• Operator’s Manual
1
P 2 Decals, displays and controls P 3 Introduction
- Key to Pictures P 4 Operation P 5 Operation (continued)
- Additional Operations P 6 Essential information P 7 Troubleshooting P 8 Additional Information P 9 Specifications
All rights reserved. QUANTIM is a trademark of SciCan,Division of Lux and Zwingenberger Ltd.
For all service and repair inquiries:
In Canada/United States: 1-800-870-7777 International: (416) 446-4500 Email (Service and Repair): techservice.ca@scican.c om
SciCan
1440 Don Mills Rd Toronto Ontario M3B 3P9 Phone: (416) 445 -1600 Fax: (416) 445 -2727
Toll Free: 1- 800 - 667 - 7733
SciCan Medtech AG
Alpenstrasse 16, 6300 Zug, Switzerland Phone: (41-41) 727 -7027 Fax: (41-41) 727 -7029
EU R
epresentative:
SciCan
MesserschmittStr. 11 D - 86368 Gersthofen Germany
Quote your model details, serial number and date of purchase when contacting your supplier
Table of Contents
Contact
Quantim B
Quantim B
2
LL22
LL11
LL33
K
B
C D
E
H
I
G
F1
F2
L
M
LL44
SS11
SS22
SS33
SS44
CC33
CC44
CC55
CC11iiii CC22iiii
CC11ii
CC22ii
LL55 LL66 LL77 LL88
J
SS11
SS55
SS55
NN
Introduction
Quantim B
• This autoclave has been qualified to sterilize loads as defined by
prEN13060:1997*.
Cycle C1i; C1ii - Type ‘B’ cycles for wrapped, pouched, solid, hollow instruments (length to diameter ratio 750:1- length may be doubled for tubes open at both ends) and porous items.
Cycle C2i* - Type ‘S’ cycles unwrapped, unpouched, solid or hollow instruments (length to diameter ratio no greater than 110:1
- length may be doubled for tubes open at both ends).
Cycle C2ii - Type ‘B’ cycle for Bowie Dick/Helix Steam Penetration Test, or loads as defined for cycle ‘C2i’
Cycle C3; C4 - Type ‘N’ cycles for unwrapped, solid instruments.
• A ‘responsible person’ must qualify other loads as suitable. Refer to ‘Additional Information’
• Maximum load per instrument tray is 2 kg. Refer to ‘Technical Specifications’ for the maximum instrument load for the autoclave
• All instruments must be cleaned before sterilizing.
• Wrapped or pouched loads should not touch adjacent loads when placed in the rack. Pouches must be used for only one item.
• When placing on a tray, ensure items are placed on the ribs of the tray (to aid drainage), they do not touch each other, and the load does not touch the other trays or chamber in any way.
• Always use the lifting device when removing trays from the autoclave as they may be hot. Long trays should be supported at their rear as they become free of the tray carrier. Do not use an unprotected hand to hold hot trays.
Symbols:
S1 - IMPORTANT: Read Operating
Instructions before use. S2 - Do Not Use Tap Water. S3 - Maximum water fill line. S4 - Fill Water tank. S5 - Hot Parts (Behind) - Do Not Touch.
(Also behind main Door - L)
General Controls/Operation:
Note: The function is selected when the LED adjacent to the button is illuminated.
A - n/a B - Standby/ready button C - Cycle start button D - Door open button
Note: This will not work for 20 seconds after door closure
E - Select time/date/cycle count display F - Adjust time/date settings (F1/F2) G - Select vacuum cycles H - Select non-vacuum cycles I - Select ‘Additional Drying’ J - Fuses (located behind access panel) K - Printer door L - Main door M - Water fill N - Fresh water tank drain
(located behind Door ‘L’)
Cycle Controls:
G - Press to select vacuum cycles
C1i -134˚C/4min Type ‘B’ Vacuum cycle
for porous loads, wrapped, pouched, solid or hollow instruments, with drying
C1ii -134˚C /18min Type ‘B’ Vacuum cycle for porous loads, wrapped, pouched, solid or hollow instruments, with drying
C2i* -134˚C/4min Type ‘S’ Rapid vacuum cycle (14 minute total cycle time) for unpouched solid or hollow instruments, without
drying.
C2ii -134˚C/4min Type ‘B’ Vacuum cycle for Bowie Dick/Helix
Steam Penetration Test
,
without drying
H - Press to select non-vacuum cycles
C3 -121˚C/16min Type ‘N’ Non-vacuum cycle for unwrapped solid instruments.
C4 -134˚C/4min Type ‘N’ Non-Vacuum cycle for unwrapped solid instruments.
I - Press to select ‘Additional Drying’
C5 - Additional drying
After every cycle the unit automatically returns to the default cycle - ‘C1i’
Display options:
L1 - Display for ‘Time’, ‘Date’, ‘Cycle
Count’
The normal display mode is time. The date or cycle count can be selected by pressing the button (E) adjacent to the display. The selection is only valid until a cycle is started or until the unit is returned to standby mode, at which point the display will revert to the current time.
L2 - Display for ‘Temperature’ L3 - Display for ‘Pressure’
The following show the cycle status: L4 - Cycle started L5 - Heating and air bleed L6 - Sterilizing L7 - Depressurization/drying L8 - Cycle complete
* Cycle C2i not included in compliance with prEN13060
Operating Symbols, Controls and Displays
The following descriptions refer to the symbols, controls, and displays shown on Page 3 of this manual.
Types of load and loading
WARNING!
Refer to the instrument manufacturer about their suitability for
autoclaving and the max temperature the instruments can withstand
3
Thank you for choosing the SciCan QUANTIM B autoclave.
All ‘Type B’ vacuum cycles on this autoclave use the triple pulse fractionated vacuum system and are fitted with an active air leak detection system.
Non-vacuum cycles use the thermodynamic air displacement system
Drying cycles employ a ‘closed door’ system
An ‘Auto-cycle Start’ option is included which allows the time of day at which a selected cycle starts, to be programmed .
Before unpacking, refer to the ‘Manual Handling’ on page 9
Whilst unpacking check the unit for transit damage. If damage is found, please report this to the shipping agent immediately, in writing, and then notify your dealer.
Product contents will contain the following:-
• Autoclave with internal furniture
• Handbook and Warranty card
• Performance test certificate and Certificate of Conformance for pressure vessel
• Bowie Dick type test pack
• Container for waste water