HYDRim and STATim are registered trademarks of SciCan Ltd. BRAVO, HIP,
and SysTM are trademarks of SciCan Ltd. All other trademarks referred to
in this manual are the property of their respective owners.
For all service and repair inquiries:
In Canada 1-800-870-7777
United States: 1-800-572-1211
Germany:
International: (416) 446-4500
Email: techservice.ca@scican.com
Manufactured by:
+49 (0)7561 98343 - 0
SciCan Ltd.
1440 Don Mills Road,
Toronto ON M3B 3P9
CANADA
Phone: (416) 445-1600
Fax: (416) 445-2727
Toll free: 1-800-667-7733
This guide provides instructions for the servicing and repair of the HYDRim® M2 G4
Instrument Washer-Disinfector. Every attempt has been made to provide accurate, detailed
instructions.
Use to prevent soil from
drying on instruments
when they will not be
washed within one hour.
P2 – Regular Cycle
(no disinfection) – Use for
moderately soiled loose
instruments.
P3 – Heavy Duty
Cycle
with disinfection –
Use for heavily soiled
instruments and cassettes.
P4 – Custom
(Rinse: A0 between A0 =
3,500 and A0 = 6,000)
<30ºC (cold)
3-10 minute s
(default 3 minutes)
<30ºC (cold)
3-10 minute s
(default 3 minutes)
<30ºC (cold)
3-10 minute s
(default 3 minutes)
<30ºC (cold)
3-10 minute s
(default 3 minutes)
<30ºC (cold)
3-10 minute s
(default 3 minutes)
N/A
N/A
50ºC
5-15 minute s
(default 9 minutes)
50ºC
5-15 minute s
(default 9 minutes)
50ºC
5-15 minute s
(default 9 minutes)
<30ºC (cold)
2 minutes
60ºC
1 minute
60ºC
1-10 m inutes
(default 1 minute)
90ºC
5 minutes
90°C
adjust from
1 to 5
minutes
N/A
N/A
20-60
minutes
(default 30 minutes)
20-60
minutes
(default 30 minutes)
20-60
minutes
(default 30 minutes)
Total Time**
w/o Drying
6
minutes
18
minutes
35
minutes
60
minutes
Variable40L***
Consumption
Water
16 L
16 L
32L
40L***
* This is not a wash cycle. Always run a wash cycle following the rinse & hold cycle.
** Cycle times depend on the temperature and pres sure of incoming water.
*** Water consumption includes drying phase.
4
Page 5
1. Introduction
1.2 HYDRIM Unit at a glance – front and rear
Blower hose
Air gap
Ethernet and
RS232 ports
Chamber
Service panel
switch
pressure
screw
for
Feed pipe
wash arms
Fuses
Condenser
EMI filter
Chemical
reservoirs
Air vents
intlet
AC Power
pump
Recirculation
Water inlets
dosing
Chemical
valve
Water
system
softener
Rear kickplate
Drain hose
valve
Regeneration
FRONTREAR
Figure 1Figure 2
Air vents
5
Sump
tube
Chemical draining
tube
Air gap draining
Moving strap
pull
Door release
Page 6
1. Introduction
1.2 HYDRIM Unit at a glance – left and right
Chamber
Capacitors
board
LCD and
LCD controller
level
switch
I/O board
Power
supply
motor
Blower
(with filter)
drawer
Chemical
Exhaust
Sump
heater
drain
assembly
assembly
Drain
pump
Dosing pumpAir gap pump
LEFTRIGHT
Figure 3Figure 4
Capacitor wires
pump
Recirculation
6
Page 7
1. Introduction
1.3 Specifications
Applicable StandardsISO 15883-1/2
Height (Standard height unit)850 mm
Height (Short height unit)830 mm
Width600 mm
Depth (Door closed)600 mm
Depth (Door open)1200 mm
Weight90 kg
Floor loading per support when full310 N
Required clearance:
Running Noise<78 dB( A)
Inlet water connectionsG 3/4” B
Inlet water pressure2-5 bar
Drain3/4”
Maximum water flow to drain47 L/min
Maximum water discharge temperature to drain95°C
Maximum water hardness30.3dH, 31.6 US GPG, 540 PPM
Maximum water conductivity844 µS/cm
pH range>6.8 and < 8.5
Water volume per process stage
Total water consumption per cycle with drying
Maximum heat transmitted through fascia2388 W
Water softener salt capacity1.0 kg
Equipment installation categoryII
Voltage230 – 240 VAC ± 10%
Frequency*50 Hz or 60 Hz
Rated load2.5 kW
Current*12A or 13A
Operating temperature range5°C – 40°C
Maximum relative humidity
Maximum operating altitude2000 m
Equipment pollution degree2
Maximum deviation from plane horizontal surface.2 mm
Suitable process chemicals for each stageHIP
Total amount of process chemical108 mL
Material Safety Data SheetRefer to website: http://www.scican.com
Cycle recording optionsUSB, G4 web por tal, external printer
*Refer to E lectr ical Su pply ta ble
Top and Sides
Rear
>10 m m
>30 mm
8 L ± 1 L
P3 has additional 8 ± 1L for drying
P1: 16 L ± 2 L
P2: 32 L ± 4 L
P3: 40 L ± 5 L
80% for temp up to 31°C
50% for temp up to 40°C
7
Page 8
1. Introduction
When ordering supplies, spare parts or requesting service, please ensure that
the information contained on the serial number plate is available (Model number,
serial number etc.).
The serial number plate is located at the bottom left on the rear panel of
the HYDRim M2 G4 unit. The serial number can also be found on a label located
on the left hand side of the chemical door.
Electrical Supply (Refer to model number for specific requirements)
Model Number
Ending
-D0112A230-240 V60 Hz1.50 m / 5 ft2.5 kW
-D0313A230-240 V60 Hz1.50 m / 5 ft2.5 kW
Current
(see Note 1)
VoltageFrequency
Maximum
power supply
location for
installation site
Rated
load
models)
(including short
HYDRIM M2 G4
Note 1: The HYDRim G4 is supplied with a domestic fused plug as standard. A dedicated hard
Note 2: Due to the power requirements of the HYDRim, (see rated load) especially during drying,
Note 3: Power supply outlet should be adjacent to the machine and NOT behind it.
-D0213A230-240 V50 Hz1.50 m / 5 ft2.5 kW
-D0413A230-240 V50 Hz1.50 m / 5 ft2.5 kW
-D1013A230 -240 V50 Hz1.50 m / 5 ft2.5 kW
wired supply can also be used.
it is advised that no other equipment is connected to the same supply outlet.
The cable should be routed away from the back panel and hot water inlet hose.
8
Page 9
1. Introduction
1.4 Safety information
The following symbols
appear in the margins
of this book.
The following symbols appear on the unit:
Caution: Hot Surface
and/or Hot Steam
A potential hazard
to the operator.
Safe operation
The following apply to both operators and service technicians:
• Exercise caution and seek assistance when lifting or carrying the unit.
• Cleaning solutions may irritate. Avoid contact with eyes, skin and mouth.
• Never lean on the open door. The unit may tip forward causing injury.
• Always turn the unit OFF before adding softener salt or solutions. Before performing
routine maintenance or servicing the unit, turn the unit OFF and unplug the power cord
from the power source.
A situation that
may lead to a
mechanical failure.
Caution: Risk of electrical
shock. Disconnect supply
before servicing.
Important
information
Caution: Refer to manual
for details.
• The operator should never remove the cover of the unit or insert objects through holes
or openings in the cabinetry. Doing so may damage the unit and/or pose a hazard to
the operator.
• If the unit is used in a manner other than that specified, the protection provided by
the equipment may be impaired.
Safe servicing
• The HYDRim M2 G4 Instrument Washer-Disinfector should only be installed and
serviced by a qualified contractor as it is an Installation Category 2 device.
The contractor should be experienced in installing equipment that requires electrical
hook-up as well as plumbing.
• SciCan shall not be liable for incidental, special or consequential damages caused by
any maintenance or services performed on the HYDRim M2 G4 by a third party or for
the use of equipment or parts manufactured by a third party, including lost profits, any
commercial loss, economic loss, or loss arising from personal injury.
• All local, regional, state, and national regulations regarding the servicing of this class
of device and safety requirements must be observed.
9
Page 10
1. Introduction
When the cover and panels are removed:
• Hazardous voltages are accessible. Disconnect the power cord before removing
the cover or any panels.
• Sharp metal edges are exposed. Be careful, and wear long sleeves and gloves.
Power main
• If the cover or panels are removed, a dielectric strength test (hi-pot) must be performed
on the unit once the cover or panels are reinstalled.
Ground
• If the cover or panels are removed, a protective bonding impedance test (ground
continuity) must be performed on the unit once the cover or panels are reinstalled.
Reporting
• It is vital for SciCan to learn of any problem in the field. This information will help
SciCan solve the problem quickly and improve product reliability in new units.
Biological waste
• Waste water in the unit may contain biological contaminants. Use a mechanical
means or absorbent material to siphon the contents from the sump. Wear disposable
rubber gloves. Dispose of absorbent material according to biological waste disposal
regulations.
1.5 Tools and hardware
Tools required for servicing include:
• Needle-nose pliers
• Screwdrivers Philips 1 & 2
• Screwdriver slot
• Wire cutters
• Small slot screwdriver
• 8 mm Nut driver
• Allen key 3.0 mm
• Channel Locks
• Spring clamp pliers
Electrical Safety test equipment:
• Hi-Pot tester
• Ground continuity tester
• Static strap
• Static bags
10
Page 11
1. Introduction
The unit contains the following types of hardware:
• Phillips pan head self-tapping metal screws
• Phillips pan head stainless steel machine screws
• Hex socket pan head stainless steel machine screws
• Spring clamps
• Band / Gear clamps
• Cable ties
1.6 Disconnecting the unit
To disconnect the unit, follow these steps:
1. Turn the unit off and disconnect it from the power supply.
2. Turn off the water supply.
3. Disconnect the drain and water intake hoses.
4. Remove the screw at the center of the kickplate and remove the kickplate to access
the moving strap.
5. Pull the unit out using the moving strap and carefully withdraw the hoses at the
same time.
Moving strap
Figure 5
11
Page 12
1. Introduction
1.7 Shipping instructions
The unit should be serviced on site. If it is necessary to send the unit back to the dealer,
follow these instructions:
• Run the ‘Prepare for Shipping’ cycle in the setup menu to remove most of the water
from the system before shipping the unit.
• Waste water in the unit may contain biological contaminants. Use a mechanical
means or absorbent material to siphon the contents from the sump. Wear disposable
rubber gloves. Dispose of absorbent material according to biological waste disposal
regulations.
• Disconnect and remove the cleaning pouch container and then drain the dosing reservoir.
• Screw in the leveling legs.
• Specify upright, heated, and insured shipping.
• Ensure unit is returned on a pallet with at least two banding straps securing the box
to the pallet. If original packaging is unavailable packaging can be ordered with
part # 01-113847S.
• Shipping outside of these conditions can affect warranty.
12
Page 13
1. Introduction
1.8 Installation
IMPORTANT INFORMATION
Pre-Installation
The machine must be installed and leveled correctly for the unit to function as described.
All electrical work must be carried out by a qualified electrician and in compliance with
all local and national electrical codes.
Voltage: 230-240V ± 10%
Frequency: 50 Hz or 60 Hz (refer to unit serial number plate)
Rated load: 2.5 kW
Circuit breaker: 13A (12A for N.A. unit)
• The outlet needs to be accessible after the unit is installed.
• The appliance must be correctly grounded! The manufacturer cannot be held respon-
sible for damage or injury caused by incorrect or missing grounding.
• The HYDRim unit is heavy (90 Kg). Exercise caution and obtain assistance when lifting
unit.
