The STATIM® 900J autoclave has been designed expressly for the rapid
sterilization of instruments intended for immediate use. This allows you to
sterilize instruments between patients. To decrease instrument turn-around
time, the Statim 900J does not have a post-sterilization drying cycle. Quick
drying and cooling of the load at the end of the cycle is achieved through
evaporation after the automatic opening of the sterilization chamber drawer.
Because of the unique operation of the Statim 900J, once the instruments are
exposed to ambient conditions, their continued sterility cannot be assured. It is
recommended to use these instruments immediately after they are comfortable
to the touch. The Statim 900J has three sterilization cycles:
The Solid Instrument Cycle is used to sterilize solid metal instruments. The
sterilization phase of this cycle is at 134°C for 3.5 minutes. For a detailed
description of this cycle, please see page 14.
The Hollow or Wrapped Instrument Cycle is used to sterilize dental handpieces.
The sterilization phase of this cycle is at 134°C for 5 minutes. For a detailed
description of the cycle please see page 15.
The Rubber or Plastic Instrument Cycle is used to gently sterilize rubber and
plastic instruments. The types of materials appropriate for this cycle are outlined
in Preparing and Loading Instruments. The sterilization phase of this cycle is at
121°C for 15 minutes. For a detailed description of this cycle please see page
15.
3
Page 4
STAT IM 900J Operator’s Manual
2. Important Information
2.1 Disclaimers
Use only steam-process distilled water in your Statim. Specially filtered
water should not be used. Never use tap water. Usage of any water other than
steam-process distilled water may damage the unit and void the warranty.
Do not permit any person other than certified personnel to supply parts for,
service or maintain your Statim. SciCan shall not be liable for incidental,
special or consequential damages caused by any maintenance or services
performed on the Statim by a third party, or for the use of equipment or parts
manufactured by a third party, including lost profits, any commercial loss,
economic loss, or loss arising from personal injury.
Never remove the cover of the unit and never insert objects through holes or
openings in the cabinetry. Doing so may damage the unit and / or pose a hazard
to the operator.
IMPORTANT: Follow local guidelines governing verification of the sterilization
procedure.
4
Page 5
STAT IM 900J Operator’s Manual
1. Reservoir cap
2. Liquid crystal display (LCD)
3. Keypad
4. Level indicator
5. Drawer
6. Sterilization chamber
7. Power switch
8. Power cord
9. Leveller feet
10. Waste bottle (condenser)
and coil assembly
TM
11. Teflon
exhaust tube
5
4
3
2
12. Exhaust fitting 13. Drain Tube
The following symbols
appear in the margins
of this book
A potential
Hazard to
the operator.
Rubber /
The following
symbols appear
on the unit
Plastic cycle
Multifunction
button
13
7
8
6
14. Communications port
A situation which
may lead to a
mechanical failure.
Hollow or Wrapped
Instrument cycle
Solid Instrument
cycles
10
1
Figure 2
Important
Information.
Caution: Hot
Surface and / or
Hot Steam
Caution: Risk of
electrical shock.
Disconnect supply
before servicing
14
12
9
11
When you recieve your Statim 900J packing carton, the items listed below will be
included.If any of the items are missing, contact your dealer immediately so the
situation can be corrected.
Waste Bottle
part number 01-100724S
Bottle Lid Fitting
part number 01-100735S
Tube Mounting Hardware
Power Cord
part number 01-101763S
Exhaust Tube
part number 01-100204S
Stat-Dri
part numbers: 2OZPLUS,
32OZPLUS, 8OZPLUS,
8OZPLUS (with trigger)
Statim 900J
Operator’s Manual
part number 95-110106
5
Page 6
STAT IM 900J Operator’s Manual
3. Installation
3.1 Environmental Considerations
There are several factors that may effect the performance of your
Statim. Please review these factors, and select a suitable location in
which to install the unit.
• Temperature and Humidity
Avoid installing your Statim in direct sunlight or close to a heat source
(e.g. vents or radiators). The recommended operating temperature is
between 15 ˚C to 25 ˚C (59 ˚F to 77 ˚F) with humidity of 25 % to 70 %.
• Spacing
The vents and openings on the Statim should remain uncovered and
unobstructed. Leave a minimum of 50 mm / 2 " between the top, sides
and back of the unit and any wall or partition.
• Venting
The Statim should be operated in a clean, dust-free environment.
• Work Surface
The Statim should be placed on a flat, level, water-resistant surface.
Never install and operate the unit on a sloped surface.
• Electromagnetic Environment
The Statim has been tested and meets applicable standards
for electromagnetic emissions. While the unit does not emit any
radiation, it may itself be affected by other equipment which does.
We recommend that the unit be kept away from potential sources of
interference.
• Electrical Requirements
Use properly grounded and fused power sources with the same voltage
rating as indicated on the label at the back of your Statim. Avoid
multiple outlet receptacles. If using a surge suppressor power bar, plug
in one Statim only.
3.2 Unit Placement
To stabilize the Statim 900J, follow these steps:
1. Align the bubble (1)
in the level indicator
(2) on the front panel
to the right front quadrant
(or 4 o’clock) of the target
by adjusting the four
leveller feet (3). This will
ensure that the Statim 900
functions properly.
