorine Dioxide
2
HCN = Hydrogen Cyanide
H2S = Hydrogen Sulfide
NH3 = Ammonia
NO2 = Nitrogen Dioxide
PH3 = Phosphine
SO2 = Sulphur Dioxide
Refer to Section 9.0 for sensor specifications.
SCOTT INSTRUMENTS
For O2 MINI-SA VERSION X.X
O2 SENSOR
INITIALIZING
After a few seconds, the instrument will sound a single “beep” and go to the Operat­ing Mode where the graphic display will show one of the following screens:
Note: If a new sensor is installed or if batteries have been removed for more than a few minutes, the initialization sequence may take longer. If the
12
sensor type has been changed or if the previous calibration was not suc­cessful, the bottom line of the display will show “CALIBRATION REQUIRED”.
ppm
<gas>
For Toxic Gases 0
For O
MENU
2
20.9% O
MENU
2
In the normal Operating Mode, the following information is displayed:
- Gas concentration in either ppm (toxic Gas) or % (oxygen)
- Battery life indication
- Backlight prompt
- Menu prompt
To turn OFF the instrument, depress and hold both push-buttons. The unit will sound three successive “beeps” accompanied by flashes of the Red LED. The instrument will then display the following screen:
DO YOU WANT TO TURN
THE UNIT OFF?
YES NO
13
6.2 MENU SELECTION
From the Operating Mode screen, there are six menus that can be selected by the user. They are “ZERO CALIBRATION,” “SPAN CALIBRATION,” “PEAK/ HOLD,” “ALARM SETPOINT,” “TWA/STEL,” and “CONFIGURATION.” De­press the pushbutton below the MENU prompt on the display and then one of the following screens will appear:
ZERO CALIBRATION
For Toxic Gases
NEXT AUTO ZERO
AUTO CALIBRATION
For O
2
NEXT CALIBRATE
The AUTO ZERO or CALIBRATE command and “NEXT” command appear above the pushbutton that activates that function.
To AUTO ZERO or CALIBRATE the sensor, depress the right pushbutton. See Section 6.3 for details on calibration.
To go to the next menu, depress the left pushbutton below the “NEXT” com­mand and the following screen will be displayed:
SPAN CALIBRATION
For toxic gases
NEXT SELECT
To Span a toxic gas sensor, depress the “SELECT” pushbutton. See Sec­tion 6.4 for details on span calibration.
NOTE: There is no span calibration screen when using the Mini-Sa portable to detect oxygen. The screen will go directly to the next menu.
To go to the next menu, depress the left pushbutton below the “NEXT” com­mand and the following screen will be displayed:
14
PEAK / HOLD
NEXT SELECT
To go to the PEAK/HOLD menu, depress the “SELECT” pushbutton.
To go to the next menu, depress the left pushbutton below the “NEXT” com­mand and one of the following screens will be displayed:
<gas> ALARM SETPOINT
For Toxic Gases
NEXT SELECT
O2 ALARM SETPOINT
For O
2
NEXT SELECT
To go to the ALARM SETPOINT menu, depress the “SELECT” pushbutton.
To go to the next menu, depress the left pushbutton below the “NEXT” com­mand and one of the following screens will be displayed:
<gas> TWA/STEL
For Toxic Gases
NEXT SELECT
NOTE: There is no TWA/STEL menu displayed when an instrument is used for oxygen detection.
To go to the TWA/STEL menu, depress the “SELECT” pushbutton. See Section 6.8 for details.
To go to the next menu, depress the left pushbutton below the “NEXT” com­mand and the following screen will be displayed:
15
CONFIGURATION
NEXT SELECT
To go to the CONFIGURATION menu, press the “SELECT” pushbutton.
To go to the next menu, depress the left pushbutton below the “NEXT” com­mand and the ZERO CALIBRATION for toxic gas or AUTO CALIBRATION for oxygen screen will again be displayed.
The MINI-SA will scroll repeatedly through these Menus each time the “NEXT” pushbutton is depressed. Each time either pushbutton is depressed, the unit will sound a single “beep”.
If no menu action is taken for fifteen (15) seconds while one of the menus is displayed, the instrument will automatically return to the operating mode, signified by a single “beep”.
16
6.3 ZERO CALIBRATION MENU
The MINI-SA instrument can be Auto Zeroed at any time. An Auto Zero should only be performed in an area known to be free of toxic gas or any known cross sensitive materials. See Section 8.0 for SCOTT INSTRUMENTS sup­plied calibration gas cylinders. When performing an auto-zero with an oxy­gen unit, be sure the area contains 20.9% oxygen.
To perform Zero Calibration of the sensor as described in Section 6.2, de­press the “AUTO ZERO” pushbutton when the ZERO CALIBRATION menu screen is displayed.
The instrument will display one of the following screens while it calibrates the appropriate sensor.
<gas> SENSOR
For Toxic Gases ZERO CALIBRATION 2
O2 SENSOR
For O2 ZERO CALIBRATION 2
The instrument display will count down 2 seconds while the sensor is ze­roed. After a successful “ZERO CALIBRATION”, the instrument will sound a “beep” and automatically go back into the Operating Mode.