• If you need to extend the water inlet and drain hoses, ensure that you use commercial
grade plumbing hose. The maximum length of the drain hose is 3.3 m.
Installation instructions
Installation should only be undertaken by a manufacturer approved technician. The use of
an unapproved installer may invalidate the warranty. A separate pre-installation checklist
should have been supplied to the user by the dealer. Please review this prior to approving
installation.
If the HYDRim M2 G4 is installed in a sterilization center, the manufacturer of the
sterilization center should allow enough space at the top, back and both sides of the
unit to facilitate installation, leveling, and service access to the unit. During installation, all
consumables should have been added to the machine as appropriate. It is important to
check that this has been undertaken before starting the machine.
The HYDRim M2 should only be installed and service by a qualified SciCan contractor, as
it is an installation Category 2 device. The contractor should be experienced in installing
equipment that requires electrical hook-up as well as plumbing.
The machine must be installed and leveled (see leveling instructions on next page) correctly
for the unit to function as described. All electrical work must be carried out by a qualified
electrician and in compliance with all local and national electrical codes.
13
Page 14
1. Introduction
Leveling the HYDRIM
The unit is standing on three supports: rollers (wheels) at the back and two legs at the front.
1. Remove the kickplate. Push the HYDRim into place while lifting the strap at the front
to allow the unit to roll on the rollers.
2. Adjust the front legs as required until the HYDRim is level. Access the legs from inside the unit.
3. The rear two legs are used only if the floor is uneven or cannot provide support to the rollers.
Electrical connection
This appliance must be correctly grounded! The manufacturer cannot be held responsible
for damage or injury caused by incorrect or missing grounding. Before making any
connections check that the voltage shown on the serial number label corresponds to your
power supply. The machine is supplied as standard for connection to 230-240V 50 Hz or
60Hz single-phase power supply and is fitted with a detachable power supply cord 2.5m.
It should be connected to the main power supply according to the information below.
Voltage: 230-240V
Frequency: 50 Hz or 60 Hz (refer to unit serial number plate)
Rated load: 2.5 kW
Circuit breaker: 13A (12A N.A. unit)
Connection to the water supply
The unit must be connected to the water supply in accordance with all local and national
plumbing codes. SciCan recommends a hard plumbing installation within 1.5m/5ft. of the
unit. If additional distance is necessary, commercial grade plumbing hose must be used to
minimize leaks. Connect inlet hoses to hot and cold water taps using the hoses connected
to the unit and in accordance with the installation instructions.
Water Pressure: 2-5 bar
Water Temperature: Cold water less than 30ºC
Hot water up to 60-84ºC
Water inlet hoses (provided): 3/4" fitting, 2m length
Drainage
The unit is supplied with a 1.5m flexible drain hose with a 2cm barb inlet. The hose should
not be shortened or attached to any fittings that would cause a reduction in water flow. The
drain system is equipped with a non-return valve that prevents dirty water from flowing back
into the unit.
The drain hose should not be further than 1.5m/5ft. from a hard plumbing drain. If this is not
possible, then commercial grade plumbing hose must be used to minimize leaks.
The hose can be attached to an existing drain through the use of a 3.5cm or larger stand
pipe/ P-trap combination. Alternatively, the hose can be connected directly to the existing
drain lines, provided the fittings or adapters used do not reduce the water flow.
The drain hose should not exceed 3.3m in length, or be attached to the main drain at
a point higher than 35cm above the floor.
14
Page 15
1. Introduction
1.9 Setting the water softener
The HYDRim M2 G4 is equipped with
a built-in water softening system that must
be adjusted according to the local water
hardness. To read local water hardness,
proceed as follows:
1. The water test kit included with your
HYDRim contains three water
hardness test strips in bags. Take a
water sample from the location where
the machine will be installed.
2. Open one of the bags, remove the
test strip and dip it into the water.
3. Compare the color of the strip with
the chart on the back of the bag
Determine the water hardness according
to the chart on the water test kit envelope.
4. Power the unit on and select the
Settings key from the main menu.
5. Go to the Setup Menu, Cycle Settings,
and select “Set Regeneration”.
7. Using the up and down arrows, set the
water softener regeneration level
according to the water hardness table
in this section. If your water hardness
falls between two settings, select the
higher setting.
8. Unscrew the water softener container
lid from the bottom left of the chamber
and pour 1 litre (1 quart) of water into
the water softener container.
9. Add 1 kg (22 lbs) of water softening salt
to the water softener container, using
Water Hardness Conversion and salt regeneration levels
°dH
Very Soft
Necessary
Typically
(values from 18-143)
No Treatment
10.1 10.5 180
11.2 11.7 200
11.8 12.3 210
16.8 17.5 300
19.6 20.5 350
20.2 21.0 360
(values from 150-535)
May Require External Treatment
22.4 23.4 400
25.2 26.3 450
28.6 29.8 510
≥30.3 ≥31.6
Required
(values >535)
External Treatment
US
GPG
1 1.0 18
2 2.1 36
3 3.1 54
4 4.2 71
5 5.2 89
5.6 5.8 100
6 6.3 107
6.2 6.4 110
7 7.3 125
8 8.3 143
8.4 8.8 150
9 9.4 161
10 10.4 178
11 11.5 196
12 12.5 214
13 13.6 232
14 14.6 250*
15 15.6 268
16 16.7 286
17 17.7 303
18 18.8 321
19 19.8 339
20 20.9 357
21 21.9 375
22 22.9 393
23 24.0 411
24 25.0 428
25 26.1 446
26 27.1 464
27 28.2 482
28 29.2 500
29 30.2 518
30 31. 3 535
PPM
(mg CaCO3 / Litre)
≥540
the supplied funnel to prevent any salt from spilling into the chamber, and close by
screwing the lid tightly back into place. An improper seal can lead to corrosion.
Regen.
1
2
3
4
5
6
7
8
Additional
Water
Treatment
Required
*Please note: The water test strip is only accurate up to 250 ppm. If the reading on the test strip exceeds
250 ppm and/or if the location in which the HYDRim is installed has known water quality problems, having
a more detailed and accurate water test done by a test lab is strongly recommended.
15
Page 16
1. Introduction
IMPORTANT: The HYDRim’s water softening system reduces the water hardness by
taking out Calcium Carbonate. If the water testing results show that the water hardness
is outside the unit's range of adjustment, or if other dissolved solids in the water cause
stains or deposits on the instruments or chamber, an external water treatment system may
be required.
1.10 Setting the language
The messages displayed by your HYDRim can be presented in a number of different
languages. To change the current language, follow these steps:
1.
2. Scroll to
Language Selection
and select.
3. From the LANGUAGE screen, press to scroll through the list of languages.
When you have found the desired language, press
to the Setup menu.
to save your selection and return
1.11 Setting the country
1.
2. Scroll to
3. Using the keypad, type the name of the country and press
to save and return to the Setup menu.
Country
and select.
EN
to select. Press
1.12 Setting the time
1.
2. Scroll to
Date/Time
and select Time Setup.
3. From the TIME screen, use the keypad to set the time. Press
to return to the Setup menu.
16
EN
to save and
Page 17
1. Introduction
NOTE: If the HYDRim is connected to a network, it is important to also enter the correct
Time Zone. Enter the Time submenu, select Time Zone and scroll and select your local
time zone.
4. To change your unit to display 12-hour time format (24-hour time format is the default
setting), go to the Setup menu and use
toggle to 12. Press
to scroll to TIME 12/24, select it and
to save and return to the Setup menu.
5. To activate daylight savings time (DST), go to the Setup menu and use
to DST ON/OFF and select. Use
to save and return to the Setup menu.
to toggle DST ON or OFF and press the
to scroll
1.13 Setting the date
1.
2. Scroll to
3. From the DATE screen, use the keypad to set the date. Press
to return to the Setup menu.
Date/Time
and select Date Setup.
EN
to save and
4. To change the format in which the date appears, return to the Setup menu and use
to scroll to DATE FORMAT. Select it, and follow the prompts to have
the date displayed in the desired format. Press
the Setup menu.
to save and return to
1.14 Assigning unit identifier number
1.
2. Scroll to
Unit No
and select.
3. Using the keypad, select a maximum of 3 digits to be used as the unit’s identier
number. Press
to save and to return to the Setup menu.
EN
1.15 Resetting the drying counter
The drying counter must be reset when the HEPA filter is changed. User will be prompted
every 750 cycles to do preventative maintenance, which is triggered by the reminder to
change the HEPA filter. To reset the drying counter, follow these steps:
1.
2. Scroll to
3. Select Default 0 to reset. This will stop the reminder to the end user.
Reset Drying Counter
and select.
17
Page 18
1. Introduction
1.16 Adjusting the screensaver delay
To change the length of time before the screensaver is activated, follow these steps:
1.
2. Scroll to
Screensaver
and select.
3. Use to scroll through your time options. When you have found the amount of
time you require, press it. Press
to save and return to the Setup menu.
1.17 Adjusting the temperature display
1.
2. Scroll to
Temperature C/F
and select.
3. Use to choose between having information displayed in degrees Celsius or
Fahrenheit. Press
to save and return to the Setup menu.
1.18 Turning the button sound ON or OFF
The HYDRim is preset to beep when a button is pressed. If you would like to turn the button
sound off, follow these steps:
NOTE: Turning OFF the button sound does NOT turn off other alarms and cycle notification
beeps.
1.
2. Scroll to
Beep ON/OFF
and select.
3. Use to scroll through your ON or OFF options and select it by pressing it.
Press
to save and move back to the Setup menu.
1.19 Adjusting the button beep volume
If you would like to adjust the beep volume, follow these steps:
1.
2. Scroll to
3. Use to scroll through the volume settings. Select the one you want by
pressing it. Press
Beep Volume
to save and move back to the Setup menu.
and select.
18
Page 19
1. Introduction
1.20 Adjusting the salt regeneration
Salt regeneration should be set according to the local water hardness. See section 1.9
Setting the water softener for instructions on determining correct settings. To set salt
regeneration, follow these steps:
1.
2. Scroll to
Set Regeneration
and select.
3. Use to change the value. The default setting is 1. Press to save and return
to the Setup menu.
1.21 Adjusting the screen contrast
The touchscreen is calibrated for the lighting condition of most sterilization centers. Should
you need to adjust the contrast for your office, follow these steps:
1.
2. Scroll to
LCD Contrast
and select.
3. Use to scroll through your contrast options. When you have found the contrast
you require, press it. Press
to save and return to the Setup menu.
1.22 Changing the touchscreen display themes
The touchscreen themes (i.e. icons and background colours) can be changed to one of the
preset options. To change themes follow these steps:
1.
2. Scroll to
3. In the
As you scroll, each theme will display on the touchscreen. Press
theme and return to the Setup menu.
Theme
Change Theme
and select.
screen, use to scroll through your available options.
19
to select your
Page 20
1. Introduction
1.23 Creating a User Name
Up to four unique User Names can be created. To assign a User Name follow these steps:
1.
2. Scroll to
3. To assign a user name, select
a name (up to 12 characters) and press
User
and select.
User Name
to save.
EN
and use the alphabetic keypad to enter
1.24 Creating a User PIN
From the User PIN screen, you can assign up to four PINs. To assign a PIN, follow these
steps:
1.
2. Scroll to
User
and select.
3. To assign a user PIN, select
a number (up to 4 digits) and select
User PIN
EN
and use the numeric keypad to enter
to save and to move to the confirmation
screen.
4. If all of the information presented in the confirmation screen is correct, press OK
to be returned to the User PIN screen. To make a correction, select the User PIN you
want to change and repeat the process described above.
20
Page 21
1. Introduction
1.25 Setting up process enforced usage
When process enforced usage is activated, users are required to enter a PIN at the end of
a cycle. For process enforced usage to function, User IDs and PINs must first be assigned.
