SciCan STATIM 2000,STATIM 2000S,STATIM 2000 G4 Service Manual

STAT
CASSETTE AUTOCLAVE
Service Manual
IM
2000/2000S/2000 G4
2000 Service Manual 96-108639 Rev 3.0. Copyright 2012 SciCan Ltd. All rights reserved.
STAT
Table of Contents
STATIM 2000 / 2000S / 2000 G4 Service Manual
96-108639 Rev 3.0
Table of Contents
About this Service Guide ....................................................... 5
Safety compliance ........................................................... 5
1.Identifying STATIM 2000 Units .......................................... 7
Identifying software versions ........................................... 7
Identifying controller board types ................................... 9
Identifying steam generator types ................................. 10
2. Tools, Maintenance Schedules,
Procedures and Testing .................................................. 13
Tools .............................................................................. 13
The control box ............................................................. 15
Calibration cassettes and reference meters................... 16
Calibration cassettes .............................................. 16
Reference meters .................................................... 17
Annual service requirements ......................................... 31
Routine maintenance schedule and procedures ........... 34
Draining the reservoir .............................................. 35
Cleaning the cassette ............................................. 35
Cleaning the water reservoir filter ........................... 36
Cleaning the water reservoir ................................... 36
Filters on the STATIM 2000/2000S ............................... 36
Bacteria retentive filter ............................................ 37
Compressor filter .................................................... 37
Filters on the STATIM 5000/5000S ............................... 37
Bacteria-retentive filter ............................................ 37
Compressor filter .................................................... 38
Replacing the cassette seal .......................................... 39
Cleaning the cover ........................................................ 40
Shipping the unit ........................................................... 40
Upgrading the firmware on STATIM G4 units ............... 41
Using the STATIM G4 remote access function ............. 42
Electrical safety testing ................................................. 44
Dielectric strength test (hi-pot) ................................ 44
Protective bonding impedance test
(ground continuity) .................................................. 44
Water conductivity testing ............................................. 45
3. Diagnostics and Troubleshooting Cycle Faults ............ 47
Basic mechanical diagnostics ....................................... 47
Using software as a diagnostic tool .............................. 47
For units with 2.x/5.x, 6.x controllers ............................ 47
For units with 7.x controllers (with service menu) .......... 48
Using the service menu on the STATIM G4
(Rev. 7 controller board) ................................................ 64
Troubleshooting cycle faults ......................................... 65
Cycle Fault Numbers and descriptions
for revision 3.x and 4.x controllers ................................ 66
Cycle Fault Numbers and descriptions on units
with revision 2.x/5.x/6.x and 7.x type controller
boards (1.xx/2.xx/4.xx/5.xx and 6.xx software .............. 67
4. STATIM 2000 Calibration ................................................ 81
Calibration procedure 1: for Rev. 3.x or 4.x boards ...... 81
Calibration procedure 2: for Rev. 2.x or 5.x boards,
software R1.xx or R2.xx ................................................ 85
Calibration procedure 3: for Rev. 6.X board,
software Rev. 4.xx and 5.xx .......................................... 89
Calibration procedure 4: for Rev. 7.x board,
software Rev. 6.xx ......................................................... 93
Calibration Procedure 8: for non-S G4 units with
Rev. 7.x Board, Software Rev. 7.xx, SL 00R1.xx .......... 97
5. STATIM 2000S Calibration ............................................ 103
Calibration procedure 5: for Rev. 2.x or 5.x
boards,software R1.xx and R2.xx ............................... 103
Calibration procedure 6: for Rev. 6.x boards,
software R4.xx and R5.xx ........................................... 108
Calibration procedure 7: for Rev. 7.x boards,
software R6.xx ............................................................. 113
Calibration Procedure 9: for S-class G4 Units with
Rev. 7.x Board,Software Rev. 7.xx, SL 00R1.xx ......... 118
6. Cover Assembly ............................................................. 123
For STAT
Removing the cover .................................................... 123
Reinstalling the cover .................................................. 125
Removing and replacing the armature gasket ............ 126
Removing and replacing the fascia ............................. 126
Removing and replacing the membrane keypad ........ 126
Removing and replacing the LCD ............................... 129
For STAT
Removing the cover .................................................... 131
Reinstalling the cover .................................................. 133
Removing and replacing the armature gasket ............ 134
Removing and replacing the LCD overlay .................... 135
Removing and replacing the LCD touchscreen ........... 136
Removing and replacing the fascia ............................. 138
IM
2000 Classic
IM
2000 G4
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 1
Table of Contents
STATIM 2000 / 2000S / 2000 G4 Service Manual
96-108639 Rev 3.0
Table of Contents
7. Electrical and Electronic Components ........................ 141
Controller board .......................................................... 141
Identifying the controller board type ..................... 141
Removing the controller board assembly ............. 143
Replacing/refitting the controller board assembly 146
Controller board fuse values ................................. 148
STAT Removing and replacing the STAT
LCD controller board ............................................ 149
Removing and replacing the STAT
USB port ............................................................... 151
Removing and replacing the STAT
speaker ................................................................. 151
Removing and replacing the STAT
Ethernet port ......................................................... 152