2. Ensure that all four feet are
in contact with the surface,
and that the Statim 900J
does not rock.
1. Bubble
2. Level indicator
3. Leveller feet
4. Reservoir
cap
6
Figure 3
2
1
4
3 (4x)
Page 7
STAT IM 900J Operator’s Manual
3. Installation
3.3 Connecting the waste bottle
To connect the waste bottle (1), follow these steps:
1. Insert one end of the
Teflon™ exhaust tube
(2) into the exhaust
fitting (3) on the back of
the Statim 900J. Push
the tube into the fitting
as far as it will go to
ensure that the tube is
connected tightly.
2. Cut the tube to length
and slide the nut, ring
and cone onto the tube.
Place the free end of
the tube into the fitting
in the lid of the waste
bottle and hand tighten.
Do not coil the exhaust
tube.
3. Place the waste bottle
near the Statim 900J.
We recommend that
you keep it in a cabinet
below the unit. For convenience, the tube can be routed through a hole, 7
mm (5/16") in diameter, drilled through the counter-top and secured with the
nylon clamps provided.
4. Unscrew the lid with the copper condenser coil assembly from the waste
bottle. The lid and coil should come out together. Fill the waste bottle with
water to the “MIN” line and replace the lid and copper condenser assembly.
5. If required, a low-level disinfectant prepared according to instructions on
the disinfectant packaging may be added to the waste bottle. This may help
prevent unpleasant odours or discoloration of the waste water if it is not
changed regularly.
1
1. waste Bottle
2. Teflon™ exhaust tube
3. exhaust fitting
2
4. nut
5. ring
6. cone
7. threaded nut
Figure 4
3
4
5
6
7
3.4 Filling the Statim Reservoir
To fill the reservoir, follow these steps:
1. Remove the reservoir cap from the top of
the Statim 900J and pour only steamprocess distilled water into the reservoir
until nearly full (a maximum of 3 litres).
Never use tap water. We recommend using
a funnel when filling the reservoir to avoid
spills. Pour slowly when filling the top portion
of the reservoir. Use only steam processed
distilled water containing less than 5 ppm
total dissolved solids (having conductivity of
less than 10 μS / cm).
7
Figure 5
Page 8
STAT IM 900J Operator’s Manual
3. Installation
2. Replace the reservoir cap to ensure that nothing other than steam-process
distilled water enters the reservoir.
3. Empty the waste bottle and refill to the MIN indicator.
3.5 Priming the Statim 900J pump
To Prime the Statim 900J pump follow
these steps:
1. Carefully turn the Statim 900J sideways,
and locate the drain tube beside the
power switch, on the back of the unit.
Gently pull the tube out as far as possible
so the free end can be positioned over a
water container.
2. Remove the stopper from the end of the
drain tube and allow water to drain from
the tube into a container for at least
30 seconds.
3. Replace the stopper, and feed the tube back into the unit until secure.
4. Plug the power cord into the power cord receptacle on the back of the unit.
5. Plug the power cord into a standard, grounded, wall outlet.
Figure 6
3.6 Language Selection
To set the language feature for the Statim 900J, follow these steps. Watch the
blinking cursor on the LCD.
1. Turn the Statim 900J power switch OFF.
2. Press and hold down the Hollow or Wrapped Instrument Cycle button on
the front of the Statim 900J, and turn the power switch at the back of the
unit ON. Release the Solid Instrument Cycle button. The current language
selection will appear.
3. To scroll to the next language press the Hollow or Wrapped Cycle button.
To scroll to the previous language press the Solid Instrument Cycle button.
4. When the desired language feature is indicated on the LCD, press
the MULTIFUNCTION button shown to save the change and return to
normal operation.
3.7 Setting the Time and Date
To set the time and date, follow these steps and watch the blinking cursor
on the LCD:
1. Power the Statim OFF.
2. Press and hold the Solid Instrument cycle button.
3. While pressing the Solid Instrument cycle button, power the Statim ON.
The LCD display reads:
14:23 06/22/2007
HH:MM DD/MM/YYYY
8
Page 9
STAT IM 900J Operator’s Manual
3. Installation
4. Use the cycle buttons to select and change the selected field’s value. To
increase a field’s value, press the Solid Instrument cycle button. Hold the
button down to increase the value.
5. To decrease the value, press the Hollow or Wrapped Instrument cycle
button.
6. To select the next field, press the RUBBER AND PLASTICS cycle button.
7. To save changes and return to the regular operating mode, press the
MULTIFUNCTION button.
8. To quit without making changes, power the Statim OFF.
3.8 Preparing Unit for Use
Once the unit is installed and before any instruments are sterilized, run two
Hollow or Wrapped instrument cycles (See section 4.5 Running a Cycle). Clean
the drawer and the instrument basket using a soft cloth to wipe the inside
surfaces and rinse thoroughly with tap water. Once the instrument basket and
sterilization chamber are clean and dry, coat the interior surfaces with Stat-Dri.
3.9 Shipping the Statim 900J
If you must ship the Statim 900J, follow these instructions:
1. Drain the water from the reservoir. Carefully turn the Statim 900J backwards.
Locate the drain tube beside the power switch on the back of the unit.
Release the drain tube from the clip.