To set up User ID and PINs, refer to sections 1.23 and 1.24 on creating a user name and
PIN. To activate process enforced usage, follow these steps:
1.
2. Scroll to
Process Enforced
and select.
3. Use to toggle process enforced function ON or OFF. Press
to save your selection and return to the Setup menu.
NOTE: Any user can stop a cycle even
with process enforced usage ON.
However, the cycle data will record
that an unauthorized user
has stopped the cycle.
1.26 Connecting to a network
The HYDRim M2 G4 has a 10/100Base-T Ethernet port located at the back of the unit.
To connect your HYDRim to a network using a router, follow these steps:
1. Connect your network cable to the Ethernet port at the back of the unit. If your office
uses a router, the router should automatically assign the unit an IP address. A red X on the
network icon means the unit is not connected. A yellow check mark means the unit has
an IP address but is not connected to the Internet and cannot send emails. A green check
mark means the Internet connection is set up properly and the unit can send out emails.
NOTE: In some circumstances, where you do not have a router, for example when using
Windows Network Sharing, you may have to assign a dedicated or ‘static’ IP address.
To assign a static IP address, contact your local network administrator.
2. From the main screen, press the Network icon. The Network screen displays information
about your HYDRim’s connectivity, including its IP address.
21
Page 22
1. Introduction
3. Type the IP address displayed on the touchscreen into the browser of any web enabled
device to access your unit’s web portal. When the Network icon is active (for example when
sending email) it will turn green.
NOTE: Use QR code if connecting to a mobile device.
NOTE: Connection time will vary depending on your network speed, and making an initial
connection can take longer.
1.27 Connecting to a wireless network
The HYDRim can be configured for wireless use by connecting the Ethernet port to an
external wireless bridge / access point. SciCan currently recommends the use of the
®
D-Link
to learn more about setting up a wireless bridge.
DAP-1522 Xtreme N® Duo Wireless Bridge. Contact your network administrator
22
Page 23
2. Routine Procedures and Maintenance
2.1 Replacing the cleaning solution
To replace the cleaning solution, follow these steps:
To prime the cleaning solution dosing pump, press the water softener/detergent icon on the
main screen. In the water softener/detergent screen, press the red X next to “Detergent”.
The unit will prime the dosing system and a green check mark will appear in place of the
red X when it is ready for use.
NOTE: The system can also be primed by simply starting a cycle and selecting “Detergent
Replaced”, when prompted.
NOTE: A cycle will not start with the red X next to the “Detergent” indicator.
23
Page 24
2. Routine Procedures and Maintenance
2.2 Refilling the water softener
If the water softener system is set to any value
above 0, the message “Salt Level Low” will appear
on the display. To add water softening salts, follow
these steps:
1. Unscrew the salt container lid.
2. On first using the HYDRim M2 G4, pour
1 liter (1 quart) of water into the salt container,
or until it is full with water. NOTE: It is not
necessary to add water during subsequent
refills of the salt container.
3. Replace the salt container lid.
After the salt has been added to the unit, the softening salt indicator will initially show
that more salt is needed. The indicator will turn off when the salt solution has become
sufficiently concentrated.
Figure 6
2.3 Filter and wash arm maintenance
Cleaning the chamber’s coarse and fine filters
Inspect the coarse and fine filters in the bottom
of the chamber daily for debris and clean if
necessary. To clean the coarse filter, remove
the filter, rinse under a tap and reinstall.
To clean the fine filter, remove the coarse filter,
remove the two screws holding the fine filter and
using a small screwdriver, gently pry the fine filter
out. To replace, reinstall with the mesh side down.
Coarse filter – part # 01-113843S
Fine filter – part # 01-113844S
coarse filter
fine filter screws
Figure 7
fine filter
24
Page 25
2. Routine Procedures and Maintenance
Removing and checking the wash arms
If you see that the wash arms are
not turning easily, remove them.
Both the upper and lower arms are
thread mounted. Turn the mounting ring
counter clockwise to remove.
Rinse under a tap, clear
obstructions from outlet holes
and reinstall.
upper
wash arm
To remove the middle wash
arm, twist the mounting ring
middle
wash arm
one half turn counter clockwise
to remove.
Lower, upper spray arm
– part # 01-109790S
Middle spray arm
– part # 01-111495S
lower
wash arm
2.4 Replacing the blower filter
The HEPA filter in the blower should be changed every
750 cycles. To change this filter, remove the top and right
panels and follow these steps:
1. Remove the blower assembly. (See section 7.9
Removing and reinstalling the blower assembly)
Figure 8
2. Remove 2 top hex bolts from the side of the blower
to open the component and remove the filter.
3. Reinstall new filter in the correct orientation,
following the direction-of-flow arrows on the side
of the component and on the side of the filter.
4. Reinstall the blower assembly.
5. Reset the drying counter. (See section 1.15
Resetting dryer counter)
Dryer filter – part # 01-111780S
25
Figure 9a
Figure 9b
Page 26
2. Routine Procedures and Maintenance
2.5 Cleaning the chamber
The HYDRim M2 G4 chamber can be cleaned using the “Cleaning” program in the User
menu. This cycle is used to periodically remove hardwater deposits from the chamber walls
and racks. Pour 1 litre of vinegar into the chamber before starting the cycle. From the User
menu, select “Cleaning” and a cleaning cycle, similar to a normal wash cycle, will run. The
user will be prompted to clean the chamber every 25 cycles. Reminder frequency can be
changed to 25, 50, or 75 cycles. To do this, enter the Technician menu.
2.6 Draining the unit for service or shipping
To drain the unit prior to shipping, or before tipping it onto its back for servicing, run the
“Prepare for Shipping” cycle. Once complete, drain any water remaining in the air gap using
the silicone tube located under the centre of the unit’s kickplate.
2.7 Upgrading the firmware
NOTE: Instructional videos are available on http://updates.scican.com
NOTE: Upgrading the Interface Software can be done from a USB drive, a MicroSD card or
a web site. The easiest and fastest method is to use a USB drive.
To upgrade the firmware using a USB drive, proceed as follows:
1. Download new firmware. The firmware will be made available on updates.scican.com or
emailed from SciCan upon request. It will be packed into a zip file (e.g. SHM2MRXXX.zip
is the name of the current revision file, but the number will change with every revision)
and must be extracted to a USB drive.
2. Check that you have the following files on the USB Drive:
• rmware.ini
• Firmware (Folder)
• SHM2MRXXX_4_100_CAA29608.sci
3. With the unit powered OFF, insert the USB drive loaded with the firmware update.
4. Power ON the unit. The firmware will be updated automatically using the USB drive.
This should take approximately 6 minutes. NOTE: The USB icon on the LCD touchscreen will flash green while it is active. Do not remove the USB key while it is active.
5. When it is complete, the “Firmware.log” file on the USB drive will include the result of
the upgrade (file name, upgrade OK, or upgrade failed, and for what reason).
26
Page 27
2. Routine Procedures and Maintenance
6. Whether the upgrade is successful or unsuccessful, the “firmware.ini” file on the
USB drive will be automatically deleted.
7. To retry or upgrade another unit, insert the USB drive into the PC’s USB port.
8. Plug in the USB drive into the computer.
9. Double-click the FIRMWARE folder on the USB drive (e.g. Z:/FIRMWARE)
10. Copy the file 'firmware.ini' file into the root directory of the USB drive (eg. copy and
paste the file 'firmeware.ini' from Z:\FIRMWARE\ to Z:\).
2.8 Using the HYDRIM remote access function
Users can allow offsite technicians to remotely access the LCD touchscreens and web
portals of HYDRim M2 units connected to a network. This can be done either from within a
network or from outside a network.
From within a network:
For local network remote access, the unit must be connected to a network. See Connecting
to a network in section 1.27 of this manual for more details. From the unit’s web portal,
proceed as follows:
From the TOOLS page, click on the LOCAL CONTROL tab.
Log in using the following credentials:
Username: scican
Password: s23can173
Click on the start button to start a local connection. It will open up a page that mirrors the
HYDRim unit’s touchscreen so that it can be controlled remotely within a local network.
From outside a local network:
For remote access of a HYDRIM web portal or touchscreen from outside a local network,
proceed as follows:
1. Someone onsite with the unit or from within the network must provide access to an
outside user by generating a ‘token’ (or access code).
2. To generate a unique token using the web portal, go to the TOOLS page and click on
the REMOTE ACCESS tab.
3. To generate a unique token using the unit’s LCD touchscreen, go to the SETUP menu
and scroll to REMOTE ACCESS and follow the prompts to enable remote access.
4. The technician attempting to access the unit from outside the network will need to go to
the following URL: http://updates.scican.com and enter their registered email address,
27
Page 28
2. Routine Procedures and Maintenance
password, token and HYDRim Serial Number (optional).
5. To create a new account to enable remote access for a HYDRim, click on the CREATE
NEW ACCOUNT link, complete the form, and click on the SUBMIT FORM link. The
system will send a confirmation email to verify the account. Once confirmed, the
account will be ready to use.
6. Use the valid user name and password to enter Updates.scican.com and enter the
token when prompted. This will bring you to the HYDRim unit’s web portal page.
7. Click on SETUP. A username and password prompt will appear. Log in using the
following credentials:
• Username: scican
• Password: s23can173
8. Upon authentication, go to TOOLS and click on LOCAL CONTROL. Click on the start
button to start a connection. It will open up a page that mirrors the HYDRim unit’s
touchscreen so that it can be controlled remotely from outside its local network.
Use your mouse to click and select touchscreen elements.
2.9 Annual Service Requirements
The HYDRim M2 G4 is designed to be maintenance free; however, it is recommended that
a SciCan-approved service technician perform an annual check. Testing protocols are available for download at MySciCan via www.scican.com.
The following checks are recommended in order to maintain optimum performance of
the unit.
Annual service schedule
• Check integrity of incoming and outgoing services (power, water supply, drain)
• Check water supply in line lters and clean as appropriate
• Check general condition of machine
• Inspect and replace main chamber seal (only if required)
• Inspect and replace lower door seal (only if required)
• Check solution container connection for leaks
• Check salt level and replenish as required
• Check water hardness (test strips) and adjust salt regeneration settings if required
• Inspect and clean sump lters (check sump for debris)
• Check wash arms for blockages and remove them for cleaning if required
• Review error history
• Software upgrade (if necessary)
• Replace dryer lter and reset dryer counter (only if required)
28
Page 29
2. Routine Procedures and Maintenance
• Check individual component functionality. Go to the technician menu (enter access
code 7919) and select ‘Diagnostic Tools’ then select ‘Component Tests’. From here
you can scroll through and check the functionality of the following components:
• be adjusted. (Dosing pump validation kit 01-113909S)
• Check thermocouple calibration and adjust if required
• Reset service cycle counter
• Clean machine
• Conduct electrical safety tests
• Check that dosing pump is dispensing. To verify:
1.
2. Scroll to
Diagnostic Tools
3. From this menu, scroll to
be activated and the predefined number of pulses will be dispensed (the screen
will display a countdown counter).
code 7919
and select. Then scroll to
Dosing Pump
and select ON. The dosing pump will
Component Test
4. With the door open, check to see if the detergent is dispensing.
NOTE:
Selecting the OFF button on this screen will activate the chemical reservoir filling
pump. This pump will automatically stop when the chemical reservoir level switch is
activated.
The dosing pump is pre-calibrated and the volume cannot be adjusted.
menu.
• Check thermocouple calibration. To calibrate:
1.
2. Scroll to
select.
Diagnostic Tools
code 7919
and select. Then scroll to
29
Set Calibration
and
Page 30
2. Routine Procedures and Maintenance
3. From
Set Calibration
, press or to enable Calibration.