Microprocessor ........................................................... 153
Processor types .................................................... 153
Removing the microprocessor .............................. 154
Replacing/refitting the microprocessor ................. 155
Microswitch ................................................................. 156
Removing the microswitch ................................... 156
Replacing the microswitch .................................... 157
Adjusting the microswitch ..................................... 157
Mains components ...................................................... 158
Removing and replacing a fixed power cord ........ 159
Removing a detachable power cord ..................... 160
Replacing/refitting a detachable power cord ........ 160
Power (on/off) switch ................................................... 161
Removing the power switch ................................. 161
Replacing/refitting the A.C. Power switch ............ 162
Receptacle / line filter .................................................. 163
Removing the receptacle / line filter ..................... 163
Replacing/refitting the receptacle / line filter ........ 163
LCD DC Power Source for STAT Removing and replacing the STAT
LCD DC power source .......................................... 165
8. Pumps, Reservoir and Compressor ............................. 167
Water pumps ............................................................... 167
Identifying pump types ......................................... 167
Testing pump flow ................................................ 168
Water pump recovery repair procedure ...................... 170
Performing filter maintenance ............................... 170
Performing internal pump filter maintenance ........ 171
Performing water reservoir filter maintenance ...... 173
Performing inline filter maintenance ...................... 174
Performing pump tube replacement ..................... 174
IM
2000 G4 Electronics ....................................... 149
IM
2000 G4
IM
2000 G4
IM
2000 G4
IM
2000 G4
IM
2000 G4 .............. 165
IM
2000 G4
Removing the Fluid-O-Tech pump assembly ....... 177
Testing the Fluid-O-Tech pump diode .................. 178
Reinstalling the Fluid-O-Tech pump assembly ..... 179
Removing the SciCan or Ulka pump assembly .... 180
Reinstalling the SciCan or Ulka pump assembly .. 180
Reservoir and water sensors ....................................... 181
Removing the reservoir ......................................... 181
Reinstalling the reservoir ....................................... 182
Troubleshooting .................................................... 183
Compressor ................................................................. 189
Removing the compressor .................................... 190
Reinstalling the compressor ................................. 191
9. Solenoid Valve and Pressure Transducer ................... 193
Solenoid valve ............................................................. 193
Inspecting and repairing the
type B solenoid valve ............................................ 193
Removing the solenoid valve ................................ 196
Refitting the solenoid valve ................................... 196
Pressure transducer .................................................... 197
Identifying and checking the
pressure transducer .............................................. 197
Removing the pressure transducer ....................... 201
Reinstalling the pressure transducer .................... 202
Removing the pressure interface board ................ 203
Reinstalling the pressure interface board ............. 203
Removing and replacing the
pressure interface board battery ........................... 205
10. Armature, Isoplate, and Probe Bracket
and Chamber Thermocouple ...................................... 207
Armature and isoplate ................................................. 207
Removing armature and isoplate ........................... 207
Reinstalling armature and isoplate ........................ 209
Removing the isoplate from the
armature-isoplate .................................................. 210
Replacing and reassembling the isoplate
into the armature .................................................... 211
Probe bracket and
chamber thermocouple ................................................ 212
Removing the probe bracket and chamber
thermocouple ........................................................ 212
Replacing and reinstalling the probe bracket
and chamber thermocouple .................................. 215
Aligning the probe bracket and chamber
thermocouple ........................................................ 218
Page 2
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
Table of Contents
STATIM 2000 / 2000S / 2000 G4 Service Manual
96-108639 Rev 3.0
Table of Contents
11. Steam Generator, Check Valve, Thermal Fuse and
Pressure Relief Valve ................................................... 221
Steam generators ......................................................... 221
Replacing 1st generation and
stainless steel steam generators (matrix) ............... 222
Removing the aluminum steam generator ............. 223
Reinstalling the aluminum steam generator ........... 224
Removing the stainless steel steam generator ...... 226
Reinstalling the stainless steel steam generator .... 227
Check valves ................................................................ 229
Testing the check valve ......................................... 229
Removing and replacing the check valve .............. 230
Steam generator thermal fuses .................................... 232
Removing the thermal fuse
on the aluminum steam generator ......................... 233
Replacing the thermal fuse
on the aluminum steam generator ......................... 235
Removing and replacing the thermal fuse
on the stainless steel steam generator .................. 236
Pressure relief valve ..................................................... 236
Checking the pressure relief valve ......................... 237
Removing and replacing the pressure relief valve ...237
12. Printer and Data Logger ............................................. 241
Printer .......................................................................... 241
Removing and replacing the STATprinter cover .... 241
Removing and replacing the printer module ......... 242
Removing and replacing the
printer interface board .......................................... 243
Removing and replacing the battery ..................... 243
Adjusting print quality ........................................... 243