2. Remove the stopper from the end of the drain tube and allow for the water
to drain freely from the reservoir. When the water no longer drips from the
drain tube replace the stopper and secure the drain tube under the clip.
9
Page 10
STAT IM 900J Operator’s Manual
4. Instructions for Use
4.1 Using The Sterilization Chamber Drawer
The Statim 900J keypad has four buttons.
The first three buttons will initiate one
of the three different sterilization cycles.
These cycles are described in detail
in section 4.4 Selecting A Cycle. The
fourth button on the keypad, the
MULTIFUNCTION button, opens and
closes the Sterilization Chamber Drawer.
Instrument
• To open the Sterilization Chamber Drawer:
Solid
Cycle
Handpieces /
Hollow
Instrument
Cycle
1. Press the MULTIFUNCTION button and the drawer will open automatically.
2. Lift the instrument basket from the sterilization chamber.
3. Load instruments to be sterilized into the instrument basket. Ensure you
follow the instrument manufacturer’s reprocessing instructions to ensure you
do not damage the instruments.
4. Replace the instrument basket into the sterilization chamber.
5. Press the MULTIFUNCTION button and the drawer will close automatically,
or press one of the sterilization cycle buttons and the drawer will close and
the selected sterilization cycle will begin automatically.
• To Remove The Sterilization Chamber Drawer:
To remove the sterilization chamber drawer, follow these steps:
1. Open the drawer by pressing the
MULTIFUNCTION button on
the keypad.
2. Turn the unit power switch to OFF.
3. Locate the green latch inside the
front opening of the machine near the
top right hand side. Push the latch to the
right and gently pull the drawer out of the unit.
Be careful. The instruments and the metal parts of the drawer may be hot.
Rubber
& Plastic
Cycle
• To Replace The Sterilization Drawer:
To replace the sterilization chamber drawer, follow these steps:
1. Center the drawer into the opening.
2. Rest the rear ends of the drawer on the front rollers in the opening.
3. Tilt the front of the drawer upwards and push the drawer into the unit. Level
the drawer and continue to push until the drawer clicks into the intermediate
position. Lift and push the drawer past the position where the green latch is
engaged by the drawer.
4. If installed properly, the drawer cannot be removed without pushing
the green latch.
5. Turn the power ON and close the drawer by pressing the MULTIFUNCTION
button.
10
Page 11
STAT IM 900J Operator’s Manual
4. Instructions for Use
4.2 Preparing and Loading Instruments
Before loading any instruments into the Statim, consult the instrument
manufacturer’s reprocessing instructions.
• Clean Instruments
Clean and rinse all instruments before loading them into the cassette.
Disinfectant residues and solid debris may inhibit sterilization and damage the
instruments, the sterilization chamber, and the Statim. Lubricated instruments
must be wiped thoroughly and any excess lubricant should be removed before
loading.
• Solid Instruments
Arrange unwrapped instruments in the instrument basket so that they do not
touch one another. This ensures that steam reaches all surfaces.
Instruments must not be stacked or piled in the instrument basket as this will
impede the sterilization process.
The Statim 900J provides cycles to sterilize instruments for both immediate and
future use.
• Immediate use
Use the Solid Instrument cycle only if the instruments are going to be used
immediately following sterilization. Handpieces should be sterilized in the Hollow
or Wrapped instrument cycle even for immediate use. Since the item is not
wrapped for processing, the sterilized item should be used immediately (as soon
as it cools off) and must be protected from re-contamination during transport to
the point of use. The rapid timing of the Solid Instrument cycle (about 6 minutes)
is sufficient to sterilize instruments between patients.
• Hollow or Wrapped Instruments
If the instruments are going to be stored for future use, place the instruments
into autoclave bags according to the instrument manufacturer’s instructions.
Pouch each instrument individually. Place one pouch on the bottom rack and
another one on the top rack of the basket. The Statim 900J unit does not
have an active drying process and handling of wet pouches will compromise
the sterility of the instruments. Following sterilization, the wet pouches may
be allowed to dry within the drawer of the unit or can be transported to the
cabinet in the basket and allowed to dry in the basket prior to further handling.
Care should be taken that the basket and contents remain protected from recontamination during this drying period. The use of cloth wraps in the Statim is
not recommended.
SciCan recommends the use of plastic / paper autoclave bags such as SPS™,
Medi-Plus™ and Chex All II™. Steri-Stik™ paper / paper may also be used. The
use of cloth wraps in the Statim is not recommended. Loosely pack instruments
in the bags to allow steam penetration to all instrument surfaces.
Care must be taken to ensure that the combined weight of the loaded bags does
not exceed 1.0 kg (2.2 lbs).
11
Page 12
STAT IM 900J Operator’s Manual
4. Instructions for Use
• Rubber and Plastic Instruments
The following materials CAN be sterilized in the Statim:
nylon, polycarbonate (Lexan™), polypropylene, PTFE (Teflon™), acetal
(Delrin™), polysulfone (Udel™), polyetherimide (Ultem™), silicone rubber, and
polyester. When loading rubber and plastic instruments in the instrument basket,
leave a space between the instruments and the chamber walls. This ensures that
steam reaches all surfaces.