If, during the “ZERO CALIBRATION”, the unit detects that the sensor is out of range, an error message will be displayed. See Section 7.0 for a description of error messages.
17
6.4 SPAN CALIBRATION MENU FOR TOXIC GAS ONLY
Note: The MINI-SA software requires that an “AUTO ZERO” be performed as described in Section 6.3 before the instrument will allow a “SPAN CALIBRA­TION”. Failure to do so will display the following screen when attempting a “SPAN CALIBRATION” of the instrument:
AUTO ZERO REQUIRED
BEFORE
SPAN CALIBRATION
ACKNOWLEDGE
Depress both pushbuttons simultaneously to acknowledge this message and “AUTO ZERO” the instrument per Section 6.3 to procede.
The MINI-SA instrument can be calibrated with span gas supplied by SCOTT INSTRUMENTS or with a calibration gas supply certified to a concentration within the following ranges:
CO = 25 to 120 ppm
Cl2 = 5 ppm only Cl
ClO2 = 5 ppm only Cl
2
2
HCN = 2.0 to 25.0
H2S = 10 to 110 ppm
NH3 = 10 to 40 ppm
NO2 = 2.0 to 15.0 ppm
PH3 = 0.5 to 4.5 ppm
SO2 = 2.0 to 18.0 ppm
See Section 8.0 for SCOTT INSTRUMENTS supplied calibration gas cylinders.
To perform “SPAN CALIBRATION” of the sensor, depress the “SELECT” pushbutton when the “SPAN CALIBRATION” screen is displayed as de­scribed in Section 6.2. The instrument will display the one of the following screens:
For Toxic Gases SPAN CALIBRATION
<gas> SENSOR
USING <_> ppm <CALGAS>
CHANGE OK
18
Calibration cylinders supplied by SCOTT INSTRUMENTS contain 25 PPM H2S or 50 PPM CO. To use another certified concentration, depress the left pushbutton below the “CHANGE” command. The one of the following screens will be displayed:
<gas> SENSOR
For Toxic Gases SPAN CALIBRATION
GAS = <_> ppm <CALGAS>
UP DOWN
Factory preset span calibration gas values:
CO = 50 ppm
Cl2 = 5.0 ppm
ClO2 = 5.0 ppm Cl
2
HCN = 10.0 ppm
H2S = 25 ppm
NH3 = 25 ppm
NO2 = 5.0 ppm
PH3 = 1.0 ppm
SO2 = 10 ppm
If you select the “UP” pushbutton, for example, one of the following screens will display with the number value flashing:
<gas> SENSOR
For Toxic Gases SPAN CALIBRATION
GAS = <_> ppm <CALGAS>
UP SET
When the desired value has been reached depress the “SET” pushbutton and the “CHANGE/OK” screen will appear again. Select the “OK” pushbutton and one of the following screens will display:
For Toxic Gases SPAN CALIBRATION
<gas> SENSOR
INJECT <_> ppm <CALGAS>
EXIT
19
This screen will remain on until the instrument detects calibration gas. If no calibration gas is detected after thirty (30) seconds, the instrument will sound a single beep and return to the operating mode.
Depressing the “EXIT” pushbutton will also return the instrument to normal operation with no calibration having occurred. The last successful calibration will remain valid.
To continue “SPAN CALIBRATION”, attach the 40014936 Calibration Adapter over the sensor cap as shown in Figure 4.
TO GAS
CYLINDER
FIGURE 4
CALIBRATION ADAPTER
20
Attach the hose of the Adapter to the calibration gas cylinder and begin flowing gas to the cap at a flow rate of 500 cc/min. (300 cc/min. for Cl2 and ClO2). When the instrument detects the gas, the display will change to one of the following screens:
<gas> SENSOR
For Toxic Gases SPAN CALIBRATION
<time> seconds
The display will count down <seconds> (See chart below for applicable sen­sor) after which the instrument will display the message “SPAN CALIBRA­TION PASSED ACKNOWLEDGE”. Press either button to return to the Oper­ating Mode showing the gas concentration. If the user does not acknowl­edge within 2 minutes, the instrument returns to Operating Mode automati­cally. In either case, the calibration is accepted as valid. If the calibration gas concentration exceeds the alarm setpoint, the alarm will operate. Turn off the gas supply and remove the Calibration Adapter.
If, during “SPAN CALIBRATION”, the unit detects that the sensor is out of range, an error message will be displayed. See Section 7.0 for a description of error messages.
Max. Span Cal Time in Seconds
300 180 180 60 90 180 60
Ammonia (NH ) Chlorine(Cl ) Chlorine Dioxide (ClO ) Carbon Monoxide (CO 500) Carbon Monoxide (CO 1000) Hydrogen Cyanide (HCN) Hydrogen Sulfide (H S 100)
Applicable
Toxic Sensor
3
2
Hydrogen Sulfide (H S 500)60 90 Nitrogen Dioxide (NO ) 90 120
Phosphine (PH )
3
Sulfur Dioxide (SO )
2
2
2
2
2
21
Note: SCOTT INSTRUMENTS supplies Teflon®-lined tubing with all chlorine and ammonia calibration kits. Sticky gases such as chlorine and ammonia require this tubing be used for calibration. Using tub­ing of another type may inhibit the automatic calibration of a MINI-SA and lead to improper readings or a false failed sensor message. All other calibration kits are supplied with Tygon (R-3603) Tubing.