Removing the printer assembly ............................ 244
Replacing the printer ............................................. 245
Installing thermal paper into the statprinter .......... 246
Removing paper jams ........................................... 247
Recommended printers ........................................ 248
Installing the external printer ................................. 249
Data Logger ................................................................. 250
Installing the SciCan Data Logger ........................ 251
Data Logger troubleshooting ................................ 254
Transferring data from the USB drive or SD card .. 256
Data Logger specifications ................................... 257
13. Installation .................................................................... 259
Environmental considerations ..................................... 259
Unit placement ............................................................ 260
Connecting the waste bottle ....................................... 260
Filling the STATIM reservoir ........................................ 261
Priming the STATIM pump .......................................... 261
Powering up and setting up the unit ........................... 261
Setting the time and date ............................................ 262
Language selection ..................................................... 262
Assigning unit identifier number .................................. 262
Setting/editing serial number ...................................... 263
Setting air drying time ................................................. 264
Connecting STAT
Setting up the G4 web portal ...................................... 266
Preparing unit for use .................................................. 271
Shipping the unit ......................................................... 271
Unit installation checklist ............................................. 271
14. STATIM Operational Diagram ..................................... 273
IM
G4 to a network ......................... 265
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 3
About this Service Guide
96-108639 Rev 3.0
About this Service Guide
This STATIM service manual was created to act as reference for the service and repair of the STAT
IM
brand. In the interest of providing one comprehensive global service manual, this book references both the STATIM 2000 and the STATIM 2000S models. Because of this, you will note that there are chapters and sections that do not apply to the STATIM unit that you may have, or are repairing.
If you have a question about the unit you are repairing, please do not hesitate to contact your local SciCan representative for confirmation.
Hazardous voltages are accessible when the cover is removed. Disconnect the power cord before servicing the power mains portion of the controller board and associated devices.
If the cover is removed, a dielectric strength test (Hi-Pot) AND a protective bonding impedance test (ground continuity) must be performed on the STATIM when the work is completed and after the cover has been returned to the unit.
The STAT
Use only steam-process distilled water in the STATIM.
The STAT working with or near printed wiring boards. In addition, use static footstraps, grounding mats and grounded work surfaces when servicing microprocessor devices. Transport boards and devices in static protected bags.
Ensure that there is sufficient steam-process distilled water in the STAT pump.
IM
is heavy. Exercise caution and seek assistance when lifting or carrying the units.
IM
contains electronic circuitry that is static sensitive. Always wear a static strap when
IM
before activating the
Safety Compliance
When a STATIM is serviced, the safety criteria as specified by applicable international safety standards and applicable national, state, provincial, and regional laws and regulations must be observed and maintained.
The following symbols appear in the margins of this book.
A potential hazard to the operator.
A situation which may lead to a mechanical failure.
Important information
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 5
1. Identifying STATIM 2000 Units
WARNINGS AND PRECAUTIONS
If you have questions about the unit you are repairing, please do not hesitate to contact your local SciCan representative for information. Also, the STATIM is heavy. Exercise caution and seek assistance when lifting or carrying units.
EXERCISE CAUTION
t Hazardous voltages are accessible when the cover is removed.
t Disconnect the power cord before servicing the power mains portion of the
controller board and associated devices.
96-108639 Rev 3.0
PERFORM TESTS
t If the cover is removed, a dielectric strength test (Hi-Pot) AND a protective
bonding impedance test (ground continuity) must be performed on the STATIM when the work is completed and after the cover has been returned to the unit.
PROTECT THE UNIT
t Use only steam-process distilled water in the STATIM.
t The STATIM contains electronic circuitry that is static sensitive. Always wear a
static strap when working with or near printed wiring boards. In addition, use static footstraps, grounding mats and grounded work surfaces when servicing microprocessor devices. Transport boards and devices in static protected bags.
t Ensure there is sufficient steam-process distilled water in the STAT
activating the pump.
t In order to ensure adherence to the applicable safety agency approvals,
state, provincial, regional and national laws, replace components with SciCan approved parts only.
IM
before
Page 6
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
1. Identifying STATIM 2000 Units
96-108639 Rev 3.0
Identifying STAT
IM
2000 Units
(Does not include ‘E’ models)
There are two main types of STATIM 2000 units, the G4 touchscreen version and Statim Classic with keypad. Each of these two versions have a North American specification or “non S” unit, and the European specification or “S” unit. In the Classic series, they are labeled respectively as “STAT
IM
2000” and “STATIM 2000S”.The G4 series models do not use the “S” designation in their names, but the G4 series does include S-class models for distribution in markets requiring these cycle types, and non S-class for the North American market.
There are many mechanical identifiers to differentiate between the set types, (e.g., pressure transducer for ‘S’ class, no pressure transducer for the non ‘S’ class). The details in the matrix further below will help you identify units based on their factory specifications.
Note that this section does not cover STAT contact SciCan Technical Service.