4. Insert external temperature reading device into sump of HYDRim and
close door.
5. Run P3 / P4 / P5 Disinfection cycle. When the temperature sensor reach-
es 90ºC use
on the external temperature reading device
IMPORTANT:
servicing of this class of device and safety requirements must be observed.
• Reset the service counter.
All local, regional, state and national regulations regarding the
1.
2. Scroll to
and select.
• Clean machine
• Conduct electrical safety tests (hi-pot and ground continuity)
Preventative Maintenance
on the touchscreen to match the values displayed
code 7919
and select. Then scroll to
Reset Service Counter
Equipment and parts requirements for annual service
• Dryer filter (HEPA) (Part number 01-111780S)
• Main chamber seal (Part number 01-113790S)
• Lower door seal (Part number 01-113789S)
• Service Manual (Part number 96-113787)
• Electrical safety test equipment
• Water hardness test strips (Part number 01-108305S)
• Calibrated independant temperature measuring device
• 100ml granulated measuring cylinder
30
Page 31
3. Diagnostics and Troubleshooting
3.1 Using the service menu
To access the service menu, select the image of the technician and enter
the service code 7919 on the keypad.
User TechnicianSetup
Wash Test
Diagnostic Tools
Error History
View IO Status
Component Tests
All Devices
AG Valve + Pump
Spare
Fan
RO Valve
Cond. Valve + Pump
Ch. Heater
Latch
Salt Valve
Dosing System
Dryer Speed
Dryer + Heater
Hot W. Valve
Cold W. Valve
LCS Sol Valve
Waste Pump
Circ Pump M1
Set Debug Screen
Set Calibration
Cycle Settings
Set Regeneration
Chemical Setup
Cycle Selection
Prewash
Wash
Rinse
Set Drying Time
RO Selection
Set RO water level
Rinse continuous
Factory Default
Repeater Mode
Production Tools
Prod Test Cycle 1
Set Regeneration Time
SW Upgrade
Network Setup
Automatic IP (DHCP)
Renew IP
IP address
Subnet Mask
Router
DNS
Dealer ID
Preventative Maintenance
Set Cleaning Warning
31
Page 32
3. Diagnostics and Troubleshooting
3.2 Using the setup menu
To access functions and settings on the setup menu, proceed as follows:
User TechnicianSetup
Language Selection
Country
Date/Time
Unit No.
Network Setup
Reset Drying Counter
Screen Saver
Temperature C/F
Beep Volume
Set Printer
Baud Rate
Set EOL CR/LF
Set Regeneration
LCD Contrast
Theme
Remote Screen
Instruction Delay
Remote Access
Prepare for Shipping
Set Button Beep
Instructions
32
Page 33
3. Diagnostics and Troubleshooting
3.3 Using the user menu
To access functions and settings on the user menu, proceed as follows:
User TechnicianSetup
Cleaning
Set Drying Time
Set Wash Time
Cycle Count
Process Enforced
User
33
Page 34
3. Diagnostics and Troubleshooting
3.4 Using software tools for diagnostics
Within the service menu, there are two useful tools for troubleshooting: Debug screen and
I/O status screen.
Debug Screen
The Debug screen is used when running a cycle to view the I/O status of components.
Active components are colored while inactive components remain white.
To access the debug screen, select Diagnostic Tools from the service menu and select
Set Debug Screen, then go to the main menu and select a cycle. The LCD screen will
display the following:
Power supply
fan
Air gap
Chamber
temperature
Air gap
pump
Fill valve
Door statusChamber pressure
switch
Blower
Dosing pump
Chemical level
Salt level
Regeneration
valve
Condenser valve
Hot and Cold
valves
R/O valve
Evacuation pump
I/O board ambient
temperature
Validation
temperature
Figure 10
34
Page 35
3. Diagnostics and Troubleshooting
I/O status screen
The I/O status screen is used when testing components and wires for functionality
without the cycle running.
Chamber temperature
less than a set point
after a timeout,
or a temperature
increase rate of 1ºC
per 2 minutes was
not achieved during
“Circulation and
heating” phase
Timeout on filling up
the chamber
Temperature reading
outside acceptable
range for primary or
secondary sensor
evacuation from
the chamber
This is caused by a water
heater malfunction:
•Check water heater
wire harness for loose
contacts.
•Check for open thermal
cut-off switch due to
high temperature.
•Check that the
heater element is not
interrupted.
•Check I/O PCB water
heater relay output.
•Water supply issue
•Water valves failure
•Air gap water pump
failure
•Air gap valve failure
•Air gap Full/Empty level
switches failure
•Chamber water level
pressure switch
malfunction
•Overflow switch
malfunction triggering
evacuation pump.
This is caused by a
temperature sensor
malfunction:
•Check temperature
sensor wires for loose
contacts.
•Replace sensor with a
good one and verify if
the CF persists.
•Replace I/O PCB.
•Blocked drain tube
•Chamber water level
switch malfunction
•Chamber water
evacuation pump failure
•Drain pump priming
connection hole in the
sump blocked
36
Page 37
3. Diagnostics and Troubleshooting
Cycle FaultEffectProblemPossible Causes
CF 5 Disinfection
Failure
CF 7Cycle aborted or
CF 9 Program
Timeout
CF 13 Temperature
Reading Failure
CF 14 Water Too HotCycle aborted or
CF 15 Water
Reservoir Overflow
CF 16 Ambient
Temperature Error
Cycle interruptedTemperature dropped
below target
temperature during
the disinfection phase
Power failure
interrupted
System FailureDefective PCB and/or
software failure
Cycle aborted or
interrupted
interrupted
Cycle aborted or
interrupted
Cycle interruptedOperating temperature
Temperature sensor
readings are out of
limits.
Water temperature
higher than the target
Operating temperature
for one or both logic
boards is too high
for one or both logic
boards is too high
The chamber or the
validation temperature
sensor is malfunctioning
or the water heater cannot
maintain the disinfection
temperature:
•Check calibration
•Check temperature
sensors
•Check water heater
•Check I/O board
(hardware failure)
•Restore power
to the unit.
•Restart program.
•Restart program.
•If message persists,
replace color LCD
controller.
The temperature sensor is
out of range:
•Check temperature
sensor wires for loose
contacts.
•Run a cycle to
monitor that the water
temperature is below
96°C.
•Replace sensor and
verify if the CF persists.
•Replace I/O board.
•Check water
connections.
The water reservoir overflow
switch was triggered:
•Restart program,
if message persists,
check the water
reservoir full switch.
The room or enclosure is too
warm and not allowing the
unit to adequately cool:
•Check that fans
are working.
37
Page 38
3. Diagnostics and Troubleshooting
Cycle FaultEffectProblemPossible Causes
CF 19 Defective
Pressure Sensor
CF 20 Condenser
System Failure
CF 21 Dosing System
Error
CF 25 Vref ErrorCycle cannot start or
CF 27 Memory ErrorHardware failureColor LCD control
Cycle aborted or
interrupted
Cycle aborted or
interrupted
Dosing System failure
Cycle interrupted
cycle interrupted
Pressure sensor failure This is caused by a stuck
Drying abortedCheck that the water hoses
Dosing system failed to
dispense the preset amount
in a predefined time (timeout is 3.5s/pulse). Dosing
reservoir level switch does
not change from Full ON to
OFF by the end of dosing
(no chemical dispensed)
Vref and VCC drift, post
CF 25 if VCC and Vref
are more than 3% apart
(power supply error)
board failure
pressure switch:
•Check pressure sensor
(stuck closed)
•Check wiring
•Check I/O board
(hardware failure).
are not kinked:
•Restart program.
•If message persists, this
may be because the
unit was not able to cool
down after disinfection
phase.
•Verify cold water
connection.
Dosing pump or switch
error:
•Verify bellows dosing
pump
•Verify bellows dosing
pump switch
The power supply 5V output
voltage is fluctuating:
•Check power supply 5V
output.
•Replace I/O board.
The internal memory of
the Color LCD Controller
is malfunctioning:
•Replace Color LCD
controller board.
38
Page 39
3. Diagnostics and Troubleshooting
Cycle FaultEffectProblemPossible Causes
CF 28 No Water
Pressure
Touchscreen is blank/
white
USB storage device
does not contain
the last print out
Unit is not sending
emails
Not receiving emails
from the unit
Cycle interruptedNot enough water
during cycle
•Check that the water
hoses are not kinked.
Check that water shut-off
valves are open. Check
for leaks. If message
persists, this could be
caused by stuck pressure
switch or not enough
water during wash phase.
•Check pressure sensor
(stuck open).
•Check wiring.
•Check I/O board
(hardware failure).
Check power source
Re-insert the USB storage
device and wait for the
data to copy over again. If
problem persists, back up all
the information on the USB
device and reformat it.
NOTE: the web portal allows
access to all of the unit’s
cycle information.
Check email settings by
using the TEST button on
the unit’s web portal.
From the SETUP web page,
select the TOOLS tab.
Click on TEST to check
your router, unit, and Internet
connections. If all settings
appear to be OK, go to
the unit’s touchscreen
and renew the IP address by
following these steps:
1. Scroll through the setup
menu to NETWORK
SETUP and select.
2. Select RENEW IP.
Check user’s spam lter.
Be certain the unit has been
identified as an accepted
email source.
39
Page 40
4. Removing and Replacing Panels
WARNINGS AND
PRECAUTIONS
If you have questions about the unit you are repairing, please do not hesitate to
contact your local SciCan representative for information. Also, the HYDRIM is
heavy. Exercise caution and seek assistance when lifting or carrying units.
EXERCISE CAUTION
• Hazardous voltages are accessible when the cover is removed.
• Disconnect the power cord before servicing the power mains portion of the
controller board and associated devices.
• Removing the panels will expose some sharp metal edges. Be careful and wear
long sleeves and gloves.
PERFORM TESTS
• If panels are removed, a dielectric strength test (Hi-Pot) AND a protective
bonding impedance test (ground continuity) must be performed on the HYDRIM
when the work is completed and after the cover has been returned to the unit.
• A dielectric strength test (hi-pot) must be performed on the unit if parts
associated with the power main are serviced or replaced.
• A protective bonding impedance test (ground continuity) must be performed
on the unit if components of the protective earthing system are changed or if
connections are broken and remade.
PROTECT THE UNIT
• The HYDRIM contains electronic circuitry that is static sensitive. Always wear
a static strap when working with or near printed wiring boards. In addition, use
static footstraps, grounding mats and grounded work surfaces when servicing
microprocessor devices. Transport boards and devices in static protected bags.
• In order to ensure adherence to the applicable safety agency approvals,
state, provincial, regional and national laws, replace components with SciCan
approved parts only.
40
Page 41
4. Removing and Replacing Panels
4.1 Removing and reinstalling the top panel
1. Turn the unit off, disconnect the power and open the door
to remove 2 screws under the top panel’s front edge.
(Figure 11a). NOTE: For units with the low profile top cover,
you must also remove the two screws at the back.
2. Pull the top panel to the front and shift it left and right
to release it from the tabs at the front. (Figure 11b)
NOTE: the top panel is attached to a ground wire
at the back right. Tip the panel up and disconnect
the ground screw to release.
screwscrew
Figure 11a
To reinstall:
1. Reattach the ground wire, place the panel on top of the unit, slightly
forward and re-engage the tabs under the front edge.
2. With the front tabs engaged, push it to the back to bring it into position.
3. Replace the 2 screws under the front edge. NOTE: For units with the low
profile top cover, replace the two screws at the back.
Cover top – part # 01-111469S
Cover top, low profile – part # 01-113832S
tab
Figure 11b
4.2 Removing and reinstalling the side and rear panels
1. Turn the unit off, disconnect the power and remove the top panel.
2. With the top panel removed, the left and right panels can be removed by pulling up to disengage them from
the tabs at the bottom, then tip them back and disconnect the ground wire to remove.