SCOTT INSTRUMENTS also recommends that the tubing and cylinder regulator used with a specific toxic gas not be used with other gases.
6.5 SENSOR REPLACEMENT MESSAGE
If a successful “AUTO CALIBRATION” reveals that the gas sensor is near the end of its service life, the MINI-SA will display one of the following REPLACE­MENT MESSAGE:
WARNING <gas> SENSOR
For Toxic Gases APPROACHING FAILURE
REPLACE SOON!
ACKNOWLEDGE
WARNING OXYGEN SENSOR
For O2 APPROACHING FAILURE
REPLACE SOON!
ACKNOWLEDGE
The user must “ACKNOWLEDGE” this message by depressing either pushbutton.
WARNING THIS SCREEN IS FOR INFORMATION ONLY AND IS NOT INTENDED TO BE AN EXACT DETERMINATION OF WHEN A SENSOR WILL FAIL AND REQUIRE REPLACEMENT. THE TIME REMAINING UNTIL A SENSOR FAILS AFTER A REPLACEMENT MESSAGE IS DISPLAYED IS BASED ON MANY FACTORS SUCH AS FREQUENCY OF CALIBRATION, LAST CALIBRATION, FREQUENCY OF USE AND LEVEL OF GAS TO WHICH SENSOR IS EXPOSED. A SENSOR MAY FAIL PRIOR TO DISPLAY OF THE REPLACEMENT MESSAGE.
22
6.6 PEAK / HOLD MENU
The MINI-SA can be set to display either the lowest and highest gas reading (PEAK) or just the highest toxic gas or lowest oxygen reading (HOLD) during use. To enter the PEAK/HOLD menu, depress the “SELECT” pushbutton when the PEAK/HOLD menu is displayed as described in Section 6.2. The following screen will appear:
SELECT MODE
PEAK HOLD
If neither pushbutton is depressed for fifteen (15) seconds after the above dis­play appears, the unit will automatically return to the Operating Mode accom­panied by a single “beep”.
6.6.1 PEAK
The PEAK feature allows the user to display both the lowest and highest readings that were detected since the instrument was last turned on, or since the readings were last reset.
In the SELECT MODE menu, depress the “PEAK” pushbutton. Then one of the following screens will appear for fifteen (15) seconds displaying the high­est reading since instrument turn on:
PEAK <gas> LEVEL
For Toxic Gases MAX = <_> ppm
RESET SHOW
PEAK PEAK
PEAK O2 LEVEL
For O
2
MAX = 20.9%
MIN = 20.2%
RESET SHOW
PEAK PEAK
If neither the “RESET PEAK” or “SHOW PEAK” pushbutton is depressed, the unit will return to the Operating Mode accompanied by a single “beep”.
If the “SHOW PEAK” pushbutton is depressed, the instrument will return to the Operating Mode and display the peak values as follows on the appropriate screen:
23
ppm
4
<gas>
For Toxic Gases
MAX = <_> ppm
MENU
%
20
O
2
For O2 MAX = 21.5 %
MIN = 20.9 %
MENU
Current and peak readings update continuously while this screen is dis­played.
If the PEAK menu is accessed from a peak reading display, the option “SHOW PEAK” will read “HIDE PEAK” instead, as follows:
PEAK <gas> LEVEL
For Toxic Gases MAX =<_> ppm
RESET HIDE
PEAK PEAK
PEAK O
For O
2
MAX =20.9 %
LEVEL
2
MIN = 20.2%
RESET HIDE
PEAK PEAK
Selecting “HIDE PEAK” will return the display to the standard Operating Mode screen with only the current reading displayed. The “SHOW PEAK” function is latching and will not change by turning the instrument OFF.
To reset the peak value, depress the “RESET PEAK” pushbutton in the PEAK menu. This will reset the MAX value to the current ambient reading. Turning the instrument OFF will also RESET the PEAK value the next time the unit is turned ON. The PEAK value will continue to be displayed until “HIDE PEAK” is selected in the PEAK menu.
24
NOTE: After an Alarm Condition occurs and the Audible Alarm is si­lenced as instructed in Section 4.0, the instrument will automatically change to the “SHOW PEAK” display to inform the user of the actual level responsible for the alarm condition. To return to the standard operating screen, depress the pushbutton under “HIDE PEAK,” which appears in place of “MENU.”
6.6.2 HOLD
The HOLD feature displays the highest toxic gas or the lowest oxygen con­centration detected while the MINI-SA instrument is in the “HOLD MODE”.
To activate the “HOLD MODE”, return to the “PEAK/HOLD” menu and select the “HOLD” pushbutton. The following screen will be displayed:
HOLD MODE
ACTIVATE EXIT
Depress the “ACTIVATE” pushbutton to turn on the HOLD feature. The “EXIT” pushbutton will return to the Operating Mode without activating the HOLD feature.