IM
2000E models. For more information on these,
Identifying Software Versions
To identify a unit’s software version, power it OFF and back ON again. During start up, a number is briefly displayed in the upper right-hand corner of the LCD that indicates which software is installed (See Figure 1). This number is also printed on the label attached to the U14 microprocessor and the U8EEPROM device on the Controller Board (See Figure 2).
Note that units with Revision 3.x or 4.x controller boards WILL NOT display the software revision on the LCD display. These units will display “SELECT A CYCLE”.
The software version number is the character to the immediate right of the R as indicated by:
Figure 1
S501R102 S501R200
example example
LCD readout on Power-up
For example:
Figure 2
Microprocessor label example:
EPROM label example:
S501R100 is software version 1.00 S501R101 is software version 1.01 S501R102 is software version 1.02 S501R200 is software version 2.00
01-xxxxxx S502R102 Sept. XX 03
52102
0319
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Product Year / week
Page 7
1. Identifying STATIM 2000 Units
Software Versions and Corresponding Calibration Procedures
96-108639 Rev 3.0
Model Approx.
Year of Manufacture
1000 1992 - 1995 N/A Aluminium
2000 1992 - 1995 N/A Aluminium
2000 1995 - 2005 S2O2R1xx
2000 2005 - 2007 S2O2R5xx Aluminium
2000
2000 G4
2000S 1995 - 2004 S2S2R1xx
2000S 2004 - 2005 S2S2R4xx Stainless
2000S 2005 - 2007 S2S2R5xx Aluminium
2000S
2000S G4
2007 Æ
2010 Æ
2007 Æ
2010 Æ
Software Revision
S2O2R2xx
S2O2R6xx Aluminium
S202R7xx Aluminium
S2S2R2xx
S2S2R6xx Aluminium
S2S2R7xx Aluminium
Steam generator Type
Internal
Internal
Aluminium Internal
External
External
External
Aluminium Internal
Steel
External
External
External
PCB Revision
3.x/4.x
3.x/4.x
2.x/5.x
6.x
7.x
7.x
2.x/5.x
6.x
6.x
7.x
7.x
Pump Compressor Pressure
transducer
3
33
33
33
33
33 N/A 3
33 3
33 3
33 3
33 3
33 3
N/A N/A N/A 1
N/A N/A 1
N/A N/A 2
N/A
N/A
(60 p.s.i.)
(68p.s.i.)
(68p.s.i.)
(68p.s.i.)
(68p.s.i.)
Validation thermocouple
3
3
N/A 5
N/A 6
N/A 6
N/A 7
N/A 9
Calibration procedure
3
4
8
Identifying Controller Board Types
There are four different types of Controller Boards that may be encountered in STATIM 2000/2000S/2000 G4 type units (See ‘PCB revision’ above). They can be identified as below (Figure 3):
Figure 3
Revision 3.x/4.x type board (1992 – 1995):
Page 8
STAT
Typical features:
t Revision number bottom right hand side. t Single, rectangular microprocessor.
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
1. Identifying STATIM 2000 Units
Revision 2.x/5.x type board (1995 – 2004):
96-108639 Rev 3.0
Typical features:
t Revision number bottom right hand
side.
t Rectangular microprocessor plus
EPROM
t ‘W1’ jumper for calibration t Blue ‘pressure interface/printer’
connector
Typical features:
t Revision number bottom right hand
side.
t Square microprocessor plus EPROM t ‘W1’ jumper for calibration t Blue ‘pressure interface/printer’
connector
Revision 6.x type board (2004 - 2007):
Revision 7.x type board (2007 to present):
Typical features:
t Up to revision 7.30, the revision
number is on the top right hand side (printed vertically)
t From revision 7.4 onwards, the
revision number is on the bottom left hand side (printed horizontally) under the connector J1
t Square microprocessor plus EPROM t NO ‘W1’ jumper for calibration t NO Blue ‘pressure interface/printer’
connector
t All components integrated on single
board
t Surface mount type component. t ‘Push in’ yellow thermocouple
connectors.
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 9
1. Identifying STATIM 2000 Units
96-108639 Rev 3.0
Identifying Steam Generator Types
There are three different types of steam generator (boiler) that may be encountered in STATIM 2000/2000S units. They can be identified as below (Figure 2):
Aluminum steam generator with internal thermocouple (1992 – 2004):
Typical features:
t 70 p.s.i. pressure relief valve (PRV)
(no ring pull)
t Thermocouple embedded in side of
steam generator
t Software required to drive this steam
generator = R1xx/R2xx (revision
3.x/4.x/2.x/5.x PCB)
Stainless steel steam generator (2004 – 2005):
Typical features:
t Software required to drive this steam
generator = R4xx (revision 6 PCB only)
Page 10
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
1. Identifying STATIM 2000 Units
Aluminum steam generator with external thermocouple, a.k.a. ‘ALEX’ boiler (2005 – present):
Typical features:
t 43.5 p.s.i. pressure relief valve (PRV)
(ring pull on S models, no ring pull on non S models)
t Thermocouple clamped to underside
of steam generator
t Blanking plug in the side of steam
generator in original thermocouple position
t Software required to drive this steam
generator is R5xx for revision 6 PCB and R6xx for revision 7 PCB
96-108639 Rev 3.0
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 11
2. Tools, Maintenance Schedules, Procedures and Testing
WARNINGS AND PRECAUTIONS
If you have questions about the unit you are repairing, please do not hesitate to contact your local SciCan representative for information. Also, the STATIM is heavy. Exercise caution and seek assistance when lifting or carrying units.