The following materials CANNOT be sterilized in the Statim:
polyethylene, ABS, styrene, cellulosics, PVC, Acrylic (Plexiglas™), PPO (Noryl™),
latex, neoprene, and similar materials. Use of these materials may lead to
instrument or equipment damage. If you are unsure of your instrument’s material
or construction, do not load into your Statim until you have checked with the
instrument manufacturer.
• All Instruments
The Statim is NOT intended for sterilizing textiles, liquids or biomedical waste.
Unwrapped instruments, once exposed to ambient or external conditions,
cannot be maintained in a sterile state. If sterile storage is desired, wrap the
instruments to be sterilized in autoclave bags, according to the instrument
manufacturer’s instructions. Then allow the wrapped cycle to run. At completion
of cycle, handle load aceptically
Best Practice: Allow instruments (wrapped or unwrapped) to dry completely
prior to handling. Wrapped or pouched instruments must not touch each other
to promote drying and enable effective sterilization.
SciCan recommends the final user carefully choose the most appropriate
sterilization cycle according to the recommendations of their leading infection
control authorities and local regulatory guidelines / recommendations.
• Routine Monitoring
SciCan recommends that chemical process indicators suitable for steam
sterilizers be used in routine monitoring of the Statim 900J. Of the many brands
on the market, it is important to choose the correct type. Use only indicators
designed for steam sterilization at 134°C for the Solid Instrument and Hollow or
Wrapped Instrument cycles, and indicators designed for steam sterilization at
121°C for the Rubber or Plastic Instrument cycle. Use at least one indicator for
each load processed. Follow the indicator manufacturer’s instructions closely.
CycleBiological
Solid
Instruments
Hollow or
wrapped
instruments
Rubber & Plastic
Instruments
Indicator
Attest™
1261
Attest™
1261
Attest™
1262
Incubation
Period
24 HoursPlace the indicator in the center of
24 HoursPlace the indicator in the center of
48 HoursPlace the indicator in the center of
Test Pack
the unloaded instrument basket.
the unloaded instrument basket.
the unloaded instrument basket.
12
Page 13
STAT IM 900J Operator’s Manual
4. Instructions for Use
For detailed instructions on how to handle, use and dispose of both the
biological and chemical indicators, please consult the product literature
accompanying the 3M Attest™ biological indicators or contact the manufacturer
directly.
To use the indicators with the Statim 900J, follow these steps:
1. Place the appropriate biological indicator in the Statim 900J chamber.
2. Process the load in the sterilizer according to your usual practice.
3. Ensure that the message “Cycle Complete” is displayed on the LCD after
the cycle is finished.
4. Recover the biological and / or chemical indicator and process further
according to the literature that accompanied the indicator.
At the first indication of a potential sterilization failure:
1. Do not process any more instruments until favourable test results have
been returned.
2. Ensure the correct indicator type was chosen.
3. Ensure the chamber was not overloaded. Consult the earlier portion of this
section for proper loading instructions.
4. If the results do not change, do not process any more instruments within the
Statim 900J and contact your SciCan dealer for further assistance.
Because the turnaround time for the 3M Attest™ is up to 48 hours, it is
recommended that the tests be conducted so that the incubation period occurs
during a period of planned downtime such a the last cycle before a weekend.
4.3 Instrument Mass Guide
InstrumentTypical Instrument Mass
Scissors30 g (0.96 ozt)
Dental Scalers20 g (0.64 ozt)
Forceps15 g (0.48 ozt)
Dental handpiece40 to 60 g (1.3 to 1.9 ozt)
Unwrapped instrument rack225 g (7.23 ozt)
Suction cannula10 g (0.32 ozt)
Plastic mouth mirror8 g (0.3 ozt)
Impression Tray15 to 45 g (0.48 to 1.5 ozt)
Plastic x-ray positioning ring20 g (0.64 ozt)
NOTE: The above weights are to be used as reference only. For exact weights
of your instruments, consult the manufacturer's specifications.
13
Page 14
STAT IM 900J Operator’s Manual
ABDEC
A Warming up
B Conditioning
C Pressurizing
D Sterilizing
E Venting
4. Instructions for Use
4.4 Selecting a Cycle
The Statim 900J Autoclave has been designed expressly for the rapid
sterilization of instruments intended for immediate use. This allows you to
sterilize instruments between patients. Quick drying and cooling of the load
at the end of the cycle is achieved through evaporation after the automatic
opening of the sterilzation chamber drawer. Because of the unique operation
of the Statim 900J, once the instruments are exposed to ambient conditions,
their continued sterility cannot be assured. It is recommended to use these
instruments immediately after they are comfortable to the touch. Each cycle can
be selected by pressing the Solid Instrument, Hollow or Wrapped Instrument, or
Rubber / Plastic buttons.
The types of instruments, sterilization requirements and a graph depicting each
cycle characteristics are described over the following pages.
The Statim 900J has three sterilization cycles, each designed to sterilize a
specific type of instrument. The instruments will remain sterile after a successful
cycle until the drawer is opened. The types of instruments, sterilization
requirements, and a graph depicting each cycle are described over the next few
pages. Consult the Instrument Mass Guide in Section 4.3 for information on how
to make up an appropriate load for the masses specified for individual cycles.
1. Solid Instrument Cycle
The Solid Instrument Cycle is used to sterilize solid metal instruments free from
deep cavities or holes, such as pliers, burrs, scalers and forceps.