In the “HOLD MODE”, the instrument will continually update and display the highest toxic gas or lowest oxygen reading. One of the following screens will appear with the words “HOLD ACTIVATED” flashing:
ppm
0
<gas>
For Toxic Gases HOLD ACTIVATED
RESET
%
O
2
For O
20.9
HOLD ACTIVATED
2
RESET
Depress the “RESET” pushbutton to turn off the HOLD feature and return the unit to the standard Operating Mode.
Note: The HOLD and PEAK function automatically reset when the calibration menu is entered.
25
6.7 ALARM SETPOINT MENU
The “ALARM SETPOINT” can be adjusted by the user in the ALARM SETPOINT menu within the following ranges:
CO = 5 to 250 ppm
Cl2 = 0.1 to 10.0 ppm
ClO2 = 0.10 to 2.00 ppm
HCN = 0.1 to 50.0 ppm
H2S = 5 to 100 ppm
NH3 = 5 to 100 ppm
NO2 = 0.2 to 5.0 ppm
O2 LO = 0.0 to 21.0%
O2 HI = 21.0 to 25.0%
PH3 = 0.10 to 5.00 ppm
SO2 = 0.5 to 20.0 ppm
Very low oxygen alarm settings are useful when the instrument is used to verify nitrogen purging. On leaving the setpoint function or on subsequent power on, the user will be prompted to verify the safety of abnormally low settings
To enter the ALARM SETPOINT menu, depress the “SELECT” pushbutton in the menu selection screen as described in Section 6.2 above. One of the following ALARM SETPOINT menu screens will appear:
<gas> ALARM SETPOINT
For Toxic Gases HI ALARM = <_> ppm
CHANGE OK
ALARM SETPOINT
For O
2
HI ALARM = 23.5%
LO ALARM = 19.5%
CHANGE OK
The setpoint shown is the factory default setting. If acceptable, depress
26
the “OK” pushbutton to return to the Operating Mode.
To change the setpoint, depress the “CHANGE” pushbutton. One of the following screens will be displayed:
<gas> ALARM SETPOINT
For Toxic Gases HI ALARM = <_ >ppm
UP DOWN
CHANGE O2 ALARM
For O
2
HI ALARM = 23.5%
LO ALARM = 19.5%
LO HI
For toxic gases, choose “UP” or “DOWN” to change the corresponding “ALARM SETPOINT”. For example, choose “UP” and one of the following screens will appear:
<gas> ALARM SETPOINT
For Toxic Gases HI ALARM = <_ >ppm
UP SET
Choose “DOWN” and one of the following screens will appear:
<gas> ALARM SETPOINT
For Toxic Gases HI ALARM = <_ >ppm
The ppm value will flash while either of these screens is displayed. Each time the “UP” or “DOWN” pushbutton is depressed, the setpoint will be changed by 1 ppm. Holding the pushbutton automatically repeats the action.
For oxygen, choose “LO” or “HI” to change the corresponding “ALARM SETPOINT”. For example, choose “LO” and the following screen will ap­pear:
DOWN SET
27
CHANGE O2 ALARM
For O
2
LO ALARM = 19.5 %
UP DOWN
To raise or lower this setpoint, select the “UP” or “DOWN” command. For example, choose “DOWN” and the following screen will appear:
CHANGE O2 ALARM
For O
2
LO ALARM = 19.5 %
DOWN SET
The % value will flash while this screen is displayed. Each time the “DOWN” pushbutton is depressed, the setpoint will be decreased by 0.1%. If the “UP” pushbutton had been chosen, the value would be increased by 0.1% each time the pushbutton was depressed.
The operation of the “HI” setpoint is exactly the same as for the “LO”. Re­member the “Oxygen LOW ALARM SETPOINT” is adjustable in increments of 0.1% from 0% to 21.0%, and the “Oxygen HIGH ALARM SETPOINT” is adjustable in increments of 0.1% from 21.0% to 25.0%.
For all sensors, when the desired setpoint is reached, depress the “SET” pushbutton. If the “SET” pushbutton is not depressed while in this screen, after fifteen (15) seconds the unit will automatically revert back to the previ­ous setpoint.
If the “SET” pushbutton is depressed, the “ALARM SETPOINT” screen will reappear and display the new setpoints in one of the following screens:
<gas> ALARM SETPOINT
For Toxic Gases HI ALARM = <_> ppm
CHANGE OK
O
ALARM SETPOINT
2
For O
2
HI ALARM = 22.0%
LO ALARM = 18.5%
CHANGE OK
28
Depressing the “OK” pushbutton will display one of the following screens:
<gas> ALARM
For Toxic Gases
NON-LATCHING
LATCH OK
O2 ALARM
For O
2
NON-LATCHING
LATCH OK
The status of the alarms will be displayed as “NON-LATCHING” or “LATCHING”. Status can be changed by depressing the left pushbutton. Depress the “OK” pushbutton to leave the status as indicated. If neither pushbutton is selected, after fifteen (15) seconds the unit will automati­cally return to the Operating Mode with the new alarm setpoints in effect and the alarms at the status indicated.