EXERCISE CAUTION
t Hazardous voltages are accessible when the cover is removed.
t Disconnect the power cord before servicing the power mains portion of the
controller board and associated devices.
96-108639 Rev 3.0
PERFORM TESTS
t If the cover is removed, a dielectric strength test (Hi-Pot) AND a protective
bonding impedance test (ground continuity) must be performed on the STATIM when the work is completed and after the cover has been returned to the unit.
PROTECT THE UNIT
t Use only steam-process distilled water in the STATIM.
t The STATIM contains electronic circuitry that is static sensitive. Always wear a
static strap when working with or near printed wiring boards. In addition, use static footstraps, grounding mats and grounded work surfaces when servicing microprocessor devices. Transport boards and devices in static protected bags.
t Ensure there is sufficient steam-process distilled water in the STAT
activating the pump.
t In order to ensure adherence to the applicable safety agency approvals,
state, provincial, regional and national laws, replace components with SciCan approved parts only.
IM
before
Page 12
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Tools, Maintenance Schedules, Procedures and Testing
Tools
STATIM specific items from SciCan
Before a STATIM can be serviced, the following special tools are required in addition to your service tool kit. These tools are available from SciCan or your nearest service depot:
1. Control box SciCan Part # 01-103141S
2. Pump tester SciCan Part # 01-100713S
3. Water conductivity tester SciCan Part # 01-103139S
4. Solenoid plunger tube wrench SciCan Part # 01-103471S
5. 9/64” ball-end allen-key (hex) screwdriver SciCan Part # 01103469S
6. Calibration cassette,
SciCan 2000 Non S models only SciCan Part # 01-103087S
7. Calibration cassette,
SciCan 2000 All models SciCan Part # 01-106366S
Generic reference devices for calibration
8. Calibrated digital thermometer with ‘K’ type probe
9. Calibrated digital voltmeter with accuracy of 0.001 Volts
10. Calibrated digital pressure meter, 0 – 400 kPa absolute/0 – 7 bar absolute More details on recommended devices can be found in section ‘Recommended Reference Meters’
Electrical safety test devices
11. Hi-Pot tester
12. Ground continuity tester
13. Static strap
14. Static bags
General tool list
t Phillips screwdriver t Flat-blade electrician’s screwdriver t Potentiometer trimmer t Needle nose pliers t Wire cutters t Wrench 3/8” A/F t Wrench 7/16” A/F t Wrench 9/16” A/F t Wrench 11/16” A/F
t Wrench 7/8” A/F
Note: This tool list is a guide and suggests what is the minimum required to remove and replace
components in the STATIM. Other tools may be required as an aid to servicing at the discretion of the individual service engineer.
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 13
2. Tools, Maintenance Schedules, Procedures and Testing
Sample equipment set up for calibration of a non-S unit
96-108639 Rev 3.0
1. Calibration cassette with detachable
thermocouple
2. Digital voltmeter
3. Digital thermometer
4. (no pressure meter required)
5. Control box
6. 2x7 ribbon cable, LCD connector
(connect to Controller Board header P3)
7. 1x7 cable, keypad connector
(connect to Controller Board header P4)
8. 2x10 molded socket, test connector
(connect to Controller Board header P1)
876
Sample equipment set up for calibration of S unit
1. Calibration cassette.
2. Digital voltmeter
3. Digital reference thermometer
with thermocouple. (Fluke 51 with 80PK-26 probe shown)
4. Digital reference pressure meter
(S only) (Druck DPI 750 R shown)
5. Control box
6. LCD connector cable (connect
to Controller Board header P3)
7. Keypad connector cable (connect
to Controller Board header P4)
8. Test connector cable (connect to
Controller Board leader P1)
8
7
6
45
1
3
2
5
Figure 1
1
3
2
Figure 2
NOTE: The following section describes in greater detail the equipment depicted in Figures 1 and 2.
Page 14
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
The Control Box
NOTE: Late model STAT
Rev.7 boards can be calibrated with the cover on using the unit’s software and keypad, making use of the Control Box as an option. For older models, use of a control box is required.
The Control Box (5) is a service tool that allows a technician to operate the unit while it is being serviced. There are two kinds of Control Boxes you may encounter: one uses toggle switches, the other uses a membrane keypad arrangement similar to the keypad on the unit.
When connecting the Control Box connectors to the Controller Board note the positions of Pin number 1 of the Control Box test connectors and Pin number 1 of the Controller Board headers.