The maximum recommended load for the Solid Instrument Cycle is 1.0 kg.
To select this cycle, press the Solid Instrument Cycle button. The cycle then
starts automatically. This cycle is illustrated in the graph below. During the
sterilizing phase of this cycle, the temperature in the chamber is maintained at
a minimun of 134ºC for 3.5 minutes. See Preparing and Loading Instruments
before running this cycle.
SOLID 134°C / 3.5 min
14
Page 15
STAT IM 900J Operator’s Manual
A B C E D
A Warming up
B Conditioning
C Pressurizing
D Sterilizing
E Venting
ABCED
A Warming up
B Conditioning
C Pressurizing
D Sterilizing
E Venting
4. Instructions for Use
2. Hollow or Wrapped Instrument Cycle
The Hollow or Wrapped Instrument Cycle is used to sterilize dental handpieces.
The maximum recommended load for the Handpiece cycle is 1.0 kg.
To select this cycle, press the Hollow or Wrapped Instrument Cycle button.
The cycle then starts automatically. This cycle is illustrated in the graph below.
During the sterilizing phase of this cycle, the temperature in the chamber is
maintained at a minimum of 134ºC for 5 minutes. See Preparing and Loading
Instruments before running this cycle.
HOLLOW OR WRAPPED 134°C / 5 min
3. Rubber or Plastic Instrument Cycle
The Rubber or Plastic Instrument Cycle is used to sterilize instruments
constructed of the materials listed under the heading Rubber or Plastic
Instruments in Preparing and Loading Instruments. The maximum recommended
load for the Rubber or Plastic Instrument Cycle is 0.1 kg.
To select this cycle, press the Rubber or Plastic Instrument Cycle button. The
cycle then starts automatically. This cycle is illustrated in the graph below.
During the sterilizing phase of this cycle, the temperature in the chamber is
maintained at a minimum of 121ºC for 15 minutes. See Preparing and Loading
Instruments before running this cycle.
RUBBER / PLASTIC 121°C / 15 min
15
Page 16
STAT IM 900J Operator’s Manual
4. Instructions for Use
4.5 Running a Cycle
To operate each cycle, follow these steps and watch the LCD:
Turn the power switch at the back of
the unit ON. The LCD display reads:
Selecting the MULTIFUNCTION
button changes the position
STATIM 900J R1.11
SELECT A CYCLE
PRESS TO CLOSE
if the drawer
is fully open
of the drawer
or
Selecting a cycle will close
the door and start that cycle.
SELECT A CYCLE
PRESS TO OPEN
SELECT A CYCLE
PRESS TO OPEN
if the drawer
is closed
if the drawer
is ajar
Select the apropriate cycle
button on the keypad located
beside the LCD.
As the drawer closes, the display reads
SOLID INSTRUMENTS
134ºC for 3.5 mins
HOLLOW OR WRAPPED
INSTRUMENTS 134ºC for 5 mins
Solid
Instrument
Cycle
Hollow or
Wrapped
Instrument
Cycle
Rubber
& Plastic
Cycle
RUBBER AND PLASTICS
121ºC for 15 mins
and the cycle begins. Messages change throughout the cycle. For example,
if the SOLID INSTRUMENT CYCLE is selected, the following LCD messages
would be displayed:
SOLID INSTRUMENTS
WARMING UP
SOLID INSTRUMENTS
WARMING UP 96ºC
SOLID INSTRUMENTS
CONDITIONING 110ºC
16
Page 17
STAT IM 900J Operator’s Manual
4. Instructions for Use
SOLID INSTRUMENTS
PRESSURIZING 130ºC
STERILIZING
134ºC 212 kPa 3:15
SOLID INSTRUMENTS
VENTING 115ºC
After venting is complete, the drawer will open automatically and the instruments
may be removed. The display reads:
SOLID INSTRUMENTS
CYCLE COMPLETE 6:02
Be careful. The instruments and metal parts of the drawer will be hot. Hot steam
will escape as the drawer opens.
This message is displayed for 30 minutes as a reminder of a successfully
completed cycle. When sterilized instruments are removed press the
MULTIFUNCTION button to clear the CYCLE COMPLETE message, and prepare
the machine for the next cycle. This ensures that the CYCLE COMPLETE
message is not displayed when a new load of unsterilized instruments is placed
in the chamber.
After 30 minutes the LCD message will automatically change to:
SELECT A CYCLE
PRESS TO CLOSE
Once exposed to a non-sterile environment at the end of a complete cycle, the
continued sterility of the chamber contents can no longer be assured. Therefore,
the Statim 900J should only be used for the prevention of cross-contamination,
unless other steps are taken to ensure continued sterility.
17
Page 18
STAT IM 900J Operator’s Manual
4. Instructions for Use
4.6 Stopping a Cycle
To stop a cycle, press the MULTIFUNCTION button. If the MULTIFUNCTION
button is pushed, or the unit detects a problem while operating, the cycle will
stop and the amber active light will flash. Once a cycle has been interrupted, the
MULTIFUNCTION button must be pressed before another cycle can be started.
The display reads the following message:
BUTTON PRESSED
PLEASE WAIT
If the display shows the message, CYCLE FAULT or NOT STERILE, the chamber
contents are not sterile! See section 6. Troubleshooting for more information.