Note: The instrument will not operate using any new setpoint for determining an alarm condition until it returns to the operating mode after a setpoint has been changed.
While in the setpoint menu the instrument remains operational. If an alarm condition occurs, the instrument will override the setpoint menu and display the alarm as appropriate. The changed setpoint will not take effect and the previous settings will be reestablished.
No user menu’s can be accessed when a instrument is in an alarm condition until the unit is turned “OFF”. When turned back “ON”, if an alarm state occurs during the first minute of operation, all user menu’s are accessible.
6.8 TWA / STEL MENU (TOXIC GASES ONLY)
The SCOTT INSTRUMENTS MINI-SA has factory preset “TWA/STEL” alarms. When the values for these alarms are exceeded, the instrument alarm will activate as described in Section 4.0. The TWA (Time Weighted Average) function of the instrument averages the level of toxic gas over an eight (8) hour period and alarms when the preset TWA value of in parts per million is exceeded.
29
The STEL (Short Term Exposure Level) function measures the level of toxic gas during a fifteen (15) minute interval and alarms when the preset STEL value in parts per million is exceeded. The alarm levels are as follows:
STEL TWA
Ammonia 35 ppm 25 ppm
Carbon monoxide 100 ppm 25 ppm
Chlorine 1 ppm 0.5 ppm
Chlorine Dioxide 0.3 ppm 0.1 ppm
Hydrogen Cyanide 4.0 ppm 2.0 ppm
Hydrogen sulfide 15 ppm 10 ppm
Nitrogen Dioxide 5.0 ppm 3.0 ppm
Phosphine 1 ppm 0.3 ppm
Suphur Dioxide 5 ppm 2 ppm
To enter the TWA/STEL Menu, depress the “SELECT” pushbutton as de­scribed in Section 6.2 when the TWA/STEL Menu screen is displayed. The instrument will display one of the following screens:
<gas> TWA/STEL MENU
For Toxic Gases STEL = 0 ppm
TWA = 0 ppm @ 0:05
RESET SHOW
DISABLE TWA/STEL
STEL and TWA values are calculated. STEL averages exposure for the latest fifteen (15) minutes. TWA shows the eight-hour average exposure. Both measures start from zero when the instrument is turned on. The TWA reading accumulates total exposure since turn-on or “RESET” and is calcu­lated as if the total exposure occurred in a standard eight-hour shift.
The “RESET/DISABLE” pushbuttons allows the user to reset the values and time on the display to “0”. One of following screens will be displayed when the “RESET/DISABLE” pushbuttons are depressed:
<gas> TWA/STEL MENU
TIME AVERAGE VALUES
For Toxic Gases
RESET DISABLE
30
Depressing the “RESET” pushbutton will reset the TWA value, the STEL value, and time all to “0”. The instrument will automatically go back into the Operating Mode and begin collecting measurement data.
The “DISABLE” pushbutton turns off the TWA/STEL functions. If the TWA/ STEL functions are off, one of the following screens will be displayed:
<gas> TWA/STEL MENU
TIME AVERAGE VALUES
For Toxic Gases
RESET ENABLE
Depress the “RESET” pushbutton to go back to the Operating Mode. De­press the “ENABLE” pushbutton to turn on the TWA/STEL functions.
When the TWA/STEL functions are enabled, selecting the TWA/STEL Menu as described in Section 6.2 will display one of the following screens:
<gas> TWA/STEL MENU
STEL = 0 ppm
For Toxic Gas TWA = 0 ppm @ 0:05
RESET SHOW
ON/OFF TWA/STEL
Depress the “SHOW TWA/STEL” pushbutton to display the STEL and TWA values on the Operating Mode screen in real time in one of the following screens:
ppm
0
<gas>
STEL = 1 ppm
For Toxic Gases TWA = 0 ppm @ 0:03
MENU
To turn off the TWA/STEL display, go back to the TWA/STEL MENU. The screen will now show “HIDE TWA/STEL” over the right pushbutton. Select this pushbutton and the standard Operating Mode screen will appear.
The TWA/STEL data measurement functions will be reset by turning off the instrument whether the function is enabled or disabled.
31
6.9 CONFIGURATION MENU
The CONFIGURATION menu of the SCOTT INSTRUMENTS MINI-SA pro­vides the user with the following two (2) optional features:
BATTERY SAVER mode can extend operational battery life to sev­eral hundred hours by turning off the instrument graphic display dur­ing normal non-alarm operation.
PASSWORD PROTECTION menu can be used to lock out all menus except ZERO CALIBRATION for toxic gas sensors or AUTO CALI­BRATION for an oxygen sensor, preventing unauthorized users from changing instrument settings.
6.9.1 BATTERY SAVER MODE
The BATTERY SAVER mode extends the operational life of the batteries by turning off the graphic display when it is not needed. To use this feature, depress the “SELECT” pushbutton when the CONFIGURATION menu screen is displayed as described in section 6.2. The following screen will appear:
CONFIGURATION
BATTERY PASSWORD
SAVER MENU
Depress the “BATTERY SAVER” pushbutton and the following screen will appear:
CONFIGURATION
ENABLE BATTERY
SAVER MODE?