The Control Box has a display (LCD) and a variety of control switches that allow the technician to:
IM
s with
Figure 3
1. Operate the unit independently of the keypad mechanism, to manually activate the pump,
valve or compressor, or run cycles.
2. Make frequently used Controller Board measurements by providing a common output and
rotary switch for signal selection during calibration. See the Calibration Instructions in chapters 4 and 5 of this service manual.
If only the display and button functions are being used, connect the LCD cable (6) to Controller Board connector P3, the keypad cable (7) to Controller Board connector P4 and power the STAT
IM
ON.
If the override or measurement functions are being used, connect the test connector cable (8) to connector P1 of the Controller Board in addition to the other cables.
When using a toggle switch model of the Control Box, turn the switch to the ON position to activate the desired device. To turn the device OFF, turn the switch to the AUTO position. To select, start and / or stop a cycle, activate the appropriate push-button switch on top of the Control Box.
Calibration uses the Select out +, Select out - jacks and a setting on the Rotary Switch to set the calibration operation desired for Revision 3.x/4.x Controller Boards (for STATIM 2000 units only). See Calibration Procedure 1 or chapter 4 of this service manual.
When using a keypad model of the Control Box, the keypad switch must be held down to turn the desired device ON. Vref measurements use test leads, the Vref + and Vref - jacks and a voltmeter hooked up in series. Calibration uses the Select out + and Select out - jacks, test leads and a setting on the Rotary Switch to set the calibration operation desired for Rev. 3.x/4.x Controller Boards (for STATIM 2000 units only). The keypad of this Control Box provides the same features as a STAT
IM
keypad. Note: The Control Box is for use with STATIM products ONLY.
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 15
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Calibration Cassettes and Reference Meters
Specially designed calibration cassettes allow technicians to take readings from within the cassette while the unit is in operation. In addition, calibration also requires the use of certain reference meters. Use the table below to identify the meters you will require.
Sterilizer type Part Number Description Meter types required
STAT
IM
2000 (non S)/
2000 G4
STATIM 5000 (non S)/
5000 G4
IM
STAT
STATIM 5000S/
2000S/
2000S G4
5000S G4
01-103087S Calibration cassette
STATIM 2000
01-103088S Calibration cassette
STAT
IM
5000
01-103088S Calibration Cassette
2000S B
01-106367S Calibration Cassette
5000S C
Temperature only
Temperature only
Temperature and
pressure
Temperature and
pressure
Calibration Cassettes
The cassette of the STATIM unit is effectively a fixed volume chamber that contains instruments during sterilization. Saturated steam at a specific temperature and pressure is introduced to produce the correct conditions for sterilization. The cassette is part of a sealed system designed to contain the pressurized steam. To calibrate the unit’s temperature and pressure (if appropriate) monitoring devices, a special calibration cassette must be used. The calibration cassette allows a technician to independently monitor the temperature and pressure inside the cassette, ensuring the unit is operating in accordance to its original specifications and to national/international standards.
Figure 4
The relevant cassette types are detailed in the chart above, and the following important information should be noted before choosing the appropriate cassette to use.
t Non S unit cassettes require a 1.6mm (1/16”) temperature probe and cannot be used with
a 3.2mm (1/8”) temperature probe. (See meter details for probe diameter).
t S unit cassettes have temperature fittings (one fitted and one in the cassette accessories
supplied) that will allow the use of both 1.6mm (1/16”) and 3.2mm (1/8”) temperature probe. (See meter details for probe diameter).
t S unit cassettes are designed for both temperature and pressure reference meters and are
suitable for all models, S and Non S. If these cassettes are acquired, the non S, single port cassette is not required.
t The non S unit cassettes should not be used to calibrate S class units.
t Please note that when calibrating non-S units with an S unit cassette that the pressure
meter will not be required and the test port will require the pressure tube supplied with the cassette to be connected to the pressure fitting to prevent steam leakage.
Page 16
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Reference Meters
Calibration reference meters are important for accurately setting the STATIM unit so that the correct sterilization conditions (temperature and pressure) occur in accordance with the original specifications of the unit and national/international standards.
When ordering any digital thermometer and temperature probe, ensure that the supplier is aware that the area where the most accuracy is required is between 130°C and 140°C.
Temperature meters and probes should always be calibrated as a matched pair.
Test equipment should be calibrated on a regular basis based on the manufacturer’s recommended calibration interval.
Calibration of reference equipment used with autoclaves should ALWAYS be to national or international standards by a certified calibration laboratory.
A number of meters are recommended by SciCan and are referenced as follows:
For non-S unit calibration
The following reference meters are recommended for use with 01-103087S and 01-103088S test cassettes:
t Omega HH81A single channel multifunction digital thermometer (www.omega.com) t Omega TJ36-CASS-116G-6-SMP-M Temperature Probe (with SMP-M miniature male
connector).
Note: the above probe is 1.6mm (1/16”) diameter.
For S unit calibration
The following options are recommended for use with 01-106366S and 01-106367S test cassettes:
t A set of two independent meters, one pressure and one temperature.
t An integrated hand-held pressure and temperature calibrator.