18
Page 19
STAT IM 900J Operator’s Manual
5. Maintenance
5.1 Cleaning the Chamber
It is good clinical practice to keep your Statim 900J sterilization chamber clean.
We recommend that the interior surface of the chamber be cleaned at least once
a week with dishwashing soap or mild detergent that does not contain chlorine.
Remove the drawer from the unit to make cleaning easier. See, section 4.1.
Afterwards, rinse thoroughly to remove all traces of soap. After the chamber is
clean treat the entire inside surface with Stat-Dri. Stat-Dri causes water to form
an even coat on treated surfaces, without beading. The water in contact with
the hot chamber surfaces evaporates more efficiently. Spotting is minimized and
instruments dry faster. Stat-Dri is available from your dealer or from SciCan.
Cleaning the inside of the chamber is very important if you regularly sterilize
lubricated instruments.
5.2 Cleaning the Exterior Surfaces
Use a soft cloth moistened with a mild cleaning solution or mild disinfectant to
clean all outer surfaces. Do not use solvents or harsh chemicals.
5.3 Preventative Maintenance Schedules
To ensure trouble-free performance, both the operator and the dealer must
follow a preventative maintenance schedule.
NOTE: Please refer to your National, Regional, State or Safety laws for any
additional reoccurring user testing that may be required.
19
Page 20
STAT IM 900J Operator’s Manual
5. Maintenance
The schedules below describes the necessary actions.
Operator
Daily Chamber Wash the interior of the sterilization chamber
Unit After a week or period of extended non-use,
Water Reservoir Drain Daily
Waste Bottle Empty the waste bottle every time you refill
technician
Once Chamber Check the drawer, lid and seal for damage.
a year Replace if necessary.
Solenoid Valve Inspect the valve and clean if dirty.
Replace the plunger if defective.
Pump Clean the filters, replace if dirty.
Water Reservoir Check the reservoir for dirt. Clean and
Calibration Calibrate the unit
with a chlorine-free dishwashing detergent
or soap. Rinse thoroughly with water.
run the first cycle with an empty load.
the reservoir. Fill with tap water up to MIN
line. You can also add some chlorine-free
disinfectant.
rinse with steam process distilled water if
necessary.
20
Page 21
5. Maintenance
5.4 Spare Parts List
01-100204S Exhaust Tube
01-101561S Packaging
01-101565S Condenser Bottle
01-101582S Replacement Cap
STAT IM 900J Operator’s Manual
01-103636S Wire Basket
01-104343S Plug - Drain Tubing
01-101562S Statim 900J Drawer
2OZPLUSStat Dri Plus 2oz.
32OZPLUSStat Dri Plus,
8OZPLUSStat Dri Plus,
8OZPLUSTStat Dri Plus, 8oz.
Complete
bottle w/sprayer
32oz. bottle
8oz. bottle
bottle w/trigge
21
Page 22
STAT IM 900J Operator’s Manual
6. Troubleshooting Problems
ProblemSolution
Unit Does not Power
ON.
Drawer does not
open / close when a
cycle button or the
multifunction button
is pressed.
Water under
the Machine
Cycle aborted,
message “NOT
STERILE”
and “CYCLE
FAULT” message
Message “CYCLE
FAULT 12”
Check that the unit is plugged into a properly
grounded outlet, that the power cord is firmly seated
at the rear of the machine, and that the power switch
is ON.
Check the condition of the line circuit breaker or fuse.
The drawer may not be installed properly. Push the
drawer further into the unit assuring that it is past
the position where the green latch is engaged by the
drawer.When installed properly the drawer may not be
pulled out of the unit without pushing the green latch.
See, section 4.1, Using the Sterilization Chamber
Drawer
Check that water was not spilled when refilling the
reservoir. We recommend that you use a funnel when
filling your Statim 900J .
Make sure the plug in the drain tube is secured. If the
leak persists, unload the chamber, unplug the unit,
and call your dealer.
Wait a few minutes and attempt another cycle before
proceeding.
Check the exhaust tube for kinks or obstructions. Try
another cycle. If the problem persists, record the cycle
fault message number and contact your dealer.
Do not attempt another cycle. There is an internal
electrical problem. Turn the unit off, unload the
sterilization chamber and call your dealer.
Be careful. The instruments and metal parts of the
drawer will be hot. Hot steam will escape as the
drawer opens.
Excessive steam
is issuing from the
front of the unit
Open and close the drawer. Attempt another cycle.
Be careful. The instruments and metal parts of the
drawer will be hot. Hot steam will escape as the
drawer opens.
If the leak persists, unload the chamber, unplug the
unit, and call
your dealer.
22
Page 23
6. Troubleshooting Problems
ProblemSolution
STAT IM 900J Operator’s Manual
Message “WATER
QUALITY IS NOT
ACCEPTABLE”.
Machine will not start
Message “REFILL
RESERVOIR” and
“EMPTY WASTE
BOTTLE”. Machine
will not start.
Inadvertently, you have used water which is not
steam-process distilled or is improperly distilled. Drain
the reservoir and refill with steam-process distilled
water To drain the unit, use the drain tube at the
back of the machine. Refer to the steps described in
Shipping the Statim 900J.