YES NO
Depress the “YES” pushbutton to enable the BATTERY SAVER mode. When the BATTERY SAVER mode is enabled, the MINI-SA will beep once and the Operating Mode screen will appear. To the right of the the battery level indicator three (3) vertical plus signs (+) will appear, which signify that the BATTERY SAVER mode is enabled. After thirty (30) seconds, the instrument will beep once and the graphic display will go blank. While the graphic dis­play is blank, the Green Status LED will flash at an increased rate of once every five (5) seconds.
While the BATTERY SAVER mode is on, depressing either pushbutton will turn on the graphic display for thirty (30) seconds.
32
If an alarm condition occurs while in the BATTERY SAVER mode, the graphic display will come on automatically and remain on until the alarm is reset. After the alarm indication clears, the graphic display will remain on for thirty (30) seconds and then the display will go blank as described above.
If the BATTERY SAVER mode had already been enabled , the graphic display will show the following:
DISABLE BATTERY
SAVER MODE?
YES NO
Depressing the “YES” pushbutton will turn off the BATTERY SAVER mode.
6.9.2 PASSWORD PROTECTION
The PASSWORD PROTECTION menu is intended to be used by authorized per­sonnel for the purpose of locking out the menus to prevent changes to the instrument settings. A four digit numeric password can be set up to restrict access to all menus except AUTO CALIBRATION or ZERO CALIBRATION. To use this feature, de­press the “SELECT” pushbutton when the CONFIGURATION MENU screen is dis­played as described in section 6.2. The following screen will appear:
CONFIGURATION
BATTERY PASSWORD
SAVER MENU
Depress the “PASSWORD MENU” pushbutton and the following screen will appear:
ENTER PASSWORD
0000
UP SET
Use the “UP” and “SET” push-buttons to enter the correct password. (The MINI-SA comes from the factory with the default password: 1234.) Entering an incorrect password will display an INCORRECT PASSWORD message and allow the user to enter the password again. When the correct password
33
has been entered, the instrument will display the following screen:
PASSWORD PROTECTION
FOR ADVANCED MENUS
TURN CHANGE
ON PASSWORD
The MINI-SA is shipped from the factory with password protection off. The “TURN ON” pushbutton will bring up the following display:
PASSWORD PROTECTION
LOCK OUT ACCESS TO
ADVANCED MENUS?
YES NO
Pressing “NO” returns the instrument to normal operation with no password required to use all menu functions. Pressing “YES” returns the instrument to normal operation with the password required to access any menu except ZERO CALIBRATION for toxic gas sensors or AUTO CALIBRATION for an oxygen sensor. Until the instrument is turned off and then on, all menus will remain accessible.
After the instrument is turned off and then on, when scrolling through the menus as described in section 6.2, the following screen will appear after the AUTO CALIBRATION or ZERO CALIBRATION menu:
PASSWORD REQUIRED
FOR ADVANCED MENUS
ENTER
EXIT PASSWORD
To disable the PASSWORD PROTECTION choose the “ENTER PASSWORD” pushbutton when the above screen appears. The instrument will display the following screen:
ENTER PASSWORD
0000
UP SET
34
Use the “UP” and “SET” push-buttons to enter the correct password. Enter­ing the incorrect password will display an INCORRECT PASSWORD mes­sage and allow the user to enter the password again. When the correct password has been entered, the instrument will display the following screen:
PASSWORD PROTECTION
FOR ADVANCED MENUS
TURN CHANGE
OFF PASSWORD
Depress the “TURN OFF” pushbutton and the instrument will display the menus as described in section 6.2.
The PASSWORD menu in the CONFIGURATION menu will still require that the password be properly entered to change the password and/or turn on the PASSWORD PROTECTION feature after it has been turned off.
To change a password, navigate through the CONFIGURATION menu to the PASSWORD menu:
PASSWORD PROTECTION
FOR ADVANCED MENUS
TURN CHANGE
ON PASSWORD
Depress the “CHANGE PASSWORD” pushbutton. The following display will appear with the first digit flashing:
ENTER NEW PASSWORD
0000
UP SET
Depressing the “UP” pushbutton will increase the value of the flashing digit from 0 through 9. Depressing the “SET” pushbutton will set the digit and advance to the next digit. For example, we will set the password to “5678”. When the fourth digit (number 8) has been set by depressing the “SET” pushbutton, the display will change to the following screen:
35
VERIFY NEW PASSWORD
5678
UP SET
Note: If the pushbutton is not depressed within fifteen (15) seconds, the instrument will beep once and automatically go back to the oper­ating mode and the password will be unchanged.
If you wish to change the password shown in the VERIFY NEW PASS­WORD screen, depress the “CHANGE” pushbutton. The previous ENTER NEW PASSWORD screen will appear and changes can be made. Be sure
to record the password in a safe pace for future use.