Note: The thermocouple entry fitting on the S class cassette will need to be changed from the
3.2mm (1/8”) diameter fitting to the 1.6mm (1/16”) diameter fitting if using the Omega meter.
The two independent meters have been included where the user may already have an Omega meter for non-S use who is upgrading to S class use, when only a pressure meter will be required.
They are also the lower cost options over the integrated device.
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 17
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Recommended separate temperature and pressure meters for S unit calibration
temperature
t Fluke 51 Series II Digital Thermometer.
t Fluke 80PK-26 SureGrip Tapered Temperature Probe.
Notes:
t The flexible thermocouple included with the Fluke digital thermometer will not be used for
calibrating SciCan sterilizers. If you wish to use the flexible thermocouple in the future, SciCan advises you to consult the digital thermometer’s manual for setting a temporary offset, if required.
t The above probe is 3.2mm (1/8”) diameter.
t Technicians already in possession of the Omega HH81A or equivalent, and the associated
probe can use this meter as a replacement for the Fluke 51 and 80PK-26 probe. Note that the probes are of different diameter and a cassette compression fitting change may be appropriate.
Pressure
t Druck DPI 705R Absolute Pressure Meter with external 0 - 7 bar absolute pressure
transducer with ¼” NPT female thread.
Page 18
STAT
Figure 5
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Combined Temperature and Pressure Meter
t Heise PTE-1 Handheld LCD digital calibrator complete with temperature and pressure
modules, and PT100 probe as follows:
Meter (without data logging capabilities) — PTEC = X X 4 4A
Pressure Module — HQS2 B A A 400 kPa A
Temperature Module — HQS RT1 PT-100
RTD Probe Pt-100 — PT5
Notes:
t These product references are very specific, and the local supplier of Heise equipment
(see www.heise.com) should be consulted prior to confirming order codes.
t The meter above is of the non data logging variety. If data logging is required, please
consult order code variations from Heise data sheet.
t The above probe is 3.2mm (1/8in) diameter.
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Figure 6
Page 19
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Setting Up your Reference Meters
With certain meter types, some sub-assembly procedures will need to be undertaken to connect the device to your Statim calibration cassette. All cassettes are supplied with the necessary accessories to enable you to convert your meter of choice from those recommended above, to the appropriate fitting on the cassette.
01-103087S and 01-103088S cassettes (non S calibration) with Omega HH81A reference meter and TJ36-CASS-116G-6-SMP-M Temperature Probe.
No sub-assembly required.
To insert the temperature probe (thermocouple) into the compression fittings, loosen the clamp nut attached to the cassette extension fitting and insert the probe into the fitting as far as it will go. A slight resistance will be felt as the probe passes through the seal. Tighten the clamp nut until a steam/air tight seal is achieved. DO NOT OVERTIGHTEN.
NOTE: The non S calibration cassette fitting is NOT compatible with the Fluke and Heise 1/8” (3.175mm) temperature probes.
01-106366S and 01-106367S cassettes (S-class calibration) with ALL recommended reference meters
S-class calibration cassette kits contain a number of accessories for use with various recommended reference meters and will need to be set up to suit the appropriate meters before use.
Contents of calibration cassette kit:
t$BTTFUUF4PS4BTBQQSPQSJBUFXJUIAHFOFSJDQSFTTVSFBOEUFNQFSBUVSFmUUJOHT
attached (see Figure 7 below)
t"DDFTTPSJFTBTGPMMPXTJOTJEFUIFDBTTFUUF
1 x Pressure tube
2 x Male Swagelok B-QC4-S-2PM ‘Quick Connect’ fittings.
1 x 1/8 NPT female to ¼ NPT male adaptor
1 x 1/8” (3.175mm) probe fitting.
Page 20
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Figure 7. S-class ‘generic’ fitting configuration (2000S cassette shown)
Setting up the temperature fittings for use with the recommended temperature reference meter(s)
The temperature fitting (right hand fitting when facing the front of the cassette) is fitted with a compression fitting that consists of a body, clamp nut and conical compression washer. (See Figure 8)
There are two fittings supplied with the cassette, one for a 1/8” (3.175mm) temperature probe and one for a 1/16” (1.6mm) temperature probe.
NOTE: The 1/16” fitting is attached to the cassette when supplied and the 1/8” fitting is in the pack of accessories supplied with the cassette.
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 21
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Figure 8. Temperature fitting cut-away
If you have the SciCan supplied Omega HH81A type temperature reference meter with the 1/16” (1.6mm) temperature probe (normally used with the Non S-class type calibration cassette) then the compression fitting attached to the S class cassette is correct and does NOT have to be removed.
If you have either the Fluke 51 type temperature reference meter, or Heise PTE1 type combined temperature and pressure reference meter then the 1/16” (1.6mm) compression fitting (complete) will need to be removed from the extension piece attached to the cassette and replaced with the 1/8” (3.175mm) compression fitting from the accessory kit, as both of these meters have 1/8” (3.175mm) thermocouple probes.