3. For the rear panel, remove the two screws at the bottom, pull it up, disconnect the ground wire and remove.
To reinstall these panels, connect the ground wires and slide them down into position, engaging the tabs at the
bottom of each panel into the slots in the chassis.
Cover left side - part # 01-111466S
Cover right side – part # 01-111468S
Cover rear – part # 01-111467S
41
Page 42
5. Front Components
WARNINGS AND
PRECAUTIONS
If you have questions about the unit you are repairing, please do not hesitate to
contact your local SciCan representative for information. Also, the HYDRIM is
heavy. Exercise caution and seek assistance when lifting or carrying units.
EXERCISE CAUTION
• Hazardous voltages are accessible when the cover is removed.
• Disconnect the power cord before servicing the power mains portion of the
controller board and associated devices.
• Removing the panels will expose some sharp metal edges. Be careful and wear
long sleeves and gloves.
PERFORM TESTS
• If panels are removed, a dielectric strength test (Hi-Pot) AND a protective
bonding impedance test (ground continuity) must be performed on the HYDRIM
when the work is completed and after the cover has been returned to the unit.
• A dielectric strength test (hi-pot) must be performed on the unit if parts
associated with the power main are serviced or replaced.
• A protective bonding impedance test (ground continuity) must be performed
on the unit if components of the protective earthing system are changed or if
connections are broken and remade.
PROTECT THE UNIT
• The HYDRIM contains electronic circuitry that is static sensitive. Always wear
a static strap when working with or near printed wiring boards. In addition, use
static footstraps, grounding mats and grounded work surfaces when servicing
microprocessor devices. Transport boards and devices in static protected bags.
• In order to ensure adherence to the applicable safety agency approvals,
state, provincial, regional and national laws, replace components with SciCan
approved parts only.
42
Page 43
5. Front Components
5.1 Removing and reinstalling the kickplate
1. Remove the screw at the center of the kickplate. (Figure 12a)
2. Insert a flat-blade screwdriver at the top to pry it back.
3. Disconnect the ground wire at right and remove.
To reinstall, connect the ground wire, tip it back into position and fasten
the center screw.
Kickplate front – part # 01-111476S.
ground wire
5.2 Opening the door with the manual door release
To open the door in case of a power
failure, turn the unit off and remove the
kickplate to access the door manual
release pull-ring in the blue tube located
on the left. (Figure 13)
NOTE: Exercise caution when opening
the door in a power failure situation.
There may be fluid remaining in the unit
and instruments may be hot.
Instruments that have not completed
the cycle should not be used and
should be reprocessed.
screw
Figure 12a
Figure 12b
ring
Figure 13
43
Page 44
5. Front Components
5.3 Removing and reinstalling the sump temperature sensor
1. Turn the unit off, disconnect the power and remove
the kickplate, top and right panels.
2. Remove the coarse and fine filters inside the chamber.
3. Remove the mounting nut from the sensor inside the sump.
4. From the outside, pull the sensor out, follow the wire with
red insulation jacket to the I/O board and disconnect the
connector at the T1 and T2 positions.
To reinstall:
1. Starting from below the sump, fish the sensor wire
to the I/O board and reattach it to the wiring harness,
ensuring it is clear of the door springs.
sensor
2. Connect the sensor to I/O board at
the T1 (red wires) and T2 (blue wires) positions.
3. Using a wrench, fasten the sensor to the sump
by tightening the nut from the inside.
4. Reinstall the panels.
Figure 14a
Dual temperature sensor – part # 01-113270S
sensor nut
Figure 14b
Figure 14c
sensor wire
44
Page 45
5. Front Components
5.4 Removing and reinstalling the chemical dosing valve
1. Disconnect the chemical
pouch, run a “Prepare for
shipping cycle” to drain
the system, turn the unit
off, and disconnect the
power cord.
2. Remove the kickplate,
top and right panels.
3. Empty the reservoir
using the dosing
reservoir draining tube.
(Figure 15a)
4. Disconnect T-connection
tubing from reservoir A
and the elbow
connection tubing
from the chemical dosing
valve. (Figure 15a)
5. Remove 2 screws and
pull out bracket. (Figure 15b)
air gap
draining tube
NOTE: The dosing reservoir
has a vent tube attached.
6. Disconnect the tubing between the dosing valve
and reservoir.
7. Remove the dosing valve from bracket (2 screws).
dosing valve
chemical
draining tube
elbow connection
T-connection
reservoir A
reservoir B
Figure 15a
To reinstall:
1. Fasten the dosing valve to bracket.
2. Connect the outlet tubing to the dosing valve.
3. Connect the wiring (wires 20, 27/28).
4. Connect the inlet tubing (ensure it is clear of door spring).
5. Connect tubing to reservoir.
6. Fasten the bracket screws and reinstall the kickplate
and panels.
Dosing valve – part # 01-113860S
45
screws
Figure 15b
Page 46
5. Front Components
5.5 Removing and reinstalling the chemical reservoirs
There are two dosing reservoirs, one with a plug (reservoir A)
and the other with a switch (reservoir B). A must be removed
before B.
To remove reservoir A:
1. Disconnect the chemical pouch, run a “Prepare for shipping
cycle” to drain the system, turn the unit off, and disconnect
the power cord.
2. Remove the kickplate, top and right panel.
3. Empty the reservoir using the dosing reservoir draining tube.
(Figure 16a)
4. Disconnect T-connection tubing from reservoir A and
the elbow connection tubing from the chemical
dosing valve. (Figure 16a)
5. Remove 2 screws and pull out the bracket.
(Figure 16b) NOTE: The dosing reservoir has
a vent tube attached.
air gap
draining tube
6. Disconnect tubing between the dosing valve
and reservoir.
7. Remove the reservoir from bracket (2 screws).
dosing valve
chemical
draining tube
reservoir B
reservoir A
elbow connection
Figure 16a
T-connection
To remove reservoir B:
1. Remove reservoir A. (See instructions above)
2. Disconnect the bracket for reservoir B by removing
the 2 screws from the unit’s right side. (Figure 16c)
3. Disconnect the tubing from the reservoir to the pump
and disconnect the switch wiring.
4. Pull the reservoir out and detach it from the bracket
by removing the 2 screws.
To reinstall:
1. Fasten reservoir B to the bracket and connect the tubing
and wiring (wires 55 and 56) then fasten to chassis.
2. Fasten reservoir A to the bracket. Connect the tubing
and fasten the reservoir to the chassis.
3. Reinstall the kickplate and panels
Dosing reservoir with plug (Reservoir A) – part # 01-113859S
Dosing reservoir with switch (Reservoir B) – part # 01-113858S
screws
Figure 16b
46
Figure 16c
Page 47
6. Door Components
WARNINGS AND
PRECAUTIONS
If you have questions about the unit you are repairing, please do not hesitate to
contact your local SciCan representative for information. Also, the HYDRIM is
heavy. Exercise caution and seek assistance when lifting or carrying units.
EXERCISE CAUTION
• Hazardous voltages are accessible when the cover is removed.
• Disconnect the power cord before servicing the power mains portion of the
controller board and associated devices.
• Removing the panels will expose some sharp metal edges. Be careful and wear
long sleeves and gloves.
PERFORM TESTS
• If panels are removed, a dielectric strength test (Hi-Pot) AND a protective
bonding impedance test (ground continuity) must be performed on the HYDRIM
when the work is completed and after the cover has been returned to the unit.
• A dielectric strength test (hi-pot) must be performed on the unit if parts
associated with the power main are serviced or replaced.
• A protective bonding impedance test (ground continuity) must be performed
on the unit if components of the protective earthing system are changed or if
connections are broken and remade.
PROTECT THE UNIT
• The HYDRIM contains electronic circuitry that is static sensitive. Always wear
a static strap when working with or near printed wiring boards. In addition, use
static footstraps, grounding mats and grounded work surfaces when servicing
microprocessor devices. Transport boards and devices in static protected bags.
• In order to ensure adherence to the applicable safety agency approvals,
state, provincial, regional and national laws, replace components with SciCan
approved parts only.
47
Page 48
6. Door Components
6.1 Removing and reinstalling the chamber seal
1. Before pulling the chamber seal, note how the bottom left and
right edges touch the bottom of the chamber.
2. Pull the seal out from the seal recess.
To reinstall:
1. Place the bottom left
and right ends of the
new seal into position,
ensuring that the ends
touch the chamber
bottom.
2. Tuck the corners into
the seal recess and
push the rest of the
seal into place,
ensuring that it
is seated evenly
throughout.
Main chamber seal – part # 01-113790S
Figure 17a
Figure 17b
Figure 17c
48
Page 49
6. Door Components
6.2 Removing and reinstalling the lower door seal
1. Turn the unit off.
2. Remove the 4 screws on the plate at the base of
the door on the inside. (Figure 18a)
3. Remove the lower door seal.
To reinstall,
1. Push the door seal on
tightly, ensuring that it
is seated properly
and evenly.
2. Re-fasten the plate to
the door.
screws
Figure 18a
Figure 18c
Figure 18b
Lower door seal – part # 01-113789S
49
Page 50
6. Door Components
6.3 Removing and reinstalling the door
1. Turn the unit off.
2. Remove the 4 screws on each side. (Figure 19a)
3. Pull the door off. CAUTION: The right hinge is spring-loaded
and will snap up.
Figure 19a
To reinstall:
1. Slide the door back
onto the two hinges,
tipping it onto the left
hinge to start and then
the right hinge. Push the
door until it connects
with the hinge.
2. Using a flat-blade
screwdriver, gently pry
the hinges out to allow
the door to slide into
position. (Figure 19c)
3. Reinstall the screws.
Door wash chamber – part # 01-113836S
Hinge left – part # 01-113840S
Hinge right – part # 01-113841S
screws
Figure 19b
spring-loaded
right hinge
Figure 19c
50
Page 51
6. Door Components
6.4 Removing and reinstalling the lower door D-seal
1. Remove the door (See 6.3 Removing and reinstalling
the door).
2. Remove the D-seal and remove any adhesive residue.
To reinstall, press the D-seal into place, ensuring it is seated
evenly, and reattach door. NOTE: D-seal should protrude from
the door frame on both sides by 1-2 mm (0.4-0.8 inches).
D-strip door seal - part # 01-113654S
D-seal
Figure 20a
6.5 Removing and reinstalling the door latch
1. Turn the unit off, disconnect the power and remove the top
panel.
2. Disconnect the door release cord and wiring from the latch
assembly. (Figure 21)
3. Remove the screws from each side and remove the latch
assembly. (Figure 21)
To reinstall
1. Insert the latch assembly into position.
2. Hold the latch tight to the front of the unit and reattach
the screws, wiring and door release cord.
3. Run a cycle to check for leaks. To adjust the fit, slide
the latch back from the door to tighten the door seal.
Door latch assembly – part # 01-111783S
Door latch cord assembly – part # 01-112080S
D-seal
door release
cord
Figure 20b
wire connection
Figure 21
screws
51
Page 52
6. Door Components
6.6 Removing and reinstalling the door springs
1. Turn the unit off and disconnect the
power.
2. Remove the top and right panels.
3. Ensure the door is closed.
4. Access the door springs from the right
side to unhook. CAUTION: Door is heavy.
To protect the technician and the unit,
ensure it is correctly closed and will not
accidentally fall open during this
procedure.
To reinstall, reverse removal instructions.
Dual spring – part # 01-112730S
springs
Figure 22
52
Page 53
7. Right Side Components
WARNINGS AND
PRECAUTIONS
If you have questions about the unit you are repairing, please do not hesitate to
contact your local SciCan representative for information. Also, the HYDRIM is
heavy. Exercise caution and seek assistance when lifting or carrying units.