NOTES:
1. The MINI-SA comes from the factory with the default password “1234”.
2. To insure the integrity of the PASSWORD PROTECTION feature of the MINI-SA, only SCOTT INSTRUMENTS Sales Representatives and Service Departments will be authorized to enter a master pass­word to access the MINI-SA menus in the event the password en­tered by the user is lost. If assistance is required, contact your SCOTT INSTRUMENTS Sales Representative or Service Department with the serial number of the instrument in question.
36
7.0 ERROR MESSAGES
WARNING
A SCOTT INSTRUMENTS MINI-SA INSTRUMENT THAT DISPLAYS ANY ERROR MESSAGE MUST BE REMOVED FROM SERVICE IM­MEDIATELY AND THE FAILURE DIAGNOSED AND REPAIRED BY AUTHORIZED PERSONNEL. DO NOT ASSUME THAT A PROBLEM HAS CORRECTED ITSELF IF A PREVIOUSLY DISPLAYED ERROR MESSAGE DOES NOT APPEAR A SECOND TIME.
During calibration, if the SCOTT INSTRUMENTS MINI-SA detects a sensor with insufficient output to be calibrated, the following screen will appear:
<gas> SENSOR
For Toxic Gases SPAN CALIBRATION
FAILURE
TURN OFF RECAL
OXYGEN SENSOR
For O2 SPAN CALIBRATION
FAILURE
TURN OFF RECAL
If one of these screens appears, verify that the proper gas is flowing to the instrument and all fittings are properly connected. Then press “RECAL” to attempt the calibration again. If calibration is not successful, press “TURN OFF” and service the instrument as needed.
On subsequent power on, the INITIALIZING line on the first screen displayed (SCOTT INSTRUMENTS MINI-SA...) will be replaced with CALIBRATION REQUIRED, until the instrument is successfully recalibrated. This may require replacing the sensor.
37
8.0 CALIBRATION GASES
Contact your SSCOTT INSTRUMENTS distributor, sales representative, or the SCOTT INSTRUMENTS Service Center for information regarding avail­ability of Calibration Gases and accessories for the SCOTT INSTRUMENTS MINI-SA Portable Gas Detection Instrument.
Calibration eguipment from SCOTT INSTRUMENTS:
Calibration Kits ( Including gas cylinder, regulator, cal adapter and tubing)
P/N 096-2389 25 ppm H2S P/N 096-2390 50 ppm CO P/N 096-2391 18.5% O P/N 096-2656 5 ppm Cl P/N 096-2658 1 ppm PH P/N 096-2660 25 ppm NH P/N 096-2662 10 ppm SO
2
2
3
3
2
P/N 096-2893 10 ppm HCN P/N 096-2894 5 ppm NO
2
Replacement Cylinders
P/N 077-0272 25 ppm H2S P/N 077-0273 50 ppm CO P/N 077-0274 18.5% O P/N 077-0276 10 ppm SO P/N 077-0277 25 ppm NH
2
2
3
P/N 077-0278 10 ppm HCN P/N 077-0279 5 ppm Cl P/N 077-0280 5 ppm NO P/N 077-1239 1 ppm PH
2
2
3
38
9.0 SPECIFICATIONS
HYDROGEN SULFIDE SPECIFICATIONS
Range:
0 to 100 ppm (for H2S 100) 0 to 500 ppm (for H2S 500)
Response Time:
T90 < 30 seconds
Operating Temperature Range:
-40 to +50 oC (-40 to +122 o F)
Graphic Display Temperature Range:
-10 to +60 oC (+14 to +140 o F)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Carbon Monoxide at 300 ppm: < 2 ppm Sulfur Dioxide, 5 ppm: - 1 ppm Nitric Oxide, 35 ppm: < 0.7 ppm Hydrogen, 10000 ppm: < 10 ppm Nitrogen Dioxide, 5 ppm: - 1 ppm
Consult Factory for additional information.
39
CARBON MONOXIDE SPECIFICATIONS
Range:
0 to 500 ppm (for CO 500) 0 to 999 ppm (for CO 999)
Response Time:
T90 < 25 seconds
Operating Temperature Range:
-20 to +50 oC (-4 to +122 o F)
Graphic Display Temperature Range:
-10 to +60 oC (+14 to +140 oF)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Filtered against H2S, 15 ppm: < 0.5 ppm Filtered against SO2, 5 ppm 0 ppm Nitric Oxide, 35 ppm: < 3 ppm Nitrogen Dioxide, 5 ppm: - 1 ppm Chlorine, 1 ppm 0 ppm Hydrogen, 100 ppm: < 40 ppm Ethylene, 100 ppm: < 50 ppm Ethanol, 200 ppm: 0 ppm
Consult Factory for additional information.
40
SULPHUR DIOXIDE SPECIFICATION
Range:
0.0 to 99.9 ppm
Response Time:
T90 < 35 seconds
Operating Temperature Range:
-20 to +50 oC (-4 to +122 o F)
Graphic Display Temperature Range:
-10 to +60 oC (+14 to +140 o F)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Carbon Monoxide at 300 ppm: < 3 ppm Hydrogen Sulfide, 15 ppm: 0 ppm Nitric Oxide, 35 ppm: 0 ppm Nitrogen Dioxide, 5 ppm: - 5 ppm
Consult Factory for additional information.