The level of water in the reservoir is low. Refill the
reservoir, using steam-process distilled water and
empty the waste bottle. Refer to the steps described
in Maintaining the Statim 900J.
23
Page 24
STAT IM 900J Operator’s Manual
7. The Communication Port
7.1 The RS232
Communication Port
All Statim units that do not have an
internal printer will have an RS232
communication port that will allow you
to connect to an external printer or to a
SciCan Data Logger. For printing, you
will need to purchase a recommended
printer (see list below) from your local
computer or electronics store. For data
storage, you can purchase SciCan’s
USB Data Logger to record and store
cycle information onto a mass storage
device (MSD) such as a USB Flash Drive
or SD memory card.
Printer
OR
Data Logger
Figure 5
Printer Model End Of
Epson
TM-U220D (C31C515603) CR/LF
Citizen
IDP-3110-40 RF 120B
Star Micro
SP212FD42-120
Star Micro
SP216FD41-120
Star Micro
SP512MD42-R
SciCan Data LoggerEnd Of Line
For mass storage device*N/A960032 [0xd2]
* Due to a variety of software programs/viewers that can be used to view the
data files stored on a mass storage device such as a USB memory stick, we
recommend the above settings.
For your Statim to communicate to a specific device, you must enable this
function through the User Setup menu. Follow the instructions in section 7.2
Installing the SciCan Data Logger below to enable communication to either an
external printer or the SciCan Data Logger.
Line
CR/LF
CR
CR
CR/LF
CR/LF
CR/LF
Serial Port
Bitrate
9600
9600
9600
9600
9600
Serial Port
Bitrate
Printer user °
character
248 [0xF8]
N/A
210 [0xd2]
210 [0xd2]
210 [0xd2]
Printer
user °
character
24
Page 25
STAT IM 900J Operator’s Manual
7. The Communication Port
7.2 Installing the SciCan
Data Logger
SciCan’s USB Data Logger can record and store cycle information onto a mass
storage device (MSD) such as a USB Flash Drive or SD memory card. These
instructions are for units with a 9-pin communication (RS232) port located at the
back of the unit.
User Setup Menu
To install the SciCan Data logger,
User Mode
Time/Date Setup
Language Setup
Unit ID Setup
Water Quality
Last Printout
RS232
End of Line CR/LF
Serial Port Bitrate
Printer User º Char
Save and Exit
Exit
access the user setup menu
by powering up the unit while
pressing the MULTIFUNCTION
button. Follow the instructions
below to complete installation
steps.
STEP 1
Selecting the USB Flash/MSD Option
RS232
1. Power up unit while pressing the MULTIFUNCTION button to access the
User Setup menu.
2. Using the Solid Instrument or Hollow or Wrapped Instrument Cycle buttons,
scroll to RS232 and select it by pressing the Rubber and Plastic Cycle
button.
3. From the RS232 menu, use the Solid Instrument or Hollow or Wrapped
Instrument Cycle buttons to scroll down to the USB/FLASH MSD option and
press the Rubber and Plastic Cycle button to select and return to the User
Setup menu.
25
USB/Flash MSD
Page 26
STAT IM 900J Operator’s Manual
7. The Communication Port
Setting the Serial Port Bitrate to 9600
Serial Port Biterate
1. From the User Setup menu, use the Solid Instrument and Hollow or
Wrapped Instrument Cycle buttons to scroll to the Serial Port Bitrate menu
and select it using the Rubber and Plastic Cycle button.
2. From the Serial Port Bitrate menu, use the Solid Instrument or Hollow or
Wrapped Instrument Cycle buttons to scroll down to 9600 and press the
Rubber and Plastic Cycle button to select and return to the User Setup
menu.
Setting the Printer user Character; (example: 134°C)
Printer User º Char
1. From the User Setup menu, use the Solid Instrument or Hollow or Wrapped
Instrument Cycle buttons to scroll to the Printer user ° char menu and select
it using the Rubber and Plastic Cycle button.
2. From the Printer user ° char menu, using the Solid Instrument cycle button
to increase the displayed value by one, and the Hollow or Wrapped
Instrument Cycle button to increase that value by ten, enter the value 32
[0x20] and press the Rubber and Plastic Cycle button to accept and return
to the User Setup menu.
Save and Exit
9600
32 [0x20]
Save and Exit
A Save and Exit must be performed after the above settings are completed. If
this is not done, the information will revert to its prior settings.
1. From the User Setup menu, use the Solid Instrument or Hollow or Wrapped
Instrument Cycle buttons to scroll to Save and Exit.
2. Select it by pressing the Rubber and Plastic Cycle button.
The LCD will display time
and date information and
a sequence of messages:
If the time and date need to be adjusted, refer to section 3.7, Setting The Time
and Date.
HH:MM DD/MM/YYYY
"MSD NOT DETECTED"/"INSERT
MSD/FLASH"/"SELECT A CYCLE"
26
Page 27
STAT IM 900J Operator’s Manual
7. The Communication Port
Step 2
Selecting the Time and Date
HH:MM DD/MM/YYYYTime/Date
Note: If the time and date were set according to section 3.7, this does not have
to be performed again.
1. Power up unit while pressing the MULTIFUNCTION button to access the
User Setup menu.