EXERCISE CAUTION
• Hazardous voltages are accessible when the cover is removed.
• Disconnect the power cord before servicing the power mains portion of the
controller board and associated devices.
• Removing the panels will expose some sharp metal edges. Be careful and wear
long sleeves and gloves.
PERFORM TESTS
• If panels are removed, a dielectric strength test (Hi-Pot) AND a protective
bonding impedance test (ground continuity) must be performed on the HYDRIM
when the work is completed and after the cover has been returned to the unit.
• A dielectric strength test (hi-pot) must be performed on the unit if parts
associated with the power main are serviced or replaced.
• A protective bonding impedance test (ground continuity) must be performed
on the unit if components of the protective earthing system are changed or if
connections are broken and remade.
PROTECT THE UNIT
• The HYDRIM contains electronic circuitry that is static sensitive. Always wear
a static strap when working with or near printed wiring boards. In addition, use
static footstraps, grounding mats and grounded work surfaces when servicing
microprocessor devices. Transport boards and devices in static protected bags.
• In order to ensure adherence to the applicable safety agency approvals,
state, provincial, regional and national laws, replace components with SciCan
approved parts only.
53
Page 54
7. Right Side Components
7.1 Removing and reinstalling the LCD touchscreen and LCD controller
NOTE: When installing a new LCD controller board,
you must manually assign the serial number and model
number. For detailed instructions on how to do this,
please contact SciCan technical service.
To remove the LCD controller board:
1. Turn the unit off and disconnect the power.
2. Remove the screw above the touchscreen and tip the
service panel forward to access the LCD controller.
(Figure 23a)
3. Remove all wire connections from the LCD controller
to the I/O board and cut cable ties affixing the wiring
harness to the LCD.
4. Remove the 4 retaining nuts on each corner of the
board. CAUTION: Lift the board gently – it is attached
to the LCD by a ribbon cable. (Figure 23b)
5. Flip the board over to expose the ribbon cable latch
fastener. Using your fingernail, gently flip up the latch
to release the ribbon cable, and separate
the LCD controller board from the LCD touchscreen.
CAUTION: Do not use a screwdriver for this.
Excess force can break the ribbon cable latch. (Figure 23c)
retaining
nuts
Figure 23a
ribbon cable latchribbon cable
Figure 23cFigure 23d
bracket retaining
nuts
Figure 23b
LCD
touchscreen
bracket
54
Page 55
7. Right Side Components
To remove the LCD touchscreen:
1. Remove the 2 retaining nuts to remove LCD
bracket. (Figure 23d)
2. Separate the LCD from the bracket. The LCD
may be glued or taped to the bracket. Push
evenly with two thumbs to remove LCD from the
bracket. (Figure 23e)
To reinstall the LCD touchscreen:
1. Remove any residual adhesive from the LCD
bracket and, using the adhesive strips supplied
with the new LCD kit, set the new LCD flush with
left and bottom tabs as shown in Figure 23f.
2. Reinstall LCD bracket onto the 2 threaded posts
and fasten with retaining nuts.
NOTE: Remember to remove protective film from LCD.
To reinstall the LCD controller board:
1. Reconnect LCD ribbon cable to controller board.
(CAUTION: ensure ribbon cable is properly seated
into ribbon cable latch).
2. Flip the board onto the four threaded posts and fasten
with retaining nuts.
3. Reconnect the wire connectors.
NOTE: When installing a new LCD controller board, you must
manually assign the serial number and model number.
For detailed instructions on how to do this, please contact
SciCan technical service.
1. Turn the unit off, disconnect from power and remove the top and right panel.
2. Disconnect all connectors on I/O board.
3. Remove the 5 screws fastening I/O board.
To reinstall:
1. Ensure the unit is disconnected from power source.
2. With I/O board in place, reinstall the 5 screws.
3. Reconnect all connectors. NOTE: all connectors have unique ports. See Figure 24 and the electrical
schematic in Appendix A for reference.
IO PCB – part # 01-113310S
Figure 24
56
Page 57
7. Right Side Components
7.3 Removing and reinstalling the power supply
1. Turn off the unit and disconnect the power cord.
2. Remove the wire contacts from the power supply.
3. Loosen the screw at the top left of the power supply bracket
and remove the screw at the bottom right.
4. Slide power supply down to remove.
To reinstall:
1. Slide the bracket back into position, tighten screws.
2. Reattach contacts according to Figure 25b.
Power supply 5V/24V – part # 01-113266S
screws
Figure 25a
L N
726/27
GRD
40
COM
72
+V2
71
+V1
74
Figure 25b
COM
73
57
Page 58
7. Right Side Components
7.4 Removing and reinstalling the dosing pump
1. Disconnect the chemical pouch, run a “Prepare for shipping
cycle” to drain the system, turn the unit off, and disconnect
the power cord.
2. Remove the kickplate and drain the detergent reservoirs
using the drain tube.
3. Remove the top and right panel.
4. Disconnect the wires from the microswitch.
5. Disconnect the pump wires.
6. Disconnect the outlet tube at the chamber connection.
7. Disconnect the inlet tube.
8. To remove the pump, remove the bracket screw on the
outside edge of the pump and loosen the bracket screw on
the chamber side.
9. Slide the pump to the outside edge and pull it out with
the outlet tube still attached.
dosing pump
Figure 26a
To reinstall:
1. Install the pump.
2. Connect the outlet and inlet tubes.
3. Connect the pump wires: yellow wire to 21 and white wire
to 28/29.
4. Connect the microswitch wires: black wire to 53 and white
wire to 54.
5. Reinstall the panels.
Dosing pump bellows – part # 01-113837S
Figure 26b
Figure 26c
58
Page 59
7. Right Side Components
7.5 Removing and reinstalling the chamber level/overflow switch
1. Turn the unit off, disconnect the power and remove the top
and right panels.
2. Remove the wiring from the switch.
3. Disconnect the tubing.
4. Remove the screw fastening the switch bracket to the unit
and remove the switch and bracket.
To reinstall:
1. Before reattaching the pressure switch’s male end to the tubing, use
a syringe to pump a small amount of air into the tube. This will purge
any fluid from the tube.
2. Reattach the bracket and switch to the unit.
3. Reattach the tubing.
4. Reconnect the wiring according to Figure 27b.
Chamber level switch – part # 01-111408S
Figure 27a
18/19
To pos. 24
16
To pos. 22
8
To pos. 21
59
51
To pos. 11
52
Figure 27b
To pos. 14
Page 60
7. Right Side Components
7.6 Removing and reinstalling the air gap pump
1. Turn the unit off and disconnect the power. Remove the top
cover, right and rear panels. CAUTION: Water will remain in
the air gap. Remove kickplate and use the air gap draining
tube to drain. (Figure 28b)
2. Disconnect the wires.
3. Cut the cable ties fastening the air gap pump to the bracket.
4. Pull the pump to the left to separate it from the outlet
adaptor.
5. Disconnect the pump from the inlet tube, and remove pump.
To reinstall, reconnect the inlet tube, reconnect the pump
to the outlet adaptor, fasten the pump to the bracket
with cable ties and reconnect the wires (red wire to red 69;
black wire to black 70).
Air gap pump – part # 01-113283S
Figure 28a
air gap
draining tube
Figure 28b
60
air gap pump
outlet adaptor
Figure 28c
Page 61
7. Right Side Components
7.7 Removing and reinstalling the air gap valves
1. Turn the unit off and disconnect the power.
Remove the top cover, right and rear panels.
CAUTION: Water will remain in the air gap.
Remove kickplate and use the air gap drain
tube to drain.
2. Disconnect the air gap pump wires and cut
the air gap pump cable ties to separate it
from the bracket. Leaving the inlet attached,
set it aside.
3. Disconnect the wires from the air gap valves.
4. Remove the 2 screws on the air gap pump
bracket (located unit right).
5. Remove the 2 screws on the valve bracket
(located unit rear).
6. Pull the air gap valve assembly out from
the back of the unit.
7. Disconnect tubing and the air gap breather tube.
(Figure 29a, 29c)
8. Unscrew the outlet adaptor from air gap valve assembly.
9. To remove the valve from the bracket, remove
the 2 fastening screws.
air gap pump
breather tube
outlet adaptor
screws
air gap valvesair gap
Figure 29a
To reinstall:
1. Fasten the valves to the bracket using the 2 screws.
2. Screw the adaptor to the air gap valve assembly.
Be sure it is tight and that the air gap breather tube
connection is rotated in the top position.
3. Connect the tubing and put the air gap valve assembly
into position from the back of the unit.
4. Replace the bracket screws at the side and back.
5. Connect the breather tube to the adaptor.
6. Reconnect the wires to the valves.
7. Put the pump back into position, connecting it to
the adaptor and fasten it with cable ties.
8. Reconnect the pump wires.
9. Reinstall the panels.
Valve 1 in, 2 out – part # 01-113331S
air gap draining tube
tubing
screws
Figure 29b
Figure 29c
61
Page 62
7. Right Side Components
7.8 Removing and reinstalling the drain pump and exhaust assembly
1. Turn the unit off and
disconnect the power.
Remove the top cover, right
and rear panels. Remove
kickplate and use the air gap
drain tube to drain.
CAUTION: Water will remain
in the drain assembly.
Suction out any remaining
water from the drain exhaust.
2. Remove the air gap pump (See
section 7.6) and air gap valve
assembly (See section 7.7).
3. Remove the exhaust
assembly bracket (2 nuts
at back of unit). (Figure 30c)
4. Loosen the hose clamp from
pump outlet (Figure 30d)
5. Remove the exhaust assembly.
6. Loosen the hose clamp from inlet of pump and
remove the 2 screws holding pump to bracket.
7. Disconnect the wires and remove the pump
draining tube
To reinstall the drain pump:
1. Put the pump into position and tighten the hose clamp connecting the inlet tube.
2. Connect the pump wiring.
3. Fasten the pump (2 screws) to the pump bracket.
4. Reattach the exhaust assembly and tighten the hose clamp on the outlet.
5. Reattach the exhaust assembly bracket to the unit rear (2 nuts).
6. Reinstall the air gap valve assembly, and air gap pump.
7. Reinstall the panels
Drain pump – part # 01-111412S
exhaust bracket
screwsinletoutletscrews
Figure 30a
air gap
exhaust bracket
exhaust hose
exhaust
assembly
air gap valves
air gap pump
Figure 30b
Figure 30c
62
Figure 30d
Page 63
7. Right Side Components
7.9 Removing and reinstalling the blower assembly
blower assembly
wire
connector
1. Turn the unit off, disconnect the power and
remove the top cover and right panel.
2. Loosen the hose clamp between the blower
hose to the blower assembly.
3. Disconnect the wiring from the wire connector.
4. Remove the 2 screws on the blower assembly
bracket to release the blower assembly and
remove.
screws
To reinstall:
1. Attach the blower assembly to the unit (2 screws)
2. Connect the blower hose to the blower assembly and tighten hose clamp.
3. Re-connect the wire connector.
4. Reinstall the panels.
Blower complete assembly – part # 01-111779S
Brushes motor blower – part # 01-113464S
Blower motor – part # 01-113535S
Dryer filter – part # 01-111780S
Figure 31
63
Page 64
7. Right Side Components
7.10 Removing and reinstalling the power switch
1. Turn off the unit and disconnect the power.
2. Remove the screw above the LCD touchscreen.
3. Open the service panel.
4. Disconnect the wires from the power switch and press
down the locking tabs to remove, pushing the switch through the panel.
To reinstall, push the switch into position and reconnect wiring according to
Figure 32b. Be sure to pull out the locking tabs at the top left and bottom
right as shown in Figure 32a.