NOTE: PTFE tape should be used when assembling these parts to ensure a steam tight seal.
To insert the relevant temperature probe (thermocouple) into either of the appropriate compression fittings, loosen the clamp nut and insert the probe into the fitting as far as it will go. A slight resistance will be felt as the probe passes through the conical seal. Tighten the clamp nut until a steam/air tight seal is achieved. DO NOT OVERTIGHTEN.
Page 22
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Setting up the pressure fittings for use with the recommended pressure reference meter(s)
NOTE: The pressure fitting (left hand fitting when facing the front of the cassette) is fitted with an adaptor to attach ¼” BSP devices such as hypodermic pressure fittings. (See Figure 9 below). This adaptor is not used with the standard SciCan pressure tube and should be removed from the cassette extension fitting before attaching the pressure tube.
Figure 9. Pressure fitting cut-away
To set up the cassette for use with the recommended Druck DPI705R and Heise PTE1 pressure reference meters, proceed as follows:
Remove the 1/8 NPT to ¼ BSP adaptor. (The accessories for the cassette include 2 male Swagelok ‘Quick connect’ fittings (see Figure 10).
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 23
2. Tools, Maintenance Schedules, Procedures and Testing
Figure 10. Male Swagelok B-QC4-S-2PM ‘Quick Connect’ fitting.
Attach and tighten the ‘Quick Connect’ fitting to the cassette extension fitting in place of the 1/8 NPT to ¼ BSP adaptor (see Figure 11).
96-108639 Rev 3.0
NOTE: PTFE tape should be used when assembling these parts to ensure a steam tight seal.
Figure 11. Male Swagelok ‘Quick Connect’ fitting and cassette extension fitting assembly.
Attaching the pressure tube to the cassette
The pressure tube (see Figure 12) supplied with the calibration cassette is designed to insulate the pressure transducer used with the pressure reference meter from the high temperatures experienced during sterilization, which may damage the transducer.
Page 24
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Figure 12
The tube is fitted with a Swagelock, self sealing ‘Quick Connect’ QC4 female connector (see Figure 13) on either end. These connectors attach to the male ‘Quick Connect’ fitting shown in Figure 10 above.
Figure 13. Swagelock ‘Quick Connect’ QC4 female connector
To attach the pressure tube to the cassette (via the male Swagelok connector), press the male and female fitting together until a ‘click’ is heard and the couplings are firmly attached. (See Figure 14)
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 25
2. Tools, Maintenance Schedules, Procedures and Testing
Figure 14. Assembled male fitting, female fitting and cassette extension fitting
96-108639 Rev 3.0
Attaching the pressure tube to the recommended pressure reference meter (s)
Important notes:
t"UUBDIJOHUIFQSFTTVSFUVCFUPUIFSFGFSFODFNFUFSJTBDIJFWFEVTJOHUIFTFDPOENBMF4XBHFMPL
b-qc4-s-2pm fitting supplied with the cassette.
t"UUBDIJOHUIFUVCFUPUIFDPOOFDUPSJTBTEFTDSJCFEJOUIFTFDUJPOBCPWFIPXFWFSUIFTFDPOE
male fitting will need to be installed onto the pressure reference meter prior to attempting this.
t)PXUIFTFDPOENBMFmUUJOHJTBUUBDIFEXJMMEFQFOEPOUIFUZQFPGSFGFSFODFNFUFSVTFE
Assembling the Swagelok b-qc4-s-2pm ‘quick connect’ fitting to the Druck dpi 705r absolute pressure meter external 0 - 7 bar absolute pressure transducer with ¼” NPT female thread.
NOTE: To complete this assembly you will need to locate the 1/8 NPT female to ¼ NPT male adapter supplied with the S-class calibration cassette. (See Figure 15 for assembly) .
Page 26
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Figure 15. Pressure transducer with adapter and Swagelok connector attached.
Install Swagelok connector (1) onto adaptor fitting (2) DO NOT OVERTIGHTEN FITTING.
NOTE: The use of P.T.F.E. tape as shown is essential to ensure the connection is free from steam leaks. If the joint leaks, steam can enter the transducer and damage the unit internally.
Install assembly (from previous step) onto external pressure sensor (3) supplied with the Druck DPI 705R. DO NOT OVERTIGHTEN FITTING.
NOTE: The use of P.T.F.E. tape as shown is essential to ensure the connection is free from steam leaks. If the joint leaks, steam can enter the transducer and damage the unit internally.
The pressure tube can now be attached to the pressure transducer.
STATIM 2000/2000S/2000 G4 Service Guide
STAT
IM
2000/2000S Service Guide
Page 27
2. Tools, Maintenance Schedules, Procedures and Testing
Figure 16, below, shows the meter, pressure transducer and pressure tube attached to a calibration cassette.
96-108639 Rev 3.0
Figure 16
Page 28
STAT
STATIM 2000/2000S/2000 G4 Service Guide
IM
2000/2000S Service Guide
Loading...
+ 245 hidden pages