2. Using Solid Instrument or Hollow or Wrapped Instrument Cycle buttons,
scroll to Time / Date Setup and press the Rubber and Plastic Cycle button
to select it.
3. From the Time / Date Setup menu, set the time and date using the Solid
Instrument or Hollow or Wrapped Instrument Cycle buttons to change the
displayed values and the Rubber and Plastic Cycle button to select. Press
the MULTIFUNCTION button when completed to save your selection.
Step 3
Setting Unit ID
001UNIT ID
1. Power up unit while pressing the MULTIFUNCTION button to access the
User Setup menu.
2. Using the Solid Instrument or Hollow or Wrapped Instrument Cycle buttons,
scroll to the Unit ID Setup and press the Rubber and Plastic Cycle button to
select it.
3. From the Setup Unit ID menu, use the Solid Instrument or Hollow or
Wrapped Instrument Cycle buttons to Change the displayed values and the
Rubber and Plastic Cycle button to select and move to the next digit. Press
the MULTIFUNCTION button when completed to save your selection.
Step 4
Connecting the SciCan Data Logger
1. Ensure that both the Statim unit and the SciCan Data Logger are OFF.
2. Connect the SciCan Data Logger to the Statim unit by using the serial cable.
3. Power up the SciCan Data Logger
4. Power up the Statim unit.
5. The LCD will display the following
sequence of messages:
6. Insert USB Flash Drive or SD memory card
7. After a few seconds the LCD will display the
following sequence of messages:
27
HH:MM DD/MM/YYYY
"MSD NOT DETECTED"/"INSERT
MSD/FLASH"/"SELECT A CYCLE"
HH:MM DD/MM/YYYY
USB/FLASH DETECTED/SAFELY
REMOVE MSD/SELECT A CYCLE
Page 28
STAT IM 900J Operator’s Manual
7. The Communication Port
7.3 Cycle Printout
Overview
1. Model: STATIM 900J
software: SJ00R604
2. Unit Identifier: autoclave
has been set up as
number 000
3. Cycle counter: the number of
cycles having been run on the
unit = 6
4. Time / Date: 4:10 pm
Dec. 4, 2007
5. Cycle Name: SOLID
INSTRUMENT CYCLE
6. Cycle Name: Cycle name
parameters 134 C for 3.5 min.
7. Cycle clock: starting
at 0:00
8. Warm up complete:
start of the conditioning
phase is 1:35 (see cycle
graph – ‘A’ phase
complete, start of
‘B’ phase)
9. Start Time of Pressurization:
2:40 (start of ‘C’ phase)
10. Start Time of Sterilization:
3:14 (start of ‘D’ phase)
11. Temp. / Press. & Time at start
of sterilization (‘D’ phase).
12- 17 Cycle pararmeters at 30
second intervals.
18. Temp. Press. & Time of end
of sterilization phase (end of D
phase.)
19. Time venting start 6:44 (start
of E phase.)
20. Sterilization was successfully
completed.
1. Statim 900J SJ00R604
2. Unit # : OOO
3. CYCLE NUMBER 00006
4. 16:10 04/12/2007
5. SOLID INSTRUMENTS
6. 134 C for 3.5 min.
7. CYCLE START 0:00
8. CONDITIONING 1:35
9. PRESSURIZING 2:40
10. STERILIZING 3:14
11. 136.5 C 227kPa 3:14
12. 135.7 C 220kPa 3:44
13. 136.4 C 226kP a 4:14
14. 136.5 C 227kPa 4:44
15. 135.9 C 221kPa 5:14
16. 136.4 C 227kPa 5:44
17. 135.9 C 221kPa 6:14
18. 136.4 C 226kPa 6:44
19. VENTING 6:44
20. CYCLE COMPLETE 6:59
28
Page 29
STAT IM 900J Operator’s Manual
8. Limited Warranty
For a period of one year, SciCan Ltd. guarantees that the STATIM 900J, when
manufactured by SciCan Ltd. in new and unused condition, will not fail during
normal service due to defects in material and workmanship that are not due to
apparent abuse, misuse, or accident. In the event of failure due to such defects
during this period of time, the exclusive remedies shall be repair or replacement,
at SciCan Ltd.’s option and without charge, of any defected part(s) (except
gasket and filters), provided SciCan Ltd. is notified in writing within thirty(30)
days of the date of such a failure and further provided that the defective part(s)
are returned to SciCan Ltd. prepaid.
This warranty shall be considered to be validated, if the product is accompanied
by the original purchase invoice from the authorized SciCan Ltd. dealer, and
such invoice identifies the item by serial number and clearly states the date of
purchase. No other validation is acceptable. After one year, all SciCan Ltd.’s
warranties and other duties with respect to the quality of the product shall
be conclusively presumed to have been satisfied, all liability therefore shall
terminate, and no action or breach of any such warranty or duty may thereafter
be commenced against SciCan Ltd.
Any express warranty not provided hereon and any implied warranty or
representation as to performance, and any remedy for breach of contract which,
but for this provision might arise by implication, operation of law, custom of
trade or course of dealing, including any implied warranty of merchantability
or of fitness for particular purpose with respect to all and any products
manufactured by SciCan Ltd. is excluded and disclaimed by SciCan Ltd. If
you would like to learn more about SciCan Ltd. products and features, visit our
website at www.scican.com.