Rocker switch – part # 01-112024S
4
3
Pull tabs out
Figure 32a
5
Figure 32b
24/25
7.11 Removing and reinstalling the USB port
1. Turn unit off and disconnect the power.
2. Remove the screw above the touchscreen to tip open
the service panel.
3. Remove the retaining nuts from the threaded posts,
remove the bracket and disconnect the USB port from
the LCD touchscreen controller board.
To reinstall, reverse removal instructions.
Cable USB – part # 01-112398S
64
Figure 33
USB portretaining nuts
Page 65
7. Right Side Components
7.12 Removing and reinstalling the power supply cooling fan
1. Turn unit off and disconnect the power.
2. Remove the top and right panels.
3. Disconnect fan wires from I/O board.
4. Fan is fixed to the bracket with 4 plastic
fasteners. Push up from below to remove
the fasteners and take out the fan.
To reinstall, reverse instructions. Ensure wires
are correctly connected to the I/O board.
(see Figure 34b)
Fan 24V – part#01-113855S
fasteners
Figure 34a
65
Figure 34b
fan connections
Page 66
8. Left Side Components
WARNINGS AND
PRECAUTIONS
If you have questions about the unit you are repairing, please do not hesitate to
contact your local SciCan representative for information. Also, the HYDRIM is
heavy. Exercise caution and seek assistance when lifting or carrying units.
EXERCISE CAUTION
• Hazardous voltages are accessible when the cover is removed.
• Disconnect the power cord before servicing the power mains portion of the
controller board and associated devices.
• Removing the panels will expose some sharp metal edges. Be careful and wear
long sleeves and gloves.
PERFORM TESTS
• If panels are removed, a dielectric strength test (Hi-Pot) AND a protective
bonding impedance test (ground continuity) must be performed on the HYDRIM
when the work is completed and after the cover has been returned to the unit.
• A dielectric strength test (hi-pot) must be performed on the unit if parts
associated with the power main are serviced or replaced.
• A protective bonding impedance test (ground continuity) must be performed
on the unit if components of the protective earthing system are changed or if
connections are broken and remade.
PROTECT THE UNIT
• The HYDRIM contains electronic circuitry that is static sensitive. Always wear
a static strap when working with or near printed wiring boards. In addition, use
static footstraps, grounding mats and grounded work surfaces when servicing
microprocessor devices. Transport boards and devices in static protected bags.
• In order to ensure adherence to the applicable safety agency approvals,
state, provincial, regional and national laws, replace components with SciCan
approved parts only.
66
Page 67
8. Left Side Components
8.1 Removing and reinstalling the sump water heater
1. Turn the unit off and disconnect the power.
CAUTION: Water will remain in the drain assembly.
Suction out any remaining water.
2. Remove the top cover and left panel.
3. Disconnect the wiring.
4. Remove the 4 retaining nuts and remove the plate.
5. From inside the chamber, remove the coarse and fine filters
and pull the heater out from the sump side.
To reinstall:
1. Reinstall the gasket on the heater’s
base and insert it into position
from the sump side.
2. From the left side of the unit,
put the heater plate into
position and fasten using
the retaining nuts.
3. Connect the wiring,
see Figure 35b.
4. Reinstall the panels.
24 to heater
10 to thermostat
and heater
Figure 35a
39 to ground
Water heater HYDRim –
part # 01-113839S
Figure 35b
9 to thermostat
67
Page 68
8. Left Side Components
8. Left Side Components
8.2 Removing and reinstalling recirculation pump
1. Turn the unit off and disconnect the power.
CAUTION: Water will remain in the drain assembly.
Suction out any remaining water.
2. Remove the top, rear and left panels.
3. Disconnect the rear kickplate by removing the screws on
either side. (Figure 36a)
4. Remove the kickplate from the front and cut the cable ties
fastening the air vent outlets.
5. From the back, pull the air vent outlets out of the way.
6. Disconnect the pump wires.
7. Place an absorbent cloth under the pump to catch residual
water and loosen hose clamp to disconnect outlet hose at
the pump side.
8. From the left side of the unit, loosen the hose clamp to
disconnect the inlet hose at the pump side.
9. From left side of the unit, disconnect the capacitor wires.
(Figure 36b)
10. From the rear of the unit, remove the 3 screws fastening
the bracket. (Figure 36c)
11. Pull the pump out and separate the pump from the bracket.
screwscrewpump
Figure 36a
To reinstall:
1. Install new pump onto the bracket.
NOTE: Use the new vibration mount that comes with the
new pump.
2. Reattach the pump to the inlet
and outlet tubings
and tighten the hose clamps.
3. Fasten the bracket into place.
4. Attach the wiring and capacitor
wires.
5. Thread the air vents back
through to the front and
fasten with new cable ties.
6. Reinstall the panels.
Circulation pump 50Hz –
part # 01-111782S
Circulation pump 60Hz –
part # 01-113201S
capacitor
wires
Figure 36b
connections
68
screwspump wire
Figure 36c
Page 69
9. Rear Components
WARNINGS AND
PRECAUTIONS
If you have questions about the unit you are repairing, please do not hesitate to
contact your local SciCan representative for information. Also, the HYDRIM is
heavy. Exercise caution and seek assistance when lifting or carrying units.
EXERCISE CAUTION
• Hazardous voltages are accessible when the cover is removed.
• Disconnect the power cord before servicing the power mains portion of the
controller board and associated devices.
• Removing the panels will expose some sharp metal edges. Be careful and wear
long sleeves and gloves.
PERFORM TESTS
• If panels are removed, a dielectric strength test (Hi-Pot) AND a protective
bonding impedance test (ground continuity) must be performed on the HYDRIM
when the work is completed and after the cover has been returned to the unit.
• A dielectric strength test (hi-pot) must be performed on the unit if parts
associated with the power main are serviced or replaced.
• A protective bonding impedance test (ground continuity) must be performed
on the unit if components of the protective earthing system are changed or if
connections are broken and remade.
PROTECT THE UNIT
• The HYDRIM contains electronic circuitry that is static sensitive. Always wear
a static strap when working with or near printed wiring boards. In addition, use
static footstraps, grounding mats and grounded work surfaces when servicing
microprocessor devices. Transport boards and devices in static protected bags.
• In order to ensure adherence to the applicable safety agency approvals,
state, provincial, regional and national laws, replace components with SciCan
approved parts only.
69
Page 70
9. Rear Components
9.1 Removing and reinstalling the Ethernet and RS232 ports
1. Disconnect the power and remove the top cover and back
panel.
2. To remove Ethernet port, remove the screw on the left of
the port, disengage the tab from the bracket, and
disconnect the wiring from the I/O board.
3. To remove the RS232 port, remove the fastening nuts,
and disconnect wiring from the LCD controller board.
To replace, reverse removal instructions.
Cable, Ethernet – part # 01-113854S
Cable, RS232 port – part #01-113260S
Ethernet portRS232 port
9.2 Removing and reinstalling the chamber pressure switch
1. Turn the unit off, disconnect the power and remove
the top and rear panels.
2. Disconnect the wiring from the chamber pressure switch.
3. Loosen the hose clamp and remove the chamber pressure
switch from the metal tubing.
To reinstall:
1. Push the chamber pressure switch onto metal tubing
and fasten with the hose clamp.
2. Reconnect wiring. (Figure 38b)
Chamber pressure switch – part # 01-111409S
60 to pos. 3
Figure 37
Figure 38a
70
59 to pos. 1
Figure 38b
Page 71
9. Rear Components
9.3 Removing and reinstalling the water softener system
1. Turn unit off and disconnect the power.
2. Remove the top, rear and left panels.
3. From inside the chamber, remove the water softener
cap and siphon the remaining water from the water
softener system.
4. From the unit’s right side, locate and disconnect
the sensor wire. (Figure 39a)
5. At the rear of the unit, number the water softener
hoses according to Figure 39c.
6. From inside the chamber, remove the water softener
system mounting nut.
7. From the rear of the unit, pull the water softener out.
8. Disconnect the hoses (hoses may need to be heated
to be removed).
sensor wire
Figure 39a
To reinstall:
1. Connect hoses matching the hose number to the number embossed
on the water softener.
2. Position the water softening system and from the chamber side,
fasten the mounting ring tightly. NOTE: The mounting nut should not
be tightened by hand. Use a tool to get a good seal.
3. Connect the sensor wire.
4. To refill the water softener, unscrew the salt container lid and pour
1 litre / 1 quart of water into the water softener. Add 1 kg / 2.2 lbs
of water softening salt in the same manner. Screw the salt container
cap on tightly.
5. Reinstall panels.
Water softener – part # 01-113857S
345
mounting nut
Figure 39b
Figure 39c
71
Figure 39d
Page 72
9. Rear Components
9.4 Removing and reinstalling the air gap
air gap wire connections
inlet tube
(not shown)
screw
(on side)
air gap pump
breather tube
screw
(behind bracket)
pump tube
screw
regeneration
valve wires
regeneration
valve screws
regeneration
valve
Figure 40
1. Turn the unit off and disconnect the power. Remove the top cover, right and rear panels.
CAUTION: Water will remain in the air gap. Remove kickplate and use air gap draining tube to drain.
2. Disconnect 3 switch wiring connections from the top of the air gap (note colour codes).
3. Disconnect the air gap pump tube.
4. Disconnect the air gap from the regeneration valve.
5. Disconnect the air gap inlet tube.
6. Disconnect air gap pump breather tube.
7. Remove the 3 mounting screws, and remove the air gap.
To reinstall:
1. Put the air gap back into position and reinstall the screws.
2. Reconnect the tubing.
3. Reconnect the switch wiring connections (green to 67/68; Yellow to 63/64; unmarked wire to 65/66).
Back flow preventer AB air gap – part # 01-113853S
72
Page 73
9. Rear Components
9.5 Removing and reinstalling the regeneration valve
1. Turn the unit off and disconnect the power. Remove the top cover, right and rear panels.
CAUTION: Water will remain in the air gap. Remove kickplate and use air gap drain tube to drain.
2. Disconnect the wires from the regeneration valve. (Figure 40)
3. Disconnect the outlet tube.
4. Remove the 2 screws from the bracket. (Figure 40)
5. Disconnect the inlet tube, and remove the valve (note valve orientation).
To reinstall, connect the inlet, connect the valve to the bracket, connect the outlet tube, and reconnect
the wires (wires 15 and 35/36). NOTE: Install the valve with directional flow arrow pointing downwards.
Regeneration valve part # 01-113861S
9.6 Removing and reinstalling the water inlet valves
1. Turn the unit off and disconnect the power.
2. Remove the top, right and rear panels.
3. Valves for hot water, cold water and RO water
inlets can be accessed from the rear and
right side of the unit.
4. Disconnect the wiring.
5. Remove the fastening screws to remove
a valve from the bracket.
6. Disconnect the tubing.
To reinstall, reverse instructions.
Valve
Cold water inletBlue1131/32
Hot water inletRed1233/34
RO water inletYellow1332/33
Cold: Valve 1 in, 1 out – part # 01-113330S
Hot: Valve 1 in, 1 out – part # 01-113330S
RO: Valve 1 in, 1 out – part # 01-113330S
Inlet hose (cold, hot) NA – part # 01-107788S
Inlet hose (RO) NA – part # 01-113863S
Inlet hose (cold, hot) Europe – part # 01-107787S
Inlet hose (RO) Europe – part # 01-113864S
Check valve EA type EN1717 – part# 01-113865S
Color Coding
Wiring
hot water inlet
cold water inlet
RO water inlet
Figure 41a
73
cold water valvehot water valve
RO water valve (below)
Figure 41b
Page 74
9. Rear Components
9.7 Removing and reinstalling the fuses and fuse holders
1. Turn the unit off, disconnect the power and remove top and rear panels.
2. For continued protection against the risk of fire, replace fuses with
15A, 250V type F only.