41
PHOSPHINE SPECIFICATIONS
Range:
0.00 to 9.99 ppm
Response Time:
T90 < 30 seconds
Operating Temperature Range:
-20 to +50 oC (-4 to +122 o F)
Graphic Display Temperature Range:
-10 to +60 oC (+14 to +140 oF)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Arsine at 150 ppb: 100 ppb Silane, 1000 ppb: 900 ppb Diborane, 300 ppb: 105 ppb Germane Dioxide, 600 ppb: 550 ppb Sulphur Dioxide, 5 ppm: 1 ppm Hydrogen, 1000 ppm: 1 ppm Ethylene, 100 ppm: 1 ppm Carbon Monoxide, 1000 ppm: 5 ppm
Consult Factory for additional information.
42
CHLORINE SPECIFICATION
Range:
0.0 to 99.9 ppm
Response Time:
T80 < 60 seconds
Operating Temperature Range:
-20 to +50 oC (-4 to +122 o F)
Graphic Display Temperature Range:
-10 to +60 oC (+14 to +140 o F)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Carbon Monoxide at 300 ppm: 0 ppm Hydrogen Sulfide, 15 ppm: -3 to 0 ppm Sulphur Dioxide, 5 ppm: 0 ppm Nitric Oxide, 35 ppm: 0 ppm
Consult Factory for additional information.
43
AMMONIA SPECIFICATIONS
Range:
0 to 200 ppm
Response Time:
T90 < 150 seconds
Operating Temperature Range:
-25 to +30 oC (-13 to +86 o F)
Graphic Display Temperature Range:
-10 to +60 oC (+14 to +140 oF)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Carbon Monoxide at 300 ppm: 0 ppm Hydrogen Sulfide, 15 ppm: 15 ppm Sulpher Dioxide, 5 ppm: 4 ppm Nitric Oxide, 35 ppm: 7 ppm Nitrogen Dioxide, 5 ppm: 0 ppm Chlorine, 1 ppm -0.5 ppm Hydrogen, 200 ppm: 0 ppm Hydrogen Cyanide, 10 ppm: 0.5 ppm Hydrogen Chloride, 5 ppm: 0 ppm Ethylene, 100 ppm: 0 ppm
Consult Factory for additional information.
44
CHLORINE DIOXIDE SPECIFICATION
Range:
0.00 to 2.00 ppm
Response Time:
T80 < 60 seconds
Operating Temperature Range:
-20o to +50 oC (-4o to +122o F)
Graphic Display Temperature Range:
-10 to +60 oC (+14 to +140 o F)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Chlorine at 3 ppm: 1 ppm Carbon Monoxide at 300 ppm: 0 ppm Hydrogen Sulfide, 15 ppm: -3 to 0 ppm Sulphur Dioxide, 5 ppm: 0 ppm Nitric Oxide, 35 ppm: 0 ppm
Consult Factory for additional information.
45
OXYGEN SPECIFICATIONS
Range:
0.0 to 25.0%
Response Time:
T95 = 15 seconds
Operating Temperature Range:
-20 to +50 oC (-4 to +122 oF)
Graphic Display Temperature Range:
-10 to +60 oC (+14 to +140 oF)
Operating Humidity:
0 to 99% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±2% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode
Consult Factory for additional information.
46
HYDROGEN CYANIDE SPECIFICATION
Range:
100 ppm
Response Time:
T90 < 200 seconds
Operating Temperature Range:
-20o to +50 oC (-4o to +122o F)
Graphic Display Temperature Range:
-10 to +60 oC (+14 to +140 o F)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Carbon Monoxide at 300 ppm: <15 ppm Hydrogen Sulfide, 15 ppm: 90 ppm Sulpher Dioxide, 20 ppm: <75 ppm Nitric Oxide, 35 ppm: <0 ppm Nitrogen Dioxide, 5 ppm: < -10 ppm Ethylene, 100 ppm: <25 ppm
Consult Factory for additional information.
47
NITROGEN DIOXIDE SPECIFICATIONS
Range:
150 ppm
Response Time:
T90 < 25 seconds
Operating Temperature Range:
-20 to +50 oC (-4 to +122 oF)
Graphic Display Temperature Range:
-10 to +60 oC (+14 to +140 oF)
Operating Humidity:
0 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Carbon Monoxide at 300 ppm: <15 ppm Hydrogen Sulfide, -1.2 ppm Sulpher Dioxide, 5 ppm: 0 ppm Nitric Oxide, 35 ppm: 0 ppm Nitrogen Dioxide, 5 ppm: 0 ppm Chlorine, 1 ppm: -1 ppm
Consult Factory for additional information.
48
NOTES
49
Telephone: (610)363-5450 • Fax: (610)363-0167
40014802 Rev. D 08/21/03 CN #3181
®
251 Welsh Pool Road
EXTON, PA 19341
50
Loading...