3. To remove fuse holders, disconnect wires and remove mounting nut
fastening holder to bracket.
To replace, reverse instructions and reference Figure 42a for
wire connections.
Fuse 15 A (2 pcs) – part # 01-103472S
3
1
4
2
Figure 42a
9.8 Removing and reinstalling the AC power inlet / EMI Filter
1. Turn unit off and disconnect power and remove top and rear panels.
2. Disconnect wiring from back of EMI Filter.
3. Remove mounting screws to release from bracket.
To replace:
1. Place EMI Filter into position in correct orientation. (Figure 42b)
2. Reconnect wiring. (Figure 43)
EMI filter – part # 01-110505S
Figure 42b
2
74
Figure 43
371
Page 75
9. Rear Components
9.9 Removing and reinstalling the condenser
1. Turn the unit off and disconnect the power.
CAUTION: Water will remain in the
drain assembly and condenser.
Suction out any remaining water
from the drain exhaust.
2. Remove the top and rear panels.
3. Disconnect the inlet tube, the outlet tube
and the drain tube.
4. Disconnect the air vent hoses.
5. Remove the fastening nut from
the bottom center of the condenser.
6. From inside the chamber, remove
the splash shield (4 screws).
7. Remove the 4 screws fastening
the condenser to the chamber.
NOTE: These 4 screws have O-rings.
8. Pull the condenser off the unit
from the back.
inlet
outlet
drain
fastener nut
(not shown)
screws
Figure 44b
air vent hoses
screws
Figure 44a
Figure 44c
75
Page 76
9. Rear Components
To reinstall:
1. Put the gasket into its position on top of the chamber.
2. Put the condenser into position.
3. Install the condenser screws
(IMPORTANT: use the 4 screws with O-rings) inside
the chamber to fasten the condenser, and install
splash shield.
4. Install fastening nut at middle centre bottom
of condenser (on unit rear).
5. Re-connect tubing and hoses.
6. Reinstall panels.
Condenser – part # 01-113912S
Figure 44d
condenser gasket
76
Page 77
10. Spare Parts & Accessories
This spare part list was last updated on the date of the release of the unit.
To see an updated spare part list, please refer to my.scican.com.
01-112051S40uF Motor Run Capacitor, Spare
01-113831SBreather Check Valve, L110w/M2 G4
01-111408SChamber Level Switch, L110w/M2/G4
01-111409SChamber Pressure Switch, L110w/M2/G4
01-111782SCirculation Pump,M2,50Hz
01-113201SCirculation Pump,M260Hz
01-111898SCoupling Kit Circular Press. Pipe, K
01-111466SCover Left Side HYDRIM L110w/M2/G4
01-111467SCover Rear HYDRIM L110w/M2 G4
01-111468SCover Right Side HYDRIM L110w/M2/G4
01-111469SCover Top HYDRIM L110w/M2/G4
01-113832SCover Top Low Profile HYDRIM M2 G4
01-113834SDecal HYDRIM M2 G4
01-111670SDetergent Coiled Tubing and Cap,K
01-113835SDoor, Chemical HYDRIM L110w/M2 G4
01-113836SDoor Wash Chamber HYDRIM L110w/M2 G4
01-111783SDoor Latch Assembly, L110w/M2/G4
01-112080SDoor Latch Cord Assy L110w/M2/G4
01-113837SDosing Pump Bellows L110w/M2 G4
01-111412SDrain Pump, HYDRIM L110w/M2/G4
01-111780SDryer Filter, M2/G4
01-111779SBlower Complete w/ Filter, M2/G4
01-112730SDual Spring Replace, Door, L110W/M2/G4
01-113838SExtension Low Spray Arm L110w/M2 G4
01-111899SFitting Adapt. Trolley Handpiece M2
01-112593SFoam Kickplate, L110/M2
01-112545SFuse Holder, Qty2, Bravo/HYDRIMs
01-113839SWater Heater HYDRIM L110w/M2 G4
01-113840SHinge Left, HYDRIM L110w/M2 G4
01-113841SHinge Right, HYDRIM L110w/M2 G4
01-113842SHYDRIM M2 G4 Reference Guide
01-108305SHYDRIM Water Test Kit,J/K
01-107787SInlet Hose Europe C61/M2/G4
01-107788SInlet Hose N.A. C61/L110w/M2/G4
01-111476SKickplate Front, HYDRIM L110/M2/G4
01-111477SKickplate Rear, HYDRIM L110w/M2/G4
01-113464SKit Brushes Motor Blower HYDRIM M2
01-112438SLower Basket Rail L110/M2 Kit
01-113843SCoarse Filter, HYDRIM L110w/M2 G4
77
Page 78
10. Spare Parts & Accessories
01-113844SFine Filter, HYDRIM L110w/M2 G4
01-113845SOperator Manual HYDRIM M2 G4
01-113847SPackaging HYDRIM M2/G4
01-112682SResistor Blower HYDRIM M2 (3pcs)
01-112024SRocker Switch Spare Kit ,L110W/M2
01-112594SSalt, Water Softener
01-111484SScrew, Back Cover & Service Door, K
01-111485SScrew Kickplate HYDRIM L110w/M2/G4
01-111483SScrew Top Cover HYDRIM L110/M2/G4
01-113849STrolley Lower HYDRIM M2 G4
01-113850STrolley Upper HYDRIM L110w/M2 G4
01-113852STubing Drain HYDRIM M2 G4
01-109790SUpper Spray Arm HYDRIM
01-110411SVertical Instrument Rack, M2, S
01-111495SWash Arm Middle L110w/M2/G4
01-113853SBackflow Preventer AB AirGap, M2 G4
01-113854SCable, Ethernet, L110w/M2 G4
01-113260SCable, RS232, C61/L110/M2 G4
01-113394SCheck Valve w/FLap, C61/M2 G4
01-113270SDual Temp. Sensor, C61wd/M2 G4
01-113855SFan, 24V, M2 G4
01-113310SIO PCB, C61/L110w/M2 G4
01-113856SLCD Assembly, L110w/M2 G4
01-110505SEMI Filter 20A/250V
01-113308SPower Cord Japan 13A/250V
01-113309SPower Cord UK 13A/250V
01-110281SPower Cord N.A. 15A/250V
01-113266SPower Supply 5V/24V,C61/L110w/M2 G4
01-113682SSpeaker Assy, C61/L110w/M2 G4
01-113331SValve,1in-2out, C61/L110w/M2 G4
01-113330SValve,1in-1out, C61/L110w/M2 G4
01-110282SPower Cord EU 16A/250V
01-110283SPower Cord Aus 15A/250V
01-110284SPower Cord Africa/India 16A/250V
01-110286SPower Cord Swiss 16A/250V
01-110287SPower Cord Israel 16A/250V
01-110361SPower Cord China 16A/250V
01-113669SColour LCD Controller, M2
96-113787Service Manual HYDRIM M2 G4
95-113756M2 G4 Installation & Maintenance
01-113790SMain Chamber Seal L110w/M2 G4
78
Page 79
10. Spare Parts & Accessories
01-113789SLower Door Seal L110w/M2 G4
01-112398SCable, USB, Statim/HYDRIML/M G4
01-113857SWater Softener, L110w/M2/G4
01-113858SDosing Reservoir w/Switch L/M G4
01-113859SDosing Reservoir w/Plug L/M G4
01-113860SDosing Valve, L110w/M2 G4
01-113861SRegeneration Valve, M2 G4
01-113862SAir Gap Pump, HYDRIM M2 G4
01-113863SInlet hose NA Straight L110w/M2 G4
01-113864SInlet hose EU Straight L110w/M2 G4
01-113865SCheck Valve EA type EN1717
01-113866SCheck Valve 3/8" C61/L110w/M2/G4
01-113867SCheck Valve 1" L110w/M2/G4
01-113868SCable, Communication L110w/M2 G4
01-113654SL110w/M2/G4 D-Strip Door Seal Spare
01-103472SFuses 15A (2 pcs), B/C/D/HYDRIMs
01-113535SBlower Motor, HYDRIM M2/G4
01-113912SCondenser, HYDRIM M2 G4
01-113909SDosing Pump Vol. Valid. C61wd/M2 G4
79
Page 80
10. Spare Parts & Accessories
This accessory part list was last updated on the date of the release of the unit.
To see an updated spare part list, please refer to my.scican.com.
01-109965S4XL Cassette Rack, 1/1, HYDRIM L/M2
01-109963S5 Cassette Rack, 1/1, HYDRIM L/M2
01-109964S5 Cassette Rack, 1/1, HYDRIM L/M2
01-111598Basket, full size HYDRIM L110W/M2,K
01-109967SBasket, Hygiene, 1/4, HYDRIM L/M2
01-109966SBasket with lid, 1/4, HYDRIM L/M2
01-110409SHinged Instrument Rack, S
01-107240Kit 2000 Basket, J
01-107241Kit 5000 Basket, J
01-108232Kit Basket Long, K
01-108294Kit Basket with Hinged Lid,K
01-108371Kit Lower Rack 4XL Cassettes, K
01-108370Kit Upper Rack 3XL Cassettes, K
01-110412STray Rack, L110/M2, S
01-113776SHandpiece Adapter Assembly M2 G4
80
Page 81
11. Appendix A
HYDRIM M2 G4 Electrical Schematic
4
5
Color LCD Controller
2
1
LB-S1
Speaker
12
11
Serial/I2C
LB to IO communication
USB connector
USB
LB-J2 USB
(10-112305)
4
Speaker
3
2
1
Power EMI filter
SALT VALVE
15
123
RS232
CONDENSER VALVE
14
456
Ethernet connector
ETHERNET
COLD WATER VALVE
HOT WATER VALVE
RO WATER VALVE
12
13
1234567
J3
51
52
Chamber_Level
53
54
Dosing_Validation
55
56
Chemical_Level
57
58
Salt
59
60
Chm_Pressure
61
62
Door_Position
T3
2
1
J13
Color LCD
11
4
3
2
1
J10 I2C
1
2
3
4
5
6
7
8
J7Extras
1
2
3
4
5
6
7
8
9
10
J8Switches 2
1
2
3
4
5
6
7
8
9
10
11
12
J1Serial/I2C
3
2
1
J11 T1
3
2
1
J12 T2
5
4
3
2
1
J14 ADC In
1
2
3
4
5
6
7
8
9
10
LB-J3Ethernet
1
LB-J7
63
65
67
Micro SD card
SD CARD
2345678910
64
AG_Full
66
AG_Overflow
68
AG EMPTY
t°
CHM TEMP SENSOR
t°
123
LB to IO communication
CHM TEMP VALIDATION SENSOR
Line
Neutral
4
F215A
F115A
Power Switch
M
AIR GAP PUMP
WATER HEATER
3
FULL/OVERFLOW SWITCH
Thermal cutoff
8
9
5
7
Fan (24V)
2
5V & 24V
Power Supply
M
CIRCULATION PUMP
1
J4
6
2
3
4
5
5
4
J15 DC Motor
7132182021
19
M
DRAIN PUMP
DOSING VALVE
M
DOSING PUMP 1
high speed
C2
1
C3
M
DRYER
1
J5
2
3
4
5
6
IO Board
low speed
22
C1
16
Air Gap Pump
Air Gap Pump
1 sehctiwS9J
1
123
2
6
5
4
3
2
1
2
1
3
4
J16 AGPump
5
6
7
8
9
10
11
12
13
14
J17 Power/Fan24
15
16
17
18
19
20
21
22
(10-112610)
Latch/Fan
1
2
3
4
J6
J2
RS232
M
AIR GAP VALVE
Door Latch
AA
BB
Serial connector
CC
DD
81
Page 82
11. Appendix B
HYDRIM M2 G4 Flow Diagram
82
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.